1 Mech. Engineering, Comp. Science, and Rad. Oncology Departments Schools of Engineering and Medicine, Bio-X Program, Stanford University 1
2 Conflict of Interest Nothing to disclose 2
3 Imaging During Beam Delivery Existing solutions are limited: Radiographic x-ray Electromagnetic Real-time marker-less soft-tissue image guidance during beam delivery is an unmet challenge 3
4 Novel Image Guidance Solution Linear Accelerator Haptic Interface Accelerator control console Treatment Intervention Patient 4D US probe Optical tracker US-guidance workstation computer US Robot Robot Control 3D US image stream US imaging system Probe position data (6 DOF) Telerobotic system enables remote probe control 4
5 Telerobotic Imaging Remote Haptic Interface Robot 5
6 Key Issues Remote ultrasound imaging during beam delivery! Robotic manipulator design! Treatment plan compatibility! Performance during radiation exposure Imaging robustness for multiple treatment sites Image guidance Temporal calibration and time delay Spatial calibration and accuracy! Details to appear in Medical Physics: Schlosser et al. (2010) 6
7 Telerobotic Imaging for Multiple Sites Prostate Volunteer Liver Volunteer Kidney Volunteer Pitch Force Time [sec] Image quality remotely maintained over 10 minutes 7
8 Image Guidance: Spatial Calibration unknown im T pr pr T tr tr T w F image F probe F tracker F treatment Goal: find im T pr Variation of planar fit method from Hartov et. al* *Hartov et al., Adaptive spatial calibration of a 3D ultrasound system, Med. Phys.,
9 Image Guidance: Spatial Calibration (1) Collect images of plate (2) Extract points on plane w p= w T tr * tr T pr * pr T im * im p (4) Optimize planar data fit (3) Convert to world frame 9
10 Temporal Calibration and System Evaluation: Experimental Method Image static target in US phantom Vary probe pitch or pressure Track target in real-time using NCC Use transformation chain to register in world frame Guidance Software Robot US probe Optical Tracker Phantom 10
11 Temporal Calibration and System Evaluation: Experimental Method Image static target in US phantom Vary probe pitch or pressure Track target in real-time using NCC Use transformation chain to register in world frame Guidance Software Robot US probe Optical Tracker Phantom 11
12 Temporal Calibration and System Evaluation: Experimental Method Image static target in US phantom Vary probe pitch or pressure Track target in real-time using NCC Use transformation chain to register in world frame Guidance Software Robot US probe Optical Tracker Phantom 12
13 Temporal Calibration and System Evaluation: Experimental Method Image static target in US phantom Vary probe pitch or pressure Track target in real-time using NCC Use transformation chain to register in world frame Guidance Software Robot US probe Optical Tracker Phantom 13
14 Temporal Calibration and System Evaluation: Experimental Method Image static target in US phantom Vary probe pitch or pressure Track target in real-time using NCC Use transformation chain to register in world frame Guidance Software Robot US probe Optical Tracker Phantom 14
15 Temporal Calibration Novel procedure: Robot pitches US probe w/ sinusoidal motion Static target imaged and tracked using NCC Optical tracker data compared to US pixel data Raw Data Sinusoid Model 92ms 15
16 Evaluation: Spatial Localization Error World Space Image Space World Space Magnified Variable Pitch Pressure Experimental Parameters -5 to Hz Tracking Error 0.4± 0.2 mm Max Tracking Error Real-time phantom localization error < 1mm 0 to 12N 0.6± 0.3mm 0.8 mm 1.3 mm 16
17 Evaluation: Spatial Localization Error World Space Image Space World Space Magnified Variable Pitch Pressure Experimental Parameters -5 to Hz Tracking Error 0.4± 0.2 mm Max Tracking Error Real-time phantom localization error < 1mm 0 to 12N 0.6± 0.3mm 0.8 mm 1.3 mm 17
18 Evaluation: Spatial Localization Error World Space Image Space World Space Magnified Variable Pitch Pressure Experimental Parameters -5 to Hz Tracking Error 0.4± 0.2 mm Max Tracking Error Real-time phantom localization error < 1mm 0 to 12N 0.6± 0.3mm 0.8 mm 1.3 mm 18
