1 Name Medical Imaging BMEN Spring 2016 Homework #4 and Nuclear Medicine Notes All questions are from the introductory Powerpoint (based on Chapter 7) and text Medical Imaging Signals and Systems, First Edition, Chapters 8-9 (scanned and available online). Due Thursday, March 31, If nuclear medicine modalities use radioactive materials and don t have very good resolution, what is their great value to medical imaging as compared to most other modalities? 2. a. What does PET stand for? b. SPECT? 2b. Explain the similar relationship between {projection radiography and CT} and {planar scintigraphy and SPECT} 3. Is PET a tomographic or a projection technique? What mechanism (emission/detection) determines this? Explain.

2 Planar scintigraphy instrumentation: 4. Unlike projection radiography and CT, in nuclear medicine we are not interested in detecting total intensity, but in the detected decay rate of the source, expressed as counts/disintegrations per time. Nuclear medicine images are thus built event-by-event. To detect these events, what instrument is used? Give both names. Note: the GOAL of this instrument is to create an image that portrays the distribution of radioactivity (sites and numbers of radioactive atoms) within the patient. That is, NUCLEAR MEDICINE MODALITIES MAP THE DISTRIBUTION OF RADIOACTIVITY IN THE BODY. 5. There are seven components to the detector named in #4. See diagram 8.1. Fill in the blank with the appropriate component following the description: a. Thick slab of lead the same dimensions as the scintillation crystal with array of holes in it. Provides an interface between the patient and the scintillation crystal by allowing only those photons traveling in an appropriate direction to interact with the crystal. b. Most commonly used detector in NM. Based on the property of certain crystals to emit light photons (to scintillate) after deposition of energy in the crystal by ionizing radiation. Most commonly used is sodium iodide with thallium doping ( NaI[T] ). c. A vacuum tube that receives the thousands of light photons from the scintillation crystal (that were produced by the incident ionizing gamma photon). This converts the light signal into an electrical signal and amplifies the electrical signal. Additional important notes (think There was an old lady who swallowed a fly ): The response (output) of the photomultiplier tube to the light photons that were produced by the scintillation crystal with the arrival of a gamma photon is electrons that form a current pulse that has a pulse height that ranges from few millivolts to few volts. There is an array of tubes that cover the entire scintillation crystal (to detect all events). More tubes means more resolution, uniformity, but higher costs, more maintenance, difficult calibration tubes are used on current cameras. d. Circuitry that determines 1) where on the face of the crystal the event occurred and 2) the combined output of all the tubes (which represents the light output from the crystal which in turn represents the energy deposited by the gamma photon). Additional important notes: the location (X,Y) of the event on the crystal can be determined to within a fraction of the tube diameter by analyzing the spatial distribution of pulse heights. The rate of (X,Y,Z) pulses coming from the camera is proportional to total radioactivity and size or height of (Z) pulses is proportional to the energy deposited in the crystal by the gamma photon. e. An analyzer that uses a scintillation counting system to obtain an energy spectrum from a radioactive source in order to sort out desired events from Compton scattered events and reject photons that have been Compton scattered Note: example 8.2 gives some more insight into this process if you are interested. f. Circuitry used to accept or reject an event based on the output from the device in (e) g. Modern interface to the camera for image capture purposes Planar scintigraphy Image Formation: 6. THE PRIMARY MECHANISM FOR CREATING IMAGES IN PLANAR SCINTIGRAPHY IS TO DETECT AND ESTIMATE THE POSITION OF INDIVIDUAL SCINTILLATION EVENTS ON THE FACE OF AN ANGER CAMERA.

3 Event position estimation is based on Hint: see summary and key points at the end of the chapter 7. If a k, k=1, K represents amplitudes of tube responses, where there are K tubes located at (x k,y k ), then K Z a k 1 k represents the of the light distribution, also denoted Z. The components of the center of mass (X,Y) are calculated as: X= Y= THE THREE VALUES (X,Y,Z) REPRESENT THE HEART OF AN ANGER CAMERA. Example 8.3 gives excellent insight into this process. 8. Image acquisition modes. Fill in the blank: a. A recording of the imaging session without images being produced immediately. Instead, (X,Y) signals and Z-pulse value are transmitted directly to computer memory, other physiological responses can be recorded, time marks are included. Allows flexibility with future data manipulation. b. The face of the Anger camera is represented as a matrix. An image is created during scan time by starting with a zero image at the beginning of scan time and incrementing the pixel values as events are accepted until a certain total count for the scan. c. A temporal succession of frame mode images. Allows the study of transient physiological processes such as uptake, washout, redistribution. d. A trigger signal (usually ECG-based) controls the distribution of data among multiple image matrices. e. A variation of static frame mode acquisition where the body is divided into a matrix and a series of static frames are acquired to cover the body. 9. Our imaging equation for planar scintigraphy can be developed (see section 8.3.3) as 0 ( x, y, z'; E) dz ' A( x, y, z) z ( x, y) e dz 2 4 z where Φ(x,y) is A(x,y,z) is μ(x,y,z ;E) is 0 The distance from (x,y,z) to a detector position is just considered a function of z. Why do we use this simplification?

