Flying Focal Spot (FFS) in Cone-Beam CT Marc Kachelrieß, Member, IEEE, Michael Knaup, Christian Penßel, and Willi A. Kalender

Size: px
Start display at page:

Download "Flying Focal Spot (FFS) in Cone-Beam CT Marc Kachelrieß, Member, IEEE, Michael Knaup, Christian Penßel, and Willi A. Kalender"

Transcription

1 1238 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 3, JUNE 2006 Flying Focal Spot (FFS) in Cone-Beam CT Marc Kachelrieß, Member, IEEE, Michael Knaup, Christian Penßel, and Willi A. Kalender Abstract In the beginning of 2004 medical spiral CT scanners that acquire up to 64 slices simultaneously became available. Most manufacturers use a straightforward acquisition principle, namely an x ray focus rotating on a circular path and an opposing cylindrical detector whose rotational center coincides with the x-ray focus. The 64-slice scanner available to us, a Somatom Sensation 64 spiral cone-beam CT scanner (Siemens, Medical Solutions, Forchheim, Germany), makes use of a flying focal spot (FFS) that allows for view-by-view deflections of the focal spot in the rotation direction ( FFS) and in the -direction ( FFS) with the goal of reducing aliasing artifacts. The FFS feature doubles the sampling density in the radial direction (channel direction, FFS) and in the longitudinal direction (detector row direction or -direction, FFS). The cost of increased radial and azimuthal sampling is a two- or four-fold reduction of azimuthal sampling (angular sampling). To compensate for the potential reduction of azimuthal sampling the scanner simply increases the number of detector read-outs (readings) per rotation by a factor two or four. Then, up to four detector readings contribute to what we define as one view or one projection. A significant reduction of in-plane aliasing and of aliasing in the -direction can be expected. Especially the latter is of importance to spiral CT scans where aliasing is known to produce so-called windmill artifacts. We have derived and analyzed the optimal focal spot deflection values and as they would ideally occur in our scanner. Based upon these we show how image reconstruction can be performed in general. A simulation study showing reconstructions of mathematical phantoms further provides evidence that image quality can be significantly improved with the FFS. Aliasing artifacts, that manifest as streaks emerging from high-contrast objects, and windmill artifacts are reduced by almost an order of magnitude with the FFS compared to a simulation without FFS. Patient images acquired with our 64-slice cone-beam CT scanner support these results. Index Terms Computed tomography, cone-beam CT, image quality, image reconstruction, spiral-ct. I. INTRODUCTION CORRECT sampling requires to satisfy the Nyquist condition: at least two sample points should be taken per full width at half maximum (FWHM) of the detector point spread function (PSF). 1 In many cases this situation is not easy to achieve. In CT the spacing of the detector samples is slightly larger than the active width of the detector pixels far from sampling the active detector area twice. One workaround is the quarter Manuscript received November 9, 2004; revised March 11, The authors are with Institute of Medical Physics (IMP), University of Erlangen-Nürnberg, D Erlangen, Germany ( marc.kachelriess@imp.uni-erlangen.de). Digital Object Identifier /TNS Actually, the sample spacing must be 1=2b or less in order to allow to exactly recover a band-limited function whose support in frequency domain lies in [0b; b]. In real systems that rarely deal with band-limited but rather with essentially band-limited functions one can only recover approximations to the true function. In our case a good approximation to the sampling theorem is to acquire at least two samples per spatial resolution element. detector offset. If no special care is taken and the detector is aligned symmetrically with respect to the central ray a 360 scan will measure each ray twice with respect to its radial position (i.e. with its distance to the origin). In that case the data are redundant. Shifting the detector array (channel direction) by one quarter of the detector sampling distance pays out since opposing rays interlace and, by combining opposing views, one effectively doubles the sampling [1], [2]. However, for cone-beam scans and for spiral scans this kind of data redundancy is not really available since opposing rays do not exist; they rather differ by their tilt-angle with respect to the rotation axis and by their -position. Further, the quarter shift does not improve the sampling in the detector row direction (longitudinal or -direction). An alternative solution is deflecting the focal spot between adjacent detector read-outs (readings) as it is done in our tube (Fig. 1) [3]. This flying focal spot (FFS) can be used to double the sampling density in both directions regardless of the coneangle and the spiral trajectory [4]. Recently, a manufacturer-dependent evaluation of the image quality achieved with the flying focal spot was presented by the engineers [5]. The present work, that was carried out in early 2004 and submitted, is independent thereof insofar as all evaluations, implementations and simulations were performed without knowing manufacturer details. Of course, the actual scan data that we use to demonstrate the potential of the FFS was acquired with the scanner provided by the manufacturer. Aliasing artifacts due to inadequate longitudinal sampling manifest themselves as so-called windmill artifacts [6]. These are streaks emerging from high-contrast objects in a star-like pattern. The streaks are caused by interpolation errors in the longitudinal direction (between adjacent detector rows or slices): every time the interpolation partners change, say from the pair of slices and to the pair of slices and, a new windmill segment is introduced. Thereby, the number of streaks is a function of the number of detector rows that are simultaneously read out during the spiral CT scan, and a function of the spiral table increment per rotation. In contrast to alternative attempts to suppress these spiral artifacts, such as the use of preferred spiral pitch values, conjugate cone-beam backprojection algorithms, or iterative image reconstruction [7], [8], where image quality improvements are achieved only close to the center of rotation the flying focal spot technique realized in our scanner provides true high density sampling and is neither restricted to special pitch values nor to certain image regions. This paper provides an analysis of the flying focal spot geometry and of the achievable image quality. We specify and optimize the FFS deflection values, we clarify the geometry, we introduce the subfan approximation and we demonstrate the effect of the FFS in a simulation study and in phantom and patient measurements. Our focus lies on the removal of spiral windmill artifacts. Additionally, noise and spatial resolution are studied /$ IEEE