19 Evaluation: Spatial Localization Error World Space Image Space World Space Magnified Variable Pitch Pressure Experimental Parameters -5 to Hz Tracking Error 0.4± 0.2 mm Max Tracking Error Real-time phantom localization error < 1mm 0 to 12N 0.6± 0.3mm 0.8 mm 1.3 mm 19
20 Evaluation: Time Lag Experimental method: Similar to time calibration Robot trajectory (1000Hz) compared to target trajectory in ultrasound sequence Raw Data Sinusoid Model 179ms 20
21 Real-Time Guidance at 5 Hz and Sub-millimeter Localization Accuracy 21
22 Conclusions Remotely-controlled soft-tissue imaging in the treatment vault is feasible for multiple abdominal sites Sub-millimeter targeting accuracy with 2D imaging indicates feasibility for accurate real-time 3D ultrasound guidance Telerobotic US guidance system could offer non-invasive localization for IGRT that truly reflects soft-tissue anatomy 22
23 Questions? Investigator s: Mech. Engineering, Comp. Science, and Rad. Oncology Departments Schools of Engineering and Medicine, Bio-X Program, Stanford University 23
The team. Disclosures. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge.
Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge Dimitre Hristov Radiation Oncology Stanford University The team Renhui Gong 1 Magdalena Bazalova-Carter 1
More informationImage Guidance and Beam Level Imaging in Digital Linacs
Image Guidance and Beam Level Imaging in Digital Linacs Ruijiang Li, Ph.D. Department of Radiation Oncology Stanford University School of Medicine 2014 AAPM Therapy Educational Course Disclosure Research
More informationSystem Integration and Preliminary In-Vivo Experiments of a Robot for Ultrasound Guidance and Monitoring during Radiotherapy
System Integration and Preliminary In-Vivo Experiments of a Robot for Ultrasound Guidance and Monitoring during Radiotherapy H. Tutkun Şen, Muyinatu A. Lediju Bell, Yin Zhang, Kai Ding, John Wong, Iulian
More informationThank-You Members of TG147 TG 147: QA for nonradiographic
Thank-You Members of TG147 TG 147: QA for nonradiographic localization and positioning systems Twyla Willoughby, M.S. Medical Physicist Clinical AAPM Meeting March 2013 Department of Radiation Oncology
More informationSelf-Calibration of Cone-Beam CT Using 3D-2D Image Registration
Self-Calibration of Cone-Beam CT Using 3D-2D Image Registration Sarah Ouadah 1 J. Webster Stayman 1 Grace Jianan Gang 1 Ali Uneri 2 Tina Ehtiati 3 Jeffrey H. Siewerdsen 1,2 1. Department of Biomedical
More informationABSTRACT 1. INTRODUCTION
Tracked ultrasound calibration studies with a phantom made of LEGO bricks Marie Soehl, Ryan Walsh, Adam Rankin, Andras Lasso, Gabor Fichtinger Queen s University, Kingston ON, Canada ABSTRACT PURPOSE:
More information8/3/2016. Image Guidance Technologies. Introduction. Outline
8/3/26 Session: Image Guidance Technologies and Management Strategies Image Guidance Technologies Jenghwa Chang, Ph.D.,2 Department of Radiation Medicine, Northwell Health 2 Hofstra Northwell School of
More informationMathematical methods and simulations tools useful in medical radiation physics
Mathematical methods and simulations tools useful in medical radiation physics Michael Ljungberg, professor Department of Medical Radiation Physics Lund University SE-221 85 Lund, Sweden Major topic 1:
More informationA Study of Motion Tracking Accuracy of Robotic Radiosurgery Using a Novel CCD Camera Based End-to-end Test System
A Study of Motion Tracking Accuracy of Robotic Radiosurgery Using a Novel CCD Camera Based End-to-end Test System Lei Wang 1, Brett Nelson 2,Youming Yang 1 1. Department of Radiation Oncology, Stanford
More informationActive Echo: A New Paradigm for Ultrasound Calibration
Active Echo: A New Paradigm for Ultrasound Calibration Xiaoyu Guo, Alexis Cheng, Haichong K. Zhang, Hyun-Jae Kang, Ralph Etienne-Cummings, and Emad M. Boctor The Johns Hopkins University, Baltimore, MD
More informationCommissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility
Technical Note Commissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility K. R. Muralidhar, Krishna Komanduri, Birendra Kumar Rout, K. K.
