AC CIRCUIT BREAKER TEST SET MODEL NUMBER HC-100 (480V) SERIAL NUMBER

Size: px
Start display at page:

Download "AC CIRCUIT BREAKER TEST SET MODEL NUMBER HC-100 (480V) SERIAL NUMBER"

Transcription

1 AC CIRCUIT BREAKER TEST SET MODEL NUMBER HC-100 (480V) SERIAL NUMBER MJH/HWR/slu March 18, 1999

2 TABLE OF CONTENTS SECTION NUMBER DANGER TITLE PAGE 1 SPECIFICATIONS 2 INTRODUCTION 3 SAFETY 4 INSTALLATION INSTRUCTIONS 5 TURN ON/OFF PROCEDURE 6 DESCRIPTION 7 CONTROLLER DESCRIPTION 8 MAIN OPERATING SCREEN 9 EDIT SCREEN 10 INITIAL TEST SCREEN 11 SYSTEM SETUP 12 TESTING MODES 13 THEORY OF OPERATION 14 TROUBLESHOOTING 15 CALIBRATION 16 SERVICE AND MAINTENANCE 17 APPENDIX RETURNED MATERIAL WARRANTY PARTS ORDERING INFORMATION RECOMMENDED SPARE PARTS PARTS LIST ELECTRICAL SCHEMATICS A B C D E F ATTACHMENTS KBIC-120 Variac Motor Drive (Speed) Control, #

3 1-1 AC CIRCUIT BREAKER TEST SET Model Number HC-100 Serial Number Customer Cutler Hammer 6024 S W Jean Rd., Bldg C10 Lake Oswego, OR Customer's Purchase Order Number Manufacturing Date

4 2-1 SPECIFICATIONS Input: 480 VAC, 1 phase, 60 Hz Output: Parallel Connection: Continuously adjustable over eight tap ranges Tap 1: VAC open circuit Tap 2: VAC open circuit Tap 3: VAC open circuit Tap 4: VAC open circuit Tap 5: VAC open circuit Tap 6: VAC open circuit Tap 7: VAC open circuit Tap 8: VAC open circuit Series Connection: Continuously adjustable over eight tap ranges Tap 1: VAC open circuit Tap 2: VAC open circuit Tap 3: VAC open circuit Tap 4: VAC open circuit Tap 5: VAC open circuit Tap 6: VAC open circuit Tap 7: VAC open circuit Tap 8: VAC open circuit Duty Cycle: Instrumentation: Parallel Connection Duty Series Connection 0-16 VAC in 8 steps 0-32 VAC in 8 steps 10,000 Amps 30 min ON/30 min OFF 5,000 Amps 20,000 Amps 30 sec ON/3 min OFF 10,000 Amps 60,000 Amps 3 sec ON/2 min OFF 30,000 Amps Currentmeter: Resolution:.1% of full scale Ranges: 100/1000/10,000/100,000 Amps Accuracy: +/-1% of full scale continuous +/-5% of full scale seconds Timer: Resolution:.002 seconds.12 cycles Ranges: seconds cycles Accuracy: +/-.004 seconds of reading +/-.25 cycles of reading

5 3-1 INTRODUCTION Phenix Technologies' High Current Test Sets represent a new era of electrical testing technology, based on time-proven concepts and the power offered by a personal computer. High reliability SCR's have been combined with a phase angle firing system to deliver precisely repeatable current pulses with minimal distortion. The output is continuously variable over the entire range by means of tap selection and vernier adjustment. Utilizing a motorized vernier and programmable tap selector the test set has auto-jog and current-hold capability. The test set incorporates a standard measurement and control interface that is operated with special cards installed in a personal computer, under software control. The combination of automated hardware and a highly intelligent monitor/controller provides an extremely versatile test instrument capable of exceptional accuracy.

6 4-1 SAFETY Because the HC-100 is designed to supply high current, safeguards have been built into it. These safeguards are only effective if safe work practices are also used. The test set and these procedures have been designed for use by knowledgeable electrical workers who have been properly trained in its use. Please note the following: 1. The test set must be grounded before input power is connected. 2. All power must be disconnected before the test set is ungrounded. 3. The tests described here are to be used only on DE-ENERGIZED and isolated circuit breakers. 4. High currents are generated during the tests and extreme caution must be used. Follow all safety procedures. 5. See page 5-1 for Installation Instructions.

7 INSTALLATION INSTRUCTIONS 5-1 The following procedure should always be followed whenever the HC-100 is installed for operation. 1. Connect #4 AWG ground cable from a good electrical ground to the chassis ground terminal. Next, connect the isolated ground cable to the power ground. Do not connect the regulator chassis ground terminal to the high current section isolated ground terminal. Each section should have an independent ground cable. See connection diagram below. 2. Mount the computer controller on top of the regulator by positioning the controller and connecting the interconnect cable. Take special care not to pinch or damage the interconnect cable. Next, plug the computer into the provided ground fault outlet and position all wires and cables behind the computer to prevent catching or snagging during movement. 3. Perform the Power On Procedure as described on page 6-1. AC INPUT GRD

8 POWER ON PRODEDURE Ensure proper ground connection. 2. CONTROL POWER switch is in the OFF position. 3. Ensure that the interconnect cable between computer and the test set has no damage or missing pins. 4. Connect the computer interconnect cable to the outlet on top of the test set. 5. Place CB1 and CB2 in the OFF position. 6. Ensure proper power to test set prior to connecting input power leads. 7. Connect input power leads. 8. Energize the power distribution breaker. 9. Place the CONTROL POWER switch in the ON position. 10. Place CB1, then CB2 in the ON position. 11. Ensure computer is turned on. (Switch located on back of computer). 12. Follow test procedure. POWER OFF PRODEDURE 1. Exit test program. 2. Place CB2 in the OFF position. 3. Place CB1 in the OFF position. 4. Return to DOS prompt in computer program. 5. Place the CONTROL POWER switch in the OFF position. 6. De-energize power distribution breaker. 7. Disconnect test set as required. WARNING: Connecting or disconnecting computer with power applied may damage test set and/or computer.

9 DESCRIPTION 7-1 Fuses and Circuit Breakers The single phase, 480 V power enters the regulator and is immediately connected to fuses, F1 and F2. These fuses provide the first level of overload protection. Fuses, F3 and F4, provide control power transformer protection. Fuse, F5, protects the 120 V control circuit. All the power transformers are protected with fuses, F6-F10. Circuit breaker, CB1, protects the main power section and is equipped with an undervoltage release coil. The undervoltage release coil is wired so that if the interlock is open or pulsed open for any reason, the main circuit breaker will open and disable the power section. Circuit breaker, CB2, protects the variable transformer. Interlock Circuit The interlock circuit is a safety mechanism that prevents the high power section from operating under unsafe conditions. In order to close the interlock the following conditions must exist: 1. Control power must be on. 2. The regulator section must be grounded at the power source from the isolated ground terminal. The cabinet must be grounded from the provided ground terminal. 3. The variable transformer circuit breaker, CB2, must be closed. 4. The computer controller must be on and reset. If the computer is displaying INTLK or OVERTEMP, the interlock will not close. NOTE: Always use #4 AWG wire or larger for grounding. After the interlock circuit is closed the panel INTLK light will go out and allow the main circuit breaker CB1 to be reset. Triggering The SCR's are controlled by a trigger board that will turn on the output when signaled by the computer. The trigger board receives its command from the input/output board and the computer controller. In this manner the computer controller can precisely time and control the phase angle of the output. Taps and Variable Transformer To achieve a variable output from minimum to maximum, a system of taps and a motorized variable transformer are employed. The tap selects the coarse range of the output and the variable transformer varies the output from zero to maximum for each tap. All of the taps and the motorized variable transformer are controlled by the computer controller. This method makes possible the current hold and auto-jog functions, as well as an easy to use, versatile, test set. A potentiometer is accessible on the front panel for varying the speed of the motorized variable transformer. All transformers and regulators are equipped with several thermocouples and a thermistor to monitor and protect against an over-temperature condition. The output of the thermistor is fed back to the computer controller so that temperature can be constantly monitored. In addition to these safeguards, the SCR's are forced air cooled to effect lower operating temperatures and longer service life.

10 7-2 DESCRIPTION (Continued) Metering There is a precision current transformer employed in the high current section to measure the output current. This transformer is placed directly around the output bus so actual output current up to 100,000 amps can be measured. Output Stabs The output stabs are configured for parallel or series connection. Parallel output connection is achieved by using the supplied stabs to short H1 and H2 for one output connection and shorting L1 and L2 for the other output connection. Series connection is achieved by shorting L1 and H2 and connecting the breaker to H1 and L2. The stabs are designed to be versatile and facilitate connection to a large variety of circuit breakers. The stabs are nickel plated for greater conductivity and long service life. Continuity Located above the output stabs are the continuity binding posts. These posts can be connected to an unused pole of the breaker so that the normally closed or normally open test modes can be used. If the continuity connections are not used, the set must be operated in current latch mode. The following is a brief description of the various parts of the HC Computer Controller. Performs all monitor and control functions. 2. Input Power Indicator, DS1. Indicates that 480 Volt power is available in the test set. 3. Control Power Switch, S1. Applies 120 Volt AC control power to the test set. 4. Fuse, F3. Control power transformer primary overload protection. 5. Fuse, F4. Control power transformer primary overload protection. 6. Fuse, F5. Control power overload protection. 7. Variable Transformer Circuit Breaker, CB2. Applies power and provides overcurrent protection for variable transformer. Must be energized for INTLK to be cleared. 8. Main Circuit Breaker, CB1. Applies 480 Volt power to the transformers and relays. An undervoltage safety release is tied to the interlock circuit for safety purposes. If the interlock light is lit on the control panel, the main circuit breaker cannot be turned on. If, during the operation of the test set, the interlock comes on, the main circuit breaker will trip. 9. Emergency Off Button, SW202. De-energizes the output and the main circuit breaker. 10. Control Power Indicator, LMP202. Indicates internal 120 Volt AC control power has been energized. 11. Output On Indicator, LMP204. Indicates that the output is energized. 12. Interlock Open Indicator, LMP203. Indicates that the safety interlock is not reset. The main breaker cannot be turned on when the indicator is illuminated. 13. Chassis Ground Terminal. Connect to the facility ground with the Ground Cable (#15). 14. Main Input Power Cables. Connect appropriate power to test set. 15. Ground Cable. Connect between the Chassis Ground Terminal and the facility ground. 16. Output Stabs. Connection to breaker under test.

