1 AAPM-SAM-2012-Das (1) Beam Data Collection and Commissioning for Linear Accelerators: Technical Considerations and Recommendations Preface Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University of School of Medicine & Indiana University Health Proton Center Indiana Med. Phys. 35(9), , 2008 Mechanical A Planning for Commissioning Data Radiation Survey Mechanical tests Light radiation Table, Collimator, Gantry Jaws Table Collimator Gantry MLC 2008 Imaging parameters Other as TG-142 4udrant Independent Jaw Test
2 AAPM-SAM-2012-Das (2) Planning for Commissioning Time Rational For Commissioning Beam Data Rational For Not Using Golden Data Time= [(PDD + 5 profiles)/beam energy] x ( open + 4 wedges) x 60 points/scan x [(1 s/pts + (1s/movement and delay)] x (15 fields x 2 energies) First, it is not evident that manufacturing procedures for all linear accelerators have produced a level of reproducibility acceptable for clinical use. For example, variations in beam parameters have been noted between beams with the same nominal energies. Second, on-site changes made during installation and acceptance of the user s accelerator e.g., changes in beam energy and/or profiles from beam steering will not be modeled in the golden data. Third, the beam characteristics of the soft wedges are made by moving jaws that depend on the speed parameters of the jaws and a deviation at site could affect the beam profile of the soft wedge. TG-106 Fourth, although acceptable agreement with the golden data set may be found in individual checks, it may be that some clinical setups will have multiple errors, which combine to produce unacceptable results. Hrbacek, et al Med. Phys. 34, , s 30 h Rational For Not Using Golden Data Definition of Detectors Setting Water Tank & Detector x i x <Δ, x i Δ =? (0.5, 1.0 0r 2%) Standard chamber 10 1 cm 3 The active volume for a standard Farmer-type ionization chamber is on average 0.6 cm 3. Minichamber 10 2 cm 3 The active volume for a mini-ionization chamber is on average 0.05 cm 3. Microchamber 10 3 cm 3 The active volume for a microionization chamber is on average cm 3 and ideally suited for small field dosimetry such as radiosurgery,gamma knife, CyberKnife, and IMRT Air Water Hrbacek, et al uantitative evaluation of a beam-matching procedure using one dimensional gamma analysis, Med. Phys. 34, , 2007.
3 AAPM-SAM-2012-Das (3) Know Your Connectors Understand Detector, Connector & Cable uality of Cables Srivastava et al, SU-GG-T-270, 2010 Setup and Possible Errors Electrometer Null Setting Cable subtraction Proper bias >300 V for ion chamber 100 V for diamond 0 v for all diodes Proper gain Proper mode Choose Consistent & Correct Polarity
4 AAPM-SAM-2012-Das (4) Chambers & Gain Selection of detector for beam data Choice of Detector Orientation Radiation beam Chamber Orientations Y Z X Z X Y Scan Direction Detector Orientation 6MV, 2x2 cm2 field, Illustration of chamber volume effects Diode, dmax 90 PTW Pinpoint, dmax 80 RK chamber, dmax Relative Dose Distance Off Axis (cm)
5 AAPM-SAM-2012-Das (5) 6 MV 60 Deg Wedge, 10 cm depth: water vs Profiler MV 60 Degree Wedge, 10 cm depth, 100 cm SAD: Water vs Profiler 200 Scanning Speed Diode Array Profile Water Profile 200 Diode Array profile Water Profile Relative Dose Relative Dose Distance Off Axis(cm) Distance Off Axis(cm) Scanning Speed Percent Depth Dose Arithmetic Mean (AM) smoothing: 60 Degree Wedge PDD Unsmoothed AM x 1 AM x 2 AM x 3 AM x Depth(cm)
6 AAPM-SAM-2012-Das (6) Future of Beam Data Commissioning Monte Carlo Codes Standardization of linear accelerators Monte Carlo based commissioning Newer Radiation Detectors & Cables Newer Scanning Systems Smart algorithms Aubin et al., Med Phys, 37(5), , 2010 Aubin et al., Med Phys, 37(5), , 2010 Simulation of intensity at target Simulated Profiles Profiles for different fields Aubin et al., Med Phys, 37(5), , 2010 Aubin et al., Med Phys, 37(5), , 2010
