Investigation of the kinematic system of a 450 kv CT scanner and its influence on dimensional CT metrology applications
|
|
- Percival Burke
- 6 years ago
- Views:
Transcription
1 Investigation of the kinematic system of a 450 kv CT scanner and its influence on dimensional CT metrology applications Frank Welkenhuyzen 1, Bart Boeckmans 1, Ye Tan 1,2, Kim Kiekens 1,2, Wim Dewulf 1,2, Jean-Pierre Kruth 1 1 KU Leuven, Mechanical Engineering Department, Celestijnenlaan 300B, 3001 Heverlee, Belgium, Frank.Welkenhuyzen@mech.kuleuven.be 2 KU Leuven, Campus Group T, Andreas Vesaliusstraat 13, 3000 Leuven, Belgium Abstract The accuracy of CT-based dimensional measurements remains yet largely uncertain, due to the high number of influencing factors related to the workpiece as well as the CT equipment, the measurement setup and subsequent data processing steps. [1] gives an overview of these unwanted effects which disturb the dimensional CT measurements. This paper focuses on the kinematic system of the CT scanner and the subsequent errors introduced by the imperfect positioning of this kinematic system. First, a kinematic model for a 450 kv CT scanner is presented. Next the geometrical error components (specified in ISO 230-1) and alignment errors between X-ray source, rotation table and detector have been determined. Finally a calibrated reference object and simulations have been used to show the influence of the geometrical errors and alignment errors on the dimensional measurements conducted using CT. Keywords: Computed Tomography, Dimensional Metrology, Simulation 1 Introduction 1.1 Problem description The accuracy of CT-based dimensional measurements is largely uncertain, due to the high number of influencing factors related to the workpiece as well as the CT equipment, the measurement setup and subsequent data processing steps. The many influencing factors, together with various additional often unknown influences (e.g. drift of the source, temperature), make it difficult to correlate separate influencing factors to observed variations in dimensional measurements when performing experimental investigations or measurements, and hence it is inconvenient to draw correct conclusions from these experimental measurements. [1] gives an overview of unwanted effects that disturb dimensional CT measurements. There is quite some literature that describes, investigates and/or provides corrections for beam hardening effects, scatter artifacts, ring artifacts, etc. Furthermore different suggested calibration objects for voxel size calibration (rescaling), beam hardening correction, etc. can be found in literature. This paper focuses on the kinematic system of the CT scanner and the subsequent errors introduced by the imperfect positioning of this kinematic system. When discussing the influencing factors on dimensional CT measurements, the main discussion topics are often related to beam hardening and edge detection. The kinematic system is another important aspect influencing the dimensional CT measurements. First of all it has a direct influence on the voxel size. Regularly a calibration object is used to rescale the voxel sizes but this is often insufficient. As a second point the kinematic system is linked to alignment errors between the rotation axis, the X-ray source and the detector: the rotation axis is supposed to be aligned parallel, resp. perpendicular to the detector and its pixel rows or columns [1]). This link will not only influence each individual measurement, but will also affect the initial machine calibration (alignment), executed by the manufacturer, to align the X-ray source, rotation ict Conference
2 table and detector. Furthermore the negative influences of the kinematic system and subsequent alignment errors play a role in two other critical steps of dimensional CT metrology: during the volume reconstruction step and during the segmentation or edge detection (determining the respective transitions between solid materials and surrounding air or between different solid materials). In addition a better knowledge of the kinematic system and subsequent alignment errors allows to perform better measurement uncertainty calculations, as the uncertainty parameters related to the kinematic system are often (incorrectly) defined by means of guessing while these errors are often repeatable. Therefore the errors of the kinematic system should be known in advance to make an accurate estimation of the measurement uncertainty. 1.2 Hardware and software The investigated CT scanner is a Nikon Metrology XT H450. With a voltage ranging up to 450 kv, the scanner is well equipped for penetrating thicker structures and materials with a high attenuation coefficient compared to standard CT scanners for dimensional metrology. The CT scanner makes use of two detectors: a flat panel detector (FPD) and a curved linear detector array (CLDA). After data acquisition, the 2D images are reconstructed using CTPro (reconstruction software of Nikon Metrology). Ensuing volume reconstruction, the volume can be loaded in VGStudio Max 2.2 for further data processing and geometrical analysis. A critical step is edge detection or segmentation: determining the respective transitions between solid materials and surrounding air or between different solid materials. The edge detection has been performed as follows in this paper: first a global threshold is determined by defining one threshold gray value. This results in a rough part contour. After this the advanced surface determination of VGStudio MAX 2.2 is used: starting from the previously defined contour, a more accurate surface determination is obtained. Further measurements have taken place in VGStudio. 2 Definition of the kinematic model 2.1 Problem description Figure 1: Kinematic system of the investigated CT scanner. The kinematic system of the investigated CT scanner is illustrated in figure 1. It consists of a turntable for stepwise or continuous rotation, two axes for horizontal translation (x and z) and a vertical translation axis (y). Furthermore the FPD can shift along a vertical axis, to be able to work with the CLDA which is positioned behind the FPD (figure 1, right). Errors of the kinematic system will unambiguously introduce errors on the dimensional measurements: The rotation axis is supposed to be aligned parallel, resp. perpendicular, to the detector and its pixel rows or columns [1]. 218
