Effective detective quantum efficiency (edqe) and effective noise equivalent quanta (eneq) for system optimization purposes in digital mammography

Size: px
Start display at page:

Download "Effective detective quantum efficiency (edqe) and effective noise equivalent quanta (eneq) for system optimization purposes in digital mammography"

Transcription

1 Effective detective quantum efficiency (edqe) and effective noise equivalent quanta (eneq) for system optimization purposes in digital mammography Elena Salvagnini a,b, Hilde Bosmans a, Lara Struelens b, Nicholas W. Marshall a a UZ Gasthuisberg, Department of Radiology, Herestraat 9, B-3 Leuven, Belgium; b SCK CEN, Boeretang, Mol, Belgium ABSTRACT Effective detective quantum efficiency (edqe) and effective noise equivalent quanta (eneq) were recently introduced to broaden the notion of DQE and NEQ by including system parameters such as focus blurring and system scatter rejection methods. This work investigates edqe and eneq normalized for mean glandular dose (eneq MGD ) as a means to characterize and select optimal exposure parameters for a digital mammographic system. The edqe was measured for three anode/filter combinations, with and without anti-scatter grid and for four thicknesses of poly(methylmethacrylate) (PMMA). The modulation transfer function used to calculate edqe and eneq was measured from an edge positioned at,,6,7 mm above the table top without scattering material in the beam. The grid-in edqe results for all A/F settings were generally larger than those for grid-out. Contrarily, the eneq MGD results were higher for grid-out than gridin, with a maximum difference of 61% among all A/F combinations and PMMA thicknesses. The W/Rh combination gave the highest eneq MGD for all PMMA thicknesses compared to the other A/F combinations (for grid-in and grid-out), supporting the results of alternative methods (e.g. the signal difference to noise ratio method). The eneq MGD was then multiplied with the contrast obtained from a.mm Al square, resulting in a normalized quantity that was higher for the W/Rh combination than for the other A/F combinations. In particular, the results for the W/Rh combination were greater for the grid-in case. Furthermore, these results showed close agreement with a non-prewhitened match filter with eye response model observer (d ) normalized for MGD. Keywords: FFDM, effective Detective Quantum Efficiency (edqe), effective Noise Equivalent Quanta (eneq), detectability index (d ), Modulation Transfer Function (MTF), SdNR, image quality. 1. DESCRIPTION OF PURPOSE The quantities effective detective quantum efficiency (edqe) and effective noise equivalent quanta (eneq) can be seen as a development of the work of Wagner et al [1], by Samei et al [] and Ertan et al [3] with application to digital radiographic systems. The edqe is an attempt to broaden the notion of the detector centric metric of detective quantum efficiency (DQE) to include system parameters such as focus blurring and system scatter rejection methods. This study applies the edqe and eneq methodology to a digital mammographic system. The formulation of edqe attempts to compare a system against an ideal imaging system with perfect x-ray detection and scatter rejection, however a patient dose marker is not included. For this reason, eneq normalized for the mean glandular dose (MGD) was included in this study (eneq MGD ), as system configurations that produce high edqe may not be the best clinical choice. It is possible that the patient dose associated with parameters that give a high value edqe could be higher than the patient dose of configuration that is suboptimal in terms of edqe. The eneq MGD describes the impact of the patient dose on the chosen configuration but this term does not give any indication on object detectability. As an example, lower patient dose configurations (eg grid-in), can lead to higher eneq MGD without leading to higher object detectability in images. For this reason, we decided to extend eneq MGD with an object contrast (C%) that is characteristic for the studied configurations, and named this quantity eneq DC.. Medical Imaging 1: Physics of Medical Imaging, edited by Norbert J. Pelc, Robert M. Nishikawa, Bruce R. Whiting, Proc. of SPIE Vol. 8313, 8313H 1 SPIE CCC code: 165-7/1/$18 doi: / Proc. of SPIE Vol H-1

2 The purpose of this work was (1) to perform a full characterization of the digital mammography system efficiency using edqe as opposed to an DQE approach that focuses on the detector, () to investigate the influence of the mean glandular dose using eneq MGD, and (3) to relate eneq DC with a well known and accepted detectability index, the nonprewhitened match filter [] with eye response [5] (NPWE) model observer (d ). The combined use of these indexes can indicate the optimal choice of beam quality and system configuration for use in clinical examinations.. METHODS The study was performed using a Siemens MAMMOMAT Inspiration (Siemens AG Healthcare, Erlangen, Germany) equipped with a-se detector with 85 µm pitch and a selenium layer thickness of µm. This system utilizes three different A/F combinations (Mo/Mo, W/Rh and Mo/Rh), has a nominal focal spot dimension of.3 mm and an SID of 65 mm. The system can be operated with and without the anti-scatter grid. The grid is a linear type, with a ratio of 5:1, lead septa and strip density of 31 lines/cm. Detector response was measured for the four PMMA thicknesses using typical tube potential values and all available A/F combinations. Air kerma at the detector was measured with a calibrated Barracuda MPD dose detector and the pixel values (PV) taken from For Processing ( Flatfield ) images. The anti-scatter grid was removed for the detector response measurements. PV was plotted against air kerma at the detector (DAK) to give four detector response curves; these were used to linearize the PV data before any calculation was performed. The edqe [] was calculated for all A/F combinations for four PMMA thicknesses:,, 6 and 7mm. The equation used was: edqe( u' ) = MTF ( u' ) (1 SF ) NNPS( u' ) TF E q (1) Where u = spatial frequency corrected to the object plane, MTF= modulation transfer function, SF=scatter fraction, NNPS = normalized noise power spectrum, TF=narrow beam transmission factor of PMMA, E= pre-phantom exposure corrected at the detector plane, q= number of photons per µgy per unit area. The scatter fraction SF was estimated using the beam stop method [6] using six Pb disks of diameter ranging from to 3 mm. The primary radiation component was estimated by placing the PMMA at the output of the tube using a narrow beam geometry to produce a low scatter image. Using the same configuration, the Pb disks were positioned in the beam at the detector and the pixel values behind the disk are used to estimate the detector glare component. Finally, acquiring full field images with the PMMA placed at the detector gave the total signal (primary, glare and scatter components). The total signal was calculated as the mean PV in the same region of interest where primary radiation and glare were estimated. Consequently subtracting the pixel values of the scatter free configuration images and the pixel values representing detector glare from the total signal gave scattered radiation component. The narrow beam transmission factor (TF) was measured by placing the Barracuda MPD on the detector cover and collimating the x-ray beam to be slightly larger than the MPD. The PMMA was placed at the output at the X-ray tube. The air kerma was than acquired for all the A/F combinations at typical tube potentials with and without the PMMA in the beam. The ratio of the measured air kerma with and without PMMA gave the narrow beam transmission factor. The NNPS was calculated from homogeneous PMMA images acquired using automatic exposure control. These measurements were made with the grid in and out of the x-ray beam. The algorithm usedto estimate the noise power spectrum is described elsewhere; [7] the final NNPS estimate was taken from a radial average of the ensemble (including the and 9 spatial frequency axes), as the NNPS was found to be isotropic. Two NNPS results were estimated for every image and averaged, each NNPS was estimated using a region size of 51 by 51 divided in record size of 56 by 56 pixels with 5% overlap in both x- and y- directions. In the first part of the study, MTF was measured using a version of the edge method [7-8], under both clinical scatter conditions (edge placed on top of PMMA blocks) and for two different low scatter geometries (edge placed on top of the detector and edge placed at different heights above the detector, without PMMA), to investigate the influence Proc. of SPIE Vol H-

