1 Signet 5700 ph/orp Monitor ph Instructions English * * Rev. K 3/06 English/pH CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents 1. Power Connections. Compatible Electrode/Preamp Wiring ma Current Output Connections 4. Relay Connections 5. Relay Operation 6. Menu Functions 7. Parts and Accessories 8. Specifi cations 9. Quick Reference Menu Parameters 10. Troubleshooting 11. Maintenance Remove terminal blocks for easy wiring Freq. IN Freq. IN Sen. Pwr. Iso. Gnd Std. Sensor Open Collector Sensor LISTED UL 77CJ 14 V 10 W Gnd Gnd PLS AUX output output 1 Total reset Patent No. D376,38 Flow 1 Freq. IN Freq. IN Sen. Pwr. Iso. Gnd Std. Sensor Open Collector Sensor LISTED UL 77CJ 14 V 10 W Gnd Gnd PLS AUX output output 1 Total reset Patent No. D376,38 Flow Power Connections. Compatible Electrode/Preamp Wiring CAUTION! Never connect 110 VAC or 0 VAC to rear power terminals. High voltage AC will damage instrument and void warranty. 14 V 10 W 5700 Terminals External power supply 1 4 VDC OR 1 4 VAC Brown Blue mv Input Brown wire Silver (earth ground) Iso. Gnd Blue wire T Green wire Green T White wire White Red Black V Red Wire V Black wire 40 ma ph/orp Preamp: Signet 70 TwistLock Electrodes 5700 Terminals DryLoc Electrodes Preamp: Signet 7601 Signet 760 = Double Insulated = DC or AC power Technical Notes: To reduce the possibility of noise interference, isolate AC power lines from signal lines. Maximum 40 ma loop impedance (sec. 3) is affected by the supply voltage. Use six conductor shielded cable for cable extensions up to 10 m (400 ft) max. (# ) Cable shield must be maintained through cable splice
2 3. 40 ma Current Output Connections Technical Notes: ** 1/8A fuse recommended (customer supplied) * 40 ma output is internally powered (nonisolated), maximum loop impedance 350 Ω with a 1 V instrument supply voltage, 950 Ω with a 4 V instrument supply voltage. To isolate output and prevent ground loop problems: 1. Use monitor device with isolated inputs, or. Use separate DC supply for 5700 and monitor device, or 3. Power 5700 with1 4 VAC step down transformer 4. Relay Connections Two internal relay contact sets (COM, NO, and NC) may be used for external device control. Front panel LED annunciators indicate the activation status of each relay. Each relay can control up to two devices simultaneously, as shown. Relay operation modes include Low alarm, High alarm, and Proportional Pulse (sec. 5). Common device connections include: Pulse mode metering pump control Pulse mode solenoid valve control Low or High mode warning lamps Low or High mode bells or sirens Low or High mode external heavyduty relay Wiring Example Right Device A IS powered when relay is deenergized (front panel LED "off"). Power is discontinued when the relay setpoint is reached (front panel LED "on"). Device B IS NOT powered when the relay is deenergized. Power is applied after the relay setpoint is reached. 5. Relay Operation 5700 Terminals Fuse** 1/8A 40 ma Relay ph/orp Technical Notes: Maximum relay contact ratings: 5 30 VDC, 5 15 VAC, or 3 50 VAC An external heavyduty relay must be used for devices with surge currents or operating currents that exceed the above specifi cations. * 5700 Terminals Relay 1 NO C NC NO C NC External AC/DC power 4 0 ma* Relay Deenergized NO C NC Device A Device B Relay Energized NO C NC A. LOW alarm mode The relay is energized when the ph drops below the setpoint, and is deenergized when the ph rises above the setpoint plus hysteresis (sec. 6.3E, 6.3F). Low setpoint= Adjustable hysteresis= Relay energized= Relay deenergized= ph B. HIGH alarm mode The relay is energized when the ph rises above the setpoint and is deengergized when the ph falls below the setpoint plus hysteresis (sec. 6.3E, 6.3F). High setpoint= Adjustable hysteresis= Relay energized= Relay deenergized= ph C. Proportional PULSE mode The proportional pulse relay configuration is primarily designed for metering pump control. The operator is prompted to enter a minimum and maximum ph setpoint and maximum pulse rate for the assigned relay (sec. 6.3G, 6.3H). Relay pulse width is fixed at 130 ms. Refer to the operation examples below. 300 pulses Metering pump chemical addition (dry contact activation type required) Example 1 (right): As the process ph rises above the minimum pulse setpoint (4.00 ph) the relay begins pulsing; triggering the metering pump for chemical addition. As the ph continues to rise, pulsing accelerates proportionally until the maximum programmed pulse rate of 300 pulses/minute and setpoint (10.00 ph) are reached, forcing the process ph back down to intended levels (e.g. 4.0 ph). ph 0 pulses 30 Relay 1 or Relay 0 ph ph Example (right): As the process ph falls below the minimum pulse setpoint (10.00 ph) the relay begins pulsing; triggering the metering pump for chemical addition. As the ph continues to decrease, pulsing accelerates proportionally until the maximum programmed pulse rate of 300 pulses/minute and setpoint (4.00 ph) are reached, forcing the process ph back up to intended levels (e.g ph). 300 pulses 3 Relay 1 or Relay 0 pulses ph 0 ph ph Signet 5700 ph/orp Monitor ph Instructions
