1 Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge Dimitre Hristov Radiation Oncology Stanford University The team Renhui Gong 1 Magdalena Bazalova-Carter 1 Ralf Bruder 2 Achim Schweikard 2 Jeff Schlosser 3 1 Stanford University, Radiation Oncology, Stanford, United States 2 Institute for Robotics and Cognitive Systems, Luebeck, Germany 3 Sonitrack Systems, Menlo Park, United States Disclosures Funding/equipment provided at various stages by: NIH STTR grant to Sonitrack Systems NSF Philips Ultrasound Interson Stanford Bio-X IIP 1
2 Anatomy position Imaging during beam delivery Add-on, real-time, volumetric, soft-tissue guidance during radiation beam delivery is largely unmet challenge Image guidance architecture Beam gating Linear accelerator LINAC Accelerator control console Optical tracker US probe Patient US probe manipulator US-guidance workstation computer High level robot commands Robot Control control (I7) US image stream US imaging system US probe position data (6 DOF) Schlosser J, Salisbury K, Hristov D, Telerobotic system concept for real-time soft-tissue imaging during radiotherapy beam delivery, Med Phys. 21 Dec;37(12): Telerobotic imaging Remote Haptic Interface Robot Schlosser J, Salisbury K, Hristov D, Telerobotic system concept for real-time soft-tissue imaging during radiotherapy beam delivery, Med Phys. 21 Dec;37(12):
3 Telerobotic imaging Challenge: Tele-robotic imaging interfering with linac and beam! Remote Haptic Interface Robot Schlosser J, Salisbury K, Hristov D, Telerobotic system concept for real-time soft-tissue imaging during radiotherapy beam delivery, Med Phys. 21 Dec;37(12):
4 Approaches to the treatment interference challenge Avoid beams through probe and robot Can we assure resulting plan is clinically acceptable for a given patient? Include probe and robot in plan Can they be modelled sufficiently accurately? Make probe and robot disappear Has Dimitre gone mad? 1 Avoid beams through probe and robot CT simulation Guide device placement to potential noninterfering imaging positions Confirm adequate image quality Augmented reality system for ultrasound guided radiation therapy Renhui Gong, Ralf Bruder, Achim Schweikard, Jeff Schlosser, and Dimitre Hristov, International Journal of Computer Assisted Radiology and Surgery, June 215, Volume 1, Issue 1 S31 11 Probe placement map generation Ralf Bruder, Floris Ernst and Achim Schweikard, SU D 22 2: Optimal transducer positions for 4D ultrasound guidance in cardiac IGRT, in: 53rd Annual Meeting of the AAPM, pages 339,
5 Probe placement map generation Define target and map attenuation to target from surface CT simulation Register probe and robot to CT during simulation by scanning optically tracked tool Guide probe placement towards non-interfering positions during simulation: probe placement map 14 Probe placement map confirmation Visualize probe in virtual space during actual imaging. 15 5
6 Probe placement map confirmation Visualize live US images fused with CT and target. 16 Avoid beams through probe and robot Planning Verify non-interference of designed plan Suggest slight changes of robot placement if necessary Design plan trajectories to avoid robot 17 Incorporate robot model in planning Custom xml descriptor defining joints and limits 18 6
7 Virtual simulation environment Robot placement 19 Robot placement in 3D 2 Recorded joint positions during simulation or interactive setup Virtual simulation environment Plan interference evaluation by loading RT objects 21 7
