Lab Introduction & Network Analyser Basics

Size: px
Start display at page:

Download "Lab Introduction & Network Analyser Basics"

Transcription

1 Lab Introduction & Network Analyser Basics Martin J. Siebers Lab Manager

2 Agenda Lab Rules and Admin Connectors and Cables Network Analyser Basics VNA Demonstration

3 Lab Rules Who may use the Labs - Permission Responsibility Available Equipment Booking Removal of Equipment

4 Lab Rules (cont ) Before Leaving the Laboratories Anechoic Chamber Work-Benches, Drawers and Cabinets No food and drinks in the Lab

5 Lab Rules (cont ) Safety, Accidents and Damage Who to talk to? Visitors If you want access to the RF and Antenna Lab you must complete a Lab Registration form and know the rules!

6 Connectors and Cables Martin J. Siebers Lab Manager

7 Connectors

8 SMA connectors SMA Male SMA Female The SMA connector is the workhorse of the RF and microwave industries. The basic design uses a 4.2 millimeter diameter outer coax, filled with Teflon (PTFE) dielectric. Their upper frequency limit is anywhere from 18 to 26 GHz, depending on the tolerances held during manufacturing. SMAs, like many other coax connector families, are sized to fit a 5/16 inch (8mm) wrench.

9 3.5mm and 2.92 connectors Both use air dielectric, and mate with each other as well as the cheaper SMA styles (see next slide) The 3.5 mm connector is the next upgrade from SMA, it performs well up to 26 GHz The 2.92 mm connector (often called "2.9 millimeter") works all the way up to 40 GHz. The K-connector is Anritsu's version of the 2.92 mm connector 3.5 mm connector (male) 2.92 mm connector (male)

10

11 Mating precision and SMA connectors SMA connectors must not be used with the 3.5mm and 2.92mm precision connectors although they physically mate. E&E allows that female SMAs may be used with male precision connectors. Male SMAs must never be used with female precision connectors.

12 7 mm connectors The lowest VSWR of any connector offered Work well up to 18 GHz Unique: sexless You will find the 7 mm connector at the ports of the 8753A and at the LC-meter but you are not supposed to do anything with them! Usually hand tighten is fine. Use a wrench if you can t open one. Fortunately, most 7 mm connectors have hex-nuts that fit a 3/4 inch wrench. E&E labs stocks a torque wrench to provides repeatable and reliable connections.

13 Type N connectors These are rugged and cheaper than most of the connectors mentioned. Therefore you will find them all over the laboratory. There is a big variety of Type N adapters available at the E&E labs Type N to SMA (male and female) Type N to BNC (male and female)

14 A male-to-male or barrel adapter Adapters A female-to-female or bullet adapter A male-to-female or connector saver SMA connector saver, barrel and bullet adapter

15 Connector torque The correct mating torque for all 5/16 hexagonal-sleeved connectors is 8 inch-pounds. Use a torque wrench for precision measurements. A torque wrench is a calibrated device and people rely on its performance. Use it with care and don t play with it. Make sure it does not drop on the floor because it may loose it s calibration. Support the not turning part of the connection with a spanner Use it only to torque connectors in the way being demonstrated

16 Connector care Connectors and adapters cost a lot of money. Show respect and follow the basic guidelines: Learn how to clean connectors. Cleaning the threads is good practice, but stay away from cleaning the center conductor of an air dielectric connectors such as 3.5mm, 2.92mm and 2.4mm Don't use pliers on a connector. There are wrenches for every size connector and adapter, even for SMA bullets. Learn how to gauge connectors to determine if they are out of spec. One bad connector can damage many.

17 Connector care (cont ) Never use any part of a calibration kit other than for calibration. If you need a special adapter or a load buy it or borrow it but don t take it from the cal kit Use a torque wrench. It's OK to use less torque, but not more. When you are tightening or loosening a connector, try not to spin the mating surfaces against each other. You should only be turning the threaded sleeve. Remember, righty-tighty, lefty-loosely! The total damage done by people turning stuff in the wrong direction is enormous. The hardware you hold in your hands could very well be worth more than your automobile. So be gentle with it.

18 How to deal with stuck connectors The female to female SMA adapter is the cause of much pain in the microwave laboratory. Why? Because it tends to get stuck. Therefore we provide special wrenches to hold the bullet. So the nuts can be opened and the connectors released Never use common slipjoint pliers to remove it. A damaged thread of a bullet will damage the threads of all connectors that it mates with

19 Storing connectors and adaptors You've just bought a nice new set of coax adapters for that new lab project. They come with some plastic caps on each end... so you take off the caps and throw them out, right? WRONG! Those plastic things are called dust caps for a reason, they keep your connectors clean.

20 Connector care (cont ) Adapters and connectors are precision items, that's why they cost a fortune. So why does someone treat them like common hardware? Check out the picture, whoever "owns" a bench that stores connectors like this should never be allowed in the lab! Those two-cents worth of stainless-steel screws are full of contaminates like oil, skin, and the worst culprit, tiny pieces of metal. Store them separately, and keep the dust caps on the adapters!

21 Connector care (cont ) Check out the photo of the 7 mm connector What is wrong? The way to protect the 7mm connector is to always spin the sleeve clockwise to extend the threads, then push on a dust cap.

22 Cleaning connectors Cleaning connectors regularly will not only make them last longer, but will also provide better results for your measurements. Convention: At E&E we clean the connector before we use them Clean connectors Cables Cal kit Adapters But NOT the ones at the test instruments Never apply the cleaning solution directly onto a connector.

23 Cleaning connectors in the E&E-Laboratories Clean the SMA connector: use a toothpick, apply the cleaning solution and carefully loosen all dirt in the thread If there is dirt on the Teflon face loosen it with a wet toothpick. Be careful on the Teflon face Don t rake it up! Repeat this procedure until the toothpick doesn t show dirt. Use a cotton swab, apply little cleaning fluid and wipe the female SMA connector. Cotton swabs are too big for the male SMA so it cant be wiped. Finally use air to dry the connector. Never shake the air can! Press the knob gently to ensure economical use. Start blowing air aside of the connector moving the flow onto it. Don t blow the thread of the female connector.