3 ENTER 6. Menu Functions To access either CALIBRATE or OPTIONS menus, press and hold the ENTER key as illustrated below: VIEW (sec. 6.1) CALIBRATE (sec. 6.3) OPTIONS (sec. 6.4) Press & hold ENTER key for access: ENTER seconds 5 seconds Menus: VIEW menu (sec. 6.1): The VIEW menu is displayed during standard operation. The operator can navigate freely through the menu by pressing either UP or DOWN arrow keys. CALIBRATE Menu (sec. 6.3): The CALIBRATE menu contains all critical display setup and output parameters. A simple security code feature prevents unauthorized tampering. The operator is required to enter a simple access code for menu access. The same code also unlocks OPTIONS menus. OPTIONS Menu (sec. 6.4): The OPTIONS menu contains setup and display features that are seldom accessed for minor display or output adjustments. Reversible Dial Your 5700 includes a reversible dial face for ORP use. See enclosed ORP manual for operation details. A B ALARM 1 ALARM 1000 C ORP 1000 ph 0 14 ALARM 1 ALARM D 6.1 VIEW Menu H A. 1 Choose: ph/temperature Change: 3 Save: Menu Displays A H: (Factory default displays shown in menu column 1) A. Active ph and temperature display B. C. mv input Loop output B. Input display: shows electrode input mv C. Loop output display: shows the loop current output level. D. ph range display: shows the programmed min and max meter range (sec. 6.3A) D. Range E. Relay 1 display: this display shows the programmed operation mode and setpoint for relay 1 (sec. 6.3E) E. Relay 1 F. Relay display: this display shows the programmed operation mode and setpoint for relay (sec. (6.3E) F. Relay G. Last calibration: this display shows a user defi ned setup date for maintenance records. This feature is not an internal timer or calender. G. H. Last calibration Buffer calibration A Press and release to access EASY CAL Procedure See EASY CAL Procedure (Section 6.) H. EASY CAL: press the right arrow key to access the EASY CAL buffer calibration procedure (sec. 6.). Signet 5700 ph/orp Monitor ph Instructions 3
4 6. EASY CAL Procedure Requirements This procedure simplifi es system calibration using standard ph buffers. Use ph 4.0, 7.0, 10.0 ph buffers only (sec. 7). If standard ph buffers are not available, the system can be calibrated using the CALIBRATE menu temperature, standard, and slope settings (sec. 6.3BD). Access CALIBRATE menu (sec. 6.3B) and set sensor temperature before performing EASY CAL for new electrode installation. 1 To Calibrate: Press keys in sequence will appear during code entry. Response: to continue, 3 To Accept: Place electrode tip in first ph buffer; Allow for stabilization Press ENTER to accept first buffer calibration. ph 4.0, 7.0 or seconds* * For best results, gently stir the submerged electrode for approximately 5 seconds during the stabilization period (step ). Place electrode tip in second (different) ph buffer; Allow for stabilization Press ENTER to accept second buffer calibration. Large temperature differences from process fl uids to buffers may require longer stabilization time. ph 4.0, 7.0 or seconds* Display returns to VIEW Menu in 3s To exit EASY CAL at any time without saving changes: quick press Technical notes: The difference between the actual mv reading and value shown in the table below is a good indication of the ph electrode condition. Differences in excess of 50 mv indicate a need to service the electrode. Refer to electrode manual for maintenance recommendations. Theoretical mv values 5 C mv Signet 5700 ph/orp Monitor ph Instructions
5 6.3 CALIBRATE Menu Requirements System calibration is possible with two known ph solutions within 0 to 14 ph and an accurate reference thermometer. The EASY CAL procedure (sec. 6.) is recommended when standard ph buffers are available. If EASY CAL is performed, manual calibration steps B D below are not required. Always calibrate solution temperature fi rst (step B below) before setting ph standard and slope (steps C D below). Single point calibration can be done by setting "Set Standard" (step C) only. Dual point calibration (recommended) can be done by setting "Set Standard" followed by "Set Slope" (steps BC). Press keys in sequence will appear during code entry. to continue, I A. B. C. D. E. F. 1 Choose: Range Electrode temp. Electrode standard Electrode slope Relay Meter min max Allow for stabilization Allow for stabilization Allow for stabilization Mode Setpoint Change: Relay Low or High Mode Selected 3 Save: ENTER "SAVING" briefly displays Menu Settings A I : (Factory default displays shown in menu column 1) A. Sets Min Max meter dial range, (factory installed dial, ). Contact factory for custom dial confi gurations. Does not effect 4 to 0 ma output B. Sets process solution temperature C. Sets electrode standard to any value from 014 ph. Standard calibration solution must be ph units from slope solution (step D) D. Sets electrode slope to any value from 014 ph. Slope calibration solution must be ph units from standard solution (step C) Menu items E H repeat for relay setup. Relay Hysteresis Relay Pulse Mode Selected, step E E. Sets relay operation mode Low or High, and setpoint, ph (sec. 5AB) G. Relay Mode Range F. Sets relay hysteresis, ph. Set to zero to disable feature (sec. 5AB) H. Relay Pulse rate G. Sets relay minimum and maximum pulse setpoint, ph (sec. 5C). I. H. Sets relay pulse rate, pulses/minute. To return to VIEW: Last calibration A quick press Optional To restore original value: quick press I. Sets user defi ned setup date for maintenance records. This feature is not an internal timer or calender Quick Reference Calibration Procedures: Point Quick Calibration (recommended): 1. Verify/enter solution temperature ±0.5 C (step B). Set solution standard (step C) 3. Set solution slope (step D) 1Point Calibration (optional): 1. Verify/enter solution temperature ±0.5 C (step B). Set solution standard (step C) Signet 5700 ph/orp Monitor ph Instructions 5
