CT Basics: Image Processing and Reconstruction Module 4

Size: px
Start display at page:

Download "CT Basics: Image Processing and Reconstruction Module 4"

Transcription

1 Module 4 For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions. Any broadcast, duplication, circulation, public viewing, conference viewing or Internet posting of this product is strictly prohibited. Purchase of the product constitutes an agreement to these terms. In return for the licensed use, the Licensee hereby releases, and waives any and all claims and/or liabilities that may arise against ASRT as a result of the product and its licensing.

2 CT Basics: Image Processing and Reconstruction Module 4 1. Title Slide Module 4: Image Processing and Reconstruction 2. License Agreement 3. Objectives Welcome to Module 4: Image Processing and Reconstruction. After completing this module, you ll be able to: 1) Describe the steps required for computed tomography image reconstruction. 2) List the postprocessing techniques needed for image enhancement. 3) Define the tools needed used to view a CT image. 4) List the workstation applications used for specialized CT scanning. 5) Describe the methods of recording and archiving CT data. 4. Image Processing and Reconstruction Image processing and reconstruction are the meat and potatoes of computed tomography imaging. A great deal of data is acquired during the scanning process, but this information is only as valuable as the quality of the image presentation. The processing algorithms and how the data are reformatted determine how easily the images can be interpreted. This module explores many important aspects of CT image processing. The module briefly reviews the components and functions of the image reconstruction process and then discusses specific postprocessing techniques. We ll also look at the basic functions of the CT technologist s workstation and the tools technologists use to alter image appearance, and display and archive image data. 5. Minicomputer and Microprocessors CT scanners use minicomputers to rapidly process the collected scan data. The minicomputer contains a microprocessor that transforms the incoming signal into a binary code. The microprocessor determines how quickly data is processed. Data processing speed is important because there are more data bits than can be handled at one time. The faster the entire data set can be processed, the more quickly the viewer can see the resulting images. Parallel processing occurs when data are handled by more than one minicomputer. When minicomputers use parallel processing, a task is split up among more than one processing unit, and each processor works on part of the scan data. The processed information from each unit is merged and then, in most cases, is sent to the analog-to-digital converter, or ADC. The ADC converts incoming analog data to a digital format. Before the data can be displayed on a monitor, however, the digital signal must be changed back to an analog signal by the digital-to-analog converter, or DAC. 6. Array Processor Computed tomography uses a type of microprocessor known as an array processor. The array processor is responsible for receiving image data and performing very high-speed calculations. It divides

3 up the various imaging tasks, such as convolution and planar reconstruction, and performs these multiple tasks simultaneously. The array processor typically works using parallel processing so that multiple tasks are performed at the same time. When parallel processing involves many tasks at once, the processor is usually referred to as a large array parallel processor, or LAPP. 7. Backprojection At first, image data acquired during the CT scanning procedure was processed using backprojection reconstruction algorithms. During backprojection the reformatted raw data goes through convolution, or the addition of filters. The data are essentially copied to create a data set that looks like the original raw data. However, backprojection doesn t produce an exact duplicate of the original reformatted raw data, but rather generates new data by making an educated guess as to what the data should look like. This process involves taking the incoming raw data and combining it with the data of a prototype scan, so that the new data set appears similar to the anticipated outcome of the scan. This process was initially difficult because of artifacts caused by the backprojection process. Backprojection is not used in today s CT scanners because of numerous limitations, such as the unsharpness of the image, which may lead to an artifact, and the length of processing time. 8. Linear Interpolation Linear interpolation is a mathematical equation applied to data acquired spirally by a singledetector row scanner. Very simply, linear interpolation reconstructs the spirally acquired images into what looks like individual slices. Interpolation is necessary because although the scan is performed in a spiral fashion, the viewer must look at the images in a flat plane. If the image was viewed in the same way that it was acquired, it would be unreadable and appear skewed. The scan would be inconclusive because there would be no way of knowing where one slice began and another slice ended. Two types of interpolation have been used for image reconstruction: 360 linear interpolation and 180 linear interpolation. A 360 linear interpolation averages measurements taken from each circular path around the patient to generate the projection data. The measurements are taken at reference points 360 apart from one another, that is, from slice one to slice two, slice two to slice three and so on. The technique then takes these measurements and approximates the information not seen between two consecutive image slices. The disadvantage of 360 linear interpolation is that the distance between each gantry revolution might be large, leading the 360 reference points to be too far apart, which in turn may cause misregistration or incorrect image reconstruction. The 180 linear interpolation method works in the same way as 360 linear interpolation, except that the reference points are 180 apart. Because the reference points now overlap, there is less distance between each of the adjacent slices. The overlap reduces misregistration and artifacts, and increases spatial resolution. Most modern CT scanners use 180 linear interpolation. Click on the play animation button to see how a two dimensional image is created from a volume of data. 9. Filtered Backprojection Filtered backprojection also uses the process of convolution, or filter assignment, much like the original backprojection process; however, filtered backprojection can automatically remove artifacts that occur in anatomy of different densities, such as soft tissues and bone. Filtered backprojection assigns the data a specific kernel, sometimes referred to as a convolution algorithm, to the scanned image data.

4 The filter enhances certain types of anatomy based on the desired appearance of the scan. After the filters are added to the incoming scan data, the newly created information is put through the traditional backprojection method to produce an artifact-free image. It s important to know that assigning filters to the image data is what creates images without processing artifacts and is why backprojection alone cannot be used. 10. Longitudinal Interpolation with Z-axis Filtering Longitudinal interpolation with z-axis filtering typically is used by scanners with multiple detector element rows. The z-axis is perpendicular to the helical path the x-ray tube and detector array follows. This reconstruction algorithm was developed because the x-ray beam in multidetector row scanners is much wider than the beam used in a single-detector row scanner. The beam is so wide that the primary path of radiation often doesn t fall directly within each scan slice. Longitudinal interpolation with z-axis filtering calculates not only the distance from one slice to the next slice, or pitch, but also how the data are displayed for each image slice within the detector rows. The pitch affects interpolation calculations because if the image slices are farther apart, the reference points for the interpolation calculation also are farther apart. The calculation that determines how the slices fall on the detector rows is referred to as the z- gap. The z-gap algorithm measures the distance between two adjacent slices and is directly related to image quality. As the pitch decreases, the space between slices narrows, and the slices overlap. The z-gap is used because the x-ray beam might not be perpendicular to each detector during image acquisition. If this is the case, the peripheral, rather than the central, portion of the x-ray beam collects the scan data because of cone-beam geometry. The interpolation algorithm uses the information provided in the z-axis to fill in unavailable data. Use the slider bar to increase or decrease the pitch. 11. Interlaced Sampling Interlaced, or interleaved, sampling refers to the ability of the CT scanner to calculate image data from multiple, overlapping acquisition data sets. CT image data are reconstructed into sections that assume the data are new or discrete, and are located immediately next to the adjacent slice. This assumption can cause problems because one slice often overlaps the previous slice, and the calculation is identical to that of the previous calculation, leading to repeated information. Modern CT scanners have algorithms that separate the acquired scan data from the reconstructed scan data. 12. Practice Question 13. Practice Question 14. Reconstruction Reconstruction can be performed at different times during the scanning process, after the scan is completed or much later in another setting. During scanning, the initial data are preprocessed to eliminate various artifacts, such those caused by faulty detector elements or beam hardening, and to ensure that there are enough data samples to fulfill the scan requirements. The preprocessed data is referred to as reformatted raw data. Filtered backprojection is used for image reconstruction immediately following the scan. At this point, a kernel is applied to the incoming scan data. The kernel provides a reference range of desired image outcomes and represents what the scan should look like, much like a set of guidelines the incoming scan data should mimic. The filter adjusts the data by accentuating the anatomy of interest.