3 Error from wrong identification of rotation center: at the beginning of the 3D reconstruction the software of the CT device will try to identify the location of the axis around which the part was rotated during CT image capturing. Failure in precise identification of the rotation center/axis will introduce reconstruction errors [1]. Positioning errors and repeatability problems of the Z-axis directly influence the measured dimensions through a change in magnification factor. Errors of the kinematic system will hinder the initial machine calibration (to align the rotation axis parallel, resp. perpendicular, to the detector and its pixel rows or columns, and to determine the source/object and source/detector distances) performed by the manufacturer. 2.2 Kinematic model The kinematic system of the CT scanner has been modelled as a kinematic chain of 4 rigid bodies connected by 3 prismatic joints. The frames have been assigned as follows: Frame {0} connected to fixed structure. Frame {1} connected to z-carriage. Frame {2} connected to y-carriage. Frame {3} connected to x-carriage. Frame {rt} which is positioned at the center of the turntable. Frame {rt} is a translation of frame {3} to the center of the turntable, as the object will be described in function of this frame. The pose of each frame with respect to the previous frame can be described with the help of homogeneous transformation matrices. Taking into account the different error motions (conventions according to [2] are used) these transformation matrices look like: ict Conference
4 With the parameters defined as follows: z enc the value read from the z-scale. ( 0 x0z, 0 y0z, 0 z0z) corresponding to the beginning (home position) of the z-scale expressed in frame {0}. exz and eyz the straightness error motions of the z-axis in x and y direction ezz the positioning error of the z-axis. eaz, ebz and ecz the angular error motions of the z-axis around x, y and z. Similar notations are used for the x and y axes. Squareness errors have not been added as a separate value but have been taken into account in the previously described errors. The coordinates of a point with respect to frame {rt} can now be calculated in frame {0} as: Determination of the error components Figure 2: Measurement of the error components with the laser interferometer (a), electronic level (b) and verification with the dial gauge and a granite square with a precise angle of 90 degrees (c). The above described errors for each axis (one positional error, two straightness errors and three angular errors) in function of the axis position, and the squareness errors between the axes, have been determined using a laser interferometer and an electronic level (figure 2). Furthermore a granite square with a precise angle of 90 degrees and dial gauge have been used to verify the measured errors (figure 2c). Figure 3 shows some of the results. Figure 3a shows the error on the magnification position (error in z of the z-axis) on a certain height (the table can move from z=-285, which is near the source, up to z=280, which is near the detector). It is clear that these errors will have a negative influence on the determination of the magnification, and therefore also on the voxelsizes and dimensional measurements. Figure 3b shows the straightness error (ezy) along the y-axis (squareness error not yet included); this already points out that scanning a calibration object for rescaling should be done at the same table height. Figure 3c shows the pitch error measurement (eaz) along the z-axis. 220
5 Figure 3: Measured error components. 2.4 Measurement of relative errors between source, turntable and detector The above explained method calculates errors/positions with respect to frame {0}, which is connected to the fixed structure. Now the errors should be determined relative to the source and detector position and orientation. The following rule has been used to estimate these values: a point object in the center(line) of the detector should stay on the centerline of the detector when moving in the plane determined by this center line and the source. Figures 4a and 4b give the example for an angular error of the rotation axis around x. In the ideal situation the point (intersection of x) should stay in the center of the detector when rotating (figure 4a). Figure 4b, illustrates the real situation where we see a shift downwards on the detector due to an inclination of the rotation axis. With this rule the position of the turntable relative to the X-ray source and detector has been estimated. Two touching spheres have been used (figure 4c), where the touch point served as the before mentioned point. The composed object has been turned 90 degrees for errors in the horizontal direction. The previously calculated errors of the kinematic system have been taken into account when estimating the relative errors. Figure 4: The influence of the inclination of the rotation axis (a and b) and an X-ray image of the object used to determine the relative errors (c). 3 Simulation program A simulation program has been developed and was presented in earlier work [3]. The simulation program generates 2D TIFF images that can be reconstructed into a 3D model as if they are real projections from a CT scan. Multiple inputs are needed and are taken into account in the simulation program. The object can be constructed out of geometrical features like (truncated) cylinders, (truncated) cones, (truncated) spheres, bars and al kind of combinations. A short overview of the input parameters: Source: spot size, position, drift in time, current, voltage, etc. Object: definition of the object, object position, material composition, etc. Turntable: source object distance, number of rotation steps, etc. Detector: number of pixels, pixelsize, distance source detector, detector attenuation processes (type and thickness of the scintillator), etc. ict Conference
6 The simulation program has been extended by taking into account the errors described in section 2. This will result in more realistic simulations. Furthermore for each input parameter there is the possibility to define an uncertainty range with respect to this input parameter. This allows to perform uncertainty calculations, which is necessary as the real input parameters are never known perfectly and can deviate from measurement to measurement. 4 Case study 4.1 Reference object A reference object is developed and produced by fixing CMM probing styli on a socket (figure 5, left), which has screw holes on distances 22.5, 37.5 and 52.5 mm to the center. The use of such styli has already been described in some literature [4]. The styli are made of ruby spheres on carbon fiber rods, with diameters 4, 6 and 8 mm. The application of the reference object is twofold. Firstly, it is possible to measure distances between sphere centers. Using sphere centers cancels out existing beam hardening and threshold errors. The sphere centers are independent of the sphere diameter. Secondly, sphere diameters can be measured to quantify beam hardening and/or threshold errors. Distance sphere diameters 1 Dist mm 2 Dist mm 3 Dist mm 4 Dist mm 5 Dist mm 6 Dist mm 7 Dist mm 8 Dist mm Figure 5: Styli calibration object (left) with schematic representation (middle) and measured distances (right). 4.2 Calibration As a reference, the sphere distances (listed in figure 5, right) between several spheres are measured with a conventional tactile CMM Mitutoyo FN 905 with specified accuracy: u1 = L/1000 µm (with L in mm, for each axis). Each investigated distance has been measured 13 times, spread over 3 days. Afterwards the standard deviation on each distance has been calculated, resulting in a maximum standard deviation value of 1 µm. Figure 6 shows the CMM measurement results for the distance between spheres 2 and 3. The specified nominal values have been used as reference values for the diameters, as the sphere diameters have sub micrometer accuracy. Figure 6: Measured CMM distance (distance 2-3) in function of measurement number. 222
