+GF+ SIGNET Conductivity/Resistivity Transmitter Instructions
|
|
- Garry Thompson
- 6 years ago
- Views:
Transcription
1 GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions ENGLISH A9/99 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents. Installation. Specifications. Electrical Connections. Menu Functions Conductivity/Resistivity 6.50 us 5.0 C Relay Relay. Installation The transmitter is available in two versions: a panel mount version, and an integral version for installation near the sensor (universal assembly). Panel Installation The Panel Mounting kits are supplied with the hardware to install instrumentation into panels and maintain a NEMA X seal.. Punch out panel and deburr edges. Recommended clearance on all sides between instruments is inch.. Place gasket on instrument, and install in panel.. Slide mounting bracket over back of instrument until quickclips snap into latches on side of instrument.. Connect wires to terminals. 5. To remove, secure instrument temporarily with tape from front or grip from rear of instrument. DO NOT RELEASE. Press quickclips outward and remove. gasket latch panel terminals mounting bracket quickclip. Universal Assembly. Install transmitter base (8050). Connect wires to transmitter.. Close unit and secure with push and twist lock. Seal cable entry. wires seal. Specifications General Compatible electrodes: GF SIGNET 8XX Standard and Certified Series Conductivity/Resistivity Electrodes Accuracy: ±% of reading Enclosure: Rating: NEMA X/IP65 front Case: PBT Window: Polyurethane coated polycarbonate Keypad: Sealed key silicone rubber Weight: Approx. 5g ( oz.) Display: Alphanumeric x 6 LCD Contrast: User selected, 5 levels Update rate: second Environmental Operating temperature: 0 to 70 C ( to 58 F) Storage temperature: 5 to 80 C (5 to 76 F) Relative humidity: 0 to 95%, noncondensing Standards and Approvals CSA, CE, UL listed Manufactured under ISO 900 Electrical Sensor input range: Conductance: to 00,000 µs Resistivity: 0 KΩ to 8.6 MΩ TDS: 0.0 to 00,000 ppm GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions Sensor Models Ultra Pure Pure Rinse GF SIGNET Sensor Ranges ,000 0,000 00,000 00,000 (8 MΩ) (0 kω) Conductivity Range (µs) Sensor input range: Temperature: PT 000, 5 to 0 C ( to 8 F) Current output: to 0 ma, isolated, fully adjustable and reversible : to VDC ±5%, regulated Max loop impedance: 50 Ω V, 5 Ω 8 V, 600 Ω V Update rate: 0.5 seconds Accuracy: ±0.0 5 C, V Relay outputs ( sets): Mechanical SPDT contacts: Hi, Lo, Pulse Programmable Maximum voltage rating: 5 0 VDC, or 5 50 VAC resistive load Hysteresis: User Adjustable page of 8
2 Panel Mount Field Mount 96mm (.8 in.) 96 mm (.8 in.) Panel Cutout 96 mm (.8 in.) Optional Rear Cover 9 x 9 mm ( 0.8, 0 mm).6 x.6 in. (0.0, 0 in.) 96 mm (.8 in.) mm (.6 in.) 97 mm (.8 in.) 56 mm (. in.) 06 mm (.8 in.). Electrical Connections Caution: Failure to fully open terminal jaws before removing wire may permanently damage instrument. Wiring Procedure. Remove in. (6 mm) of insulation from wire end.. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.. Insert exposed (noninsulated) wire end in terminal hole until it bottoms out.. Release orange terminal lever to secure wire in place. Gently pull on each wire to ensure a good connection. Wiring Removal Procedure. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.. When fully open, remove wire from terminal... System /Loop. System /Loop. System /Loop Description VDC VDC ±5%, system power and current loop connections. Max. loop impedence: 50 Ω V, 600 Ω V. Relays 5. Relay NC contact 6. Relay COM contact 7. Relay NO contact 8. Relay NC contact 9. Relay COM contact 0. Relay NO contact Sensor Input. Red (Signal IN). White (Temp IN). Black (ISO GND). Shield (Sensr GND) Relay / contact sets programmable as: High/Low alarm with adjustable hysteresis Proportional pulse output Disable (Off) selection System Pwr Loop System Pwr Loop Relay (NO) Relay (COM) Relay (NC) Relay (NO) Relay (COM) Relay (NC) Sensr Gnd (SHIELD) Iso. Gnd (BLACK) Temp. IN (WHITE) Signal IN (RED) Wiring Tips: Do not route sensor cable in conduit containing AC power wiring electrical noise may interfere with sensor signal. Routing sensor cabling in grounded metal conduit may prevent moisture damage, electrical noise, and mechanical damage. Seal cable entry points to prevent moisture damage. When placing two wire ends into a single terminal, solder or crimp ends together. page of 8 GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions
3 . System /Loop Connections Standalone application, no current loop used Connection to a PLC with builtin power supply Transmitter Loop Loop DC V Transmitter Loop Loop Internal PLC Connection PLC Ground Loop Input 0 ma Loop Input 0 ma Connection to a PLC/Recorder, separate supply Transmitter Loop Loop PLC or Recorder DC V Loop Input 0 ma in Loop Input 0 ma in Example: Two transmitters connected to PLC/Recorder with separate power supply Transmitter DC V Loop Loop Transmitter Loop Loop PLC or Recorder Channel 0 ma in Channel 0 ma in Channel 0 ma in Channel 0 ma in. Sensor Input Connections Wiring Tip: Do not route sensor cable in any conduit containing AC power wiring electrical noise may interfere with the signal. Shield (Sensor Gnd) Black (Iso Gnd) White (Temp In) Red (Signal In) GF SIGNET 8XX Standard and Certified Cells. Relay Functions Low: Output triggers when process variable is less than setpoint. High: Output triggers when process variable is