Stratos Pro A2.. CONDI Conductivity Measurement with Toroidal Sensors

Size: px
Start display at page:

Download "Stratos Pro A2.. CONDI Conductivity Measurement with Toroidal Sensors"

Transcription

1 The Art of Measuring. User Manual English Stratos Pro A2.. CONDI Conductivity Measurement with Toroidal Sensors Latest Product Information:

2 About This Manual Return of products Please contact our Service Team before returning a defective device. Ship the cleaned device to the address you have been given. If the device has been in contact with process fluids, it must be decontaminated/disinfected before shipment. In that case, please attach a corresponding certificate, for the health and safety of our service personnel. Disposal Please observe the applicable local or national regulations concerning the disposal of waste electrical and electronic equipment. About this manual: This manual is intended as a reference guide to your device You don't have to read the book from front to back. Take a look at the Table of Contents or the Index to find the function you are interested in. Each topic is explained on a double-page spread with step-by-step instructions on how to configure the desired function. Clearly legible page numbers and headlines help you to quickly find the information: Configuration 5 Configuration Left page: How do I get to the function Alarm Settings CONTROL input (FLOW MIN, FLOW MAX) 1 1) Press menu key. 2) Select CONF using keys, press enter. 3) Select parameter set using keys, press enter. 2 4) Select ALARM menu using keys, press enter. 5) All items of this menu group are indicated by the ALA: code. Press enter to select menu, enter edit using arrow keys (see next page). Confirm (and proceed) using enter. 6) End: Press meas key until the [meas] mode 3 indicator is displayed. enter 4 5 enter ALARM: Delay enter Alarm: Sensocheck 5 Alarm: CONTROL input For flow monitoring: Max. flow alarm For flow monitoring: meas Min. flow alarm Menu item Action Choices CONTROL input The CONTROL input can generate an alarm when assigned to FLOW (flow monitoring) in the CONF menu: FLOW CNTR Flow measurement: allows monitoring the minimum and maximum flow (pulse counter) Alarm Specify value Default: liters/h Minimum flow FLOW MIN Alarm Specify value Default: liters/h Maximum flow FLOW MAX Right page: Which settings are provided for this function

3 CD-ROM Complete documentation: User manuals Safety instructions Certificates Quickstart guides Safety Instructions In official EU languages and others. EC Declarations of Conformity Knick Corporate Design 39 Manual / Bedienungsanleitungen Betriebsanleitung für Sauerstoff-Sensoren der Reihe SE 706-Memosens Å SENS MEMO MEMO SENS SENS MEMO Betriebsanleitung deutsch SE 706 SENS MEMO Berlin SE 706X/1-NMSN MEMO SENS Berlin Berlin BVS 10 ATEXXE089 X BVS 10 ATEX E089 II 1G ia IIC T3/T4/T6 II 1GSE Ex 706X/1-NMSN ia IIC Ex T3/T4/T SE 706X/1-NMSN SE 706X/2-NMSN BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T6 Safety Instructions Berlin SE 706X/2-NMSN BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T6 SE 706X/2-NMSN BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T6 Stratos Pro/MS/Evo Series Pg SE 706X/2-NMSN BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T6 Safety Instructions Betriebsanleitung Installation and Use in Hazardous Locations EC Declarations of Conformity Stratos Pro A2... ph Stratos Pro A2... PH deutsch Installation and Use in Hazardous Locations (shipped with hazardous-area devices only) Serial No Berlin Betriebsanleitung II 1 G Ex ia IIC T3/T4/T6 BVS 10 ATEX E 089 X 120 / 225 mm Knick Corporate Design Membrankappe Manual / Bedienungsanleitungen Es gibt u.a. folgende Formate: DIN A5 DIN A6 70 x 297 mm 70 x 230 mm 98 x 110 mm 39 Å MEMO SENS MEMO Membran Betriebsanleitung für Sauerstoff-Sensoren der Reihe SE 706-Memosens Formate Die große Produktbreite des Unternehmens mit ihren unterschiedlichsten Verpackungsgrößen erfordert entsprechende Formatvarianten der Bedienungsanleitungen. orange-red ring The Art of Measuring Berlin BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T6 BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T Berlin Berlin 0044 SE 706X/1-NMSN Berlin 706X/1-NMSN SE 706X/2-NMSN Memosens Pg Steckkopf BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T BerlinSE MEMO SENS Faltungen SENS Ø 12 mm Betriebsanleitung deutsch Aktuelle Produktinformation: Latest Product Information: Aktuelle Produktinformation: MEMO SE 706X/1-NMSN MEMO II 1G ia IIC T3/T4/T6 II 1GSE Ex 706X/1-NMSN ia IIC Ex T3/T4/T Berlin BVS 10 ATEXXE089 X BVS 10 ATEX E SE 706X/1-NMSN SE 706X/2-NMSN BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T6 SE 706X/2-NMSN BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T6 orange-red ring Berlin 0044 The Art of Measuring. BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T6 SE 706X/2-NMSN Pg 13.5 BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T SE 706X/2-NMSN Conductivity Measurement Stratos Pro A2... ph with Toroidal Sensors Stratos Pro A2... PH Betriebsanleitung II 1 G Ex ia IIC T3/T4/T6 BVS 10 ATEX E 089 X Quickstart Guide / 225 mm Kurzübersicht Guide rapide Inicio rápido Membrankappe Início rápido Guida rapida Faltungen Ø 12 mm Aktuelle Produktinformation: Latest Product Information: Aktuelle Produktinformation: First steps after installation: Operation Menu structure Calibration Error messages and recommended actions In German, English, French, Spanish, Portuguese, and Italian. More languages on CD-ROM and on our website: Serial No Berlin BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T6 Stratos Pro A2.. CONDI Abhängig von Format und Umfang der Bedienungsanleitungen werden unterschiedliche Faltungen effektiv eingesetzt: Klammerheftung, ZickzackFormate falz und Kreuzfalz. Die große Produktbreite des Unternehmens mit ihren unterschiedlichsten VerFarben packungsgrößen erfordert Die Bedienungsanleitungen entsprechende Formatsind grundsätzlich schwarz/ varianten der Bedienungsweiß angelegt. Farben anleitungen. werden nur zur Codierung, Es gibt u.a. folgende Darstellung von Farbdisplays Formate: und wichtiger Features eingesetzt. DIN A5 DIN A6 70 x 297 mm 70 x 230 mm 98 x 110 mm Quickstart Guide Berlin Berlin Berlin 0044 SE 706X/1-NMSN 0044 BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T6 BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T6 SENS Berlin Berlin 706X/1-NMSN SE 706X/2-NMSN Memosens Pg Steckkopf BVS 10 ATEX E089 X II 1G Ex ia IIC T3/T4/T BerlinSE Betriebsanleitung deutsch Quickstart SENS Berlin SE 706 Membran G Ex ia IIC T3/T4/T6 S 10 ATEX E 089 X Specific Test Report 0044 G Ex ia IIC T3/T4/T6 S 10 ATEX E 089 X Documents Supplied Abhängig von Format und Umfang der Bedienungsanleitungen werden unterschiedliche Faltungen effektiv eingesetzt: Klammerheftung, Zickzackfalz und Kreuzfalz. Farben Die Bedienungsanleitungen sind grundsätzlich schwarz/ weiß angelegt. Farben werden nur zur Codierung, Darstellung von Farbdisplays und wichtiger Features eingesetzt. 3

4 Contents Documents Supplied... 3 Introduction... 7 Intended Use... 7 Safety Information...10 Overview of Stratos Pro A2... CONDI...11 Assembly...12 Package Contents...12 Mounting Plan, Dimensions...13 Pipe Mounting, Protective Hood...14 Panel Mounting...15 Installation...16 Installation Instructions...16 Rating Plates / Terminal Assignments...16 Wiring of Stratos Pro A2... CONDI...17 Cable Preparation SE 655 / SE Wiring Example: SE 655 / SE Wiring Example: SE Wiring Example: Yokogawa ISC Wiring Example: Yokogawa IC 40S...22 Wiring Example: SE 670, SE Connecting a SE 670 / SE 680 Sensor...25 User Interface, Keypad...26 Display...27 Signal Colors (Display Backlighting)...27 Measuring Mode...28 Selecting the Mode / Entering Values...29 Color-Coded User Interface...31 Operating Modes...32 Menu Structure of Modes and Functions...33 HOLD Mode...34 Alarm...35 Alarm and HOLD Messages...36 Generating a Message via the CONTROL Input

5 Contents Configuration...38 Menu Structure of Configuration...38 Manual Switchover of Parameter Sets A/B...40 Configuration (Original for Copy)...45 Sensor...48 Sensor Verification (TAG, GROUP)...54 Current Output Current Output Temperature Compensation...68 CONTROL Input...72 Alarm Settings...74 Time and Date...78 Measuring Points (TAG/GROUP)...78 Calibration...81 Selecting a Calibration Mode...81 Calibration with Calibration Solution...82 Product Calibration...84 Calibration by Input of Cell Factor...86 Zero Calibration in Air / with Calibration Solution...87 Temp Probe Adjustment...88 Measurement...89 Diagnostics...90 Service...95 Operating States...98 A2...X: Supply Units and Connection Product Line and Accessories Specifications Calibration Solutions Concentration Measurement Concentration Curves Error Handling Error Messages

6 Contents Sensoface FDA 21 CFR Part Electronic Signature Passcodes Audit Trail Index Trademarks Passcodes

7 Introduction Intended Use Stratos Pro A2... CONDI is a 2-wire device for measurement of electrical conductivity and temperature in liquids using electrodeless (toroidal) sensors. Fields of application are: biotechnology, chemical industry, environment, food processing, water/wastewater treatment. Enclosure and mounting possibilities The sturdy molded enclosure is rated IP 67/NEMA 4X for outdoor use. It is made of glass-reinforced PBT / PC and measures 148 mm x 148 mm x 117 mm (H x W x D). It is provided with knockouts to allow: wall mounting (with sealing plugs to seal the enclosure), see page 13 post/pipe mounting (Ø mm, mm) see page 14 panel mounting (138 mm x 138 mm cutout to DIN 43700), see page 15 Protective hood (accessory) The weather protector, which is available as accessory, provides additional protection against direct weather exposure and mechanical damage, see page 14. Connection of sensors, cable glands For connecting the cables, the enclosure provides 3 knockouts for cable glands M20x1.5 2 knockouts for NPT 1/2 or rigid metallic conduit For quasi-stationary installations with Memosens sensors, we recommend using the M12 device socket (accessory ZU 0822) instead of a cable gland it allows simple replacement of the sensor cable without opening the device. Sensors The device has been designed for application with toroidal sensors, in particular for Models SE 655/656, SE 660, SE 670 and SE 680 (Knick). 7