2 KACHELRIEß et al.: FLYING FOCAL SPOT (FFS) IN CONE-BEAM CT 1239 Fig. 1. A rotating envelope tube with focal spot deflection capability. Due to the anode angle the periodic motion of the electron beam, whose position is controlled by the magnetic field B, translates into a focal spot deflection 6@z in the z-direction and in a change 6@R of the focal spot s distance to the isocenter. II. SOURCE AND DETECTOR GEOMETRY To define the focal spot positions we avoid introducing a local tube coordinate system. We rather use the fact that the vertex positions can be described by small corrections in the view angle and in the -position of a circular or spiral scan. With being the distance of the undeflected focal spot to the isocenter, being the table increment per rotation,, and being the view angle let us define the spiral source trajectory as the scanner s rotation axis a focal spot deflection in the -direction will effectively change by where is the anode angle (in our case 7 ). The focal spot s radial motion is significant and this side-effect must be accounted for during image reconstruction. Note that the variation in due to is of second order in and therefore negligible. The detector is a cylindrical detector whose symmetry axis coincides with. We parameterize it as Circular scans are a special case of spiral scans and can be obtained by setting. The vector is the source trajectory vector of a spiral CT scanner without FFS. Note that the detector is centered about while the focal spot can be electromagnetically deflected to the positions. The new parameters and are the focal spot deflection angle and deflection length that are used to improve the in-plane and the axial sampling properties of CT. These deflection parameters are of very small magnitude; using the small angle approximation (first order Taylor series in the two deflection parameters) is justified and will be implicitly used below when deriving the ideal values for and. The remaining parameter is not independent (see Fig. 1). It accounts for small variations in the radius of the deflected focal spot to the isocenter. Since the rotation axis of the x-ray tube is parallel to where is the distance focal spot to detector, is the angle within the fan and is the detector s longitudinal coordinate. We also define the distance isocenter to detector such that holds. Note that the flying focal spot deflection parameters do not enter the definition of the detector coordinates. To treat sampling we must go into discretization. We assume equidistant sampling in all three coordinates and let, and denote the sample spacing. The angle is the angular increment of the gantry between two read-outs. The angle is the transaxial angle between two detector elements as seen from ; the actual distance between two detector elements (in the transaxial direction) is then given by. The sampling distance is the longitudinal distance between two detector rows. It is related to the nominal slice thickness by scaling to the isocenter:. The samples are located at for the gantry angle, at for the angle within the fan and at for the longitudinal detector coordinate. Thereby and denote the starting values and and count the readings, channels and detector rows, respectively.

3 1240 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 3, JUNE 2006 Fig. 2. Geometry of the FFS deflection (gantry rotation not shown). Physically, the detectors are spaced by R 1 ^ which makes R 1 ^ at the isocenter. To halve this sampling distance the focal spot deflection between two adjacent detector read-outs must be chosen such that it corresponds to (1)=(2)R 1 ^ at the isocenter. III. FLYING FOCAL SPOT DEFLECTION Now, the optimal focal spot deflection parameters are derived. To do so, we note that both effects can be separated and consider the FFS first, and then the FFS. Also note that azimuthal sampling (number of acquired angular positions during one rotation) itself is not an issue since it can be, and actually is, kept constant by simply increasing the number of detector read-outs per rotation by a factor of 2 if either the FFS or the FFS are used or by a factor of 4 if both FFS are active. A. In the fan or channel direction the CT detector elements are arranged on a circle of radius centered at. Their angular spacing yields a physical distance of. In the isocenter this corresponds to a sampling distance of. The aim of the FFS is to reduce this sampling distance by a factor of two. This is achieved by deflecting the focal spot with a deflection parameter of alternating sign such that between two focal spot positions yields 0.85 mm approximately. Be aware of the fact that the FFS improves the radial sampling which means that only the distance of the rays to the origin has to be considered and not their angle (which would be the azimuthal sampling). This is the reason why Fig. 2 does not show the angular increment of the detector between the two readings acquired. B. FFS The longitudinal sampling distance ( -direction) is given by ; scaled to the isocenter this gives, the value is also called the nominal slice thickness (although it rather corresponds to the longitudinal sampling distance in multi-slice CT). The FFS aims at doubling the longitudinal sampling density. This is achieved by deflecting the focal spot with a deflection parameter of alternating sign such that is the focal spot position of reading. The situation is illustrated in Fig. 2. The alternating sign yields a total deflection difference of which corresponds to a spacing of of the two focal spot positions. After scaling to the isocenter this becomes and should equal (half the sampling distance of the original fan). This is the case when one chooses is the focal spot position of reading. The situation is illustrated in Fig. 3. The rays shall be longitudinally separated by when intersecting the isocenter. Neglecting the table increment per reading which is about three orders of magnitude smaller than we find by simple geometric scaling that should be just the same distance which means The numerical value given applies to our scanner where mm, mm. The distance The numerical value applies to our scanner where mm, mm, mm. Obviously, both focal spot positions are separated by about 0.66 mm in the -direction. Due

4 KACHELRIEß et al.: FLYING FOCAL SPOT (FFS) IN CONE-BEAM CT 1241 Fig. 3. Geometry of the zffs deflection. to the anode angle we get mm which means that the radial separation of both focal spots is 5.4 mm. C. Both FFSs If both the FFS and the FFS are switched on simultaneously the deflection values and remain the same as derived in the previous sections. However, the focal spot itself will jump in a rectangular fashion relative to the anode The symbols indicate that the sign of the deflection values depends on. The order in which the focal spot positions are visited depends on the implementation; for our scanner applies. D. View Definition The question arises what to define as a projection (or view). Up to here, we have learned to know the readings which comprise a set of simultaneous detector read-outs. Without FFS one reading makes up what is usually called a projection. With the FFS switched on up to four readings collect new sample positions. Therefore we define the view or projection to be the collection of adjacent readings that make up a total FFS cycle. Thus, the view consists of two readings if either the FFS or the FFS is switched on and it consists of four readings if both flying focal spots are used. IV. RECONSTRUCTION Three types of image reconstruction algorithms can be found in today s 16- and 64-slice clinical CT scanners: -interpolation, single-slice rebinning and Feldkamp-type image reconstruction. To reconstruct stationary objects (standard reconstruction) manufacturers use Feldkamp-based algorithms [9] [14] or they use so-called single-slice rebinning algorithms such as the advanced single-slice rebinning algorithm (ASSR) [15] [20]. The algorithms can be grouped into two categories. Either they are based on some fan-beam filtered backprojection or they perform a rebinning of the data to fan-parallel or parallel geometry. According to this classification image reconstruction from FFS data can either use the subfan approximation or rebinning. A. Subfan Approximation To be able to perform fan-beam filtered backprojection it is necessary to bring the data into a format that allows for shift-invariant filtering; only a certain class of geometries, including the fan-beam, allows to do so [21]. In this paper s context this means that the complicated geometry that results from the focal spot deflection must be approximated by true fan beams where the focal spot lies in the center of the detector arc. 1) -Subfan Approximation: Since the -focal spot deflection is only a tiny angle one can approximate each of the two deflected subfans by an ideal subfan whose focal spot coincides with the rotational center of the detector cylinder. Thus, we try to replace the deflected system by the ideal fan-beam system-that is a system where the focal spot is fixed and centered relative to the detector arc Regarding Fig. 2 and using the small angle approximation we find that yields the desired compensation, and