7 4.3 Results CT measurements are repeated for different magnifications (figure 7, left). After performing a CT scan and reconstructing the volume, this volume is loaded in VGStudio. Because of the inaccuracy of the magnification axis, it is necessary to rescale the volume. The volume can be rescaled by adjusting the voxelsize in VGStudio. Determining the correct voxelsize is done by compensating the CT measured distance between spheres 2 and 3 to the calibration value (CMM measurement of distance between spheres 2 and 3) Distances The investigated distances are listed and schematically represented in figure 7. Figure 7, right, gives the deviations between the measured CT values and the calibrated CMM values (blue rhombus) in function of the measured distance (see figure 5) for measurement 1 (magnification 2.66). The red squares give the results of the simulated measurement with the same input parameters. However these input parameters are never exactly known. Therefore 30 additional simulations have been done where each simulation input parameter (errors of the kinematic system, source spotsize, source drift, object position, ) was chosen out of a small uncertainty interval related to that parameter: this way a 95% uncertainty interval has been determined (green triangle and the purple x representing the upper and lower limit of the 95% interval). Distance number 1 (distance between spheres 2 and 3) has zero deviation as this distance is used for rescaling. It is clear that the result obtained from real CT measurement is similar to the simulated measurement, although there is still a small difference, but it is contained in the 95% uncertainty interval. Further research will be performed to determine the relative errors between source, turntable and detector more accurately. Figure 7: The magnification and the resulting voxel size of each CT measurement (left) and deviations (in mm) to the nominal distances for measurement 1 (right) Diameters When evaluating the diameters of the 4 mm spheres, there was a trend visible for all magnifications (figure 8a): the diameter of the center sphere (sphere 12, located near the rotation axis of the CT scanner while scanning) is always largest, the sphere furthest away from the center (sphere 1) always has the smallest diameter and the other spheres seem to have comparable diameters. As a test, a new CT scan has been executed with sphere 6 closest to the rotation axis. The results are shown in figure 8b: a similar trend is present. Again the measured sphere diameters depend of the distance to the axis of rotation. Figure 8c shows the deviations of a large number of points on a CT measured sphere (real or simulated CT measurement) with respect to its least square fitted sphere. This least square fit is assumed to represent the real physical sphere since this sphere is highly accurate with very small unroundness. Those deviation plots allow to evaluate the sphere form errors introduced by the (real or ict Conference
8 simulated) CT measurement: the real sphere being assumed to have no or minimal form errors. Figure 8c compares the sphere near the axis of rotation (left) and another sphere (right). It is clear that the latter sphere (top right) suffers from a form deviation introduced by the CT measurement. The results of real CT measurements (two top spheres) have been compared to the results of simulated CT measurements: Figure 8c, bottom spheres, shows the deviations of the sphere closest to the axis of rotation and another sphere when no alignment errors are implemented in the simulation and when there are alignment errors implemented. It is clear that the deviations are similar to these of the real measurements (top right) in case of the simulation with errors (middle right). In case of the simulation without alignment errors (bottom), all spheres have similar small deviations that fail to represent the effect of kinematic errors of the CT equipment. Figure 8d shows the calculated diameters for the 4 mm sphere diameters in case of the simulation with errors (sphere 12 on the rotation axis and magnification 2.66). A similar trend can be seen as for the real CT measurements (figure 8a). In case of the simulation without errors, this trend was not present. Figure 8: Figures a and b give the measured sphere diameters in case of sphere 12 near the axis of rotation (a) and in case of sphere 6 near the axis of rotation (b, only measurements 2 to 5 have been executed). Figure c gives the fitpoint deviations to the fitted least square spheres for the real measurement and the simulation with and without errors for the sphere near the axis of rotation and another sphere. Figure d gives the measured sphere diameters in case of sphere 12 near the axis of rotation for a simulated measurement with errors. 5 Conclusions This paper presented a kinematic model for a 450 kv CT scanner. The error components (specified in ISO 230-1) have been determined using a laser interferometer and an electronic level. Furthermore the alignment between the X-ray source, rotation table and detector has been investigated with a calibration object. Results of dimensional CT measurements are shown to validate the results and prove the importance. Real-life CT measurements as well as simulated results are presented. References [1] J.P. Kruth, M. Bartscher, S. Carmignato, R. Schmitt, L. De Chiffre, A. Weckenmann, 2011, Computed Tomography for Dimensional Metrology, CIRP Annals - Manufacturing Technology, 60/2:
9 [2] ISO 230-1:1996. Test code for machine tools Part 1: Geometric accuracy of machines operating under no-load or finishing conditions. ISO, [3] F. Welkenhuyzen, B. Boeckmans, J.P. Kruth, W. Dewulf, A. Voet, Simulation of X-ray projection images for dimensional CT metrology, 2012, Proc. 5th Intern. Conf. on optical measurement techniques for structures and systems pages: [4] H. Lettenbauer, B. Georgi, and D. Weiß. Means to verify the accuracy of ct systems for metrology applications (in the absence of established international standards). DIR - International symposium on digital industrial radiology and computed tomography, 4, ict Conference