8 Collision evaluation Plan interference evaluation 22 Beam evaluation Beam eye-view to monitor entrance through probe/robot Include probe and robot in plan No probe present Matrix array transducer Philips X6-1 3D/4D matrix array probe Severe CT artifacts preclude CT-based modeling. 8
9 Megavoltage CT for electron density calibration Electron density models of X6-1 and C5-2 ultrasound probes built with a Tomotherapy 3.5 MV CT scan. No probe present Matrix array transducer Monte Carlo modeling of ultrasound probes for image guided radiotherapy Magdalena Bazalova-Carter, Jeffrey Schlosser, Josephine Chen, and Dimitre Hristov Submitted to Medical Physics Experimental setup No probe present Note detail differences between MC, film, and 2D array. X6-1 Probe Horizontal Position No probe present 9
10 MC versus 2D array No probe present X6-1 Probe Vertical Position No probe present Cable placement introduces dose discrepancies. Remotely-Actuated Ultrasound Scanning (RUSS) Stepper motor DC motor Flexible shaft Rigid linkage system Make probe and robot disappear (well partially) Axis 1 (lateral rotation) Axis 2 (elevational rotation) Transducer element 3D US volume Elevational sweep 2D US planes Lateral sweep 1
11 Frame rate and Field of View For ~1 slice per elevational degree: Elevational Sweep Angle Lateral Imaging Sweep Depth Angle Axial Slice Field of View 15⁰ 19⁰ 15 cm 5. cm x 4. cm 23⁰ 29⁰ 1 cm 5. cm x 4. cm 45⁰ 6⁰ 5 cm 5. cm x 4. cm 15⁰ 6⁰ 15 cm 15. cm x 4. cm 23⁰ 6⁰ 1 cm 1. cm x 4. cm 45⁰ 6⁰ 5 cm 5. cm x 4. cm # of Elevational Slices Volumes Per Second Planes Per Second Spatial Resolution Method: Lateral/elevational: point spread using -3dB peak drop-off Axial: minimum resolvable spacing 11
12 Target Displacement Error [mm] Target Displacement Error [mm] Target Displacement Error [mm] Tracked Target Displacement [mm] Tracked Target Displacement [mm] Tracked Target Displacement [mm] Resolution [mm] Elevational Resolution [mm] Spatial Resolution: Results Depth [mm] Depth [mm] 1 Slices 16 Slices 2 Slices 24 Slices 3 Slices 4 Slices Axial Lateral Elevational 5 cm depth 1 mm 5.1 mm 3.7 mm 1 cm depth 1 mm 7.9 mm 5.7 mm Tracking Resolution Method: US phantom placed on motion platform Targets automatically tracked using normalized cross correlation Tracking Resolution: Results Actual Target Displacement [mm] Actual Target Displacement [mm] Actual Target Displacement [mm] Actual Target Displacement [mm] Actual Target Displacement [mm] Actual Target Displacement [mm] Axial Lateral Elevational Mean Target Tracking Error.2 mm.4 mm.3 mm Max Target Tracking Error.4 mm 1.7 mm.9 mm 12
13 CT Compatibility No probe present Philips X6-1 3D/4D matrix array probe RUSS probe Radiotherapy Beam Compatibility Method 1: Compare planned and delivered dose through RUSS probe Method 2: Image-based tracking during beam delivery Experimental Setup: Accuros Plan: Radiotherapy Beam Compatibility: Results 6MV PTW Measurements 15MV PTW Measurements All points met 2. mm / 3.% dose deviation criteria 13
14 Tracked Target Displament (Normalized) Radiotherapy Beam Compatibility: Results 6MV PTW Measurements 15MV PTW Measurements Tracking Performance with beam on/off Beam off Beam on Relative Probe Displacement [mm] All points met 2. mm / 3.% dose deviation criteria RUSS summary Metric Framerate and Field of View (FOV) Target: Intrafractional Liver Radiotherapy Guidance 6.5 cm x 4 cm FOV at 5-15 cm depth; 1 volume per second Result with RUSS 1 cm depth: 5. cm x 4. cm FOV, 2.2 Hz volume rate, 48 Hz plane rate Tracking Resolution 2 mm in each direction.4 mm mean resolution Imaging during radiation delivery Radiotherapy planning compatibility No statistically significant difference between tracking performance with beam on/off ±3.