higher than setpoint. Example: In Low Alarm Mode Operation, the output becomes active when the process drops below the setpoint, and becomes inactive when the process rises above the setpoint plus hysteresis. The opposite is true for High Alarm Mode. Hysteresis Setpoint Process Output Active Output Inactive Time GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions Off: Disables output. Proportional Pulse Mode Operation The output emits a 00 ms pulse (simulated contact closure) at rate defined by the Output, Pulse Range, Output Rate, and the process condition (0 to 00 pulses/minute, as programmed) Example: As the process falls below 0 the output will start pulsing in relation to the process value, the max pulse endpoint and the programmed pulses/min. Pulse rate will increase as the process approaches the programmed endpoint. 0 = pulses/min. 7.5 = 50 pulses/min. 5 = 00 pulses/min. Pulse rate 5 00 pulses Relay or Relay Output Rate: 00 Pulses/min> Relay : Pulse > 0 Start max. pulse endpoint 0 pulses Process variable page of 8
4 . Menu Functions VIEW Menu: is displayed during standard operation. Press UP or DOWN buttons to view process parameters. Press UP and DOWN buttons at the same time, to exit any other display and return to VIEW menu. Display will return to VIEW menu in 0 minutes unless a key is pressed. CALIBRATE Menu: contains display setup and output parameters. A security code feature prevents unauthorized tampering. To access CALIBRATE menu: Press button for seconds to display: Press UP, UP, UP, DOWN buttons in sequence to display: CALIBRATE: Enter Key Code CALIBRATE: XXXX Enter Key Code OPTIONS Menu: contains setup and display features for minor display or output adjustments. To access OPTIONS menu: Press button for 5 seconds to display: Press UP, UP, UP, DOWN buttons in sequence to display: OPTIONS: Enter Key Code OPTIONS: XXXX Enter Key Code Menu Tips Right button scrolls to right, from top to bottom row, and allows editing when ">" symbol is shown. In CALIBRATE or OPTIONS menus, the transmitter will continue to measure and control outputs. When > is pressed, the input value is held at the last measured process value. When sensor is not connected, unit will display CHECK SENSOR and any output controlled by sensor will be at.6 ma or OFF. Example To change date, first enter CALIBRATE menu (Press button for seconds; Press UP, UP, UP, DOWN buttons in sequence) Once in CALIBRATE menu, press UP button time.. Display shows right arrow. Press RIGHT button to display "0" blinking. Press buttons to scroll through numbers.. Press button to save 5. Display now reads new date LAST CAL: 0099 > LAST CAL: 0099 LAST CAL: SAVING LAST CAL: GF SIGNET GF SIGNET GF SIGNET GF SIGNET GF SIGNET Menu Functions View Menu Range Calibrate Menu Range Preset Options Menu Range Preset Cond. unit sensor Cell Constant Standard Standard Contrast: 5 Temperature based Standard > Custom Level > Loop Output: 0 ma Cell: Standard 0.0 Cond Decimal *.**** to ****.* ma > 0. ****.* *****. Last Cal: to Averaging: Off Off Date Off > Low (secs) 0 High(8secs) Cell: Custom to.0000 Loop Adjust:.8 to > ma > 5.0 ma ma Cond Units: us us/cm Loop Adjust: 9.0 to 0.00 us > ms 0.00 ma >.0 ma ma PPM* Temp Display: C C KΩ C > F MΩ Temperature 0.00 to.00% If PPM selected Comp %:.00 > 0.00% PPM Factor: 0.00 to Test Loop: 0 ma N/A.00 > *(TDS ppmf) > Test Relay: On or N/A Set: ± 0 C N/A > Off Temperature > Test Relay: On or N/A Set: to N/A > Off Conductivity > Loop Source: Cond Cond Cond > Temp Loop Range: us to 0.00 to 00 us/cm 0.00 > > (ma) (0 ma) Relay Source: Cond Cond (Relay) Cond > Temp. Temp (Relay) Relay Mode: Off Low (Relay) Low > Low Hi (Relay) Hi Pulse Low or High Selected Relay Setpnt: to 0 us (Relay) Settings repeat for Relay 0 us > C (Relay) Relay Hys: to 0.5 us (Relay) 0.50 us > C (Relay) Pulse Selected Relay Rng: us to 00 us (Relay) 0.00 > 0.00 > C (Relay) (Start>Endpoint) Relay PlsRate: pulses/min > pulses/min pulses/min Last Cal: to > 9999 page of 8 GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions
5 Calibration Procedure. Requirements Electronic calibration is performed to exacting standards by GF SIGNET. System calibration will reduce errors which may be caused by sensor wire lengths longer than the standard 5 ft. length. Wire lengths of 00 feet are acceptable; cable shield must be maintained through cable splice. Calibration may be done by known solution value (A), or by resistance simulation (B). A) Calibration with NIST Traceable Solutions: When using NIST traceable standards, ensure the sensor and test solution are at the solution temperature specified on the test solution label. Prevent contamination of the calibration solution. Thoroughly rinse the sensor in a small amount of test solution before placing in any test solution for calibration purposes. B) Optional Verification with Precision Resistors: The use of precision resistors (±0.%) connected to the rear "Temp In", "Signal IN", and "Iso Gnd" terminals in place of the GF SIGNET sensor, yields quick and accurate electronic instrument calibration. Calibration is completed as follows: ) Select a standard cell constant based on desired range of operation. ) Place a 096 Ω (5 C) resistor between "Temp IN" and "Iso. Gnd" terminals as shown. Note: Temperature simulation errors can adversely effect calibration:.85 Ω = C. ) Calculate the required simulation resistor that represents a value within the selected cell range. The formula for determining the required simulation resistance is: Iso. Gnd (BLACK) Temp. IN (WHITE) Signal IN (RED) TC resistor Simulation resistor Resistance = Sensor Cell : e.g. 0. Cell = 5,000 Ω or 5 KΩ Desired conductivity (Siemens*) (Siemens*) Conductivity = Sensor cell : e.g. 0. Cell = Siemens* Simulation resistance (Ω) 00,000 (Ω) or µs/cm (*Conversion: µs = X 0 6 Siemens or Siemens) ) Place the calculated simulation resistance between the "Signal IN" and "Iso Gnd" terminals as shown. 