TTR500 Series Vector Network Analyzers Demonstration Guide

Size: px
Start display at page:

Download "TTR500 Series Vector Network Analyzers Demonstration Guide"

Transcription

1 xx ZZZ TTR500 Series Vector Network Analyzers Demonstration Guide *P *

2

3 xx ZZZ TTR500 Series Vector Network Analyzers Demonstration Guide Register now! Click the following link to protect your product

4 Copyright Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specifications and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. Contacting Tektronix Tektronix, Inc SW Karl Braun Drive P.O. Box 500 Beaverton, OR USA For product information, sales, service, and technical support: In North America, call Worldwide, visit to find contacts in your area.

5 Contents Check List for the Demo... 2 Setup... 3 One-Port Calibration... 4 STIMULUS SETUP... 4 ONE PORT CALIBRATION (REFLECTION TEST)... 5 SAVE AND RECALL CALIBRATION... 6 What You Will Learn With this guide, you will explore how to use Tektronix USB vector network analyzers to Set up the VNA One Port Measurements... 7 RETURN LOSS... 7 VSWR... 7 SMITH CHART / IMPEDANCE MEASUREMENT... 8 Two-Port Calibration STIMULUS SETUP TWO PORT CALIBRATION: 2-PORT 2-PATH SOLT APPLY PORT EXTENSIONS Make one port and two port calibrations Perform simple one port and two measurements on a bandpass filter. Two Port Measurements TRANSMISSION/ INSERTION LOSS BANDWIDTH/ CUT-OFF FREQUENCIES PHASE RESPONSE GROUP DELAY Offline Analysis Using Touchstone SnP files SAVE AND LOAD SNP FILE Bias Tee Setup The screenshots in this demo guide may not match the displays on your instrument due to the different instruments, devices under test, and test environments.

6 Check List for the Demo Before beginning the demo, please be sure you have the following items Tektronix TTR500 Series Vector Network Analyzer 4-in-1 OSLT Compact Cal Kit (Nmale) ( ) 60 cm N-male to N-male cables ( ) Type N Female to Female adapter ( INMET 5003) Type N Male to Female Adapter ( INMET 5005) 1560 to 1620 MHz BP Filter ( Mini-Circuits NBP-1560+) Type B USB 2.0 cable (174- ` ) 2

7 Setup 1 Set up the VNA a) Connect the TTR500 to a power source using the DC adapter. b) Use the USB 2.0 cable to connect the VNA to your Windows PC. The driver files are installed automatically and the installation wizard opens. Power Outlet USB 2.0 cable Windows-based PC TTR Install VectorVu-PC if not previously installed. Follow the prompts in the installation wizard to install VectorVu-PC software. You can also download and install VectorVu-PC from TekVISA is NOT necessary to run VectorVu-PC with TTR500. Connect VectorVu-PC to the VNA a) Open VectorVu-PC. b) The software should identify and connect to the TTR500 device automatically. If this does not happen, open Connection Manager (System > More > Connection) and connect the device manually. c) The connection indicator is green when there is an active connection (or when the device is in simulator mode). 3

8 One-Port Calibration To reduce errors in measurement, calibrate the VNA before you use it to measure device parameters. Calibration works to minimize Systematic Errors in the test setup, maximizing measurement accuracy. Stimulus setup Quick Tip: User interface of VectorVu- PC There are multiple ways to control VectorVu-PC. Use the icon to expand or contract the soft key panels. 1. Full Soft Key panel 1. Start the calibration with a factory preset. Under System, click Preset > OK. 2. In Stimulus, set Start and Stop frequencies to 100 MHz and 4.1 GHz. Press Enter or click the checkmark to confirm your selection. 2. Compact soft key panel 3. You can also access these menus from the menu bar. Under Stimulus, click Sweep Setup > Points and set points to 201. Quick tip: Frequency range 3. Under Stimulus, click Sweep Setup > Power Menu > Power Level and set Power level to 3 dbm A wider range of calibration increases calibration time. However, a short range puts you at the risk of making inaccurate measurements outside your calibration range. 4

9 One port calibration (reflection test) The calibration procedure fine-tunes the measurements made by a VNA by comparing measurements taken by the instrument with those taken by a known standard, in this case the Spinner 4-in-1 Cal Kit. This is a crucial step for all VNA measurements. 1. Under Response, click Cal > Cal Kit. 2. Select your desired calibration kit. In this example, we use OSLT N BN Repeat steps 6 and 7 using the Short and Load ports of the calibration kit. See images. When the measurement is complete, the VNA produces a beep and the instrument status bar displays. A check mark appears next to the standard measured in the port dialog box. 9. Apply the calibration data. The VNA creates and applies the calibration coefficients. The error correction function is also enabled. 3. Under Response, click Cal > Calibrate > 1- port SOL. 4. Click the port you want to calibrate. In this example, we use Port In the 1-port SOL Dialog Box, select the subclass of the calibration kit. In this example, we use 1 Male-Male. Quick Tip: Component Connection When working with precision RF cables and connectors it is important to minimize wear on the gold center conductor pin. The best practice for connecting and disconnecting cables is to isolate the conductor and spin the barrel to secure the threads, rather than spinning the entire cable and putting added wear on the center conductor. Connections should be snug but not over-tight. 6. Connect the Open port of the calibration kit to Port 1 of the VNA. See the images on the right side of the page for steps In the 1-port SOL dialog box click Open. The VNA records a trace measurement. 10. Disconnect the calibration kit and verify a flat 0 db S11 response. 5

10 Save and recall calibration Quick Tip: Limitations when recalling data To save a calibration file: 1. Under System, click Save/Recall > Save 2. In the Save dialog box select State + calibration (*.cstate) from the dropdown as the file format. 3. Click Save to store the calibration file. To recall a calibration file: 4. Click Preset > OK. 5. Under System, click Save/Recall > Recall. 6. In the Recall dialog box navigate to the saved.cstate file and click open to recall saved state. Calibration and state information are unique to an individual VNA at a particular time and in a specific environment. You cannot share calibration data (*.cstate files) between different units. The data saved in a VNA is specific to the environment of the instrument. When you recall the saved state of an instrument (*.state or *.dstate file) and reinstate these settings, the VNA can behave differently if the operating conditions have changed since the time the information was saved. 6

11 One Port Measurements Return Loss 4. Under Markers/Analysis, click Setup > Marker 1. Return loss is a linear reflection measurement. Lower return loss is better when designing a circuit intended to transmit signals. This is important for RF components, such as mixers, filters, and amplifiers. In this example, we use the band pass filter (Mini- Circuits 15542) provided in the demo kit. 1. Either continue from the last example, or recall One_Port_Cal.state from: 2. Connect the device under test to Port 1 of the VNA as depicted below: 5. Click and drag to move the marker on the display to answer these questions: What is the approximate return loss at the center of the passband? At what frequency is the magnitude of return loss the greatest? Band Pass Filter Load Note: Connect filter to 50 Ω load. VSWR The 1-port Measurement data can also be displayed linearly as Voltage Standing Wave Ratio (VSWR), which is the ratio of Forward/Reflected vs frequency. 1. Continue from the last example, or recall Return_Loss.state from: 3. Under Response, click Scale > Auto Scale. The return loss (S11 Log Mag) of the filter should be displayed on Trace 1 with a passband roughly between GHz as shown below: 2. Under Channel/Trace, click Trace > Trace Layout and select D1_2 as the layout. 3. Under Channel/Trace, click Trace > Num Traces and select 2 as the number of traces. The second trace appears as shown below: 7

