ARTCOLL PACKAGE FOR GAMMA ART6000 TM ROTATING GAMMA SYSTEM EMISSION SPECTRA CALCULATION

Size: px
Start display at page:

Download "ARTCOLL PACKAGE FOR GAMMA ART6000 TM ROTATING GAMMA SYSTEM EMISSION SPECTRA CALCULATION"

Transcription

1 ARTCOLL PACKAGE FOR GAMMA ART6000 TM ROTATING GAMMA SYSTEM EMISSION SPECTRA CALCULATION Radovan D Ili}. PhD Institute of Nuclear Sciences Vin~a Belgrade, Serbia and Tomasz K. Helenowski, M.D. 936 Burnham Court GLENVIEW, IL 60025, USA Author's rasacale@gmail.com in Serbia tkh@tkh.com in USA Belgrade - Chicago 2006

2 The Asymmetric Rotating Gamma System Gamma ART 6000 Figure 1. The RGS ART 6000 System view. The GAMMA ART 6000 Rotating Gamma System s design is based on the USA patents No. 5,528,653; No. 5,757,866; and 6,512,813 B1. The system utilizes concentric 360 degree arcs formed by convergent gamma-ray beams that are emitted from 30 Cobalt-60 cylindrical sources. The design of the GAMMA ART 6000 combines the accuracy of the static gamma design and the arching of the accelerator based stereotactic radiosurgery system. The rotation of the Cobalt-60 gamma-ray beams during treatment is the major enhancement resulting in many advantages. By rotating the source body, primary collimators and secondary collimators together, a smaller number of radiation sources are utilized. Because the beams converge from greater solid angle as compared with the static design, the radiation dose to normal, healthy tissue surrounding the target is reduced, since the dose to the normal tissue is spread over a greater volume. The secondary collimators are built-in, eliminating the need for secondary collimator helmets, simplifying setup and changing of the treatment spot size. The use of fewer Cobalt-60 sources also reduces the cost for changing the sources.

3 Compared to the linear accelerator based radiosurgery systems, the RGS design employs much larger number of arcs in a single fraction to minimize dose to organs and tissues surrounding the izocenter location, without increasing the treatment time and the positional uncertainties. The RGS system delivers dose from 6 groups of cobalt sources spaced at 60 degree increments around the central axis of the treatment machine. Each group contains 5 sources increasingly inclined from the transverse plane. The first source is located at an incline of 13 degrees which increases to 43 degrees for the fifth source. Additionally, each of the 6 groups contains a slight offset of incline angle. Figure 2. The helmets with holes for Cobalt-60 sources and a secondary collimators plugging. The combined effect of this geometry, when rotated during treatment is to produce a region of dose that can be thought of as five conical wedges whose axis is coincident with the rotation of the treatment machine. The treatment planning system Explorer 3D is implemented on a graphic computer workstation running Windows 2000 / NT. Dose calculation is accomplished in two steps. First, for each shot, a normalized dose rate matrix is generated. The individual shot matrix is then scaled, weighted and summed into a single dose rate matrix. Actual dose times are derived from the prescription dose and the dose rate matrix, and then adjusted for the age of the radioactive sources.

4 ARTCOLL PACKAGE Overview of Package ARTCOLL package calculates emission spectra at end of collimator for Gamma ART6000 TM Rotating Gamma System, using the FOTELP-COLL and RED-COLL versions of Monte Carlo FOTELP code ( The ARTCOLL package is a Windows and LINUX application and consists of two programs. First, the FOTELP-COLL, with full physics and real shape and with dimensions of the photon source from C0-60 source, is simulating their transport and gives an emission spectra of primary and secondary particles (photon and electron) at the end of primary or secondary collimators. This program is providing the experiments with collimators for obtaining 3D distribution of absorbed dose in phantoms. Second, the RED-COLL treating only photon transport from suorce to end of primary collimator. Program RED-COLL servs for fast, routine simulations of the emission spectra with good statistics (more then billion photons from source). The FOTELP-KNF code with full physics and real shapes and materials, uses preprocessed emission spectra for calculating 3D dose during single rotation of ART6000 (see FOTART package). Programs FOTELP-COLL and RED-COLL can simulate collimators RGS ART 6000 as independent packages. They are providing simulations of collimators different from those in ART 6000, such as collimators Leksell GammaKnife or collimators of therapeut accelerators, and also with the phantoms of various forms, dimensions and material composition. 1. Collimator configuration Program FOTELP-COLL with Monte Carlo techniques is simulating photon transport from Cobalt-60 source through that source and all parts of primary and secondary collimator, and in the end through the experimental phantom. The emission spectra of all primary and secondary particles that had reached the bases of primary or secondary collimator can be used for the continuation of simulation in other parts of collimator or experimental phantom. This opportunity was realized primarily to reduce the need of long-term simulation from the source to phantom by using the emission spectra. Fig 3. shows the cross section of ART 6000 collimator, cross section 4 of the photon source

5 and water phantom. The figure is illustrating all problems JDET Figure 3. The ART 6000 conllimators cross section of photon flow simulation, electrons and positrons, and their secondary particles through all the elements of collimator construction The FOTELP-COLL usage The preparation of FOTEL-COLL program for simulation begins with writing of geometry model in accordance to a used geometry module RFG. User already has previously prepared geometry models for all ART 6000 collimator types, and in folder INP2K5 they have the names: THA2K5... Input files for "4mm" collimator THB2K5... Input files for "8mm" collimator THC2K5... Input files for "14mm" collimator THD2K5... Input files for "18mm" collimator Each geometry form can be controlled by program rfgt.for in folder RFGTEST, and the procedure is described in fot-2k3.pdf. Program FOTELP-COLL has three simulation opportunities that user can choose by doing input of flag values ISURF and JDET in fotelp.inp. The flag values have following meanings: ISURF=10 for the simulation of emission spectra at the end of primary collimator with JDET=20 or secondary collimator with JDET=14 with the spectra written in file ASURF.DAT. Sometimes it is useful to know the emission spectra close under the photon source, what can be obtained with flags ISUR=10 and JDET=12.

