1 Technical Publications SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET for ARTIS Q / Q.ZEN / ONE ARTIS ZEEGO / ZEEGO.ZEN ARTIS ZEE MULTIPURPOSE SINGLE PLANE IECAX480V125A Main Disconnect Panel

2 Bevco Engineering Company, Inc. CONFIDENTIAL AND PROPRIETARY This Bevco Engineering Company document contains confidential and proprietary information of Bevco Engineering Company. The intellectual properties contained herein may not be used or copied to the detriment of Bevco Engineering Company. The information contained in this document may not be copied without the express written approval of Bevco Engineering Company. CAUTION! Troubleshooting and servicing should be performed by a qualified electrician. WARNING! De-energize and lockout the main disconnect before servicing this panel. Table of Contents 1.0 DESCRIPTION OF OPERATION SYSTEM CONFIGURATION SPECIFICATIONS... 5 ELECTRICAL SPECIFICATIONS:... 5 SEISMIC SPECIFICATIONS:... 5 ENVIORNMENTAL SPECIFICATIONS:... 5 ELECTRICAL WIRING DIAGRAM: PANEL DIMENSIONS... 6 CENTER OF GRAVITY... 7 MOUNTING HOLE DIMENSIONS INSTALLATION TESTING and TROUBLESHOOTING MAINTENANCE REPLACEMENT PARTS DOCUMENT REVISION HISTORY COMPONENTS SUPPLIED WITH EACH PANEL: The IEC Main Disconnect Panel This Installation, Operations & Service Manual (3) sets of Emergency Power Off pushbuttons with 2NC contacts on each EPO Drawings and Electrical Schematics IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 2 of 14

3 1.0 DESCRIPTION OF OPERATION INTEGRATED ELECTRICAL CABINET This panel incorporates several features desirable for system installations to minimize down time, protect the X-Ray Generator electronics, and to reduce operational delays after a power outage. This product replaces the power panel, the lighting control relay box, and the Emergency Power Off controls. The panel has a main circuit breaker, Q1, provides fully integrated X-Ray ON warning light control and a relay to reduce the room lighting during the procedure. When the main circuit breaker is turned off, all power circuits within the panel will be de-energized. Q1 provides the disconnect means and lock-out and tag-out (LOTO) the X-Ray Generator power circuit for maintenance purposes. The K2 contactor will open with any loss of power or by pressing any Emergency Power Off (EPO) pushbutton. The contactor control circuit is factory configured to automatically re-energize the X-Ray Generator upon restoration of facilities power. The control circuit may be re-configured to require the operator to manually restart the equipment once the incoming power has been restored. This protects the sensitive electronic circuits of the X-Ray Generator from sags and surges that immediately follow power loss from blackouts, storms, utility reclosure operations, and out of phase automatic transfer switch operations. The SC1 cabinet is protected by an electronic circuit breaker, Q3, and controlled by contactor K6. The contactor will open with any loss of power or by pressing any Emergency Power Off (EPO) pushbutton. The control circuits for the EPOs are low voltage 24 VDC and are fully powered from within the panel. The restart functionality and EPO circuitry is controlled with a safety relay, K10. See page 11 for detail. The white SAFETIES OK indicator light on the front of the panel is illuminated when none of the EPOs are pressed. When the white light is active, pressing the green START pushbutton will cause the X- Ray Generator to be energized. The green START button will illuminate, and the white SAFETIES OK light will go off. Pressing the STOP button will de-energize the system. Any EPO pressed while the system is energized will result in the immediate de-energizing of the X-Ray Generator system. If an EPO is pressed at any time, the EPO must be reset which will cause the SAFETIES OK light to activate. Then the START button will activate the system. The EPOs provided with the IEC include additional Normally Closed contacts for connections to a UPS system. IMPORTANT: If building power is removed from the panel while the X-Ray system is energized, the power to the X- Ray system will be restored when building power returns without any human interaction. The X-Ray system can then be restarted normally. Additional provisions are made to integrate the X-Ray ON warning lights and room lighting with the X- Ray Equipment. The facility lighting panel provides 120- or 277-volt power that is controlled by contacts relays of K4 and K5 mounted in the IEC. The signal controlling the relays comes directly from the Siemens Generator/Power Cabinet. The IEC will accept signals from the generator as either Dry Contacts or 24 Volts AC or DC. Note the maximum connected load to the X-Ray On lights is 20 Amps. IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 3 of 14

4 2.0 SYSTEM CONFIGURATION UTILITY SUPPLY 480V 3-PHASE 100A MAX Q1 125A WIRESIZE: 6-3/0AWG KEY TO COMPONENTS: Q1 MAIN CIRCUIT BREAKER Q2 GENERATOR 2 CIRCUIT BREAKER Q3 GENERATOR 1 CIRCUIT BREAKER F2 POWER SUPPLY FUSES F4 DC CIRCUIT BREAKER G01 DC POWER SUPPLY K1 SAFETY CONTACTOR K2 X-RAY GENERATOR CONTACTOR K4 ROOM LIGHT RELAY K5 X-RAY ON WARNING LIGHT RELAY K6 SYSTEM CABINET CONTACTOR Q2 100A F2 4A Q3 ADJUSTABLE 30A G01 DC POWER SUPPLY K6 30A WIRE SIZE: 14-8AWG F4 5A SC1 DOOR SWITCH CONTROL ROOM LIGHT CONTROL X-RAY ON CONTROL 30A MAX K2 185A MAX WIRE SIZE: 10-2/0AWG DC CONTROL CIRCUITS K1 SAFETY CONTACTOR K4 ROOM LIGHT RELAY K5 WARNING LIGHT RELAY EMERGENCY POWER OFF CIRCUITS X-RAY ON WARNING LIGHT 20A MAX LIGHTING POWER PANEL 120/277V GENERATOR 1 100A MAX EPO1 EPO2 EPO3 ROOM LIGHTS 20A MAX IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 4 of 14

5 3.0 SPECIFICATIONS ELECTRICAL SPECIFICATIONS: UL design and ratings of this panel require the application of wire at the 75 C ampacity ratings of NEC table Wire rated at higher temperature ratings, such as 90 C, may be used but ONLY applied at the 75 C ampacity ratings of NEC Table Catalog Number Incoming Voltage & Frequency Incoming Power Configuration Main Circuit Breaker Rating Regulatory Markings IECAX480V125A 480VAC, 50/60Hz 3Ø + Ground + Neutral Amps culus Short Circuit Current Rating is 35,000 Amperes RMS 480V SEISMIC SPECIFICATIONS: This Integrated Electrical Cabinet has been certified by an independent California structural engineer in conformance with the shake testing requirements of ICC-AC 156. The California OSHPD number is OSP The seismic performance characteristics are as follows: SDS(g) 2.56; z/h 1.0 ; Ip 1.5 ENVIORNMENTAL SPECIFICATIONS: FOR INDOOR USE ONLY Temperature F (15-32 C) Humidity 30-75%, NON-CONDENSING ELECTRICAL WIRING DIAGRAM: The Electrical Wiring Diagram is shipped along with the disconnect panel inside the enclosure door for convenient reference to the circuitry. PARTS, TECHNICAL SUPPORT AND SERVICE FOR INQUIRIES DURING INSTALLATION CONTACT BEVCO ENGINEERING, SUSSEX WI (262) FOR ISSUES FOLLOWING INSTALLATION, CONTACT SIEMENS UPTIME CENTER, (800) IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 5 of 14