7 AAPM-SAM-2012-Das (7) Depth Dose Simulation Aubin et al., Med Phys, 37(5), , 2010 Detectors Detector Manufacturer Type volume SFD Scanditronix Photon diode 1.7x10-5 cm 3 PFD Scanditronix Photon diode 1.9x10-4 cm 3 Exradin A-16 Standard Imaging Ion chamber 0.007cm 3 Wellhofer-IC4 Scanditronix Ion chamber 0.40 cm 3 Pinpoint PTW Ion chamber 0.015cm cc PTW Ion chamber 0.125cm 3 0.3cc PTW Ion chamber 0.3 cm 3 0.6cc PTW Ion chamber 0.6 cm 3 Diamond PTW Diamond 0.003cm 3 Markus PTW Parallel plate 0.055cm 3 Edge Detector Sun Nuclear Diode 10-5 cm 3 Other Relative sensitivity Sensitivity vs Volume of Detectors PFD cc SFD 0.3cc cc 10 A-16 PinPoint Markus IC E E E E E-01 Volume (cm 3 ) Magna-Fields 200x200 cm 2 Treatment Fields Traditional Fields 40x40 cm 2 4x4 cm 2 Small Field 4x4 cm 2 0.3x0.3 cm 2 Advance Therapy Fields SRS/SRT Gamma Knife Cyber-Knife Tomotherapy IMRT What is a Small Field? Lack of charged particle Dependent on the range of secondary electrons Photon energy Collimator setting that obstructs the source size Detector is comparable to the field size Views of Source Sizes Jaffray et al, Med Phys 20, (1993).
8 AAPM-SAM-2012-Das (8) Definition of Small Fields Calculation of Fluence Map Elements Dosimetry Raytrace to calculate attenuation or use approximate penumbra shape For each element, find the contributions from the relevant sources Dominating effect: The width and shape of the source Absolute Dose Relative Depth Dose [D(r,d)/D(r,dm)] TMR Profiles Output, S cp (total scatter factor), [D(r)/D(ref)] Das et al, Med. Phys. 35, , 2008 Courtesy Ahnesjö POI-eye-view of the source! Relative dose at dmax 6 MV; Central Axis Scanditronix-SFD Scanditronix-PFD 0.5 Exradin-A PTW-Pinpoint PTW-0.125cc 0.3 PTW-0.3cc PTW-0.6cc 0.2 PTW-Markus Wellhofer-IC Field Size (cm) Das et al, TG-106, Med Phys, 35, 4186, 2008 Relative dose at dmax Field Size Limit for Accurate Dose Measurements with Available Detectors 15 MV; Central Axis Scanditronix-SFD Scanditronix-PFD Exradin-A16 PTW-Pinpoint PTW-0.125cc PTW-0.3cc PTW-0.6cc PTW-Markus Wellhofer-IC Field Size (cm) D = M Ω k k f w, f clin, f clin, Relative Dosimetry f N DW, o k, o M ( D ) ( ) clin w, M clin M clin f f f = M ( D ) ( ) w, / M M f ( Dw, )/( M ) ( Output ) clin w, clin rel ( Dw, )/( M w, ) (Re ading rel =, f clin, = = ) ( S = ( S w, air ) w, air ) f k f, f ref, / clin, f k f clin, P P New Data on Correction Factor Pantelis et al, Med Phys, 37(6), , 2010
9 AAPM-SAM-2012-Das (9) clin clin f, f, k Correction factor , f k clin, Siemens; PTW diode Elekta; PTW diode Siemens; Exradin A16 Elekta; Exradin A16 Siemens; Sun Nuclear Dedge "Elekta; Sun Nuclear Dedge" Siemens; PTW Pinpoint Elekta; PTW Pinpoint Siemens; PTW microlion Elekta; PTW microlion Field size (mm) New Data on Correction Factor Chung et al, Med Phys, 37(6), , 2010 k is not Constant in Small Field Kawachi et al, Med Phys, 35, , 2008 Depth Dose & Source Size Profile & Source Size Comparison of Large Tank and Small SRS Cylinder Tank for SRS, TMR & Profiles Moving Tank System for TMR Sham et al, Med Phys, 35, , 2008 Sham et al, Med Phys, 35, , 2008 ARM Inc., Port Saint Lucie, FL 34983
10 AAPM-SAM-2012-Das (10) Direct TMR Data Acquisition No SSD to SAD calculations required, No cubic spline fit of a limited number of fixed data points needed Calculated TPR ~2% less at depth Cubic spline fit of 12 data points 3D Scanner (Sun Nuclear) Setup Subjectivity Automatic leveling, water surface detection and beam center detection No tank shifts Detector orientation/resolution 3D Scanner design always using the short dimension of chamber to scan Time Setup is faster and more accurate No tank shifts Smaller tank fills and drains faster Scanner Orientation Advantage Nikesch et al, CyberKnife Center, Palm Beach, FL Sun Nuclear 3D Scanner Ring drive maintains consistent scanning direction 2. Diameter drive has maximum scanning range of 640mm 3. Vertical drive has maximum travel of 400mm Conclusions Golden Data should be taken as a reference only Understand time and amount of data to be taken View each parameters properly, double check by another individual Use proper detector for each type of data collection Set optimum speed for scanning, do not rush -Conclusions Understand the limits and measuring condition uestion every unusual data set Do not smooth data too much Write report for future reference Future technology & resources could help commissioning simpler
11 AAPM-SAM-2012-Das (11) Thanks
UNCOMPROMISING QUALITY
ION CHAMBERS UNCOMPROMISING QUALITY Designed with over 30 years of scientific integrity for a broad range of dosimetry measurements in diverse radiation beams Farmer-type Chambers For absolute dosimetry
More informationComparison of Two Phantoms for End-to-End SRS/SBRT Testing. Vikren Sarkar, Ph.D. Associate Professor University of Utah Huntsman Cancer Center