6 ISURF=20 with the emission spectra of 3D dose simulations in water or some other phantom. This simulation is performed by program when file ASURF.DAT is used with name ASURF.INP. Option ISURF=20 is implementing for the emission spectra at the end of primary or secondary collimator. ISURF=30 for simulation of 3D absorbed doses in phantom. FEPDAT - Transition probabilities For materials used in collimators construction, FEPDAT.EXE creates files of PEC type, that are needed for the simulation with collimator materials contained in FEPDAT.INP. FOTELP - Input data preparation In the FOTCOM.FOR file there are maximal values declarated in PARAMETER statements. IGZON = 22 - number of geometry zones in RFG.INP; MATER = 8 - number of different materials in FEPDAT.INP; LEL = MATER*5 number of elements in all materials; LNE = 150 number of energy bins; MSUR = 2, number of lines in file ASURF.INP; MPIX, MPIY, MPIZ = 161 x 161 x161 voxel numbers along X,Y,Z axis. After adjusment of these values, you should create FOTELP.EXE. User must carefully check out FOTELP.INP according to the FOTINF.DOC file and set the values in blue, red and magenta color in this file. First and second steps in simulation Start FOTELP.EXE and after few minutes you will have the simulation test with ISURF=10 for NPOC=1E6 photons. After copying ASURF.INP file=asurf.dat file, do the simulation on target with ISURF=20 that you will change into FOTELP.INP. Result of simulation Result of these simulation provide file FIGDOS.DAT. That one is the matrix of deposited energies in target EMEP(k,i,j) in MeV/kg. The matrix user can be read for independent graphical presentations with dimension emep(kzk,ixi,jyj) open(8,file='figdos.dat') read(8,*) kzk,ixi,jyj do k=1,kzk

7 do j=1,jyj read(8,108) (emep(k,i,j),i=1,ixi) 108 format(64e11.3) close(8) Note: From the FOTELP.INP file take your dimensions of EMEP matrix. In ARTCOLL package the user is provided with program acdose.for for calculating the absorbed dose around izocenter of "tumor" in phantom. This program uses files 2 figdose.dat ( E ) and summa2.dat ( E ) for that purpose. Change of target distance i When you want new simulation with the variable target position, you should keep ASURF.INP file as referent, and then do the changes in files RFG.INP and FOTELP.INP. In RFG.INP file, move the lines D31 and D32 for DZ by writing new values in magenta of the field. For that same DZ, change Zt value in magnet of the field in FOTELP.INP. Change of collimator materials If you wish to change collimator materials but without the change of geometry of collimator types, you should choose the material codes in file material.dat, and after compiling fmatsel.for user can start fmatsel.exe. The further work should be done by following the previously procedures from, under the condition to carefully create input file FOTELP.INP. The trick is that the fields marked in the italic blue in FOTELP.INP should correspondent to the indexes of new materials written in FEPDAT.INP file. That is easy thing to do if you consult FOTELP.INP before making changes. Geometry file RFG.INP C [ THD2K5 August 17, 2005 ] C [ RGS ART 6000 '18 mm' collimator, Chicago - Tomasz Helenowski ] C [ Cylinders ] D1 [ OBLIK=6; JED=2; KOEF = 2.5, 2.5 ] D2 [ OBLIK=6; JED=2; KOEF = 1.3, 1.3 ] D3 [ OBLIK=6; JED=2; KOEF = 0.25, 0.25 ] D5 [ OBLIK=6; JED=2; KOEF = 1.3, 1.3 ] D6 [ OBLIK=6; JED=2; KOEF = 0.3, 0.3 ] D7 [ OBLIK=6; JED=2; KOEF = 1.0, 1.0 ] D8 [ OBLIK=8; JED=2; KOEF = 1.0, 1.0, ; TRANS=0.0, 0.0, ] D9 [ OBLIK=6; JED=2; KOEF = 8.0, 8.0 ] D10 [OBLIK=6;JED=2;KOEF=0.05,0.05;TRANS=-0.075, ] D11 [OBLIK=6;JED=2;KOEF=0.05,0.05;TRANS= 0.000, ] D12 [OBLIK=6;JED=2;KOEF=0.05,0.05;TRANS= 0.075, ] D13 [OBLIK=6;JED=2;KOEF=0.05,0.05;TRANS= 0.000, ] D14 [OBLIK=3;JED=2;KOEF=100.0, 100.0, ] C [ Planes ] i

8 D20 [ OBLIK=2; JED=1; KOEF = 0.0, 0.0, 1.0, ] D21 [ OBLIK=2; JED=1; KOEF = 0.0, 0.0, 1.0, ] D22 [ OBLIK=2; JED=1; KOEF = 0.0, 0.0, 1.0, ] D23 [ OBLIK=2; JED=1; KOEF = 0.0, 0.0, 1.0, ] D24 [ OBLIK=2; JED=1; KOEF = 0.0, 0.0, 1.0, ] D25 [ OBLIK=2; JED=1; KOEF = 0.0, 0.0, 1.0, ] D26 [ OBLIK=2; JED=1; KOEF = 0.0, 0.0, 1.0, ] D27 [ OBLIK=2; JED=1; KOEF = 0.0, 0.0, 1.0, ] D28 [ OBLIK=2; JED=1; KOEF = 0.0, 0.0, 1.0, ] D29 [ OBLIK=2; JED=1; KOEF = 0.0, 0.0, 1.0, ] D30 [ OBLIK=2; JED=1; KOEF = 0.0, 0.0, 1.0, ] D31 [ OBLIK=2; JED=1; KOEF = 0.0, 0.0, 1.0, 2.5 ] D32 [ OBLIK=2; JED=1; KOEF = 0.0, 0.0, 1.0, 18.5 ] C [ Complex space ] D51 [ OBLIK=1; LOGIC= 10I(26 I-27) ] D52 [ OBLIK=1; LOGIC= 11I(26 I-27) ] D53 [ OBLIK=1; LOGIC= 12I(26 I-27) ] D54 [ OBLIK=1; LOGIC= 13I(26 I-27) ] D55 [ OBLIK=1; LOGIC= (3I(26I-27))I-51I-52I-53I-54 ] D56 [ OBLIK=1; LOGIC= 2I(28I-30) ] D57 [ OBLIK=1; LOGIC= 2I(27I-28) ] D58 [ OBLIK=1; LOGIC= 6I(24I-25) ] D59 [ OBLIK=1; LOGIC= (5I(24I-25))I-58] D60 [ OBLIK=1; LOGIC= 8I(22I-23) ] D61 [ OBLIK=1; LOGIC= (7I(22I-23))I-60 ] D62 [ OBLIK=1; LOGIC= 9I(32I-31) ] D71 [ OBLIK=1; LOGIC= 3I(26I-27) ] D72 [ OBLIK=1; LOGIC= 5I(24I-25) ] D88 [ OBLIK=1; LOGIC= 2I(23I-24) ] D89 [ OBLIK=1; LOGIC= 2I(26I-27) ] D85 [ OBLIK=1; LOGIC= 89I-71 ] D73 [ OBLIK=1; LOGIC= 7I(22I-23) ] D74 [ OBLIK=1; LOGIC= 5I(25I-26) ] D63 [ OBLIK=1; LOGIC= (1I(22I-30)I-56I-57I-89I-74I-72I-73I-88) ] D64 [ OBLIK=1; LOGIC= 1I(20I-21) ] D65 [ OBLIK=1; LOGIC= 1I(21I-22) ] D66 [ OBLIK=1; LOGIC= 1I(31I-20) ] D67 [ OBLIK=1; LOGIC= (14I-(1I(22I-30))I-64I-62I-65I-66) ] C [ Material zones ] Z1 [ DOMEN = 51 ] Z2 [ DOMEN = 52 ] Z3 [ DOMEN = 53 ] Z4 [ DOMEN = 54 ] Z5 [ DOMEN = 55 ] Z6 [ DOMEN = 56 ] Z7 [ DOMEN = 57 ] Z8 [ DOMEN = 58 ] Z9 [ DOMEN = 59 ] Z10 [ DOMEN = 60 ] Z11 [ DOMEN = 61 ] Z12 [ DOMEN = 74 ] Primary collimator end Z13 [ DOMEN = 65 ] Z14 [ DOMEN = 64 ] Lexan detektor Z15 [ DOMEN = 66 ] Z16 [ DOMEN = 62 ] Z17 [ DOMEN = 63 ] Z18 [ DOMEN = 67 ]