6 4.0 PANEL DIMENSIONS Refer to the Table and Illustrations below for the physical dimensions of the panel. The panel may be recessed into the wall up to 5 inches for semi-flush installations. See Illustrations below for typical front and side views with the mounting hole pattern. Refer to the drawing set provided with the panel for total weight and center of gravity dimensions. PANEL EXTERNAL PHYSICAL DIMENSIONS Parameter Dimensions Height 30.0 inches 762 mm Width 20.0 inches 508 mm Total Depth 9.44 inches 244 mm Enclosure Depth 8.0 inches 203 mm Mounting Hole Width inches 467 mm Mounting Hole Height 28.5 inches 740 mm Weight 67 pounds 30.4 kg IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 6 of 14

7 CENTER OF GRAVITY IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 7 of 14

8 MOUNTING HOLE DIMENSIONS INTEGRATED ELECTRICAL CABINET IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 8 of 14

9 5.0 INSTALLATION The enclosure must be mounted at a height so the top of the main circuit breaker handle, when in the ON position, does not exceed 6ft-7in. (2m) above the floor. (Per National Electric Code (NEC # 404.8). Incoming power is connected to the main circuit breaker located at the upper right of the Main Disconnect Panel. Refer to the schematic drawing for the incoming conductor size range and tightening torques. There are no conduit knockouts provided in the panel. The installer must punch the conduit holes in the enclosure walls in the desired locations. The conduits may enter on the top, bottom, either side, or in designated areas from the back of the enclosure. NOTE: When drilling or punching the conduit entry holes, protect the internal components from the falling metal chips. All conductors must be sized per the wiring shown on the installation drawings or in the system configuration diagram. 1. Determine the system configuration. 2. Note the model of X-Ray Generator. It will have a rating of 65kW or 80kW. 3. Connect the incoming power to the Line side of the Main CB, Q1, terminals L1, L2, and L3. 4. Connect the X-Ray Generator directly to the load side of contactor K2, terminals 2T1, 4T2 and 6T3. 5. Connect the System Cabinet SC1 directly to the load side of contactor K6, terminals 2T1, 4T2 and 6T3. 6. Connect the IEC Emergency Power Off (EPO) circuits to terminals 1 and 2 of X02. All EPOs should have normally closed contacts wired in series. See EPO wiring diagram. 7. Room light controls are on terminals X05 1 through 4. Verify first if the Siemens System Cabinet is providing a dry contact or 24V AC or DC. If 24V AC or DC is provided, connect the two wires coming from the Siemens System Cabinet to terminals 1 and 2 of X05. If a dry contact is provided, connect the two wires coming from the Siemens System Cabinet to terminals 3 and 4 of X05. See Lighting Wiring Diagram. 8. Connect Room light and lighting panel to terminals 5 and 6 of X05. See Lighting Wiring Diagram. 9. X-Ray On controls are on terminals X06 1 through 4. Verify first if the Siemens System Cabinet is providing a dry contact or 24V AC or DC. If 24V AC or DC is provided, connect the two wires coming from the Siemens System Cabinet to terminals 1 and 2 of X06. If a dry contact is provided, connect the two wires coming from the Siemens System Cabinet to terminals 3 and 4 of X06. See Lighting Wiring Diagram. 10. Connect Warning light and lighting panel to terminals 5 and 6 of X06. See Lighting Wiring Diagram. 11. Connect the Door Switch DSW1 to terminals 1 and 2 of X07. See Door Switch Wiring Diagram. 12. Connect the Door Switch Connections of the Siemens System Cabinet SC1 to the terminals 3 and 4 of X07. See Door Switch Wiring Diagram. IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 9 of 14

10 INSTALLATION (continued) TERMINAL LAYOUT (Located in the bottom left of the Disconnect Panel) EPO WIRING IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 10 of 14

11 LIGHTING WIRING DOOR SWITCH WIRING IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 11 of 14

12 6.0 TESTING and TROUBLESHOOTING INTEGRATED ELECTRICAL CABINET WARNING! PRIOR TO TESTING THIS PANEL, ENSURE NO ONE IS WORKING ON OR NEAR THE EQUIPMENT THAT THIS PANEL FEEDS. PARTS, TECHNICAL SUPPORT AND SERVICE FOR INQUIRIES DURING INSTALLATION CONTACT BEVCO ENGINEERING, SUSSEX WI (262) FOR ISSUES FOLLOWING INSTALLATION, CONTACT SIEMENS UPTIME CENTER, (800) Symptom No Power to the X-Ray Generator Possible Solutions If attempting to energize the X-Ray immediately after applying power, the system will not turn on; ten seconds must pass before starting the system in order to allow the safety relay to start up. Verify the 3Ø incoming power is present at the top of Q1. Verify the Main Circuit Breaker is ON. Verify the DC power supply circuit breaker (F2) is on and the DC power supply is producing 24VDC. Verify the control power circuit breaker (F4) is on. Verify S11 (Stop) is not depressed. Verify no Emergency Stop Pushbuttons are depressed and they are correctly wired to terminals X02. Check the four lights on the front of the safety relay, when ready, the DEVICE light should be solid green. When all e-stops are released, the IN light should be solid green and the Safeties OK light on the front of the panel should be lit. Press the Start button, contactor K2 should pull in. If not, check the OUT light on the safety relay, if not solid green, press and hold the RESET button for five seconds then release. The lights should flash and the DEVICE and IN light should be green. Press the Start button, contactor K2 should pull in. If the DEVICE light is red, the safety relay is defective and must be replaced. For other Safety Relay Conditions, refer to the diagram and chart next page. PDA Off circuit inoperable Verify the PDA field wiring is not crossed between terminals X03 and X04. Verify 24Vdc is present at X03 Verify Safeties OK is lit on the front door. If not, see above. X-Ray ON Warning Light does not turn on as expected Room lights does not turn off as expected Verify the circuit breaker from the lighting panel is turned on. Verify the signal, either 24 Volts AC or 24 Volts DC, is present on terminals 1 and 2 of terminal block X06. Verify Continuity between terminals 3 and 4 of terminal bloc X06 if dry contact provided by Siemens System Cabinet. Verify K5 is off Verify the circuit breaker from the lighting panel is turned on. Verify the signal, either 24 Volts AC or 24 Volts DC, is present on terminals 1 and 2 of terminal block X05. Verify Continuity between terminals 3 and 4 of terminal bloc X05 if dry contact provided by Siemens System Cabinet. Verify K4 is on. IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 12 of 14

