CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI
|
|
- Melinda Oliver
- 5 years ago
- Views:
Transcription
1 Figure 1. In the brain, the gray matter has substantially more blood vessels and capillaries than white matter. The magnified image on the right displays the rich vasculature in gray matter forming porous, sponge-like structures. The high concentration of blood in the gray matter vasculature makes the gray matter highly sensitive to the blood oxygen leveldependent (BOLD) effect.
2 Figure 2. The major air-containing structures included in the model are the frontal sinus (turquoise), the sphenoid sinus (pink), the maxillary sinuses (blue), mastoid air cells (red), nasal and oral cavities, and trachea (yellow). Calculated (top) and measured (bottom) B 0 distributions demonstrate local gradients in the sagittal (a), coronal (b), and axial (c) planes. Magnetic field deviations are color coded in parts per million (ppm). Reprinted with permission from [30].
3 Figure 3. Calculated B 0 distributions inside the human head model demonstrate the remarkable distortion by the air-containing structures inside the human head. Illustration courtesy of Drs. Shizhe Li, C.M. Collins, Q.X. Yang, and M.B. Smith. Figure 4. Axial plane view of numerically calculated B 0 field magnitudes in ppm deviation from applied field. The gray scale is set to reveal the B 0 deviation due to the susceptibility difference between gray and white matter in the brain. Reprinted with permission from [35].
4 Figure 5. Timing diagram of 2D gradient echo sequence for a voxel. A, slice selection gradient; B, slice rewinding gradient. During TE the local gradient in the slice selection dimension, G z l, causes intravoxel phase dispersion, which results in signal loss in GE images. A compensation gradient, G c, or a tailored RF pulse, α[φ 0 (z)] can be applied to cancel the phase dispersion by G z l. Figure 6. The GESEPI image sets of the human brain obtained at 3.0 Tesla before (2D-FT data set) and after (3D-FT) a Fourier transform along the slice direction with respect to k z by the compensation gradient. The 2D-FT image set was obtained by sequentially incrementing G c to obtain N = 16 frames. Note the removal of the dark artifacts in the images from 4 6 sub-slices (ca. 1 mm). Reprinted with permission from [46]).
5 Figure 7. A coronal view of the phantom and images of the axial slice along the midline between the two spheres with the GE and GESEPI methods. The phantom consists of two air-filled spheres positioned inside a gelatin-filled cylindrical container with its longitudinal axis perpendicular to B 0. The voxels P, Q, and R indicated in the GE image are labeled P', Q', and R' in the corresponding GESEPI images. The phantom images were obtained with a flip angle of 25º, TR/TE = 80 ms/25 ms, FOV = cm 2, image matrix = , slice thickness = 5 mm. Adapted with permission from [46]. Figure 8. Serial signal intensity plots from the representative voxels P, Q, and R (see Figure 7) in the 2D-FT data set, and from the corresponding voxels P', Q', and R' in the 3D-FT data set. The image number axis is shown above each plot, corresponding to either k z (2D- FT, top row) or position z along the slice direction (3D-FT, bottom row). The dashed lines in the top row plots indicate k = 0, where 2D images were acquired. Reprinted with permission from [46].
6 Figure 9. GESEPI effectively removed signal-loss artifacts in heavily T 2 *-weighted images at 3.0 and 7.0 Tesla. All the images have the same slice thickness 5 mm. The GESEPI images were obtained with 8 G c steps. The intensities of the globus pallidus (GP) and substantia nigra (SN) are obscured in the GE images and enhanced in the GESEPI images. Note that the signal nonuniformity in the GESEPI image at 7.0 Tesla is caused by the B 1 field distortion due to the wave effect. Figure 10. T 2 *-weighted images of a mouse brain at 14.0 Tesla. By removing the artifacts in the GE image, the GESEPI image improves contrast resolution and displays anatomic details not observable in spin-echo images. Reprinted with permission from [50].
7 Figure 11. GE (top row) and GESEPI (bottom row) images of an immature rat brain as a function of TE taken on a 9.4T system. TR = 80 ms, matrix = , FOV = mm 2, slice thickness = 1 mm. Reprinted with permission from [ 46]. Figure 12. Average R 2 * map of the human brain with GESEPI.
