Signet 5700 ph/orp Monitor - ph Instructions

Size: px
Start display at page:

Download "Signet 5700 ph/orp Monitor - ph Instructions"

Transcription

1 Signet 5700 ph/orp Monitor ph Instructions English * * Rev. K 3/06 English/pH CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents 1. Power Connections. Compatible Electrode/Preamp Wiring ma Current Output Connections 4. Relay Connections 5. Relay Operation 6. Menu Functions 7. Parts and Accessories 8. Specifi cations 9. Quick Reference Menu Parameters 10. Troubleshooting 11. Maintenance Remove terminal blocks for easy wiring Freq. IN Freq. IN Sen. Pwr. Iso. Gnd Std. Sensor Open Collector Sensor LISTED UL 77CJ 14 V 10 W Gnd Gnd PLS AUX output output 1 Total reset Patent No. D376,38 Flow 1 Freq. IN Freq. IN Sen. Pwr. Iso. Gnd Std. Sensor Open Collector Sensor LISTED UL 77CJ 14 V 10 W Gnd Gnd PLS AUX output output 1 Total reset Patent No. D376,38 Flow Power Connections. Compatible Electrode/Preamp Wiring CAUTION! Never connect 110 VAC or 0 VAC to rear power terminals. High voltage AC will damage instrument and void warranty. 14 V 10 W 5700 Terminals External power supply 1 4 VDC OR 1 4 VAC Brown Blue mv Input Brown wire Silver (earth ground) Iso. Gnd Blue wire T Green wire Green T White wire White Red Black V Red Wire V Black wire 40 ma ph/orp Preamp: Signet 70 TwistLock Electrodes 5700 Terminals DryLoc Electrodes Preamp: Signet 7601 Signet 760 = Double Insulated = DC or AC power Technical Notes: To reduce the possibility of noise interference, isolate AC power lines from signal lines. Maximum 40 ma loop impedance (sec. 3) is affected by the supply voltage. Use six conductor shielded cable for cable extensions up to 10 m (400 ft) max. (# ) Cable shield must be maintained through cable splice

2 3. 40 ma Current Output Connections Technical Notes: ** 1/8A fuse recommended (customer supplied) * 40 ma output is internally powered (nonisolated), maximum loop impedance 350 Ω with a 1 V instrument supply voltage, 950 Ω with a 4 V instrument supply voltage. To isolate output and prevent ground loop problems: 1. Use monitor device with isolated inputs, or. Use separate DC supply for 5700 and monitor device, or 3. Power 5700 with1 4 VAC step down transformer 4. Relay Connections Two internal relay contact sets (COM, NO, and NC) may be used for external device control. Front panel LED annunciators indicate the activation status of each relay. Each relay can control up to two devices simultaneously, as shown. Relay operation modes include Low alarm, High alarm, and Proportional Pulse (sec. 5). Common device connections include: Pulse mode metering pump control Pulse mode solenoid valve control Low or High mode warning lamps Low or High mode bells or sirens Low or High mode external heavyduty relay Wiring Example Right Device A IS powered when relay is deenergized (front panel LED "off"). Power is discontinued when the relay setpoint is reached (front panel LED "on"). Device B IS NOT powered when the relay is deenergized. Power is applied after the relay setpoint is reached. 5. Relay Operation 5700 Terminals Fuse** 1/8A 40 ma Relay ph/orp Technical Notes: Maximum relay contact ratings: 5 30 VDC, 5 15 VAC, or 3 50 VAC An external heavyduty relay must be used for devices with surge currents or operating currents that exceed the above specifi cations. * 5700 Terminals Relay 1 NO C NC NO C NC External AC/DC power 4 0 ma* Relay Deenergized NO C NC Device A Device B Relay Energized NO C NC A. LOW alarm mode The relay is energized when the ph drops below the setpoint, and is deenergized when the ph rises above the setpoint plus hysteresis (sec. 6.3E, 6.3F). Low setpoint= Adjustable hysteresis= Relay energized= Relay deenergized= ph B. HIGH alarm mode The relay is energized when the ph rises above the setpoint and is deengergized when the ph falls below the setpoint plus hysteresis (sec. 6.3E, 6.3F). High setpoint= Adjustable hysteresis= Relay energized= Relay deenergized= ph C. Proportional PULSE mode The proportional pulse relay configuration is primarily designed for metering pump control. The operator is prompted to enter a minimum and maximum ph setpoint and maximum pulse rate for the assigned relay (sec. 6.3G, 6.3H). Relay pulse width is fixed at 130 ms. Refer to the operation examples below. 300 pulses Metering pump chemical addition (dry contact activation type required) Example 1 (right): As the process ph rises above the minimum pulse setpoint (4.00 ph) the relay begins pulsing; triggering the metering pump for chemical addition. As the ph continues to rise, pulsing accelerates proportionally until the maximum programmed pulse rate of 300 pulses/minute and setpoint (10.00 ph) are reached, forcing the process ph back down to intended levels (e.g. 4.0 ph). ph 0 pulses 30 Relay 1 or Relay 0 ph ph Example (right): As the process ph falls below the minimum pulse setpoint (10.00 ph) the relay begins pulsing; triggering the metering pump for chemical addition. As the ph continues to decrease, pulsing accelerates proportionally until the maximum programmed pulse rate of 300 pulses/minute and setpoint (4.00 ph) are reached, forcing the process ph back up to intended levels (e.g ph). 300 pulses 3 Relay 1 or Relay 0 pulses ph 0 ph ph Signet 5700 ph/orp Monitor ph Instructions

3 ENTER 6. Menu Functions To access either CALIBRATE or OPTIONS menus, press and hold the ENTER key as illustrated below: VIEW (sec. 6.1) CALIBRATE (sec. 6.3) OPTIONS (sec. 6.4) Press & hold ENTER key for access: ENTER seconds 5 seconds Menus: VIEW menu (sec. 6.1): The VIEW menu is displayed during standard operation. The operator can navigate freely through the menu by pressing either UP or DOWN arrow keys. CALIBRATE Menu (sec. 6.3): The CALIBRATE menu contains all critical display setup and output parameters. A simple security code feature prevents unauthorized tampering. The operator is required to enter a simple access code for menu access. The same code also unlocks OPTIONS menus. OPTIONS Menu (sec. 6.4): The OPTIONS menu contains setup and display features that are seldom accessed for minor display or output adjustments. Reversible Dial Your 5700 includes a reversible dial face for ORP use. See enclosed ORP manual for operation details. A B ALARM 1 ALARM 1000 C ORP 1000 ph 0 14 ALARM 1 ALARM D 6.1 VIEW Menu H A. 1 Choose: ph/temperature Change: 3 Save: Menu Displays A H: (Factory default displays shown in menu column 1) A. Active ph and temperature display B. C. mv input Loop output B. Input display: shows electrode input mv C. Loop output display: shows the loop current output level. D. ph range display: shows the programmed min and max meter range (sec. 6.3A) D. Range E. Relay 1 display: this display shows the programmed operation mode and setpoint for relay 1 (sec. 6.3E) E. Relay 1 F. Relay display: this display shows the programmed operation mode and setpoint for relay (sec. (6.3E) F. Relay G. Last calibration: this display shows a user defi ned setup date for maintenance records. This feature is not an internal timer or calender. G. H. Last calibration Buffer calibration A Press and release to access EASY CAL Procedure See EASY CAL Procedure (Section 6.) H. EASY CAL: press the right arrow key to access the EASY CAL buffer calibration procedure (sec. 6.). Signet 5700 ph/orp Monitor ph Instructions 3