Comparison of Two Phantoms for End-to-End SRS/SBRT Testing Vikren Sarkar, Ph.D. Associate Professor University of Utah Huntsman Cancer Center Disclosure Travel and Speaker Honorarium from Sun Nuclear Corp.
More informationFAST, precise. qa software
qa software FAST, precise Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, rotational or brachytherapy plans no film, no phantoms, no linac
More informationIMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE
QA SOFTWARE FAST, PRECISE Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, VMAT, SRS and brachytherapy plans no film, no phantoms, no linac
More informationA DOSIMETRIC MODEL FOR SMALL-FIELD ELECTRON RADIATION THERAPY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL
A DOSIMETRIC MODEL FOR SMALL-FIELD ELECTRON RADIATION THERAPY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE MASTER OF
More informationAnalysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code
American Journal of Biomedical Engineering 216, 6(4): 124-131 DOI: 1.5923/j.ajbe.21664.3 Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code Ankit Kajaria 1,*, Neeraj Sharma
More informationDisclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions
LINAC and MLC QA for IMRT Ping Xia, Ph.D., Department of Radiation Oncology Disclosure Received research support from Siemens Medical Solutions University of California San Francisco Therapy Series (SAM)
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Evaluation of dosimetric effect of leaf position in a radiation field of an 80-leaf multileaf collimator fitted to the LINAC
More informationMachine and Physics Data Guide
WWW..COM Machine and Physics Data Guide STANDARD IMAGING, INC. 3120 Deming Way Middleton, WI 53562-1461 May / 2008 2008 Standard Imaging, Inc. TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 www.standardimaging.com
More information3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools
3D SCANNER 3D Scanning Comes Full Circle Your Most Valuable QA and Dosimetry Tools BETTER OUTCOMES THROUGH TECHNOLOGY The 3D SCANNER is different by design. 3D SCANNER has been developed from the ground
More informationA SYSTEM OF DOSIMETRIC CALCULATIONS
A SYSTEM OF DOSIMETRIC CALCULATIONS INTRODUCTION Dose calculation based on PDD and TAR have Limitations The dependence of PDD on SSD Not suitable for isocentric techniques TAR and SAR does not depend on
More informationOutput factor comparison of Monte Carlo and measurement for Varian TrueBeam 6 MV and 10 MV flattening filter-free stereotactic radiosurgery system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 3, 2016 Output factor comparison of Monte Carlo and measurement for Varian TrueBeam 6 MV and 10 MV flattening filter-free stereotactic radiosurgery
More informationDevelopment a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code
Iran. J. Radiat. Res., 2006; 4 (1): 7-14 Development a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code. Mesbahi * Department of Medical Physics, Medical School,
More informationMeasurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code
reports of practical oncology and radiotherapy 1 5 (2 0 1 0) 64 68 available at www.sciencedirect.com journal homepage: http://www.rpor.eu/ Original article Measurement of depth-dose of linear accelerator
More informationIMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays
IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays Sotiri Stathakis Outline Why IMRT/VMAT QA AAPM TG218 UPDATE Tolerance Limits and Methodologies for IMRT Verification QA Common sources
More informationTPS COMMISSIONING. Laurence Court University of Texas MD Anderson Cancer Center 4/3/2017 1
TPS COMMISSIONING Laurence Court University of Texas MD Anderson Cancer Center 4/3/2017 1 Conflicts of interest Court receives funding from NIH, CPIRT, Varian and Elekta 4/3/2017 2 Resources your first
More informationImprovements in Monte Carlo simulation of large electron fields
SLAC-PUB-12908 October 2007 Improvements in Monte Carlo simulation of large electron fields Bruce A Faddegon 1, Joseph Perl 2 and Makoto Asai 2 1 University of California San Francisco Comprehensive Cancer
More informationMegan A. Wood, M.S. Under the direction of Larry DeWerd, Ph.D. University of Wisconsin Medical Radiation Research Center (UWMRRC)