5 During the convolution process, the scan data are manipulated to show the desired, enhanced anatomy. Convolution removes artifacts present during the original backprojection method. After the appropriate windows are applied to the information, the data continue through the interpolation process. Reconstruction that is performed at a much later time is known as retrospective reconstruction. This type of reconstruction can create a variety of images, from 3-dimensional images to multiplanar reconstructions. 15. Reformatting The term reformatting is sometimes applied to the postprocessing technique of multiplanar reconstruction, or MPR. Multiplanar reconstruction rearranges the axial data set so that it can be seen from front to back, left to right and in any plane desired. We ll discuss MPR in another part of this module. Reformatting is used when the axial data set needs to be manipulated for methods such as for maximum intensity projection, or MIP and minimum intensity projection, or minip. We ll also discuss these reformatting methods in much greater detail at a later time. A disadvantage of reformatting images is the loss of image detail. Loss of image detail can occur for several reasons such as patient motion during image acquisition; however, loss of detail in reformatted images is a postprocessing issue that occurs after scanning. One reason for a loss of detail is that multiplanar reconstructions use only 10% of the data set for reformatting. If that part of the data set is not of sufficient quality, the user might have to switch to a reconstruction method that uses 100% of the data. 16. Specific Postprocessing Techniques CT technologists use many postprocessing techniques to manipulate imaging data. The techniques we ll discuss in this section affect how the final image is viewed. Although these techniques are designed to enhance the appearance of the displayed image, it s important to understand that they also remove image data during the process. The raw, unformatted data set collected during the acquisition process contains the most image information. 17. Maximum Intensity Pixels Maximum intensity pixels, more commonly known as a maximum intensity projections, or MIPs, are created using less than 10% of the total data set. To create a MIP, the computer software allows only the brightest, or whitest, shades to remain and removes all other densities not needed. The brightest shades are now the focus of attention for the viewer. This type of processing is especially useful in computed tomography angiography, or CTA. A CTA scan enhances the arteries with a positive iodinated contrast agent. The IV contrast appears as the brightest pixel values on the scan after the bones. In addition to artery reconstructions, this method sometimes is used for exams to evaluate the kidneys, ureters and bladder. 18. Minimum Intensity Pixels Minimum intensity pixels, more commonly known as a minimum intensity projections, or minips, also are created using less than 10% of the total data set. The computer software enhances the darkest shades within the scan, while removing the brighter shades. The darker shades are now the focus of attention for the viewer. This type of processing is especially useful in CT virtual colonography exams. Virtual colonography uses a negative contrast medium, usually carbon dioxide, to enhance the colon. The minip software highlights the darker pixels, leaving the carbon dioxide-filled colon and

6 subtracting the rest of the tissues. A minip reconstruction also may be used for dedicated lung studies. Lung tissues fall in the darker part of the Hounsfield number spectrum because the lungs are filled with air. 19. Image Smoothing The image smoothing filter, referred to as a smoothing filter, is applied to scans that must demonstrate subtle soft tissue differences. The filter evens out areas that look blurred or have subtle changes within the shades of gray. The smoothing filter also helps improve low-contrast resolution. Lowcontrast resolution is the ability for the CT scanner to differentiate between very minor changes in shades of gray. Image smoothing takes place during the assignment of a kernel, or convolution algorithm. The use of a smoothing filter is by far the preferred technique for a soft tissue study. 20. Edge Enhancement An edge enhancement kernel, also known as an edge enhancement convolution algorithm, adds detail to the denser anatomy within a scan. This filter usually is applied to studies in which it s important to show bony structures. Kernels and algorithms are related in that the kernel can only be applied to the scanned data through a series of algorithms, or mathematical processes. Used during the convolution process, the kernel is preset according to a histogram. The kernel is what gives the scanned data a bony, a soft tissue or a cleaned-up appearance. A histogram is a curved line that indicates what the scan parameters should resemble. The histogram on the left has a narrow slope, with curvatures only at the top and bottom of the Hounsfield scale. In other words, the computer software should only use the highest and lowest Hounsfield numbers for image reconstruction. The histogram in this example indicates how an exam yielding blacks and whites should appear. The histogram on the right has a much more rounded curve overall. This means the entire Hounsfield scale should be used for image reconstruction rather than just the upper and lower Hounsfield numbers. 21. Gray-scale Manipulation Gray-scale manipulation is the process of adjusting the range of grays within an image. Grayscale manipulation also occurs when filters are applied to the scan data. The Hounsfield scale represents the given range of numbers the computer can assign to each pixel element. The numbers range from +1000, which represents white or bone, to -1000, which represents black or air. Each pixel within the CT scan is assigned a CT/Hounsfield number to give images a desired appearance. Images with a wide window width contain many shades between black and white, while those with a narrow window width have a smaller range of shades. Gray-scale manipulation is necessary because the appearance of the image can be changed drastically based on which pixels are enhanced and which pixels are minimized. 22. Gray-scale Manipulation Window Width The window width (WW) is the total number of CT/Hounsfield numbers available for any given exam. Window width typically is described as either narrow or wide. A narrow WW is associated with scans requiring fewer shades of gray. Each pixel within the image is assigned a different shade of gray than the pixel next to it. This is especially important when imaging anatomy like the brain because the densities within the brain are so similar. Gray and white matters differ from one another by only a few CT/Hounsfield numbers. A wide WW is used when an exam needs many shades of gray to adequately show the changes in the different tissues.

7 23. Gray-scale Manipulation Window Level Another important aspect of gray-scale manipulation is the window level (WL). The window level is the midpoint number in the CT/Hounsfield range for any given exam. For example, the window level for a CT/Hounsfield range of 0 to 100 is 50. The appearance of an image is affected in part by the window width but also by the WL, or midpoint of the range. 24. Gray-scale Manipulation WW/WL Equation The gray-scale range is determined using the window width/window level equation. Let s look at an example. Assume that the scan parameters are set to a window width of +800 and a window level, or midpoint, of The first step in this equation is to take the WW number and divide it in half. In this example, we divide the WW of +800 by 2, which equals Next, the new WW number is both added and subtracted from the window level value. In this case, we add +400 to +200, which equals +600, and we subtract +400 from +200, which equals So, for this example, the entire range for a window width of +800 and a window level of +200 is -200 to Gray-scale Manipulation WW/WL Equation Let s look at one more example of the WW/WL equation. What is the range of CT/Hounsfield numbers if a window width of +800 is set, but a different window level, or midpoint number, of +400 is used? As you remember, the first step of the equation is to take the given WW number and divide it in half. So, for this equation, +800 is divided by 2. Next, the new WW number is both added and subtracted from the WL value. So for this example, the entire range for a window width of +800 and a window level of +400 is 0 to Gray-scale Manipulation Grayer shades usually fall in the -200 to +200 area of the CT/Hounsfield number range. Blacker shades typically start around the -100 area of the scale. The window width in the first example included CT/Hounsfield numbers from white shades to black shades, so the resulting image could correctly display more variations in tissue densities. The window width ranges in each of the previous examples show their relationship to the CT/Hounsfield range. If a window width range falls higher up on the CT /Hounsfield scale, the resulting image will appear brighter, and vice versa. Radiologists use the window width range to determine how the CT/Hounsfield numbers are assigned for the scan, and thus demonstrate what pathology is present. 27. Shaded Surface Rendering Shaded surface rendering is used to create the illusion of three dimensions on a twodimensional computer screen. Shaded surface rendering uses 10% of the total data set. For this type of reconstruction, the data set should be acquired using thin slices, or a pitch of less than 1. Shaded surface rendering gives shape and depth to the contour or periphery of an image for a more life-like appearance. In effect, the technique manipulates the position of a simulated light source so that some regions of the image are illuminated while other regions are in shadow. The voxels farthest from the area of interest are darkened to create the illusion of roundness, shape and texture. By moving the light source in 90 increments around the object of interest, different parts of anatomy appear more detailed. Shaded surface display, or SSD, is the end result of shaded surface rendering. Often the terms are used interchangeably, but shaded surface rendering involves the actual creation of the shaded surface, and shaded surface display consists of the presentation of the image on the 2-D computer screen.