12 4. Click on the title of Trace 2 to select it. It will be outlined in blue once selected. 8. Click and drag to move the marker on Trace 2 to answer the following questions: What is the SWR at the passband? What is the SWR at 3 db return loss? Smith Chart / Impedance measurement The Smith chart maps reflection coefficients of measurement data from the DUT to normalized impedances. Frequencies are indicated by markers. 5. Under Response, click Format > SWR 1. Continue from the last example, or recall VSWR.state from: 2. Click the title of Trace 2. Under Response, click Format > More > Smith > R+jX. 6. Click Scale > Auto Scale. Now you can see the SWR measurement displayed in the bottom display as shown below. Now you can see the smith chart in trace 2 with the impedance displayed. 7. With Trace 2 selected, under Markers/Analysis, click Setup > Marker 1. 8

13 3. Click on the title of Trace 1 to select it. It will be outlined in blue once selected. 4. Click and drag the marker on trace 1 to see the impedance change on trace 2, outlined in red below. Quick Tip: Invert Color/ Ink Saver Mode You can print screen images with a white background. Under Response, click Display > Invert Color and select Invert to change the background color from black to white. 5. With Trace 1 selected, under Markers/ Analysis, click Search. 6. Click Max to search the frequency point with the max return loss. Observe the impedance readout of the marker on Trace Click Min to search the frequency point with the minimum return loss. Observe the impedance readout of the marker on the Smith Chart on Trace 2. 9

14 Two-Port Calibration To reduce errors in measurement, calibrate the VNA before you use it to measure device parameters. Calibration works to minimize Systematic Errors in the test setup, maximizing measurement accuracy. Stimulus setup Quick Tip: User interface of VectorVu- PC There are multiple ways to control VectorVu-PC. Use the icon to expand or contract the soft key panels. 4. Full Soft Key panel 1. Start the calibration with a factory preset. Under System, click Preset > OK. 2. In Stimulus, set Start and Stop frequencies to 100 MHz and 4.1 GHz. Press Enter or click the checkmark to confirm your selection. 5. Compact soft key panel 6. You can also access these menus from the menu bar. 3. In Stimulus click Sweep Setup > Points and set points to 201. Quick tip: Frequency range 4. From Sweep Setup click Power Menu > Power Level and set Power level to 3 dbm A wider range of calibration increases calibration time. However, a short range puts you at the risk of making inaccurate measurements outside your calibration range. 10

15 Two port calibration: 2-port 2-path SOLT The 2-Port calibration is an Open, Short, Load and Thru measurement on both ports. 1. Under Response, click Cal > Cal Kit. Select the OSLT N BN cal kit. 2. Under Response, click Cal > Calibrate > 2- port SOLT. The 2-Port SOLT dialog box pops open. 3. Connect the Open port of the calibration kit to Port 1 of the VNA. See the image below. 4. In the 2-Port SOLT dialog box, click Open under Port 1. The VNA records a trace measurement. 9. Connect the N-type cable provided in the Demo Kit to Port Connect the N(f) to N(f) adapter (INMET 5003) to the other end of the N-type cable. 11. Connect the Open port of the calibration kit to the other end of the N(f) to N(f) adapter, as depicted below. 12. In the 2-Port SOLT dialog box, click Open under Port 2. Open N(f)-N(f) INMET Open 5. Connect the Short port of the calibration kit to Port 1 of the VNA. See the image below. 6. In the 2-Port SOLT dialog box, click Short under Port Connect the Short port of the calibration kit to other end of the N(f) to N(f) adapter, as depicted below. 14. In the 2-Port SOLT dialog box, click Short under Port 2. Short Short 7. Connect the Load port of the calibration kit to Port 1 of the VNA. See the image below. 8. In the 2-Port SOLT dialog box, click Load under Port 1. N(f)-N(f) INMET Connect the Load port of the calibration kit to other end of the N(f) to N(f) adapter, as depicted below. 16. In the 2-Port SOLT dialog box, click Load under Port 2. Load Load N(f)-N(f) INMET

16 17. Connect the Thru standard of the calibration kit between Port 1 and the N(f) to N(f) adapter, as depicted below. 18. In the 2-Port SOLT dialog box, click Thru. 19. Click Apply. The 2-Port SOLT calibration is now complete. Thru Thru N(f)-N(f) INMET Isolation Calibration The isolation calibration process is removes the errors due to crosstalk between test ports in a transmission measurement (S21/S12). Crosstalk can be a problem when testing high dynamic range devices such as filters with a high level stopband. In a two-port calibration, you have the option to omit the isolation calibration. Since the Demo Kit does not contain the required amount of load standards to perform this step, it is omitted here. Please refer to the TTR500 User Manual for more information. Save calibration To save a calibration file: 1. Under System, click Save/Recall > Save 2. In the Save dialog box select State + calibration (*.cstate) from the dropdown as the file format. 3. Click Save to store the calibration file. Quick Tip: Limitations when recalling data Calibration and state information are unique to an individual VNA at a particular time and in a specific environment. You cannot share calibration data (*.cstate files) between different units. The data saved in a VNA is specific to the environment of the instrument. When you recall the saved state of an instrument (*.state or *.dstate file) and reinstate these settings, the VNA can behave differently if the operating conditions have changed since the time the information was saved. 12

17 Apply Port Extensions In some cases, you may need to apply a port extension to compensate for connection changes going from the user calibration to the measurement setup. For example: The DUT may be in a test fixture. The port gender for the cal kit may be different from the DUT s. You performed a calibration but need to add a length of transmission line to the test setup. Port extension is a function usually done after a user calibration. It compensates for the phase response change that occurs when the calibration reference plane is not the same as the measurement plane of the DUT. In this example, we use port extensions to compensate for the difference in length between the N(f)-N(f) adaptor (used in calibration) and the N(f)- N(m) adaptor (used for the filter measurements). Due to the electrical length difference between the N(f) to N(f) adaptor used during the calibration and the N(f) to N(m) adaptor we will use to make two port measurements, the phase response is not completely flat at 0 degrees. An additional calibration step, known as port extension, must be performed to remove the linear portion of the phase response. 1. Continue from the last example. 2. Remove the Cal kit and the N(f) to N(f) adaptor from the test setup. 5. Under Response, click Format > Phase (Deg) to measure the Phase (Deg) response of S21. Note that the phase response of S21 without any DUT is linear, but not completely flat (0 degrees). 6. With Trace 2 selected, click Format > Group Delay. 7. Under Response, click Scale > Auto Scale. 8. Under Response, click Avg > Enable Smoothing. 9. Estimate and record the average group delay. In this example, it is about 25 ps. 3. Connect the N(f) to N(m) adaptor between port 1 of VNA and the other end of the cable. N(f)-N(m) INMET Click on the title of Trace 2 to select it. It will be outlined in blue once selected. 13