4 6. EASY CAL Procedure Requirements This procedure simplifi es system calibration using standard ph buffers. Use ph 4.0, 7.0, 10.0 ph buffers only (sec. 7). If standard ph buffers are not available, the system can be calibrated using the CALIBRATE menu temperature, standard, and slope settings (sec. 6.3BD). Access CALIBRATE menu (sec. 6.3B) and set sensor temperature before performing EASY CAL for new electrode installation. 1 To Calibrate: Press keys in sequence will appear during code entry. Response: to continue, 3 To Accept: Place electrode tip in first ph buffer; Allow for stabilization Press ENTER to accept first buffer calibration. ph 4.0, 7.0 or seconds* * For best results, gently stir the submerged electrode for approximately 5 seconds during the stabilization period (step ). Place electrode tip in second (different) ph buffer; Allow for stabilization Press ENTER to accept second buffer calibration. Large temperature differences from process fl uids to buffers may require longer stabilization time. ph 4.0, 7.0 or seconds* Display returns to VIEW Menu in 3s To exit EASY CAL at any time without saving changes: quick press Technical notes: The difference between the actual mv reading and value shown in the table below is a good indication of the ph electrode condition. Differences in excess of 50 mv indicate a need to service the electrode. Refer to electrode manual for maintenance recommendations. Theoretical mv values 5 C mv Signet 5700 ph/orp Monitor ph Instructions

5 6.3 CALIBRATE Menu Requirements System calibration is possible with two known ph solutions within 0 to 14 ph and an accurate reference thermometer. The EASY CAL procedure (sec. 6.) is recommended when standard ph buffers are available. If EASY CAL is performed, manual calibration steps B D below are not required. Always calibrate solution temperature fi rst (step B below) before setting ph standard and slope (steps C D below). Single point calibration can be done by setting "Set Standard" (step C) only. Dual point calibration (recommended) can be done by setting "Set Standard" followed by "Set Slope" (steps BC). Press keys in sequence will appear during code entry. to continue, I A. B. C. D. E. F. 1 Choose: Range Electrode temp. Electrode standard Electrode slope Relay Meter min max Allow for stabilization Allow for stabilization Allow for stabilization Mode Setpoint Change: Relay Low or High Mode Selected 3 Save: ENTER "SAVING" briefly displays Menu Settings A I : (Factory default displays shown in menu column 1) A. Sets Min Max meter dial range, (factory installed dial, ). Contact factory for custom dial confi gurations. Does not effect 4 to 0 ma output B. Sets process solution temperature C. Sets electrode standard to any value from 014 ph. Standard calibration solution must be ph units from slope solution (step D) D. Sets electrode slope to any value from 014 ph. Slope calibration solution must be ph units from standard solution (step C) Menu items E H repeat for relay setup. Relay Hysteresis Relay Pulse Mode Selected, step E E. Sets relay operation mode Low or High, and setpoint, ph (sec. 5AB) G. Relay Mode Range F. Sets relay hysteresis, ph. Set to zero to disable feature (sec. 5AB) H. Relay Pulse rate G. Sets relay minimum and maximum pulse setpoint, ph (sec. 5C). I. H. Sets relay pulse rate, pulses/minute. To return to VIEW: Last calibration A quick press Optional To restore original value: quick press I. Sets user defi ned setup date for maintenance records. This feature is not an internal timer or calender Quick Reference Calibration Procedures: Point Quick Calibration (recommended): 1. Verify/enter solution temperature ±0.5 C (step B). Set solution standard (step C) 3. Set solution slope (step D) 1Point Calibration (optional): 1. Verify/enter solution temperature ±0.5 C (step B). Set solution standard (step C) Signet 5700 ph/orp Monitor ph Instructions 5

6 6.4 OPTIONS Menu Press keys in sequence will appear during code entry. to continue, F A. 1 Choose: Change: 3 Save: Menu Settings A F: (Factory default displays shown in menu column 1) Contrast Low High A. Selects display contrast: 5 levels B. Loop output 4 ma setpoint B. Sets 4 ma setpoint: ph. 4 ma and 0 ma setpoints are reversible C. D. Loop output 0 ma setpoint ENTER "SAVING" briefly displays C. Sets 0 ma setpoint: ph. 0 ma and 4 ma setpoints are reversible D. Adjusts 4 ma output: 3.0 to 5.0 ma (overrides factory 4.00 ma calibration) E. Loop output 4 ma adjust E. Adjusts 0 ma output: 19 to 1 ma (overrides factory 0.00 ma calibration) F. Loop output 0 ma adjust F. Selects temperature display: C or F. Recalibration is not required when switching from Celsius to Fahrenheit. To return to VIEW: Temp. display A quick press Display, C or F To restore original value: quick press 7. Parts and Accessories Splashproof rear cover # (code ) 5 x 5 inch adapter plate for Signet retrofi t # (code ) Optional surface mount bracket # (code ) Power supply, 10 VAC 4 VAC, # ph buffer kit, ph 4.0, 7.0, 10.0, # (code ) Front snapon bezel, # (code ) 5700 ph/orp instruction sheet, # (code ) 6 Signet 5700 ph/orp Monitor ph Instructions

7 8. Specifications General Compatible electrodes: Accuracy: ±0.% of full scale Input range: ph: Signet 370 ph/orp Preamplifi er/electrode System 0 to 14, optically isolated (standard 0 14 ph dial included) Temperature input: 3 kω Balco, 5 to 10 C (13 to 48 F), optically isolated Enclosure: Rating: NEMA 4X/IP65 front Dimensions: 1/4 DIN, 96 x 96 x 88 mm (3.8 x 3.8 x 3.5 in.) Case: ABS plastic Keypad: Sealed 4key silicone rubber Weight: Approximately 500 g (18 oz.) Electrical Power requirements: 1 to 4 VDC or 1 to 4 VAC, unregulated, 5060 Hz, 10 W max. Relay contacts ( sets): Mechanical SPDT contacts Max. voltage rating: 5 30 VDC, 5 15 VAC, or 3 50 VAC, (power factor = 1.0) Hysteresis: User adjustable Current output: 4 to 0 ma, nonisolated, internally powered, fully adjustable and reversible Update rate: <1s Max loop impedance: 350 Ω with a 1 V instrument supply voltage, 950 Ω with a 4 V instrument supply voltage Accuracy: ±0.1% of max. range Display: Type: Microprocessor controlled aircore meter movement and backlit Alphanumeric x 16 LCD Update rate: <1s Contrast: User selected Relay annunciators: LEDs Noise immunity: Noise emissions: Safety: EN5008 EN55011 EN Environmental Operating temp.: 10 to 55 C (14 to 131 F), 50 C (1 F) max. with optional rear cover Storage temp.: 15 to 80 C (5 to 176 F) Relative humidity: 0 to 95%, noncondensing Altitude: 4000 m max. Pollution degree: Agency Approvals CE, UL listed Manufactured under ISO 9001 Dimensions 96 mm (3.8 in.) Panel Gasket Optional Splashproof Rear Cover # ph 10 RELAY 1 RELAY 1 Mounting Panel Mounting Clamp ( ea.) mm (3.8 in.) 91 mm (3.6 in.) 88 mm (3.5 in.) Side View ENTER Signet Monitor Front View 76 mm (3 in.) 96 mm (3.8 in.) 88 mm (3.5 in.) (Continued) Signet 5700 ph/orp Monitor ph Instructions 7