5 1242 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 3, JUNE 2006 Fig. 4. The penumbra cannot be used for imaging. Therefore the pre-patient collimators must open wide enough. Two dose profiles offset by a distance O result when using the zffs. The offset O and the penumbra width P are geometrically related to R 1, to@z, to the focal spot size F, and to the collimator distance D. results. If the FFS is not used implies. Note that the minus sign is introduced since a clockwise deflection of the focal spot can be approximately compensated by a counterclockwise deflection of the detector channels. 2) -Subfan Approximation: For the FFS a similar approximation can be established. Instead of assuming a shifted focal spot during reconstruction one may rather assume a shifted -axis, i.e. a detector array that is shifted by. The deflected system shall be replaced with the ideal system Regarding Fig. 3 we find that desired compensation and yields the results. If the FFS is not used implies. The minus sign shows that a forward deflection of the focal spot can be approximately compensated by a backward deflection of the detector rows. B. Rebinning An alternative to using the subfan approximation is rebinning. This means that the data are resampled to correspond to some more convenient geometry. During the resampling step one can exactly account for the geometry of each individual ray and no approximation is required. Note that rebinning always involves some kind of interpolation along the ray coordinates, and, however. In many cases rebinning to parallel beam geometry is performed, including the Feldkamp-type extended parallel backprojection (EPBP, see [22]) that is used in this paper. It can be split into azimuthal, longitudinal and radial rebinning. Azimuthal rebinning means resorting the rays into non-equidistant parallel geometry (fan-parallel geometry) by performing a resampling in the -direction. Radial rebinning is the step to convert the fan-parallel data to parallel data; the latter are characterized by equidistant parallel rays. During azimuthal and radial rebinning one can exactly account for the FFS geometry. Many popular reconstruction algorithms, such as spiral -interpolation (single-slice CT), spiral -filtering (up to 4-slice CT), ASSR-type algorithms and EPBP (cone-beam spiral CT) perform a resampling of the data where only the longitudinal coordinate is involved. Since this means taking into account the focal spot s -position longitudinal rebinning is the step to account for. V. DOSE ISSUES The FFS requires to open the pre-patient collimator a bit wider than it is necessary without a flying focal spot: the collimator s mechanical inertia prohibits to follow the movement of the flying focal spot with such high frequency (in the order of 10 khz). Basic geometric considerations help to get an idea of the magnitude of this additionally required collimator opening that directly converts to a patient dose increase. A schematic illustration is given in Fig. 4. It basically shows the trapezoidal dose profile at the axis of rotation. The area of the dose profile that can be used for imaging is the collimated slices area. Other parts of the dose profile, such as the penumbra and the offset, are relevant for the patient dose but irrelevant for imaging.

6 KACHELRIEß et al.: FLYING FOCAL SPOT (FFS) IN CONE-BEAM CT 1243 Fig. 5. Point spread functions and slice sensitivity profiles obtained from the reconstructions of the simulated delta peak. Without zffs one obtains the expected broadening of the SSP due to linear interpolation between adjacent detector rows. The zffs improves sampling and the effective slice width (FWHM) equals the nominal slice width. The same behavior is observed in-plane where the radial point spread function is plotted. (a) Slice sensitivity profiles SSP(z) with and without zffs. (b) Point spread functions PSF(r) with and without FFS. Thus, the patient dose is proportional to which is a factor of compared to an ideal scanner. For scanners without FFS there is only one trapezoidal dose profile and the offset would be zero which means a dose of. Evidently, the dose ratio of scans with FFS and those without is given by. Assuming a focal spot size mm and a collimator distance mm and our scanner s mm collimation ( mm) this evaluates to about This implies a dose increase of 6% with the FFS. VI. SIMULATIONS For our simulation we use the Sensation 64 geometry with a slice thickness of mm. The number of readings per rotation is set to, the number of detector channels to and the number of detector rows to. Further, mm and mm. The resulting fan angle and cone angle yield and, respectively. A spiral pitch of, that corresponds to a table increment of 25 mm per rotation, was simulated. To mimic a finite detector area, focal spot area and integration time we used a 3 3 subsampling for the detector and the focal spot and a three-fold subsampling for the view angle. Altogether 243 needle beams were computed and averaged to obtain one measured value. A monochromatic source was assumed to avoid beam hardening effects. The simulation software used was ImpactSim (

7 1244 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 3, JUNE 2006 Fig. 6. With double z-sampling only half of the quanta contribute to each detector reading when a constant total dose is assumed. Consequently, image noise (standard deviation) will increase by p 2. TABLE I FFS COMBINATIONS To assess spatial resolution a delta object was simulated and its response was evaluated by computing radial and longitudinal profiles. To study the artifact behavior the FORBILD head phantom ( was simulated. The FFS combinations, among which only the last two are available at our scanner were simulated, are shown in shown in Table1. Here is the number of projections per rotation, which is the number of independent azimuthal samples per rotation. Each projection is made up of 1, 2, or 4 of the readings depending on whether no FFS, either FFS, or both FFS are used. denotes the number of independent radial samples per projection and it is equal to the number of channels when the FFS is inactive or it equals with the FFS switched on. Similarly, holds the number of independent longitudinal samples and is also known as the number of simultaneously acquired slices. equals the number of active detector rows with the FFS being switched off and whenever the FFS is used. Image reconstruction of simulated and measured data was performed with the Feldkamp-type extended parallel backprojection (EPBP) algorithm [22]. VII. RESULTS Fig. 5 shows the axial and longitudinal profiles through the delta object. Apparently, the double - and the double -sampling improve spatial resolution by about 20 to 30% each. It should be noted that these improvements have been achieved without modifying the kind of interpolation used: in all cases standard linear interpolation was applied. However, implicitly the FWHM of the triangle function that represents linear interpolation shrunk from 0.6 mm (standard sampling) to 0.3 mm (double sampling) when switching from standard to FFS reconstruction. We also evaluated image noise and found a two-fold increase of noise variance with increasing spatial resolution. Quantitatively, the increased noise corresponds to the theoretically expected cost for the increased resolution. This is illustrated in Fig. 6. It compares a standard scanner with single -sampling to a FFS scanner with double -sampling. At constant patient dose (constant number of quanta) each detector receives only half of the number of quanta when using double sampling compared to the standard scanner. Since the longitudinal interpolation algorithm is a linear interpolation between adjacent samples in both cases one finds the said two-fold increase in noise variance or a increase in the noise standard deviation. The figure further shows a similar consideration for a hypothetical high-resolution scanner where the same 0.3 mm sampling has been achieved by halving the detector aperture and spacing. This scanner could yield a higher spatial resolution than the FFS scanner while keeping the same image noise level. On the other hand one could use this scanner together with some smoother -interpolation algorithm to obtain the same SSP as with the FFS scanner. In that case image noise would be significantly lower which implies a better dose usage with the high-resolution scanner. Note that the same point spread function, which implies the same spatial resolution, and the same image noise as with FFS