11 DESCRIPTION (Continued) Continuity Binding Post. For use in contact mode, i.e., normally closed or normally open. 18. Regulator Speed Control. Controls vernier speed.

12 8-1 CONTROLLER DESCRIPTION The controller consists of a laptop-style portable computer with floppy and hard disk drives, a fullfeatured keyboard, and an EGA compatible gas-plasma screen. A custom-designed interface card installed in the computer, combined with an I/O card mounted in the test set, allows all measurement and control functions. The disk-based software allows very simple modifications and upgrades in the future. Binding posts for continuity operating modes are provided near the output stabs. A power switch and indicators for input power, control power, interlock, and output ON/OFF status are located on the control panel. The main menu for controller operation indicates the present date and time, as well as the test type, and circuit breaker information, including serial number, and phase being tested. Test mode is displayed, including MOM/MAINT operation, N.O./N.C./C.L. mode, and normal/autojog/current hold modes. Coarse tap, vernier position, range, and output connection are displayed. The output status, continuity, breaker status, and temperature are indicated. Measurement data, consisting of continuous and pulse current, and pulse time, are displayed. A menu of commands is accessible by means of moving the cursor and pressing the ENTER key, or by pressing the key letter for the command. The commands available in the main menu, with their associated key letters, are: 1. B (Base): Selects timer units in seconds or cycles 2. W (Waveform): Selects waveform analysis 3. R (Range): Selects the currentmeter range 4. E (Edit): Advances to test configuration menu for editing the test and breaker data 5. Q (Quit): Exits run menu 6. T (Tap Increase): Selects Tap ALT & T (Tap Decrease): Selects Tap H (Help): Displays help menu 9. I or + (Init): Initiates the output 10. S or (Stop): Stops the test in progress 11. Z (Zero): Resets currentmeter, timer, and clears interlock 12. L (Latch): Selects momentary or maintained output mode 13. C (Contacts): Selects normally open, normally closed or current latch mode 14. M (Mode): Selects Normal Test Mode, JogH Mode, Curr Hold Mode or JogT Mode 15. PG UP (Raise): Raises variable transformer while depressed 16. PG DN (Lower): Lowers variable transformer while depressed 17. Y (type): Type of test Long Time, Short Time and Instantaneous 18. P (Pole): Pole of breaker under test A,B or C

13 9-1 MAIN OPERATING SCREEN Timer: Displays the trip time in either seconds or cycles. This may be selected before or after the test, using the B key. 2. Currentmeter Range: Displays the currentmeter range. The proper range for the test current should be selected before the output is initiated. Select the lowest range that is greater than the maximum current expected for the test. If too low a range is selected, the controller will autorange to a correct value, but the accuracy of the reading will be impaired for the first pulse of current from the output of test set. If too high a range is selected, the accuracy will be diminished, and current flow may not be detected properly in current latch mode causing the test set to turn off. Select using the R key. 3. Output Currentmeter: Displays the current to the device under test. 4. Temperature Meter: Displays the primary winding temperature of the high current transformer, T305. Must be less than 75 degrees C. 5. Date/Time: Displays the current date and time derived from the computer s clock. 6. Circuit Breaker Test Type: Displays the type of test being performed, Long Time, Short Time or Instantaneous. Select using the Y key. Displays the pole of breaker under test, A, B or C. Select using the P key.

14 9-2 MAIN OPERATING SCREEN (Continued) 7. Output Tap Indicator: Displays the current tap selection. Press T to increment; press ALT & T to decrement. Each keystroke changes the tap setting by one. 8. Vernier Position Indicator: Displays the position of the variable transformer wiper arm as a percent of maximum. The output voltage range of the variable transformer varies from minimum to maximum for each tap resulting in a continuously variable total output voltage. Depressing the PG UP key raises the variable transformer output; depressing the PG DN key lowers it. 9. Output, Breaker and Continuity Status Indicators: Displays the status of these items; active status indicated by reverse video. 10. Test Set Status: Displays the current status of the test set. READY: Will allow initiation of a test. RUN: Test in progress. RESET: Test is completed; awaiting system reset (Press Z). INTLK: Interlock is not cleared. Press Z to clear. OVER TEMP: HC Transformer temperature too high. Allow transformer to cool. 11. Output Latch Indicator: Displays the MOMENTARY or MAINTAINED condition of the Output Latch Mode. In MAINTAINED, the output latches ON when the I (INIT) key is pressed and remains on until the current stops, the contacts change state, or another key is pressed. In MOMENTARY, the I key must be depressed and held to keep the output ON. 12. Mode/Output Connection: Displays the current operating mode of the test set with the parameters as defined by the EDIT screen. Modes include NORMAL, JOGH, CURR HOLD, and JOGT. Selected by pressing the M key. See Section 13 for a description of each mode. 13. Contact Mode Indicator: Displays the current Contact Mode: NORM OPEN, NORM CLOSED, or CURR LATCH. Selected by pressing the C key. In the Contact Modes, the Continuity Binding Posts are connected to an unused pole or an auxiliary contact of the circuit breaker under test. Important: Under no circumstances should the Continuity Binding Posts be connected to a pole or contact which is being tested. Normally closed refers to a contact which is closed when the main contacts are closed. The output may be initiated only when the circuit breaker contacts are in the position indicating that the main contacts are closed. Current latch mode must be used for single pole breakers and may be used for any other series connected breakers. Contact modes must be used to test shunt trip breakers. 14. WAVEFORM: Selects waveform analysis. Also selected by pressing W. See Section 13-4 for more detailed explanation. This displays the waveform of the current test and allows viewing of portions of the waveform. 15. EDIT: Advances to test configuration menu for editing the test and breaker data. Also selected by pressing E. See Section 10-1 for detailed explanation. 16. QUIT: Exits run menu. Also selected by pressing Q. You will be prompted if you want to print and save the results. Press Y if you wish to save the results to the database or to print the results.

15 EDIT SCREEN 10-1 Audible Continuity: When enabled, the computer beeps while continuity is sensed on the contacts binding posts. The indicator says CONT'Y ON anytime there is continuity. Connection Mode: Parallel or series connection on output of test set. Current Hold Amps: Used in Hold mode. Set to desired output current. Current Hold Tol: Set to approximately 1% of Current Hold Amps. Speed control knob on front panel may be adjusted to prevent the variable transformer from hunting. Jog Off Seconds: This selects how long the variable transformer will raise with the output of the test set de-energized. Jog On Cycles: This will select how many cycles the test set will apply current on the output of the test set. Maximum setting 9.9. Maximum Amperes: specify the maximum current allowed on the output of the test set for the breaker under test. If set to zero, maximum amp protection will be disabled. Maximum Seconds: specify the maximum amount of time the output will be energized. If set to zero, maximum second protection will be disabled. Maximum Cycles: Specify the maximum amount of cycles the output will be energized. If set to zero, maximum cycle protection will be disabled. Phase Angle: Set to desired setting, used for repeatability of tests. Test Mode: Change test modes. Also change on main operating screen by pressing M until desired mode of test is selected. Other information from breaker under test. Pressing PG UP or PG DN will bring you to resistance measurement information. Input all relevant data.

16 11-1 INITIAL TEST SCREEN System Setup: The SETUP screen may only be accessed if the proper password was typed on the command line when the program was run. The information in this menu includes calibration factors, sampling rates, and analog to digital converter calibration factors. This information is preset during calibration and normally does not need readjusted. Test Configuration: Also accessible from main operating screen by pressing E for the edit menu. See Section Perform a Test: Access to main operating screen. See Section 9-1. View/Print Results: View or print previously stored test data. Database Maintenance: Manage test data. Quit: Quits program and returns to DOS prompt.

17 12-1 SYSTEM SETUP The SETUP screen may only be accessed if the proper password was typed on the command line when the program was run. The information in this menu includes calibration factors, sampling rates, and analog to digital converter calibration factors. This information is preset during calibration and normally does not need readjusted. The SETUP screen displays data in reverse video, and starts with the edit cursor in the first selection. The <UP> and <DOWN> cursor keys move the edit cursor to successive data, which is edited as alphanumeric strings, with no range checking. Data that is out of range can produce strange results. The function and range of values for SETUP data are as follows: DESCRIPTION MIN NORM MAX PURPOSE Average Sample Number of Counts for Average Cal Factor AC AC Calibration Multiplier Cal Factor DC DC 1kA Calibration Multiplier Cal Factor DC DC 10kA Calibration Multiplier Clock Factor System Timer Speedup Divisor Debug Low Level Debug Flag Freq AC Line Frequency Licensee Software Owner ID Max ADC Maximum ADC Waveform Data Queue Max CL Delay Msec Wait For Curr After INIT Max Data Maximum test results Max Intlk ADC Interlock Max Threshold Max Sample Maximum ADC Sample Queue Max Tap Maximum Test Set Coarse Taps Max Temp Maximum Temperature Degrees C Max Vern Pos Maximum Vernier Position % CONTROLLER OPERATION AND PROGRAMMING DESCRIPTION MIN NORM MAX PURPOSE Meas Mode 'A' 'A' 'D' AC or DC Measurement Modes Min Intlk ADC Interlock Min Threshold Min Threshold Current Sense Threshold Counts Min Vern Pos Minimum Vernier Position % Print Filename - "PRN" - Print Output File Name Range Delay Msec Wait After Range Change Set Impedance Output Volts/Short Circuit Amps Spline Samples Waveform Smoothing Function Tap Mode = BUCK/BOOST 1 = BOOST 2 = SEQUENTIAL (DC SET) 3 = HC100 Tap Volts Open Ckt Voltage Per Tap Vernier Factor ADC Counts For Vernier Span Vernier Offset ADC Counts At Vernier 0% Vernier Volts Open Ckt Vo ltage For Vernier