8 Figure 13. The voxel signal intensity from the excited slice of the GESEPI data is plotted as a function of k c and acquisition time t along the trajectory (a) and its Fourier transform (b). The plots were calculated using Eqs. (13) and (14) with T 2 * = 30 ms, G z l = 0.5 ppm/cm, z 0 = 5 mm, and B 0 = 3.0 Tesla. The plots in (c) are standard 2D voxel signal intensity along the trajectory with k l = 0 (Plot I), k l 0 (Plot II), and GESEPI intensity with k l 0 (Plot III), respectively. The Plot II in (c), obtained by intersecting the two-dimensional plot in (a) with k z = 0, shows a sinc-function modulation of T 2 * decay. Plot III, obtained by intersecting the plot in (b) with z = 0, coincides with Plot I of normal T 2 * relaxation. Figure 14. Top row: the signal intensities acquired without phase-encoding in the left right direction. Center row: image intensity plot, representing point-spread function (PSF) profile in phase-encoding direction. Bottom row: images acquired with phaseencoding, demonstrating the reduction of blurring artifacts in EPI using the GESEPI method. The quasi-point source phantom for axial images consists of two 5-mm diameter tubes filled with gelatin, with a 1.8-mm distance between the tubes. The tubes are aligned with the B 0 field. The image intensities are adjusted to facilitate visualization of PSF. Adapted with permission from [71].
9 Figure 15. Spin-echo EPI images acquired without (a) and with SENSE with a reduction factor r = 3 (b) from an inferior brain slice. In spin echo images, the signal-loss artifact is minimal so that the geometric distortion due to the in-plane gradient can be demonstrated clearly. In order to highlight the brain geometry, a diffusion gradient was applied to reduce the CSF signals. Both images were acquired from a Philips Intera 3.0T MRI system with the same acquisition parameters. Reprinted with permission from [75].
10 Figure 16. The EPI images were acquired from the same human brain volume with SENSE (r = 3) using the GE (a), GESEPI (b), and SE (c) methods. The typical signal-loss artifact in the GE images (a) in the inferior frontal and temporal brain regions indicated by the arrows is reduced significantly with GESEPI (b). Reprinted with permission from [71].
11 Figure 17. Relationship of T 2 * decay with Forward and Reversed spiral acquisition. Illustration provided courtesy of Dr. V. Andrew Stenger. Figure 18. Simulated k-space signals of a point source with reversed and forward spiral acquisitions (a) and the corresponding image profiles (b). Illustration provided courtesy of Dr. V. Andrew Stenger.
12 Figure 19. Simulated (a) and experimental (b) image profiles with different data acquisition methods in the absence and presence of local gradients. Illustration provided courtesy of Dr. V. Andrew Stenger. Figure 20. Spiral images taken from the inferior brain region show remarkable signal-loss artifact. As indicated by the arrows, the artifact is mitigated by reversed spiral and removed when GESEPI is incorporated into the reversed spiral technique.
13 Figure 21. Olfactory fmri activation at orbitofrontal and olfactory tract with GESEPIreversed-spiral imaging.
Functional MRI. Jerry Allison, Ph. D. Medical College of Georgia
Functional MRI Jerry Allison, Ph. D. Medical College of Georgia BOLD Imaging Technique Blood Oxygen Level Dependent contrast can be used to map brain function Right Hand Motor Task Outline fmri BOLD Contrast
More informationSlide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.
Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing
More informationFollowing on from the two previous chapters, which considered the model of the
Chapter 5 Simulator validation Following on from the two previous chapters, which considered the model of the simulation process and how this model was implemented in software, this chapter is concerned
More informationMRI Physics II: Gradients, Imaging
MRI Physics II: Gradients, Imaging Douglas C., Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor Magnetic Fields in MRI B 0 The main magnetic field. Always on (0.5-7 T) Magnetizes
More informationWhite Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space
White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different
More informationHST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006
MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.
