Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Size: px
Start display at page:

Download "Improvement of Efficiency and Flexibility in Multi-slice Helical CT"

Transcription

1 J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): DOI: /s x Improvement of Efficiency and Flexibility in Multi-slice Helical CT SUN Wen-wu 1 ( ), CHEN Si-ping 2 ( ), ZHUANG Tian-ge 3 ( ) (1. Nanchang Institute of Aeronautical Technology, Nanchang , China; 2. Shenhen University, Shenhen , China; 3. Department of Biomedical Engineering, Shanghai Jiaotong University, Shanghai , China) Abstract: One of the main aspects in computed tomography (CT) development is to make CT rapidly scan a large longitudinal volume with high -axis resolution. The combination of helical scanning with multi-slice CT is a promising approach. Image reconstruction in multi-slice CT becomes, therefore, the major challenge. Known algorithms need to derive the complementary data or work only for certain range of pitches. A reconstruction algorithm was presented that works with the direct data as well as arbitrary pitches. Filter interpolation based on the proposed method was implemented easy. The results of computer simulations under kinds of conditions for four-slice CT were presented. The proposed method can obtain higher efficiency than the conventional method. Key words: multi-slice computed tomography; helical interpolation; rebinning; filter interpolation CLC number: R Document code: A Introduction Step-and-shoot CT and helical CT are two modes used generally for a CT scan. Different from stepand-shoot CT where the table remains stationary while acquiring projection data, helical CT keeps the table transporting at a constant speed as well as continuous gantry rotation. Compared with step-and-shoot CT, helical CT shortens greatly the time of examination and improves the volume coverage speed performance [1]. Single-slice helical CT, however, is not able to obtain high -axis resolution while scanning large longitudinal volume due to a few reasons, e.g., the limitations of the tube loading and pitch, where the pitch is defined as the ratio of table feed per rotation versus entire collimated beam width. In order to meet the needs for many medical CT applications where both high volume coverage speed performance and high image quality, i.e., high - axis resolution and low image artifacts, are necessary, multi-row helical CT, equipping with a multi-row detector array, is a promising approach. Multi-slice CT allows for simultaneous scan of multiple slices at different locations, and enhances the -axis resolution. Due to the distinct differences in scanning geometry, multi-slice CT exhibits complex imaging characteristics and calls for new reconstruction algorithms. In multislice CT, the X-ray beam fan is extended in the -axis, as well as the gantry plane, the reconstruction algorithms for single-slice CT may no longer be adequate to maintain image quality. Therefore, 3D reconstruction algorithms may be required. In last few years, the cone- Received date: sun anke@yahoo.com.cn beam reconstruction algorithms [2-9] and the nutating surface reconstruction algorithms [10,11] have been addressed in order to eliminate the artifacts induced by beam divergence. Cone-beam back-projection, however, requires expensive calculations, and cannot be implemented efficiently. Fortunately, for multi-slice CT with less than four slices, the effect of beam divergence on image quality is small so that it may be neglected, and the cone-beam geometry is, therefore, approximated as multiple, parallel fan-beams. Several reconstruction algorithms have been addressed in multi-slice CT [1,12-14] that all do not take into account of beam divergent effect. However, the algorithms need to derive the complementary data or work only for certain range of pitches. The purpose of this paper is to propose a reconstruction algorithm in multi-slice helical CT. The proposed method works with the direct data as well as arbitrary pitches. Furthermore, it is also easy to implement filter interpolation based on the proposed method to reduce the artifacts and noise. 1 Multi-slice 180 Linear Interpolation The 360 and 180 linear interpolations in multi-slice CT (see Fig. 1(a)) are, respectively, extended from the 360LI and 180LI for single-slice CT (see Fig. 1(b)) [15,16]. The rays used by the 180 linear interpolation in multislice CT are not restricted from the same slice, both direct and complementary rays of different slices are used. Figure 2 shows the schematic sketch of helical scan in multi-slice CT in the case of pitch P = Let β and γ denote, respectively, the view and channel angles, the direct and complementary data for the nth

2 J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): Helical orbit Helical orbit 2D detector 1D detector Axis of rotation Axis of rotation Fig. 1 X-ray source X-ray source (a) Multi-slice CT (b) Single-slice CT The helical scanning geometry for multi-slice and single-slice CT Fig. 2 Filter width Schematic sketch of helical scan in multi-slice CT with pitch P = The slice position of direct data (solid line) and complementary data at γ = 0 (dashed line) are shown in whole helical scan. Filters may be used to average data along the -axis and provide more trade-off options between effective slice thickness and image noise slice are defined as p d (β, γ, n) and p c (β, γ, n), respectively, and d (β, n) and c (β, γ, n) are the corresponding slice positions. For the slice = 0, the 180 linear interpolation in multi-slice CT is described as where p(β, γ) = w(β, γ) p h (β, γ, q + 1) w(β, γ) = + (1 w(β, γ) p h (β, γ, q), (1) 0 h (β, γ, q) h (β, γ, q + 1) h (β, γ, q) ; p h (β, γ, q) and p h (β, γ, q + 1) are the lower and upper data sets closest to the slice, respectively; and h (β, γ, q) and h (β, γ, q+1) are the corresponding slice positions. As Fig. 2 indicates, the data p h (β, γ, q) and p h (β, γ, q + 1) can be the direct data p d (β, γ, n) or the complementary data p c (β, γ, n). 2 Proposed Method 2.1 Helical Interpolation According to the 180 linear interpolation in multislice CT given by Eq. (1), the complementary data have to be derived from the direct data, which makes the interpolation expensive. Actually, not only does there exist the relationship between the direct and complementary data as p c (β, γ, n) = p d (β + 2γ ± π, γ, n), (2) but it is easy to prove that the weights w(β, γ) have the following relationship: w(β, γ) = w(β + 2γ ± π, γ). (3) The contribution w(β, γ) p c (β, γ, n) of the complementary data p c (β, γ, n) to projection measurement p(β, γ) is, therefore, same as that of the direct data p d (β + 2γ ± π, γ, n) to p(β + 2γ ± π, γ), i.e., w(β + 2γ ± π, γ) p d (β + 2γ ± π, γ, n). The contributions of the projections from the nth slice to projection measurement p(β, γ) are then known by combining the contributions of both direct data p d (β, γ, n) to p(β, γ) and p d (β + 2γ ± π, γ, n) to p(β + 2γ ± π, γ). As a consequence, the derivations of the complementary data p c (β, γ, n) can be omitted so as to reduce the computation load while the weights of the direct data p d (β + 2γ ± π, γ, n) are doubled over the interval β [0, 2π].

