Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J.
|
|
- Kerrie Crawford
- 6 years ago
- Views:
Transcription
1 Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging Stephen J. Glick* J. Michael O Connor**, Clay Didier**, Mini Das*, * University of Massachusetts Medical School ** University of Massachusetts, Lowell Background CT Breast Imaging (CTBI) and Breast Tomosynthesis (BT) imaging systems are currently being developed and studied by a number of researchers and commercial vendors. Preliminary evidence suggests that these tomographic breast imaging systems have potential for improving visualization of breast masses at approximately equivalent dose to mammography. Some BT and CTBI system design and acquisition parameters CsI thickness Pixel size Electronic noise levels kvp setting X-ray tube filter Reconstructed voxel size Reconstruction filter, # of iterations Magnification (focal spot blurring) Objective assessment of image quality Barrett and Myers (Foundations of Image Science), any meaningful approach to optimizing an imaging system must include definitions of 1. the specific task to be performed, 2. the observer, 3. an object model representing the objects to be imaged, and 4. a figure-of-merit used to evaluate task performance 1
2 3D Digital Breast Models 3D power-law noise, (Burgess et al, 2001, Reiser et al 2008) Clustered lumpy background (Rowland et al, 1992). Bakic 3D breast phantom (Bakic et al, 2002, Zhang et al, 2008) - Geometrical shapes Bliznakova 3D breast phantom (Bliznakova et al, 2003, 2010) - Geometrical shapes, power law noise etc. Li et al, 3D breast phantom (Li et al, 2009) - Based on clinical breast CT images Proposed UMASS model based on surgical mastectomy Power-law specimens noiseclustered Lumpy Background UMMS prototype bench-top CTBI system Varian Rad94 X-ray tube Bow-tie filter Rotary Stage Varian 2520 Flat- Panel Detector UMMS Benchtop CT Imaging System Pendant Breast Holder Mastectomy Imaging Varian Rad94 X-ray tube Varian 2520 Flat- Panel Detector Uncompressed Breast Holder Compressed Breast Holder Bow-tie filter Rotary Stage 2
3 #19 - Invasive Ductal Carcinoma Goal of Study Use CT images of mastectomy specimens to develop an ensemble of 3D breast phantoms that can be used as input to computer simulation software to generate realistic simulated CT slices Simulation Methodology Reconstruction 3D Breast Model Phantom Generation Processing Steps Simulation Reconstruction Generate Model CT Reconstruction CTBI or BT Simulation 1. Projection Averaging 2. Cupping artifact compensation 3. Anisotropic diffusion filtering 4. Segmentation - classification of voxels a. Binary method b. Fuzzy mixture method Projections 3
4 Projection Averaging Acquire 10 projections per angle Cupping Correction by Altunbas Method* RATP (Radial Adipose Tissue Profile) Sample Radials to determine RATP * Med. Phys., (7) Weighted Fit to RATP Clinical Dose = 2.3 mgy High Dose = 23.0 mgy RATP fit treated as additive noise. Assuming circular symmetry, the correction is applied to entire slice in order to mitigate intraslice variation Preprocessing Step Anisotropic Diffusion Filtering* I (p( (G I) ) I) t Example Histograms Clinical Dose = 2.3 mgy High Dose = 23.0 mgy Reconstruction After ADF Filtering *Perona and Malik et al, 1990 High Dose with ADF Filter 4
5 Method 1 : Classify each Voxel as Adipose or Fibroglandular Tissue (Binary Method) Histogram after ADF Filtering Method 1 : Classify each Voxel as Adipose or Fibroglandular Tissue (Binary Method) Adipose Tissue Fibroglandular Tissue reconstruction after ADF filtering Breast object model using Binary Method Method 2 : Voxels can be Weighted Mixture of Adipose and Fibroglandular Tissue (Fuzzy Mixture Method) Method 2 : Voxels can be Weighted Mixture of Adipose and Fibroglandular Tissue (Fuzzy Mixture Method) Adipose Tissue Histogram after TV Filtering Weighted Mixture Fibroglandular Tissue reconstruction after TV filtering Breast object model using Fuzzy Mixture Method 5
