Evaluation of Penalty Design in Penalized Maximum- likelihood Image Reconstruction for Lesion Detection
|
|
- Gloria Floyd
- 6 years ago
- Views:
Transcription
1 Evaluation of Penalty Design in Penalized Maximum- likelihood Image Reconstruction for Lesion Detection Li Yang, Andrea Ferrero, Rosalie J. Hagge, Ramsey D. Badawi, and Jinyi Qi Supported by NIBIB under grant no. R01EB B I O M E D I C A L E N G I N E E R I N G
2 Introduction PET is widely used in oncology Increasingly being used for staging and treatment monitoring Challenging to detect small tumors (< 10 mm) Numerous efforts have been focused on improving detection of small tumors by developing New PET tracers New PET scanners: time- of- flight, dedicated systems, etc. Optimized image reconstruction methods (our focus)
3 PML Image Reconstruction PML image reconstruction ˆx(y) = arg max[l(y x) (x)] x 0 Quadratic penalty function (x) = Shift- invariant penalty kernel NX j=1 X ln j jl(x j x l ) = x t Rx Rx =Ker R (m, n, o) x(m, n, o) D 1 st order quadratic penalty 4 Ker R Parameterization (Stayman et al 000) Ker R = X l Penalty design: to find the optimal lb l = b 1 b
4 Optimize Penalty Function Quadratic penalty function has been optimized for: Achieve uniform resolution Optimize local contrast to noise ratio Improve lesion detectability in D (Stayman et al 000, 004, Qi 000) (Qi 1999) (Qi et al 006, Yang et al 01) In the previous work [1], we have designed the penalty function to improve the lesion detectability in 3D image, and validated it using computer simulations Goal: to evaluate the proposed penalty function for lesion detection using real data [1] Yang et al, 014, PMB,
5 Lesion Detection To study lesion detectability, numerical observers are often used è Normal (H0) è Abnormal (H1) Human Observer Figure of merit (FOM) Numerical Observer 1 ROC CURVE TP AUC Popular numerical observers: D: channelized Hotelling observer (CHO) 0 1 FP (Yao and Barrett 199) 3D: single- slice CHO, volumetric CHO, multislice CHO, multislice multiview CHO, multiview CHO ( Liang et al 008, Kim et al 004, Platisa et al 011, Chen et al 00, Gifford et al 006) 5
6 Multiview CHO D channels 3D image ˆx Channel outputs U ˆx 1 3 FOM : SNR = z 0 U 0 K 1 Uz Internal noise n with zero mean and covariance K n Test statistics z 0 U 0 K 1 ( ) U ˆx + n ( : mean reconstructed lesion profile) Hotelling Observer U ˆx + n 6
7 Multiview CHO SNR of PML FOM: SNR = z 0 U 0 K 1 Uz For lesion at a given location, SNR of PML reconstruction can be evaluated by using locally shift invariant approximation: (Qi 004) where 8 8 >< { Ũ 8 >< i } i=1 { { i } >: N i=1 > 8 >< >< { i } N i=1 {{µ >: i } N i=1 >: : the Fourier transform of the channels : the Fourier transform of the column vector of the Fisher information matrix corresponding to the lesion location : the Fourier transform of the expected lesion profile : the Fourier transform of the penalty kernel 7
8 Penalty Design Objective: to maximize for every voxel 1 st order quadratic penalty: b 1 = , and b = Equivalent to finding the optimum weights γ : ˆ = arg max SNR ( ) We used 9 nearest neighboring voxels ( ) in 3D To reduce computational cost, the penalty weights were computed for a subset of preselected voxels and the nearest neighbor interpolation was used to form the overall penalty function 8
9 Patient Background Data We used a 60- minute dynamic PET scan of a female patient on a GE DST whole- body PET scanner 5 mci FDG injection FOV: 700 mm (transaxial) x 157 mm (axial) Crystals size: 6.54 x 6.54 x 30 mm The FOV covered the heart, breasts, and part of the lungs and liver The last 45 minutes data were summed to create a high- count sinogram with ~800M events (9 of a normal scan) The reconstructed patient image was verified by a radiologist to be free of lesion and the sinogram data were used as the noise- free normal background 9
10 Lesion- Present Data A Na- point source was scanned in air at 7 locations Attenuated by the patient body Added to the patient sinogram 7 implausible positions were excluded Sample reconstruction with a superimposed lesion in the liver Independent Poisson noise was introduced to generate 00 independent noisy realizations, each with around 90M total counts, mimicking a 5- minute scan 10
11 Image Reconstruction Image matrix size: 19 x 19 x 47 Voxel size: 3.64 x 3.64 x 3.7 mm Randoms, scatters, and normalization factors were estimated from the patient data using a manufacturer provided software and included in the reconstruction Reconstruction algorithm: PML with the 1 st order quadratic penalty PML with the optimized penalty Both reconstruction methods used the same forward/back projectors and correction factors 11