More informationHST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008
MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.
More informationHST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008
MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.
More informationDiffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford
Diffusion MRI Acquisition Karla Miller FMRIB Centre, University of Oxford karla@fmrib.ox.ac.uk Diffusion Imaging How is diffusion weighting achieved? How is the image acquired? What are the limitations,
More information(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions
EE225E/BIOE265 Spring 2013 Principles of MRI Miki Lustig Assignment 8 Solutions 1. Nishimura 7.1 P)jc%z 5 ivcr3. 1. I Due Wednesday April 10th, 2013 (a Scrhon5 R2iwd b 0 ZOms Xn,s r cx > qs 4-4 8ni6 4
More informationRole of Parallel Imaging in High Field Functional MRI
Role of Parallel Imaging in High Field Functional MRI Douglas C. Noll & Bradley P. Sutton Department of Biomedical Engineering, University of Michigan Supported by NIH Grant DA15410 & The Whitaker Foundation
More informationACQUIRING AND PROCESSING SUSCEPTIBILITY WEIGHTED IMAGING (SWI) DATA ON GE 3.0T
ACQUIRING AND PROCESSING SUSCEPTIBILITY WEIGHTED IMAGING (SWI) DATA ON GE 3.0T Revision date: 12/13/2010 Overview Susceptibility Weighted Imaging (SWI) is a relatively new data acquisition and processing
More informationCOBRE Scan Information
COBRE Scan Information Below is more information on the directory structure for the COBRE imaging data. Also below are the imaging parameters for each series. Directory structure: var/www/html/dropbox/1139_anonymized/human:
More informationCORRECTION OF IMAGE DISTORTION IN ECHO PLANAR IMAGE SERIES USING PHASE AND INTENSITY. Ning Xu. Dissertation. Submitted to the Faculty of the
CORRECTION OF IMAGE DISTORTION IN ECHO PLANAR IMAGE SERIES USING PHASE AND INTENSITY By Ning Xu Dissertation Submitted to the Faculty of the Graduate School of Vanderbilt University in partial fulfillment
More informationMR Advance Techniques. Vascular Imaging. Class III
MR Advance Techniques Vascular Imaging Class III 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology
More informationLucy Phantom MR Grid Evaluation
Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is
More informationLab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street
Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street MRI is located in the sub basement of CC wing. From Queen or Victoria, follow the baby blue arrows and ride the CC south
More informationThis exercise uses one anatomical data set (ANAT1) and two functional data sets (FUNC1 and FUNC2).
Exploring Brain Anatomy This week s exercises will let you explore the anatomical organization of the brain to learn some of its basic properties, as well as the location of different structures. The human
More informationA more accurate account of the effect of k-space sampling and signal decay on the effective spatial resolution in functional MRI
A more accurate account of the effect of k-space sampling and signal decay on the effective spatial resolution in functional MRI Denis Chaimow 1 and Amir Shmuel 1,2 1 Centre for Magnetic Resonance Research
More informationsurface Image reconstruction: 2D Fourier Transform
2/1/217 Chapter 2-3 K-space Intro to k-space sampling (chap 3) Frequenc encoding and Discrete sampling (chap 2) Point Spread Function K-space properties K-space sampling principles (chap 3) Basic Contrast
More informationA Transmission Line Matrix Model for Shielding Effects in Stents
A Transmission Line Matrix Model for hielding Effects in tents Razvan Ciocan (1), Nathan Ida (2) (1) Clemson University Physics and Astronomy Department Clemson University, C 29634-0978 ciocan@clemon.edu
More informationSupplementary Figure 1
Supplementary Figure 1 BOLD and CBV functional maps showing EPI versus line-scanning FLASH fmri. A. Colored BOLD and CBV functional maps are shown in the highlighted window (green frame) of the raw EPI
More informationMRI. When to use What sequences. Outline 2012/09/19. Sequence: Definition. Basic Principles: Step 2. Basic Principles: Step 1. Govind Chavhan, MD
MRI When to use What sequences Govind Chavhan, MD Assistant Professor and Staff Radiologist The Hospital For Sick Children, Toronto Planning Acquisition Post processing Interpretation Patient history and
More informationFmri Spatial Processing
Educational Course: Fmri Spatial Processing Ray Razlighi Jun. 8, 2014 Spatial Processing Spatial Re-alignment Geometric distortion correction Spatial Normalization Smoothing Why, When, How, Which Why is
More informationMB-EPI PCASL. Release Notes for Version February 2015
MB-EPI PCASL Release Notes for Version 1.0 20 February 2015 1 Background High-resolution arterial spin labeling (ASL) imaging is highly desirable in both neuroscience research and clinical applications
More informationModule 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution.
