Acknowledgments. Nesterov s Method for Accelerated Penalized-Likelihood Statistical Reconstruction for C-arm Cone-Beam CT.
|
|
- Walter Potter
- 6 years ago
- Views:
Transcription
1 June 5, Nesterov s Method for Accelerated Penalized-Likelihood Statistical Reconstruction for C-arm Cone-Beam CT Adam S. Wang, J. Webster Stayman, Yoshito Otake, Gerhard Kleinszig, Sebastian Vogt, Jeffrey H. Siewerdsen Biomedical Engineering, Johns Hopkins University XP Division, Siemens Healthcare Johns Hopkins University Schools of Medicine and Engineering Acknowledgments The I-STAR Laboratory Imaging for Surgery, Therapy, and Radiology Hopkins Collaborators Surgery Z. Gokaslan (Neuro Spine) A. Khanna (Ortho Spine) G. Gallia (Neuro Skull Base) D. Reh (Otolaryngology) Funding Support NIH Fellowship F3-EB757 AAPM Research Seed Funding Siemens Healthcare (XP Division) Johns Hopkins University Schools of Medicine and Engineering Salt Lake City, UT (June )
2 June 5, Motivation Intraoperative C-arm cone-beam CT High-precision surgical guidance Detection of complications OR quality assurance Model-based reconstruction for C-arm CBCT Model physics and statistics in imaging chain Reduce dose, noise, and artifact Extend CBCT to soft-tissue imaging Enable novel source-detector trajectories Task-driven imaging Incorporate a wealth of prior information Planning CT, previous CBCT scans CT CBCT Motivation Iterative Reconstruction Tends to be Slow Typical PL reconstruction: ~ iterations, ~3 hrs Incomplete Data Slows Convergence Further a) Cone-beam artifacts (e.g., away from central slice) b) Longitudinal truncation (e.g., 5 cm coverage) c) Incomplete orbit (e.g., ~ orbit) d) Lateral truncation (e.g., 5 cm FOV) Proposed Solution Apply momentum-based methods to accelerate reconstruction time by an order of magnitude, Utilize fast, GPU-based projectors RMSE Time (Hours) z (a) (b) Kim, Ramani, and Fessler, Fully3D 3 (c) y x (d) Salt Lake City, UT (June )
3 June 5, Anthropomorphic Head Phantom Human skull encased in tissueequivalent plastic.7 mm diameter spheres for contrast Benchtop CBCT Untruncated projections Varian 33CB detector SAD = cm, SDD = cm kvp, mas orbit, 9 projections # subsets M {, 33,, } Mobile C-arm CBCT Truncated projections Varian 33+ detector Same geometry and technique Methods Bench C-arm Penalized Likelihood Model-based image reconstruction Basic Poisson statistical model of quantum noise Edge-preserving Huber penalty = ; Bench: =, C-arm: = Methods Objective Log-likelihood Regularization =argmax{φ ;" $ ;" % } % =' ' ( ) * ) ) *, - =,, > Penalty Recon Quadratic Linear Huber -δ δ Difference Quadratic Linear Reconstructions Separable footprints () projector in CUDA... mm 3 isotropic voxels Recon FOV encompasses entire object Head: voxels Bench C-arm Salt Lake City, UT (June ) 3
4 June 5, Conventional SQS-M Initialize < For >,,3,,A iterations For B,,3,,C subsets ) DE F forward project ) $G C E H F IG likelihood gradient 3) JC E H F (K M(D++ likelihood curvature ) N% *,- G ) * reg. gradient P Q R ) * reg. curvature μ *,- 5) Δ T GUVWX YUVZ \Δ U 5 Δ PL Reconstruction compute update update image 5 μ * $G N% J μ * μ P Accelerated Reconstruction Conventional SQS-M Initialize < For >,,3,,A For B,,3,,C Compute Δ \Δ U update image Nesterov Acceleration (Nes-M+ Initialize <,`,g For >,,3,,A iterations For B,,3,,C subsets Compute Δ image update ` `\gδ accumulated updates g (\ \g +/ momentum weight j \Δ U conventional update \ < \` j U momentum image Δ ` 5 \Δ (<+ \` Salt Lake City, UT (June )
5 June 5, 5 3 Accelerated Reconstruction Momentum weight t # Subiterations Nesterov Acceleration (Nes-M) Initialize = <,`=,g= For >=,,3,,A iterations For B=,,3,,C subsets Compute Δ image update ` `+gδ accumulated updates g (+ +g )/ momentum weight j +Δ U + j < +` U conventional update momentum image Δ ` 5 +Δ (<) +` Acceleration Factor SQS- Equivalent SQS- Acceleration Factor Use objective function Φ to compare progress of algorithms A and B: min> k s.t. Φ k m n ;" Φ p m q ;" Errstsuvwxyz {vrwyu: AF > p => p /> k Baseline algorithm: SQS- (no subsets) Limit-cycle / unstable when C too large Nesterov AF increases with more iterations Faster convergence rate than SQS SQS-M Acceleration SQS- Acceleration Factor 3 Nes-M Acceleration Nes- Nes-9 Nes- Equivalent SQS- Salt Lake City, UT (June ) 5