18 10. Under Response, click Cal > Port Extension. 11. Enter the group delay measured in Step 9 as the Delay Time of Port Check Enable Port Extension, and click OK to apply settings. Note: Once the port extension is applied, the group delay of S21 should be close to zero. 13. Click Format > Phase (Deg), the phase response of S21 should be a flat 0 response. 14

19 Two Port Measurements Transmission/ Insertion Loss Insertion loss is often used to describe the characteristics of a filter. It is usually considered to be the same as the inverse of the log magnitude format of S21. The value of insertion loss is usually a positive number. 1. Continue from the last example. 2. Attach the male end of the filter to Port 1 and attach the female end of the filter to the cable on Port 2. Use the male-to-female adapter included in the demo kit. What is the type of this filter? (bandpass, low pass, high pass?) Band Pass Filter N(f)-N(m) INMET 5005 Option 1: Connect the other end of filter to the cable using the N(f)-N(m) adapter Under Markers/Analysis, click Setup > Marker Move the Marker to measure insertion loss at the pass band. 9. Set the Start frequency to 850 MHz and the Stop frequency to 2450 MHz to zoom in. 3. Under Channel/Trace, click Trace > Num Traces and set the number of traces to 1 to see only Trace Click the title of Trace Under Response, click Measure. Select S21 as the measurement. 6. Under Response, click Scale > Auto Scale. 15

20 Bandwidth/ Cut-off Frequencies In VectorVu-PC, you can perform a bandwidth search to determine the bandwidth of a trace, and identify the parameters associated with a stopband or passband. 1. Continue from the last example. 2. Click Search under Markers/Analysis. 3. Click Bandwidth to make the bandwidth measurement automatically. Phase Response For components with free linear distortion transmission, other than observing the magnitude response, phase response is also a very important measurement. The phase response must be linear over the desired transmission bandwidth. 1. Continue from the last example. 2. Under Channel/Trace, click Trace > Num Traces and set the number of traces to Under Channel/Trace, click Trace > Trace Layout and select D1_2 as the layout. 4. Under Markers/Analysis, click Search > Bandwidth to turn off the Bandwidth measurement, and click the title of Trace Under Response, click Measure > S Under Markers/Analysis, click Setup > Marker Under Response, click Format > Phase (Deg). 4. You can also click More > Bandwidth Value, to define a measurement bandwidth. In this example, we use 3 db. The VNA will display high and low frequencies, the center frequency, bandwidth, insertion loss, and Q value in the upper left corner of the trace window. How does the general shape of the phase response correlate with the magnitude response shown above? 16

21 Group Delay Group delay is the time delay measuring the phase distortion of RF components. Ideally, all frequencies present in the signal should have the same time delay so that the signal will not be distorted. 1. Continue from the last example. 2. Click the title of Trace Under Response, click Avg. 8. Click Smoothing Aperture, and enter 20 as the aperture value. 9. Click Enable Smoothing. Observe the difference after applying the large aperture smoothing. 3. Under Response, click Format. 4. Select Group Delay as the measurement format. 5. Under Response, click Scale > Auto Scale. Group delay aperture Increasing the aperture of the group delay smoothing settings makes the measurement less susceptible to noise, but reduces the resolution of the group delay measurement. The minimum aperture is equal to the frequency span divided by the number of points minus one. 6. Observe the average group delay to answer the following questions: How would you specify this filter in terms of the group delay across the passband? Group delay across passband = ns 17

22 Offline Analysis Using Touchstone SnP files Save and Load SnP File You can save s-parameter information from the VNA for 1 port and 2 port models in the touchstone format. Touchstone files contain the s-parameters of the DUT in a space-separated format. The number of ports used in the measurement determines the matrix of s-parameters for the DUT and consequently the format of the touchstone file. Save SnP File 1. Under System, click Save / Recall. 2. Click Save SnP. 3. Select the format of the touchstone file (S1P or S2P). In this example, we save the file as S2P. Touchstone file format In a touchstone file, the behavior of the DUT is completely represented by the collection of its S- parameters. You can import the touchstone file into a data modeling program or circuit simulator like MATLAB. In order to save all the S-parameters, you first need to run the calibration of the desired port(s). For example, to save the S2P file, a full 2-port 2-path calibration needs to be performed first. Once the calibration is complete, run any measurement you like, and the S-parameters will be able to be saved to an S2P file. Load SnP File 5. Under System, click More. 6. Click Connections. 7. In the Connection Manager, select the simulator, and click Connect to use the offline S2P file in VectorVu-PC. 8. Under System, click More. 9. Click Simulator. 10. Click Load SnP and open the S2P file. 4. Save the file. Go to the folder location and open the saved file as a text file. Observe the header and S-parameters in the file. 18

23 Bias Tee Setup Sometimes measuring active DUTs requires a DC bias voltage. With the TTR500, a DC bias can easily be fed through the bias tee inputs located on the rear panel. The DC voltage is supplied to the DUTs via the inner conductor of the VNA and can be supplied to either port. This is very useful when both DC and RF signals must be applied to a DUT, eliminating the need for external bias tee circuitry. To Turn on the Bias Tee, 1. Under System, Click More > Bias Tee. The warning popup should appear. The figures below show the simplified power amplifier test setup with and without using the internal bias tee of the VNA. Power Supply Bias 2 Resistor Power amplifier (DUT) Using the built-in bias tee of the VNA 2. After you have read and followed the directions outlined in the popup window, select the box and click OK. Power Supply The bias tee is now able to source power to either port 1 or port 2. C (DC Block) Power amplifier (DUT) C (DC Block) L (RF Choke) R Using an external bias tee 19

DSA8300 Digital Serial Analyzer TekScope and Connectivity Software Installation

DSA8300 Digital Serial Analyzer TekScope and Connectivity Software Installation xx ZZZ DSA8300 Digital Serial Analyzer TekScope and Connectivity Software Installation *P071289803* 071-2898-03 Copyright Tektronix. All rights reserved. Licensed software products are owned by Tektronix

More information

Keysight MOI for USB 2.0 Connectors & Cable Assemblies Compliance Tests

Keysight MOI for USB 2.0 Connectors & Cable Assemblies Compliance Tests Revision 1.10 October 18, 2016 Universal Serial Bus Specification Revision 2.0 Keysight Method of Implementation (MOI) for USB 2.0 Connectors and Cables Assemblies Compliance Tests Using Keysight E5071C

More information

FCA3000, FCA3100, and MCA3000 Series Connectivity Installation Manual

FCA3000, FCA3100, and MCA3000 Series Connectivity Installation Manual xx ZZZ FCA3000, FCA3100, and MCA3000 Series Connectivity Installation Manual *P077049700* 077-0497-00 xx ZZZ FCA3000, FCA3100, and MCA3000 Series Connectivity Installation Manual www.tektronix.com 077-0497-00

More information

BSXUSB31 & BSXPCIE4CEM Receiver Testing Applications Instructions

BSXUSB31 & BSXPCIE4CEM Receiver Testing Applications Instructions xx ZZZ BSXUSB31 & BSXPCIE4CEM Receiver Testing Applications Instructions *P077136501* 077-1365-01 xx ZZZ BSXUSB31 & BSXPCIE4CEM Receiver Testing Applications Instructions Register now! Click the following