More informationAutomated Quality Assurance for Image-Guided Radiation Therapy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 1, WINTER 2009 Automated Quality Assurance for Image-Guided Radiation Therapy Eduard Schreibmann, a Eric Elder, Tim Fox Department of Radiation
More informationCommissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility
0 0 0 0 JMP R Commissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility K. R. Muralidhar, Krishna Komanduri, Birendra Kumar Rout, K. K. D.
More information7/31/ D spatial compounding with the Clarity system: Improving ultrasound image quality in gynaecological image-guided radiotherapy
in partnership with 3D spatial compounding with the Clarity system: Improving ultrasound image quality in gynaecological image-guided radiotherapy Sarah Mason, Ingrid White, Helen McNair, Susan Lalondrelle,
More informationNew Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC
New Technology in Radiation Oncology James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC Technology s s everywhere From the imaging chain To the planning system To the linac To QA..it..it
More informationCALIBRATION ALGORITHM FOR FREEHAND ULTRASOUND
CALIBRAION ALGORIHM FOR FREEHAND ULRASOUND 1 YAO RAO, YAO PENG 1 Aeronautic College, Shanghai University of Engineering Science, Shanghai 0160, China School of Economics & Management, Beihua University,
More informationFacility Questionnaire PART I (General Information for 3DCRT and IMRT)
Facility Questionnaire PART I (General Information for 3DCRT and IMRT) The following items are required before you can enter cases on any RTOG protocol that requires data submission to the Image-Guided
More informationCalibration Method for Determining the Physical Location of the Ultrasound Image Plane
Calibration Method for Determining the Physical Location of the Ultrasound Image Plane Devin V. Amin, Ph.D. 1, Takeo Kanade, Ph.D 1., Branislav Jaramaz, Ph.D. 2, Anthony M. DiGioia III, MD 2, Constantinos
More informationSuturing in Confined Spaces: Constrained Motion Control of a Hybrid 8-DoF Robot
Suturing in Confined Spaces: Constrained Motion Control of a Hybrid 8-DoF Robot Ankur Kapoor 1, Nabil Simaan 2, Russell H. Taylor 1 1 ERC-CISST Department of Computer Science Johns Hopkins University 2
More informationBME I5000: Biomedical Imaging
1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.