3 of different MTF curves on edqe. Following initial analysis, the MTF required for edqe was measured using a low scatter geometry, with the edge placed at different heights above the detector. The latter MTF was measured at 8 kv and W/Rh A/F setting with mm Al at the x-ray tube exit port. A steel plate of dimension 1 x 1 mm and thickness 1 mm with machined straight edges was positioned at,, 6 and 7 mm from the detector top without scatter material in the beam (referred to as Free-Air MTF). The edge was angled slightly (1 to 5 ) against the pixel matrix and images were acquired at a DAK of about 5µGy. The eneq was calculated for the same conditions as the edqe and normalized to the mean glandular dose (MGD) using the equation: MTF ( u') (1 SF ) 1 eneq MGD ( u') = () NNPS( u') MGD where MGD = mean glandular dose and u and SF, MTF and NNPS are the same terms used for the calculation of the edqe. The MGD was calculated using the method of Dance et al [9] : MGD = Kgsc (3) where K is the incident air kerma at the upper surface of the breast, measured without backscatter, the factor g is the incident air kerma to mean glandular dose conversion factor for a glandularity of 5%, the factor c corrects for the difference in breast composition from 5% glandularity and the factor s corrects for the x-ray spectrum used. The eneq MGD was than corrected for the object contrast factor (C%). The contrast was estimated using. mm Al square of 1x1 mm, applying equation () PVBkg PVAl C% = () PVBkg Where PV Bkg is the mean pixel value of a region of interest (ROI) of 5x5mm in the PMMA and PV Al is the mean pixel value of a region of interest (ROI) of 5x5mm in the Al. Through the text we will refer to the eneq MGD corrected for object contrast as eneq DC., with DC we want to indicate that the term is weighed for patient Dose and object Contrast. Equation 5 was used to calculate eneq DC. eneqdc ( u' ) = eneqmgd ( u') C% (5) The detectability in this study was calculated using the NPWE observer in equation (6): d' = π C S ( f ) MTF S( f ) MTF ( f ) VTF( f ) ( f ) VTF fdf ( f ) NNPS( f ) fdf (6) where C is the contrast measured using the. mm Al square, f is the spatial frequency, MTF(f) is the detector modulation transfer function, NNPS(f) is the normalized noise power spectrum and VTF(f) is the Visual Transfer Function. The VTF given by Kelly [1] was used for a nominal image magnification of 1.5 and viewing distance of mm 7. The signal was defined for a disc-like object of.1 mm diameter using equation (7): d J ( df ) S( f ) 1 π = (7) f Proc. of SPIE Vol H-3

4 where J 1 is a Bessel function of the first kind and d is the object diameter. Calculation of d requires measurement of the MTF and NNPS. These parameters were calculated as described above. In particular for the MTF calculation a low scatter geometry (detector MTF) was used at 8 kv and W/Rh A/F setting. Finally d was squared and normalized to MGD for comparison with the eneq DC results. 3. RESULTS AND DISCUSSION Before performing a full characterization of the system, an investigation of the influence of the MTF used in the edqe equation was performed. In the works of Samei et al [] and Ertan et al [3] the MTF was calculated from an edge placed on the top of the phantom. This configuration introduces a low frequency drop (LFD) in the MTF curve, even if a small region of interest (ROI) is used to calculate the MTF. Taken together with the SF correction, this can lead to an over-estimation of the scatter effect on the efficiency of the system thus a reduction in the edqe curves. For this reason it was decided to investigate the influence of three different MTF methods on the edqe equation: i) MTF measured with the edge placed at the phantom entrance (with scatter) (Samei et al) for two ROI sizes (7x7mm and xmm) ii) the detector MTF (edge on the detector) iii) the Free-Air MTF. The results are presented in figure 1 for and 7 mm of PMMA ( and 6 mm presented the same trends). The values of SF, TF, q and E used to obtain these results are presented in table 1. Figure 1 shows that the detector MTF gives the highest edqe, as might be expected, and the MTF estimated with the edge on top of the PMMA with a x mm ROI gives the lowest edqe. This is due to the absence of the focus blurring effects in the detector MTF and to the strong influence of scatter (large LFD) of the MTF calculated from the x mm ROI. The results obtained for Free-Air MTF and 7x7 mm MTF for mm of PMMA are quite similar (maximum difference 1%). Both methods include the focal blurring effects while scattered radiation at mm PMMA does not influence substantially the LFD of the MTF calculated with a 7x7 mm ROI. The influence of the LFD, even using the 7x7 mm ROI, became progressively larger (MTF became lower) as PMMA increased, compared to the Free-Air MTF - again due to the influence of scattered radiation (maximum difference 1%)...3 detector MTF PMMA mm Free-Air MTF PMMA mm 7x7 mm MTF PMMA mm x mm MTF PMMA mm..3 detector MTF PMMA 7mm Free-Air MTF PMMA 7mm 7x7 mm MTF PMMA 7mm x mm MTF PMMA 7mm edqe. edqe Fig. 1 shows edqe calculated using four different MTF methods: detector MTF, Free-Air MTF estimated without scattering material in the beam, and the MTF estimated with the edge placed on top of the PMMA with two different ROIs of 7x7 mm and x mm. The results are presented for mm of PMMA on the left and for 7 mm PMMA on the right. These results highlight the influence of scattered radiation on the MTF; when MTF estimated under clinical condition (edge on top of PMMA) is implemented in the equation 1, which also include a correction for SF, this leads to an overestimation of the scatter effects and consequentially a drop in edqe. The Free-Air MTF was considered a better choice for the edqe calculation as focal spot blurring effects are included without an additional contribution due to Proc. of SPIE Vol H-