18 SYSTEM SETUP (Continued) 12-2 System Configuration The CONFIG screen displays the data in reverse video, and uses a moving reverse video cursor on the description. When <ENTER> is pressed, the selected data is either toggled among its valid conditions, or the data may be edited as alphanumeric information. If the data is numeric, it is evaluated and compared to preset limits. If outside the limits, an ERROR message is displayed, and the original value reappears for editing. If <ESC> is pressed, the edit of the selected data is terminated, and the original value is restored. If the edited text (e.g., "X123"), evaluates to zero, and zero is a valid value, zero will be displayed. Function and range of values for editable CONFIG data are as follows: DESCRIPTION MIN NORM MAX PURPOSE Audible Mode 'A' 'S' 'S' Audible or Silent Continuity Connection Mode SER, S/P, or PAR Connection Contacts Mode 'C' 'I' 'O' NC, CL, or NO modes Current Hold Current Hold Amperes Hold Tolerance Tolerance +/- For Hold Jog Off Seconds Off Time (Vern Raise) for JOG Jog On Cycles On Time for JOG mode Maximum Amperes Maximum Current Allowed Maximum Seconds Maximum Seconds ON Allowed Maximum Cycles Maximum Cycles ON Allowed Phase Angle Initial Firing Phase Angle Test Mode 'J' 'N' 'H' Jog, Normal, or Hold Modes Technician Technician ID Location Circuit Breaker Location Code Customer Circuit Breaker Customer Code Serial Number Circuit Breaker Serial Number Manufacturer Circuit Breaker Mfgr Code Type Circuit Breaker Type Frame Circuit Breaker Frame Size CONTROLLER OPERATION AND PROGRAMMING DESCRIPTION MIN NORM MAX PURPOSE Rated Amperes Circuit Breaker Rating Pole 'A' 'A' 'G' Pole Under Test or GFI Test Type Test Description Voltmeter Option The voltmeter option provides convenient panel display of input voltage, output voltage, or an external voltage connected to the binding posts above the output panel. The value displayed is TRUE RMS VOLTS AC. To use, simply select the desired mode (INPUT, OUTPUT, or EXTERNAL) and select the voltage range (2, 20, 200, 600 volts AC) best suited for the voltage being measured. If the voltage level is unknown, start with the highest range. If the display reads less than 200 counts, switch to the next lower range for greater accuracy.

19 13-1 TESTING MODES Basic Operation Before initiating the output, it is important to set up the controller for the proper mode of operation. First, select normally open (NO), normally closed (NC), or current latch mode. The mode is indicated in the TEST MODE box. The contact modes are selected sequentially by pressing the C key or moving the highlight bar to this selection and pressing the ENTER key. For most purposes, CURRENT LATCH operation is most convenient. MOMENTARY mode is best used at first. The MAINTAINED mode, toggled with the L (Latch) key, may be more convenient for long duration testing. Before initiating a test, the status indicator in the center of the screen must read READY, indicating breaker CLOSED, output OFF, and interlock CLEAR. To Test Circuit Breaker 1. Connect breaker under test to output of test set. 2. Energize test set as described in Section Select PERFORM A TEST on the computer screen. 4. Press Z on keyboard to clear INTLK. 5. Energize main breaker CB1. 6. Ensure variac is at 0% and Tap 1 is selected. If the breaker under test has been tested in the past, variac and Tap may be run up to previous setting. PG DN runs variac down, PG UP runs variac up, T changes taps up, ALT & T changes taps down. 7. Press M to select the desired mode of operation. 8. Press E to access the EDIT screen and input test parameters. Normal Mode Testing All functions are controlled by the test technician. Select a range that is appropriate for the test current expected, and select a tap and vernier position that is likely to produce the desired current. Make sure that the circuit breaker is closed, select MOMENTARY and briefly press and release the I or + key to initiate the output. The current-meter will read the value of the current pulse, and the timer will read the duration in seconds or cycles as desired. To repeat the test, the currentmeter and timer must be reset by pressing Z. Jog to Hold Mode The JOG to HOLD mode is used to reach a desired breaker holding current with minimal breaker stress. Best if used just prior to HOLD Mode to reach desired current. Hold Mode Testing The HOLD Mode is used to maintain a given current value for a long period of time, for testing constant current breaker trip times.

20 TESTING MODES (Continued) 13-2 Jog to Trip Mode The Jog to Trip mode is used for determining the instantaneous trip point. JOG MODE TESTING Jog to Hold Mode While in the EDIT screen. (Section 10-1) 1. Select CURRENT HOLD AMPS. Input the desired output current from test set. 2. Select JOG OFF SECONDS. Input the time period that the variac will increase its output with the output of the test set de-energized. 3. Select JOG ON CYCLES. Input the duration in cycles for which the output of the test set will be energized. Maximum setting Select MAXIMUM AMPS. Input the maximum current allowed on the output of the test set for the breaker under test. If set to zero, MAXIMUM AMP protection will be disabled. 5. Select MAXIMUM SECONDS. Input the maximum amount of time the output will be energized. If set to zero, MAXIMUM SECONDS protection will be disabled. 6. Select MAXIMUM CYCLES. Input the maximum number of cycles for which the output will be energized. If set to zero, MAXIMUM CYCLE protection will be disabled. 7. Enter information for breaker under test. 8. Press ESC to return to the main operating screen. 9. Press M to change mode to JOGH. 10. Press C to select contact sense from NC, NO or CURRENT SENSE. If using NC or NO sense, connect the sense leads to the unused set of appropriate contacts on the breaker under test. 11. Press L to select MAINTAINED output. 12. Press R to select the current range. Select the appropriate range for the expected current. Test set will auto range if the output current is exceeded on the selected range. 13. Press Y to select the desired type of test being performed. Select LONG TIME or SHORT TIME. 14. Press P to select the pole of the breaker under test. Select A, B or C. 15. Press I or + to initiate the test set and begin the test. 16. Test set will automatically run up to the desired current output and turn off the output when this current is reached. Press S or - to stop the test at any time. Jog to Trip Mode To adjust the output of the test set to a specific current, the Jog Cycles should be set to a time long enough to get a good reading, but short enough to avoid overheating. A good value is 5. The JOG OFF Seconds should be long enough to move the vernier a reasonable amount, such as 1 to 2 seconds. The Max Current may be set to the current desired. Now, the test set may be initiated, and the unit will produce increasing pulses until the Max Current value is reached. Now the test mode may be switched to NORMAL or HOLD, and Max Current should be set to a value higher than expected. As an additional safety precaution, the maximum cycles should be set to slightly more than the JOG ON Cycles.

21 13-3 TESTING MODES (Continued) Set JOG ON Cycles to a value at least twice the expected instantaneous trip time and set Max Current to about 20% more than the maximum expected current. The vernier and taps will automatically increase the output until the circuit breaker trips; this is the instantaneous trip current. 1. Press M to change the mode to JOGT. 2. Press E to EDIT desired settings. (Section 10-1) 3. Select JOG OFF SECONDS. Input the time period that the variac will increase its output with the output of the test set de-energized. 4. Select JOG ON CYCLES. Input the duration in cycles for which the output of the test set will be energized. Maximum setting Select MAXIMUM AMPS. Input the maximum current allowed on the output of the test set for the breaker under test. If set to zero, maximum amp protection will be disabled. 6. Select MAXIMUM SECONDS. Input the maximum amount of time the output will be energized. If set to zero, MAXIMUM SECONDS protection will be disabled. 7. Select MAXIMUM CYCLES. Input the maximum number of cycles for which the output will be energized. If set to zero, MAXIMUM CYCLE protection will be disabled. 8. Press ESC to return to the main operating screen. 9. Press Y to type of test being performed. Select INSTANTANEOUS. 10. Press L to select MAINTAINED output. 11. Press R to select the current range. Select the appropriate range for the expected current. Test set will auto range if the output current is exceeded on the selected range. 12. Press P to select the pole of the breaker under test. Select A, B or C. 13. Press I or + to initiate the test set and begin the test. 14. Test set will automatically run until the breaker trips. Press S or - to stop the test at any time. Hold Mode Testing Set MAXIMUM CYCLES to zero or the test set will shut off on MAXIMUM CYCLES instead of MAXIMUM SECONDS. Set the HOLD CURRENT to the value desired, and set the HOLD TOLERANCE to the maximum variation that is allowable. If HOLD TOLERANCE is set too close, the test set will constantly adjust the variable transformer to try to hold the same current. If this occurs adjust the speed control knob on front panel down until the variable transformer movement is acceptable. If the desired current is obtained at a low tap and vernier position, it may be impossible to hold the set value accurately. Better results will be obtained by adding impedance (such as long cables), to raise the tap and vernier settings. Once the desired current is produced, the test set will hold that current until the breaker trips or the test is interrupted by the technician.

22 13-4 TESTING MODES (Continued) Testing in HOLD Mode 1. Press M to change mode to CURR HOLD. 2. Press E to change to the EDIT screen. (Section 10-1) 3. Select desired CURRENT HOLD AMPS. 4. Select desired CURRENT HOLD TOL. Set to approximately 1% of desired current. 5. Press ESC to return to the main operating screen. 6. Press Y to select the type of test to be performed. 7. Press I or + to begin test. 8. Test set will run until breaker under test trips. Press S or - to stop the test at any time. Waveform Analysis One of the advanced features of this test set is the ability to look at the waveform of the current that passed through the circuit breaker during the test. This can be particularly useful in instantaneous tests, where the current waveform can be badly distorted because of poor connections, DC offset or other problems, producing erroneous or misleading test results. After a test, and before the meters are reset (or zeroed), the first several seconds of the waveform may be observed by pressing W. The run menu will be saved and the screen will clear. A graph will appear, with a vertical axis in counts (maximum and minimum values for the 12 bit A/D), and a horizontal axis scaled in seconds, initially zero to about Each sample of the converter is plotted with an asterik (*), and the rough outline of the AC current waveform will be visible. Across the bottom of the screen is a menu of ten choices, which are selected by using the <LEFT> and <RIGHT> cursor keys. Their functions are as follows: <START START> <REF1 REF1> <REF2 REF2> <END END> SMOOTH QUIT Decrements the starting point of the display Increments the starting point of the display Decrements the position of the first reference line Increments the position of the first reference line Decrements the position of the second reference line Increments the position of the second reference line Halves total number of samples (Magnifies display) Doubles total number of samples (Contracts display) Connects samples with smooth line for easy viewing Erases display and returns to previous menu At the top of the screen are values for duration and RMS value. They correspond to the time interval between the REF1 and REF2 cursors, and the calculated true RMS value between them. Note that the reference cursors move outside the graph when their position cannot be displayed within the endpoints. If a cursor outside the endpoints is moved, it will first move to the end of the graph. The cursors move about 1/20 of the width of the graph, or one sample, depending on magnification.

23 13-5 TESTING MODES (Continued) Exiting Program and Saving Results After test is performed and while test results are still displayed press Q. Message will be displayed in lower left corner DO YOU WANT TO EXIT FROM THIS RUN MENU [Y/N]?. Pressing N will return to main operating screen. Pressing Y will display SAVE RESULTS TO DATABASE [Y/N]?. Pressing Y will save results to database. Pressing N will exit from main operating screen. To print or view saved results see Section Trademarks * "MS-DOS" is a trademark of Microsoft Corporation.