5) Set temperature and set conductivity. (Optional: enter zero to reset factory calibration (zero must be reentered if currently displayed to reset factory calibration.). Temperature Coefficient Conductivity measurement is highly dependent on temperature. Temperature dependence is expressed as the relative change per C, commonly known as percent/ C change from 5 C, or slope of the solution. Slopes can very significantly depending on process solution type. The factory default temperature compensation factor is.00%/ C. Process solutions may require adjustment for maximum accuracy. To determine the optimum temperature compensation factor for a process:. Disable the 8850 temperature comp % factor by entering Heat the sample solution close to the maximum process temperature. Place sensor in the sample solution allowing several minutes for stabilization. Access the VIEW menu and record the displayed temperature and conductivity values in the spaces provided: Displayed temperature: T = C Displayed conductivity: C = C (Do not use this procedure for solutions from µs to 0. µs (0 MΩ to 8 MΩ). An internal pure water curve is used for these ranges. The factory default setting of.00%/ C should be used.). Cool the sample solution close to the minimum process temperature. Place sensor in the sample solution allowing several minutes for stabilization. Record displayed temperature and conductivity values in the spaces provided: Displayed temperature: T = C Displayed conductivity: C = C (A 0% change in conductivity between steps and is recommended.). Substitute recorded readings (steps and ) into the following formula: TC Slope = 00 x (C C) (C x (T 5)) (C x (T 5)) Example: A sample solution has a conductivity of 05 8 C. After cooling the solution, the conductivity was measured at 50 C. (C = 05, T = 8, C = 50, T = ) The TC is calculated as follows: TC Slope = 00 x (05 50) = 5500 =.%/ C (50 x (8 5)) (05 x ( 5)) 860 GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions page 5 of 8
6 . Parts Per Million (PPM) Factor This feature is only applicable when PPM display units are selected. The programmable PPM Factor is adjustable from 0.0 to 9.99 (factory default =.00). Determine the best PPM Factor for a process solution by calculating the solution's conductivity (µs) and the percent of total dissolved solids (PPM). PPM Factor = TDS (PPM) = Solution conductivity (µs) Total dissolved solids (PPM) Solution conductivity (µs) PPM Factor Example: Solution conductivity = 00 µs TDS = 00 PPM (mg/l) PPM Factor = 00 µs = PPM Troubleshooting Display Check Sensor? Value Must be less than 8. Value must be or more Value must be 00 or less SETUP READ ERROR Press Any Key Problem No sensor detected. Note: You may enter the CALIBRATION and OPTIONS menu to program setpoint values even though Check Sensor? is displayed. Applies to Mohm display only. Setpoint value entered is not achievable Applies to us/cm display only. Setpoint value entered is not achievable Setpoint value greater than 00. Memory fault occurred. Solution Connect sensor. Reduce setpoint value. Set Temp Coefficient to Increase setpoint value. Set Temp Coefficient to 0.00 Use a pulse rate value less than 00 Press any key to reload presets, then reprogram setpoints. page 6 of 8 GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions
7 GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions page 7 of 8
8 GF SIGNET Signet Scientific Company, 0 Aerojet Avenue, El Monte, CA 9788 U.S.A. Tel. (66) Fax (66) For Worldwide Sales and Service, visit our website: gfsignet.com Or call (in the U.S.): (800) GEORGE FISCHER GF Piping Systems /(A9/99) English Scientific Company 999 Printed in U.S.A. on Recycled Paper page 8 of 8 GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions
+GF+ SIGNET ph/orp Transmitter Instructions
GF SIGNET 8750 ph/orp Transmitter Instructions ENGLISH 8750.090 A9/99 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury.
More informationSignet Pressure Transmitter
Signet 850 Pressure English *850.090* 850.090 Rev. H /06 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.
More informationContents. 1. Installation
ENGLISH 8750.090 C/0 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents. Installation. Specifications. Electrical
More information+GF+ SIGNET 5800CR Conductivity/Resistivity Monitor Instructions
GF SIGNET 5800CR Conductivity/Resistivity Monitor Instructions ENGLISH 35800CR.0901 D1/98 CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections.
More informationSignet Flow Transmitter
Signet 8550 Flow *8550.090* English 8550.090 Rev. L 0/ English WARNING! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.
More informationSignet Temperature Transmitter
Loop Output Output 7 Sensr IN (RED) Sensr V (BLACK) Signet 80 Temperature 80.090 English 80.090 Rev. J 0/ English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions
More informationSIGNET ph/orp Transmitter
SIGNET 8750- ph/orp Transmitter ENGLISH 3-8750.090- D-4/03 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.
