1 nd International Symposium on NDT in Aerospace - We..A. Comparison of Probing Error in Dimensional Measurement by Means of D Computed Tomography with Circular and Helical Sampling Jochen HILLER, Stefan KASPERL, Tobias SCHÖN, Stefan SCHRÖPFER * Daniel WEISS ** * Fraunhofer IIS, Development Center for X-ray Technology, Dr.-Mack-Str. 8, 97 Fürth, Germany ** Carl Zeiss Industrielle Messtechnik, Carl-Zeiss-Str. -, 77 Oberkochen, Germany Abstract. In this article the parameter probing error according to the national guideline VDI/VDE is determined for the circular and helical sampling of spherical objects using cone-beam computed tomography. In order to be able to evaluate exclusively the acquisition geometry, simulated data sets are used for the determination of the parameters. Due to the completeness of the data a significantly lower probing error is to be expected when applying helical CT, especially in the upper and lower border areas.. Introduction Axial cone-beam computed tomography systems (D CT) for industrial applications use a circular sampling curve, well aware of the fact that using this acquisition geometry it is not possible to obtain all information necessary for an artifact-free reconstruction. Namely in the upper and lower border areas of the measured volume high contrast losses are to be expected due to the incomplete data resulting from the angle of aperture. When applying the helical acquisition geometry, however, the data for the reconstruction of the measured object are complete. The parameters described in the national guideline VDI/VDE are suitable for a quantitative comparison of the measurement characteristics of both acquisition geometries. These parameters are defined for specification test, supervision and comparison of measurement characteristics of CT systems following DIN EN ISO - closely. Referring to optical and tactile measuring systems the parameter probing error is described as the three-dimensional deviation behaviour of the entire system in a very small measured volume, differentiating between probing error form PF and probing error size PS. In this article the parameter probing error is determined for the circular and helical sampling of spherical objects according to the national guideline VDI/VDE. In order to be able to evaluate exclusively the acquisition geometry, simulated data sets are used for the determination of the parameters. Due to the completeness of the data a significantly lower probing error is to be expected when applying helical CT, especially in the upper and lower border areas. The paper is divided as followed. Section compares conventional and helical D CT systems. In section the content of the national guideline VDI/VDE regarding the definition of the probing error is described. Subsequently the parameters of License:

2 the computer simulation of the CT scanning process are described in section. After the presentation of the results in section the paper ends with a short summary.. Comparison of circular and helical cone-beam sampling geometry Conventional D CT techniques for industrial applications are usually using a circular sampling geometry, even though not all necessary information for an artifact-free reconstruction of the object to be inspected can be provided this way. During the circular sampling (standard D CT) the object to be inspected is rotated only around the rotational axis (see Figure a). The reconstruction method is normally based on the Feldkamp algorithm (FDK) [] or derivates. According to the so-called Tuy-Smith sufficiency condition [] an exact reconstruction without losses of spatial resolution using a circular acquisition mode is possible if all surfaces intersecting the object intersect the trajectory of the X-ray source at least once. This condition is only true for the central x-y plane of the measuring field using standard D CT. In a circular cone-beam scan, Radon data are known only inside a torus as shown in figure. rotation axis shadow zone Figure : Radon torus in the circular cone-beam scanning case. Radon data are known only inside the torus. Losses of spatial resolution in the direction of the ration axis using circular conebeam CT can lead to rhombus-like artifacts (so called Feldkamp artifacts). These artifacts increase with the cone beam angle []. The trend-setting helical CT is based on a helical sampling geometry. When using helical data acquisition, the object is also shifted along the rotation axis. From the object s perspective, the source of radiation is moving on a helical trajectory. For the helical acquisition mode the Tuy-Smith sufficiency condition is fulfilled for all x-y planes. So the measuring object can be reconstructed exactly. Figure shows the two sampling geometries in comparison. Axial shift direction Object z Object z Source x y Source x y Axis of rotation Flat-panel detector Axis of rotation Flat-panel detector a) Circular (standard) D CT b) Helical D CT Figure : Comparison of the two D CT scanning techniques. Figure a) displays the circular (standard) D CT scanning trajectory. Using helical D CT displayed in b) the object is also shifted along the rotation axis.

3 . Probing Error according to the guideline VDI/VDE Aim of the guideline VDI/VDE of the society for metrology and automation (GMA) is the description of methods for the reproducible testing and calibration of measurements using industrial CT. The guideline also addresses techniques for the determination of measurement uncertainty. In addition to that, fields of use for industrial CT are to be specified and described in order to achieve optimal usage of systems available []. The guideline is supposed to facilitate a common approach towards testing and calibration for producers and users of such CT systems. Also an aid is given to the competent users for evaluating their systems in a better way. On part -. some parameters like probing error size PS or form PF are defined. Also acceptance tests are defined in order to check, whether the specified critical values are adhered to. The requirements of the guideline towards the test specimens are: A calibrated test specimen consisting of six spheres is recommended to evaluate PS and PF both on- and off-axis and at three different heights in the volume. Sphere diameter should be -% of the measurement space diagonal. At least surface points for the determination of the deviations for size and form are to be used. It is recommended to perform the measurement under two "significantly" different magnifications. The probing error size PS describes the difference between measured diameter D and reference value D ref : PS = D - D ref () The probing error form PF describes the range of radial deviations and is defined as the difference between maximal and minimal distance of the measured surface points from the centre of the compensating element: PF = R max - R min () Figure illustrates the measuring procedure on the calculated surface of the reconstructed and segmented ball specimen. Figure : Measurement of the diameter D and the minimal and maximum values R min and R max using a compensating element (marked red). For the adjustment of the compensating element in this case a best-fit method was used.