24 14-1 THEORY OF OPERATION All of the control functions originate from the computer controller, and all status data is reported back to the controller. In this way, when a test is initiated, the computer controller is in full control and can monitor and record all test data automatically hundreds of times per second. The operation of the computer controller is explained in the Controller Description section. Therefore, only the operation of the regulator and high current sections will be discussed at this time. Fuses and Circuit Breakers The first level of protection for the entire test set is two 400 Amp fuses, labeled F1 and F2. These fuses are connected directly to the 480 V input. The control section is protected by two 10 Amp fuses, F3 and F4, and the AC HI line is protected by a 15 Amp Fuse, F5. Two 3 Amp fuses, F11 and F12, protect the variable transformer motor drive controller. For the protection of the variable transformer regulator, there is a 50 Amp circuit breaker, CB2. The main power circuit breaker, CB1, 350 Amp, protects the high power section. Circuit breaker CB1 is also equipped with an undervoltage release coil that is connected to the interlock. This breaker must be reset the interlock opens. Interlock Circuit The interlock circuit will be held open by any one of several types of fault conditions. The indicator lamp on the control panel will light to indicate that the interlock is open and must be closed before the test can continue. The interlock can be held open by two basic sources, a failure to properly ground the test set or a computer detected fault. The ground fault is detected by relay K11. The computer controlled interlock relay K14A which is brought out to J12 on the I/O board can at any time open the interlock. The computer will hold the interlock open if any thermocouple closes or if the output bus temperature exceeds 75 degrees C. The computer also holds the interlock open during power on procedures to ensure safe and smooth operation. Triggering The SCR trigger board serves only one task, that is to turn on the SCR's. The computer controller sends the initiate output pulse to the trigger board when the output is to turn on and then terminates the signal when the output is to turn off. When the output is initiated, relay K6 is activated and the Output On indicator is lit. To prevent timing errors that may occur with this method, a phase reference signal is generated from the input voltage. This is done on the I/O board. The phase reference signal is optically isolated and conditioned to the form of a digital pulse. This signal is fed to the computer and used to control the turn on phase angle and increase accuracy. The various conditions that will turn the output on and off are explained in the Controller Description section. Taps and Vernier To achieve a variable output from zero to maximum, a system of taps and a motorized vernier are employed. Both the taps and vernier are controlled only through the computer. This can be done in several modes that are explained in the Controller Description section. The tap relays are controlled by relays K2, K3, K4 and K8 on the I/O board. These relays also provide isolation from the digital circuitry. The vernier is also controlled by relays on the I/O board. Relays K9 and K10 control both direction and start or stop. The vernier also has a variable resistor attached to its shaft that is used for position sensing. This position signal is fed to the computer as an analog signal called VER POS and is present on J11, pin 2.

25 14-2 THEORY OF OPERATION (Continued) Timer The timer is used to measure the width of the applied current pulse. The timer always starts at zero and begins counting at the start of the test. The timer can count either seconds or cycles and can be changed before or after the test. In the normally closed (NC) or normally open (NO) modes of operation, the timer will start when the computer sends the initiate signal to the trigger board and stops when either the contacts change state or the computer terminates the initiate signal. In the current latch mode, the timer will start when current flow is detected and stop when current flow stops. This mode must be used for single-phase breakers. Currentmeter To measure the output current, two precision current transformers are placed around the output bus. These transformers are wired in series for maximum output voltage and can measure up to 100,000 Amps. This voltage is fed to the interface board which is located inside the computer. The signal is then integrated and amplified. The gain of the amplifier is set by the computer depending on the range that is being used. The amplifier is a precision, differential instrumentation amplifier with a very high common mode rejection ratio for noise immunity. The signal is then fed to the analog-to-digital converter and then to the computer. The current is then calculated as an average of the current readings to reduce the effects of noise. Other than the initial adjustments of DC offset and common mode rejection there are no other adjustments that can be made on the interface board. The final adjustments are made from the system setup menu in the form of cal-factors. This procedure is explained in detail in the Calibration section of this manual. Continuity The continuity binding posts are located above the output stabs and can be used to indicate whether the breaker is open or closed. This is done by applying 15 VDC to an opto isolator through the binding posts. When the posts are shorted, the opto isolator conducts and the continuity signal becomes active. This signal is fed to the computer and is reported on the screen. The continuity signal is also used by the timer to detect when the breaker opens and the precise time to stop the timer.

26 15-1 TROUBLESHOOTING If the unit is suspected to be malfunctioning, refer to the following guide for testing the unit and isolating the problem. Some minor repairs can be made in the field. Otherwise, call the factory for assistance. NOTE: SOME TROUBLESHOOTING PROCEDURES REQUIRE OPENING THE TEST SET, WHICH HAS POTENTIALLY LETHAL VOLTAGES ON EXPOSED CONDUCTORS. ONLY FULLY EXPERIENCED TECHNICIANS SHOULD ATTEMPT TO PERFORM THESE PROCEDURES. IF YOU HAVE ANY DOUBTS, CONTACT THE FACTORY FOR ASSISTANCE. 1. Insufficient Output When the unit does not seem to yield sufficient current through a breaker, the problem could be one of the following: a. Insufficient power supply. b. Loose connection in unit. c. Incorrect controller range or output tap configuration setting. d. Poor connections or excessive contact resistance at the breaker. Check cable or stab connections. e. Improper selection of output configuration. 2. Cannot Energize Main Circuit Breaker a. Main breaker can only be energized when the interlock is clear. Check the status on the controller. b. Remove power and inspect the under voltage release mechanism for the main breaker. 3. No Output Current When Initiated a. Check to see that main circuit breaker is on. b. Verify that the breaker is reset and not in the trip position. c. Verify continuity of the breaker contacts. d. Check all internal fuses of the test set. e. Check LED indicator on SCR trigger board. It should light when initiated. f. Check voltage out of SCR's when initiated.

27 16-1 CALIBRATION Calibration is recommended at one year intervals. A High Current standard is required. The standard should have an accuracy of +/- 1/4%. A digital millivolt meter with an accuracy of +/-.1 mv is also required. Use the following procedure once for each range of the test set. 1. Connect the standard across the output and the millivolt meter to the metering points on the standard. 2. Initiate the unit at several current values within the range and record the current measured on the standard meter and the test set. 3. If the unit is out of tolerance, the new cal-factor can be calculated as follows: (Test Set Reading) - (Standard Reading) % Error = (Full Scale) Calibration is accomplished by changing the cal-factor. If the observed error was -.01, multiply the appropriate cal-factor by 1.01; if the error was +.01, then multiply the cal-factor by.99. The cal-factor should be adjusted until readings are within tolerance. If the cal factor must be adjusted by more than 5%, internal adjustments are required which require factory assistance. 4. To access the cal-factor, the proper password must be entered. Once the password is entered, enter the system setup menu and change the appropriate cal-factor. 5. The time base of the digital timer is derived from the computer s internal crystal controlled oscillator and does not require periodic recalibration.

28 17-1 SERVICE AND MAINTENANCE Periodic Maintenance Periodically, the following maintenance should be performed on the unit to insure that the test set maintains proper output and operation. 1. Disconnect the unit from the input supply. 2. Open all panels on the unit. Using a vacuum cleaner, remove any dust and debris from the interior of the unit, including wiring and components. Use caution when cleaning printed circuit boards and small components. 3. Tighten all connections, especially those on transformers, contactors, and input and output connectors. Special attention should be given to tightening all connections in the high current section every six months or less. If the unit is transported often, this procedure should be performed more often. 4. Remove the front cover on the main circuit breaker and clean the main contacts. Replace the contacts if they are beyond repair. 5. Inspect the powerstat brushes for wear. Replace if they appear worn. 6. Perform basic operational tests to determine proper operation. The above maintenance procedures should be performed at least once a year depending on the amount of use and the environment.

29 RETURNED MATERIAL A-1 If for any reason it should become necessary to return this equipment to the factory, the Service Department of Phenix Technologies, Inc. must be given the following information: Name Plate Information Model Number Serial Number Reason for Return Cause of Defect If Phenix Technologies, Inc. deems return of the part appropriate, it will then issue an "Authorization for Return". If return is not deemed advisable, other inspection arrangements will be made. NOTE: Material received at this plant without the proper authorization shall be held as "Customer's Property" with no service until such time as the proper steps have been taken. Your cooperation is requested in order to ensure prompt service.

30 B-1 WARRANTY Phenix Technologies, Inc. warrants to the original purchaser of any new merchandise that the merchandise is free from defects in material and workmanship under normal use and service for a period of 18 months from the date of shipment. The obligation of Phenix Technologies, Inc. under this warranty is limited, in its exclusive option, to repair, replace, or issue credit for parts or materials which prove to be defective, and is subject to purchaser's compliance with the Phenix Technologies, Inc. warranty claim procedure as set forth below. The happening of any one or more of the following events will serve to void this warranty and defect or damage resulting therefrom is specifically excluded from warranty coverage: A. Defects due to accident, negligence, alteration, modification, faulty installation by purchaser or purchaser's agents or employees, abuse, or misuse. B. Attempted or actual dismantling, disassembling, service or repair by any person, firm, or corporation not specifically authorized in writing by Phenix Technologies, Inc. C. Defects caused by or due to handling by carrier, or incurred during shipment, transshipment, or other move. This warranty covers only those parts and/or materials deemed by Phenix Technologies, Inc. to be defective within the meaning of this warranty. The liability of Phenix Technologies, Inc. shall be limited to the repair, replacement, or issuance of credit for parts deemed defective within the meaning of this warranty. Costs incurred by purchaser for labor or other expenses incidental to the inspection, repair, replacement, or issuance of credit for such parts and/or material shall be the sole responsibility of purchaser. This warranty shall not apply to any accessories, parts, or materials not manufactured or supplied by Phenix Technologies, Inc. and if, in the sole discretion of Phenix Technologies, Inc., purchaser's claim relates to any materials or workmanship manufactured or performed by the supplier of a component part, or of the manufacturer of a device of which the defective part is a component, Phenix Technologies, Inc. reserves the right to disclaim liability under this warranty and to direct that the purchaser deal directly with such supplier or manufacturer. Phenix Technologies, Inc. agrees to assist the purchaser in processing or settling any such claims without prejudicing its position as to liability. Warranty Claim Procedure Compliance with the following warranty claim procedure is a condition precedent to the obligation of Phenix Technologies, Inc. under this warranty. A. Purchaser must notify Phenix Technologies, Inc. in writing by certified or registered mail, of the defect claimed within 18 months after date of original shipment. Said notice shall describe in detail the defect, the defective parts and/or part, and the alleged cause of defect. B. At the exclusive option of Phenix Technologies, Inc., purchaser shall dismantle or disassemble at purchaser's cost and expense and shall ship the defective part or material prepaid to and from Phenix Technologies, Inc., Accident, Maryland 21520, for inspection, or permit an authorized service representative of Phenix Technologies, Inc. to inspect the defective part or material at purchaser's premises. If Phenix Technologies, Inc. shall inspect the part or material at the purchaser's premises, purchaser shall provide facilities for, and at purchaser's cost and expense, dismantle, disassemble, or otherwise make accessible the subject part of material whether or not same is a component of or installed in a device other than that manufactured or supplied by Phenix Technologies, Inc. If disclosure shows that the defect is not one for which Phenix