Megan A. Wood, M.S. Under the direction of Larry DeWerd, Ph.D. University of Wisconsin Medical Radiation Research Center (UWMRRC) NCCAAPM Spring Meeting May 3, 2013 Introduction FFF background Project
More informationInvestigation of photon beam output factors for conformal radiation therapy Monte Carlo simulations and measurements
INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N133 N143 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)32395-9 NOTE Investigation of photon beam output factors for conformal radiation therapy
More informationVolumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT
Volumetric Modulated Arc Therapy - Clinical Implementation Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA Acknowledgement David M. Shepard, Ph.D. Muhammad K. N. Afghan, Ph.D. Fan Chen, Ph.D.
More informationMonte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters
Monte Carlo Simulation for Neptun 1 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters M.T. Bahreyni Toossi a, M. Momen Nezhad b, S.M. Hashemi a a Medical Physics Research Center,
More informationNEW METHOD OF COLLECTING OUTPUT FACTORS FOR COMMISSIONING LINEAR ACCELERATORS WITH SPECIAL EMPHASIS
NEW METHOD OF COLLECTING OUTPUT FACTORS FOR COMMISSIONING LINEAR ACCELERATORS WITH SPECIAL EMPHASIS ON SMALL FIELDS AND INTENSITY MODULATED RADIATION THERAPY by Cindy D. Smith A Thesis Submitted to the
More informationQalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004
Qalitätssicherung an Multileafkollimatoren Dr. Lutz Müller Würzburg jan 2004 IMRT Verification - present Target Volume Constraints Inverse Planning Algorithm Fluence Map Leaf Sequencer Leaf & Gantry sequence
More informationQuick Reference Datasheet For All RIT113 Packages
Quick Reference Datasheet For All RIT113 Packages For Rotational Therapies, IMRT & TG142 Highlights and selected product information only. A complete TG142 brochure is available. For more information on
More informationCBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona
CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona Introduction CBCT has become a routine procedure for
More informationCommissioning and verification of the collapsed cone convolution superposition algorithm for SBRT delivery using flattening filter-free beams
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 2, 2014 Commissioning and verification of the collapsed cone convolution superposition algorithm for SBRT delivery using flattening filter-free
More informationIndrin Chetty Henry Ford Hospital Detroit, MI. AAPM Annual Meeting Houston 7:30-8:25 Mon 08/07/28 1/30
Review of TG105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics,
More informationGoals. Scope/exclusions. Outline: follow outline of MPPG (plus rationale & some implementation experiences)
AAPM MEDICAL PHYSICS PRACTICE GUIDELINE # : Commissioning and QA of Treatment Planning Dose Calculations: Megavoltage Photon and Electron Beams Jennifer Smilowitz (Chair), University of Wisconsin-Madison
More informationMCNP4C3-BASED SIMULATION OF A MEDICAL LINEAR ACCELERATOR
Computational Medical Physics Working Group Workshop II, Sep 3 Oct 3, 7 University of Florida (UF), Gainesville, Florida USA on CD-ROM, American Nuclear Society, LaGrange Park, IL (7) MCNP4C3-BASED SIMULATION
More informationA secondary monitor unit calculation algorithm using superposition of symmetric, open fields for IMRT plans
Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2013 A secondary monitor unit calculation algorithm using superposition of symmetric, open fields for IMRT plans Adam
More informationUse of a commercial ion chamber detector array for the measurement of high spatial resolution photon beam profiles
Received: 17 January 2018 Revised: 27 August 2018 Accepted: 31 August 2018 DOI: 10.1002/acm2.12466 TECHNICAL NOTE Use of a commercial ion chamber detector array for the measurement of high spatial resolution
More informationUse of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta
Use of Monte Carlo modelling in radiotherapy linac design David Roberts, PhD Senior Physicist Elekta Contents Overview of Elekta What we do Where we use Monte Carlo Codes and resources Example : Agility
More informationClinical implementation of photon beam flatness measurements to verify beam quality
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Clinical implementation of photon beam flatness measurements to verify beam quality Simon Goodall, a Nicholas Harding, Jake Simpson,