8 Dimensions 88 mm (3.5 in.) LISTED UL 77CJ Signet Model # V 10 W mv Input Brown wire Iso. Gnd Blue wire T Green wire T White wire V Red Wire Relay V Black wire NO C NC NO C NC 40 ma Relay 1 Patent No. D 376,38 ph/orp 88 mm (3.5 in.) 9 x 9 mm (3.6 x 3.6 in.) Rear View Panel cutout 9. Quick Reference Menu Parameters 9.1 VIEW Menu Setup Parameters (sec. 6.1) Menu Parameters Display Description Range Factory Default A. Process ph Process temperature Process temperature B. Electrode mv input mv C. Current loop output 3 1 ma D. Min Max meter and dial range E. Relay 1 mode Relay 1 setpoint Low, High, or Pulse, Low 4.00 ph F. Relay mode Relay setpoint Low, High, or Pulse, High ph G. Last calibration date H. EASY CAL procedure Two buffers required: ph 4, 7, or 10 (sec. 6.) 8 Signet 5700 ph/orp Monitor ph Instructions

9 9. CALIBRATE Menu Setup Parameters (sec. 6.3) Menu Parameters Display Description Range Factory Default A. Min max meter dial range B. Process temperature Solution temperature 30 C (86 F) C. Manual electrode standard calibration (must be ph units from Slope D. E. F. Manual electrode slope calibration Relay 1 mode Relay 1 setpoint Relay 1 hysteresis (must be ph units from Standard) Low or High Low 4.00 ph 1.00 ph Relay mode and setpoint displays repeat for relay setup G. Relay 1 mode Relay 1 range Pulse ph H. Relay 1 pulse rate pulses/minute 10 pulses/minute I. Last calibration date OPTIONS Menu Setup Parameters (sec. 6.4) Menu Parameters Display Description Range Factory Default A. Display contrast 0 to 5 3 B. 4 ma setpoint ph 0.00 ph C. 0 ma setpoint ph D. 4 ma adjust 3.0 to 5.0 ma 4.00 ma E. 0 ma adjust 19 to 1 ma 0.00 ma F. Temperature display Celsius or Fahrenheit C Signet 5700 ph/orp Monitor ph Instructions 9

10 10. Troubleshooting Display Problem Solution 1. ORP: 1000 mv Input: 998 mv (example ) ORP electrode installed in preamplifi er Remove ORP electrode and replace with recommended ph electrode (sec. ) A. Out Of Range CHECK SENSOR B. CHECK SENSOR? C. Temperature Bad CHECK SENSOR Electrode not installed in preamplifi er Temperature input miscalibrated or out of range during power up Temperature input missing or out of range during power up Temperature input missing or out of range during EASY CAL procedure (sec. 6.) 1) Verify temperature input connections and electrode installation (sec. ) ) Disconnect Green and White preamplifi er wires from rear T and T terminals. Measure resistance across disconnected wires for.9 kω 3.1 kω at 5 C (77 F). If measured resistance is out of range: Wrong electrode installed in preamplifi er (ORP electrode=10 kω). Replace ORP electrode with ph electrode. Faulty preamplifi er contacts, see preamplifi er manual for additional information. Faulty electrode, see electrode manual for additional information. 3) To verify instrument input: Disconnect Green and White preamplifi er wires from rear T and T terminals, then install a 3 kω fi xed resistor across the terminals. Power instrument and verify 5 C (77 F) on display. If 5 C (77 F) is not displayed, calibrate temperature input for 5 C (sec. 6.3B). If instrument displays correctly after calibration, instrument is ok. If error condition persists, instrument requires factory service. Refer to solution steps 1 3 above. Refer to solution steps 1 3 above 3. Out Of Range Use Manual Cal ph buffers other than ph 4, 7, 10 used during EASY CAL procedure or electrode has excess offset 4.! Same Buffer Same ph buffer used for EASY CAL buffer 1 and ph 5.0 C Excessive mv input from preamplifi er Use ph 4, 7, 10 buffers for EASY CAL procedure (sec. 6.). Use manual calibration to set standard and slope (sec. 6.3C, 6.3D) Clean electrode (see electrode manual) then retry EASY CAL (sec. 6.) Use different ph 4, 7, 10 buffer for EASY CAL procedure solution #1 and solution # (sec. 6.3). Faulty preamplifi er, refer to preamplifi er manual for additional information To verify instrument input: Disconnect Brown and Blue preamplifi er wires and Install shorting strap (jumper) across mv input and Iso Gnd terminals. Power instrument and verify 7.00 ph on display. If 7.00 ph is not displayed, calibrate Set Standard input for 7.00 ph (sec. 6.3C) If instrument displays correctly after calibration, instrument is ok. If error condition persists, instrument requires factory service. continued 10 Signet 5700 ph/orp Monitor ph Instructions

11 Display Problem Solution 6.! ph Must Be Or Less ph input out of range or improperly entered during Standard or Slope calibration A) Use two solutions of known ph within range for manual Standard and Slope calibration (sec. 6.3C, 6.3D). B) Enter ph value only within ph during standard or slope calibration (sec. 6.3C 6.3D) 7. Slope Too Close To Standard 8. Value Must Be Or Less ph slope calibration too close to ph standard calibration Relay, 4 ma, or 0 ma setpoint greater than ph (CALIBRATE Menu sec. 6.3D only) Calibrate ph slope using a ph solution ph units from the ph standard solution (sec. 6.3C, 6.3D). Enter setpoint within range (sec. 6.3E) 9. Value Must Be 300 Or Less 10. SETUP READ ERROR Press any Key Relay pulse setting greater than 300 pulses per minute Power fault occurred while saving setup menu entry Enter relay pulse setting within pulses per minute (sec. 6.3H) Press any key to reload factory defaults then reprogram system setup parameters. 11. Maintenance Clean the instrument case and front panel with a soft cloth and a mild liquid soap solution. Notes: Signet 5700 ph/orp Monitor ph Instructions 11

