1 SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg
2 Disclosures I have the following financial relationships to disclose Consultant to CT Imaging GmbH Managing director of RayConStruct GmbH Grant supports from AiF, DFG, Intel, Siemens, Varian, Ziehm I will discuss the following off-label use in my presentation Partial scan artifact reduction Dynamic iterative beam hardening correction Blooming artifact reduction Temporal resolution improvement Low-dose phase-correlated image reconstruction
3 Early Cardiac Spiral CT Single Slice CT (RSNA 1997) Standard 4-Slice CT (RSNA 1998) Univ. Erlangen Achenbach, Kalender, Kachelrieß ECG-correlated 1998: First 4D images of the heart! M. Kachelrieß, and W.A. Kalender. Electrocardiogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart. Med. Phys., 25(12): , December 1998.
4 Outline Cardiac beam hardening correction Iterative reconstruction and restoration Blooming artifact reduction Temporal resolution improvement Low-dose phase-correlated CT Motion compensation techniques Improved DECT reconstruction
5 Perfusion Analysis in CT Time-attenuation curves (TAC) C(t) Healthy Peak enhancement Ischemic Time to peak t Beam hardening artifacts cause an underestimation of the CT-values leading to incorrect perfusion parameters! 130/130
6 DIBHC: Beam Hardening Correction of Three Materials using Temporal Dependence Where is soft tissue? Where is bone? Where is Iodine? soft tissue bone+iodine σ-image bone iodine Stenner et al., Dynamic iterative beam hardening correction (DIBHC) in myocardial perfusion imaging using contrast-enhanced computed tomography. Investigative Radiology 45(6): , June 2010.
7 Original Corrected Difference 4 Application of DIBHC /200 1 Beam hardening Perfusion defect 80/200 0/50 ROI Original CT-values / HU Corrected CT-values / HU / HU (end-diastolic trigger) Rawdata courtesy of Prof. Stephan Achenbach, Department of Cardiology, University of Erlangen-Nürnberg, Erlangen, Germany
8 Iterative Reconstruction Iterative Restoration Iterative image reconstruction comparison in rawdata domain reduces noise reduces cone-beam artifacts enhances spatial resolution reduces beam hardening artifacts time consuming Iterative image restoration comparison in raw data domain not needed for noise reduction operating on an initial FBP image computationally highly efficient FBP (Standard) FBP + Image Restoration Images provided by Siemens Healthcare, Forchheim, Germany
9 Blooming Artifact Reduction High density objects calcifications high iodine concentrations in small vessels Limited spatial resolution blurrs high density regions
10 Blooming Artifact Reduction (BAR) Low resolution image f L (r) High resolution image f H (r) Weight function Bilateral filtered w (r) image f BH (r) Mixing ( 1 w( r) ) f ( r) w( r) f ( ) f ( r) = + r BAR L BH BAR image f BAR (r)
11 Is this a stenosis? B26f B26f Patient-data provided by Prof. Dr. Stephan Achenbach, Universitätsklinikum Erlangen C 150 W 1000 C 150 W 1000
12 Is this a stenosis? B26f B26f+B50f BAR C 150 W 1000 C 150 W 1000 Patient-data provided by Prof. Dr. Stephan Achenbach, Universitätsklinikum Erlangen
13 Stent visibility B26f B26f C 150 W 1000 C 150 W 1000 Patient-data provided by Prof. Dr. Stephan Achenbach, Universitätsklinikum Erlangen
14 Stent visibility B26f B26f+B75f BAR C 150 W 1000 C 150 W 1000 Patient-data provided by Prof. Dr. Stephan Achenbach, Universitätsklinikum Erlangen
15 Visibility of stent lumen B26f B26f Patient-data provided by Prof. Dr. Stephan Achenbach, Universitätsklinikum Erlangen C W C 400 W 1500