9 Z19 [ DOMEN = 85 ] Z20 [ DOMEN = 88 ] Secondary collimator end Z21 [ DOMEN =-14 ] K [ End of data ] FEPDAT.INP Material data file Input file fepdat.inp is prepared by user starting from the material being used for photon source, collimators and phantom, For RGG ART 6000 these materials were accepted: Co, Fe, W, Air, Lexan and Tissue Soft (ICRU 4 components), and their index values in file material.dat are 26, 27, 74, 104, 219 and 262. Preserving that sequence, one starts program fmatsel.exe which prepares input fepdat.inp. After running program fepdat.exe one obtaines file fepdat.out which becomes the first part of input file fotelp.inp. From the end of fepdat.out, here on the user is writing the remaining of fotelp.inp file: 0.010, 0.06, 0.06 Cutoff Energy: photon, electron, positron [MeV] (These values should be chosen carefully due to time of simulation) , 0, 0, 0 Atomic relaxation, Delta electron, Bremsstrahlung (If any number is greater then zero, the program is following the process coresponding to that number). Zones number, JGEM, ISURF, JDET 21, 202, 10, 12 JGEM = 101 for one cylindrical source; JGEM=202 for 4 x 26 cylindrical sources Zones, Material index, JAWA, Material (comment) JAWA = 0 electron history to cutoff energy JAWA = 1 electron history in following material (see fot-2k3.pdf) 1, 2, 1 Co 2, 2, 1 Co 3, 2, 1 Co 4, 2, 1 Co 5, 1, 1 Fe 6, 3, 1 W 7, 1, 1 W 8, 4, 0 Air 9, 3, 1 W 10, 4, 0 Air 11, 3, 1 W 12, 6, 0 Air 13, 1, 1 Fe 14, 5, 1 Lexan 15, 4, 0 Air 16, 4, 0 TISSUE 17, 1, 1 Fe 18, 4, 0 Air

10 19, 3, 1 W 20, 4, 0 Air 21, -2, 0 Vacuum 0.0, 0.0, 10.5 Target center (X tm,y tm,z tm ) 0.001, 0.050, , 2.60, , , , , , 2, 0, , Number of photon, Cutoff time [minutes] 1.33, , , 1.0, X_base, 0.5 Y_base and Z_target [cm] 40, 40, 60 Number of bins alog above lines The scheme of photon source, primary and secondary collimator with phantom is shown at Figure.3. In planning the routine therapy, fotelp-coll is to be used only at the beginning or when necessary in order to obtain an emission spectra at the bottom of primary collimator with option ISURF=10 and flag JDET=20. (See details in fotinp.doc). Sometimes there is a need of knowing the distribution of deposed energy in phantom, when any of 4 collimator types is to be used. Program fotelp-coll has such an opportunity, and a following text is related to that usage which user starts from the command line. a) The choice of source material, primary collimator and protection. Number of materials (first line) is to be chosen by starting program fmatsel.exe, then the codes of materials from file material.dat (second line), lower and upper energy of particles (MeV), the factor of scale MFAK=8 (third line) and parameters of angular distributions 1, 1, 1 (fourth line). At the end, the file fepdat.inp appears for starting program fepdat.exe which is calculating cross section and transition probabilities for the chosen materials. b) Data preparation for the simulation starts with putting data into file fotelp.inp. Then the program fotelp.exe simulates photon transport and obtaining of ASURF.DAT file with particles emission spectra. This program uses any rfg.inp geometry file for simulation for collimators of 4mm, 8mm, 14mm or 18mm type (THA2K5, THB2K5,THC2K5 or THD2K5), because any of them has an identical primary collimator. c) For this way of using program FOTELP-COLL one should consult fot-3k3.pdf where all the details of data preparation are described.

11 1.2. The RED-COLL code At the beginning of FOTELP-COLL program description we pointed that collimator simulations run with full-physics model so that the entire physical picture of photon and secondary particles transport is to be preserved. Since very small number of particles arrive at target - patients head, the simulation of that picture is demanding lots of time. For the acceptable result of simulation it is needed to take between 10 8 and 10 9 photons in 4 As the emission spectra is simulation characteristics, it's long duration can be tolerated because the spectra can be obtained independetly from it's usage in therapy planning. However, that approach may be abandoned when certain processes that occur with charged particles in construction materials are not of interest for tumor radiating. If we ignore processes of secondary interest and focus main attention at the simulation of primary and scattered photon transport, than the program RED-COLL can be written with reduced physical picture which provides only emission spectra at the end of primary collimator. With program RED-COLL, simulation of RGS ART 6000 collimator takes much less time than simulation with programs FOTEL-COLL. In paragraph 1.1. The procedures for preparation and usage of program FOTELP-COLL are described. For the program RAD-COLL these procedures and all input files are prepared in the same way Single collimator and phantom's experiments A single RGS ART 6000 channel consists of photon source, primary and secondary collimator. When the phantom is placed at some distance at collimator axis, one gets an experimental channel that is suitable for numerical experiments in radiotherapy. In all phases of developing program FOTELP-COLL such need was taken in consideration. Here we describe two opportunities in using that program. First one is regarding homogenuos and second heterogenuos phantoms. Homogenous phantoms is obtained when instead of cylindrical phantom, which is limited by cylinder D9 and planes D31 and D32, we place a box in RFG.INP. Phantoms can be filled with water or some other material. For those configurations one can use the characteristic of program FOTELP-COLL to provide 3D absorbed dose in voxels. Heterogenous phantoms is obtained when instead of cylindrical phantom, which is limited by cylinder D9 and planes D31 and D32, we place a dish with shape of box having the wall in RFG.INP. Phantoms can be filled with water or some other material, and the wall can be of material equal to scull. Exploration of such phantoms is supported by the characteristic of program FOTELP-COLL to provide 3D absorbed dose in voxels. Procession of simulation results with various phantoms is supported by program acdose.for for calculating the absorbed dose around "tumor" izocenter in phantom. Program acdose.for uses files figdose.dat 2 E ) and summa2.dat ( E ) to calculate dose when user chooses voxel numbers around ( i izocenter. i

Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations

Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations A. Al-Basheer, G. Sjoden, M. Ghita Computational Medical Physics Team Nuclear & Radiological Engineering University

More information

ELECTRON DOSE KERNELS TO ACCOUNT FOR SECONDARY PARTICLE TRANSPORT IN DETERMINISTIC SIMULATIONS

ELECTRON DOSE KERNELS TO ACCOUNT FOR SECONDARY PARTICLE TRANSPORT IN DETERMINISTIC SIMULATIONS Computational Medical Physics Working Group Workshop II, Sep 30 Oct 3, 2007 University of Florida (UF), Gainesville, Florida USA on CD-ROM, American Nuclear Society, LaGrange Park, IL (2007) ELECTRON DOSE

More information

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization

More information

Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code

Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code American Journal of Biomedical Engineering 216, 6(4): 124-131 DOI: 1.5923/j.ajbe.21664.3 Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code Ankit Kajaria 1,*, Neeraj Sharma

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Monte Carlo Methods 1 Monte Carlo! Most accurate at predicting dose distributions! Based on

More information

Comparison of measured and Monte Carlo calculated electron beam central axis depth dose in water

Comparison of measured and Monte Carlo calculated electron beam central axis depth dose in water Original article UDC: 681.324:519.245 Archive of Oncology 2001;9(2):83-7. Comparison of measured and Monte Carlo calculated electron beam central axis depth dose in water Darko LALIÆ 1 Radovan D. ILIÆ

More information

Assesing multileaf collimator effect on the build-up region using Monte Carlo method

Assesing multileaf collimator effect on the build-up region using Monte Carlo method Pol J Med Phys Eng. 2008;14(3):163-182. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0014-0 website: http://www.pjmpe.waw.pl M. Zarza Moreno 1, 2, N. Teixeira 3, 4, A. P. Jesus 1, 2, G. Mora 1 Assesing multileaf

More information

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them Dose Distributions George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Purpose To familiarize the resident with dose distributions and the factors that affect them

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

Study of scattered photons from the collimator system of Leksell Gamma Knife using the EGS4 Monte Carlo Code

Study of scattered photons from the collimator system of Leksell Gamma Knife using the EGS4 Monte Carlo Code Study of scattered photons from the collimator system of Leksell Gamma Knife using the EGS4 Monte Carlo Code Joel Y. C. Cheung Gamma Knife Centre, Canossa Hospital, 1 Old Peak Road, Hong Kong K. N. Yu

More information

Basic Radiation Oncology Physics

Basic Radiation Oncology Physics Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides

More information

Implementation of the EGSnrc / BEAMnrc Monte Carlo code - Application to medical accelerator SATURNE43

Implementation of the EGSnrc / BEAMnrc Monte Carlo code - Application to medical accelerator SATURNE43 International Journal of Innovation and Applied Studies ISSN 2028-9324 Vol. 6 No. 3 July 2014, pp. 635-641 2014 Innovative Space of Scientific Research Journals http://www.ijias.issr-journals.org/ Implementation

More information

I. INTRODUCTION. Figure 1. Radiation room model at Dongnai General Hospital

I. INTRODUCTION. Figure 1. Radiation room model at Dongnai General Hospital International Journal of Computational Engineering Research Vol, 04 Issue, 4 Simulation of Photon and Electron dose distributions 5 code for the treatment area using the linear electron accelerator (LINAC)

More information

Indrin Chetty Henry Ford Hospital Detroit, MI. AAPM Annual Meeting Houston 7:30-8:25 Mon 08/07/28 1/30

Indrin Chetty Henry Ford Hospital Detroit, MI.   AAPM Annual Meeting Houston 7:30-8:25 Mon 08/07/28 1/30 Review of TG105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics,

More information

gpmc: GPU-Based Monte Carlo Dose Calculation for Proton Radiotherapy Xun Jia 8/7/2013

gpmc: GPU-Based Monte Carlo Dose Calculation for Proton Radiotherapy Xun Jia 8/7/2013 gpmc: GPU-Based Monte Carlo Dose Calculation for Proton Radiotherapy Xun Jia xunjia@ucsd.edu 8/7/2013 gpmc project Proton therapy dose calculation Pencil beam method Monte Carlo method gpmc project Started

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!

More information

Outline. Monte Carlo Radiation Transport Modeling Overview (MCNP5/6) Monte Carlo technique: Example. Monte Carlo technique: Introduction

Outline. Monte Carlo Radiation Transport Modeling Overview (MCNP5/6) Monte Carlo technique: Example. Monte Carlo technique: Introduction Monte Carlo Radiation Transport Modeling Overview () Lecture 7 Special Topics: Device Modeling Outline Principles of Monte Carlo modeling Radiation transport modeling with Utilizing Visual Editor (VisEd)

More information

Michael Speiser, Ph.D.

Michael Speiser, Ph.D. IMPROVED CT-BASED VOXEL PHANTOM GENERATION FOR MCNP MONTE CARLO Michael Speiser, Ph.D. Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX September 1 st, 2012 CMPWG Workshop Medical

More information

Dose Calculations: Where and How to Calculate Dose. Allen Holder Trinity University.