8 28. Multiplanar Reconstruction Multiplanar reconstruction is performed after the patient has left the department. The scan data is manipulated in different ways to enhance certain characteristics. These reconstructions do not add any information to the data set, but rather take the information already acquired and present the data in alternative ways. Multiplanar reconstructions use 10% of the image data, changing the image from an axial projection to a coronal, sagittal or oblique projection. To perform the reconstruction, the computer software rearranges the pixels and changes the pattern of image slices. 29. Curved Multiplanar Reconstruction Like multiplanar reconstruction, curved multiplanar reconstruction, or curved MPR, is a postprocessing technique that allows the viewer to look at images in coronal, sagittal or oblique projections. The technique works in the same way as traditional multiplanar reconstruction except that the edges of the reconstructed images can be curved or irregular rather than just straight. The only difference between traditional and curved MPR is that for traditional MPR, the computer selects the direction of the reconstruction based on the desired viewing position. In a curved MPR, the viewer sets the boundaries for reconstruction. Only 10% of the axial data set is used for a curved MPR. In the example on this page, the broad black lines show where the viewer wants to begin the image reconstruction. The computer then determines the voxels that will be used for the reconstruction process. 30. Volume Rendering Volume rendering is a complex technique used to create images that appear three-dimensional. Reconstructions can be adjusted to show the surface of an object or deep into the most internal structures. This rendering method uses 100% of the data set, so more information is available for reconstruction and the cost of the computer software is higher. Volume rendering sometimes is referred to as a four-dimensional, or 4-D, reconstruction in that the added pixels give the illusion of a transparent layer on top of the image. In volume rendering, the voxel elements go through the process of segmentation, much like in shaded surface rendering. Segmentation involves creating a boundary of Hounsfield numbers in which the anatomy of interest is highlighted. In addition to a number boundary applied to the voxels, there s an additional pixel element added to the structures demonstrated on the scan. These additional pixels give the illusion of partial visibility, similar to a soapy film left on a dish. The film is visible, but everything under the film also is visible. The image has the appearance of a shaded surface rendered image, except that the light source used for the periphery of the image is used for the rest of the image. In other words, the viewer is able to remove the outermost surface of the image, revealing the internal anatomy. The volume rendering technique continues to use the light source calculations to add depth to the image, as though the viewer is looking at the image head-on D Reconstruction Like other postprocessing reconstruction techniques, 3-D is used to give the acquired images a more life-like appearance. The 3-D technique requires a data set acquired in thin, overlapping slices, although the computer software only uses 10% of the data set. The software takes each pixel and creates what looks like coronal, sagittal and oblique projections in other words, multiplanar reconstructions. The axial data set is simply restacked so that the projection of interest determines the direction the pixel elements are viewed. The 3-D software either enhances the brightest pixels while minimizing

9 or removing the darker pixels, as in maximum intensity projections, or enhances the darker pixels and removes the brighter shades, as in minimum intensity projections. With 3-D viewing, the radiologist and other medical professionals can better understand the relationship of anatomical structures and can locate pathology or trauma more precisely. 32. Virtual Reality Virtual reality, or VR, imaging is a postprocessing application used to demonstrate what the inside of structures look like. The abbreviation VR can sometimes be mistaken for the term volume rendering, but the two techniques are not identical. Essentially, VR software allows the viewer to travel through the inside, or lumen, of a tubular structure, as if the patient had undergone an actual procedure such as an endoscopy or colonoscopy. Virtual reality imaging uses 100% of the axial data set to create images. The data set should be acquired using thin slices and a pitch of less than 1. When there s greater slice overlap, more data is available for reconstruction, although this comes at the cost of increased radiation exposure. The viewing route can be created automatically by a set of established protocols or drawn on top of the image by the user. For an automatic reconstruction, the computer identifies a reference point within the lumen to start the process. For example, if a colon is visualized, the computer singles out the Hounsfield number assigned to the lumen s opening. The software calculates a set of reference numbers and then uses all numbers that resemble the reference set to create the reconstruction. Any numbers that vary greatly from the reference numbers are excluded from the reconstruction, or may be used as the wall of the structure. The viewer also can hand draw the reconstruction pathway while viewing the axial data set in a multiplanar reconstruction format. After the images are displayed in the coronal, sagittal and axial planes, the viewer traces the target line onto the images with the mouse cursor. The MPR images allow the viewer to trace just the tube-like structure as it appears. After the viewer has outlined the structure, the computer includes only the Hounsfield numbers that are close in value to the drawn line. Hounsfield numbers that vary greatly from the viewer s handdrawn line are excluded from the reconstruction. These numbers can be included as the wall of the lumen, or if many are present, may be shown as an abnormal mass or pathology within the lumen. 33. Practice Question 34. Practice Question 35. Image Display, Manipulation, Recording and Archiving As you ve just seen, image reconstruction and postprocessing are two very important parts of CT imaging. The processes we ve looked at so far affect the quality and integrity of the image data. However, the CT technologist has other tools that play an equally important role in creating a quality image for interpretation. The final section of this module explores the various tools, viewing modes, functions and applications of the computed tomography workstation. If you re familiar with the workstation functions of cassette-less and cassette-based imaging systems in diagnostic radiology, you might recognize some of these functions. Many of these tools are similar to the those used by technologists who perform digital imaging.

10 36. Cathode Ray Tube Regardless of what type of CT scanner or processing software your department uses, the final images are displayed on either a cathode ray tube, or CRT, or liquid crystal display monitor. A CRT monitor contains a cathode gun that interprets the densities within the scan and fires a stream of electrons at a phosphor screen. The intensity of the electron beam varies based on the densities of the scan. The electrons strike various points on the phosphor screen at different intensities. When the electrons hit the screen, the screen s phosphor crystals emit light. The brightness of the light produced by the crystals represents the densities in the scan. The Hounsfield scale guides the crystals brightness. 37. Liquid Crystal Display In a liquid crystal display, or LCD, a liquid crystal matrix is placed between electrodes. When voltage is applied to the electrodes, they change the brightness of each unit within the liquid crystal matrix. The electrodes alter their intensity based on the voltage signal received, and the voltage signal changes depending on the densities present within the scan. 38. Pan Image pan is a display tool that allows the user to move the image on the display field in any direction desired. The viewer uses the mouse to electronically select the image and move it in the desired direction. Panning is an especially useful tool when the image is enlarged using the zoom tool. When the image is enlarged, only a small section can be seen on the monitor. The pan tool allows the user to move the image around until the desired anatomy is displayed on the monitor. Pan the image on this page by clicking on it with the left mouse button. While holding this button down, move the mouse in any direction and the image will move in that direction. 39. Zoom Viewers use the zoom tool to increase the size of an image. Zoom typically enlarges the image but doesn t add information to the scan. The pixels in the image appear larger because the image is brought closer to the viewer s perspective. The image matrix doesn t change either, but image quality may be degraded if the viewer zooms too much. As the pixels within the image are viewed more closely, small details are lost. Although zooming the image may decrease the resolution, smaller anatomical structures are easier to measure, and sometimes it s easier to relate anatomy to surrounding tissues. Use the slider bar on the side of the image to see how increasing or decreasing the zoom setting affects the image appearance. 40. Image Scrolling The image scrolling tool typically is controlled via the computer mouse. The viewer rolls a raised wheel in the middle of the mouse to move the images in sequential order. When the viewer places the mouse cursor over the dog-ear at one of the corners of the data set, the images can be viewed from beginning to end. Image scrolling gives viewers a fast way to flip forward and backward through the image slices so that the data set can be interpreted as a whole. Image scrolling also allows the CT technologist to see where to set the parameters for postprocessing reconstructions.