More information

and above SSD Remote Mounting Kit Instructions

and above SSD Remote Mounting Kit Instructions 016-2102-00 and above SSD Remote Mounting Kit Instructions Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you

More information

TDSUSBF USB 2.0 Test Fixture Socket Upgrade Kit

TDSUSBF USB 2.0 Test Fixture Socket Upgrade Kit Instructions 040-1717-00 TDSUSBF USB 2.0 Test Fixture Socket Upgrade Kit 075-0901-00 Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform

More information

DPO-UP Upgrades for MSO70000/C/DX, DSA70000/B/C/D, DPO70000/B/C/ D/DX/SX, DPO7000/C, MSO5000/B, and DPO5000/B Series Instruments Instructions

DPO-UP Upgrades for MSO70000/C/DX, DSA70000/B/C/D, DPO70000/B/C/ D/DX/SX, DPO7000/C, MSO5000/B, and DPO5000/B Series Instruments Instructions DPO-UP Upgrades for MSO70000/C/DX, DSA70000/B/C/D, DPO70000/B/C/ D/DX/SX, DPO7000/C, MSO5000/B, and DPO5000/B Series Instruments Instructions Warning The servicing instructions are for use by qualified

More information

DMM4020, DMM4040, and DMM4050 Connectivity Installation Manual

DMM4020, DMM4040, and DMM4050 Connectivity Installation Manual DMM4020, DMM4040, and DMM4050 Connectivity Installation Manual *P077039502* 077-0395-02 DMM4020, DMM4040, and DMM4050 Connectivity Installation Manual www.tektronix.com 077-0395-02 Copyright Tektronix.

More information

ArbExpress AXW100 Waveform Creation and Editing Tool for Tektronix AWG/AFG Installation Manual

ArbExpress AXW100 Waveform Creation and Editing Tool for Tektronix AWG/AFG Installation Manual xx ZZZ ArbExpress AXW100 Waveform Creation and Editing Tool for Tektronix AWG/AFG Installation Manual *P077069700* 077-0697-00 xx ZZZ ArbExpress AXW100 Waveform Creation and Editing Tool for Tektronix

More information

PQA600A and PQASW Picture Quality Analyzer Application Install

PQA600A and PQASW Picture Quality Analyzer Application Install xx ZZZ PQA600A and PQASW Picture Quality Analyzer Application Install *P071312100* 071-3121-00 Copyright Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries

More information

SourceXpress Waveform Creation Application Printable Help Document

SourceXpress Waveform Creation Application Printable Help Document xx ZZZ SourceXpress Waveform Creation Application Printable Help Document *P077114502* 077-1145-02 ZZZ SourceXpress Waveform Creation Application Printable Help Document www.tek.com 077-1145-02 Copyright

More information

DPO70000/B and DSA70000/B and Above Power Supply Replacement Kit Instructions

DPO70000/B and DSA70000/B and Above Power Supply Replacement Kit Instructions xx ZZZ DPO70000/B and DSA70000/B 050-3719-00 and Above Power Supply Replacement Kit Instructions www.tektronix.com *P075099500* 075-0995-00 Copyright Tektronix. All rights reserved. Licensed software products

More information

AWG701AUP AWG702AUP Options 01, 03, and SSD-0x AWG70000A Series Upgrades Instructions

AWG701AUP AWG702AUP Options 01, 03, and SSD-0x AWG70000A Series Upgrades Instructions xx ZZZ AWG701AUP AWG702AUP Options 01, 03, and SSD-0x AWG70000A Series Upgrades Instructions www.tek.com *P075104903* 075-1049-03 Copyright Tektronix. All rights reserved. Licensed software products are

More information

Tektronix Bench Instruments PC Connectivity Installation Manual

Tektronix Bench Instruments PC Connectivity Installation Manual xx ZZZ Tektronix Bench Instruments PC Connectivity Installation Manual *P077052700* 077-0527-00 xx ZZZ Tektronix Bench Instruments PC Connectivity Installation Manual www.tektronix.com 077-0527-00 Copyright

More information

TDSPTD Protocol Trigger and Decode Application Online Help

TDSPTD Protocol Trigger and Decode Application Online Help xx ZZZ TDSPTD Protocol Trigger and Decode Application Online Help *P077002502* 077-0025-02 ZZZ TDSPTD Protocol Trigger and Decode Application Online Help www.tektronix.com 077-0025-02 Copyright Tektronix.

More information

VNA0440 VNA0440e VNA0460 VNA0460e User Manual Version 3.0

VNA0440 VNA0440e VNA0460 VNA0460e User Manual Version 3.0 User Manual Version 3.0 - 2 - www.megiq.com How to use the MegiQ VNA0440 (e) VNA0460 (e) Vector Network Analyzers o General information o User software o Using the Instrument MegiQ BV Weegschaalstraat

More information

5 Series MSO Option SUP5-WIN SSD with Microsoft Windows 10 Upgrade Kit Installation Instructions

5 Series MSO Option SUP5-WIN SSD with Microsoft Windows 10 Upgrade Kit Installation Instructions 5 Series MSO Option SUP5-WIN SSD with Microsoft Windows 10 Upgrade Kit Installation Instructions Supports Product Firmware V1.4 and above *P071352400* 071-3524-00 Copyright Tektronix. All rights reserved.

More information

HP 8714B Calibration

HP 8714B Calibration Wireless Circuits and Systems Laboratory Procedure #7 Calibrating the HP 8714 Network Analyzer This procedure describes the different types of calibrations that are available with the HP 8714 vector network

More information

DSA8200 Digital Serial Analyzer

DSA8200 Digital Serial Analyzer DVD DSA8200 Digital Serial Analyzer Microsoft Windows XP Operating System Restore 071-2052-00 www.tektronix.com Copyright Tektronix. All rights reserved. Licensed software products are owned by Tektronix

More information

AWG5000 and AWG7000 Series Windows XP Operating System Restore

AWG5000 and AWG7000 Series Windows XP Operating System Restore DVD AWG5000 and AWG7000 Series Windows XP Operating System Restore 071-2084-01 www.tektronix.com Copyright Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries

More information

RSA5100B Series Real-Time Signal Analyzers RSA5BUP Option 300 High Performance Real Time Upgrade Instructions

RSA5100B Series Real-Time Signal Analyzers RSA5BUP Option 300 High Performance Real Time Upgrade Instructions xx ZZZ RSA5100B Series Real-Time Signal Analyzers RSA5BUP Option 300 High Performance Real Time Upgrade Instructions www.tektronix.com *P075105800* 075-1058-00 Copyright Tektronix. All rights reserved.