More informationOptical Guidance. Sanford L. Meeks. July 22, 2010
Optical Guidance Sanford L. Meeks July 22, 2010 Optical Tracking Optical tracking is a means of determining in real-time the position of a patient relative to the treatment unit. Markerbased systems track
More informationData. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy
Data ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy The ModuLeaf Mini Multileaf
More information3D Ultrasound System Using a Magneto-optic Hybrid Tracker for Augmented Reality Visualization in Laparoscopic Liver Surgery
3D Ultrasound System Using a Magneto-optic Hybrid Tracker for Augmented Reality Visualization in Laparoscopic Liver Surgery Masahiko Nakamoto 1, Yoshinobu Sato 1, Masaki Miyamoto 1, Yoshikazu Nakamjima
More informationValidation System of MR Image Overlay and Other Needle Insertion Techniques
Submitted to the Proceedings fro Medicine Meets Virtual Reality 15 February 2007, Long Beach, California To appear in Stud. Health Inform Validation System of MR Image Overlay and Other Needle Insertion
More informationEnabling Technologies for Robot Assisted Ultrasound Tomography
Enabling Technologies for Robot Assisted Ultrasound Tomography Seminar Presentation By: Fereshteh Aalamifar Team Members: Rishabh Khurana, Fereshteh Aalamifar Mentors: Emad Boctor, Iulian Iordachita, Russell
More informationAn Accuracy Approach to Robotic Microsurgery in the Ear
An Accuracy Approach to Robotic Microsurgery in the Ear B. Bell¹,J.Salzmann 1, E.Nielsen 3, N.Gerber 1, G.Zheng 1, L.Nolte 1, C.Stieger 4, M.Caversaccio², S. Weber 1 ¹ Institute for Surgical Technologies
More informationA Radiometry Tolerant Method for Direct 3D/2D Registration of Computed Tomography Data to X-ray Images
A Radiometry Tolerant Method for Direct 3D/2D Registration of Computed Tomography Data to X-ray Images Transfer Function Independent Registration Boris Peter Selby 1, Georgios Sakas 2, Stefan Walter 1,
More informationLecture 6: Medical imaging and image-guided interventions
ME 328: Medical Robotics Winter 2019 Lecture 6: Medical imaging and image-guided interventions Allison Okamura Stanford University Updates Assignment 3 Due this Thursday, Jan. 31 Note that this assignment
More informationState-of-the-Art IGRT
in partnership with State-of-the-Art IGRT Exploring the Potential of High-Precision Dose Delivery and Real-Time Knowledge of the Target Volume Location Antje-Christin Knopf IOP Medical Physics Group Scientific
More informationFOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION
Philips J. Res. 51 (1998) 197-201 FOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION This special issue of Philips Journalof Research includes a number of papers presented at a Philips
More informationIntroduction. Quality Assurance for Image- Guided Radiation Therapy. Justification for IGRT. Image-Guided Radiation Therapy
Introduction Quality Assurance for Image- Guided Radiation Therapy Jean-Pierre Bissonnette, Ph.D., MCCPM Princess Margaret Hospital, Toronto, Canada IGRT What is it? Rationale Equipment Quality Assurance
More informationAvailable online at ScienceDirect. Physics Procedia 63 (2015 )
Available online at www.sciencedirect.com ScienceDirect Physics Procedia 63 (2015 ) 189 194 43 rd Annual Symposium of the Ultrasonic Industry Association, UIA Symposium 2014 Real time target tracking in
More informationInvestigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers
Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Krista Chytyk MSc student Supervisor: Dr. Boyd McCurdy Introduction The objective of cancer radiotherapy
More informationDigital Tomosynthesis for Target Localization
Digital Tomosynthesis for Target Localization Fang-Fang Yin, Devon Godfrey, Lei Ren Jacqueline Maurer, Jackie Q-L Wu Duke University Medical Center Acknowledgements Duke Radiation Oncology faculty and
More informationBiomet PMI. Patient-Matched Implants. CT Protocols
CT Protocols One Surgeon. One Patient. Over 1 million times per year, Biomet helps one surgeon provide personalized care to one patient. The science and art of medical care is to provide the right solution
More informationEndoscopic Reconstruction with Robust Feature Matching
Endoscopic Reconstruction with Robust Feature Matching Students: Xiang Xiang Mentors: Dr. Daniel Mirota, Dr. Gregory Hager and Dr. Russell Taylor Abstract Feature matching based 3D reconstruction is a
More informationICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588
ICARO Vienna April 27-29 2009 Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588 M. Saiful Huq, Ph.D., Professor and Director, Dept. of Radiation
More informationClinical implementation of photon beam flatness measurements to verify beam quality
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Clinical implementation of photon beam flatness measurements to verify beam quality Simon Goodall, a Nicholas Harding, Jake Simpson,
More informationA Non-Linear Image Registration Scheme for Real-Time Liver Ultrasound Tracking using Normalized Gradient Fields
A Non-Linear Image Registration Scheme for Real-Time Liver Ultrasound Tracking using Normalized Gradient Fields Lars König, Till Kipshagen and Jan Rühaak Fraunhofer MEVIS Project Group Image Registration,
More informationProstate Detection Using Principal Component Analysis
Prostate Detection Using Principal Component Analysis Aamir Virani (avirani@stanford.edu) CS 229 Machine Learning Stanford University 16 December 2005 Introduction During the past two decades, computed
More informationTumor motion during liver SBRT
Tumor motion during liver SBRT - projects at Aarhus University Hospital - Per Poulsen, Esben Worm, Walther Fledelius, Morten Høyer Aarhus University Hospital, Denmark SBRT: Stereotactic Body Radiation
More informationIMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays
IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays Sotiri Stathakis Outline Why IMRT/VMAT QA AAPM TG218 UPDATE Tolerance Limits and Methodologies for IMRT Verification QA Common sources
More informationph fax Deming Way Middleton WI USA
www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 3120 Deming Way Middleton WI 53562-1461 USA 2007 Standard Imaging, Inc. 1239-21 (03.07) beam qa LINAC AND TOMOTHERAPY QA The NEW
More informationIntraoperative Prostate Tracking with Slice-to-Volume Registration in MR
Intraoperative Prostate Tracking with Slice-to-Volume Registration in MR Sean Gill a, Purang Abolmaesumi a,b, Siddharth Vikal a, Parvin Mousavi a and Gabor Fichtinger a,b,* (a) School of Computing, Queen
More informationLOGIQ. V2 Ultrasound. Part of LOGIQ Vision Series. Imagination at work LOGIQ is a trademark of General Electric Company.
TM LOGIQ V2 Ultrasound Part of LOGIQ Vision Series Imagination at work The brilliance of color. The simplicity of GE. Now you can add the advanced capabilities of color Doppler to patient care with the
More informationMonte Carlo methods in proton beam radiation therapy. Harald Paganetti
Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More informationValidation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition
Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.
More informationReal-time self-calibration of a tracked augmented reality display
Real-time self-calibration of a tracked augmented reality display Zachary Baum, Andras Lasso, Tamas Ungi, Gabor Fichtinger Laboratory for Percutaneous Surgery, Queen s University, Kingston, Canada ABSTRACT
More informationRobot Control for Medical Applications and Hair Transplantation
Dr. John Tenney Director of Research Restoration Robotics, Inc. Robot Control for Medical Applications and Hair Transplantation Presented to the IEEE Control Systems Society, Santa Clara Valley 18 November
More informationFreehand 3-D Sonographic Measurement of the Superficial Femoral Artery
Freehand 3-D Sonographic Measurement of the Superficial Femoral Artery Dennis Sandkühler, Christian Sobotta, Matthias Samsel, Heinrich Martin Overhoff Medical Engineering Laboratory, University of Applied
More informationAutomated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator
Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4
More informationFinancial disclosure. Onboard imaging modality for IGRT
Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty
More informationINTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory
INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1 Modifications for P551 Fall 2013 Medical Physics Laboratory Introduction Following the introductory lab 0, this lab exercise the student through
More informationMotion Tracking in 2D Ultrasound Using Vessel Models and Robust Optic-Flow
Motion Tracking in 2D Ultrasound Using Vessel Models and Robust Optic-Flow Maxim Makhinya and Orcun Goksel ETH Zürich, Computer Vision Laboratory, Switzerland, {makhinya,ogoksel}@vision.ee.ethz.ch Abstract.