13 K10 Safety Relay configurations The Safety Relay DIP switches are set as follows from the factory On On Off On On On On Off IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 13 of 14

14 7.0 MAINTENANCE No components within this panel require adjustment or periodic servicing. 8.0 REPLACEMENT PARTS PARTS, TECHNICAL SUPPORT AND SERVICE FOR INQUIRIES DURING INSTALLATION CONTACT BEVCO ENGINEERING, SUSSEX WI (262) FOR ISSUES FOLLOWING INSTALLATION, CONTACT SIEMENS UPTIME CENTER, (800) DOCUMENT REVISION HISTORY REV DATE AUTHOR DESCRIPTION OF CHANGE 1 10/03/2013 RJB Initial Release 2 02/27/2015 RJB Added Artis Tradenames on first page 3 03/09/2015 XYUE Update the manual based on schematic 50 03/30/16 DRO IEC Redesign 51 07/10/18 JT Update of system config data IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 14 of 14

SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET

SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET Product Data Sheet Technical Publications SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET for ARTIS Q / Q.ZEN / ZEE BIPLANE IECAX480V225A Main Disconnect Panel Bevco Engineering Company, Inc. CONFIDENTIAL

More information

User s Manual. ACS550-PD Stock 3R Irrigation Packaged Drive Supplement to ACS550-U1 User s Manual

User s Manual. ACS550-PD Stock 3R Irrigation Packaged Drive Supplement to ACS550-U1 User s Manual User s Manual ACS550-PD Stock 3R Irrigation Packaged Drive Supplement to ACS550-U1 User s Manual 2 ACS550-PD 3R Irrigation Packaged Drive ACS550 Drive Manuals GENERAL MANUALS ACS550-U1 User s Manual (1

More information

User s Manual. ACS550-PD 3R Irrigation Packaged Drive Supplement to ACS550-U1 User s Manual

User s Manual. ACS550-PD 3R Irrigation Packaged Drive Supplement to ACS550-U1 User s Manual User s Manual ACS550-PD 3R Irrigation Packaged Drive Supplement to ACS550-U1 User s Manual 2 ACS550-PD 3R Irrigation Packaged Drive ACS550 Drive Manuals GENERAL MANUALS ACS550-U1 User s Manual (1 200 HP)

More information

HUBBELL SPIKESHIELD Wired-In Branch Panels Surge Protective Device (SPD)

HUBBELL SPIKESHIELD Wired-In Branch Panels Surge Protective Device (SPD) HUBBELL SPIKESHIELD Wired-In Branch Panels Surge Protective Device (SPD) Instruction Bulletin 8222-0513B Retain for future use. SPIKESHIELD Wired-In Branch Panels 8222-0513B 05/2012 Table of Contents Precautions...

More information

Quick Start Guide TS A

Quick Start Guide TS A Quick Start Guide TS 930 125-630A DANGER HAZARD OF ELECTRICAL SHOCK, EXPLOSION, OR ARC FLASH Read and understand this quick start guide before installing and operating the transfer switch The installer

More information

APPLICATION CONTROL GUIDELINES. IntelliROL Power Supply PN Revision Date: March 15, 2017

APPLICATION CONTROL GUIDELINES. IntelliROL Power Supply PN Revision Date: March 15, 2017 APPLICATION CONTROL GUIDELINES IntelliROL Power Supply PN 1176718 Revision Date: March 15, 2017 Table of Contents List of Tables...3 TGW Safety Recommendation...4 Warnings and Safety Instructions...5 Introduction...6

More information

Specifications, Installation, and Operating Instructions

Specifications, Installation, and Operating Instructions Specifications, Installation, and Operating Instructions Model: Power Vantage AC Panel Protector CAUTION: The installation of a surge protection device () must be done by qualified electrical personnel.

More information

Surge Protective Devices Installation, Operation and Maintenance Manual. LoadTrack LTL

Surge Protective Devices Installation, Operation and Maintenance Manual. LoadTrack LTL LoadTrack LTL Surge Protective Devices Installation, Operation and Maintenance Manual P.O. Box 3760 Winter Park, FL 32790 USA TEL: 800-647-1911 www.tpssurge.com LOADTRACK LTL INSTALLATION, OPERATION AND

More information

Expandable Power Systems

Expandable Power Systems Expandable Power Systems Installation Guide Models Include: Maximal11EV 12VDC @ 4A or 24VDC @ 3A. 12VDC @ 4A or 24VDC @ 3A. Maximal33EV 12VDC or 24VDC @ 6A. 12VDC or 24VDC @ 6A. Maximal37EV 24VDC @ 10A.

More information

DANGER DANGER: is used in this manual to warn of high voltages capable of causing shock, burns, or death.

DANGER DANGER: is used in this manual to warn of high voltages capable of causing shock, burns, or death. Installation/Operation Manual TABLE OF CONTENTS Page UNPACKING AND INSTALLATION Unpacking and Preliminary Inspection...2, 3 Model Number...9 LOCATION CONSIDERATIONS Environment, Equipment Performance,

More information

Always "On" UPS Systems Inc. USER MANUAL

Always On UPS Systems Inc. USER MANUAL Always "On" UPS Systems Inc. SMART HOME PROTECTOR USER MANUAL Bldg 1-150 Campion Road, Kelowna, BC, Canada, V1X 7S8 Phone: (250) 491-9777 Ext. 451, Fax: (250) 491-9775, Email: sales@alwaysonups.com Web

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS LIGHTING CONTROL PANELS 16 AND 24 RELAYS INSTALLATION INSTRUCTIONS INSTALLATION OVERVIEW The installation instructions contained in this document are provided as a guide for proper and reliable installation.

More information

Models 120LS, 200LS, 300LS, 400LS, 560LS Installation Instructions

Models 120LS, 200LS, 300LS, 400LS, 560LS Installation Instructions 299-600-96 Rev. F MCG Surge Protection Models 120LS, 200LS, 300LS, 400LS, 560LS Installation Instructions Important Warranty Information: MCG surge protectors are designed to work at specific voltages

More information

920 Remote Control (RC) Switches

920 Remote Control (RC) Switches 920 Remote Control (RC) Switches ASCO is the global leader in power switching and controls ASCO 920 PRODUCT SPECIFICATIONS SWITCH ONLY (PANELBOARD MOUNTING) 1,2,3 () DIMENSIONS, IN.(MM) WIDTH HEIGHT DEPTH

More information

PDI Modular Compact Remote Power Panel

PDI Modular Compact Remote Power Panel Guide Specifications (Revision 000, 6/17/2016) 1 GENERAL 1.1 Summary These specifications describe requirements for a remote power panel (RPP) distributing power to sensitive loads. These specifications