6 Using Breast Object Models to Simulate Cone-beam Projections UMMS Cone Beam Simulation Software (CBSS) - (Vedula et. al. SPIE 5030, 2003, Gong et al, Med Phys, 2006) TASMIP (tungsten anode spectral model, Boone et al, Med. Phys. 24, 1997) or other spectral models. Spectra scaled to provide specified mean glandular dose [DgN] (Thacker et. al. Phys. Med. Biol. 49, 2004) to pendant breast X-ray transport through breast modeled using Siddon s ray-tracing, and attenuation coefficients defined by Johns and Yaffe PMB 1987 Scatter component from Monte Carlo simulation Signal and noise propagation through CsI indirect detector using parameters from a serial cascade model Reconstruct projections using either Feldkamp FBP or a penalized maximum likelihood iterative reconstruction algorithm Reconstruction Validation Reconstruction -Binary Reconstruction - Fuzzy Reconstruction: Feldkamp FBP, no roll-off of ramp filter, voxel.254mm 3 Simulation parameters approximate measured technique: 40kVp, 0.5mAs, 300 projections, no external filter, approximate Mean Glandular Dose of 2.3mGy Validation - Quantitative Validation - Quantitative Reconstruction Reconstruction -Binary Reconstruction - Fuzzy Reconstruction Reconstruction -Binary Reconstruction - Fuzzy FG % Adipose % FG % Adipose % FG % Adipose % FG % Adipose % FG % Adipose % FG % Adipose % Contrast 24.8% Contrast 26.2% Contrast 24.2% Contrast 28.4% Contrast 25.0% Contrast 23.5% 6
7 Sagital Coronal Breast CT Simulation SPECIMEN PHANTOM SIMULATION Research Power Law Analysis Validation of β for CT slices # of Specs β σ R 2 Range Freq end Freq end Avg. Burgess β CT = β Mammo - 1 P (f ) Metheany O Connor a f Freq start =.1 Generation of Compressed Breast Phantom (for DBT simulations) Reconstruction of compressed mastectomy specimen Another approach for generating compressed phantoms - use template Compressed breast phantom (binary method) Template based on CIRS Stereotactic Needle Biopsy Training Phantom 7
8 Mammograms from Compressed Breast Phantoms Example Usage of Breast Phantoms 98 tomosynthesis reconstructions, half with calc cluster present Two reconstruction methods compared Four observers (physicist) selected location and confidence of presence Average LROC Area Comparisons PML FBP 1 mgy aquisition PML FBP mgy 1.0 mgy 0.7 mgy p= p=0.014 p=
9 Summary Two methods for generating breast objects based on high-dose CT imaging of mastectomy specimens An ensemble of uncompressed and compressed breast phantoms and simulation software can be used to explore effect of acquisition and design parameters on image quality Acknowledgements This work is supported by NIH/NIBIB EB02133 ( Feasibility of CT Mammography Using Flat-Panel Detectors ) and NIH/NCI CA ("Iterative Reconstruction for Breast Tomosynthesis ). The contents are solely the responsibility of the authors and do not necessarily represent the official views of the National Institutes of Health. Patients (who consent during difficult period of their life). Clinical Team of Comprehensive Breast Clinic at Levine Cancer Center, UMass Memorial Health Care (UMMHC). Dr. Robert Quinlan, Medical Director. Pathology Department and Cytology Lab, UMass Medical School (UMMS). Dr. Ashraf Khan. UMMS Tomographic Breast Imaging Lab (TBIL) Team 3D Digital Breast Models 3D power-law noise, (Burgess et al, 2001, Metheany et al, 2007) Clustered lumpy background (Rowland et al, 1992). Bakic breast phantom (Bakic et al, 2002, Zhang et al, 2008) mammogram Proposed UMASS model based on CTBI specimens Power-law Bakic noise Clustered breast phantom Lumpy Background 9
Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images
Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),
More informationAcknowledgments and financial disclosure
AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories
More informationDigital Scatter Removal in Mammography to enable Patient Dose Reduction
Digital Scatter Removal in Mammography to enable Patient Dose Reduction Mary Cocker Radiation Physics and Protection Oxford University Hospitals NHS Trust Chris Tromans, Mike Brady University of Oxford