8 Introduction Display Plain-text messages in a large, backlit LC display allow intuitive operation. You can specify which values are to be displayed in standard measuring mode ( Main Display, see page 30). Color-coded user interface The colored display backlighting signals different operating states (e.g., alarm: red, HOLD mode: orange, see page 31). Diagnostics functions Sensocheck automatically monitors sensor and cables and the Sensoface function clearly indicates the sensor condition, see page 120. Data logger The internal logbook (TAN SW-A002) can handle up to 100 entries up to 200 with AuditTrail (TAN SW-A003), see page parameter sets A, B The device provides two parameter sets which can be switched manually or via a control input for different process adaptations or different process conditions. For an overview of parameter sets (table for copy), see page 46. Password protection Password protection (passcodes) for granting access rights during operation can be configured, see page 97. TC process medium: selecting the compensation method The following temperature compensation methods are provided: linear (by entering a temperature coefficient), natural waters (nlf), ultrapure water with traces of NaCl, HCl, NH 3, NaOH, see page 68. 8

9 Control inputs Introduction Input + Input 5 6 Current input I input The analog (0) ma current input can be used for external temperature compensation (TAN required). See page 70. HOLD + HOLD HOLD input HOLD (floating digital control input) The HOLD input can be used for external activation of the HOLD mode, see page 35. CONTROL + CONTROL CONTROL input CONTROL (floating digital control input) The CONTROL input can be used either for parameter set selection (A/B) or for flow monitoring, see page 72. Signal outputs The device provides two current outputs (for transmission of measured value and temperature, for example). The output curve is adjustable (linear, bilinear or logarithmic), see page 56 et seq. Options Additional functions can be activated by entering a TAN (page 97). 9

10 Safety Information Be sure to read and observe the following safety instructions! The device has been manufactured using state of the art technology and it complies with applicable safety regulations. When operating the device, certain conditions may nevertheless lead to danger for the operator or damage to the device. See also the following documents (page 3): Safety Instructions Safety Instructions / Installation and Use in Hazardous Locations CAUTION! Commissioning must only be performed by trained personnel authorized by the operating company! Whenever it is likely that protection has been impaired, the device shall be made inoperative and secured against unintended operation. The protection is likely to be impaired if, for example: the device shows visible damage the device fails to perform the intended measurements after prolonged storage at temperatures below 30 C / 22 F or above +70 C / +158 F after severe transport stresses Before recommissioning the device, a professional routine test must be performed. This test must be carried out at the manufacturer's factory. Note: Before commissioning you must prove that the device may be connected with other equipment. 10

11 Overview of Stratos Pro A2... CONDI Overview receive hi receive lo A B CONDI input Output send lo C send hi D Output 2 RTD (GND) E 17 RTD F RTD (Sense) G Shield H RS Input + 5 Current Input 6 input HOLD + 10 HOLD HOLD 11 input CONTROL + 13 CONTROL CONTROL 14 input 11

12 Assembly Package Contents Check the shipment for transport damage and completeness! The package should contain: Front unit, rear unit, bag containing small parts Specific test report Documentation (cf p. 3) CD-ROM Fig.: Assembling the enclosure 1) Jumper (3 x) 2) Washer (1 x), for conduit mounting: Place washer between enclosure and nut 3) Cable tie (3 x) 4) Hinge pin (1 x), insertable from either side 5) Enclosure screw (4 x) 12 6) Sealing insert (1 x) 7) Rubber reducer (1 x) 8) Cable gland (3 x) 9) Filler plug (3 x) 10) Hexagon nut (5 x) 11) Sealing plug (2 x), for sealing in case of wall mounting

13 Assembly Mounting Plan, Dimensions ) Cable gland (3 x) 2) Knockouts for cable gland or ½ conduit, 21.5 mm dia. (2 knockouts) Conduit couplings not included! 3) Knockout for pipe mounting (4 x) 4) Knockout for wall mounting (2 x) Fig.: Mounting plan (All dimensions in mm!) 13

14 Assembly Pipe Mounting, Protective Hood ø40...ø ) Hose clamp with worm gear drive to DIN 3017 (2 x) 2) Pipe-mount plate (1 x) 3) For vertical or horizontal posts or pipes 4) Self-tapping screw (4 x) 4 Fig.: Pipe-mount kit, accessory ZU 0274 (All dimensions in mm!) Fig.: Protective hood for wall and pipe mounting, accessory ZU 0737 (All dimensions in mm!) 14

15 Assembly Panel Mounting < ) Circumferential sealing (1 x) 2) Screws (4 x) 3) Position of control panel 4) Span piece (4 x) 5) Threaded sleeve (4 x) Cutout 138 x 138 mm (DIN 43700) Fig.: Panel-mount kit, accessory ZU 0738 (All dimensions in mm!) 15

16 Installation Installation Instructions Installation of the device must be carried out by trained experts in accordance with this user manual and as per applicable local and national codes. Be sure to observe the technical specifications and input ratings during installation! Be sure not to notch the conductor when stripping the insulation! The supplied current must be galvanically isolated. If not, connect an isolator module. All parameters must be set by a system administrator prior to commissioning! Terminals: suitable for single wires / flexible leads up to 2.5 mm 2 (AWG 14) Rating Plates / Terminal Assignments Fig.: Terminal assignments of Stratos Pro A2... Fig.: Stratos Pro A2...N rating plate at outside bottom of front 16

17 Wiring of Stratos Pro A2... CONDI Installation Sensor connection MK-CONDI module Areas for placing the screwdriver to pull out the terminals Terminal row 1 Terminal row 2 Fig.: MK-CONDI module terminal assignments In addition: 2 HART pins (between terminal row 1 and 2) Fig.: Terminals, device opened, back of front unit 17

18 Cable Preparation SE 655 / SE 656 Preparing the shield connection Pre-assembled special cable for SE 655 / SE 656 sensors Insert the special cable through the cable entry into the terminal compartment. Remove the already separated part of the cable insulation (1). 3 Turn the shielding mesh (2) over the cable insulation (3). 4 Then shift the crimp ring (4) over the shielding mesh and tighten it using a pince (5) The pre-assembled special cable: Coax RD Coax WH BU RD RD BU GN WH YE GN/YE 18

19 Wiring Example: SE 655 / SE 656 Measuring task: Sensors: Conductivity, temperature SE 655/SE 656 sensor Connecting the pre-assembled cable receive hi Blue A receive lo Coax red Red send lo send hi RTD (GND) RTD RTD (sense) shield B C D E F G H Red Coax white Blue Green White Yellow Shield Green/yellow Cable Device 19

20 Wiring Example: SE 660 Measuring task: Sensor: Conductivity, temperature SE 660 sensor receive hi Blue A receive lo Coax red Red send lo send hi RTD (GND) RTD RTD (sense) shield B C D E F G H Red Coax black Blue Green White Yellow Shield Green/yellow Cable Device 20

21 Wiring Example: Yokogawa ISC 40 Measuring task: Conductivity, temperature Sensor: Yokogawa ISC40 (Pt 1000) receive hi receive lo send lo send hi RTD (GND) RTD RTD (sense) shield A B C D E F G H Jumper! Black Black Blue Red Black Coax BN Coax WH Cable Device Configuration settings for this sensor: Conductivity, temperature Sensor: 21

22 Wiring Example: Yokogawa IC 40S Measuring task: Conductivity, temperature Sensor: Yokogawa IC40S (NTC 30k) receive hi receive lo send lo send hi RTD (GND) RTD RTD (sense) shield A B C D E F G H Black Black Blue Red Black Coax BN Coax WH Cable Device Configuration settings for this sensor: Conductivity, temperature Sensor: Approx

23 Wiring Example: SE 670, SE 680 Measuring task: Conductivity, temperature Sensor: SE 670 NOTICE! Connection to RS-485 interface! Remove the measuring module. Sensor Brown Green Yellow White Transparent Cable Device The SE 670 / SE 680 sensors are connected to the RS-485 interface of the device for an A2... Series (2-wire) device, the measuring module slot must be empty. Therefore, first remove the measuring module from the slot (see next page). When the sensor model SE 670 / SE 680 is selected in the Configuration menu, the default values are taken as calibration data. They can then be modified by calibration. All calibration data of the SE 680-M sensor with Memosens protocol are stored in the sensor. 23

24 24

25 Connecting a SE 670 / SE 680 Sensor NOTICE! The slot for the MK-CONDI module must be empty be sure to remove the module! Areas for placing the screwdriver to pull out the terminals Connection of SE 670 / SE 680: Wire color Brown Green Yellow White, transparent shield 25

26 MEMO SENS User Interface, Keypad Display 2 Keypad 3 Rating plate (enclosure bottom) 3 Key Arrow keys up / down Arrow keys left / right Function Return to last menu level Directly to measuring mode (press > 2 s) Measuring mode: other display Retrieve information Show error messages Configuration: Confirm entries, next configuration step Calibration: Continue program flow Measuring mode: Call menu Menu: Increase/decrease a numeral Menu: Selection Previous/next menu group Number entry: Move between digits 26

27 Display 23 MEMO SENS Temperature 13 Info available 2 Sensocheck 14 Hold mode active 3 Interval/response time 15 Main display 4 Sensor data 16 Secondary display 5 Not used 17 Proceed using enter 6 Limit message: 18 Not used Limit 1 or Limit 2 19 Diagnostics 7 Alarm 20 Configuration mode 8 Service 21 Calibration mode 9 Parameter set 22 Measuring mode 10 Calibration 23 Sensoface 11 Memosens 24 Unit symbols 12 Waiting time running Signal Colors (Display Backlighting) Red Red blinking Orange Turquoise Green Magenta Alarm (in case of fault: display values blink) Input error: illegal value or wrong passcode HOLD mode (Calibration, Configuration, Service) Diagnostics Info Sensoface message 27

28 Measuring Mode After the operating voltage has been connected, the analyzer automatically goes to Measuring mode. To call the measuring mode from another operating mode (e.g. Diagnostics, Service): Hold meas key depressed (> 2 s). Sensoface indicator (sensor status) Time (or flow) Mode indicator (measuring) Hold meas key depressed for calling the measuring mode (pressing once more switches the display) Active parameter set (configuration) Display indicates OUT1: e.g. process variable Display indicates OUT2: e.g. temperature enter key Depending on the configuration, one of the following displays can be set as standard display for the measuring mode (see page 30): Measured value, time and temperature (default setting) Measured value and selection of parameter set A/B or flow Measured value and measuring point ( TAG ) Time and date Note: By pressing the meas key in measuring mode you can view the displays for approx. 60 sec. 28 You must configure the device for the respective measurement task!