Dose Calculations: Where and How to Calculate Dose. Allen Holder Trinity University. Dose Calculations: Where and How to Calculate Dose Trinity University www.trinity.edu/aholder R. Acosta, W. Brick, A. Hanna, D. Lara, G. McQuilen, D. Nevin, P. Uhlig and B. Slater Dose Calculations - Why

More information

Monte Carlo simulations. Lesson FYSKJM4710 Eirik Malinen

Monte Carlo simulations. Lesson FYSKJM4710 Eirik Malinen Monte Carlo simulations Lesson FYSKJM4710 Eirik Malinen MC simulations 1 Simulations of stochastic processes Interactions are stochastic: the path of a single ionizing particle may not be predicted Interactions

More information

Measurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code

Measurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code reports of practical oncology and radiotherapy 1 5 (2 0 1 0) 64 68 available at www.sciencedirect.com journal homepage: http://www.rpor.eu/ Original article Measurement of depth-dose of linear accelerator

More information

Monte Carlo simulations

Monte Carlo simulations MC simulations Monte Carlo simulations Eirik Malinen Simulations of stochastic processes Interactions are stochastic: the path of a single ioniing particle may not be predicted Interactions are quantified

More information

COMPARISON OF DOSE CALCULATION ALGORITHMS FOR LEKSELL GAMMA KNIFE PERFEXION USING MONTE CARLO VOXEL PHANTOMS

COMPARISON OF DOSE CALCULATION ALGORITHMS FOR LEKSELL GAMMA KNIFE PERFEXION USING MONTE CARLO VOXEL PHANTOMS COMPARISON OF DOSE CALCULATION ALGORITHMS FOR LEKSELL GAMMA KNIFE PERFEXION USING MONTE CARLO VOXEL PHANTOMS Jan Pipek 1, Josef Novotný Jr. 1,2,3, Josef Novotný 1, Petra Kozubíková 1 1 Faculty of Nuclear

More information

Photon beam dose distributions in 2D

Photon beam dose distributions in 2D Photon beam dose distributions in 2D Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2014 Acknowledgments! Narayan Sahoo PhD! Richard G Lane (Late) PhD 1 Overview! Evaluation

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

Use of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta

Use of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta Use of Monte Carlo modelling in radiotherapy linac design David Roberts, PhD Senior Physicist Elekta Contents Overview of Elekta What we do Where we use Monte Carlo Codes and resources Example : Agility

More information

THE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION

THE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION 2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 THE SIMULATION OF

More information

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

C a t p h a n / T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice

More information

A SYSTEM OF DOSIMETRIC CALCULATIONS

A SYSTEM OF DOSIMETRIC CALCULATIONS A SYSTEM OF DOSIMETRIC CALCULATIONS INTRODUCTION Dose calculation based on PDD and TAR have Limitations The dependence of PDD on SSD Not suitable for isocentric techniques TAR and SAR does not depend on

More information

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Artifact Mitigation in High Energy CT via Monte Carlo Simulation PIERS ONLINE, VOL. 7, NO. 8, 11 791 Artifact Mitigation in High Energy CT via Monte Carlo Simulation Xuemin Jin and Robert Y. Levine Spectral Sciences, Inc., USA Abstract The high energy (< 15 MeV) incident

More information

MCNP4C3-BASED SIMULATION OF A MEDICAL LINEAR ACCELERATOR

MCNP4C3-BASED SIMULATION OF A MEDICAL LINEAR ACCELERATOR Computational Medical Physics Working Group Workshop II, Sep 3 Oct 3, 7 University of Florida (UF), Gainesville, Florida USA on CD-ROM, American Nuclear Society, LaGrange Park, IL (7) MCNP4C3-BASED SIMULATION

More information

A dedicated tool for PET scanner simulations using FLUKA

A dedicated tool for PET scanner simulations using FLUKA A dedicated tool for PET scanner simulations using FLUKA P. G. Ortega FLUKA meeting June 2013 1 Need for in-vivo treatment monitoring Particles: The good thing is that they stop... Tumour Normal tissue/organ

More information

Source Model Tuning for a 6 MV Photon Beam used in Radiotherapy

Source Model Tuning for a 6 MV Photon Beam used in Radiotherapy Journal of Physics: Conference Series Source Model Tuning for a 6 MV Photon Beam used in Radiotherapy To cite this article: Lukas A Hirschi et al 2007 J. Phys.: Conf. Ser. 74 021008 View the article online

More information

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. 2-D D Dose-CT Mapping in Geant4 Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. Table of Contents Background & Purpose Materials Methods

More information

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters Monte Carlo Simulation for Neptun 1 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters M.T. Bahreyni Toossi a, M. Momen Nezhad b, S.M. Hashemi a a Medical Physics Research Center,

More information

EGS5 Monte Carlo Code: Installation

EGS5 Monte Carlo Code: Installation EGS5 Monte Carlo Code: Installation What do you need to install? EGS5 Cygwin CGView gvim (or any editor) EGS5 Monte Carlo Code: Installation Installing CYGWIN Installing EGS5 Running an EGS5 TEST Any trouble

More information

Development a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code

Development a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code Iran. J. Radiat. Res., 2006; 4 (1): 7-14 Development a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code. Mesbahi * Department of Medical Physics, Medical School,

More information

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE QA SOFTWARE FAST, PRECISE Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, VMAT, SRS and brachytherapy plans no film, no phantoms, no linac

More information

Proton dose calculation algorithms and configuration data

Proton dose calculation algorithms and configuration data Proton dose calculation algorithms and configuration data Barbara Schaffner PTCOG 46 Educational workshop in Wanjie, 20. May 2007 VARIAN Medical Systems Agenda Broad beam algorithms Concept of pencil beam

More information

Code characteristics

Code characteristics The PENELOPE Computer code M.J. Anagnostakis Nuclear Engineering Department National Technical University of Athens The PENELOPE code system PENetration and Energy LOss of Positrons and Electrons in matter

More information

Monte Carlo simulation of photon and electron transport

Monte Carlo simulation of photon and electron transport First Barcelona Techno Week Course on semiconductor detectors ICCUB, 11-15th July 2016 Monte Carlo simulation of photon and electron transport Francesc Salvat Monte Carlo 1 Simulations performed with the

More information

Comparison of Predictions by MCNP and EGSnrc of Radiation Dose

Comparison of Predictions by MCNP and EGSnrc of Radiation Dose Comparison of Predictions by MCNP and EGSnrc of Radiation Dose Computational Medical Physics Working Group Workshop II, Sep 30 Oct 3, 2007 at UF imparted to various Material Targets by Beams and small

More information

Synrad3D Photon propagation and scattering simulation

Synrad3D Photon propagation and scattering simulation Synrad3D Photon propagation and scattering simulation G. Dugan, D. Sagan CLASSE, Cornell University, Ithaca, NY 14853 USA Abstract As part of the Bmad software library, a program called Synrad3D has been

More information

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Krista Chytyk MSc student Supervisor: Dr. Boyd McCurdy Introduction The objective of cancer radiotherapy