24 Cleaning connectors in the E&E-Laboratories Clean the 3.5 mm precision connector: use a toothpick, apply the cleaning solution and carefully loosen all dirt in the thread of the male and the female connector. Repeat this procedure until the toothpick doesn t show dirt. There are special foam swabs for the 3.5 connectors. They may be used to very carefully wipe the male and female connector under the supervision of the lab manager. Never wipe the inside of a precision connector on your own! Finally use air to dry the connector. Never shake the air can! Press the knob gently to ensure economical use. Start blowing air aside of the connector moving the flow onto it. Don t blow the thread of the female connector.

25 Cleaning connectors (cont ) The inside of a female SMA should be wiped with a cotton swab. Don t wipe the inside of 3.5mm PC connectors Outside threads of female connectors may be cleaned using a swab but the toothpick or the paper method are more efficient.

26 Cleaning connectors (cont ) Swabs may be reused but Throw out the swab when it looks like the swab on the left in the photo!

27 Connector Handling Never clean the precision connectors fitted at the instruments. Lab maintenance is responsible for that! When connector threads are cleaned, you will be able to hand-tighten all connections to within 1/2 turn of the proper torque If you are not able to hand tighten a connector, something is wrong. Usually it is an aligning problem, or one of the connectors in the pair is already damaged. Connector damage is often the result of bad cleaning or a mechanical force being applied, causing the connector to become slightly "d-shaped" instead of perfectly round.

28 Transmission lines: Coaxial Cables The dimensions d, D, and the dielectric determine: Z0 (characteristic impedance) fc (the cutoff frequency) capacitance and inductance per meter Temperature Bending radius Connector transition

29 Network Analyser Basics Martin J. Siebers Lab Manager

30 Overview Network analysers fall into two categories The Vector Network Analyser (VNA) is capable of measuring complex reflection and transmission The scalar analyser can only measure magnitude A VNA can measure S-parameter and VSWR, loss, gain, isolation, and group delay of any two (or more) ports of a given network Key manufacturers of VNAs are Agilent (former HP), Anritsu (Wiltron), and Rohde & Schwarz. VNAs also called "ANA automated network analyser.

31 History The original network analyser (HP 8409) No Error Correction build in Error correction was done by hand Then the first automated network analysers were introduced. A minicomputer processed the vector data delivered error corrected and accurate magnitude and phase of the four S-parameters The next step was to build the error correction into the TE and display the error-corrected measurements in nearly real time (the original HP 8510, circa 1982).

32 The 8510 VNA The "classic" vector network analyser is the Agilent (HP) Depending on how much one spend, this analyser can make measurements from 45 MHz to 110 GHz. E&E has two 8510C 45MHz to 20GHz 45MHz to 50GHz

33 Simplified VNA Block Diagram Incident DUT Transmitted SOURCE Reflected SIGNAL SEPARATION INCIDENT (R) REFLECTED (A) TRANSMITTED (B) RECEIVER / DETECTOR PROCESSOR / DISPLAY

34 Characterizing Devices Incident R Reflected A REFLECTION Transmitted B TRANSMISSION Reflected Incident = A R Transmitted Incident = B R SWR S-Parameters S11,S22 Reflection Coefficient Γ, ρ Return Loss Impedance, Admittance R+jX, G+jB Gain / Loss S-Parameters S21,S12 Transmission Coefficient Τ,τ Insertion Phase Group Delay

35 S-Parameters Relatively easy to obtain at high frequencies Measure voltage traveling waves with a vector network analyzer Don't need shorts/opens which can cause active devices to oscillate or self-destruct Relate to familiar measurements (gain, loss, reflection coefficient...) Can cascade S-parameters of multiple devices to predict system performance Can easily import and use S-parameter files in electronic simulation tools Incident a1 S11 Reflected b 1 Transmitted S 21 DUT Port 1 Port 2 S12 b1 = S11 a 1 + S 12 a 2 b 2 = S21 a1 + S22 a 2 Transmitted b 2 S22 Reflected a 2 Incident

36 S-Parameters Relate to Common Measurement Terms S11 = forward reflection coefficient (input match) S22 = reverse reflection coefficient (output match) S21 = forward transmission coefficient (gain or loss) S12 = reverse transmission coefficient (isolation) Remember, S-parameters are inherently linear quantities -- however, we often express them in a log-magnitude format

37 Errors Systematic errors due to imperfections in the analyzer and test setup are assumed to be time invariant (predictable) can be characterized (during calibration process) and mathematically removed during measurements Random errors vary with time in random fashion (unpredictable) cannot be removed by calibration main contributors: instrument noise (source phase noise, IF noise floor, etc.) switch repeatability connector repeatability Drift errors are due to instrument or test-system Measured Data performance changing after a calibration has been done are primarily caused by temperature variation can be removed by further calibration(s) Errors: SYSTEMATIC RANDOM DRIFT Unknown Device

38 Systematic Measurement Errors R Directivity A Crosstalk B DUT Frequency response reflection tracking (A/R) transmission tracking (B/R) Source Mismatch Load Mismatch Six forward and six reverse error terms yields 12 error terms for two-port devices

39 What is Calibration? Measure known device Note measured value Generate error table based on the difference between known and measured values Subtract error from subsequent measurements

40 Standards Definitions Definitions are the parameters of an equivalent circuit network that represents the physical network A standard may have a combination of a delay, a loss, a capacitance, an inductance, a frequency range and a characteristic impedance Standards definitions must be known to the VNA Enter manually Load from disc A Cal Kits come with the definitions of their standards.

41 Calibration Kits Coaxial calibration kits come in type N, 7 mm, 3.5 mm, 2.92 mm, 2.4 mm, 1.0 mm and skripsie. (Bold available at E&E) Be sure not to exceed the frequency capability of the cables, adapters and calibration kit (remember session on connectors). There are waveguide calibration kits for every waveguide band. E&E has one for S-Band and one for X-Band. Remember that cal kits are expensive, and pieces of the cal kit should never be used as adapters or terminations. Always put the little plastic covers onto the cal kit pieces.

42 Verification - Validation Verification means to measure known standards which haven t been used for the actual calibration. Therefore a verification kit is used. To validate the calibration and the general health of the test system, look at a few things after calibration Check the residual error in a "through" connection. Ideally the magnitude should be 0 db and the phase should be 0.0 degrees. The reflection at both ports should be better than -40dB. Usually it is at about -60dB or lower. Depending on the cable quality and the frequency range a variation of the transmission and reflection parameters will be noticed upon the movement of the cables. If there is an issue with the calibration, figure out the source of the problem before performing another calibration. Otherwise you will be wasting your time!