More informationGeneralized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction
Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction Klaus Erhard a, Michael Grass a, Sebastian Hitziger b, Armin Iske b and Tim Nielsen a a Philips Research Europe Hamburg,
More informationAssessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT
Assessment of 3D performance metrics D and 3D Metrics of Performance Towards Quality Index: Volumetric imaging systems X-ray based Volumetric imaging systems: CBCT/CT Tomosynthesis Samuel Richard and Ehsan
More informationPower Spectrum Analysis of an Anthropomorphic Breast Phantom Compared to Patient Data in 2D Digital Mammography and Breast Tomosynthesis
Power Spectrum Analysis of an Anthropomorphic Breast Phantom Compared to Patient Data in 2D Digital Mammography and Breast Tomosynthesis Lesley Cockmartin 1,*, Predrag R. Bakic 2, Hilde Bosmans 1, Andrew
More informationDesign and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon
Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization
More informationNoise power spectrum and modulation transfer function analysis of breast tomosynthesis imaging
Noise power spectrum and modulation transfer function analysis of breast tomosynthesis imaging Weihua Zhou a, Linlin Cong b, Xin Qian c, Yueh Z. Lee d, Jianping Lu c,e, Otto Zhou c,e, *Ying Chen a,b a
More informationPartial Volume Simulation in Software Breast Phantoms
Partial Volume Simulation in Software Breast Phantoms Feiyu Chen, David Pokrajac, Xiquan Shi, Fengshan Liu, Andrew D.A. Maidment a, Predrag R. Bakic a Delaware State University, 1200 N DuPont Hwy, Dover
More informationCT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION
CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology
More informationBackground. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /
Radiographic Tomosynthesis: Image Quality and Artifacts Reduction Baojun Li, Ph.D Department of Radiology Boston University Medical Center 2012 AAPM Annual Meeting Background Linear Trajectory Tomosynthesis
More informationDetection of microcalcification clusters by 2D-mammography and narrow and wide angle digital breast tomosynthesis
Detection of microcalcification clusters by 2D-mammography and narrow and wide angle digital breast tomosynthesis Andria Hadjipanteli a, Premkumar Elangovan b, Padraig T Looney a, Alistair Mackenzie a,
More information4/19/2016. Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT
Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT 1 Three manufacturers approved for Tomo Hologic and GE, and Siemens Why 2D Digital Mammography 2D FFDM it appears to
More informationACMP 25th Annual Meeting
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation ACMP 25th Annual Meeting Seattle, WA May 3, 2008 Melissa C. Martin, M.S., FACR Therapy Physics, Inc. 879 West 190th St., Ste 419,
More information3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH
3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk
More informationRefraction Corrected Transmission Ultrasound Computed Tomography for Application in Breast Imaging
Refraction Corrected Transmission Ultrasound Computed Tomography for Application in Breast Imaging Joint Research With Trond Varslot Marcel Jackowski Shengying Li and Klaus Mueller Ultrasound Detection
More informationAverage glandular dose coefficients for pendant-geometry breast CT using realistic breast phantoms
Average glandular dose coefficients for pendant-geometry breast CT using realistic breast phantoms Andrew M. Hernandez Department of Radiology, Biomedical Engineering Graduate Group, University of California
More informationEmpirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography
Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Marc Kachelrieß, a Katia Sourbelle, and Willi A. Kalender Institute of Medical Physics, University of
More informationUnique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc.