More information

PowerFlex 400 AC Drive Guide Specification

PowerFlex 400 AC Drive Guide Specification PowerFlex 400 AC Drive Guide Specification Adjustable Frequency Drives with Bypass 3.0 50HP @ 208V AC 3.0 350HP @ 480V AC PART 1 GENERAL 1.01 Quality Assurance A. The manufacturer shall have minimum 5

More information

ASCO 920 Remote Control (RC) Switches

ASCO 920 Remote Control (RC) Switches Maximum reliability and excellent value ASCO 920 Remote Control (RC) Switches ASCO 920 Remote Control Switch in Type Enclosure The ASCO 920 is designed as a feeder disconnect switch for load capacities

More information

(1986; Rev. 1989) Molded Case Circuit Breakers and Molded Case Switches. (1985; Rev. 1988) Enclosures for Electrical Equipment (1000 Volts Maximum)

(1986; Rev. 1989) Molded Case Circuit Breakers and Molded Case Switches. (1985; Rev. 1988) Enclosures for Electrical Equipment (1000 Volts Maximum) Section 16306 50/60-HERTZ (HZ) SOLID STATE FREQUENCY CONVERTER Index PART 1 GENERAL PART 2 PRODUCTS 1.1 References 2.1 Frequency Converters 1.2 Factory Tests 2.2 Conduit and Fittings 1.3 Submittals 2.3

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS TM REPLACEMENT PARTS FOR USE WITH CX SERIES PANELS INSTALLATION INSTRUCTIONS Hubbell Control Solutions 9601 Dessau Road Building One Suite 100 Austin, TX 78754 Toll Free: 888-698-3242 Fax: 512-450-1215

More information

Installation Manual. APEX Series. Silicon and MOV Panel Protection. APEX Series

Installation Manual. APEX Series. Silicon and MOV Panel Protection. APEX Series Installation Manual APEX Series Silicon and MOV Panel Protection APEX Series APEX Series Installation Guide 2 Table of Contents Mechanical Installation 3 Table One - Application Rating 3 Figure 1 - Mounting

More information

PowerLogic High Density Metering System 1-Meter Enclosure

PowerLogic High Density Metering System 1-Meter Enclosure PowerLogic High Density Metering System 1-Meter Enclosure Installation Guide 63230-508-211A1 Safety information PowerLogic High Density Metering System 1-Meter Enclosure Important information Read these

More information

ETC Installation Guide

ETC Installation Guide Unison Foundry Overview The Unison Foundry provides fully-rated 20A relays for switched power control with 0 10V dimming for compatible fluorescent ballasts and LED drivers. The provides two outputs of

More information

Installing the Tracer MP503 Input/Output Module Ordering number: ,

Installing the Tracer MP503 Input/Output Module Ordering number: , Installing the Tracer MP503 Input/Output Module Ordering number: 4950 0490, 4950 0590 Product overview The Tracer MP503 input/output (I/O) module is a field-installed device used to monitor inputs and

More information

ControlKeeper 4. General Information. Connecting Relay Loads. Installation Sheet. Getting Started. Power Supply Wiring. Mounting the Cabinet

ControlKeeper 4. General Information. Connecting Relay Loads. Installation Sheet. Getting Started. Power Supply Wiring. Mounting the Cabinet General Information ControlKeeper 4 Installation Sheet Model# CK4-120NO- Model# CK4-277NO The ControlKeeper-4 model is shipped in one package and is configured with either a 120V or a 277V transformer.

More information

Temperature. Power requirements. Weight. Dimensions

Temperature. Power requirements. Weight. Dimensions Installing the Tracer Summit BMTX Building Control Unit Ordering numbers: BMTX001AAB000, BMTX001AAB010, BMTX001AAB001, BMTX001AAB011, BMTX001BAB000, BMTX001BAB010, BMTX001BAB001, BMTX001BAB011, BMTX001DAB000

More information

Line Impedance Stabilization Network (LISN)

Line Impedance Stabilization Network (LISN) Model 3816/2 Line Impedance Stabilization Network (LISN) User Manual ETS-Lindgren Inc. reserves the right to make changes to any product described herein in order to improve function, design, or for any

More information

Total Home Surge Protection for Point-of-Entry. FirstSurge TM. User Manual - Canada (English)

Total Home Surge Protection for Point-of-Entry. FirstSurge TM. User Manual - Canada (English) Total Home Surge Protection for Point-of-Entry FirstSurge TM User Manual - Canada (English) www.siemens.ca/surge Introduction Thank you for choosing the Siemens FirstSurge Residential Surge Protective

More information

Power Supply (48Vdc, 2.5A)

Power Supply (48Vdc, 2.5A) Telecommunications Group Section 855-180-201 Equipment Issue 1 Fourth Printing, September 2006 8551-80 Power Supply (48Vdc, 2.5A) Compliant with UL Standard 60950, Second Edition* CONTENTS PAGE Part 1.

More information

Table of Contents. Safety. Attention. Warnings

Table of Contents. Safety. Attention. Warnings Solstart Miniature Soft Starter 8-58A, 220-600V Instruction Manual Ver. 21.2. 2002 Table of Contents Page Subject 3 Starter Selection 4 Installation Notes 5 Wiring 6 Starter Settings & Start-up Procedure

More information

INSTALLATION INSTRUCTIONS STZ-R SERIES (ENGLISH)

INSTALLATION INSTRUCTIONS STZ-R SERIES (ENGLISH) Surge-Trap STZ-R series Surge Protective Device (SPD) Danger Hazard of Electrical Shock, Explosion or Arc Flash Only qualified licensed electricians should install or service SPDs Verify that all power

More information

MOBILE CONNECTOR - GEN 2 OWNER'S MANUAL

MOBILE CONNECTOR - GEN 2 OWNER'S MANUAL MOBILE CONNECTOR - GEN 2 OWNER'S MANUAL UNITED STATES Contents Safety Information... 2 Save These Important Safety Instructions... 2 Warnings...2 Cautions...3 General Information... 4 Mobile Connector

More information

PanelView Plus/VersaView CE Terminals and Display Modules

PanelView Plus/VersaView CE Terminals and Display Modules Installation Instructions PanelView Plus/VersaView CE Terminals and Display Modules (Catalog Numbers 2711P-xxxxxx, 6182H-xxxxxx) English Inside: Overview...2 For More Information...2 Modular Components...3

More information

Catalog Number Identification with Elements Explained V V V V X 347 V Hz

Catalog Number Identification with Elements Explained V V V V X 347 V Hz Owner s Manual 917 & 918 Remote Control Switches TABLE OF CONTENTS page Ratings & Installation... 1 2 Optional Accessories... 3 4 Troubleshooting & Parts Kits... 5 Wiring & Outline Diagrams... back of

More information

Liebert AccuVar (ACV Series) Surge Protective Device (SPD) (With Noise Filtering) GUIDE SPECIFICATIONS for a Parallel Surge Suppression System