5 scattered radiation. A full evaluation of edqe and eneq MGD was then performed using the Free-Air MTF. The values used for the analysis are presented in table 1. Table 1. PMMA thickness, tube potential, scatter fraction (SF) estimated with and without the grid are indicated, these values are common for all the A/F combinations. Number of photons per µgy unit area (q), transmission factor (TF) and prephantom exposure corrected at the detector plane (E) for the AEC mode with and without the grid for each A/F combination are also indicated. PMMA (mm) Tube kv A/F Mo/Mo Mo/Rh W/Rh SF grid IN SF grid OUT q TF grid IN grid OUT q TF grid IN grid OUT q TF grid IN grid OUT Figure shows the results of edqe for all the A/F combinations, with and without antiscatter grid in place for and 7 mm of PMMA. The curves for and 6 mm of PMMA showed trends similar to that of 7mm PMMA. Figure shows that the values of edqe for the digital mammographic system (between 1 and 3%) are higher than those for digital radiographic systems (typically in the range of to 1%) (Samei et al, Ertan et al). This is probably due to the higher system resolution (MTF) and the lower SF. Ertan et al reports a SF of about. and.6 with and without grid respectively, while this work shows a maximum SF of.157 and.59 with and without grid respectively. For mm of PMMA, the edqe results are slightly higher in absence of grid (9% maximum difference), however the grid in and grid out results are quite similar, reflecting the lower values of scattered radiation at mm. The use of W/Rh gives slightly higher edqe values than the other A/F settings. From to 7 mm of PMMA, the edqe values with grid in are higher than grid out (maximum difference of 35%) for all A/F combinations due to increased quantities of scattered radiation., Mo/Mo_IN mm, Mo/Mo_IN 7mm Mo/Rh_IN mm Mo/Rh_IN 7mm W/Rh_IN mm W/Rh_IN 7mm,3 Mo/Mo_OUT mm Mo/Rh_OUT mm,3 Mo/Mo_OUT 7mm Mo/Rh_OUT 7mm W/Rh_OUT mm W/Rh_OUT 7mm edqe, edqe,,1,1,, Fig. shows the edqe results for all the A/F combination of the digital mammographic system, with and without the presence of the antiscatter grid for mm of PMMA (left) and 7 mm of PMMA (right). Proc. of SPIE Vol H-5

6 eneq MGD (x1 5 ) Mo/Mo_IN mm Mo/Rh_IN mm W/Rh_IN mm Mo/Mo_OUT mm Mo/Rh_OUT mm W/Rh_OUT mm eneq MGD (x1 5 ) Mo/Mo_IN 7mm Mo/Rh_IN 7mm W/Rh_IN 7mm Mo/Mo_OUT 7mm Mo/Rh_OUT 7mm W/Rh_OUT 7mm 1 1 Fig. 3 shows the eneq results normalized for MGD for all A/F combinations of the digital mammographic system, with and without the presence antiscatter grid, for mm of PMMA (left) and 7 mm of PMMA (right). Figure 3 shows the results of eneq MGD for all A/F combinations, with and without the anti-scatter grid in place, for and 7 mm of PMMA (the results for and 6 mm of PMMA showed the same trends). Somewhat surprisingly, eneq MGD is higher for grid out than grid in, with a maximum difference of 61% for all the A/F combinations and PMMA thicknesses. This behaviour is due to the lower mas used for the grid out case leading to lower MGD. The results for MGD are shown in figure where the increasing difference in eneq MGD between grid in and grid out geometries can be seen as a function of PMMA. MGD has, in fact, very similar values for all A/F combinations for mm of PMMA, resulting in similar eneq MGD for grid in and grid out for the same A/F. Conversely, MGD is more than doubled for grid in compared to configurations without grid for 7 mm of PMMA. This explains the higher eneq MGD results in absence of anti-scatter grid. The eneq MGD values include patient dose and hence give an indication of system efficiency weighted or normalized for a patient risk related quantity. The eneq MGD curves show that the W/Rh combination gives the highest result for a given PMMA thickness and grid setting, compared to the other A/F settings. The fact that the W/Rh setting gives the highest edqe and eneq values is consistent with the optimal A/F setting found with standard signal difference to noise methods (SDNR) [11]. The eneq MGD does not include aspects related to object detectability and in order to relate eneq MGD to object detectability, the term was corrected according to equation 5. The values of contrast as a function of PMMA thickness for different A/F combinations with and without the anti-scatter grid in place are shown in figure 5. The results of eneq DC are shown in figure 6, at the left, for mm at the left of PMMA and figure 6, at the right, for 7mm of PMMA. Figure 6 shows that the W/Rh A/F combination gives the highest results but in this case the grid in configuration gives higher results (peak difference of 1%) than grid out. This is due to the higher contrast (figure 5) for the grid in geometry because of reduced scattered radiation; the maximum difference is 1% for mm of PMMA and 3% for 7 mm of PMMA for W/Rh A/F combination. For and 6 mm of PMMA the same trend was seen. Proc. of SPIE Vol H-6

7 8 Mo/Mo IN 5% 7 Mo/Rh IN 6 W/Rh IN Mo/Mo OUT % MGD [mgy] 5 3 Mo/Rh OUT W/Rh OUT Acceptable Reachable contrast [%] 15% 1% Mo/Mo IN Mo/Rh IN W/Rh IN 1 5% Mo/Mo OUT Mo/Rh OUT thickness [mm] Fig. shows the MGD results A/F combinations of the digital mammographic system, with and without the presence antiscatter grid. Acceptable and reachable threshold of the European Guideline for mammographic doses are reported for completeness. % W/Rh OUT 6 8 thickness [mm] Fig. 5 shows the contrast as function of thickness for different A/F combination with and without the anti-scatter grid in place.9 Mo/Mo_IN mm.9 Mo/Mo_IN 7mm eneq DC (x1 5 ).6.3 Mo/Rh_IN mm W/Rh_IN mm Mo/Mo_OUT mm Mo/Rh_OUT mm W/Rh_OUT mm eneq DC (x1 5 ).6.3 Mo/Rh_IN 7mm W/Rh_IN 7mm Mo/Mo_OUT 7mm Mo/Rh_OUT 7mm W/Rh_OUT 7mm.. Fig. 6 shows the eneq results normalized for MGD and corrected for contrast for all A/F combinations of the digital mammographic system, with and without the presence antiscatter grid, for mm of PMMA (left) and 7 mm of PMMA (right). Finally as a means of validation of the new figure of merit proposed we compared the results with the well known detectability index d [1]. In order to perform an appropriate (normalized) comparison, the d index was squared and divided by the MGD. The results of this operation are shown in figure 7(a), where the trend follows that obtained for eneq DC, as illustrated in figure 7(b). In fact d /MGD shows larger differences in the results for mm of PMMA for difference A/F combinations than in the results for 7 mm PMMA thickness. Moreover also this index confirms the superiority of the W/Rh A/F combination for,6,7 mm. For mm PMMA the highest results were for the Mo/Mo A/F combination and grid out, but the difference between Mo/Mo grid out and W/Rh grid in was just 6%. Proc. of SPIE Vol H-7

