1 DICOM Supplement 188 Multi-Energy CT Imaging DICOM Working Group 21 Computed Tomography

2 Rationale Short introduction of Multi Energy (ME) s Overview: Imaging techniques, including scanning, reconstruction, processing, when the scanner utilizes multiple energies from the X-Ray beam spectrum, as opposed to the conventional CT imaging, when a single (accumulated) X-Ray spectrum is used. The existing CT and Enhanced CT (ect) IODs do not adequately describe the new CT multienergy imaging. Although different vendors apply different scanning and detection techniques to achieve multi-energy images, there is large commonality in the generated diagnostic images. Goals for ME implementation: provide new essential ME information (acquisition, reconstruction and processing attributes) within the IOD. facilitate fast and easy adoption of standard based ME imaging across the imaging community, both modalities and PACS/Displays. address (or at least to minimize) the risk of mis-interpretation when the ME images are displayed by a display does not support the new attributes of the ME-image, including incorrect measurements adapt existing attributes of the CT / Enhanced CT IOD to fit ME techniques. 2

3 Overview Multi Energy Imaging Standard CT Objective Quantification Labeling Visualization Virtual Monoenergetic (VMI) Effective Atomic Number (Z) Electron Density - Specific Iodine Map; Bone Density - Removed Virt. Non- Contrast; Virt. Non-Ca; Proportional Map Discrete Labeling Probability Map Value based Map Gout crystals CT IOD - Modified Highlighted; Partially- Suppressed Other IOD Color Overlay Color Blending Color Map

4 Organization Structure Multi-energy CT Macro (new added to CT IODs) Multi-energy Flag (Y/N) Multi-energy CT Acquisition Sequence (1C, 1 item) Multi-energy CT Acquisition Sequence (1-N) Multi-Energy Source Technique Multi-Energy Detection Technique Other ME-specific attributes Common CT Acquisition Macro Multi-energy CT Processing Sequence (1C, 1 item) Decomposition Method Decomposition Basis Sequence (2-N items, one for each basis) Other decomposition attributes Multi-energy CT Characteristics Sequence (1C) Monochromatic Energy Equivalent (for Virtual Monochromatic ) Multi-energy Quantification CT Macro Specific Code Sequence Modification Sequence Multi-energy Labeling CT Sequence (1 item) Labeling Type Modification Sequence 4

5 Multi-Energy CT Acquisition Techniques X-Ray Source Scanned Object Detector Generating different X-Ray Energies: Discriminating Energies: KVP Filter Other Parameters Multiple Layers Photon Counting Methods to separate at least two energies include Multiple Scans of the same area with diff parameters Multiple X-Ray Sources and/or Detectors Switch KVP during the rotation One source with Multi-Layer Detector One source with Photon Counting Detector

6 Objective s Data Acquisition Decomposition to Base Components Generation of Diagnostic images A1 A2 An M1 M2 Mn Virtual Monochromatic Effective Atomic Number Electron Density Acquisition Sequence Processing Sequence Characteristics Sequence (currently just kev for VMI)

7 s Data Acquisition Decomposition to and/or Classification of two or more s Generation of Diagnostic images A1 A2 An M1 M2 Mn Quantification Labeling Visualization Acquisition Sequence Processing Sequence Characteristics Sequence

8 s Generation M1 M2 Mn May be ignored / Not described -Specific Proportional Map Labeling s Examples:: - Iodine Map - Bone (Ca) Density Map Conventional CT or VMI o o o o o Subtract Suppress Highlight Compensate Recalculate M1 M2 Mn - Removed Visualization s Examples:: - Virtual Non-Contrast : COMPENSATED - Virtual Non-Calcium : COMPENSATED - Bone or Uric Acid Removed : SUBTRACTED (0) Examples:: - Tendon Enhancement: HIGHLIGHTED

9 Labeling s Discrete Labeling (most-probable material): Proportional: A BC 2 0: None 0.3 D 1: A 2: B contains 30% of material A 3: C Value based: Probability: A = 0-15 B = % chance containing material A 0.2 A BC D 15 0 C = 18-50

10 Visualization s Color Overlay A B Overlay Blended C Structural A) A CT image that is windowed to highlight a particular material with a color map applied. It also may have a translucency applied to be able to see the image underneath. (E.g. Iodine image, Effective Z image) B) A structural image showing the anatomical structure. (E.g. Monochromatic image) C) The result image (combined information e.g. Secondary Capture, Blending, )