4 10. The imaging equation as a whole is a type of projection, just like projection radiography. There are two sources of depth-dependent attenuation between the radiation source inside the body and the detector: 1) 2) Because of these effects, we expect that 1. Activity close to the camera will contribute a larger photon fluence rate than equivalent activity farther from the camera. 2. Viewing the identical activity from one side of the body or the other will result in dramatically different observed fluence rates (** SEE EXAMPLE FROM INTRODUCTORY PPT **). Radiologists are trained to correct for this, just as they do for depth-dependent magnification in projection radiography. 11. Rewrite equation 8.11 and explain it as a simplification of the imaging equation in # The basic imaging equation includes terms for both (the desired parameter) and (an undesired, but extremely important, additional factor). These two terms are not separable. Hint: see summary and key concepts Planar scintigraphy Image Quality: 13. Image quality in planar scintigraphy depends on three factors: (see summary and key concepts ) a. b. c. 14. The two factors that primarily affect image resolution are: a. b. THESE TWO FACTORS COMBINE TO YIELD THE SYSTEM RESOLUTION OF THE ANGER CAMERA. THE RESOLUTION OF THE CAMERA CAN BE CONSIDERED TO BE ITS ABILITY TO LOCALIZE A PHOTON INTERACTION AND IS DEFINED BY A FWHM VALUE THAT DEPENDS ON THE PSFs OF THE COLLIMATOR AND THE CAMERA. 15. It is common to assume that the PSF of the collimator, h C (x,y; z 0 ), is Gaussian with FWHM equal to R C, defined as collimator resolution. (Equation for R C is eqn. 8.12, but it is not necessary to know these details right now). The collimator resolution is dependent on the depth of the source activity being detected, z 0, and the

5 energy of the radiotracer used. *As always, our blurring can be expressed as a convolution with our developed imaging equation.* Rewrite Equation 8.14 to reinforce this. 16. Blurring that takes place in the scintillator is characterized by the intrinsic resolution of the Anger camera. There are two sources of the blurring [inaccuracy in estimation of (X,Y) ] in the scintillator: 1. Compton scatter events 2. Noise (statistical fluctuation) We express the blurring due to intrinsic camera effects with a PSF, h I (x,y), a Gaussian with FWHM equal to R I. Factor intrinsic resolution into our imaging equation by rewriting equation 8.18: 17. As we have discussed, a collimator experiences trade-off in sensitivity vs. resolution. A collimator with longer slits will yield better resolution (stop more scatter events) but will also be less sensitive. While it is not necessary to know the exact sensitivity equations, the example of 8.6 shows us that if we double the collimator hole length, we reduce the sensitivity by %. 18. Define field uniformity: 19. The intrinsic SNR of the planar scintigraphy image reflects a situation exactly like we discussed in radiography: The SNR is made larger by increasing the number of detected photons. The count rate, however, is affected by several factors. One of these factors is the resolving time of the camera. The inverse of the resolving time is the. Exceeding the maximum counting rate will cause, where the photons are absorbed too close together in the crystal to be counted as an event.