MRES 7005 - Fast Imaging Techniques Module 5: Dynamic Imaging and Phase Sharing (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review Improving Temporal Resolution True-FISP (I) True-FISP (II) Keyhole
More informationSIEMENS MAGNETOM Verio syngo MR B17
\\USER\Dr. Behrmann\routine\Ilan\ep2d_bold_PMU_resting TA: 8:06 PAT: Voxel size: 3.03.03.0 mm Rel. SNR: 1.00 USER: ep2d_bold_pmu Properties Special sat. Prio Recon System Before measurement Body After
More informationHST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008
MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.
More informationScan Acceleration with Rapid Gradient-Echo
Scan Acceleration with Rapid Gradient-Echo Hsiao-Wen Chung ( 鍾孝文 ), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ. Dept. Radiology, Tri-Service General Hospital 1 of 214 The Need
More informationParallel Imaging. Marcin.
Parallel Imaging Marcin m.jankiewicz@gmail.com Parallel Imaging initial thoughts Over the last 15 years, great progress in the development of pmri methods has taken place, thereby producing a multitude
More informationSupplementary Information
Supplementary Information Magnetic resonance imaging reveals functional anatomy and biomechanics of a living dragon tree Linnea Hesse 1,2,*, Tom Masselter 1,2,3, Jochen Leupold 4, Nils Spengler 5, Thomas
More informationMagnetic Resonance Elastography (MRE) of Liver Disease
Magnetic Resonance Elastography (MRE) of Liver Disease Authored by: Jennifer Dolan Fox, PhD VirtualScopics Inc. jennifer_fox@virtualscopics.com 1-585-249-6231 1. Overview of MRE Imaging MRE is a magnetic
More informationPage 1 of 9. Protocol: adult_other_adni3basichumanprotocol25x_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION
3 Localizer FOV 26.0 Slice Thickness 5.0 Slice Spacing 0.0 Freq 256 Phase 128 3-PLANE 3 Localizer Unswap Phase Correction Gradient Echo Imaging Options Seq, Fast Recon All Images 3 Localizer Pause / SCIC
More informationImaging Notes, Part IV
BME 483 MRI Notes 34 page 1 Imaging Notes, Part IV Slice Selective Excitation The most common approach for dealing with the 3 rd (z) dimension is to use slice selective excitation. This is done by applying
More informationDynamic Contrast enhanced MRA
Dynamic Contrast enhanced MRA Speaker: Yung-Chieh Chang Date : 106.07.22 Department of Radiology, Taichung Veterans General Hospital, Taichung, Taiwan 1 Outline Basic and advanced principles of Diffusion
More informationPartial k-space Reconstruction
Chapter 2 Partial k-space Reconstruction 2.1 Motivation for Partial k- Space Reconstruction a) Magnitude b) Phase In theory, most MRI images depict the spin density as a function of position, and hence
More informationNew Technology Allows Multiple Image Contrasts in a Single Scan
These images were acquired with an investigational device. PD T2 T2 FLAIR T1 MAP T1 FLAIR PSIR T1 New Technology Allows Multiple Image Contrasts in a Single Scan MR exams can be time consuming. A typical
More informationPartial k-space Recconstruction
Partial k-space Recconstruction John Pauly September 29, 2005 1 Motivation for Partial k-space Reconstruction a) Magnitude b) Phase In theory, most MRI images depict the spin density as a function of position,
More informationAcquisition Methods for fmri at 7 Tesla
Acquisition Methods for fmri at 7 Tesla Peter J. Koopmans, Nuffield department for clinical neurosciences, University of Oxford, UK fmri at 7 Tesla is mostly used in the context of high spatial resolution
More informationSIEMENS MAGNETOM Avanto syngo MR B15
\\USER\INVESTIGATORS\Ravi\ADNI-Subject\Localizer TA: 0:10 PAT: Voxel size: 1.9 1.5 8.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer Inline movie
More informationSIEMENS MAGNETOM Skyra syngo MR D13