AC CIRCUIT BREAKER TEST SET OPERATOR S MANUAL Version 2.0 MODEL HC-75C (480V) SERIAL NUMBER:

AC CIRCUIT BREAKER TEST SET OPERATOR S MANUAL Version 2.0 MODEL HC-75C (480V) SERIAL NUMBER: AC CIRCUIT BREAKER TEST SET OPERATOR S MANUAL Version 2.0 MODEL HC-75C (480V) SERIAL NUMBER: TABLE OF CONTENTS SECTION NUMBER DANGER TITLE PAGE SPECIFICATIONS 1 INTRODUCTION 2 SAFETY 3 INSTALLATION INSTRUCTIONS

More information

AC CIRCUIT BREAKER TEST SET OPERATOR S MANUAL Version 3.1

AC CIRCUIT BREAKER TEST SET OPERATOR S MANUAL Version 3.1 AC CIRCUIT BREAKER TEST SET OPERATOR S MANUAL Version 3.1 MODEL HC-75C (480V) SERIAL NUMBER: Phenix Technologies, Inc. 75 Speicher Drive Accident, MD 21520 USA Copyright Phenix Technologies, Inc. Rev 1/18/2018

More information

Phenix Technologies, Inc 75 Speicher Drive Accident, MD 21520

Phenix Technologies, Inc 75 Speicher Drive Accident, MD 21520 AC CIRCUIT BREAKER TEST SET OPERATOR S MANUAL Version 2.0 MODEL Phenix Technologies, Inc 75 Speicher Drive Accident, MD 21520 Copyright Phenix Technologies, Inc. Rev 6/5/2017 nab TABLE OF CONTENTS SECTION

More information

Phenix Technologies Inc. 75 Speicher Drive Accident, Maryland 21520

Phenix Technologies Inc. 75 Speicher Drive Accident, Maryland 21520 USER S MANUAL PORTABLE HIGH CURRENT TEST SET MODEL NUMBER Version 4.1 Phenix Technologies Inc. 75 Speicher Drive Accident, Maryland 21520 Copyright Phenix Technologies, Inc. Rev 11/14/2016 nab TABLE OF

More information

AirTest Model CN9000 Series Sensor Controller

AirTest Model CN9000 Series Sensor Controller AirTest Model CN9000 Series Sensor Controller AirTest Model CN9000 Series Sensor Controller THEORY OF OPERATION A basic CN9000 configuration consists of Input/Process/Display combination modules, a 3 relay

More information

OPERATING AND SERVICE MANUAL. Universal Interface Device 47

OPERATING AND SERVICE MANUAL. Universal Interface Device 47 OPERATING AND SERVICE MANUAL Universal Interface Device 47 MAGNA-POWER ELECTRONICS, INC. 39 ROYAL ROAD, FLEMINGTON, NJ 08822 May 24, 2012 SAFETY NOTICE Universal Interface Device 47 (UID46) connects

More information

MAINTENANCE MANUAL. EDACS REDUNDANT POWER SUPPLY SYSTEM 350A1441P1 and P2 POWER MODULE CHASSIS 350A1441P3, P4, AND P5 POWER MODULES TABLE OF CONTENTS

MAINTENANCE MANUAL. EDACS REDUNDANT POWER SUPPLY SYSTEM 350A1441P1 and P2 POWER MODULE CHASSIS 350A1441P3, P4, AND P5 POWER MODULES TABLE OF CONTENTS MAINTENANCE MANUAL EDACS REDUNDANT POWER SUPPLY SYSTEM 350A1441P1 and P2 POWER MODULE CHASSIS 350A1441P3, P4, AND P5 POWER MODULES TABLE OF CONTENTS SPECIFICATIONS*... 2 INTRODUCTION... 3 DESCRIPTION...

More information

PHASETRONICS. SCR Power Control Specialists. EP1 Series Power Control Single Phase SCR Amps OPERATION & SERVICE MANUAL

PHASETRONICS. SCR Power Control Specialists. EP1 Series Power Control Single Phase SCR Amps OPERATION & SERVICE MANUAL PHASETRONICS Specialists EP1 Series Power Control Single Phase SCR 10-50 Amps OPERATION & SERVICE MANUAL Phasetronics, Inc. P.O. Box 5988 1600 Sunshine Drive Clearwater, FL 33765 (727)573-1900 FAX(727)573-1803

More information

Contents. HP E1586A Rack Mount Terminal Panel User s Manual

Contents. HP E1586A Rack Mount Terminal Panel User s Manual Contents HP E1586A Rack Mount Terminal Panel User s Manual Description... 5 Connecting to VXIbus Instruments... 5 Interconnect Cables... 5 Terminal Block Connections... 6 Using the Terminal Panel for Reference

More information

OPERATING AND SERVICE MANUAL. Universal Interface Device 47

OPERATING AND SERVICE MANUAL. Universal Interface Device 47 OPERATING AND SERVICE MANUAL Universal Interface Device 47 MAGNA-POWER ELECTRONICS, INC. 39 ROYAL ROAD, FLEMINGTON, NJ 08822 May 24, 202 SAFETY NOTICE Universal Interface Device 47 (UID47) connects two

More information

MYRIAD QLC 4-CHANNEL MONITOR/CONTROLLER INSTRUCTION MANUAL

MYRIAD QLC 4-CHANNEL MONITOR/CONTROLLER INSTRUCTION MANUAL MYRIAD QLC 4-CHANNEL MONITOR/CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM MYR QLC MANUAL 013114 2 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features WIRING

More information

TC100 Precision Thermocouple Calibrator

TC100 Precision Thermocouple Calibrator TC100 Precision Thermocouple Calibrator Table of Contents 1. Introduction.....................1 2. Accessories....................5 3. Set-Up Basic/Advanced...........5 4. Operating Procedure.............6

More information

S Reclosers Form 4C Microprocessor-Based Recloser Control KME (120V) and KME (240V) Analog Current Metering Accessory Design II

S Reclosers Form 4C Microprocessor-Based Recloser Control KME (120V) and KME (240V) Analog Current Metering Accessory Design II Reclosers Form 4C Microprocessor-Based Recloser Control KME4-88-1 (120V) and KME4-88-2 (240V) Analog Current Metering Accessory Design II Service Information S280-77-6 961015, 961016 Figure 1. Kyle KME4-88

More information

ATE QUICK START GUIDE. KEPCO An ISO 9001 Company. VOLTAGE/CURRENT STABILIZED POWER SUPPLY I INTRODUCTION

ATE QUICK START GUIDE. KEPCO An ISO 9001 Company. VOLTAGE/CURRENT STABILIZED POWER SUPPLY I INTRODUCTION QUICK START GUIDE KEPCO An ISO 9001 Company. ATE VOLTAGE/CURRENT STABILIZED POWER SUPPLY I INTRODUCTION 1.1. SCOPE OF MANUAL. This Quick Start Guide covers simple installation and local operation of the

More information

R & D SPECIALTIES SERIES 100 RO CONTROLLER USERS MANUAL. 2004, by R & D Specialties, Inc. All Rights Reserved.

R & D SPECIALTIES SERIES 100 RO CONTROLLER USERS MANUAL. 2004, by R & D Specialties, Inc. All Rights Reserved. R & D SPECIALTIES 2004, by R & D Specialties, Inc. All Rights Reserved. No part of this document may be copied or reproduced in any form or by any means without the prior written permission of R & D Specialties.

More information

MICRO GROW GREENHOUSE SYSTEMS, INC

MICRO GROW GREENHOUSE SYSTEMS, INC MICRO GROW GREENHOUSE SYSTEMS, INC 4065 ZEVO DR., UNIT B-, TEMECULA, CA 9590 PHONE (95) 96-3340 FAX (95) 96-3350 www.microgrow.com Revision., -09-0 Growstat Control Series INSTALLATION PROCEDURES SERIES

More information

Plus-X 300. Installation and Operation Manual

Plus-X 300. Installation and Operation Manual Plus-X 300 Installation and Operation Manual Table of Contents Introduction... 1 Compatibility... 1 Installation... 1 Configuration... 2 Operation... 5 Getting Help... 6 Warranty... 6 Appendix A: Specifications...

More information

SMD Series Integrated Stepper Driver and Motor Revision 1.3

SMD Series Integrated Stepper Driver and Motor Revision 1.3 The AMCI Integrated Stepper Motor and Microstepping Drive Combination represents the future of Stepper Motor Control applications. The SMD is a self-contained stepper motor and driver package, capable

More information

ControlKeeper 4. General Information. Connecting Relay Loads. Installation Sheet. Getting Started. Power Supply Wiring. Mounting the Cabinet

ControlKeeper 4. General Information. Connecting Relay Loads. Installation Sheet. Getting Started. Power Supply Wiring. Mounting the Cabinet General Information ControlKeeper 4 Installation Sheet Model# CK4-120NO- Model# CK4-277NO The ControlKeeper-4 model is shipped in one package and is configured with either a 120V or a 277V transformer.