12 Detection Performance Lesion detectability of PML reconstructions at two representative locations Proposed penalty 1.8 Location 1 SNR st order quadratic penalty log10( ) 1.8 Location SNR log10( ) SNR of theoretical (curves) and Monte Carlo results ( x & o ) 1
13 Human Observer One human observer was asked to perform AFC experiment The resulting percent correct (PC) can be converted to SNR: (Burgess 1995) SNR at two representative locations: Location 1 Location Proposed penalty SNR 1 SNR st order quadratic penalty log10( ) log10( ) SNR of theoretical (curves) and human observer results ( x & o ) 13
14 Human Observer AFC Results The β value of the 1 st order penalty was tuned for each tumor individually The optimized penalty improves the SNR by up to 15%.5 Maximum SNR Optimal penalty 1st order quadratic Tumor # 14
15 Statistical Test Each AFC experiment was considered as a Bernoulli experiment with possible outcomes Sort the AFC results of two penalty functions into 4 categories 1 st order penalty Proposed penalty Number of cases 1 Correct Correct N 1 = 991 Wrong Wrong N = 56 3 Correct Wrong N 3 = Wrong Correct N 4 = 14 Under the null hypothesis, categories 3 and 4 should happen with same probability (N3 N4) Using the McNemar test, we obtained a p- value <
16 Summary We designed the penalty function in PML image reconstruction for 3D lesion detection, and evaluated the detection performance using real data Both numerical and human observer results show that the proposed penalty outperforms the 1 st order quadratic penalty In the future we will investigate the dynamic PET for lesion detection 16
17 Thank you!
Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system
Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation
More informationSTATISTICAL positron emission tomography (PET) image
IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 23, NO. 9, SEPTEMBER 2004 1057 Accurate Estimation of the Fisher Information Matrix for the PET Image Reconstruction Problem Quanzheng Li, Student Member, IEEE,
More informationScatter Correction for Positron Emission Mammography
Scatter Correction for Positron Emission Mammography Jinyi Qi and Ronald H. Huesman Center for Functional Imaging, Lawrence Berkeley National Laboratory, Berkeley, CA 9472, USA. E-mail: {jqi,rhhuesman}@lbl.gov
More information8/2/2016. Acknowledgement. Common Clinical Questions. Presumption Images are Good Enough to accurately answer clinical questions
Image Quality Assessment using Model Observers: Clinical Implementation and Practical Considerations Shuai Leng, PhD Associate Professor, Department of Radiology Mayo Clinic, Rochester MN Acknowledgement
More informationAssessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT
Assessment of 3D performance metrics D and 3D Metrics of Performance Towards Quality Index: Volumetric imaging systems X-ray based Volumetric imaging systems: CBCT/CT Tomosynthesis Samuel Richard and Ehsan
More informationQ.Clear. Steve Ross, Ph.D.
Steve Ross, Ph.D. Accurate quantitation (SUV - Standardized Uptake Value) is becoming more important as clinicians seek to utilize PET imaging for more than just diagnosing and staging disease, but also
More informationModeling and Incorporation of System Response Functions in 3D Whole Body PET
Modeling and Incorporation of System Response Functions in 3D Whole Body PET Adam M. Alessio, Member IEEE, Paul E. Kinahan, Senior Member IEEE, and Thomas K. Lewellen, Senior Member IEEE University of
More informationIntroduction to Positron Emission Tomography
Planar and SPECT Cameras Summary Introduction to Positron Emission Tomography, Ph.D. Nuclear Medicine Basic Science Lectures srbowen@uw.edu System components: Collimator Detector Electronics Collimator
More informationResolution and Noise Properties of MAP Reconstruction for Fully 3-D PET
IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 19, NO. 5, MAY 2000 493 Resolution and Noise Properties of MAP Reconstruction for Fully 3-D PET Jinyi Qi*, Member, IEEE, and Richard M. Leahy, Member, IEEE Abstract
More informationWorkshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET
Workshop on Quantitative SPECT and PET Brain Studies 14-16 January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Físico João Alfredo Borges, Me. Corrections in SPECT and PET SPECT and
More informationIntroduc)on to PET Image Reconstruc)on. Tomographic Imaging. Projec)on Imaging. Types of imaging systems
Introduc)on to PET Image Reconstruc)on Adam Alessio http://faculty.washington.edu/aalessio/ Nuclear Medicine Lectures Imaging Research Laboratory Division of Nuclear Medicine University of Washington Fall
More informationNIH Public Access Author Manuscript J Nucl Med. Author manuscript; available in PMC 2010 February 9.