Digital Imaging and Communications in Medicine (DICOM) Supplement 188: Multi-energy CT Images

Digital Imaging and Communications in Medicine (DICOM) Supplement 188: Multi-energy CT Images Supplement 188: Multi-energy CT Images Page 1 2 4 6 Digital Imaging and Communications in Medicine (DICOM) 8 Supplement 188: Multi-energy CT Images 10 12 14 16 18 20 Prepared by: 22 DICOM Standards Committee,

More information

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

Cardiac Dual Energy CT: Technique

Cardiac Dual Energy CT: Technique RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de

More information

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Joonki Noh, Jeffrey A. Fessler EECS Department, The University of Michigan Paul E. Kinahan Radiology Department,

More information

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland Metal Artifact Reduction CT Techniques R S S S Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland N. 1 v o 4 1 0 2. Postoperative CT Metal Implants CT is accurate for assessment

More information

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Stefan C. Saur 1, Hatem Alkadhi 2, Luca Regazzoni 1, Simon Eugster 1, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision

More information

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology Diagnostic imaging techniques Krasznai Zoltán University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology 1. Computer tomography (CT) 2. Gamma camera 3. Single Photon

More information

Enhanced material contrast by dual-energy microct imaging

Enhanced material contrast by dual-energy microct imaging Enhanced material contrast by dual-energy microct imaging Method note Page 1 of 12 2 Method note: Dual-energy microct analysis 1. Introduction 1.1. The basis for dual energy imaging Micro-computed tomography

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

Technical Publications

Technical Publications GE Medical Systems Technical Publications Direction 2188003-100 Revision 0 Tissue Volume Analysis DICOM for DICOM V3.0 Copyright 1997 By General Electric Co. Do not duplicate REVISION HISTORY REV DATE

More information

Medical Imaging Introduction

Medical Imaging Introduction Medical Imaging Introduction Jan Kybic February 16, 2010 Medical imaging: a collaborative paradigm picture from Atam P. Dhawan: Medical Imaging From physiology to information processing (what we should

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

DICOM Correction Item

DICOM Correction Item Specify contents for DICOMDIR records DICOM Correction Item Correction Number CP 688 Log Summary: Specify contents for DICOMDIR records Type of Modification Modification Rationale for Correction Name of

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

Digital Imaging and Communications in Medicine (DICOM) Supplement 167: X-Ray 3D Angiographic IOD Informative Annex

Digital Imaging and Communications in Medicine (DICOM) Supplement 167: X-Ray 3D Angiographic IOD Informative Annex 5 Digital Imaging and Communications in Medicine (DICOM) Supplement 167: X-Ray 3D Angiographic IOD Informative Annex 10 15 Prepared by: DICOM Standards Committee, Working Group 2, Projection Radiography

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging

More information

Digital Imaging and Communications in Medicine (DICOM) Supplement 116: 3D X-Ray Storage SOP Class

Digital Imaging and Communications in Medicine (DICOM) Supplement 116: 3D X-Ray Storage SOP Class 2 4 6 Digital Imaging and Communications in Medicine (DICOM) 8 Supplement 116: 3D X-Ray Storage SOP Class 10 12 14 16 18 20 22 Prepared by: 24 DICOM Standards Committee, Working Groups 02, Projection Imaging

More information

Optimization of scanner parameters for dual energy micro-ct

Optimization of scanner parameters for dual energy micro-ct Optimization of scanner parameters for dual energy micro-ct E. PAUWELS* 1, J. DHAENE 1, A. DE MUYNCK 1 E., M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics and Astronomy, Ghent University, Proeftuinstraat

More information

Technical Publications

Technical Publications g GE Medical Systems Technical Publications Direction 2264272-100 Revision 1 DICOM for DICOM V3.0 Copyright ª 2000 By General Electric Co. Do not duplicate THIS PAGE LEFT INTENTIONALLY BLANK TABLE OF

More information

Physical bases of X-ray diagnostics

Physical bases of X-ray diagnostics Physical bases of X-ray diagnostics Dr. István Voszka Possibilities of X-ray production (X-ray is produced, when charged particles of high velocity are stopped) X-ray tube: Relatively low accelerating