8 KACHELRIEß et al.: FLYING FOCAL SPOT (FFS) IN CONE-BEAM CT 1245 can be achieved from a standard scan: just replace the linear in by a conterpolation which is a convolution with, where denotes a volution with a slim triangle triangle function of maximum and full width at half maximum 1. What is the FFS for, if we can obtain the same PSF or SSP from a standard scan with a slim interpolation function? It is the sampling which makes the difference. A slim -filter applied to a standard scan would amplify the windmill artifacts whereas using double -sampling significantly attenuates these artifacts. The main reason why the manufacturers introduced double sampling techniques was the expected reduction of aliasing artifacts. This reduction is clearly visible in the reconstructions of our simulated head phantom that are shown in Fig. 7. Streaks emerging from high contrast objects are strongly suppressed by the Nyquist-conform double sampling. The figure also shows that in-plane aliasing is not influenced by the FFS. Windmill artifacts are present in real patient data, too. They become especially annoying when slicing through a complete volume since the windmill-type streaks that emerge from sharp edges seem to rotate about their origin when switching from one image to the next (hence the name windmill). However, the windmill artifacts without FFS and their suppression with FFS becomes less impressive in static images such as the slice presented in Fig. 8. Nevertheless, we can clearly see low amplitude and low frequency streaks emerging from the high contrast bone. Sliding through adjacent images would show these streaks rotating around their origin with increasing or decreasing -position. An electronic version of this comparison with the possibility to zoom, pan, slide and window is provided at uni-erlangen.de/zffs. The Sensation 64 scanner does not allow to switch off the double -sampling feature. Therefore our reconstruction software was modified to simply leave out every other reading to mimic a standard scan and we did the full reconstruction including all FFS position to obtain FFS reconstructions. The alert reader might thus have observed that, due to our manipulation, image noise is the same with and without FFS in Fig. 8. instead of showing the expected factor VIII. SUMMARY AND DISCUSSION We have given a theoretical treatment of the focal spot deflection technique of our cone-beam spiral CT scanner. The flying focal spot allows to double the sampling in the axial and in the longitudinal direction. The corresponding equations of optimized focal spot placement have been derived. For reconstruction based on rebinning to parallel data this is the case for the ASSR and for the EPBP cone beam algorithms that we prefer to use the exact equations can be conveniently used. If fan-beam filtered backprojection techniques shall be used for image reconstruction subfan approximation is the method of choice. It has been shown that dose due to penumbra effects may insignificantly increase when using the double -sampling. Since our scanner provides only the mm collimation where mm scan mode without the FFS is switched on but no FFS we cannot provide dose measurements that support this result. Our results mainly concentrate upon the new FFS feature since the FFS is available in Somatom scanners for more than Fig. 7. Reconstructions of the FORBILD head phantom. Aliasing artifacts apparent on the left side are greatly reduced with the z FFS. An exception is the second row of images that also shows significant in-plane aliasing due to the periodic high-contrast bubbles in the phantom s inner ear. These in-plane aliasing artifacts are not influenced by the double z -sampling. The increase in noise by a factor of 2 in the right column is due to the higher z -sharpness of the z FFS-reconstruction (50=90). p one decade. Basically, the results show that both, the SSP and the in-plane PSF benefit from the improved sampling assuming that the interpolation between adjacent samples, linear interpolation in our case, is scaled with the sample spacing. Then, when switching from single noise increases by a factor of

9 1246 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 3, JUNE 2006 The most significant improvement obtained with the FFS is the removal of windmill artifacts that have been present for years in multi-slice spiral CT. These streaks that emerge from high-density objects and appear to be rotating when scrolling through the volume (see are due to insufficient longitudinal sampling. The reduction of the windmill artifacts might have strong clinical impact since it now allows to go down to the finest effective slice thicknesses without impairing image quality. REFERENCES Fig. 8. Head patient, spiral acquisition with p = 1 and :6 mm collimation. A significant reduction of windmill artifacts present in the standard scanner (left) is achieved when utilizing a zffs (right). The fact that noise is equivalent in both images is due to skipping every other reading for the standard reconstruction. Thereby the images in the left column received only 50% of the quanta that contributed to those in the right column. For an animated version see (0/200) to double sampling at constant patient dose since each channel receives only half of the quanta with double sampling. However, this noise increase is correctly balanced by the spatial resolution gain. We also indicated that a hypothetical high resolution scanner that achieves double sampling by using more but smaller detector elements would be able to achieve the same resolution with significantly less noise than possible with the FFS approach. This finding is a special case of the general observation that a given PSF or SSP can be best achieved with detectors far smaller than absolutely necessary [23], [24]. [1] P. M. Joseph, Artifacts in computed tomography, in Radiology of the skull and brain-technical aspects of computed tomography. St. Louis, MO: Mosby, 1981, pp [2] W. A. Kalender, Computed Tomography, 2nd ed. New York: Wiley, 2005, ISBN [3] P. Schardt, J. Deuringer, J. Freudenberger, E. Hell, W. Kniipfer, D. Mattern, and M. Schild, New x-ray tube performance in computed tomography by introducing the rotating envelope tube technology, Med. Phys., vol. 31, no. 9, pp , [4] M. Kachelrieß, M. Knaup, C. Penßel, and W. A. Kalender, Flying focal spot (FFS) in cone-beam CT, in IEEE Medical Imaging Conf. Record, Apr. 2005, pp [5] T. Flohr, K. Stierstorfer, S. Ulzheimer, H. Bruder, A. Primak, and C. McCollough, Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot, Med. Phys., vol. 32, pp , Aug [6] M. Silver, K. Taguchi, I. Hein, B. Ciang, M. Kazama, and I. Mori, Windmill artifact in multi-slice helical CT, in SPIE Medical Imaging Proc., 2003, pp [7] J. Hsieh, X. Tang, J.-B. Thibault, C. Shaughnessy, and R. A. Nilsen, Conjugate cone beam backprojection for high z-resolution reconstruction, Proc. 8th Int. Meeting on Fully 3D Image Reconstruction, pp , Jul [Online]. Available: [8] J.-B. Thibault, K. Sauer, C. Bouman, and J. Hsieh, Three-dimensional statistical modeling for image quality improvements in multi-slice helical CT, in Proc. 8th Int. Meeting on Fully 3D Image Reconstruction, July 2005, pp [Online]. Available: [9] L. Feldkamp, L. Davis, and J. Kress, Practical cone-beam algorithm, J. Opt. Soc. Amer., vol. 1, pp , Jun [10] X.-H. Yan and R. Leahy, Derivation and analysis of a filtered backprojection algorithm for cone beam projection data, IEEE Trans. Med. Imaging, vol. 10, pp , Sep [11] X. Yan and R. Leahy, Cone beam tomography with circular, elliptical and spiral orbits, Phys. Med. Biol., vol. 37, pp , Mar [12] G. Wang, T.-H. Lin, P. Cheng, and D. Shinozaki, A general conebeam reconstruction algorithm, IEEE Trans. Med. Imaging, vol. 12, pp , Sep [13] T. Köhler, C. Bontus, M. Grass, and J. Timmer, Artifact analysis of approximate helical cone-beam CT reconstruction algorithms, Med. Phys., vol. 29, pp , Jan [14] H. Turbell, Cone-Beam Reconstruction Using Filtered Backprojection, Ph.D. dissertation (No. 672), Linkoping Studies in Science and Technology, Linkoping Universitet, Linkoping, Germany, 2001, ISBN [15] M. Kachelrieß, S. Schaller, and W. A. Kalender, Advanced single-slice rebinning in cone-beam spiral CT, Med. Phys., vol. 27, pp , Apr [16] H. Bruder, M. Kachelrieß, S. Schaller, K. Stierstorfer, and T. Flohr, Single-slice rebinning reconstruction in spiral cone-beam computed tomography, IEEE Trans. Med. Imaging, vol. 19, pp , Sep [17] M. Kachelrieß, T. Fuchs, S. Schaller, and W. A. Kalender, Advanced single-slice rebinning for tilted spiral cone-beam CT, Med. Phys., vol. 28, pp , Jun [18] S. Schaller, K. Stierstorfer, H. Bruder, M. Kachelrieß, and T. Flohr, Novel approximate approach for high-quality image reconstruction in helical cone beam CT at arbitrary pitch, SPIE Med. Imaging Conf. Proc., vol. 4322, pp , [19] K. Stierstorfer, T. Flohr, and H. Bruder, Segmented multiple plane reconstruction-a novel approximate reconstruction for multi-slice spiral CT, Phys. Med. Biol., vol. 47, pp , 2002.