Unique Features of the GE Senoclaire Tomosynthesis System Tyler Fisher, M.S., DABR Therapy Physics, Inc. Conflict of Interest Disclosure I have no conflicts to disclose. Learning Objectives Overview of
More informationA comparative study of limited-angle cone-beam reconstruction methods for breast tomosynthesis
A comparative study of limited-angle cone-beam reconstruction methods for breast tomosynthesis Yiheng Zhang, a Heang-Ping Chan, Berkman Sahiner, Jun Wei, Mitchell M. Goodsitt, Lubomir M. Hadjiiski, Jun
More informationInvestigating Oblique Reconstructions with Super-Resolution in Digital Breast Tomosynthesis
Investigating Oblique Reconstructions with Super-Resolution in Digital Breast Tomosynthesis Raymond J. Acciavatti, Stewart B. Mein, and Andrew D.A. Maidment University of Pennsylvania, Department of Radiology,
More informationABSTRACT. Keywords: Digital mammography, ultrasound tomography, anthropomorphic breast phantom, breast density analysis, validation. 1.
Comparison of 3D and 2D Breast Density Estimation from Synthetic Ultrasound Tomography Images and Digital Mammograms of Anthropomorphic Software Breast Phantoms Predrag R. Bakic, Cuiping Li*, Erik West*,
More informationDesign Considerations in Optimizing a Breast Tomosynthesis System
Design Considerations in Optimizing a Breast Tomosynthesis System Andrew Smith, Ph.D., Vice President - Imaging Science, Hologic Introduction Breast tomosynthesis, also referred to as three-dimensional
More informationDigital breast tomosynthesis: comparison of different methods to calculate patient doses
Digital breast tomosynthesis: comparison of different methods to calculate patient doses Poster No.: C-2220 Congress: ECR 2011 Type: Scientific Paper Authors: A. Jacobs 1, L. Cockmartin 1, D. R. Dance
More informationTask-based Assessment of X-ray Breast Imaging Systems Using Insilico
55 th : Virtual Tools for Validation of X-ray Breast Imaging Systems Task-based Assessment of X-ray Breast Imaging Systems Using Insilico Modeling Tools Rongping Zeng Division of Imaging and Applied Mathematics
More informationOptimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago
Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic
More informationScaling Calibration in the ATRACT Algorithm
Scaling Calibration in the ATRACT Algorithm Yan Xia 1, Andreas Maier 1, Frank Dennerlein 2, Hannes G. Hofmann 1, Joachim Hornegger 1,3 1 Pattern Recognition Lab (LME), Friedrich-Alexander-University Erlangen-Nuremberg,
More informationEstimation of adipose compartment volumes in CT images of a mastectomy specimen
Estimation of adipose compartment volumes in CT images of a mastectomy specimen Abdullah-Al-Zubaer Imran a, David D. Pokrajac a, Andrew D.A. Maidment b, Predrag R. Bakic b a Department of Computer & Information
More information8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit
Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant
More information2008 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 5, OCTOBER G. Mettivier, N. Lanconelli, S. Lo Meo, and P. Russo, Member, IEEE
2008 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 5, OCTOBER 2012 Scatter Correction in Cone-Beam Breast Computed Tomography: Simulations and Experiments G. Mettivier, N. Lanconelli, S. Lo Meo, and
More informationCertificate Extension II 8/1/2016. Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role
Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role Kish Chakrabarti, Ph.D., FAAPM Division of Mammography Quality Standards Center for Devices and Radiological
More information8/2/2016. Acknowledgement. Common Clinical Questions. Presumption Images are Good Enough to accurately answer clinical questions
Image Quality Assessment using Model Observers: Clinical Implementation and Practical Considerations Shuai Leng, PhD Associate Professor, Department of Radiology Mayo Clinic, Rochester MN Acknowledgement
More information7/13/2015 EVALUATION OF NONLINEAR RECONSTRUCTION METHODS. Outline. This is a decades-old challenge
EVALUATION OF NONLINEAR RECONSTRUCTION METHODS Kyle J. Myers, Ph.D. Director, Division of Imaging, Diagnostics, and Software Reliability Office of Science and Engineering Laboratories, CDRH, FDA 2 Outline
More informationPhoton counting spectral CT versus conventional CT: comparative evaluation for breast imaging application
Physics in Medicine & Biology Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application To cite this article: Polad M Shikhaliev and Shannon G Fritz 2011
More informationABSTRACT 1. INTRODUCTION
Estimating breast thickness for dual-energy subtraction in contrastenhanced digital mammography using calibration phantoms Kristen C. Lau 1, Young Joon Kwon, Moez Karim Aziz, Raymond J. Acciavatti, Andrew