6 June 5, Head Phantom (Benchtop Study) Minimal truncation Can use high number of subsets RMSD =. HU Nesterov uses 9.3 fewer iters Artifact from incomplete orbit slow to converge } Bench RMSD 5 ( ƒ) } ~~ SQS- SQS- RMSD RMSD =. HU ( ) } s Nes- Nes-9 Nes- Difference 5 RMSD } Almost fully truncated Slower convergence, especially outside C- arm FOV Use fewer subsets for stability Choose higher RMSD Head Phantom (C-arm Study) RMSD } ( Š ) } ~~ SQS- SQS- RMSD RMSD =. HU 9. fewer iterations (Œ ) } s Nes- Nes-9 Nes C-arm -5 Salt Lake City, UT (June )
7 June 5, Projector Selection High-Fidelity Projector Separable Footprints TT Voxel-driven Cast shadow onto detector, trapezoid basis functions Accurate (but slow) Source Faster Projectors Forward-Project (Siddon) Ray-driven Voxel weighted by intersection length with ray Voxel Siddon Detector Back-Project (Peters) 3 Voxel-driven Match voxel size (. mm ~. mag) with pixel size (.3 mm) Convolve detector with 3 3 window prior to backprojection Peters Long et al, IEEE TMI Siddon, Med Phys () 95 3 Peters et al, IEEE TNS 9 Faster Projectors -SP able to achieve same RMSD Mismatched projectors assisted by smoothing and regularization RMSD SQS- SQS- RMSD Nes- Nes-9 Nes- -SP } ( Š ) } ~~ 9. fewer iterations RMSD =. HU (Œ ) } s.5 more iterations ( Œ) } s ~ Difference Salt Lake City, UT (June ) 7
8 June 5, Reconstruction Time CUDA implementation on x GeForce GTX Titan Black (NVIDIA) data volume ( Š ) } ~~ (Œ ) } s ( Œ) } s ~ 5 5 = Separable Footprints, trapezoidal basis SP = Siddon forward-, Peters backprojector -5-5 Time/Iter (sec) # SP +7.9% +.3% -7.3% Forward Project Backproject (likelihood gradient) Backproject (likelihood curvature) Regularization Update Volume / Variables SP Reconstruction Time CUDA implementation on x GeForce GTX Titan Black (NVIDIA) data volume ( Š ) } ~~ (Œ ) } s ( Œ) } s ~ 5 5 = Separable Footprints, trapezoidal basis SP = Siddon forward-, Peters backprojector -5-5 Total Time (sec),, SP Salt Lake City, UT (June )
9 June 5, Cadaver Abdomen Realistic Soft Tissue Imaging Fresh cadaver (non-fixed) Larger recon volume ( voxels) to cover entire object kvp, mas (3. mgy) PL parameters: β =, δ = e-3 } MDCT CBCT Incomplete data slows convergence Choose higher RMSD Larger M unstable Excellent agreement in soft-tissue Some differences at high-contrast edges Cadaver Reconstructions RMSD SQS- SQS- RMSD Nes- Nes-9 Nes- -SP } (Œƒ ) } ~~ C-arm RMSD =. HU (ŒŒ) } s ( ) } s ~ 95 sec 959 sec 97 sec 5 5 Difference Salt Lake City, UT (June ) 9
10 June 5, Discussion Accelerated PL Reconstructions SQS: ~ iterations Nesterov: ~ iterations SP projectors: ~3- iterations much faster per iteration Further Acceleration Is Possible Use pre-computed curvature (save BP) Spatially non-uniform updates Relaxed momentum (increased stability) Optimized momentum method 3 Multi-GPU Speedup observed for Siddon/Peters fast enough that overhead costs may limit benefit Multi-resolution reconstruction (especially outside imaging FOV) (e.g., Abdomen: ~% of voxels outside FOV) SP Total Time (sec),, Cadaver C-arm Recon Kim et al, IEEE TMI 3 Kim and Fessler, IEEE MIC 3 3 Kim and Fessler, CT Meeting Conclusions Nesterov-Accelerated PL Reconstruction Helps overcome slower convergence of incomplete data Nesterov s method ~ speedup Fast projectors same image quality with ~5 speedup ~-3 minute iterative reconstructions possible Increased compatibility with clinical workflow Increased adoption of MBIR for C-arm CBCT sec 97 sec Future Work Further acceleration Automatic selection of M (# subsets) Incorporation of convergence criteria Salt Lake City, UT (June )
Self-Calibration of Cone-Beam CT Using 3D-2D Image Registration