More information7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT
CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance
More informationA Novel Phantom-Less Spatial and Temporal Ultrasound Calibration Method
A Novel Phantom-Less Spatial and Temporal Ultrasound Calibration Method Ali Khamene and Frank Sauer Imaging and Visualization Dept., Siemens Corporate Research, 755 College Road East, Princeton NJ 08540,
More informationVision-based endoscope tracking for 3D ultrasound image-guided surgical navigation [Yang et al. 2014, Comp Med Imaging and Graphics]
Vision-based endoscope tracking for 3D ultrasound image-guided surgical navigation [Yang et al. 2014, Comp Med Imaging and Graphics] Gustavo Sato dos Santos IGI Journal Club 23.10.2014 Motivation Goal:
More informationImprovement and Evaluation of a Time-of-Flight-based Patient Positioning System
Improvement and Evaluation of a Time-of-Flight-based Patient Positioning System Simon Placht, Christian Schaller, Michael Balda, André Adelt, Christian Ulrich, Joachim Hornegger Pattern Recognition Lab,
More informationReal-time Tissue Tracking with B-Mode Ultrasound Using Speckle and Visual Servoing
Real-time Tissue Tracking with B-Mode Ultrasound Using Speckle and Visual Servoing Alexandre Krupa, Gabor Fichtinger, and Gregory D. Hager IRISA - INRIA Rennes, France, alexandre.krupa@irisa.fr Engineering
More informationThe MSKCC Approach to IMRT. Outline
The MSKCC Approach to IMRT Spiridon V. Spirou, PhD Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York, NY Outline Optimization Field splitting Delivery Independent verification
More informationIntroduction to Digitization Techniques for Surgical Guidance
Introduction to Digitization Techniques for Surgical Guidance Rebekah H. Conley, MS and Logan W. Clements, PhD Department of Biomedical Engineering Biomedical Modeling Laboratory Outline Overview of Tracking
More informationMR-guided radiotherapy: Vision, status and research at the UMC Utrecht. Dipl. Ing. Dr. Markus Glitzner
MR-guided radiotherapy: Vision, status and research at the UMC Utrecht Dipl. Ing. Dr. Markus Glitzner About myself Training Medizintechnik TU Graz PhD UMC Utrecht Clinical work Software implementation
More informationA SYSTEM OF DOSIMETRIC CALCULATIONS
A SYSTEM OF DOSIMETRIC CALCULATIONS INTRODUCTION Dose calculation based on PDD and TAR have Limitations The dependence of PDD on SSD Not suitable for isocentric techniques TAR and SAR does not depend on
More informationImage Acquisition Systems
Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS
More informationC-mode Real Time Tomographic Reflection for a Matrix Array Ultrasound Sonic Flashlight
C-mode Real Time Tomographic Reflection for a Matrix Array Ultrasound Sonic Flashlight George Stetten 1,2,3, Aaron Cois 1,2,, Wilson Chang 1,2,3, Damion Shelton 2, Robert Tamburo 1,2, John Castellucci
More informationUNCOMPROMISING QUALITY
ION CHAMBERS UNCOMPROMISING QUALITY Designed with over 30 years of scientific integrity for a broad range of dosimetry measurements in diverse radiation beams Farmer-type Chambers For absolute dosimetry
More informationMotion estimation of unmanned marine vehicles Massimo Caccia
Motion estimation of unmanned marine vehicles Massimo Caccia Consiglio Nazionale delle Ricerche Istituto di Studi sui Sistemi Intelligenti per l Automazione Via Amendola 122 D/O, 70126, Bari, Italy massimo.caccia@ge.issia.cnr.it
More informationTakayama, Kenji; Mizowaki, Takashi; Nagano, Kazuo; Kamino, Yuichiro; Hi. Rightc 2009 Elsevier Ireland Ltd.