12 George Fischer Signet, Inc., 3401 Aerojet Avenue, El Monte, CA U.S.A. Tel. (66) Fax (66) For Worldwide Sales and Service, visit our website: Or call (in the U.S.): (800) Rev. K 3/06 English/pH George Fischer Signet, Inc Printed in U.S.A. on recycled paper

Signet 5700 ph/orp Monitor - ORP Instructions

Signet 5700 ph/orp Monitor - ORP Instructions Signet 5700 ph/ Monitor Instructions English *5700.0901* 5700.0901 Rev. K /06 English/ Contents 1. Power Connections. Compatible Electrode/Preamp Wiring. 4 0 ma Current Output Connections 4. Relay Connections

More information

SIGNET 5700 ph/orp Monitor - ORP Instructions

SIGNET 5700 ph/orp Monitor - ORP Instructions SIGNET 5700 ph/orp Monitor ORP Instructions ENGLISH 35700.0901 H (6/01) English CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections. Follow

More information

Signet 5900 Salinity Monitor

Signet 5900 Salinity Monitor Signet 5900 Salinity Monitor *35900.0901* 35900.0901 Rev. E 3/06 English Contents 1. Power Connections 2. Compatible Sensor Wiring 3. 4 20 ma Current Output Connections 4. Relay Connections 5. Relay Operation

More information

+GF+ SIGNET 5800CR Conductivity/Resistivity Monitor Instructions

+GF+ SIGNET 5800CR Conductivity/Resistivity Monitor Instructions GF SIGNET 5800CR Conductivity/Resistivity Monitor Instructions ENGLISH 35800CR.0901 D1/98 CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections.

More information

+GF+ SIGNET ph/orp Transmitter Instructions

+GF+ SIGNET ph/orp Transmitter Instructions GF SIGNET 8750 ph/orp Transmitter Instructions ENGLISH 8750.090 A9/99 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury.

More information

Signet 5800CR Conductivity/Resistivity Monitor

Signet 5800CR Conductivity/Resistivity Monitor Signet 5800CR Conductivity/Resistivity Monitor *35800CR.0901* 35800CR.0901 Rev. H 3/06 English Contents 1. Power Connections 2. Compatible Sensor Wiring 3. 4 20 ma Current Output Connections 4. Relay Connections

More information

Contents. 1. Installation

Contents. 1. Installation ENGLISH 8750.090 C/0 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents. Installation. Specifications. Electrical

More information

Signet Pressure Transmitter

Signet Pressure Transmitter Signet 850 Pressure English *850.090* 850.090 Rev. H /06 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.

More information

+GF+ SIGNET 5900 Salinity Monitor Instructions

+GF+ SIGNET 5900 Salinity Monitor Instructions +GF+ SIGNET 5900 Salinity Monitor Instructions ENGLISH 35900.0901 (B8/99), English CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections.

More information

SIGNET ph/orp Transmitter

SIGNET ph/orp Transmitter SIGNET 8750- ph/orp Transmitter ENGLISH 3-8750.090- D-4/03 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.

More information

Signet Temperature Transmitter

Signet Temperature Transmitter Loop Output Output 7 Sensr IN (RED) Sensr V (BLACK) Signet 80 Temperature 80.090 English 80.090 Rev. J 0/ English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions

More information

Signet ph/orp Transmitter

Signet ph/orp Transmitter Signet 87501 ph/orp *38750.0901* 38750.0901 Rev. H 06/10 English English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury.

More information

Signet Flow Transmitter

Signet Flow Transmitter Signet 8550 Flow *8550.090* English 8550.090 Rev. L 0/ English WARNING! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.

More information

Signet ph/orp Transmitter * *

Signet ph/orp Transmitter * * Loop 7 6 5 Signet 8750- ph/orp *-8750.090-* CAUTION! Contents 0.0 ph 5.0C Signet ph/orp. Installation. Panel Installation gasket panel terminals mounting bracket 96 mm (.8 in.) 8050 8 mm (. in.) Optional

More information

+GF+ SIGNET Conductivity/Resistivity Transmitter Instructions

+GF+ SIGNET Conductivity/Resistivity Transmitter Instructions GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions ENGLISH 8850.090 A9/99 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to

More information

User s Guide PHCN ph/orp indicator ph Instructions. Shop online at. ENTER RELAY 1 RELAY 2

User s Guide PHCN ph/orp indicator ph Instructions. Shop online at.    ENTER RELAY 1 RELAY 2 User s Guide 4 6 7 8 10 RELAY 1 RELAY 2 2 12 0 14 ENTER Shop online at www.omega.com email: info@omega.com PHCN5700 /ORP indicator Instructions OMEGAnet Online Service www.omega.com Internet email info@omega.com

More information

Table of Contents. 1 Introduction Introduction Front Panel Description Rear Panel Description 3

Table of Contents. 1 Introduction Introduction Front Panel Description Rear Panel Description 3 Chapter Page 1 Introduction 1 1.1 Introduction 1 1.2 Front Panel Description 2 1.3 Rear Panel Description 3 Table of Contents 2 Installation and Operation 4 2.1 Mounting Instructions 4 2.2 Power Connections

More information

FLS M9.06 PH/ORP MONITOR. safety instructions. packing List

FLS M9.06 PH/ORP MONITOR. safety instructions. packing List FLS M9.06 PH/ORP MONITOR safety instructions General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other instruments

More information

SIGNET Level Transmitter

SIGNET Level Transmitter ENGLISH 850.090 A8/0 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents. Description. Specifications. Panel

More information

FLS M9.06. ph/orp MONITOR

FLS M9.06. ph/orp MONITOR FLS M9.06 /ORP MONITOR Safety Instructions General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other instruments

More information

Signet 9900 Transmitter

Signet 9900 Transmitter Signet 9900 Transmitter Member of the SmartPro Family of Instruments Features Panel Mount Field Mount NEW! Modbus Module supports RS485 Serial Modbus Communications Multiple sensor types supported with

More information

FLOW MONITOR & TRANSMITTER

FLOW MONITOR & TRANSMITTER FLS M9.02 FLOW MONITOR & TRANSMITTER SAFETY INSTRUCTIONS General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other

More information

Signet 2750 DryLoc ph/orp Sensor Electronics

Signet 2750 DryLoc ph/orp Sensor Electronics Signet 750 DryLoc ph/orp Sensor Electronics *750.090* 750.090 Rev. G /05 English WARNING! SAFETY INSTRUCTIONS. Depressurize and vent system prior to installation or removal.. Confi rm chemical compatibility

More information

+GF+ SIGNET 9050CR Conductivity/Resistivity Controller

+GF+ SIGNET 9050CR Conductivity/Resistivity Controller +GF+ SIGNET 9050CR Conductivity/Resistivity Controller Instruction Manual GEORGE FISCHER +GF+ Important Safety Information! CAUTION: Remove AC power to unit prior to wiring input and output connections.