8 d' /MGD Mo/Mo IN Mo/Rh IN W/Rh IN Mo/Mo OUT 1 Mo/Rh OUT W/Rh OUT 6 8 (a) thickness [mm] (b) Fig. 7 shows (a) d / MGD for all A/F combinations of the digital mammographic system, with and without the presence antiscatter grid as a function of PMMA thickness (b) d /MGD plotted against eneq DC with PMMA thickness as a parameter. d' /MGD Mo/Mo grid IN Mo/Rh grid IN W/Rh grid IN Mo/Mo grid OUT Mo/Rh grid OUT W/Rh grid OUT eneq DC (x1 5 ). CONCLUSIONS A complete characterization of a mammographic system in term of edqe and eneq was performed. The established MTF method for edqe, measured in the presence of scattered radiation, was substituted for a Free-Air MTF measured without scattered radiation but including the focal spot blurring. The edqe method produced results that are largely consistent with standard methods such as signal difference to noise ratio (SdNR) in terms of system characterization [11], but does not include patient dose. For this reason eneq normalized for the MGD was calculated to account for patient risk. After eneq was normalized for MGD, we extended the concept to include object contrast. This quantity, which includes system sharpness and noise characteristics, patient dose and contrast, closely followed results calculated using the established non-prewhitened match filter with eye response (NPWE) model observer (d ). REFERENCES [1] Wagner, R.F., Barnes, G.T., and Askins, B.S., Effect of reduced scatter on radiographic information content and patient exposure: a quantitative demonstration, Med. Phys. 7(1), (198). [] Samei, E., Ranger, N.T., MacKenzie, A., Honey, I.D., Dobbins, J.T., and Ravin, C.E., Effective DQE (edqe) and speed of digital radiographic systems: an experimental methodology., Med. Phys. 36(8), (9). [3] Ertan, F., Mackenzie, A., Urbanczyk, H.J., Ranger, N.T., and Samei, E., Use of effective detective quantum efficiency to optimise radiographic exposures for chest imaging with computed radiography, Proceedings of SPIE 758(1), 7585O-7585O-1 (9). [] Wagner, R.F., and Brown, D.G., Unified SNR analysis of medical imaging systems, Phys. Med. Biol. 3(6), (1985). [5] Monnin, P., Marshall, N.W., Bosmans, H., Bochud, F.O., and Verdun, F.R., Image quality assessment in digital mammography: part II. NPWE as a validated alternative for contrast detail analysis., Phys. Med. Biol. 56(1), 1-38 (11). [6] Carton, A.K., Acciavatti, R., Kuo, J., and Maidment, A.D.A., The effect of scatter and glare on image quality in contrast-enhanced breast imaging using an a-si/csi(tl) full-field flat panel detector, Med. Phys. 36(3), 9 (9). [7] Marshall, N.W., A comparison between objective and subjective image quality measurements for a full field digital mammography system., Phys. Med. Biol. 51(1), 1-63 (6). Proc. of SPIE Vol H-8

9 [8] Samei, E., Flynn, M.J., and Reimann, D.A., A method for measuring the presampled MTF of digital radiographic systems using an edge test device., Med. Phys. 5(1), 1-13 (1998). [9] Dance, D.R., Skinner, C.L., Young, K.C., Beckett, J.R., and Kotre, C.J., Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol., Phys. Med. Biol. 5(11), 35-3 (). [1] Kelly, D.H., Motion and vision. II. Stabilized spatio-temporal threshold surface., J. Opt. Soc. Am. 69(1), (1979). [11] Bernhardt, P., Mertelmeier, T., and Hoheisel, M., X-ray spectrum optimization of full-field digital mammography: simulation and phantom study., Med. Phys. 33(11), (6). [1] Eckstein, M.P., Abbey, C.K., and Bochud, F.O., [A practical guide to model observers for visual detection in synthetic and natural noisy images], in Handbook of Medical Imaging 1, SPIE Press, Bellingham, WA, (). Proc. of SPIE Vol H-9

Nicholas Marshall Department of Radiology, University Hospitals Leuven, Herestraat 49, 3000 Leuven, Belgium

Nicholas Marshall Department of Radiology, University Hospitals Leuven, Herestraat 49, 3000 Leuven, Belgium Practical Methods for Assessing Image Quality in Diagnostic Radiology Nicholas Marshall Department of Radiology, University Hospitals Leuven, Herestraat 49, 3000 Leuven, Belgium general x-ray conebeam

More information

Validation of a digital mammography image simulation chain with automated scoring of CDMAM images

Validation of a digital mammography image simulation chain with automated scoring of CDMAM images Validation of a digital mammography image simulation chain with automated scoring of CDMAM images Mary Yip 1, Abdulaziz Alsager 2, Emma Lewis 1, Kevin Wells 1, and Kenneth C. Young 3 1 Centre for Vision,

More information

Power Spectrum Analysis of an Anthropomorphic Breast Phantom Compared to Patient Data in 2D Digital Mammography and Breast Tomosynthesis

Power Spectrum Analysis of an Anthropomorphic Breast Phantom Compared to Patient Data in 2D Digital Mammography and Breast Tomosynthesis Power Spectrum Analysis of an Anthropomorphic Breast Phantom Compared to Patient Data in 2D Digital Mammography and Breast Tomosynthesis Lesley Cockmartin 1,*, Predrag R. Bakic 2, Hilde Bosmans 1, Andrew

More information

Simulation of Internal Backscatter Effects on MTF and SNR of Pixelated Photon-counting Detectors

Simulation of Internal Backscatter Effects on MTF and SNR of Pixelated Photon-counting Detectors Simulation of Internal Backscatter Effects on MTF and SNR of Pixelated Photon-counting Detectors Alexander Korn, Juergen Giersch a and Martin Hoheisel b a Physikalisches Institut Universitaet Erlangen-Nuernberg,

More information

Digital breast tomosynthesis: comparison of different methods to calculate patient doses

Digital breast tomosynthesis: comparison of different methods to calculate patient doses Digital breast tomosynthesis: comparison of different methods to calculate patient doses Poster No.: C-2220 Congress: ECR 2011 Type: Scientific Paper Authors: A. Jacobs 1, L. Cockmartin 1, D. R. Dance

More information

Noise power spectrum and modulation transfer function analysis of breast tomosynthesis imaging

Noise power spectrum and modulation transfer function analysis of breast tomosynthesis imaging Noise power spectrum and modulation transfer function analysis of breast tomosynthesis imaging Weihua Zhou a, Linlin Cong b, Xin Qian c, Yueh Z. Lee d, Jianping Lu c,e, Otto Zhou c,e, *Ying Chen a,b a

More information

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT Assessment of 3D performance metrics D and 3D Metrics of Performance Towards Quality Index: Volumetric imaging systems X-ray based Volumetric imaging systems: CBCT/CT Tomosynthesis Samuel Richard and Ehsan

More information

Detection of microcalcification clusters by 2D-mammography and narrow and wide angle digital breast tomosynthesis

Detection of microcalcification clusters by 2D-mammography and narrow and wide angle digital breast tomosynthesis Detection of microcalcification clusters by 2D-mammography and narrow and wide angle digital breast tomosynthesis Andria Hadjipanteli a, Premkumar Elangovan b, Padraig T Looney a, Alistair Mackenzie a,

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

Lorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc.

Lorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc. Lorad FFDM QC Procedures for Medical Physicist Tao Wu, Ph.D. Hologic, Inc. Lorad Selenia FFDM Tests Following Lorad QC Manual Collimation Assessment Artifact Evaluation System Resolution Phantom Image

More information

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Wenjuan Sun and Stephen Brown 10 th May 2016 1 Possible factors influencing XCT measurements Components Influencing variables Possible

More information

Monte Carlo simulation of scatter field for calculation of contrast of discs in synthetic CDMAM images

Monte Carlo simulation of scatter field for calculation of contrast of discs in synthetic CDMAM images Monte Carlo simulation of scatter field for calculation of contrast of discs in synthetic CDMAM images Oliver Díaz 1, Mary Yip 1, Jorge Cabello 1, David R. Dance 2,3, Kenneth C. Young 2,3, and Kevin Wells

More information

Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors

Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors Elena Marimon 1,2*, Hammadi Nait-Charif 1, Asmar Khan 2, Philip A. Marsden 3, Oliver Diaz 4 1 Centre for

More information

Z-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL

Z-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL Control-X Medical Z-MOTION Compact design, low ceiling height requirement Motorized and manual movement capability Wide motion / SID range Best-in-class image quality Flexible connectivity to PACS systems

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc.

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc. Unique Features of the GE Senoclaire Tomosynthesis System Tyler Fisher, M.S., DABR Therapy Physics, Inc. Conflict of Interest Disclosure I have no conflicts to disclose. Learning Objectives Overview of

More information

Certificate Extension II 8/1/2016. Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role

Certificate Extension II 8/1/2016. Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role Kish Chakrabarti, Ph.D., FAAPM Division of Mammography Quality Standards Center for Devices and Radiological

More information

Digital Scatter Removal in Mammography to enable Patient Dose Reduction

Digital Scatter Removal in Mammography to enable Patient Dose Reduction Digital Scatter Removal in Mammography to enable Patient Dose Reduction Mary Cocker Radiation Physics and Protection Oxford University Hospitals NHS Trust Chris Tromans, Mike Brady University of Oxford

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

8/2/2016. Hands-on GE SenoClaire DBT: technical characteristics & quality control. Disclosures and Acknowledgements

8/2/2016. Hands-on GE SenoClaire DBT: technical characteristics & quality control. Disclosures and Acknowledgements Hands-on GE SenoClaire DBT: technical characteristics & quality control Razvan Iordache, Ph.D. GE Healthcare Imagination at work Disclosures and Acknowledgements Razvan Iordache Employee of GE Healthcare

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

Detector Noise evaluation by means of Continue Wavelet Transform. Comparison with Fourier Transform methods

Detector Noise evaluation by means of Continue Wavelet Transform. Comparison with Fourier Transform methods Detector Noise evaluation by means of Continue Wavelet Transform. Comparison with Fourier Transform methods Poster No.: C-0215 Congress: ECR 2014 Type: Scientific Exhibit Authors: N. Kalyvas 1, S. Angelakis

More information

A Software Tool for Measurement of the Modulation Transfer Function

A Software Tool for Measurement of the Modulation Transfer Function A Software Tool for Measurement of the Modulation Transfer Function Yen-Hong Kao a, Michael Albert a, Ann-Katherine Carton a, Hilde Bosmans b, Andrew D. A. Maidment a a Department of Radiology, University

More information

Scatter reduction for grid-less mammography using the convolutionbased image post-processing technique

Scatter reduction for grid-less mammography using the convolutionbased image post-processing technique Scatter reduction for grid-less mammography using the convolutionbased image post-processing technique Elena Marimón *a,b, Hammadi Nait-Charif a, Asmar Khan b, Philip A. Marsden c, Oliver Diaz d a CDE

More information

Methodology of NEQ (f) analysis for optimization and comparison of digital breast tomosynthesis acquisition techniques and reconstruction algorithms

Methodology of NEQ (f) analysis for optimization and comparison of digital breast tomosynthesis acquisition techniques and reconstruction algorithms Methodology of NEQ (f) analysis for optimization and comparison of digital breast tomosynthesis acquisition techniques and reconstruction algorithms Ying Chen a,b, Joseph Y. Lo a,b,c,d, Nicole T. Ranger

More information

Simple method for modulation transfer function determination of digital imaging detectors from edge images

Simple method for modulation transfer function determination of digital imaging detectors from edge images Simple method for modulation transfer function determination of digital imaging detectors from edge images Egbert Buhr *a, Susanne Günther-Kohfahl b, Ulrich Neitzel b a Physikalisch-Technische Bundesanstalt,

More information

Quality control phantoms and protocol for a tomography system

Quality control phantoms and protocol for a tomography system Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Investigating Oblique Reconstructions with Super-Resolution in Digital Breast Tomosynthesis

Investigating Oblique Reconstructions with Super-Resolution in Digital Breast Tomosynthesis Investigating Oblique Reconstructions with Super-Resolution in Digital Breast Tomosynthesis Raymond J. Acciavatti, Stewart B. Mein, and Andrew D.A. Maidment University of Pennsylvania, Department of Radiology,

More information

Linköping University Post Print A MONTE CARLO-BASED MODEL FOR SIMULATION OF DIGITAL CHEST TOMOSYNTHESIS

Linköping University Post Print A MONTE CARLO-BASED MODEL FOR SIMULATION OF DIGITAL CHEST TOMOSYNTHESIS Linköping University Post Print A MONTE CARLO-BASED MODEL FOR SIMULATION OF DIGITAL CHEST TOMOSYNTHESIS Gustaf Ullman, David R. Dance, Michael Sandborg, Gudrun Alm Carlsson, Angelica Svalkvist and Magnus