More informationRaising the Bar in IMRT QA
MapCHECK 2TM Raising the Bar in IMRT QA The leader in quick and precise measurement of modulated radiotherapy beams Benefits Proven solution for film-less rotational delivery and IMRT QA - More than 1500
More informationAssesing multileaf collimator effect on the build-up region using Monte Carlo method
Pol J Med Phys Eng. 2008;14(3):163-182. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0014-0 website: http://www.pjmpe.waw.pl M. Zarza Moreno 1, 2, N. Teixeira 3, 4, A. P. Jesus 1, 2, G. Mora 1 Assesing multileaf
More informationDynalog data tool for IMRT plan verification
Dynalog data tool for IMRT plan verification Poster No.: R-0051 Congress: 2014 CSM Type: Scientific Exhibit Authors: V. Sashin; FOOTSCRAY/AU Keywords: Computer applications, Radiation physics, Experimental,
More informationAcknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF
Page 1 Quality Assurance of IMRT Delivery Systems - Siemens Lynn J. Verhey, Ph.D. Professor and Vice-Chair UCSF Dept. of Radiation Oncology AAPM 22 Annual Meeting, Montreal Acknowledgments Ping Xia, Ph.D.,
More informationAn Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields
Int. J. Cancer (Radiat. Oncol. Invest): 96, 140 145 (2001) 2001 Wiley-Liss, Inc. Publication of the International Union Against Cancer An Investigation of a Model of Percentage Depth Dose for Irregularly
More informationMonitor Unit (MU) Calculation
Monitor Unit (MU) Calculation Timothy C. Zhu 1, Haibo Lin 1, and JiaJian Shen 2 1 University of Pennsylvania, Philadelphia, PA 2 Mayo Clinic, Phoenix, AZ Introduction Pencil-beam based dose/mu algorithms
More informationBasic Radiation Oncology Physics
Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides
More informationImplementation of a double Gaussian source model for the BEAMnrc Monte Carlo code and its influence on small fields dose distributions
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Implementation of a double Gaussian source model for the BEAMnrc Monte Carlo code and its influence on small fields dose distributions
More informationMapCHECK 2 & 3DVH The Gold Standard for 2D Arrays
MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.
More informationProton dose calculation algorithms and configuration data
Proton dose calculation algorithms and configuration data Barbara Schaffner PTCOG 46 Educational workshop in Wanjie, 20. May 2007 VARIAN Medical Systems Agenda Broad beam algorithms Concept of pencil beam
More information3D Scanning Comes Full Circle
3D SCANNER TM 3D Scanning Comes Full Circle Relative 3D Dosimetry offering the easiest setup, most objectivity, and best consistency available Cross-Plane Advanced Design Ring and diameter drive mechanism
More informationMonte Carlo methods in proton beam radiation therapy. Harald Paganetti
Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization
More informationMapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays
MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.
More informationCalibration and quality assurance for rounded leaf-end MLC systems
Calibration and quality assurance for rounded leaf-end MLC systems Maria N. Graves, a) Antoinette V. Thompson, b) Mary K. Martel, Daniel L. McShan, and Benedick A. Fraass Department of Radiation Oncology,
More informationDetermination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation
Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation Danial Seifi Makrani 1, Hadi Hasanzadeh 2*, Tayyeb Allahverdi Pourfallah 3, Arash Ghasemi 4, Majid
More informationAcceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems. Bruce Curran University of Michigan Medical Center Ann Arbor, MI
Acceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems Bruce Curran University of Michigan Medical Center Ann Arbor, MI Disclosure Some of the work discussed in this talk was supported
More informationCLINICAL ASPECTS OF COMPACT GANTRY DESIGNS
CLINICAL ASPECTS OF COMPACT GANTRY DESIGNS J. Heese, J. Wulff, A. Winnebeck, A. Huggins, M. Schillo VARIAN PARTICLE THERAPY JUERGEN HEESE New gantry developments Viewpoint from user and vendor perspective
More informationValidation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition
Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.
More informationTomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS
Tomotherapy Physics Machine Twinning and Quality Assurance Emilie Soisson, MS Tomotherapy at UW- Madison Treating for nearly 5 years Up to ~45 patients a day on 2 tomo units Units twinned to facilitate
More informationCNMC 865 Easthagan Drive, Nashville, TN USA phone
CNMC 865 Easthagan Drive, Nashville, TN 37217 USA BMI.v.04.2010 Electrometers CNMC Model 206 The CNMC Model 206 electrometer maintains the simplicity and ease of use of classic electrometers. Designed
More informationTransitioning from pencil beam to Monte Carlo for electron dose calculations
Transitioning from pencil beam to Monte Carlo for electron dose calculations Jessie Huang-Vredevoogd (jyhuang4@wisc.edu) University of Wisconsin NCC AAPM October 12, 2019 1 Topics to cover Background RayStation
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More informationPhoton beam dose distributions in 2D
Photon beam dose distributions in 2D Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2014 Acknowledgments! Narayan Sahoo PhD! Richard G Lane (Late) PhD 1 Overview! Evaluation
More informationComparison of absorbed dose distribution 10 MV photon beam on water phantom using Monte Carlo method and Analytical Anisotropic Algorithm
Journal of Physics: Conference Series PAPER OPEN ACCESS Comparison of absorbed dose distribution 1 MV photon beam on water phantom using Monte Carlo method and Analytical Anisotropic Algorithm To cite
More informationVerification of Varian enhanced dynamic wedge implementation in MasterPlan treatment planning system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 2, SPRING 2009 Verification of Varian enhanced dynamic wedge implementation in MasterPlan treatment planning system Massimo Pasquino, 1,a
More informationDAILY LINAC QA BEAM QA
BEAM QA DAILY LINAC QA The QA BeamChecker Plus allows for fast, reliable, and uncomplicated daily QA of Varian, Elekta, Siemens, and Accuray Treatment Machines. The QA BeamChecker Plus is specifically
More informationMonte Carlo Methods for Accelerator Simulation and Photon Beam Modeling
Monte Carlo Methods for Accelerator Simulation and Photon Beam Modeling AAPM Summer School 2006 Windsor, ON Part I Daryoush Sheikh-Bagheri, PhD Allegheny General Hospital Pittsburgh, PA A sample of work
More informationAn investigation into the use of MMCTP to tune accelerator source parameters and testing its clinical application
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 An investigation into the use of MMCTP to tune accelerator source parameters and testing its clinical application Elaine Conneely,
More informationI. INTRODUCTION. Figure 1. Radiation room model at Dongnai General Hospital
International Journal of Computational Engineering Research Vol, 04 Issue, 4 Simulation of Photon and Electron dose distributions 5 code for the treatment area using the linear electron accelerator (LINAC)
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Measurement and Monte Carlo simulation for energy- and intensity-modulated electron radiotherapy delivered by a computer-controlled
More informationNew Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC
New Technology in Radiation Oncology James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC Technology s s everywhere From the imaging chain To the planning system To the linac To QA..it..it
More information3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending
3DVH : On The Accuracy Of The Planned Dose Perturbation Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools *Patent Pending introduction State-of-the-art IMRT QA of static gantry
More information87.55.K-, Gh, km, Ca. Agility, analytic model, Elekta, Monaco, Monte Carlo, virtual source
Received: 14 May 2018 Revised: 21 September 2018 Accepted: 27 September 2018 DOI: 10.1002/acm2.12485 RADIATION ONCOLOGY PHYSICS Monte Carlo and analytic modeling of an Elekta Infinity linac with Agility
More informationVALIDATION OF IN-HOUSE DOSE CALCULATION SOFTWARE FOR SUPERFICIAL X-RAY THERAPY. A Thesis. Presented to the. Faculty of. San Diego State University
VALIDATION OF IN-HOUSE DOSE CALCULATION SOFTWARE FOR SUPERFICIAL X-RAY THERAPY A Thesis Presented to the Faculty of San Diego State University In Partial Fulfillment of the Requirements for the Degree
More informationVerification measurements of an emc algorithm using a 2D ion chamber array
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Verification measurements of an emc algorithm using a 2D ion chamber array Mark D. Wanklyn, 1a Ghirmay Kidane, 2 and Liz Crees 2 Medical
More informationVerification of dose calculations with a clinical treatment planning system based on a point kernel dose engine
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 Verification of dose calculations with a clinical treatment planning system based on a point kernel dose engine Lars Weber*
More informationRoom scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation study
Received: 9 November 2016 Revised: 9 November 2016 Accepted: 28 November 2016 DOI: 10.1002/acm2.12039 RADIATION ONCOLOGY PHYSICS Room scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation
More informationDelta 4 PT. True Volumetric Pre-treatment Verification. The difference is clear
Delta 4 PT True Volumetric Pre-treatment Verification The difference is clear ScandiDos innovators in advanced volumetric dosimetry solutions ScandiDos ScandiDos is the innovative company that introduced
More informationA virtual photon energy fluence model for Monte Carlo dose calculation
A virtual photon energy fluence model for Monte Carlo dose calculation Matthias Fippel, a) Freddy Haryanto, Oliver Dohm, and Fridtjof Nüsslin Abteilung für Medizinische Physik, Universitätsklinikum Tübingen,
More informationTHE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!
THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER! Confidence in complex treatments Modern radiation therapy uses complex plans with techniques such as IMRT, VMAT and Tomotherapy.
More informationPROVEN TECHNOLOGY IMPROVED
PHANTOMS PROVEN TECHNOLOGY IMPROVED The DoseView 3D is a new 3-axis water phantom for performing comprehensive linear accelerator beam commissioning and quality assurance measurements quickly, accurately
More informationAdvanced Radiotherapy
Advanced Radiotherapy IBA Concept for IMRT Verification Dr. Lutz Müller 1 IMRT Verification BIS BIS ADAS Fluence Model Delivered fluence Target Volume Constraints Inverse Backprojection Fluence Map Leaf
More informationMonte Carlo Treatment Planning: Implementation of Clinical Systems
Monte Carlo Treatment Planning: Implementation of Clinical Systems Richard Popple 1 and Joanna E. Cygler 2,3,4 1 Department of Radiation Oncology, The University of Alabama at Birmingham 2 The Ottawa Hospital
More informationMonte Carlo modeling of a Novalis TX Varian 6 MV with HD-120 multileaf collimator
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Monte Carlo modeling of a Novalis TX Varian 6 MV with HD-120 multileaf collimator Luis Alberto Vazquez-Quino, 1 Brian Massingill,
More informationComparison Study of Monte Carlo Simulations and Measurements of Relative Output Factors of 6 MV Photon Beam
Comparison Study of Monte Carlo Simulations and Measurements of Relative Output Factors of 6 MV Photon Beam Nakorn Phaisangittisakul 1*, Wanchaloem Saprangsi 2, Chirapha Tannanonta 2 and Sivalee Suriyapee
More informationFacility Questionnaire PART I (General Information for 3DCRT and IMRT)
Facility Questionnaire PART I (General Information for 3DCRT and IMRT) The following items are required before you can enter cases on any RTOG protocol that requires data submission to the Image-Guided
More informationDose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them
Dose Distributions George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Purpose To familiarize the resident with dose distributions and the factors that affect them
More informationMedical Dosimetry 37 (2012) Medical Dosimetry. journal homepage:
Medical Dosimetry 37 (2012) 53-60 Medical Dosimetry journal homepage: www.meddos.org Characterization of responses of 2d array seven29 detector and its combined use with octavius phantom for the patient-specific
More informationVMAT for dummies: Concepts, Clinical Implementation and Treatment Planning. Rajat Kudchadker, Ph.D. Associate Professor
VMAT for dummies: Concepts, Clinical Implementation and Treatment Planning Rajat Kudchadker, Ph.D Associate Professor 2012 AAPM Spring Clinical Meeting Disclaimer I hold no financial interest nor have
More informationEXTERNAL PHOTON BEAMS: PHYSICAL ASPECTS
EXTERNAL PHOTON BEAMS: PHYSICAL ASPECTS E.B. PODGORSAK Department of Medical Physics, McGill University Health Centre, Montreal, Quebec, Canada 6.1. INTRODUCTION Radiotherapy procedures fall into two main
More informationQuality assurance of a helical tomotherapy machine
INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 49 (2004) 2933 2953 PHYSICS IN MEDICINE AND BIOLOGY PII: S0031-9155(04)71892-8 Quality assurance of a helical tomotherapy machine J D Fenwick 1,2,WATomé
More informationTable of Contents. About AQUA
User Manual About AQUA Table of Contents 1 About AQUA... 6 1.1 Radiation Therapy... 6 1.1.1 Workflow manager... 6 1.1.2 Reports... 7 1.1.3 Tests... 7 1.1.4 Many Ways to Work... 7 1.2 Technology... 7 1.3
More informationph fax Deming Way Middleton WI USA
www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 3120 Deming Way Middleton WI 53562-1461 USA 2007 Standard Imaging, Inc. 1239-21 (03.07) beam qa LINAC AND TOMOTHERAPY QA The NEW
More informationEvaluation of 3D Gamma index calculation implemented in two commercial dosimetry systems
