CT Reconstruction Using Spectral and Morphological Prior Knowledge: Application to Imaging the Prosthetic Knee

Size: px
Start display at page:

Download "CT Reconstruction Using Spectral and Morphological Prior Knowledge: Application to Imaging the Prosthetic Knee"

Transcription

1 CT Reconstruction Using Spectral and Morphological Prior Knowledge: Application to Imaging the Prosthetic Knee Wojciech Zbijewski, J. Webster Stayman, Abdullah Muhit, John Yorkston, John A. Carrino and Jeffrey H. Siewerdsen Abstract Imaging in the presence of prosthetic implants presents a notoriously difficult challenge to CT reconstrucion. Such hardware is made of alloys that are highly attenuating (e.g., Co-Cr-Mo) and impart severe degradation in image quality due to photon starvation, beam hardening, etc. An important clinical example is in the proliferation of total knee replacement, increasing the need for technologies capable of imaging in the presence of knee prostheses. The usefulness of CT in follow-up to knee replacement surgery is, however, extremely limited due to severe artifacts associated with the implant. Recent developments in likelihood-based CT reconstruction offer a potential solution to the problem. In particular, we exploit the fact that exact models of the shape and composition of prostheses are often available. A framework is proposed that extends earlier work [1,2] on known component reconstruction (KCR) to account for polyenergetic beam hardening and apply to the case of a large, highly attenuating object such as a knee implant. The proposed algorithm uses a polyenergetic object model to simultaneously estimate the unknown background density volume and the position and orientation of the known implant. We test the approach in studies emulating a recently developed, dedicated cone-beam CT scanner for extremities imaging. The results indicate substantial reduction of image artifacts and significant improvements in the visualization of areas adjacent to the implant. The KCR approach is found to outperform traditional filtered-backprojection and penalizedlikelihood methods that do not account for the implant model or polyenergetic object attenuation. The method suggests promising new capability to assess implant integrity, loosening, and tissue disease (osteolysis and soft-tissue derangement). Index Terms CT Reconstruction, Extremities Imaging, Implant Imaging, Metal Artifact Reduction, Penalized- Likelihood Estimation. I. INTRODUCTION The prevalence on knee replacement surgery is rapidly increasing, with some predictions suggesting over a million replacements per year by 2015 [3]. Consequently, there is a growing need for imaging technologies that enable follow- This work was supported in part by academic-industry partnership with Carestream Health (Rochester NY) and by NIH 2R01-CA W. Zbijewski, J. W. Stayman, A. Muhit and J. H. Siewerdsen are with the Department of Biomedical Engineering, Johns Hopkins University, Baltimore, MD USA (phone: ; fax: ; e- mail: wzbijewski@jhu.edu). J. Yorkston is with Carestream Health, Rochester, NY. J. A. Carrino is with Department of Radiology, Johns Hopkins University, Baltimore, MD up of knee prostheses, which requires reliable visualization in the direct vicinity of the implant. X-ray CT would be a compelling candidate thanks to its excellent spatial resolution and numerous existing applications in orthopedic radiology. However, its application to implant imaging remains a major and largely unsolved challenge due to severe artifacts caused primarily by beam hardening and photon starvation. An example knee implant is shown in Fig. 1(A), along with CT images of a prosthetic knee in (B) and (C), clearly demonstrating the magnitude of the associated artifacts. (CT images are courtesy of E. Fishman, MD, [4]). As illustrated in Fig. 1(D), typical alloys used in manufacturing knee prostheses (Co-Cr-Mo, consisting of ~60% Co, ~30% Cr and ~5% Mo) are significantly more attenuating to diagnostic x-rays than other common highly x-ray opaque materials, such as Ti or cortical bone. As a result, x-ray beams traversing even a short pathlength (~1-2 mm) in such alloys suffer not only from severe photon starvation but also significant beam hardening. This is shown in Fig. 1(E), where 110 kvp x-ray spectra are attenuated by 2 mm of cortical bone, Ti, and Co-Cr-Mo, with the spectrum attenuated by Co-Cr-Mo exhibiting the most significant shift to high energies. Fig. 1. (A) Example of a knee implant: DePuy Sigma [5]. Such prostheses present a large volume of highly x-ray attenuating alloy, often Cr-Co-Mo containing ~60% Cobalt, and thus pose a major challenge for x-ray CT. Artifacts associated with the presence of knee implants in CT scans are shown in (B) and (C) [4]. In (D), x-ray attenuation of Co-Cr-Mo (solid line) is compared to Titanium (dotted line) and cortical bone (dashed line). Significant loss of photon flux and substantial beam hardening are observed even for short path lengths through the alloy, as shown in (E), where x-ray spectra attenuated by 2 mm of Co-Cr-Mo (solid line), Ti (dotted line) and cortical bone (dashed line) are compared to the unattenuated 110 kvp beam. Page 434

2 These extremely strong attenuation characteristics of the Co-Cr-Mo alloy, combined with the large size of the implants [Fig. 1(A)] lead to large areas of missing data (zero or near-zero counts) and low signal-to-noise ratio in the projections. They are thus extremely challenging for conventional metal artifact reduction methods, such as those based on interpolation in the projection domain [6]. While the development of CT-compatible implants is an area of ongoing research, the ability to image heavy metallic prosthetics would be of significant immediate benefit to a growing population of patients. The ability to image the prosthetic knee could be especially beneficial in the context of recently introduced dedicated cone-beam CT (CBCT) systems for extremities, such as the prototype shown in Fig. 2 [7]. Providing novel capabilities (e.g., load-bearing imaging), these systems promise to significantly expand the scope and quality of CT applications in orthopedics. Increasing prevalence of knee replacement implies that the ability to image in the presence of implants will be an important aspect of broad utilization of such dedicated devices. Moreover, the open system architecture of the prototype in Fig. 2 provides an ideal platform for development of novel reconstruction algorithms. monoenergetic x-ray beam, which may be sufficient when polyenergetic effects in the object under consideration are relatively weak. However, as illustrated above, the heavy alloys used in knee implants exhibit significant beam hardening, and thus the KCR algorithm needs to be expanded to polyenergetic beams and energy-dependent attenuation. A number of approaches to include spectral effects in penalized-likelihood reconstruction have been proposed [8-10]. Here, we combine the algorithm of Elbakri et al. [9] with KCR and apply the resulting method to imaging in the presence of a knee prostheses. A. Forward Model II. METHODS We assume an energy-integrating x-ray detector with uniform detection efficiency and write the measured projection value for detector element i: (1) where is the total photon flux for pixel i, is the x-ray energy, is the spectral density of the x-ray beam, is the mass-attenuation coefficient of the object, is the object density, and x is the spatial dimension. We express the line integral in (1) for a voxelized object following Elbakri et al. [9], where we assume that the object consists of K materials (massattenuation of k-th material given by ) and that the fraction of k-th material in voxel j is known and denoted by. The mean signal at pixel i is then: Fig. 2. The recently developed flat-panel cone-beam CT system for extremities imaging. The scanner allows for scanning in both a weightbearing, standing (A) configuration and in a non-weight-bearing, sitting (B) configuration. Studies in cadaveric specimens indicate soft-tissue visibility comparable to conventional CT (C) and exquisite spatial resolution (D). Images in (C) and (D) were obtained from a single 80 kvp, 9 mgy acquisition using soft-tissue and bone reconstruction kernels, respectively. One of the unique features of implant imaging is that the knowledge of their shape and composition is often available (e.g., from CAD models). If properly leveraged in the reconstruction, such prior knowledge can be used to alleviate the effects of missing data and low signal-to-noise ratio. This is especially true if combined with a penalizedlikelihood (PL) approach, which properly accounts for projection noise, such as in the recently proposed Known Component Reconstruction (KCR) [1,2]. In KCR, the position of the known components is estimated jointly with the unknown background image (i.e., the underlying anatomy) using an alternating joint optimization method. This algorithm has been shown to yield excellent, near artifact-free reconstructions in imaging of pedicle screws under conditions of severe photon starvation. KCR has been previously developed under the assumption of a where is the system matrix. By assuming that the fraction is known throughout the volume, the formulation in (2) allows the likelihood-based objective function for image reconstruction to have only one unknown per voxel, namely the density (as opposed to a number of unknowns equal to the number of energy bins without this simplification). Elbakri et al. [9] assumed that the material fractions are known from segmentation of an initial FBP reconstruction. Analogous to KCR, we will now further parameterize the object as a superposition of an unknown background image, given by density distribution, and a known implant volume undergoing an arbitrary transformation (registration). The implant transformation is characterized by an unknown vector (e.g., rotation and translation). This leads to the following substitution in (2): where m is a mask representing the support of the implant. (2) (3) Page 435

3 B. Reconstruction algorithms Equation (2) with the line integral given by (3) provides a relationship between mean measurement at pixel i and the object volume. To estimate the background density and the parameters of the implant transformation, we invoke a Poisson noise model and write the loglikelihood: (4) The unknowns can be estimated by maximizing the penalized-likelihood estimator: (5) where is a regularization term to penalize noisy images (e.g., a pair-wise quadratic penalty). The estimator in (5) provides a general form that encompasses a variety of reconstruction object models (mono and polyenergetic, with and without the known components). Different iterative algorithms are used to solve (5), depending on the exact choice of the reconstruction object model. The simulated projection data in this study were obtained using the full spectral model. We consider the following cases: PL-Mono: if the x-ray spectrum is assumed to be monoenergetic ( and no implant is included in the reconstruction model, (5) becomes the familiar penalized-likelihood estimator for monoenergetic x-ray CT, which is solved iteratively as in [11]. KCR-Mono: in this case, the spectrum is still assumed monoenergetic, but the implant volume is included in the model, as in (3). Since polyenergetic effects are neglected, the implant attenuation is assumed equal to its attenuation at 90 kev, chosen to reflect the beam hardening effect of the Co alloy for the 110 kvp beam. The KCR algorithm of Stayman et al. [1,2] is used to solve this objective by an alternating minimization of and. PL-Poly: when the reconstruction uses a polyenergetic object model, but without knowledge of the implant volume, the segmentation-based polyenergetic PL algorithm of Elbakri et al. [9] is employed. Two cases are considered: PL-Poly-Single, where the segmentation ( in (2)) assumes that all the voxels are composed of soft-tissue (represented by of muscle) and PL-Poly-Full with oracle segmentation into soft tissue, bone, and implant. Obtaining such segmentation from the initial FBP corrupted by artifacts caused by the prosthesis may be difficult; thus, PL-Poly-Full provides an upper bound on the performance of PL-Poly for such data. KCR-Poly: the full model of (3) is assumed. We estimate the unknown background density volume and the position of the known implant using a polyenergetic model of x-ray transmission. The same update equation as in PL- Poly is employed, but with the terms depending on the forward projection computed using the separation of the volume into the density map and the registered implant volume, as in (3). This extension of the KCR methodology (aside from application to extremity imaging) is the main novel contribution of the current work. Similar to PL-Poly, segmentation needs to be provided to constrain the estimation of Note however that the segmentation of the implant is no longer required in KCR since the a priori knowledge of the implant is now part of the object model. Two approaches are tested: a simple segmentation assuming that all voxels consist of soft tissue (KCR-Poly- Single) and oracle segmentation into soft tissue and bone (KCR-Poly-Full, again providing the upper bound on algorithm performance). The initial studies were performed under the assumption that the implant transformation W is fixed - i.e., that the implant location is known. The minimization over in (5) is therefore omitted. While this is a simplification, the results presented in [1,2] indicate that the registration step is extremely robust for a wide range of conditions, including the presence of multiple known objects. We therefore expect that inclusion of this step in KCR-Poly (ongoing work) will not alter the conclusions. C. Phantoms, simulations, and reconstruction settings Fig. 3. (left) Phantoms used in the study. The knee implant (pink overlay) was simulated by replacing bone voxels with the Co-based alloy. Note the circular soft-tissue contrast inserts in the joint spaces and inside the condyles. (center) FBP reconstruction of the phantom with no implant, illustrating the magnitude of beam hardening due to bone. (right) Beam hardening artifacts are efficiently removed with the polyenergetic PL algorithm of Elbakri et al. [10] This algorithm was combined with Known-Component Reconstruction (KCR) [1,2] to yield a polyenergetic algorithm robust to the presence of implants. Fig. 3 (left) demonstrates the phantoms used in this study. Two slices through segmented knee volumes from the Virtual Population dataset [12] were used. The base materials included four soft tissues (skin, fat, muscle, and cartilage), bone marrow, and cortical bone. Circular soft tissue contrast inserts (4 mm radius) were placed inside the joint space and inside the condyles; each insert was made from the same material as its background, but at a 1.2x higher density. The knee implant was simulated by replacing a region of bone voxels with a Co-Cr-Mo alloy, shown as pink overlay in Fig. 3. Following the basic geometry of such implants (Fig. 1), all bone voxels were replaced with the alloy in the distal region of the condyles (top row of Fig. 2); in the proximal region of the condyles (bottom row) only the inner and anterior bone surfaces were replaced with the alloy. The geometry of the extremities CBCT scanner (Fig. 2) was simulated. For simplicity, a single slice, fan-beam Page 436

4 acquisition was considered in this preliminary study. The SDD was 550 mm, and the SAD was 430 mm; there were 384 detector pixels at mm pitch. A circular orbit with 360 projections at 1 o increment was assumed. The voxel size was 1 mm. Polyenergetic projections were simulated using Eq. (1) with 10 6 photons per detector element. The x- ray spectrum for a 110 kvp beam with 2 mm Al and 0.2 mm Cu filtration (as in the prototype scanner) was computed using SPEKTR [13]. Reconstruction by the PL and KCR methods above involved the same x-ray spectrum, pixel size, and voxel size as in the simulation. 200 iterations with 60 subsets were performed in each case. The regularization parameter β was varied to achieve similar resolution in all reconstructions. Matched forward- and back-projector based on separable footprints [14] were implemented in a CUDA-based library for nvidia GPUs. somewhat alleviated when a likelihood-based algorithm is used (PL-Mono). This demonstrates the value of including a noise model that penalizes low-count projections in the reconstruction. Nevertheless, the areas surrounding the implant are still plagued by artifacts. Including knowledge of the implant morphology (KCR-Mono) is not sufficient to reduce the artifacts because of the mismatch between the monoenergetic model of implant attenuation and the polyenergetic nature of the measured data. III. RESULTS Fig. 4. Monoenergetic reconstructions (from polyenergetic projection data) in the presence of a knee implant. (top) FBP, showing severe streaks due to photon starvation and beam-hardening in the presence of the implant. (center) PL-Mono reconstruction, demonstrating a degree of improvement in artifacts caused by poor photon statistics; however, significant streak artifacts are persistent. (bottom) KCR-Mono reconstruction, where the implant (pink overlay) was modeled using its attenuation at 90 kev. Neglecting spectral effects results in little or no improvement over PL- Mono. FBP and PL-Poly reconstructions of the two phantoms without the prostheses are shown Fig. 3 (center and right), demonstrating the strength of PL-Poly in addressing beam hardening from common materials such as cortical bone. In Fig. 4, reconstructions of polyenergetic projections of the phantoms with the implant included are compared for monoenergetic algorithms. The FBP images are severely compromised by streak artifacts. These artifacts are Fig. 5. Polyenergetic PL and KCR reconstructions. (top) PL-Poly reconstruction of the implanted knee, where object segmentation consisted of only soft tissue. The single-component model is not sufficient to suppress streak artifacts. (second row) PL-Poly with oracle segmentation that included soft tissue, bone, and the implant. While significantly improved from either PL-Mono or PL-Poly-Single, the image still suffers from artifacts. Such artifacts are largely removed with KCR-Poly-Single (third row), which applies knowledge of the material and morphological characteristics of the implant and estimates the unknown underlying anatomy simultaneously with the position of the known implant. (bottom) KCR-Poly-Full gives further improvement by including segmentation of bone and soft tissue. In Fig. 5, reconstruction algorithms that employ a polyenergetic object model are compared for the same projection data. A simple model of object attenuation used in PL-Poly-Single (all voxels consisting of soft-tissue) fails to produce significant reduction in artifacts. Even when the implant is included in the initial segmentation (PL-Poly- Full), artifacts persist in the vicinity of the prosthesis. They are significantly reduced with KCR-Poly-Single (third row in Fig. 5), as evident by the improved visualization of the Page 437

5 contrast inserts. The artifacts are further diminished when bone is also included in segmentation (KCR-Poly-Full, bottom row on Fig. 5). While traditional algorithms attempt to extract information from data with very few or zero counts (in effect a data null-space), KCR uses a priori implant knowledge to better condition the problem and select solutions from within potential null-spaces. Note also that KCR-Poly-Single achieves excellent artifact reduction while not requiring a complete segmentation of the object, which is likely difficult to obtain in the presence of artifacts caused by the prosthesis. IV. DISCUSSION A novel reconstruction algorithm (KCR-Poly) that combines the knowledge of the material and morphological characteristics of the prosthesis with a polyenergetic model of x-ray propagation was introduced. The algorithm was shown to yield excellent results in reduction of severe image artifacts around such implants in conventional reconstructions. The success of KCR-Poly owes to both the use of a priori knowledge of the implant and the account of the polyenergetic nature of object attenuation, as indicated by the persistence of artifacts for the monoenergetic version of KCR. Ongoing work includes inclusion of the implant registration step as part of the joint optimization [1,2] in KCR, extension to segmentation-free polyenergetic reconstruction [10], and testing in real data, including an analysis of the effects of scatter [15]. The algorithm will be tested on the prototype extremities scanner in realistic scenarios and extended to other applications involving large metallic implants. REFERENCES [1] Stayman J. W., Otake Y., Prince J. L., Siewerdsen J. H., "Likelihoodbased CT Reconstruction of Objects Containing Known Components," Int l Mtg. Fully 3D Image Recon., (2011). [2] Stayman J. W., Otake Y., Prince J. L., Siewerdsen J. H., "Model- Based Tomographic Reconstruction of Objects Containing Known Components", IEEE Trans. Med. Im., in press (2012). [3] Kim, S., "Changes in Surgical Loads and Economic Burden of Hip and Knee Replacements in the US: ," Arthritis & Rheumatism (Arthritis Care & Research), 59(4): (2008). [4] Fishman E. K., "Case 3563: Total Knee Replacement," [5] DePuy Companies, "Sigma Rotating Platform Knee," [6] Kalender W. A., Hebel R., Ebersberger J., "Reduction of CT artifacts caused by metallic implants," Radiology, 164: (1987). [7] Zbijewski, W., De Jean P., Prakash P., Ding Y., Stayman J. W., Packard N., Senn R., Yang D., Yorkston J., Machado A., Carrino J. A., Siewerdsen J. H., "A dedicated cone-beam CT system for musculoskeletal extremities imaging: Design, optimization, and initial performance characterization," Med. Phys. 38(8): (2011) [8] De Man, B., Nuyts, J., Dupont, P., Marchal, G., Suetens, P., "An iterative maximum-likelihood polychromatic algorithm for CT," IEEE Trans. Med. Im., 20(10): (2001). [9] Elbakri, I. A., Fessler, J. A., "Statistical image reconstruction for polyenergetic X-ray computed tomography," IEEE Trans. Med. Im. 21(2):89-99 (2002). [10] Elbakri, I. A., Fessler, J. A., " Segmentation-free statistical image reconstruction for polyenergetic X-ray computed tomography with experimental validation, Phys. Med. Biol., 48(15): (2003) [11] Erdogan, H., Fessler, J. A., "Ordered subsets algorithms for transmission tomography," Phys. Med. Biol. 44: (1999). [12] IT'IS Foundation, "The Virtual Population," [13] Siewerdsen, J. H., Waese, A. M., Moseley, D. J., Richard, S., Jaffray, D. A., " Spektr: A computational tool for x-ray spectral analysis and imaging system optimization," Med. Phys., 31:3057 (2004). [14] Long, Y., Fessler, J. A., Balter, J. M., "3D forward and backprojection or X-ray CT using separable footprints," IEEE Trans. Med. Im., 29: (2010). [15] Zbijewski, W., Sisniega, A., Vaquero, J. J., Packard, N., Senn, R., Yang, D., J. Yorkston, Carrino, J. A., Siewerdsen, J. H., Dose and Scatter Characteristics of a Novel Cone Beam CT system for Musculoskeletal Extremities, Proc. of SPIE Medical Imaging (2011). Page 438

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Qiao Yang 1,4, Meng Wu 2, Andreas Maier 1,3,4, Joachim Hornegger 1,3,4, Rebecca Fahrig

More information

Acknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury

Acknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury A. Sisniega et al. (presented at RSNA 214) High Performance Cone-Beam CT of Acute Traumatic Brain Injury A. Sisniega 1 W. Zbijewski 1, H. Dang 1, J. Xu 1 J. W. Stayman 1, J. Yorkston 2, N. Aygun 3 V. Koliatsos

More information

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Joonki Noh, Jeffrey A. Fessler EECS Department, The University of Michigan Paul E. Kinahan Radiology Department,

More information

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland Metal Artifact Reduction CT Techniques R S S S Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland N. 1 v o 4 1 0 2. Postoperative CT Metal Implants CT is accurate for assessment

More information

Acknowledgments. Nesterov s Method for Accelerated Penalized-Likelihood Statistical Reconstruction for C-arm Cone-Beam CT.

Acknowledgments. Nesterov s Method for Accelerated Penalized-Likelihood Statistical Reconstruction for C-arm Cone-Beam CT. June 5, Nesterov s Method for Accelerated Penalized-Likelihood Statistical Reconstruction for C-arm Cone-Beam CT Adam S. Wang, J. Webster Stayman, Yoshito Otake, Gerhard Kleinszig, Sebastian Vogt, Jeffrey

More information

ML reconstruction for CT

ML reconstruction for CT ML reconstruction for CT derivation of MLTR rigid motion correction resolution modeling polychromatic ML model dual energy ML model Bruno De Man, Katrien Van Slambrouck, Maarten Depypere, Frederik Maes,

More information

Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2)

Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2) Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2) Author: Levon Vogelsang Introduction Cone beam computed tomography (CBCT), or cone beam CT technology, offers considerable promise

More information

Simplified statistical image reconstruction algorithm for polyenergetic X-ray CT. y i Poisson I i (E)e R } where, b i

Simplified statistical image reconstruction algorithm for polyenergetic X-ray CT. y i Poisson I i (E)e R } where, b i implified statistical image reconstruction algorithm for polyenergetic X-ray C omesh rivastava, tudent member, IEEE, and Jeffrey A. Fessler, enior member, IEEE Abstract In X-ray computed tomography (C),

More information

Superiorized polyenergetic reconstruction algorithm for reduction of metal artifacts in CT images

Superiorized polyenergetic reconstruction algorithm for reduction of metal artifacts in CT images 1 Superiorized polyenergetic reconstruction algorithm for reduction of metal artifacts in CT images T. Humphries 1 and A. Gibali 2 Abstract Artifacts caused by metal objects such as dental fillings, hip

More information

Incorporation of Prior Knowledge for Region of Change Imaging from Sparse Scan Data in Image-Guided Surgery

Incorporation of Prior Knowledge for Region of Change Imaging from Sparse Scan Data in Image-Guided Surgery Incorporation of Prior Knowledge for Region of Change Imaging from Sparse Scan Data in Image-Guided Surgery J. Lee a, J. W. Stayman b, Y. Otake c, S. Schafer b, W. Zbijewski b, A. J. Khanna b,d, J. L.

More information

Spatial Resolution Properties in Penalized-Likelihood Reconstruction of Blurred Tomographic Data

Spatial Resolution Properties in Penalized-Likelihood Reconstruction of Blurred Tomographic Data Spatial Resolution Properties in Penalized-Likelihood Reconstruction of Blurred Tomographic Data Wenying Wang, Grace J. Gang and J. Webster Stayman Department of Biomedical Engineering, Johns Hopkins University,

More information

Frequency split metal artifact reduction (FSMAR) in computed tomography

Frequency split metal artifact reduction (FSMAR) in computed tomography The Johns Hopkins University Advanced Computer Integrated Surgery Group 4 Metal Artifact Removal in C-arm Cone-Beam CT Paper Seminar Critical Review of Frequency split metal artifact reduction (FSMAR)

More information

Penalized-Likelihood Reconstruction for Sparse Data Acquisitions with Unregistered Prior Images and Compressed Sensing Penalties

Penalized-Likelihood Reconstruction for Sparse Data Acquisitions with Unregistered Prior Images and Compressed Sensing Penalties Penalized-Likelihood Reconstruction for Sparse Data Acquisitions with Unregistered Prior Images and Compressed Sensing Penalties J. W. Stayman* a, W. Zbijewski a, Y. Otake b, A. Uneri b, S. Schafer a,

More information

Scatter Correction Methods in Dimensional CT

Scatter Correction Methods in Dimensional CT Scatter Correction Methods in Dimensional CT Matthias Baer 1,2, Michael Hammer 3, Michael Knaup 1, Ingomar Schmidt 3, Ralf Christoph 3, Marc Kachelrieß 2 1 Institute of Medical Physics, Friedrich-Alexander-University

More information

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION 2009 IEEE Nuclear Science Symposium Conference Record M03-7 FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION Wonseok Huh, Jeffrey A. Fessler, Adam M. Alessio, and Paul E. Kinahan Department

More information

Reconstruction of CT Images from Sparse-View Polyenergetic Data Using Total Variation Minimization

Reconstruction of CT Images from Sparse-View Polyenergetic Data Using Total Variation Minimization 1 Reconstruction of CT Images from Sparse-View Polyenergetic Data Using Total Variation Minimization T. Humphries and A. Faridani Abstract Recent work in CT image reconstruction has seen increasing interest

More information

Self-Calibration of Cone-Beam CT Using 3D-2D Image Registration

Self-Calibration of Cone-Beam CT Using 3D-2D Image Registration Self-Calibration of Cone-Beam CT Using 3D-2D Image Registration Sarah Ouadah 1 J. Webster Stayman 1 Grace Jianan Gang 1 Ali Uneri 2 Tina Ehtiati 3 Jeffrey H. Siewerdsen 1,2 1. Department of Biomedical

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy T. Rohlfing a, D. Zerfowski b, J. Beier a, P. Wust a, N. Hosten a, R. Felix a a Department of

More information

Artifacts in Spiral X-ray CT Scanners: Problems and Solutions

Artifacts in Spiral X-ray CT Scanners: Problems and Solutions Artifacts in Spiral X-ray CT Scanners: Problems and Solutions Mehran Yazdi, and Luc Beaulieu International Science Index, Electrical and Computer Engineering waset.org/publication/12966 Abstract Artifact

More information

Acknowledgments and financial disclosure

Acknowledgments and financial disclosure AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Artifact Mitigation in High Energy CT via Monte Carlo Simulation PIERS ONLINE, VOL. 7, NO. 8, 11 791 Artifact Mitigation in High Energy CT via Monte Carlo Simulation Xuemin Jin and Robert Y. Levine Spectral Sciences, Inc., USA Abstract The high energy (< 15 MeV) incident

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

Hot Topics in Imaging Physics A Key Interdisciplinary Science for 21 st Century Medicine

Hot Topics in Imaging Physics A Key Interdisciplinary Science for 21 st Century Medicine Hot Topics in Imaging Physics A Key Interdisciplinary Science for 21 st Century Medicine Jeff Siewerdsen, PhD FAAPM Department of Biomedical Engineering Department of Computer Science Russell H. Morgan

More information

Splitting-Based Statistical X-Ray CT Image Reconstruction with Blind Gain Correction

Splitting-Based Statistical X-Ray CT Image Reconstruction with Blind Gain Correction Splitting-Based Statistical X-Ray CT Image Reconstruction with Blind Gain Correction Hung Nien and Jeffrey A. Fessler Department of Electrical Engineering and Computer Science University of Michigan, Ann

More information

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Improvement of Efficiency and Flexibility in Multi-slice Helical CT J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): 408 412 DOI: 10.1007/s12204-008-0408-x Improvement of Efficiency and Flexibility in Multi-slice Helical CT SUN Wen-wu 1 ( ), CHEN Si-ping 2 ( ), ZHUANG Tian-ge

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare

More information

Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J.

Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging Stephen J. Glick* J. Michael O Connor**, Clay Didier**, Mini Das*, * University of Massachusetts Medical

More information

Optimization of scanner parameters for dual energy micro-ct

Optimization of scanner parameters for dual energy micro-ct Optimization of scanner parameters for dual energy micro-ct E. PAUWELS* 1, J. DHAENE 1, A. DE MUYNCK 1 E., M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics and Astronomy, Ghent University, Proeftuinstraat

More information

Advanced Image Reconstruction Methods for Photoacoustic Tomography

Advanced Image Reconstruction Methods for Photoacoustic Tomography Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic

More information

Enhanced material contrast by dual-energy microct imaging

Enhanced material contrast by dual-energy microct imaging Enhanced material contrast by dual-energy microct imaging Method note Page 1 of 12 2 Method note: Dual-energy microct analysis 1. Introduction 1.1. The basis for dual energy imaging Micro-computed tomography

More information

Spectral CT reconstruction with an explicit photon-counting detector model: a one-step approach

Spectral CT reconstruction with an explicit photon-counting detector model: a one-step approach Spectral CT reconstruction with an explicit photon-counting detector model: a one-step approach Pierre-Antoine Rodesch, V. Rebuffel, C. Fournier, Florence Forbes, L. Verger To cite this version: Pierre-Antoine

More information

arxiv: v1 [physics.med-ph] 18 Apr 2016

arxiv: v1 [physics.med-ph] 18 Apr 2016 Segmentation-free x-ray energy spectrum estimation for computed tomography arxiv:164.4986v1 [physics.med-ph] 18 Apr 216 Wei Zhao a, Qiude Zhang a, Tianye Niu b a Department of Biomedical Engineering, Huazhong

More information

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Marc Kachelrieß, a Katia Sourbelle, and Willi A. Kalender Institute of Medical Physics, University of

More information

DUE to beam polychromacity in CT and the energy dependence

DUE to beam polychromacity in CT and the energy dependence 1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact

More information

Fast iterative beam hardening correction based on frequency splitting in computed tomography

Fast iterative beam hardening correction based on frequency splitting in computed tomography Fast iterative beam hardening correction based on frequency splitting in computed tomography Qiao Yang a,b, Matthias Elter b, Ingo Schasiepen b, Nicole Maass b and Joachim Hornegger a,c a Pattern Recognition

More information

Segmentation-free statistical image reconstruction for polyenergetic x-ray computed tomography with experimental validation

Segmentation-free statistical image reconstruction for polyenergetic x-ray computed tomography with experimental validation Segmentation-free statistical image reconstruction for polyenergetic x-ray computed tomography with experimental validation Idris A. Elbakri and Jeffrey A. Fessler Electrical Engineering and Computer Science

More information

Axial block coordinate descent (ABCD) algorithm for X-ray CT image reconstruction

Axial block coordinate descent (ABCD) algorithm for X-ray CT image reconstruction Axial block coordinate descent (ABCD) algorithm for X-ray CT image reconstruction Jeffrey A. Fessler and Donghwan Kim EECS Department University of Michigan Fully 3D Image Reconstruction Conference July

More information

[PDR03] RECOMMENDED CT-SCAN PROTOCOLS

[PDR03] RECOMMENDED CT-SCAN PROTOCOLS SURGICAL & PROSTHETIC DESIGN [PDR03] RECOMMENDED CT-SCAN PROTOCOLS WORK-INSTRUCTIONS DOCUMENT (CUSTOMER) RECOMMENDED CT-SCAN PROTOCOLS [PDR03_V1]: LIVE 1 PRESCRIBING SURGEONS Patient-specific implants,

More information

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Workshop on Quantitative SPECT and PET Brain Studies 14-16 January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Físico João Alfredo Borges, Me. Corrections in SPECT and PET SPECT and

More information

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh et al. : Proceedings of the National Seminar & Exhibition on Non-Destructive Evaluation DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh V Acharya, Umesh Kumar, Gursharan

More information

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004 Spiral ASSR Std p = 1.0 Spiral EPBP Std p = 1.0 Kachelrieß et al., Med. Phys. 31(6): 1623-1641, 2004 256 slices (0/300) Advantages of Cone-Beam Spiral CT Image quality nearly independent of pitch Increase

More information

Metal artifacts reduction in x-ray CT based on segmentation and forward-projection

Metal artifacts reduction in x-ray CT based on segmentation and forward-projection Bio-Medical Materials and Engineering 24 (2014) 3287 3293 DOI 10.3233/BME-141151 IOS Press 3287 Metal artifacts reduction in x-ray CT based on segmentation and forward-projection Shoukat Nawaz, Jian Fu

More information

CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona

CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona Introduction CBCT has become a routine procedure for

More information

Acknowledgements. Atlas-based automatic measurements of the morphology of the tibiofemoral joint

Acknowledgements. Atlas-based automatic measurements of the morphology of the tibiofemoral joint Atlas-based automatic measurements of the morphology of the tibiofemoral joint M Brehler 1, G Thawait 2, W Shyr 1, J Ramsay 3, JH Siewerdsen 1,2, W Zbijewski 1 1 Dept. of Biomedical Engineering, Johns

More information

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings)

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings) Marquette University e-publications@marquette Biomedical Engineering Faculty Research and Publications Engineering, College of 1-1-1999 An Iterative Approach to the Beam Hardening Correction in Cone Beam

More information

GPU implementation for rapid iterative image reconstruction algorithm

GPU implementation for rapid iterative image reconstruction algorithm GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial

More information

Tomographic Reconstruction

Tomographic Reconstruction Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)

More information

Non-Stationary CT Image Noise Spectrum Analysis

Non-Stationary CT Image Noise Spectrum Analysis Non-Stationary CT Image Noise Spectrum Analysis Michael Balda, Björn J. Heismann,, Joachim Hornegger Pattern Recognition Lab, Friedrich-Alexander-Universität Erlangen Siemens Healthcare, Erlangen michael.balda@informatik.uni-erlangen.de

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!

More information

Proton dose calculation algorithms and configuration data

Proton dose calculation algorithms and configuration data Proton dose calculation algorithms and configuration data Barbara Schaffner PTCOG 46 Educational workshop in Wanjie, 20. May 2007 VARIAN Medical Systems Agenda Broad beam algorithms Concept of pencil beam

More information

AIDR 3D Iterative Reconstruction:

AIDR 3D Iterative Reconstruction: Iterative Reconstruction: Integrated, Automated and Adaptive Dose Reduction Erin Angel, PhD Manager, Clinical Sciences, CT Canon Medical Systems USA Iterative Reconstruction 1 Since the introduction of

More information

Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage. 29 April - 3 May, 2013

Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage. 29 April - 3 May, 2013 2455-5 Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage 29 April - 3 May, 2013 Lecture NoteBasic principles of X-ray Computed Tomography Diego Dreossi Elettra, Trieste

More information

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),

More information

Efficient Monte Carlo based scatter artefact reduction in cone-beam micro-ct

Efficient Monte Carlo based scatter artefact reduction in cone-beam micro-ct Chapter 6 Efficient Monte Carlo based scatter artefact reduction in cone-beam micro-ct Abstract Cupping and streak artefacts caused by the detection of scattered photons may severely degrade the quantitative

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study

Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study B. De Man, J. Nuyts, P. Dupont, G. Marchal and P. Suetens Medical Image Computing, ESAT-PSI, K.U.Leuven, B-3000 Leuven, Belgium Department

More information

Central Slice Theorem

Central Slice Theorem Central Slice Theorem Incident X-rays y f(x,y) R x r x Detected p(, x ) The thick line is described by xcos +ysin =R Properties of Fourier Transform F [ f ( x a)] F [ f ( x)] e j 2 a Spatial Domain Spatial

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

Cardiac Dual Energy CT: Technique

Cardiac Dual Energy CT: Technique RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation 26 IEEE Nuclear Science Symposium Conference Record M6-349 Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation Habib Zaidi, Senior Member, IEEE and Mohammad

More information

Accelerated quantitative multi-material beam hardening correction(bhc) in cone-beam CT

Accelerated quantitative multi-material beam hardening correction(bhc) in cone-beam CT Accelerated quantitative multi-material beam hardening correction(bhc) in cone-beam CT Award: Poster No.: C-2161 Certificate of Merit Congress: ECR 2012 Type: Authors: Scientific Exhibit Q. Yang, M. Elter,

More information

(RMSE). Reconstructions showed that modeling the incremental blur improved the resolution of the attenuation map and quantitative accuracy.

(RMSE). Reconstructions showed that modeling the incremental blur improved the resolution of the attenuation map and quantitative accuracy. Modeling the Distance-Dependent Blurring in Transmission Imaging in the Ordered-Subset Transmission (OSTR) Algorithm by Using an Unmatched Projector/Backprojector Pair B. Feng, Member, IEEE, M. A. King,

More information

Estimating 3D Respiratory Motion from Orbiting Views

Estimating 3D Respiratory Motion from Orbiting Views Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct. 2005 Funding provided by NIH Grant P01 CA59827 Motivation Free-breathing

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

GE Healthcare. Agile Ultrasound. The Next Revolution in Ultrasound Imaging

GE Healthcare. Agile Ultrasound. The Next Revolution in Ultrasound Imaging Agile Ultrasound The Next Revolution in Ultrasound Imaging Abstract Diagnostic use of ultrasound has greatly expanded over the past couple of decades because it offers many advantages as an imaging modality.

More information

A new calibration-free beam hardening reduction method for industrial CT

A new calibration-free beam hardening reduction method for industrial CT A new calibration-free beam hardening reduction method for industrial CT ECC 2 for industrial CT Tobias Würfl 1, Nicole Maaß 2, Frank Dennerlein 2, Andreas K. Maier 1 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg;

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

Scaling Calibration in the ATRACT Algorithm

Scaling Calibration in the ATRACT Algorithm Scaling Calibration in the ATRACT Algorithm Yan Xia 1, Andreas Maier 1, Frank Dennerlein 2, Hannes G. Hofmann 1, Joachim Hornegger 1,3 1 Pattern Recognition Lab (LME), Friedrich-Alexander-University Erlangen-Nuremberg,

More information

Refraction Corrected Transmission Ultrasound Computed Tomography for Application in Breast Imaging

Refraction Corrected Transmission Ultrasound Computed Tomography for Application in Breast Imaging Refraction Corrected Transmission Ultrasound Computed Tomography for Application in Breast Imaging Joint Research With Trond Varslot Marcel Jackowski Shengying Li and Klaus Mueller Ultrasound Detection

More information

An approximate cone beam reconstruction algorithm for gantry-tilted CT

An approximate cone beam reconstruction algorithm for gantry-tilted CT An approximate cone beam reconstruction algorithm for gantry-tilted CT Ming Yan a, Cishen Zhang ab, Hongzhu Liang a a School of Electrical & Electronic Engineering, Nanyang Technological University, Singapore;

More information

Reduction of metal streak artifacts in x-ray computed tomography using a transmission maximum a posteriori algorithm

Reduction of metal streak artifacts in x-ray computed tomography using a transmission maximum a posteriori algorithm Reduction of metal streak artifacts in x-ray computed tomography using a transmission maximum a posteriori algorithm B. De Man 1, Student Member, IEEE, J. Nuyts 2, Member, IEEE,. Dupont 2, G. Marchal 3,

More information

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel Proc. Of SPIE Vol 0626, Application of Optical Instrumentation in Medicine XIV and Picture Archiving and Communication Systems(PACS IV) for Medical Applications IV, ed. S J Dwyer/R H Schneider (Jan 1986)