4 . Numerical simulation of the CT scanning process To compare the dimensional measuring properties of circular and helical sampling computer simulations were used. The CT scanning process was simulated deterministically. A monoenergetic point source was assumed. A sketch of the virtual ball specimen used is displayed in Figure. The diameter of the balls is mm. This value was used as the reference value D ref. axis of rotation z z = mm mm Figure : Sketch of the ball specimen used in the simulation. The diameter of the balls is mm. The balls are numbered as indicated. The virtual detector consists of a 78 9 pixel matrix with a pixel pitch (= pixel size) of µm. The focus to detector distance is 8 mm and the magnification factor 7., which implies a focus to isocenter distance of 7 mm, a fan beam angle of.9 and a cone beam angle of 7.. For the circular (FDK) and helical reconstruction the following parameters were used. We only investigated the measurement properties for one magnification factor. Parameters for the circular (FDK) reconstruction: Number of projections : Voxel size in reconstruction (kubic) : µm Parameters for the helical reconstruction: Number of projections : Angle increment :. Axial shift increment : µm Voxel size in reconstruction (kubic) : µm The reconstructed volume size is

5 . Results For the determination of the ball s surface a local adaptive threshold method is used []. The probing error size and form was determined using the compensating element sphere and a best-fit method to adjust the element into the measuring points on the ball s surface. Per ball measuring points were used. The local measurement deviation as the difference of the measuring points to the compensating element fitted is plotted in the angle range of -π π and π in spherical coordinates to form an unrolled ball surface. Values between local measuring points are interpolated. The values are standardised in the range of - µm. Figure shows the diagrams for the balls reconstructed. For the circular CT the high deviations at the poles for balls beyond the central plane (ball number,,, ) are caused by missing Radon data. The results for helical CT show, that high measurement accuracy is achievable even if the object is placed outside of the central plane (ball and ) due to complete Radon data. In comparison to circular CT the results clearly show the almost homogeneous measurement properties of the setup when using a helical sampling geometry. In all fields the measuring deviations are almost comparable using this scanning mode. FDK, central, top (ball ) FDK, central, center (ball ) FDK, central, lower (ball ) Helical, central, top (ball ) Helical, central, center (ball ) Helical, central, lower (ball ) (µm) FDK, outer, top (ball ) FDK, outer, center (ball ) FDK, outer, lower (ball ) Helical, outer, top (ball ) Helical, outer, center (ball ) Helical, outer, lower (ball ) Figure : Visualization of the local measurement deviation using the circular and the helical sampling geometry. The difference of the measuring points to the compensating element fitted is plotted in the angle range of -π π and π in spherical coordinates.

6 . Probing error size and form The probing errors size and form for the FDK and the helical reconstructions of the virtual ball specimen were determined according to the formulas () and (). The probing errors size and form are plotted against the ball number as shown in figure. The deviation of probing error size for the FDK and the helical reconstruction is about.7 till. µm at a voxel pitch of µm. This higher deviation is primarily caused by the higher interpolation effort in the helical reconstruction algorithm at the back projection filtering step. For this, further investigations have to be arranged whether this effect can be neglected in case of real projection data or not. Outside the central plane the probing error form for the FDK reconstruction is about three times higher than for the helical reconstruction. The deviation is caused by incomplete data. The Tuy-Smith condition is not fulfilled for the circular sampling geometry. As a consequence of that, the image sharpness at the poles of the balls outside the central plane is less than in the central plane. The probing error form for the helical reconstruction is about µm to 8 µm for all balls. As assumed, only in the central plane the probing error form is comparable to the circular sampling. Probing error size PS (µm) - Circular CT (FDK) Helical CT - Ball number Probing error form PF (µm) Circular CT (FDK) Helical CT Ball number Figure : Probing error size (top) and form (lower) plotted against the ball number.

7 In Table the results for the probing errors size and form of both circular and helical sampling geometry are listed. Circular CT (FDK) Probing error Helical CT Probing error Ball Size PS (µm) Form PF (µm) Ball Size PS (µm) Form PF (µm) Table : Results for the probing errors size and form for the circular and helical reconstruction of the virtual ball specimen Summary and outlook In this article the probing errors size and form were determined for a circular and a helical CT scan and reconstruction of a ball specimen in order to compare the measuring properties of these two CT techniques according to the national guideline VDI/VDE. A deterministic CT simulator was used to calculate the projection images of the virtual ball specimen. The results clearly show a small probing error form in all measurement positions considered due to complete Radon data using the helical sampling geometry. Using a circular sampling geometry probing error form is only comparable to the helical sampling in the central plane. Because of the total avoidance of Feldkamp artifacts, high precision dimensional measuring with minimum geometrical faults at reproduction, helical sampling should be preferred to the classical circular sampling geometry - especially when scanning very long objects - to achieve highest precision of form measurements. Due to the higher interpolation effort in the helical reconstruction algorithm, further investigations regarding the accuracy of size measurements using helical CT and real projection data will be arranged next. References [] Feldkamp, L. A.; Davis, L. C.; Kress, J. W.: Practical Cone-Beam Algorithm. J. Opt. Soc Amer,, A, 98, p. -9. [] Tuy, H.: An Inversion Formula for Cone-Beam Reconstruction. SIAM Journal of Applied Mathematics,, 98, p. -. [] Buzug, T. M.: Computed Tomography: From Photon Statistics to Modern Cone-Beam CT. Springer,. [] VDI/VDE -.: Computertomografie in der dimensionellen Messtechnik. Leitfaden zur Anwendung von DIN EN ISO für Koordinatenmessgeräte mit CT-Sensoren. Technische Regel, Entwurf, 8/9. [] Reinhart, C.: Industrial Computer Tomography - A universal Inspection. Proceedings of 7th World Conference on Nondestructive Testing, 8. 7