Liebert AccuVar (ACV Series) Surge Protective Device (SPD) (With Noise Filtering) GUIDE SPECIFICATIONS for a Parallel Surge Suppression System Liebert AccuVar (ACV Series) Surge Protective Device (SPD) (With Noise Filtering) GUIDE SPECIFICATIONS for a Parallel Surge Suppression System Part 1 General 1.01 Summary A. These specifications describe

More information

60W Power over Ethernet Waterproof Adapter PoE IEEE BT Single Port Injector for Outdoor Application

60W Power over Ethernet Waterproof Adapter PoE IEEE BT Single Port Injector for Outdoor Application WWW.PHIHONG.COM 60W Power over Ethernet Waterproof Adapter PoE IEEE BT Single Port Injector for Outdoor Application Features Compliant with the IEEE802.3bt Standard Non-Vented Case with Mounting Bracket

More information

PGR-3200 MANUAL INSULATION MONITOR

PGR-3200 MANUAL INSULATION MONITOR Tel: +1-800-832-3873 E-mail: techline@littelfuse.com www.littelfuse.com PGR-3200 MANUAL INSULATION MONITOR REVISION 3-E-040918 Copyright 2018 by Littelfuse, Inc. All rights reserved. Document Number: PM-1025-EN

More information

Installation, Testing, and Operating Procedures 30 AMP PORTABLE AND PERMANENT SERIES GFCI SINGLE and MULTIPHASE

Installation, Testing, and Operating Procedures 30 AMP PORTABLE AND PERMANENT SERIES GFCI SINGLE and MULTIPHASE IMPORTANT! Please read all the information on this sheet. SAVE THESE INSTRUCTIONS! NOTICE BEFORE USING READ INSTRUCTIONS COMPLETELY. TO BE INSTALLED BY A QUALIFIED ELECTRICIAN IN ACCORDANCE WITH NATIONAL

More information

Maximum Reliability & Excellent Value. Features. UL Listed Withstand & Close-On Ratings. Series 300 Power Transfer Switches

Maximum Reliability & Excellent Value. Features. UL Listed Withstand & Close-On Ratings. Series 300 Power Transfer Switches Series 300 Power Transfer Switches Maximum Reliability & Excellent Value With a Series 300 Transfer Switch, you get a product backed by ASCO Power Technologies, the industry leader responsible for virtually

More information

Overview These instructions are presented as a guideline for installing and setting up the LX Series Breaker Control Panel.

Overview These instructions are presented as a guideline for installing and setting up the LX Series Breaker Control Panel. LX Breaker Control Panels Installation and Setup Procedure Hubbell Building Automation, Inc. 9601 Dessau Road Building One Suite 100 Austin, Texas 78754 512-450-1100 512-450-1215 Fax www.hubbell-automation.com

More information

IIM Warwick Court. Williamsburg VA REV 5

IIM Warwick Court. Williamsburg VA REV 5 1 1-757-258-3939 100 Warwick Court Williamsburg VA 23185 REV 5 2 Technical Specification >Accuracy Better than + 1% over a range of 1.5 to 200 amperes with power factor ranging from 1 to 0.5 Approved by

More information

A0/A1 Controller Assembly Operation/Installation Manual READ AND FOLLOW ALL SAFETY INSTRUCTIONS

A0/A1 Controller Assembly Operation/Installation Manual READ AND FOLLOW ALL SAFETY INSTRUCTIONS Document Number: 9100-127-2301-99 Release Date: 2/2/15 A0/A1 Controller Assembly Operation/Installation Manual READ AND FOLLOW ALL SAFETY INSTRUCTIONS DO NOT let any supply cords touch hot surfaces higher

More information

PLT. Breaker-based transfer switch open, closed or soft transition amps. Specification sheet

PLT. Breaker-based transfer switch open, closed or soft transition amps. Specification sheet Specification sheet PLT Breaker-based transfer switch open, closed or soft transition 800 3000 amps Description PLT breaker-based transfer switches are designed for operation and switching of electrical

More information

SAVE THESE INSTRUCTIONS

SAVE THESE INSTRUCTIONS READ AND FOLLOW ALL SAFETY INSTRUCTIONS! SAVE THESE INSTRUCTIONS AND DELIVER TO OWNER AFTER INSTALLATION IMPORTANT SAFEGUARDS! When using electrical equipment, basic safety precautions should always be

More information

QuantumR Light Management Hub

QuantumR Light Management Hub Quantum 369375e 1 04.02.2015 QuantumR Light Management Hub The Quantum light management hub provides a centralized connection point for Lutron EcoSystem digital ballast modules, Lutron power panels, GRAFIK

More information

PGR-4300 MANUAL GENERATOR GROUND-FAULT RELAY

PGR-4300 MANUAL GENERATOR GROUND-FAULT RELAY Tel: +1-800-832-3873 E-mail: techline@littelfuse.com www.littelfuse.com PGR-4300 MANUAL GENERATOR GROUND-FAULT RELAY REVISION 3-B-041318 Copyright 2018 by Littelfuse, Inc. All rights reserved. Document

More information

Model DMX8DIM Installation and Operations Manual

Model DMX8DIM Installation and Operations Manual Model DMX8DIM Installation and Operations Manual Doug Fleenor Design, Inc. 396 Corbett Canyon Road Arroyo Grande, CA 93420 (805) 481-9599 Voice and FAX Manual revision June, 2018 Product description The

More information

AcquiSuite Ally 12 & 48 Advanced Multi-Circuit Meter

AcquiSuite Ally 12 & 48 Advanced Multi-Circuit Meter AcquiSuite Ally 12 & 48 Advanced Multi-Circuit Meter Install Guide Revision C (12/18) DANGER HAZARD OF ELECTRIC SHOCK, EXPLOSION, OR ARC FLASH Revision C (12/18) Turn off all power supplying equipment

More information

QuantumR Light Management Hub (QP2)

QuantumR Light Management Hub (QP2) Quantum 369375f 1 11.05.15 QuantumR Hub (QP2) The Quantum light management hub provides a centralized connection point for Lutron EcoSystem digital ballast modules, Lutron power panels, and QS devices

More information

Installation Instructions & User Manual

Installation Instructions & User Manual Installation Instructions & User Manual ZONEMASTER PRO AC Panel Transient Voltage Surge Suppressors The Wiremold Company In US: 60 Woodlawn Street, West Hartford, CT 06110 800-621-0049 FAX 860-232-2062

More information

Quick Start Guide TS 870 ATS

Quick Start Guide TS 870 ATS Quick Start Guide TS 870 ATS 100A - 800A (with TSC 900 Controller) DANGER HAZARD OF ELECTRICAL SHOCK, EXPLOSION, OR ARC FLASH Read and understand this quick start guide before installing and operating