% / 2. mm agreement between computed and measured dose distributions p=.52 All points agree within ±3.% / 2. mm RUSS performance meets requirements for intrafractional radiotherapy motion management Low CT artifacts, beam compatibility, and low cost Conclusions Addressing treatment interferences by tele-robotic US imaging requires simulation tools for avoidance or inclusion strategies Inclusion strategy feasible but possibly less practical than avoidance Dedicated radiolucent ultrasound probes may greatly facilitate RT ultrasound guidance but careful evaluation of performance trade-offs is necessary 14
Mech. Engineering, Comp. Science, and Rad. Oncology Departments. Schools of Engineering and Medicine, Bio-X Program, Stanford University
Mech. Engineering, Comp. Science, and Rad. Oncology Departments Schools of Engineering and Medicine, Bio-X Program, Stanford University 1 Conflict of Interest Nothing to disclose 2 Imaging During Beam
More informationImage Guidance and Beam Level Imaging in Digital Linacs
Image Guidance and Beam Level Imaging in Digital Linacs Ruijiang Li, Ph.D. Department of Radiation Oncology Stanford University School of Medicine 2014 AAPM Therapy Educational Course Disclosure Research
More information7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT
CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance
More informationThank-You Members of TG147 TG 147: QA for nonradiographic
Thank-You Members of TG147 TG 147: QA for nonradiographic localization and positioning systems Twyla Willoughby, M.S. Medical Physicist Clinical AAPM Meeting March 2013 Department of Radiation Oncology
More informationMonte Carlo methods in proton beam radiation therapy. Harald Paganetti
Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization
More informationREAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT
REAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT Anand P Santhanam Assistant Professor, Department of Radiation Oncology OUTLINE Adaptive radiotherapy for head and
More informationFinancial disclosure. Onboard imaging modality for IGRT
Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More informationSystem Integration and Preliminary In-Vivo Experiments of a Robot for Ultrasound Guidance and Monitoring during Radiotherapy
System Integration and Preliminary In-Vivo Experiments of a Robot for Ultrasound Guidance and Monitoring during Radiotherapy H. Tutkun Şen, Muyinatu A. Lediju Bell, Yin Zhang, Kai Ding, John Wong, Iulian
More informationState-of-the-Art IGRT
in partnership with State-of-the-Art IGRT Exploring the Potential of High-Precision Dose Delivery and Real-Time Knowledge of the Target Volume Location Antje-Christin Knopf IOP Medical Physics Group Scientific
More informationTG-148 overview. Introduction. System Overview. System Overview. QA for Helical Tomotherapy: Report of the AAPM Task Group 148. Conflict of Interest:
QA or Helical Tomotherapy: Report o the AAPM Tas Group 148 Members: Conlict o Interest: r. John Balog owns TomoTherapy stoc. Katja Langen (Co-chair) Nio Papaniolaou (Co-chair) Walter Grant Richard Crilly
More informationThe IORT Treatment Planning System. radiance. GMV, 2012 Property of GMV All rights reserved
The IORT Treatment Planning System radiance Property of GMV All rights reserved WHY RADIANCE? JUSTIFICATION Property of GMV All rights reserved ADVANTAGES OF IORT PRECISION: RT guided by direct vision.