3 410 J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): Parallel Rebinning Different from the fan-beam geometry where β and γ are used, two parameters θ and t are usually denoted as the aimuthal angle and radial position of a ray, respectively, for the parallel geometry. There are two simple relationships between the two sets of parameters (θ, t) and (β, γ) as θ = β + γ, (4) t = R sin γ, (5) where R is the source-to-isocenter distance. Using Eqs. (4) and (5), we can perform rebinning from sampled fan-beam projections p(β i, γ j ) to parallel projections. In general, the rebinning process can be separated into the following two steps: first, data are interpolated in the aimuth to obtain samples on radial lines that are arranged at equiangular intervals [17] ; second, a radial interpolation is performed to obtain equidistant samples on the radial lines. Let β 1,=0 (n) and β 2,=0 (γ j, n) be, respectively, the view angles when d (β, n) and c (β, γ j, n) equal to 0, and g(θ l, t j ) be the equiangular sampled parallel projections, where t j corresponds to γ j by Eq. (5). In order to maintain image quality, the rebinning in the proposed method is different from that given by Ref. [17]. Not only the data p(β i, γ j ) but the interpolated data p d (β 1,=0 (n), γ j, n) are used for the linear interpolation for g(θ l, t j ) as given by g(θ l, t j ) = h(β, γ j ) f(β, γ j )+(1 h(β, γ j )) f( ˆβ, γ j ), (6) where h(β, γ j ) is the weight, and β and ˆβ are the two view angles adjacent to the view angle θ l γ j β and ˆβ may be β i, β 1,=0 (n) and β 2,=0 (γ j, n). The interpolated data f(β, γ j ) for the cases are p(β i, γ j ), β =β i f(β, γ j )= 2p d (β 1,=0 (n), γ j, n), β =β 1,=0 (n). 0, β = β 2,=0 (δ j, n) However, if the channel angle increment γ equals to half the view angle increment β and there is the following relationship for all slice index n : β 1,=0 (n) = k n β, where k n is an integer, then, the view angles β 1,=0 (n) and β 2,=0 (γ j, n) coincide with β i, the proposed rebinning is simplified into that given by Ref. [17]. 2.3 Full-scan Parallel Reconstruction The normal full-scan parallel reconstruction technique, which uses filtered back-projection, is applied to obtain slice image. 2.4 Filter Interpolation The problem with helical interpolation in multi-slice CT is the existence of several discontinuous changeovers in pairs of data samples in the interpolated data sets obtained for one rotation. The most effective method of eliminating these changeover effects is to use filter interpolation that is a filtering process performed in the -axis with several data sets [13]. The concept is shown in Fig. 1. A width defined as the filter width (FW) is assumed in the direction, and several data points lie within FW. Filter interpolation can be divided into the following two steps: 1 resampling by linear interpolation using adjacent data points; 2 filtering the resampled data set. In the proposed method, it seems that the weights of the direct data for helical interpolation must be calculated at each resampling position when using filter interpolation based on the proposed method. Let v be the projection index, the weights w(β v, γ j, n) for the direct data of the nth slice are found to have the following relationship between the positions at = 0 and = 0 + as w =0+ (β v, γ j, n) = w =0 (β v v, γ j, n), where v is the projection spacing. Furthermore, the weights h =0 (β, γ j ) and h =0+ (β, γ j ) for rebinning the projections from fan-beam to equiangular parallel corresponding to the positions = 0 and = 0 +, respectively, also show the following characteristic similar to Eq. (7): h =0+ (β, γ j ) = h =0 (β v β, γ j ). As a consequence of the observations, it is sufficient to know the weights w =0 (β v, γ j, n) for helical interpolation and the weights h = 0 (β, γ j ) for rebinning at the position = 0. The convolution for filtering can be operated on the equiangular parallel data g(θ l, t j ). 3 Experiments and Discussions The results are simulated by the conventional and proposed methods. In the simulations, four ball phantoms ( = 20 mm) locate uniformly at the plane perpendicular to the -axis along a circle ( = 150 mm). The contrasts of the ball and background are 0 and HU, respectively. The beam width is fixed at 2 mm. The ratios of the proposed versus conventional method in terms of the time of reconstruction are given in Table 1 under kinds of conditions. The proposed method can obtain higher efficiency than the conventional method, and the advantages of the proposed over conventional method become more obvious with the increase of the number of resampling points.