16 Visibility of stent lumen B26f B26f+B50f BAR C 400 W 1500 C 400 W 1500 Patient-data provided by Prof. Dr. Stephan Achenbach, Universitätsklinikum Erlangen
17 Temporal Resolution Improvement Temporal resolution can be improved by increasing the gantry rotation speed increasing the number of source detector units using reduced angular coverage for reconstruction Recent improvements in compressed sensing1, 2, 3 appear to enable reconstruction from less data are said to have the potential to improve temporal resolution may allow to reconstruct from only 90 or 2 45 of data 1 Sidky et al., Accurate image reconstruction from few-views and limited-angle data in divergent-beam CT, J. X-ray Sci. Tech., 14: , Chen et al., Temporal resolution improvement using PICCS in MDCT cardiac imaging, Med. Phys. 36: , Ritschl et al., Improved Sparsity-Constrained Image Reconstruction Applied to Clinical CT Data, submitted to IEEE Transactions on Medical Imaging, 2010
18 Approach for itri-piccs and itri-cs Cost function: Constraint α ( f f prior ) + (1 α) f 2 Rf p < ε General proceeding 1. Short scan to initialize ART 2. ART iteration 3. Minimize cost function 4. Continue with step 2 until result is satisfying Improved update strategy itri-piccs α = 0.4 / itri-cs α = 0
19 Example Short Scan itri-piccs itri-cs Dual Source Single Source C/W 200/1000
20 Low-Dose Phase-Correlated (LDPC) Double-Gated Reconstruction 1,2 Small animal cardiac CT imaging requires flat panel detector technology implies rather slow rotation times forces us to do respiratory (R) and cardiac (C) gating suffers from very high respiratory rates (around 150 rpm) and from very high heart rates (around 300 bpm) RC-gating (double gating) results in a very small amount of usable data e.g. 2% for a 10% respiratory and a 20% cardiac window High image quality requires either to scan at very high dose values (> 1 Gy) or to use sophisticated image reconstruction techniques 1 Sawall et al., Boosting image quality in low-dose RC-gated 5D cone-beam micro-ct, 5th European Molecular Imaging Meeting, May Sawall et al., Low Dose Phase Correlated Cone Beam Micro CT of Small Animals, IEEE Medical Imaging Conference, November 2010.
21 Mouse Double-Gating: Prior Art 2800 mgy 1840 mgy 500 mgy M. Drangova, N. L. Ford, S. A. Detombe, A. R. Wheatley, and D. W. Holdsworth, Fast retrospectively gated quantitative four dimensional (4D) cardiac micro computed tomography imaging of free breathing mice, Investigative Radiology, vol. 42, no. 2, pp , Feb C. Badea, B. Fubara, L. Hedlund, and G. Johnson, 4D micro CT of the mouse heart, Molecular Imaging, vol. 4, no. 2, pp , Apr./Jun Conventional, but at low dose
22 LDPC Algorithm edge-preserving smoothing f std =X -1 p p diff =p-xf std f LDPC =f std +X -1 PC p diff use image that is based on all projections as prior (standard image) calculate rawdata difference for desired motion phases perform correction include edge preserving anisotropic filtering Standard image f std reconstructed from all projections. LDPC reconstruction f LDPC.
23 Example, R=0%, Cardiac-Loop Axial Sagittal Coronal σ=125 HU LDPC PC σ=26 HU R=0%, R=10%, C=20% (400 HU / 700 HU)
24 Dose Reduction Study Phase-Correlated Low Dose Phase-Correlated Axial Sagittal Coronal Axial Sagittal Coronal σ=354 HU σ=89 HU 60 mgy σ=198 HU σ=57 HU 250 mgy σ=125 HU σ=26 HU 500 mgy R=0%, C=0%, R=10%, C=20% (400 HU / 800 HU)
25 What to expect is: lower dose less image noise reduced blooming artifacts improved temporal resolution lower beam hardening artifacts more quantitative images (including DE)