More information8/3/2016. Image Guidance Technologies. Introduction. Outline
8/3/26 Session: Image Guidance Technologies and Management Strategies Image Guidance Technologies Jenghwa Chang, Ph.D.,2 Department of Radiation Medicine, Northwell Health 2 Hofstra Northwell School of
More information연구용유방초음파질관리 원광대학병원김혜원
연구용유방초음파질관리 원광대학병원김혜원 Why US QC? Quality control (QC) testing of ultrasound scanners is important to verify the proper and consistent operation of these devices. main goal ; quality improvement Guidelines
More informationINTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory
INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1 Modifications for P551 Fall 2013 Medical Physics Laboratory Introduction Following the introductory lab 0, this lab exercise the student through
More informationOutline. Outline 7/24/2014. Fast, near real-time, Monte Carlo dose calculations using GPU. Xun Jia Ph.D. GPU Monte Carlo. Clinical Applications
Fast, near real-time, Monte Carlo dose calculations using GPU Xun Jia Ph.D. xun.jia@utsouthwestern.edu Outline GPU Monte Carlo Clinical Applications Conclusions 2 Outline GPU Monte Carlo Clinical Applications
More information1. Learn to incorporate QA for surface imaging
Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago ***No disclosures*** 1. Learn to incorporate QA for surface imaging into current QA procedures for IGRT. 2. Understand
More informationADVANCING CANCER TREATMENT
The RayPlan treatment planning system makes proven, innovative RayStation technology accessible to clinics that need a cost-effective and streamlined solution. Fast, efficient and straightforward to use,
More informationCARS 2008 Computer Assisted Radiology and Surgery
Online External Beam Radiation Planning and Training Felix Hamza-Lup a, Ivan Sopin a, Omar Zeidan b a Computer Science, Armstrong Atlantic State University, Savannah, Georgia, USA b MD Anderson Cancer
More informationVolumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT
Volumetric Modulated Arc Therapy - Clinical Implementation Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA Acknowledgement David M. Shepard, Ph.D. Muhammad K. N. Afghan, Ph.D. Fan Chen, Ph.D.
More informationDigital Tomosynthesis for Target Localization
Digital Tomosynthesis for Target Localization Fang-Fang Yin, Devon Godfrey, Lei Ren Jacqueline Maurer, Jackie Q-L Wu Duke University Medical Center Acknowledgements Duke Radiation Oncology faculty and
More informationNew Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC
New Technology in Radiation Oncology James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC Technology s s everywhere From the imaging chain To the planning system To the linac To QA..it..it
More informationMR-guided radiotherapy: Vision, status and research at the UMC Utrecht. Dipl. Ing. Dr. Markus Glitzner
MR-guided radiotherapy: Vision, status and research at the UMC Utrecht Dipl. Ing. Dr. Markus Glitzner About myself Training Medizintechnik TU Graz PhD UMC Utrecht Clinical work Software implementation
More informationSignificance of time-dependent geometries for Monte Carlo simulations in radiation therapy. Harald Paganetti
Significance of time-dependent geometries for Monte Carlo simulations in radiation therapy Harald Paganetti Modeling time dependent geometrical setups Key to 4D Monte Carlo: Geometry changes during the
More informationIndrin Chetty Henry Ford Hospital Detroit, MI. AAPM Annual Meeting Houston 7:30-8:25 Mon 08/07/28 1/30
Review of TG105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning D. W. O. Rogers, Carleton Laboratory for Radiotherapy Physics,
More informationBasics of treatment planning II
Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!
More informationPhoton beam dose distributions in 2D
Photon beam dose distributions in 2D Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2014 Acknowledgments! Narayan Sahoo PhD! Richard G Lane (Late) PhD 1 Overview! Evaluation
More informationImage Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD
Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image
More informationTHE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 THE SIMULATION OF
More informationIntroduction. Quality Assurance for Image- Guided Radiation Therapy. Justification for IGRT. Image-Guided Radiation Therapy
Introduction Quality Assurance for Image- Guided Radiation Therapy Jean-Pierre Bissonnette, Ph.D., MCCPM Princess Margaret Hospital, Toronto, Canada IGRT What is it? Rationale Equipment Quality Assurance
More informationVirtual Phantoms for IGRT QA
TM Virtual Phantoms for IGRT QA Why ImSimQA? ImSimQA was developed to overcome the limitations of physical phantoms for testing modern medical imaging and radiation therapy software systems, when there
More informationEstimating 3D Respiratory Motion from Orbiting Views
Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct. 2005 Funding provided by NIH Grant P01 CA59827 Motivation Free-breathing
More informationMichael Speiser, Ph.D.