6 6.4 OPTIONS Menu Press keys in sequence will appear during code entry. to continue, F A. 1 Choose: Change: 3 Save: Menu Settings A F: (Factory default displays shown in menu column 1) Contrast Low High A. Selects display contrast: 5 levels B. Loop output 4 ma setpoint B. Sets 4 ma setpoint: ph. 4 ma and 0 ma setpoints are reversible C. D. Loop output 0 ma setpoint ENTER "SAVING" briefly displays C. Sets 0 ma setpoint: ph. 0 ma and 4 ma setpoints are reversible D. Adjusts 4 ma output: 3.0 to 5.0 ma (overrides factory 4.00 ma calibration) E. Loop output 4 ma adjust E. Adjusts 0 ma output: 19 to 1 ma (overrides factory 0.00 ma calibration) F. Loop output 0 ma adjust F. Selects temperature display: C or F. Recalibration is not required when switching from Celsius to Fahrenheit. To return to VIEW: Temp. display A quick press Display, C or F To restore original value: quick press 7. Parts and Accessories Splashproof rear cover # (code ) 5 x 5 inch adapter plate for Signet retrofi t # (code ) Optional surface mount bracket # (code ) Power supply, 10 VAC 4 VAC, # ph buffer kit, ph 4.0, 7.0, 10.0, # (code ) Front snapon bezel, # (code ) 5700 ph/orp instruction sheet, # (code ) 6 Signet 5700 ph/orp Monitor ph Instructions
7 8. Specifications General Compatible electrodes: Accuracy: ±0.% of full scale Input range: ph: Signet 370 ph/orp Preamplifi er/electrode System 0 to 14, optically isolated (standard 0 14 ph dial included) Temperature input: 3 kω Balco, 5 to 10 C (13 to 48 F), optically isolated Enclosure: Rating: NEMA 4X/IP65 front Dimensions: 1/4 DIN, 96 x 96 x 88 mm (3.8 x 3.8 x 3.5 in.) Case: ABS plastic Keypad: Sealed 4key silicone rubber Weight: Approximately 500 g (18 oz.) Electrical Power requirements: 1 to 4 VDC or 1 to 4 VAC, unregulated, 5060 Hz, 10 W max. Relay contacts ( sets): Mechanical SPDT contacts Max. voltage rating: 5 30 VDC, 5 15 VAC, or 3 50 VAC, (power factor = 1.0) Hysteresis: User adjustable Current output: 4 to 0 ma, nonisolated, internally powered, fully adjustable and reversible Update rate: <1s Max loop impedance: 350 Ω with a 1 V instrument supply voltage, 950 Ω with a 4 V instrument supply voltage Accuracy: ±0.1% of max. range Display: Type: Microprocessor controlled aircore meter movement and backlit Alphanumeric x 16 LCD Update rate: <1s Contrast: User selected Relay annunciators: LEDs Noise immunity: Noise emissions: Safety: EN5008 EN55011 EN Environmental Operating temp.: 10 to 55 C (14 to 131 F), 50 C (1 F) max. with optional rear cover Storage temp.: 15 to 80 C (5 to 176 F) Relative humidity: 0 to 95%, noncondensing Altitude: 4000 m max. Pollution degree: Agency Approvals CE, UL listed Manufactured under ISO 9001 Dimensions 96 mm (3.8 in.) Panel Gasket Optional Splashproof Rear Cover # ph 10 RELAY 1 RELAY 1 Mounting Panel Mounting Clamp ( ea.) mm (3.8 in.) 91 mm (3.6 in.) 88 mm (3.5 in.) Side View ENTER Signet Monitor Front View 76 mm (3 in.) 96 mm (3.8 in.) 88 mm (3.5 in.) (Continued) Signet 5700 ph/orp Monitor ph Instructions 7
8 Dimensions 88 mm (3.5 in.) LISTED UL 77CJ Signet Model # V 10 W mv Input Brown wire Iso. Gnd Blue wire T Green wire T White wire V Red Wire Relay V Black wire NO C NC NO C NC 40 ma Relay 1 Patent No. D 376,38 ph/orp 88 mm (3.5 in.) 9 x 9 mm (3.6 x 3.6 in.) Rear View Panel cutout 9. Quick Reference Menu Parameters 9.1 VIEW Menu Setup Parameters (sec. 6.1) Menu Parameters Display Description Range Factory Default A. Process ph Process temperature Process temperature B. Electrode mv input mv C. Current loop output 3 1 ma D. Min Max meter and dial range E. Relay 1 mode Relay 1 setpoint Low, High, or Pulse, Low 4.00 ph F. Relay mode Relay setpoint Low, High, or Pulse, High ph G. Last calibration date H. EASY CAL procedure Two buffers required: ph 4, 7, or 10 (sec. 6.) 8 Signet 5700 ph/orp Monitor ph Instructions
9 9. CALIBRATE Menu Setup Parameters (sec. 6.3) Menu Parameters Display Description Range Factory Default A. Min max meter dial range B. Process temperature Solution temperature 30 C (86 F) C. Manual electrode standard calibration (must be ph units from Slope D. E. F. Manual electrode slope calibration Relay 1 mode Relay 1 setpoint Relay 1 hysteresis (must be ph units from Standard) Low or High Low 4.00 ph 1.00 ph Relay mode and setpoint displays repeat for relay setup G. Relay 1 mode Relay 1 range Pulse ph H. Relay 1 pulse rate pulses/minute 10 pulses/minute I. Last calibration date OPTIONS Menu Setup Parameters (sec. 6.4) Menu Parameters Display Description Range Factory Default A. Display contrast 0 to 5 3 B. 4 ma setpoint ph 0.00 ph C. 0 ma setpoint ph D. 4 ma adjust 3.0 to 5.0 ma 4.00 ma E. 0 ma adjust 19 to 1 ma 0.00 ma F. Temperature display Celsius or Fahrenheit C Signet 5700 ph/orp Monitor ph Instructions 9
10 10. Troubleshooting Display Problem Solution 1. ORP: 1000 mv Input: 998 mv (example ) ORP electrode installed in preamplifi er Remove ORP electrode and replace with recommended ph electrode (sec. ) A. Out Of Range CHECK SENSOR B. CHECK SENSOR? C. Temperature Bad CHECK SENSOR Electrode not installed in preamplifi er Temperature input miscalibrated or out of range during power up Temperature input missing or out of range during power up Temperature input missing or out of range during EASY CAL procedure (sec. 6.) 1) Verify temperature input connections and electrode installation (sec. ) ) Disconnect Green and White preamplifi er wires from rear T and T terminals. Measure resistance across disconnected wires for.9 kω 3.1 kω at 5 C (77 F). If measured resistance is out of range: Wrong electrode installed in preamplifi er (ORP electrode=10 kω). Replace ORP electrode with ph electrode. Faulty preamplifi er contacts, see preamplifi er manual for additional information. Faulty electrode, see electrode manual for additional information. 