More information

Moscow-Bavarian Joint Advanced Student School 2006 / Medical Imaging Principles of Computerized Tomographic Imaging and Cone-Beam Reconstruction

Moscow-Bavarian Joint Advanced Student School 2006 / Medical Imaging Principles of Computerized Tomographic Imaging and Cone-Beam Reconstruction Line Integrals Line integrals represent the integral of some parameter of the object along the line (e.g. attenuation of x-rays) Object: f(x,y) Line: x cosθ + y sinθ = t Line integral / Radon transform:

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

ML reconstruction for CT

ML reconstruction for CT ML reconstruction for CT derivation of MLTR rigid motion correction resolution modeling polychromatic ML model dual energy ML model Bruno De Man, Katrien Van Slambrouck, Maarten Depypere, Frederik Maes,

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT Assessment of 3D performance metrics D and 3D Metrics of Performance Towards Quality Index: Volumetric imaging systems X-ray based Volumetric imaging systems: CBCT/CT Tomosynthesis Samuel Richard and Ehsan

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

DICOM Conformance Statement. CT AV Series Product Line. Document Number February 1997

DICOM Conformance Statement. CT AV Series Product Line. Document Number February 1997 Philips Medical Systems DICOM Conformance Statement CT AV Series Product Line Document Number 4522 983 65911 28 February 1997 Copyright Philips Medical Systems Nederland B.V. 1997 Philips Medical Systems

More information

Medical Image Processing: Image Reconstruction and 3D Renderings

Medical Image Processing: Image Reconstruction and 3D Renderings Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,

More information

Technical Publications

Technical Publications g GE Medical Systems Technical Publications Direction 2275362-100 Revision 0 DICOM for DICOM V3.0 Copyright 2000 By General Electric Co. Do not duplicate REVISION HISTORY REV DATE REASON FOR CHANGE 0 May

More information

Future Topics. Projection Imaging Dose Reporting, XA 3D Volume Objects. for DICOM WG-02. Presented by Heinz Blendinger, Siemens Medical Solutions

Future Topics. Projection Imaging Dose Reporting, XA 3D Volume Objects. for DICOM WG-02. Presented by Heinz Blendinger, Siemens Medical Solutions Future Topics for Projection Imaging Dose Reporting, XA 3D Volume Objects DICOM WG-02 Presented by Heinz Blendinger, Siemens Medical Solutions 1 Presentation outline Dose Reporting Why Dose Reporting?

More information

Micro-CT Methodology Hasan Alsaid, PhD

Micro-CT Methodology Hasan Alsaid, PhD Micro-CT Methodology Hasan Alsaid, PhD Preclinical & Translational Imaging LAS, PTS, GlaxoSmithKline 20 April 2015 Provide basic understanding of technical aspects of the micro-ct Statement: All procedures

More information

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings)

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings) Marquette University e-publications@marquette Biomedical Engineering Faculty Research and Publications Engineering, College of 1-1-1999 An Iterative Approach to the Beam Hardening Correction in Cone Beam

More information

TEP Hounsfield units. Related topics Attenuation coefficient, Hounsfield units

TEP Hounsfield units. Related topics Attenuation coefficient, Hounsfield units Hounsfield units TEP Related topics Attenuation coefficient, Hounsfield units Principle Depending on the type of CT scanner and the settings, the result of a CT scan of the same material can be different

More information

Digital Imaging and Communications in Medicine (DICOM) Supplement 121: CT Protocol Storage

Digital Imaging and Communications in Medicine (DICOM) Supplement 121: CT Protocol Storage DICOM Supplement 121 : CT Protocol Storage Page 1 2 4 6 Digital Imaging and Communications in Medicine (DICOM) 8 Supplement 121: CT Protocol Storage 10 12 14 16 18 20 Prepared by: 22 DICOM Standards Committee,

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Medical Images Analysis and Processing

Medical Images Analysis and Processing Medical Images Analysis and Processing - 25642 Emad Course Introduction Course Information: Type: Graduated Credits: 3 Prerequisites: Digital Image Processing Course Introduction Reference(s): Insight

More information

DUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography

DUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography Marc Kachelrieß, Member, IEEE, Timo Berkus, Philip Stenner, Willi A. Kalender Abstract Material selective imaging using dual energy