More informationSignet 5900 Salinity Monitor
Signet 5900 Salinity Monitor *35900.0901* 35900.0901 Rev. E 3/06 English Contents 1. Power Connections 2. Compatible Sensor Wiring 3. 4 20 ma Current Output Connections 4. Relay Connections 5. Relay Operation
More informationSIGNET Level Transmitter
ENGLISH 850.090 A8/0 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents. Description. Specifications. Panel
More informationSIGNET 5700 ph/orp Monitor - ORP Instructions
SIGNET 5700 ph/orp Monitor ORP Instructions ENGLISH 35700.0901 H (6/01) English CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections. Follow
More information+GF+ SIGNET 5900 Salinity Monitor Instructions
+GF+ SIGNET 5900 Salinity Monitor Instructions ENGLISH 35900.0901 (B8/99), English CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections.
More informationSignet 5800CR Conductivity/Resistivity Monitor
Signet 5800CR Conductivity/Resistivity Monitor *35800CR.0901* 35800CR.0901 Rev. H 3/06 English Contents 1. Power Connections 2. Compatible Sensor Wiring 3. 4 20 ma Current Output Connections 4. Relay Connections
More informationSignet ph/orp Transmitter
Signet 87501 ph/orp *38750.0901* 38750.0901 Rev. H 06/10 English English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury.
More informationSignet 5700 ph/orp Monitor - ORP Instructions
Signet 5700 ph/ Monitor Instructions English *5700.0901* 5700.0901 Rev. K /06 English/ Contents 1. Power Connections. Compatible Electrode/Preamp Wiring. 4 0 ma Current Output Connections 4. Relay Connections
More informationSignet ph/orp Transmitter * *
Loop 7 6 5 Signet 8750- ph/orp *-8750.090-* CAUTION! Contents 0.0 ph 5.0C Signet ph/orp. Installation. Panel Installation gasket panel terminals mounting bracket 96 mm (.8 in.) 8050 8 mm (. in.) Optional
More informationSignet 5700 ph/orp Monitor - ph Instructions
Signet 5700 ph/orp Monitor ph Instructions English *35700.0901* 35700.0901 Rev. K 3/06 English/pH CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output
More informationConductivity Transmitters
Conductivity Transmitters CDTX-90 Series Starts at $ 576 Electronics Only Display in µs, ms, kω, MΩ, ppm (TDS) Simulate Function Programmable Temperature Compensation Relay Options Dual Output Option Allows
More information+GF+ SIGNET 9050CR Conductivity/Resistivity Controller
+GF+ SIGNET 9050CR Conductivity/Resistivity Controller Instruction Manual GEORGE FISCHER +GF+ Important Safety Information! CAUTION: Remove AC power to unit prior to wiring input and output connections.
More informationSignet Conductivity Electrodes
Signet 2839-2842 Conductivity Electrodes Features ± 1% accuracy - Custom calibration certificate provided Dual-threaded Compact electrode length for easy in-line installation in small pipe sizes Triple
More informationSignet Conductivity/Resistivity Electrodes
Signet 2819-2823 Conductivity/Resistivity Electrodes 2819 2820 2821 2822 2823 Sanitary Features Standard process connections - ¾ in. NPT Polypro - Tri-clamp 1-1½ in., 2 - Opt. ½ in. NPT 316 SS 316 SS or
More informationSignet 9900 Transmitter
Signet 9900 Transmitter Member of the SmartPro Family of Instruments Features Panel Mount Field Mount NEW! Modbus Module supports RS485 Serial Modbus Communications Multiple sensor types supported with
More informationFLS M9.06. ph/orp MONITOR
FLS M9.06 /ORP MONITOR Safety Instructions General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other instruments
More informationFLS M9.06 PH/ORP MONITOR. safety instructions. packing List
FLS M9.06 PH/ORP MONITOR safety instructions General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other instruments
More informationSignet Conductivity/Resistivity Electrodes
Signet 2819-2823 Conductivity/Resistivity Electrodes 2819 2820 2821 2822 2823 Sanitary Features Standard process connections - ¾ in. NPT Polypro - Tri-clamp 1-1½ in., 2 - Opt. ½ in. NPT 316 SS 316 SS or
More informationFLOW MONITOR & TRANSMITTER
FLS M9.02 FLOW MONITOR & TRANSMITTER SAFETY INSTRUCTIONS General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other
More informationCX3100 Intelligent Conductivity Transmitter
CX3100 Intelligent Conductivity Transmitter Operation Manual CX3100 CONTENTS Precautions for Installation Brief Instruction 1 1. Specifications 4 2. Assembly and Installation 2.1 Transmitter Installation
More informationTable of Contents. 1 Introduction Introduction Front Panel Description Rear Panel Description 3
Chapter Page 1 Introduction 1 1.1 Introduction 1 1.2 Front Panel Description 2 1.3 Rear Panel Description 3 Table of Contents 2 Installation and Operation 4 2.1 Mounting Instructions 4 2.2 Power Connections
More informationMicroprocessor Resistivity & Conductivity Monitor/Controller EC410 / EC430. Operation Manual
Microprocessor Resistivity & Conductivity Monitor/Controller EC410 / EC430 Operation Manual INDEX 1 Specifications 2 Assembly and installation 2.1 Precautions for installation 2.2 Installation of controller
More informationCDD Carbon Dioxide Transmitter
Introduction The OSA CO2 transmitter uses Infrared Technology to monitor CO2 levels within a range of 0 2000 ppm and outputs a linear 4-20 ma or 0-5/0-10 Vdc signal. The enclosure is designed to operate