10 KACHELRIEß et al.: FLYING FOCAL SPOT (FFS) IN CONE-BEAM CT 1247 [20] L. Chen, Y. Liang, and D. Heuscher, General surface reconstruction for cone-beam multislice spiral computed tomography, Med. Phys., vol. 30, pp , Oct [21] G. Besson, CT fan-beam parameterizations leading to shift-invariant filtering, Inverse Problems, vol. 12, pp , [22] M. Kachelrieß, M. Knaup, and W. A. Kalender, Extended parallel backprojection for standard 3D and phase-correlated 4D axial and spiral cone-beam CT with arbitrary pitch and 100% dose usage, Med. Phys., vol. 31, pp , Jun [23] M. Kachelrieß and W. A. Kalender, Presampling, algorithm factors and noise: Considerations for CT in particular and for medical imaging in general, Med. Phys., vol. 32, pp , May [24] M. Kachelrieß and W. A. Kalender, Optimizing detector size in x-ray imaging, IEEE Medical Imaging Conf. Rec., pp , Oct

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Improvement of Efficiency and Flexibility in Multi-slice Helical CT J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): 408 412 DOI: 10.1007/s12204-008-0408-x Improvement of Efficiency and Flexibility in Multi-slice Helical CT SUN Wen-wu 1 ( ), CHEN Si-ping 2 ( ), ZHUANG Tian-ge

More information

An approximate cone beam reconstruction algorithm for gantry-tilted CT

An approximate cone beam reconstruction algorithm for gantry-tilted CT An approximate cone beam reconstruction algorithm for gantry-tilted CT Ming Yan a, Cishen Zhang ab, Hongzhu Liang a a School of Electrical & Electronic Engineering, Nanyang Technological University, Singapore;

More information

Advanced single slice rebinning in cone beam spiral CT: Theoretical considerations and medical applications

Advanced single slice rebinning in cone beam spiral CT: Theoretical considerations and medical applications Advanced single slice rebinning in cone beam spiral CT: Theoretical considerations and medical applications Marc Kachelrieß a, Stefan Schaller b and Willi A. Kalender a a Institute of Medical Physics,

More information

Artifact analysis of approximate helical cone-beam CT reconstruction algorithms

Artifact analysis of approximate helical cone-beam CT reconstruction algorithms Artifact analysis of approximate helical cone-beam CT reconstruction algorithms Th. Köhler, a) R. Proksa, C. Bontus, and M. Grass Philips Research Laboratories, Sector Technical Systems, Roentgenstrasse

More information

DUE to beam polychromacity in CT and the energy dependence

DUE to beam polychromacity in CT and the energy dependence 1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact

More information

Key words: Cone-beam CT CBCT, cardiac imaging, 4-D reconstruction, image quality

Key words: Cone-beam CT CBCT, cardiac imaging, 4-D reconstruction, image quality Extended parallel backprojection for standard three-dimensional and phase-correlated four-dimensional axial and spiral cone-beam CT with arbitrary pitch, arbitrary cone-angle, and 100% dose usage Marc

More information

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA Computed Tomography Principles, Design, Artifacts, and Recent Advances THIRD EDITION Jiang Hsieh SPIE PRESS Bellingham, Washington USA Table of Contents Preface Nomenclature and Abbreviations xi xv 1 Introduction

More information

Advanced single-slice rebinning for tilted spiral cone-beam CT

Advanced single-slice rebinning for tilted spiral cone-beam CT Advanced single-slice rebinning for tilted spiral cone-beam CT Marc Kachelrieß a) and Theo Fuchs Institute of Medical Physics, University of Erlangen Nürnberg, Germany Stefan Schaller Siemens AG, Medical

More information

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004 Spiral ASSR Std p = 1.0 Spiral EPBP Std p = 1.0 Kachelrieß et al., Med. Phys. 31(6): 1623-1641, 2004 256 slices (0/300) Advantages of Cone-Beam Spiral CT Image quality nearly independent of pitch Increase

More information

Scaling Calibration in the ATRACT Algorithm

Scaling Calibration in the ATRACT Algorithm Scaling Calibration in the ATRACT Algorithm Yan Xia 1, Andreas Maier 1, Frank Dennerlein 2, Hannes G. Hofmann 1, Joachim Hornegger 1,3 1 Pattern Recognition Lab (LME), Friedrich-Alexander-University Erlangen-Nuremberg,

More information

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Benedikt Lorch 1, Martin Berger 1,2, Joachim Hornegger 1,2, Andreas Maier 1,2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

Non-Stationary CT Image Noise Spectrum Analysis

Non-Stationary CT Image Noise Spectrum Analysis Non-Stationary CT Image Noise Spectrum Analysis Michael Balda, Björn J. Heismann,, Joachim Hornegger Pattern Recognition Lab, Friedrich-Alexander-Universität Erlangen Siemens Healthcare, Erlangen michael.balda@informatik.uni-erlangen.de

More information

Feldkamp-type image reconstruction from equiangular data

Feldkamp-type image reconstruction from equiangular data Journal of X-Ray Science and Technology 9 (2001) 113 120 113 IOS Press Feldkamp-type image reconstruction from equiangular data Ben Wang a, Hong Liu b, Shiying Zhao c and Ge Wang d a Department of Elec.