More information

MAXIMA+ Series Rotary Level Indicator

MAXIMA+ Series Rotary Level Indicator MAXIMA+ Series Rotary Level Indicator BinMaster: Division of Garner Industries 7201 N. 98th St., Lincoln, NE 68507 402-434-9102 email: info@binmaster.com www.binmaster.com OPERATING INSTRUCTIONS PLEASE

More information

Model HM-535 Power Supply Installation and Service Instructions

Model HM-535 Power Supply Installation and Service Instructions Model HM-535 Power Supply Installation and Service Instructions 430-535 0104 2004 Heritage MedCall, Inc SENTRY INSTALLATION & SERVICE INSTRUCTIONS POWER SUPPLY UNIT Model HM-535 IMPORTANT SAFETY INSTRUCTIONS

More information

READ AND FOLLOW ALL SAFETY INSTRUCTIONS! SAVE THESE INSTRUCTIONS AND DELIVER TO OWNER AFTER INSTALLATION IMPORTANT SAFETY INSTRUCTIONS

READ AND FOLLOW ALL SAFETY INSTRUCTIONS! SAVE THESE INSTRUCTIONS AND DELIVER TO OWNER AFTER INSTALLATION IMPORTANT SAFETY INSTRUCTIONS LRC0 8X0A Relay Control Panel BACnet MS/TP Enabled INSTALLATI INSTRUCTIS READ AND FOLLOW ALL SAFETY INSTRUCTIS! SAVE THESE INSTRUCTIS AND DELIVER TO OWNER AFTER INSTALLATI IMPORTANT SAFETY INSTRUCTIS!

More information

CS XL XM. EcoSystem Bus Supply

CS XL XM. EcoSystem Bus Supply EcoSystem An EcoSystem lighting network containing more than one ballast, module, or driver requires an EcoSystem. This supply powers one or two independent EcoSystem buses with up to 64 ballasts or ballast

More information

Surge Protective Devices. STV100K Series. Instruction Manual

Surge Protective Devices. STV100K Series. Instruction Manual Surge Protective Devices STV100K Series Instruction Manual Contents 1.0 Installation... 4 1.1 Environment... 4 1.2 Mounting... 4 1.3 Proper Connection of SPDs... 4 1.4 Mechanical Dimensions... 5 1.5 Voltage

More information

BCM2 Series Branch Circuit Monitors Quick Setup Guide

BCM2 Series Branch Circuit Monitors Quick Setup Guide BCM2 Series Branch Circuit Monitors Quick Setup Guide Safety Information DANGER! HAZARD OF ELECTRIC SHOCK, EXPLOSION, OR ARC FLASH Follow safe electrical work practices. See NFPA 70E in the USA, or applicable

More information

DTS 310. DIN Rail Mounted, Indoor Rated Revenue Grade Electrical Sub-meter. Page 1. Measurlogic MQ A

DTS 310. DIN Rail Mounted, Indoor Rated Revenue Grade Electrical Sub-meter. Page 1. Measurlogic MQ A DTS 310 DIN Rail Mounted, Indoor Rated Revenue Grade Electrical Sub-meter Page 1 1 PRODUCT OVERVIEW... 3 1.1 SUPPLIED ITEMS... 3 1.2 DOCUMENT CONVENTIONS... 4 1.3 PRODUCT SPECIFICATION... 4 1.3.1 Current

More information

Surge Protective Devices. STV100K Series. Instruction Manual

Surge Protective Devices. STV100K Series. Instruction Manual Surge Protective Devices STV100K Series Instruction Manual Contents 1.0 Installation... 4 1.1 Environment... 4 1.2 Mounting... 4 1.3 Proper Connection of SPDs... 4 1.4 Mechanical Dimensions... 5 1.5 Voltage

More information

Models beginning with 2M, 2L or 2X. Product Description. Technical Specifications. Installation Instructions. Series 2000 Multiple Meter Units (MMUs)

Models beginning with 2M, 2L or 2X. Product Description. Technical Specifications. Installation Instructions. Series 2000 Multiple Meter Units (MMUs) Models beginning with 2M, 2L or 2X Series 2000 Multiple Meter Units (MMUs) Product Description Technical Specifications Installation Instructions February 28 th, 2013 List of Figures...2 List of Tables...2

More information

Sync Control. Installation/Operation Manual Rev. E

Sync Control. Installation/Operation Manual Rev. E 9 315 Sync Control Installation/Operation Manual 164201140 Rev. E ------------------------------------------------------------------------ ------------------------------------------------------------------------

More information

PGR-4700 MANUAL SENSITIVE GROUND-FAULT RELAY

PGR-4700 MANUAL SENSITIVE GROUND-FAULT RELAY Tel: +1-800-832-3873 E-mail: techline@littelfuse.com www.littelfuse.com/pgr-4700 PGR-4700 MANUAL SENSITIVE GROUND-FAULT RELAY REVISION 2-B-032218 Copyright 2018 by Littelfuse, Inc. All rights reserved.

More information

MAXIMA + Series ROTARY LEVEL CONTROL

MAXIMA + Series ROTARY LEVEL CONTROL Price $5.00 MAXIMA + Series ROTARY LEVEL CONTROL OPERATING INSTRUCTIONS PLEASE READ CAREFULLY Division of Garner Industries 7201 North 98th Street Lincoln, NE 68507-9741 (402) 434-9102 925-0268 Rev. A

More information

INSTALLATION DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER. Before installation:

INSTALLATION DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER. Before installation: DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER The DKM-409 is a precision instrument designed for displaying various AC parameters in 3-phase distribution panels. Thanks to its isolated

More information

Avanti HP Series Centrifuges Preinstallation

Avanti HP Series Centrifuges Preinstallation J3HP-TB-001AC May 2009 Avanti HP Series Centrifuges Preinstallation INTRODUCTION Purchase of an Avanti HP series high performance centrifuge includes installation and Basic Instrument Training for two

More information

to 48V DC Power Supply

to 48V DC Power Supply Telecommunications Group Section 855-080-201 Equipment Issue 1 Second Printing, July 2007 8550-80 24 to 48V DC Power Supply Compliant with UL Standard 60950, Second Edition* CONTENTS PAGE Part 1. GENERAL.............................................................................