NIH Public Access Author Manuscript Published in final edited form as: J Nucl Med. 2010 February ; 51(2): 237. doi:10.2967/jnumed.109.068098. An Assessment of the Impact of Incorporating Time-of-Flight
More informationMAXIMUM a posteriori (MAP) or penalized ML image
42 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 25, NO. 1, JANUARY 2006 Mean and Covariance Properties of Dynamic PET Reconstructions From List-Mode Data Evren Asma and Richard M. Leahy* Abstract We derive
More informationComparison of 3D PET data bootstrap resampling methods for numerical observer studies
2005 IEEE Nuclear Science Symposium Conference Record M07-218 Comparison of 3D PET data bootstrap resampling methods for numerical observer studies C. Lartizien, I. Buvat Abstract Bootstrap methods have
More informationPerformance Evaluation of radionuclide imaging systems
Performance Evaluation of radionuclide imaging systems Nicolas A. Karakatsanis STIR Users meeting IEEE Nuclear Science Symposium and Medical Imaging Conference 2009 Orlando, FL, USA Geant4 Application
More informationValidation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition
Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.
More informationPURE. ViSION Edition PET/CT. Patient Comfort Put First.
PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion
More informationTask-based Assessment of X-ray Breast Imaging Systems Using Insilico
55 th : Virtual Tools for Validation of X-ray Breast Imaging Systems Task-based Assessment of X-ray Breast Imaging Systems Using Insilico Modeling Tools Rongping Zeng Division of Imaging and Applied Mathematics
More informationDeviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies
g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,
More informationPET Image Reconstruction using Anatomical Information through Mutual Information Based Priors
2005 IEEE Nuclear Science Symposium Conference Record M11-354 PET Image Reconstruction using Anatomical Information through Mutual Information Based Priors Sangeetha Somayajula, Evren Asma, and Richard
More informationProjection Space Maximum A Posterior Method for Low Photon Counts PET Image Reconstruction
Proection Space Maximum A Posterior Method for Low Photon Counts PET Image Reconstruction Liu Zhen Computer Department / Zhe Jiang Wanli University / Ningbo ABSTRACT In this paper, we proposed a new MAP
More informationNIH Public Access Author Manuscript J Nucl Med Technol. Author manuscript; available in PMC 2013 December 09.
NIH Public Access Author Manuscript Published in final edited form as: J Nucl Med Technol. 2013 December ; 41(4):. doi:10.2967/jnmt.113.131904. Effect of Varying Number of OSEM Subsets on PET Lesion Detectability
More informationUvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication
UvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication Citation for published version (APA): Kruis, M. F. (2014). Motion compensation for 4D PET/CT General
More informationImprovement of contrast using reconstruction of 3D Image by PET /CT combination system
Available online at www.pelagiaresearchlibrary.com Advances in Applied Science Research, 2013, 4(1):285-290 ISSN: 0976-8610 CODEN (USA): AASRFC Improvement of contrast using reconstruction of 3D Image
More informationPET Quantification using STIR
PET Quantification using STIR STIR User s Meeting Charalampos Tsoumpas, PhD King s College London Hammersmith Imanet 1 PET Quantification Image elements should correspond to concentration of the injected
More informationPET image reconstruction algorithms based on maximum
IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 18, NO. 4, APRIL 1999 293 A Theoretical Study of the Contrast Recovery and Variance of MAP Reconstructions from PET Data Jinyi Qi,* Member, IEEE, and Richard
More informationREMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO
REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO Abstract C.S. Levin, Y-C Tai, E.J. Hoffman, M. Dahlbom, T.H. Farquhar UCLA School of Medicine Division
More informationBiomedical Imaging. Computed Tomography. Patrícia Figueiredo IST
Biomedical Imaging Computed Tomography Patrícia Figueiredo IST 2013-2014 Overview Basic principles X ray attenuation projection Slice selection and line projections Projection reconstruction Instrumentation
More informationSplitting-Based Statistical X-Ray CT Image Reconstruction with Blind Gain Correction
Splitting-Based Statistical X-Ray CT Image Reconstruction with Blind Gain Correction Hung Nien and Jeffrey A. Fessler Department of Electrical Engineering and Computer Science University of Michigan, Ann
More informationSemi-Quantitative Metrics in Positron Emission Tomography. Michael Adams. Department of Biomedical Engineering Duke University.