43 General Considerations Before performing a calibration, answer the following questions: Frequency range? Do the cal kit, the cables and any adapters operate at the desired band? Are the cables fit for this application? Notice the effect of gently bending cables. Notice the effect of moving cables. Will the cables reach the DUT without stressing them? If possible, perform the calibration with cables in the position needed to be connected to DUT

44 General Considerations Averaging Averaging will improve the accuracy of your data, as long as you do it during the calibration as well as the actual measurement. It will slow down the measurement process Smoothing Smoothing is cheating! Smoothing reduces the "bumpiness" of a frequency response by averaging data across a couple of frequency points and using the result at one frequency.

45 General Considerations Electrostatic Discharge Always discharge yourself! Use the Static Control Wrist Straps provided. Be extremely careful when executing antenna measurements (centre conductor is often exposed) Torque The torque of the connections will have an influence on the quality and repeatability of the calibration and the actual measurements. Always apply the right torque using a torque wrench. Using less torque is your choice. Never over-torque!

46 Any Questions? Thanks for your attention! See you upstairs R551 for VNA Demonstrations!

spinner Measurement & Calibration equipment for network analyzers

spinner Measurement & Calibration equipment for network analyzers spinner Measurement & Calibration equipment for network analyzers Edition B 2011 High Frequency Performance Worldwide www.spinner-group.com SPINNER test & calibration equipment for best measurement results

More information

11525A 11524A

11525A 11524A 8 Typical Configuration 11900A 11901A 11904A 8059A 1250-1159 1250-1748 85058-60007 11900C 11901C 11901D 11904C 11904D 8059C 1250-1462 85058-60009 1190A 1250-166 1250-174 11525A 11524A 11852B 11852B Option

More information

01-161, -162, -163, AND -164 GAUGING SETS USER S GUIDE

01-161, -162, -163, AND -164 GAUGING SETS USER S GUIDE 01-161, -162, -163, AND -164 GAUGING SETS USER S GUIDE 1. INTRODUCTION This user s guide provides a description and instructions on how to use the 01-161, -162, -163, and -164 Gauging Sets to measure pin

More information

RF Connector Handling Guidelines

RF Connector Handling Guidelines RF Connector Handling Guidelines A compilation of information found on the internet Author: Jeffrey Landry Electrical and Computer Engineering, Concordia University, Sept 2012 Connector handling (SMA,

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) OPERATION AND MAINTENANCE MANUAL FOR SWR AUTOTESTERS AND BRIDGES 1. INTRODUCTION This manual provides product descriptions

More information

TNC VNA Calibration Kits

TNC VNA Calibration Kits TNC VNA Calibration Kits DATA SHEET / 2Z-062 Models: 8650CK10 Fixed Load Kit 8650CK11 Fixed Load Kit Plus Adapters 8650CK20 Sliding Load Kit 8650CK21 Sliding Load Kit Plus Adapters // JANUARY 2018 TNC

More information

Keysight M8000 Series BER Test Solutions

Keysight M8000 Series BER Test Solutions Keysight M8000 Series BER Test Solutions J-BERT M8020A High-Performance BERT M8030A Multi-Channel BERT M8040A High-Performance BERT M8041A, M8051A, M8061A, M8062A, M8045A, M8046A & M8057A Tips for Preventing

More information

365XA, 365XA-1, 365XD, and 365XD-1 Mechanical Calibration Kits

365XA, 365XA-1, 365XD, and 365XD-1 Mechanical Calibration Kits Reference Manual 365XA, 365XA-1, 365XD, and 365XD-1 Mechanical Calibration Kits 3650A Calibration Kit, SMA/3.5 mm Connectors 3650A-1 Calibration Kit, SMA/3.5 mm Connectors with Sliding Loads 3651A Calibration

More information

HP 8714B Calibration

HP 8714B Calibration Wireless Circuits and Systems Laboratory Procedure #7 Calibrating the HP 8714 Network Analyzer This procedure describes the different types of calibrations that are available with the HP 8714 vector network

More information

What We Do. Fixturing and Fixture Removal for Multiport Devices with Non-Standard RF Interfaces. Fully automate for: Speed, Accuracy, Ease of Use

What We Do. Fixturing and Fixture Removal for Multiport Devices with Non-Standard RF Interfaces. Fully automate for: Speed, Accuracy, Ease of Use Fixturing and Fixture Removal for Multiport Devices with Non-Standard RF Interfaces What We Do Fully automate for: Speed, Accuracy, Ease of Use Calibration Validation To give confidence in test results

More information

3.5mm VNA Calibration Kits

3.5mm VNA Calibration Kits 3.5mm VNA Calibration Kits DATA SHEET / 2Z-059 Models: 8050CK10 Fixed Load Kit 8050CK11 Fixed Load Kit Plus Adapters 8050CK20 Sliding Load Kit - Basic 8050CK21 Sliding Load Kit Plus Adapters 8050CK30 TRL

More information

2.92mm VNA Calibration Kits

2.92mm VNA Calibration Kits 2.92mm VNA Calibration Kits DATA SHEET / 2Z-058B Models: 8770CK10 Fixed Load Kit 8770CK11 Fixed Load Kit Plus Adapters 8770CK20 Sliding Load Kit 8770CK21 Sliding Load Kit Plus Adapters 8770CK30 TRL Kit

More information

Precision N Coaxial Calibration Kit

Precision N Coaxial Calibration Kit User Guide Precision N Coaxial Calibration Kit DC to 18 GHz Models: 8850CK10/11 8850CK20/21 8850-511 (A) 3/15 User Guide Precision N Coaxial Calibration Kit DC to 18 GHz Models: 8850CK10/11 8850CK20/21

More information

Adapter characterization up to 110 GHz

Adapter characterization up to 110 GHz Adapter characterization up to 110 GHz EMRP HF Circuits project: 6 th European ANAMET meeting 28.06.2016 Adapter characterization: Motivation The challenges of an adapter characterization: Sheet Series

More information

FMCK1013 DATA SHEET. Features: Applications: Configuration. Mechanical Specifications

FMCK1013 DATA SHEET. Features: Applications: Configuration. Mechanical Specifications Short Open Load Thru (SOLT) Analyzer Calibration Kit with 3.5mm Male Connections in a portable 4-in-1 Design and Operating to 26.5 GHz Short Open Load Thru (SOLT) Analyzer Calibration Kit, part number

