Tumor motion during liver SBRT

Size: px
Start display at page:

Download "Tumor motion during liver SBRT"

Transcription

1 Tumor motion during liver SBRT - projects at Aarhus University Hospital - Per Poulsen, Esben Worm, Walther Fledelius, Morten Høyer Aarhus University Hospital, Denmark

2 SBRT: Stereotactic Body Radiation Therapy Few fractions (1-5) with high daily doses (12-25 Gy) Steep dose gradients High precision needed Traditionally an external stereotactic coordinate system is used for precise targeting but this tends to be replaced by IGRT

3 IGRT for liver SBRT Liver tumors are very difficult to image in-room [contrast required] Tumor position surrogates: Bony anatomy, diaphragm, implanted markers Implanted markers are more accurate than anatomical land marks [if marker is less than ~8 cm from tumor ( * ) ] ( * ) Seppenwoolde PMB 2011

4 Outline Liver SBRT planning and setup Monitoring intrafraction tumor motion Tumor motion characteristics Trajectory based patient setup Dose reconstruction Simulated Dynamic MLC tracking Real-time marker segmentation

5 Liver SBRT planning and setup Marker implantation Fixation CT scanning Target definition Treatment planning Setup 2-3 gold markers. 1 mm x 3 mm Implanted percutaneously guided by ultrasound Ideally close to and surrounding the tumor Often limited implantation directions

6 Liver SBRT planning and setup Marker implantation Fixation CT scanning Target definition Treatment planning Setup Stereotactic Body Frame (Elekta) If possible: Photo from Berbeco et al. IJORBP 2007 Abdominal compression to reduce breathing motion

7 Liver SBRT planning and setup Marker implantation Fixation CT scanning Target definition Treatment planning Setup CT1: Expiration breath-hold with contrast [optimal contrast timing window ~15 seconds] Good contrast timing - and a bit later

8 Liver SBRT planning and setup Marker implantation Fixation CT scanning Target definition Treatment planning Setup CT2: 4DCT c

9 Liver SBRT planning and setup Marker implantation Fixation CT scanning Target definition Treatment planning Setup 1. Match breath-hold CT and 4DCT mid-ventilation based on markers 2. Delineate GTV in 4DCT mid-vent assisted by contrast enhanced breath-hold CT 3. CTV = GTV 4. PTV = CTV + 5mm 5mm 10mm

10 Liver SBRT planning and setup Marker implantation Fixation CT scanning Target definition Treatment planning Setup Risk organs: Healthy liver (Liver CTV) Spinal cord (with 10 mm PRV margin) Esophagus (with 10 mm PRV margin) Stomach, duodenum, bowel, kidneys, heart

11 Liver SBRT planning and setup Marker implantation Fixation CT scanning Target definition Treatment planning Setup Most often conformal plan with IM segments IMRT or VMAT may be used for complicated cases CTV covered with 95% dose, PTV with 67% dose

12 Liver SBRT planning and setup Marker implantation Fixation CT scanning Target definition Treatment planning Setup Risk adapted dose prescription for metastasis: Use highest dose among Gy, Gy, 3 15Gy and Gy that fulfills constraints:

13 Liver SBRT planning and setup Marker implantation Fixation CT scanning Target definition Treatment planning Setup 1. Laser guided positioning of patient in Stereotactic Body Frame 2. Room laser guided positioning of Stereotactic Body Frame relative to linac isocenter 3. Cone-beam CT with match to planning CT (=midventilation 4DCT phase): a. Automatic match on bony anatomy b. Manual match on markers (max 10 mm shift)

14 Outline Liver SBRT planning and setup Monitoring intrafraction tumor motion Tumor motion characteristics Trajectory based patient setup Dose reconstruction Simulated Dynamic MLC tracking Real-time marker segmentation

15 Monitoring intrafraction liver tumor motion PhD project, Esben Worm, started June 2010 Aim: Detailed characterization of tumor motion in liver SBRT Method: Monitoring liver marker motion during the entire treatment delivery NB: Esben to visit U Sydney in April-May 2012

16 Monitoring intrafraction liver tumor motion Planning: 3 x treatment: 4DCT time Record external motion (RPM) + kv/mv images during the entire treatment RPM: CBCT1 (setup) CBCT2 (control) Field 1 MV+kV Field 2 MV+kV Field n MV+kV CBCT3 time

17 Intra-treatment imaging kv, 5 Hz RPM, 25 Hz MV, Hz

18 Example 1: Intra-treatment motion MV kv

19 Example 2: Motion in MV images DRR Mid-ventilation 7.5 Hz portal imaging Cranio-caudal geometrical error

20 Example 2: Motion in MV images Fraction 1 Fraction 2 Fraction 3 Note the large day-to-day variation in marker positions

21 Pre- and post-treatment imaging RPM CBCT: 1. Acquire ~650 projection images in ~1 minute during a full gantry rotation 2. Reconstruct 3D image from projection images

22 Examples: Liver CBCT scans: Motion artifacts Example 1: Diaphragm 15 mm CBCT CT Example 2: Gold marker (3 mm) 12 mm

23 Examples: Liver CBCT projections 1 CBCT projection (out of 629) Marker motion Marker motion

24 2D 3D?