Semi-Quantitative Metrics in Positron Emission Tomography by Michael Adams Department of Biomedical Engineering Duke University Date: Approved: Timothy G. Turkington, Supervisor Adam P. Wax Terence Z.
More informationQuantitative Statistical Methods for Image Quality Assessment
Quantitative Statistical Methods for Image Quality Assessment The Harvard community has made this article openly available. Please share how this access benefits you. Your story matters Citation Dutta,
More informationCorrelation between Model and Human Observer Performance on a Lesion Shape Discrimination Task in CT
Correlation between Model and Human Observer Performance on a Lesion Shape Discrimination Task in CT Yi Zhang, Shuai Leng, Lifeng Yu and Cynthia McCollough Department of Radiology Mayo Clinic, Rochester
More informationUniform Spatial Resolution for Statistical Image Reconstruction for 3D Cone-Beam CT
Uniform Spatial Resolution for Statistical Image Reconstruction for 3D Cone-Beam CT Jang Hwan Cho, Member, IEEE, and Jeffrey A. Fessler, Fellow, IEEE Department of EECS, University of Michigan, Ann Arbor,
More informationClassification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging
1 CS 9 Final Project Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging Feiyu Chen Department of Electrical Engineering ABSTRACT Subject motion is a significant
More informationEffects of the difference in tube voltage of the CT scanner on. dose calculation
Effects of the difference in tube voltage of the CT scanner on dose calculation Dong Joo Rhee, Sung-woo Kim, Dong Hyeok Jeong Medical and Radiological Physics Laboratory, Dongnam Institute of Radiological
More informationOp#miza#on of a Fast CT System for More Accurate Radiotherapy. Erica Cherry, Meng Wu, and Rebecca Fahrig
Op#miza#on of a Fast CT System for More Accurate Radiotherapy Erica Cherry, Meng Wu, and Rebecca Fahrig Problem: mo#on during treatment Two overlapping thorax views during normal breathing liver lung spine
More informationCorso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET
Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 5 SPECT, PET Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes
More informationTechnological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI. Floris Jansen, GE Healthcare July, 2015
Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI Floris Jansen, GE Healthcare July, 2015 PET/MR 101 : challenges Thermal Workflow & Apps RF interactions?!!
More informationMonte-Carlo-Based Scatter Correction for Quantitative SPECT Reconstruction
Monte-Carlo-Based Scatter Correction for Quantitative SPECT Reconstruction Realization and Evaluation Rolf Bippus 1, Andreas Goedicke 1, Henrik Botterweck 2 1 Philips Research Laboratories, Aachen 2 Fachhochschule
More informationReview of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA
Review of PET Physics Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Chart of Nuclides Z (protons) N (number of neutrons) Nuclear Data Evaluation Lab.
More informationCT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION
CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology
More informationMulti-slice CT Image Reconstruction Jiang Hsieh, Ph.D.
Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced
More informationIterative SPECT reconstruction with 3D detector response
Iterative SPECT reconstruction with 3D detector response Jeffrey A. Fessler and Anastasia Yendiki COMMUNICATIONS & SIGNAL PROCESSING LABORATORY Department of Electrical Engineering and Computer Science
More informationAxial block coordinate descent (ABCD) algorithm for X-ray CT image reconstruction
Axial block coordinate descent (ABCD) algorithm for X-ray CT image reconstruction Jeffrey A. Fessler and Donghwan Kim EECS Department University of Michigan Fully 3D Image Reconstruction Conference July
More informationThe Effects of PET Reconstruction Parameters on Radiotherapy Response Assessment. and an Investigation of SUV peak Sampling Parameters.