Page 1 of 8 SIEMENS MAGNETOM Skyra syngo MR D13 \\USER\CIND\StudyProtocols\PTSA\*dm_ep2d_mono70_b0_p2_iso2.0 TA:1:05 PAT:2 Voxel size:2.0 2.0 2.0 mm Rel. SNR:1.00 :epse Properties Routine Prio Recon Load
More informationSources of Distortion in Functional MRI Data
Human Brain Mapping 8:80 85(1999) Sources of Distortion in Functional MRI Data Peter Jezzard* and Stuart Clare FMRIB Centre, Department of Clinical Neurology, University of Oxford, Oxford, UK Abstract:
More informationImage Acquisition Systems
Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS
More informationField Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005
Field Maps John Pauly October 5, 25 The acquisition and reconstruction of frequency, or field, maps is important for both the acquisition of MRI data, and for its reconstruction. Many of the imaging methods
More informationBasic fmri Design and Analysis. Preprocessing
Basic fmri Design and Analysis Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial filtering
More informationM R I Physics Course
M R I Physics Course Multichannel Technology & Parallel Imaging Nathan Yanasak, Ph.D. Jerry Allison Ph.D. Tom Lavin, B.S. Department of Radiology Medical College of Georgia References: 1) The Physics of
More informationTOF-MRA Using Multi-Oblique-Stack Acquisition (MOSA)
JOURNAL OF MAGNETIC RESONANCE IMAGING 26:432 436 (2007) Technical Note TOF-MRA Using Multi-Oblique-Stack Acquisition (MOSA) Ed X. Wu, PhD, 1,2 * Edward S. Hui, BEng, 1,2 and Jerry S. Cheung, BEng 1,2 Purpose:
More informationXI Signal-to-Noise (SNR)
XI Signal-to-Noise (SNR) Lecture notes by Assaf Tal n(t) t. Noise. Characterizing Noise Noise is a random signal that gets added to all of our measurements. In D it looks like this: while in D
More informationSIEMENS MAGNETOM Avanto syngo MR B15
\\USER\INVESTIGATORS\Ravi\ADNI-phantom\QC Phantom-Localizer TA: 0:10 PAT: Voxel size: 1.9 1.5 8.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer
More information3D z-shim Method for Reduction of Susceptibility Effects in BOLD fmri
3D z-shim Method for Reduction of Susceptibility Effects in BOLD fmri Gary H. Glover* Magnetic Resonance in Medicine 42:290 299 (1999) Susceptibility-induced magnetic field gradients (SFGs) perpendicular
More informationPhilips MRI Protocol Dump Created on Comment Software Stream
Page 1 of 5 Philips MRI Protocol Dump Created on 2/17/2011 4:11:01 PM Comment Created by ExamCard_to_XML with inputs: "J:\ADNI GO - ADNI 2 Phantom5.ExamCard" on system (BU SCHOOL OF MEDICINE :: 192.168.71.10)
More informationHead motion in diffusion MRI
Head motion in diffusion MRI Anastasia Yendiki HMS/MGH/MIT Athinoula A. Martinos Center for Biomedical Imaging 11/06/13 Head motion in diffusion MRI 0/33 Diffusion contrast Basic principle of diffusion
More informationSupplementary methods
Supplementary methods This section provides additional technical details on the sample, the applied imaging and analysis steps and methods. Structural imaging Trained radiographers placed all participants
More informationImproved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm
Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm L.P. Panych 1,3, B. Madore 1,3, W.S. Hoge 1,3, R.V. Mulkern 2,3 1 Brigham and
More informationBreast MRI Accreditation Program Clinical Image Quality Guide
Breast MRI Accreditation Program Clinical Image Quality Guide Introduction This document provides guidance on breast MRI clinical image quality and describes the criteria used by the ACR Breast MRI Accreditation
More informationChapter 3 Set Redundancy in Magnetic Resonance Brain Images
16 Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 3.1 MRI (magnetic resonance imaging) MRI is a technique of measuring physical structure within the human anatomy. Our proposed research focuses