More information

Application of MCNP Code in Shielding Design for Radioactive Sources

Application of MCNP Code in Shielding Design for Radioactive Sources Application of MCNP Code in Shielding Design for Radioactive Sources Ibrahim A. Alrammah Abstract This paper presents three tasks: Task 1 explores: the detected number of as a function of polythene moderator

More information

Deliverable D10.2. WP10 JRA04 INDESYS Innovative solutions for nuclear physics detectors

Deliverable D10.2. WP10 JRA04 INDESYS Innovative solutions for nuclear physics detectors MS116 Characterization of light production, propagation and collection for both organic and inorganic scintillators D10.2 R&D on new and existing scintillation materials: Report on the light production,

More information

Monaco VMAT. The Next Generation in IMRT/VMAT Planning. Paulo Mathias Customer Support TPS Application

Monaco VMAT. The Next Generation in IMRT/VMAT Planning. Paulo Mathias Customer Support TPS Application Monaco VMAT The Next Generation in IMRT/VMAT Planning Paulo Mathias Customer Support TPS Application 11.05.2011 Background What is Monaco? Advanced IMRT/VMAT treatment planning system from Elekta Software

More information

MONTE CARLO MODELING OF THE GAMMA KNIFE PERFEXION RYAN C.M. BEST. A Thesis Submitted to the Graduate Faculty of

MONTE CARLO MODELING OF THE GAMMA KNIFE PERFEXION RYAN C.M. BEST. A Thesis Submitted to the Graduate Faculty of MONTE CARLO MODELING OF THE GAMMA KNIFE PERFEXION BY RYAN C.M. BEST A Thesis Submitted to the Graduate Faculty of WAKE FOREST UNIVERSITY GRADUATE SCHOOL OF ARTS AND SCIENCES in Partial Fulfillment of the

More information

Absorbed Fractions for Multi-Region Models of the Kidneys in ICRP/ICRU Voxel Phantoms

Absorbed Fractions for Multi-Region Models of the Kidneys in ICRP/ICRU Voxel Phantoms Absorbed Fractions for Multi-Region Models of the Kidneys in ICRP/ICRU Voxel Phantoms Sakae KINASE *, Masanori KIMURA, Shogo TAKAHARA and Toshimitsu HOMMA * 2-4 Shirane, Shirakata, Tokai-mura, Naka-gun,

More information

Disclosure 7/24/2014. Validation of Monte Carlo Simulations For Medical Imaging Experimental validation and the AAPM Task Group 195 Report

Disclosure 7/24/2014. Validation of Monte Carlo Simulations For Medical Imaging Experimental validation and the AAPM Task Group 195 Report Validation of Monte Carlo Simulations For Medical Imaging Experimental validation and the AAPM Task Group 195 Report Ioannis Sechopoulos, Ph.D., DABR Diagnostic Imaging Physics Lab Department of Radiology

More information

A CT-based Monte Carlo Dose Calculations for Proton Therapy Using a New Interface Program

A CT-based Monte Carlo Dose Calculations for Proton Therapy Using a New Interface Program World Academy of Science, Engineering and Technology 53 29 A CT-based Monte Carlo Dose Calculations for Proton Therapy Using a New Interface Program A. Esmaili Torshabi, A. Terakawa, K. Ishii, H. Yamazaki,

More information

Detector simulations for in-beam PET with FLUKA. Francesco Pennazio Università di Torino and INFN, TORINO

Detector simulations for in-beam PET with FLUKA. Francesco Pennazio Università di Torino and INFN, TORINO Detector simulations for in-beam PET with FLUKA Francesco Pennazio Università di Torino and INFN, TORINO francesco.pennazio@unito.it Outline Why MC simulations in HadronTherapy monitoring? The role of

More information

The Monte Carlo simulation of a Package formed by the combination of three scintillators: Brillance380, Brillance350, and Prelude420.

The Monte Carlo simulation of a Package formed by the combination of three scintillators: Brillance380, Brillance350, and Prelude420. EURONS I3 506065 JRA9 RHIB Report made during stay IEM-CSIC Madrid december 2006 MINISTERIO DE ASUNTOS EXTERIORES Y DE COOPERACIÓN AECI VICESECRETARÍA GENERAL The Monte Carlo simulation of a Package formed

More information

Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation

Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation Danial Seifi Makrani 1, Hadi Hasanzadeh 2*, Tayyeb Allahverdi Pourfallah 3, Arash Ghasemi 4, Majid

More information

Evaluation of RayXpert for shielding design of medical facilities

Evaluation of RayXpert for shielding design of medical facilities Evaluation of Raypert for shielding design of medical facilities Sylvie Derreumaux 1,*, Sophie Vecchiola 1, Thomas Geoffray 2, and Cécile Etard 1 1 Institut for radiation protection and nuclear safety,

More information

Click to edit Master title style

Click to edit Master title style New features in Serpent 2 for fusion neutronics 5th International Serpent UGM, Knoxville, TN, Oct. 13-16, 2015 Jaakko Leppänen VTT Technical Research Center of Finland Click to edit Master title Outline

More information

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms.

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms. Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms. D.Leone, A.Häußler Intitute for Nuclear Waste Disposal, Karlsruhe Institute for Technology,

More information

Comparison of absorbed dose distribution 10 MV photon beam on water phantom using Monte Carlo method and Analytical Anisotropic Algorithm

Comparison of absorbed dose distribution 10 MV photon beam on water phantom using Monte Carlo method and Analytical Anisotropic Algorithm Journal of Physics: Conference Series PAPER OPEN ACCESS Comparison of absorbed dose distribution 1 MV photon beam on water phantom using Monte Carlo method and Analytical Anisotropic Algorithm To cite

More information

Limitations in the PHOTON Monte Carlo gamma transport code

Limitations in the PHOTON Monte Carlo gamma transport code Nuclear Instruments and Methods in Physics Research A 480 (2002) 729 733 Limitations in the PHOTON Monte Carlo gamma transport code I. Orion a, L. Wielopolski b, * a St. Luke s/roosevelt Hospital, Columbia

More information

Accuracy of treatment planning calculations for conformal radiotherapy van 't Veld, Aart Adeodatus

Accuracy of treatment planning calculations for conformal radiotherapy van 't Veld, Aart Adeodatus University of Groningen Accuracy of treatment planning calculations for conformal radiotherapy van 't Veld, Aart Adeodatus IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's

More information

Improvements in Monte Carlo simulation of large electron fields

Improvements in Monte Carlo simulation of large electron fields SLAC-PUB-12908 October 2007 Improvements in Monte Carlo simulation of large electron fields Bruce A Faddegon 1, Joseph Perl 2 and Makoto Asai 2 1 University of California San Francisco Comprehensive Cancer