25 f 3 f 2 Probability-Based 3D Trajectory Estimation Image 1 f 1 Image 3 Target is known to be located on a sequence of ray lines

26 Assume 3D Gaussian Probability Density f 3 f 2 Image 1 Gaussian PDF f 1 Image 3 Fit 3D Gaussian PDF to observed projections

27 Estimate 3D target position from PDF f 3 f 2 Image 1 Gaussian PDF f 1 Image 3 Select most likely position along each ray line

28 Result: 3D trajectory 1 Image Image 3

29 Single-imager trajectory estimation Accurate: 3D RMS error < 1mm in 99.1% of cases for lung/abdomen and in 99.7% of cases for prostate Can be used in real-time with modest loss in accuracy as compared to retrospective ( Tracking) Poulsen et al: IJROBP 2008, PMB 2009

30 Example: Motion during CBCT acquisition External RPM signal External Internal Internal tumor motion (from CBCT projections)

31 Amending missing intra-treatment motion data Amend with single imager method when: kv images are missing [not recorded, marker not segmentable due to MV scatter] MV images are missing [not recorded, marker outside field, IMRT, VMAT] Amend with RPM signal when: Both MV and kv are momentarily missing [less than one breathing cycle at a time for MV] [only a few breathing cycles for kv]

32 Monitoring intrafraction liver tumor motion Planning: 3 x treatment: 4DCT time Synchronized 1D external abdominal motion and 3D internal tumor surrogate motion during the entire treatment RPM: CBCT1 (setup) CBCT2 (control) Field 1 MV+kV Field 2 MV+kV Field n MV+kV CBCT3 time

33 Fraction 1 Fraction 2 Fraction 3 Example: 3D trajectories for one patient

34 Outline Liver SBRT planning and setup Monitoring intrafraction tumor motion Tumor motion characteristics Trajectory based patient setup Dose reconstruction Simulated Dynamic MLC tracking Real-time marker segmentation

35 Status and perspectives of liver motion study 15 patients recruited Allows addressing a wide range of questions about: Tumor motion characteristics (magnitude, directionality, stability, predictability, correlation with 4DCT, correlation with RPM, etc, on all relevant time scales) Use of adaptive margins, gating & tracking based on limited information (e.g kv alone, RPM+kV...) Delivered tumor dose distribution (in actual treatment, in simulated gating or tracking treatment, etc.)

36 Motion (mm) CBCT AP (mm) M CB Example: Tumor trajectory during CBCT scan Time (s) RPM AP (mm) CBCT AP RPM RPM vs CBCT AP Pearsons R = Time (s) RPM AP (mm) Example: Correlation between internal motion (black) and external motion (blue). Could be used for proper gating window setting based on CBCT.

37 LR (mm) Example: Variation in breathing motion Max respiration Min respiration CC (mm) Smallest and largest respiratory cycle for a patient during beam-on AP (mm) 6

38 field5 field5 field5 Intra-treatment time field4 field4 field4 field3 field3 field3 field2 field2 field2 field1 field1 field1 CBCT CBCT CBCT Example: 3D PDF evolvement for single fraction mm AP Position (mm) CC Position (mm) LR Position (mm)

39 Example: Intrafraction mean position drift (9 pat) CBCT First field First Field Last field CBCT Last field

40 Example: Motion directionality, Pat1 Sagittal view

41 Example: Motion directionality, 9 patients Spherical view

42 Outline Liver SBRT planning and setup Monitoring intrafraction tumor motion Tumor motion characteristics Trajectory based patient setup Dose reconstruction Simulated Dynamic MLC tracking Real-time marker segmentation

43 CBCT based setup in liver SBRT Respiratory motion during CBCT scannning Motion blurring of markers Difficult to find the mean marker position for CBCT planct match Trajectory based setup would be better

44 Trajectory based setup in liver SBRT CC (mm) LR (mm) AP (mm) Segment 2D trajectory 2 0 3D trajectory Mean marker position can be found unambiguously from the trajectory Estimated motion Estimated mean CBCT marker match CBCT bony match Time (seconds) Esben Worm IJROPB 2012

45 Online trajectory based setup Custom made semiautomatic segmentation program Esben Worm IJROPB 2012

46 Online trajectory based setup Custom made semiautomatic segmentation program Esben Worm IJROPB 2012

47 Trajectory based setup in liver SBRT Status: Used on-line for 2 patients Tested off-line (post-treatment) for 5 other patients Next: Go from hand-held semi-automatic method to fully automatic method integrated into the clinical workflow

48 Outline Liver SBRT planning and setup Monitoring intrafraction tumor motion Tumor motion characteristics Trajectory based patient setup Dose reconstruction Simulated Dynamic MLC tracking Real-time marker segmentation

49 Dose reconstruction in DMLC tracking CIRS lung phantom with 1D moving rod (CC direction) Solid water tumor insert (24mm high, 25mm Ø) Sandwiched vertical Gafchromic EBT film Left lateral conformal treatment field Static or sinusoidal motion (20mm peak-to-peak, T = 4 sec) MV image-based tracking using an embedded gold marker Mai Lykkegaard Schmidt, Lone Hoffmann

50 Static Measured dose distributions Motion Tracking Static Motion Tracking

51 Dose reconstruction I: Conformal, no tracking TPS Experiment Dicom plan Target motion Motion mimicking Dicom plan Dose reconstruction in TPS Motion mimicking treatment plan Divide target motion into 1 mm position bins Create a sub-beam for each position bin Shift the sub-beam isocenter to mimick target motion Method equivalent to Berbeco, Med Phys 2008

52 Dose reconstruction II: Tracking TPS Experiment Experiment Dicom plan Target motion MLC positions Motion mimicking Dicom plan Dose reconstruction in TPS The sub-beam for each target position bin is now a step-and-shoot IMRT beam

53 Dose reconstruction characteristics Includes effects of: MLC shape changes 3D target translations Interplay effects Physical path length changes Does not include: Radiological path length changes Target deformations Target motion relative to the marker

54 Reconstructed and measured dose distributions Reconstructed Reconstructed Measured Measured

55 Reconstructed and measured dose distributions Comparison of measured and reconstructed doses: 2%/2mm gamma pass rates: % Mai Lykkegaard Schmidt Lone Hoffmann

56 Reconstructed and measured dose distributions Comparison of measured and reconstructed doses: 2%/2mm gamma pass rates: % Motion parallel to MLC GTV DVHs Motion perpendicular to MLC GTV DVHs

57 Outline Liver SBRT planning and setup Monitoring intrafraction tumor motion Tumor motion characteristics Trajectory based patient setup Dose reconstruction Simulated Dynamic MLC tracking Real-time marker segmentation

58 Simulated MV based DMLC tracking for liver SBRT Background Detailed knowledge of prototype DMLC tracking system (latencies, prediction errors, refitting of MLC) Detailed measurements of MLC response to leaf motion requests Examples of leaf adaptation dynamics measured on millisecond time scale