More information

USING cone-beam geometry with pinhole collimation,

USING cone-beam geometry with pinhole collimation, IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 687 A Backprojection-Based Parameter Estimation Technique for Skew-Slit Collimation Jacob A. Piatt, Student Member, IEEE, and Gengsheng L.

More information

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh et al. : Proceedings of the National Seminar & Exhibition on Non-Destructive Evaluation DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh V Acharya, Umesh Kumar, Gursharan

More information

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Marc Kachelrieß, a Katia Sourbelle, and Willi A. Kalender Institute of Medical Physics, University of

More information

The Near Future in Cardiac CT Image Reconstruction

The Near Future in Cardiac CT Image Reconstruction SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de

More information

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling nd International Symposium on NDT in Aerospace - We..A. Comparison of Probing Error in Dimensional Measurement by Means of D Computed Tomography with Circular and Helical Sampling Jochen HILLER, Stefan

More information

TESTING OF THE CIRCLE AND LINE ALGORITHM IN THE SETTING OF MICRO-CT

TESTING OF THE CIRCLE AND LINE ALGORITHM IN THE SETTING OF MICRO-CT SCA2016-080 1/7 TESTING OF THE CIRCLE AND LINE ALGORITHM IN THE SETTING OF MICRO-CT Alexander Katsevich 1, 2 and Michael Frenkel 1 1 itomography Corp., 2 University of Central Florida (UCF) This paper

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation

Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation Bastian Bier 1, Chris Schwemmer 1,2, Andreas Maier 1,3, Hannes G. Hofmann 1, Yan Xia 1, Joachim Hornegger 1,2, Tobias Struffert

More information

A study on the section sensitivity profile in multi-row-detector spiral CT

A study on the section sensitivity profile in multi-row-detector spiral CT Journal of X-Ray Science and Technology (3) IOS Press A study on the section sensitivity profile in multi-row-detector spiral CT Ge Wang a, Mark Madsen a, Katherine Redford a, Shiying Zhao b and Michael

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Spiral-CT. Benjamin Keck. 21. March 2006

Spiral-CT. Benjamin Keck. 21. March 2006 Spiral-CT Benjamin Keck 21. March 2006 1 Motivation Spiral-CT offers reconstruction of long objects compared to circular filtered backprojection, where reconstruction is limited in z-direction. While the

More information

An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography

An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography 4th International Symposium on NDT in Aerospace 2012 - Tu.3.A.3 An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography Jonathan HESS *, Patrick KUEHNLEIN *, Steven OECKL

More information

Introduction to Medical Imaging. Cone-Beam CT. Klaus Mueller. Computer Science Department Stony Brook University

Introduction to Medical Imaging. Cone-Beam CT. Klaus Mueller. Computer Science Department Stony Brook University Introduction to Medical Imaging Cone-Beam CT Klaus Mueller Computer Science Department Stony Brook University Introduction Available cone-beam reconstruction methods: exact approximate algebraic Our discussion:

More information

The n-pi-method for Helical Cone-Beam CT

The n-pi-method for Helical Cone-Beam CT 848 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 19, NO. 9, SEPTEMBER 2000 The n-pi-method for Helical Cone-Beam CT R. Proksa*, Th. Köhler, M. Grass, and J. Timmer Abstract A new class of acquisition schemes

More information

Motion Compensation from Short-Scan Data in Cardiac CT

Motion Compensation from Short-Scan Data in Cardiac CT Motion Compensation from Short-Scan Data in Cardiac CT Juliane Hahn 1,2, Thomas Allmendinger 1, Herbert Bruder 1, and Marc Kachelrieß 2 1 Siemens Healthcare GmbH, Forchheim, Germany 2 German Cancer Research

More information

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings)

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings) Marquette University e-publications@marquette Biomedical Engineering Faculty Research and Publications Engineering, College of 1-1-1999 An Iterative Approach to the Beam Hardening Correction in Cone Beam

More information

Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset

Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset Andreas Maier 1, Patrick Kugler 2, Günter Lauritsch 2, Joachim Hornegger 1 1 Pattern Recognition Lab and SAOT Erlangen,

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Computer-Tomography II: Image reconstruction and applications

Computer-Tomography II: Image reconstruction and applications Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de

More information

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE / Radiographic Tomosynthesis: Image Quality and Artifacts Reduction Baojun Li, Ph.D Department of Radiology Boston University Medical Center 2012 AAPM Annual Meeting Background Linear Trajectory Tomosynthesis

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

FDK Half-Scan with a Heuristic Weighting Scheme on a Flat Panel Detector-Based Cone Beam CT (FDKHSCW)

FDK Half-Scan with a Heuristic Weighting Scheme on a Flat Panel Detector-Based Cone Beam CT (FDKHSCW) Biomedical Imaging Volume 26, Article ID 83983, Pages 8 DOI.55/IJBI/26/83983 FDK Half-Scan with a Heuristic Weighting Scheme on a Flat Panel Detector-Based Cone Beam CT (FDKHSCW) Dong Yang and Ruola Ning

More information

Suitability of a new alignment correction method for industrial CT

Suitability of a new alignment correction method for industrial CT Suitability of a new alignment correction method for industrial CT Matthias Elter 1, Nicole Maass 1, Peter Koch 2 1 Siemens AG, Healthcare Sector, Erlangen, Germany, e-mail: matthias.elter@siemens.com,

More information

CIRCULAR scanning trajectory has been widely used in

CIRCULAR scanning trajectory has been widely used in IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 25, NO. 7, JULY 2006 869 Region of Interest Reconstruction From Truncated Data in Circular Cone-Beam CT Lifeng Yu, Yu Zou, Emil Y. Sidky, Charles A. Pelizzari,

More information

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam Home Search Collections Journals About Contact us My IOPscience Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam projections This content has been downloaded from IOPscience.