4 J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): Fig. 3 Reconstructed images of ball phantoms with the diameter of 20 mm for various scan and filter parameters Furthermore, as reconstructed images that are 8 mm away from the centers of the balls show (see Fig. 3, the window level and width are and 150 HU, respectively), image quality is improved by filter interpolation, and enhanced significantly with the increase of FW as indicated in the bottom two rows. Table 1 The ratios of the proposed versus conventional method in terms of the reconstruction time while the number of resampling points is fixed at 11 and 21 Ratios of time P FW=2 mm FW=4 mm num 11 num 21 num 11 num Conclusion The reconstruction algorithm for multi-slice CT works well with the direct data as well as arbitrary pitches. Filter interpolation based on the proposed method can also be implemented easily. The results of computer simulations under kinds of conditions for four slice CT are presented. The quantitative gains show that the proposed method can obtain higher efficiency than the conventional method. References [1] Hu H. Multi-slice helical CT: scan and reconstruction [J]. Med Phys, 1999, 26(1): [2] Feldkamp L A, Davis L C, Kress J W. Practical conebeam algorithm [J]. J Opt Soc Am, 1984, 1(6): [3] Flohr T, Stierstorfer K, Bruder H, et al. Image reconstruction and image quality evaluation for a 16-slice CT scanner [J]. Med Phys, 2003, 30(5): [4] Silver M D, Taguchi K, Han K S. Field-of-view dependent helical pitch in multi-slice CT [C]//Proc SPIE. San Diego CA: [s.n.], 2001: [5] Katsevich A, Lauritsch G, Bruder H, et al. Evaluation and empirical analysis of an exact FBP algorithm for spiral cone-beam CT. [C]//Proc SPIE. San Diego, California: [s.n.], 2003: [6] Zhao S, Wang G. Feldkamp-type cone-beam tomography in the wavelet framework [J]. IEEE Transactions on Medical Imaging, 2000, 19(9): [7] Tam K, Samarasekera S, Sauer F. Exact cone beam CT with a spiral scan [J]. Phys Med Biol, 1998, 43(4): [8] Proksa R, Kohler T, Grass M, et al. The n-pi-method for helical cone-beam CT [J]. IEEE Transactions on Medical Imaging, 2000, 19(9): [9] Wang G, Lin T H, Cheng P C, et al. A general conebeam reconstruction algorithm [J]. IEEE Transactions on Medical Imaging, 1993, 12(9): [10] Kachelriess M, Schaller S, Kalender W A. Advanced single-slice rebinning in cone-beam helical CT [J]. Med Phys, 2000, 27(4):

5 412 J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): [11] Bruder H, Kachelriess M, Schaller S, et al. Single-slice rebinning reconstruction in helical cone-beam computed tomography [J]. IEEE Transactions on Medical Imaging, 2000, 19(9): [12] Liang Y, Kruger R. Dual-slice spiral versus single-slice spiral scanning: comparison of the physical performance of two computed tomography scanners [J]. Med Phys, 1996, 23(2): [13] Taguchi K, Aradate H. Algorithm for image reconstruction in multi-slice helical CT [J]. Med Phys, 1998, 25(4): [14] Schaller S, Flohr T, Klingenbeck K, et al. Helical interpolation algorithm for multi-slice helical CT- Part I: Theory [J]. IEEE Transactions on Medical Imaging, 1999, 19(9): [15] Kalender W, Seissler W, Klot E, et al. Spiral volumetric CT with single-breath-hold technique, continuous transport, and continuous scanner rotation [J]. Radiology, 1990, 176(1): [16] Crawford C, King K. Computed tomography scanning with simultaneous patient translation [J]. Med Phys, 1990, 17(6): [17] Besson G. CT image reconstruction from fan-parallel data [J]. Med Phys, 1999, 26(3):

An approximate cone beam reconstruction algorithm for gantry-tilted CT

An approximate cone beam reconstruction algorithm for gantry-tilted CT An approximate cone beam reconstruction algorithm for gantry-tilted CT Ming Yan a, Cishen Zhang ab, Hongzhu Liang a a School of Electrical & Electronic Engineering, Nanyang Technological University, Singapore;

More information

Feldkamp-type image reconstruction from equiangular data

Feldkamp-type image reconstruction from equiangular data Journal of X-Ray Science and Technology 9 (2001) 113 120 113 IOS Press Feldkamp-type image reconstruction from equiangular data Ben Wang a, Hong Liu b, Shiying Zhao c and Ge Wang d a Department of Elec.

More information

A study on the section sensitivity profile in multi-row-detector spiral CT

A study on the section sensitivity profile in multi-row-detector spiral CT Journal of X-Ray Science and Technology (3) IOS Press A study on the section sensitivity profile in multi-row-detector spiral CT Ge Wang a, Mark Madsen a, Katherine Redford a, Shiying Zhao b and Michael

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

Flying Focal Spot (FFS) in Cone-Beam CT Marc Kachelrieß, Member, IEEE, Michael Knaup, Christian Penßel, and Willi A. Kalender

Flying Focal Spot (FFS) in Cone-Beam CT Marc Kachelrieß, Member, IEEE, Michael Knaup, Christian Penßel, and Willi A. Kalender 1238 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 3, JUNE 2006 Flying Focal Spot (FFS) in Cone-Beam CT Marc Kachelrieß, Member, IEEE, Michael Knaup, Christian Penßel, and Willi A. Kalender Abstract