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system ECNDT 2014 Prague October 6-10, 2014 Dr. Eberhard Neuser Dr. Alexander Suppes Imagination

More information

An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography

An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography 4th International Symposium on NDT in Aerospace 2012 - Tu.3.A.3 An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography Jonathan HESS *, Patrick KUEHNLEIN *, Steven OECKL

More information

Introduction to Medical Imaging. Cone-Beam CT. Klaus Mueller. Computer Science Department Stony Brook University

Introduction to Medical Imaging. Cone-Beam CT. Klaus Mueller. Computer Science Department Stony Brook University Introduction to Medical Imaging Cone-Beam CT Klaus Mueller Computer Science Department Stony Brook University Introduction Available cone-beam reconstruction methods: exact approximate algebraic Our discussion:

More information

Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset

Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset Andreas Maier 1, Patrick Kugler 2, Günter Lauritsch 2, Joachim Hornegger 1 1 Pattern Recognition Lab and SAOT Erlangen,

More information

Comparison of Reconstruction Methods for Computed Tomography with Industrial Robots using Automatic Object Position Recognition

Comparison of Reconstruction Methods for Computed Tomography with Industrial Robots using Automatic Object Position Recognition 19 th World Conference on Non-Destructive Testing 2016 Comparison of Reconstruction Methods for Computed Tomography with Industrial Robots using Automatic Object Position Recognition Philipp KLEIN 1, Frank

More information

MULTI-PURPOSE 3D COMPUTED TOMOGRAPHY SYSTEM

MULTI-PURPOSE 3D COMPUTED TOMOGRAPHY SYSTEM MULTI-PURPOSE 3D COMPUTED TOMOGRAPHY SYSTEM M. Simon, C. Sauerwein, I. Tiseanu, S. Burdairon Hans Wälischmiller GmbH Klingleweg 8, D-88709 Meersburg, Germany e-mail: ms@hwm.com ABSTRACT A new flexible

More information

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications 4th International Symposium on NDT in Aerospace 2012 - Tu.3.A.1 Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications Malte KURFISS 1, Gerd STRECKENBACH 2 1 YXLON International

More information

THREE DIMENSIONAL EXAMINATION OF DIRECTIVITY PATTERN IN IMMERSION TANK TESTING

THREE DIMENSIONAL EXAMINATION OF DIRECTIVITY PATTERN IN IMMERSION TANK TESTING 7 th International Symposium on NDT in Aerospace We.3.A.7 THREE DIMENSIONAL EXAMINATION OF DIRECTIVITY PATTERN IN IMMERSION TANK TESTING Daniel KOTSCHATE 1, Thomas HECKEL 1 and Dirk GOHLKE 1 1 Federal

More information

Traceable 3D X-Ray Measurements ZEISS METROTOM

Traceable 3D X-Ray Measurements ZEISS METROTOM 20 Traceable 3D X-Ray Measurements With a computer tomograph, you can successfully perform measuring and inspection jobs yourself with only one X-ray scan. The standard acceptance test, the precision engineering

More information

International Symposium on Digital Industrial Radiology and Computed Tomography - Mo.2.2

International Symposium on Digital Industrial Radiology and Computed Tomography - Mo.2.2 International Symposium on Digital Industrial Radiology and Computed Tomography - Mo.2.2 Accuracy Evaluation and Exploration of Measurement Uncertainty for Exact Helical Cone Beam Reconstruction Using

More information

Advanced Reconstruction Techniques Applied to an On-Site CT System

Advanced Reconstruction Techniques Applied to an On-Site CT System 2nd International Symposium on NDT in Aerospace 2010 - We.1.A.4 Advanced Reconstruction Techniques Applied to an On-Site CT System Jonathan HESS, Markus EBERHORN, Markus HOFMANN, Maik LUXA Fraunhofer Development

More information

Coordinate Measuring Machines with Computed Tomography

Coordinate Measuring Machines with Computed Tomography Always a Step Ahead with Quality Coordinate Measuring Machines with Computed Tomography Multisensor Coordinate Measuring Machines with Computed Tomography Computed Tomography in Coordinate Measuring Machines

More information

Registration concepts for the just-in-time artefact correction by means of virtual computed tomography

Registration concepts for the just-in-time artefact correction by means of virtual computed tomography DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Registration concepts for the just-in-time artefact correction by means of virtual

More information

Translational Computed Tomography: A New Data Acquisition Scheme

Translational Computed Tomography: A New Data Acquisition Scheme 2nd International Symposium on NDT in Aerospace 2010 - We.1.A.3 Translational Computed Tomography: A New Data Acquisition Scheme Theobald FUCHS 1, Tobias SCHÖN 2, Randolf HANKE 3 1 Fraunhofer Development

More information

ASSESSMENT OF MEASUREMENT UNCERTAINTY CAUSED IN THE PREPARATION OF MEASUREMENTS USING COMPUTED TOMOGRAPHY

ASSESSMENT OF MEASUREMENT UNCERTAINTY CAUSED IN THE PREPARATION OF MEASUREMENTS USING COMPUTED TOMOGRAPHY XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 2009, Lisbon, Portugal ASSESSMENT OF MEASUREMENT UNCERTAINTY CAUSED IN THE PREPARATION OF MEASUREMENTS USING COMPUTED TOMOGRAPHY

More information

Theoretically-exact CT-reconstruction from experimental data

Theoretically-exact CT-reconstruction from experimental data Theoretically-exact CT-reconstruction from experimental data T Varslot, A Kingston, G Myers, A Sheppard Dept. Applied Mathematics Research School of Physics and Engineering Australian National University