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

RECENTLY, biomedical imaging applications of

RECENTLY, biomedical imaging applications of 1190 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 24, NO. 9, SEPTEMBER 2005 A General Exact Reconstruction for Cone-Beam CT via Backprojection-Filtration Yangbo Ye*, Shiying Zhao, Hengyong Yu, Ge Wang, Fellow,

More information

THE FAN-BEAM scan for rapid data acquisition has

THE FAN-BEAM scan for rapid data acquisition has 190 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 26, NO. 2, FEBRUARY 2007 Hilbert Transform Based FBP Algorithm for Fan-Beam CT Full Partial Scans Jiangsheng You*, Member, IEEE, Gengsheng L. Zeng, Senior

More information

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

C a t p h a n / T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice

More information

Accelerated C-arm Reconstruction by Out-of-Projection Prediction

Accelerated C-arm Reconstruction by Out-of-Projection Prediction Accelerated C-arm Reconstruction by Out-of-Projection Prediction Hannes G. Hofmann, Benjamin Keck, Joachim Hornegger Pattern Recognition Lab, University Erlangen-Nuremberg hannes.hofmann@informatik.uni-erlangen.de

More information

ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart

ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart Marc Kachelrieß, a) Stefan Ulzheimer, and Willi A. Kalender University of Erlangen-Nürnberg, Institute of Medical

More information

Scatter Correction Methods in Dimensional CT

Scatter Correction Methods in Dimensional CT Scatter Correction Methods in Dimensional CT Matthias Baer 1,2, Michael Hammer 3, Michael Knaup 1, Ingomar Schmidt 3, Ralf Christoph 3, Marc Kachelrieß 2 1 Institute of Medical Physics, Friedrich-Alexander-University

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 3, JUNE 2003 405 Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging Edward J. Soares, Kevin W. Germino,

More information

Cardiac Dual Energy CT: Technique

Cardiac Dual Energy CT: Technique RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de

More information

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy T. Rohlfing a, D. Zerfowski b, J. Beier a, P. Wust a, N. Hosten a, R. Felix a a Department of

More information

Translational Computed Tomography: A New Data Acquisition Scheme

Translational Computed Tomography: A New Data Acquisition Scheme 2nd International Symposium on NDT in Aerospace 2010 - We.1.A.3 Translational Computed Tomography: A New Data Acquisition Scheme Theobald FUCHS 1, Tobias SCHÖN 2, Randolf HANKE 3 1 Fraunhofer Development

More information

DUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography

DUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography Marc Kachelrieß, Member, IEEE, Timo Berkus, Philip Stenner, Willi A. Kalender Abstract Material selective imaging using dual energy

More information

F3-D: Computationally Efficient Simultaneous Segmentation and Image Reconstruction for CT X-ray Based EDS Systems Screening

F3-D: Computationally Efficient Simultaneous Segmentation and Image Reconstruction for CT X-ray Based EDS Systems Screening F3-D: Computationally Efficient Simultaneous Segmentation and Image Reconstruction for CT X-ray Based EDS Systems Screening Abstract Despite recent technological advances, reliable detection of explosives

More information

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Beam Attenuation Grid Based Scatter Correction Algorithm for More Info at Open Access Database

More information

ECG-Correlated Imaging of the Heart with Subsecond Multislice Spiral CT

ECG-Correlated Imaging of the Heart with Subsecond Multislice Spiral CT 888 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 19, NO. 9, SEPTEMBER 2000 ECG-Correlated Imaging of the Heart with Subsecond Multislice Spiral CT Marc Kachelrieß*, Stefan Ulzheimer, and Willi A. Kalender

More information

J. Ling Chinese Academy of Science Institute for Ceramics Shanghai, China

J. Ling Chinese Academy of Science Institute for Ceramics Shanghai, China DEVELOPMENT AND APPLICATION OF LOCAL 3-D CT RECONSTRUCTION SOFTWARE FOR IMAGING CRITICAL REGIONS IN LARGE CERAMIC TURBINE ROTORS E. A. Sivers, D. L. Holloway, and W. A. Ellingson Argonne National Laboratory

More information

An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy

An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy Chenyang Xu 1, Siemens Corporate Research, Inc., Princeton, NJ, USA Xiaolei Huang,

More information

Separate CT-Reconstruction for Orientation and Position Adaptive Wavelet Denoising

Separate CT-Reconstruction for Orientation and Position Adaptive Wavelet Denoising Separate CT-Reconstruction for Orientation and Position Adaptive Wavelet Denoising Anja Borsdorf 1,, Rainer Raupach, Joachim Hornegger 1 1 Chair for Pattern Recognition, Friedrich-Alexander-University

More information

A sufficient condition for spiral cone beam long object imaging via backprojection

A sufficient condition for spiral cone beam long object imaging via backprojection A sfficient condition for spiral cone beam long object imaging via backprojection K. C. Tam Siemens Corporate Research, Inc., Princeton, NJ, USA Abstract The response of a point object in cone beam spiral

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study

Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study B. De Man, J. Nuyts, P. Dupont, G. Marchal and P. Suetens Medical Image Computing, ESAT-PSI, K.U.Leuven, B-3000 Leuven, Belgium Department

More information

Recognition and Measurement of Small Defects in ICT Testing

Recognition and Measurement of Small Defects in ICT Testing 19 th World Conference on Non-Destructive Testing 2016 Recognition and Measurement of Small Defects in ICT Testing Guo ZHIMIN, Ni PEIJUN, Zhang WEIGUO, Qi ZICHENG Inner Mongolia Metallic Materials Research

More information

Central Slice Theorem

Central Slice Theorem Central Slice Theorem Incident X-rays y f(x,y) R x r x Detected p(, x ) The thick line is described by xcos +ysin =R Properties of Fourier Transform F [ f ( x a)] F [ f ( x)] e j 2 a Spatial Domain Spatial

More information

Index. aliasing artifacts and noise in CT images, 200 measurement of projection data, nondiffracting

Index. aliasing artifacts and noise in CT images, 200 measurement of projection data, nondiffracting Index Algebraic equations solution by Kaczmarz method, 278 Algebraic reconstruction techniques, 283-84 sequential, 289, 293 simultaneous, 285-92 Algebraic techniques reconstruction algorithms, 275-96 Algorithms

More information

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction 2D Fan Beam Reconstruction 3D Cone Beam Reconstruction Mario Koerner March 17, 2006 1 2D Fan Beam Reconstruction Two-dimensional objects can be reconstructed from projections that were acquired using parallel

More information

Material for Chapter 6: Basic Principles of Tomography M I A Integral Equations in Visual Computing Material

Material for Chapter 6: Basic Principles of Tomography M I A Integral Equations in Visual Computing Material Material for Chapter : Integral Equations in Visual Computing Material Basic Principles of Tomography c 00 Bernhard Burgeth 0 Source: Images Figure : Radon Transform: ttenuation http://en.wikimedia.org/wiki/image:radon_transform.png

More information

A Backprojection-Filtration Algorithm for Nonstandard. Spiral Cone-beam CT with an N-PI Window

A Backprojection-Filtration Algorithm for Nonstandard. Spiral Cone-beam CT with an N-PI Window A Backprojection-Filtration Algorithm for Nonstandard Spiral Cone-beam CT with an N-PI Window Hengyong Yu, Yangbo Ye,, Shiying Zhao, Ge Wang, CT/Micro-CT Laboratory, Department of Radiology, Department

More information

Enhancement Image Quality of CT Using Single Slice Spiral Technique

Enhancement Image Quality of CT Using Single Slice Spiral Technique Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,