More information

SERIES MS462XX VECTOR NETWORK MEASUREMENT SYSTEM VERIFICATION KITS OPERATION AND MAINTENANCE MANUAL

SERIES MS462XX VECTOR NETWORK MEASUREMENT SYSTEM VERIFICATION KITS OPERATION AND MAINTENANCE MANUAL SERIES MS462XX VECTOR NETWORK MEASUREMENT SYSTEM VERIFICATION KITS OPERATION AND MAINTENANCE MANUAL 490 JARVIS DRIVE l MORGAN HILL, CA 95037-2809 P/N: 10100-00036 REVISION: A PRINTED: MAY 1999 COPYRIGHT

More information

UNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY

UNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY UNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY HIGH TEMPERATURE & HIGH FREQUENCY TESTING SYSTEM LABORATORY USER MANUAL 4/29/2011 Overview

More information

Agilent 85130C NMD 3.5 mm to Type-N Adapter Kit

Agilent 85130C NMD 3.5 mm to Type-N Adapter Kit Agilent 85130C NMD 3.5 mm to Type-N Adapter Kit Operating and Service Manual Agilent Technologies Notices Agilent Technologies, Inc. 1987, 2004, 2008 No part of this manual may be reproduced in any form

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Operating and Service Manual Agilent Technologies 85131E/F/H NMD-3.5 mm -f- to 3.5 mm Flexible Test Port Return Cables

More information

TN Vector Network Analyzer Calibrations and Their Effect on Measurement Results. Bernie McDermott. May 2010

TN Vector Network Analyzer Calibrations and Their Effect on Measurement Results. Bernie McDermott. May 2010 TN 201005 Vector Network Analyzer Calibrations and Their Effect on Measurement Results May 2010 Bernie McDermott Lead Metrologist Calibration Products and Services Measurement Examples and Technical Notes

More information

ASSEMBLY, INSTALLATION, AND REMOVAL OF CONTACTS AND MODULES

ASSEMBLY, INSTALLATION, AND REMOVAL OF CONTACTS AND MODULES ASSEMBLY, INSTALLATION, AND REMOVAL OF S AND MODULES FOR 40 GHz COAXIAL S AND MODULES Table of Contents SECTION 1 INSTALLATION AND REMOVAL SECTION 2 MODULE INSTALLATION AND REMOVAL SECTION 3 CROSS REFERENCE

More information

Keysight 11636B APC-3.5 mm Power Divider DC to 26.5 GHz

Keysight 11636B APC-3.5 mm Power Divider DC to 26.5 GHz Keysight 11636B APC-3.5 mm Power Divider DC to 26.5 GHz Operating and Service Manual Notices Keysight Technologies 1985, 2013, 2014 No part of this manual may be reproduced in any form or by any means

More information

User s and Service Guide

User s and Service Guide User s and Service Guide Agilent Technologies 11644A X, P, and K Waveguide Calibration Kits This manual applies to 11644A series calibration kits with serial number prefix 3032A. Agilent Part Number: 11644-90371

More information

Keysight 85130D NMD 3.5 mm to 3.5 mm Adapter Kit

Keysight 85130D NMD 3.5 mm to 3.5 mm Adapter Kit Keysight 85130D NMD 3.5 mm to 3.5 mm Adapter Kit Operating and Service Manual Notices Keysight Technologies 1987-2015 No part of this manual may be reproduced in any form or by any means (including electronic

More information

User s and Service Guide

User s and Service Guide User s and Service Guide Agilent Technologies 85053B 3.5 mm Verification Kit Agilent Part Number: 85053-90028 Printed in USA Print Date: January 2002 Supersedes: October 2000 Copyright 1995, 2000, 2002

More information

laboratories and production environments.

laboratories and production environments. Improved, user-friendly RF&MW calibration techniques for calibration laboratories and production environments. Reducing calibration time & uncertainties by using known, characterized standards. Harald

More information

Keysight 11878A 50 Ohm 3.5 mm Adapter Kit

Keysight 11878A 50 Ohm 3.5 mm Adapter Kit Keysight 11878A 50 Ohm 3.5 mm Adapter Kit Operating and Service Manual Notices Keysight Technologies 1989, 2004, 2013, 2014 No part of this manual may be reproduced in any form or by any means (including

More information

Connector Repeatability of Type-N Coaxial Adapters

Connector Repeatability of Type-N Coaxial Adapters Connector Repeatability of Type-N Coaxial Adapters Nowadays we have in Finnish Defence Forces a lot of different automatic testing systems. Many of these systems mainly use Type-N coaxial connectors Because

More information

Lecture 10. Vector Network Analyzers and Signal Flow Graphs

Lecture 10. Vector Network Analyzers and Signal Flow Graphs HP8510 Lecture 10 Vector Network Analyzers and Signal Flow Graphs Sections: 6.7 and 6.11 Homework: From Section 6.13 Exercises: 4, 5, 6, 7, 9, 10, 22 Acknowledgement: Some diagrams and photos are from

More information

Keysight Technologies 85050B 7 mm Calibration Kit

Keysight Technologies 85050B 7 mm Calibration Kit Keysight Technologies 85050B 7 mm Calibration Kit User s and Service Guide 85050B i Documentation Warranty THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED,

More information

Operations and Installation Manual. Keysight 85132E/F 3.5 mm to 7 mm Test Port Return Cables

Operations and Installation Manual. Keysight 85132E/F 3.5 mm to 7 mm Test Port Return Cables Operations and Installation Manual Keysight 85132E/F 3.5 mm to 7 mm Test Port Return Cables Notices Keysight Technologies, Inc. 1987-2016 No part of this manual may be reproduced in any form or by any

More information

Applications & Features, Type SSMP When using blind mate connectors in an application, careful consideration has to be given to choosing the right con

Applications & Features, Type SSMP When using blind mate connectors in an application, careful consideration has to be given to choosing the right con SSMP Push-On Connectors & Adapters THE SSMP (Sub-Sub-Miniature-Push-On) is a 1.7 mm connector, using solid dielectric interface. The connector is 70% of the size of the SMP and therefore allows even higher

More information

Practical hints: splitter characterization. J. Ruefenacht, M. Wollensack, J. Hoffmann, M. Zeier