More information

Investigation on the influence of image quality in X-ray CT metrology

Investigation on the influence of image quality in X-ray CT metrology Investigation on the influence of image quality in X-ray CT metrology Pavel Müller 1, Jochen Hiller 1, Angela Cantatore 1, Markus Bartscher 2, Leonardo De Chiffre 1 1 Technical University of Denmark, Department

More information

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Improvement of Efficiency and Flexibility in Multi-slice Helical CT J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): 408 412 DOI: 10.1007/s12204-008-0408-x Improvement of Efficiency and Flexibility in Multi-slice Helical CT SUN Wen-wu 1 ( ), CHEN Si-ping 2 ( ), ZHUANG Tian-ge

More information

Industrial Computed Tomography Innovations

Industrial Computed Tomography Innovations GE Inspection Technologies Industrial Computed Tomography Innovations Premium performance for premium quality and speed. gemeasurement.com/ct Premium CT technologies. Faster than ever before. Every industrial

More information

Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography

Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography Natalia Grozmani 1, Andrea Buratti 1, Robert H. Schmitt 1 More info about this article:

More information

Frequency-based method to optimize the number of projections for industrial computed tomography

Frequency-based method to optimize the number of projections for industrial computed tomography More Info at Open Access Database www.ndt.net/?id=18739 Frequency-based method to optimize the number of projections for industrial computed tomography Andrea Buratti 1, Soufian Ben Achour 1, Christopher

More information

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh et al. : Proceedings of the National Seminar & Exhibition on Non-Destructive Evaluation DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh V Acharya, Umesh Kumar, Gursharan

More information

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS 23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS P. Helberg 1,

More information

Characterization of microshells experimented on Laser Megajoule using X-Ray tomography

Characterization of microshells experimented on Laser Megajoule using X-Ray tomography Characterization of microshells experimented on Laser Megajoule using X-Ray tomography More info about this article: http://www.ndt.net/?id=20881 Alexandre Choux, Lise Barnouin, Ludovic Reverdy, Marc Theobald

More information

Wieblinger Weg 92a, Heidelberg, Germany, Phone: , Fax: ;

Wieblinger Weg 92a, Heidelberg, Germany, Phone: , Fax: ; HOW INDUSTRIAL COMPUTER TOMOGRAPHY ACCELERATES PRODUCT DEVELOPMENT IN THE LIGHT METAL CASTING AND INJECTION MOULDING INDUSTRY. Christof REINHART 1, Christoph POLIWODA 1, Thomas GÜNTHER 1 1 Volume Graphics

More information

Quality control phantoms and protocol for a tomography system

Quality control phantoms and protocol for a tomography system Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing

More information

THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS

THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS A. Campos a*, A. Zorrilla a, F. Lasagni a a Materials & Processes dept., Center for Advanced Aerospace Technologies

More information

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Wenjuan Sun and Stephen Brown 10 th May 2016 1 Possible factors influencing XCT measurements Components Influencing variables Possible

More information

On the impact of probing form error on form measurement in Computed Tomography

On the impact of probing form error on form measurement in Computed Tomography On the impact of probing form error on form measurement in Computed Tomography Jan Lasson Andreasen 1, Filippo Montanari 2, Leonardo De Chiffre 2 More info about this article: http://www.ndt.net/?id=23729

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Scaling Calibration in the ATRACT Algorithm

Scaling Calibration in the ATRACT Algorithm Scaling Calibration in the ATRACT Algorithm Yan Xia 1, Andreas Maier 1, Frank Dennerlein 2, Hannes G. Hofmann 1, Joachim Hornegger 1,3 1 Pattern Recognition Lab (LME), Friedrich-Alexander-University Erlangen-Nuremberg,

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

TESTING OF THE CIRCLE AND LINE ALGORITHM IN THE SETTING OF MICRO-CT

TESTING OF THE CIRCLE AND LINE ALGORITHM IN THE SETTING OF MICRO-CT SCA2016-080 1/7 TESTING OF THE CIRCLE AND LINE ALGORITHM IN THE SETTING OF MICRO-CT Alexander Katsevich 1, 2 and Michael Frenkel 1 1 itomography Corp., 2 University of Central Florida (UCF) This paper

More information

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China 7th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China Image Registration Combining Digital Radiography and Computer-Tomography Image Data Frank HEROLD YXLON International X-Ray

More information

Feldkamp-type image reconstruction from equiangular data

Feldkamp-type image reconstruction from equiangular data Journal of X-Ray Science and Technology 9 (2001) 113 120 113 IOS Press Feldkamp-type image reconstruction from equiangular data Ben Wang a, Hong Liu b, Shiying Zhao c and Ge Wang d a Department of Elec.

More information

Uncertainty Evaluation of Pore Analysis for Additively Manufactured Parts using Cross Sections

Uncertainty Evaluation of Pore Analysis for Additively Manufactured Parts using Cross Sections Uncertainty Evaluation of Pore Analysis for Additively Manufactured Parts using Cross Sections More info about this article: http://www.ndt.net/?id=23659 Abstract Leonard Schild 1, Manuel Fülling 1, Benjamin

More information

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction 2D Fan Beam Reconstruction 3D Cone Beam Reconstruction Mario Koerner March 17, 2006 1 2D Fan Beam Reconstruction Two-dimensional objects can be reconstructed from projections that were acquired using parallel

More information

A cone-beam CT geometry correction method based on intentional misalignments to render the projection images correctable

A cone-beam CT geometry correction method based on intentional misalignments to render the projection images correctable A cone-beam CT geometry correction method based on intentional misalignments to render the projection images correctable Felix Meli, Benjamin A. Bircher, Sarah Blankenberger, Alain Küng, Rudolf Thalmann