More information

RSA6000A Series TDS5000B Series DSA8200 Hard Disk Drive Adapter Upgrade Kit XX Instructions

RSA6000A Series TDS5000B Series DSA8200 Hard Disk Drive Adapter Upgrade Kit XX Instructions xx ZZZ RSA6000A Series TDS5000B Series DSA8200 Hard Disk Drive Adapter Upgrade Kit 050-3803-XX Instructions www.tektronix.com *P075105100* 075-1051-00 Copyright Tektronix. All rights reserved. Licensed

More information

Keysight MOI for USB Type-C Cable Assemblies Compliance Tests Using Keysight M937XA Multiport PXIe VNA

Keysight MOI for USB Type-C Cable Assemblies Compliance Tests Using Keysight M937XA Multiport PXIe VNA Revision 1.01 Feb-24, 2017 Universal Serial Bus Type-C TM Specification Revision 1.1 Keysight Method of Implementation (MOI) for USB Type-C TM Cables Assemblies Compliance Tests Using Keysight For Type-C

More information

*P * ZZZ

*P * ZZZ xx ZZZ CTRL7100UP X-RAID-B and X-RAID-C Option B. 10 1-TB drives and Option C.16 2-TB drives for the CTRL7100A RSA7100A Real-Time Spectrum Analyzer Upgrade Kit Instructions www.tek.com *P071355300* 071-3553-00

More information

Direct Connection of OML Frequency Extenders to Keysight s PNA-C (E836xC) VNA for 1-Path 2-Port S-Parameter Measurements

Direct Connection of OML Frequency Extenders to Keysight s PNA-C (E836xC) VNA for 1-Path 2-Port S-Parameter Measurements Direct Connection of OML Frequency Extenders to Keysight s PNA-C (E836xC) VNA for 1-Path 2-Port S-Parameter Measurements Introduction For full two-port S-parameter measurements to extend PNA-C vector network

More information

TDS5000B Series SATA Hard Disk Drive Replacement Kit Instructions

TDS5000B Series SATA Hard Disk Drive Replacement Kit Instructions xx ZZZ TDS5000B Series 050-3708-00 SATA Hard Disk Drive Replacement Kit Instructions www.tektronix.com *P075099100* 075-0991-00 Copyright Tektronix. All rights reserved. Licensed software products are

More information

AKELA Vector Network Analyzer (VNA) Quick Start Guide

AKELA Vector Network Analyzer (VNA) Quick Start Guide AKELA Vector Network Analyzer (VNA) Quick Start Guide Copyright AKELA, Inc. 2012, all rights reserved http:\\akelainc.com LabVIEW and LabWindows are registered trademarks of National Instruments Incorporated

More information

UNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY

UNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY UNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY HIGH TEMPERATURE & HIGH FREQUENCY TESTING SYSTEM LABORATORY USER MANUAL 4/29/2011 Overview

More information

DSA8300 Digital Serial Analyzer Microsoft Windows 7 Ultimate Operating System Restore Instructions

DSA8300 Digital Serial Analyzer Microsoft Windows 7 Ultimate Operating System Restore Instructions xx ZZZ DSA8300 Digital Serial Analyzer Microsoft Windows 7 Ultimate Operating System Restore Instructions *P071289900* 071-2899-00 Copyright Tektronix. All rights reserved. Licensed software products are

More information

TLA5Fxxx, TLA62F0x, & TLA7Fxxx PowerFlex Field Upgrade Kit Logic Analyzers Instructions

TLA5Fxxx, TLA62F0x, & TLA7Fxxx PowerFlex Field Upgrade Kit Logic Analyzers Instructions xx ZZZ TLA5Fxxx, TLA62F0x, & TLA7Fxxx PowerFlex Field Upgrade Kit Logic Analyzers Instructions www.tektronix.com *P077059110* 077-0591-10 Copyright Tektronix. All rights reserved. Licensed software products

More information

PC Courseware Editor User Manual

PC Courseware Editor User Manual xx ZZZ PC Courseware Editor User Manual *P077099300* 077-0993-00 xx ZZZ PC Courseware Editor User Manual Register now! Click the following link to protect your product. www.tektronix.com/register www.tektronix.com

More information

Tektronix DPO Demo 1 Board Instruction Manual

Tektronix DPO Demo 1 Board Instruction Manual xx ZZZ Tektronix DPO Demo 1 Board Instruction Manual www.tektronix.com *P071253900* 071-2539-00 Copyright Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries

More information

Service Manual. P GHz Differential Probe

Service Manual. P GHz Differential Probe Service Manual P7330 3.5 GHz Differential Probe 071-1057-00 Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you

More information

, , and and above Rackmount Kits Instructions

, , and and above Rackmount Kits Instructions xx ZZZ 016-1983-01, 016-1985-01, and 016-1988-01 and above Rackmount Kits Instructions Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform

More information

TDS820 Digitizing Oscilloscope Declassification and Security Instructions

TDS820 Digitizing Oscilloscope Declassification and Security Instructions xx ZZZ TDS820 Digitizing Oscilloscope Declassification and Security Instructions www.tektronix.com 077-0109-00 Copyright Tektronix. All rights reserved. Licensed software products are owned by Tektronix

More information

Keysight MOI for USB Type-C Cable Assemblies Compliance Tests Using Keysight M937XA Multiport PXIe VNA

Keysight MOI for USB Type-C Cable Assemblies Compliance Tests Using Keysight M937XA Multiport PXIe VNA Revision 1.00 Apr-14, 2016 Universal Serial Bus Type-C TM Specification Revision 1.1 Keysight Method of Implementation (MOI) for USB Type-C TM Cables Assemblies Compliance Tests Using Keysight For Type-C

More information

RF Characterization Report

RF Characterization Report RF058 Series Cable Assemblies RF058-01BJ1-01BJ1-0150 RF058-01SB1-01SB1-0150 RF058-01SP1-01SP1-0150 Description: RF Cable Assembly, 50Ohm, RG058 Coaxial Cable Samtec Inc. WWW.SAMTEC.COM Phone: 812-944-6733

More information

Agilent Technologies S-Parameter and TDR Impedance Measurement Solution Summary

Agilent Technologies S-Parameter and TDR Impedance Measurement Solution Summary Agilent Technologies S-Parameter and TDR Impedance Measurement Solution Summary Built-in S-parameter testing Easy and accurate transmission channel/media characterization Transmission lines for high speed

More information

Software Release Notes

Software Release Notes Fixed a bug related to IFBW settings below 10 Hz in the S5048 and S7530 models Changed the application icon Version 15.4.0 IMPORTANT: All instruments in the (reversing 2-Port VNAs) now share a common installer

More information

Report Generator for DPOPWR

Report Generator for DPOPWR Online Help Report Generator for DPOPWR Adapted from the Report Generator for DPOPWR Online Help www.tektronix.com Copyright Tektroni x. All rights reserved. Licensed software products are owned by Tektronix

More information

ArbExpressr AXW100 Waveform Creation and Editing Tool for Tektronix AWG/AFG Version

ArbExpressr AXW100 Waveform Creation and Editing Tool for Tektronix AWG/AFG Version Manual ArbExpressr AXW100 Waveform Creation and Editing Tool for Tektronix AWG/AFG Version 2.3 061-4288-03 www.tektronix.com *P061428803* 061428803 Copyright Tektronix. All rights reserved. Licensed software

More information

TekExpress DiiVA Automated Solution

TekExpress DiiVA Automated Solution x TekExpress DiiVA Automated Solution ZZZ Quick Start User Manual *P077050800* 077-0508-00 xx TekExpress DiiVA Automated Solution ZZZ Quick Start User Manual www.tektronix.com 077-0508-00 Copyright Tektronix.