Practical hints: splitter characterization. J. Ruefenacht, M. Wollensack, J. Hoffmann, M. Zeier Practical hints: splitter characterization J. Ruefenacht, M. Wollensack, J. Hoffmann, M. Zeier Motivation once upon a time at METAS Juerg Ruefenacht HF-Circuits, European ANAMET Meeting, METAS 03.06.2015

More information

Keysight Technologies 85131C/D/G NMD-3.5 mm -fto 3.5 mm Semi-Rigid Test Port Return Cables

Keysight Technologies 85131C/D/G NMD-3.5 mm -fto 3.5 mm Semi-Rigid Test Port Return Cables Keysight Technologies 85131C/D/G NMD-3.5 mm -fto 3.5 mm Semi-Rigid Test Port Return Cables Operating and Service Manual 85131C/D/G i Documentation Warranty THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED

More information

Keysight 85032F Type-N 50 Calibration Kit

Keysight 85032F Type-N 50 Calibration Kit Keysight 85032F Type-N 50 Calibration Kit User s and Service Guide User s and Service Guide Keysight Technologies 85032F Type-N 50 Calibration Kit Keysight Part Number: 85032-90021 Printed in USA Print

More information

VNA0440 VNA0440e VNA0460 VNA0460e User Manual Version 3.0

VNA0440 VNA0440e VNA0460 VNA0460e User Manual Version 3.0 User Manual Version 3.0 - 2 - www.megiq.com How to use the MegiQ VNA0440 (e) VNA0460 (e) Vector Network Analyzers o General information o User software o Using the Instrument MegiQ BV Weegschaalstraat

More information

TTR500 Series Vector Network Analyzers Demonstration Guide

TTR500 Series Vector Network Analyzers Demonstration Guide xx ZZZ TTR500 Series Vector Network Analyzers Demonstration Guide *P071349301* 071-3493-01 xx ZZZ TTR500 Series Vector Network Analyzers Demonstration Guide Register now! Click the following link to protect

More information

ATEC (2832) Models 8757A, 85027A/B/C, 85020A/B

ATEC (2832) Models 8757A, 85027A/B/C, 85020A/B Established 1981 Advanced Test NETWORK Equipment ANALYZERS Rentals 8757/8756 System Accessories wwwatecorpcom 800-404-ATEC (2832) Models 8757A, 85027A/B/C, 85020A/B UP]] HP 8757A Specifications Amplitude

More information

User s and Service Guide. Agilent Technologies 85052C 3.5 mm Precision Calibration Kit

User s and Service Guide. Agilent Technologies 85052C 3.5 mm Precision Calibration Kit User s and Service Guide Agilent Technologies 85052C 3.5 mm Precision Calibration Kit Agilent Part Number: 85052-90078 Printed in USA Print Date: March 2005 Supersedes: January 2002 Copyright Agilent Technologies,

More information

Keysight Technologies 85052B 3.5 mm Calibration Kit

Keysight Technologies 85052B 3.5 mm Calibration Kit Keysight Technologies 85052B 3.5 mm Calibration Kit User s and Service Guide 85052B WARRANTY STATEMENT THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED AS IS, AND IS SUBJECT TO BEING CHANGED, WITHOUT

More information

Keysight Technologies 85031B 7 mm Calibration Kit

Keysight Technologies 85031B 7 mm Calibration Kit Keysight Technologies 85031B 7 mm Calibration Kit User s and Service Guide 85031B Documentation Warranty THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT

More information

User s and Service Guide

User s and Service Guide User s and Service Guide Agilent Technologies 85056A 2.4 mm Precision Calibration Kit Agilent Part Number: 85056-90020 Printed in USA Print Date: May 2012 Supersedes: August 2010 Copyright 1996, 2000,

More information

N5532B Sensor Module User s and Service Guide

N5532B Sensor Module User s and Service Guide N5532B Sensor Module User s and Service Guide for the Agilent N5531S Measuring Receiver Systems Manufacturing Part Number: N5532-90008 Printed in USA April 2010 Copyright 2010 Agilent Technologies Notice

More information

Keysight Second Source, Combiner, and Mechanical Switches Upgrade Kit

Keysight Second Source, Combiner, and Mechanical Switches Upgrade Kit Keysight Second Source, Combiner, and Mechanical Switches Upgrade Kit To Upgrade PNA-X N5241A, N5242A, or N5249A Option 419 to Option 423 Upgrade Kit Order Number: N5241AU- 927, N5242AU- 927, and N5249AU-

More information

Labs. Introduction. Application Notes

Labs. Introduction. Application Notes Introduction Application Notes Florida RF has provided an extensive collection of Application Notes in this section to help designers select and specify cable assemblies. These Application Notes cover

More information

User s and Service Guide. Agilent Technologies 85033E 3.5 mm Calibration Kit

User s and Service Guide. Agilent Technologies 85033E 3.5 mm Calibration Kit User s and Service Guide Agilent Technologies 3.5 mm Calibration Kit Copyright 2000 2002, 2004, 2008-2010 Agilent Part Number: 85033-90028 Printed in USA Print Date: August 2010 Supersedes: April 2010

More information

Agilent Upgrade Guide for the 8510 Vector Network Analyzer Product Note

Agilent Upgrade Guide for the 8510 Vector Network Analyzer Product Note Agilent Upgrade Guide for the 8510 Vector Network Analyzer Product Note Discontinued Product Information For Support Reference Only Information herein, may refer to products/services no longer supported.

More information

OPERATION MANUAL SERIES BORESCOPES INSTRUMENT TECHNOLOGY, INC. POB 381, Westfield, MA Airport Road, Westfield, MA 01085

OPERATION MANUAL SERIES BORESCOPES INSTRUMENT TECHNOLOGY, INC. POB 381, Westfield, MA Airport Road, Westfield, MA 01085 OPERATION MANUAL SERIES 123000 BORESCOPES INSTRUMENT TECHNOLOGY, INC. POB 381, Westfield, MA 01086 33 Airport Road, Westfield, MA 01085 Tel: (413) 562-3606 Fax: (413) 568-9809 Email: iti@scopes.com www.scopes.com

More information

User s and Service Guide

User s and Service Guide User s and Service Guide Agilent Technologies 85056K 2.4 mm/2.92 mm Calibration Kit Agilent Part Number: 85056-90019 Printed in USA Print Date: November 2005 Supersedes: August 2004 Copyright 1996, 2001,

More information

Operating and Service Manual

Operating and Service Manual Operating and Service Manual Agilent Technologies 85130F NMD 2.4 mm to 3.5 mm Adapter Kit Manufacturing Part Number: 85130-90030 Printed in USA Print Date: May 1988 Agilent Technologies, Inc. 1988 Hewlett-Packard

More information

VectorStar ME7838A4 Multiport Broadband/Banded Millimeter-Wave System

VectorStar ME7838A4 Multiport Broadband/Banded Millimeter-Wave System Quick Start Guide VectorStar ME7A4 Multiport Broadband/Banded Millimeter-Wave System High Performance Modular Broadband/Banded Millimeter-Wave Vector Network Analyzer (VNA) Multiport Measurement System

More information

This guide covers common precision and related connectors used at DC, Low Frequency, Radio Frequency, Microwave and Millimetre-wave Frequencies.