More information

Parameter Dependent Thresholding for Dimensional X-ray Computed Tomography

Parameter Dependent Thresholding for Dimensional X-ray Computed Tomography International Symposium on Digital Industrial Radiology and Computed Tomography - Poster 12 Parameter Dependent Thresholding for Dimensional X-ray Computed Tomography Kim KIEKENS *, Ye TAN *, Jean-Pierre

More information

Adapted acquisition trajectory and iterative reconstruction for few-views CT inspection

Adapted acquisition trajectory and iterative reconstruction for few-views CT inspection Adapted acquisition trajectory and iterative reconstruction for few-views CT inspection Caroline Vienne 1, Marius Costin 1 More info about this article: http://www.ndt.net/?id=21917 1 CEA, LIST, Département

More information

Efficient 3D Crease Point Extraction from 2D Crease Pixels of Sinogram

Efficient 3D Crease Point Extraction from 2D Crease Pixels of Sinogram Efficient 3D Crease Point Extraction from 2D Crease Pixels of Sinogram Ryo Jinnouchi 1, Yutaka Ohtake 1, Hiromasa Suzuki 1, Yukie Nagai 1 1 The University of Tokyo, Japan, e-mail: {jinnouchi, yu-ohtake,

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

An approximate cone beam reconstruction algorithm for gantry-tilted CT

An approximate cone beam reconstruction algorithm for gantry-tilted CT An approximate cone beam reconstruction algorithm for gantry-tilted CT Ming Yan a, Cishen Zhang ab, Hongzhu Liang a a School of Electrical & Electronic Engineering, Nanyang Technological University, Singapore;

More information

CIVA Computed Tomography Modeling

CIVA Computed Tomography Modeling CIVA Computed Tomography Modeling R. FERNANDEZ, EXTENDE, France S. LEGOUPIL, M. COSTIN, D. TISSEUR, A. LEVEQUE, CEA-LIST, France page 1 Summary Context From CIVA RT to CIVA CT Reconstruction Methods Applications

More information

X-ray Industrial Computed Laminography (ICL) Simulation Study of Planar Objects: Optimization of Laminographic Angle

X-ray Industrial Computed Laminography (ICL) Simulation Study of Planar Objects: Optimization of Laminographic Angle More info about this article: http://www.ndt.net/?id=21086 X-ray Industrial Computed Laminography (ICL) Simulation Study of Planar Objects: Optimization of Laminographic Angle Lakshminarayana Yenumula

More information

Evaluation strategies in CT scanning

Evaluation strategies in CT scanning Downloaded from orbit.dtu.dk on: Dec 20, 2017 Evaluation strategies in CT scanning Hiller, Jochen Publication date: 2012 Document Version Publisher's PDF, also known as Version of record Link back to DTU

More information

Implementation of a backprojection algorithm on CELL

Implementation of a backprojection algorithm on CELL Implementation of a backprojection algorithm on CELL Mario Koerner March 17, 2006 1 Introduction X-ray imaging is one of the most important imaging technologies in medical applications. It allows to look

More information

STUDY ON PARAMETER VARIATION OF AN INDUSTRIAL COMPUTED TOMOGRAPHY SIMULATION TOOL CONCERNING DIMENSIONAL MEASUREMENT DEVIATIONS

STUDY ON PARAMETER VARIATION OF AN INDUSTRIAL COMPUTED TOMOGRAPHY SIMULATION TOOL CONCERNING DIMENSIONAL MEASUREMENT DEVIATIONS STUDY ON PARAMETER VARIATION OF AN INDUSTRIAL COMPUTED TOMOGRAPHY SIMULATION TOOL CONCERNING DIMENSIONAL MEASUREMENT DEVIATIONS Michael REITER a, Christoph HEINZL a, Dietmar SALABERGER a, Daniel WEISS

More information

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY David K. Lehmann 1, Kathleen Brockdorf 1 and Dirk Neuber 2 1 phoenix x-ray Systems + Services Inc. St. Petersburg, FL, USA 2 phoenix x-ray Systems + Services

More information

In-Situ Surface Roughness Measurement of Laser Beam Melted Parts a Feasibility Study of Layer Image Analysis

In-Situ Surface Roughness Measurement of Laser Beam Melted Parts a Feasibility Study of Layer Image Analysis In-Situ Surface Roughness Measurement of Laser Beam Melted Parts a Feasibility Study of Layer Image Analysis Joschka zur Jacobsmühlen 1, Stefan Kleszczynski 2, Alexander Ladewig 3, Gerd Witt 2 and Dorit

More information

A RADIAL WHITE LIGHT INTERFEROMETER FOR MEASUREMENT OF CYLINDRICAL GEOMETRIES

A RADIAL WHITE LIGHT INTERFEROMETER FOR MEASUREMENT OF CYLINDRICAL GEOMETRIES A RADIAL WHITE LIGHT INTERFEROMETER FOR MEASUREMENT OF CYLINDRICAL GEOMETRIES Andre R. Sousa 1 ; Armando Albertazzi 2 ; Alex Dal Pont 3 CEFET/SC Federal Center for Technological Education of Sta. Catarina

More information

High-resolution X-ray CT Inspection of Honeycomb Composites Using Planar Computed Tomography Technology

High-resolution X-ray CT Inspection of Honeycomb Composites Using Planar Computed Tomography Technology 2nd International Symposium on NDT in Aerospace 2010 - We.4.B.4 High-resolution X-ray CT Inspection of Honeycomb Composites Using Planar Computed Tomography Technology Tong LIU, Andrew A. MALCOLM, and

More information

Advanced Image Reconstruction Methods for Photoacoustic Tomography

Advanced Image Reconstruction Methods for Photoacoustic Tomography Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic

More information

3D Computed Tomography (CT) Its Application to Aerospace Industry

3D Computed Tomography (CT) Its Application to Aerospace Industry 3D Computed Tomography (CT) Its Application to Aerospace Industry C. Muralidhar, M. P. Subramanian, V. Ravi Shankar and G. Chandrasekhar Directorate of Non Destructive Evaluation, Defence Research & Development

More information

CT Systems and their standards

CT Systems and their standards CT Systems and their standards Stephen Brown Engineering Measurement 11 th April 2012 Industrial X-ray computed tomography: The future of co-ordinate metrology? Burleigh Court, Loughborough University

More information

Methods for Quantitative Characterization of Large-Scale High Energy Computed Tomography Systems

Methods for Quantitative Characterization of Large-Scale High Energy Computed Tomography Systems More Info at Open Access Database www.ndt.net/?id=16704 Methods for Quantitative Characterization of Large-Scale High Energy Computed Tomography Systems Michael BÖHNEL 1, Andreas GRUBER 2, Nils REIMS 1,

More information

Image Reconstruction 3 Fully 3D Reconstruction

Image Reconstruction 3 Fully 3D Reconstruction Image Reconstruction 3 Fully 3D Reconstruction Thomas Bortfeld Massachusetts General Hospital, Radiation Oncology, HMS HST.S14, February 25, 2013 Thomas Bortfeld (MGH, HMS, Rad. Onc.) Image Reconstruction

More information

Visual Optimality and Stability Analysis of 3DCT Scan Positions

Visual Optimality and Stability Analysis of 3DCT Scan Positions Visual Optimality and Stability Analysis of 3DCT Scan Positions Artem Amirkhanov, Christoph Heinzl, Michael Reiter, and M. Eduard Gröller X-ray source Flatpanel detector X-ray source Flatpanel detector

More information

AUTOMATED HIGH SPEED VOLUME COMPUTED TOMOGRAPHY FOR INLINE QUALITY CONTROL

AUTOMATED HIGH SPEED VOLUME COMPUTED TOMOGRAPHY FOR INLINE QUALITY CONTROL AUTOMATED HIGH SPEED VOLUME COMPUTED TOMOGRAPHY FOR INLINE QUALITY CONTROL R. Hanke 1, A. Kugel 2, and P. Troup 2 1 Fraunhofer-Entwicklungszentrum für Röntgentechnik, Fürth, Germany; 2 Austria Aluguss

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

Investigation of the kinematic system of a 450 kv CT scanner and its influence on dimensional CT metrology applications

Investigation of the kinematic system of a 450 kv CT scanner and its influence on dimensional CT metrology applications Investigation of the kinematic system of a 450 kv CT scanner and its influence on dimensional CT metrology applications Frank Welkenhuyzen 1, Bart Boeckmans 1, Ye Tan 1,2, Kim Kiekens 1,2, Wim Dewulf 1,2,

More information

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon WHITE PAPER Introduction An image guidance system based on Cone Beam CT (CBCT) is included in Leksell

More information

Central Slice Theorem

Central Slice Theorem Central Slice Theorem Incident X-rays y f(x,y) R x r x Detected p(, x ) The thick line is described by xcos +ysin =R Properties of Fourier Transform F [ f ( x a)] F [ f ( x)] e j 2 a Spatial Domain Spatial

More information

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1 Modifications for P551 Fall 2013 Medical Physics Laboratory Introduction Following the introductory lab 0, this lab exercise the student through

More information

UNITEST - A new device for the static and dynamic testing of unconventional machine structures

UNITEST - A new device for the static and dynamic testing of unconventional machine structures UNITEST - A new device for the static and dynamic testing of unconventional machine structures K.E. Meitz Department of Production Engineering, Technical University Graz, Kopernikusgasse 24, A-8010 Graz,

More information

Comparision of contact and contactless measuring methods for form evaluation

Comparision of contact and contactless measuring methods for form evaluation Available online at www.sciencedirect.com Procedia Engineering 48 (2012 ) 273 279 MMaMS 2012 Comparision of contact and contactless measuring methods for form evaluation Peter Ka uch a, Miroslav Dovica

More information

A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY

A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY BACKSCATTER DATA Young S. Ham and C.F. Poranski Materials Chemistry Branch Naval Research Laboratory Washington DC 20375 E.C. Greenawald Geo-Centers,

More information

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction. Mario Koerner

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction. Mario Koerner 2D Fan Beam Reconstruction 3D Cone Beam Reconstruction Mario Koerner Moscow-Bavarian Joint Advanced Student School 2006 March 19 2006 to March 29 2006 Overview 2D Fan Beam Reconstruction Shortscan Reconstruction

More information

Accurate surface extraction on CT volume using analytical gradient of FDK formula

Accurate surface extraction on CT volume using analytical gradient of FDK formula Accurate surface extraction on CT volume using analytical gradient of FDK formula Yukie Nagai 1, Yutaka Ohtake 2, and Hiromasa Suzuki 2 1 Tokyo Metropolitan University, Asahigaoka 6-6, Hino, Tokyo, Japan,

More information

New Method for Master Pattern Tool Correction Based on Computed Tomography Data and Intelligent Reverse Engineering

New Method for Master Pattern Tool Correction Based on Computed Tomography Data and Intelligent Reverse Engineering New Method for Master Pattern Tool Correction Based on Computed Tomography Data and Intelligent Reverse Engineering Dominik Schmid 1, Jochen Merz 1, Jochen Hiller 2 1 Carl Zeiss Holometric Technologies