More informationMidterm Review
Midterm Review - 2017 EE369B Concepts Noise Simulations with Bloch Matrices, EPG Gradient Echo Imaging 1 About the Midterm Monday Oct 30, 2017. CCSR 4107 Up to end of C2 1. Write your name legibly on this
More informationSPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing
SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial
More informationQualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs
Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Doug Dean Final Project Presentation ENGN 2500: Medical Image Analysis May 16, 2011 Outline Review of the
More informationTransform Introduction page 96 Spatial Transforms page 97
Transform Introduction page 96 Spatial Transforms page 97 Pad page 97 Subregion page 101 Resize page 104 Shift page 109 1. Correcting Wraparound Using the Shift Tool page 109 Flip page 116 2. Flipping
More informationSampling, Ordering, Interleaving
Sampling, Ordering, Interleaving Sampling patterns and PSFs View ordering Modulation due to transients Temporal modulations Slice interleaving Sequential, Odd/even, bit-reversed Arbitrary Other considerations:
More informationMeasuring baseline whole-brain perfusion on GE 3.0T using arterial spin labeling (ASL) MRI
Measuring baseline whole-brain perfusion on GE 3.0T using arterial spin labeling (ASL) MRI Revision date: 11/20/2006 Overview This document describes the procedure for measuring baseline whole-brain perfusion
More informationComputational Medical Imaging Analysis
Computational Medical Imaging Analysis Chapter 2: Image Acquisition Systems Jun Zhang Laboratory for Computational Medical Imaging & Data Analysis Department of Computer Science University of Kentucky
More informationMRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health
MRI image formation Indiana University School of Medicine and Indiana University Health Disclosure No conflict of interest for this presentation 2 Outlines Data acquisition Spatial (Slice/Slab) selection
More informationFunctional MRI in Clinical Research and Practice Preprocessing
Functional MRI in Clinical Research and Practice Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization
More informationMeasuring baseline whole-brain perfusion on GE 3.0T using arterial spin labeling (ASL) MRI
Measuring baseline whole-brain perfusion on GE 3.0T using arterial spin labeling (ASL) MRI Revision date: 09/15/2008 Overview This document describes the procedure for measuring baseline whole-brain perfusion
More informationIMAGE reconstruction in conventional magnetic resonance
IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 24, NO. 3, MARCH 2005 325 Conjugate Phase MRI Reconstruction With Spatially Variant Sample Density Correction Douglas C. Noll*, Member, IEEE, Jeffrey A. Fessler,
More informationThE ultimate, INTuITIVE Mr INTErFAcE
ThE ultimate, INTuITIVE Mr INTErFAcE Empowering you to do more The revolutionary Toshiba M-power user interface takes Mr performance and flexibility to levels higher than ever before. M-power is able to
More informationThe cost of parallel imaging in functional MRI of the human brain
Magnetic Resonance Imaging 24 (2006) 1 5 Original contributions The cost of parallel imaging in functional MRI of the human brain Henry Lqtcke4, Klaus-Dietmar Merboldt, Jens Frahm Biomedizinische NMR Forschungs
More informationSIEMENS MAGNETOM Verio syngo MR B15V
\\USER\ZAHID_RESEARCH\MS\No Name\3D SWI TA: 6:39 PAT: 2 Voxel size: 1.0 0.5 2.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer Inline movie Auto
More informationEPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing
Functional Connectivity Preprocessing Geometric distortion Head motion Geometric distortion Head motion EPI Data Are Acquired Serially EPI Data Are Acquired Serially descending 1 EPI Data Are Acquired
More information2.1 Signal Production. RF_Coil. Scanner. Phantom. Image. Image Production
An Extensible MRI Simulator for Post-Processing Evaluation Remi K.-S. Kwan?, Alan C. Evans, and G. Bruce Pike McConnell Brain Imaging Centre, Montreal Neurological Institute, McGill University, Montreal,
More informationSPM8 for Basic and Clinical Investigators. Preprocessing
SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial
More informationCSImage Tutorial v August 2000
CSImage Tutorial v 1.0 8 August 2000 To use CSImage you need to have Java2 installed. Development was done using JDK 1.2.2 from Sun Microsystems. The program has been run on Digital Unix (Compaq Tru64
More informationA Virtual MR Scanner for Education
A Virtual MR Scanner for Education Hackländer T, Schalla C, Trümper A, Mertens H, Hiltner J, Cramer BM Hospitals of the University Witten/Herdecke, Department of Radiology Wuppertal, Germany Purpose A
More informationMRI Imaging Options. Frank R. Korosec, Ph.D. Departments of Radiology and Medical Physics University of Wisconsin Madison
MRI Imaging Options Frank R. Korosec, Ph.D. Departments of Radiolog and Medical Phsics Universit of Wisconsin Madison f.korosec@hosp.wisc.edu As MR imaging becomes more developed, more imaging options
More informationContinuation Format Page
C.1 PET with submillimeter spatial resolution Figure 2 shows two views of the high resolution PET experimental setup used to acquire preliminary data [92]. The mechanics of the proposed system are similar
More informationUnaliasing by Fourier-Encoding the Overlaps Using the Temporal Dimension (UNFOLD), Applied to Cardiac Imaging and fmri
1999 ISMRM YOUNG INVESTIGATORS MOORE AWARD PAPERS Magnetic Resonance in Medicine 42:813 828 (1999) Unaliasing by Fourier-Encoding the Overlaps Using the Temporal Dimension (UNFOLD), Applied to Cardiac
More informationSuper-Resolution Reconstruction of Diffusion-Weighted Images from Distortion Compensated Orthogonal Anisotropic Acquisitions.
Super-Resolution Reconstruction of Diffusion-Weighted Images from Distortion Compensated Orthogonal Anisotropic Acquisitions. Benoit Scherrer Ali Gholipour Simon K. Warfield Children s Hospital Boston,
More informationBasic principles of MR image analysis. Basic principles of MR image analysis. Basic principles of MR image analysis
Basic principles of MR image analysis Basic principles of MR image analysis Julien Milles Leiden University Medical Center Terminology of fmri Brain extraction Registration Linear registration Non-linear
More informationCompressed Sensing for Rapid MR Imaging
Compressed Sensing for Rapid Imaging Michael Lustig1, Juan Santos1, David Donoho2 and John Pauly1 1 Electrical Engineering Department, Stanford University 2 Statistics Department, Stanford University rapid
More informationA Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data
A Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data Alexey Samsonov, Julia Velikina Departments of Radiology and Medical
More informationCorrection for EPI Distortions Using Multi-Echo Gradient-Echo Imaging
Correction for EPI Distortions Using Multi-Echo Gradient-Echo Imaging Nan-kuei Chen and Alice M. Wyrwicz* Magnetic Resonance in Medicine 41:1206 1213 (1999) A novel and effective technique is described
More informationGE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals.
GE Healthcare CLINICAL GALLERY Discovery * MR750w 3.0T This brochure is intended for European healthcare professionals. NEURO PROPELLER delivers high resolution, motion insensitive imaging in all planes.