3) To verify instrument input: Disconnect Green and White preamplifi er wires from rear T and T terminals, then install a 3 kω fi xed resistor across the terminals. Power instrument and verify 5 C (77 F) on display. If 5 C (77 F) is not displayed, calibrate temperature input for 5 C (sec. 6.3B). If instrument displays correctly after calibration, instrument is ok. If error condition persists, instrument requires factory service. Refer to solution steps 1 3 above. Refer to solution steps 1 3 above 3. Out Of Range Use Manual Cal ph buffers other than ph 4, 7, 10 used during EASY CAL procedure or electrode has excess offset 4.! Same Buffer Same ph buffer used for EASY CAL buffer 1 and ph 5.0 C Excessive mv input from preamplifi er Use ph 4, 7, 10 buffers for EASY CAL procedure (sec. 6.). Use manual calibration to set standard and slope (sec. 6.3C, 6.3D) Clean electrode (see electrode manual) then retry EASY CAL (sec. 6.) Use different ph 4, 7, 10 buffer for EASY CAL procedure solution #1 and solution # (sec. 6.3). Faulty preamplifi er, refer to preamplifi er manual for additional information To verify instrument input: Disconnect Brown and Blue preamplifi er wires and Install shorting strap (jumper) across mv input and Iso Gnd terminals. Power instrument and verify 7.00 ph on display. If 7.00 ph is not displayed, calibrate Set Standard input for 7.00 ph (sec. 6.3C) If instrument displays correctly after calibration, instrument is ok. If error condition persists, instrument requires factory service. continued 10 Signet 5700 ph/orp Monitor ph Instructions
11 Display Problem Solution 6.! ph Must Be Or Less ph input out of range or improperly entered during Standard or Slope calibration A) Use two solutions of known ph within range for manual Standard and Slope calibration (sec. 6.3C, 6.3D). B) Enter ph value only within ph during standard or slope calibration (sec. 6.3C 6.3D) 7. Slope Too Close To Standard 8. Value Must Be Or Less ph slope calibration too close to ph standard calibration Relay, 4 ma, or 0 ma setpoint greater than ph (CALIBRATE Menu sec. 6.3D only) Calibrate ph slope using a ph solution ph units from the ph standard solution (sec. 6.3C, 6.3D). Enter setpoint within range (sec. 6.3E) 9. Value Must Be 300 Or Less 10. SETUP READ ERROR Press any Key Relay pulse setting greater than 300 pulses per minute Power fault occurred while saving setup menu entry Enter relay pulse setting within pulses per minute (sec. 6.3H) Press any key to reload factory defaults then reprogram system setup parameters. 11. Maintenance Clean the instrument case and front panel with a soft cloth and a mild liquid soap solution. Notes: Signet 5700 ph/orp Monitor ph Instructions 11
12 George Fischer Signet, Inc., 3401 Aerojet Avenue, El Monte, CA U.S.A. Tel. (66) Fax (66) For Worldwide Sales and Service, visit our website: Or call (in the U.S.): (800) Rev. K 3/06 English/pH George Fischer Signet, Inc Printed in U.S.A. on recycled paper
Signet 5700 ph/orp Monitor - ORP Instructions
Signet 5700 ph/ Monitor Instructions English *5700.0901* 5700.0901 Rev. K /06 English/ Contents 1. Power Connections. Compatible Electrode/Preamp Wiring. 4 0 ma Current Output Connections 4. Relay Connections
More informationSIGNET 5700 ph/orp Monitor - ORP Instructions
SIGNET 5700 ph/orp Monitor ORP Instructions ENGLISH 35700.0901 H (6/01) English CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections. Follow
More informationSignet 5900 Salinity Monitor
Signet 5900 Salinity Monitor *35900.0901* 35900.0901 Rev. E 3/06 English Contents 1. Power Connections 2. Compatible Sensor Wiring 3. 4 20 ma Current Output Connections 4. Relay Connections 5. Relay Operation
More information+GF+ SIGNET 5800CR Conductivity/Resistivity Monitor Instructions
GF SIGNET 5800CR Conductivity/Resistivity Monitor Instructions ENGLISH 35800CR.0901 D1/98 CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections.
More information+GF+ SIGNET ph/orp Transmitter Instructions
GF SIGNET 8750 ph/orp Transmitter Instructions ENGLISH 8750.090 A9/99 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury.
More informationSignet 5800CR Conductivity/Resistivity Monitor
Signet 5800CR Conductivity/Resistivity Monitor *35800CR.0901* 35800CR.0901 Rev. H 3/06 English Contents 1. Power Connections 2. Compatible Sensor Wiring 3. 4 20 ma Current Output Connections 4. Relay Connections
More informationContents. 1. Installation
ENGLISH 8750.090 C/0 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents. Installation. Specifications. Electrical
More informationSignet Pressure Transmitter
Signet 850 Pressure English *850.090* 850.090 Rev. H /06 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.
More information+GF+ SIGNET 5900 Salinity Monitor Instructions
+GF+ SIGNET 5900 Salinity Monitor Instructions ENGLISH 35900.0901 (B8/99), English CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections.
More informationSIGNET ph/orp Transmitter
SIGNET 8750- ph/orp Transmitter ENGLISH 3-8750.090- D-4/03 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.
More informationSignet Temperature Transmitter
Loop Output Output 7 Sensr IN (RED) Sensr V (BLACK) Signet 80 Temperature 80.090 English 80.090 Rev. J 0/ English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions
More informationSignet ph/orp Transmitter
Signet 87501 ph/orp *38750.0901* 38750.0901 Rev. H 06/10 English English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury.
More informationSignet Flow Transmitter
Signet 8550 Flow *8550.090* English 8550.090 Rev. L 0/ English WARNING! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.