More information

Recognition and Measurement of Small Defects in ICT Testing

Recognition and Measurement of Small Defects in ICT Testing 19 th World Conference on Non-Destructive Testing 2016 Recognition and Measurement of Small Defects in ICT Testing Guo ZHIMIN, Ni PEIJUN, Zhang WEIGUO, Qi ZICHENG Inner Mongolia Metallic Materials Research

More information

The new generation of industrial computed tomography. The Desktop CT exact S

The new generation of industrial computed tomography. The Desktop CT exact S The new generation of industrial computed tomography The Desktop CT S Volume scanning technology Innovation with a family tradition Founded in 1968, the family-owned, WEN- ZEL Metrology Group is one of

More information

Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage. 29 April - 3 May, 2013

Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage. 29 April - 3 May, 2013 2455-5 Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage 29 April - 3 May, 2013 Lecture NoteBasic principles of X-ray Computed Tomography Diego Dreossi Elettra, Trieste

More information

Multi-sensor measuring technology. O-INSPECT The best of optical and contact measuring technology for true 3D measurements.

Multi-sensor measuring technology. O-INSPECT The best of optical and contact measuring technology for true 3D measurements. Multi-sensor measuring technology O-INSPECT The best of optical and contact measuring technology for true 3D measurements. 2 // multifunctionality made BY CarL Zeiss The moment you realize that new requirements

More information

Two Local FBP Algorithms for Helical Cone-beam Computed Tomography

Two Local FBP Algorithms for Helical Cone-beam Computed Tomography Digital Industrial Radiology and Computed Tomography (DIR 215) 22-25 June 215, Belgium, Ghent - www.ndt.net/app.dir215 More Info at Open Access Database www.ndt.net/?id=187 Two Local FBP Algorithms for

More information

Fiber Composite Material Analysis in Aerospace Using CT Data

Fiber Composite Material Analysis in Aerospace Using CT Data 4th International Symposium on NDT in Aerospace 2012 - We.2.A.3 Fiber Composite Material Analysis in Aerospace Using CT Data Dr. Tobias DIERIG, Benjamin BECKER, Christof REINHART, Thomas GÜNTHER Volume

More information

Non-Stationary CT Image Noise Spectrum Analysis

Non-Stationary CT Image Noise Spectrum Analysis Non-Stationary CT Image Noise Spectrum Analysis Michael Balda, Björn J. Heismann,, Joachim Hornegger Pattern Recognition Lab, Friedrich-Alexander-Universität Erlangen Siemens Healthcare, Erlangen michael.balda@informatik.uni-erlangen.de

More information

Non-Destructive Failure Analysis and Measurement for Molded Devices and Complex Assemblies with X-ray CT and 3D Image Processing Techniques

Non-Destructive Failure Analysis and Measurement for Molded Devices and Complex Assemblies with X-ray CT and 3D Image Processing Techniques SINCE2013 Singapore International NDT Conference & Exhibition 2013, 19-20 July 2013 Non-Destructive Failure Analysis and Measurement for Molded Devices and Complex Assemblies with X-ray CT and 3D Image

More information

The PolyCT Increasing the sample throughput using multiple rotation axes

The PolyCT Increasing the sample throughput using multiple rotation axes The PolyCT Increasing the sample throughput using multiple rotation axes More info about this article: http://www.ndt.net/?id=21946 1 Dr. Nicolas Thiemeyer, Johannes Leisner 2, Michael Salamon 2, Carina

More information

Japan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field. Kouichi Inagaki ICC / IHI Corporation

Japan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field. Kouichi Inagaki ICC / IHI Corporation Japan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field Kouichi Inagaki ICC / IHI Corporation 13 th WORLD CONFERENCE ON INVESTMENT CASTING Paper: T3 Copyright

More information

ULTRASONIC TESTING AND FLAW CHARACTERIZATION. Alex KARPELSON Kinectrics Inc., Toronto, Canada

ULTRASONIC TESTING AND FLAW CHARACTERIZATION. Alex KARPELSON Kinectrics Inc., Toronto, Canada ULTRASONIC TESTING AND FLAW CHARACTERIZATION Alex KARPELSON Kinectrics Inc., Toronto, Canada 1. Introduction Ultrasonic Testing (UT) is a commonly used inspection method. Various techniques are employed

More information

CIRCULAR scanning trajectory has been widely used in

CIRCULAR scanning trajectory has been widely used in IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 25, NO. 7, JULY 2006 869 Region of Interest Reconstruction From Truncated Data in Circular Cone-Beam CT Lifeng Yu, Yu Zou, Emil Y. Sidky, Charles A. Pelizzari,

More information

Reconstruction Methods for Coplanar Translational Laminography Applications

Reconstruction Methods for Coplanar Translational Laminography Applications Reconstruction Methods for Coplanar Translational Laminography Applications U. EWERT, K.-U. THIESSENHUSEN, A. DERESCH, C. BELLON, S. HOHENDORF, S. KOLKOORI, N. WROBEL, B. REDMER, M. TSCHAIKNER, BAM, Berlin

More information

Industrial Metrology. Multisensor Measuring Machines O-INSPECT 322/442

Industrial Metrology. Multisensor Measuring Machines O-INSPECT 322/442 Industrial Metrology Multisensor Measuring Machines O-INSPECT 322/442 The moment the answer is right in front of your eyes. This is the moment we work for. // Certainty Made By Zeiss 2 Table of Contents

More information

MULTIPLE-SENSOR INTEGRATION FOR EFFICIENT REVERSE ENGINEERING OF GEOMETRY

MULTIPLE-SENSOR INTEGRATION FOR EFFICIENT REVERSE ENGINEERING OF GEOMETRY Proceedings of the 11 th International Conference on Manufacturing Research (ICMR2013) MULTIPLE-SENSOR INTEGRATION FOR EFFICIENT REVERSE ENGINEERING OF GEOMETRY Feng Li, Andrew Longstaff, Simon Fletcher,

More information

Suitability of a new alignment correction method for industrial CT

Suitability of a new alignment correction method for industrial CT Suitability of a new alignment correction method for industrial CT Matthias Elter 1, Nicole Maass 1, Peter Koch 2 1 Siemens AG, Healthcare Sector, Erlangen, Germany, e-mail: matthias.elter@siemens.com,

More information

FDK Half-Scan with a Heuristic Weighting Scheme on a Flat Panel Detector-Based Cone Beam CT (FDKHSCW)

FDK Half-Scan with a Heuristic Weighting Scheme on a Flat Panel Detector-Based Cone Beam CT (FDKHSCW) Biomedical Imaging Volume 26, Article ID 83983, Pages 8 DOI.55/IJBI/26/83983 FDK Half-Scan with a Heuristic Weighting Scheme on a Flat Panel Detector-Based Cone Beam CT (FDKHSCW) Dong Yang and Ruola Ning

More information

Quality Digest Magazine 1 of 10 DLP-Based Projected Fringe Measuring Technology This technology eases the burden of optical 3-D digitization and coordinate measurement of parts. by M. Kaestner, G. Frankowski,

More information

The n-pi-method for Helical Cone-Beam CT

The n-pi-method for Helical Cone-Beam CT 848 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 19, NO. 9, SEPTEMBER 2000 The n-pi-method for Helical Cone-Beam CT R. Proksa*, Th. Köhler, M. Grass, and J. Timmer Abstract A new class of acquisition schemes

More information

Industrial Computer Tomography for Dimensional Metrology: Overview of Influence Factors and Improvement Strategies

Industrial Computer Tomography for Dimensional Metrology: Overview of Influence Factors and Improvement Strategies 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 Industrial Computer Tomography for Dimensional Metrology: Overview of Influence Factors and

More information

2. Equipment, specimens and inspection procedures

2. Equipment, specimens and inspection procedures 4th International CANDU In-service Inspection Workshop and NDT in Canada 2012 Conference, 2012 June 18-21, Toronto, Ontario Improved Flaw Characterization and Sizing in Pressure Tubes Alex KARPELSON Kinectrics

More information

AUTOMATED 4 AXIS ADAYfIVE SCANNING WITH THE DIGIBOTICS LASER DIGITIZER

AUTOMATED 4 AXIS ADAYfIVE SCANNING WITH THE DIGIBOTICS LASER DIGITIZER AUTOMATED 4 AXIS ADAYfIVE SCANNING WITH THE DIGIBOTICS LASER DIGITIZER INTRODUCTION The DIGIBOT 3D Laser Digitizer is a high performance 3D input device which combines laser ranging technology, personal

More information

Analysis of Combined Probing Measurement Error and Length Measurement Error Test for Acceptance Testing in Dimensional Computed Tomography

Analysis of Combined Probing Measurement Error and Length Measurement Error Test for Acceptance Testing in Dimensional Computed Tomography Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18038 Analysis of Combined Probing

More information

Optimization of Cone Beam CT Reconstruction Algorithm Based on CUDA

Optimization of Cone Beam CT Reconstruction Algorithm Based on CUDA Sensors & Transducers 2013 by IFSA http://www.sensorsportal.com Optimization of Cone Beam CT Reconstruction Algorithm Based on CUDA 1 Wang LI-Fang, 2 Zhang Shu-Hai 1 School of Electronics and Computer

More information

Advanced automated High Speed Computed Tomography for production process control. DGZfP 2011, Bremen

Advanced automated High Speed Computed Tomography for production process control. DGZfP 2011, Bremen DGZfP-Jahrestagung 2011 - Mo.3.B.3 Advanced automated High Speed Computed Tomography for production process control DGZfP 2011, Bremen Dr. Ingo Stuke* and Dr. Oliver Brunke** GE Sensing & Inspection Technologies

More information

LABORATORY SYSTEM FOR X-RAY NANOTOMOGRAPHY

LABORATORY SYSTEM FOR X-RAY NANOTOMOGRAPHY 79 LABORATORY SYSTEM FOR X-RAY NANOTOMOGRAPHY Alexander Sasov, SkyScan, Vluchtenburgstraat 3, Aartselaar B2630, Belgium, www.skyscan.be. ABSTRACT Using advanced X-ray technologies and X-ray scattering

More information

Exterior Orientation Parameters

Exterior Orientation Parameters Exterior Orientation Parameters PERS 12/2001 pp 1321-1332 Karsten Jacobsen, Institute for Photogrammetry and GeoInformation, University of Hannover, Germany The georeference of any photogrammetric product

More information

A new concept for high-speed atline and inline CT for up to 100% mass production process control allowing both 3D metrology and failure analysis

A new concept for high-speed atline and inline CT for up to 100% mass production process control allowing both 3D metrology and failure analysis A new concept for high-speed atline and inline CT for up to 100% mass production process control allowing both 3D metrology and failure analysis Oliver Brunke 1, Ferdinand Hansen 2, Ingo Stuke 3, Friedrich

More information

On the Characteristics of Helical 3-D X-ray Dark-field Imaging

On the Characteristics of Helical 3-D X-ray Dark-field Imaging On the Characteristics of Helical 3-D X-ray Dark-field Imaging Lina Felsner 1, Shiyang Hu 1, Veronika Ludwig 2, Gisela Anton 2, Andreas Maier 1, Christian Riess 1 1 Pattern Recognition Lab, Computer Science,

More information

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam Home Search Collections Journals About Contact us My IOPscience Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam projections This content has been downloaded from IOPscience.

More information