More information

PLT. Breaker-based transfer switch open, closed or soft transition amps. Specification sheet

PLT. Breaker-based transfer switch open, closed or soft transition amps. Specification sheet Specification sheet PLT Breaker-based transfer switch open, closed or soft transition 800 3000 amps Description PLT breaker-based transfer switches are designed for operation and switching of electrical

More information

PIECAL 322 Automated Thermocouple Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

PIECAL 322 Automated Thermocouple Calibrator Operating Instructions. Product Description. Practical Instrument Electronics PIECAL 322 Automated Thermocouple Calibrator Operating Instructions Product Description Easy to use With the PIECAL 322-1 you can check & calibrate all your thermocouple instruments and measure thermocouple

More information

Relay Board. User Manual. 1.0, Oct 2013

Relay Board. User Manual. 1.0, Oct 2013 Relay Board User Manual 1.0, Oct 2013 This work is licensed under the Creative Commons Attribution-Share Alike 2.5 India License. To view a copy of this license, visit http://creativecommons.org/licenses/by-sa/2.5/in/

More information

DCS-E 1kW Series, DLM-E 3kW & 4kW Power Supplies

DCS-E 1kW Series, DLM-E 3kW & 4kW Power Supplies DCS-E 1kW Series, DLM-E 3kW & 4kW Power Supplies M51A Option: Isolated Analog Programming Manual Power Supplies Elgar Electronics Corporation 9250 Brown Deer Road San Diego, CA 92121-2294 1-800-73ELGAR

More information

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER SAFETY INFORMATION The following safety information must be observed to ensure maximum personal safety during the operation of this meter: This meter is

More information

INSTRUCTION and OPERATIONS MANUAL. for

INSTRUCTION and OPERATIONS MANUAL. for INSTRUCTION and OPERATIONS MANUAL for CAB SIGNAL MASTER MODEL NUMBER 15100-00 CAUTION Be sure to read and become thoroughly familiar with the entire contents of this manual before attempting to operate

More information

EPS Power Supply

EPS Power Supply EPS - 600 Power Supply Installation and Operation Manual Version 1.0 *This instrument is intended for laboratory use only Index A. Important Notice ----------------------------------------------------------------

More information

PCM-7140 Pulsed Current Source Operation Manual

PCM-7140 Pulsed Current Source Operation Manual PCM-7140 Pulsed Current Source Operation Manual Directed Energy, Inc. 1609 Oakridge Dr., Suite 100, Fort Collins, CO 80525 (970) 493-1901 sales@ixyscolorado.com www.ixyscolorado.com Manual Document 7650-0031

More information

Voltage Dip-Proofing Inverter

Voltage Dip-Proofing Inverter Voltage Dip-Proofing Inverter DPI54S / 54L Series Models 120V & 208/230 50/60Hz Contents: Introduction...3 Theory of operation...3 Specifications...5 Up-time considerations...6 Installation Guide...7 Test

More information

Install Motor Controller User Manual

Install Motor Controller User Manual Property of Motion Laboratories, Inc. Install Motor Controller User Manual 2014 Motion Laboratories, Inc. Created By: Michael Shaw Approved By: John Coppolecchia Page: 1 Page: 2 2014 Motion Laboratories,

More information

OPERATING PROCEDURES for the UPS-S2 Universal Power Supply Series 2

OPERATING PROCEDURES for the UPS-S2 Universal Power Supply Series 2 OPERATING PROCEDURES for the Universal Power Supply Series 2 Vanguard Instruments Co., Inc. 1520 S. Hellman Ave. Ontario, California 91761 TEL: 909-923-9390 November 2013 FAX: 909-923-9391 REV. 2 SAFETY

More information

MCCB-500 MOLDED-CASE CIRCUIT BREAKER TESTER

MCCB-500 MOLDED-CASE CIRCUIT BREAKER TESTER MCCB-500 MOLDED-CASE CIRCUIT BREAKER TESTER USER S MANUAL Vanguard Instruments Company, Inc. 1520 S. Hellman Ave. Ontario, California 91761, USA TEL: (909) 923-9390 FAX: (909) 923-9391 January 2015 Revision

More information

INSTRUCTION MANUAL. Model Dual Input RTD Thermometer. Measures two temperatures simultaneously. Dual RTD probe inputs

INSTRUCTION MANUAL. Model Dual Input RTD Thermometer. Measures two temperatures simultaneously. Dual RTD probe inputs INSTRUCTION MANUAL Model 421504 Dual Input RTD Thermometer Measures two temperatures simultaneously Dual RTD probe inputs Clock and Elapsed Timer functions Special functions include Data Hold, MIN/MAX/AVG,

More information

Integrated Stepper Drive & Motor

Integrated Stepper Drive & Motor SMD23 Integrated Stepper Drive & Motor Manual #: 940-0S050 User Manual AMCI Motion Control Products Important User Information The products and application data described in this manual are useful in a

More information

Pulse LED Instruction Guide

Pulse LED Instruction Guide PARTS LIST Light Fixture Aquarium Frame Mounts Instruction Guide WARNING: To guard against injury, basic precautions should be observed, including the following: A) READ AND FOLLOW ALL SAFETY INSTRUCTIONS.

More information

INSTALLATION AND MAINTENANCE MANUAL FOR GROUND MONITOR GM-300

INSTALLATION AND MAINTENANCE MANUAL FOR GROUND MONITOR GM-300 INSTALLATION AND MAINTENANCE MANUAL FOR GROUND MONITOR GM-300 COPYRIGHT 1988 American Mine Research, Inc Manual Part Number 160-0094 Original: 07-19-88 Revised: 03-28-90 NOT TO BE CHANGED WITHOUT MSHA

More information

DFI 7000 DIGITAL FORCE INDICATOR USER S GUIDE 1.0 INTRODUCTION 2.0 DESCRIPTION. 2.1 Configuration. 2.2 Specifications

DFI 7000 DIGITAL FORCE INDICATOR USER S GUIDE 1.0 INTRODUCTION 2.0 DESCRIPTION. 2.1 Configuration. 2.2 Specifications DFI 7000 DIGITAL FORCE INDICATOR USER S GUIDE 1.0 INTRODUCTION The DFI 7000 Signal Conditioner-Indicator is a complete 4-1/2-digit signal conditioner and indicator housed in a 1/8 DIN case. This unit provides

More information

Sonic Ruptor 400. User Manual

Sonic Ruptor 400. User Manual Sonic Ruptor 400 User Manual Data herein has been verified and validated. It is believed adequate for the intended use of the instrument. If the instrument or procedures are used for purposes over and

More information

Resolver to Digital Expansion Board

Resolver to Digital Expansion Board Resolver to Digital Expansion Board Catalog No. EXB009A01 Installation and Operating Manual 6/98 MN1313 Table of Contents Section 1 General Information............................. 1-1 Introduction....................................

More information

Product Description. S&C 5800 Series Automatic Switch Control S&C ELECTRIC COMPANY TABLE OF CONTENTS

Product Description. S&C 5800 Series Automatic Switch Control S&C ELECTRIC COMPANY TABLE OF CONTENTS S&C 5800 Series Automatic Switch Control Product Description TABLE OF CONTENTS Section Page INTRODUCTION Qualified Persons............................... 2 Read this Instruction Sheet.......................

More information

EASON TECHNOLOGY. IO8 & IO24 Break-Out Module

EASON TECHNOLOGY. IO8 & IO24 Break-Out Module EASON TECHNOLOGY IO8 & IO24 Break-Out Module p/n 50-00180-01 Revision1.2 Eason Technology, Inc. 7975 Cameron Dr. Bldg 300 Windsor, CA 95492 Phone (707) 837-0120 FAX (707) 837-2742 http://www.eason.com

More information

SA SERIES Arlingate Lane, Columbus, Ohio (614)

SA SERIES Arlingate Lane, Columbus, Ohio (614) SA SERIES 2080 Arlingate Lane, Columbus, Ohio 43228 (614) 850-5000 Sensotec, Inc. SA Series Instruction Manual Sensotec Part Number: 008-0131 - 00 Rev. A: January 1996 Copyright Notice: 2080 Arlingate

More information

Innovative Circuit Technology Ltd.

Innovative Circuit Technology Ltd. Innovative Circuit Technology Ltd. Pro Series DC Power Supply INSTRUCTION MANUAL 855-343-001 Models: ICT690-12S/ICT690-12SB ICT690-24S/ICT690-24SB ICT690-48S/ICT690-48SB ICT1190-12S/ICT1190-12SB ICT1190-24S/ICT1190-24SB

More information

4300 WINDFERN RD #100 - HOUSTON TX VOICE (713) FAX (713) web: IMPORTANT!!!

4300 WINDFERN RD #100 - HOUSTON TX VOICE (713) FAX (713) web:   IMPORTANT!!! 4300 WINDFERN RD #100 - HOUSTON TX 77041-8943 VOICE (713) 973-6905 - FAX (713) 973-9352 web: www.twrlighting.com IMPORTANT!!! PLEASE TAKE THE TIME TO FILL OUT THIS FORM COMPLETELY. FILE IT IN A SAFE PLACE.

More information

User's Guide. MiniTec TM Series Model MN25 MultiMeter

User's Guide. MiniTec TM Series Model MN25 MultiMeter User's Guide MiniTec TM Series Model MN25 MultiMeter Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a

More information

AB0602D ARCHITECTURAL BALLAST DRIVER. Version Date 06/25/2012

AB0602D ARCHITECTURAL BALLAST DRIVER. Version Date 06/25/2012 AB0602D ARCHITECTURAL BALLAST DRIVER Version 0.10 Date 06/25/2012 Page 2 of 10 DESCRIPTION AND FEATURES The AB-0602D is a 6 circuit, 2400 Watts per circuit ballast driver intended for dimmable fluorescent

More information

DCM Digital Control Modules

DCM Digital Control Modules DCM Digital Control Modules TECHNICAL MANUAL Version 1.2 November 2011 Safety Precautions Caution Read Instructions: Read and understand all safety and operating instructions before using the equipment.

More information

Installation Instructions for Eaton Surge Protective Device XXCFXXX10-DIN and XXCFXXX10-DIN2

Installation Instructions for Eaton Surge Protective Device XXCFXXX10-DIN and XXCFXXX10-DIN2 Supersedes 6/2015 Surge Protective Device XXCFXXX10-DIN and XXCFXXX10-DIN2 XXCFXXX10-DIN Contents Description Page 1.0 Setup...2 1.1 Before Installation...2 1.2 Installation...3 For AC Applications...3

More information

AD-7830 Servo Amplifier

AD-7830 Servo Amplifier Instruction Manual IM-0606 AD-7830 Servo Amplifier Table of Contents General Information... 2 Introduction... 2 Receiving/Inspection... 2 Storage... 2 Equipment Return... 2 Abbreviations Used in This Manual...