The Effects of PET Reconstruction Parameters on Radiotherapy Response Assessment and an Investigation of SUV peak Sampling Parameters by Leith Rankine Graduate Program in Medical Physics Duke University
More informationFast Projectors for Iterative 3D PET Reconstruction
2005 IEEE Nuclear Science Symposium Conference Record M05-3 Fast Projectors for Iterative 3D PET Reconstruction Sanghee Cho, Quanzheng Li, Sangtae Ahn and Richard M. Leahy Signal and Image Processing Institute,
More informationHigh-resolution 3D Bayesian image reconstruction using the micropet small-animal scanner
Phys. Med. Biol. 43 (1998) 1001 1013. Printed in the UK PII: S0031-9155(98)90627-3 High-resolution 3D Bayesian image reconstruction using the micropet small-animal scanner Jinyi Qi, Richard M Leahy, Simon
More informationClinically feasible reconstruction of 3D whole-body PET/CT data using blurred anatomical labels
INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (00) 0 PHYSICS INMEDICINE AND BIOLOGY PII: S003-955(0)56-9 Clinically feasible reconstruction of 3D whole-body PET/CT data using blurred anatomical labels
More informationComputer-Tomography II: Image reconstruction and applications
Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de
More informationMEDICAL IMAGING 2nd Part Computed Tomography
MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional
More information3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH
3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk
More informationReconstruction in CT and relation to other imaging modalities
Reconstruction in CT and relation to other imaging modalities Jørgen Arendt Jensen November 1, 2017 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering Center for Fast Ultrasound
More informationPhilips SPECT/CT Systems
Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET
More informationTomographic Reconstruction
Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)
More informationThe Design and Implementation of COSEM, an Iterative Algorithm for Fully 3-D Listmode Data
IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 20, NO. 7, JULY 2001 633 The Design and Implementation of COSEM, an Iterative Algorithm for Fully 3-D Listmode Data Ron Levkovitz, Dmitry Falikman*, Michael Zibulevsky,
More informationContinuation Format Page
C.1 PET with submillimeter spatial resolution Figure 2 shows two views of the high resolution PET experimental setup used to acquire preliminary data [92]. The mechanics of the proposed system are similar
More informationWhole-body PET using the glucose analog 18 F-FDG
Comparison of 2-Dimensional and 3-Dimensional Acquisition for 18 F-FDG PET Oncology Studies Performed on an LSO-Based Scanner Martin A. Lodge, PhD 1 ; Ramsey D. Badawi, PhD 2 ; Richard Gilbert, AA 3 ;
More informationMedical Image Processing: Image Reconstruction and 3D Renderings
Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,
More informationSingle and Multipinhole Collimator Design Evaluation Method for Small Animal SPECT
Single and Multipinhole Collimator Design Evaluation Method for Small Animal SPECT Kathleen Vunckx, Dirk Bequé, Michel Defrise and Johan Nuyts Abstract High resolution functional imaging of small animals
More information8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial
8//0 AAPM0 Scientific Symposium: Emerging and New Generation PET: Instrumentation, Technology, Characteristics and Clinical Practice Aug Wednesday 0:4am :pm Solid State Digital Photon Counting PET/CT Instrumentation
More informationAutomatic Lesion Detection for Measuring Response using Dynamic FDG-PET
Automatic Lesion Detection for Measuring Response using Dynamic FDG-PET Xiujuan Zheng a,b, Guangjian Tian a, Shaoli Song b, Gang Huang b, David Dagan Feng a,c a Department of Electronic and Information
More informationNuclear Medicine Imaging
Introduction to Medical Engineering (Medical Imaging) Suetens 5 Nuclear Medicine Imaging Ho Kyung Kim Pusan National University Introduction Use of radioactive isotopes for medical purposes since 1920
More information3-D PET Scatter Correction
Investigation of Accelerated Monte Carlo Techniques for PET Simulation and 3-D PET Scatter Correction C.H. Holdsworth, Student Member, IEEE, C.S. Levin", Member, IEEE, T.H. Farquhar, Student Member, IEEE,
More informationImproving Reconstructed Image Quality in a Limited-Angle Positron Emission
Improving Reconstructed Image Quality in a Limited-Angle Positron Emission Tomography System David Fan-Chung Hsu Department of Electrical Engineering, Stanford University 350 Serra Mall, Stanford CA 94305
More informationMassively Parallel GPU-friendly Algorithms for PET. Szirmay-Kalos László, Budapest, University of Technology and Economics
Massively Parallel GPU-friendly Algorithms for PET Szirmay-Kalos László, http://cg.iit.bme.hu, Budapest, University of Technology and Economics (GP)GPU: CUDA (OpenCL) Multiprocessor N Multiprocessor 2
More informationImage reconstruction for PET/CT scanners: past achievements and future challenges
Review Image reconstruction for PET/CT scanners: past achievements and future challenges PET is a medical imaging modality with proven clinical value for disease diagnosis and treatment monitoring. The
More informationCentral Slice Theorem
Central Slice Theorem Incident X-rays y f(x,y) R x r x Detected p(, x ) The thick line is described by xcos +ysin =R Properties of Fourier Transform F [ f ( x a)] F [ f ( x)] e j 2 a Spatial Domain Spatial
More informationDiagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology
Diagnostic imaging techniques Krasznai Zoltán University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology 1. Computer tomography (CT) 2. Gamma camera 3. Single Photon
More informationarxiv: v1 [cs.cv] 6 Jun 2017
Volume Calculation of CT lung Lesions based on Halton Low-discrepancy Sequences Liansheng Wang a, Shusheng Li a, and Shuo Li b a Department of Computer Science, Xiamen University, Xiamen, China b Dept.