More information

Artifact analysis of approximate helical cone-beam CT reconstruction algorithms

Artifact analysis of approximate helical cone-beam CT reconstruction algorithms Artifact analysis of approximate helical cone-beam CT reconstruction algorithms Th. Köhler, a) R. Proksa, C. Bontus, and M. Grass Philips Research Laboratories, Sector Technical Systems, Roentgenstrasse

More information

Spiral-CT. Benjamin Keck. 21. March 2006

Spiral-CT. Benjamin Keck. 21. March 2006 Spiral-CT Benjamin Keck 21. March 2006 1 Motivation Spiral-CT offers reconstruction of long objects compared to circular filtered backprojection, where reconstruction is limited in z-direction. While the

More information

FDK Half-Scan with a Heuristic Weighting Scheme on a Flat Panel Detector-Based Cone Beam CT (FDKHSCW)

FDK Half-Scan with a Heuristic Weighting Scheme on a Flat Panel Detector-Based Cone Beam CT (FDKHSCW) Biomedical Imaging Volume 26, Article ID 83983, Pages 8 DOI.55/IJBI/26/83983 FDK Half-Scan with a Heuristic Weighting Scheme on a Flat Panel Detector-Based Cone Beam CT (FDKHSCW) Dong Yang and Ruola Ning

More information

Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset

Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset Andreas Maier 1, Patrick Kugler 2, Günter Lauritsch 2, Joachim Hornegger 1 1 Pattern Recognition Lab and SAOT Erlangen,

More information

A Backprojection-Filtration Algorithm for Nonstandard. Spiral Cone-beam CT with an N-PI Window

A Backprojection-Filtration Algorithm for Nonstandard. Spiral Cone-beam CT with an N-PI Window A Backprojection-Filtration Algorithm for Nonstandard Spiral Cone-beam CT with an N-PI Window Hengyong Yu, Yangbo Ye,, Shiying Zhao, Ge Wang, CT/Micro-CT Laboratory, Department of Radiology, Department

More information

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA Computed Tomography Principles, Design, Artifacts, and Recent Advances THIRD EDITION Jiang Hsieh SPIE PRESS Bellingham, Washington USA Table of Contents Preface Nomenclature and Abbreviations xi xv 1 Introduction

More information

Key words: Cone-beam CT CBCT, cardiac imaging, 4-D reconstruction, image quality

Key words: Cone-beam CT CBCT, cardiac imaging, 4-D reconstruction, image quality Extended parallel backprojection for standard three-dimensional and phase-correlated four-dimensional axial and spiral cone-beam CT with arbitrary pitch, arbitrary cone-angle, and 100% dose usage Marc

More information

TESTING OF THE CIRCLE AND LINE ALGORITHM IN THE SETTING OF MICRO-CT

TESTING OF THE CIRCLE AND LINE ALGORITHM IN THE SETTING OF MICRO-CT SCA2016-080 1/7 TESTING OF THE CIRCLE AND LINE ALGORITHM IN THE SETTING OF MICRO-CT Alexander Katsevich 1, 2 and Michael Frenkel 1 1 itomography Corp., 2 University of Central Florida (UCF) This paper

More information

Advanced single-slice rebinning for tilted spiral cone-beam CT

Advanced single-slice rebinning for tilted spiral cone-beam CT Advanced single-slice rebinning for tilted spiral cone-beam CT Marc Kachelrieß a) and Theo Fuchs Institute of Medical Physics, University of Erlangen Nürnberg, Germany Stefan Schaller Siemens AG, Medical

More information

Advanced single slice rebinning in cone beam spiral CT: Theoretical considerations and medical applications

Advanced single slice rebinning in cone beam spiral CT: Theoretical considerations and medical applications Advanced single slice rebinning in cone beam spiral CT: Theoretical considerations and medical applications Marc Kachelrieß a, Stefan Schaller b and Willi A. Kalender a a Institute of Medical Physics,

More information

Improvements of FDK-type Helical Cone-beam CT Imaging

Improvements of FDK-type Helical Cone-beam CT Imaging Improvements of FDK-type Helical Cone-beam CT Imaging Liang Hong Zhu School of Electrical & Electronic Engineering A thesis submitted to the in fulfillment of the requirements for the degree of Doctor

More information

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh et al. : Proceedings of the National Seminar & Exhibition on Non-Destructive Evaluation DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh V Acharya, Umesh Kumar, Gursharan

More information

The n-pi-method for Helical Cone-Beam CT

The n-pi-method for Helical Cone-Beam CT 848 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 19, NO. 9, SEPTEMBER 2000 The n-pi-method for Helical Cone-Beam CT R. Proksa*, Th. Köhler, M. Grass, and J. Timmer Abstract A new class of acquisition schemes

More information

Enhancement Image Quality of CT Using Single Slice Spiral Technique

Enhancement Image Quality of CT Using Single Slice Spiral Technique Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,

More information

GPU-based cone-beam CT reconstruction with extended FOV

GPU-based cone-beam CT reconstruction with extended FOV GPU-based cone-beam CT reconstruction with extended FOV Tamás Huszár 1, Gábor Jakab 12, and Attila Rácz 1 1 Mediso Medical Equipment Developing and Service Ltd. Budapest, 1022 Hungary http://www.mediso.hu,

More information

RECENTLY, biomedical imaging applications of

RECENTLY, biomedical imaging applications of 1190 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 24, NO. 9, SEPTEMBER 2005 A General Exact Reconstruction for Cone-Beam CT via Backprojection-Filtration Yangbo Ye*, Shiying Zhao, Hengyong Yu, Ge Wang, Fellow,

More information

A sufficient condition for spiral cone beam long object imaging via backprojection