Self-Calibration of Cone-Beam CT Using 3D-2D Image Registration Sarah Ouadah 1 J. Webster Stayman 1 Grace Jianan Gang 1 Ali Uneri 2 Tina Ehtiati 3 Jeffrey H. Siewerdsen 1,2 1. Department of Biomedical
More informationAcknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury
A. Sisniega et al. (presented at RSNA 214) High Performance Cone-Beam CT of Acute Traumatic Brain Injury A. Sisniega 1 W. Zbijewski 1, H. Dang 1, J. Xu 1 J. W. Stayman 1, J. Yorkston 2, N. Aygun 3 V. Koliatsos
More informationAccelerated statistical reconstruction for C-arm cone-beam CT using Nesterov s method
Accelerated statistical reconstruction for C-arm cone-beam CT using Nesterov s method Adam S. Wang, J. Webster Stayman, and Yoshito Otake Department of Biomedical Engineering, Johns Hopkins University,
More informationIncorporation of Prior Knowledge for Region of Change Imaging from Sparse Scan Data in Image-Guided Surgery
Incorporation of Prior Knowledge for Region of Change Imaging from Sparse Scan Data in Image-Guided Surgery J. Lee a, J. W. Stayman b, Y. Otake c, S. Schafer b, W. Zbijewski b, A. J. Khanna b,d, J. L.
More informationPenalized-Likelihood Reconstruction for Sparse Data Acquisitions with Unregistered Prior Images and Compressed Sensing Penalties
Penalized-Likelihood Reconstruction for Sparse Data Acquisitions with Unregistered Prior Images and Compressed Sensing Penalties J. W. Stayman* a, W. Zbijewski a, Y. Otake b, A. Uneri b, S. Schafer a,
More informationSpatial Resolution Properties in Penalized-Likelihood Reconstruction of Blurred Tomographic Data
Spatial Resolution Properties in Penalized-Likelihood Reconstruction of Blurred Tomographic Data Wenying Wang, Grace J. Gang and J. Webster Stayman Department of Biomedical Engineering, Johns Hopkins University,
More informationEstimating 3D Respiratory Motion from Orbiting Views
Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct. 2005 Funding provided by NIH Grant P01 CA59827 Motivation Free-breathing
More informationCT Reconstruction Using Spectral and Morphological Prior Knowledge: Application to Imaging the Prosthetic Knee
CT Reconstruction Using Spectral and Morphological Prior Knowledge: Application to Imaging the Prosthetic Knee Wojciech Zbijewski, J. Webster Stayman, Abdullah Muhit, John Yorkston, John A. Carrino and
More informationDesign and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon
Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization
More informationCT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION
CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology
More informationHot Topics in Imaging Physics A Key Interdisciplinary Science for 21 st Century Medicine
Hot Topics in Imaging Physics A Key Interdisciplinary Science for 21 st Century Medicine Jeff Siewerdsen, PhD FAAPM Department of Biomedical Engineering Department of Computer Science Russell H. Morgan
More informationSpiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004
Spiral ASSR Std p = 1.0 Spiral EPBP Std p = 1.0 Kachelrieß et al., Med. Phys. 31(6): 1623-1641, 2004 256 slices (0/300) Advantages of Cone-Beam Spiral CT Image quality nearly independent of pitch Increase
More informationSpiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA
Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization
More informationAcknowledgements. Atlas-based automatic measurements of the morphology of the tibiofemoral joint
Atlas-based automatic measurements of the morphology of the tibiofemoral joint M Brehler 1, G Thawait 2, W Shyr 1, J Ramsay 3, JH Siewerdsen 1,2, W Zbijewski 1 1 Dept. of Biomedical Engineering, Johns
More informationPhilips SPECT/CT Systems
Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET
More informationUSING cone-beam geometry with pinhole collimation,
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 687 A Backprojection-Based Parameter Estimation Technique for Skew-Slit Collimation Jacob A. Piatt, Student Member, IEEE, and Gengsheng L.