26 Thank You!
Motion Compensation from Short-Scan Data in Cardiac CT
Motion Compensation from Short-Scan Data in Cardiac CT Juliane Hahn 1,2, Thomas Allmendinger 1, Herbert Bruder 1, and Marc Kachelrieß 2 1 Siemens Healthcare GmbH, Forchheim, Germany 2 German Cancer Research
More informationSpiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004
Spiral ASSR Std p = 1.0 Spiral EPBP Std p = 1.0 Kachelrieß et al., Med. Phys. 31(6): 1623-1641, 2004 256 slices (0/300) Advantages of Cone-Beam Spiral CT Image quality nearly independent of pitch Increase
More informationCardiac Dual Energy CT: Technique
RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de
More informationAutomatic Intrinsic Cardiac and Respiratory Gating from Cone-Beam CT Scans of the Thorax Region
Automatic Intrinsic Cardiac and Respiratory Gating from Cone-Beam CT Scans of the Thorax Region Andreas Hahn 1, Sebastian Sauppe 1, Michael Lell 2, and Marc Kachelrieß 1 1 German Cancer Research Center
More informationArtefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT
Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT Marcus Brehm 1,2, Thorsten Heußer 1, Pascal Paysan 3, Markus Oehlhafen 3, and Marc Kachelrieß 1,2 1 German Cancer Research Center
More informationMoving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region
Moving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region Andreas Hahn 1,2, Sebastian Sauppe 1,2, Michael Knaup 1, and Marc Kachelrieß 1,2 1 German Cancer Research Center (DKFZ),
More informationRespiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data
Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data Christopher M Rank 1, Thorsten Heußer 1, Andreas Wetscherek 1, and Marc Kachelrieß 1 1 German Cancer
More informationTotal Variation Regularization Method for 3D Rotational Coronary Angiography
Total Variation Regularization Method for 3D Rotational Coronary Angiography Haibo Wu 1,2, Christopher Rohkohl 1,3, Joachim Hornegger 1,2 1 Pattern Recognition Lab (LME), Department of Computer Science,
More informationDUE to beam polychromacity in CT and the energy dependence
1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact
More informationTotal Variation Regularization Method for 3-D Rotational Coronary Angiography
Total Variation Regularization Method for 3-D Rotational Coronary Angiography Haibo Wu 1,2, Christopher Rohkohl 1,3, Joachim Hornegger 1,2 1 Pattern Recognition Lab (LME), Department of Computer Science,
More informationCT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION
CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology
More informationNIH Public Access Author Manuscript Med Phys. Author manuscript; available in PMC 2009 March 13.
NIH Public Access Author Manuscript Published in final edited form as: Med Phys. 2008 February ; 35(2): 660 663. Prior image constrained compressed sensing (PICCS): A method to accurately reconstruct dynamic
More informationEmpirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography
Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Marc Kachelrieß, a Katia Sourbelle, and Willi A. Kalender Institute of Medical Physics, University of
More informationComputer-Tomography II: Image reconstruction and applications
Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de
More informationMake the Most of Time
Make the Most of Time Temporal Extension of the itv Algorithm for 4D Cardiac C-Arm CT Viktor Haase 1,5, Oliver Taubmann 2,3, Yixing Huang 2, Gregor Krings 4, Günter Lauritsch 5, Andreas Maier 2,3, Alfred
More informationSeparate CT-Reconstruction for Orientation and Position Adaptive Wavelet Denoising
Separate CT-Reconstruction for Orientation and Position Adaptive Wavelet Denoising Anja Borsdorf 1,, Rainer Raupach, Joachim Hornegger 1 1 Chair for Pattern Recognition, Friedrich-Alexander-University
More informationCT Basics Principles of Spiral CT Dose. Always Thinking Ahead.
1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central
More informationMulti-slice CT Image Reconstruction Jiang Hsieh, Ph.D.
Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced
More informationDUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography
Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography Marc Kachelrieß, Member, IEEE, Timo Berkus, Philip Stenner, Willi A. Kalender Abstract Material selective imaging using dual energy
More informationEdge-Preserving Denoising for Segmentation in CT-Images
Edge-Preserving Denoising for Segmentation in CT-Images Eva Eibenberger, Anja Borsdorf, Andreas Wimmer, Joachim Hornegger Lehrstuhl für Mustererkennung, Friedrich-Alexander-Universität Erlangen-Nürnberg
More information3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH
3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk
More informationEstimating 3D Respiratory Motion from Orbiting Views
Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct. 2005 Funding provided by NIH Grant P01 CA59827 Motivation Free-breathing
More informationPhilips SPECT/CT Systems
Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET
More informationProjection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT
Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Benedikt Lorch 1, Martin Berger 1,2, Joachim Hornegger 1,2, Andreas Maier 1,2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg
More informationECG-Correlated Imaging of the Heart with Subsecond Multislice Spiral CT
888 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 19, NO. 9, SEPTEMBER 2000 ECG-Correlated Imaging of the Heart with Subsecond Multislice Spiral CT Marc Kachelrieß*, Stefan Ulzheimer, and Willi A. Kalender
More informationSpatial-temporal Total Variation Regularization (STTVR) for 4D-CT Reconstruction
Spatial-temporal Total Variation Regularization (STTVR) for 4D-CT Reconstruction Haibo Wu a, b, Andreas Maier a, Rebecca Fahrig c, and Joachim Hornegger a, b a Pattern Recognition Lab (LME), Department
More informationComputed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA
Computed Tomography Principles, Design, Artifacts, and Recent Advances THIRD EDITION Jiang Hsieh SPIE PRESS Bellingham, Washington USA Table of Contents Preface Nomenclature and Abbreviations xi xv 1 Introduction
More informationRespiratory Motion Estimation using a 3D Diaphragm Model
Respiratory Motion Estimation using a 3D Diaphragm Model Marco Bögel 1,2, Christian Riess 1,2, Andreas Maier 1, Joachim Hornegger 1, Rebecca Fahrig 2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg 2
More informationAcknowledgments and financial disclosure
AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories
More informationAIDR 3D Iterative Reconstruction:
Iterative Reconstruction: Integrated, Automated and Adaptive Dose Reduction Erin Angel, PhD Manager, Clinical Sciences, CT Canon Medical Systems USA Iterative Reconstruction 1 Since the introduction of
More informationConvolution-Based Truncation Correction for C-Arm CT using Scattered Radiation
Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation Bastian Bier 1, Chris Schwemmer 1,2, Andreas Maier 1,3, Hannes G. Hofmann 1, Yan Xia 1, Joachim Hornegger 1,2, Tobias Struffert
More informationSingle Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction
Single Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction Felix Lugauer 1,3, Jens Wetzl 1, Christoph Forman 2, Manuel Schneider 1, Berthold Kiefer
More informationRADIOLOGY AND DIAGNOSTIC IMAGING
Day 2 part 2 RADIOLOGY AND DIAGNOSTIC IMAGING Dr hab. Zbigniew Serafin, MD, PhD serafin@cm.umk.pl 2 3 4 5 CT technique CT technique 6 CT system Kanal K: RSNA/AAPM web module: CT Systems & CT Image Quality
More information7/31/ D Cone-Beam CT: Developments and Applications. Disclosure. Outline. I have received research funding from NIH and Varian Medical System.
4D Cone-Beam CT: Developments and Applications Lei Ren, PhD, DABR Department of Radiation Oncology Duke University Medical Center Disclosure I have received research funding from NIH and Varian Medical
More informationScaling Calibration in the ATRACT Algorithm
Scaling Calibration in the ATRACT Algorithm Yan Xia 1, Andreas Maier 1, Frank Dennerlein 2, Hannes G. Hofmann 1, Joachim Hornegger 1,3 1 Pattern Recognition Lab (LME), Friedrich-Alexander-University Erlangen-Nuremberg,
More informationReconstruction of CT Images from Sparse-View Polyenergetic Data Using Total Variation Minimization
1 Reconstruction of CT Images from Sparse-View Polyenergetic Data Using Total Variation Minimization T. Humphries and A. Faridani Abstract Recent work in CT image reconstruction has seen increasing interest
More information8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit
Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant
More informationCarestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2)
Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2) Author: Levon Vogelsang Introduction Cone beam computed tomography (CBCT), or cone beam CT technology, offers considerable promise
More informationFinancial disclosure. Onboard imaging modality for IGRT
Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty
More informationGuided Noise Reduction with Streak Removal for High Speed Flat Detector CT Perfusion
Guided Noise Reduction with Streak Removal for High Speed Flat Detector CT Perfusion Michael T. Manhart, André Aichert, Markus Kowarschik, Yu Deuerling-Zheng, Tobias Struffert, Arnd Doerfler, Andreas K.