More informationCDD4 Series Room CO2 Transmitter Installation Instructions
CDD4 Series Room CO2 Transmitter Installation Instructions Introduction The CO2 transmitter uses Infrared Technology to monitor CO2 levels and outputs a linear 4-20 ma or 0-5/0-10 Vdc signal. Options include
More information200CRS Single-Channel Conductivity/Resistivity Measurement System THORNTON. Leading Pure Water Analytics
THORNTON Leading Pure Water Analytics - NIST Traceable Calibration - Two Analog Outputs Standard - Three-Wire Temperature Measurement - Wide Range of Sensor and Cable Selections 200CRS Single-Channel Conductivity/Resistivity
More informationSignet 2552 Metal Magmeter Flow Sensors
Signet 2552 Metal Magmeter Flow s Features Test certificate included Award winning hot-tap magnetic flow sensor up to DN2550 (102 in.) Patented Magmeter technology* Operating range 0.05 to 10 m/s (0.15
More informationHigh Voltage DC Meter
High Voltage DC Meter Javelin D PD644 0-300 VDC input NEMA 4X, IP65 front Scale in engineering units Sunlight readable LED display 4-20 ma analog output Two form C 3 A relays option RS-485 serial communications
More informationMODEL: M5DY SEN TRONIC AG CURRENT LOOP SUPPLY. [2] POWER INPUT AC Power M: V AC (Operational voltage range V, Hz)
Super-mini Terminal Block Signal Conditioners M5-UNIT CURRENT LOOP SUPPLY Functions & Features Powering a 0 ma DC current loop Applicable to smart transmitters Isolation between the input and output Fast
More informationX Year mfg
4to20 ma Output S L Loop PWR Loop2 PWR N/C S L DATA 7 8 9 6,*(7 S L Output 8 S L SIGNET 8900 MultiParameter Controller Supplemental Manual *8900.090SUP* 8900.090SUP Rev. A 2/04 (English). Description This
More informationHIGH-VOLTAGE DC METER
HIGH-VOLTAGE DC METER Javelin D PD644 0-300 VDC input NEMA 4X, IP65 front Scale in engineering units Sunlight readable LED display 4-20 ma analog output Two form C 3 A relays option RS-485 serial communications
More informationOperation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter
Operation Manual MODEL 2TX 2-wire Isolated ph/orp Transmitter 0 2TX CONTENTS INITIAL INSPECTION.....2 INTRODUCTION......2 ASSEMBLY...3 PREPARATION....4 CONNECTING THE ELECTRODE...4 CONNECTING THE TEMPERATURE
More informationOPERATION MANUAL MODEL TT-LP
TURTLE TOUGH OPERATION MANUAL MODEL TT-LP 2-Wire Isolated ph/orp Transmitter turtletough.com.au CONTENT INITIAL INSPECTION 3 INTRODUCTION 3 ASSEMBLY 4 PREPARATION 5 Connection Diagram of Turtle Tough Sensors
More informationCDD4 Duct Carbon Dioxide Transmitter
Drill or punch a 1-1/8 or 1-1/4 hole in the duct at the preferred location and insert the probe into the hole to mark the enclosure mounting holes. Remove the unit and drill the four mounting holes. Clean
More informationHWS Series Wall Mount with Relay Setpoints, LCD, and Humidistat Control
Wall Mount with Relay Setpoints, LCD, and Humidistat Control NOTICE This product is not intended for life or safety applications. Do not install this product in hazardous or classified locations. Read
More informationGlobal Water globalw.com. Global Water. Instrumentation, Inc.
Global Water Instrumentation, Inc. 151 Graham Road P.O. Box 9010 College Station, TX 77842-9010 T: 800-876-1172 Int l: (979) 690-5560, Fax: (979) 690-0440 E-mail : globalw@globalw.com PC300: Process Controller
More information- SELF-TUNING PID ALGORITHM - INTUITIVE COLOR DISPLAY WITH TEXT MESSAGING - UNIVERSAL PROCESS AND TC/RTD INPUTS - MULTI- FUNCTION RAMP- DWELL/PROCESS
ELK. ETK COMPACT BASIC CONTROL - SELF-TUNING PID ALGORITHM - INTUITIVE COLOR DISPLAY WITH TEXT MESSAGING - UNIVERSAL PROCESS AND TC/RTD INPUTS - MULTI- FUNCTION RAMP- DWELL/PROCESS TIMER - SOFT START OUTPUT
More informationTB84TE Conductivity transmitter 2-Electrode. Superior accuracy in low conductivity and pure water. applications
Data sheet DS/TB84TE EN Rev. C TB84TE Conductivity transmitter 2-Electrode Superior accuracy in low conductivity and pure water applications Smart key menu programming Superior accuracy in low conductivity
More informationLCD Readout Digital Pressure Switch
LCD Readout Digital Pressure Switch Series 3 (For )/ 3 (For Positive Pressure) For General Pneumatics Easy pressure setting with the digital display Can be integrated with a vacuum unit, Series ZX Built-in
More informationOperation Manual CX10 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller
Operation Manual CX10 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller 11751 Markon Drive Garden Grove, CA. 92841 U.S.A. pg. 1 CONTENTS GENERAL INTRODUCTION.....3 INITIAL INSPECTION......3
More informationINDICATOR FOR PROCESS CONTROL. Love Controls. LCI108J-0x INSTRUCTION MANUAL. EDITION: January Valid for models with software version P-1
Love Controls INDICATOR FOR PROCESS CONTROL LCI108-0x LCI108J-0x INSTRUCTION MANUAL EDITION: January 2001 Valid for models with software version P-1 INTRODUCTION TO THE LOVE CONTROLS LCI108 SERIES This
More informationDIGITAL PANEL THERMOMETER
Love Controls DIGITAL PANEL THERMOMETER LCI108-2x LCI108J-2x INSTRUCTIONS MANUAL EDITION: February 2001 INTRODUCTION TO THE LOVE CONTROLS LCI108 SERIES This manual does not constitute a formal agreement.