6 Emission Computed Tomography (PET and SPECT): PLANAR SCINTIGRAPHY IS THE ONLY DIRECT NUCLEAR MEDICINE IMAGING METHOD, JUST AS PROJECTION RADIOGRAPHY IS A DIRECT RADIOGRAPHIC IMAGING METHOD. SPECT AND PET, LIKE CT, REQUIRE COMPUTED TOMOGRAPHIC IMAGE RECONSTRUCTIONS. SPECT instrumentation: SPECT image formation refers to true transaxial tomography with standard nuclear medicine pharmaceuticals (the same as those used in planar scintigraphy that emit a single gamma photon upon decay). The options for image detection in SPECT consist of a ring of detectors (expensive, never used) or a single detector (in this case, the Anger camera), rotating around the patient. A single Anger camera rotating around the patient is the most popular method. (See instrumentation photos in the introductory ppt) **unlike CT acquisition, where we collimated our x-ray beam source into a slice in order to detect information from a slice of the patient, the Anger camera detects from an area (think planar scintigraphy). After the series of projections are acquired then, (lots of scintigraphy images from different angles), cross sections are extracted, yielding a reconstructed volume. (See intro ppt for illustration) PET instrumentation: 19. In PET, positron (antielectron: +1 charge, same rest mass as electron) emitting radionuclides are used. The positron travels up to several mm in the tissue, meets a free electron, and mutual annihilation occurs. From conservation of, two gamma photons appear and from conservation of, they are emitted back-to-back. 20. It is possible to use an Anger camera to detect just one of the emitted photons. Images could then be reconstructed using planar scintigraphy or SPECT reconstruction (if the camera rotates). HOWEVER, what sharply distinguishes PET from SPECT is the use of rings of detectors and circuitry for Annihilation Coincidence Detection. With this detection, only simultaneous gamma rays are declared events, and all others are rejected. If two opposing detectors simultaneously detect the gamma rays, the annihilation event must have occurred somewhere along that line. That line is called the. *This concept makes detector collimation not only unnecessary, but undesirable.* Note: Because PET and CT both require detector rotation and PET does not need collimation, it is state-of-theart to fuse anatomic and functional imaging with PET/CT systems, where detectors are rotated on the same gantry, which leads to effective image registration between the two modalities. In addition, the CT maps attenuation, which is a confounding factor in nuclear medicine. The CT image can be used then for image correction on the nuclear medicine image. SPECT Image Formation : Notes on frame of reference (see Figure9.6): In developing the imaging equation for planar scintigraphy, the x-y plane represented the detector surface coordinates above the patient and z the depth into the patient. To study SPECT and PET, we will return to notation we used when developing CT reconstruction. X-Y space is in the axial slice of the patient and z runs head-to-toe on the patient. As with CT, l represents the lateral position of the line integral and θ corresponds to the angle of detector rotation. See figure 9.7 for clarity. 21. Our imaging equation with appropriate spatial variable changes from above, can be written as eqn 9.5 (, ) A( x( s), y( s)) s 2 e ds 4 ( s R) R ( x( s'), y( s'); E) ds' R

7 keeping in mind that the parallel reconstructions along z are a separate computing issue after data acquisition. What are the difficulties with reconstructing an image from the above equation? 22. What simplifications are commonly used to get us to equation 9.6? Rewrite Equation Use the line impulse function to rewrite the imaging equation exactly in the form of the CT imaging equation. (see eqn 9.7) 24.. where we have the following relationships: Desired object to reconstruct: in CT (attenuation map) in SPECT (activity and location) Data acquired (projection filled space from many angles): in CT in SPECT One important difference: SPECT uses actual data recorded by detectors (number of photons) in reconstruction. CT requires first a logarithmic translation of intensities. (see equation 6.3 and 6.4). 25. Therefore, the reconstruction technique of choice is the convolution backprojection (the typically implemented form of filtered backprojection). Rewrite Equation 9.10, the reconstructed activity map, A(x,y): PET Image Formation : Notes on frame of reference: In developing the imaging equations for PET, we will be considering only 2D PET, since that is the natural form it takes. We therefore assume a fixed ring of detectors receiving data from a given z plane. The cross-sectional geometry then remains the same as CT and SPECT.

8 As with planar scintigraphy and SPECT, PET counts the number of detected photons, BUT these counts are now based on ACDs rather than single gamma ray detections. Note that it is impossible to tell where the event occurred on the line of response between two detectors 180 degrees apart. Therefore, like SPECT and CT, PET measures over line integrals. 26. The imaging equation for PET can be written as 9.16 R ( x( s'), y( s'); E) ds' R (, ) K A( x( s), y( s)) e ds R Explain briefly the source of the difference in the integration limits between this equation and equation 9.5 that was developed (and unsolvable) for SPECT. R 27. Ignoring attenuation, as we did for SPECT, the activity map reconstruction equation can be written, just as with SPECT as the inverse radon transform using the convolution backprojection method: A( x, y) (, ) c( xcos ysin ) d d 0 where we have the following relationships: Desired object to reconstruct: in CT (attenuation map) in PET (activity and location) Data acquired (projection filled space from many angles): in CT in PET 28. Unlike SPECT, PET has a straightforward approach for handling attenuation due to the fact that the integration limits are different (see your answer to 26), giving a closed-form solution for attenuation correction. i.e. In the case of PET (unlike SPECT) the attenuation term and the activity term can be separated. Corrected coincidence allows for a more accurate reconstruction of the integral of activity along the line of response, unaffected by attenuation in the body. See equation 9.22 for clarification. How is this correction data acquired for PET?

9 29. is a newer, more computer intensive approach to reconstruction in PET and SPECT that implicitly takes the random nature of decay into account and can incorporate models of attenuation, scatter, and blur. Image Quality in PET and SPECT : 30. Image quality in SPECT and PET is limited by (see summary and key points ) a. b. c. 31. Rewrite the final expression for the reconstructed (blurred) image as related to the object (eqn. 9.26): where h(r) represents the cascaded blurring sources. 32. This blurring in PET and SPECT also affects the axial resolution (note that this is mistakenly referred to as slice thickness, which we will discuss in MRI). This effect is sometimes referred to as (we will use this same term later in MRI). 33. The deterministic effects (attenuation, scatter, random coincidences) influence the signal component of SNR; in other words, they affect an image s. Noise arises from random statistical nature of radioactive decay and (somewhat) the hardware, influencing the noise component of SNR; in other words, affecting the of the technique.