More information

Copyright ANRITSU. All rights reserved.

Copyright ANRITSU. All rights reserved. General Caution --------------- Please use a USB Memory Stick for firmware updates. If there is a firmware update question, please contact Anritsu service support at https://techsupport.anritsu.com, and

More information

TDS5BUP and TDS5UP Option 1P TDS5000B and TDS5000 Series Integrated Thermal Printer Upgrade and Thermal Printer Replacement Kit

TDS5BUP and TDS5UP Option 1P TDS5000B and TDS5000 Series Integrated Thermal Printer Upgrade and Thermal Printer Replacement Kit Instructions TDS5BUP and TDS5UP Option 1P TDS5000B and TDS5000 Series Integrated Thermal Printer Upgrade and Thermal Printer Replacement Kit 075-0814-03 Warning The servicing instructions are for use by

More information

TN Vector Network Analyzer Calibrations and Their Effect on Measurement Results. Bernie McDermott. May 2010

TN Vector Network Analyzer Calibrations and Their Effect on Measurement Results. Bernie McDermott. May 2010 TN 201005 Vector Network Analyzer Calibrations and Their Effect on Measurement Results May 2010 Bernie McDermott Lead Metrologist Calibration Products and Services Measurement Examples and Technical Notes

More information

Measurement Accuracy of Vector Network Analysis. Thilo Bednorz

Measurement Accuracy of Vector Network Analysis. Thilo Bednorz Measurement Accuracy of Vector Network Analysis Thilo Bednorz EDICON 2019 VNA Measurement and Accuracy Error sources occuring with VNA measurement measurement errors stochastic errors systematic errors

More information

Copyright ANRITSU. All rights reserved.

Copyright ANRITSU. All rights reserved. --------------- ShockLine VNA Software If there is a software/firmware update question, please contact Anritsu service support at https://techsupport.anritsu.com, and then click on "Technical Support"

More information

Online Help. Protocol Trigger and Decode Adapted from the Protocol Trigger and Decode Online Help (May, 2005)

Online Help. Protocol Trigger and Decode Adapted from the Protocol Trigger and Decode Online Help (May, 2005) Online Help Protocol Trigger and Decode 077-0009-01 Adapted from the Protocol Trigger and Decode Online Help (May, 2005) www.tektronix.com Table of Contents General Safety Summary 3 Welcome to the Protocol

More information

spinner Measurement & Calibration equipment for network analyzers

spinner Measurement & Calibration equipment for network analyzers spinner Measurement & Calibration equipment for network analyzers Edition B 2011 High Frequency Performance Worldwide www.spinner-group.com SPINNER test & calibration equipment for best measurement results

More information

Service Manual. P GHz Active Probe

Service Manual. P GHz Active Probe Service Manual P7240 4 GHz Active Probe 071-1056-00 Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified

More information

S2 Family Reversing 2-port VNAs of Planar, S, and Cobalt Series

S2 Family Reversing 2-port VNAs of Planar, S, and Cobalt Series Version 18.3.1 Fixed a bug with the Vector Mixer Calibration menu. Now the *.s2p file of the mixer automatically de-embeds upon completing the calibration process Version 18.2.4 Added T4311 and Z5411 to

More information

Advanced Test Equipment Rentals ATEC (2832) Manual Supplement. Agilent E5070B/E5071B ENA Series RF Network Analyzers

Advanced Test Equipment Rentals ATEC (2832) Manual Supplement. Agilent E5070B/E5071B ENA Series RF Network Analyzers Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Agilent E5070B/E5071B ENA Series RF Network Analyzers Manual Supplement Second Edition Agilent Part No. E5070-90045

More information

TDSHT3 HDMI Compliance Test Software Programmer Online Help

TDSHT3 HDMI Compliance Test Software Programmer Online Help xx ZZZ TDSHT3 HDMI Compliance Test Software Programmer Online Help *P077035301* 077-0353-01 ZZZ TDSHT3 HDMI Compliance Test Software Programmer Online Help www.tektronix.com 077-0353-01 Copyright Tektronix.

More information

RF Probing With Rohde & Schwarz ZNB VNA. PacketMicro, Inc Wyatt Drive, Suite 9, Santa Clara, CA

RF Probing With Rohde & Schwarz ZNB VNA. PacketMicro, Inc Wyatt Drive, Suite 9, Santa Clara, CA RF Probing With Rohde & Schwarz ZNB VNA PacketMicro, Inc. 1900 Wyatt Drive, Suite 9, Santa Clara, CA 95054 www.packetmicro.com Outline Why RF Probing Page 3 4 S-Probe Overview Page 5-8 RF Probing Tips

More information

Copyright ANRITSU. All rights reserved.

Copyright ANRITSU. All rights reserved. ShockLine VNA Software If there is a software/firmware update question, please contact Anritsu service support at https://techsupport.anritsu.com, and then click on "Technical Support" for further information.

More information

WFM520UP and WVR520UP WFM5200 Waveform Monitor and WVR5200 Waveform Rasterizer Upgrade Kit Instructions

WFM520UP and WVR520UP WFM5200 Waveform Monitor and WVR5200 Waveform Rasterizer Upgrade Kit Instructions xx ZZZ WFM520UP and WVR520UP WFM5200 Waveform Monitor and WVR5200 Waveform Rasterizer Upgrade Kit Instructions Warning The servicing instructions are for use by qualified personnel only. To avoid personal

More information

TLA700 Series Tektronix Logic Analyzer Declassification and Security Instructions

TLA700 Series Tektronix Logic Analyzer Declassification and Security Instructions TLA700 Series Tektronix Logic Analyzer Declassification and Security Instructions www.tektronix.com 077-0038-00 Copyright Tektronix. All rights reserved. Licensed software products are owned by Tektronix

More information

Application Note FOR SMA Calibration Kit for VNA. LCAL06A A female calibration kit and used for DC ~ 6.0 GHz VNA test cable with male end connector.

Application Note FOR SMA Calibration Kit for VNA. LCAL06A A female calibration kit and used for DC ~ 6.0 GHz VNA test cable with male end connector. Application Note FOR SMA Calibration Kit for VNA LCAL06A, LCAL06B, LCAL06C and LCAL09A are precision SMA calibration kits which provide accurate and reliable Short-Open-Load-Thru (SOLT) or Load-Reflect-Match

More information

RM3100 Rackmount Kit for the AFG3000 and AFG3000C Series Arbitrary Function Generators

RM3100 Rackmount Kit for the AFG3000 and AFG3000C Series Arbitrary Function Generators xx RM3100 Rackmount Kit for the AFG3000 and AFG3000C Series Arbitrary Function Generators ZZZ Instructions Register now! Click the following link to protect your product. www.tektronix.com/register www.tektronix.com

More information

3.5mm VNA Calibration Kits

3.5mm VNA Calibration Kits 3.5mm VNA Calibration Kits DATA SHEET / 2Z-059 Models: 8050CK10 Fixed Load Kit 8050CK11 Fixed Load Kit Plus Adapters 8050CK20 Sliding Load Kit - Basic 8050CK21 Sliding Load Kit Plus Adapters 8050CK30 TRL