More informationSignet ph/orp Transmitter * *
Loop 7 6 5 Signet 8750- ph/orp *-8750.090-* CAUTION! Contents 0.0 ph 5.0C Signet ph/orp. Installation. Panel Installation gasket panel terminals mounting bracket 96 mm (.8 in.) 8050 8 mm (. in.) Optional
More information+GF+ SIGNET Conductivity/Resistivity Transmitter Instructions
GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions ENGLISH 8850.090 A9/99 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to
More informationUser s Guide PHCN ph/orp indicator ph Instructions. Shop online at. ENTER RELAY 1 RELAY 2
User s Guide 4 6 7 8 10 RELAY 1 RELAY 2 2 12 0 14 ENTER Shop online at www.omega.com email: info@omega.com PHCN5700 /ORP indicator Instructions OMEGAnet Online Service www.omega.com Internet email info@omega.com
More informationTable of Contents. 1 Introduction Introduction Front Panel Description Rear Panel Description 3
Chapter Page 1 Introduction 1 1.1 Introduction 1 1.2 Front Panel Description 2 1.3 Rear Panel Description 3 Table of Contents 2 Installation and Operation 4 2.1 Mounting Instructions 4 2.2 Power Connections
More informationFLS M9.06 PH/ORP MONITOR. safety instructions. packing List
FLS M9.06 PH/ORP MONITOR safety instructions General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other instruments
More informationSIGNET Level Transmitter
ENGLISH 850.090 A8/0 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents. Description. Specifications. Panel
More informationFLS M9.06. ph/orp MONITOR
FLS M9.06 /ORP MONITOR Safety Instructions General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other instruments
More informationSignet 9900 Transmitter
Signet 9900 Transmitter Member of the SmartPro Family of Instruments Features Panel Mount Field Mount NEW! Modbus Module supports RS485 Serial Modbus Communications Multiple sensor types supported with
More informationFLOW MONITOR & TRANSMITTER
FLS M9.02 FLOW MONITOR & TRANSMITTER SAFETY INSTRUCTIONS General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other
More informationSignet 2750 DryLoc ph/orp Sensor Electronics
Signet 750 DryLoc ph/orp Sensor Electronics *750.090* 750.090 Rev. G /05 English WARNING! SAFETY INSTRUCTIONS. Depressurize and vent system prior to installation or removal.. Confi rm chemical compatibility
More information+GF+ SIGNET 9050CR Conductivity/Resistivity Controller
+GF+ SIGNET 9050CR Conductivity/Resistivity Controller Instruction Manual GEORGE FISCHER +GF+ Important Safety Information! CAUTION: Remove AC power to unit prior to wiring input and output connections.
More informationX Year mfg
4to20 ma Output S L Loop PWR Loop2 PWR N/C S L DATA 7 8 9 6,*(7 S L Output 8 S L SIGNET 8900 MultiParameter Controller Supplemental Manual *8900.090SUP* 8900.090SUP Rev. A 2/04 (English). Description This
More informationAutoranging True RMS Multimeter User Manual
Autoranging True RMS Multimeter User Manual Please read this manual before switching the unit on. Important safety information inside. Contents Page 1. Safety Information... 4 2. Safety Symbols... 5 3.
More informationOPERATION MANUAL MODEL TT-LP
TURTLE TOUGH OPERATION MANUAL MODEL TT-LP 2-Wire Isolated ph/orp Transmitter turtletough.com.au CONTENT INITIAL INSPECTION 3 INTRODUCTION 3 ASSEMBLY 4 PREPARATION 5 Connection Diagram of Turtle Tough Sensors
More information4801P ph/orp controller
4801P ph/orp controller Introduction: Thank you for choosing this ph/orp controller. This 1/16 DIN controller offers dual backlight LCD display, high accuracy, LED indicators, two control relays, and automatic
More informationAnalyzer/Controller. ph/orp HART. Model 54e ph/orp ESSENTIAL INSTRUCTIONS WARNINGS RISK OF ELECTRICAL SHOCK
Instruction Sheet PN 51A-54epH/rev J October 2010 ph/orp HART Model 54e ph/orp Analyzer/Controller For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. ESSENTIAL
More informationMini Digital Multimeter
User's Guide Mini Digital Multimeter Model MN15 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Back to the Extech MN15/MN16 Series
More informationMini Digital Multimeter
User Manual Mini Digital Multimeter Model MN15A Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the Extech MN15A MultiMeter. The MN15A offers
More informationOperation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter
Operation Manual MODEL 2TX 2-wire Isolated ph/orp Transmitter 0 2TX CONTENTS INITIAL INSPECTION.....2 INTRODUCTION......2 ASSEMBLY...3 PREPARATION....4 CONNECTING THE ELECTRODE...4 CONNECTING THE TEMPERATURE
More informationJUMO ecotrans ph 03 Microprocessor transmitter/ switching device for ph/redox voltage and temperature
Page 1/7 JUMO ecotrans 03 Microprocessor transmitter/ switching device for /Redox voltage and temperature with a 2-line LCD for mounting on a 35 mm DIN rail Brief description Depending on the configuration,
More informationHigh Voltage DC Meter
High Voltage DC Meter Javelin D PD644 0-300 VDC input NEMA 4X, IP65 front Scale in engineering units Sunlight readable LED display 4-20 ma analog output Two form C 3 A relays option RS-485 serial communications
More informationHIGH-VOLTAGE DC METER
HIGH-VOLTAGE DC METER Javelin D PD644 0-300 VDC input NEMA 4X, IP65 front Scale in engineering units Sunlight readable LED display 4-20 ma analog output Two form C 3 A relays option RS-485 serial communications
More informationUDC 1000 and UDC 1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS
UDC 1000 and UDC 1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS EN0I-6041 12/99 PRODUCT SPECIFICATION SHEET OVERVIEW The UDC 1000 and UDC 1500 are microprocessor-based 1/16 DIN and 1/8 DIN controllers
More information1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882
1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882 This manual is intended for use in support of installation, commissioning and configuration of the 1/32-DIN Temperature
More informationOPERATING INSTRUCTION
OPERATING INSTRUCTION AUTORANGING MULTIMETER MAX Ω F C 10A MAX every 15 min. COM V SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation
More information5450 NW 33rd Ave, Suite 104 Fort Lauderdale, FL Fruitland Ave Los Angeles, CA SST7000 SST7100. Speed Switch / Transmitter
5450 NW 33rd Ave, Suite 104 Fort Lauderdale, FL 33309 3211 Fruitland Ave Los Angeles, CA 90058 SST7000 SST7100 Speed Switch / Transmitter Installation and Operation Manual Rev. C P/N145F-13112 PCO 00009270
More informationMYRIAD QLC 4-CHANNEL MONITOR/CONTROLLER INSTRUCTION MANUAL
MYRIAD QLC 4-CHANNEL MONITOR/CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM MYR QLC MANUAL 013114 2 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features WIRING
More informationLAUREL ELECTRONICS, INC.