Emission Computed Tomography Notes

Emission Computed Tomography Notes Noll (24) ECT Notes: Page 1 Emission Computed Tomography Notes Introduction Emission computed tomography (ECT) is the CT applied to nuclear medicine. There are two varieties of ECT: 1. SPECT single-photon

More information

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 5 SPECT, PET Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes

More information

Introduction to Positron Emission Tomography

Introduction to Positron Emission Tomography Planar and SPECT Cameras Summary Introduction to Positron Emission Tomography, Ph.D. Nuclear Medicine Basic Science Lectures srbowen@uw.edu System components: Collimator Detector Electronics Collimator

More information

Constructing System Matrices for SPECT Simulations and Reconstructions

Constructing System Matrices for SPECT Simulations and Reconstructions Constructing System Matrices for SPECT Simulations and Reconstructions Nirantha Balagopal April 28th, 2017 M.S. Report The University of Arizona College of Optical Sciences 1 Acknowledgement I would like

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Workshop on Quantitative SPECT and PET Brain Studies 14-16 January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Físico João Alfredo Borges, Me. Corrections in SPECT and PET SPECT and

More information

Positron Emission Tomography

Positron Emission Tomography Physics 656 Seminar on Physical Fundamentals of Medical Imaging Positron Emission Tomography Ahmed Qamesh Outline What is PET? PET mechanism Radionuclide and its synthesis Detection concept and Development

More information

Continuation Format Page

Continuation Format Page C.1 PET with submillimeter spatial resolution Figure 2 shows two views of the high resolution PET experimental setup used to acquire preliminary data [92]. The mechanics of the proposed system are similar

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

Introduction to Emission Tomography

Introduction to Emission Tomography Introduction to Emission Tomography Gamma Camera Planar Imaging Robert Miyaoka, PhD University of Washington Department of Radiology rmiyaoka@u.washington.edu Gamma Camera: - collimator - detector (crystal

More information

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology Diagnostic imaging techniques Krasznai Zoltán University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology 1. Computer tomography (CT) 2. Gamma camera 3. Single Photon

More information

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar Doctoral Thesis Time-of-Flight Time-of-Flight Positron Positron Emission Emission Tomography Tomography Using Using Cherenkov Cherenkov Radiation Radiation Rok Dolenec Supervisor: Prof. Dr. Samo Korpar

More information

Fits you like no other

Fits you like no other Fits you like no other BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase in room

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

Central Slice Theorem

Central Slice Theorem Central Slice Theorem Incident X-rays y f(x,y) R x r x Detected p(, x ) The thick line is described by xcos +ysin =R Properties of Fourier Transform F [ f ( x a)] F [ f ( x)] e j 2 a Spatial Domain Spatial

More information

Fits you like no other

Fits you like no other Fits you like no other Philips BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase

More information

MEDICAL IMAGE ANALYSIS

MEDICAL IMAGE ANALYSIS SECOND EDITION MEDICAL IMAGE ANALYSIS ATAM P. DHAWAN g, A B IEEE Engineering in Medicine and Biology Society, Sponsor IEEE Press Series in Biomedical Engineering Metin Akay, Series Editor +IEEE IEEE PRESS

More information

Positron. MillenniumVG. Emission Tomography Imaging with the. GE Medical Systems

Positron. MillenniumVG. Emission Tomography Imaging with the. GE Medical Systems Positron Emission Tomography Imaging with the MillenniumVG GE Medical Systems Table of Contents Introduction... 3 PET Imaging With Gamma Cameras PET Overview... 4 Coincidence Detection on Gamma Cameras...

More information

Medical Image Analysis

Medical Image Analysis Computer assisted Image Analysis VT04 29 april 2004 Medical Image Analysis Lecture 10 (part 1) Xavier Tizon Medical Image Processing Medical imaging modalities XRay,, CT Ultrasound MRI PET, SPECT Generic

More information

CHAPTER 11 NUCLEAR MEDICINE IMAGING DEVICES

CHAPTER 11 NUCLEAR MEDICINE IMAGING DEVICES CHAPTER 11 M.A. LODGE, E.C. FREY Russell H. Morgan Department of Radiology and Radiological Sciences, Johns Hopkins University, Baltimore, Maryland, United States of America 11.1. INTRODUCTION Imaging

More information

Computational Medical Imaging Analysis

Computational Medical Imaging Analysis Computational Medical Imaging Analysis Chapter 2: Image Acquisition Systems Jun Zhang Laboratory for Computational Medical Imaging & Data Analysis Department of Computer Science University of Kentucky

More information

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Review of PET Physics Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Chart of Nuclides Z (protons) N (number of neutrons) Nuclear Data Evaluation Lab.