Parameter Dependent Thresholding for Dimensional X-ray Computed Tomography
International Symposium on Digital Industrial Radiology and Computed Tomography - Poster 12 Parameter Dependent Thresholding for Dimensional X-ray Computed Tomography Kim KIEKENS *, Ye TAN *, Jean-Pierre
More informationTHREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS
THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS A. Campos a*, A. Zorrilla a, F. Lasagni a a Materials & Processes dept., Center for Advanced Aerospace Technologies
More informationDevelopments in Dimensional Metrology in X-ray Computed Tomography at NPL
Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Wenjuan Sun and Stephen Brown 10 th May 2016 1 Possible factors influencing XCT measurements Components Influencing variables Possible
More informationIndustrial Computer Tomography for Dimensional Metrology: Overview of Influence Factors and Improvement Strategies
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 Industrial Computer Tomography for Dimensional Metrology: Overview of Influence Factors and
More informationFrequency-based method to optimize the number of projections for industrial computed tomography
More Info at Open Access Database www.ndt.net/?id=18739 Frequency-based method to optimize the number of projections for industrial computed tomography Andrea Buratti 1, Soufian Ben Achour 1, Christopher
More informationComputed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system
Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system ECNDT 2014 Prague October 6-10, 2014 Dr. Eberhard Neuser Dr. Alexander Suppes Imagination
More informationInvestigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography
Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography Natalia Grozmani 1, Andrea Buratti 1, Robert H. Schmitt 1 More info about this article:
More informationOn the impact of probing form error on form measurement in Computed Tomography
On the impact of probing form error on form measurement in Computed Tomography Jan Lasson Andreasen 1, Filippo Montanari 2, Leonardo De Chiffre 2 More info about this article: http://www.ndt.net/?id=23729
More informationMulti-spectral (W, Mo, Cu, and Ag) XCT measurements
Multi-spectral (W, Mo, Cu, and Ag) XCT measurements Marko Katić 1, Damir Markučič 1, Gorana Baršić 1 Faculty of mechanical engineering and naval architecture, Ivana Lucica 5, Zagreb, Croatia, marko.katic@fsb.hr
More informationJapan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field. Kouichi Inagaki ICC / IHI Corporation
Japan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field Kouichi Inagaki ICC / IHI Corporation 13 th WORLD CONFERENCE ON INVESTMENT CASTING Paper: T3 Copyright
More informationInvestigation on the influence of image quality in X-ray CT metrology
Investigation on the influence of image quality in X-ray CT metrology Pavel Müller 1, Jochen Hiller 1, Angela Cantatore 1, Markus Bartscher 2, Leonardo De Chiffre 1 1 Technical University of Denmark, Department
More informationEfficient 3D Crease Point Extraction from 2D Crease Pixels of Sinogram
Efficient 3D Crease Point Extraction from 2D Crease Pixels of Sinogram Ryo Jinnouchi 1, Yutaka Ohtake 1, Hiromasa Suzuki 1, Yukie Nagai 1 1 The University of Tokyo, Japan, e-mail: {jinnouchi, yu-ohtake,
More informationComparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling
nd International Symposium on NDT in Aerospace - We..A. Comparison of Probing Error in Dimensional Measurement by Means of D Computed Tomography with Circular and Helical Sampling Jochen HILLER, Stefan
More informationMore Info at Open Access Database
More Info at Open Access Database www.ndt.net/?id=18750 Abstract Enhancing the Accuracy of Computed Tomography Measurements using Data Filtering Alessandro Stolfi a, Maarja-Helena Kallasse b, Lorenzo Carli
More informationINVESTIGATION OF MEASURING STRATEGIES IN COMPUTED TOMOGRAPHY
NEWTECH 2011 NEW TECHNOLOGIES IN MANUFACTURING ISBN 978-80-214-4267-2 INVESTIGATION OF MEASURING STRATEGIES IN COMPUTED TOMOGRAPHY Pavel Müller, Jochen Hiller, Angela Cantatore, Leonardo De Chiffre Technical
More informationExperimental investigation of new multi-material gap reference standard for testing computed tomography systems
More info about this article: http://www.ndt.net/?id=20837 Experimental investigation of new multi-material gap reference standard for testing computed tomography systems Petr Hermanek 1, Fabrício Borges
More informationHow beam hardening influences the dimensional geometry in X-ray computed tomography
How beam hardening influences the dimensional geometry in X-ray computed tomography More info about this article: http://www.ndt.net/?id=20809 Hannah Corcoran 1, 2, Stephen Brown 2, Stuart Robson 1, Robert
More informationThickness Measurement of Metal Plate Using CT Projection Images and Nominal Shape
Thickness Measurement of Metal Plate Using CT Projection Images and Nominal Shape Tasuku Ito 1, Yutaka Ohtake 1, Yukie Nagai 2, Hiromasa Suzuki 1 More info about this article: http://www.ndt.net/?id=23662
More informationCreating a Multi-material Probing Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems
Creating a Multi-material Probing Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems More info about this article: http://www.ndt.net/?id=20848 Fabricio Borges de Oliveira
More informationA cone-beam CT geometry correction method based on intentional misalignments to render the projection images correctable
A cone-beam CT geometry correction method based on intentional misalignments to render the projection images correctable Felix Meli, Benjamin A. Bircher, Sarah Blankenberger, Alain Küng, Rudolf Thalmann
More informationHIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY
HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY David K. Lehmann 1, Kathleen Brockdorf 1 and Dirk Neuber 2 1 phoenix x-ray Systems + Services Inc. St. Petersburg, FL, USA 2 phoenix x-ray Systems + Services
More informationCharacterization of microshells experimented on Laser Megajoule using X-Ray tomography
Characterization of microshells experimented on Laser Megajoule using X-Ray tomography More info about this article: http://www.ndt.net/?id=20881 Alexandre Choux, Lise Barnouin, Ludovic Reverdy, Marc Theobald
More informationASSESSMENT OF MEASUREMENT UNCERTAINTY CAUSED IN THE PREPARATION OF MEASUREMENTS USING COMPUTED TOMOGRAPHY
XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 2009, Lisbon, Portugal ASSESSMENT OF MEASUREMENT UNCERTAINTY CAUSED IN THE PREPARATION OF MEASUREMENTS USING COMPUTED TOMOGRAPHY
More informationINFLUENCE OF GEOMETRICAL MAGNIFICATION ON COMPUTED TOMOGRAPHY DIMENSIONAL MEASUREMENTS
DOI: 10.2507/27th.daaam.proceedings.090 INFLUENCE OF GEOMETRICAL MAGNIFICATION ON COMPUTED TOMOGRAPHY DIMENSIONAL MEASUREMENTS Amalija Horvatić Novak, Biserka Runje, Danijel Butković This Publication has
More informationCoordinate Measuring Machines with Computed Tomography
Always a Step Ahead with Quality Coordinate Measuring Machines with Computed Tomography Multisensor Coordinate Measuring Machines with Computed Tomography Computed Tomography in Coordinate Measuring Machines
More informationGood Practice guide to measure roundness on roller machines and to estimate their uncertainty
Good Practice guide to measure roundness on roller machines and to estimate their uncertainty Björn Hemming, VTT Technical Research Centre of Finland Ltd, Finland Thomas Widmaier, Aalto University School
More informationUse of reference objects for correction of measuring errors in X-ray computed tomography
Downloaded from orbit.dtu.dk on: Nov 19, 2018 Use of reference objects for correction of measuring errors in X-ray computed tomography Müller, Pavel Publication date: 2010 Document Version Publisher's
More informationDigital Laminography and Computed Tomography with 600 kv for Aerospace Applications
4th International Symposium on NDT in Aerospace 2012 - Tu.3.A.1 Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications Malte KURFISS 1, Gerd STRECKENBACH 2 1 YXLON International
More informationVoxel dataset noise reduction algorithm based on CMM reference measurements for geometrical tolerance analysis of prismatic test pieces
Voxel dataset noise reduction algorithm based on CMM reference measurements for geometrical tolerance analysis of prismatic test pieces More info about this article: http://www.ndt.net/?id=21929 Crhistian
More informationRELIABILITY OF PARAMETRIC ERROR ON CALIBRATION OF CMM
RELIABILITY OF PARAMETRIC ERROR ON CALIBRATION OF CMM M. Abbe 1, K. Takamasu 2 and S. Ozono 2 1 Mitutoyo Corporation, 1-2-1, Sakato, Takatsu, Kawasaki, 213-12, Japan 2 The University of Tokyo, 7-3-1, Hongo,
More informationCreating a Multi-Material Length Measurement Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems
Creating a Multi-Material Length Measurement Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems More info about this article: http://www.ndt.net/?id=20847 Fabrício Borges
More informationSquareness C0Y A0Z B0Z
ETALON offers system solutions for highly accurate and complete geometric analysis, monitoring and accuracy improvement of machine tools, measuring machines, robots and structures. Innovative technologies
More informationTest Piece for 5-axis Machining Centers
Test Piece for 5-axis Machining Centers S. Bossoni, W. Knapp, K. Wegener Institute of Machine Tools and Manufacturing, ETH Zurich, Switzerland, and inspire AG für mechatronische Produktionssysteme und
More informationRegistration concepts for the just-in-time artefact correction by means of virtual computed tomography
DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Registration concepts for the just-in-time artefact correction by means of virtual
More informationSelection of items for InteraqCT Comparison on Assemblies
More Info at Open Access Database www.ndt.net/?id=18776 Selection for InteraqCT Comparison on Assemblies Alessandro Stolfi, Leonardo De Chiffre Technical University of Denmark, Produktionstorvet Building
More informationAn initial experimental study on the influence of beam hardening in X-ray CT for dimensional metrology
An initial experimental study on the influence of beam hardening in X-ray CT for dimensional metrology Joseph J. Lifton 1,2, Andrew A. Malcolm 2, John W. McBride 3 1 The University of Southampton, United
More informationDigital Image Processing
Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture
More informationDESIGN AND VERIFICATION OF THE TRINANO ULTRA PRECISION CMM
URN (Paper): urn:nbn:de:gbv:ilm1-2011iwk-092:8 56 TH INTERNATIONAL SCIENTIFIC COLLOQUIUM Ilmenau University of Technology, 12 16 September 2011 URN: urn:nbn:gbv:ilm1-2011iwk:5 DESIGN AND VERIFICATION OF
More informationCase Studies in the Use of Computed Tomography for Non-Destructive Testing, Inspection and Measurement
SINCE2013 Singapore International NDT Conference & Exhibition 2013, 19-20 July 2013 Case Studies in the Use of Computed Tomography for Non-Destructive Testing, Inspection and Measurement Andrew A. MALCOLM,
More informationCIVA Computed Tomography Modeling
CIVA Computed Tomography Modeling R. FERNANDEZ, EXTENDE, France S. LEGOUPIL, M. COSTIN, D. TISSEUR, A. LEVEQUE, CEA-LIST, France page 1 Summary Context From CIVA RT to CIVA CT Reconstruction Methods Applications
More informationThree-Dimensional Laser Scanner. Field Evaluation Specifications
Stanford University June 27, 2004 Stanford Linear Accelerator Center P.O. Box 20450 Stanford, California 94309, USA Three-Dimensional Laser Scanner Field Evaluation Specifications Metrology Department
More informationWieblinger Weg 92a, Heidelberg, Germany, Phone: , Fax: ;
HOW INDUSTRIAL COMPUTER TOMOGRAPHY ACCELERATES PRODUCT DEVELOPMENT IN THE LIGHT METAL CASTING AND INJECTION MOULDING INDUSTRY. Christof REINHART 1, Christoph POLIWODA 1, Thomas GÜNTHER 1 1 Volume Graphics
More informationCh 22 Inspection Technologies
Ch 22 Inspection Technologies Sections: 1. Inspection Metrology 2. Contact vs. Noncontact Inspection Techniques 3. Conventional Measuring and Gaging Techniques 4. Coordinate Measuring Machines 5. Surface
More informationUNIT IV - Laser and advances in Metrology 2 MARKS
UNIT IV - Laser and advances in Metrology 2 MARKS 81. What is interferometer? Interferometer is optical instruments used for measuring flatness and determining the lengths of slip gauges by direct reference
More informationRecognition and Measurement of Small Defects in ICT Testing
19 th World Conference on Non-Destructive Testing 2016 Recognition and Measurement of Small Defects in ICT Testing Guo ZHIMIN, Ni PEIJUN, Zhang WEIGUO, Qi ZICHENG Inner Mongolia Metallic Materials Research
More informationCalypso Construction Features. Construction Features 1
Calypso 1 The Construction dropdown menu contains several useful construction features that can be used to compare two other features or perform special calculations. Construction features will show up
More informationAdvanced Non-Destructive Testing by High Resolution Computed Tomography for 3D analysis of Automotive Components
Advanced Non-Destructive Testing by High Resolution Computed Tomography for 3D analysis of Automotive Components J. Luebbehuesen GE Sensing & Inspection Technologies GmbH, phoenix x-ray, Niels-Bohr-Str.