59 Simulated MV based DMLC tracking for liver SBRT Simulate MLC motion by modeling the effects of MV tracking latency (290 ms) Prediction for latency compensation Refitting of leaves to the estimated target position Re-adjustment of leaves to the refitted positions

60 Simulated MV based DMLC tracking for liver SBRT Assume correct real-time marker segmentation in all MV images 2D DMLC tracking Prediction starts after 8 seconds treatment Assume marker visibility in all MV images of main fields No tracking for IM segments Similar to pig tracking experiment, Poulsen IJROBP 2011

61 Results example: Patient #2/5: MV images MV. Pat02. Fx1. Field1 Beam s eye view error Green: Gold Marker Red: CTV

62 Results example: Patient #2/5: MV images MV BEV error for all fields at fraction 1 Field 1 Field 2 Field 3 Field 4 Field 5 Mean CC error: -4.1 mm

63 Results example: Patient #2/5: Tracking Simulation MV. Pat02. Fx1. Field1 MLC motion in simulated DMLC tracking treatment guided by these MV images The residual errors with tracking tend to be random Green: Gold Marker Red: CTV

64 Dose reconstruction: As Treated sagittal view BEV 2/299 MV images MU 16/299 MV images MU Original iso-c

65 Dose reconstruction: Simulated tracking sagittal view BEV 2/299 MV images MU Step&shoot IMRT With 2 segments 16/299 MV images MU Original iso-c Step&shoot IMRT With 16 segments

66 Motion including treatment plan: All subfields As treated Simulated tracking Motion speed doubled

67 Reconstructed dose distributions Dosewash: >95% dose Red = CTV Patient 2 Fraction 1

68 Reconstructed dose distribution DVH for CTV Fraction 1 Delivered CTV dose at first fraction: D 95 = 7%

69 Reconstructed dose distribution DVH for CTV Fractions 1-3 Delivered CTV dose at all three fractions: D 95 = 7%, D 95 = 5.6%, D 95 = 2.5% Tracking would have restored the planned CTV dose

70 Results: Minimum dose to 95% of the CTV Reduction in D 95 (%-point) As treated Simulated tracking Mean Maximum Mean Maximum

71 Results: Minimum dose to 95% of the CTV? Reduction in D 95 (%-point) As treated Simulated tracking Mean Maximum Mean Maximum

72 Increased D 95 with motion?? Planned (static) Treated (w/motion) Color dose wash: 97.3% ( D 95 ) Motion blurring increases V 97.3%

73 D 95 correlates with mean geometrical error p <

74 D 95 correlates with mean geometrical error p <

75 Conclusion for DMLC tracking simulations Tumor motion can deteriorate the target dose in liver SBRT even with daily image-guided localization DMLC tracking can restore the planned dose DMLC tracking made no difference for 3 out of 5 investigated treatment courses The tracking would be a safety net for the outliers and would not harm the inliers (no imaging dose).

76 DMLC tracking simulations: Possible usage Dosimetry evaluation Dosimetric gain of tracking Tracking QA: Test robustness of a given DMLC plan against a wide range of tumor motions Development of new tracking strategies MRI, multiple targets, deformations,.

77 Outline Liver SBRT planning and setup Monitoring intrafraction tumor motion Tumor motion characteristics Trajectory based patient setup Dose reconstruction Simulated Dynamic MLC tracking Real-time marker segmentation

78 Multiple marker segmentation Challenges for segmentation: Changing shape with projection angle Marker overlap, trace entanglement Poor contrast in lateral pelvic images Motion of markers relative to each other

79 Multiple marker segmentation Aim: Fully automatic No prior knowledge For both abdomen/thorax and pelvis Fledelius MedPhys 2011

80 Step 1: Identify marker candidates in each frame Single frame of pelvic CBCT scan. Four markers present

81 Step 1: Identify marker candidates in each frame Marker candidates identified in the frame (blobs)

82 Step 2: Identify trace candidates - by following marker candidates in subsequent frames

83 Step 3: Reject trace candidates - that are inconsistent with a 3D marker position

84 Step 4: Construct 3D marker constellation model - by using the most reliable remaining traces

85 Step 5: Auto-segment i all images - using projections of the 3D model as templates 3D model Anterior view Anterior image Project

86 Step 5: Auto-segment i all images - using projections of the 3D model as templates 3D model Lateral view Lateral image Project

87 Step 5: Auto-segment i all images - using projections of the 3D model as templates Segmentations failures: In 0.2 % of prostate cases In 0.1 % of liver cases Fledelius MedPhys 2011

88 Real-time auto-segmentation: Liver SBRT Pre-treatment CBCT scan: 3D marker constellation model Clouds of visited 3D positions Fledelius ESTRO 2012

89 Real-time auto-segmentation: Liver SBRT Projection image Templates Search areas Fledelius ESTRO 2012

90 Real-time auto-segmentation: Liver SBRT In CBCT projections In intra-treatment kv Fledelius ESTRO 2012

91 Summary 1. Liver SBRT planning: Risk adapted dose prescription 2. Intrafraction motion monitoring: OBI/kV/MV/RPM 3. Tumor motion characteristics: magnitude, direction, stability, correlation. 4. Trajectory based patient setup: Improves accuracy 5. Dose reconstruction: Generalized to any treatment type. 6. Simulated Dynamic MLC tracking: Dosimetric gain with tracking? 7. Multiple marker segmentation: Going real-time.

92 Thank you Esben Worm Walther Fledelius Thomas Ravkilde Morten Høyer Mai Lykkegaard Schmidt Lone Hoffmann Paul Keall & research group, Sydney Varian, Varian ilab

Image Guidance and Beam Level Imaging in Digital Linacs

Image Guidance and Beam Level Imaging in Digital Linacs Image Guidance and Beam Level Imaging in Digital Linacs Ruijiang Li, Ph.D. Department of Radiation Oncology Stanford University School of Medicine 2014 AAPM Therapy Educational Course Disclosure Research

More information

8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques

8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery Lei Ren, Ph.D. Duke University Medical Center 2016 AAPM 58 th annual meeting, Educational Course, Therapy Track Disclosure I have

More information

State-of-the-Art IGRT

State-of-the-Art IGRT in partnership with State-of-the-Art IGRT Exploring the Potential of High-Precision Dose Delivery and Real-Time Knowledge of the Target Volume Location Antje-Christin Knopf IOP Medical Physics Group Scientific

More information

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance

More information

Brilliance CT Big Bore.