LAUREL ELECTRONICS, INC. Laureate RTD Temperature Panel Meter / Controller Features Factory calibrated for 100Ω platinum, 10Ω copper & 120Ω nickel RTDs 2, 3 or 4-wire connection with lead resistance compensation
More informationMODEL: M5DY SEN TRONIC AG CURRENT LOOP SUPPLY. [2] POWER INPUT AC Power M: V AC (Operational voltage range V, Hz)
Super-mini Terminal Block Signal Conditioners M5-UNIT CURRENT LOOP SUPPLY Functions & Features Powering a 0 ma DC current loop Applicable to smart transmitters Isolation between the input and output Fast
More informationDM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER
DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER SAFETY INFORMATION The following safety information must be observed to ensure maximum personal safety during the operation of this meter: This meter is
More informationMini Digital Multimeter
User's Guide Mini Digital Multimeter Model MN15 Introduction Congratulations on your purchase of the Extech MN15 MultiMeter. The MN15 offers AC/DC Voltage, AC/DC Current, Resistance, Diode, and Continuity
More informationOwner's Manual. True RMS Multimeter. Model No Safety Operation Maintenance Español
Owner's Manual True RMS Multimeter Model No. 82023 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance Español
More informationMYRIAD LC1 LEVEL CONTROLLER INSTRUCTION MANUAL
MYRIAD LC1 LEVEL CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM MYR LC1 MANUAL 062114 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features Dimensions WIRING
More informationTABLE I RANGE, POWER, OUTPUT (CONDUCTIVITY MEASUREMENTS)
Issue 11 Page 1 of 5 Model Selection Guide Instructions Consult Steps to Selecting Appropriate Conductivity Instrumentation and Cells before making selections below. Select the desired key number. The
More informationS-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter
S-14 Compact Digital Multimeter S-14 Compact Digital Multimeter SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation at this meter
More informationTempco Instruction Manual
Tempco Instruction Manual 1/16 DIN Solid State Temperature Controller Relay Output Solid State Output For Heating Model Numbers: TEC-901, TEC-902, TEC-905 Temperature controls in this series are designed
More informationDIGITAL PANEL THERMOMETER
Love Controls DIGITAL PANEL THERMOMETER LCI108-2x LCI108J-2x INSTRUCTIONS MANUAL EDITION: February 2001 INTRODUCTION TO THE LOVE CONTROLS LCI108 SERIES This manual does not constitute a formal agreement.
More information700 SERIES DPC DUPLEX PUMP CONTROLLER INSTRUCTION MANUAL
1 700 SERIES DPC DUPLEX PUMP CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM 700 DPC MANUAL 072114 2 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features WIRING
More informationConductivity Transmitters
Conductivity Transmitters CDTX-90 Series Starts at $ 576 Electronics Only Display in µs, ms, kω, MΩ, ppm (TDS) Simulate Function Programmable Temperature Compensation Relay Options Dual Output Option Allows
More informationTABLE OF CONTENTS INTRODUCTION. 3. Analog Input Analog Output Digital Input Digital Output OPERATIONAL DESCRIPITON.. 7 PROGRAMMING AND INITIAL SETUP.
DIVERSIFIED HEAT TRANSFER SERIES 700 STEAM GENERATOR CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE AT SIGMACONTROLS.COM SERIES 700 DHT STEAM GENERATOR MANUAL 042514 2 TABLE OF CONTENTS INTRODUCTION.
More informationLAUREL. Laureate RTD Temperature Panel Meter / Controller ELECTRONICS, INC. Features. Description. Specifications
LAUREL ELECTRONICS, INC. Laureate RTD Temperature Panel Meter / Controller Features Factory calibrated for 100Ω platinum, 10Ω copper & 120Ω nickel RTDs 2, 3 or 4-wire connection with lead resistance compensation
More informationUser's Guide. Extech AM A AC Analog Clamp Meter
User's Guide Extech AM300 300A AC Analog Clamp Meter Introduction Congratulations on your purchase of the Extech AM300 Analog Clamp Meter. This device measure AC Voltage and Current, DC Voltage, and Resistance.
More informationABB MEASUREMENT & ANALYTICS DATA SHEET. TB84PH ph /ORP / pion transmitter
ABB MEASUREMENT & ANALYTICS DATA SHEET TB84PH ph /ORP / pion transmitter 2 TB 8 4PH P H /ORP / P I ON TR AN S MIT TE R DS/ TB8 4PH-E N RE V. D Measurement made easy Hazardous area rated transmitter with
More informationRXTH DUAL ROOM SENSOR / SWITCH
DUAL ROOM SENSOR / SWITCH FOR TEMPERATURE AND RELATIVE HUMIDITY Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE
More informationCole-Parmer Instrument Company
Operating Instructions Cole-Parmer ph/orp 400 ph/orp Controller Cole-Parmer Instrument Company 625 East Bunker Court, Vernon Hills, Illinois 60061 Tel: 1-847-549-7600 or Toll-free: 1-800-323-4340 Fax:
More informationUser Guide True RMS Multimeter Extech EX205T
User Guide Extech EX205T True RMS Digital Multimeter Extech EX210T True RMS Digital Multimeter IR True RMS Multimeter Extech EX205T Introduction Thank you for selecting the Extech EX205T True RMS Auto-ranging
More informationSignet BC Batch Controller System
Signet 9900-1BC Batch Controller System *3-9900-1BC.090* 3-9900-1BC.090-1 Rev. A1 03/13 English Product Manual Look for the Quick Start icon to quickly set up your new 9900-1BC. Safety WARNING! Panel Mount
More informationUser's Guide. MiniTec TM Series Model MN25 MultiMeter
User's Guide MiniTec TM Series Model MN25 MultiMeter Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a
More informationOPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS
OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS GB Contents Page 1 Introduction 4 2 Safety 5 3 Use 6 4 Checking on Delivered Equipment 6 5 General Specification and Dimensions 7 6 Positioning 10 7 Operating
More informationTC Type Range Conformity Error. J -210 C to +760 C (-347 F to F) ±0.09 C (±0.16 F) K -244 C to C (-408 F to F) ±0.1 C (±0.