More informationQuality control phantoms and protocol for a tomography system
Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing
More informationScienceDirect. Comparison of Different Method of Measurement Geometry using CMM, Optical Scanner and Computed Tomography 3D
Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 69 (2014 ) 255 262 24th DAAAM International Symposium on Intelligent Manufacturing and Automation, 2013 Comparison of Different
More informationA geometry measurement system for a dimensional cone-beam CT
A geometry measurement system for a dimensional cone-beam CT More info about this article: http://www.ndt.net/?id=21904 Benjamin A. Bircher, Felix Meli, Alain Küng, Rudolf Thalmann Federal Institute of
More informationSimulation-based correction of systematic errors for CT measurements
Simulation-based correction of systematic errors for CT measurements Andreas Michael Müller 1, Florian Wohlgemuth 1, Tino Hausotte 1 1 Institute of Manufacturing Metrology, Friedrich-Alexander-Universität
More informationDiscrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset
Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset Andreas Maier 1, Patrick Kugler 2, Günter Lauritsch 2, Joachim Hornegger 1 1 Pattern Recognition Lab and SAOT Erlangen,
More informationSUPPLEMENTARY INFORMATION
SUPPLEMENTARY INFORMATION doi:10.1038/nature10934 Supplementary Methods Mathematical implementation of the EST method. The EST method begins with padding each projection with zeros (that is, embedding
More informationCT Measurement Realization Process: The Need for Advanced Correction Methods
CT Measurement Realization Process: The Need for Advanced Correction Methods Crhistian R Baldo 1, Thiago L Fernandes 2, Gustavo D Donatelli 2, Gergely Szikszay-Molnar 3, Kostadin Doytchinov 3 1 Federal
More informationTEP Hounsfield units. Related topics Attenuation coefficient, Hounsfield units
Hounsfield units TEP Related topics Attenuation coefficient, Hounsfield units Principle Depending on the type of CT scanner and the settings, the result of a CT scan of the same material can be different
More informationAvailable online at ScienceDirect
Available online at www.sciencedirect.com ScienceDirect Procedia CIRP 13 ( 2014 ) 43 48 2 nd CIRP 2nd CIRP Conference on Surface Integrity (CSI) Computed Tomography as a Tool for Examining Surface Integrity
More informationCOMPARISON OF ERROR MAPPING TECHNIQUES FOR COORDINATE MEASURING MACHINES USING THE PLATE METHOD AND LASER TRACER TECHNIQUE
I IMEKO World Congress Fundamental and Applied Metrology September 6, 9, Lisbon, Portugal COMPARISON OF ERROR MAPPING TECHNIQUES FOR COORDINATE MEASURING MACHINES USING THE PLATE METHOD AND LASER TRACER
More informationUncertainty Evaluation of Pore Analysis for Additively Manufactured Parts using Cross Sections
Uncertainty Evaluation of Pore Analysis for Additively Manufactured Parts using Cross Sections More info about this article: http://www.ndt.net/?id=23659 Abstract Leonard Schild 1, Manuel Fülling 1, Benjamin
More informationAccurate surface extraction on CT volume using analytical gradient of FDK formula
Accurate surface extraction on CT volume using analytical gradient of FDK formula Yukie Nagai 1, Yutaka Ohtake 2, and Hiromasa Suzuki 2 1 Tokyo Metropolitan University, Asahigaoka 6-6, Hino, Tokyo, Japan,
More informationA RADIAL WHITE LIGHT INTERFEROMETER FOR MEASUREMENT OF CYLINDRICAL GEOMETRIES
A RADIAL WHITE LIGHT INTERFEROMETER FOR MEASUREMENT OF CYLINDRICAL GEOMETRIES Andre R. Sousa 1 ; Armando Albertazzi 2 ; Alex Dal Pont 3 CEFET/SC Federal Center for Technological Education of Sta. Catarina
More informationForm Measurement ROUNDTEST RA-1600
Form Measurement ROUNDTEST RA-1600 A new PC-compliant Roundness and Cylindricity form measuring instrument with extensive analysis features to enable measurement of a wide variety of workpieces. Powerful
More informationMICHELSON S INTERFEROMETER
MICHELSON S INTERFEROMETER Objectives: 1. Alignment of Michelson s Interferometer using He-Ne laser to observe concentric circular fringes 2. Measurement of the wavelength of He-Ne Laser and Na lamp using
More informationACTA TECHNICA CORVINIENSIS Bulletin of Engineering Tome X [2017] Fascicule 2 [April June] ISSN:
Tome X [2017] Fascicule 2 [April June] ISSN: 2067 3809 1. Yung-Cheng WANG, 2. Bean-Yin LEE, 1. Chih-Hao HUANG, 3. Chi-Hsiang CHEN DEVELOPMENT OF THE AXIAL PRECISION INSPECTION SYSTEM FOR SPINDLES IN TOOL