More information

MTS-2000 USER S MANUAL

MTS-2000 USER S MANUAL MTS-2000 USER S MANUAL USER s MANUAL - March 2010 MTS-2000 METER TEST SYSTEM Pay special attention to the warnings and safety instructions that accompany the above symbol wherever it is found within this

More information

SOLARIMMERSION IV Advanced Installation Manual v1.9

SOLARIMMERSION IV Advanced Installation Manual v1.9 SOLARIMMERSION IV Advanced Installation Manual v1.9 1 Contents 1. Overview 2. Technical Specifications 3. Installation Mounting Electrical Installation Clamp Installation Wiring Diagrams 4. Installation

More information

POWER SERIES Plus. 3 in 1 AC Voltage. Digital Switchboard Meter. User s Manual IM2493VVV-2

POWER SERIES Plus. 3 in 1 AC Voltage. Digital Switchboard Meter. User s Manual IM2493VVV-2 POWER SERIES Plus 3 in 1 AC Voltage Digital Switchboard Meter User s Manual General Description The POWER SERIES Plus digital switchboard meters incorporate the latest DSP microprocessor technology. Careful

More information

Hybrid AC Driver [GCNC-1110]

Hybrid AC Driver [GCNC-1110] Page 1 Installation Manual and Datasheet Page 2 Key Features Smooth and quiet operation at all speeds and extremely low motor heating Industrial grade performance for an alternating current servo motor

More information

Model E100RC. DC Millivolt Reference Source ±1µV to ±10Vdc - 10ppm Resolution. Operating Manual

Model E100RC. DC Millivolt Reference Source ±1µV to ±10Vdc - 10ppm Resolution. Operating Manual Model E100RC DC Millivolt Reference Source ±1µV to ±10Vdc - 10ppm Resolution Operating Manual This page intentionally left blank. MANUALLY OPERATED DC VOLTAGE CALIBRATOR Model E-100-RC Serial No. This

More information

MCCB-250 MOLDED-CASE CIRCUIT BREAKER TESTER

MCCB-250 MOLDED-CASE CIRCUIT BREAKER TESTER MCCB-250 MOLDED-CASE CIRCUIT BREAKER TESTER USER S MANUAL Vanguard Instruments Company, Inc. 1520 S. Hellman Ave. Ontario, California 91761, USA TEL: (909) 923-9390 FAX: (909) 923-9391 January 2015 Revision

More information

LCI User Manual mantracourt.com

LCI User Manual mantracourt.com LCI User Manual mantracourt.com LCI Load Cell Junction Box with Fault Monitor Contents Chapter 1 Introduction to the LCI... 2 Chapter 2 Installing the LCI... 3 Chapter 3 Setting up the LCI... 4 Sequence

More information

SECTION SOLID-STATE REDUCED VOLTAGE STARTERS

SECTION SOLID-STATE REDUCED VOLTAGE STARTERS SECTION 26 29 13.16 PART 1 - GENERAL 1.1 THE REQUIREMENT A. General: The CONTRACTOR shall provide solid-state reduced voltage motor starters, complete and operable, in accordance with the Contract Documents.

More information

CMSA-100 Manual. Cinema Media Server Automation

CMSA-100 Manual. Cinema Media Server Automation CMSA-100 Manual Cinema Media Server Automation 181 Bonetti Drive San Luis Obispo, CA 93401-7397 USA Phone: +1 805 549 0161 Fax: +1 805 549 0163 www.uslinc.com Table of Contents 1. Safety Notice... 3 2.

More information

4300 WINDFERN RD #100 HOUSTON TX VOICE (713) FAX (713) web: IMPORTANT!!!

4300 WINDFERN RD #100 HOUSTON TX VOICE (713) FAX (713) web:   IMPORTANT!!! 4300 WINDFERN RD #100 HOUSTON TX 77041-8943 VOICE (713) 973-6905 FAX (713) 973-9352 web: www.twrlighting.com IMPORTANT!!! PLEASE TAKE THE TIME TO FILL OUT THIS FORM COMPLETELY. FILE IT IN A SAFE PLACE.

More information

INSTRUCTION MANUAL A START-UP AND SHUT DOWN LOAD SEQUENCER AND AUTOMATIC LOAD SHEDDING SYSTEM WITH INTEGRAL RELAYS FOR 8 LOADS MODEL #091-60A

INSTRUCTION MANUAL A START-UP AND SHUT DOWN LOAD SEQUENCER AND AUTOMATIC LOAD SHEDDING SYSTEM WITH INTEGRAL RELAYS FOR 8 LOADS MODEL #091-60A FILE: 09-60A.PMD DATE: 07-30-09 REV. B INSTRUCTI MANUAL LOAD MANAGER MARK II A START-UP AND SHUT DOWN LOAD SEQUENCER AND AUTOMATIC LOAD SHEDDING SYSTEM WITH INTEGRAL RELAYS FOR 8 LOADS MODEL #09-60A INPUT:

More information

Tempco Instruction Manual

Tempco Instruction Manual Tempco Instruction Manual 1/16 DIN Solid State Temperature Controller Relay Output Solid State Output For Heating Model Numbers: TEC-901, TEC-902, TEC-905 Temperature controls in this series are designed

More information

MiT. MOVING image TECHNOLOGIES INSTRUCTIONS FOR INSTALLATION, OPERATION, AND MAINTENANCE

MiT. MOVING image TECHNOLOGIES INSTRUCTIONS FOR INSTALLATION, OPERATION, AND MAINTENANCE MiT MOVING image TECHNOLOGIES INSTRUCTIONS FOR INSTALLATION, OPERATION, AND MAINTENANCE OF A-LMS 24 Architectural Lighting Management System Part number A000230-xxx Manual Version 1.0 MOVING image TECHNOLOGIES,

More information

PMDX-105. I/O Option Riser Board User s Manual. Document Revision: 1.1 Date: 7 September 2004 PCB Revision: PCB-443A

PMDX-105. I/O Option Riser Board User s Manual. Document Revision: 1.1 Date: 7 September 2004 PCB Revision: PCB-443A PMDX-105 I/O Option Riser Board User s Manual Date: 7 September 2004 PMDX Web: http://www.pmdx.com 7432 Alban Station Blvd., A105 Phone: +1 (703) 912-4991 Springfield, VA 22150-2321 USA FAX: +1 (703) 912-5849

More information

MYRIAD LC1 LEVEL CONTROLLER INSTRUCTION MANUAL

MYRIAD LC1 LEVEL CONTROLLER INSTRUCTION MANUAL MYRIAD LC1 LEVEL CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM MYR LC1 MANUAL 062114 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features Dimensions WIRING

More information

POWER SERIES Plus Watt / VAR / Power Factor Digital Switchboard Meter User s Manual IM2493WVP-3

POWER SERIES Plus Watt / VAR / Power Factor Digital Switchboard Meter User s Manual IM2493WVP-3 POWER SERIES Plus Watt / VAR / Power Factor Digital Switchboard Meter User s Manual General Description The POWER SERIES Plus digital switchboard meters incorporate the latest DSP microprocessor technology.

More information

POWER SERIES Plus Watt-VAR / Watt-Power Factor Digital Switchboard Meter User s Manual IM2492WVPF-1

POWER SERIES Plus Watt-VAR / Watt-Power Factor Digital Switchboard Meter User s Manual IM2492WVPF-1 POWER SERIES Plus Watt-VAR / Watt-Power Factor Digital Switchboard Meter User s Manual General Description The POWER SERIES Plus digital switchboard meters incorporate the latest DSP microprocessor technology.

More information

Global Water globalw.com. Global Water. Instrumentation, Inc.

Global Water globalw.com. Global Water. Instrumentation, Inc. Global Water Instrumentation, Inc. 151 Graham Road P.O. Box 9010 College Station, TX 77842-9010 T: 800-876-1172 Int l: (979) 690-5560, Fax: (979) 690-0440 E-mail : globalw@globalw.com PC300: Process Controller

More information

********SERVICE MANUAL******** MODELS:

********SERVICE MANUAL******** MODELS: ********SERVICE MANUAL******** MODELS: 75211-10 Variable Speed Drive, 5000 RPM, 115 V 75211-15 Variable Speed Drive, 5000 RPM, 230 V, CE Mark 75211-60 Variable Speed Drive w/pump, 9000 RPM, 115 V 75211-62

More information

Operation Manual. Concorde 600 Power Supply. *This instrument is intended for laboratory use only.

Operation Manual. Concorde 600 Power Supply. *This instrument is intended for laboratory use only. Concorde 600 Power Supply Operation Manual Cat.no. R10-1001011 *This instrument is intended for laboratory use only http://www.recenttec.com E-mail : support@recenttec.com Version 1.1 Packing List x 1

More information

ATE-DMG QUICK START GUIDE. KEPCO An ISO 9001 Company. VOLTAGE/CURRENT STABILIZED POWER SUPPLY WARNINGS I INTRODUCTION II SAFETY.

ATE-DMG QUICK START GUIDE. KEPCO An ISO 9001 Company. VOLTAGE/CURRENT STABILIZED POWER SUPPLY WARNINGS I INTRODUCTION II SAFETY. QUICK START GUIDE KEPCO An ISO 9001 Company. ATE-DMG VOLTAGE/CURRENT STABILIZED POWER SUPPLY I INTRODUCTION 1.1. SCOPE OF MANUAL. This Quick Start Guide covers simple installation and local operation of

More information

TM0-1 AMPLIFIER / CONDITIONER MODULE OPERATORS MANUAL. Transducer Techniques R

TM0-1 AMPLIFIER / CONDITIONER MODULE OPERATORS MANUAL. Transducer Techniques R TM0-1 AMPLIFIER / CONDITIONER MODULE OPERATORS MANUAL Transducer Techniques R 42480 RIO NEDO, TEMECULA, CA 92590 (909) 719-3965 FAX (909) 719-3900 URL: http://www.ttloadcells.com E-mail: tti@ttloadcells.com

More information

SECTION ENCLOSED SWITCHES AND CIRCUIT BREAKERS

SECTION ENCLOSED SWITCHES AND CIRCUIT BREAKERS SECTION 262816 - ENCLOSED SWITCHES AND CIRCUIT BREAKERS PART 1 - GENERAL 1.1 RELATED DOCUMENTS A. Drawings and general provisions of the Contract, including General and Supplementary Conditions and Division

More information

Installation, Testing, and Operating Procedures 30 AMP PORTABLE AND PERMANENT SERIES GFCI SINGLE and MULTIPHASE

Installation, Testing, and Operating Procedures 30 AMP PORTABLE AND PERMANENT SERIES GFCI SINGLE and MULTIPHASE IMPORTANT! Please read all the information on this sheet. SAVE THESE INSTRUCTIONS! NOTICE BEFORE USING READ INSTRUCTIONS COMPLETELY. TO BE INSTALLED BY A QUALIFIED ELECTRICIAN IN ACCORDANCE WITH NATIONAL

More information

Broadband Automatic Disconnect Switch. User Manual

Broadband Automatic Disconnect Switch. User Manual Reset/Test Primary/ Primary Broadband Automatic Disconnect Switch User Manual Local Power Remote Pwer Local 63V Fault Secondary Select Secondary 220V Normal 990-1929 09/2004 Introduction Introduction

More information

TABLE OF CONTENTS TABLE OF CONTENTS... 1 IMPORTANT SAFETY NOTICE...