Title Initial validations for pursuing ir tracking system. Takayama, Kenji; Mizowaki, Takashi; Author(s) Kawada, Noriyuki; Nakayama, Hiroshi Nagano, Kazuo; Kamino, Yuichiro; Hi Citation Radiotherapy and
More informationDirect 3D Ultrasound to Video Registration Using Photoacoustic Effect
Direct 3D Ultrasound to Video Registration Using Photoacoustic Effect Alexis Cheng, Jin U. Kang, Russell H. Taylor, and Emad M. Boctor The Johns Hopkins University, Baltimore, Maryland 21211, United States
More informationradiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li NCAAPM Spring Meeting 2010 Madison, WI
GPU-Accelerated autosegmentation for adaptive radiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li agodley@mcw.edu NCAAPM Spring Meeting 2010 Madison, WI Overview Motivation Adaptive
More informationSpatial-temporal Total Variation Regularization (STTVR) for 4D-CT Reconstruction
Spatial-temporal Total Variation Regularization (STTVR) for 4D-CT Reconstruction Haibo Wu a, b, Andreas Maier a, Rebecca Fahrig c, and Joachim Hornegger a, b a Pattern Recognition Lab (LME), Department
More informationDisclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions
LINAC and MLC QA for IMRT Ping Xia, Ph.D., Department of Radiation Oncology Disclosure Received research support from Siemens Medical Solutions University of California San Francisco Therapy Series (SAM)
More informationOptimization of Image Guided Radiation Therapy Using. Limited Angle Projections
Optimization of Image Guided Radiation Therapy Using Limited Angle Projections by Lei Ren Medical Physics Graduate Program Duke University Date: Approved: Fang-Fang Yin, Supervisor Shiva K. Das Q. Jackie
More information724. Simulation of the radiation therapy system for respiratory movement compensation
724. Simulation of the radiation therapy system for respiratory movement compensation Tomas Krilavičius 1, Daiva Vitkutė-Adžgauskienė 2, Kęstutis Šidlauskas 3 Vytautas Magnus University, Kaunas, Lithuania
More informationAn Improved Tracking Technique for Assessment of High Resolution Dynamic Radiography Kinematics
Copyright c 2008 ICCES ICCES, vol.8, no.2, pp.41-46 An Improved Tracking Technique for Assessment of High Resolution Dynamic Radiography Kinematics G. Papaioannou 1, C. Mitrogiannis 1 and G. Nianios 1
More informationA Simulation Study and Experimental Verification of Hand-Eye-Calibration using Monocular X-Ray
A Simulation Study and Experimental Verification of Hand-Eye-Calibration using Monocular X-Ray Petra Dorn, Peter Fischer,, Holger Mönnich, Philip Mewes, Muhammad Asim Khalil, Abhinav Gulhar, Andreas Maier
More informationRecent Advances in Non-Invasive Thermometry Using Changes in Backscattered Ultrasound
Recent Advances in Non-Invasive Thermometry Using Changes in Backscattered Ultrasound R. Martin Arthur 1, Jason W. Trobaugh 1, William L. Straube 2 and Eduardo G. Moros 2 1 Electrical & Systems Engineering
More informationDevelopment and validation of a short-lag spatial coherence theory for photoacoustic imaging
Development and validation of a short-lag spatial coherence theory for photoacoustic imaging Michelle T. Graham 1 and Muyinatu A. Lediju Bell 1,2 1 Department of Electrical and Computer Engineering, Johns
More informationA finite state model for respiratory motion analysis in image guided radiation therapy
INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 49 (2004) 5357 5372 PHYSICS IN MEDICINE AND BIOLOGY PII: S0031-9155(04)85065-6 A finite state model for respiratory motion analysis in image guided radiation
More informationA novel-integrated quality assurance phantom for radiographic and nonradiographic radiotherapy localization and positioning systems
A novel-integrated quality assurance phantom for radiographic and nonradiographic radiotherapy localization and positioning systems Amy S. Yu a) Tyler L. Fowler, and Piotr Dubrowski Department of Radiation
More informationIntroducing Computer-Assisted Surgery into combined PET/CT image based Biopsy
Introducing Computer-Assisted Surgery into combined PET/CT image based Biopsy Santos TO(1), Weitzel T(2), Klaeser B(2), Reyes M(1), Weber S(1) 1 - Artorg Center, University of Bern, Bern, Switzerland 2
More informationPhysiological Motion Compensation in Minimally Invasive Robotic Surgery Part I