More informationDetermination of Aspect Ratio Limitations, Accuracy and Repeatability of a Laser Line Scanning CMM Probe
Paper: Determination of Aspect Ratio Limitations, Accuracy and Repeatability of a Laser Line Scanning CMM Probe Bart Boeckmans,MinZhang,, Frank Welkenhuyzen, and Jean-Pierre Kruth Department Mechanical
More informationHIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS
23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS P. Helberg 1,
More informationNovel evaluation method of low contrast resolution performance of dimensional X-ray CT
More Info at Open Access Database www.ndt.net/?id=18754 Novel evaluation method of low contrast resolution performance of dimensional X-ray CT Makoto Abe 1, Hiroyuki Fujimoto 1, Osamu Sato 1, Kazuya Matsuzaki
More informationIND62 TIM CALIBRATION OF FREE-FORM STANDARD AND THEIR APPLICATIONS FOR IN-PROCESS MEASUREMENT ON MACHINE TOOLS
IND62 TIM CALIBRATION OF FREE-FORM STANDARD AND THEIR APPLICATIONS FOR IN-PROCESS MEASUREMENT ON MACHINE TOOLS VÍT ZELENÝ, IVANA LINKEOVÁ, PAVEL SKALNÍK (LENGTH MEASUREMENT DEPARTMENT) 5th November 2014
More informationThe development of a laser array measurement system for three dimensional positioning testing in machine tool
The development of a laser array measurement system for three dimensional ing testing in machine tool Wenyuh Jywe 1, Fong-Zhi Chen 2, Chun-Jen Chen 2, Hsin Hong Jwo 1, Jhih-Ming Pan 1 1 National Formosa
More informationAP Physics: Curved Mirrors and Lenses
The Ray Model of Light Light often travels in straight lines. We represent light using rays, which are straight lines emanating from an object. This is an idealization, but is very useful for geometric
More information3D Computed Tomography (CT) Its Application to Aerospace Industry
3D Computed Tomography (CT) Its Application to Aerospace Industry C. Muralidhar, M. P. Subramanian, V. Ravi Shankar and G. Chandrasekhar Directorate of Non Destructive Evaluation, Defence Research & Development
More informationEMRP JRP IND62 TIM: Use of on-board metrology systems for area-scanning on machine tools
EMRP JRP IND62 TIM: Use of on-board metrology systems for area-scanning on machine tools Author: Co-authors: doc. Ing. Vít Zelený, CSc. doc. Ing. Ivana Linkeová, Ph.D. Ing. Jakub Sýkora Pavel Skalník Jaromír
More informationEvaluation strategies in CT scanning
Downloaded from orbit.dtu.dk on: Dec 20, 2017 Evaluation strategies in CT scanning Hiller, Jochen Publication date: 2012 Document Version Publisher's PDF, also known as Version of record Link back to DTU
More informationROUNDTEST RA Roundness/Cylindricity Measurment
Form Measurement Roundness/Cylindricity Measurment ROUNDTEST RA-1600 Catalog No.E15000 A new PC-compliant roundness and cylindrical-form measuring instrument with extensive analysis features to enable
More informationStraightness measurement
Page 1 of 45 Straightness measurement User guide This manual is available in English only 2000-2001 Renishaw - 8 August 2001 Issue 5.1 Procedure for performing straightness measurement Straightness measurements
More informationAnalysis of Combined Probing Measurement Error and Length Measurement Error Test for Acceptance Testing in Dimensional Computed Tomography
Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18038 Analysis of Combined Probing
More informationThe Virtual CMM a software tool for uncertainty evaluation practical application in an accredited calibration lab
The Virtual CMM a software tool for uncertainty evaluation practical application in an accredited calibration lab Michael Trenk *, Matthias Franke **, Dr. Heinrich Schwenke ** * FEINMESS GmbH&Co. KG **
More informationLABORATORY SYSTEM FOR X-RAY NANOTOMOGRAPHY
79 LABORATORY SYSTEM FOR X-RAY NANOTOMOGRAPHY Alexander Sasov, SkyScan, Vluchtenburgstraat 3, Aartselaar B2630, Belgium, www.skyscan.be. ABSTRACT Using advanced X-ray technologies and X-ray scattering
More informationLength, Germany, PTB (Physikalisch-Technische Bundesanstalt)
Laser radiations Laser radiations Laser radiations Laser radiations Length Length Frequency stabilized laser (He- Ne): vacuum wavelength Frequency stabilized laser (He- Ne): absolute frequency Frequency
More informationOUT OF MACHINE CALIBRATION TECHNIQUE FOR ANALOG PROBES
XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil OUT OF MACHINE CALIBRATION TECHNIQUE FOR ANALOG PROBES J.J. Aguilar 1, J.A. Yagüe 1, J.A.