Brilliance CT Big Bore. 1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.

More information

1. Learn to incorporate QA for surface imaging

1. Learn to incorporate QA for surface imaging Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago ***No disclosures*** 1. Learn to incorporate QA for surface imaging into current QA procedures for IGRT. 2. Understand

More information

Using a research real-time control interface to go beyond dynamic MLC tracking

Using a research real-time control interface to go beyond dynamic MLC tracking in partnership with Using a research real-time control interface to go beyond dynamic MLC tracking Dr. Simeon Nill Joint Department of Physics at The Institute of Cancer Research and the Royal Marsden

More information

MR-guided radiotherapy: Vision, status and research at the UMC Utrecht. Dipl. Ing. Dr. Markus Glitzner

MR-guided radiotherapy: Vision, status and research at the UMC Utrecht. Dipl. Ing. Dr. Markus Glitzner MR-guided radiotherapy: Vision, status and research at the UMC Utrecht Dipl. Ing. Dr. Markus Glitzner About myself Training Medizintechnik TU Graz PhD UMC Utrecht Clinical work Software implementation

More information

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image

More information

8/3/2016. Image Guidance Technologies. Introduction. Outline

8/3/2016. Image Guidance Technologies. Introduction. Outline 8/3/26 Session: Image Guidance Technologies and Management Strategies Image Guidance Technologies Jenghwa Chang, Ph.D.,2 Department of Radiation Medicine, Northwell Health 2 Hofstra Northwell School of

More information

Volumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT

Volumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT Volumetric Modulated Arc Therapy - Clinical Implementation Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA Acknowledgement David M. Shepard, Ph.D. Muhammad K. N. Afghan, Ph.D. Fan Chen, Ph.D.

More information

IMRT site-specific procedure: Prostate (CHHiP)

IMRT site-specific procedure: Prostate (CHHiP) IMRT site-specific procedure: Prostate (CHHiP) Scope: To provide site specific instructions for the planning of CHHIP IMRT patients Responsibilities: Radiotherapy Physicists, HPC Registered Therapy Radiographers

More information

Chapter 9 Field Shaping: Scanning Beam

Chapter 9 Field Shaping: Scanning Beam Chapter 9 Field Shaping: Scanning Beam X. Ronald Zhu, Ph.D. Department of Radiation Physics M. D. Anderson Cancer Center Houston, TX June 14-18, 2015 AAPM - Summer School 2015, Colorado Spring Acknowledgement

More information

The MSKCC Approach to IMRT. Outline

The MSKCC Approach to IMRT. Outline The MSKCC Approach to IMRT Spiridon V. Spirou, PhD Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York, NY Outline Optimization Field splitting Delivery Independent verification

More information

Experimental investigation of a general real-time 3D target localization method using sequential kv imaging combined with respiratory monitoring

Experimental investigation of a general real-time 3D target localization method using sequential kv imaging combined with respiratory monitoring IOP PUBLISHING Phys. Med. Biol. 57 (2012) 7395 7407 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/57/22/7395 Experimental investigation of a general real-time 3D target localization method using

More information

MapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays

MapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.

More information

FIRST DEMONSTRATION OF COMBINED KV/MV IMAGE-GUIDED REAL-TIME DYNAMIC MULTILEAF-COLLIMATOR TARGET TRACKING

FIRST DEMONSTRATION OF COMBINED KV/MV IMAGE-GUIDED REAL-TIME DYNAMIC MULTILEAF-COLLIMATOR TARGET TRACKING doi:10.1016/j.ijrobp.2009.02.012 Int. J. Radiation Oncology Biol. Phys., Vol. 74, No. 3, pp. 859 867, 2009 Copyright Ó 2009 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/09/$ see front

More information

MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays

MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.

More information

A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study

A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study Indrin J. Chetty, a) Mihaela Rosu, Neelam Tyagi, Lon H. Marsh, Daniel

More information

7/31/ D Cone-Beam CT: Developments and Applications. Disclosure. Outline. I have received research funding from NIH and Varian Medical System.

7/31/ D Cone-Beam CT: Developments and Applications. Disclosure. Outline. I have received research funding from NIH and Varian Medical System. 4D Cone-Beam CT: Developments and Applications Lei Ren, PhD, DABR Department of Radiation Oncology Duke University Medical Center Disclosure I have received research funding from NIH and Varian Medical

More information

An experimental comparison of conventional two-bank and novel four-bank dynamic MLC tracking

An experimental comparison of conventional two-bank and novel four-bank dynamic MLC tracking An experimental comparison of conventional two-bank and novel four-bank dynamic MLC tracking G A Davies, P Clowes, D McQuaid, P M Evans, S Webb and G Poludniowski Joint Department of Physics, Institute

More information

Real Time Tumor Motion Tracking with CyberKnife

Real Time Tumor Motion Tracking with CyberKnife Real Time Tumor Motion Tracking with CyberKnife Martina Descovich, Ph.D University of California San Francisco July 16, 2015 Learning objectives Review the principles of real-time tumor motion tracking

More information

Acknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF

Acknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF Page 1 Quality Assurance of IMRT Delivery Systems - Siemens Lynn J. Verhey, Ph.D. Professor and Vice-Chair UCSF Dept. of Radiation Oncology AAPM 22 Annual Meeting, Montreal Acknowledgments Ping Xia, Ph.D.,

More information

Estimating 3D Respiratory Motion from Orbiting Views

Estimating 3D Respiratory Motion from Orbiting Views Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct. 2005 Funding provided by NIH Grant P01 CA59827 Motivation Free-breathing

More information

Is deformable image registration a solved problem?