Physiological Motion Compensation in Minimally Invasive Robotic Surgery Part I Tobias Ortmaier Laboratoire de Robotique de Paris 18, route du Panorama - BP 61 92265 Fontenay-aux-Roses Cedex France Tobias.Ortmaier@alumni.tum.de
More informationReal-Time Data Acquisition for Cardiovascular Research
Real-Time Data Acquisition for Cardiovascular Research Andras Lasso, PhD Laboratory for Percutaneous Surgery School of Computing, Queen s University, Kingston Questions / challenges What hardware/software
More informationTomographic Reconstruction
Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)
More informationThe theory and practical aspects of proton imaging proton radiography (prad) and proton tomography (pct)
The theory and practical aspects of proton imaging proton radiography (prad) and proton tomography (pct) Fritz DeJongh, ProtonVDA Inc August 30 2018 Stay away from negative people. They have a problem
More informationAdvanced Image Reconstruction Methods for Photoacoustic Tomography
Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic
More informationULTRASOUND (US) has become one of the standard
538 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 36, NO. 2, FEBRUARY 2017 Towards MRI-Based Autonomous Robotic US Acquisitions: A First Feasibility Study Christoph Hennersperger, Bernhard Fuerst, Salvatore
More informationClosed-Loop Control in Fused MR-TRUS Image-Guided Prostate Biopsy
Closed-Loop Control in Fused MR-TRUS Image-Guided Prostate Biopsy Sheng Xu 1, Jochen Kruecker 1, Peter Guion 2, Neil Glossop 3, Ziv Neeman 2, Peter Choyke 2, Anurag K. Singh 2, and Bradford J. Wood 2 1
More informationClinical Prospects and Technological Challenges for Multimodality Imaging Applications in Radiotherapy Treatment Planning
Clinical Prospects and Technological Challenges for Multimodality Imaging Applications in Radiotherapy Treatment Planning Issam El Naqa, PhD Assistant Professor Department of Radiation Oncology Washington
More informationEndoscopic Navigation for Minimally Invasive Suturing
Endoscopic Navigation for Minimally Invasive Suturing Christian Wengert 1, Lukas Bossard 1, Armin Häberling 1, Charles Baur 2, Gábor Székely 1, and Philippe C. Cattin 1 1 Computer Vision Laboratory, ETH
More informationTowards deformable registration for augmented reality in robotic assisted partial nephrectomy
Towards deformable registration for augmented reality in robotic assisted partial nephrectomy Supervisor: Elena De Momi, PhD Co-supervisor: Dott. Ing. Sara Moccia Author: Anna Morelli, 853814 Academic
More informationIntroduction to Biomedical Imaging
Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray
More informationPosition accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon
Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon WHITE PAPER Introduction An image guidance system based on Cone Beam CT (CBCT) is included in Leksell
More informationNew 4-D Imaging for Real-Time Intraoperative MRI: Adaptive 4-D Scan
New 4-D Imaging for Real-Time Intraoperative MRI: Adaptive 4-D Scan Junichi Tokuda 1, Shigehiro Morikawa 2, Hasnine A. Haque 3, Tetsuji Tsukamoto 3, Kiyoshi Matsumiya 1, Hongen Liao 4, Ken Masamune 1,
More informationProtocol. Technical evaluation of X-ray tomographic image-guided radiotherapy devices CEP10070
Protocol Technical evaluation of X-ray tomographic image-guided radiotherapy devices CEP10070 March 2010 Contents 2 Introduction... 3 Protocol design and validation... 4 General information... 7 Technical
More informationCT vs. VolumeScope: image quality and dose comparison
CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,
More informationBasic principles of MR image analysis. Basic principles of MR image analysis. Basic principles of MR image analysis
Basic principles of MR image analysis Basic principles of MR image analysis Julien Milles Leiden University Medical Center Terminology of fmri Brain extraction Registration Linear registration Non-linear
More informationDAILY LINAC QA BEAM QA
BEAM QA DAILY LINAC QA The QA BeamChecker Plus allows for fast, reliable, and uncomplicated daily QA of Varian, Elekta, Siemens, and Accuray Treatment Machines. The QA BeamChecker Plus is specifically
More informationChoosing and Commissioning a Video Based Motion Management System
Choosing and Commissioning a Video Based Motion Management System David Shepard Swedish Cancer Institute 2 1 Acknowledgments Daliang Cao, PhD, SCI Mohammed Jermoumi, PhD, SCI Roger Xie, PhD, SCI Malin
More information