This guide covers common precision and related connectors used at DC, Low Frequency, Radio Frequency, Microwave and Millimetre-wave Frequencies. Guidance on using Precision Coaxial Connectors in Measurement Photograph Nick Ridler NPL Compiled by: Doug Skinner MBE 3rd Edition (August 2007) For this edition the sections on the various connectors

More information

Fd PACKARD. Application Note 326. Coaxial Systems. Principles of microwave connector care HEWLETT. ,/ (for higher reliability and better measurements)

Fd PACKARD. Application Note 326. Coaxial Systems. Principles of microwave connector care HEWLETT. ,/ (for higher reliability and better measurements) t ' -J Application Note 326 Coaxial Systems Principles of microwave connector care,/ (for higher reliability and better measurements) HEWLETT Fd PACKARD Recent advances in measurement capabilities have

More information

Installation Note. Configurable Test Set Upgrade Kit. For PNA Series RF Network Analyzers (E8801A, E8802A, and E8803A) Network Analyzer Model Number

Installation Note. Configurable Test Set Upgrade Kit. For PNA Series RF Network Analyzers (E8801A, E8802A, and E8803A) Network Analyzer Model Number Installation Note Configurable Test Set Upgrade Kit For PNA Series RF Network Analyzers (E8801A, E8802A, and E8803A) Network Analyzer Model Number Upgrade Kit Part Number E8801A, E8802A, E8803A E8801-60103

More information

OPERATION MANUAL SERIES MICRO BORESCOPES INSTRUMENT TECHNOLOGY, INC. POB 381, Westfield, MA Airport Road, Westfield, MA 01085

OPERATION MANUAL SERIES MICRO BORESCOPES INSTRUMENT TECHNOLOGY, INC. POB 381, Westfield, MA Airport Road, Westfield, MA 01085 OPERATION MANUAL SERIES 124000 MICRO BORESCOPES INSTRUMENT TECHNOLOGY, INC. POB 381, Westfield, MA 01086 33 Airport Road, Westfield, MA 01085 Tel: (413) 562-3606 Fax: (413) 568-9809 Email: iti@scopes.com

More information

Massachusetts Institute of Technology Department of Electrical Engineering and Computer Science Introductory Digital Systems Laboratory

Massachusetts Institute of Technology Department of Electrical Engineering and Computer Science Introductory Digital Systems Laboratory Massachusetts Institute of Technology Department of Electrical Engineering and Computer Science 6.111 -- Introductory Digital Systems Laboratory NUBUS LABORATORY KIT For your pleasure and convenience,

More information

RF Characterization Report

RF Characterization Report RF058 Series Cable Assemblies RF058-01BJ1-01BJ1-0150 RF058-01SB1-01SB1-0150 RF058-01SP1-01SP1-0150 Description: RF Cable Assembly, 50Ohm, RG058 Coaxial Cable Samtec Inc. WWW.SAMTEC.COM Phone: 812-944-6733

More information

Configuring a Precision System for On-Wafer Capacitance

Configuring a Precision System for On-Wafer Capacitance Application Note Innovating Test Technologies Configuring a Precision System for On-Wafer Capacitance Development of new aggressively scaled MOS processes and devices depends on high accuracy CV measurements.

More information

GMS Interconnect Series Product Information

GMS Interconnect Series Product Information GMS Interconnect Series Product Information Microwave Products Product Features: Frequency range: DC up to 23 GHz, Force to engage/disengage: 10 ounces min./2.5 pounds max., Male connectors have full self-centering

More information

Installation Note. Source Attenuators and Bias Tees Upgrade Kit. For E8362B/C PNA Series Microwave Network Analyzers. Network Analyzer Model Number

Installation Note. Source Attenuators and Bias Tees Upgrade Kit. For E8362B/C PNA Series Microwave Network Analyzers. Network Analyzer Model Number Installation Note Source Attenuators and Bias Tees Upgrade Kit For E8362B/C PNA Series Microwave Network Analyzers Network Analyzer Model Number Upgrade Kit Part Number E8362B/C E8362-60115 Agilent Part

More information

Installation Note. IF Access Upgrade Kit. For PNA Series Microwave Network Analyzers. Network Analyzer Model Number. Upgrade Kit Part Number

Installation Note. IF Access Upgrade Kit. For PNA Series Microwave Network Analyzers. Network Analyzer Model Number. Upgrade Kit Part Number Installation Note IF Access Upgrade Kit For PNA Series Microwave Network Analyzers Network Analyzer Model Number E8361A/C, E8362B/C, E8363B/C, E8364B/C Upgrade Kit Part Number E8362-60116 Agilent Part

More information

SMA Series. Specification. Description. Applications Civil & Military Telecommunication Instrumentation SMA SMA. Interface Mating Dimensions

SMA Series. Specification. Description. Applications Civil & Military Telecommunication Instrumentation SMA SMA. Interface Mating Dimensions Series Specification 50 ohm 0-18 GHZ connectors are adaptable to interconnection requirements of both systems and components. S-Conn offers a wide variety of cable connectors, receptacles, feed thrus,

More information

Keysight 85052D 3.5 mm Economy Calibration Kit

Keysight 85052D 3.5 mm Economy Calibration Kit Keysight 85052D 3.5 mm Economy Calibration Kit User s and Service Guide Notices Keysight Technologies 1994-2014 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Why Calibrate to Center of Thru

Why Calibrate to Center of Thru Why Calibrate to Center of Thru How to read this document : This document might be very hard to read pending your VNWA skill level. Are you pretty Noval, then read Appendix B first to see what it is all