More information

A Method for Producing Simulated Mammograms: Observer Study

A Method for Producing Simulated Mammograms: Observer Study A Method for Producing Simulated Mammograms: Observer Study Payam Seifi M.Sc. Michael R. Chinander Ph.D. Robert M. Nishikawa Ph.D., FAAPM Carl J. Vyborny Translational Laboratory for Breast Imaging Research

More information

Partial Volume Simulation in Software Breast Phantoms

Partial Volume Simulation in Software Breast Phantoms Partial Volume Simulation in Software Breast Phantoms Feiyu Chen, David Pokrajac, Xiquan Shi, Fengshan Liu, Andrew D.A. Maidment a, Predrag R. Bakic a Delaware State University, 1200 N DuPont Hwy, Dover

More information

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Stefan C. Saur 1, Hatem Alkadhi 2, Luca Regazzoni 1, Simon Eugster 1, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision

More information

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube T. Donath* a, F. Pfeiffer a,b, O. Bunk a, W. Groot a, M. Bednarzik a, C. Grünzweig a, E. Hempel c, S. Popescu c, M. Hoheisel

More information

DX-D 100 with wireless detector

DX-D 100 with wireless detector M O B I L E D R S O L U T I O N DX-D 100 with wireless detector Patients who most need imaging exams may lack the mobility necessary to move to the X-ray room or to position themselves properly for optimum

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

ACMP 25th Annual Meeting

ACMP 25th Annual Meeting Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation ACMP 25th Annual Meeting Seattle, WA May 3, 2008 Melissa C. Martin, M.S., FACR Therapy Physics, Inc. 879 West 190th St., Ste 419,

More information

DX-D 100 with WITH ITS EXCELLENT IMAGE QUALITY AND

DX-D 100 with WITH ITS EXCELLENT IMAGE QUALITY AND M O B I L E D R S O L U T I O N DX-D 100 with wireless detector Patients who most need imaging exams may lack the mobility necessary to move to the X-ray room or to position themselves properly for optimum

More information

Acknowledgments and financial disclosure

Acknowledgments and financial disclosure AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories

More information

Extracting and Mathematical Identifying Form of Stationary Noise in X-ray Images

Extracting and Mathematical Identifying Form of Stationary Noise in X-ray Images Original Paper Forma, 29, S37 S43, 2014 Extracting and Mathematical Identifying Form of Stationary Noise in X-ray Images Akihiro Sugiura 1,2, Kiyoko Yokoyama 1, Hiroki Takada 3,Akiko Ihori 2, Naruomi Yasuda

More information

Towards full-body X-ray images

Towards full-body X-ray images Towards full-body X-ray images Christoph Luckner 1,2, Thomas Mertelmeier 2, Andreas Maier 1, Ludwig Ritschl 2 1 Pattern Recognition Lab, FAU Erlangen-Nuernberg 2 Siemens Healthcare GmbH, Forchheim christoph.luckner@fau.de

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters Monte Carlo Simulation for Neptun 1 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters M.T. Bahreyni Toossi a, M. Momen Nezhad b, S.M. Hashemi a a Medical Physics Research Center,

More information

Task-based Assessment of X-ray Breast Imaging Systems Using Insilico

Task-based Assessment of X-ray Breast Imaging Systems Using Insilico 55 th : Virtual Tools for Validation of X-ray Breast Imaging Systems Task-based Assessment of X-ray Breast Imaging Systems Using Insilico Modeling Tools Rongping Zeng Division of Imaging and Applied Mathematics

More information

Non-Stationary CT Image Noise Spectrum Analysis

Non-Stationary CT Image Noise Spectrum Analysis Non-Stationary CT Image Noise Spectrum Analysis Michael Balda, Björn J. Heismann,, Joachim Hornegger Pattern Recognition Lab, Friedrich-Alexander-Universität Erlangen Siemens Healthcare, Erlangen michael.balda@informatik.uni-erlangen.de

More information

Optimization of optical systems for LED spot lights concerning the color uniformity

Optimization of optical systems for LED spot lights concerning the color uniformity Optimization of optical systems for LED spot lights concerning the color uniformity Anne Teupner* a, Krister Bergenek b, Ralph Wirth b, Juan C. Miñano a, Pablo Benítez a a Technical University of Madrid,

More information

FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300

FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300 The DX-D 300 DR system unites excellent image quality with complete convenience. It offers top-of-the-line technology, a single detector and a fully-motorized

More information

Scatter Correction Methods in Dimensional CT

Scatter Correction Methods in Dimensional CT Scatter Correction Methods in Dimensional CT Matthias Baer 1,2, Michael Hammer 3, Michael Knaup 1, Ingomar Schmidt 3, Ralf Christoph 3, Marc Kachelrieß 2 1 Institute of Medical Physics, Friedrich-Alexander-University

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

Industrial Radiography Simulation by MCNPX for Pipeline Corrosion Detection

Industrial Radiography Simulation by MCNPX for Pipeline Corrosion Detection More Info at Open Access Database www.ndt.net/?id=18672 Industrial Radiography Simulation by MCNPX for Pipeline Corrosion Detection Mostafa KHODABANDELOU 1, Amir MOVAFEGHI 2 *, Peyman ROSTAMI 1 and Jahanghir

More information

RECENTLY APPROVED GE FFDM DBT TESTING PROCEDURES.

RECENTLY APPROVED GE FFDM DBT TESTING PROCEDURES. RECENTLY APPROVED GE FFDM DBT TESTING PROCEDURES. S. G U R U P R A S A D, P H. D., D A B R, E M E R I T U S, M AT H E W H A L L M. S. N O RT H S H O R E U N I V E R S I T Y H E A LT H S Y S T E M A N D

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology.