More informationA Fast GPU-Based Approach to Branchless Distance-Driven Projection and Back-Projection in Cone Beam CT
A Fast GPU-Based Approach to Branchless Distance-Driven Projection and Back-Projection in Cone Beam CT Daniel Schlifske ab and Henry Medeiros a a Marquette University, 1250 W Wisconsin Ave, Milwaukee,
More informationAxial block coordinate descent (ABCD) algorithm for X-ray CT image reconstruction
Axial block coordinate descent (ABCD) algorithm for X-ray CT image reconstruction Jeffrey A. Fessler and Donghwan Kim EECS Department University of Michigan Fully 3D Image Reconstruction Conference July
More informationFrom DQE to TRE: Image Science, Image Guidance, and Cone-Beam CT
W task = F task C task Magnification Focal Spot Size Scatter-to-Primary Ratio kvp, Dose Scintillator, Detector Pixel Aperture, 2D Sampling Reconstruction Filter Backprojection Voxel Size 3D Sampling PW,
More informationEvaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT
Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Qiao Yang 1,4, Meng Wu 2, Andreas Maier 1,3,4, Joachim Hornegger 1,3,4, Rebecca Fahrig
More informationStatistical iterative reconstruction using fast optimization transfer algorithm with successively increasing factor in Digital Breast Tomosynthesis
Statistical iterative reconstruction using fast optimization transfer algorithm with successively increasing factor in Digital Breast omosynthesis Shiyu Xu a and Zhenxi Zhang b and Ying Chen a,* a Department
More informationLow-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration
Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Joonki Noh, Jeffrey A. Fessler EECS Department, The University of Michigan Paul E. Kinahan Radiology Department,
More information(RMSE). Reconstructions showed that modeling the incremental blur improved the resolution of the attenuation map and quantitative accuracy.
Modeling the Distance-Dependent Blurring in Transmission Imaging in the Ordered-Subset Transmission (OSTR) Algorithm by Using an Unmatched Projector/Backprojector Pair B. Feng, Member, IEEE, M. A. King,
More information3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH
3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk
More information8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit
Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant
More informationIterative and analytical reconstruction algorithms for varying-focal-length cone-beam
Home Search Collections Journals About Contact us My IOPscience Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam projections This content has been downloaded from IOPscience.
More informationSimplified statistical image reconstruction algorithm for polyenergetic X-ray CT. y i Poisson I i (E)e R } where, b i
implified statistical image reconstruction algorithm for polyenergetic X-ray C omesh rivastava, tudent member, IEEE, and Jeffrey A. Fessler, enior member, IEEE Abstract In X-ray computed tomography (C),
More informationAcknowledgments and financial disclosure
AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories
More informationAccelerating CT Iterative Reconstruction Using ADMM and Nesterov s Methods
Accelerating CT Iterative Reconstruction Using ADMM and Nesterov s Methods Jingyuan Chen Meng Wu Yuan Yao June 4, 2014 1 Introduction 1.1 CT Reconstruction X-ray Computed Tomography (CT) is a conventional
More informationSplitting-Based Statistical X-Ray CT Image Reconstruction with Blind Gain Correction
Splitting-Based Statistical X-Ray CT Image Reconstruction with Blind Gain Correction Hung Nien and Jeffrey A. Fessler Department of Electrical Engineering and Computer Science University of Michigan, Ann
More informationBackground 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J.
Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging Stephen J. Glick* J. Michael O Connor**, Clay Didier**, Mini Das*, * University of Massachusetts Medical
More informationDiscrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset
Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset Andreas Maier 1, Patrick Kugler 2, Günter Lauritsch 2, Joachim Hornegger 1 1 Pattern Recognition Lab and SAOT Erlangen,
More informationGPU implementation for rapid iterative image reconstruction algorithm
GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial
More informationFinancial disclosure. Onboard imaging modality for IGRT
Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty
More informationPerformance Evaluation of CT Systems
8/2/206 Performance Evaluation of CT Systems AAPM Task Group 233 Foundations, Methods, and Prospects Ehsan Samei Duke University Medical Center Disclosures Research grant: NIH R0 EB00838 Research grant:
More informationAIDR 3D Iterative Reconstruction:
Iterative Reconstruction: Integrated, Automated and Adaptive Dose Reduction Erin Angel, PhD Manager, Clinical Sciences, CT Canon Medical Systems USA Iterative Reconstruction 1 Since the introduction of
More informationUniform Spatial Resolution for Statistical Image Reconstruction for 3D Cone-Beam CT
Uniform Spatial Resolution for Statistical Image Reconstruction for 3D Cone-Beam CT Jang Hwan Cho, Member, IEEE, and Jeffrey A. Fessler, Fellow, IEEE Department of EECS, University of Michigan, Ann Arbor,
More informationUnion of Learned Sparsifying Transforms Based Low-Dose 3D CT Image Reconstruction
Union of Learned Sparsifying Transforms Based Low-Dose 3D CT Image Reconstruction Xuehang Zheng 1, Saiprasad Ravishankar 2, Yong Long 1, Jeff Fessler 2 1 University of Michigan - Shanghai Jiao Tong University
More informationCARDIAC CT reconstruction requires high temporal resolution
1 Model Based Iterative Reconstruction With Spatially Adaptive Sinogram Weights for Wide-Cone Cardiac CT Amirkoushyar Ziabari, member, IEEE, Dong Hye Ye, member, IEEE, Lin Fu, Somesh Srivastava, member,
More informationImage Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD
Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image
More informationML reconstruction for CT
ML reconstruction for CT derivation of MLTR rigid motion correction resolution modeling polychromatic ML model dual energy ML model Bruno De Man, Katrien Van Slambrouck, Maarten Depypere, Frederik Maes,
More information1. INTRODUCTION ABSTRACT
Incorporating Tissue Excision in Deformable Image Registration: A Modified Demons Algorithm for Cone-Beam CT-Guided Surgery S.Nithiananthan, a D. Mirota, b A. Uneri, b S. Schafer, a Y. Otake, b J.W. Stayman,
More informationSimulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images
Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),
More informationCarbondale, IL USA; University of North Carolina Chapel Hill, NC USA; USA; ABSTRACT
Pre-computed backprojection based penalized-likelihood (PPL) reconstruction with an edge-preserved regularizer for stationary Digital Breast Tomosynthesis Shiyu Xu a, Christy Redmon Inscoe b, Jianping
More informationGPU acceleration of 3D forward and backward projection using separable footprints for X-ray CT image reconstruction
GPU acceleration of 3D forward and backward projection using separable footprints for X-ray CT image reconstruction Meng Wu and Jeffrey A. Fessler EECS Department University of Michigan Fully 3D Image
More informationCT vs. VolumeScope: image quality and dose comparison
CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,
More informationOptimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago
Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic
More informationComputer-Tomography II: Image reconstruction and applications
Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de
More informationMulti-slice CT Image Reconstruction Jiang Hsieh, Ph.D.
Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced
More information7/31/ D Cone-Beam CT: Developments and Applications. Disclosure. Outline. I have received research funding from NIH and Varian Medical System.
4D Cone-Beam CT: Developments and Applications Lei Ren, PhD, DABR Department of Radiation Oncology Duke University Medical Center Disclosure I have received research funding from NIH and Varian Medical
More informationBackground. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /
Radiographic Tomosynthesis: Image Quality and Artifacts Reduction Baojun Li, Ph.D Department of Radiology Boston University Medical Center 2012 AAPM Annual Meeting Background Linear Trajectory Tomosynthesis
More informationThe Near Future in Cardiac CT Image Reconstruction
SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de
More informationAdvanced Image Reconstruction Methods for Photoacoustic Tomography
Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic
More informationIEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 34, NO. 1, JANUARY
IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 34, NO. 1, JANUARY 2015 203 Axially Extended-Volume C-Arm CT Using a Reverse Helical Trajectory in the Interventional Room Zhicong Yu*, Andreas Maier, Günter
More informationReconstruction of CT Images from Sparse-View Polyenergetic Data Using Total Variation Minimization
1 Reconstruction of CT Images from Sparse-View Polyenergetic Data Using Total Variation Minimization T. Humphries and A. Faridani Abstract Recent work in CT image reconstruction has seen increasing interest
More informationConvolution-Based Truncation Correction for C-Arm CT using Scattered Radiation
Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation Bastian Bier 1, Chris Schwemmer 1,2, Andreas Maier 1,3, Hannes G. Hofmann 1, Yan Xia 1, Joachim Hornegger 1,2, Tobias Struffert
More information2005 IEEE Nuclear Science Symposium Conference Record M10-2
25 IEEE Nuclear Science Symposium Conference Record M1-2 Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng 1, Jeffrey A. Fessler 1, and James M. Balter 2 rzeng@eecs.umich.edu, fessler@eecs.umich.edu,
More informationComparison of High-Speed Ray Casting on GPU
Comparison of High-Speed Ray Casting on GPU using CUDA and OpenGL November 8, 2008 NVIDIA 1,2, Andreas Weinlich 1, Holger Scherl 2, Markus Kowarschik 2 and Joachim Hornegger 1 1 Chair of Pattern Recognition
More informationradiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li NCAAPM Spring Meeting 2010 Madison, WI
GPU-Accelerated autosegmentation for adaptive radiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li agodley@mcw.edu NCAAPM Spring Meeting 2010 Madison, WI Overview Motivation Adaptive
More informationEffects of the difference in tube voltage of the CT scanner on. dose calculation
Effects of the difference in tube voltage of the CT scanner on dose calculation Dong Joo Rhee, Sung-woo Kim, Dong Hyeok Jeong Medical and Radiological Physics Laboratory, Dongnam Institute of Radiological
More informationEnhancement Image Quality of CT Using Single Slice Spiral Technique
Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,
More informationNOVEL METHODS FOR SCATTER CORRECTION AND DUAL ENERGY IMAGING IN CONE-BEAM CT
NOVEL METHODS FOR SCATTER CORRECTION AND DUAL ENERGY IMAGING IN CONE-BEAM CT A Dissertation Presented to The Academic Faculty by Xue Dong In Partial Fulfillment of the Requirements for the Degree Doctor
More informationUnmatched Projector/Backprojector Pairs in an Iterative Reconstruction Algorithm
548 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 19, NO. 5, MAY 2000 Unmatched Projector/Backprojector Pairs in an Iterative Reconstruction Algorithm Gengsheng L. Zeng*, Member, IEEE, and Grant T. Gullberg,
More informationOptimization of Cone Beam CT Reconstruction Algorithm Based on CUDA
Sensors & Transducers 2013 by IFSA http://www.sensorsportal.com Optimization of Cone Beam CT Reconstruction Algorithm Based on CUDA 1 Wang LI-Fang, 2 Zhang Shu-Hai 1 School of Electronics and Computer
More informationQ.Clear. Steve Ross, Ph.D.