More information

Tomography. Forward projectionsp θ (r) are known as a Radon transform. Objective: reverse this process to form the original image

Tomography. Forward projectionsp θ (r) are known as a Radon transform. Objective: reverse this process to form the original image C. A. Bouman: Digital Image Processing - January 9, 217 1 Tomography Many medical imaging systems can only measure projections through an object with density f(x,y). Projections must be collected at every

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

An educational tool for demonstrating the TOF-PET technique

An educational tool for demonstrating the TOF-PET technique Nuclear Instruments and Methods in Physics Research A 471 (2001) 200 204 An educational tool for demonstrating the TOF-PET technique T.Bȧack a, *, J. Cederkȧall a, B. Cederwall a, A. Johnson a, A. Kerek

More information

Lecture 6: Medical imaging and image-guided interventions

Lecture 6: Medical imaging and image-guided interventions ME 328: Medical Robotics Winter 2019 Lecture 6: Medical imaging and image-guided interventions Allison Okamura Stanford University Updates Assignment 3 Due this Thursday, Jan. 31 Note that this assignment

More information

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC What is needed? Why? How often? Who says? QA and QC in Nuclear Medicine QA - collective term for all the efforts made to produce

More information

Medical Imaging Modalities

Medical Imaging Modalities Image Science Introduction Medical Imaging Modalities Ho Kyung Kim hokyung@pusan.ac.kr Pusan National University Projection Radiography Routine diagnostic radiography Chest x rays, fluoroscopy, mammography,

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!

More information

SNIC Symposium, Stanford, California April The Hybrid Parallel Plates Gas Counter for Medical Imaging

SNIC Symposium, Stanford, California April The Hybrid Parallel Plates Gas Counter for Medical Imaging The Hybrid Parallel Plates Gas Counter for Medical Imaging F. Anulli, G. Bencivenni, C. D Ambrosio, D. Domenici, G. Felici, F. Murtas Laboratori Nazionali di Frascati - INFN, Via E. Fermi 40, I-00044 Frascati,

More information

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging

More information

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives 1. Be familiar with recommendations

More information

ISO ISO ISO OHSAS ISO

ISO ISO ISO OHSAS ISO ISO 9001 ISO 13485 ISO 14001 OHSAS 18001 ISO 27001 Pro-NM Performance 08-101 - standard version 08-103 - version with the PET Lid Phantom for NM and PET systems performance evaluation (collimator, artifacts,

More information

Tomographic Reconstruction

Tomographic Reconstruction Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)

More information

UNIVERSITY OF SOUTHAMPTON

UNIVERSITY OF SOUTHAMPTON UNIVERSITY OF SOUTHAMPTON PHYS2007W1 SEMESTER 2 EXAMINATION 2014-2015 MEDICAL PHYSICS Duration: 120 MINS (2 hours) This paper contains 10 questions. Answer all questions in Section A and only two questions

More information

ANALYSIS OF CT AND PET/SPECT IMAGES FOR DOSIMETRY CALCULATION

ANALYSIS OF CT AND PET/SPECT IMAGES FOR DOSIMETRY CALCULATION 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 ANALYSIS OF

More information

Biomedical Imaging. Computed Tomography. Patrícia Figueiredo IST

Biomedical Imaging. Computed Tomography. Patrícia Figueiredo IST Biomedical Imaging Computed Tomography Patrícia Figueiredo IST 2013-2014 Overview Basic principles X ray attenuation projection Slice selection and line projections Projection reconstruction Instrumentation

More information

Development and Performance of a Sparsity- Exploiting Algorithm for Few-View Single Photon Emission Computed Tomogrpahy (SPECT) Reconstruction

Development and Performance of a Sparsity- Exploiting Algorithm for Few-View Single Photon Emission Computed Tomogrpahy (SPECT) Reconstruction Marquette University e-publications@marquette Master's Theses (2009 -) Dissertations, Theses, and Professional Projects Development and Performance of a Sparsity- Exploiting Algorithm for Few-View Single

More information

Medical Imaging Introduction

Medical Imaging Introduction Medical Imaging Introduction Jan Kybic February 16, 2010 Medical imaging: a collaborative paradigm picture from Atam P. Dhawan: Medical Imaging From physiology to information processing (what we should

More information

Joint CI-JAI advanced accelerator lecture series Imaging and detectors for medical physics Lecture 1: Medical imaging

Joint CI-JAI advanced accelerator lecture series Imaging and detectors for medical physics Lecture 1: Medical imaging Joint CI-JAI advanced accelerator lecture series Imaging and detectors for medical physics Lecture 1: Medical imaging Dr Barbara Camanzi barbara.camanzi@stfc.ac.uk Course layout Day AM 09.30 11.00 PM 15.30

More information

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS 23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS P. Helberg 1,

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 61675-2 First edition 1998-01 Radionuclide imaging devices Characteristics and test conditions Part 2: Single photon emission computed tomographs Dispositifs d imagerie par radionucléides