Is deformable image registration a solved problem? Is deformable image registration a solved problem? Marcel van Herk On behalf of the imaging group of the RT department of NKI/AVL Amsterdam, the Netherlands DIR 1 Image registration Find translation.deformation

More information

Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT

Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT Marcus Brehm 1,2, Thorsten Heußer 1, Pascal Paysan 3, Markus Oehlhafen 3, and Marc Kachelrieß 1,2 1 German Cancer Research Center

More information

Digital Tomosynthesis for Target Localization

Digital Tomosynthesis for Target Localization Digital Tomosynthesis for Target Localization Fang-Fang Yin, Devon Godfrey, Lei Ren Jacqueline Maurer, Jackie Q-L Wu Duke University Medical Center Acknowledgements Duke Radiation Oncology faculty and

More information

Creating a Knowledge Based Model using RapidPlan TM : The Henry Ford Experience

Creating a Knowledge Based Model using RapidPlan TM : The Henry Ford Experience DVH Estimates Creating a Knowledge Based Model using RapidPlan TM : The Henry Ford Experience Karen Chin Snyder, MS, DABR AAMD Region V Meeting October 4, 2014 Disclosures The Department of Radiation Oncology

More information

URGENT IMPORTANT FIELD SAFETY NOTIFICATION

URGENT IMPORTANT FIELD SAFETY NOTIFICATION Subject: Incorrect Movement of the Treatment Table Product: MOSAIQ Scope: Sites affected will be those: 1. Running MOSAIQ and, 2. Treating on linear accelerators with the RATM license Notification Released:

More information

Anthropomorphic Verification of the Eclipse AAA Algorithm for Spine SBRT Treatments Involving Titanium Implants

Anthropomorphic Verification of the Eclipse AAA Algorithm for Spine SBRT Treatments Involving Titanium Implants Anthropomorphic Verification of the Eclipse AAA Algorithm for Spine SBRT Treatments Involving Titanium Implants Mike Grams Clinical Medical Physicist Mayo Clinic, Rochester, MN NCCAAPM Meeting, April 17,

More information

doi: /j.ijrobp

doi: /j.ijrobp doi:10.1016/j.ijrobp.2010.02.052 Int. J. Radiation Oncology Biol. Phys., Vol. 79, No. 1, pp. 269 278, 2011 Copyright Ó 2011 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/$ see front matter

More information

3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending

3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending 3DVH : On The Accuracy Of The Planned Dose Perturbation Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools *Patent Pending introduction State-of-the-art IMRT QA of static gantry

More information

Moving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region

Moving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region Moving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region Andreas Hahn 1,2, Sebastian Sauppe 1,2, Michael Knaup 1, and Marc Kachelrieß 1,2 1 German Cancer Research Center (DKFZ),

More information

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS Tomotherapy Physics Machine Twinning and Quality Assurance Emilie Soisson, MS Tomotherapy at UW- Madison Treating for nearly 5 years Up to ~45 patients a day on 2 tomo units Units twinned to facilitate

More information

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE QA SOFTWARE FAST, PRECISE Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, VMAT, SRS and brachytherapy plans no film, no phantoms, no linac

More information

Hugues Mailleux Medical Physics Department Institut Paoli-Calmettes Marseille France. Sunday 17 July 2016

Hugues Mailleux Medical Physics Department Institut Paoli-Calmettes Marseille France. Sunday 17 July 2016 Hugues Mailleux Medical Physics Department Institut Paoli-Calmettes Marseille France Sunday 17 July 2016 AGENDA 1. Introduction 2. Material 3. Optimization process 4. Results 5. Comments 6. Conclusion

More information

Raising the Bar in IMRT QA

Raising the Bar in IMRT QA MapCHECK 2TM Raising the Bar in IMRT QA The leader in quick and precise measurement of modulated radiotherapy beams Benefits Proven solution for film-less rotational delivery and IMRT QA - More than 1500

More information

radiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li NCAAPM Spring Meeting 2010 Madison, WI

radiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li NCAAPM Spring Meeting 2010 Madison, WI GPU-Accelerated autosegmentation for adaptive radiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li agodley@mcw.edu NCAAPM Spring Meeting 2010 Madison, WI Overview Motivation Adaptive

More information

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon WHITE PAPER Introduction An image guidance system based on Cone Beam CT (CBCT) is included in Leksell

More information

IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays

IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays Sotiri Stathakis Outline Why IMRT/VMAT QA AAPM TG218 UPDATE Tolerance Limits and Methodologies for IMRT Verification QA Common sources

More information

Laser Systems for Patient Alignment

Laser Systems for Patient Alignment Laser Systems for Patient Alignment GALAXY Patient TOPOGRAPHY Laser System Page: 2 Date: 28 March 2009 The GALAXY System NEW METHOD FOR HIGHER ALIGNMENT ACCURACY. Patient Positioning System for Linac Simulator

More information

Initial Clinical Experience with 3D Surface Image Guidance

Initial Clinical Experience with 3D Surface Image Guidance Initial Clinical Experience with 3D Surface Image Guidance Amanda Havnen-Smith, Ph.D. Minneapolis Radiation Oncology Ridges Radiation Therapy Center Burnsville, MN April 20 th, 2012 Non-funded research

More information

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER! THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER! Confidence in complex treatments Modern radiation therapy uses complex plans with techniques such as IMRT, VMAT and Tomotherapy.