More information

Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application

Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application Physics in Medicine & Biology Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application To cite this article: Polad M Shikhaliev and Shannon G Fritz 2011

More information

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Qiao Yang 1,4, Meng Wu 2, Andreas Maier 1,3,4, Joachim Hornegger 1,3,4, Rebecca Fahrig

More information

Technical Publications

Technical Publications g GE Healthcare Technical Publications 2391078-2-800 Revision 1 Seno Advantage2 WorkStation for DICOM V3.0 do not duplicate Copyright 2006 By General Electric Co. THIS PAGE LEFT INTENTIONALLY BLANK REVISION

More information

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct J. DHAENE* 1, E. PAUWELS 1, T. DE SCHRYVER 1, A. DE MUYNCK 1, M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics

More information

DICOM Correction Item

DICOM Correction Item DICOM Correction Item Correction Number CP-776 Log Summary: Type of Modification Correction Name of Standard PS 3.3 2007 Rationale for Correction For some time it has been the convention not to make attributes

More information

Simulation of Beam Hardening in Industrial CT with x-ray and Monoenergetic Source by Monte Carlo Code

Simulation of Beam Hardening in Industrial CT with x-ray and Monoenergetic Source by Monte Carlo Code 2012, TextRoad Publication ISSN 2090-4304 Journal of Basic and Applied Scientific Research www.textroad.com Simulation of Beam Hardening in Industrial CT with x-ray and Monoenergetic Source by Monte Carlo

More information

Financial disclosure. Onboard imaging modality for IGRT

Financial disclosure. Onboard imaging modality for IGRT Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty

More information

Photon counting spectroscopic CT with dynamic beam attenuator

Photon counting spectroscopic CT with dynamic beam attenuator Photon counting spectroscopic CT with dynamic beam attenuator Haluk Atak 1) and Polad M. Shikhaliev a) 1) Department of Nuclear Engineering, Hacettepe University, Ankara, Turkey Abstract Purpose: Photon

More information

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION 2009 IEEE Nuclear Science Symposium Conference Record M03-7 FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION Wonseok Huh, Jeffrey A. Fessler, Adam M. Alessio, and Paul E. Kinahan Department

More information

State of the Standards Kevin O Donnell, MASc. Canon Medical Research

State of the Standards Kevin O Donnell, MASc. Canon Medical Research State of the Standards 2018 Kevin O Donnell, MASc. Canon Medical Research If you build it, they will come Yeah, right. Draft & Publish Standard SDOs Implement & Release in Product Vendors Install & Use

More information

Emission Computed Tomography Notes

Emission Computed Tomography Notes Noll (24) ECT Notes: Page 1 Emission Computed Tomography Notes Introduction Emission computed tomography (ECT) is the CT applied to nuclear medicine. There are two varieties of ECT: 1. SPECT single-photon

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Technical Publications

Technical Publications g GE Medical Systems Technical Publications 2391078-800 Revision 1 Seno Advantage WorkStation for DICOM V3.0 do not duplicate Copyright 2003 By General Electric Co. THIS PAGE LEFT INTENTIONALLY BLANK REVISION

More information

Index. aliasing artifacts and noise in CT images, 200 measurement of projection data, nondiffracting

Index. aliasing artifacts and noise in CT images, 200 measurement of projection data, nondiffracting Index Algebraic equations solution by Kaczmarz method, 278 Algebraic reconstruction techniques, 283-84 sequential, 289, 293 simultaneous, 285-92 Algebraic techniques reconstruction algorithms, 275-96 Algorithms

More information

VJ Technologies Inspection Services Division. By Bob Maziuk, February 2015

VJ Technologies Inspection Services Division. By Bob Maziuk, February 2015 White Paper on the Inspection of Underground Electrical Conductors, Insulation, and Splices in Oil-Filled Piping Using High Energy Computed Tomography (CT) VJ Technologies Inspection Services Division

More information

Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage. 29 April - 3 May, 2013

Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage. 29 April - 3 May, 2013 2455-5 Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage 29 April - 3 May, 2013 Lecture NoteBasic principles of X-ray Computed Tomography Diego Dreossi Elettra, Trieste

More information

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic

More information

Tomographic Reconstruction

Tomographic Reconstruction Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)