More informationSignet 2750 DryLoc ph/orp Sensor Electronics
Signet 750 DryLoc ph/orp Sensor Electronics *750.090* 750.090 Rev. G /05 English WARNING! SAFETY INSTRUCTIONS. Depressurize and vent system prior to installation or removal.. Confi rm chemical compatibility
More informationNN1 PH1 CC1 DM1 DB1. TABLE V - Options. Mounting Hardware. Instruction Books NN2 PH2 CC2 DM2 DB2. Certificates. PID Control NN C3.
0---0 January 00 Quick Start Guide for UDA Universal Dual Analyzer For detailed instructions see UDA Universal Dual Analyzer Product Manual 0---. Step. Model Number Interpretation Write your analyzer model
More information5450 NW 33rd Ave, Suite 104 Fort Lauderdale, FL Fruitland Ave Los Angeles, CA SST7000 SST7100. Speed Switch / Transmitter
5450 NW 33rd Ave, Suite 104 Fort Lauderdale, FL 33309 3211 Fruitland Ave Los Angeles, CA 90058 SST7000 SST7100 Speed Switch / Transmitter Installation and Operation Manual Rev. C P/N145F-13112 PCO 00009270
More informationBXD17 Series. Features. Analytical instrumentation for your measurement applications.
BXD17 Series Analytical instrumentation for your measurement applications. Features Single parameter instrument with temperature measurement capabilities Custom IP66 NEMA 4X (144 x 144mm) enclosure Versatile
More informationProSense DPM2 Series. Digital Panel Meter. User Manual. for process, temperature and resistance input signals. Models: DPM2-AT-HL DPM2-AT-2R-HL
User Manual ProSense DPM2 Series Digital Panel Meter for process, temperature and resistance input signals Models: DPM2-AT-HL DPM2-AT-2R-HL Scan or click the QR code for a series of Configuration and Programming
More informationConductivity-Panelmeter LF 9648 Conductivity measurement with 2-and 4-electrode cells
Conductivity-Panelmeter LF 9648 Conductivity measurement with 2-and 4-electrode cells Features M LED-Display 14.2 mm red M Indicating range 2000(0) Digit M Measuring ranges programmable from 0... 2.000
More informationAnalyzer/Controller. ph/orp HART. Model 54e ph/orp ESSENTIAL INSTRUCTIONS WARNINGS RISK OF ELECTRICAL SHOCK
Instruction Sheet PN 51A-54epH/rev J October 2010 ph/orp HART Model 54e ph/orp Analyzer/Controller For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. ESSENTIAL
More informationThe PM1000 series is a universal 4 digit LED plug-on display for transmitters with 4-20mA 2 wire output and fitted with DIN43650 connector.
PM1000 SERIES PLUG-ON DISPLAY BRIGHT LED DISPLAY INDICATION RANGE -999 TO +9999 FITS TO DIN 43650 CONNECTOR PLUG-ON TO ANY TRANSMITTER WITH 4-20MA OUTPUT EASY TO SCALE ON SITE ROBUST DESIGN SET POINT OPTION
More informationTempco Instruction Manual
Tempco Instruction Manual 1/16 DIN Solid State Temperature Controller Relay Output Solid State Output For Heating Model Numbers: TEC-901, TEC-902, TEC-905 Temperature controls in this series are designed
More informationModel T28 Transmitter
Model T28 Transmitter Product Training Presented by: Joe Bradley October 2009 ELECTRO-CHEMICAL DEVICES Model T28 Transmitter Two Wire Transmitter Loop Powered 24 VDC @ 500 Ω Measurement Choices ph ORP
More informationSignet BC Batch Controller System
Signet 9900-1BC Batch Controller System *3-9900-1BC.090* 3-9900-1BC.090-1 Rev. A1 03/13 English Product Manual Look for the Quick Start icon to quickly set up your new 9900-1BC. Safety WARNING! Panel Mount
More informationOperation Manual CX20 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller
Operation Manual CX20 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller 11751 Markon Drive Garden Grove, CA. 92841 U.S.A. pg. 1 CONTENTS GENERAL INTRODUCTION.....3 INITIAL INSPECTION......3
More informationLAUREL ELECTRONICS, INC.
LAUREL ELECTRONICS, INC. Laureate RTD Temperature Panel Meter / Controller Features Factory calibrated for 100Ω platinum, 10Ω copper & 120Ω nickel RTDs 2, 3 or 4-wire connection with lead resistance compensation
More informationDescription, continued
DL424/425 DirectLine Sensor Module for DL5000 Dissolved Oxygen Probes Overview 70-82-03-48 4/02 Page of 8 Specification The DirectLine DL424/425 Sensor Modules are part of a family of products developed
More informationModel Q45C2 Two-Electrode Conductivity Monitor/Analyzer
Model Q45C2 Two-Electrode Conductivity Monitor/Analyzer Home Office European Office Analytical Technology, Inc. ATI (UK) Limited 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park Collegeville, PA 19426
More information7511 / / / 711 / / 731 / Preset 27
22 Timers Timers Electronic 23 7511 / 3410 23 6320 / 720-6300 24 ElectRo-MeCH 25 710 / 711 / 720 25 722 / 731 / 732 26 Preset 27 7932 27 Electronic 23 7511 Series 8 digit self powered LCD electronic timer
More informationMODEL 692E TWO-WIRE ELECTRODELESS CONDUCTIVITY TRANSMITTER
Manual No. 692E Revision 5-598 MODEL 692E TWO-WIRE ELECTRODELESS CONDUCTIVITY TRANSMITTER (selectable for conductivity or % concentration measurement) INSTRUMENT SETUP GUIDE This manual contains detailed
More informationMulti-Point Gas Detection and Control System
Multi-Point Gas Detection and Control System Specifications subject to change without notice. USA 40804 Page of 7 DESCRIPTION Wall mounted, microprocessor-based, multi-point, analog electronic control
More informationHT Series Wall Mount with Relay Setpoints, LCD, and Humidistat and Thermostat Control
Wall Mount with Relay Setpoints, LCD, and Humidistat and Thermostat Control NOTICE This product is not intended for life or safety applications. Do not install this product in hazardous or classified locations.