A sufficient condition for spiral cone beam long object imaging via backprojection A sfficient condition for spiral cone beam long object imaging via backprojection K. C. Tam Siemens Corporate Research, Inc., Princeton, NJ, USA Abstract The response of a point object in cone beam spiral

More information

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction 2D Fan Beam Reconstruction 3D Cone Beam Reconstruction Mario Koerner March 17, 2006 1 2D Fan Beam Reconstruction Two-dimensional objects can be reconstructed from projections that were acquired using parallel

More information

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling nd International Symposium on NDT in Aerospace - We..A. Comparison of Probing Error in Dimensional Measurement by Means of D Computed Tomography with Circular and Helical Sampling Jochen HILLER, Stefan

More information

CIRCULAR scanning trajectory has been widely used in

CIRCULAR scanning trajectory has been widely used in IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 25, NO. 7, JULY 2006 869 Region of Interest Reconstruction From Truncated Data in Circular Cone-Beam CT Lifeng Yu, Yu Zou, Emil Y. Sidky, Charles A. Pelizzari,

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

A practical software platform of integrated reconstruction algorithms and forward algorithms for 2D industrial CT

A practical software platform of integrated reconstruction algorithms and forward algorithms for 2D industrial CT Journal of X-Ray Science and Technology 13 (2005) 9 21 9 IOS Press A practical software platform of integrated reconstruction algorithms and forward algorithms for 2D industrial CT Hui Li a, Zhaotian Zhang

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid by Yingxuan Chen Graduate Program in Medical Physics Duke University Date: Approved: Lei Ren, Supervisor Fang Fang Yin, Chair

More information

Novel C-arm based cone-beam CT using a source trajectory of two concentric arcs

Novel C-arm based cone-beam CT using a source trajectory of two concentric arcs Novel C-arm based cone-beam CT using a source trajectory of two concentric arcs Joseph Zambelli a, Brian E. Nett a,shuaileng a, Cyril Riddell c, Barry Belanger d,guang-hong Chen a,b a Department of Medical

More information

Non-Stationary CT Image Noise Spectrum Analysis

Non-Stationary CT Image Noise Spectrum Analysis Non-Stationary CT Image Noise Spectrum Analysis Michael Balda, Björn J. Heismann,, Joachim Hornegger Pattern Recognition Lab, Friedrich-Alexander-Universität Erlangen Siemens Healthcare, Erlangen michael.balda@informatik.uni-erlangen.de

More information

An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography

An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography 4th International Symposium on NDT in Aerospace 2012 - Tu.3.A.3 An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography Jonathan HESS *, Patrick KUEHNLEIN *, Steven OECKL

More information

Image Reconstruction 3 Fully 3D Reconstruction

Image Reconstruction 3 Fully 3D Reconstruction Image Reconstruction 3 Fully 3D Reconstruction Thomas Bortfeld Massachusetts General Hospital, Radiation Oncology, HMS HST.S14, February 25, 2013 Thomas Bortfeld (MGH, HMS, Rad. Onc.) Image Reconstruction

More information

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam Home Search Collections Journals About Contact us My IOPscience Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam projections This content has been downloaded from IOPscience.

More information

Principles of Computerized Tomographic Imaging

Principles of Computerized Tomographic Imaging Principles of Computerized Tomographic Imaging Parallel CT, Fanbeam CT, Helical CT and Multislice CT Marjolein van der Glas August 29, 2000 Abstract The total attenuation suffered by one beam of x-rays

More information

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings)

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings) Marquette University e-publications@marquette Biomedical Engineering Faculty Research and Publications Engineering, College of 1-1-1999 An Iterative Approach to the Beam Hardening Correction in Cone Beam

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction. Mario Koerner

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction. Mario Koerner 2D Fan Beam Reconstruction 3D Cone Beam Reconstruction Mario Koerner Moscow-Bavarian Joint Advanced Student School 2006 March 19 2006 to March 29 2006 Overview 2D Fan Beam Reconstruction Shortscan Reconstruction

More information

Research Article Cone-Beam Composite-Circling Scan and Exact Image Reconstruction for a Quasi-Short Object

Research Article Cone-Beam Composite-Circling Scan and Exact Image Reconstruction for a Quasi-Short Object Hindawi Publishing Corporation International Journal of Biomedical Imaging Volume 27, Article ID 87, pages doi:.55/27/87 Research Article Cone-Beam Composite-Circling Scan and Exact Image Reconstruction

More information

Translational Computed Tomography: A New Data Acquisition Scheme

Translational Computed Tomography: A New Data Acquisition Scheme 2nd International Symposium on NDT in Aerospace 2010 - We.1.A.3 Translational Computed Tomography: A New Data Acquisition Scheme Theobald FUCHS 1, Tobias SCHÖN 2, Randolf HANKE 3 1 Fraunhofer Development

More information

GRAPHICAL USER INTERFACE (GUI) TO STUDY DIFFERENT RECONSTRUCTION ALGORITHMS IN COMPUTED TOMOGRAPHY

GRAPHICAL USER INTERFACE (GUI) TO STUDY DIFFERENT RECONSTRUCTION ALGORITHMS IN COMPUTED TOMOGRAPHY GRAPHICAL USER INTERFACE (GUI) TO STUDY DIFFERENT RECONSTRUCTION ALGORITHMS IN COMPUTED TOMOGRAPHY A thesis submitted in partial fulfillment of the requirements for the degree of Master of Science in Engineering

More information

A comparison of FBP and BPF reconstruction methods for circular X-ray tomography with off-center detector