More information

DICOM Correction Proposal Form

DICOM Correction Proposal Form DICOM Correction Proposal Form Tracking Information - Administration Use Only Correction Proposal Number STATUS CP-1418 Assigned Date of Last Update 2016.03.17 Person Assigned Submitter Name Kevin O Donnell

More information

Superiorized polyenergetic reconstruction algorithm for reduction of metal artifacts in CT images

Superiorized polyenergetic reconstruction algorithm for reduction of metal artifacts in CT images 1 Superiorized polyenergetic reconstruction algorithm for reduction of metal artifacts in CT images T. Humphries 1 and A. Gibali 2 Abstract Artifacts caused by metal objects such as dental fillings, hip

More information

Reconstruction in CT and relation to other imaging modalities

Reconstruction in CT and relation to other imaging modalities Reconstruction in CT and relation to other imaging modalities Jørgen Arendt Jensen November 1, 2017 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering Center for Fast Ultrasound

More information

Computer-Tomography I: Principles, History, Technology

Computer-Tomography I: Principles, History, Technology Computer-Tomography I: Principles, History, Technology Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de History

More information

Quality control phantoms and protocol for a tomography system

Quality control phantoms and protocol for a tomography system Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing

More information

MEDICAL IMAGE ANALYSIS

MEDICAL IMAGE ANALYSIS SECOND EDITION MEDICAL IMAGE ANALYSIS ATAM P. DHAWAN g, A B IEEE Engineering in Medicine and Biology Society, Sponsor IEEE Press Series in Biomedical Engineering Metin Akay, Series Editor +IEEE IEEE PRESS

More information

Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications 1

Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications 1 Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Cynthia H. McCollough,

More information

Dual-Energy CT Based Monochromatic Imaging

Dual-Energy CT Based Monochromatic Imaging Dual-Energy CT Review Yu et al. Monochromatic Imaging Based on Dual-Energy CT Dual-Energy CT Review Lifeng Yu 1 Shuai Leng Cynthia H. McCollough Yu L, Leng S, McCollough CH Keywords: CT, dual-energy CT,

More information

Lecture 6: Medical imaging and image-guided interventions

Lecture 6: Medical imaging and image-guided interventions ME 328: Medical Robotics Winter 2019 Lecture 6: Medical imaging and image-guided interventions Allison Okamura Stanford University Updates Assignment 3 Due this Thursday, Jan. 31 Note that this assignment

More information

19 CP Replace uses of "data set" that do not refer to the PS3.5 defined meaning

19 CP Replace uses of data set that do not refer to the PS3.5 defined meaning 19 CP-1817 - Replace uses of "data set" that do not refer to the PS3.5 defined meaning Page 1 1 Status Assigned 2 Date of Last Update 2018/06/04 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com

More information

A New Approach for the Enhancement. of Dual-energy Computed Tomography Images. Kyung Kook Park

A New Approach for the Enhancement. of Dual-energy Computed Tomography Images. Kyung Kook Park A New Approach for the Enhancement of Dual-energy Computed Tomography Images by Kyung Kook Park A Dissertation Presented in Partial Fulfillment of the Requirements for the Degree Doctor of Philosophy Approved

More information

Some reference material

Some reference material Some reference material Physics reference book on medical imaging: A good one is The Essential Physics of Medical Imaging, 3 rd Ed. by Bushberg et al. ($170! new). However, there are several similar books

More information

Technical Publications

Technical Publications g GE Medical Systems Technical Publications Direction 2206867 Revision 1.0 DICOM Software V2.0 for DICOM V3.0 Copyright ª 1997 By General Electric Co. Do not duplicate 1 g GE Medical Systems THIS PAGE

More information

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology.

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology. Photon Counting Detectors and Their Applications in Medical Imaging Principles, Pitfalls and Progress in Photon-Counting-Detector Technology Taly Gilat Schmidt, PhD Associate Professor Department of Biomedical

More information

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system ECNDT 2014 Prague October 6-10, 2014 Dr. Eberhard Neuser Dr. Alexander Suppes Imagination

More information

Technical Publications

Technical Publications g GE Medical Systems Technical Publications Direction 5198554-100 Revision 1 Cortex ID for DICOM Copyright ª 2007 By General Electric Co. Do not duplicate Including: REVISION HISTORY REV Version DATE REASON

More information

DICOM CONFORMANCE STATEMENT MEDIA STORAGE FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A. with XIDF-037A and XIDF-038A or SRS-1000A