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology. Photon Counting Detectors and Their Applications in Medical Imaging Principles, Pitfalls and Progress in Photon-Counting-Detector Technology Taly Gilat Schmidt, PhD Associate Professor Department of Biomedical

More information

Application of MCNP Code in Shielding Design for Radioactive Sources

Application of MCNP Code in Shielding Design for Radioactive Sources Application of MCNP Code in Shielding Design for Radioactive Sources Ibrahim A. Alrammah Abstract This paper presents three tasks: Task 1 explores: the detected number of as a function of polythene moderator

More information

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE / Radiographic Tomosynthesis: Image Quality and Artifacts Reduction Baojun Li, Ph.D Department of Radiology Boston University Medical Center 2012 AAPM Annual Meeting Background Linear Trajectory Tomosynthesis

More information

Limitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods

Limitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods Stereoscopic Breast Imaging Andrew D. A. Maidment, Ph.D. Chief, Physics Section Department of Radiology University of Pennsylvania Limitations of Projection Radiography Mammography is a projection imaging

More information

Investigation of the effect of varying scatter-toprimary ratios on nodule contrast in chest tomosynthesis

Investigation of the effect of varying scatter-toprimary ratios on nodule contrast in chest tomosynthesis Investigation of the effect of varying scatter-toprimary ratios on nodule contrast in chest tomosynthesis Angelica Svalkvist, Gustaf Ullman, Markus Håkansson, David Dance, Michael Sandborg, Gudrun Alm

More information

8/2/2016. Acknowledgement. Common Clinical Questions. Presumption Images are Good Enough to accurately answer clinical questions

8/2/2016. Acknowledgement. Common Clinical Questions. Presumption Images are Good Enough to accurately answer clinical questions Image Quality Assessment using Model Observers: Clinical Implementation and Practical Considerations Shuai Leng, PhD Associate Professor, Department of Radiology Mayo Clinic, Rochester MN Acknowledgement

More information

Performance Evaluation of CT Systems

Performance Evaluation of CT Systems 8/2/206 Performance Evaluation of CT Systems AAPM Task Group 233 Foundations, Methods, and Prospects Ehsan Samei Duke University Medical Center Disclosures Research grant: NIH R0 EB00838 Research grant:

More information

Design Considerations in Optimizing a Breast Tomosynthesis System

Design Considerations in Optimizing a Breast Tomosynthesis System Design Considerations in Optimizing a Breast Tomosynthesis System Andrew Smith, Ph.D., Vice President - Imaging Science, Hologic Introduction Breast tomosynthesis, also referred to as three-dimensional

More information

Application of Tatian s Method to Slanted-Edge MTF Measurement

Application of Tatian s Method to Slanted-Edge MTF Measurement Application of s Method to Slanted-Edge MTF Measurement Peter D. Burns Eastman Kodak Company, Rochester, NY USA 465-95 ABSTRACT The 33 method for the measurement of the spatial frequency response () of

More information

Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction

Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction Klaus Erhard a, Michael Grass a, Sebastian Hitziger b, Armin Iske b and Tim Nielsen a a Philips Research Europe Hamburg,

More information

A fast scatter field estimator for Digital Breast Tomosynthesis

A fast scatter field estimator for Digital Breast Tomosynthesis A fast scatter field estimator for Digital Breast Tomosynthesis Oliver Díaz a, David R. Dance b,c, Kenneth C. Young b,c, Premkumar Elangovan a, Predrag R. Bakic d and Kevin Wells a a Centre for Vision,

More information

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT.

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT. Progress report: Development of assessment and predictive metrics for quantitative imaging in chest CT Subaward No: HHSN6801000050C (4a) PI: Ehsan Samei Reporting Period: month 1-18 Deliverables: 1. Deployment

More information

ABSTRACT 1. INTRODUCTION

ABSTRACT 1. INTRODUCTION Estimating breast thickness for dual-energy subtraction in contrastenhanced digital mammography using calibration phantoms Kristen C. Lau 1, Young Joon Kwon, Moez Karim Aziz, Raymond J. Acciavatti, Andrew

More information

Fujifilm DR Solution. FDR AcSelerate. The new pinnacle in diagnostic imaging from Fujifilm ISS. CsI. Dynamic Visualization. Technology.

Fujifilm DR Solution. FDR AcSelerate. The new pinnacle in diagnostic imaging from Fujifilm ISS. CsI. Dynamic Visualization. Technology. Fujifilm DR Solution FDR AcSelerate The new pinnacle in diagnostic imaging from Fujifilm CsI Scintillator ISS Technology Dynamic Visualization Welcome to the X-ray room of the future! A streamlined solution

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60601-2-44 2001 AMENDMENT 1 2002-09 Amendment 1 Medical electrical equipment Part 2-44: Particular requirements for the safety of X-ray equipment for computed tomography Amendement

More information

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling nd International Symposium on NDT in Aerospace - We..A. Comparison of Probing Error in Dimensional Measurement by Means of D Computed Tomography with Circular and Helical Sampling Jochen HILLER, Stefan

More information

Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.

Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran. DXRaySMCS First User Friendly Interface Developed for Prediction of Diagnostic Radiology X-Ray Spectra Produced by Monte Carlo (MCNP-4C) Simulation in Iran M.T. Bahreyni Toosi a*, H. Moradi b, H. Zare

More information

Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J.

Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging Stephen J. Glick* J. Michael O Connor**, Clay Didier**, Mini Das*, * University of Massachusetts Medical

More information

Development and Evaluation of 2D and 3D Image Quality Metrics. Simon Nicholas Murphy. Graduate Program in Medical Physics Duke University

Development and Evaluation of 2D and 3D Image Quality Metrics. Simon Nicholas Murphy. Graduate Program in Medical Physics Duke University Development and Evaluation of 2D and 3D Image Quality Metrics by Simon Nicholas Murphy Graduate Program in Medical Physics Duke University Date: Approved: Ehsan Samei, Supervisor James T. Dobbins, III

More information

Scaling Calibration in the ATRACT Algorithm

Scaling Calibration in the ATRACT Algorithm Scaling Calibration in the ATRACT Algorithm Yan Xia 1, Andreas Maier 1, Frank Dennerlein 2, Hannes G. Hofmann 1, Joachim Hornegger 1,3 1 Pattern Recognition Lab (LME), Friedrich-Alexander-University Erlangen-Nuremberg,

More information

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications 4th International Symposium on NDT in Aerospace 2012 - Tu.3.A.1 Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications Malte KURFISS 1, Gerd STRECKENBACH 2 1 YXLON International

More information

Wissenschaftliche Highlights der GSF 2007

Wissenschaftliche Highlights der GSF 2007 H Forschungszentrum für Umwelt und Gesundheit GmbH in der Helmholtzgemeinschaft Wissenschaftlich-Technische Abteilung Wissenschaftliche Highlights der GSF 2007 Abfrage April 2007 Institut / Selbst. Abteilung

More information

Pipe Inspection with Digital Radiography

Pipe Inspection with Digital Radiography NDT INNOVATIVE COMPACT RUGGED Portable Digital Radiography Systems Pipe Inspection with Digital Radiography Railways www. NOVO-DR.com Pipe Inspection with Digital Radiography Digital Radiography or DR

More information

DIGITALIZATION OF A MAMMOGRAPHIC PHANTOM VIEW THROUGH A MONTE CARLO SIMULATION

DIGITALIZATION OF A MAMMOGRAPHIC PHANTOM VIEW THROUGH A MONTE CARLO SIMULATION DIGITALIZATION OF A MAMMOGRAPHIC PHANTOM VIEW THROUGH A MONTE CARLO SIMULATION M. Ramos 1, S. Ferrer 1, J.I. Villaescusa 2, G. Verdú 1, J.M. Campayo 3 1 Departamento de Ingeniería Química y Nuclear. Universidad