More information

Monaco VMAT. The Next Generation in IMRT/VMAT Planning. Paulo Mathias Customer Support TPS Application

Monaco VMAT. The Next Generation in IMRT/VMAT Planning. Paulo Mathias Customer Support TPS Application Monaco VMAT The Next Generation in IMRT/VMAT Planning Paulo Mathias Customer Support TPS Application 11.05.2011 Background What is Monaco? Advanced IMRT/VMAT treatment planning system from Elekta Software

More information

FAST, precise. qa software

FAST, precise. qa software qa software FAST, precise Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, rotational or brachytherapy plans no film, no phantoms, no linac

More information

ROBUST OPTIMIZATION THE END OF PTV AND THE BEGINNING OF SMART DOSE CLOUD. Moe Siddiqui, April 08, 2017

ROBUST OPTIMIZATION THE END OF PTV AND THE BEGINNING OF SMART DOSE CLOUD. Moe Siddiqui, April 08, 2017 ROBUST OPTIMIZATION THE END OF PTV AND THE BEGINNING OF SMART DOSE CLOUD Moe Siddiqui, April 08, 2017 Agenda Background IRCU 50 - Disclaimer - Uncertainties Robust optimization Use Cases Lung Robust 4D

More information

Choosing and Commissioning a Video Based Motion Management System

Choosing and Commissioning a Video Based Motion Management System Choosing and Commissioning a Video Based Motion Management System David Shepard Swedish Cancer Institute 2 1 Acknowledgments Daliang Cao, PhD, SCI Mohammed Jermoumi, PhD, SCI Roger Xie, PhD, SCI Malin

More information

Quick Reference Datasheet For All RIT113 Packages

Quick Reference Datasheet For All RIT113 Packages Quick Reference Datasheet For All RIT113 Packages For Rotational Therapies, IMRT & TG142 Highlights and selected product information only. A complete TG142 brochure is available. For more information on

More information

8/2/2017. PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA

8/2/2017. PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA Jennifer R. Clark, M.S., DABR Conflict of Interest Statement: I am an employee of Sun Nuclear Corporation Outline: Platform overview

More information

Patient Set-ups and Tumor Localizations

Patient Set-ups and Tumor Localizations Patient Set-ups and Tumor Localizations Amy S. Harrison Patient Positioning Prior to starting any localization or simulation procedure patients need to be positioned and immobilized Patients disease location

More information

ADVANCING CANCER TREATMENT

ADVANCING CANCER TREATMENT 3 ADVANCING CANCER TREATMENT SUPPORTING CLINICS WORLDWIDE RaySearch is advancing cancer treatment through pioneering software. We believe software has un limited potential, and that it is now the driving

More information

Volumetric Cine Imaging for On-board Target Localization in Radiation Therapy. Wendy Beth Harris. Graduate Program in Medical Physics Duke University

Volumetric Cine Imaging for On-board Target Localization in Radiation Therapy. Wendy Beth Harris. Graduate Program in Medical Physics Duke University Volumetric Cine Imaging for On-board Target Localization in Radiation Therapy by Wendy Beth Harris Graduate Program in Medical Physics Duke University Date: Approved: Lei Ren, Supervisor Fang-Fang Yin,

More information

8/3/2016. MR-guided RT: Commissioning and Quality Control. Topics for MR-guided RT system commissioning & QC

8/3/2016. MR-guided RT: Commissioning and Quality Control. Topics for MR-guided RT system commissioning & QC MR-guided RT: Commissioning and Quality Control T. Stanescu, PhD, MCCPM Medical Physicist, Princess Margaret Cancer Centre, UHN Assistant Professor, Radiation Oncology, University of Toronto Affiliated

More information

Disclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions

Disclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions LINAC and MLC QA for IMRT Ping Xia, Ph.D., Department of Radiation Oncology Disclosure Received research support from Siemens Medical Solutions University of California San Francisco Therapy Series (SAM)

More information

7/29/2017. Making Better IMRT Plans Using a New Direct Aperture Optimization Approach. Aim of Radiotherapy Research. Aim of Radiotherapy Research

7/29/2017. Making Better IMRT Plans Using a New Direct Aperture Optimization Approach. Aim of Radiotherapy Research. Aim of Radiotherapy Research Making Better IMRT Plans Using a New Direct Aperture Optimization Approach Dan Nguyen, Ph.D. Division of Medical Physics and Engineering Department of Radiation Oncology UT Southwestern AAPM Annual Meeting

More information

ADVANCING CANCER TREATMENT

ADVANCING CANCER TREATMENT The RayPlan treatment planning system makes proven, innovative RayStation technology accessible to clinics that need a cost-effective and streamlined solution. Fast, efficient and straightforward to use,

More information

GPU applications in Cancer Radiation Therapy at UCSD. Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC

GPU applications in Cancer Radiation Therapy at UCSD. Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC GPU applications in Cancer Radiation Therapy at UCSD Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC Conventional Radiotherapy SIMULATION: Construciton, Dij Days PLANNING:

More information

REAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT

REAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT REAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT Anand P Santhanam Assistant Professor, Department of Radiation Oncology OUTLINE Adaptive radiotherapy for head and

More information

Real-time monitoring of linac performance using RT plans and logfiles

Real-time monitoring of linac performance using RT plans and logfiles Real-time monitoring of linac performance using RT plans and logfiles NCCAAPM Wisconsin Dells 7 April 2017 Mark Wiesmeyer, PhD, DABR Technical Product Manager Technical talk with insights and results based

More information

Qalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004

Qalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004 Qalitätssicherung an Multileafkollimatoren Dr. Lutz Müller Würzburg jan 2004 IMRT Verification - present Target Volume Constraints Inverse Planning Algorithm Fluence Map Leaf Sequencer Leaf & Gantry sequence

More information

Introduction. Quality Assurance for Image- Guided Radiation Therapy. Justification for IGRT. Image-Guided Radiation Therapy

Introduction. Quality Assurance for Image- Guided Radiation Therapy. Justification for IGRT. Image-Guided Radiation Therapy Introduction Quality Assurance for Image- Guided Radiation Therapy Jean-Pierre Bissonnette, Ph.D., MCCPM Princess Margaret Hospital, Toronto, Canada IGRT What is it? Rationale Equipment Quality Assurance

More information

Advanced Radiotherapy

Advanced Radiotherapy Advanced Radiotherapy IBA Concept for IMRT Verification Dr. Lutz Müller 1 IMRT Verification BIS BIS ADAS Fluence Model Delivered fluence Target Volume Constraints Inverse Backprojection Fluence Map Leaf