More informationPenalized-Likelihood Reconstruction for Sparse Data Acquisitions with Unregistered Prior Images and Compressed Sensing Penalties
Penalized-Likelihood Reconstruction for Sparse Data Acquisitions with Unregistered Prior Images and Compressed Sensing Penalties J. W. Stayman* a, W. Zbijewski a, Y. Otake b, A. Uneri b, S. Schafer a,
More information10/9/2018. Deborah Thames BSRS RT (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT
Deborah Thames BSRS RT (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT 1 Mammography Five FFDM approved for Tomo Hologic, Ge Senoclair and GE Pristina, Siemens, and Fujifilm 2 Why 2D Digital
More informationA Method for Producing Simulated Mammograms: Observer Study
A Method for Producing Simulated Mammograms: Observer Study Payam Seifi M.Sc. Michael R. Chinander Ph.D. Robert M. Nishikawa Ph.D., FAAPM Carl J. Vyborny Translational Laboratory for Breast Imaging Research
More informationHidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.
2-D D Dose-CT Mapping in Geant4 Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. Table of Contents Background & Purpose Materials Methods
More informationAcknowledgments. Nesterov s Method for Accelerated Penalized-Likelihood Statistical Reconstruction for C-arm Cone-Beam CT.
June 5, Nesterov s Method for Accelerated Penalized-Likelihood Statistical Reconstruction for C-arm Cone-Beam CT Adam S. Wang, J. Webster Stayman, Yoshito Otake, Gerhard Kleinszig, Sebastian Vogt, Jeffrey
More informationShadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS
Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic
More informationDeep Scatter Estimation (DSE): Accurate Real-Time Scatter Estimation for X-Ray CT using a Deep Convolutional Neural Network
Deep Scatter Estimation (DSE): Accurate Real-Time Scatter Estimation for X-Ray CT using a Deep Convolutional Neural Network Joscha Maier 1,2, Stefan Sawall 1,2, and Marc Kachelrieß 1,2 1 German Cancer
More informationFAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION
2009 IEEE Nuclear Science Symposium Conference Record M03-7 FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION Wonseok Huh, Jeffrey A. Fessler, Adam M. Alessio, and Paul E. Kinahan Department
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More informationImage quality of microcalcifications in digital breast tomosynthesis: Effects of projection-view distributions
Image quality of microcalcifications in digital breast tomosynthesis: Effects of projection-view distributions Yao Lu, a) Heang-Ping Chan, Jun Wei, Mitch Goodsitt, Paul L. Carson, and Lubomir Hadjiiski
More informationPosition accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon
Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon WHITE PAPER Introduction An image guidance system based on Cone Beam CT (CBCT) is included in Leksell
More informationSelf-Calibration of Cone-Beam CT Using 3D-2D Image Registration
Self-Calibration of Cone-Beam CT Using 3D-2D Image Registration Sarah Ouadah 1 J. Webster Stayman 1 Grace Jianan Gang 1 Ali Uneri 2 Tina Ehtiati 3 Jeffrey H. Siewerdsen 1,2 1. Department of Biomedical
More informationBreast Tissue Classification in Digital Tomosynthesis Images Based on Global Gradient Minimization and Texture Features
Breast Tissue Classification in Digital Tomosynthesis Images Based on Global Gradient Minimization and Texture Features Xulei Qin, Emory University Guolan Lu, Emory University Ioannis Sechopoulos, Emory
More informationUniversità degli Studi di Napoli Federico II
Università degli Studi di Napoli Federico II DOTTORATO DI RICERCA IN FISICA Ciclo XXIX Coordinatore: Prof. Salvatore Capozziello Dose and image quality in X-ray phase contrast breast imaging Settore Scientifico
More informationCBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona
CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona Introduction CBCT has become a routine procedure for
More informationDetailed analysis of scatter contribution from different simulated geometries of X-ray detectors
Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors Elena Marimon 1,2*, Hammadi Nait-Charif 1, Asmar Khan 2, Philip A. Marsden 3, Oliver Diaz 4 1 Centre for
More informationScatter reduction for grid-less mammography using the convolutionbased image post-processing technique
Scatter reduction for grid-less mammography using the convolutionbased image post-processing technique Elena Marimón *a,b, Hammadi Nait-Charif a, Asmar Khan b, Philip A. Marsden c, Oliver Diaz d a CDE
More informationRECENTLY APPROVED GE FFDM DBT TESTING PROCEDURES.