11 41. Swivel The swivel tool rotates an image to the left or right, as seen by the viewer. This action is similar to shaking your head when saying no. The swivel tool is especially useful when viewing images that have been loaded into a 3-dimensional software program. 42. Roll Rolling is similar to moving an image head over heels. This tool is used to show different parts of anatomy and is especially useful in viewing reconstructed 3-D images. Use the slider bar to adjust the image. 43. Rotate The rotate tool allows the image to be moved in any direction. The viewer can work with the image as if holding the anatomy in his or her hand. The ability to rotate anatomy allows the user to view an area of interest in greater detail and at all angles. The rotate tool is used most often with a volumerendered data set. Volume rendering creates a cube-like display of individual pixels. The cubed pixels, now called voxels, are used to stack the image slices, so that the image no longer looks flat, but has a more textured appearance. Rotation is used for a variety of exams, but is seen most frequently in head-over-heels and left-to-right spin reconstructions for CT angiography. 44. Measurement The measurement tool calculates the size of abnormal pathologies seen within the anatomy. The tool usually measures distance and diameter. The radiologist uses the distance measurement tool to determine the length of an object or how close an object is to surrounding tissues. The diameter measurement tool is used to calculate an object s circumference. Measurements typically are made during postprocessing, but measurements also can be compared from one scan to the next. Examinations that use measurement tools include chest scans to evaluate pulmonary nodules or scans to rule out an abdominal aortic aneurysm. 45. Magnify The magnification tool is used to look more closely at an image. When the tool is activated with the mouse, a small square appears over the image. The part of the image inside the square appears closer than the surrounding anatomy. This tool is similar to zoom, but instead of the entire image being brought closer to the viewer and appearing larger, only the portion under the magnification square appears bigger. Click on the image to activate the magnify tool. 46. Viewing Modes The CT technologist s workstation has several types of viewing modes. Depending on their needs, radiologists or technologists can select two-dimensional, three-dimensional, slab, planar or cine viewing modes D and 3-D Two-dimensional, or 2-D, refers to an object displayed as a horizontal and vertical representation. Two-dimensional images have no depth or roundness, but appear flat. The ability to view objects in 2-D is important because axial scanning acquires and displays image data in 2-D. CT image data acquired by helical scanning can be reconstructed in other ways.

12 The ability to collect volumetric data led to the development of 3-D reconstructions. To create 3- D images, the computer software changes each pixel element into a cube. The cubed pixel is referred to as a voxel. 48. Slab A slab is similar to a multiplanar reconstruction in that the voxels of each image slice are stacked in a desired order based on the anatomy of interest. A slab takes a chunk of the scan and creates coronal, sagittal or oblique perspectives. The viewer can manipulate the direction of the slab by changing the position of the voxel elements. Individual voxel elements are selected as a reference or starting line. Simply reassigning or changing the reference line moves or changes the direction of the slab. 49. Planar Planar refers to the point of view of the data set. Images appear in different planes depending on the anatomy of interest. To produce an image in the axial plane, spirally acquired data first must undergo 180 linear interpolation and multiplanar reconstruction. During interpolation, the images are converted from a spiral, skewed appearance, to flat individual slices of data. Multiplanar reconstructions change the direction of the data set to present a coronal, sagittal or oblique projection. If a coronal reconstruction is needed, the voxels from each slice are stacked on top of each other, starting with the first horizontal row of slice one, the first horizontal row of slice two and so on. The first row of all the slices are then stacked on each other, producing an image seen from front to back. This same process occurs for images seen in a sagittal projection, in that the entire vertical row of voxels in slice one is taken and stacked on the vertical row of voxels in slice two and so on. Click on the images to view the different planes. 50. Cine Cine, or loop viewing, refers to the ability of the viewer to watch the series of images move sequentially from the beginning to the end of the data set. This type of viewing mode typically is created during postprocessing, most often using 3-D reconstruction. A cine film usually requires the entire data set, which entails the use of an advanced computing system at an increased cost. 51. Technologist Workstation The technologist workstation is the control center of the entire CT scanning process. Besides controlling data acquisition, image display and specific reconstruction software, the workstation performs the administrative duties associated with handling patient and image data. 52. Directory or Patient List Computed tomography uses a type of microprocessor known as an array processor. The array processor is responsible for receiving image data and performing very high-speed calculations. It divides up the various imaging tasks, such as convolution and planar reconstruction, and performs these multiple tasks simultaneously. The array processor typically works using parallel processing so that multiple tasks are performed at the same time. When parallel processing involves many tasks at once, the processor is usually referred to as a large array parallel processor, or LAPP. The directory, or patient list, is a register of patients who are scheduled for an exam in the radiology department. The list is generated from the radiology information system, or RIS. When patients arrive at a hospital, their information typically is entered into the hospital information system,

13 or HIS, database. Radiology examination requests are entered into an ordering system and automatically routed to the RIS. The RIS and HIS are integrated so that care is not delayed, and the patient data is identical in both databases. The RIS assigns each exam a unique number, similar to a bar code, which typically is referred to as an accession number. Accession numbers can be filtered according to each modality in the radiology department. 53. Film/Print Radiology images can be reproduced on a sheet of film, a process that gradually is becoming outdated in the radiology department. These films are hung on the view box for the radiologist to interpret. Images also can be printed as a hard copy using a laser printer. The resolution required for printing images must be of diagnostic quality so that critical information is not missed during interpretation. Resolution typically is given in line pairs per millimeter. The more line pairs per millimeter, or information within each millimeter of film, the more information seen on the image. 54. Delete The CT technologist can remove unwanted images, text or incorrect information from the examination; however, it s important to be careful and certain of any decision to delete information. This is especially true if the raw data has been removed from the database. The only way to regain the deleted data is if, and only if, the raw data is still on the scanner. 55. Archive Button Archiving is the process of saving examinations to a database so that images can be recalled when needed. Archiving images is essential, and an archiving system should be set up in such a way that the images are easy to retrieve and have the same quality as the initial images. Archived information should be stored in a separate location such as a backup server so that images can be used for comparison at a later time. 56. Query The query process is based on a question-and-answer format. The user requests specific information from available databases and receives the answer via the computer. Another example of the query process is when the computer requests patient information updates from the radiology information system database. This query can be set up automatically, or the user can make updating the work list a selectable option. 57. Send/Network The CT technologist selects the send option after completing the exam. The technologist chooses the appropriate image data sets and then exports them to the desired location. 58. Copy/CD A copy is an exact duplicate of the original image data. Copies allow more than one person to have the images. It s important to remember that a copy can become degraded if it s created from a duplicate image rather than the original data set. As copies are made from other duplicates, image data can become degraded, and the image detail will be poor. A copy of an examination typically is sent to a storage area in a digital format, although the digital archiving of a scan is not a true copy. Copies of the scan can then be created at a later time using