More information

2.92mm VNA Calibration Kits

2.92mm VNA Calibration Kits 2.92mm VNA Calibration Kits DATA SHEET / 2Z-058B Models: 8770CK10 Fixed Load Kit 8770CK11 Fixed Load Kit Plus Adapters 8770CK20 Sliding Load Kit 8770CK21 Sliding Load Kit Plus Adapters 8770CK30 TRL Kit

More information

RSA5100A Series Real-Time Signal Analyzers Declassification and Security Instructions

RSA5100A Series Real-Time Signal Analyzers Declassification and Security Instructions xx ZZZ RSA5100A Series Real-Time Signal Analyzers Declassification and Security Instructions www.tektronix.com *P077052100* 077-0521-00 Copyright Tektronix. All rights reserved. Licensed software products

More information

Multiport Testing with CMT VNAs and Off-the-shelf RF Switches

Multiport Testing with CMT VNAs and Off-the-shelf RF Switches Introduction Copper Mountain Technologies offers VNA with one port, two ports or four ports. But sometimes, more ports are required, for example for testing multi-port devices. There are many options to

More information

DSA70000D, DPO70000D/DX, MSO70000C/DX, DSA70000C, DPO70000C, and DPO7000C High Performance Oscilloscopes Best Practices Manual

DSA70000D, DPO70000D/DX, MSO70000C/DX, DSA70000C, DPO70000C, and DPO7000C High Performance Oscilloscopes Best Practices Manual xx ZZZ DSA70000D, DPO70000D/DX, MSO70000C/DX, DSA70000C, DPO70000C, and DPO7000C High Performance Oscilloscopes Best Practices Manual *P071298901* 071-2989-01 xx ZZZ DSA70000D, DPO70000D/DX, MSO70000C/DX,

More information

FieldFox N9923A RF Vector Network Analyzer

FieldFox N9923A RF Vector Network Analyzer Quick Reference Guide FieldFox N9923A RF Vector Network Analyzer Contents Do You Have Everything?... 2 The Power Button and LED... 2 Battery Usage... 3 Measure S-Parameters... 4 Multi-Trace Configurations...

More information

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED

More information

New Features in Firmware V2.02 (Compared to V2.01) New Features in Firmware V2.01 (Compared to V2.00)

New Features in Firmware V2.02 (Compared to V2.01) New Features in Firmware V2.01 (Compared to V2.00) What's New in this Revision This help describes version V2.10 of the R&S ZVA firmware. Compared to the previous version V2.02, the following True differential mode (option R&S ZVA-K6) Automatic calibration

More information

NPRT 2200 Operation Manual

NPRT 2200 Operation Manual NPRT 2200 Operation Manual Module Version NPRT 2200 4.05 NprGraph 2.11 SwpEditor 2.00 12/02/2016 Table of Contents 1. Introduction...1 1.1. NPR Measurement...1 1.2. NPRT Block Diagram...1 1.3. Power Sweep

More information

*P *

*P * TBS1000, TBS1000B/TBS1000B-EDU, TDS1000B/TDS2000B, TDS1000C-EDU/TDS2000C, and TPS2000B Series Digital Storage Oscilloscope Declassification and Security Instructions *P077077102* 077-0771-02 TBS1000,

More information

Tektronix 4000 Series Oscilloscopes Declassification and Security Instructions

Tektronix 4000 Series Oscilloscopes Declassification and Security Instructions xx ZZZ Tektronix 4000 Series Oscilloscopes Declassification and Security Instructions Revision B www.tektronix.com 071-2196-00 Copyright Tektronix. All rights reserved. Licensed software products are owned

More information

FMCK1013 DATA SHEET. Features: Applications: Configuration. Mechanical Specifications

FMCK1013 DATA SHEET. Features: Applications: Configuration. Mechanical Specifications Short Open Load Thru (SOLT) Analyzer Calibration Kit with 3.5mm Male Connections in a portable 4-in-1 Design and Operating to 26.5 GHz Short Open Load Thru (SOLT) Analyzer Calibration Kit, part number

More information

Differential Probes TDP1500 TDP3500 Data Sheet

Differential Probes TDP1500 TDP3500 Data Sheet Differential Probes TDP1500 TDP3500 Data Sheet Features & Benefits Outstanding Electrical Performance 3.5 GHz and 1.5 GHz Bandwidth Models Accurate Measurements for Serial and Digital Applications Excellent

More information

+14dBm to +20dBm LO Buffers/Splitters with ±1dB Variation

+14dBm to +20dBm LO Buffers/Splitters with ±1dB Variation -24; Rev 2; 3/4 +dbm to +dbm LO Buffers/Splitters General Description The MAX9987 and MAX9988 LO buffers/splitters each integrate a passive two-way power splitter with highisolation input and output buffer

More information

Keysight Technologies On-Wafer Testing of Opto-Electronic Components Using the Lightwave Component Analyzers. Application Note

Keysight Technologies On-Wafer Testing of Opto-Electronic Components Using the Lightwave Component Analyzers. Application Note Keysight Technologies On-Wafer Testing of Opto-Electronic Components Using the Lightwave Component Analyzers Application Note Introduction When measurements of optoelectronic components are performed on

More information

S3101 Cable & Antenna Tester Datasheet

S3101 Cable & Antenna Tester Datasheet S3101 Cable & Antenna Tester Datasheet Saluki Technology Inc. The datasheet applies to the cable and antenna analyzer of the following models: S3101A cable and antenna analyzer (1MHz - 4GHz). S3101B cable

More information

100BASE-T1 EMC Test Specification for ESD suppression devices

100BASE-T1 EMC Test Specification for ESD suppression devices IEEE 100BASE-T1 EMC Test Specification for ESD suppression devices Version 1.0 Author & Company Dr. Bernd Körber, FTZ Zwickau Title 100BASE-T1 EMC Test Specification for ESD suppression devices Version

More information

TNC VNA Calibration Kits

TNC VNA Calibration Kits TNC VNA Calibration Kits DATA SHEET / 2Z-062 Models: 8650CK10 Fixed Load Kit 8650CK11 Fixed Load Kit Plus Adapters 8650CK20 Sliding Load Kit 8650CK21 Sliding Load Kit Plus Adapters // JANUARY 2018 TNC

More information

WFM6120, WFM7020, and WFM7120 Waveform Monitors Declassification and Security Instructions

WFM6120, WFM7020, and WFM7120 Waveform Monitors Declassification and Security Instructions WFM6120, WFM7020, and WFM7120 Waveform Monitors Declassification and Security Instructions www.tektronix.com 071-2295-00 Copyright Tektronix. All rights reserved. Licensed software products are owned by

More information

Instructions. RM2000B Rackmount Kit

Instructions. RM2000B Rackmount Kit Instructions RM2000B Rackmount Kit 071-1984-00 Warning These instructions are for qualified personnel only. To avoid personal injury, do not assemble or install this rack mount unless you are qualified

More information

TG8000 Multiformat Test Signal Generator Declassification and Security Instructions

TG8000 Multiformat Test Signal Generator Declassification and Security Instructions xx ZZZ TG8000 Multiformat Test Signal Generator Declassification and Security Instructions www.tek.com *P077068801* 077-0688-01 Copyright Tektronix. All rights reserved. Licensed software products are

More information

Digital Pre-emphasis Processor BERTScope DPP Series Datasheet

Digital Pre-emphasis Processor BERTScope DPP Series Datasheet Digital Pre-emphasis Processor BERTScope DPP Series Datasheet New microcontroller to provide more processing power RS-232 interface enhancement to speed up PCIe receiver equalization link training BERTScope

More information

LeCroy SPARQ S-Parameter Measurement Methodology

LeCroy SPARQ S-Parameter Measurement Methodology LeCroy SPARQ S-Parameter Measurement Methodology Dr. Alan Blankman, Product Manager Summary The SPARQ Signal Integrity Network Analyzer uses TDR and TDT to characterize a network s electrical behavior.