More information

Dosimetry Simulations with the UF-B Series Phantoms using the PENTRAN-MP Code System

Dosimetry Simulations with the UF-B Series Phantoms using the PENTRAN-MP Code System Dosimetry Simulations with the UF-B Series Phantoms using the PENTRAN-MP Code System A. Al-Basheer, M. Ghita, G. Sjoden, W. Bolch, C. Lee, and the ALRADS Group Computational Medical Physics Team Nuclear

More information

Image-based Monte Carlo calculations for dosimetry

Image-based Monte Carlo calculations for dosimetry Image-based Monte Carlo calculations for dosimetry Irène Buvat Imagerie et Modélisation en Neurobiologie et Cancérologie UMR 8165 CNRS Universités Paris 7 et Paris 11 Orsay, France buvat@imnc.in2p3.fr

More information

New Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC

New Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC New Technology in Radiation Oncology James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC Technology s s everywhere From the imaging chain To the planning system To the linac To QA..it..it

More information

UNIVERSITY OF CALGARY. Monte Carlo model of the Brainlab Novalis Classic 6 MV linac using the GATE simulation. platform. Jared K.

UNIVERSITY OF CALGARY. Monte Carlo model of the Brainlab Novalis Classic 6 MV linac using the GATE simulation. platform. Jared K. UNIVERSITY OF CALGARY Monte Carlo model of the Brainlab Novalis Classic 6 MV linac using the GATE simulation platform by Jared K. Wiebe A THESIS SUBMITTED TO THE FACULTY OF GRADUATE STUDIES IN PARTIAL

More information

Chapter 9 Field Shaping: Scanning Beam

Chapter 9 Field Shaping: Scanning Beam Chapter 9 Field Shaping: Scanning Beam X. Ronald Zhu, Ph.D. Department of Radiation Physics M. D. Anderson Cancer Center Houston, TX June 14-18, 2015 AAPM - Summer School 2015, Colorado Spring Acknowledgement

More information

Tutorial for PENELOPE (version 2014)

Tutorial for PENELOPE (version 2014) Tutorial for PENELOPE (version 2014) The distribution package has the following directory structure and contents: penelope fsource pendbase other mains doc pdfiles gview tables emfields shower pencyl penmain

More information

Shielding factors for traditional safety glasses

Shielding factors for traditional safety glasses Shielding factors for traditional safety glasses Malcolm McEwen, Hong Shen and Ernesto Mainegra-Hing Ionizing Radiation Standards, National Research Council Canada Alan DuSautoy, Radiation and Health Sciences

More information

Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.

Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran. DXRaySMCS First User Friendly Interface Developed for Prediction of Diagnostic Radiology X-Ray Spectra Produced by Monte Carlo (MCNP-4C) Simulation in Iran M.T. Bahreyni Toosi a*, H. Moradi b, H. Zare

More information

Transitioning from pencil beam to Monte Carlo for electron dose calculations

Transitioning from pencil beam to Monte Carlo for electron dose calculations Transitioning from pencil beam to Monte Carlo for electron dose calculations Jessie Huang-Vredevoogd (jyhuang4@wisc.edu) University of Wisconsin NCC AAPM October 12, 2019 1 Topics to cover Background RayStation

More information

On compensator design for photon beam intensity-modulated conformal therapy

On compensator design for photon beam intensity-modulated conformal therapy On compensator design for photon beam intensity-modulated conformal therapy Steve B. Jiang a) and Komanduri M. Ayyangar Department of Radiation Therapy, Medical College of Ohio, 3000 Arlington Avenue,

More information

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS Tomotherapy Physics Machine Twinning and Quality Assurance Emilie Soisson, MS Tomotherapy at UW- Madison Treating for nearly 5 years Up to ~45 patients a day on 2 tomo units Units twinned to facilitate

More information

Supercomputing the Cascade Processes of Radiation Transport

Supercomputing the Cascade Processes of Radiation Transport 19 th World Conference on Non-Destructive Testing 2016 Supercomputing the Cascade Processes of Radiation Transport Mikhail ZHUKOVSKIY 1, Mikhail MARKOV 1, Sergey PODOLYAKO 1, Roman USKOV 1, Carsten BELLON

More information

THESIS NEUTRON PRODUCTION AND TRANSPORT AT A MEDICAL LINEAR ACCELERATOR. Submitted by. Amber Allardice

THESIS NEUTRON PRODUCTION AND TRANSPORT AT A MEDICAL LINEAR ACCELERATOR. Submitted by. Amber Allardice THESIS NEUTRON PRODUCTION AND TRANSPORT AT A MEDICAL LINEAR ACCELERATOR Submitted by Amber Allardice Department of Environmental and Radiological Health Sciences In partial fulfillment of the requirements

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

Image Guided Multibeam Radiotherapy

Image Guided Multibeam Radiotherapy José Luis Freijo a* Image Guided Multibeam Radiotherapy a Comisión Nacional de Energía Atómica, Av. del Libertador 8250, 1429, Buenos Aires, Argentina. Abstract. This paper provides an outlook of the status

More information

NKS GammaUser

NKS GammaUser Uncertainties Marie-Christine Lépy Laboratoire National Henri Becquerel CEA Saclay, F-91191 Gif-sur-Yvette cedex, France E-mail: marie-christine.lepy@cea.fr NKS GammaUser 2014 Helsinki 6-8 October 2014

More information

CHAPTER 9 INFLUENCE OF SMOOTHING ALGORITHMS IN MONTE CARLO DOSE CALCULATIONS OF CYBERKNIFE TREATMENT PLANS: A LUNG PHANTOM STUDY

CHAPTER 9 INFLUENCE OF SMOOTHING ALGORITHMS IN MONTE CARLO DOSE CALCULATIONS OF CYBERKNIFE TREATMENT PLANS: A LUNG PHANTOM STUDY 148 CHAPTER 9 INFLUENCE OF SMOOTHING ALGORITHMS IN MONTE CARLO DOSE CALCULATIONS OF CYBERKNIFE TREATMENT PLANS: A LUNG PHANTOM STUDY 9.1 INTRODUCTION 9.1.1 Dose Calculation Algorithms Dose calculation

More information

Digital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities

Digital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities sup176_pc_newrtradiations.doc: Sup 176: 2 nd Generation RT Additional RT Treatment Modalities Page 1 2 4 Digital Imaging and Communications in Medicine (DICOM) 6 Supplement 176: Second Generation Radiotherapy

More information

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon WHITE PAPER Introduction An image guidance system based on Cone Beam CT (CBCT) is included in Leksell