29 Selecting the Mode / Entering Values To select the operating mode: 1) Hold meas key depressed (> 2 s) (directly to measuring mode) 2) Press menu key: the selection menu appears 3) Select operating mode using left / right arrow key 4) Press enter to confirm the selected mode 1 Selection menu Selected mode (blinks) To enter a value: 5) Select numeral: left / right arrow key 6) Change numeral: up / down arrow key 7) Confirm entry by pressing enter

30 Display in Measuring Mode The MAIN DISPLAY is the display which is shown in measuring mode. To call the measuring mode from any other mode, hold the meas key depressed for at least 2 sec. meas key enter key meas meas By pressing meas briefly you can step through further displays such as tag number (TAG) or flow (L/h). These displays are turquoise. After 60 sec they switch back to the main display. enter approx. 2 s Press enter to select a display as MAIN DISPLAY the secondary display shows MAIN DISPLAY NO. Use the UP / DOWN arrow keys to select MAIN DISPLAY YES and confirm by pressing enter. The display color changes to white. This display is now shown in measuring mode. 30

31 Color-Coded User Interface The color-coded user interface guarantees increased operating safety. Operating modes are clearly signaled. The normal measuring mode is white. Information text appears on a green screen and the diagnostic menu appears on turquoise. The orange HOLD mode (e.g. during calibration) is quickly visible as is the magenta screen which indicates asset management messages for predictive diagnostics such as maintenance request, pre-alarm and sensor wear. The alarm status has a particularly noticeable red display color and is also signaled by flashing display values. Invalid inputs or false passcodes cause the entire display to blink red so that operating errors are noticeably reduced. White: Measuring mode Red blinking: Alarm, errors Orange: HOLD mode Magenta: Maintenance request Turquoise: Diagnostics Green: Information texts 31

32 Operating Modes Diagnostics Display of calibration data, display of sensor data, performing a device self-test, viewing the logbook entries, display of hardware/software versions of the individual components. The logbook can store 100 events ( ). They can be displayed directly on the device. The logbook can be extended to 200 entries using a TAN (Option). HOLD Manual activation of HOLD mode, e.g. for servicing. The signal outputs adopt a defined state. Calibration Every sensor has typical characteristic values. Calibration is required to supply a correct measured value. The device checks which value the sensor delivers when measuring in a known solution. When there is a deviation, the device can be adjusted. In that case, the device displays the actual value and internally corrects the measurement error of the sensor. During calibration the device is in HOLD mode. During calibration the device remains in the HOLD mode until it is stopped by the operator. Configuration The analyzer must be configured for the respective measurement task. In the Configuration mode you select the connected sensor, the measuring range to be transmitted, and the conditions for warning and alarm messages. During configuration the device is in HOLD mode. Configuration mode is automatically exited 20 minutes after the last keystroke. The device returns to measuring mode. Service Maintenance functions (current source), passcode assignment, reset to factory settings, enabling of options (TAN). 32

33 Menu Structure of Modes and Functions Meas. mode (main display selectable) TAG display after 60 s CLK display after 60 s (Access via code, factory setting: 5555) Pressing the menu key (down arrow) opens the selection menu. Select the menu group using the left/right arrow keys. Pressing enter opens a menu item. Press meas to return. Display of calibration data Display of sensor data Self test: RAM, ROM, EEPROM, module Logbook: 100 events with date and time Display of direct, uncorrected sensor signals Display of software version, model designation, serial number Manual activation of HOLD mode, e.g. for sensor replacement. The signal outputs behave as configured (e.g. last measured value, 21 ma) Calibration with calibration solution Calibration by input of cell factor Zero calibration Product calibration Adjustment of temperature probe Configuring parameter set A Configuring parameter set B Display of measured values for validation (simulators) Current source, output 1 Current source, output 2 Specifying access codes for operating modes Reset to factory setting Enabling an option via TAN 33

34 HOLD Mode The HOLD mode is a safety state during configuration and calibration. Output current is frozen (LAST) or set to a fixed value (FIX). The HOLD mode is indicated by orange display backlighting. HOLD mode, display icon: Output signal response LAST: The output current is frozen at its last value. Recommended for short configuration procedures. The process should not change decisively during configuration. Changes are not noticed with this setting! FIX: The output current is set to a value that is noticeably different from the process value to signal the control system that the device is being worked at. Output signal during HOLD: Output current [ma] 21 Output signal for HOLD FIX setting = 21.0 ma Output signal for HOLD LAST setting 4 HOLD active HOLD active Terminating the HOLD mode The HOLD mode is ended by switching to measuring mode (hold meas key depressed). The display reads Good Bye, after that, the HOLD mode is exited. When the calibration mode is exited, a confirmation prompt ensures that the installation is ready for operation (e.g.: sensor reinstalled, located in process). 34

35 Alarm External activation of HOLD (SW-A005) The HOLD mode can be activated from outside by sending a signal to the HOLD input (e.g. from the process control system). Power supply V AC/DC HOLD input Stratos Pro A2... Process control system HOLD inactive HOLD active V AC/DC V AC/DC Manual activation of HOLD The HOLD mode can be activated manually from the HOLD menu. This allows checking or replacing a sensor, for example, without provoking unintended reactions at the outputs. Press meas key to return to selection menu. Alarm When an error has occurred, Err xx is displayed immediately. Only after expiry of a user-defined delay time will the alarm be registered and entered in the logbook. During an alarm the display blinks, the display backlighting turns red. Error messages can also be signaled by a 22 ma output current (see Configuration). 2 sec after the failure event is corrected, the alarm status will be deleted. 35

36 Alarm and HOLD Messages Message Released by Cause Alarm Sensocheck Polarization / Cable (22 ma) Error Messages Flow (CONTROL input) ERR 10: Conductance > 3500 ms HOLD HOLD via menu or input (Last/Fix) Configuration Calibration Service 36

37 Alarm and HOLD Messages Generating a Message via the CONTROL Input (min. flow / max. flow) The CONTROL input can be used for parameter set selection or for flow measurement (pulse principle), depending on its assignment in the Configuration menu. When preset to flow measurement an alarm can be generated when the measured flow exceeds a specified range: (specify value, default: 5 liters/h) (specify value, default: 25 liters/h) Power supply V AC/DC CONTROL input 37

38 Configuration Menu Structure of Configuration The device provides 2 parameter sets A and B. By switching between the parameter sets you can adapt the device to different measurement situations, for example. Parameter set B only permits setting of process-related parameters. The configuration steps are assigned to different menu groups. With the left/right arrow keys you can jump between the individual menu groups. Each menu group contains menu items for setting the parameters. Pressing enter opens a menu item. Use the arrow keys to edit a value. Press enter to confirm/save the settings. Return to measurement: Hold meas key depressed (> 2 s). Select menu group Menu group Code Display Sensor selection Menu item 1 Select menu item... Current output 1 Current output 2 Compensation Control input (parameter set or flow measurement) Alarm mode Setting the clock Measuring point Menu item... 38

39 Configuration Parameter set A/B: configurable menu groups The device provides 2 parameter sets A and B. By switching between the parameter sets you can adapt the device to different measurement situations, for example. Parameter set B only permits setting of process-related parameters. Menu group Parameter set A Parameter set B Sensor selection --- Current output 1 Current output 1 Current output 2 Current output 2 Compensation Compensation Control input --- Alarm mode Alarm mode Parameter set selection --- Setting the clock --- TAG TAG of measuring point --- GROUP GROUP of measuring points --- External switchover of parameter sets A/B You can switch between parameter sets A and B by applying a signal to the CONTROL input (parameter setting: ). Max. 30 V AC/DC 13 CONTROL 14 input e.g. process control system Stratos Pro A2... Parameter set A active Parameter set B active V AC/DC V AC/DC 39

40 Configuration Manual Switchover of Parameter Sets A/B Display Action Remark To switch between Manual selection of parameter sets: parameter sets must have been preset in CONFIG Press meas. mode. Default setting is a fixed parameter set A. Wrong settings change the measurement properties! PARSET blinks in the lower line. Select parameter set using and keys Select PARSET A / PARSET B Press enter to confirm. Cancel by pressing meas 40

41 Configuration Configuration Choices Default Sensor (SENSOR) SNS: SE 655, SE 656 SE 660, SE 670, SE680-K, SE680-M MEMOSENS OTHER SE 655 OTHER RTD TYPE 100PT / 1000PT / 1000PT 30 NTC CELL FACTOR XX.XXx TRANS RATIO XXX.Xx MEAS MODE Cond Conc % Sal Cond Cond MEAS RANGE xxx.x μs/cm x.xxx ms/cm x.xxx ms/cm xx.xx ms/cm xxx.x ms/cm x.xxx S/m xx.xx S/m Conc Solution -01- (NaCl) -01- (NaCl) -02- (HCl) -03- (NaOH) -04- (H2SO4) -05- (HNO3) -06- (H2SO4) -07- (HCl) -08- (HNO3) -09- (H2SO4) -10- (NaOH) TEMP UNIT C / F C CIP COUNT ON/OFF OFF ON CYCLES SIP COUNT ON/OFF OFF ON CYCLES CHECK TAG ON/OFF OFF CHECK GROUP ON/OFF OFF 41