DICOM CONFORMANCE STATEMENT MEDIA STORAGE FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A. with XIDF-037A and XIDF-038A or SRS-1000A DICOM CONFORMANCE STATEMENT MEDIA STORAGE FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A with XIDF-037A and XIDF-038A or SRS-1000A (MIIXR0002EAC) 2001 IMPORTANT! (1) No part of this manual may

More information

IHE Radiation Oncology Technical Framework Supplement. Treatment Delivery Plan Content (TDPC) Rev. 1.1 Trial Implementation

IHE Radiation Oncology Technical Framework Supplement. Treatment Delivery Plan Content (TDPC) Rev. 1.1 Trial Implementation Integrating the Healthcare Enterprise 5 IHE Radiation Oncology Technical Framework Supplement 10 Treatment Delivery Plan Content 15 Rev. 1.1 Trial Implementation 20 Date: November 16, 2016 Author: IHE

More information

A Comparative Study of Two Methods for the Correction of Beam Hardening Artifacts in X-ray Computed Tomography

A Comparative Study of Two Methods for the Correction of Beam Hardening Artifacts in X-ray Computed Tomography A Comparative Study of Two Methods for the Correction of Beam Hardening Artifacts in X-ray Computed Tomography by Maryam Khalid Alarfaj submitted to Oregon State University in partial fulfillment of the

More information

Continuous and Discrete Image Reconstruction

Continuous and Discrete Image Reconstruction 25 th SSIP Summer School on Image Processing 17 July 2017, Novi Sad, Serbia Continuous and Discrete Image Reconstruction Péter Balázs Department of Image Processing and Computer Graphics University of

More information

Medical Image Reconstruction Term II 2012 Topic 6: Tomography

Medical Image Reconstruction Term II 2012 Topic 6: Tomography Medical Image Reconstruction Term II 2012 Topic 6: Tomography Professor Yasser Mostafa Kadah Tomography The Greek word tomos means a section, a slice, or a cut. Tomography is the process of imaging a cross

More information

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel Proc. Of SPIE Vol 0626, Application of Optical Instrumentation in Medicine XIV and Picture Archiving and Communication Systems(PACS IV) for Medical Applications IV, ed. S J Dwyer/R H Schneider (Jan 1986)

More information

Medical Imaging BMEN Spring 2016

Medical Imaging BMEN Spring 2016 Name Medical Imaging BMEN 420-501 Spring 2016 Homework #4 and Nuclear Medicine Notes All questions are from the introductory Powerpoint (based on Chapter 7) and text Medical Imaging Signals and Systems,

More information

DICOM Enhanced XA/XRF Object New Dimensions for X-Ray Projection Imaging. (DICOM Supplement 83)

DICOM Enhanced XA/XRF Object New Dimensions for X-Ray Projection Imaging. (DICOM Supplement 83) DICOM Enhanced XA/XRF Object New Dimensions for X-Ray Projection Imaging (DICOM Supplement 83) Authors: Heinz Blendinger Bas Revet Francisco Sureda Rainer Thieme Speaker: Heinz Blendinger Siemens Medical

More information

Philips SPECT/CT Systems

Philips SPECT/CT Systems Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET

More information

Japan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field. Kouichi Inagaki ICC / IHI Corporation

Japan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field. Kouichi Inagaki ICC / IHI Corporation Japan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field Kouichi Inagaki ICC / IHI Corporation 13 th WORLD CONFERENCE ON INVESTMENT CASTING Paper: T3 Copyright

More information

Integrating Modalities to Enterprise PACS through Thinking Systems

Integrating Modalities to Enterprise PACS through Thinking Systems Integrating Modalities to Enterprise PACS through 1 The Need For most enterprise PACS implementations, traditional CT, MR, CR, DR and general Ultrasound are of primary concerns. Not much attention has

More information

PURE. ViSION Edition PET/CT. Patient Comfort Put First.