More information

Calibration kit coefficients -the saga continues

Calibration kit coefficients -the saga continues Calibration kit coefficients -the saga continues 22nd ANAMET Meeting 9 th September 2004 Jürg Rüfenacht Swiss Federal Office of Metrology and Accreditation metas Lindenweg 50, CH-3003 Bern-Wabern, Switzerland

More information

Model GHz 50 Ohm RF Module

Model GHz 50 Ohm RF Module Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 tek.com/keithley Model 7711 2 GHz 50 Ohm RF Module Instructions for use with DAQ6510 Introduction The 7711 plug-in module provides

More information

TG700 TV Signal Generator Platform Module Installation

TG700 TV Signal Generator Platform Module Installation Instructions TG700 TV Signal Generator Platform Module Installation TG700 TV 075-0986-01 www.tektronix.com 075-0986-01 Copyright Tektronix. All rights reserved. Licensed software products are owned by

More information

User s Guide. RP7000S Series Single-Ended Active Probe. Nov RIGOL Technologies, Inc.

User s Guide. RP7000S Series Single-Ended Active Probe. Nov RIGOL Technologies, Inc. User s Guide RP7000S Series Single-Ended Active Probe Nov. 2013 RIGOL Technologies, Inc. Guaranty and Declaration Copyright 2013 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL

More information

Digital Pre-emphasis Processor BERTScope DPP Series Datasheet

Digital Pre-emphasis Processor BERTScope DPP Series Datasheet Digital Pre-emphasis Processor BERTScope DPP Series Datasheet Overview The DPP125C is a nonlinear signal conditioner capable of adding controllable amounts of pre-emphasis to a signal. It takes in single-ended

More information

TLA7XX/TLA7XXX Series Tektronix Logic Analyzer Declassification and Security Instructions

TLA7XX/TLA7XXX Series Tektronix Logic Analyzer Declassification and Security Instructions TLA7XX/TLA7XXX Series Tektronix Logic Analyzer Declassification and Security Instructions www.tektronix.com 077-0037-00 Copyright Tektronix. All rights reserved. Licensed software products are owned by

More information

WFM8200 and WFM8300 Waveform Monitors Release Notes

WFM8200 and WFM8300 Waveform Monitors Release Notes xx ZZZ WFM8200 and WFM8300 Waveform Monitors www.tektronix.com *P077026405* 077-0264-05 Copyright Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or

More information

ENA Multiport Solution Setup Procedure of MWA Software for SP9T (10-port) Switch

ENA Multiport Solution Setup Procedure of MWA Software for SP9T (10-port) Switch ENA Multiport Solution Setup Procedure of MWA Software for SP9T (10-port) Switch Step 1. Define a test plan Step 2. Create a setup file Step 3. Perform measurement Rev. 01.00 Step 1. Define a Test Plan

More information

MAX2009/MAX2010 Evaluation Kits

MAX2009/MAX2010 Evaluation Kits 19-2972; Rev 0; 9/03 MAX2009/MAX2010 Evaluation Kits General Description The MAX2009/MAX2010 evaluation kits (EV kits) simplify the evaluation of the MAX2009 and MAX2010. These kits are fully assembled

More information

S3601 Series Vector Network Analyzer User Manual

S3601 Series Vector Network Analyzer User Manual S3601 Series Vector Network Analyzer User Manual Saluki Technology Inc. The manual applies to the vector network analyzers of the following models: S3601A vector network analyzer (100kHz-3GHz). S3601B

More information

Keysight Technologies On-Wafer Testing of Opto-Electronic Components Using the Lightwave Component Analyzers. Application Note

Keysight Technologies On-Wafer Testing of Opto-Electronic Components Using the Lightwave Component Analyzers. Application Note Keysight Technologies On-Wafer Testing of Opto-Electronic Components Using the Lightwave Component Analyzers Application Note Introduction When measurements of optoelectronic components are performed on

More information

Practical hints: splitter characterization. J. Ruefenacht, M. Wollensack, J. Hoffmann, M. Zeier

Practical hints: splitter characterization. J. Ruefenacht, M. Wollensack, J. Hoffmann, M. Zeier Practical hints: splitter characterization J. Ruefenacht, M. Wollensack, J. Hoffmann, M. Zeier Motivation once upon a time at METAS Juerg Ruefenacht HF-Circuits, European ANAMET Meeting, METAS 03.06.2015

More information

8000 Series Instruments Windows 2000 Operating System Restore

8000 Series Instruments Windows 2000 Operating System Restore 8000 Series Instruments Windows 2000 Operating System Restore 071-1381-01 CD ROM 8000 Series Instruments Windows 2000 Operating System Restore 071-1381-01 www.tektronix.com Copyright Tektronix, Inc. All

More information

P5200A Series High Voltage Differential Probes. Installation and Safety Instructions & Product Documentation CD

P5200A Series High Voltage Differential Probes. Installation and Safety Instructions & Product Documentation CD xx P5200A Series High Voltage Differential Probes ZZZ Installation and Safety Instructions & Product Documentation CD *P071288902* 071-2889-02 Copyright Tektronix. All rights reserved. Licensed software

More information

PQA600A Picture Quality Analyzer Declassification and Security Instructions

PQA600A Picture Quality Analyzer Declassification and Security Instructions xx ZZZ PQA600A Picture Quality Analyzer Declassification and Security Instructions *P077078800* 077-0788-00 xx ZZZ PQA600A Picture Quality Analyzer Declassification and Security Instructions www.tektronix.com

More information

PicoVNA. USB vector network analyzer. Quick Start Guide

PicoVNA. USB vector network analyzer. Quick Start Guide PicoVNA USB vector network analyzer Quick Start Guide CONTENTS 1. Introduction... 1 2. Safety information... 1 2.1 Symbols... 2 2.2 Maximum input and output ranges... 2 2.3 Grounding... 3 2.4 External

More information

TG8000 Multiformat Test Signal Generator Declassification and Security Instructions

TG8000 Multiformat Test Signal Generator Declassification and Security Instructions xx ZZZ TG8000 Multiformat Test Signal Generator Declassification and Security Instructions www.tektronix.com *P077068800* 077-0688-00 Copyright Tektronix. All rights reserved. Licensed software products

More information

Decompensated Operational Amplifiers

Decompensated Operational Amplifiers Decompensated Operational Amplifiers Abstract This paper discusses the what, why, and where of decompensated op amps in section one. The second section of the paper describes external compensation techniques,

More information