More information

Optical Guidance. Sanford L. Meeks. July 22, 2010

Optical Guidance. Sanford L. Meeks. July 22, 2010 Optical Guidance Sanford L. Meeks July 22, 2010 Optical Tracking Optical tracking is a means of determining in real-time the position of a patient relative to the treatment unit. Markerbased systems track

More information

Secondary 3D Dose QA Fully Automated using MOSAIQ's IQ Engine. MOSAIQ User Meeting May Antwerp

Secondary 3D Dose QA Fully Automated using MOSAIQ's IQ Engine. MOSAIQ User Meeting May Antwerp Secondary 3D Dose QA Fully Automated using MOSAIQ's IQ Engine MOSAIQ User Meeting May 31 2013 - Antwerp Contents Project goal and collaboration Secondary 3D Dose QA project justification Secondary 3D Dose

More information

Facility Questionnaire PART I (General Information for 3DCRT and IMRT)

Facility Questionnaire PART I (General Information for 3DCRT and IMRT) Facility Questionnaire PART I (General Information for 3DCRT and IMRT) The following items are required before you can enter cases on any RTOG protocol that requires data submission to the Image-Guided

More information

Comparison of Two Phantoms for End-to-End SRS/SBRT Testing. Vikren Sarkar, Ph.D. Associate Professor University of Utah Huntsman Cancer Center

Comparison of Two Phantoms for End-to-End SRS/SBRT Testing. Vikren Sarkar, Ph.D. Associate Professor University of Utah Huntsman Cancer Center Comparison of Two Phantoms for End-to-End SRS/SBRT Testing Vikren Sarkar, Ph.D. Associate Professor University of Utah Huntsman Cancer Center Disclosure Travel and Speaker Honorarium from Sun Nuclear Corp.

More information

Delta 4 PT. True Volumetric Pre-treatment Verification. The difference is clear

Delta 4 PT. True Volumetric Pre-treatment Verification. The difference is clear Delta 4 PT True Volumetric Pre-treatment Verification The difference is clear ScandiDos innovators in advanced volumetric dosimetry solutions ScandiDos ScandiDos is the innovative company that introduced

More information

3DVH FAQs. What is PDP questions

3DVH FAQs. What is PDP questions 3DVH FAQs What is PDP questions 1. Explain the PDP in layman terms. How does PDP work? a. Very simply, PDP uses measured diode data, and compares it to the expected treatment plan data. The differences

More information

Thank-You Members of TG147 TG 147: QA for nonradiographic

Thank-You Members of TG147 TG 147: QA for nonradiographic Thank-You Members of TG147 TG 147: QA for nonradiographic localization and positioning systems Twyla Willoughby, M.S. Medical Physicist Clinical AAPM Meeting March 2013 Department of Radiation Oncology

More information

Monaco Concepts and IMRT / VMAT Planning LTAMON0003 / 3.0

Monaco Concepts and IMRT / VMAT Planning LTAMON0003 / 3.0 and IMRT / VMAT Planning LTAMON0003 / 3.0 and Planning Objectives By the end of this presentation you can: Describe the cost functions in Monaco and recognize their application in building a successful

More information

Simultaneous MV-kV imaging for intrafractional motion management during volumetric-modulated arc therapy delivery *

Simultaneous MV-kV imaging for intrafractional motion management during volumetric-modulated arc therapy delivery * JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Simultaneous MV-kV imaging for intrafractional motion management during volumetric-modulated arc therapy delivery * Margie A. Hunt,

More information

Takayama, Kenji; Mizowaki, Takashi; Nagano, Kazuo; Kamino, Yuichiro; Hi. Rightc 2009 Elsevier Ireland Ltd.

Takayama, Kenji; Mizowaki, Takashi; Nagano, Kazuo; Kamino, Yuichiro; Hi.   Rightc 2009 Elsevier Ireland Ltd. Title Initial validations for pursuing ir tracking system. Takayama, Kenji; Mizowaki, Takashi; Author(s) Kawada, Noriyuki; Nakayama, Hiroshi Nagano, Kazuo; Kamino, Yuichiro; Hi Citation Radiotherapy and

More information

ICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588

ICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588 ICARO Vienna April 27-29 2009 Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588 M. Saiful Huq, Ph.D., Professor and Director, Dept. of Radiation

More information

Thoracic target volume delineation using various maximum-intensity projection computed tomography image sets for stereotactic body radiation therapy

Thoracic target volume delineation using various maximum-intensity projection computed tomography image sets for stereotactic body radiation therapy Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 8-2010 Thoracic target volume delineation using various maximum-intensity

More information

VALIDATION OF DIR. Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology

VALIDATION OF DIR. Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology VALIDATION OF DIR Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology Overview Basics: Registration Framework, Theory Discuss Validation techniques Using Synthetic CT data & Phantoms What metrics to

More information

An investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition

An investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 An investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition Yildirim D. Mutaf

More information

Simple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field

Simple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Simple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field Hideharu Miura, 1a Shuichi

More information

MRI-LINAC Dynamic Phantom

MRI-LINAC Dynamic Phantom MRI-LINAC Dynamic Phantom Model 008M IMAGE ACQUISITION TREATMENT PLANNING DOSE DELIVERY Strict QA procedures for the imaging, planning and delivery of radiotherapy using respiratory management devices

More information

Performance and characteristics of an IR localizing system for radiation therapy

Performance and characteristics of an IR localizing system for radiation therapy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 2, SPRING 2006 Performance and characteristics of an IR localizing system for radiation therapy Yulia Lyatskaya, 1 Hsiao-Ming Lu, 2 and Lee

More information

The IORT Treatment Planning System. radiance. GMV, 2012 Property of GMV All rights reserved

The IORT Treatment Planning System. radiance. GMV, 2012 Property of GMV All rights reserved The IORT Treatment Planning System radiance Property of GMV All rights reserved WHY RADIANCE? JUSTIFICATION Property of GMV All rights reserved ADVANTAGES OF IORT PRECISION: RT guided by direct vision.