RECENTLY APPROVED GE FFDM DBT TESTING PROCEDURES. S. G U R U P R A S A D, P H. D., D A B R, E M E R I T U S, M AT H E W H A L L M. S. N O RT H S H O R E U N I V E R S I T Y H E A LT H S Y S T E M A N D
More informationSpatial Resolution Properties in Penalized-Likelihood Reconstruction of Blurred Tomographic Data
Spatial Resolution Properties in Penalized-Likelihood Reconstruction of Blurred Tomographic Data Wenying Wang, Grace J. Gang and J. Webster Stayman Department of Biomedical Engineering, Johns Hopkins University,
More informationScatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University
Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid by Yingxuan Chen Graduate Program in Medical Physics Duke University Date: Approved: Lei Ren, Supervisor Fang Fang Yin, Chair
More informationContrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis
Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Stefan C. Saur 1, Hatem Alkadhi 2, Luca Regazzoni 1, Simon Eugster 1, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision
More informationPhilips SPECT/CT Systems
Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET
More informationThe Near Future in Cardiac CT Image Reconstruction
SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de
More informationAcknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury
A. Sisniega et al. (presented at RSNA 214) High Performance Cone-Beam CT of Acute Traumatic Brain Injury A. Sisniega 1 W. Zbijewski 1, H. Dang 1, J. Xu 1 J. W. Stayman 1, J. Yorkston 2, N. Aygun 3 V. Koliatsos
More informationLorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc.
Lorad FFDM QC Procedures for Medical Physicist Tao Wu, Ph.D. Hologic, Inc. Lorad Selenia FFDM Tests Following Lorad QC Manual Collimation Assessment Artifact Evaluation System Resolution Phantom Image
More informationAdvanced Image Reconstruction Methods for Photoacoustic Tomography
Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic
More informationProton dose calculation algorithms and configuration data
Proton dose calculation algorithms and configuration data Barbara Schaffner PTCOG 46 Educational workshop in Wanjie, 20. May 2007 VARIAN Medical Systems Agenda Broad beam algorithms Concept of pencil beam
More informationLimitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods
Stereoscopic Breast Imaging Andrew D. A. Maidment, Ph.D. Chief, Physics Section Department of Radiology University of Pennsylvania Limitations of Projection Radiography Mammography is a projection imaging
More informationGPU implementation for rapid iterative image reconstruction algorithm
GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial
More informationMetal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland
Metal Artifact Reduction CT Techniques R S S S Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland N. 1 v o 4 1 0 2. Postoperative CT Metal Implants CT is accurate for assessment
More informationCollection of Work Projects
EUTEMPE-RX: European Training and Education for Medical Physics Experts in Radiology MPE05: Anthropomorphic phantoms Collection of Work Projects Varna, September 2015 1 2 Introduction: The present booklet
More informationRADIOMICS: potential role in the clinics and challenges
27 giugno 2018 Dipartimento di Fisica Università degli Studi di Milano RADIOMICS: potential role in the clinics and challenges Dr. Francesca Botta Medical Physicist Istituto Europeo di Oncologia (Milano)
More informationScatter Correction Methods in Dimensional CT
Scatter Correction Methods in Dimensional CT Matthias Baer 1,2, Michael Hammer 3, Michael Knaup 1, Ingomar Schmidt 3, Ralf Christoph 3, Marc Kachelrieß 2 1 Institute of Medical Physics, Friedrich-Alexander-University
More informationIncorporation of Prior Knowledge for Region of Change Imaging from Sparse Scan Data in Image-Guided Surgery
Incorporation of Prior Knowledge for Region of Change Imaging from Sparse Scan Data in Image-Guided Surgery J. Lee a, J. W. Stayman b, Y. Otake c, S. Schafer b, W. Zbijewski b, A. J. Khanna b,d, J. L.