More information

AMP Rack User Manual. Publication Rev G

AMP Rack User Manual. Publication Rev G Publication Introduction This manual contains information for the AMP Rack from Forney Corporation, 16479 Dallas Parkway, Suite 600, Addison, Texas. www.forneycorp.com All personnel should become thoroughly

More information

Girard Awnings R ACMC Motor Controller (Last revised on May 10, 2008) A Visionary Awning Control by

Girard Awnings R ACMC Motor Controller (Last revised on May 10, 2008) A Visionary Awning Control by Girard Awnings R ACMC Motor Controller (Last revised on May 10, 2008) R A Visionary Awning Control by Girard AC Motor Controller Installation Guide ACMC Revision 2.08 May 10, July 2008 ACMC Installation

More information

Specialized Lighting Solutions, Inc. SLS DALI CONTROLLERC

Specialized Lighting Solutions, Inc. SLS DALI CONTROLLERC DCP-A Panel Network DALI Control Panel for DALI Ballast, DALI Relays and DALI Dimmers 24 20 6 The DCP-A electronic network control panel with Clipsal DCDALCIP250-2 serial interfaces supports applications

More information

hp uninterruptible power system r12000 xr models installation instructions

hp uninterruptible power system r12000 xr models installation instructions hp uninterruptible power system r000 xr models installation instructions Overview These instructions show how to install an uninterruptible power system (UPS). For detailed information about the UPS, refer

More information

Installation and Operation Manual

Installation and Operation Manual Installation and Operation Manual APC 2X Series Hardwire Products PMP2X-A PMF2X-A PMG2X-A PMH2X-A APC hardwired products are designed to provide specification-grade performance at the service entrance,

More information

SAVE THESE INSTRUCTIONS

SAVE THESE INSTRUCTIONS OUTDOOR HARDWIRE INSTALLATION INSTRUCTIONS Please read and save these instructions. Read carefully before using product. Protect yourself and others by observing all safety information, warnings and cautions.

More information

Dual-Feed 80A Auto Low Voltage Disconnect Panel - 80LVD02

Dual-Feed 80A Auto Low Voltage Disconnect Panel - 80LVD02 Dual-Feed 80A Auto Low Voltage Disconnect Panel - 80LVD02 User Manual 1.1 Overview Telect s 80A, -48 Vdc Low-Voltage Disconnect (LVD) panel is a rack-mounted component with coil-operated contactors. These

More information

Electrical Demand Specification (Reference SOP: )

Electrical Demand Specification (Reference SOP: ) Project: Equipment Description: Location: Equipment No.: Project No: Protocol No.: Content Index 1. GENERAL...3 Design Standards...3 1.1. Standards...3 2. DESIGN...3 2.1. Safety...3 2.2. Circuit protection...3

More information

PGR-6101 MANUAL GROUND-FAULT & INSULATION MONITOR

PGR-6101 MANUAL GROUND-FAULT & INSULATION MONITOR Tel: +1-800-832-3873 E-mail: techline@littelfuse.com www.littelfuse.com PGR-6101 MANUAL GROUND-FAULT & INSULATION MONITOR Revision 0-C-041918 Copyright 2018 by Littelfuse, Inc. All rights reserved. Document

More information

Model# CKM18 Model# CKM36 Model# CKM48 WARNING

Model# CKM18 Model# CKM36 Model# CKM48 WARNING Installation Instructions Model# CKM18 Model# CKM36 Model# CKM48 WARNING HAZARDOUS VOLTAGES. DISCONNECT FROM SUPPLY BEFORE REMOVING COVERS. NO USER SERVICEABLE PARTS INSIDE. SERVICE BY QUALIFIED PERSONNEL

More information

1. Section Lightning Protection for lightning protection system. D. The installation shall be in conformance with NEC Article 285.

1. Section Lightning Protection for lightning protection system. D. The installation shall be in conformance with NEC Article 285. SECTION 26 04 75 - SURGE PROTECTION DEVICES PART 1 - GENERAL 1.1 SUMMARY A. Applicable provisions of the General Conditions, Supplementary General Conditions and Special Conditions shall govern work performed

More information

MODELS DPDT DPDT DPDT-60MS DPDT-60MS

MODELS DPDT DPDT DPDT-60MS DPDT-60MS INSTALLATION INSTRUCTIONS Revision B1 Rapid City, SD, USA, 03/2010 MODELS 201-100-DPDT 201-200-DPDT 201-100-DPDT-60MS 201-200-DPDT-60MS II_201-XXX-DPDT_B1 II-201-DPDT-B BE SURE POWER IS DISCONNECTED PRIOR

More information

SPECIAL INSTRUCTIONS FOR CAPACITORS COMPACT GENERATORS

SPECIAL INSTRUCTIONS FOR CAPACITORS COMPACT GENERATORS SPECIAL INSTRUCTIONS FOR CAPACITORS COMPACT GENERATORS (WITH CAPACITOR CHARGER BOARD A3517-02) The process depends on Generator and System configuration. This document applies to installation of Capacitors

More information

Multi Unit Enclosed Meter

Multi Unit Enclosed Meter Contents Description Page Introduction Preliminary comments and safety precautions............................... 2 Warranty and liability information................ 2 Safety precautions..............................

More information

SECTION 16496E ATS MOLDED CASE WITH ATC 900 CONTROLLER

SECTION 16496E ATS MOLDED CASE WITH ATC 900 CONTROLLER ATS MOLDED CASE WITH ATC 900 CONTROLLER PART 1 GENERAL 01 02 03 04 SCOPE Furnish and install the low voltage automatic transfer switches having the ratings, features/accessories and enclosures as specified

More information

Installation Instructions

Installation Instructions 50ES, 50EZ, 50GL, 50GS, 50GX, 50JS, 50JX, 50JZ, 50SD, 50SZ, 50VL, 50VT, 50XP, 50XZ 601A, 602A, 602B, 604A, 604B, 604D, 607C, 701A, 702A, 702B, 704A, 704B, 704D, 707C PA1P, PA2P, PA3P, PH1P, PH2P, PH3P

More information

BSPD Surge Protective Devices

BSPD Surge Protective Devices BUSSMANN Supersedes December 2016 SERIES BSPD Surge Protective Devices High capacity Type 1 and Type 2 SPDs for overvoltage protection on Contents Description Page Introduction 2 Applications 2 Features

More information

QUICK SETUP GUIDE. BCM2 Series Branch Circuit Monitors. Safety Information. Equipment Maintenance and Service. Raritan DANGER!

QUICK SETUP GUIDE. BCM2 Series Branch Circuit Monitors. Safety Information. Equipment Maintenance and Service. Raritan DANGER! QUICK SETUP GUIDE BCM2 Series Branch Circuit Monitors Safety Information DANGER! HAZARD OF ELECTRIC SHOCK, EXPLOSION, OR ARC FLASH Follow safe electrical work practices. See NFPA 70E in the USA, or applicable

More information

Page Housing Plastic Metal Mounting DIN Rail or surface mount Panel or bracket mount Wattage Range 10W to 240W 7.5W to 240W 10W to 300W

Page Housing Plastic Metal Mounting DIN Rail or surface mount Panel or bracket mount Wattage Range 10W to 240W 7.5W to 240W 10W to 300W Selection Guide Selection Guide PS5R Slim Line PS5R PS3L Operator Interfaces Communication & Networking PLCs NEW Models Appearance Automation Software Page 107 113 118 Housing Plastic Metal Mounting DIN

More information

ICS Entrance Management Sign Installation Guide. Version 1.0

ICS Entrance Management Sign Installation Guide. Version 1.0 ICS Entrance Management Sign Installation Guide Version 1.0 Thank you for purchasing your new ICS Entrance Management System (EMS) from Innovative Control Systems, Inc. Installation Overview This document

More information

Hubbell Surge Protective DeviceS

Hubbell Surge Protective DeviceS Hubbell Surge Protective DeviceS Hubbell Modular Panel Products Surge Protective Devices (SPD) The Hubbell Modular Panel Product Surge Protective Devices (SPD) are designed to provide protection at the

More information

Voltage Sag Simulator

Voltage Sag Simulator For use with all DPI s & VDC s 120/208/240V 50/60Hz LEADERS IN VOLTAGE-DIP PROOFING Contents Introduction to the DPI... 3 Theory of DPI operation... 3 Description -... 4 Operation... 5 Sag Simulation...