More information8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition
Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare
More informationIterative CT Reconstruction Using Curvelet-Based Regularization
Iterative CT Reconstruction Using Curvelet-Based Regularization Haibo Wu 1,2, Andreas Maier 1, Joachim Hornegger 1,2 1 Pattern Recognition Lab (LME), Department of Computer Science, 2 Graduate School in
More informationML reconstruction for CT
ML reconstruction for CT derivation of MLTR rigid motion correction resolution modeling polychromatic ML model dual energy ML model Bruno De Man, Katrien Van Slambrouck, Maarten Depypere, Frederik Maes,
More informationA new calibration-free beam hardening reduction method for industrial CT
A new calibration-free beam hardening reduction method for industrial CT ECC 2 for industrial CT Tobias Würfl 1, Nicole Maaß 2, Frank Dennerlein 2, Andreas K. Maier 1 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg;
More informationTomographic Reconstruction
Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)
More informationCorso di laurea in Fisica A.A Fisica Medica 4 TC
Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure
More informationDeep Scatter Estimation (DSE): Accurate Real-Time Scatter Estimation for X-Ray CT using a Deep Convolutional Neural Network
Deep Scatter Estimation (DSE): Accurate Real-Time Scatter Estimation for X-Ray CT using a Deep Convolutional Neural Network Joscha Maier 1,2, Stefan Sawall 1,2, and Marc Kachelrieß 1,2 1 German Cancer
More informationAn Efficient Technique For Multi-Phase Model Based Iterative Reconstruction
1 An Efficient Technique For Multi-Phase Model Based Iterative Reconstruction Shiyu Xu, Debashish Pal and Jean-Baptiste Thibault Abstract Multi-phase scan is a fundamental CT acquisition technology used
More informationReduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy
Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy T. Rohlfing a, D. Zerfowski b, J. Beier a, P. Wust a, N. Hosten a, R. Felix a a Department of
More informationEnhancement Image Quality of CT Using Single Slice Spiral Technique
Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,
More informationMedical Image Processing: Image Reconstruction and 3D Renderings
Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,
More informationLow-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration
Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Joonki Noh, Jeffrey A. Fessler EECS Department, The University of Michigan Paul E. Kinahan Radiology Department,
More informationECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart
ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart Marc Kachelrieß, a) Stefan Ulzheimer, and Willi A. Kalender University of Erlangen-Nürnberg, Institute of Medical
More informationOptimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago
Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic
More informationRespiratory Motion Compensation for C-arm CT Liver Imaging
Respiratory Motion Compensation for C-arm CT Liver Imaging Aline Sindel 1, Marco Bögel 1,2, Andreas Maier 1,2, Rebecca Fahrig 3, Joachim Hornegger 1,2, Arnd Dörfler 4 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg
More informationImprovement of Efficiency and Flexibility in Multi-slice Helical CT
J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): 408 412 DOI: 10.1007/s12204-008-0408-x Improvement of Efficiency and Flexibility in Multi-slice Helical CT SUN Wen-wu 1 ( ), CHEN Si-ping 2 ( ), ZHUANG Tian-ge
More informationBrilliance CT Big Bore.
1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.
More informationContrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis
Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Stefan C. Saur 1, Hatem Alkadhi 2, Luca Regazzoni 1, Simon Eugster 1, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision
More informationCardiac C-arm CT: Efficient Motion Correction for 4D-FBP
2006 IEEE Nuclear Science Symposium Conference Record M11-235 Cardiac C-arm CT: Efficient Motion Correction for 4D-FBP M. Prümmer 1, L. Wigström 2,4, J. Hornegger 1, J. Boese 3, G. Lauritsch 3, N. Strobel
More informationEmpirical dual energy calibration EDEC for cone-beam computed tomography
Empirical dual energy calibration EDEC for cone-beam computed tomography Philip Stenner, Timo Berkus, and Marc Kachelriess Institute of Medical Physics, University of Erlangen-Nürnberg, Henkestrasse 91,
More informationMultidimensional Noise Reduction in C-arm Cone-beam CT via 2D-based Landweber Iteration and 3D-based Deep Neural Networks
Multidimensional Noise Reduction in C-arm Cone-beam CT via 2D-based Landweber Iteration and 3D-based Deep Neural Networks Dahim Choi 1, Juhee Kim 2, Seung-Hoon Chae 3, Byeongjoon Kim 4, Jongduk Baek 4,
More informationFast iterative beam hardening correction based on frequency splitting in computed tomography
Fast iterative beam hardening correction based on frequency splitting in computed tomography Qiao Yang a,b, Matthias Elter b, Ingo Schasiepen b, Nicole Maass b and Joachim Hornegger a,c a Pattern Recognition
More informationPURE. ViSION Edition PET/CT. Patient Comfort Put First.
PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion
More informationAn approximate cone beam reconstruction algorithm for gantry-tilted CT
An approximate cone beam reconstruction algorithm for gantry-tilted CT Ming Yan a, Cishen Zhang ab, Hongzhu Liang a a School of Electrical & Electronic Engineering, Nanyang Technological University, Singapore;
More informationMEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah
MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography
More informationFrequency split metal artifact reduction (FSMAR) in computed tomography
The Johns Hopkins University Advanced Computer Integrated Surgery Group 4 Metal Artifact Removal in C-arm Cone-Beam CT Paper Seminar Critical Review of Frequency split metal artifact reduction (FSMAR)
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More informationSimulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images
Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),
More information4DM Packages. 4DM Packages & License Types. Information to help you order the appropriate licenses for your site.
4DM Packages 4DM Packages & License Types. Information to help you order the appropriate licenses for your site. Nuclear Cardiac Quantification, Review, and Reporting Select Your 4DM Package and corresponding
More informationComputer-Tomography I: Principles, History, Technology
Computer-Tomography I: Principles, History, Technology Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de History
More informationHelical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice Guido Hilgers, a Tonnis Nuver, André Minken Department
More informationMetal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland
Metal Artifact Reduction CT Techniques R S S S Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland N. 1 v o 4 1 0 2. Postoperative CT Metal Implants CT is accurate for assessment
More informationArtifacts in Spiral X-ray CT Scanners: Problems and Solutions
Artifacts in Spiral X-ray CT Scanners: Problems and Solutions Mehran Yazdi, and Luc Beaulieu International Science Index, Electrical and Computer Engineering waset.org/publication/12966 Abstract Artifact
More informationSpiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA
Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization
More informationKymogram detection and kymogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart
Kymogram detection and kymogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart Marc Kachelrieß, a) Dirk-Alexander Sennst, Wolfgang Maxlmoser, and Willi A.
More informationDigital Image Processing
Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture
More informationShadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS
Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic
More informationDiagnostic quality of time-averaged ECG-gated CT data
Diagnostic quality of time-averaged ECG-gated CT data Almar Klein a, Luuk J. Oostveen b, Marcel J.W. Greuter c, Yvonne Hoogeveen b, Leo J. Schultze Kool b, Cornelis H. Slump a and W. Klaas Jan Renema b
More informationMOTION of a patient induces artifacts such as streaks
Motion-compensated Image Reconstruction for Cardiac CT with Sinogram-Based Motion Estimation JangHwanChoMember IEEE and Jeffrey A. Fessler Fellow IEEE Abstract Motion-compensated image reconstruction ()
More informationAcknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury
A. Sisniega et al. (presented at RSNA 214) High Performance Cone-Beam CT of Acute Traumatic Brain Injury A. Sisniega 1 W. Zbijewski 1, H. Dang 1, J. Xu 1 J. W. Stayman 1, J. Yorkston 2, N. Aygun 3 V. Koliatsos
More informationFlying Focal Spot (FFS) in Cone-Beam CT Marc Kachelrieß, Member, IEEE, Michael Knaup, Christian Penßel, and Willi A. Kalender
1238 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 3, JUNE 2006 Flying Focal Spot (FFS) in Cone-Beam CT Marc Kachelrieß, Member, IEEE, Michael Knaup, Christian Penßel, and Willi A. Kalender Abstract
More informationAssessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT
Assessment of 3D performance metrics D and 3D Metrics of Performance Towards Quality Index: Volumetric imaging systems X-ray based Volumetric imaging systems: CBCT/CT Tomosynthesis Samuel Richard and Ehsan
More information[PDR03] RECOMMENDED CT-SCAN PROTOCOLS
SURGICAL & PROSTHETIC DESIGN [PDR03] RECOMMENDED CT-SCAN PROTOCOLS WORK-INSTRUCTIONS DOCUMENT (CUSTOMER) RECOMMENDED CT-SCAN PROTOCOLS [PDR03_V1]: LIVE 1 PRESCRIBING SURGEONS Patient-specific implants,