LAUREL ELECTRONICS, INC. Laureate Thermocouple Panel Meter / Controller Features Factory calibrated for thermocouple types J, K, T, E, N, R, S Entire range of each thermocouple in one scale Highly accurate
More informationAdapter Kit - QSI Electronics to TM Meter Installation Instructions
Adapter Kit - QSI Electronics to TM Meter Installation Instructions These installation instructions cover the adapter kit installation of the QSI1, QSI2 and QSI3 versions of communications electronics
More informationUser's Guide. 1/16 th DIN ph Controller. Model 48PH2. Introduction
User's Guide 1/16 th DIN ph Controller Model 48PH2 Introduction Thank you for selecting the Extech Model 48PH2 controller. This 1/16 th DIN-size controller offers dual display, high resolution measurements,
More informationINSTALLATION DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER. Before installation:
DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER The DKM-409 is a precision instrument designed for displaying various AC parameters in 3-phase distribution panels. Thanks to its isolated
More informationOperating Manual. Dynisco /8 DIN Indicator Concise Product Manual
Dynisco 1490 1/8 DIN Indicator Concise Product Manual 59476-9 Operating Manual -1- CAUTION: Installation should be only performed by technically competent personnel. Local Regulations regarding electrical
More informationph/orp 300 (Cat. No ) MICROCOMPUTER BASED OPERATION MANUAL ph /ORP/ Temperature CONTROLLER Phone: , Fax:
OPERATION MANUAL /ORP 300 (Cat. No. 0580212) MICROCOMPUTER BASED /ORP/ Temperature CONTROLLER 625 East Bunker CT, Vernon Hills, IL 60061-1844 Phone: 800-32-4340, Fax: 847-247-2929 E-mail: info@coleparmer.com,
More informationOperating instructions. Speed monitor D / / 2014
Operating instructions Speed monitor D200 80005257 / 00 05 / 2014 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5 2.2 Target group...5
More information12-36 VDC/12-24 VAC Power Option 4-Digit Display, 0.56 (14.2 mm) or 1.20 (30.5 mm)
4-20 ma & Relay Output Features 4-20 ma, ± 10 V, TC & RTD Inputs 12-36 VDC/12-24 VAC Power Option 4-Digit Display, 0.56 (14.2 mm) or 1.20 (30.5 mm) 24 VDC @ 200 ma Transmitter Power Supply Options Type
More information1. Installation. 2. Configuration - Operation. 3. Specifications. Safety
Safety 1. Installation 1.1 OP_ext: Services 1.2 Driver Card Installation 1.3 Sensor Types 1.4 Sensor Wiring 1.5 Controller Wiring 2. Configuration - Operation 2.1 Replaces ph Sensor 2.2 AS -Flex Series
More informationUser Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at
User Manual 400Amp AC Clamp Meter + NCV Model MA430 Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of this Extech MA430 Clamp Meter. This
More informationDigital Proportional Valve Amplifier
Digital Proportional Valve Amplifier The Amplifier is a compact, low profile DIN mounted controller for use with proportional solenoid valves. The provides current to a valve coil in proportion to an input
More informationUser's Guide. Digital Multimeter. Model MN42
User's Guide Digital Multimeter Model MN42 Introduction Congratulations on your purchase of the Extech MN42 MultiMeter. The MN42 offers AC/DC Voltage, DC Current, and Resistance testing. Proper use and
More informationR5050. Model. Instruction Manual. TRMS AC/DC Clamp Meter. reedinstruments. www. com
Model R5050 TRMS AC/DC Clamp Meter Instruction Manual reedinstruments com Table of Contents Safety... 3 Features... 4 Specifications...4-6 Instrument Description...7-8 Measurement Procedures...9-12 Battery
More informationINDICATOR FOR PROCESS CONTROL. Love Controls. LCI108J-0x INSTRUCTION MANUAL. EDITION: January Valid for models with software version P-1
Love Controls INDICATOR FOR PROCESS CONTROL LCI108-0x LCI108J-0x INSTRUCTION MANUAL EDITION: January 2001 Valid for models with software version P-1 INTRODUCTION TO THE LOVE CONTROLS LCI108 SERIES This
More informationOperation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller
Operation Manual TX10 1/8 DIN Microcomputer Based /ORP Controller 11751 Markon Drive Garden Grove, CA. 92841 U.S.A. pg. 1 CONTENTS GENERAL INTRODUCTION.....3 INITIAL INSPECTION......3 USING THE Sensorex
More informationSHARK MULTI-PARAMETER CONTROLLER & ANALYZER USER S MANUAL
SHARK MULTI-PARAMETER CONTROLLER & ANALYZER USER S MANUAL Rev 3 AquaMetrix Inc. 1245 Maple Hill Ct., Unit 7 Newmarket, ON Canada, L3Y 9E8 Tel: (800) 742-1413 (905) 954-0841 Fax: (905) 954-0415 www.aquametrix.com
More informationAnalog outputs Output 1 + 2: 0(4) 20 ma or 0 10 V. Configurable as - analog process value output - continuous controller output (PID action)
Page 1/10 JUMO AQUIS 500 ph Transmitter/Controller for ph, ORP, NH 3 (ammonia) concentration and temperature Brief description The instrument is used for measuring/controlling the ph, ORP or NH 3 (ammonia)
More informationUser's Guide. 800 Amp Clamp Meters. EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter
User's Guide 800 Amp Clamp Meters EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter Introduction Congratulations on your purchase of the EX710, EX720, or EX730 Clamp DMM.
More informationModel: Pro93 TRUE RMS LEAKAGE CURRENT TESTER
Model: Pro93 TRUE RMS LEAKAGE CURRENT TESTER CONTENTS TITLE PAGE I. Safety Information.......1 Environmental Conditions... 1 Explanation of Symbols..... 1 II. Specification..... 2 General Specification...