More information

User s and Service Guide

User s and Service Guide User s and Service Guide Agilent Technologies 85032F Type-N 50 Ω Calibration Kit Agilent Part Number: 85032-90021 Printed in USA Print Date: August 2005 Supersedes: May 2005 Copyright 2000, 2002, 2004,

More information

User s and Service Guide

User s and Service Guide User s and Service Guide Agilent Technologies 85056K 2.4 mm/2.92 mm Calibration Kit Agilent Part Number: 85056-90019 Printed in USA Print Date: August 2010 Supersedes: November 2008 Copyright 1996, 2001,

More information

Labs. Introduction. Application Notes

Labs. Introduction. Application Notes Introduction Application Notes Florida RF has provided an extensive collection of Application Notes in this section to help designers select and specify cable assemblies. These Application Notes cover

More information

Agilent PNA Series Microwave Network Analyzers

Agilent PNA Series Microwave Network Analyzers Agilent PNA Series Microwave Network Analyzers Configuration Guide PNA-L N5230A PNA-L N5230A PNA E8362B PNA E8363B PNA E8364B PNA E8361A PNA N5250A 300 khz to 6, 13.5, or 20 GHz 10 MHz to 20, 40, or 50

More information

Dimensions (inch) 3.74 x 1.45 x.95. PMA1-03S 70 MHz - 3 GHz Inline Power Monitor with SMA Connectors 1

Dimensions (inch) 3.74 x 1.45 x.95. PMA1-03S 70 MHz - 3 GHz Inline Power Monitor with SMA Connectors 1 The is a thru-line microwave power monitor, designed to non-invasively measure and display the power passing through a coaxial line. Battery life is exceptional due to proprietary low power design and

More information

PicoVNA. USB vector network analyzer. Quick Start Guide

PicoVNA. USB vector network analyzer. Quick Start Guide PicoVNA USB vector network analyzer Quick Start Guide CONTENTS 1. Introduction... 1 2. Safety information... 1 2.1 Symbols... 2 2.2 Maximum input and output ranges... 2 2.3 Grounding... 3 2.4 External

More information

Keysight 909A/D Coaxial Terminations

Keysight 909A/D Coaxial Terminations Keysight 909A/D Coaxial Terminations Operating and Service Manual Notices Keysight Technologies 2000-2014 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Extended Frequency Range Upgrade Kit (50 GHz to 67 GHz)

Extended Frequency Range Upgrade Kit (50 GHz to 67 GHz) Installation Note Extended Frequency Range Upgrade Kit (50 GHz to 67 GHz) Upgrade Kit Number: E8364-60105 For E8364B/C Microwave Network Analyzers WITHOUT the Configurable Test Set Option (Option 014)

More information

HS-27-Mini. Headstage preamplifier providing 27 channels of unity gain amplification.

HS-27-Mini. Headstage preamplifier providing 27 channels of unity gain amplification. HS-27-Mini Headstage preamplifier providing 27 channels of unity gain amplification. 9/16/2013 Neuralynx, Inc. 105 Commercial Drive, Bozeman, MT 59715 Phone 406.585.4542 Fax 866.585.1743 www.neuralynx.com

More information

Keysight 85130F NMD 2.4 mm to 3.5 mm Adapter Kit

Keysight 85130F NMD 2.4 mm to 3.5 mm Adapter Kit Keysight 85130F NMD 2.4 mm to 3.5 mm Adapter Kit Operating and Service Manual Notices Keysight Technologies 1988-2015 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Models 1417 and 1437 User s Manual. High-Speed Photodetector Modules

Models 1417 and 1437 User s Manual. High-Speed Photodetector Modules Models 1417 and 1437 User s Manual High-Speed Photodetector Modules Handling Precautions The detector is sensitive to electrostatic discharges and could be permanently damaged if subjected even to small

More information

Mounting Instructions for MTP Modules

Mounting Instructions for MTP Modules VISHAY SEMICONDUCTORS www.vishay.com Modules By Kevin Liu This application note introduces Vishay s MTP rectifier-switch modules and discusses the assembly and PCB issues involved in their use. MTP modules

More information

Keysight N9030A Signal Analyzer Option BBA Analog Baseband IQ Inputs Retrofit Kit

Keysight N9030A Signal Analyzer Option BBA Analog Baseband IQ Inputs Retrofit Kit Keysight N9030A Signal Analyzer Option BBA Analog Baseband IQ Inputs Retrofit Kit Notice: This document contains references to Agilent. Please note that Agilent s Test and Measurement business has become

More information

5. Operating instructions: Octagon 200CL

5. Operating instructions: Octagon 200CL 1. Technical specifications 5. Operating instructions: Octagon 200CL Copyright 2015 by Endecotts Ltd. 33 1. Setting up Technical specifications SIEVE SHAKER MODEL: Octagon 200CL General Information The

More information

User s and Service Guide

User s and Service Guide User s and Service Guide Agilent Technologies 85056D 2.4 mm Economy Calibration Kit Agilent Part Number: 85056-90021 Printed in USA Print Date: January 2002 Supersedes: October 2000 Copyright 1996, 2000,

More information

Pin diameter specifications and challenges

Pin diameter specifications and challenges Federal Department of Justice and Police FDJP Federal Office of Metrology METAS Pin diameter specifications and challenges 32nd ANAMET meeting (NPL, Teddington) Juerg Ruefenacht 16 October 2009 Pin diameter

More information

Techniques for Precise Measurement Calibrations in the Field

Techniques for Precise Measurement Calibrations in the Field Techniques for Precise Measurement s in the Field Using FieldFox handheld analyzers up to 26.5 GHz Application Note This application note, which applies to cable and antenna and vector network analysis,

More information

Measurement Accuracy of Vector Network Analysis. Thilo Bednorz

Measurement Accuracy of Vector Network Analysis. Thilo Bednorz Measurement Accuracy of Vector Network Analysis Thilo Bednorz EDICON 2019 VNA Measurement and Accuracy Error sources occuring with VNA measurement measurement errors stochastic errors systematic errors

More information

Forté Technology, Inc. World leaders in moisture measurement

Forté Technology, Inc. World leaders in moisture measurement Forté Technology, Inc. Company History 1956 - Established to measure moisture in cashmere. 1965 - Expanded into pulp industry. 2008 - Facility moved to Taunton, MA. Present - Serving clients worldwide