TABLE OF CONTENTS TABLE OF CONTENTS... 1 IMPORTANT SAFETY NOTICE... TABLE OF CONTENTS TABLE OF CONTENTS... 1 IMPORTANT SAFETY NOTICE... 2 1.0 General Information... 3 1.1 System Components... 3 1.2 Specifications... 3 1.2.1 Torque Ranges... 3 1.2.2 Electrical Specifications...

More information

OPERATING INSTRUCTIONS. For the KNOPP TYPE KCTS-8000 CURRENT TRANSFORMER TESTING SYSTEM. Serial Number: Leaflet Number G.

OPERATING INSTRUCTIONS. For the KNOPP TYPE KCTS-8000 CURRENT TRANSFORMER TESTING SYSTEM. Serial Number: Leaflet Number G. OPERATING INSTRUCTIONS For the KNOPP TYPE KCTS-8000 CURRENT TRANSFORMER TESTING SYSTEM Serial Number: Leaflet Number 2 015 G October 2016 Knopp Incorporated 1307 66th Street Emeryville, California 94608

More information

PIM-Mini Pulsed Current Source Operation Manual

PIM-Mini Pulsed Current Source Operation Manual PIM-Mini Pulsed Current Source Operation Manual Directed Energy, Inc. 1609 Oakridge Dr., Suite 100, Fort Collins, CO 80525 (970) 493-1901 sales@ixyscolorado.com www.ixyscolorado.com Manual Document 7650-0007

More information

V-9908 MESSAGE/PAGE PANEL

V-9908 MESSAGE/PAGE PANEL Issue 4 V-9908 MESSAGE/PAGE PANEL Introduction These instructions contain the specifications and guidelines necessary to install, operate, and maintain the V-9908, /Page Panel. The V-9908 /Page Panel provides

More information

RE-82 RACK MOUNT DIMMER OWNERS MANUAL. 8 X 2400Watts. Revision /29/2007

RE-82 RACK MOUNT DIMMER OWNERS MANUAL. 8 X 2400Watts. Revision /29/2007 RE-82 RACK MOUNT DIMMER 8 X 2400Watts OWNERS MANUAL Revision 2.4 11/29/2007 Page 2 of 8 RE-82 CONTROL PANEL DESCRIPTION The RE-82 is an 8 channel dimmer with a maximum capacity of 2,400 watts per channel

More information

BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL

BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL Issue 2011 ASL Intercom BV DESIGNED & MANUFACTURED BY: ASL Intercom B.V. Zonnebaan 42 3542 EG Utrecht The Netherlands Tel: +31 (0)30 2411901 Fax: +31 (0)30

More information

4301 Operation Manual

4301 Operation Manual Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) 4301 Operation Manual Complete Solution for ANSI-T1.315-2001 Protective Device Output Transient Testing Limited One-Year

More information

CM-220 True RMS AC CLAMP METER INSTRUCTION MANUAL

CM-220 True RMS AC CLAMP METER INSTRUCTION MANUAL CM-220 True RMS AC CLAMP METER INSTRUCTION MANUAL Safety International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information.

More information

Model 3526-B Digi-Probe Multimeter

Model 3526-B Digi-Probe Multimeter Model 3526-B Digi-Probe Multimeter CAUTION: Read, understand and follow all Safety Rules and Operating Instructions in this instruction manual before using this product. 84-797 Rev A The Model 3526-B has

More information

Installation & Operation

Installation & Operation LED Readout Installation & Operation WARRANTY Accurate Technology, Inc. warrants the ProScale Systems against defective parts and workmanship for 1 year commencing from the date of original purchase. Upon

More information

CRM Control Panel and Weather Detection System CRM3 for DC actuators (part # 40821J) CRM4 for AC actuators (part # 40822K)

CRM Control Panel and Weather Detection System CRM3 for DC actuators (part # 40821J) CRM4 for AC actuators (part # 40822K) CRM Control Panel and Weather Detection System CRM3 for DC actuators (part # 40821J) CRM4 for AC actuators (part # 40822K) CRM Applications CRM includes a control relay box and a power supply; additional

More information

Quantum III. Compact DC Drive Package. Slitter DC Drive Package. Quantum III

Quantum III. Compact DC Drive Package. Slitter DC Drive Package. Quantum III Compact DC Drive Package The delivers a DC drive package that integrates the intelligence of the Mentor II with a space saving design that incorporates many accessories typically required in the North

More information

Voltage Dip-Proofing Inverter

Voltage Dip-Proofing Inverter Voltage Dip-Proofing Inverter DPI53S / 54L Series Models 120V & 208/230 50/60Hz LEADERS IN VOLTAGE DIP-PROOFING Contents: Introduction...3 Theory of operation...3 Specifications...5 Up-time considerations...6

More information

700 Series 200 Amp Clamp Meters

700 Series 200 Amp Clamp Meters 700 Series 200 Amp Clamp Meters #61-700 #61-701 #61-702 1 2 3 6 5 7 4 8 1. Non-contact voltage (NCV) (#61-701 and #61-702) With the NCV tab on the tip of the clamp close to an AC voltage, press the NCV

More information

Regenerative Power Module And Common Bus RPM Panels Instruction Manual

Regenerative Power Module And Common Bus RPM Panels Instruction Manual Regenerative Power Module And Common Bus RPM Panels Instruction Manual Part Number September 2010 Copyright 2010 Magnetek Material Handling PRODUCT MANUAL SAFETY INFORMATION Magnetek, Inc. (Magnetek)

More information

User Guide. Control Box. RoscoLED TM.

User Guide. Control Box. RoscoLED TM. RoscoLED TM Control Box User Guide This guide applies to the following RoscoLED Control Box models: RoscoLED Control Box 300W/Static White (293 22250 0000) RoscoLED Control Box 400W/VariWhite (293 22260

More information

Voltage Sag Simulator

Voltage Sag Simulator For use with all DPI s & VDC s 120/208/240V 50/60Hz LEADERS IN VOLTAGE-DIP PROOFING Contents Introduction to the DPI... 3 Theory of DPI operation... 3 Description -... 4 Operation... 5 Sag Simulation...

More information

OPERATING AND SERVICE MANUAL. Universal Interface Device 45

OPERATING AND SERVICE MANUAL. Universal Interface Device 45 OPERATING AND SERVICE MANUAL Universal Interface Device 45 MAGNA-POWER ELECTRONICS, INC. 81 FULTON STREET, BOONTON, NJ 07005 September 7, 2005 SAFETY NOTICE Universal Interface Device 45 (UID45) connects

More information

Agilent Technologies E5339A 38-Pin Low-Voltage Probe

Agilent Technologies E5339A 38-Pin Low-Voltage Probe Agilent Technologies E5339A 38-Pin Low-Voltage Probe Installation Note The 38-pin low-voltage probe provides a convenient way to connect two Agilent Technologies logic analyzer probe cables to a small

More information

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389 User s Guide 600A True RMS AC/DC Clamp Meter Model 38389 Safety International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further

More information

Digital Lighting Systems, Inc. PD216. Two Channel Dimmer and Switch Packs PROTOCOL USER'S MANUAL. PD216-UM Rev. E - 02/03

Digital Lighting Systems, Inc. PD216. Two Channel Dimmer and Switch Packs PROTOCOL USER'S MANUAL. PD216-UM Rev. E - 02/03 Digital Lighting Systems, Inc. PD26 Two Channel Dimmer and Switch Packs PROTOCOL PD26 S2 S USER'S MANUAL PD26-UM Rev. E - 02/03 Digital Lighting Systems PD26 User's Manual - Page GENERAL DESCRIPTION The

More information

HITACHI. EH-150 series PLC EH-RTD8 Resistance Temperature Detective input module Instruction manual. Safety precautions

HITACHI. EH-150 series PLC EH-RTD8 Resistance Temperature Detective input module Instruction manual. Safety precautions HITACHI EH-150 series PLC Resistance Temperature Detective input module Instruction manual Thank you for purchasing a Hitachi Programmable Logic Controller. To operate it safely, please read this instruction

More information

Profibus DP Expansion Board

Profibus DP Expansion Board Profibus DP Expansion Board Catalog No. EXB014A01 Installation and Operating Manual 8/03 MN1323 Table of Contents Section 1 General Information................................... 1-1 Introduction.........................................

More information

COOPER POWER SERIES. DC motor operator control installation and operation instructions. Padmounted Switchgear MN285018EN

COOPER POWER SERIES. DC motor operator control installation and operation instructions. Padmounted Switchgear MN285018EN Padmounted Switchgear MN285018EN Effective May 2017 Supersedes October 1994 (S285-80-1) COOPER POWER SERIES DC motor operator control installation and operation instructions DISCLAIMER OF WARRANTIES AND

More information

Analog Monitor Installation Manual

Analog Monitor Installation Manual Analog Monitor Installation Manual Part Number: 144-23919 Copyright 2011 Magnetek 1. Preface and Safety Magnetek manufactures products used as components in a wide variety of industrial systems and equipment.

More information

PMDX-108-Output. 8-Channel Isolated Output Board for PC parallel port pins 2-9. User s Manual

PMDX-108-Output. 8-Channel Isolated Output Board for PC parallel port pins 2-9. User s Manual PMDX-108-Output 8-Channel Isolated Output Board for PC parallel port pins 2-9 User s Manual Date: 25 February 2010 PMDX Web: http://www.pmdx.com 9704-D Gunston Cove Rd Phone: +1 (703) 372-2975 Lorton,

More information

PWRguard PLUS Spring City Drive Waukesha, WI

PWRguard PLUS Spring City Drive Waukesha, WI PWRguard PLUS www.westmountainradio.com 1020 Spring City Drive Waukesha, WI 53186 262-522-6503 sales@westmountainradio.com 2016, All rights reserved. All trademarks are the property of their respective

More information