More informationMulti-Point Gas Detection and Control System
Multi-Point Gas Detection and Control System DESCRIPTION Wall mounted, microprocessor-based, multi-point, analog electronic control system for various gas, temperature and humidity detection, control and
More informationABB MEASUREMENT & ANALYTICS DATA SHEET. TB84PH ph /ORP / pion transmitter
ABB MEASUREMENT & ANALYTICS DATA SHEET TB84PH ph /ORP / pion transmitter 2 TB 8 4PH P H /ORP / P I ON TR AN S MIT TE R DS/ TB8 4PH-E N RE V. D Measurement made easy Hazardous area rated transmitter with
More informationDIGITAL PANEL THERMOMETER
Love Controls DIGITAL PANEL THERMOMETER LCI108-1x LCI108J-1x INSTRUCTION MANUAL EDITION: January 2001 INTRODUCTION TO THE LOVE CONTROLS LCI108 SERIES This manual does not constitute a formal agreement.
More informationThe IQ240 panel mount load cell indicator is a precision digital indicator for load cell and strain gauge applications.
IQ240 Panel Mount Load Cell Indicator Data sheet English 1.01 Introduction The IQ240 panel mount load cell indicator is a precision digital indicator for load cell and strain gauge applications. The high
More information1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882
1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882 This manual is intended for use in support of installation, commissioning and configuration of the 1/32-DIN Temperature
More informationMYRIAD QLC 4-CHANNEL MONITOR/CONTROLLER INSTRUCTION MANUAL
MYRIAD QLC 4-CHANNEL MONITOR/CONTROLLER INSTRUCTION MANUAL VISIT OUR WEBSITE SIGMACONTROLS.COM MYR QLC MANUAL 013114 2 TABLE OF CONTENTS INTRODUCTION 3 Ordering Information Specifications Features WIRING
More informationTrident and Trident X2 Digital Process and Temperature Panel Meter
Sign In New User ISO 9001:2008 Certified Quality System Home Products Online Tools Videos Downloads About Us Store Contact Policies Trident and Trident X2 Digital Process and Temperature Panel Meter Products
More information12-36 VDC/12-24 VAC Power Option 4-Digit Display, 0.56 (14.2 mm) or 1.20 (30.5 mm)
4-20 ma & Relay Output Features 4-20 ma, ± 10 V, TC & RTD Inputs 12-36 VDC/12-24 VAC Power Option 4-Digit Display, 0.56 (14.2 mm) or 1.20 (30.5 mm) 24 VDC @ 200 ma Transmitter Power Supply Options Type
More informationTable of Contents. Introduction... 2 Description...2 Specifications...3 Installation.. 5 Calibration.. 6 General Maintenance...7
User's Manual: Series 330I Model 330I DC-Powered Three-Way Isolator Table of Contents Page Introduction... 2 Description....2 Specifications....3 Installation.. 5 Calibration.. 6 General Maintenance...7
More informationC600 2-wire Conductivity Transmitter
C600 2-wire Conductivity Transmitter Installation and Operating Instructions 900596-C - Sensolution C600 Isolated 2-wire transmitter instruction manual.doc 1 of 24 1. Contents 1. Contents... 3 2. Introduction...
