1 Dear Customer, A problem has been found with the Philips IntelliVue MX40 Instructions for Use (IFU) for software revisions B.05, B.06 and B.06.5X. Your IntelliVue MX40 remains safe to use. These IFUs are missing warning statements related to monitoring paced patients and the interpretation of QT/QTc measurements that were present in earlier revisions of the IFU. A Warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to observe a warning may result in death or serious injury to the user or patient. This Field Safety Notice is intended to inform you about: what the problem is and under what circumstances it can occur the actions that should be taken by the customer/user in order to prevent risks for patients the actions planned by Philips to correct the problem. This document contains important information for the continued safe and proper use of your equipment Please review the following information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication. Please refer to the following pages, which provide information on the missing warnings and instructions for actions to be taken. Follow the Action to be taken by Customer/User section of this notice. This issue has been reported to the appropriate regulatory agencies. I sincerely regret the inconvenience that this may cause you. Your satisfaction with Philips products and with our response to this issue is very important to us. Please contact your local Philips representative under with questions or concerns about this correction. Sincerely, Kristen Phillips Head of Quality & Regulatory Affairs, Andover
2 AFFECTED PRODUCTS The Philips IntelliVue MX40 revisions B.05, B.06 and B.06.5X. Model Product Number IntelliVue MX PROBLEM DESCRIPTION HAZARD INVOLVED HOW TO IDENTIFY AFFECTED PRODUCTS ACTIONS PLANNED BY PHILIPS Five warning statements are missing from the IntelliVue MX40 IFU for software revisions B.05, B.06 and B.06.5X. If users are unaware of the hazards or limitations described in the missing warning statements, they may not properly assess or provide appropriate treatment to a patient being monitored using an IntelliVue MX40. IntelliVue MX40 Instructions for Use for software revisions B.05, B.06 and B.06.5X. Philips is voluntarily initiating a correction consisting of: Distribution of this Field Safety Notice (). Release of an IFU Errata that provides the missing warnings A representative will contact customers with affected IFU revisions to provide a copy of the errata. This FSN includes the Errata providing the missing warning statements. ACTION TO BE TAKEN BY CUSTOMER / USER The enclosed MX40 IFU Errata Sheet must be attached to the first page of Chapter 6 of the Instructions for Use for ready reference. Complete and return the attached Customer Reply Form.
3 Contact Name Telephone Number Address Facility Name Street Address City, State, Zip Customer Reply for IntelliVue MX40 IFU Please complete and fax to Please fax or this completed form to the number or address provided above. CUSTOMER ACKNOWLEDGEMENT: The MX40 IFU Errata must be attached to the first page of Chapter 6 to ensure that it is not misplaced and is stored with the Instructions for Use for ready reference. CUSTOMER NAME (please print) TITLE CUSTOMER SIGNATURE DATE Please fax or the completed reply form to customercare.ch@philips.com. If you experience difficulty carrying out the instructions contained in this communication, contact your local Philips representative:
4 IntelliVue MX40 Release B.05 or Later Instructions for Use Errata This Errata is a supplement to the IntelliVue MX40 Instructions for Use Release B.05 and later. It contains important safety information. Attach this Errata to the first page of Chapter 6 to ensure that it is not misplaced and is stored with the Instructions for Use for ready reference. Part Number Printed in the U.S.A. First Edition * *
5 6 ECG and Arrhythmia Monitoring English 6 ECG and Arrhythmia Monitoring ECG Safety Information Warning The device provides QT and QTc interval change information; the clinical significance of the QT and QTc interval change information should be determined by a clinician. For more information, see the QT Interval Monitoring Application Note, p/n For Paced Patients Warning The output power of the MX40 and other sources of radio frequency energy, when used in the proximity of a pacemaker, can be sufficient to interfere with pacemaker performance. Due to the shielding effects of the body, internal pacemakers are somewhat less vulnerable than external pacemakers. However, caution should be exercised when monitoring any paced patient. In order to minimize the possibility of interference, position electrodes, electrode wires, and the MX40 as far away from the pacemaker as possible. Consult the pacemaker manufacturer for information on the RF susceptibility of their products and the use of their products with the MX40. See the Patient Information Center Instructions for Use for additional information on monitoring paced patients. When an external pacemaker is being used on a patient, arrhythmia monitoring is severely compromised due to the high energy level in the pacer pulse. This may result in the arrhythmia algorithm's failure to detect pacemaker non-capture or asystole. Pacemakers that create fusion beats (pace pulse on top of the QRS complex) cannot be detected by the monitor's QRS detector. For paced patients who exhibit only intrinsic rhythm, the monitor can erroneously count pace pulses as QRS complexes when the algorithm first encounters them, resulting in missed detection of cardiac arrest. The risk of missing cardiac arrest can be reduced by monitoring these patients with the low heart rate limit at or slightly above the basic/ demand pacemaker rate. A low heart rate alarm notifies you when the patient begins pacing. Proper detection and classification of the paced rhythm can then be determined. Note During defibrillation, monitoring may be temporarily interrupted or distorted. It may take several seconds for the ECG trace to reappear on the screen. After defibrillation, the device will continue to monitor as before; the device settings will not be affected. 6-1
URGENT Field Safety Notice Philips HeartStart MRx Monitor/Defibrillator
Philips Healthcare - 1/5 - FSN86100160A Dear Customer, Philips has identified issues that could impact the safety and/or performance of certain MRx monitor/defibrillators. These issues are further detailed
More informationThis document contains important information for the continued safe and proper use of your equipment
Philips Healthcare - 1/6 - FSN86100179A Dear Customer, Philips has identified the possibility of abnormal device behaviors on the HeartStart MRx Monitor/Defibrillator. Under exceptionally rare circumstances,
More informationURGENT FIELD SAFETY NOTICE
URGENT FIELD SAFETY NOTICE Affected Devices: Rythmic Evolution product family Type of Action: Field safety corrective action Ref.: FSN2018-01 Date: 12 March 2018 Attention: Distributors, Biomedical Professionals,
More informationThis document contains important information for the continued safe and proper use of your equipment
Philips Healthcare - 1/6 - FSN86100179A Dear Customer, Philips has identified the possibility of abnormal device behaviors on the HeartStart MRx Monitor/Defibrillator. Under exceptionally rare circumstances,
More informationAlways in touch. IntelliVue Telemetry System with Smart-hopping technology, surveillance of ambulatory cardiac patients
Always in touch with Smart-hopping technology, surveillance of ambulatory cardiac patients Real clinical network strength includes flexibility When you have mobile cardiac patients who need constant monitoring,
More informationURGENT FIELD SAFETY NOTICE Reserve battery and new instructions for users of the PAD 300/PAD 300P (Public Access Defibrillator)
14 th September 2012 URGENT FIELD SAFETY NOTICE Reserve battery and new instructions for users of the PAD 300/PAD 300P (Public Access Defibrillator) Dear Sir or Madam, The purpose of this letter is to
More informationPhilips HeartStart XL Defibrillator Manual
Philips HeartStart XL Defibrillator Manual Get an original copy of the Philips HeartStart XL Defibrillator Operators Guide for manufacturer information about service, available accessories and how to use
More informationCARELINK NETWORK UPDATES SUMMARY OF ENHANCEMENTS May 2017
CARELINK NETWORK UPDATES SUMMARY OF ENHANCEMENTS May 2017 Quick Look Update for Reveal LINQ ICM Patient Monitoring CareLink Network Experience With feedback from over 200 clinicians worldwide, Medtronic
More informationImportant Medical Device Advisory
Important Medical Device Advisory Battery Performance Alert and Cybersecurity Firmware Updates for Certain ICD & CRT-D Devices April 16, 2018 Dear Doctor, As part of a planned series of system updates
More informationProduct codes with prefix (all variants): 8001, 8002, 8003, 8004, and ATTENTION: Clinical Personnel, Risk Managers, Biomedical Personnel
CareFusion Switzerland 317 Sàrl A-One Business Centre Zone d activités, Vers-la-Pièce No 10 Rolle, CH-1180 Switzerland carefusion.com Name Address URGENT FIELD SAFETY NOTICE Product Name: Alaris / Asena
More informationEC-12RT Resting ECG Machine. Technical Specification
2017 ECG input Input impedance Input bias current Input dynamic range 10 MOhm 0,1μА from 0,03mV up to 10mV (peak-to-peak amplitude) Nonlinearity ±2% Electrode offset tolerance Common-mode rejection Internal
More informationImportant Medical Device Advisory NM Implantable Pulse Generator Inaccurate Elective Replacement Indicator
Important Medical Device Advisory NM Implantable Pulse Generator Inaccurate Elective Replacement Indicator 12 September 2017 Dear Physician, In an effort to keep you continually informed, and to ensure
More informationUse Of Mobile Communication Devices Within Healthcare Premises Policy
Use Of Mobile Communication Devices Within Healthcare Premises Policy Co-ordinator: Director of Facilities Reviewer: Working Group chaired by Director of Facilities Approver: GAPF Signature Signature Signature
More information!"# $ # # $ $ % $ &% $ '"# $ ()&*&)+(( )+(( )
!"# # # % &% '"# ) !#, ' "# " "# -. / # 0 0 0 0 0 "0 "# " # 1 #! " " 0 0 0 0 0 0 2# 0 # # 3 ' 4 56 7-56 87 9# 5 6 7 6 & 0 " : 9 ; 4 " #! 0 - '% # % "# " "# " < 4 "! % " % 4 % % 9# 4 56 87 = 4 > 0 " %!#
More informationImportant Medical Device Advisory NM Implantable Pulse Generator Inaccurate Elective Replacement Indicator
Important Medical Device Advisory NM Implantable Pulse Generator Inaccurate Elective Replacement Indicator 12 September 2017 Dear Physician, In an effort to keep you continually informed, and to ensure
More information4P S. Wireless LAN Connecting Adapter <BRP069A42> Installation Manual
4P359542-2S Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com For details
More informationURGENT MEDICAL DEVICE FIELD SAFETY NOTICE- REMOVAL
June 19, 2017 To: Subject: Surgeons/ Hospitals/ Clinics URGENT MEDICAL DEVICE FIELD SAFETY NOTICE- REMOVAL Affected Product: Trauma, Guide Wires 70cm Image 1. Guide wire with protector Image 2. 3mm*70cms
More informationIntelliVue Patient Monitor MX450, MX500, MX550. Remote Application Setup. Release K.2 with Software Revision K.2. Patient Monitoring
Installation and Configuration Guide IntelliVue Patient Monitor MX450, MX500, MX550 Remote Application Setup Release K.2 with Software Revision K.2 Patient Monitoring 1Table of Contents 1 Introduction
More informationURGENT: MEDICAL DEVICE RECALL
Wednesday, September 20, 2017 [Contact Name] Andrew Lindsay QM Manager [Business Name] [Customer Address] [City, State, Zip Code] Carl Zeiss Ltd. 509 Coldhams Lane Cambridge Cambridgeshire CB1 3JS U.K.
More informationFIELD CORRECTIVE ACTION NOTICE
FIELD CORRECTIVE ACTION NOTICE Leica M525 F20 Microscope System FCA Identifier: CAPA-HER-MD-18-001 Enhancement of instructions for use and additional markings to prevent inadvertent tipping Customer Name
More informationcubestress your profession our mission
The CARDIOLINE wireless stress testing system combines combina the high performance of the software and of the acquisition unit clickecgbt. is the complete solution to manage ECG Stress tests. combines
More informationUrgent Field Safety Notice
Urgent Field Safety Notice Commercial name/model: WATO EX-65, WATO EX-55, A5 and A7 anesthesia machines FSCA-identifier: CP17-MZ0036 Type of action: Safety Notice and Device modification - June, 2017 Attention:
More informationSpO 2 Upgrade Instructions
SpO 2 Upgrade Instructions M3535A/M3536A HeartStart MRx Notice About This Edition Edition 3 Printed in the USA Publication number M3535-90905 The information in this document applies to the HeartStart
More informationTIP CARD MODEL TEMPORARY EXTERNAL PACEMAKER. Model Single Chamber Temporary Pacemaker
TIP CARD MODEL 53401 TEMPORARY EXTERNAL PACEMAKER Model 53401 Single Chamber Temporary Pacemaker PRE-USE What to do... What NOT to do... Battery Replace the battery for each new patient. Do NOT reuse battery.
More informationCardiostart. Biphasic Defibrillator Monitor
Cardiostart Biphasic Defibrillator Monitor Standard Accessories Carrying Case Rechargeable lithium ion battery Set of accessories Other accessories upon request Internal Pads Adult / Pediatric / