A software tool for the quantitative evaluation of 3D dose calculation algorithms

Size: px
Start display at page:

Download "A software tool for the quantitative evaluation of 3D dose calculation algorithms"

Transcription

1 A software tool for the quantitative evaluation of 3D dose calculation algorithms William B. Harms, Sr., Daniel A. Low, John W. Wong, a) and James A. Purdy Washington University School of Medicine, Mallinckrodt Institute of Radiology, 510 S. Kingshighway Boulevard, St. Louis, Missouri Received 30 October 1997; accepted for publication 17 July 1998 Current methods for evaluating modern radiation therapy treatment planning RTP systems include the manual superposition of calculated and measured isodose curves and the comparison of a limited number of calculated and measured point doses. Both techniques have significant limitations in providing quantitative evaluations of the large number of dose data generated by modern RTP systems. More sophisticated comparison techniques have been presented in the literature, including dose-difference and distance-to-agreement DTA analyses. A software tool has been developed that uses superimposed isodose plots, dose-difference, and DTA distributions to quantify errors in computed dose distributions. Dose-difference and DTA analyses are overly sensitive in regions of high- and low-dose gradient, respectively. The logical union of locations that fail both dose-difference and DTA acceptance criteria, termed the composite evaluation, is calculated and displayed. The composite evaluation provides a method for the physicist to efficiently identify regions that fail both the dose-difference and DTA acceptance criteria. The tool provides a computer platform for the quantitative comparison of calculated and measured dose distributions American Association of Physicists in Medicine. S Key words: 3D radiation therapy treatment planning, dose calculation, dose calculation verification Dose computation is one of the most important features of radiation therapy treatment planning systems. As the technologies of both computer hardware and software used to model radiation therapy beams improve, it becomes apparent that improved quantitative methods for the evaluation of three-dimensional 3D dose calculation programs are required. The technique used most frequently by radiation oncology physicists to compare measured and calculated dose distributions is the superposition of measured and calculated isodose curves, with a subsequent qualitative assessment of the acceptability of the calculation algorithm. This method does not provide a quantitative assessment of the accuracy of the calculation program. Determination and presentation of the numerical difference between the measured and calculated dose distributions highlight regions of disagreement. Mah et al. 1 used dosedifference distributions to evaluate two-dimensional 2D and 3D electron beam dose calculation algorithms. Others have used color wash presentations and dose-difference histograms. 2 4 One difficulty with dose-difference distributions is that they are very sensitive in high-dose gradient regions. In the beam penumbra region, a small spatial offset between the calculated and measured dose distributions will result in a large, but clinically insignificant, dose difference. To provide a more suitable evaluation of the dose calculation algorithm in regions of high-dose gradient, the distance-to-agreement DTA concept was developed. 5 8 The DTA is the distance between a measured dose point and the nearest point in the calculated distribution containing the same dose value. The DTA distribution provides an excellent measure of the calculation quality in regions of high-dose gradient but is overly sensitive in regions of low-dose gradient, a slight difference in dose e.g., due to a difference in normalization or block transmission yields a large DTA value. Van Dyk et al. 9 provided guidelines for the evaluation of treatment planning systems. They recommended that the percent dose difference and DTA be less than 3% and 4 mm for photon beams in regions of low- and high-dose gradient 30% cm 1, respectively. For electrons, the recommended criteria were 4% and 4 mm, respectively, with the same gradient specification and definition. In each case, these criteria can be applied to the dose-difference and DTA distributions to identify regions that fail the criteria. These are consistent with the aforementioned regions the dose-difference and DTA distributions are not overly sensitive. If these guidelines are followed, a logical extension to the dose difference and DTA is a simultaneous examination of the dosedifference and DTA criteria. 8,10 A software tool has been developed to support 3D radiation therapy treatment planning system testing 11 by implementing the dose-distribution superposition, dose-difference, and DTA analyses described above. In addition, a composite distribution is developed based on the concepts of Shiu et al. 8 as applied by Cheng et al. 10 The dose-difference and DTA tests are overly sensitive in regions of high- and lowdose gradients, respectively. The composite distribution analysis shows only regions that fail both criteria, and is therefore is not overly sensitive in these regions. For example, in regions of high-dose gradient, the dose-difference 1830 Med. Phys , October /98/25 10 /1830/7/$ Am. Assoc. Phys. Med. 1830

2 1831 Harms et al.: Tool for the evaluation of algorithms 1831 distribution yields large values for small spatial offsets between the tested dose distributions. The DTA analysis will return the approximate spatial difference between the two distributions. If the spatial offset is less than the criteria, the DTA analysis will pass, and the algorithm will have passed the composite analysis. The software tool compares measured and calculated planar dose distributions for identical beam and phantom geometries. Because of limitations of dose-distribution measurement and display devices, comparisons are made on a planar basis and calculated dose distributions are not examined beyond the comparison plane. Because the measured distribution will, in general, have fewer data points, comparisons are made for each measurement point, and the comparison quantities are determined on the measurement coordinate system. The dose-difference, (r m ), at position r m on the measured distribution is equal to r m r m,r c r m r c, r m,r c D m r m D c r c is the general difference between dose values on the measured, D m (r m ), and calculated, D c (r c ), distributions at r m and r c, respectively. The DTA, d(r m ), is given by d r m min r 0 r m,r c r c, r 0 r m,r c r r m,r c r m,r c 0 is the distance between the set of points that have a difference of zero between calculated and measured doses and r r m,r c r m r c. The composite distribution c(r m ) is a binary distribution formed by the points that fail both the dose-difference and DTA criteria, D M and d M, respectively. c r m f r m d f r m, and f r m d f r m r m D M 7 1 r m D M 0 d r m d M 8 1 d r m d M are the points that fail the dose-difference and DTA criteria, respectively. The criteria D M and d M selected for the examples are 3% and 3 mm, respectively, the dose difference is in percent of a selected normalization value e.g., the dose at d max. The selection of d M 3mm is based on our internal quality assurance criteria. A binary distribution does not easily lend itself to interpretation, so the dose-difference distribution is presented in the regions c(r m ) 1. The analysis programs are written within the confines of a commercial scanning dosimetry system software platform DynaScan, Computerized Medical Systems, Inc., St. Louis, MO. It provides the ability to develop only the tools necessary for computing the required distributions while allowing the use of existing software to handle the tasks of displaying and plotting the isodose, isodose difference, iso-dta, and composite distributions. An additional program is written on a UNIX platform to extract the desired planar dose matrices from computed 3D dose distributions. Ideally, the entire 3D calculated dose distribution would be used to generate the comparisons. However, software limitations restrict the comparison to 2D planar extractions. The calculated dose-distribution planar extraction routines include software that 1 allows the user to specify the dose plane to be extracted from the 3D dose matrix, 2 extracts the dose for the specified plane, and 3 writes an intermediate text data file containing the planar dose array for import by the dosimetry system software. The data files are transferred from the 3D RTP system 11 to the dosimetry system by means of a computer network. To facilitate the dose-difference computations, the measured and calculated dose-distribution datasets are resampled onto a common rectilinear grid. This common grid is constructed from the spatial intersection of the grids from both datasets. For transverse photon distributions, the spacing of the grid points is 0.2 cm across the beam and 0.5 cm along the beam. A grid spacing of 0.2 cm is used along both axes for electron beams and beam s eye view photon distributions, and will be used in subsequent tests of intensity modulated dose distributions. The coarser spacing for transverse photon distributions is used due to the relatively small dose gradient along the depth direction. To ensure correct divergence for transverse distributions, the resampling is performed along ray lines intersecting the rectilinear matrix points. An additional feature of this resampling is that both matrices are rescaled so that their dose values are a percentage of the normalization dose. The dose-difference distribution is calculated for each point on the common grid. Software programs are added to the dosimetry system software to facilitate implementation of the quantitative dose analysis. These routines include programs to 1 convert the common text file format into the internal file format of the dosimetry system, 2 resample the measured and calculated planar dose matrices onto a common, regularly spaced grid, 3 compute dose-difference distributions (r m ), 4 compute DTA distributions d(r m ), and 5 generate composite distributions c(r m ) to demonstrate regions outside the selected acceptance limits. The DTA calculation is made point-by-point on the common grid. The DTA value for a specific measurement dose point is determined by searching the calculated dose matrix for the same dose value. In general, the calculated dose points bracket the measured dose value. A method is used that rapidly interpolates the location of the calculated dose value on the calculation matrix. Figure 1 shows an example of the DTA calculation, the DTA is to be determined for point r m (x m,y m ) with dose value D m (x m,y m ). The

3 1832 Harms et al.: Tool for the evaluation of algorithms 1832 FIG. 1. Drawing of the distance-to-agreement calculation geometry. The ideal isodose line is shown, along with the intersections, I x and I y, used to calculate the list of distances, r 0 (r m,r c ), from which the minimum, d(r m ), is selected. FIG. 2. Beam s eye view of mantle portal blocking geometry. The tic marks indicate 1 cm spacing projected to the plane of isocenter. The dashed cut line shows the intersection with the transverse plane evaluated in Figs. 3 and 4. The dark and light lung blocks are 1 HVL and 5 HVL, respectively. calculation points r c under investigation in this example are (x c,y c ), (x c x,y c ), and (x c,y c y) with doses equal to D c (x c,y c ), D c (x c x,y c ), and D c (x c,y c y), respectively, x and y are the matrix spacing in the x and y directions, respectively. The interpolated intersections of the D m (x m,y m ) isodose line with the calculation matrix are identified in Fig. 1 by I x and I y. The positions of I x and I y are determined by linear interpolation of the calculation matrix and I x x c D m D c x c,y c M x I y y c D m D c x c,y c M y, 10 and M x D c x c x,y c D c x c,y c x 9 11 M y D c x c,y c y D c x c,y c. 12 y It is possible that the intersection coordinates I x and I y of the isodose lines with the matrix segments do not lie between the points (x c,y c ), (x c x,y c ), and (x c,y c y). A test is applied to I x and I y to determine if they are suitable for inclusion in the list of values r 0 (r m,r c ) by inspecting their values and determining if they lie between (x c,y c ), (x c x,y c ), and (x c,y c y). The DTA search is not conducted over the entire calculation space. A DTA value greater than 1.0 cm is considered excessive usually associated with a low dose-gradient region, so the search is terminated when no suitable point in the calculation matrix is located within 1.0 cm of the measurement point. The value of 1.0 cm is used as the DTA in these cases. To show the tool s effectiveness, two photon beam dose calculation algorithms are presented. 12,13 The measured dose distribution from a 6 MV photon beam with a mantle field portal shown in Fig. 2 is compared against calculations using the convolution adapted ratio tissue air ratio CAR- TAR algorithm 13 and the ratio-tar RTAR algorithm. The CARTAR algorithm models block penumbras, while the block penumbra of the RTAR algorithm is a discontinuous change in dose level. The mantle block contains both fully attenuating 5 half-value layers HVL and partial transmission 1 HVL lung blocks. The results for the CARTAR mantle calculation are shown in Fig. 3. Regions of interest have been identified by letters A to C. Region A is at the left-most collimator-defined penumbra; region B lies at the left edge of the central open field portion of the beam, and region C lies beneath the 1 HVL lung block. Figure 3 a shows the superimposed isodose distributions. Quantitative assessment of the dose calculation quality is difficult using only this information. The dose-difference distribution is shown in Fig. 3 b and highlights regions the dose distributions disagree by more than 3%, principally within the block and collimator penumbra regions A and B. The DTA distribution shown in Fig. 3 c appears more complex than the dose-difference distribution. Region A exhibits a DTA of more than 3 mm, while region B does not, indicating that a discrepancy of more than 3 mm exists between the calculated and measured collimator penumbra positions, but that the block positions and penumbra are accurately modeled. Region C contains DTA values of up to the 1.0 cm limit the software is written to display the results in mm due to the relatively small dose gradient in this region. The composite distribution is shown in Fig. 3 d. As previously mentioned, the dose difference is selected as the plotted quantity. The composite distribution clearly shows that the dose calculation fails in region A, but passes in regions B and C. This identifies the need to further investigate the collimator positions and penumbra. Additional regions of disagreement lie near the surface within the build-up region.

4 1833 Harms et al.: Tool for the evaluation of algorithms 1833 Figure 4 shows the RTAR algorithm results. Figure 4 a shows the superimposed isodose distributions. Region A appears similar to that for the CARTAR algorithm, due to the collimator penumbra model in the RTAR algorithm. The isodose lines near the block penumbra, however, have an unrealistic square shape. It is clear that the RTAR algorithm does not model the block penumbra as well as the CARTAR algorithm. As in Fig. 3, quantitative assessment is difficult. FIG. 3. a Superimposed measured and calculated isodoses for the transverse, off-axis plane shown in Fig. 2. The convolution-adapted ratio-tar algorithm is used, and doses are shown as a percent of the central axis dose at d max. The measurement is plotted with a solid line and the calculation with a dashed line. b Dosedifference plot for distributions shown in a. The curves indicate the percentage error relative to the normalization dose value of 100%. c Distance-toagreement distribution in mm for dose distributions shown in a. d Composite distribution demonstrating regions in which the dose difference and distance-to-agreement both exceed acceptance limits of 3% and 0.3 cm, respectively.

5 1834 Harms et al.: Tool for the evaluation of algorithms 1834 FIG. 3 Continued. Figure 4 b shows the composite distribution. Region A appears similar to that of the CARTAR algorithm see Fig. 3 d. The CARTAR algorithm passes in region B, but the RTAR algorithm does not pass, a fact made clear by the composite distribution. The RTAR algorithm passes in region C due to the relatively low dose level and subsequent small dose difference. This software tool has been extremely effective for plane-

6 1835 Harms et al.: Tool for the evaluation of algorithms 1835 FIG. 4. a Superimposed measured and calculated isodoses, for the transverse, off-axis plane shown in Fig. 2. The calculation uses the ratio-tar algorithm. The measurement is plotted with a solid line and the calculation with a dashed line. b Composite distribution demonstrating areas in which the dose difference and distance-toagreement both exceed the acceptance limits of 3% and 0.3 cm, respectively. The curves indicate percent dose difference relative to the normalization dose value of 100%. by-plane analysis of 3D dose calculation software. The traditional method of dose calculation software review, in which the evaluation is limited to a subjective review of the superposition of isodoses, is inadequate for commissioning 3D treatment planning systems. This tool provides an effective means for the physicist to clearly identify and document the commissioning of dose calculation software. While the current software compares only calculation planes that coin-

7 1836 Harms et al.: Tool for the evaluation of algorithms 1836 cide with the measured planes, in the future, the DTA distribution will consider calculation points within the entire 3D dose calculation volume. Manufacturers of isodose scanning systems are encouraged to work with 3D treatment planning manufacturers to implement the type of quantitative analysis tools described in this report. a Also at William Beaumont Hospital, Royal Oak, MI. 1 E. Mah, J. Antolak, J. W. Scrimger, and J. J. Battista, Experimental evaluation of a 2D and 3D electron pencil beam algorithm, Phys. Med. Biol. 34, B. A. Fraass, Quality assurance for 3D treatment planning, in Teletherapy: Present and Future Advanced Medical Publishing, Vancouver, B.C., B. A. Fraass and D. McShan, Three-dimensional photon beam treatment planning, in Radiation Therapy Physics, edited by A. R. Smith Springer-Verlag, New York, B. A. Fraass, M. K. Martel, and D. L. McShan, Tools for dose calculation verification and QA for conformal therapy treatment techniques, in XIth International Conference on the use of Computers in Radiation Therapy Medical Physics Publishing, Madison, WI, K. R. Hogstrom, M. D. Mills, J. A. Meyer, J. R. Palta, D. E. Mellenberg, R. T. Moez, and R. S. Fields, Dosimetric evaluation of a pencil-beam algorithm for electrons employing a two-dimensional heterogeneity correction, Int. J. Radiat. Oncol., Biol., Phys. 10, ICRU, Report 42, Use of Computers in External Beam Radiotherapy Procedures with High-Energy Photons and Electrons ICRU, H. Dahlin, I. L. Lamm, T. Landberg, S. Levernes, and N. Ulso, User requirements on CT based computerized dose planning systems in radiotherapy, Acta Radiol.: Oncol. 22, A. S. Shiu, S. Tung, K. R. Hogstrom, J. W. Wong, R. L. Gerber, W. B. Harms, J. A. Purdy, R. K. TenHaken, D. L. McShan, and B. A. Fraass, Verification data for electron beam dose algorithms, Med. Phys. 19, J. Van Dyk, R. B. Barnett, J. E. Cygler, and P. C. Shragge, Commissioning and quality assurance of treatment planning computers, Med. Phys. 26, A. Cheng, W. B. Harms, R. L. Gerber, J. W. Wong, and J. A. Purdy, Systematic verification of a three-dimensional electron beam dose calculation algorithm, Med. Phys. 23, J. A. Purdy, W. B. Harms, J. W. Matthews, R. E. Drzymala, B. Emami, J. R. Simpson, J. M. Manolis, F. U. Rosenburger, Advances in 3- dimensional radiation treatment planning systems: Room-view display with real-time interactivity, Int. J. Radiat. Oncol., Biol., Phys. 27, D. A. Low, X. R. Zhu, W. B. Harms, J. W. Matthews, J. A. Purdy, A. Cheng, and R. L. Gerber, Evaluation of ratio-tar photon and electron pencil beam dose calculation algorithms for 3D treatment planning, in XI International Conference on the Use of Computers in Radiation Therapy Manchester, England, X. R. Zhu, D. A. Low, W. B. Harms, and J. A. Purdy, A convolution adapted ratio-tar algorithm for 3D photon beam treatment planning, Med. Phys. 22,

Gamma Dose Distribution Evaluation Tool

Gamma Dose Distribution Evaluation Tool Journal of Physics: Conference Series Gamma Dose Distribution Evaluation Tool To cite this article: Daniel A Low PhD 2010 J. Phys.: Conf. Ser. 250 012071 View the article online for updates and enhancements.

More information

Basic Radiation Oncology Physics

Basic Radiation Oncology Physics Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides

More information

4 Measurement. and Analysis. 4.1 Overview and Underlying Principles 4-1

4 Measurement. and Analysis. 4.1 Overview and Underlying Principles 4-1 Measurement and Analysis.1 Overview and Underlying Principles.1.1 Introductory Remarks The physics and setup for film dosimetry have been described in the previous chapters. The measurement setup for IMRT

More information

Proton dose calculation algorithms and configuration data

Proton dose calculation algorithms and configuration data Proton dose calculation algorithms and configuration data Barbara Schaffner PTCOG 46 Educational workshop in Wanjie, 20. May 2007 VARIAN Medical Systems Agenda Broad beam algorithms Concept of pencil beam

More information

Calibration and quality assurance for rounded leaf-end MLC systems

Calibration and quality assurance for rounded leaf-end MLC systems Calibration and quality assurance for rounded leaf-end MLC systems Maria N. Graves, a) Antoinette V. Thompson, b) Mary K. Martel, Daniel L. McShan, and Benedick A. Fraass Department of Radiation Oncology,

More information

A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study

A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study Indrin J. Chetty, a) Mihaela Rosu, Neelam Tyagi, Lon H. Marsh, Daniel

More information

Calculating percent depth dose with the electron pencilbeam redefinition algorithm

Calculating percent depth dose with the electron pencilbeam redefinition algorithm JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 8, NUMBER 2, SPRING 2007 Calculating percent depth dose with the electron pencilbeam redefinition algorithm Michael J. Price, 1,2,a Kenneth R. Hogstrom,

More information

FAST, precise. qa software

FAST, precise. qa software qa software FAST, precise Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, rotational or brachytherapy plans no film, no phantoms, no linac

More information

A simple method to test geometrical reliability of digital reconstructed radiograph (DRR)

A simple method to test geometrical reliability of digital reconstructed radiograph (DRR) JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 1, WINTER 2010 A simple method to test geometrical reliability of digital reconstructed radiograph (DRR) Stefania Pallotta, a Marta Bucciolini

More information

Determination of rotations in three dimensions using two-dimensional portal image registration

Determination of rotations in three dimensions using two-dimensional portal image registration Determination of rotations in three dimensions using two-dimensional portal image registration Anthony E. Lujan, a) James M. Balter, and Randall K. Ten Haken Department of Nuclear Engineering and Radiological

More information

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters Monte Carlo Simulation for Neptun 1 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters M.T. Bahreyni Toossi a, M. Momen Nezhad b, S.M. Hashemi a a Medical Physics Research Center,

More information

Effect of Statistical Fluctuation in Monte Carlo Based Photon Beam Dose Calculation on Gamma Index Evaluation

Effect of Statistical Fluctuation in Monte Carlo Based Photon Beam Dose Calculation on Gamma Index Evaluation Effect of Statistical Fluctuation in Monte Carlo Based Photon Beam Dose Calculation on Gamma Index Evaluation Yan Jiang Graves 1,2, Xun Jia 1, Steve B. Jiang 1 1 Center for Advanced Radiotherapy Technologies

More information

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them Dose Distributions George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Purpose To familiarize the resident with dose distributions and the factors that affect them

More information

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Krista Chytyk MSc student Supervisor: Dr. Boyd McCurdy Introduction The objective of cancer radiotherapy

More information

Three-dimensional gamma criterion for patient-specific quality assurance of spot scanning proton beams

Three-dimensional gamma criterion for patient-specific quality assurance of spot scanning proton beams JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 5, 2015 Three-dimensional gamma criterion for patient-specific quality assurance of spot scanning proton beams Chang Chang, a Kendra L. Poole,

More information

Facility Questionnaire PART I (General Information for 3DCRT and IMRT)

Facility Questionnaire PART I (General Information for 3DCRT and IMRT) Facility Questionnaire PART I (General Information for 3DCRT and IMRT) The following items are required before you can enter cases on any RTOG protocol that requires data submission to the Image-Guided

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

Clinical implementation of photon beam flatness measurements to verify beam quality

Clinical implementation of photon beam flatness measurements to verify beam quality JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Clinical implementation of photon beam flatness measurements to verify beam quality Simon Goodall, a Nicholas Harding, Jake Simpson,

More information

Photon beam dose distributions in 2D

Photon beam dose distributions in 2D Photon beam dose distributions in 2D Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2014 Acknowledgments! Narayan Sahoo PhD! Richard G Lane (Late) PhD 1 Overview! Evaluation

More information

A SYSTEM OF DOSIMETRIC CALCULATIONS

A SYSTEM OF DOSIMETRIC CALCULATIONS A SYSTEM OF DOSIMETRIC CALCULATIONS INTRODUCTION Dose calculation based on PDD and TAR have Limitations The dependence of PDD on SSD Not suitable for isocentric techniques TAR and SAR does not depend on

More information

Michael Speiser, Ph.D.

Michael Speiser, Ph.D. IMPROVED CT-BASED VOXEL PHANTOM GENERATION FOR MCNP MONTE CARLO Michael Speiser, Ph.D. Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX September 1 st, 2012 CMPWG Workshop Medical

More information

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization

More information

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS Tomotherapy Physics Machine Twinning and Quality Assurance Emilie Soisson, MS Tomotherapy at UW- Madison Treating for nearly 5 years Up to ~45 patients a day on 2 tomo units Units twinned to facilitate

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

The MSKCC Approach to IMRT. Outline

The MSKCC Approach to IMRT. Outline The MSKCC Approach to IMRT Spiridon V. Spirou, PhD Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York, NY Outline Optimization Field splitting Delivery Independent verification

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

Verification of dose calculations with a clinical treatment planning system based on a point kernel dose engine

Verification of dose calculations with a clinical treatment planning system based on a point kernel dose engine JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 Verification of dose calculations with a clinical treatment planning system based on a point kernel dose engine Lars Weber*

More information

IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays

IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays Sotiri Stathakis Outline Why IMRT/VMAT QA AAPM TG218 UPDATE Tolerance Limits and Methodologies for IMRT Verification QA Common sources

More information

Evaluation of 3D Gamma index calculation implemented in two commercial dosimetry systems

Evaluation of 3D Gamma index calculation implemented in two commercial dosimetry systems University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2015 Evaluation of 3D Gamma index calculation implemented

More information

The Jacobian as a measure of planar dose congruence

The Jacobian as a measure of planar dose congruence The Jacobian as a measure of planar dose congruence L. D.Paniak 1 and P. M. Charland 2 1 4 Solutions Inc, Kitchener, Ontario 2 Grand River Hospital, Kitchener, Ontario and the University of Waterloo, Waterloo,

More information

A graphical user interface for an electron monitor unit calculator using a sector-integration algorithm and exponential curve-fitting method

A graphical user interface for an electron monitor unit calculator using a sector-integration algorithm and exponential curve-fitting method JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 A graphical user interface for an electron monitor unit calculator using a sector-integration algorithm and exponential curve-fitting

More information

ICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588

ICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588 ICARO Vienna April 27-29 2009 Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588 M. Saiful Huq, Ph.D., Professor and Director, Dept. of Radiation

More information

Analysis and evaluation of planned and delivered dose distributions: practical concerns with γ- and χ- Evaluations

Analysis and evaluation of planned and delivered dose distributions: practical concerns with γ- and χ- Evaluations Journal of Physics: Conference Series OPEN ACCESS Analysis and evaluation of planned and delivered dose distributions: practical concerns with γ- and χ- Evaluations To cite this article: L J Schreiner

More information

PyCMSXiO: an external interface to script treatment plans for the Elekta CMS XiO treatment planning system

PyCMSXiO: an external interface to script treatment plans for the Elekta CMS XiO treatment planning system Journal of Physics: Conference Series OPEN ACCESS PyCMSXiO: an external interface to script treatment plans for the Elekta CMS XiO treatment planning system To cite this article: Aitang Xing et al 2014

More information

A DOSIMETRIC MODEL FOR SMALL-FIELD ELECTRON RADIATION THERAPY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

A DOSIMETRIC MODEL FOR SMALL-FIELD ELECTRON RADIATION THERAPY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL A DOSIMETRIC MODEL FOR SMALL-FIELD ELECTRON RADIATION THERAPY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE MASTER OF

More information

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE QA SOFTWARE FAST, PRECISE Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, VMAT, SRS and brachytherapy plans no film, no phantoms, no linac

More information

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. 2-D D Dose-CT Mapping in Geant4 Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. Table of Contents Background & Purpose Materials Methods

More information

An Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields

An Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields Int. J. Cancer (Radiat. Oncol. Invest): 96, 140 145 (2001) 2001 Wiley-Liss, Inc. Publication of the International Union Against Cancer An Investigation of a Model of Percentage Depth Dose for Irregularly

More information

Disclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions

Disclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions LINAC and MLC QA for IMRT Ping Xia, Ph.D., Department of Radiation Oncology Disclosure Received research support from Siemens Medical Solutions University of California San Francisco Therapy Series (SAM)

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Evaluation of dosimetric effect of leaf position in a radiation field of an 80-leaf multileaf collimator fitted to the LINAC

More information

ELECTRON DOSE KERNELS TO ACCOUNT FOR SECONDARY PARTICLE TRANSPORT IN DETERMINISTIC SIMULATIONS

ELECTRON DOSE KERNELS TO ACCOUNT FOR SECONDARY PARTICLE TRANSPORT IN DETERMINISTIC SIMULATIONS Computational Medical Physics Working Group Workshop II, Sep 30 Oct 3, 2007 University of Florida (UF), Gainesville, Florida USA on CD-ROM, American Nuclear Society, LaGrange Park, IL (2007) ELECTRON DOSE

More information

3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending

3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending 3DVH : On The Accuracy Of The Planned Dose Perturbation Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools *Patent Pending introduction State-of-the-art IMRT QA of static gantry

More information

Dynalog data tool for IMRT plan verification

Dynalog data tool for IMRT plan verification Dynalog data tool for IMRT plan verification Poster No.: R-0051 Congress: 2014 CSM Type: Scientific Exhibit Authors: V. Sashin; FOOTSCRAY/AU Keywords: Computer applications, Radiation physics, Experimental,

More information

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Artifact Mitigation in High Energy CT via Monte Carlo Simulation PIERS ONLINE, VOL. 7, NO. 8, 11 791 Artifact Mitigation in High Energy CT via Monte Carlo Simulation Xuemin Jin and Robert Y. Levine Spectral Sciences, Inc., USA Abstract The high energy (< 15 MeV) incident

More information

Acceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems. Bruce Curran University of Michigan Medical Center Ann Arbor, MI

Acceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems. Bruce Curran University of Michigan Medical Center Ann Arbor, MI Acceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems Bruce Curran University of Michigan Medical Center Ann Arbor, MI Disclosure Some of the work discussed in this talk was supported

More information

Verification of Varian enhanced dynamic wedge implementation in MasterPlan treatment planning system

Verification of Varian enhanced dynamic wedge implementation in MasterPlan treatment planning system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 2, SPRING 2009 Verification of Varian enhanced dynamic wedge implementation in MasterPlan treatment planning system Massimo Pasquino, 1,a

More information

Advantages of multiple algorithm support in treatment planning system for external beam dose calculations

Advantages of multiple algorithm support in treatment planning system for external beam dose calculations Review Article Free full text available from www.cancerjournal.net Advantages of multiple algorithm support in treatment planning system for external beam dose calculations Animesh Kirloskar Theratronics

More information

An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy

An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy Chenyang Xu 1, Siemens Corporate Research, Inc., Princeton, NJ, USA Xiaolei Huang,

More information

Implementation of an integrated quality assurance program for CT-TPS process

Implementation of an integrated quality assurance program for CT-TPS process Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2002 Implementation of an integrated quality assurance program for CT-TPS process Junqing Wu Louisiana State University

More information

Assesing multileaf collimator effect on the build-up region using Monte Carlo method

Assesing multileaf collimator effect on the build-up region using Monte Carlo method Pol J Med Phys Eng. 2008;14(3):163-182. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0014-0 website: http://www.pjmpe.waw.pl M. Zarza Moreno 1, 2, N. Teixeira 3, 4, A. P. Jesus 1, 2, G. Mora 1 Assesing multileaf

More information

Data. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy

Data. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy Data ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy The ModuLeaf Mini Multileaf

More information

15 Dose Calculation Algorithms

15 Dose Calculation Algorithms Dose Calculation Algorithms 187 15 Dose Calculation Algorithms Uwe Oelfke and Christian Scholz CONTENTS 15.1 Introduction 187 15.2 Model-Based Algorithms 188 15.3 Modeling of the Primary Photon Fluence

More information

Multi-Purpose Body Phantom

Multi-Purpose Body Phantom Multi-Purpose Body Phantom Legal & Copyright Notice Information in this manual is subject to change without notice. Permission is granted to Owners of the to reproduce the Worksheets included in this manual

More information

Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code

Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code American Journal of Biomedical Engineering 216, 6(4): 124-131 DOI: 1.5923/j.ajbe.21664.3 Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code Ankit Kajaria 1,*, Neeraj Sharma

More information

Influence of electron density spatial distribution and X-ray beam quality during CT simulation on dose calculation accuracy

Influence of electron density spatial distribution and X-ray beam quality during CT simulation on dose calculation accuracy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 3, summer 2011 Influence of electron density spatial distribution and X-ray beam quality during CT simulation on dose calculation accuracy

More information

Adaptive algebraic reconstruction technique

Adaptive algebraic reconstruction technique Adaptive algebraic reconstruction technique Wenkai Lua) Department of Automation, Key State Lab of Intelligent Technology and System, Tsinghua University, Beijing 10084, People s Republic of China Fang-Fang

More information

Helical Tomotherapy Qualitative dose Delivery Verification

Helical Tomotherapy Qualitative dose Delivery Verification University of Tennessee, Knoxville Trace: Tennessee Research and Creative Exchange University of Tennessee Honors Thesis Projects University of Tennessee Honors Program 5-2003 Helical Tomotherapy Qualitative

More information

UNCOMPROMISING QUALITY

UNCOMPROMISING QUALITY ION CHAMBERS UNCOMPROMISING QUALITY Designed with over 30 years of scientific integrity for a broad range of dosimetry measurements in diverse radiation beams Farmer-type Chambers For absolute dosimetry

More information

IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques

IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques Report of the Coordinated Research Project (CRP) on Development of Procedures

More information

A novel method to correct for pitch and yaw patient setup errors in helical tomotherapy

A novel method to correct for pitch and yaw patient setup errors in helical tomotherapy A novel method to correct for pitch and yaw patient setup errors in helical tomotherapy Sarah A. Boswell a Department of Medical Physics, University of Wisconsin, Madison, Wisconsin 53706 Robert Jeraj

More information

An approach for measuring the spatial orientations of a computed-tomography simulation system

An approach for measuring the spatial orientations of a computed-tomography simulation system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 2, 2014 An approach for measuring the spatial orientations of a computed-tomography simulation system Meng Chia Wu, a Ramani Ramaseshan Department

More information

New Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC

New Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC New Technology in Radiation Oncology James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC Technology s s everywhere From the imaging chain To the planning system To the linac To QA..it..it

More information

A fast, independent dose check of HDR plans

A fast, independent dose check of HDR plans JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 2, SPRING 2003 A fast, independent dose check of HDR plans Martin E. Lachaine* Department of Radiation Oncology, The University of Arizona,

More information

Release Notes for Dosimetry Check with Convolution-Superposition Collapsed Cone Algorithm (CC)

Release Notes for Dosimetry Check with Convolution-Superposition Collapsed Cone Algorithm (CC) Dosimetry Check for convolution/superposition collapsed cone (CC), Page 1 of 11 Release Notes for Dosimetry Check with Convolution-Superposition Collapsed Cone Algorithm (CC) File: CCReleaseNotes.doc Date:

More information

Accuracy of treatment planning calculations for conformal radiotherapy van 't Veld, Aart Adeodatus

Accuracy of treatment planning calculations for conformal radiotherapy van 't Veld, Aart Adeodatus University of Groningen Accuracy of treatment planning calculations for conformal radiotherapy van 't Veld, Aart Adeodatus IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's

More information

Source Model Tuning for a 6 MV Photon Beam used in Radiotherapy

Source Model Tuning for a 6 MV Photon Beam used in Radiotherapy Journal of Physics: Conference Series Source Model Tuning for a 6 MV Photon Beam used in Radiotherapy To cite this article: Lukas A Hirschi et al 2007 J. Phys.: Conf. Ser. 74 021008 View the article online

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!

More information

THE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION

THE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION 2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 THE SIMULATION OF

More information

Use of a commercial ion chamber detector array for the measurement of high spatial resolution photon beam profiles

Use of a commercial ion chamber detector array for the measurement of high spatial resolution photon beam profiles Received: 17 January 2018 Revised: 27 August 2018 Accepted: 31 August 2018 DOI: 10.1002/acm2.12466 TECHNICAL NOTE Use of a commercial ion chamber detector array for the measurement of high spatial resolution

More information

Chapter 9 Field Shaping: Scanning Beam

Chapter 9 Field Shaping: Scanning Beam Chapter 9 Field Shaping: Scanning Beam X. Ronald Zhu, Ph.D. Department of Radiation Physics M. D. Anderson Cancer Center Houston, TX June 14-18, 2015 AAPM - Summer School 2015, Colorado Spring Acknowledgement

More information

GPU-based fast gamma index calcuation

GPU-based fast gamma index calcuation 1 GPU-based fast gamma index calcuation 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 Xuejun Gu, Xun Jia, and Steve B. Jiang Center for Advanced Radiotherapy Technologies

More information

Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations

Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations A. Al-Basheer, G. Sjoden, M. Ghita Computational Medical Physics Team Nuclear & Radiological Engineering University

More information

Machine and Physics Data Guide

Machine and Physics Data Guide WWW..COM Machine and Physics Data Guide STANDARD IMAGING, INC. 3120 Deming Way Middleton, WI 53562-1461 May / 2008 2008 Standard Imaging, Inc. TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 www.standardimaging.com

More information

Room scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation study

Room scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation study Received: 9 November 2016 Revised: 9 November 2016 Accepted: 28 November 2016 DOI: 10.1002/acm2.12039 RADIATION ONCOLOGY PHYSICS Room scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation

More information

Mutual information based CT registration of the lung at exhale and inhale breathing states using thin-plate splines

Mutual information based CT registration of the lung at exhale and inhale breathing states using thin-plate splines Mutual information based CT registration of the lung at exhale and inhale breathing states using thin-plate splines Martha M. Coselmon, a) James M. Balter, Daniel L. McShan, and Marc L. Kessler Department

More information

Acknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF

Acknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF Page 1 Quality Assurance of IMRT Delivery Systems - Siemens Lynn J. Verhey, Ph.D. Professor and Vice-Chair UCSF Dept. of Radiation Oncology AAPM 22 Annual Meeting, Montreal Acknowledgments Ping Xia, Ph.D.,

More information

8/3/2017. Contour Assessment for Quality Assurance and Data Mining. Objective. Outline. Tom Purdie, PhD, MCCPM

8/3/2017. Contour Assessment for Quality Assurance and Data Mining. Objective. Outline. Tom Purdie, PhD, MCCPM Contour Assessment for Quality Assurance and Data Mining Tom Purdie, PhD, MCCPM Objective Understand the state-of-the-art in contour assessment for quality assurance including data mining-based techniques

More information

On compensator design for photon beam intensity-modulated conformal therapy

On compensator design for photon beam intensity-modulated conformal therapy On compensator design for photon beam intensity-modulated conformal therapy Steve B. Jiang a) and Komanduri M. Ayyangar Department of Radiation Therapy, Medical College of Ohio, 3000 Arlington Avenue,

More information

An experimental investigation on the effect of beam angle optimization on the reduction of beam numbers in IMRT of head and neck tumors

An experimental investigation on the effect of beam angle optimization on the reduction of beam numbers in IMRT of head and neck tumors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 An experimental investigation on the effect of beam angle optimization on the reduction of beam numbers in IMRT of head and neck tumors

More information

The team. Disclosures. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge.

The team. Disclosures. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge Dimitre Hristov Radiation Oncology Stanford University The team Renhui Gong 1 Magdalena Bazalova-Carter 1

More information

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER! THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER! Confidence in complex treatments Modern radiation therapy uses complex plans with techniques such as IMRT, VMAT and Tomotherapy.