IMPROVED CT-BASED VOXEL PHANTOM GENERATION FOR MCNP MONTE CARLO Michael Speiser, Ph.D. Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX September 1 st, 2012 CMPWG Workshop Medical
More informationEvaluation of 3D Gamma index calculation implemented in two commercial dosimetry systems
University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2015 Evaluation of 3D Gamma index calculation implemented
More informationABSTRACT 1. INTRODUCTION
Tracked ultrasound calibration studies with a phantom made of LEGO bricks Marie Soehl, Ryan Walsh, Adam Rankin, Andras Lasso, Gabor Fichtinger Queen s University, Kingston ON, Canada ABSTRACT PURPOSE:
More informationDelta 4 PT. True Volumetric Pre-treatment Verification. The difference is clear
Delta 4 PT True Volumetric Pre-treatment Verification The difference is clear ScandiDos innovators in advanced volumetric dosimetry solutions ScandiDos ScandiDos is the innovative company that introduced
More informationHelical Tomotherapy Qualitative dose Delivery Verification
University of Tennessee, Knoxville Trace: Tennessee Research and Creative Exchange University of Tennessee Honors Thesis Projects University of Tennessee Honors Program 5-2003 Helical Tomotherapy Qualitative
More informationIMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays
IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays Sotiri Stathakis Outline Why IMRT/VMAT QA AAPM TG218 UPDATE Tolerance Limits and Methodologies for IMRT Verification QA Common sources
More informationC-mode Real Time Tomographic Reflection for a Matrix Array Ultrasound Sonic Flashlight
C-mode Real Time Tomographic Reflection for a Matrix Array Ultrasound Sonic Flashlight George Stetten 1,2,3, Aaron Cois 1,2,, Wilson Chang 1,2,3, Damion Shelton 2, Robert Tamburo 1,2, John Castellucci
More informationInitial Clinical Experience with 3D Surface Image Guidance
Initial Clinical Experience with 3D Surface Image Guidance Amanda Havnen-Smith, Ph.D. Minneapolis Radiation Oncology Ridges Radiation Therapy Center Burnsville, MN April 20 th, 2012 Non-funded research
More informationEffects of the difference in tube voltage of the CT scanner on. dose calculation
Effects of the difference in tube voltage of the CT scanner on dose calculation Dong Joo Rhee, Sung-woo Kim, Dong Hyeok Jeong Medical and Radiological Physics Laboratory, Dongnam Institute of Radiological
More informationMutual information based CT registration of the lung at exhale and inhale breathing states using thin-plate splines
Mutual information based CT registration of the lung at exhale and inhale breathing states using thin-plate splines Martha M. Coselmon, a) James M. Balter, Daniel L. McShan, and Marc L. Kessler Department
More information8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques
Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery Lei Ren, Ph.D. Duke University Medical Center 2016 AAPM 58 th annual meeting, Educational Course, Therapy Track Disclosure I have
More informationADVANCING CANCER TREATMENT
3 ADVANCING CANCER TREATMENT SUPPORTING CLINICS WORLDWIDE RaySearch is advancing cancer treatment through pioneering software. We believe software has un limited potential, and that it is now the driving
More informationClinical implementation of photon beam flatness measurements to verify beam quality
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Clinical implementation of photon beam flatness measurements to verify beam quality Simon Goodall, a Nicholas Harding, Jake Simpson,
More informationUse of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta
Use of Monte Carlo modelling in radiotherapy linac design David Roberts, PhD Senior Physicist Elekta Contents Overview of Elekta What we do Where we use Monte Carlo Codes and resources Example : Agility
More informationOptimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago
Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic
More informationArtefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT
Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT Marcus Brehm 1,2, Thorsten Heußer 1, Pascal Paysan 3, Markus Oehlhafen 3, and Marc Kachelrieß 1,2 1 German Cancer Research Center
More informationUNCOMPROMISING QUALITY
ION CHAMBERS UNCOMPROMISING QUALITY Designed with over 30 years of scientific integrity for a broad range of dosimetry measurements in diverse radiation beams Farmer-type Chambers For absolute dosimetry