More information

Accelerated quantitative multi-material beam hardening correction(bhc) in cone-beam CT

Accelerated quantitative multi-material beam hardening correction(bhc) in cone-beam CT Accelerated quantitative multi-material beam hardening correction(bhc) in cone-beam CT Award: Poster No.: C-2161 Certificate of Merit Congress: ECR 2012 Type: Authors: Scientific Exhibit Q. Yang, M. Elter,

More information

Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction

Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction Klaus Erhard a, Michael Grass a, Sebastian Hitziger b, Armin Iske b and Tim Nielsen a a Philips Research Europe Hamburg,

More information

A Novel Two-step Method for CT Reconstruction

A Novel Two-step Method for CT Reconstruction A Novel Two-step Method for CT Reconstruction Michael Felsberg Computer Vision Laboratory, Dept. EE, Linköping University, Sweden mfe@isy.liu.se Abstract. In this paper we address the parallel beam 2D

More information

Moving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region

Moving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region Moving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region Andreas Hahn 1,2, Sebastian Sauppe 1,2, Michael Knaup 1, and Marc Kachelrieß 1,2 1 German Cancer Research Center (DKFZ),

More information

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications 4th International Symposium on NDT in Aerospace 2012 - Tu.3.A.1 Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications Malte KURFISS 1, Gerd STRECKENBACH 2 1 YXLON International

More information

Ultrasonic Multi-Skip Tomography for Pipe Inspection

Ultrasonic Multi-Skip Tomography for Pipe Inspection 18 th World Conference on Non destructive Testing, 16-2 April 212, Durban, South Africa Ultrasonic Multi-Skip Tomography for Pipe Inspection Arno VOLKER 1, Rik VOS 1 Alan HUNTER 1 1 TNO, Stieltjesweg 1,

More information

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS 23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS P. Helberg 1,

More information

Constructing System Matrices for SPECT Simulations and Reconstructions

Constructing System Matrices for SPECT Simulations and Reconstructions Constructing System Matrices for SPECT Simulations and Reconstructions Nirantha Balagopal April 28th, 2017 M.S. Report The University of Arizona College of Optical Sciences 1 Acknowledgement I would like

More information

Registration concepts for the just-in-time artefact correction by means of virtual computed tomography

Registration concepts for the just-in-time artefact correction by means of virtual computed tomography DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Registration concepts for the just-in-time artefact correction by means of virtual

More information

Iterative SPECT reconstruction with 3D detector response

Iterative SPECT reconstruction with 3D detector response Iterative SPECT reconstruction with 3D detector response Jeffrey A. Fessler and Anastasia Yendiki COMMUNICATIONS & SIGNAL PROCESSING LABORATORY Department of Electrical Engineering and Computer Science

More information

Improvements of FDK-type Helical Cone-beam CT Imaging

Improvements of FDK-type Helical Cone-beam CT Imaging Improvements of FDK-type Helical Cone-beam CT Imaging Liang Hong Zhu School of Electrical & Electronic Engineering A thesis submitted to the in fulfillment of the requirements for the degree of Doctor

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice Guido Hilgers, a Tonnis Nuver, André Minken Department

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

GRAPHICAL USER INTERFACE (GUI) TO STUDY DIFFERENT RECONSTRUCTION ALGORITHMS IN COMPUTED TOMOGRAPHY

GRAPHICAL USER INTERFACE (GUI) TO STUDY DIFFERENT RECONSTRUCTION ALGORITHMS IN COMPUTED TOMOGRAPHY GRAPHICAL USER INTERFACE (GUI) TO STUDY DIFFERENT RECONSTRUCTION ALGORITHMS IN COMPUTED TOMOGRAPHY A thesis submitted in partial fulfillment of the requirements for the degree of Master of Science in Engineering

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

A study on image quality provided by a kilovoltage cone-beam computed tomography

A study on image quality provided by a kilovoltage cone-beam computed tomography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 1, 2013 A study on image quality provided by a kilovoltage cone-beam computed tomography Julia Garayoa a and Pablo Castro Servicio de Radiofísica,

More information

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube T. Donath* a, F. Pfeiffer a,b, O. Bunk a, W. Groot a, M. Bednarzik a, C. Grünzweig a, E. Hempel c, S. Popescu c, M. Hoheisel

More information

Cone-beam reconstruction by backprojection and filtering

Cone-beam reconstruction by backprojection and filtering Andrei V. Bronnikov Vol. 7, No. / November 000/ J. Opt. Soc. Am. A 993 Cone-beam reconstruction by backprojection and filtering Andrei V. Bronnikov KEMA, Utrechtseweg 30, P.O. Box 9035, 6800 ET Arnhem,

More information

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid by Yingxuan Chen Graduate Program in Medical Physics Duke University Date: Approved: Lei Ren, Supervisor Fang Fang Yin, Chair

More information

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction. Mario Koerner

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction. Mario Koerner 2D Fan Beam Reconstruction 3D Cone Beam Reconstruction Mario Koerner Moscow-Bavarian Joint Advanced Student School 2006 March 19 2006 to March 29 2006 Overview 2D Fan Beam Reconstruction Shortscan Reconstruction

More information

Frequency split metal artifact reduction (FSMAR) in computed tomography

Frequency split metal artifact reduction (FSMAR) in computed tomography The Johns Hopkins University Advanced Computer Integrated Surgery Group 4 Metal Artifact Removal in C-arm Cone-Beam CT Paper Seminar Critical Review of Frequency split metal artifact reduction (FSMAR)

More information

Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners

Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners Poster No.: C-1748 Congress: ECR 2012 Type: Educational Exhibit Authors: S. Miyashita

More information

Advanced Image Reconstruction Methods for Photoacoustic Tomography

Advanced Image Reconstruction Methods for Photoacoustic Tomography Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic

More information

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC What is needed? Why? How often? Who says? QA and QC in Nuclear Medicine QA - collective term for all the efforts made to produce

More information

Application of Tatian s Method to Slanted-Edge MTF Measurement

Application of Tatian s Method to Slanted-Edge MTF Measurement Application of s Method to Slanted-Edge MTF Measurement Peter D. Burns Eastman Kodak Company, Rochester, NY USA 465-95 ABSTRACT The 33 method for the measurement of the spatial frequency response () of

More information

ULTRASONIC TESTING AND FLAW CHARACTERIZATION. Alex KARPELSON Kinectrics Inc., Toronto, Canada

ULTRASONIC TESTING AND FLAW CHARACTERIZATION. Alex KARPELSON Kinectrics Inc., Toronto, Canada ULTRASONIC TESTING AND FLAW CHARACTERIZATION Alex KARPELSON Kinectrics Inc., Toronto, Canada 1. Introduction Ultrasonic Testing (UT) is a commonly used inspection method. Various techniques are employed

More information