A comparison of FBP and BPF reconstruction methods for circular X-ray tomography with off-center detector A comparison of FBP and BPF reconstruction methods for circular X-ray tomography with off-center detector Dirk Schäfer, Michael Grass Philips Technologie GmbH Forschungslaboratorien, Röntgenstrasse, D

More information

Medical Image Processing: Image Reconstruction and 3D Renderings

Medical Image Processing: Image Reconstruction and 3D Renderings Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

An FDK-like cone-beam SPECT reconstruction algorithm for non-uniform attenuated

An FDK-like cone-beam SPECT reconstruction algorithm for non-uniform attenuated Home Search Collections Journals About Contact us My IOPscience An FK-like cone-beam SPECT reconstruction algorithm for non-uniform attenuated projections acquired using a circular trajectory This content

More information

Scaling Calibration in the ATRACT Algorithm

Scaling Calibration in the ATRACT Algorithm Scaling Calibration in the ATRACT Algorithm Yan Xia 1, Andreas Maier 1, Frank Dennerlein 2, Hannes G. Hofmann 1, Joachim Hornegger 1,3 1 Pattern Recognition Lab (LME), Friedrich-Alexander-University Erlangen-Nuremberg,

More information

MOTION of a patient induces artifacts such as streaks

MOTION of a patient induces artifacts such as streaks Motion-compensated Image Reconstruction for Cardiac CT with Sinogram-Based Motion Estimation JangHwanChoMember IEEE and Jeffrey A. Fessler Fellow IEEE Abstract Motion-compensated image reconstruction ()

More information

MULTI-PURPOSE 3D COMPUTED TOMOGRAPHY SYSTEM

MULTI-PURPOSE 3D COMPUTED TOMOGRAPHY SYSTEM MULTI-PURPOSE 3D COMPUTED TOMOGRAPHY SYSTEM M. Simon, C. Sauerwein, I. Tiseanu, S. Burdairon Hans Wälischmiller GmbH Klingleweg 8, D-88709 Meersburg, Germany e-mail: ms@hwm.com ABSTRACT A new flexible

More information

Artifacts in Spiral X-ray CT Scanners: Problems and Solutions

Artifacts in Spiral X-ray CT Scanners: Problems and Solutions Artifacts in Spiral X-ray CT Scanners: Problems and Solutions Mehran Yazdi, and Luc Beaulieu International Science Index, Electrical and Computer Engineering waset.org/publication/12966 Abstract Artifact

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Biomedical Imaging. Computed Tomography. Patrícia Figueiredo IST

Biomedical Imaging. Computed Tomography. Patrícia Figueiredo IST Biomedical Imaging Computed Tomography Patrícia Figueiredo IST 2013-2014 Overview Basic principles X ray attenuation projection Slice selection and line projections Projection reconstruction Instrumentation

More information

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice Guido Hilgers, a Tonnis Nuver, André Minken Department

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Geometric studies on variable radius spiral cone-beam scanning

Geometric studies on variable radius spiral cone-beam scanning Geometric studies on variable radius spiral cone-beam scanning Yangbo Ye a) Jiehua Zhu Department of Mathematics, The University of Iowa, Iowa City, Iowa 52242 Ge Wang Department of Radiology, The University

More information

Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners

Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners Poster No.: C-1748 Congress: ECR 2012 Type: Educational Exhibit Authors: S. Miyashita

More information

Central Slice Theorem

Central Slice Theorem Central Slice Theorem Incident X-rays y f(x,y) R x r x Detected p(, x ) The thick line is described by xcos +ysin =R Properties of Fourier Transform F [ f ( x a)] F [ f ( x)] e j 2 a Spatial Domain Spatial

More information

Digital phantoms for the evaluation of a software used for an automatic analysis of the Winston-Lutz test in image guided radiation therapy

Digital phantoms for the evaluation of a software used for an automatic analysis of the Winston-Lutz test in image guided radiation therapy Author manuscript, published in "Medical Imaging 008: Physics of Medical Imaging, San Diego, CA, USA : United States (008)" DOI : 10.1117/1.768668 Digital phantoms for the evaluation of a software used

More information

X-ray Industrial Computed Laminography (ICL) Simulation Study of Planar Objects: Optimization of Laminographic Angle

X-ray Industrial Computed Laminography (ICL) Simulation Study of Planar Objects: Optimization of Laminographic Angle More info about this article: http://www.ndt.net/?id=21086 X-ray Industrial Computed Laminography (ICL) Simulation Study of Planar Objects: Optimization of Laminographic Angle Lakshminarayana Yenumula

More information

Grangeat-type helical half-scan computerized tomography algorithm for reconstruction of a short object

Grangeat-type helical half-scan computerized tomography algorithm for reconstruction of a short object Grangeat-type helical half-scan computerized tomography algorithm for reconstruction of a short object Seung Wook Lee a) CT/Micro-CT Laboratory, Department of Radiology, University of Iowa, Iowa City,

More information

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare

More information

An Efficient Technique For Multi-Phase Model Based Iterative Reconstruction

An Efficient Technique For Multi-Phase Model Based Iterative Reconstruction 1 An Efficient Technique For Multi-Phase Model Based Iterative Reconstruction Shiyu Xu, Debashish Pal and Jean-Baptiste Thibault Abstract Multi-phase scan is a fundamental CT acquisition technology used