More information7/31/ D Cone-Beam CT: Developments and Applications. Disclosure. Outline. I have received research funding from NIH and Varian Medical System.
4D Cone-Beam CT: Developments and Applications Lei Ren, PhD, DABR Department of Radiation Oncology Duke University Medical Center Disclosure I have received research funding from NIH and Varian Medical
More informationEvaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT
Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Qiao Yang 1,4, Meng Wu 2, Andreas Maier 1,3,4, Joachim Hornegger 1,3,4, Rebecca Fahrig
More informationComputer-Tomography II: Image reconstruction and applications
Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de
More informationEvaluation of Back Projection Methods for Breast Tomosynthesis Image Reconstruction
DOI 10.1007/s10278-014-9736-6 Evaluation of Back Projection Methods for Breast Tomosynthesis Image Reconstruction Weihua Zhou & Jianping Lu & Otto Zhou & Ying Chen # Society for Imaging Informatics in
More informationImproving Image Reconstruction for Digital Breast Tomosynthesis
Improving Image Reconstruction for Digital Breast Tomosynthesis by Jiabei Zheng A dissertation submitted in partial fulfillment of the requirements for the degree of Doctor of Philosophy (Electrical Engineering:
More informationArtifact Mitigation in High Energy CT via Monte Carlo Simulation
PIERS ONLINE, VOL. 7, NO. 8, 11 791 Artifact Mitigation in High Energy CT via Monte Carlo Simulation Xuemin Jin and Robert Y. Levine Spectral Sciences, Inc., USA Abstract The high energy (< 15 MeV) incident
More informationAIDR 3D Iterative Reconstruction:
Iterative Reconstruction: Integrated, Automated and Adaptive Dose Reduction Erin Angel, PhD Manager, Clinical Sciences, CT Canon Medical Systems USA Iterative Reconstruction 1 Since the introduction of
More informationGPU-based Fast Cone Beam CT Reconstruction from Undersampled and Noisy Projection Data via Total Variation
GPU-based Fast Cone Beam CT Reconstruction from Undersampled and Noisy Projection Data via Total Variation 5 10 15 20 25 30 35 Xun Jia Department of Radiation Oncology, University of California San Diego,
More informationMEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah
MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography
More informationDUAL energy X-ray radiography [1] can be used to separate
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy
More informationFinancial disclosure. Onboard imaging modality for IGRT
Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty
More informationCarbondale, IL USA; University of North Carolina Chapel Hill, NC USA; USA; ABSTRACT
Pre-computed backprojection based penalized-likelihood (PPL) reconstruction with an edge-preserved regularizer for stationary Digital Breast Tomosynthesis Shiyu Xu a, Christy Redmon Inscoe b, Jianping
More informationA Generation Methodology for Numerical Phantoms with Statistically Relevant Variability of Geometric and Physical Properties
A Generation Methodology for Numerical Phantoms with Statistically Relevant Variability of Geometric and Physical Properties Steven Dolly 1, Eric Ehler 1, Yang Lou 2, Mark Anastasio 2, Hua Li 2 (1) University
More informationValidation of a digital mammography image simulation chain with automated scoring of CDMAM images
Validation of a digital mammography image simulation chain with automated scoring of CDMAM images Mary Yip 1, Abdulaziz Alsager 2, Emma Lewis 1, Kevin Wells 1, and Kenneth C. Young 3 1 Centre for Vision,
More informationSpiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004
Spiral ASSR Std p = 1.0 Spiral EPBP Std p = 1.0 Kachelrieß et al., Med. Phys. 31(6): 1623-1641, 2004 256 slices (0/300) Advantages of Cone-Beam Spiral CT Image quality nearly independent of pitch Increase
More informationRADIOLOGY AND DIAGNOSTIC IMAGING
Day 2 part 2 RADIOLOGY AND DIAGNOSTIC IMAGING Dr hab. Zbigniew Serafin, MD, PhD serafin@cm.umk.pl 2 3 4 5 CT technique CT technique 6 CT system Kanal K: RSNA/AAPM web module: CT Systems & CT Image Quality
More informationMulti-slice CT Image Reconstruction Jiang Hsieh, Ph.D.
Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced
More informationDEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE
Rajesh et al. : Proceedings of the National Seminar & Exhibition on Non-Destructive Evaluation DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh V Acharya, Umesh Kumar, Gursharan
More informationCIVA CT, an advanced simulation platform for NDT
More Info at Open Access Database www.ndt.net/?id=18774 CIVA CT, an advanced simulation platform for NDT Marius Costin 1, David Tisseur 1, Caroline Vienne 1, Ronan Guillamet 1, Hussein Banjak 1, Navnina
More informationCT imaging using energy-sensitive photon-counting detectors. Disclosure. Vision 20/20 paper
CT imaging using energy-sensitive photon-counting detectors Katsuyuki Ken Taguchi, Ph.D. ktaguchi@jhmi.edu Division of Medical Imaging Physics Department of Radiology Johns Hopkins University No financial
More informationMIXTURE MODELING FOR DIGITAL MAMMOGRAM DISPLAY AND ANALYSIS 1
MIXTURE MODELING FOR DIGITAL MAMMOGRAM DISPLAY AND ANALYSIS 1 Stephen R. Aylward, Bradley M. Hemminger, and Etta D. Pisano Department of Radiology Medical Image Display and Analysis Group University of
More informationSpiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA
Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization
More informationDUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography
Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography Marc Kachelrieß, Member, IEEE, Timo Berkus, Philip Stenner, Willi A. Kalender Abstract Material selective imaging using dual energy
More informationCone-beam computed tomography with a flat-panel imager: Magnitude and effects of x-ray scatter
Cone-beam computed tomography with a flat-panel imager: Magnitude and effects of x-ray scatter Jeffrey H. Siewerdsen a) and David A. Jaffray Department of Radiation Oncology, William Beaumont Hospital,
More informationEffects of the difference in tube voltage of the CT scanner on. dose calculation
Effects of the difference in tube voltage of the CT scanner on dose calculation Dong Joo Rhee, Sung-woo Kim, Dong Hyeok Jeong Medical and Radiological Physics Laboratory, Dongnam Institute of Radiological
More informationMedical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.
DXRaySMCS First User Friendly Interface Developed for Prediction of Diagnostic Radiology X-Ray Spectra Produced by Monte Carlo (MCNP-4C) Simulation in Iran M.T. Bahreyni Toosi a*, H. Moradi b, H. Zare
More information8/2/2016. Hands-on GE SenoClaire DBT: technical characteristics & quality control. Disclosures and Acknowledgements
Hands-on GE SenoClaire DBT: technical characteristics & quality control Razvan Iordache, Ph.D. GE Healthcare Imagination at work Disclosures and Acknowledgements Razvan Iordache Employee of GE Healthcare
More informationFits you like no other
Fits you like no other BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase in room
More informationBasics of treatment planning II
Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!
More informationA framework of modeling detector systems for computed tomography simulations
Home Search Collections Journals About Contact us My IOPscience A framework of modeling detector systems for computed tomography simulations This content has been downloaded from IOPscience. Please scroll
More information8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition
Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare
More information