42 Configuration Configuration Choices Default Output 1 (OUT1) OT1: Channel Cond/TMP Cond Output LIN / BiLIN / LOG LIN LIN BEGIN 4mA xxxx ms/cm END 20 ma xxxx ms/cm BiLIN BEGIN 4 ma END 20 ma CORNER X Input range: selected CHANNEL Vertex X : BEGIN CORNER X END (rising) BEGIN CORNER X END (falling) CORNER Y Input range: selected CHANNEL Default: 12 ma Vertex Y: (0) 4 ma CORNER Y 20 ma LOG BEGIN (0) 4mA Decades END 20 ma Decades TMP BEGIN (0) 4mA C C END 20 ma C TMP F BEGIN (0) 4mA F END 20 ma F FILTERTIME SEC 0000 SEC 22mA FAIL ON/OFF OFF 22mA FACE ON/OFF OFF HOLD MODE LAST/FIX LAST FIX HOLD-FIX ma ma Monitoring the sensor lines for breakage The sensor lines are monitored for breakage when the temperature is used for calculating the conductivity or concentration. If the sensor or line is broken, an alarm will be generated (output current FIX or 22 ma, depending on the configuration). If you want to output a conductivity value that is independent of the measured temperature (uncompensated), you can monitor the sensor lines for breakage by setting TEMP CHECK to ON in the Alarm menu. (See following page, ALARM menu) 42

43 Configuration Configuration Choices Default Output 2 (OUT2)... other steps like output 1 Temperature compensation (CORRECTION) Compensation for ultrapure water: NaCl, HCL, NH3 LIN TC LIQUID %/K 00.00%/K REF TEMP C C I-INPUT (only with TEMP EXT selected The I-input option must be enabled via TAN) I-INPUT 0 20 ma / 4 20 ma ma C BEGIN 4 ma C C END 20 ma C C F BEGIN 4 ma F END 20 ma F Control input (CNTR_IN) IN: CONTROL Parameter-set switchover (PARSET) or flow measurement (FLOW) PARSET FLOW FLOW ADJUST pulses/liter pulses/liter Alarm (ALARM) ALA: DELAYTIME SEC 0010 SEC SENSOCHECK ON/OFF OFF TEMP CHECK ON/OFF OFF FLOW CNTR *) ON/OFF OFF ON FLOW MIN **) L/h L/h FLOW MAX **) L/h L/h *) These menu items appear only if selected. **) Hysteresis fixed at 5% of threshold value 43

44 Configuration Configuration Choices Default Parameter set (PARSET) Select fixed parameter set (A) or switch between A/B via control input or manually in measuring mode Real-time clock (CLOCK) CLK: FORMAT 24 h / 12 h 24 h TIME hh/mm : h TIME hh/mm :59 AM / :59 PM DAY/MONTH / YEAR Measuring points (TAG / GROUP) TAG: (Input in text line) A...Z, 0...9, + < >? GROUP: (Input in text line) (fixed parameter set A) Sensor Verification (TAG, GROUP) When Memosens sensors are calibrated in the lab, it is often useful and sometimes even mandatory that these sensors will be operated again at the same measuring points or at a defined group of measuring points. To ensure this, you can save the respective measuring point (TAG) or group of measuring points (GROUP) in the sensor. TAG and GROUP can be specified by the calibration tool or automatically entered by the transmitter. When connecting an MS sensor to the transmitter, it can be checked if the sensor contains the correct TAG or belongs to the correct GROUP. If not, a message will be generated, Sensoface gets sad, and the display backlighting turns purple. The sad Sensoface icon can also be signaled by a 22 ma error current. Sensor verification can be switched on in the Configuration in two steps as TAG and GROUP if required. When no measuring point or group of measuring points is saved in the sensor, e.g., when using a new sensor, Stratos enters its own TAG and GROUP. When sensor verification is switched off, Stratos always enters its own measuring point and group. A possibly existing TAG/GROUP will be overwritten. 44

45 Configuration (Original for Copy) Two complete parameter sets are stored in the EEPROM. As delivered, the two sets are identical but can be edited. Note: Fill in your configuration data on the following pages or use them as original for copy. 45

46 Configuration (Original for Copy) Parameter Parameter set A Parameter set B SNS: Sensor type SNS: RTD type SNS: Cell factor SNS: Transfer ratio SNS: Measuring mode SNS: Measuring range SNS: Concentration determination SNS: Temperature unit SNS: CIP counter SNS: SIP counter SNS: CHECK TAG SNS: CHECK GROUP OT1: Process variable OT1: Lin/bilin/log output OT1: Current start OT1: Current end OT1: Vertex X (bilinear curve only) OT1: Vertex Y (bilinear curve only) OT1: Filter time OT1: FAIL 22 ma (error messages) OT1: FACE 22 ma (Sensoface messages) OT1: HOLD mode OT1: HOLD FIX current 46

47 Configuration (Original for Copy) Parameter Set A Set B OT2: Process variable OT2: Lin/bilin/log output OT2: Current start OT2: Current end OT2: Vertex X (bilinear curve only) OT2: Vertex Y (bilinear curve only) OT2: Filter time OT2: FAIL 22 ma (error messages) OT2: FACE 22 ma (Sensoface messages) OT2: HOLD mode OT2: HOLD FIX current COR: TC SELECT COR: Temp coefficient COR: Reference temperature COR: Current range COR: Current start COR: Current end IN: Parameter set A/B or flow IN: (Flow meter) Adjusting pulses/liter ALA: Delay ALA: Sensocheck on/off ALA: Tempcheck on/off ALA: Flow control FLOW CNTR on/off ALA: Minimum flow (hysteresis fixed at 5 %) ALA: Maximum flow (hysteresis fixed at 5 %) CLK: Time & Date TAG: Measuring point (tag number) GROUP: Group of measuring points *) These parameters cannot be adjusted in parameter set B, the values are the same as in parameter set A. 47

48 Configuration Sensor Select: Sensor type, temperature probe, cell factor, transfer ratio, measuring mode, range 1) Press menu key. 2) Select CONF using, press enter. 3) Select parameter set using keys, press enter. 4) Select SENSOR menu using keys, press enter. 5) All items of this menu group are indicated by the SNS: code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) by pressing enter. 6) Exit: Press meas key until the [meas] mode indicator is displayed. 5 6 Sensor type Temperature probe Cell factor Transfer ratio Measuring mode Measuring range Concentration determination Temperature unit Cleaning cycles Sterilization cycles CHECK TAG CHECK GROUP 5 48

49 5 Menu item Action Choices Sensor type Select sensor type using keys. Press enter to confirm. Configuration SE 655 SE 656, SE 660 SE 670, SE 680-K, SE 680-M, MEMOSENS, OTHER Temperature probe Cell factor Only with OTHER: Select type of temperature probe using keys. Press enter to confirm. Enter cell factor using keys. Press enter to confirm. 1000PT 100PT 30 NTC XX.XXx Transfer ratio Enter transfer ratio using keys. Press enter to confirm XXX.Xx Measuring mode Select desired measuring mode using keys. Press enter to confirm. Cond Conc % Sal Measuring range For cond measurement only Select desired range using keys. Press enter to confirm. x.xxx ms/cm, xx.xx ms/cm xxx.x ms/cm, x.xxx S/m xx.xx S/m 49

50 Configuration Sensor Select: Concentration determination, temperature unit 1) Press menu key. 2) Select CONF using keys, press enter. 3) Select parameter set using, press enter. 4) Select SENSOR menu using keys, press enter. 5) All items of this menu group are indicated by the SNS: code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter. 6) Exit: Press meas key until the [meas] mode indicator is displayed. 5 6 Sensor type Temperature probe Cell factor Transfer ratio Measuring mode Measuring range Concentration determination Temperature unit Cleaning cycles Sterilization cycles CHECK TAG CHECK GROUP 5 50

51 5 Menu item Action Choices Concentration For conc measurement determination only Temperature unit Select desired concentration solution using (see appendix for ranges). Press enter to confirm. Select C or F using keys. Configuration -01- (NaCl) -02- (HCl) -03- (NaOH) -04- (H 2 SO 4 ) -05- (HNO 3 ) -06- (H 2 SO 4 ) -07- (HCl) -08- (HNO 3 ) -09- (H 2 SO 4 ) -10- (NaOH) C / F Press enter to confirm. 51

52 Configuration Sensor (ISM and Memosens only) Adjust: Cleaning cycles, sterilization cycles 1) Press menu key. 2) Select CONF using keys, press enter. 3) Select parameter set using, press enter. 4) Select SENSOR menu using keys, press enter. 5) All items of this menu group are indicated by the SNS: code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter. 6) Exit: Press meas key until the [meas] mode indicator is displayed. 5 6 Select sensor type Select temp probe Select cell factor Select transfer ratio Select measuring mode Select range Concentration determination Temperature unit Cleaning cycles Sterilization cycles CHECK TAG CHECK GROUP 5 52

53 5 Menu item Action Choices CIP / SIP (ISM and Memosens only) Cleaning cycles on/off Select ON or OFF using keys. Activates/deactivates logging in extended logbook Press enter to confirm. Configuration ON/OFF Sterilization cycles on/off Select ON or OFF using keys. Activates/deactivates logging in extended logbook Press enter to confirm. ON/OFF The cleaning and sterilization cycles are logged to measure the load on the sensor. Suitable for biochemical applications (process temp approx C / F, CIP temp > +55 C / +131 F, SIP temp > +115 C / +239 F). Note: A CIP or SIP cycle is only entered into the logbook 2 hours after the start to ensure that the cycle is complete. With Memosens (e.g., SE680-M), an entry is also made in the sensor. 53

54 Configuration Memosens Sensor Sensor Verification (TAG, GROUP) 1 2 enter 3 1) Press menu key. 2) Select CONF using, press enter. 3) Select parameter set using keys, press enter. 4) Select SENSOR menu using keys, press enter. 5) All items of this menu group are indicated by the SNS: code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) by pressing enter. 6) Exit: Press meas key until the [meas] mode indicator is displayed. enter 4 enter 5 meas 6 Select sensor type Select temp probe Select cell factor Select transfer ratio Select measuring mode Select range Concentration determination Temperature unit Cleaning cycles Sterilization cycles CHECK TAG CHECK GROUP 5 54

55 5 Menu item Action Choices TAG ON/OFF Select ON or OFF using keys. Press enter to confirm. Configuration GROUP When switched on, the entry for TAG in the Memosens sensor is compared to the entry in the analyzer. If the entries differ, a message will be generated. Select ON or OFF using keys. Press enter to confirm. Function as described above ON/OFF Sensor Verification (TAG, GROUP) When Memosens sensors are calibrated in the lab, it is often useful and sometimes even mandatory that these sensors will be operated again at the same measuring points or at a defined group of measuring points. To ensure this, you can save the respective measuring point (TAG) or group of measuring points (GROUP) in the sensor. TAG and GROUP can be specified by the calibration tool or automatically entered by the transmitter. When connecting an MS sensor to the transmitter, it can be checked if the sensor contains the correct TAG or belongs to the correct GROUP. If not, a message will be generated, Sensoface gets sad, and the display backlighting turns purple. The sad Sensoface icon can also be signaled by a 22 ma error current. Sensor verification can be switched on in the Configuration in two steps as TAG and GROUP if required. When no measuring point or group of measuring points is saved in the sensor, e.g., when using a new sensor, Stratos enters its own TAG and GROUP. When sensor verification is switched off, Stratos always enters its own measuring point and group. A possibly existing TAG/GROUP will be overwritten. 55