More information

J. Ling Chinese Academy of Science Institute for Ceramics Shanghai, China

J. Ling Chinese Academy of Science Institute for Ceramics Shanghai, China DEVELOPMENT AND APPLICATION OF LOCAL 3-D CT RECONSTRUCTION SOFTWARE FOR IMAGING CRITICAL REGIONS IN LARGE CERAMIC TURBINE ROTORS E. A. Sivers, D. L. Holloway, and W. A. Ellingson Argonne National Laboratory

More information

Adaptive algebraic reconstruction technique

Adaptive algebraic reconstruction technique Adaptive algebraic reconstruction technique Wenkai Lua) Department of Automation, Key State Lab of Intelligent Technology and System, Tsinghua University, Beijing 10084, People s Republic of China Fang-Fang

More information

Introduction to Medical Imaging. Cone-Beam CT. Klaus Mueller. Computer Science Department Stony Brook University

Introduction to Medical Imaging. Cone-Beam CT. Klaus Mueller. Computer Science Department Stony Brook University Introduction to Medical Imaging Cone-Beam CT Klaus Mueller Computer Science Department Stony Brook University Introduction Available cone-beam reconstruction methods: exact approximate algebraic Our discussion:

More information

ECG-Correlated Imaging of the Heart with Subsecond Multislice Spiral CT

ECG-Correlated Imaging of the Heart with Subsecond Multislice Spiral CT 888 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 19, NO. 9, SEPTEMBER 2000 ECG-Correlated Imaging of the Heart with Subsecond Multislice Spiral CT Marc Kachelrieß*, Stefan Ulzheimer, and Willi A. Kalender

More information

THE FAN-BEAM scan for rapid data acquisition has

THE FAN-BEAM scan for rapid data acquisition has 190 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 26, NO. 2, FEBRUARY 2007 Hilbert Transform Based FBP Algorithm for Fan-Beam CT Full Partial Scans Jiangsheng You*, Member, IEEE, Gengsheng L. Zeng, Senior

More information

Cone-beam reconstruction by backprojection and filtering

Cone-beam reconstruction by backprojection and filtering Andrei V. Bronnikov Vol. 7, No. / November 000/ J. Opt. Soc. Am. A 993 Cone-beam reconstruction by backprojection and filtering Andrei V. Bronnikov KEMA, Utrechtseweg 30, P.O. Box 9035, 6800 ET Arnhem,

More information

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),

More information

Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction

Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction Klaus Erhard a, Michael Grass a, Sebastian Hitziger b, Armin Iske b and Tim Nielsen a a Philips Research Europe Hamburg,

More information

Advanced Image Reconstruction Methods for Photoacoustic Tomography

Advanced Image Reconstruction Methods for Photoacoustic Tomography Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60601-2-44 2001 AMENDMENT 1 2002-09 Amendment 1 Medical electrical equipment Part 2-44: Particular requirements for the safety of X-ray equipment for computed tomography Amendement

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

A prototype table-top inverse-geometry volumetric CT system

A prototype table-top inverse-geometry volumetric CT system A prototype table-top inverse-geometry volumetric CT system Taly Gilat Schmidt a Department of Radiology, Stanford University, Stanford, California 94305 Josh Star-Lack NexRay, Inc., Los Gatos, California

More information

Planar tomosynthesis reconstruction in a parallel-beam framework via virtual object reconstruction

Planar tomosynthesis reconstruction in a parallel-beam framework via virtual object reconstruction Planar tomosynthesis reconstruction in a parallel-beam framework via virtual object reconstruction Brian E. Nett a,shuaileng a and Guang-Hong Chen a,b a Department of Medical Physics, University of Wisconsin-Madison,

More information

Accelerated C-arm Reconstruction by Out-of-Projection Prediction

Accelerated C-arm Reconstruction by Out-of-Projection Prediction Accelerated C-arm Reconstruction by Out-of-Projection Prediction Hannes G. Hofmann, Benjamin Keck, Joachim Hornegger Pattern Recognition Lab, University Erlangen-Nuremberg hannes.hofmann@informatik.uni-erlangen.de

More information

Optimization of Cone Beam CT Reconstruction Algorithm Based on CUDA

Optimization of Cone Beam CT Reconstruction Algorithm Based on CUDA Sensors & Transducers 2013 by IFSA http://www.sensorsportal.com Optimization of Cone Beam CT Reconstruction Algorithm Based on CUDA 1 Wang LI-Fang, 2 Zhang Shu-Hai 1 School of Electronics and Computer

More information

A Reconstruction Algorithm for Helical Cone-Beam SPECT

A Reconstruction Algorithm for Helical Cone-Beam SPECT 1092 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 40, NO. 4, AUGUST 1993 A Reconstruction Algorithm for Helical Cone-Beam SPECT Y. Weng, G. L. Zeng and G. T. Gullberg Medical Imaging Research Laboratory

More information

A closer look at CT scanning

A closer look at CT scanning Vet Times The website for the veterinary profession https://www.vettimes.co.uk A closer look at CT scanning Author : Charissa Lee, Natalie Webster Categories : General, Vets Date : April 3, 2017 A basic

More information

The Near Future in Cardiac CT Image Reconstruction

The Near Future in Cardiac CT Image Reconstruction SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de

More information

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct J. DHAENE* 1, E. PAUWELS 1, T. DE SCHRYVER 1, A. DE MUYNCK 1, M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics

More information

A half-scan error reduction based algorithm for cone-beam CT

A half-scan error reduction based algorithm for cone-beam CT Journal of X-Ray Science an Technology 12 (2004) 73 82 73 IOS Press A half-scan error reuction base algorithm for cone-beam CT Kai Zeng a, Zhiqiang Chen a, Li Zhang a an Ge Wang a,b a Department of Engineering