More information

X Year mfg

X Year mfg 4to20 ma Output S L Loop PWR Loop2 PWR N/C S L DATA 7 8 9 6,*(7 S L Output 8 S L SIGNET 8900 MultiParameter Controller Supplemental Manual *8900.090SUP* 8900.090SUP Rev. A 2/04 (English). Description This

More information

Autoranging True RMS Multimeter User Manual

Autoranging True RMS Multimeter User Manual Autoranging True RMS Multimeter User Manual Please read this manual before switching the unit on. Important safety information inside. Contents Page 1. Safety Information... 4 2. Safety Symbols... 5 3.

More information

OPERATION MANUAL MODEL TT-LP

OPERATION MANUAL MODEL TT-LP TURTLE TOUGH OPERATION MANUAL MODEL TT-LP 2-Wire Isolated ph/orp Transmitter turtletough.com.au CONTENT INITIAL INSPECTION 3 INTRODUCTION 3 ASSEMBLY 4 PREPARATION 5 Connection Diagram of Turtle Tough Sensors

More information

4801P ph/orp controller

4801P ph/orp controller 4801P ph/orp controller Introduction: Thank you for choosing this ph/orp controller. This 1/16 DIN controller offers dual backlight LCD display, high accuracy, LED indicators, two control relays, and automatic

More information

Analyzer/Controller. ph/orp HART. Model 54e ph/orp ESSENTIAL INSTRUCTIONS WARNINGS RISK OF ELECTRICAL SHOCK

Analyzer/Controller. ph/orp HART. Model 54e ph/orp ESSENTIAL INSTRUCTIONS WARNINGS RISK OF ELECTRICAL SHOCK Instruction Sheet PN 51A-54epH/rev J October 2010 ph/orp HART Model 54e ph/orp Analyzer/Controller For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. ESSENTIAL

More information

Mini Digital Multimeter

Mini Digital Multimeter User's Guide Mini Digital Multimeter Model MN15 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Back to the Extech MN15/MN16 Series

More information

Mini Digital Multimeter

Mini Digital Multimeter User Manual Mini Digital Multimeter Model MN15A Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the Extech MN15A MultiMeter. The MN15A offers

More information

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter Operation Manual MODEL 2TX 2-wire Isolated ph/orp Transmitter 0 2TX CONTENTS INITIAL INSPECTION.....2 INTRODUCTION......2 ASSEMBLY...3 PREPARATION....4 CONNECTING THE ELECTRODE...4 CONNECTING THE TEMPERATURE

More information

JUMO ecotrans ph 03 Microprocessor transmitter/ switching device for ph/redox voltage and temperature

JUMO ecotrans ph 03 Microprocessor transmitter/ switching device for ph/redox voltage and temperature Page 1/7 JUMO ecotrans 03 Microprocessor transmitter/ switching device for /Redox voltage and temperature with a 2-line LCD for mounting on a 35 mm DIN rail Brief description Depending on the configuration,

More information

High Voltage DC Meter

High Voltage DC Meter High Voltage DC Meter Javelin D PD644 0-300 VDC input NEMA 4X, IP65 front Scale in engineering units Sunlight readable LED display 4-20 ma analog output Two form C 3 A relays option RS-485 serial communications

More information

HIGH-VOLTAGE DC METER

HIGH-VOLTAGE DC METER HIGH-VOLTAGE DC METER Javelin D PD644 0-300 VDC input NEMA 4X, IP65 front Scale in engineering units Sunlight readable LED display 4-20 ma analog output Two form C 3 A relays option RS-485 serial communications

More information

UDC 1000 and UDC 1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS

UDC 1000 and UDC 1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS UDC 1000 and UDC 1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS EN0I-6041 12/99 PRODUCT SPECIFICATION SHEET OVERVIEW The UDC 1000 and UDC 1500 are microprocessor-based 1/16 DIN and 1/8 DIN controllers

More information

1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882

1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882 1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882 This manual is intended for use in support of installation, commissioning and configuration of the 1/32-DIN Temperature

More information

OPERATING INSTRUCTION

OPERATING INSTRUCTION OPERATING INSTRUCTION AUTORANGING MULTIMETER MAX Ω F C 10A MAX every 15 min. COM V SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation

More information

5450 NW 33rd Ave, Suite 104 Fort Lauderdale, FL Fruitland Ave Los Angeles, CA SST7000 SST7100. Speed Switch / Transmitter

5450 NW 33rd Ave, Suite 104 Fort Lauderdale, FL Fruitland Ave Los Angeles, CA SST7000 SST7100. Speed Switch / Transmitter 5450 NW 33rd Ave, Suite 104 Fort Lauderdale, FL 33309 3211 Fruitland Ave Los Angeles, CA 90058 SST7000 SST7100 Speed Switch / Transmitter Installation and Operation Manual Rev. C P/N145F-13112 PCO 00009270

More information

MYRIAD QLC 4-CHANNEL MONITOR/CONTROLLER INSTRUCTION MANUAL

MYRIAD QLC 4-CHANNEL MONITOR/CONTROLLER INSTRUCTION MANUAL MYRIAD QLC 4-CHANNEL MONITOR/CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM MYR QLC MANUAL 013114 2 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features WIRING

More information

LAUREL ELECTRONICS, INC.

LAUREL ELECTRONICS, INC. LAUREL ELECTRONICS, INC. Laureate RTD Temperature Panel Meter / Controller Features Factory calibrated for 100Ω platinum, 10Ω copper & 120Ω nickel RTDs 2, 3 or 4-wire connection with lead resistance compensation

More information

MODEL: M5DY SEN TRONIC AG CURRENT LOOP SUPPLY. [2] POWER INPUT AC Power M: V AC (Operational voltage range V, Hz)

MODEL: M5DY SEN TRONIC AG CURRENT LOOP SUPPLY. [2] POWER INPUT AC Power M: V AC (Operational voltage range V, Hz) Super-mini Terminal Block Signal Conditioners M5-UNIT CURRENT LOOP SUPPLY Functions & Features Powering a 0 ma DC current loop Applicable to smart transmitters Isolation between the input and output Fast

More information

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER SAFETY INFORMATION The following safety information must be observed to ensure maximum personal safety during the operation of this meter: This meter is

More information

Mini Digital Multimeter

Mini Digital Multimeter User's Guide Mini Digital Multimeter Model MN15 Introduction Congratulations on your purchase of the Extech MN15 MultiMeter. The MN15 offers AC/DC Voltage, AC/DC Current, Resistance, Diode, and Continuity

More information

Owner's Manual. True RMS Multimeter. Model No Safety Operation Maintenance Español

Owner's Manual. True RMS Multimeter. Model No Safety Operation Maintenance Español Owner's Manual True RMS Multimeter Model No. 82023 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance Español

More information

MYRIAD LC1 LEVEL CONTROLLER INSTRUCTION MANUAL

MYRIAD LC1 LEVEL CONTROLLER INSTRUCTION MANUAL MYRIAD LC1 LEVEL CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM MYR LC1 MANUAL 062114 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features Dimensions WIRING