More informationInstruction manual for T3DS calculator software. Analyzer for terahertz spectra and imaging data. Release 2.4
Instruction manual for T3DS calculator software Release 2.4 T3DS calculator v2.4 16/02/2018 www.batop.de1 Table of contents 0. Preliminary remarks...3 1. Analyzing material properties...4 1.1 Loading data...4
More informationIdentification of Shielding Material Configurations Using NMIS Imaging
Identification of Shielding Material Configurations Using NMIS Imaging B. R. Grogan, J. T. Mihalczo, S. M. McConchie, and J. A. Mullens Oak Ridge National Laboratory, P.O. Box 2008, MS-6010, Oak Ridge,
More informationThe PolyCT Increasing the sample throughput using multiple rotation axes
The PolyCT Increasing the sample throughput using multiple rotation axes More info about this article: http://www.ndt.net/?id=21946 1 Dr. Nicolas Thiemeyer, Johannes Leisner 2, Michael Salamon 2, Carina
More informationgood check of volumetric accuracy. However, if the mea error components. However, if the errors measured are large,
REVIEW OF SCIENTIFIC INSTRUMENTS VOLUME 71, NUMBER 10 OCTOBER 2000 Laser vector measurement technique for the determination and compensation of volumetric positioning errors. Part I: Basic theory Charles
More informationCT Reconstruction with Good-Orientation and Layer Separation for Multilayer Objects
17th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China CT Reconstruction with Good-Orientation and Layer Separation for Multilayer Objects Tong LIU 1, Brian Stephan WONG 2, Tai
More informationImproving dimensional accuracy by error modelling and its compensation for 3-axis Vertical Machining Centre
Improving dimensional accuracy by error modelling and its compensation for 3-axis Vertical Machining Centre H M Dobariya, Y D Patel 2, D A Jani 3 Abstract In today s era, machining centres are very important
More informationPosition accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon
Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon WHITE PAPER Introduction An image guidance system based on Cone Beam CT (CBCT) is included in Leksell
More informationGeometric Accuracy Evaluation, DEM Generation and Validation for SPOT-5 Level 1B Stereo Scene
Geometric Accuracy Evaluation, DEM Generation and Validation for SPOT-5 Level 1B Stereo Scene Buyuksalih, G.*, Oruc, M.*, Topan, H.*,.*, Jacobsen, K.** * Karaelmas University Zonguldak, Turkey **University
More informationMulti-sensor measuring technology. O-INSPECT The best of optical and contact measuring technology for true 3D measurements.
Multi-sensor measuring technology O-INSPECT The best of optical and contact measuring technology for true 3D measurements. 2 // multifunctionality made BY CarL Zeiss The moment you realize that new requirements
More informationCT in Dimensional Metrology, Engineering and Manufacture
CT in Production Quality Control - Leuven, June 4, 2013 CT in Dimensional Metrology, Engineering and Manufacture Prof. Dr. Ir. Jean-Pierre KRUTH Katholieke Universiteit Leuven (K.U.Leuven) Mechanical Engineering
More informationExperimental accuracy assessment of different measuring sensors on workpieces with varying properties
Experimental accuracy assessment of different measuring sensors on workpieces with varying properties Rauf Oezden 1,*, Metin Aclan 1 and Saliba Danho 2 1 Technical University of Cluj-Napoca, Ph.D. Stud.,
More information501, , 1052, , 1602, 1604 EXCEL EXCEL 1602UC EXCEL 1052UC EXCEL 501HC. Micro-Vu Corporation. Precision Measurement Systems
501, 502 1051, 1052, 1054 1601, 1602, 1604 1602UC 1052UC 501HC Precision Measurement Systems 501, 502 1051, 1052, 1054 1601, 1602, 1604 Excel 501 HM/HC Excel 502 HM/HC Excel 501 Excel 502 Scale Resolution
More informationCT for Industrial Metrology Accuracy and Structural Resolution of CT Dimensional Measurements
CT for Industrial Metrology Accuracy and Structural Resolution of CT Dimensional Measurements Simone Carmignato 1, Anna Pierobon 2, Paola Rampazzo 1, Matteo Parisatto 3, Enrico Savio 2 1 University of
More informationAdvanced Reconstruction Techniques Applied to an On-Site CT System
2nd International Symposium on NDT in Aerospace 2010 - We.1.A.4 Advanced Reconstruction Techniques Applied to an On-Site CT System Jonathan HESS, Markus EBERHORN, Markus HOFMANN, Maik LUXA Fraunhofer Development
More informationFast checking of CMM geometry with a patented tool
17 International Congress of Metrology, 13012 (2015) DOI: 10.1051/ metrolo gy/201513012 C Owned by the authors, published by EDP Sciences, 2015 Fast checking of CMM geometry with a patented tool Jean-François
More informationDUE to beam polychromacity in CT and the energy dependence
1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact
More informationTraceable 3D X-Ray Measurements ZEISS METROTOM
20 Traceable 3D X-Ray Measurements With a computer tomograph, you can successfully perform measuring and inspection jobs yourself with only one X-ray scan. The standard acceptance test, the precision engineering
More informationIndustrial Metrology. Multisensor Measuring Machines O-INSPECT 322/442
Industrial Metrology Multisensor Measuring Machines O-INSPECT 322/442 The moment the answer is right in front of your eyes. This is the moment we work for. // Certainty Made By Zeiss 2 Table of Contents
More informationINSPECTION OF THE TURBINE BLADES USING SCANNING TECHNIQUES
INSPECTION OF THE TURBINE BLADES USING SCANNING TECHNIQUES H. Nieciag, M. Traczynski and Z. Chuchro Department of Geometrical Quanities Metrology The Institute of Metal Cutting, 30-011 Cracow, Poland Abstract:
More informationResearch on error detection technology of numerical control machine tool. Cao YongJie
Joint International Mechanical, Electronic and Information Technology Conference (JIMET 2015) Research on error detection technology of numerical control machine tool Cao YongJie Shanghai University of
More informationHigh-resolution X-ray CT Inspection of Honeycomb Composites Using Planar Computed Tomography Technology
2nd International Symposium on NDT in Aerospace 2010 - We.4.B.4 High-resolution X-ray CT Inspection of Honeycomb Composites Using Planar Computed Tomography Technology Tong LIU, Andrew A. MALCOLM, and
More information