56 Configuration Current Output 1 Output current range. Linear/Logarithmic. Current start. 1) Press menu key. 2) Select CONF using keys, press enter. 3) Select parameter set using, press enter. 4) Select OUT1 menu using keys, press enter. 5) All items of this menu group are indicated by the OT1: code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter. 6) Exit: Press meas key until the [meas] mode indicator is displayed. 5 Process variable LIN/biLIN/LOG output Current start Current end Time averaging filter Output current for error message Output current for Sensoface Output current during HOLD Output current for HOLD FIX 5 enter 6 56

57 5 Menu item Action Remark Process variable Select using keys: Cond: Conductivity TMP: Temperature Press enter to confirm. Then select characteristic (LIN/biLIN/LOG). Configuration Selectable decades with logarithmic setting (LOG): S/cm: ms/cm, 0.01 ms/cm, 0.1 ms/cm, 1.0 ms/cm, 10.0 ms/ cm, ms/cm, 1000 ms/cm S/M: S/m, 0.01 S/m, 0.1 S/m, 1.0 S/m, 10.0 S/m, 100 S/m Current start Current end Modify digit using keys, select next digit using keys. Press enter to confirm. Enter value using keys. Press enter to confirm. Entered value applies to selected process variable/ range If the adjusted range is exceeded, the device automatically switches to the next higher range (Autorange) Entered value applies to selected process variable/ range If the adjusted range is exceeded, the device automatically switches to the next higher range (Autorange) Assignment of measured values: Current start and current end Example 1: Range ms/cm Example 2: Range ms/cm [ms/cm] Advantage: Higher resolution in range [ms/cm] of interest Output current 20 [ma] Output current 20 [ma] 57

58 Configuration Current Output 1 Output current curve, bilinear 1) Press menu key. 2) Select CONF using keys, press enter. 3) Select parameter set using, press enter. 4) Select OUT1 menu using keys, press enter. 5) All items of this menu group are indicated by the OT1: code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter. 6) Exit: Press meas key until the [meas] mode indicator is displayed. 5 6 Process variable LIN/biLIN/LOG output Current start Current end Bilinear: Vertex X Bilinear: Vertex Y Time averaging filter Output current for error message Output current for Sensoface Output current during HOLD Output current for HOLD FIX 5 enter 58

59 5 Configuration Menu item Action Choices Output current Select using keys. LIN curve Press enter to confirm. Linear characteristic bilin Bilinear curve LOG Logarithmic curve Current start and current end Bilinear curve: Vertex X/Y Enter value using keys. Press enter to confirm. Enter value using keys. Press enter to confirm. Entered value applies to selected process variable/ range If the adjusted range is exceeded, the device automatically switches to the next higher range (Autorange) Entered value applies to selected vertex of bilinear curve "Corner X (process variable) and "Corner Y (output current) see figure below. Vertex of bilinear curve Output current Example: Current range ma, Current start: 0 µs/cm Current end: 200 µs/cm Vertex: "CORNER X : 10 µs/cm (process variable) "CORNER Y : 12 ma (output current) Result: The output current change in the range µs/cm is much greater than in the range µs/cm. Process variable 59

60 Logarithmic Curve Nonlinear output current characteristic: allows measurements over several decades, e.g. measuring very low values with a high resolution and high values with a low resolution. Parameters required: Start and end value Possible start and end values The start value must be at least one decade lower than the end value. Start value and end value must be specified in the same units (either in µs/cm or in S/m, see listing): ms/cm S/m 0.01 ms/cm 0.01 S/m 0.1 ms/m 0.1 S/m 1.0 S/m 10.0 S/m 100 S/m The start value is the next decade value below the lowest measured value. The end value is the next decade value above the highest measured value. The number of decades results from: Number of decades = log (end value) log (start value) The output current value is defined as follows: Output current = log(measured value) log(start value) Number of decades Measured values Start value 60 End value Decades

61 5 Configuration Menu item Action Choices Logarithmic curve Select using keys. LOG of output current Press enter to confirm. Logarithmic curve bilin Bilinear curve Start value Enter value using keys. LIN Linear characteristic Start value of logarithmic output curve End value Press enter to confirm. Enter value using keys. End value of logarithmic output curve Press enter to confirm. Possible start and end values for the logarithmic curve S/cm: S/m: 61

62 Configuration Current Output 1 Adjusting the time interval of the output filter 1) Press menu key. 2) Select CONF using keys, press enter. 3) Select parameter set using, press enter. 4) Select OUT1 menu using keys, press enter. 5) All items of this menu group are indicated by the OT1: code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter. 6) Exit: Press meas key until the [meas] mode indicator is displayed. 5 Process variable LIN/biLIN/LOG output Current start Current end Time averaging filter Output current for error message Output current for Sensoface Output current during HOLD Output current for HOLD FIX 5 enter 6 62

63 5 Menu item Action Choices Time averaging filter Enter value using keys. Configuration SEC (0000 SEC) Press enter to confirm. Time averaging filter To smoothen the current output, a low-pass filter with adjustable filter time constant can be switched on. When there is a jump at the input (100 %), the output level is at 63 % after the time interval has been reached. The time interval can be set from 0 to 120 sec. If the time interval is set to 0 sec, the current output directly follows the input. Note: The filter only acts on the current output, not on the display or the limit value! During HOLD the filter is not applied. This prevents a jump at the output. Display COND Time interval s 0/4-20 ma TMP Time interval s 63

Stratos Pro A2... PH. Betriebsanleitung. Stratos Pro A2.. MSPH. ph Measurement. Produktinformation: User Manual.

Stratos Pro A2... PH. Betriebsanleitung. Stratos Pro A2.. MSPH. ph Measurement. Produktinformation:  User Manual. II 1G Ex ia IIC T3/T4/T6 14163 Berlin 14163 Berlin SE 706X/1-NMSN SE 706X/2-NMS 14163 Berlin BVS 10 ATEXXE0 BVS 10 ATEX E089 II 1G ia IIC T II 1GSE Ex 706X/1-NM ia IIC Ex T3/T4/T6 The Art of Measuring.

More information

The Art of Measuring. Stratos MS A405. User Manual English. Conductivity Measurement. Latest Product Information:

The Art of Measuring. Stratos MS A405. User Manual English. Conductivity Measurement. Latest Product Information: The Art of Measuring. User Manual English Stratos MS A405 Conductivity Measurement Latest Product Information: www.knick.de Table of Contents 3 About This Manual...5 Documents Supplied...6 Introduction...

More information

Stratos Pro A2.. CC Dual Conductivity Measurement

Stratos Pro A2.. CC Dual Conductivity Measurement The Art of Measuring. User Manual English Stratos Pro A2.. CC Dual Conductivity Measurement Latest Product Information: www.knick.de Return of Products, Disposal Return of products Please contact our Service

More information

M420 Cond Ind. Instruction Manual.

M420 Cond Ind. Instruction Manual. M420 Cond Ind Instruction Manual www.mt.com/pro Warranty Warranty Defects occurring within 1 year from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender).

More information

The Art of Measuring. Stratos Evo A402. User Manual English. Conductivity Measurement. Latest Product Information:

The Art of Measuring. Stratos Evo A402. User Manual English. Conductivity Measurement. Latest Product Information: The Art of Measuring. User Manual English Stratos Evo A402 Conductivity Measurement Latest Product Information: www.knick.de Table of Contents 3 Basics...5 Documents Supplied...6 Introduction... 7 Overview

More information

Stratos Pro A4... MSCOND

Stratos Pro A4... MSCOND Stratos Pro A4... MSCOND Instruction Manual Latest Product Information: www.knick.de Warranty Warranty Defects occurring within 3 years from delivery date shall be remedied free of charge at our plant

More information

Process Analysis Systems. Stratos Pro A4 MSPH/MSPH. Chem Energy Pharm Food Water. Connection

Process Analysis Systems. Stratos Pro A4 MSPH/MSPH. Chem Energy Pharm Food Water. Connection Process Analysis Systems Chem Energy Pharm Food Water Stratos Pro A4 MSPH/MSPH Connection Connection of Memosens interface of 4-wire device with 2 Memosens sensors Model used: Stratos Pro A401N MSPH/MSPH

More information

Stratos Pro A2... MSPH

Stratos Pro A2... MSPH Stratos Pro A2... MSPH Instruction Manual Latest Product Information: www.knick.de Warranty Warranty Defects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage

More information

Process Analysis Systems. Stratos Pro A2 CONDI. Chem Energy Pharm Food Water. Connection

Process Analysis Systems. Stratos Pro A2 CONDI. Chem Energy Pharm Food Water. Connection Process Analysis Systems hem Energy Pharm Food Water Stratos Pro A2 ONDI onnection onnection of ONDI module with electrodeless sensors Model used: Stratos Pro A201N ONDI 1 ONDI module ontrol room Module

More information

Stratos Pro A4 PH. Connection

Stratos Pro A4 PH. Connection For up-to-date information, please visit www.m4knick.com Stratos Pro A4 PH Connection Connection of PH module with any desired analog sensor or with ISM or ISFET sensors. Operation of Memosens sensors

More information

Stratos Pro A2... PH. Instruction Manual. Latest Product Information:

Stratos Pro A2... PH. Instruction Manual.   Latest Product Information: Stratos Pro A2... PH Instruction Manual Latest Product Information: www.knick.de Warranty Warranty Defects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage

More information

Connection of OXY module with oxygen sensor (SE 706, Mettler Toledo InPro 6800, Hamilton Oxyferm) Model used: Stratos Pro A401N OXY/ V AC/DC