More information

Reconstruction in CT and relation to other imaging modalities

Reconstruction in CT and relation to other imaging modalities Reconstruction in CT and relation to other imaging modalities Jørgen Arendt Jensen November 1, 2017 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering Center for Fast Ultrasound

More information

James R Halama, PhD Loyola University Medical Center

James R Halama, PhD Loyola University Medical Center James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives Be familiar with the tests recommended

More information

Fast Timing and TOF in PET Medical Imaging

Fast Timing and TOF in PET Medical Imaging Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part

More information

Extremely Fast Detector for 511 kev Gamma

Extremely Fast Detector for 511 kev Gamma Extremely Fast Detector for 511 kev Gamma V. Sharyy, D. Yvon, G. Tauzin, E.Delagnes, Ph. Abbon, J P. Bard, M. Kebbiri, M. Alokhina, C. Canot IRFU, CEA D. Breton, J. Maalmi LAL,IN2P3 Journée 2015 du Labex

More information

FRONT-END DATA PROCESSING OF NEW POSITRON EMIS- SION TOMOGRAPHY DEMONSTRATOR

FRONT-END DATA PROCESSING OF NEW POSITRON EMIS- SION TOMOGRAPHY DEMONSTRATOR SOUDABEH MORADI FRONT-END DATA PROCESSING OF NEW POSITRON EMIS- SION TOMOGRAPHY DEMONSTRATOR Master of Science Thesis Examiners: Prof. Ulla Ruotsalainen MSc Defne Us Examiners and topic approved by the

More information

Computational Medical Imaging Analysis

Computational Medical Imaging Analysis Computational Medical Imaging Analysis Chapter 1: Introduction to Imaging Science Jun Zhang Laboratory for Computational Medical Imaging & Data Analysis Department of Computer Science University of Kentucky

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

Medical Images Analysis and Processing

Medical Images Analysis and Processing Medical Images Analysis and Processing - 25642 Emad Course Introduction Course Information: Type: Graduated Credits: 3 Prerequisites: Digital Image Processing Course Introduction Reference(s): Insight

More information

Evaluation of Centrally Located Sources in. Coincidence Timing Calibration for Time-of-Flight PET

Evaluation of Centrally Located Sources in. Coincidence Timing Calibration for Time-of-Flight PET Evaluation of Centrally Located Sources in Coincidence Timing Calibration for Time-of-Flight PET by Richard Ryan Wargo Graduate Program in Medical Physics Duke University Date: Approved: Timothy G. Turkington,

More information

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh et al. : Proceedings of the National Seminar & Exhibition on Non-Destructive Evaluation DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh V Acharya, Umesh Kumar, Gursharan

More information

Index. aliasing artifacts and noise in CT images, 200 measurement of projection data, nondiffracting

Index. aliasing artifacts and noise in CT images, 200 measurement of projection data, nondiffracting Index Algebraic equations solution by Kaczmarz method, 278 Algebraic reconstruction techniques, 283-84 sequential, 289, 293 simultaneous, 285-92 Algebraic techniques reconstruction algorithms, 275-96 Algorithms

More information

Recognition and Measurement of Small Defects in ICT Testing

Recognition and Measurement of Small Defects in ICT Testing 19 th World Conference on Non-Destructive Testing 2016 Recognition and Measurement of Small Defects in ICT Testing Guo ZHIMIN, Ni PEIJUN, Zhang WEIGUO, Qi ZICHENG Inner Mongolia Metallic Materials Research

More information

COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER

COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER By RAMI RIMON ABU-AITA A THESIS PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY

More information

Reconstruction from Projections

Reconstruction from Projections Reconstruction from Projections M.C. Villa Uriol Computational Imaging Lab email: cruz.villa@upf.edu web: http://www.cilab.upf.edu Based on SPECT reconstruction Martin Šámal Charles University Prague,

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

SPECT reconstruction

SPECT reconstruction Regional Training Workshop Advanced Image Processing of SPECT Studies Tygerberg Hospital, 19-23 April 2004 SPECT reconstruction Martin Šámal Charles University Prague, Czech Republic samal@cesnet.cz Tomography

More information

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed

More information

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic

More information

Midterm Review. Yao Wang Polytechnic University, Brooklyn, NY 11201

Midterm Review. Yao Wang Polytechnic University, Brooklyn, NY 11201 Midterm Review Yao Wang Polytechnic University, Brooklyn, NY 11201 Based on J. L. Prince and J. M. Links, Medical maging Signals and Systems, and lecture notes by Prince. Figures are from the textbook.