PURE. ViSION Edition PET/CT. Patient Comfort Put First. PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement DICOM Pro Plus as an SCU Array Corporation Index 0. INTRODUCTION...3 1. IMPLEMENTATION MODEL...4 1.1 APPLICATION DATA FLOW DIAGRAM...4 1.2 FUNCTIONAL DEFINITION OF APPLICATION

More information

Improvement of contrast using reconstruction of 3D Image by PET /CT combination system

Improvement of contrast using reconstruction of 3D Image by PET /CT combination system Available online at www.pelagiaresearchlibrary.com Advances in Applied Science Research, 2013, 4(1):285-290 ISSN: 0976-8610 CODEN (USA): AASRFC Improvement of contrast using reconstruction of 3D Image

More information

Michal E. Kulon, MD 1,2 1. Peter Komlosi, MD, PhD 3 3. Radiology Universe Institute, 2

Michal E. Kulon, MD 1,2 1. Peter Komlosi, MD, PhD 3 3. Radiology Universe Institute, 2 Blend+Proximity, a novel algorithm achieves high suppression of metallic streak artifacts and maximal preservation of contrast between soft tissues and iodinated contrast material on dual-energy CT scans

More information

ENHANCED IMAGING OF CORROSION IN AIRCRAFT STRUCTURES WITH REVERSE GEOMETRY X-RAY

ENHANCED IMAGING OF CORROSION IN AIRCRAFT STRUCTURES WITH REVERSE GEOMETRY X-RAY ENHANCED IMAGING OF CORROSION IN AIRCRAFT STRUCTURES WITH REVERSE GEOMETRY X-RAY William P. Winfree a,noreen A. Cmar-Mascis b, F. Raymond Parker a a NASA Langley Research Center, MS 231, Hampton VA 23681-2199

More information

Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2)

Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2) Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2) Author: Levon Vogelsang Introduction Cone beam computed tomography (CBCT), or cone beam CT technology, offers considerable promise

More information

Technical Publications

Technical Publications GE Medical Systems Technical Publications Direction 2259903-100 Revision 1 Advantage 3D XR 2.0 for DICOM V3.0 Copyright 2000 By General Electric Co. Do not duplicate GE Medical Systems REVISION HISTORY

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement ECLOS Tokyo, Japan (SN-C357E) Copyright Hitachi Medical Corporation. 2007, 2008. All rights reserved. History Revision Description Date Rev. 1.0 Initial. 2007/02/01 Rev 2.0

More information

DICOM Conformance Statement. DSI Release 4.4. Document Number June Copyright Philips Medical Systems Nederland B.V.

DICOM Conformance Statement. DSI Release 4.4. Document Number June Copyright Philips Medical Systems Nederland B.V. Philips Medical Systems DICOM Conformance Statement DSI Release 4.4 Document Number 4522 220 84061 6 June 1997 Copyright Philips Medical Systems Nederland B.V. 1997 Philips Medical Systems apple PHILIPS

More information

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling nd International Symposium on NDT in Aerospace - We..A. Comparison of Probing Error in Dimensional Measurement by Means of D Computed Tomography with Circular and Helical Sampling Jochen HILLER, Stefan

More information

Applying Hounsfield unit density calibration in SkyScan CT-analyser

Applying Hounsfield unit density calibration in SkyScan CT-analyser 1 Bruker-microCT Method note Applying Hounsfield unit density calibration in SkyScan CT-analyser Hounsfield units (HU) are a standard unit of x-ray CT density, in which air and water are ascribed values

More information

Optimization of Beam Spectrum and Dose for Lower-Cost CT. Mary Esther Braswell. Graduate Program in Medical Physics Duke University.

Optimization of Beam Spectrum and Dose for Lower-Cost CT. Mary Esther Braswell. Graduate Program in Medical Physics Duke University. Optimization of Beam Spectrum and Dose for Lower-Cost CT by Mary Esther Braswell Graduate Program in Medical Physics Duke University Date: Approved: James Dobbins, Supervisor Anuj Kapadia Robert Reiman

More information

The Near Future in Cardiac CT Image Reconstruction

The Near Future in Cardiac CT Image Reconstruction SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de

More information

The CT Synergy (V3.1/V3.2/V3.3/V3.4/V3.5) Conformance Statement

The CT Synergy (V3.1/V3.2/V3.3/V3.4/V3.5) Conformance Statement The CT Synergy (V3.1/V3.2/V3.3/V3.4/V3.5) Conformance Statement for DICO V3.0 Direction 2210079 Revision 1.0 do not duplicate Copyright 2000 by General Electric Company GE EDICAL SYSTES STATEENT DIR 2210079

More information