Steve Ross, Ph.D. Accurate quantitation (SUV - Standardized Uptake Value) is becoming more important as clinicians seek to utilize PET imaging for more than just diagnosing and staging disease, but also
More informationResearch Article CT Image Reconstruction from Sparse Projections Using Adaptive TpV Regularization
Hindawi Publishing Corporation Computational and Mathematical Methods in Medicine Volume 2015, Article ID 354869, 8 pages http://dx.doi.org/10.1155/2015/354869 Research Article CT Image Reconstruction
More informationIEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 32, NO. 11, NOVEMBER
IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 32, NO. 11, NOVEMBER 2013 1965 Accelerating Ordered Subsets Image Reconstruction for X-ray CT Using Spatially Nonuniform Optimization Transfer Donghwan Kim*,
More information8/2/2016. Acknowledgement. Common Clinical Questions. Presumption Images are Good Enough to accurately answer clinical questions
Image Quality Assessment using Model Observers: Clinical Implementation and Practical Considerations Shuai Leng, PhD Associate Professor, Department of Radiology Mayo Clinic, Rochester MN Acknowledgement
More informationGPU applications in Cancer Radiation Therapy at UCSD. Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC
GPU applications in Cancer Radiation Therapy at UCSD Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC Conventional Radiotherapy SIMULATION: Construciton, Dij Days PLANNING:
More informationNoise weighting with an exponent for transmission CT
doi:10.1088/2057-1976/2/4/045004 RECEIVED 13 January 2016 REVISED 4 June 2016 ACCEPTED FOR PUBLICATION 21 June 2016 PUBLISHED 27 July 2016 PAPER Noise weighting with an exponent for transmission CT Gengsheng
More informationAccelerated C-arm Reconstruction by Out-of-Projection Prediction
Accelerated C-arm Reconstruction by Out-of-Projection Prediction Hannes G. Hofmann, Benjamin Keck, Joachim Hornegger Pattern Recognition Lab, University Erlangen-Nuremberg hannes.hofmann@informatik.uni-erlangen.de
More informationImage Acquisition Systems
Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS
More informationSpectral analysis of non-stationary CT noise
Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available
More informationCIVA CT, an advanced simulation platform for NDT
More Info at Open Access Database www.ndt.net/?id=18774 CIVA CT, an advanced simulation platform for NDT Marius Costin 1, David Tisseur 1, Caroline Vienne 1, Ronan Guillamet 1, Hussein Banjak 1, Navnina
More informationRespiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data
Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data Christopher M Rank 1, Thorsten Heußer 1, Andreas Wetscherek 1, and Marc Kachelrieß 1 1 German Cancer
More informationOutline. Outline 7/24/2014. Fast, near real-time, Monte Carlo dose calculations using GPU. Xun Jia Ph.D. GPU Monte Carlo. Clinical Applications
Fast, near real-time, Monte Carlo dose calculations using GPU Xun Jia Ph.D. xun.jia@utsouthwestern.edu Outline GPU Monte Carlo Clinical Applications Conclusions 2 Outline GPU Monte Carlo Clinical Applications
More informationArtefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT
Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT Marcus Brehm 1,2, Thorsten Heußer 1, Pascal Paysan 3, Markus Oehlhafen 3, and Marc Kachelrieß 1,2 1 German Cancer Research Center
More informationACCELERATED OPTIMIZATION ALGORITHMS FOR STATISTICAL 3D X-RAY COMPUTED TOMOGRAPHY IMAGE RECONSTRUCTION
ACCELERATED OPTIMIZATION ALGORITHMS FOR STATISTICAL 3D X-RAY COMPUTED TOMOGRAPHY IMAGE RECONSTRUCTION by Donghwan Kim A dissertation submitted in partial fulfillment of the requirements for the degree
More informationAn Efficient Technique For Multi-Phase Model Based Iterative Reconstruction
1 An Efficient Technique For Multi-Phase Model Based Iterative Reconstruction Shiyu Xu, Debashish Pal and Jean-Baptiste Thibault Abstract Multi-phase scan is a fundamental CT acquisition technology used
More informationAutomated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator
Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4
More informationCT Basics Principles of Spiral CT Dose. Always Thinking Ahead.
1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central
More informationCT imaging using energy-sensitive photon-counting detectors. Disclosure. Vision 20/20 paper
CT imaging using energy-sensitive photon-counting detectors Katsuyuki Ken Taguchi, Ph.D. ktaguchi@jhmi.edu Division of Medical Imaging Physics Department of Radiology Johns Hopkins University No financial
More informationPortability of TV-Regularized Reconstruction Parameters to Varying Data Sets
Portability of TV-Regularized Reconstruction Parameters to Varying Data Sets Mario Amrehn 1, Andreas Maier 1,2, Frank Dennerlein 1, Joachim Hornegger 1,2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg
More informationAutomated Quality Assurance for Image-Guided Radiation Therapy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 1, WINTER 2009 Automated Quality Assurance for Image-Guided Radiation Therapy Eduard Schreibmann, a Eric Elder, Tim Fox Department of Radiation
More informationCombined algorithmic and GPU acceleration for ultra-fast circular conebeam backprojection
Combined algorithmic and GPU acceleration for ultra-fast circular conebeam backprojection Jeffrey Brokish a, Paul Sack a, Yoram Bresler ab a InstaRecon, Inc., Champaign, IL USA 61820 b Dept. of Electrical
More informationThe team. Disclosures. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge.
Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge Dimitre Hristov Radiation Oncology Stanford University The team Renhui Gong 1 Magdalena Bazalova-Carter 1
More informationA Multi-view Active Contour Method for Bone Cement Reconstruction from C-Arm X-Ray Images
A Multi-view Active Contour Method for Bone Cement Reconstruction from C-Arm X-Ray Images Blake C. Lucas 1,2, Yoshito Otake 1, Mehran Armand 2, Russell H. Taylor 1 1 Dept. of Computer Science, Johns Hopkins
More informationOptimization of GPU-Accelerated Iterative CT Reconstruction Algorithm for Clinical Use
Washington University in St. Louis Washington University Open Scholarship Engineering and Applied Science Theses & Dissertations Engineering and Applied Science Spring 5-18-2018 Optimization of GPU-Accelerated
More informationFeldkamp-type image reconstruction from equiangular data
Journal of X-Ray Science and Technology 9 (2001) 113 120 113 IOS Press Feldkamp-type image reconstruction from equiangular data Ben Wang a, Hong Liu b, Shiying Zhao c and Ge Wang d a Department of Elec.
More informationEvaluation of Low-Dose CT Perfusion for the Liver using Reconstruction of Difference
1 Evaluation of Low-Dose CT Perfusion for the Liver using Reconstruction of Difference Saeed Seyyedi, Eleni Liapi, Tobias Lasser, Robert Ivkov, Rajeev Hatwar, J. Webster Stayman Abstract CT perfusion imaging
More informationNon-Homogeneous Updates for the Iterative Coordinate Descent Algorithm
Non-Homogeneous Updates for the Iterative Coordinate Descent Algorithm Zhou Yu a, Jean-Baptiste Thibault b, Charles A. Bouman a, Ken D. Sauer c, and Jiang Hsieh b a School of Electrical Engineering, Purdue
More informationSlab-by-Slab Blurring Model for Geometric Point Response Correction and Attenuation Correction Using Iterative Reconstruction Algorithms
2168 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL 45, NO. 4, AUGUST 1998 Slab-by-Slab Blurring Model for Geometric Point Response Correction and Attenuation Correction Using Iterative Reconstruction Algorithms
More informationShiyu Xu a, Henri Schurz b, Ying Chen a,c, Abstract
Parameter Optimization of relaxed Ordered Subsets Pre-computed Back Projection (BP) based Penalized-Likelihood (OS-PPL) Reconstruction in Limited-angle X-ray Tomography Shiyu Xu a, Henri Schurz b, Ying
More informationDisclosures 7/21/2014. AAPM 2014 Challenges and opportunities in assessment of image quality
7/2/24 AAPM 24 Challenges and opportunities in assessment of image quality Ehsan Samei Department of Radiology Duke University Medical Center Disclosures Research grant: NIH R EB838 Research grant: General
More informationCUDA and OpenCL Implementations of 3D CT Reconstruction for Biomedical Imaging
CUDA and OpenCL Implementations of 3D CT Reconstruction for Biomedical Imaging Saoni Mukherjee, Nicholas Moore, James Brock and Miriam Leeser September 12, 2012 1 Outline Introduction to CT Scan, 3D reconstruction
More informationFAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION
2009 IEEE Nuclear Science Symposium Conference Record M03-7 FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION Wonseok Huh, Jeffrey A. Fessler, Adam M. Alessio, and Paul E. Kinahan Department
More informationCombining Ordered Subsets and Momentum for Accelerated X-Ray CT Image Reconstruction
IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 34, NO. 1, JANUARY 2015 167 Combining Ordered Subsets and Momentum for Accelerated X-Ray CT Image Reconstruction Donghwan Kim*, Member, IEEE, Sathish Ramani,
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More informationAn FDK-like cone-beam SPECT reconstruction algorithm for non-uniform attenuated
Home Search Collections Journals About Contact us My IOPscience An FK-like cone-beam SPECT reconstruction algorithm for non-uniform attenuated projections acquired using a circular trajectory This content
More informationCone-beam computed tomography with a flat-panel imager: Initial performance characterization
Cone-beam computed tomography with a flat-panel imager: Initial performance characterization D. A. Jaffray a) and J. H. Siewerdsen Department of Radiation Oncology, William Beaumont Hospital, Royal Oak,
More information