14 a digital storage medium such as a compact disc, DVD or optical disc. We ll cover optical discs and DVDs later in this module. Compact discs are commonly used to archive or copy medical images. A CD is a small, relatively inexpensive optical disc that has been digitally imprinted with the scan information. To view the images, an optical scanning device reads and displays the digital information. CDs can store and retain several exams for extended periods of time. 59. Online Help Equipment manufacturers often provide an online help service. This assistance can be interactive; that is, the CT technologist can submit questions to a knowledgeable vendor s representative at an offsite location. Or, help may be available through a manufacturer s Web site. A dedicated Web site often contains frequently asked questions, also known as FAQs, or short online tutorials that provide a step-by-step guide for users. 60. Workstation Applications Workstations can be set up to perform specific functions depending on their location in the hospital. Special software for various CT applications that perform highly specialized functions may be installed on a specific scanner. This page lists some of the major applications available on CT scanners. We ll briefly discuss a few of these different uses; however the final module in this series examines the applications in greater detail. 61. Ejection Fraction Ejection fraction is an advanced coronary application used to determine how much blood is pumped from chamber to chamber. Ejection fraction also may be used to evaluate blood flow volume from each side of the heart. The heart scan is performed using gating software that tracks the patient s heartbeat during the exam. Scanning takes place during the resting portion of the patient s heartbeat. The gating software tracks the heartbeat on what looks like an electrocardiogram strip, and then the scanner acquires the images when the patient s heart is at rest. After the examination, the scanner creates a postprocessing reconstruction of the heart cycle. These reconstructions, when loaded into the appropriate software, give the illusion of a beating heart. The images help determine which phase of the heartbeat is not pumping sufficient blood. A color map is applied to the heart reconstruction to indicate deficient function at certain times during the heart cycle. 62. Calcium Scoring Calcium scoring is an advanced cardiac CT application used to calculate the amount of calcium in the coronary arteries. It is performed retrospectively, after the patient has left the department. The software assigns a preset Hounsfield number to calcium and then evaluates the areas that appear like calcium buildup. Calcium possesses a range of numbers at the upper end of the Hounsfield unit scale. Because calcium is one of the building blocks for bone, the range is near +1000, or the Hounsfield number for bone. Next, the software calculates the amount of calcium in the artery based on the given threshold. This number is subtracted from the remainder of the coronary arteries. Calcium scoring is important because calcium buildup leads to a lack of blood flow within the arteries. Depending on the total amount of calcium blocking the arteries, the radiologist can determine whether the patient has, or is at risk for, developing arteriosclerosis.

15 63. Stereotaxis Stereotaxis is a tool used for CT-guided biopsies. Stereotactic CT equipment consists of a CT scanner with real-time imaging, which means that images are seen at virtually the same time they are acquired. Real-time imaging often is referred to as CT fluoroscopy. The images are similar to conventional spiral CT images; however, there are differences in how many images are acquired at one time and where the patient is during the procedure. The x-ray tube and detector array rotate around the patient in a 360 path, and an image is produced for each rotation. Instead of acquiring as many slices as there are rows of detectors, the detector array produces a single image, and the patient stays in exactly the same position for the entire procedure. In conventional CT, the table moves at the speed needed to gather the images in relation to the rotation of the x-ray tube and detector array. During CT fluoroscopy, the images are acquired in one area only, and a single image is taken from each 60 portion of the 360 rotation. So, a total of six single images in the same position are acquired for each complete rotation. The series of images looks like a live action sequence that allows the radiologist to see the exact location of the biopsy. In addition, the patient remains stationary within the gantry because the position for the very first image is used to gauge exactly where the biopsy will take place. The process of image reconstruction varies significantly in CT fluoroscopy. An interpolation algorithm isn t required because image acquisition is similar to a single-slice scanner. The x-ray tube and detectors don t follow a helical path, and the patient doesn t move through the CT gantry. The images are reconstructed using the initial 360 image as a point of reference. This first image contains the full amount of scan data from an entire rotation. The next image is reconstructed from the first image based on the complete data set collected in one full rotation. The second image acquired uses only 60 of the 360 rotation. The rest of the information pertinent to the image is averaged based on what the initial image reveals. After the first 60 image, the second 60 image is acquired at the next rotation increment. Reconstruction occurs by subtracting the first 60 image data from the new data of the second image. The CT technologist then scrolls through the image data set, allowing the radiologist to verify where the biopsy will take place. When the exact location is determined, the technologist sets the biopsy position to zero so that if the scanner needs to be repositioned, the zero setting will move the patient back to the selected spot. The radiologist places a stereotactic tool on the patient and determines what anatomy lies beneath the device. The stereotactic tool transmits an exact coordinate of its position to the computer. The computer then uses the initial, full scan of the area of interest to create an image of what that anatomy will look like with a biopsy needle. 64. Radiation Oncology Treatment Planning CT imaging is used to help plan where radiation is targeted during treatment. The patient is positioned for the CT scan in exactly the same position as the set up for the radiation therapy table. Occasionally the patient s skin is marked to indicate the treatment set up, and sometimes the radiation therapy immobilization masks or fiducial markers are used for the CT scan to ensure reproducible positioning when the patient undergoes treatment. After the CT scan is performed, the image data are loaded into the treatment planning software. The CT scan is used to define and measure the radiation treatment volume, with the treatment plan superimposed on the CT scan. The CT scan also helps to define beam parameters and radiation dose distribution.

16 65. Fusion Fusion is a technique that retrospectively combines images from two modalities. Positron emission tomography and computed tomography, or PET-CT, are commonly merged as a fusion study. The first step is to acquire the images from both modalities. For PET-CT, the radiopharmaceutical is administered first and then the CT scan is performed, with or without IV contrast, followed by PET imaging. Although images can be acquired independently by a CT scanner and nuclear medicine camera and then fused, hybrid machines are now manufactured that can perform both imaging methods. During the examination, it s important for the patient to maintain the same position for both the CT scan and PET imaging. If the patient moves out of position, it will be difficult to correctly align the anatomy in the CT and nuclear medicine images. This is the reason patients usually don t hold their breath during the CT scan. The nuclear medicine scan takes approximately 15 minutes to complete, and it would be impossible for patients to hold their breath for that length of time. After image acquisition, the technologist uses a computer software program to perform retrospective reconstructions, loading the data from the CT scan first and then placing the PET images over the CT scan. 66. Neurology Neurology applications are used for many different reasons, including fracture assessment, cerebral artery evaluation and assessment of brain damage and recovery potential. Volume rendering and advanced subtraction methods are two of the techniques used in neurology applications to demonstrate anatomy. Volume rendering creates images that appear three-dimensional and can show anatomy from the periphery of an object down to the most internal structures. For example, volume rendering provides a clear representation of the skull to evaluate fractures. Subtraction methods typically are used for evaluating the arteries of the brain. Subtraction works by determining the change in two separate data sets acquired at different times. The change usually is demonstrated using IV contrast. For example, the CT technologist performs a head scan without a contrast agent and then performs the same scan using contrast. The computer software removes all the identical Hounsfield numbers in the two studies, leaving the remaining, different numbers to be displayed. Xenon studies are performed to determine if the patient is receiving sufficient blood flow to the brain. After a baseline CT scan of the head is performed, the patient inhales xenon gas. The computer software calculates the Hounsfield numbers before the xenon gas was inhaled, the Hounsfield numbers during inhalation and finally the Hounsfield numbers at the washout phase. The washout phase occurs when the xenon gas leaves the brain and the original Hounsfield numbers begin to reappear. 67. Vessel Analysis A vessel often may be obscured by the anatomy around it, especially in the cranial bones. Vessel analysis helps demonstrate the exact location of a vessel, while excluding all nonessential anatomy. The technique works by determining the threshold number for the vessel. Vessels have a certain range of Hounsfield numbers, and based on those numbers, the computer software can remove, or suppress, the numbers for all other tissues, leaving the vessels of interest visible. Vessel analysis simply gives the computer software a guideline of which Hounsfield numbers it should keep and which numbers it should discard.

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

Medical Image Processing: Image Reconstruction and 3D Renderings

Medical Image Processing: Image Reconstruction and 3D Renderings Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,

More information

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed

More information

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic

More information

Fundamentals of CT imaging

Fundamentals of CT imaging SECTION 1 Fundamentals of CT imaging I History In the early 1970s Sir Godfrey Hounsfield s research produced the first clinically useful CT scans. Original scanners took approximately 6 minutes to perform

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead.