More information

HydroGuard 4 to 20 ma Output Module. For use with HydroGuard Water Quality Monitor and Controller

HydroGuard 4 to 20 ma Output Module. For use with HydroGuard Water Quality Monitor and Controller HydroGuard 4 to 20 ma Output Module For use with HydroGuard Water Quality Monitor and Controller Technician's Manual Installation, Operation, and Maintenance Guide Chapter 1: Preface... 3 1.1 Intended

More information

SECTION SURGE PROTECTION DEVICES. A. Section includes Surge Protection Devices for low-voltage power, control, and communication equipment.

SECTION SURGE PROTECTION DEVICES. A. Section includes Surge Protection Devices for low-voltage power, control, and communication equipment. SECTION 26 43 00 SURGE PROTECTION DEVICES PART 1 - GENERAL 1.1 SUMMARY A. Section includes Surge Protection Devices for low-voltage power, control, and communication equipment. 1.2 DEFINITIONS A. ATS:

More information

SAVE THESE INSTRUCTIONS

SAVE THESE INSTRUCTIONS READ AND FOLLOW ALL SAFETY INSTRUCTIS! SAVE THESE INSTRUCTIS AND DELIVER TO OWNER AFTER INSTALLATI IMPORTANT SAFETY INSTRUCTIS WARNING To reduce the risk of death, injury or property damage from fire,

More information

BT300 HVAC Drives Electronic Bypass (E-Bypass) Options

BT300 HVAC Drives Electronic Bypass (E-Bypass) Options BT300 HVAC Drives Electronic Bypass (E-Bypass) Options Description The BT300 Electronic Bypass (E-Bypass) Options are companion packages for the family of BT300 Variable Frequency Drives (VFDs). For information

More information

Installation Manual for D244X Series 24-volt Power Supplies

Installation Manual for D244X Series 24-volt Power Supplies Installation Manual for D244X Series 24-volt Power Supplies D2441-B D2441-C D2443-B D2443-C D2445-B D2445-C D2400 Series 1A 24 volts 1A 24 volts 3A 24 volts 3A 24 volts 5A 24 volts 5A 24 volts Dycon Power

More information

PG Softstarter Installation Manual for 5A, 9A and 16A Controllers

PG Softstarter Installation Manual for 5A, 9A and 16A Controllers PG Softstarter Installation Manual for 5A, 9A and 16A Controllers Important User Information The examples and diagrams in this manual are included solely for illustrative purposes. Because of the many

More information

SAVE THESE INSTRUCTIONS

SAVE THESE INSTRUCTIONS READ AND FOLLOW ALL SAFETY INSTRUCTIS! SAVE THESE INSTRUCTIS AND DELIVER TO OWNER AFTER INSTALLATI IMPORTANT SAFETY INSTRUCTIS WARNING To reduce the risk of death, injury or property damage from fire,

More information

Digital Lighting Systems, Inc. PD216. Two Channel Dimmer and Switch Packs PROTOCOL USER'S MANUAL. PD216-UM Rev. E - 02/03

Digital Lighting Systems, Inc. PD216. Two Channel Dimmer and Switch Packs PROTOCOL USER'S MANUAL. PD216-UM Rev. E - 02/03 Digital Lighting Systems, Inc. PD26 Two Channel Dimmer and Switch Packs PROTOCOL PD26 S2 S USER'S MANUAL PD26-UM Rev. E - 02/03 Digital Lighting Systems PD26 User's Manual - Page GENERAL DESCRIPTION The

More information

Voltage Dip-Proofing Inverter

Voltage Dip-Proofing Inverter Voltage Dip-Proofing Inverter DPI54S / 54L Series Models 120V & 208/230 50/60Hz Contents: Introduction...3 Theory of operation...3 Specifications...5 Up-time considerations...6 Installation Guide...7 Test

More information

MAINTENANCE MANUAL. EDACS REDUNDANT POWER SUPPLY SYSTEM 350A1441P1 and P2 POWER MODULE CHASSIS 350A1441P3, P4, AND P5 POWER MODULES TABLE OF CONTENTS

MAINTENANCE MANUAL. EDACS REDUNDANT POWER SUPPLY SYSTEM 350A1441P1 and P2 POWER MODULE CHASSIS 350A1441P3, P4, AND P5 POWER MODULES TABLE OF CONTENTS MAINTENANCE MANUAL EDACS REDUNDANT POWER SUPPLY SYSTEM 350A1441P1 and P2 POWER MODULE CHASSIS 350A1441P3, P4, AND P5 POWER MODULES TABLE OF CONTENTS SPECIFICATIONS*... 2 INTRODUCTION... 3 DESCRIPTION...

More information

ICS Regent. Relay Output Modules. Low Power and High Power (T3446L and T3446H) PD-6017

ICS Regent. Relay Output Modules. Low Power and High Power (T3446L and T3446H) PD-6017 ICS Regent PD-6017 Relay Output Modules Low Power and High Power (T3446L and T3446H) Issue 1, March, 06 Relay output modules provide control of eight user output loads. Two types of relay output modules

More information

UNT1100 Series. Binary Output Jumpers AO2 AO1 AO3 AO4 AOCM AOCM AOCM AOCM AO1. AI Switches Job Information N2 Address. Ref N2+ N2- ADDR 0 = ALL OPEN

UNT1100 Series. Binary Output Jumpers AO2 AO1 AO3 AO4 AOCM AOCM AOCM AOCM AO1. AI Switches Job Information N2 Address. Ref N2+ N2- ADDR 0 = ALL OPEN 1 2 8 4 AI6 R R Installation Bulletin UNT1100 Issue Date 0309 UNT1100 Series Introduction The Unitary (UNT) controller (UNT1100 Series) is a digital controller with applications for air handling units,

More information

Installation. MGE Galaxy 3500 and Smart-UPS VT. Maintenance Bypass, Distribution and Transformer Cabinet

Installation. MGE Galaxy 3500 and Smart-UPS VT. Maintenance Bypass, Distribution and Transformer Cabinet Installation MGE Galaxy 3500 and Smart-UPS VT Maintenance Bypass, Distribution and Transformer Cabinet American Power Conversion Legal Disclaimer The information presented in this manual is not warranted

More information