More information

Reconstruction in CT and relation to other imaging modalities

Reconstruction in CT and relation to other imaging modalities Reconstruction in CT and relation to other imaging modalities Jørgen Arendt Jensen November 16, 2015 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering Center for Fast Ultrasound

More information

High-performance tomographic reconstruction using graphics processing units

High-performance tomographic reconstruction using graphics processing units 18 th World IMACS / MODSIM Congress, Cairns, Australia 13-17 July 29 http://mssanz.org.au/modsim9 High-performance tomographic reconstruction using graphics processing units Ya.I. esterets and T.E. Gureyev

More information

Image Reconstruction from Projection

Image Reconstruction from Projection Image Reconstruction from Projection Reconstruct an image from a series of projections X-ray computed tomography (CT) Computed tomography is a medical imaging method employing tomography where digital

More information

Recognition and Measurement of Small Defects in ICT Testing

Recognition and Measurement of Small Defects in ICT Testing 19 th World Conference on Non-Destructive Testing 2016 Recognition and Measurement of Small Defects in ICT Testing Guo ZHIMIN, Ni PEIJUN, Zhang WEIGUO, Qi ZICHENG Inner Mongolia Metallic Materials Research

More information

GPU-based Fast Cone Beam CT Reconstruction from Undersampled and Noisy Projection Data via Total Variation

GPU-based Fast Cone Beam CT Reconstruction from Undersampled and Noisy Projection Data via Total Variation GPU-based Fast Cone Beam CT Reconstruction from Undersampled and Noisy Projection Data via Total Variation 5 10 15 20 25 30 35 Xun Jia Department of Radiation Oncology, University of California San Diego,

More information

Modeling and Incorporation of System Response Functions in 3D Whole Body PET

Modeling and Incorporation of System Response Functions in 3D Whole Body PET Modeling and Incorporation of System Response Functions in 3D Whole Body PET Adam M. Alessio, Member IEEE, Paul E. Kinahan, Senior Member IEEE, and Thomas K. Lewellen, Senior Member IEEE University of

More information

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Marc Kachelrieß, a Katia Sourbelle, and Willi A. Kalender Institute of Medical Physics, University of

More information

Micro-CT in situ study of carbonate rock microstructural evolution for geologic CO2 storage

Micro-CT in situ study of carbonate rock microstructural evolution for geologic CO2 storage Downloaded from orbit.dtu.dk on: Apr 10, 2018 Micro-CT in situ study of carbonate rock microstructural evolution for geologic CO2 storage Zheng, Yi; Yang, Yan; Rogowska, M.; Gundlach, Carsten Published

More information

Emission Computed Tomography Notes

Emission Computed Tomography Notes Noll (24) ECT Notes: Page 1 Emission Computed Tomography Notes Introduction Emission computed tomography (ECT) is the CT applied to nuclear medicine. There are two varieties of ECT: 1. SPECT single-photon

More information

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Beam Attenuation Grid Based Scatter Correction Algorithm for More Info at Open Access Database

More information

Parallel Implementation of Katsevich s FBP Algorithm

Parallel Implementation of Katsevich s FBP Algorithm Hindawi Publishing Corporation International Journal of Biomedical Imaging Volume 26, Article ID 7463, Pages 8 DOI.55/IJBI/26/7463 Parallel Implementation of Katsevich s FBP Algorithm Jiansheng Yang, Xiaohu

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

Two Local FBP Algorithms for Helical Cone-beam Computed Tomography

Two Local FBP Algorithms for Helical Cone-beam Computed Tomography Digital Industrial Radiology and Computed Tomography (DIR 215) 22-25 June 215, Belgium, Ghent - www.ndt.net/app.dir215 More Info at Open Access Database www.ndt.net/?id=187 Two Local FBP Algorithms for

More information

7/31/ D Cone-Beam CT: Developments and Applications. Disclosure. Outline. I have received research funding from NIH and Varian Medical System.

7/31/ D Cone-Beam CT: Developments and Applications. Disclosure. Outline. I have received research funding from NIH and Varian Medical System. 4D Cone-Beam CT: Developments and Applications Lei Ren, PhD, DABR Department of Radiation Oncology Duke University Medical Center Disclosure I have received research funding from NIH and Varian Medical

More information

Clinical Evaluation of GPU-Based Cone Beam Computed Tomography.

Clinical Evaluation of GPU-Based Cone Beam Computed Tomography. Clinical Evaluation of GPU-Based Cone Beam Computed Tomography. Peter B. Noël 1,2, Alan M. Walczak 2, Kenneth R. Hoffmann 1,2, Jinhui Xu 1, Jason J. Corso 1, and Sebastian Schafer 2 1 Department of Computer

More information

Metal artifacts reduction in x-ray CT based on segmentation and forward-projection

Metal artifacts reduction in x-ray CT based on segmentation and forward-projection Bio-Medical Materials and Engineering 24 (2014) 3287 3293 DOI 10.3233/BME-141151 IOS Press 3287 Metal artifacts reduction in x-ray CT based on segmentation and forward-projection Shoukat Nawaz, Jian Fu

More information

Adaptive region of interest method for analytical micro-ct reconstruction

Adaptive region of interest method for analytical micro-ct reconstruction Journal of X-Ray Science and Technology 19 (2011) 23 33 23 DOI 10.3233/XST-2010-0274 IOS Press Adaptive region of interest method for analytical micro-ct reconstruction Wanneng Yang, Xiaochun Xu, Kun Bi,

More information