More information

in PET Medical Imaging

in PET Medical Imaging Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part

More information

3-D PET Scatter Correction

3-D PET Scatter Correction Investigation of Accelerated Monte Carlo Techniques for PET Simulation and 3-D PET Scatter Correction C.H. Holdsworth, Student Member, IEEE, C.S. Levin", Member, IEEE, T.H. Farquhar, Student Member, IEEE,

More information

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

C a t p h a n / T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Biophysical Techniques (BPHS 4090/PHYS 5800)

Biophysical Techniques (BPHS 4090/PHYS 5800) Biophysical Techniques (BPHS 4090/PHYS 5800) Instructors: Prof. Christopher Bergevin (cberge@yorku.ca) Schedule: MWF 1:30-2:30 (CB 122) Website: http://www.yorku.ca/cberge/4090w2017.html York University

More information

Nuclear Instruments and Methods in Physics Research A

Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 621 (2010) 590 594 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

The great interest shown toward PET instrumentation is

The great interest shown toward PET instrumentation is CONTINUING EDUCATION PET Instrumentation and Reconstruction Algorithms in Whole-Body Applications* Gabriele Tarantola, BE; Felicia Zito, MSc; and Paolo Gerundini, MD Department of Nuclear Medicine, Ospedale

More information

Detection of Lesions in Positron Emission Tomography

Detection of Lesions in Positron Emission Tomography Detection of Lesions in Positron Emission Tomography Bachelor Thesis Nina L.F. Bezem Study: Physics and Astronomy Faculty of Science Supervised by: Dr. Andre Mischke Utrecht University, Institute for Subatomic

More information

Some reference material

Some reference material Some reference material Physics reference book on medical imaging: A good one is The Essential Physics of Medical Imaging, 3 rd Ed. by Bushberg et al. ($170! new). However, there are several similar books

More information

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam Home Search Collections Journals About Contact us My IOPscience Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam projections This content has been downloaded from IOPscience.

More information

Ultrasonic Multi-Skip Tomography for Pipe Inspection

Ultrasonic Multi-Skip Tomography for Pipe Inspection 18 th World Conference on Non destructive Testing, 16-2 April 212, Durban, South Africa Ultrasonic Multi-Skip Tomography for Pipe Inspection Arno VOLKER 1, Rik VOS 1 Alan HUNTER 1 1 TNO, Stieltjesweg 1,

More information

GPU implementation for rapid iterative image reconstruction algorithm

GPU implementation for rapid iterative image reconstruction algorithm GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

Radon Transform and Filtered Backprojection

Radon Transform and Filtered Backprojection Radon Transform and Filtered Backprojection Jørgen Arendt Jensen October 13, 2016 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering Center for Fast Ultrasound Imaging Department

More information

Moscow-Bavarian Joint Advanced Student School 2006 / Medical Imaging Principles of Computerized Tomographic Imaging and Cone-Beam Reconstruction

Moscow-Bavarian Joint Advanced Student School 2006 / Medical Imaging Principles of Computerized Tomographic Imaging and Cone-Beam Reconstruction Line Integrals Line integrals represent the integral of some parameter of the object along the line (e.g. attenuation of x-rays) Object: f(x,y) Line: x cosθ + y sinθ = t Line integral / Radon transform:

More information

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry Sasha A. Philips, Frazier Bronson, Ram Venkataraman, Brian M. Young Abstract--Activity measurements require knowledge of the

More information

Small Angle Gamma Ray Scatter: What Is The Impact On Image Quality

Small Angle Gamma Ray Scatter: What Is The Impact On Image Quality ISPUB.COM The Internet Journal of Medical Technology Volume 4 Number 2 Small Angle Gamma Ray Scatter: What Is The Impact On Image Quality G CUrrie, J Wheat Citation G CUrrie, J Wheat.. The Internet Journal

More information

CT Systems and their standards

CT Systems and their standards CT Systems and their standards Stephen Brown Engineering Measurement 11 th April 2012 Industrial X-ray computed tomography: The future of co-ordinate metrology? Burleigh Court, Loughborough University

More information

Computed tomography (Item No.: P )

Computed tomography (Item No.: P ) Computed tomography (Item No.: P2550100) Curricular Relevance Area of Expertise: Biology Education Level: University Topic: Modern Imaging Methods Subtopic: X-ray Imaging Experiment: Computed tomography

More information

CIVA Computed Tomography Modeling

CIVA Computed Tomography Modeling CIVA Computed Tomography Modeling R. FERNANDEZ, EXTENDE, France S. LEGOUPIL, M. COSTIN, D. TISSEUR, A. LEVEQUE, CEA-LIST, France page 1 Summary Context From CIVA RT to CIVA CT Reconstruction Methods Applications

More information

AN ELLIPTICAL ORBIT BACKPROJECTION FILTERING ALGORITHM FOR SPECT

AN ELLIPTICAL ORBIT BACKPROJECTION FILTERING ALGORITHM FOR SPECT 1102 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 40, NO. 4, AUGUST 1993 AN ELLIPTICAL ORBIT BACKPROJECTION FILTERING ALGORITHM FOR SPECT Grant. T. Gullberg and Gengsheng L. Zeng, Department of Radiology,