More information

Verification measurements of an emc algorithm using a 2D ion chamber array

Verification measurements of an emc algorithm using a 2D ion chamber array JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Verification measurements of an emc algorithm using a 2D ion chamber array Mark D. Wanklyn, 1a Ghirmay Kidane, 2 and Liz Crees 2 Medical

More information

Investigation of photon beam output factors for conformal radiation therapy Monte Carlo simulations and measurements

Investigation of photon beam output factors for conformal radiation therapy Monte Carlo simulations and measurements INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N133 N143 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)32395-9 NOTE Investigation of photon beam output factors for conformal radiation therapy

More information

Quick Reference Datasheet For All RIT113 Packages

Quick Reference Datasheet For All RIT113 Packages Quick Reference Datasheet For All RIT113 Packages For Rotational Therapies, IMRT & TG142 Highlights and selected product information only. A complete TG142 brochure is available. For more information on

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 8, NUMBER 1, WINTER 2007

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 8, NUMBER 1, WINTER 2007 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 8, NUMBER 1, WINTER 2007 Dosimetric characteristics of a cubic-block-piled compensator for intensity-modulated radiation therapy in the Pinnacle radiotherapy

More information

Implementation of the EGSnrc / BEAMnrc Monte Carlo code - Application to medical accelerator SATURNE43

Implementation of the EGSnrc / BEAMnrc Monte Carlo code - Application to medical accelerator SATURNE43 International Journal of Innovation and Applied Studies ISSN 2028-9324 Vol. 6 No. 3 July 2014, pp. 635-641 2014 Innovative Space of Scientific Research Journals http://www.ijias.issr-journals.org/ Implementation

More information

Implementation of a double Gaussian source model for the BEAMnrc Monte Carlo code and its influence on small fields dose distributions

Implementation of a double Gaussian source model for the BEAMnrc Monte Carlo code and its influence on small fields dose distributions JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Implementation of a double Gaussian source model for the BEAMnrc Monte Carlo code and its influence on small fields dose distributions

More information

Dose Calculation and Optimization Algorithms: A Clinical Perspective

Dose Calculation and Optimization Algorithms: A Clinical Perspective Dose Calculation and Optimization Algorithms: A Clinical Perspective Daryl P. Nazareth, PhD Roswell Park Cancer Institute, Buffalo, NY T. Rock Mackie, PhD University of Wisconsin-Madison David Shepard,

More information

ADVANCING CANCER TREATMENT

ADVANCING CANCER TREATMENT 3 ADVANCING CANCER TREATMENT SUPPORTING CLINICS WORLDWIDE RaySearch is advancing cancer treatment through pioneering software. We believe software has un limited potential, and that it is now the driving

More information

Development a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code

Development a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code Iran. J. Radiat. Res., 2006; 4 (1): 7-14 Development a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code. Mesbahi * Department of Medical Physics, Medical School,

More information

CARS 2008 Computer Assisted Radiology and Surgery

CARS 2008 Computer Assisted Radiology and Surgery Online External Beam Radiation Planning and Training Felix Hamza-Lup a, Ivan Sopin a, Omar Zeidan b a Computer Science, Armstrong Atlantic State University, Savannah, Georgia, USA b MD Anderson Cancer

More information

Measurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code

Measurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code reports of practical oncology and radiotherapy 1 5 (2 0 1 0) 64 68 available at www.sciencedirect.com journal homepage: http://www.rpor.eu/ Original article Measurement of depth-dose of linear accelerator

More information

UNIVERSITY OF SOUTHAMPTON

UNIVERSITY OF SOUTHAMPTON UNIVERSITY OF SOUTHAMPTON PHYS2007W1 SEMESTER 2 EXAMINATION 2014-2015 MEDICAL PHYSICS Duration: 120 MINS (2 hours) This paper contains 10 questions. Answer all questions in Section A and only two questions

More information

Image-based Monte Carlo calculations for dosimetry

Image-based Monte Carlo calculations for dosimetry Image-based Monte Carlo calculations for dosimetry Irène Buvat Imagerie et Modélisation en Neurobiologie et Cancérologie UMR 8165 CNRS Universités Paris 7 et Paris 11 Orsay, France buvat@imnc.in2p3.fr

More information

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization

More information

Improvements in Monte Carlo simulation of large electron fields

Improvements in Monte Carlo simulation of large electron fields SLAC-PUB-12908 October 2007 Improvements in Monte Carlo simulation of large electron fields Bruce A Faddegon 1, Joseph Perl 2 and Makoto Asai 2 1 University of California San Francisco Comprehensive Cancer

More information

Automatic measurement of air gap for proton therapy using orthogonal x ray imaging with radiopaque wires

Automatic measurement of air gap for proton therapy using orthogonal x ray imaging with radiopaque wires Received: 18 February 2018 Revised: 13 June 2018 Accepted: 5 November 2018 DOI: 10.1002/acm2.12509 TECHNICAL NOTE Automatic measurement of air gap for proton therapy using orthogonal x ray imaging with

More information

Raising the Bar in IMRT QA

Raising the Bar in IMRT QA MapCHECK 2TM Raising the Bar in IMRT QA The leader in quick and precise measurement of modulated radiotherapy beams Benefits Proven solution for film-less rotational delivery and IMRT QA - More than 1500

More information

2D DOSE MEASUREMENT USING A FLAT PANEL EPID

2D DOSE MEASUREMENT USING A FLAT PANEL EPID 2D DOSE MEASUREMENT USING A FLAT PANEL EPID by Seng Boh Lim B.A.Sc. (Hons.), University of British Columbia, 1994 M.A.Sc., University of British Colombia, 1996 M.Eng, University of British Columbia, 2002

More information

Verification of dynamic and segmental IMRT delivery by dynamic log file analysis

Verification of dynamic and segmental IMRT delivery by dynamic log file analysis JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 Verification of dynamic and segmental IMRT delivery by dynamic log file analysis Dale W. Litzenberg,* Jean M. Moran, and Benedick

More information