More information

Reference Guide. Agilent Technologies Electronic Calibration Modules. RF Two-Port 85091C 85092C 85093C 85096C 85098C 85099C. RF Four-Port N4431B

Reference Guide. Agilent Technologies Electronic Calibration Modules. RF Two-Port 85091C 85092C 85093C 85096C 85098C 85099C. RF Four-Port N4431B Reference Guide Agilent Technologies Electronic Calibration Modules RF Two-Port 8509C 85092C 85093C 85096C 85098C 85099C RF Four-Port N443B Microwave Two-Port N4690B N469B N4692A N4693A N4694A N4696B Microwave

More information

Installation Note. Removable Hard Disk Drive Upgrade Kit. For All PNA Series RF Network Analyzers. Network Analyzer Model Number

Installation Note. Removable Hard Disk Drive Upgrade Kit. For All PNA Series RF Network Analyzers. Network Analyzer Model Number Installation Note Removable Hard Disk Drive Upgrade Kit For All PNA Series RF Network Analyzers Network Analyzer Model Number E8356A, E8357A, E8358A E8801A, E8802A, E8803A N3381A, N3382A, N3383A Upgrade

More information

Operation Manual for OPM-1315 Optical Power Meter

Operation Manual for OPM-1315 Optical Power Meter Operation Manual for OPM-1315 Optical Power Meter www.tonercable.com Table of Contents 1. Introduction... 1,2 1.1 Summary... 1 1.2 Product Features... 2 1.3 Technical Parameters... 2 1.4 Main Applications...

More information

Mini Probe. user manual. DM Models

Mini Probe. user manual. DM Models Mini Probe user manual DM Models 1.0: Index Section Title Page 1.0 Index...............................2 2.0 Safety Summary......................3 3.0 Introduction..........................4 4.0 Components

More information

Keysight Technologies N4000A, N4001A, N4002A SNS Series Noise Sources

Keysight Technologies N4000A, N4001A, N4002A SNS Series Noise Sources Keysight Technologies N4000A, N4001A, N4002A SNS Series Noise Sources 10 MHz to 26.5 GHz Technical Overview 02 Keysight N4000A, N4001A, N4002A SNS Series Noise Sources - Technical Overview Noise Sources

More information

RF Probing With Rohde & Schwarz ZNB VNA. PacketMicro, Inc Wyatt Drive, Suite 9, Santa Clara, CA

RF Probing With Rohde & Schwarz ZNB VNA. PacketMicro, Inc Wyatt Drive, Suite 9, Santa Clara, CA RF Probing With Rohde & Schwarz ZNB VNA PacketMicro, Inc. 1900 Wyatt Drive, Suite 9, Santa Clara, CA 95054 www.packetmicro.com Outline Why RF Probing Page 3 4 S-Probe Overview Page 5-8 RF Probing Tips

More information

Elecraft K3 KREF3 Output Level Modification

Elecraft K3 KREF3 Output Level Modification Elecraft K3 Revision A, September 15, 2015 Copyright 2015, Elecraft, Inc. All Rights Reserved Introduction This modification increases the output levels from the KREF3 Reference Oscillator to provide proper

More information

SMA Series - Crimp Plug and Jack

SMA Series - Crimp Plug and Jack Plug Features: SMA connectors are semi-precision, subminiature devices that provide repeatable electrical performance. These devices offer broadband performance with low reflection. These properties, along

More information

Economizer Regulator Filter Upgrade

Economizer Regulator Filter Upgrade Overview Chart LNG has released a new and improved economizer regulator, and filtering system. The new economizer filter system is designed to enhance reliability, and offer lower operating costs. In 2012

More information

User s and Service Guide

User s and Service Guide User s and Service Guide Agilent Technologies 85052D 3.5 mm Economy Calibration Kit This manual directly applies to 85052D calibration kits with serial number prefix 3106A. Agilent Part Number: 85052-90079

More information

INCHES (MILLIMETERS) CUSTOMER DRAWINGS AVAILABLE ON REQUEST

INCHES (MILLIMETERS) CUSTOMER DRAWINGS AVAILABLE ON REQUEST TNC Connectors Flexible Cable... 3 In-Series Adapters... 4 Specifications... 2 RF Adapter Kit Adapters... 6 N Connectors Panel Mount and In-Series Adapters... 10 Flexible Cable... 8 Specifications... 7

More information

To connect the AC adapter:

To connect the AC adapter: Replacing the AC Adapter Replacing the AC Adapter 3 Plug the power cord into a wall outlet. The power indicator turns on. To connect the AC adapter: Connect the power cord to the AC adapter. Power indicator

More information

Keysight Add Source and Receiver Attenuators Upgrade Kit

Keysight Add Source and Receiver Attenuators Upgrade Kit Keysight Add Source and Receiver Attenuators Upgrade Kit To Upgrade PNA N5221A or N5222A Option 401 to Option 417 Upgrade Kit Order Numbers: N5221AU-417 or N5222AU-417 Keysight Kit Number: N5222-60107

More information

TCM-4 Data Logger Battery Replacement Instructions

TCM-4 Data Logger Battery Replacement Instructions TCM-4 Data Logger Battery Replacement Instructions These instructions apply to the following product: TCM-4 Shallow Water Current Meter These instructions assume that the user has purchased a TCM-4 Battery

More information

Dimensions (inch) 3.74 x 1.45 x.95. PMA1-08S 70 MHz -10 GHz Inline Power Monitor with SMA Connectors 1

Dimensions (inch) 3.74 x 1.45 x.95. PMA1-08S 70 MHz -10 GHz Inline Power Monitor with SMA Connectors 1 PMA-8 The PMA-8 is a thru-line microwave power monitor, designed to non-invasively measure and display the power passing through a coaxial line. Battery life is exceptional due to proprietary low power

More information

Series PC 3.5 precision subminiature connectors

Series PC 3.5 precision subminiature connectors Series PC 3.5 precision subminiature connectors Description Content HUBER+SUHNER PC 3.5 connectors are precision connectors for use in microwave applications up to 26.5 GHz. They are especially suitable

More information

M-Line Series MN Multi-Pin Plugs and Receptacles Vac, Vdc

M-Line Series MN Multi-Pin Plugs and Receptacles Vac, Vdc Applications Designed for the roughest industrial applications, providing reliable, rugged service for power, control and instrumentation. For use in the roughest industrial applications, ranging from:

More information