More informationAnalog Voltage Combination Module
Analog Voltage Combination Module T1F-8AD4DA-2 8-channel analog voltage input 4-channel analog voltage output The combination 8-in and 4-out voltage module uses a T1K-8B or T1K-8B-1 base, which is purchased
More informationMulti-Point Gas Detection and Control System
Multi-Point Gas Detection and Control System Specifications subject to change without notice. USA 50729 Page of 9 DESCRIPTION Wall mounted, microprocessor-based, multi-point, analog electronic control
More informationLAUREL. Laureate RTD Temperature Panel Meter / Controller ELECTRONICS, INC. Features. Description. Specifications
LAUREL ELECTRONICS, INC. Laureate RTD Temperature Panel Meter / Controller Features Factory calibrated for 100Ω platinum, 10Ω copper & 120Ω nickel RTDs 2, 3 or 4-wire connection with lead resistance compensation
More informationLAKEWOOD INSTRUMENTS MODEL 1575e WATER TREATMENT CONTROLLER
LAKEWOOD INSTRUMENTS MODEL 1575e WATER TREATMENT CONTROLLER The Model 1575e uses the latest in microprocessor capability, giving the user a high level of application flexibility. A large illuminated graphics
More informationH3CR-F H3CR-F8 H3CR-F-300 H3CR-F8-300
R Solid-State Repeat-Cycle Timer 1/1 DIN Solid-State Repeat-Cycle Timer Wide power supply ranges of 100 to 240 VAC 24 VAC/VDC, 12VDC Combinations of independent long or short / time settings are possible
More informationMODEL: 43AL1. Digital Panel Meters 43 Series
Digital Panel Meters 43 Series LOOP POWERED DIGITAL PANEL METER (process meter) Functions & Features Loop powered -1999 to 9999 compact digital meter Scaling function Screwless spring terminal 24 (.94)
More informationCDTX-45T Portable Toroidal Conductivity Meter
CDTX-45T Portable Toroidal Conductivity Meter Table of Contents TABLE OF FIGURES... 4 PART 1 - INTRODUCTION... 5 1.1 General... 5 1.2 Features... 5 1.3 CDTX-45T System Specifications... 7 1.4 CDTX-45T
More informationCarbon Monoxide (CO) Gas Detection and Control System
Carbon Monoxide (CO) Gas Detection and Control System DESCRIPTION Gas monitor with built-in carbon monoxide (CO) sensor, wall-mounted, accepts inputs from remote devices such as other gas sensors, temperature
More informationInstruments and Capabilities (Panel Mount and Standalone) By: Thornton Leading Pure Water Analytics
Instruments and Capabilities (Panel Mount and Standalone) Table of Instrument Models and Capabilities. ph & Dissolved Instrument System Resistivity Conductivity ORP Oxygen Flow 770MAX Multiparameter Analyzer/Transmitter
More informationThe information in this chapter will enable you to: Understand the product's basic functions & features
C H A P T E R ➀ Chapter Objective Product Description The information in this chapter will enable you to: Understand the product's basic functions & features Introduction The RP240 is designed to operate
More informationThermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter
English Cond Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the appropriate wall socket
More informationDINALOG A 144 x 36 A1400 Light-Strip Indicator
DINALOG A x 6 A00 Light-Strip Indicator -9-0-0 /.00 Front panel dimensions: x 6 mm Light-strip indicator with 7 high-contrast LEDs Red LED display color Digital display range for portrait format: 999 to
More informationZone Pressure Touch (ZPT) Sensor - Standard Range
Product Identification BAPI s Touch Pressure Sensor is an accurate, rugged and economical solution for measuring and reporting duct/ building static pressure, room-to-room differential pressure or air
More informationLAUREL. Laureate Digital Panel Meter for Process and Ratiometric Signals ELECTRONICS, INC. Features. Description
LAUREL ELECTRONICS, INC. Laureate Digital Panel Meter for Process and Ratiometric Signals Features Reads process signals from ±200 mv to ±600V or ±2 ma to ±5A full scale Ratiometric mode for bridges and
More informationOperating Manual. Dynisco /8 DIN Indicator Concise Product Manual
Dynisco 1490 1/8 DIN Indicator Concise Product Manual 59476-9 Operating Manual -1- CAUTION: Installation should be only performed by technically competent personnel. Local Regulations regarding electrical
More informationV E1B Snap-in I/O Module
V200-18-E1B Snap-in I/O Module The V200-18-E1B plugs directly into the back of compatible Unitronics OPLCs, creating a selfcontained PLC unit with a local I/O configuration. Features 16 isolated digital
More informationH7ER H7ER-V H7ER-V1 H7ER-V2 H7ER-V3
Tachometers Subminiature Tachometer Requires No External Power Supply Subminiature 8 x 2 mm (1.89 x.9 in) Improved noise immunity Screw terminal and wire-wrap types available DC, and no-voltage input Printed
More informationV E2B Snap-in I/O Module
V200-18-E2B Snap-in I/O Module The V200-18-E2B plugs directly into the back of compatible Unitronics OPLCs, creating a selfcontained PLC unit with a local I/O configuration. Features 16 isolated digital
More informationGFC114. Low Power Flow Computer FEATURES
Low Power Flow Computer FEATURES l UNIVERSAL INPUT, 5 khz, CAN POWER THE l SIMPLIFIED MENUS AND PROGRAMMING l NO NEED OF AMPLIFIERS OR LINEARIZERS l SEVEN DIGIT RATE WITH PROGRAMMABLE DECIMAL PLACES l
More informationHT Series Wall Mount with Analog Setpoints, LCD, and Humidistat and Thermostat Control
Wall Mount with Analog Setpoints, LCD, and Humidistat and Thermostat Control NOTICE This product is not intended for life or safety applications. Do not install this product in hazardous or classified
More informationUser Guide. ExStik II EC400. Conductivity / TDS / Salinity / Temperature Meter
User Guide ExStik II EC400 Conductivity / TDS / Salinity / Temperature Meter Introduction Congratulations on your purchase of the ExStik II EC400 Conductivity/Total Dissolved Solids (TDS) / Salinity /
More informationAdapter Kit - QSI Electronics to TM Meter Installation Instructions
Adapter Kit - QSI Electronics to TM Meter Installation Instructions These installation instructions cover the adapter kit installation of the QSI1, QSI2 and QSI3 versions of communications electronics
More informationSelf-Powered Counters H7EC
Self-Powered Counters Subminiature Count Totalizers Require No External Power Supply Subminiature 8 x mm (1.89 x.9 in) Improved noise immunity Screw terminal and wire-wrap types available AC, DC, and no-voltage
More informationPLEASE READ INSTRUCTIONS CAREFULLY BEFORE INSTALLATION!
TUC2 / TUCH2 Model Series Installation Instructions PLEASE READ INSTRUCTIONS CAREFULLY BEFORE INSTALLATION! APPLICATION The TUC2 provides temperature space monitoring with a backlit LCD. The TUCH2 provides
More information