Connection of OXY module with oxygen sensor (SE 706, Mettler Toledo InPro 6800, Hamilton Oxyferm) Model used: Stratos Pro A401N OXY/ V AC/DC Process Analysis Systems Chem Energy Pharm Food Water Stratos Pro A4 OXY Connection Connection of OXY module with oxygen sensor (SE 706, Mettler Toledo InPro 6800, Hamilton Oxyferm) Model used: Stratos

More information

Stratos Pro A2... OXY

Stratos Pro A2... OXY Stratos Pro A2... OXY Instruction Manual Latest Product Information: www.knick.de Warranty Warranty Defects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage

More information

Instruction Manual DULCOMETER 2201 X Cond

Instruction Manual DULCOMETER 2201 X Cond Instruction Manual DULCOMETER 2201 X Cond ProMinent Dosiertechnik GmbH D-69123 Heidelberg Germany ProMinent Dosiertechnik GmbH Im Schuhmachergewann 5 11 D-69123 Heidelberg Postfach 10 17 60 D-69007 Heidelberg

More information

Stratos Pro A4... PH Instruction Manual

Stratos Pro A4... PH Instruction Manual Stratos Pro A4... PH Instruction Manual Latest Product Information: www.knick.de Warranty Warranty Defects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage

More information

Stratos Pro A2... PH. Betriebsanleitung. Produktinformation: English. Betriebsanleitung deutsch. Ex 706X/1-NM SE 706X/2-NM

Stratos Pro A2... PH. Betriebsanleitung. Produktinformation:  English. Betriebsanleitung deutsch. Ex 706X/1-NM SE 706X/2-NM II 1G Ex ia IIC T3/T4/T6 14163 Berlin 14163 Berlin SE 706X/1-NMSN SE 706X/2-NMS 14163 Berlin BVS 10 ATEXXE0 BVS 10 ATEX E089 II 1G ia IIC T II 1GSE Ex 706X/1-NM ia IIC Ex T3/T4/T6 The Art of Measuring.

More information

Portavo 904 X Multi. Multiparameter Ex

Portavo 904 X Multi. Multiparameter Ex Multiparameter Ex Portavo 904 X Multi The world's only portable device for liquid analysis with ATEX approval for measurements up to zone 0/1. Ideal for applications in the process industry. For the first

More information

Instruction Manual DULCOMETER 2201 X ph

Instruction Manual DULCOMETER 2201 X ph Instruction Manual DULCOMETER 2201 X ph ProMinent Dosiertechnik GmbH D-69123 Heidelberg Germany ProMinent Dosiertechnik GmbH Im Schuhmachergewann 5 11 D-69123 Heidelberg Postfach 10 17 60 D-69007 Heidelberg

More information

Instruction manual. O 2 Transmitter 4100 e/2(x)h

Instruction manual. O 2 Transmitter 4100 e/2(x)h Instruction manual O 2 Transmitter 4100 e/2(x)h Warranty Defects occurring within 1 year from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender). Subject

More information

Transmitter Your Consultant: 02/

Transmitter Your Consultant: 02/ Transmitter 7500 Your Consultant: 02/99 52 120 435 Mettler-Toledo GmbH, Process, Postfach, 8902 Urdorf, Switzerland Phone +41 (01) 736 22 11, Fax +41 (01) 736 26 36 Transmitter 7500 Warranty Defects occurring

More information

The Art of Measuring. User Manual English. Portavo 904(X) COND. Latest Product Information:

The Art of Measuring. User Manual English. Portavo 904(X) COND. Latest Product Information: The Art of Measuring. User Manual English Portavo 904(X) COND Latest Product Information: www.knick.de Basics 3 Return of products under warranty Please contact our Service Team before returning a defective

More information

The Art of Measuring. Multiparameter analyzers with Fieldbus technology. For analog, digital and Memosens sensors. Stratos

The Art of Measuring. Multiparameter analyzers with Fieldbus technology. For analog, digital and Memosens sensors. Stratos The Art of Measuring. Multiparameter analyzers with Fieldbus technology. For analog, digital and Memosens sensors Multi-Talents with Bus Communication With their exceptional functionality, reliability

More information

Stratos Pro A2... PH. Betriebsanleitung. Produktinformation: English. Betriebsanleitung deutsch. Ex 706X/1-NM SE 706X/2-NM

Stratos Pro A2... PH. Betriebsanleitung. Produktinformation:  English. Betriebsanleitung deutsch. Ex 706X/1-NM SE 706X/2-NM II 1G Ex ia IIC T3/T4/T6 14163 Berlin 14163 Berlin SE 706X/1-NMSN SE 706X/2-NMS 14163 Berlin BVS 10 ATEXXE0 BVS 10 ATEX E089 II 1G ia IIC T II 1GSE Ex 706X/1-NM ia IIC Ex T3/T4/T6 The Art of Measuring.

More information

Measuring Module M 700 Cond Ind 7700(X)

Measuring Module M 700 Cond Ind 7700(X) Measuring Module M 700 Cond Ind 7700(X) For Conductivity Measurement with Electrodeless Sensors 52121224 Warranty Defects occurring within 1 year from delivery date shall be remedied free of charge at

More information

Instruction manual. ph Transmitter 2100 e/2(x)h

Instruction manual. ph Transmitter 2100 e/2(x)h Instruction manual ph Transmitter 2100 e/2(x)h 70090 Warranty Defects occurring within 1 year from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender). Subject

More information

Conductivity Measuring System

Conductivity Measuring System Brief Operating Instructions Smartec S CLD134 Conductivity Measuring System These instructions are Brief Operating Instructions. For detailed information, please read the Operating Instructions. The complete

More information

Stratos PROFIBUS 2221 X Cond. Instruction Manual

Stratos PROFIBUS 2221 X Cond. Instruction Manual Stratos PROFIBUS 2221 X Cond Instruction Manual Warranty Defects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender). Accessories:

More information

Protos 3400(X) Process Analysis System

Protos 3400(X) Process Analysis System The Art of Measuring. Protos 3400(X) Process Analysis System User Manual English Measuring Module Protos COND 3400(X)-041 For Conductivity Measurement with 2- or 4-Electrode Sensors Latest Product Information:

More information

Indicators. Hazardous-Area Indicators. 830 X R Loop-Powered 16 Process Indicator. 830 X S1 Loop-Powered 16 Process Indicator

Indicators. Hazardous-Area Indicators. 830 X R Loop-Powered 16 Process Indicator. 830 X S1 Loop-Powered 16 Process Indicator Hazardous-Area 830 X R Loop-Powered 16 830 X S1 Loop-Powered 16 830 X S2 Loop-Powered 16 15 Loop-Powered Digital 830 X Loop-Powered 830 X R 830 X S1 830 X S2 The 830 X loop-powered digital indicators are

More information

Extremely robust process analyzers with unique color-coded user interface, for all common analog and digital sensors

Extremely robust process analyzers with unique color-coded user interface, for all common analog and digital sensors Process Analysis Systems Chem Energy Pharm Food Water Stratos Pro Extremely robust process analyzers with unique color-coded user interface, for all common analog and digital sensors Process analytics

More information

Portavo. Portavo. Applications

Portavo. Portavo. Applications Portavo Portavo The new portable meters for use either with analog or Memosens sensors. With graphical color display, data logger, Li-ion battery and USB port. Robust enclosure, protected against strong

More information

O i(X) module. Instruction manual. Order number:

O i(X) module. Instruction manual. Order number: O 2 4700i(X) module Instruction manual Order number: 52 121 275 Warranty Defects occurring within 1 year from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by

More information

Mettler-Toledo Smart Communications Protocol Revison 1.0: June 9, 2004 Cond Ind 7100/2(X)H Command Specification Page: 1 of 16

Mettler-Toledo Smart Communications Protocol Revison 1.0: June 9, 2004 Cond Ind 7100/2(X)H Command Specification Page: 1 of 16 Cond Ind 7100/2(X)H Command Specification Page: 1 of 16 Cond Ind 7100/2(X)H Transmitter Specific Command Specification using the HART Communications Protocol Revision 1.1 Initial Release: February 13,

More information

ph 2700i(X) module Instruction manual Order number:

ph 2700i(X) module Instruction manual Order number: ph 2700i(X) module Instruction manual Order number: 52 121 276 Warranty Defects occurring within 1 year from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by

More information

OPT2011. High-performance distance sensor. Operating Instructions

OPT2011. High-performance distance sensor. Operating Instructions OPT2011 High-performance distance sensor Operating Instructions Status: 15/07/2013 2 Table of Contents 1. Use for Intended Purpose 4 2. Safety Precautions 4 2.1. Safety Precautions 4 2.2. Laser/LED warning

More information

Technical Information Liquiline To Go CYM290 and Liquiline To Go Ex CYM291

Technical Information Liquiline To Go CYM290 and Liquiline To Go Ex CYM291 TI01198C/07/EN/02.15 71288268 Products Solutions Services Technical Information Liquiline To Go CYM290 and Liquiline To Go Ex CYM291 Portable multiparameter device for Memosens ph, conductivity and oxygen

More information

User Manual. Portamess 913 Cond and Portamess 913 X Cond

User Manual. Portamess 913 Cond and Portamess 913 X Cond User Manual Portamess 913 Cond and Portamess 913 X Cond Warranty Defects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender).

More information

Stratos Eco 2405 ph. User Manual. Latest Product Information:

Stratos Eco 2405 ph. User Manual. Latest Product Information: Stratos Eco 2405 ph User Manual Latest Product Information: www.knick.de 093783 Subject to change without notice. Return of Products Under Warranty Please contact our Service Team before returning a defective

More information

703 Laboratory Conductivity Meter

703 Laboratory Conductivity Meter 703 Laboratory Conductivity Meter Sensoface Sensoface monitors the sensor and measuring equipment and provides information on sensor selection and handling. It reports clock memory loss and requests regular

More information

Protos 3400(X) Process Analysis System

Protos 3400(X) Process Analysis System The Art of Measuring. Protos 3400(X) Process Analysis System User Manual English System Components FRONT and BASE Modular measuring system for liquid analysis with slots for up to 3 measuring and communication

More information

Transmitter Cond 7100 e /2 (X)H

Transmitter Cond 7100 e /2 (X)H Transmitter Cond 7100 e/2(x)h Command Specification Page: 1 of 16 Transmitter Cond 7100 e /2(X)H Transmitter Specific Command Specification using the HART Communications Protocol Revision 1.1 Initial Release:

More information

Conductivity-Panelmeter LF 9648 Conductivity measurement with 2-and 4-electrode cells

Conductivity-Panelmeter LF 9648 Conductivity measurement with 2-and 4-electrode cells Conductivity-Panelmeter LF 9648 Conductivity measurement with 2-and 4-electrode cells Features M LED-Display 14.2 mm red M Indicating range 2000(0) Digit M Measuring ranges programmable from 0... 2.000

More information

Operating Instructions VEGAMET 381

Operating Instructions VEGAMET 381 Operating Instructions VEGAMET 381 in out Contents Contents 1 About this document... 4 1.1 Function... 4 1.2 Target group... 4 1.3 Symbolism used... 4 2 For your safety... 6 2.1 Authorised personnel...