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

SPATIALLY FRACTIONATED PROTON THERAPY: A MONTE CARLO VERIFICATION. A Thesis Presented to The Academic Faculty. Jenna Fair

SPATIALLY FRACTIONATED PROTON THERAPY: A MONTE CARLO VERIFICATION. A Thesis Presented to The Academic Faculty. Jenna Fair SPATIALLY FRACTIONATED PROTON THERAPY: A MONTE CARLO VERIFICATION A Thesis Presented to The Academic Faculty By Jenna Fair In Partial Fulfillment Of the Requirements for the Degree Master of Science in

More information

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 3, JUNE 2003 405 Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging Edward J. Soares, Kevin W. Germino,

More information

Measurement of Skin Dose

Measurement of Skin Dose Measurement of Skin Dose Sources of Uncertainty Kenneth A. Fetterly, Ph.D. William Pavlicek, Ph.D. Dan Bednarek, PhD 2014 AAPM Annual Meeting, Austin Texas 2013 MFMER slide-1 Purpose 1. Present a framework

More information

Combining Analytical and Monte Carlo Modelling for Industrial Radiology

Combining Analytical and Monte Carlo Modelling for Industrial Radiology 19 th World Conference on Non-Destructive Testing 2016 Combining Analytical and Monte Carlo Modelling for Industrial Radiology Carsten BELLON, Gerd-Rüdiger JAENISCH, Andreas DERESCH BAM Bundesanstalt für

More information

Op#miza#on of CUDA- based Monte Carlo Simula#on for Radia#on Therapy. GTC 2014 N. Henderson & K. Murakami

Op#miza#on of CUDA- based Monte Carlo Simula#on for Radia#on Therapy. GTC 2014 N. Henderson & K. Murakami Op#miza#on of CUDA- based Monte Carlo Simula#on for Radia#on Therapy GTC 2014 N. Henderson & K. Murakami The collabora#on Geant4 @ Special thanks to the CUDA Center of Excellence Program Makoto Asai, SLAC

More information

Release Notes for Dosimetry Check with Convolution-Superposition Collapsed Cone Algorithm (CC)

Release Notes for Dosimetry Check with Convolution-Superposition Collapsed Cone Algorithm (CC) Dosimetry Check for convolution/superposition collapsed cone (CC), Page 1 of 11 Release Notes for Dosimetry Check with Convolution-Superposition Collapsed Cone Algorithm (CC) File: CCReleaseNotes.doc Date:

More information

Particle track plotting in Visual MCNP6 Randy Schwarz 1,*

Particle track plotting in Visual MCNP6 Randy Schwarz 1,* Particle track plotting in Visual MCNP6 Randy Schwarz 1,* 1 Visual Editor Consultants, PO Box 1308, Richland, WA 99352, USA Abstract. A visual interface for MCNP6 has been created to allow the plotting

More information

Constructing System Matrices for SPECT Simulations and Reconstructions

Constructing System Matrices for SPECT Simulations and Reconstructions Constructing System Matrices for SPECT Simulations and Reconstructions Nirantha Balagopal April 28th, 2017 M.S. Report The University of Arizona College of Optical Sciences 1 Acknowledgement I would like

More information

Modeling the ORTEC EX-100 Detector using MCNP

Modeling the ORTEC EX-100 Detector using MCNP Modeling the ORTEC EX-100 Detector using MCNP MCNP is a general-purpose Monte Carlo radiation transport code for modeling the interaction of radiation with materials based on composition and density. MCNP

More information

Medical Imaging BMEN Spring 2016

Medical Imaging BMEN Spring 2016 Name Medical Imaging BMEN 420-501 Spring 2016 Homework #4 and Nuclear Medicine Notes All questions are from the introductory Powerpoint (based on Chapter 7) and text Medical Imaging Signals and Systems,

More information

UNCOMPROMISING QUALITY

UNCOMPROMISING QUALITY ION CHAMBERS UNCOMPROMISING QUALITY Designed with over 30 years of scientific integrity for a broad range of dosimetry measurements in diverse radiation beams Farmer-type Chambers For absolute dosimetry

More information

VALIDATION OF A MONTE CARLO DOSE CALCULATION ALGORITHM FOR CLINICAL ELECTRON BEAMS IN THE PRESENCE OF PHANTOMS WITH COMPLEX HETEROGENEITIES

VALIDATION OF A MONTE CARLO DOSE CALCULATION ALGORITHM FOR CLINICAL ELECTRON BEAMS IN THE PRESENCE OF PHANTOMS WITH COMPLEX HETEROGENEITIES VALIDATION OF A MONTE CARLO DOSE CALCULATION ALGORITHM FOR CLINICAL ELECTRON BEAMS IN THE PRESENCE OF PHANTOMS WITH COMPLEX HETEROGENEITIES by Shayla Landfair Enright A Thesis Submitted to the Faculty

More information

Getting Started with EGSnrc

Getting Started with EGSnrc Getting Started with EGSnrc R. Townson, F. Tessier, E. Mainegra, B. Walters Ionizing Radiation Standards National Research Council of Canada, Ottawa Printed: June 1, 2018 Abstract This manual is a quick

More information

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Effects of the difference in tube voltage of the CT scanner on. dose calculation Effects of the difference in tube voltage of the CT scanner on dose calculation Dong Joo Rhee, Sung-woo Kim, Dong Hyeok Jeong Medical and Radiological Physics Laboratory, Dongnam Institute of Radiological

More information

Radiation therapy treatment plan optimization

Radiation therapy treatment plan optimization H. Department of Industrial and Operations Engineering The University of Michigan, Ann Arbor, Michigan MOPTA Lehigh University August 18 20, 2010 Outline 1 Introduction Radiation therapy delivery 2 Treatment

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 61675-2 First edition 1998-01 Radionuclide imaging devices Characteristics and test conditions Part 2: Single photon emission computed tomographs Dispositifs d imagerie par radionucléides

More information

Data. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy

Data. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy Data ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy The ModuLeaf Mini Multileaf

More information

Click to edit Master title style

Click to edit Master title style Introduction to Serpent Code Fusion neutronics workshop, Cambridge, UK, June 11-12, 2015 Jaakko Leppänen VTT Technical Research Center of Finland Click to edit Master title Outline style Serpent overview

More information

Digital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities

Digital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities Modalities Page 1 2 4 Digital Imaging and Communications in Medicine (DICOM) 6 Supplement 176: Second Generation Radiotherapy Additional RT Treatment Modalities 8 10 12 14 16 DICOM Standards Committee,

More information