More informationImage Guidance and Beam Level Imaging in Digital Linacs
Image Guidance and Beam Level Imaging in Digital Linacs Ruijiang Li, Ph.D. Department of Radiation Oncology Stanford University School of Medicine 2014 AAPM Therapy Educational Course Disclosure Research
More informationClinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis
Clinical Importance Rapid cardiovascular flow quantitation using sliceselective Fourier velocity encoding with spiral readouts Valve disease affects 10% of patients with heart disease in the U.S. Most
More informationComputed tomography - outline
Computed tomography - outline Computed Tomography Systems Jørgen Arendt Jensen and Mikael Jensen (DTU Nutech) October 6, 216 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering
More informationLimits of Dose Reduction in CT: How low is too low? Acknowledgement. Disclosures 8/2/2012
8/2/22 Limits of Dose Reduction in CT: How low is too low? Ehsan Samei Duke University Acknowledgement Disclosures Research grant: NIH Research grant: DHS Research grant: RSNA, QIBA Research grant: General
More informationImage Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD
Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image
More informationGeneralized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction
Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction Klaus Erhard a, Michael Grass a, Sebastian Hitziger b, Armin Iske b and Tim Nielsen a a Philips Research Europe Hamburg,
More informationKey words: Cone-beam CT CBCT, cardiac imaging, 4-D reconstruction, image quality
Extended parallel backprojection for standard three-dimensional and phase-correlated four-dimensional axial and spiral cone-beam CT with arbitrary pitch, arbitrary cone-angle, and 100% dose usage Marc
More informationCombination of Markerless Surrogates for Motion Estimation in Radiation Therapy
Combination of Markerless Surrogates for Motion Estimation in Radiation Therapy CARS 2016 T. Geimer, M. Unberath, O. Taubmann, C. Bert, A. Maier June 24, 2016 Pattern Recognition Lab (CS 5) FAU Erlangen-Nu
More information664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005
664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005 Attenuation Correction for the NIH ATLAS Small Animal PET Scanner Rutao Yao, Member, IEEE, Jürgen Seidel, Jeih-San Liow, Member, IEEE,
More informationBackground. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /
Radiographic Tomosynthesis: Image Quality and Artifacts Reduction Baojun Li, Ph.D Department of Radiology Boston University Medical Center 2012 AAPM Annual Meeting Background Linear Trajectory Tomosynthesis
More informationCorrecting organ motion artifacts in x-ray CT systems based on tracking of motion phase by the spatial overlap correlator. II. Experimental study
Correcting organ motion artifacts in x-ray CT systems based on tracking of motion phase by the spatial overlap correlator. II. Experimental study Amar C. Dhanantwari a) Defence and Civil Institute of Environmental
More informationEvaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT
Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Qiao Yang 1,4, Meng Wu 2, Andreas Maier 1,3,4, Joachim Hornegger 1,3,4, Rebecca Fahrig
More informationImage Acquisition Systems
Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS
More informationGradient-Based Differential Approach for Patient Motion Compensation in 2D/3D Overlay
Gradient-Based Differential Approach for Patient Motion Compensation in 2D/3D Overlay Jian Wang, Anja Borsdorf, Benno Heigl, Thomas Köhler, Joachim Hornegger Pattern Recognition Lab, Friedrich-Alexander-University
More informationThe team. Disclosures. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge.
Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge Dimitre Hristov Radiation Oncology Stanford University The team Renhui Gong 1 Magdalena Bazalova-Carter 1
More informationImprovement and Evaluation of a Time-of-Flight-based Patient Positioning System
Improvement and Evaluation of a Time-of-Flight-based Patient Positioning System Simon Placht, Christian Schaller, Michael Balda, André Adelt, Christian Ulrich, Joachim Hornegger Pattern Recognition Lab,
More informationAdvanced Image Reconstruction Methods for Photoacoustic Tomography
Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic
More information