More information

Some reference material

Some reference material Some reference material Physics reference book on medical imaging: A good one is The Essential Physics of Medical Imaging, 3 rd Ed. by Bushberg et al. ($170! new). However, there are several similar books

More information

Towards full-body X-ray images

Towards full-body X-ray images Towards full-body X-ray images Christoph Luckner 1,2, Thomas Mertelmeier 2, Andreas Maier 1, Ludwig Ritschl 2 1 Pattern Recognition Lab, FAU Erlangen-Nuernberg 2 Siemens Healthcare GmbH, Forchheim christoph.luckner@fau.de

More information

Segmentation-Free Quasi-Newton Method for Polyenergetic CT Reconstruction

Segmentation-Free Quasi-Newton Method for Polyenergetic CT Reconstruction Segmentation-Free Quasi-Newton Method for Polyenergetic CT Reconstruction T. Humphries and A. Faridani Abstract X-ray polychromaticity is a well-known source of artifacts in clinical CT imaging. As a polyenergetic

More information

F3-A5: Toward Model-Based Reconstruction in Scanned Baggage Security Applications

F3-A5: Toward Model-Based Reconstruction in Scanned Baggage Security Applications F3-A5: Toward Model-Based Reconstruction in Scanned Baggage Security Applications Abstract While traditional direct reconstruction algorithms such as filtered back projection depend on the analytic inversion

More information

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct J. DHAENE* 1, E. PAUWELS 1, T. DE SCHRYVER 1, A. DE MUYNCK 1, M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics

More information

A Curvelet based Sinogram Correction Method for Metal Artifact Reduction

A Curvelet based Sinogram Correction Method for Metal Artifact Reduction A based Sinogram Correction Method for Metal Artifact Reduction Kiwan Jeon 1 and Hyoung Suk Park 1 More info about this article: http://www.ndt.net/?id=3715 1 National Institute for Mathematical Sciences,

More information

Adaptive algebraic reconstruction technique

Adaptive algebraic reconstruction technique Adaptive algebraic reconstruction technique Wenkai Lua) Department of Automation, Key State Lab of Intelligent Technology and System, Tsinghua University, Beijing 10084, People s Republic of China Fang-Fang

More information

Unmatched Projector/Backprojector Pairs in an Iterative Reconstruction Algorithm

Unmatched Projector/Backprojector Pairs in an Iterative Reconstruction Algorithm 548 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 19, NO. 5, MAY 2000 Unmatched Projector/Backprojector Pairs in an Iterative Reconstruction Algorithm Gengsheng L. Zeng*, Member, IEEE, and Grant T. Gullberg,

More information

Noise weighting with an exponent for transmission CT

Noise weighting with an exponent for transmission CT doi:10.1088/2057-1976/2/4/045004 RECEIVED 13 January 2016 REVISED 4 June 2016 ACCEPTED FOR PUBLICATION 21 June 2016 PUBLISHED 27 July 2016 PAPER Noise weighting with an exponent for transmission CT Gengsheng

More information

Prospects for spectral CT with Medipix detectors

Prospects for spectral CT with Medipix detectors Nikhef, Amsterdam, The Netherlands E-mail: e.schioppa@nikhef.nl Jan Visser Nikhef, Amsterdam, The Netherlands E-mail: janvs@nikhef.nl Els Koffeman Nikhef, Amsterdam, The Netherlands E-mail: e.koffeman@nikhef.nl

More information

Philips SPECT/CT Systems

Philips SPECT/CT Systems Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET

More information

Accelerating CT Iterative Reconstruction Using ADMM and Nesterov s Methods

Accelerating CT Iterative Reconstruction Using ADMM and Nesterov s Methods Accelerating CT Iterative Reconstruction Using ADMM and Nesterov s Methods Jingyuan Chen Meng Wu Yuan Yao June 4, 2014 1 Introduction 1.1 CT Reconstruction X-ray Computed Tomography (CT) is a conventional

More information

Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation

Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation Bastian Bier 1, Chris Schwemmer 1,2, Andreas Maier 1,3, Hannes G. Hofmann 1, Yan Xia 1, Joachim Hornegger 1,2, Tobias Struffert

More information

RADIOLOGY AND DIAGNOSTIC IMAGING

RADIOLOGY AND DIAGNOSTIC IMAGING Day 2 part 2 RADIOLOGY AND DIAGNOSTIC IMAGING Dr hab. Zbigniew Serafin, MD, PhD serafin@cm.umk.pl 2 3 4 5 CT technique CT technique 6 CT system Kanal K: RSNA/AAPM web module: CT Systems & CT Image Quality

More information

Translational Computed Tomography: A New Data Acquisition Scheme

Translational Computed Tomography: A New Data Acquisition Scheme 2nd International Symposium on NDT in Aerospace 2010 - We.1.A.3 Translational Computed Tomography: A New Data Acquisition Scheme Theobald FUCHS 1, Tobias SCHÖN 2, Randolf HANKE 3 1 Fraunhofer Development

More information

Quantitative Attenuation Correction for PET/CT Using Iterative Reconstruction of Low-Dose Dual-Energy CT

Quantitative Attenuation Correction for PET/CT Using Iterative Reconstruction of Low-Dose Dual-Energy CT Quantitative Attenuation Correction for PET/CT Using Iterative Reconstruction of Low-Dose Dual-Energy CT Paul E. Kinahan, Senior Member, IEEE, Jeffrey A. Fessler, Senior Member, IEEE, Adam M. Alessio,

More information

Carbondale, IL USA; University of North Carolina Chapel Hill, NC USA; USA; ABSTRACT

Carbondale, IL USA; University of North Carolina Chapel Hill, NC USA; USA; ABSTRACT Pre-computed backprojection based penalized-likelihood (PPL) reconstruction with an edge-preserved regularizer for stationary Digital Breast Tomosynthesis Shiyu Xu a, Christy Redmon Inscoe b, Jianping

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information