More informationLow-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration
Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Joonki Noh, Jeffrey A. Fessler EECS Department, The University of Michigan Paul E. Kinahan Radiology Department,
More informationCorso di laurea in Fisica A.A Fisica Medica 4 TC
Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure
More informationAssessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy
Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy Physics Department, Faculty of Applied Science,Umm Al-Qura
More informationCS6375: Machine Learning Gautam Kunapuli. Mid-Term Review
Gautam Kunapuli Machine Learning Data is identically and independently distributed Goal is to learn a function that maps to Data is generated using an unknown function Learn a hypothesis that minimizes
More informationComputational Medical Imaging Analysis Chapter 4: Image Visualization
Computational Medical Imaging Analysis Chapter 4: Image Visualization Jun Zhang Laboratory for Computational Medical Imaging & Data Analysis Department of Computer Science University of Kentucky Lexington,
More informationHigh Temporal-Resolution Dynamic PET Image Reconstruction Using A New Spatiotemporal Kernel Method
1 High Temporal-Resolution Dynamic PET Image Reconstruction Using A New Spatiotemporal Kernel Method Guobao Wang Abstract Current clinical dynamic PET has an effective temporal resolution of 5-10 seconds,
More informationSuper-resolution Reconstruction of Fetal Brain MRI
Super-resolution Reconstruction of Fetal Brain MRI Ali Gholipour and Simon K. Warfield Computational Radiology Laboratory Children s Hospital Boston, Harvard Medical School Worshop on Image Analysis for
More informationUnion of Learned Sparsifying Transforms Based Low-Dose 3D CT Image Reconstruction
Union of Learned Sparsifying Transforms Based Low-Dose 3D CT Image Reconstruction Xuehang Zheng 1, Saiprasad Ravishankar 2, Yong Long 1, Jeff Fessler 2 1 University of Michigan - Shanghai Jiao Tong University
More informationAcknowledgments and financial disclosure
AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories
More informationMichael Speiser, Ph.D.
IMPROVED CT-BASED VOXEL PHANTOM GENERATION FOR MCNP MONTE CARLO Michael Speiser, Ph.D. Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX September 1 st, 2012 CMPWG Workshop Medical
More informationEnhancement Image Quality of CT Using Single Slice Spiral Technique
Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,
More informationQIBA PET Amyloid BC March 11, Agenda
QIBA PET Amyloid BC March 11, 2016 - Agenda 1. QIBA Round 6 Funding a. Deadlines b. What projects can be funded, what cannot c. Discussion of projects Mechanical phantom and DRO Paul & John? Any Profile
More information664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005
664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005 Attenuation Correction for the NIH ATLAS Small Animal PET Scanner Rutao Yao, Member, IEEE, Jürgen Seidel, Jeih-San Liow, Member, IEEE,
More informationTime-of-Flight Technology
Medical Review Time-of-Flight Technology Bing Bai, PhD Clinical Sciences Manager, PET/CT Canon Medical Systems INTRODUCTION Improving the care for every patient while providing a high standard care to
More informationLimits of Dose Reduction in CT: How low is too low? Acknowledgement. Disclosures 8/2/2012
8/2/22 Limits of Dose Reduction in CT: How low is too low? Ehsan Samei Duke University Acknowledgement Disclosures Research grant: NIH Research grant: DHS Research grant: RSNA, QIBA Research grant: General
More informationAssessment of image quality for the new CT: Statistical reconstruction methods
Assessment of image quality for the new CT: Statistical reconstruction methods Jeffrey A. Fessler EECS Dept., BME Dept., Dept. of Radiology University of Michigan web.eecs.umich.edu/ fessler AAPM: Interactive
More informationLimitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods
Stereoscopic Breast Imaging Andrew D. A. Maidment, Ph.D. Chief, Physics Section Department of Radiology University of Pennsylvania Limitations of Projection Radiography Mammography is a projection imaging
More informationBy choosing to view this document, you agree to all provisions of the copyright laws protecting it.