More informationMonte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters
Monte Carlo Simulation for Neptun 1 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters M.T. Bahreyni Toossi a, M. Momen Nezhad b, S.M. Hashemi a a Medical Physics Research Center,
More informationThe Near Future in Cardiac CT Image Reconstruction
SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de
More informationRecent Advances in Non-Invasive Thermometry Using Changes in Backscattered Ultrasound
Recent Advances in Non-Invasive Thermometry Using Changes in Backscattered Ultrasound R. Martin Arthur 1, Jason W. Trobaugh 1, William L. Straube 2 and Eduardo G. Moros 2 1 Electrical & Systems Engineering
More informationOptical Guidance. Sanford L. Meeks. July 22, 2010
Optical Guidance Sanford L. Meeks July 22, 2010 Optical Tracking Optical tracking is a means of determining in real-time the position of a patient relative to the treatment unit. Markerbased systems track
More informationMotion Compensation from Short-Scan Data in Cardiac CT
Motion Compensation from Short-Scan Data in Cardiac CT Juliane Hahn 1,2, Thomas Allmendinger 1, Herbert Bruder 1, and Marc Kachelrieß 2 1 Siemens Healthcare GmbH, Forchheim, Germany 2 German Cancer Research
More informationDose Calculation and Optimization Algorithms: A Clinical Perspective
Dose Calculation and Optimization Algorithms: A Clinical Perspective Daryl P. Nazareth, PhD Roswell Park Cancer Institute, Buffalo, NY T. Rock Mackie, PhD University of Wisconsin-Madison David Shepard,
More informationSuturing in Confined Spaces: Constrained Motion Control of a Hybrid 8-DoF Robot
Suturing in Confined Spaces: Constrained Motion Control of a Hybrid 8-DoF Robot Ankur Kapoor 1, Nabil Simaan 2, Russell H. Taylor 1 1 ERC-CISST Department of Computer Science Johns Hopkins University 2
More informationFacility Questionnaire PART I (General Information for 3DCRT and IMRT)
Facility Questionnaire PART I (General Information for 3DCRT and IMRT) The following items are required before you can enter cases on any RTOG protocol that requires data submission to the Image-Guided
More informationI. INTRODUCTION. Figure 1. Radiation room model at Dongnai General Hospital
International Journal of Computational Engineering Research Vol, 04 Issue, 4 Simulation of Photon and Electron dose distributions 5 code for the treatment area using the linear electron accelerator (LINAC)
More informationCBCT: Past, Present and Future. Disclosures. Computed Tomography. ATec Education Course Feb 28-Mar3, Douglas Moseley PhD, DABR
CBCT: Past, Present and Future Douglas Moseley PhD, DABR Disclosures License Agreement Modus Medical Educational Consultant Elekta Oncology Systems Computed Tomography First CT Scanner Third Generation
More informationMRI-LINAC Dynamic Phantom
MRI-LINAC Dynamic Phantom Model 008M IMAGE ACQUISITION TREATMENT PLANNING DOSE DELIVERY Strict QA procedures for the imaging, planning and delivery of radiotherapy using respiratory management devices
More informationIMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE
QA SOFTWARE FAST, PRECISE Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, VMAT, SRS and brachytherapy plans no film, no phantoms, no linac
More information3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending
3DVH : On The Accuracy Of The Planned Dose Perturbation Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools *Patent Pending introduction State-of-the-art IMRT QA of static gantry
More informationHidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.
2-D D Dose-CT Mapping in Geant4 Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. Table of Contents Background & Purpose Materials Methods
More informationImage Acquisition Systems
Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS
More informationDosimetry Simulations with the UF-B Series Phantoms using the PENTRAN-MP Code System
Dosimetry Simulations with the UF-B Series Phantoms using the PENTRAN-MP Code System A. Al-Basheer, M. Ghita, G. Sjoden, W. Bolch, C. Lee, and the ALRADS Group Computational Medical Physics Team Nuclear
More informationDisclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions
LINAC and MLC QA for IMRT Ping Xia, Ph.D., Department of Radiation Oncology Disclosure Received research support from Siemens Medical Solutions University of California San Francisco Therapy Series (SAM)
More informationTHE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!
THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER! Confidence in complex treatments Modern radiation therapy uses complex plans with techniques such as IMRT, VMAT and Tomotherapy.
More informationFAST, precise. qa software
qa software FAST, precise Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, rotational or brachytherapy plans no film, no phantoms, no linac
More informationDynalog data tool for IMRT plan verification
Dynalog data tool for IMRT plan verification Poster No.: R-0051 Congress: 2014 CSM Type: Scientific Exhibit Authors: V. Sashin; FOOTSCRAY/AU Keywords: Computer applications, Radiation physics, Experimental,
More informationLecture 6: Medical imaging and image-guided interventions
ME 328: Medical Robotics Winter 2019 Lecture 6: Medical imaging and image-guided interventions Allison Okamura Stanford University Updates Assignment 3 Due this Thursday, Jan. 31 Note that this assignment
More informationICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588
ICARO Vienna April 27-29 2009 Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588 M. Saiful Huq, Ph.D., Professor and Director, Dept. of Radiation
More informationOp#miza#on of an On- Board Imaging System for Rapid Radiotherapy. Erica Kemmerling, Meng Wu, He Yang, and Rebecca Fahrig
Op#miza#on of an On- Board Imaging System for Rapid Radiotherapy Erica Kemmerling, Meng Wu, He Yang, and Rebecca Fahrig Problem: mo#on during treatment Two overlapping thorax views during normal breathing
More informationInvestigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers
Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Krista Chytyk MSc student Supervisor: Dr. Boyd McCurdy Introduction The objective of cancer radiotherapy
More informationBasic Radiation Oncology Physics
Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides
More informationPET-CT in Radiation Treatment Planning
PET-CT in Radiation Treatment Planning TA van de Water, Radiotherapeutic Institute Friesland, Leeuwarden JA van Dalen, Isala, Zwolle ACKNOWLEDGEMENTS A. van der Schaaf (University of Groningen, University
More informationJournal of Visualized Experiments
Video Article Real-time Monitoring of High Intensity Focused Ultrasound (HIFU) Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound (HMIFU) Julien Grondin 1, Thomas Payen
More informationBiomet PMI. Patient-Matched Implants. CT Protocols
CT Protocols One Surgeon. One Patient. Over 1 million times per year, Biomet helps one surgeon provide personalized care to one patient. The science and art of medical care is to provide the right solution
More informationradiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li NCAAPM Spring Meeting 2010 Madison, WI
GPU-Accelerated autosegmentation for adaptive radiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li agodley@mcw.edu NCAAPM Spring Meeting 2010 Madison, WI Overview Motivation Adaptive
More informationAutomated Quality Assurance for Image-Guided Radiation Therapy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 1, WINTER 2009 Automated Quality Assurance for Image-Guided Radiation Therapy Eduard Schreibmann, a Eric Elder, Tim Fox Department of Radiation
More informationCardiac Dual Energy CT: Technique
RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de
More informationValidation System of MR Image Overlay and Other Needle Insertion Techniques
Submitted to the Proceedings fro Medicine Meets Virtual Reality 15 February 2007, Long Beach, California To appear in Stud. Health Inform Validation System of MR Image Overlay and Other Needle Insertion
More informationReal-time self-calibration of a tracked augmented reality display
Real-time self-calibration of a tracked augmented reality display Zachary Baum, Andras Lasso, Tamas Ungi, Gabor Fichtinger Laboratory for Percutaneous Surgery, Queen s University, Kingston, Canada ABSTRACT
More informationComparison of absorbed dose distribution 10 MV photon beam on water phantom using Monte Carlo method and Analytical Anisotropic Algorithm
Journal of Physics: Conference Series PAPER OPEN ACCESS Comparison of absorbed dose distribution 1 MV photon beam on water phantom using Monte Carlo method and Analytical Anisotropic Algorithm To cite
More informationSource Model Tuning for a 6 MV Photon Beam used in Radiotherapy
Journal of Physics: Conference Series Source Model Tuning for a 6 MV Photon Beam used in Radiotherapy To cite this article: Lukas A Hirschi et al 2007 J. Phys.: Conf. Ser. 74 021008 View the article online
More informationA fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study
A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study Indrin J. Chetty, a) Mihaela Rosu, Neelam Tyagi, Lon H. Marsh, Daniel
More informationRobot Control for Medical Applications and Hair Transplantation
Dr. John Tenney Director of Research Restoration Robotics, Inc. Robot Control for Medical Applications and Hair Transplantation Presented to the IEEE Control Systems Society, Santa Clara Valley 18 November
More informationAcceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems. Bruce Curran University of Michigan Medical Center Ann Arbor, MI
Acceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems Bruce Curran University of Michigan Medical Center Ann Arbor, MI Disclosure Some of the work discussed in this talk was supported
More informationPreface. Med. Phys. 35(9), , Mechanical QA. Radiation Survey Mechanical tests Light radiation Table, Collimator, Gantry Jaws.