More information

TABLE I RANGE, POWER, OUTPUT (CONDUCTIVITY MEASUREMENTS)

TABLE I RANGE, POWER, OUTPUT (CONDUCTIVITY MEASUREMENTS) Issue 11 Page 1 of 5 Model Selection Guide Instructions Consult Steps to Selecting Appropriate Conductivity Instrumentation and Cells before making selections below. Select the desired key number. The

More information

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter S-14 Compact Digital Multimeter S-14 Compact Digital Multimeter SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation at this meter

More information

Tempco Instruction Manual

Tempco Instruction Manual Tempco Instruction Manual 1/16 DIN Solid State Temperature Controller Relay Output Solid State Output For Heating Model Numbers: TEC-901, TEC-902, TEC-905 Temperature controls in this series are designed

More information

DIGITAL PANEL THERMOMETER

DIGITAL PANEL THERMOMETER Love Controls DIGITAL PANEL THERMOMETER LCI108-2x LCI108J-2x INSTRUCTIONS MANUAL EDITION: February 2001 INTRODUCTION TO THE LOVE CONTROLS LCI108 SERIES This manual does not constitute a formal agreement.

More information

700 SERIES DPC DUPLEX PUMP CONTROLLER INSTRUCTION MANUAL

700 SERIES DPC DUPLEX PUMP CONTROLLER INSTRUCTION MANUAL 1 700 SERIES DPC DUPLEX PUMP CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM 700 DPC MANUAL 072114 2 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features WIRING

More information

Conductivity Transmitters

Conductivity Transmitters Conductivity Transmitters CDTX-90 Series Starts at $ 576 Electronics Only Display in µs, ms, kω, MΩ, ppm (TDS) Simulate Function Programmable Temperature Compensation Relay Options Dual Output Option Allows

More information

TABLE OF CONTENTS INTRODUCTION. 3. Analog Input Analog Output Digital Input Digital Output OPERATIONAL DESCRIPITON.. 7 PROGRAMMING AND INITIAL SETUP.

TABLE OF CONTENTS INTRODUCTION. 3. Analog Input Analog Output Digital Input Digital Output OPERATIONAL DESCRIPITON.. 7 PROGRAMMING AND INITIAL SETUP. DIVERSIFIED HEAT TRANSFER SERIES 700 STEAM GENERATOR CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE AT SIGMACONTROLS.COM SERIES 700 DHT STEAM GENERATOR MANUAL 042514 2 TABLE OF CONTENTS INTRODUCTION.

More information

LAUREL. Laureate RTD Temperature Panel Meter / Controller ELECTRONICS, INC. Features. Description. Specifications

LAUREL. Laureate RTD Temperature Panel Meter / Controller ELECTRONICS, INC. Features. Description. Specifications LAUREL ELECTRONICS, INC. Laureate RTD Temperature Panel Meter / Controller Features Factory calibrated for 100Ω platinum, 10Ω copper & 120Ω nickel RTDs 2, 3 or 4-wire connection with lead resistance compensation

More information

User's Guide. Extech AM A AC Analog Clamp Meter

User's Guide. Extech AM A AC Analog Clamp Meter User's Guide Extech AM300 300A AC Analog Clamp Meter Introduction Congratulations on your purchase of the Extech AM300 Analog Clamp Meter. This device measure AC Voltage and Current, DC Voltage, and Resistance.

More information

ABB MEASUREMENT & ANALYTICS DATA SHEET. TB84PH ph /ORP / pion transmitter

ABB MEASUREMENT & ANALYTICS DATA SHEET. TB84PH ph /ORP / pion transmitter ABB MEASUREMENT & ANALYTICS DATA SHEET TB84PH ph /ORP / pion transmitter 2 TB 8 4PH P H /ORP / P I ON TR AN S MIT TE R DS/ TB8 4PH-E N RE V. D Measurement made easy Hazardous area rated transmitter with

More information

RXTH DUAL ROOM SENSOR / SWITCH

RXTH DUAL ROOM SENSOR / SWITCH DUAL ROOM SENSOR / SWITCH FOR TEMPERATURE AND RELATIVE HUMIDITY Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE

More information

Cole-Parmer Instrument Company

Cole-Parmer Instrument Company Operating Instructions Cole-Parmer ph/orp 400 ph/orp Controller Cole-Parmer Instrument Company 625 East Bunker Court, Vernon Hills, Illinois 60061 Tel: 1-847-549-7600 or Toll-free: 1-800-323-4340 Fax:

More information

User Guide True RMS Multimeter Extech EX205T

User Guide True RMS Multimeter Extech EX205T User Guide Extech EX205T True RMS Digital Multimeter Extech EX210T True RMS Digital Multimeter IR True RMS Multimeter Extech EX205T Introduction Thank you for selecting the Extech EX205T True RMS Auto-ranging

More information

Signet BC Batch Controller System

Signet BC Batch Controller System Signet 9900-1BC Batch Controller System *3-9900-1BC.090* 3-9900-1BC.090-1 Rev. A1 03/13 English Product Manual Look for the Quick Start icon to quickly set up your new 9900-1BC. Safety WARNING! Panel Mount

More information

User's Guide. MiniTec TM Series Model MN25 MultiMeter

User's Guide. MiniTec TM Series Model MN25 MultiMeter User's Guide MiniTec TM Series Model MN25 MultiMeter Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a

More information

OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS

OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS GB Contents Page 1 Introduction 4 2 Safety 5 3 Use 6 4 Checking on Delivered Equipment 6 5 General Specification and Dimensions 7 6 Positioning 10 7 Operating

More information

TC Type Range Conformity Error. J -210 C to +760 C (-347 F to F) ±0.09 C (±0.16 F) K -244 C to C (-408 F to F) ±0.1 C (±0.

TC Type Range Conformity Error. J -210 C to +760 C (-347 F to F) ±0.09 C (±0.16 F) K -244 C to C (-408 F to F) ±0.1 C (±0. LAUREL ELECTRONICS, INC. Laureate Thermocouple Panel Meter / Controller Features Factory calibrated for thermocouple types J, K, T, E, N, R, S Entire range of each thermocouple in one scale Highly accurate

More information

Adapter Kit - QSI Electronics to TM Meter Installation Instructions

Adapter Kit - QSI Electronics to TM Meter Installation Instructions Adapter Kit - QSI Electronics to TM Meter Installation Instructions These installation instructions cover the adapter kit installation of the QSI1, QSI2 and QSI3 versions of communications electronics

More information

User's Guide. 1/16 th DIN ph Controller. Model 48PH2. Introduction

User's Guide. 1/16 th DIN ph Controller. Model 48PH2. Introduction User's Guide 1/16 th DIN ph Controller Model 48PH2 Introduction Thank you for selecting the Extech Model 48PH2 controller. This 1/16 th DIN-size controller offers dual display, high resolution measurements,

More information

INSTALLATION DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER. Before installation:

INSTALLATION DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER. Before installation: DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER The DKM-409 is a precision instrument designed for displaying various AC parameters in 3-phase distribution panels. Thanks to its isolated

More information

Operating Manual. Dynisco /8 DIN Indicator Concise Product Manual

Operating Manual. Dynisco /8 DIN Indicator Concise Product Manual Dynisco 1490 1/8 DIN Indicator Concise Product Manual 59476-9 Operating Manual -1- CAUTION: Installation should be only performed by technically competent personnel. Local Regulations regarding electrical

More information

ph/orp 300 (Cat. No ) MICROCOMPUTER BASED OPERATION MANUAL ph /ORP/ Temperature CONTROLLER Phone: , Fax:

ph/orp 300 (Cat. No ) MICROCOMPUTER BASED OPERATION MANUAL ph /ORP/ Temperature CONTROLLER Phone: , Fax: OPERATION MANUAL /ORP 300 (Cat. No. 0580212) MICROCOMPUTER BASED /ORP/ Temperature CONTROLLER 625 East Bunker CT, Vernon Hills, IL 60061-1844 Phone: 800-32-4340, Fax: 847-247-2929 E-mail: info@coleparmer.com,

More information

Operating instructions. Speed monitor D / / 2014

Operating instructions. Speed monitor D / / 2014 Operating instructions Speed monitor D200 80005257 / 00 05 / 2014 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5 2.2 Target group...5

More information

12-36 VDC/12-24 VAC Power Option 4-Digit Display, 0.56 (14.2 mm) or 1.20 (30.5 mm)

12-36 VDC/12-24 VAC Power Option 4-Digit Display, 0.56 (14.2 mm) or 1.20 (30.5 mm) 4-20 ma & Relay Output Features 4-20 ma, ± 10 V, TC & RTD Inputs 12-36 VDC/12-24 VAC Power Option 4-Digit Display, 0.56 (14.2 mm) or 1.20 (30.5 mm) 24 VDC @ 200 ma Transmitter Power Supply Options Type

More information

1. Installation. 2. Configuration - Operation. 3. Specifications. Safety

1. Installation. 2. Configuration - Operation. 3. Specifications. Safety Safety 1. Installation 1.1 OP_ext: Services 1.2 Driver Card Installation 1.3 Sensor Types 1.4 Sensor Wiring 1.5 Controller Wiring 2. Configuration - Operation 2.1 Replaces ph Sensor 2.2 AS -Flex Series

More information

User Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at

User Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at User Manual 400Amp AC Clamp Meter + NCV Model MA430 Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of this Extech MA430 Clamp Meter. This

More information

Digital Proportional Valve Amplifier

Digital Proportional Valve Amplifier Digital Proportional Valve Amplifier The Amplifier is a compact, low profile DIN mounted controller for use with proportional solenoid valves. The provides current to a valve coil in proportion to an input

More information

User's Guide. Digital Multimeter. Model MN42

User's Guide. Digital Multimeter. Model MN42 User's Guide Digital Multimeter Model MN42 Introduction Congratulations on your purchase of the Extech MN42 MultiMeter. The MN42 offers AC/DC Voltage, DC Current, and Resistance testing. Proper use and

More information

R5050. Model. Instruction Manual. TRMS AC/DC Clamp Meter. reedinstruments. www. com

R5050. Model. Instruction Manual. TRMS AC/DC Clamp Meter. reedinstruments. www. com Model R5050 TRMS AC/DC Clamp Meter Instruction Manual reedinstruments com Table of Contents Safety... 3 Features... 4 Specifications...4-6 Instrument Description...7-8 Measurement Procedures...9-12 Battery

More information

INDICATOR FOR PROCESS CONTROL. Love Controls. LCI108J-0x INSTRUCTION MANUAL. EDITION: January Valid for models with software version P-1

INDICATOR FOR PROCESS CONTROL. Love Controls. LCI108J-0x INSTRUCTION MANUAL. EDITION: January Valid for models with software version P-1 Love Controls INDICATOR FOR PROCESS CONTROL LCI108-0x LCI108J-0x INSTRUCTION MANUAL EDITION: January 2001 Valid for models with software version P-1 INTRODUCTION TO THE LOVE CONTROLS LCI108 SERIES This

More information

Operation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller

Operation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller Operation Manual TX10 1/8 DIN Microcomputer Based /ORP Controller 11751 Markon Drive Garden Grove, CA. 92841 U.S.A. pg. 1 CONTENTS GENERAL INTRODUCTION.....3 INITIAL INSPECTION......3 USING THE Sensorex

More information

SHARK MULTI-PARAMETER CONTROLLER & ANALYZER USER S MANUAL

SHARK MULTI-PARAMETER CONTROLLER & ANALYZER USER S MANUAL SHARK MULTI-PARAMETER CONTROLLER & ANALYZER USER S MANUAL Rev 3 AquaMetrix Inc. 1245 Maple Hill Ct., Unit 7 Newmarket, ON Canada, L3Y 9E8 Tel: (800) 742-1413 (905) 954-0841 Fax: (905) 954-0415 www.aquametrix.com

More information

Analog outputs Output 1 + 2: 0(4) 20 ma or 0 10 V. Configurable as - analog process value output - continuous controller output (PID action)

Analog outputs Output 1 + 2: 0(4) 20 ma or 0 10 V. Configurable as - analog process value output - continuous controller output (PID action) Page 1/10 JUMO AQUIS 500 ph Transmitter/Controller for ph, ORP, NH 3 (ammonia) concentration and temperature Brief description The instrument is used for measuring/controlling the ph, ORP or NH 3 (ammonia)

More information

User's Guide. 800 Amp Clamp Meters. EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter

User's Guide. 800 Amp Clamp Meters. EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter User's Guide 800 Amp Clamp Meters EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter Introduction Congratulations on your purchase of the EX710, EX720, or EX730 Clamp DMM.

More information

Model: Pro93 TRUE RMS LEAKAGE CURRENT TESTER

Model: Pro93 TRUE RMS LEAKAGE CURRENT TESTER Model: Pro93 TRUE RMS LEAKAGE CURRENT TESTER CONTENTS TITLE PAGE I. Safety Information.......1 Environmental Conditions... 1 Explanation of Symbols..... 1 II. Specification..... 2 General Specification...

More information

Trident and Trident X2 Digital Process and Temperature Panel Meter

Trident and Trident X2 Digital Process and Temperature Panel Meter Sign In New User ISO 9001:2008 Certified Quality System Home Products Online Tools Videos Downloads About Us Store Contact Policies Trident and Trident X2 Digital Process and Temperature Panel Meter Products

More information

RТTH DUAL ROOM SWITCH

RТTH DUAL ROOM SWITCH DUAL ROOM SWITCH FOR TEMPERATURE AND RELATIVE HUMIDITY Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS PRODUCT DESCRIPTION ARTICLE CODES INTENDED AREA OF USE TECHNICAL DATA

More information

PUMP DRIVE OPERA TING MANUAL: A Edition 01

PUMP DRIVE OPERA TING MANUAL: A Edition 01 OPERA TING MANUAL: PUMP DRIVE 77420-00 Pump Drive, shown with 77600-62 High Performance I/P Pump Head. 77420-00 A-1288-0990 Edition 01 1-800-637-3739 (U.S. and Canada only) 11 (847) 381-7050 (Outside U.S.)