More information

Use of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta

Use of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta Use of Monte Carlo modelling in radiotherapy linac design David Roberts, PhD Senior Physicist Elekta Contents Overview of Elekta What we do Where we use Monte Carlo Codes and resources Example : Agility

More information

ATTENTION! The whole content of the lecture with all the animations can be dowloaded from:

ATTENTION! The whole content of the lecture with all the animations can be dowloaded from: ATTENTION! The whole content of the lecture with all the animations can be dowloaded from: http://www.onko.szote.uszeged.hu/letoltes/radioter_phys_basis/radiother_phys_basis.zip Size > 300 M!!! The dowloaded.zip

More information

Integrated proton-photon treatment planning

Integrated proton-photon treatment planning Pinnacle 3 Proton Planning Integrated proton-photon treatment planning Philips Pinnacle 3 Proton Planning specifications Pinnacle 3 Proton Planning is designed to simplify treatment planning for proton

More information

RITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package.

RITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package. So much is asked of medical physicists these days We Can Help Automated Radiation Therapy Phantom Analysis in Seconds QC for Therapy OBI s and CT simulators. Radia s Catphan/OBI module performs analysis

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Virtual Phantoms for IGRT QA

Virtual Phantoms for IGRT QA TM Virtual Phantoms for IGRT QA Why ImSimQA? ImSimQA was developed to overcome the limitations of physical phantoms for testing modern medical imaging and radiation therapy software systems, when there

More information

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization

More information

Auto-Segmentation Using Deformable Image Registration. Disclosure. Objectives 8/4/2011

Auto-Segmentation Using Deformable Image Registration. Disclosure. Objectives 8/4/2011 Auto-Segmentation Using Deformable Image Registration Lei Dong, Ph.D. Dept. of Radiation Physics University of Texas MD Anderson Cancer Center, Houston, Texas AAPM Therapy Educational Course Aug. 4th 2011

More information

Ini$al Performance Test Results on an Elekta MR-Linac

Ini$al Performance Test Results on an Elekta MR-Linac Ini$al Performance Test Results on an Elekta MR-Linac Eric Paulson, PhD DABR Radia$on Oncology, Radiology, and Biophysics Froedtert and Medical College of Wisconsin Milwaukee, Wisconsin, United States

More information

Verification of Gated Radiation Therapy: Dosimetric Impact of Residual Motion

Verification of Gated Radiation Therapy: Dosimetric Impact of Residual Motion Original Article PROGRESS in MEDICAL PHYSICS Vol. 25, No. 3, September, 2014 http://dx.doi.org/10.14316/pmp.2014.25.3.128 Verification of Gated Radiation Therapy: Dosimetric Impact of Residual Motion Inhwan

More information

Financial disclosure. Onboard imaging modality for IGRT

Financial disclosure. Onboard imaging modality for IGRT Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty

More information

Measurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images

Measurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images Measurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images T E Marchant, A M Amer 1 and C J Moore North Western Medical Physics, Christie Hospital NHS Foundation

More information

Use of Deformable Image Registration in Radiation Therapy. Colin Sims, M.Sc. Accuray Incorporated 1

Use of Deformable Image Registration in Radiation Therapy. Colin Sims, M.Sc. Accuray Incorporated 1 Use of Deformable Image Registration in Radiation Therapy Colin Sims, M.Sc. Accuray Incorporated 1 Overview of Deformable Image Registration (DIR) Algorithms that can deform one dataset to another have

More information

CLINICAL ASPECTS OF COMPACT GANTRY DESIGNS

CLINICAL ASPECTS OF COMPACT GANTRY DESIGNS CLINICAL ASPECTS OF COMPACT GANTRY DESIGNS J. Heese, J. Wulff, A. Winnebeck, A. Huggins, M. Schillo VARIAN PARTICLE THERAPY JUERGEN HEESE New gantry developments Viewpoint from user and vendor perspective

More information

Dynamic Image Prediction Using Principal Component and Multi-Channel Singular Spectral Analysis: A Feasibility Study

Dynamic Image Prediction Using Principal Component and Multi-Channel Singular Spectral Analysis: A Feasibility Study Open Journal of Medical Imaging, 205, 5, 33-42 Published Online September 205 in SciRes. http://www.scirp.org/journal/ojmi http://dx.doi.org/0.4236/ojmi.205.5307 Dynamic Image Prediction Using Principal

More information

Retrospective Spiral Respiratory Correlated Imaging with Varian RPM

Retrospective Spiral Respiratory Correlated Imaging with Varian RPM Retrospective Spiral Respiratory Correlated Imaging with Varian RPM This is a Quick Step Guide for Retrospective Spiral Respiratory Correlated Imaging (4D CT) using the Varian RPM device v1.7 with the

More information

Analysis of 4D CT cine images for the characterization of organ motion due to breathing

Analysis of 4D CT cine images for the characterization of organ motion due to breathing -243- Analysis of 4D CT cine images for the characterization of organ motion due to breathing Maria Francesca Spadea 1,2, Marta Peroni 2, Marco Riboldi 2, Guido Baroni 2, George TY Chen 3, Gregory Sharp

More information

Basic Radiation Oncology Physics

Basic Radiation Oncology Physics Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides

More information

Mech. Engineering, Comp. Science, and Rad. Oncology Departments. Schools of Engineering and Medicine, Bio-X Program, Stanford University

Mech. Engineering, Comp. Science, and Rad. Oncology Departments. Schools of Engineering and Medicine, Bio-X Program, Stanford University Mech. Engineering, Comp. Science, and Rad. Oncology Departments Schools of Engineering and Medicine, Bio-X Program, Stanford University 1 Conflict of Interest Nothing to disclose 2 Imaging During Beam

More information

TEPZZ Z754_7A_T EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (51) Int Cl.: A61N 5/10 ( )

TEPZZ Z754_7A_T EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (51) Int Cl.: A61N 5/10 ( ) (19) TEPZZ Z74_7A_T (11) EP 3 07 417 A1 (12) EUROPEAN PATENT APPLICATION (43) Date of publication: 0..16 Bulletin 16/ (1) Int Cl.: A61N / (06.01) (21) Application number: 16163147.8 (22) Date of filing:

More information