More information

D&S Technical Note 09-2 D&S A Proposed Correction to Reflectance Measurements of Profiled Surfaces. Introduction

D&S Technical Note 09-2 D&S A Proposed Correction to Reflectance Measurements of Profiled Surfaces. Introduction Devices & Services Company 10290 Monroe Drive, Suite 202 - Dallas, Texas 75229 USA - Tel. 214-902-8337 - Fax 214-902-8303 Web: www.devicesandservices.com Email: sales@devicesandservices.com D&S Technical

More information

Measurement of Skin Dose

Measurement of Skin Dose Measurement of Skin Dose Sources of Uncertainty Kenneth A. Fetterly, Ph.D. William Pavlicek, Ph.D. Dan Bednarek, PhD 2014 AAPM Annual Meeting, Austin Texas 2013 MFMER slide-1 Purpose 1. Present a framework

More information

4/19/2016. Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT

4/19/2016. Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT 1 Three manufacturers approved for Tomo Hologic and GE, and Siemens Why 2D Digital Mammography 2D FFDM it appears to

More information

DX-D 300. Flexible Direct Radiography System

DX-D 300. Flexible Direct Radiography System DX-D 300 Flexible Direct Radiography System MUSICA 2 processing provides superior contrast detail and exam-independent, consistent image quality Cesium Iodide DR detector technology offers potential for

More information

International Conference on Space Optics ICSO 2008 Toulouse, France October 2008

International Conference on Space Optics ICSO 2008 Toulouse, France October 2008 ICSO 008 14 17 October 008 Edited by Josiane Costeraste, Errico Armandillo, and Nikos Karafolas Coupled thermo-elastic and optical performance analyses of a reflective baffle for the BepiColombo laser

More information

Registration concepts for the just-in-time artefact correction by means of virtual computed tomography

Registration concepts for the just-in-time artefact correction by means of virtual computed tomography DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Registration concepts for the just-in-time artefact correction by means of virtual

More information

DX-D 100 WITH WIRELESS DETECTOR

DX-D 100 WITH WIRELESS DETECTOR DX-D 100 WITH WIRELESS DETECTOR MOBILE DR SOLUTION WITH ITS EXCELLENT IMAGE QUALITY AND FLEXIBLE HANDLING, THE MOBILE DX-D 100 WITH WIRELESS DETECTOR OFFERS FAST IMAGING THAT CAN BE VALIDATED IMMEDIATELY.

More information

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Beam Attenuation Grid Based Scatter Correction Algorithm for More Info at Open Access Database

More information

The digital EVOlution in cassette size format

The digital EVOlution in cassette size format http://cr-pacs.com/ Platzhalter Wireless Mobile Fast Retrofit Light Low Dose Highest DQE Highest MTF CsI GOS The digital EVOlution in cassette size format Your wireless entry to the world of Digital Radiography

More information

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct J. DHAENE* 1, E. PAUWELS 1, T. DE SCHRYVER 1, A. DE MUYNCK 1, M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics

More information

Product data. Mammographic System

Product data. Mammographic System Technologically advanced mammographic system, offering the best quality/price ratio. Lightweight and easy-to-use, facilitating the operator s job. Excellent diagnostic image quality, similar to the top-of

More information

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing ImPACT Information Leaflet No. 1: CT Scanner Acceptance Testing Version 1.02, 18/05/01 CONTENTS: 1. SCOPE OF LEAFLET 2. GENERAL PRINCIPLES OF ACCEPTANCE AND COMMISSIONING 2.1 PHANTOMS 2.2 EXPOSURE AND

More information

Optimisation of Toshiba Aquilion ONE Volume Imaging

Optimisation of Toshiba Aquilion ONE Volume Imaging Optimisation of Toshiba Aquilion ONE Volume Imaging Jane Edwards, RPRSG Royal Free London NHS Foundation Trust Dr Mufudzi Maviki, Plymouth Hospitals NHS Trust Background In 2011/12 Radiology at RFH was

More information

RaySafe Solo. Specifications

RaySafe Solo. Specifications RaySafe Solo Specifications RAYSAFE SOLO GENERAL EMC TESTED According to EN 61000-6-1:2007 and EN 61000-6-3:2007 EXPOSURE NEEDED One RESET Automatic TEMP. 15 35 C (59 95 F) DETECTOR CABLE LENGTH 2 and

More information

DUE to beam polychromacity in CT and the energy dependence

DUE to beam polychromacity in CT and the energy dependence 1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact

More information

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel Proc. Of SPIE Vol 0626, Application of Optical Instrumentation in Medicine XIV and Picture Archiving and Communication Systems(PACS IV) for Medical Applications IV, ed. S J Dwyer/R H Schneider (Jan 1986)

More information

Refined Measurement of Digital Image Texture Loss Peter D. Burns Burns Digital Imaging Fairport, NY USA

Refined Measurement of Digital Image Texture Loss Peter D. Burns Burns Digital Imaging Fairport, NY USA Copyright 3 SPIE Refined Measurement of Digital Image Texture Loss Peter D. Burns Burns Digital Imaging Fairport, NY USA ABSTRACT Image texture is the term given to the information-bearing fluctuations

More information

Department of Biomedical Physics and Bioengineering, University of Aberdeen, Foresterhill, Aberdeen AB25 2ZD (UK)

Department of Biomedical Physics and Bioengineering, University of Aberdeen, Foresterhill, Aberdeen AB25 2ZD (UK) Technical Note Physica Medica - Vol. XVIII, N. 1, January-March 2002 Calculation of the Modulation Transfer Function from the Square Wave Response Function data with an interactive curve fitting software

More information

DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300

DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300 DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM MUSICA processing provides excellent contrast detail and exam-independent, consistent image quality Cesium Iodide DR detector technology offers potential for

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

UNCOMPROMISING QUALITY

UNCOMPROMISING QUALITY ION CHAMBERS UNCOMPROMISING QUALITY Designed with over 30 years of scientific integrity for a broad range of dosimetry measurements in diverse radiation beams Farmer-type Chambers For absolute dosimetry

More information

Adaptive detection mechanisms in globally statistically nonstationary-oriented noise

Adaptive detection mechanisms in globally statistically nonstationary-oriented noise Zhang et al. Vol. 23, No. 7/July 2006/J. Opt. Soc. Am. A 1549 Adaptive detection mechanisms in globally statistically nonstationary-oriented noise Yani Zhang, Craig K. Abbey, and Miguel P. Eckstein Vision

More information