More information

Signet Conductivity/Resistivity Electrodes

Signet Conductivity/Resistivity Electrodes Signet 2819-2823 Conductivity/Resistivity Electrodes 2819 2820 2821 2822 2823 Sanitary Features Standard process connections - ¾ in. NPT Polypro - Tri-clamp 1-1½ in., 2 - Opt. ½ in. NPT 316 SS 316 SS or

More information

NKQB24-1, NKQX24-1 On/Off, Electronic Fail-Safe, Direct Coupled, 24V

NKQB24-1, NKQX24-1 On/Off, Electronic Fail-Safe, Direct Coupled, 24V NKQB24-1, NKQX24-1 On/Off, Electronic Fail-Safe, Direct Coupled, 24V orque min. 54 in-lb for control damper surfaces up to 12 sq ft. Application For on/off, fail-safe control of dampers in HVAC systems.

More information

EA401 SERIES PULSE INPUT FLOW RATEMETER/TOTALIZER INSTALLATION & OPERATING INSTRUCTIONS West Stetson Avenue Hemet, CA USA

EA401 SERIES PULSE INPUT FLOW RATEMETER/TOTALIZER INSTALLATION & OPERATING INSTRUCTIONS West Stetson Avenue Hemet, CA USA EA401 SERIES PULSE INPUT FLOW RATEMETER/TOTALIZER INSTALLATION & OPERATING INSTRUCTIONS 3255 West Stetson Avenue Hemet, CA 92545-7799 USA Lit# 24508-67 99773 Rev. 1.1 02/25/0 Tel: 951-652-6811 Fax: 951-652-3078

More information

Color Mark Photoelectric Sensor E3S-VS

Color Mark Photoelectric Sensor E3S-VS Sensing Supply voltage Output 12 to 80 ma, with 1.5 to 4 ma constant current source; 1.2 cm, 5 cm 100 ma, Color Mark Photoelectric Sensor Small Color Mark Sensor With Built-In DC Amplifier Fast, 1 ms maximum

More information

AutoRanging Digital MultiMeter

AutoRanging Digital MultiMeter Owner's Manual AutoRanging Digital MultiMeter Model No. 82175 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance

More information

700 SERIES TPC TRIPLEX PUMP CONTROLLER INSTRUCTION MANUAL

700 SERIES TPC TRIPLEX PUMP CONTROLLER INSTRUCTION MANUAL 1 700 SERIES TPC TRIPLEX PUMP CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM 700SERIESTPCMANUAL071414 2 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features

More information

ProSense DPM2 Series. Digital Panel Meter. User Manual. for process, temperature and resistance input signals. Models: DPM2-AT-HL DPM2-AT-2R-HL

ProSense DPM2 Series. Digital Panel Meter. User Manual. for process, temperature and resistance input signals. Models: DPM2-AT-HL DPM2-AT-2R-HL User Manual ProSense DPM2 Series Digital Panel Meter for process, temperature and resistance input signals Models: DPM2-AT-HL DPM2-AT-2R-HL Scan or click the QR code for a series of Configuration and Programming

More information

I-7520: RS-232 to RS-485. Input: RS-232 protocol Output:RS-485/RS-422 Speed: Self Tuner 'inside, auto switching baud rate, 300~ BPS

I-7520: RS-232 to RS-485. Input: RS-232 protocol Output:RS-485/RS-422 Speed: Self Tuner 'inside, auto switching baud rate, 300~ BPS I-7520 RS-232 to RS-485 Converter I-7510 RS-485 Repeater RS-232 RS-485 I-7000 SERIES Common Specifications Isolation voltage: 3000VDC Speed: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115000 Dual watchdog

More information

M A C 3 Wind Speed Alarm & Controller

M A C 3 Wind Speed Alarm & Controller M A C 3 Wind Speed Alarm & Controller Installation Instructions Thank you for purchasing the MAC3 wind speed alarm and controller. This manual is designed to lead you through a step-by-step process to

More information

LCD LCDS LCDRI CONTROLLER OPERATING MANUAL

LCD LCDS LCDRI CONTROLLER OPERATING MANUAL This manual contains important safety information about installation and use of this equipment. Ignoring this information could result in injuries or damages. It is strictly forbidden to use this equipment

More information

DIGITAL PANEL THERMOMETER

DIGITAL PANEL THERMOMETER Love Controls DIGITAL PANEL THERMOMETER LCI108-1x LCI108J-1x INSTRUCTION MANUAL EDITION: January 2001 INTRODUCTION TO THE LOVE CONTROLS LCI108 SERIES This manual does not constitute a formal agreement.

More information

INSTALLATION MANUAL. LC 200 Electronic Overload Guard. Software versione PW0501 R 0.3

INSTALLATION MANUAL. LC 200 Electronic Overload Guard. Software versione PW0501 R 0.3 INSTALLATION MANUAL LC 200 Electronic Overload Guard Software versione PW0501 R 0.3 CONTENTS MAIN FEATURES LC 200 TECHNICAL FEATURES Page 2 SYMBOLS Page 3 WARNINGS Page 3 IDENTIFICATION DATA PLATE Page

More information

RTT2 ROOM TEMPERATURE SWITCH. Mounting and operating instructions

RTT2 ROOM TEMPERATURE SWITCH. Mounting and operating instructions Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS PRODUCT DESCRIPTION ARTICLE CODES INTENDED AREA OF USE TECHNICAL DATA STANDARDS OPERATIONAL DIAGRAMS WIRING AND CONNECTIONS

More information

Table of Contents. General Information. Document Sure-Aire Flow Monitoring System. User and Service Manual WARNING CAUTION

Table of Contents. General Information. Document Sure-Aire Flow Monitoring System. User and Service Manual WARNING CAUTION Document 476092 User and Service Manual Installation, Operation and Maintenance Manual Please read and save these instructions for future reference. Read carefully before attempting to assemble, install,

More information

Instruction Manual. M Pump Motor Controller. For file reference, please record the following data:

Instruction Manual. M Pump Motor Controller. For file reference, please record the following data: Instruction Manual M Pump Motor Controller For file reference, please record the following data: Model No: Serial No: Installation Date: Installation Location: When ordering replacement parts for your

More information

NKQB24-1, NKQX24-1 On/Off, Fail-Safe, Operation, Direct Coupled, 24V

NKQB24-1, NKQX24-1 On/Off, Fail-Safe, Operation, Direct Coupled, 24V NKQB24-1, NKQX24-1 On/Off, Fail-Safe, Operation, Direct Coupled, 24V orque min. 54 in-lb for control damper surfaces up to 12 sq ft. Application For proportional modulation of dampers in HVAC systems.

More information