Copyright 2007 IEEE. Reprinted from 2007 IEEE Nuclear Science Symposium Conference Record, Paper M13-5, p. 2838-41. This material is posted here with permission of the IEEE. Internal or personal use of
More informationQuantitative comparison of FBP, EM, and Bayesian reconstruction algorithms, including the impact of accurate system modeling, for the IndyPET scanner
Quantitative comparison of,, and Bayesian reconstruction algorithms, including the impact of accurate system modeling, for the IndyPET scanner Thomas Frese,,NedC.Rouze, Charles A. Bouman, Ken Sauer, and
More informationA Weighted Least Squares PET Image Reconstruction Method Using Iterative Coordinate Descent Algorithms
A Weighted Least Squares PET Image Reconstruction Method Using Iterative Coordinate Descent Algorithms Hongqing Zhu, Huazhong Shu, Jian Zhou and Limin Luo Department of Biological Science and Medical Engineering,
More informationMathematical methods and simulations tools useful in medical radiation physics
Mathematical methods and simulations tools useful in medical radiation physics Michael Ljungberg, professor Department of Medical Radiation Physics Lund University SE-221 85 Lund, Sweden Major topic 1:
More informationBackground 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J.
Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging Stephen J. Glick* J. Michael O Connor**, Clay Didier**, Mini Das*, * University of Massachusetts Medical
More informationImage Acquisition Systems
Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS
More informationPositron. MillenniumVG. Emission Tomography Imaging with the. GE Medical Systems
Positron Emission Tomography Imaging with the MillenniumVG GE Medical Systems Table of Contents Introduction... 3 PET Imaging With Gamma Cameras PET Overview... 4 Coincidence Detection on Gamma Cameras...
More informationImproving Positron Emission Tomography Imaging with Machine Learning David Fan-Chung Hsu CS 229 Fall
Improving Positron Emission Tomography Imaging with Machine Learning David Fan-Chung Hsu (fcdh@stanford.edu), CS 229 Fall 2014-15 1. Introduction and Motivation High- resolution Positron Emission Tomography
More informationRICE UNIVERSITY. Optimization of Novel Developments in Positron Emission Tomography (PET) Imaging. Tingting Chang
RICE UNIVERSITY Optimization of Novel Developments in Positron Emission Tomography (PET) Imaging by Tingting Chang A THESIS SUBMITTED IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE Doctor of
More information7/13/2015 EVALUATION OF NONLINEAR RECONSTRUCTION METHODS. Outline. This is a decades-old challenge
EVALUATION OF NONLINEAR RECONSTRUCTION METHODS Kyle J. Myers, Ph.D. Director, Division of Imaging, Diagnostics, and Software Reliability Office of Science and Engineering Laboratories, CDRH, FDA 2 Outline
More informationBME I5000: Biomedical Imaging
1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.
More informationIntroduction to Emission Tomography
Introduction to Emission Tomography Gamma Camera Planar Imaging Robert Miyaoka, PhD University of Washington Department of Radiology rmiyaoka@u.washington.edu Gamma Camera: - collimator - detector (crystal
More informationImaging-based Classification Algorithms on Clinical Trial Data with Injected Tumour Responses
Imaging-based Classification Algorithms on Clinical Trial Data with Injected Tumour Responses Yunpeng Li, Emily Porter, Mark Coates Dept. of Electrical and Computer Engineering, McGill University, Montréal,
More informationAccelerating CT Iterative Reconstruction Using ADMM and Nesterov s Methods
Accelerating CT Iterative Reconstruction Using ADMM and Nesterov s Methods Jingyuan Chen Meng Wu Yuan Yao June 4, 2014 1 Introduction 1.1 CT Reconstruction X-ray Computed Tomography (CT) is a conventional
More information