More informationFast Imaging Trajectories: Non-Cartesian Sampling (1)
Fast Imaging Trajectories: Non-Cartesian Sampling (1) M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.05.03 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business
More informationThe organization of the human cerebral cortex estimated by intrinsic functional connectivity
1 The organization of the human cerebral cortex estimated by intrinsic functional connectivity Journal: Journal of Neurophysiology Author: B. T. Thomas Yeo, et al Link: https://www.ncbi.nlm.nih.gov/pubmed/21653723
More informationSampling, Ordering, Interleaving
Sampling, Ordering, Interleaving Sampling patterns and PSFs View ordering Modulation due to transients Temporal modulations Timing: cine, gating, triggering Slice interleaving Sequential, Odd/even, bit-reversed
More informationExam 8N080 - Introduction MRI
Exam 8N080 - Introduction MRI Friday January 23 rd 2015, 13.30-16.30h For this exam you may use an ordinary calculator (not a graphical one). In total there are 6 assignments and a total of 65 points can
More informationBlipped-Controlled Aliasing in Parallel Imaging for Simultaneous Multislice Echo Planar Imaging With Reduced g-factor Penalty
IMAGING METHODOLOGY - Full Papers Magnetic Resonance in Medicine 67:1210 1224 (2012) Blipped-Controlled Aliasing in Parallel Imaging for Simultaneous Multislice Echo Planar Imaging With Reduced g-factor
More informationAdvanced Imaging Trajectories
Advanced Imaging Trajectories Cartesian EPI Spiral Radial Projection 1 Radial and Projection Imaging Sample spokes Radial out : from k=0 to kmax Projection: from -kmax to kmax Trajectory design considerations
More informationCS/NEUR125 Brains, Minds, and Machines. Due: Wednesday, April 5
CS/NEUR125 Brains, Minds, and Machines Lab 8: Using fmri to Discover Language Areas in the Brain Due: Wednesday, April 5 In this lab, you will analyze fmri data from an experiment that was designed to
More informationImaging Guide Pancreatic Imaging
Imaging Guide Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Imaging System Ver 1.0 Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Imaging System Course Objectives: This guide
More informationADNI, ADNI_QH, SURVEY. Geometry. connection
ADNI, ADNI_QH, SURVEY Geometry Coil selection = Head connection = d Multi coil Homogeneity correction ne FOV (mm) = 250.00 RFOV (%) = 100.00 Foldover suppression Matrix scan = 256 reconstruction = 256
More informationFunctional MRI data preprocessing. Cyril Pernet, PhD
Functional MRI data preprocessing Cyril Pernet, PhD Data have been acquired, what s s next? time No matter the design, multiple volumes (made from multiple slices) have been acquired in time. Before getting
More informationSIEMENS MAGNETOM TrioTim syngo MR B17
\\USER\KNARRGROUP\MultiBand\LavretskyMultiBand\trufi localizer 3-plane TA: 5.1 s PAT: Voxel size: 1.2 1.2 5. Rel. SNR: 1.00 SIEMENS: trufi Load to stamp Slice group 1 Slices 1 Dist. factor 20 % Phase enc.
More informationThree-dimensional numerical simulations of susceptibility-induced magnetic field inhomogeneities in the human head
Magnetic Resonance Imaging 20 (2002) 759 770 Three-dimensional numerical simulations of susceptibility-induced magnetic field inhomogeneities in the human head Trong-Kha Truong a, Bradley D. Clymer b,
More information45 µm polystyrene bead embedded in scattering tissue phantom. (a,b) raw images under oblique
Phase gradient microscopy in thick tissue with oblique back-illumination Tim N Ford, Kengyeh K Chu & Jerome Mertz Supplementary Figure 1: Comparison of added versus subtracted raw OBM images 45 µm polystyrene
More informationMagnetic Susceptibility Artefact Correction of Spin-Echo and Gradient- Echo EPI Images
Magnetic Susceptibility Artefact Correction of Spin-Echo and Gradient- Echo EPI Images Gary McGinley Submission date: 12 th June 2015 Supervisor: Atle Bjørnerud Co-supervisor: Øystein Bech Gadmar University
More informationSteen Moeller Center for Magnetic Resonance research University of Minnesota
Steen Moeller Center for Magnetic Resonance research University of Minnesota moeller@cmrr.umn.edu Lot of material is from a talk by Douglas C. Noll Department of Biomedical Engineering Functional MRI Laboratory
More informationAdvanced Image Reconstruction Methods for Photoacoustic Tomography
Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic
More information