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead. 1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

FINDING THE TRUE EDGE IN CTA

FINDING THE TRUE EDGE IN CTA FINDING THE TRUE EDGE IN CTA by: John A. Rumberger, PhD, MD, FACC Your patient has chest pain. The Cardiac CT Angiography shows plaque in the LAD. You adjust the viewing window trying to evaluate the stenosis

More information

icatvision Quick Reference

icatvision Quick Reference icatvision Quick Reference Navigating the i-cat Interface This guide shows how to: View reconstructed images Use main features and tools to optimize an image. REMINDER Images are displayed as if you are

More information

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1 Modifications for P551 Fall 2013 Medical Physics Laboratory Introduction Following the introductory lab 0, this lab exercise the student through

More information

RADIOLOGY AND DIAGNOSTIC IMAGING

RADIOLOGY AND DIAGNOSTIC IMAGING Day 2 part 2 RADIOLOGY AND DIAGNOSTIC IMAGING Dr hab. Zbigniew Serafin, MD, PhD serafin@cm.umk.pl 2 3 4 5 CT technique CT technique 6 CT system Kanal K: RSNA/AAPM web module: CT Systems & CT Image Quality

More information

How to Adjust the 16-Bit CLUT (Color Look Up Table) Editor in OsiriX

How to Adjust the 16-Bit CLUT (Color Look Up Table) Editor in OsiriX How to Adjust the 16-Bit CLUT (Color Look Up Table) Editor in OsiriX 1. Import the series you are interested in into the 2D/3D Viewer. I m using the OBELIX data set, slices 1-395, available for download

More information

Tomographic Reconstruction

Tomographic Reconstruction Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)

More information

VieW 3D. 3D Post-Processing WorKstation THE THIRD DIMENSION. Version 3.1

VieW 3D. 3D Post-Processing WorKstation THE THIRD DIMENSION. Version 3.1 VieW 3D 3D Post-Processing WorKstation THE THIRD DIMENSION Version 3.1 iq-view 3D THE FULLY-FEATURED 3D MEDICAL IMAGING SOLUTION FOR RADIOLOGISTS iq-view 3D contains all components of iq-view with the

More information

4DM Packages. 4DM Packages & License Types. Information to help you order the appropriate licenses for your site.

4DM Packages. 4DM Packages & License Types. Information to help you order the appropriate licenses for your site. 4DM Packages 4DM Packages & License Types. Information to help you order the appropriate licenses for your site. Nuclear Cardiac Quantification, Review, and Reporting Select Your 4DM Package and corresponding

More information

Graphics Hardware and Display Devices

Graphics Hardware and Display Devices Graphics Hardware and Display Devices CSE328 Lectures Graphics/Visualization Hardware Many graphics/visualization algorithms can be implemented efficiently and inexpensively in hardware Facilitates interactive

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

Modifications for P551 Fall 2014

Modifications for P551 Fall 2014 LAB DEMONSTRATION COMPUTED TOMOGRAPHY USING DESKCAT 1 Modifications for P551 Fall 2014 Introduction This lab demonstration explores the physics and technology of Computed Tomography (CT) and guides the

More information

RSNA, /rg

RSNA, /rg RSNA, 2015 10.1148/rg.2015140320 Appendix As noted in the main article, DICOM image files cannot be used directly for 3D printing; further steps are necessary to make them readable by 3D printers. The

More information

LAB DEMONSTRATION COMPUTED TOMOGRAPHY USING DESKCAT Lab Manual: 0

LAB DEMONSTRATION COMPUTED TOMOGRAPHY USING DESKCAT Lab Manual: 0 LAB DEMONSTRATION COMPUTED TOMOGRAPHY USING DESKCAT Lab Manual: 0 Introduction This lab demonstration explores the physics and technology of Computed Tomography (CT) and guides the student and instructor

More information

Copyright 2017 Medical IP - Tutorial Medip v /2018, Revision

Copyright 2017 Medical IP - Tutorial Medip v /2018, Revision Copyright 2017 Medical IP - Tutorial Medip v.1.0.0.9 01/2018, Revision 1.0.0.2 List of Contents 1. Introduction......................................................... 2 2. Overview..............................................................

More information

RT_Image v0.2β User s Guide

RT_Image v0.2β User s Guide RT_Image v0.2β User s Guide RT_Image is a three-dimensional image display and analysis suite developed in IDL (ITT, Boulder, CO). It offers a range of flexible tools for the visualization and quantitation

More information

Enhanced material contrast by dual-energy microct imaging

Enhanced material contrast by dual-energy microct imaging Enhanced material contrast by dual-energy microct imaging Method note Page 1 of 12 2 Method note: Dual-energy microct analysis 1. Introduction 1.1. The basis for dual energy imaging Micro-computed tomography

More information

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA Computed Tomography Principles, Design, Artifacts, and Recent Advances THIRD EDITION Jiang Hsieh SPIE PRESS Bellingham, Washington USA Table of Contents Preface Nomenclature and Abbreviations xi xv 1 Introduction

More information

A closer look at CT scanning

A closer look at CT scanning Vet Times The website for the veterinary profession https://www.vettimes.co.uk A closer look at CT scanning Author : Charissa Lee, Natalie Webster Categories : General, Vets Date : April 3, 2017 A basic

More information

3DMMVR REFERENCE MANUAL V 0.81

3DMMVR REFERENCE MANUAL V 0.81 3DMMVR REFERENCE MANUAL V 0.81 Page 1 of 30 Index: 1.0 System Requirements...5 1.1 System Processor...5 1.2 System RAM...5 1.3 Graphics Card...5 1.4 Operating System...5 2.0 Conventions...6 2.1 Typographic

More information

Instructions for Display of Imaging Studies Using Stentor isite

Instructions for Display of Imaging Studies Using Stentor isite Instructions for Display of Imaging Studies Using Stentor isite Open http://isite.rad.tju.edu in Internet Explorer Log-In: Enter your user name (case Sensitive) and password (case sensitive) and press

More information

Brilliance CT Big Bore.

Brilliance CT Big Bore. 1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.

More information

Lecture 4 Image Enhancement in Spatial Domain

Lecture 4 Image Enhancement in Spatial Domain Digital Image Processing Lecture 4 Image Enhancement in Spatial Domain Fall 2010 2 domains Spatial Domain : (image plane) Techniques are based on direct manipulation of pixels in an image Frequency Domain

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

Voxar 3D ColonMetrix. Reference Guide

Voxar 3D ColonMetrix. Reference Guide Voxar 3D ColonMetrix Reference Guide The software described in this document is furnished under a license, and may be used or copied only according to the terms of such license. Toshiba means, Toshiba

More information

An Intuitive Explanation of Fourier Theory

An Intuitive Explanation of Fourier Theory An Intuitive Explanation of Fourier Theory Steven Lehar slehar@cns.bu.edu Fourier theory is pretty complicated mathematically. But there are some beautifully simple holistic concepts behind Fourier theory

More information

Gamepad Controls. Figure 1: A diagram of an Xbox controller. Figure 2: A screenshot of the BodyViz Controller Panel. BodyViz 3 User Manual 1

Gamepad Controls. Figure 1: A diagram of an Xbox controller. Figure 2: A screenshot of the BodyViz Controller Panel. BodyViz 3 User Manual 1 BodyViz User Manual Gamepad Controls The first step in becoming an expert BodyViz user is to get acquainted with the Xbox gamepad, also known as a controller, and the BodyViz Controller Panel. These can

More information

Using TeraRecon intuition Viewer with STAT Quick Reference Guide

Using TeraRecon intuition Viewer with STAT Quick Reference Guide Using TeraRecon intuition Viewer with STAT Quick Reference Guide 1. Launching intuition 2. 3D Assessment of Coronary Arteries for Stenosis 3. Time Volume Analysis (TVA) for Determination of Left Ventricular