More information

The Art of Measuring. User Manual English. Paraly SW 112. Latest Product Information:

The Art of Measuring. User Manual English. Paraly SW 112. Latest Product Information: The Art of Measuring. User Manual English Paraly SW 112 Latest Product Information: www.knick.de Basics 3 Registered trademarks The following names are registered trademarks. For practical reasons they

More information

The Model 703 offers a wide range of practical features to meet the numerous requirements of everyday measuring tasks.

The Model 703 offers a wide range of practical features to meet the numerous requirements of everyday measuring tasks. Laboratory Meters The requirements for lab measurements become stricter every day. Quality assurance and measurement documentation in accordance with GLP are a must in many areas. 703 Laboratory Conductivity

More information

Measuring Module Protos OXY 3400(X)-067

Measuring Module Protos OXY 3400(X)-067 Measuring Module Protos OXY 3400(X)-067 For Trace Oxygen Measurement in Liquids and Gases Latest Product Information: www.knick.de 87404 Warranty Defects occurring within 3 years from delivery date shall

More information

Protos 3400(X) Process Analysis System

Protos 3400(X) Process Analysis System The Art of Measuring. Protos 3400(X) Process Analysis System User Manual English Measuring Module Protos PH 3400(X)-033 For Simultaneous Measurement of ph Value and Temperature with Pfauder Differential

More information

Product Line Specifications English. Portavo 904(X) PH. Latest Product Information:

Product Line Specifications English. Portavo 904(X) PH. Latest Product Information: Product Line Specifications English Portavo 904(X) PH Latest Product Information: www.knick.de 2 Product Line Accessories Item Order No. Robust field case (for meter, sensor, ZU 0934 various small parts

More information

Instruction manual. O 2 Transmitter 4100e FF

Instruction manual. O 2 Transmitter 4100e FF Instruction manual O 2 Transmitter 41e FF 69961 Warranty Defects occurring within 1 year from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender). Subject

More information

Modular analysis transmitter for ph/orp, conductivity measurement

Modular analysis transmitter for ph/orp, conductivity measurement Modular analysis transmitter for ph/orp, conductivity measurement Type 8285 can be combined with... High flexibility provided by modular concept for several measuring parameters Outstanding features for

More information

Bargraph for quick range overview The digital indicators provide a bargraph in addition to the digital display. This gives you all information

Bargraph for quick range overview The digital indicators provide a bargraph in addition to the digital display. This gives you all information Loop-Powered Digital Indicators 830 Loop-Powered Process Indicator The 830 loop-powered digital indicators are universally applicable. The range (either 0 to 20 ma or 4 to 20 ma) is simply selected via

More information

The Art of Measuring. Compact Transmitter with Modbus Interface. MemoRail Modbus

The Art of Measuring. Compact Transmitter with Modbus Interface. MemoRail Modbus The Art of Measuring. Compact Transmitter with Modbus Interface MemoRail Modbus MEMO SENS MemoRail Modbus A 1405 N Compact and Inexpensive. MemoRail Modbus is a space-saving transmitter with Modbus RTU

More information

DST550 Digi-Stem Thermometer

DST550 Digi-Stem Thermometer DST550 Digi-Stem Thermometer User Manual 09/16/09 WD1045 Rev C Wahl Instruments, Inc. 234 Old Weaverville Road Asheville, NC 28804 Toll Free 800-421-2853 Phone 828-658-3131 Fax 828-658-0728 www.palmerwahl.com

More information

Instruction Sheet PN 51A-6081C/rev.A June Model 6081-C Wireless Contacting Conductivity Transmitter

Instruction Sheet PN 51A-6081C/rev.A June Model 6081-C Wireless Contacting Conductivity Transmitter Instruction Sheet PN 51A-6081C/rev.A June 2009 Model 6081-C Wireless Contacting Conductivity Transmitter ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures,

More information

Communication Module M 700 EC 700(X)

Communication Module M 700 EC 700(X) Communication Module M 700 EC 700(X) For Actuating Probe Controllers (EC 400) - Automatic ph Measurement - Warranty Defects occurring within 1 year from delivery date shall be remedied free of charge at

More information

NN1 PH1 CC1 DM1 DB1. TABLE V - Options. Mounting Hardware. Instruction Books NN2 PH2 CC2 DM2 DB2. Certificates. PID Control NN C3.

NN1 PH1 CC1 DM1 DB1. TABLE V - Options. Mounting Hardware. Instruction Books NN2 PH2 CC2 DM2 DB2. Certificates. PID Control NN C3. 0---0 January 00 Quick Start Guide for UDA Universal Dual Analyzer For detailed instructions see UDA Universal Dual Analyzer Product Manual 0---. Step. Model Number Interpretation Write your analyzer model

More information

Instrumentation for Process Analytics

Instrumentation for Process Analytics Instrumentation for Process Analytics Transmitters, Analyzers, & Automated Systems digital and analog analytical loops Analytical Measurements Can Be Painful We re Changing the Status Quo of Process Analytics

More information

The Art of Measuring. User Manual English. Portavo 902 PH. Latest Product Information:

The Art of Measuring. User Manual English. Portavo 902 PH. Latest Product Information: The Art of Measuring. User Manual English Portavo 902 PH Latest Product Information: www.knick.de Basics 3 Warranty Defects occurring within 3 years from delivery date shall be remedied free of charge

More information

MODEL: 6DV-B. [2] TERMINAL BLOCK 0: None (Euro terminal block at the rear of indicator module)

MODEL: 6DV-B. [2] TERMINAL BLOCK 0: None (Euro terminal block at the rear of indicator module) Field-mounted Two-wire Signal Conditioners 6-UNIT -DIGIT LOOP POWERED INDICATOR (outdoor enclosure, intrinsically safe/explosion-proof) Functions & Features -digit LED display No external power source

More information

MODEL: 6DV-B. [2] TERMINAL BLOCK 0: None (Euro terminal block at the rear of indicator module)

MODEL: 6DV-B. [2] TERMINAL BLOCK 0: None (Euro terminal block at the rear of indicator module) Field-mounted Two-wire Signal Conditioners 6-UNIT -DIGIT LOOP POWERED INDICATOR (outdoor enclosure, intrinsically safe/explosion-proof) Functions & Features -digit LED display No external power source

More information

+GF+ SIGNET ph/orp Transmitter Instructions

+GF+ SIGNET ph/orp Transmitter Instructions GF SIGNET 8750 ph/orp Transmitter Instructions ENGLISH 8750.090 A9/99 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury.

More information

C100 Handheld Cond./TDS/Sal. Meter. Instruction Manual

C100 Handheld Cond./TDS/Sal. Meter. Instruction Manual C100 Handheld Cond./TDS/Sal. Meter Instruction Manual Table of Contents 1. Introduction ---------------------------------------------------------------------------------------------------------------------------

More information

FLOW MONITOR & TRANSMITTER

FLOW MONITOR & TRANSMITTER FLS M9.02 FLOW MONITOR & TRANSMITTER SAFETY INSTRUCTIONS General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other

More information

+GF+ SIGNET Conductivity/Resistivity Transmitter Instructions

+GF+ SIGNET Conductivity/Resistivity Transmitter Instructions GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions ENGLISH 8850.090 A9/99 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to

More information

CombiLyz AFI4 / AFI5 Inductive conductivity transmitter

CombiLyz AFI4 / AFI5 Inductive conductivity transmitter Main features Range from 500 µs/cm to 1000 ms/cm ll hygienic design Built in graphical display CombiView DFON Very fast temperature compensation Easy and full programmable with FlexProgrammer 9701 FI5

More information

CONDUCELL 4USF ARC SENSORS. Operating Instructions

CONDUCELL 4USF ARC SENSORS. Operating Instructions CONDUCELL 4USF ARC SENSORS Operating Instructions Important note Copyright 2009 HAMILTON Bonaduz AG, Bonaduz Switzerland. All rights reserved. The reproduction of any part of this document in any form

More information

Technical Information Memobase Plus CYZ71D

Technical Information Memobase Plus CYZ71D TI00502C/07/EN/01.12 71162017 Products Solutions Services Technical Information Multichannel multiparameter software for liquid analysis with digital Memosens sensors Measure - Calibrate - Document - Improve

More information

Instruction manual. Cond Ind Transmitter 7100e FF

Instruction manual. Cond Ind Transmitter 7100e FF Instruction manual Cond Ind Transmitter 71e FF 69967 Warranty Defects occurring within 1 year from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender). Subject

More information

Portavo 908 Multi. Comprehensive Data Logger

Portavo 908 Multi. Comprehensive Data Logger Multiparameter Portable multiparameter analyzer for digital ph/orp, conductivity and oxygen sensors with Memosens technology. Direct printer control via USB. The offers new features for the pharmaceutical

More information

CombiLyz AFI4 / AFI5 Inductive conductivity transmitter

CombiLyz AFI4 / AFI5 Inductive conductivity transmitter Main features Range from 500 µs/cm to 1000 S/cm ll hygienic design Built in graphical display CombiView DFON Very fast temperature compensation Easy and full programmable with FlexProgrammer 9701 FI5 split

More information

Output Signal Active Note DC V. Electrical Data Power Supply DC V, ±10%, 1.4 W

Output Signal Active Note DC V. Electrical Data Power Supply DC V, ±10%, 1.4 W Differential Pressure Sensor (Air) Differential pressure transmitter with 8 selectable ranges 0 to 5/10 V, 4 to 20 ma outputs and Modbus functionality. NEMA 4X / IP65 rated enclosure. For monitoring the