University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2015 Evaluation of 3D Gamma index calculation implemented
More informationValidation of GEANT4 Monte Carlo Simulation Code for 6 MV Varian Linac Photon Beam
Validation of GEANT4 Monte Carlo Code for 6 MV Varian Linac Photon Beam E. Salama ab*, A.S. Ali c, N. Emad d and A. Radi a a Physics Department, Faculty of Science, Ain Shams University, Cairo, Egypt;
More informationFeasibility study of performing IGRT system daily QA using a commercial QA device
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 3, summer 2011 Feasibility study of performing IGRT system daily QA using a commercial QA device Jean L.Peng, 1 Darren Kahler, 2 Jonathan
More informationThe IORT Treatment Planning System. radiance. GMV, 2012 Property of GMV All rights reserved
The IORT Treatment Planning System radiance Property of GMV All rights reserved WHY RADIANCE? JUSTIFICATION Property of GMV All rights reserved ADVANTAGES OF IORT PRECISION: RT guided by direct vision.
More informationA method for determining multileaf collimator transmission and scatter for dynamic intensity modulated radiotherapy a
A method for determining multileaf collimator transmission and scatter for dynamic intensity modulated radiotherapy a Mark R. Arnfield, b) Jeffrey V. Siebers, Jong O. Kim, Qiuwen Wu, Paul J. Keall, and
More informationUNIVERSITY OF CALGARY. Monte Carlo model of the Brainlab Novalis Classic 6 MV linac using the GATE simulation. platform. Jared K.
UNIVERSITY OF CALGARY Monte Carlo model of the Brainlab Novalis Classic 6 MV linac using the GATE simulation platform by Jared K. Wiebe A THESIS SUBMITTED TO THE FACULTY OF GRADUATE STUDIES IN PARTIAL
More informationTHE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 THE SIMULATION OF
More informationCalculating percent depth dose with the electron pencilbeam redefinition algorithm
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 8, NUMBER 2, SPRING 2007 Calculating percent depth dose with the electron pencilbeam redefinition algorithm Michael J. Price, 1,2,a Kenneth R. Hogstrom,
More informationRelease Notes for Dosimetry Check with Convolution-Superposition Collapsed Cone Algorithm (CC)
Dosimetry Check for convolution/superposition collapsed cone (CC), Page 1 of 11 Release Notes for Dosimetry Check with Convolution-Superposition Collapsed Cone Algorithm (CC) File: CCReleaseNotes.doc Date:
More informationShielding factors for traditional safety glasses
Shielding factors for traditional safety glasses Malcolm McEwen, Hong Shen and Ernesto Mainegra-Hing Ionizing Radiation Standards, National Research Council Canada Alan DuSautoy, Radiation and Health Sciences
More informationChapter 9 Field Shaping: Scanning Beam
Chapter 9 Field Shaping: Scanning Beam X. Ronald Zhu, Ph.D. Department of Radiation Physics M. D. Anderson Cancer Center Houston, TX June 14-18, 2015 AAPM - Summer School 2015, Colorado Spring Acknowledgement
More informationOn compensator design for photon beam intensity-modulated conformal therapy
On compensator design for photon beam intensity-modulated conformal therapy Steve B. Jiang a) and Komanduri M. Ayyangar Department of Radiation Therapy, Medical College of Ohio, 3000 Arlington Avenue,
More informationImplementation of Monte Carlo Dose calculation for CyberKnife treatment planning
Journal of Physics: Conference Series Implementation of Monte Carlo Dose calculation for CyberKnife treatment planning To cite this article: C-M Ma et al 2008 J. Phys.: Conf. Ser. 102 012016 View the article
More informationImplementation of the EGSnrc / BEAMnrc Monte Carlo code - Application to medical accelerator SATURNE43
International Journal of Innovation and Applied Studies ISSN 2028-9324 Vol. 6 No. 3 July 2014, pp. 635-641 2014 Innovative Space of Scientific Research Journals http://www.ijias.issr-journals.org/ Implementation
More information