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

Computational Medical Imaging Analysis Chapter 4: Image Visualization

Computational Medical Imaging Analysis Chapter 4: Image Visualization Computational Medical Imaging Analysis Chapter 4: Image Visualization Jun Zhang Laboratory for Computational Medical Imaging & Data Analysis Department of Computer Science University of Kentucky Lexington,

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

Prostate Detection Using Principal Component Analysis

Prostate Detection Using Principal Component Analysis Prostate Detection Using Principal Component Analysis Aamir Virani (avirani@stanford.edu) CS 229 Machine Learning Stanford University 16 December 2005 Introduction During the past two decades, computed

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

AUTOMATED 4 AXIS ADAYfIVE SCANNING WITH THE DIGIBOTICS LASER DIGITIZER

AUTOMATED 4 AXIS ADAYfIVE SCANNING WITH THE DIGIBOTICS LASER DIGITIZER AUTOMATED 4 AXIS ADAYfIVE SCANNING WITH THE DIGIBOTICS LASER DIGITIZER INTRODUCTION The DIGIBOT 3D Laser Digitizer is a high performance 3D input device which combines laser ranging technology, personal

More information

Medical Images Analysis and Processing

Medical Images Analysis and Processing Medical Images Analysis and Processing - 25642 Emad Course Introduction Course Information: Type: Graduated Credits: 3 Prerequisites: Digital Image Processing Course Introduction Reference(s): Insight

More information

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology Diagnostic imaging techniques Krasznai Zoltán University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology 1. Computer tomography (CT) 2. Gamma camera 3. Single Photon

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

Advances in MicroStation 3D

Advances in MicroStation 3D MW1HC515 Advances in MicroStation 3D Hands-on class sponsored by the Bentley Institute Presenter: Sam Hendrick, Senior MicroStation Product Consultant Bentley Systems, Incorporated 685 Stockton Drive Exton,

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

MR Advance Techniques. Vascular Imaging. Class III

MR Advance Techniques. Vascular Imaging. Class III MR Advance Techniques Vascular Imaging Class III 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology

More information

Medical Image Analysis

Medical Image Analysis Computer assisted Image Analysis VT04 29 april 2004 Medical Image Analysis Lecture 10 (part 1) Xavier Tizon Medical Image Processing Medical imaging modalities XRay,, CT Ultrasound MRI PET, SPECT Generic

More information

Equipment Specification

Equipment Specification MULTISLICE CT SCANNER ( 16 Slices ) Merk : Hitachi Japan Model : SUPRIA 5 MHU Price : Rp 6.512.360.215,27 No Equipment Specification 1 2 3 Scanner Gantry Object for scanning : Whole body including head

More information

New Technology Allows Multiple Image Contrasts in a Single Scan

New Technology Allows Multiple Image Contrasts in a Single Scan These images were acquired with an investigational device. PD T2 T2 FLAIR T1 MAP T1 FLAIR PSIR T1 New Technology Allows Multiple Image Contrasts in a Single Scan MR exams can be time consuming. A typical

More information

Computer-Tomography II: Image reconstruction and applications

Computer-Tomography II: Image reconstruction and applications Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de

More information

3D Design with 123D Design

3D Design with 123D Design 3D Design with 123D Design Introduction: 3D Design involves thinking and creating in 3 dimensions. x, y and z axis Working with 123D Design 123D Design is a 3D design software package from Autodesk. A

More information

Medical Imaging BMEN Spring 2016

Medical Imaging BMEN Spring 2016 Name Medical Imaging BMEN 420-501 Spring 2016 Homework #4 and Nuclear Medicine Notes All questions are from the introductory Powerpoint (based on Chapter 7) and text Medical Imaging Signals and Systems,

More information

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging

More information

Effect of Scattering on the Image. Reducing Compton Scatter with a Grid

Effect of Scattering on the Image. Reducing Compton Scatter with a Grid Effect of Scattering on the Image Increasing Compton scattering degrades image. Webb 21 Reducing Compton Scatter with a Grid Grids Parallel (focused at infinity) Linear Focused (see figure) Moving grids

More information

Topics and things to know about them:

Topics and things to know about them: Practice Final CMSC 427 Distributed Tuesday, December 11, 2007 Review Session, Monday, December 17, 5:00pm, 4424 AV Williams Final: 10:30 AM Wednesday, December 19, 2007 General Guidelines: The final will

More information

Imaging protocols for navigated procedures

Imaging protocols for navigated procedures 9732379 G02 Rev. 1 2015-11 Imaging protocols for navigated procedures How to use this document This document contains imaging protocols for navigated cranial, DBS and stereotactic, ENT, and spine procedures

More information

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory Introduction In this lab exercise, you will investigate the linearity of the DeskCAT scanner by making measurements

More information

Computer-Tomography I: Principles, History, Technology

Computer-Tomography I: Principles, History, Technology Computer-Tomography I: Principles, History, Technology Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de History

More information

IMPAX Volume Viewing 3D Visualization & Segmentation

IMPAX Volume Viewing 3D Visualization & Segmentation Getting started guide IMPAX Volume Viewing 3D Visualization & Segmentation This guide outlines the basic steps to perform and manipulate a 3D reconstruction of volumetric image data using IMPAX Volume

More information

A Study of Medical Image Analysis System

A Study of Medical Image Analysis System Indian Journal of Science and Technology, Vol 8(25), DOI: 10.17485/ijst/2015/v8i25/80492, October 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 A Study of Medical Image Analysis System Kim Tae-Eun

More information

Constructing System Matrices for SPECT Simulations and Reconstructions

Constructing System Matrices for SPECT Simulations and Reconstructions Constructing System Matrices for SPECT Simulations and Reconstructions Nirantha Balagopal April 28th, 2017 M.S. Report The University of Arizona College of Optical Sciences 1 Acknowledgement I would like

More information

Chapter 1. Getting to Know Illustrator

Chapter 1. Getting to Know Illustrator Chapter 1 Getting to Know Illustrator Exploring the Illustrator Workspace The arrangement of windows and panels that you see on your monitor is called the workspace. The Illustrator workspace features

More information

Whole Body Submillimeter Scan

Whole Body Submillimeter Scan 128 Whole Body Submillimeter Scan The real value of 64ch/128slice CT systems does not come from the ability to perfom cardiac scans but the capability to scan all parts of the body in high definition submillimeter

More information

Physics 1CL WAVE OPTICS: INTERFERENCE AND DIFFRACTION Fall 2009

Physics 1CL WAVE OPTICS: INTERFERENCE AND DIFFRACTION Fall 2009 Introduction An important property of waves is interference. You are familiar with some simple examples of interference of sound waves. This interference effect produces positions having large amplitude

More information

Working with the BCC Z Space I Filter

Working with the BCC Z Space I Filter Working with the BCC Z Space I Filter Normally, if you create an effect with multiple DVE layers, each layer is rendered separately. The layer that is topmost in the timeline overlaps all other layers,

More information

Fluorescence Tomography Source Reconstruction and Analysis

Fluorescence Tomography Source Reconstruction and Analysis TECHNICAL NOTE Pre-clinical in vivo imaging Fluorescence Tomography Source Reconstruction and Analysis Note: This Technical Note is part of a series for Fluorescence Imaging Tomography (FLIT). The user

More information

2 Michael E. Leventon and Sarah F. F. Gibson a b c d Fig. 1. (a, b) Two MR scans of a person's knee. Both images have high resolution in-plane, but ha

2 Michael E. Leventon and Sarah F. F. Gibson a b c d Fig. 1. (a, b) Two MR scans of a person's knee. Both images have high resolution in-plane, but ha Model Generation from Multiple Volumes using Constrained Elastic SurfaceNets Michael E. Leventon and Sarah F. F. Gibson 1 MIT Artificial Intelligence Laboratory, Cambridge, MA 02139, USA leventon@ai.mit.edu