More informationTrident and Trident X2 Digital Process and Temperature Panel Meter
Sign In New User ISO 9001:2008 Certified Quality System Home Products Online Tools Videos Downloads About Us Store Contact Policies Trident and Trident X2 Digital Process and Temperature Panel Meter Products
More informationRТTH DUAL ROOM SWITCH
DUAL ROOM SWITCH FOR TEMPERATURE AND RELATIVE HUMIDITY Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS PRODUCT DESCRIPTION ARTICLE CODES INTENDED AREA OF USE TECHNICAL DATA
More informationPUMP DRIVE OPERA TING MANUAL: A Edition 01
OPERA TING MANUAL: PUMP DRIVE 77420-00 Pump Drive, shown with 77600-62 High Performance I/P Pump Head. 77420-00 A-1288-0990 Edition 01 1-800-637-3739 (U.S. and Canada only) 11 (847) 381-7050 (Outside U.S.)
More informationSignet Conductivity/Resistivity Electrodes
Signet 2819-2823 Conductivity/Resistivity Electrodes 2819 2820 2821 2822 2823 Sanitary Features Standard process connections - ¾ in. NPT Polypro - Tri-clamp 1-1½ in., 2 - Opt. ½ in. NPT 316 SS 316 SS or
More informationNKQB24-1, NKQX24-1 On/Off, Electronic Fail-Safe, Direct Coupled, 24V
NKQB24-1, NKQX24-1 On/Off, Electronic Fail-Safe, Direct Coupled, 24V orque min. 54 in-lb for control damper surfaces up to 12 sq ft. Application For on/off, fail-safe control of dampers in HVAC systems.
More informationEA401 SERIES PULSE INPUT FLOW RATEMETER/TOTALIZER INSTALLATION & OPERATING INSTRUCTIONS West Stetson Avenue Hemet, CA USA
EA401 SERIES PULSE INPUT FLOW RATEMETER/TOTALIZER INSTALLATION & OPERATING INSTRUCTIONS 3255 West Stetson Avenue Hemet, CA 92545-7799 USA Lit# 24508-67 99773 Rev. 1.1 02/25/0 Tel: 951-652-6811 Fax: 951-652-3078
More informationColor Mark Photoelectric Sensor E3S-VS
Sensing Supply voltage Output 12 to 80 ma, with 1.5 to 4 ma constant current source; 1.2 cm, 5 cm 100 ma, Color Mark Photoelectric Sensor Small Color Mark Sensor With Built-In DC Amplifier Fast, 1 ms maximum
More informationAutoRanging Digital MultiMeter
Owner's Manual AutoRanging Digital MultiMeter Model No. 82175 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance
More information700 SERIES TPC TRIPLEX PUMP CONTROLLER INSTRUCTION MANUAL
1 700 SERIES TPC TRIPLEX PUMP CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM 700SERIESTPCMANUAL071414 2 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features
More informationProSense DPM2 Series. Digital Panel Meter. User Manual. for process, temperature and resistance input signals. Models: DPM2-AT-HL DPM2-AT-2R-HL
User Manual ProSense DPM2 Series Digital Panel Meter for process, temperature and resistance input signals Models: DPM2-AT-HL DPM2-AT-2R-HL Scan or click the QR code for a series of Configuration and Programming
More informationI-7520: RS-232 to RS-485. Input: RS-232 protocol Output:RS-485/RS-422 Speed: Self Tuner 'inside, auto switching baud rate, 300~ BPS
I-7520 RS-232 to RS-485 Converter I-7510 RS-485 Repeater RS-232 RS-485 I-7000 SERIES Common Specifications Isolation voltage: 3000VDC Speed: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115000 Dual watchdog
More informationM A C 3 Wind Speed Alarm & Controller
M A C 3 Wind Speed Alarm & Controller Installation Instructions Thank you for purchasing the MAC3 wind speed alarm and controller. This manual is designed to lead you through a step-by-step process to
More informationLCD LCDS LCDRI CONTROLLER OPERATING MANUAL
This manual contains important safety information about installation and use of this equipment. Ignoring this information could result in injuries or damages. It is strictly forbidden to use this equipment
More informationDIGITAL PANEL THERMOMETER
Love Controls DIGITAL PANEL THERMOMETER LCI108-1x LCI108J-1x INSTRUCTION MANUAL EDITION: January 2001 INTRODUCTION TO THE LOVE CONTROLS LCI108 SERIES This manual does not constitute a formal agreement.
More informationINSTALLATION MANUAL. LC 200 Electronic Overload Guard. Software versione PW0501 R 0.3
INSTALLATION MANUAL LC 200 Electronic Overload Guard Software versione PW0501 R 0.3 CONTENTS MAIN FEATURES LC 200 TECHNICAL FEATURES Page 2 SYMBOLS Page 3 WARNINGS Page 3 IDENTIFICATION DATA PLATE Page
More informationRTT2 ROOM TEMPERATURE SWITCH. Mounting and operating instructions
Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS PRODUCT DESCRIPTION ARTICLE CODES INTENDED AREA OF USE TECHNICAL DATA STANDARDS OPERATIONAL DIAGRAMS WIRING AND CONNECTIONS
More informationTable of Contents. General Information. Document Sure-Aire Flow Monitoring System. User and Service Manual WARNING CAUTION
Document 476092 User and Service Manual Installation, Operation and Maintenance Manual Please read and save these instructions for future reference. Read carefully before attempting to assemble, install,
More informationInstruction Manual. M Pump Motor Controller. For file reference, please record the following data:
Instruction Manual M Pump Motor Controller For file reference, please record the following data: Model No: Serial No: Installation Date: Installation Location: When ordering replacement parts for your
More informationNKQB24-1, NKQX24-1 On/Off, Fail-Safe, Operation, Direct Coupled, 24V
NKQB24-1, NKQX24-1 On/Off, Fail-Safe, Operation, Direct Coupled, 24V orque min. 54 in-lb for control damper surfaces up to 12 sq ft. Application For proportional modulation of dampers in HVAC systems.
More information