More information

Product manual plug, sockets and coupler 16 A.- Z0251

Product manual plug, sockets and coupler 16 A.- Z0251 Product manual plug, sockets and coupler 16 A.- Z0251 INDEX ELEKTRO BV Harregatplein 15 3214 VP Zuidland Netherlands T. +31(0)181 452120 http/www.indexelektro.nl email info@indexelektro.nl Purpose of these

More information

DSTHM-2 COMBINED T AND RH DUCT TRANSMITTER. Mounting and operating instructions

DSTHM-2 COMBINED T AND RH DUCT TRANSMITTER. Mounting and operating instructions Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4 STANDARDS 4 OPERATIONAL DIAGRAMS 5 WIRING AND

More information

Technical Information Memobase Plus CYZ71D

Technical Information Memobase Plus CYZ71D TI00502C/07/EN/02.13 71228167 Products Solutions Services Technical Information Memobase Plus CYZ71D Multichannel multiparameter software for liquid analysis with digital Memosens sensors Measure - Calibrate

More information

RST ROOM TEMPERATURE TRANSMITTER. Mounting and operating instructions

RST ROOM TEMPERATURE TRANSMITTER. Mounting and operating instructions Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4 STANDARDS 4 OPERATIONAL DIAGRAM 5 WIRING AND

More information

Modular analysis transmitter for ph/orp, conductivity measurement

Modular analysis transmitter for ph/orp, conductivity measurement Modular analysis transmitter for ph/orp, conductivity measurement Type 8285 can be combined with... High flexibility provided by modular concept for several measuring parameters Outstanding features for

More information

Adapter Kit - QSI Electronics to G-Series Meter Installation Instructions

Adapter Kit - QSI Electronics to G-Series Meter Installation Instructions Adapter Kit - QSI Electronics to G-Series Meter Installation Instructions These installation instructions cover the adapter kit installation of the QSI1, QSI2 and QSI3 versions of communications electronics

More information

OFP401P0189. Color Sensor. Operating Instructions

OFP401P0189. Color Sensor. Operating Instructions OFP401P0189 Color Sensor Operating Instructions Last update: 23 May 2013 2 Table of contents 1. Proper Use 4 2. Safety Precautions 4 3. EC Declaration of Conformity 4 4. Technical Data 5 4.1. Connection

More information

CX3100 Intelligent Conductivity Transmitter

CX3100 Intelligent Conductivity Transmitter CX3100 Intelligent Conductivity Transmitter Operation Manual CX3100 CONTENTS Precautions for Installation Brief Instruction 1 1. Specifications 4 2. Assembly and Installation 2.1 Transmitter Installation

More information

GREISINGER electronic GmbH. D Regenstauf, Hans-Sachs-Straße 26. T-Logg 120 W -... T-Logg 120 K -...

GREISINGER electronic GmbH. D Regenstauf, Hans-Sachs-Straße 26. T-Logg 120 W -... T-Logg 120 K -... E39.0.1X.6C-01 Data logger for standard signals as of version 1.0 Operating Instruction T-Logg 120... T-Logg 120 W -... T-Logg 120 K -... GREISINGER electronic GmbH D - 93128 Regenstauf, Hans-Sachs-Straße

More information

Versatile Transmitter Series For Intelligent Measurement Solutions. INGOLD Leading Process Analytics

Versatile Transmitter Series For Intelligent Measurement Solutions. INGOLD Leading Process Analytics INGOLD Leading Process Analytics Transmitter Series M400 / M420 / M300 Transmitters Extended ISM functions Easy handling Intuitive user interface Multi-parameter Versatile Transmitter Series For Intelligent

More information

RTT2 ROOM TEMPERATURE SWITCH. Mounting and operating instructions

RTT2 ROOM TEMPERATURE SWITCH. Mounting and operating instructions Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS PRODUCT DESCRIPTION ARTICLE CODES INTENDED AREA OF USE TECHNICAL DATA STANDARDS OPERATIONAL DIAGRAMS WIRING AND CONNECTIONS

More information

THERMASGARD TM 65. Immersion screw-in duct temperature measuring transducer, calibratable, with multi-range switching and active output TECHNICAL DATA

THERMASGARD TM 65. Immersion screw-in duct temperature measuring transducer, calibratable, with multi-range switching and active output TECHNICAL DATA Patented quality product (Immersion sensor patent no. DE 10 2012 017 500.0) Calibratable temperature measuring transducer THERMASGARD with eight switchable measuring ranges, continuous linear output, enclosure

More information

Part 2: Operation. Liquiline M CM42. Operating Instructions. Two-wire transmitter for conductivity (digital sensors)

Part 2: Operation. Liquiline M CM42. Operating Instructions. Two-wire transmitter for conductivity (digital sensors) Operating Instructions Liquiline M CM42 Two-wire transmitter for conductivity (digital sensors) Part 2: Operation BA382C/07/en/POD Valid as of: Software version 13.04.01 Operating concept Fig. 1: Press

More information

Operating manual LRD 325. Conductivity measuring cell to be installed in pipelines

Operating manual LRD 325. Conductivity measuring cell to be installed in pipelines Operating manual Conductivity measuring cell to be installed in pipelines ba55321de02 10/2010 Copyright Weilheim 2010, WTW GmbH Reprinting - even as excerpts - is only allowed with the explicit written

More information

SensoTrans DMS P 32200

SensoTrans DMS P 32200 ProLine Interface Technology Strain Gauge Transmitters The transmitters for strain gauge full bridges in a 6 mm housing with infrared interface and SIL approval. The Task In many different industrial applications

More information

Flow Measurement SITRANS F US Inline

Flow Measurement SITRANS F US Inline Overview Function Displays and keypad Operation of the SITRANS FUS060 transmitter can be carried out using: Keypad and display unit HART communicator PC/laptop and SIMATIC PDM software via HART communication

More information

Multichannel, multifunction Transmitter/Controller

Multichannel, multifunction Transmitter/Controller Multichannel, multifunction Transmitter/Controller Type 8619 can be combined with... Compatible with most common flow, ph/o.r.p. and conductivity sensors, directly connected Analogue and digital input

More information

Indicating Digital Transmitter for Differential Pressure and Flow. Media 6 with LCD Media 6 with LED

Indicating Digital Transmitter for Differential Pressure and Flow. Media 6 with LCD Media 6 with LED Indicating Digital Transmitter for Differential Pressure and Flow Media 6 with LCD Media 6 with LED Application Microprocessor-controlled transmitter for measuring and indicating the differential pressure

More information

Operating manual. Series 450. Compact thermometer. BA01235O/09/en/

Operating manual. Series 450. Compact thermometer. BA01235O/09/en/ Operating manual Series 450 Compact thermometer BA01235O/09/en/01.13 71220279 2 Table of contents 1 Safety instructions... 3 2 Installation... 4 3 Wiring overview... 7 4 Commissioning... 7 5 Operation...

More information

Insertion Flowmeter for compressed air and gases DN50 - DN700 (2-28 )

Insertion Flowmeter for compressed air and gases DN50 - DN700 (2-28 ) The flow meter is based on the thermal mass flow measurement and is ideal for measuring the flow of compressed air and gases in pipes from DN50 (2 ) to DN700 (28 ). With the, the consumption of compressed

More information

KOBOLD TDA Series Digital Temperature Transmitter

KOBOLD TDA Series Digital Temperature Transmitter KOBOLD TDA Series Digital Temperature Transmitter TDA-15 Series User Instructions KOBOLD Instruments Inc. 1801 Parkway View Drive Pittsburgh, PA 15205 Phone (412) 788-2830 Fax (412)-788-4890 www.koboldusa.com

More information

TRANSMITTERS. HygroFlex & HygroClip-EX. Applications. Highlights

TRANSMITTERS. HygroFlex & HygroClip-EX. Applications. Highlights HygroFlex & HygroClip-EX HygroFlex series may be used together with the intrinsically safe HygroClip-EX probes. The relative humidity and temperature can be displayed and output as analog signals. Calculated

More information

The Art of Measuring. User Manual English. Portavo 904(X) PH. Latest Product Information:

The Art of Measuring. User Manual English. Portavo 904(X) PH. Latest Product Information: The Art of Measuring. User Manual English Portavo 904(X) PH Latest Product Information: www.knick.de Basics 3 Warranty Defects occurring within 3 years from delivery date shall be remedied free of charge

More information

Flow Measurement SITRANS F US Inline

Flow Measurement SITRANS F US Inline Overview Function Displays and keypad Operation of the SITRANS FUS060 transmitter can be carried out using: Keypad and display unit HART communicator PC/laptop and SIMATIC PDM software via HART communication

More information

Siemens AG Continuous level measurement - Radar transmitters. Configuration. Mounting on a nozzle. Installation min.

Siemens AG Continuous level measurement - Radar transmitters. Configuration. Mounting on a nozzle. Installation min. Overview Configuration Mounting on a nozzle Min. clearance: 10 (0.4) Installation min. 1 000 (39) is a 2-wire 2 GHz pulse radar level transmitter for continuous monitoring of solids in storage vessels

More information

EE360. High-End Moisture in Oil Transmitter. Typical applications. Features EE360 EE360. (+34)

EE360. High-End Moisture in Oil Transmitter. Typical applications. Features EE360 EE360. (+34) is dedicated for reliable monitoring of lubrication, hydraulic and insulation oils as well as diesel fuel. In addition to highly accurate measurement of water activity (a w ) and temperature (T), calculates

More information

The Art of Measuring. Compact. Cost-efficient. Digital. Plug & Measure with the new analyzers for Memosens. MemoRail

The Art of Measuring. Compact. Cost-efficient. Digital. Plug & Measure with the new analyzers for Memosens. MemoRail The Art of Measuring. Compact. Cost-efficient. Digital. Plug & Measure with the new analyzers for Memosens The Essence of Measuring is the first genuinely compact, digital analyzer for measuring ph values,

More information

SITRANS F C. Transmitter MASS 6000 Ex d compact/remote 3/195

SITRANS F C. Transmitter MASS 6000 Ex d compact/remote 3/195 Overview MASS 6000 is based on the latest developments within digital signal processing technology engineered for high performance, fast flow step response, fast batching applications, high immunity against

More information

Operating instructions Electronic pressure sensor. PM205x /01 09/2011

Operating instructions Electronic pressure sensor. PM205x /01 09/2011 Operating instructions Electronic pressure sensor PM205x 706071/01 09/2011 Contents 1 Preliminary note...3 1.1 Symbols used...3 2 Safety instructions...3 3 Functions and features...4 3.1 Applications...4

More information