More information

Physical bases of X-ray diagnostics

Physical bases of X-ray diagnostics Physical bases of X-ray diagnostics Dr. István Voszka Possibilities of X-ray production (X-ray is produced, when charged particles of high velocity are stopped) X-ray tube: Relatively low accelerating

More information

Learning to use the drawing tools

Learning to use the drawing tools Create a blank slide This module was developed for Office 2000 and 2001, but although there are cosmetic changes in the appearance of some of the tools, the basic functionality is the same in Powerpoint

More information

Transform Introduction page 96 Spatial Transforms page 97

Transform Introduction page 96 Spatial Transforms page 97 Transform Introduction page 96 Spatial Transforms page 97 Pad page 97 Subregion page 101 Resize page 104 Shift page 109 1. Correcting Wraparound Using the Shift Tool page 109 Flip page 116 2. Flipping

More information

Selective Space Structures Manual

Selective Space Structures Manual Selective Space Structures Manual February 2017 CONTENTS 1 Contents 1 Overview and Concept 4 1.1 General Concept........................... 4 1.2 Modules................................ 6 2 The 3S Generator

More information

New Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC

New Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC New Technology in Radiation Oncology James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC Technology s s everywhere From the imaging chain To the planning system To the linac To QA..it..it

More information

Computed tomography - outline

Computed tomography - outline Computed tomography - outline Computed Tomography Systems Jørgen Arendt Jensen and Mikael Jensen (DTU Nutech) October 6, 216 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering

More information

Chapter 12 Notes: Optics

Chapter 12 Notes: Optics Chapter 12 Notes: Optics How can the paths traveled by light rays be rearranged in order to form images? In this chapter we will consider just one form of electromagnetic wave: visible light. We will be

More information

Quick Reference Guide

Quick Reference Guide Key functions Quick Reference Guide Philips isite Enterprise Version 3.6 Enhance mouse cursor mode Function Wheel mouse 2-button mouse Adjust WW/WL Left click and drag Left click and drag Scroll/Cine Wheel

More information

Dynamic Reconstruction for Coded Aperture Imaging Draft Unpublished work please do not cite or distribute.

Dynamic Reconstruction for Coded Aperture Imaging Draft Unpublished work please do not cite or distribute. Dynamic Reconstruction for Coded Aperture Imaging Draft 1.0.1 Berthold K.P. Horn 2007 September 30. Unpublished work please do not cite or distribute. The dynamic reconstruction technique makes it possible

More information

Athena Radiology Medical Workstation

Athena Radiology Medical Workstation High productivity and integration Athena DICOM Viewer is being designed according with the suggestions and the necessities of medical radiologists. From the design of its interface to the advanced image

More information

Shadows in the graphics pipeline

Shadows in the graphics pipeline Shadows in the graphics pipeline Steve Marschner Cornell University CS 569 Spring 2008, 19 February There are a number of visual cues that help let the viewer know about the 3D relationships between objects

More information

Shading Techniques Denbigh Starkey

Shading Techniques Denbigh Starkey Shading Techniques Denbigh Starkey 1. Summary of shading techniques 2 2. Lambert (flat) shading 3 3. Smooth shading and vertex normals 4 4. Gouraud shading 6 5. Phong shading 8 6. Why do Gouraud and Phong

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!

More information

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist CT Protocol Review: Practical Tips for the Imaging Physicist Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center August 8, 2013 AAPM Annual Meeting Goals Understand purpose and importance

More information

Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2)

Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2) Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2) Author: Levon Vogelsang Introduction Cone beam computed tomography (CBCT), or cone beam CT technology, offers considerable promise

More information

First CT Scanner. How it Works. Contemporary CT. Before and After CT. Computer Tomography: How It Works. Medical Imaging and Pattern Recognition

First CT Scanner. How it Works. Contemporary CT. Before and After CT. Computer Tomography: How It Works. Medical Imaging and Pattern Recognition Computer Tomography: How t Works Medical maging and Pattern Recognition Lecture 7 Computed Tomography Oleh Tretiak Only one plane is illuminated. Source-subject motion provides added information. 2 How

More information

Excel Core Certification

Excel Core Certification Microsoft Office Specialist 2010 Microsoft Excel Core Certification 2010 Lesson 6: Working with Charts Lesson Objectives This lesson introduces you to working with charts. You will look at how to create

More information

Transforming Objects and Components

Transforming Objects and Components 4 Transforming Objects and Components Arrow selection Lasso selection Paint selection Move Rotate Scale Universal Manipulator Soft Modification Show Manipulator Last tool used Figure 4.1 Maya s manipulation

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

EE795: Computer Vision and Intelligent Systems

EE795: Computer Vision and Intelligent Systems EE795: Computer Vision and Intelligent Systems Spring 2012 TTh 17:30-18:45 WRI C225 Lecture 04 130131 http://www.ee.unlv.edu/~b1morris/ecg795/ 2 Outline Review Histogram Equalization Image Filtering Linear

More information

The Wireframe Update Buttons. The Frontface and Backface Buttons. The Project Designer 265

The Wireframe Update Buttons. The Frontface and Backface Buttons. The Project Designer 265 The Wireframe Update Buttons The speed at which objects can be manipulated in the Project Designer viewport depends in part on the complexity of the object being moved. An object that is made up of many

More information

12 APPLYING EFFECTS. Lesson overview

12 APPLYING EFFECTS. Lesson overview 12 APPLYING EFFECTS Lesson overview In this lesson, you ll learn how to do the following: Use various effects, such as Pathfinder, Scribble, and Drop Shadow. Use Warp effects to distort type. Create three-dimensional

More information

Visualisation : Lecture 1. So what is visualisation? Visualisation

Visualisation : Lecture 1. So what is visualisation? Visualisation So what is visualisation? UG4 / M.Sc. Course 2006 toby.breckon@ed.ac.uk Computer Vision Lab. Institute for Perception, Action & Behaviour Introducing 1 Application of interactive 3D computer graphics to

More information

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,

More information

3 Polygonal Modeling. Getting Started with Maya 103

3 Polygonal Modeling. Getting Started with Maya 103 3 Polygonal Modeling In Maya, modeling refers to the process of creating virtual 3D surfaces for the characters and objects in the Maya scene. Surfaces play an important role in the overall Maya workflow

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

CTA HEAD Perfusion AqONE without and with IV Contrast

CTA HEAD Perfusion AqONE without and with IV Contrast CTA HEAD Perfusion AqONE without and with IV Contrast Patient Position Adult Contrast Adult Injection Rate Supine IOML perpendicular to table top. IV: 100 ml with helical head CTA 50 ml without helical

More information

Limitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods

Limitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods Stereoscopic Breast Imaging Andrew D. A. Maidment, Ph.D. Chief, Physics Section Department of Radiology University of Pennsylvania Limitations of Projection Radiography Mammography is a projection imaging

More information

Motivation. Intensity Levels

Motivation. Intensity Levels Motivation Image Intensity and Point Operations Dr. Edmund Lam Department of Electrical and Electronic Engineering The University of Hong ong A digital image is a matrix of numbers, each corresponding

More information

3D Space is one of the most rewarding areas

3D Space is one of the most rewarding areas Lesson 8 206 3D Space Add a new dimension to your animations. In This Lesson 207 enabling layers for 3D 208 moving and rotating layers in 3D space 210 multiplaning effects 212 3D motion paths 213 multiple

More information

/5 Stacks. Displays the slice that follows the currently displayed slice. As a shortcut, press the > key.

/5 Stacks. Displays the slice that follows the currently displayed slice. As a shortcut, press the > key. 20-02-2018 1/5 Stacks Stacks This submenu contains commands that work with stacks. Add Slice Inserts a blank slice after the currently displayed slice. Hold down the Alt key to add the slice before the

More information