More information

CoE4TN4 Image Processing. Chapter 5 Image Restoration and Reconstruction

CoE4TN4 Image Processing. Chapter 5 Image Restoration and Reconstruction CoE4TN4 Image Processing Chapter 5 Image Restoration and Reconstruction Image Restoration Similar to image enhancement, the ultimate goal of restoration techniques is to improve an image Restoration: a

More information

Computed Tomography. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Sep 26th/Oct 3rd, 2016

Computed Tomography. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Sep 26th/Oct 3rd, 2016 Computed Tomography Principles of Medical Imaging Prof. Dr. Philippe Cattin MIAC, University of Basel Contents Abstract 1 Computed Tomography Basics Introduction Computed Tomography Hounsfield's CT Prototype

More information

Iterative SPECT reconstruction with 3D detector response

Iterative SPECT reconstruction with 3D detector response Iterative SPECT reconstruction with 3D detector response Jeffrey A. Fessler and Anastasia Yendiki COMMUNICATIONS & SIGNAL PROCESSING LABORATORY Department of Electrical Engineering and Computer Science

More information

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE / Radiographic Tomosynthesis: Image Quality and Artifacts Reduction Baojun Li, Ph.D Department of Radiology Boston University Medical Center 2012 AAPM Annual Meeting Background Linear Trajectory Tomosynthesis

More information

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Joonki Noh, Jeffrey A. Fessler EECS Department, The University of Michigan Paul E. Kinahan Radiology Department,

More information

Engineered Diffusers Intensity vs Irradiance

Engineered Diffusers Intensity vs Irradiance Engineered Diffusers Intensity vs Irradiance Engineered Diffusers are specified by their divergence angle and intensity profile. The divergence angle usually is given as the width of the intensity distribution

More information

Effect of Scattering on the Image. Reducing Compton Scatter with a Grid

Effect of Scattering on the Image. Reducing Compton Scatter with a Grid Effect of Scattering on the Image Increasing Compton scattering degrades image. Webb 21 Reducing Compton Scatter with a Grid Grids Parallel (focused at infinity) Linear Focused (see figure) Moving grids

More information

Quantitative capabilities of four state-of-the-art SPECT-CT cameras

Quantitative capabilities of four state-of-the-art SPECT-CT cameras Seret et al. EJNMMI Research 2012, 2:45 ORIGINAL RESEARCH Open Access Quantitative capabilities of four state-of-the-art SPECT-CT cameras Alain Seret 1,2*, Daniel Nguyen 1 and Claire Bernard 3 Abstract

More information

Image Reconstruction Methods for Dedicated Nuclear Medicine Detectors for Prostate Imaging

Image Reconstruction Methods for Dedicated Nuclear Medicine Detectors for Prostate Imaging Image Reconstruction Methods for Dedicated Nuclear Medicine Detectors for Prostate Imaging Mark F. Smith, Ph.D. Detector and Imaging Group, Physics Division Thomas Jefferson National Accelerator Facility

More information

Computed Tomography January 2002 KTH A.K.

Computed Tomography January 2002 KTH A.K. CT A.K. Computed Tomography January KTH 1 Introduction X-ray was discovered (accidentally) by a German physicist, Wilhelm Konrad Röntgen in 1895. A few years later, in 191, Röntgen was awarded the first

More information

Image reconstruction for PET/CT scanners: past achievements and future challenges

Image reconstruction for PET/CT scanners: past achievements and future challenges Review Image reconstruction for PET/CT scanners: past achievements and future challenges PET is a medical imaging modality with proven clinical value for disease diagnosis and treatment monitoring. The

More information

Mediso AnyScan. Single-Head and Dual-Head SPECT

Mediso AnyScan. Single-Head and Dual-Head SPECT Mediso AnyScan S Single-Head and Dual-Head SPECT ANYSCAN S (SINGLE-HEAD & DUAL-HEAD SPECT) Low Cost of Ownership Associated Imaging Services is the exclusive U.S. source in the Midwest for Mediso Medical

More information

MOHAMMAD MINHAZ AKRAM THE EFFECT OF SAMPLING IN HISTOGRAMMING AND ANALYTICAL RECONSTRUCTION OF 3D AX-PET DATA

MOHAMMAD MINHAZ AKRAM THE EFFECT OF SAMPLING IN HISTOGRAMMING AND ANALYTICAL RECONSTRUCTION OF 3D AX-PET DATA MOHAMMAD MINHAZ AKRAM THE EFFECT OF SAMPLING IN HISTOGRAMMING AND ANALYTICAL RECONSTRUCTION OF 3D AX-PET DATA Master of Science Thesis Examiners: Prof. Ulla Ruotsalainen M.Sc. Uygar Tuna Examiners and

More information