AAPM-SAM-2012-Das (1) Beam Data Collection and Commissioning for Linear Accelerators: Technical Considerations and Recommendations Preface Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation
More informationMeasurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images
Measurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images T E Marchant, A M Amer 1 and C J Moore North Western Medical Physics, Christie Hospital NHS Foundation
More information3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH
3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk
More informationUse of MRI in Radiotherapy: Technical Consideration
Use of MRI in Radiotherapy: Technical Consideration Yanle Hu, PhD Department of Radiation Oncology, Mayo Clinic Arizona 04/07/2018 2015 MFMER slide-1 Conflict of Interest: None 2015 MFMER slide-2 Objectives
More informationAnalysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code
American Journal of Biomedical Engineering 216, 6(4): 124-131 DOI: 1.5923/j.ajbe.21664.3 Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code Ankit Kajaria 1,*, Neeraj Sharma
More informationDetermination of rotations in three dimensions using two-dimensional portal image registration
Determination of rotations in three dimensions using two-dimensional portal image registration Anthony E. Lujan, a) James M. Balter, and Randall K. Ten Haken Department of Nuclear Engineering and Radiological
More informationCT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION
CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology
More informationTomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS
Tomotherapy Physics Machine Twinning and Quality Assurance Emilie Soisson, MS Tomotherapy at UW- Madison Treating for nearly 5 years Up to ~45 patients a day on 2 tomo units Units twinned to facilitate
More informationAcknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF
Page 1 Quality Assurance of IMRT Delivery Systems - Siemens Lynn J. Verhey, Ph.D. Professor and Vice-Chair UCSF Dept. of Radiation Oncology AAPM 22 Annual Meeting, Montreal Acknowledgments Ping Xia, Ph.D.,
More informationImplementation of the EGSnrc / BEAMnrc Monte Carlo code - Application to medical accelerator SATURNE43
International Journal of Innovation and Applied Studies ISSN 2028-9324 Vol. 6 No. 3 July 2014, pp. 635-641 2014 Innovative Space of Scientific Research Journals http://www.ijias.issr-journals.org/ Implementation
More informationLOGIQ. V2 Ultrasound. Part of LOGIQ Vision Series. Imagination at work LOGIQ is a trademark of General Electric Company.
TM LOGIQ V2 Ultrasound Part of LOGIQ Vision Series Imagination at work The brilliance of color. The simplicity of GE. Now you can add the advanced capabilities of color Doppler to patient care with the
More informationNuclear Associates and
Nuclear Associates 84-317 and 84-317-7000 Multipurpose Tissue/Cyst Ultrasound Phantoms Users Manual February 2005 Manual No. 84-317-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in
More informationComprehensive treatment planning for brachytherapy. Advanced planning made easy
Comprehensive treatment planning for brachytherapy Advanced planning made easy Oncentra Brachy offers a variety of smart tools that facilitate many of the repetitive tasks for you. In contemporary brachytherapy,
More information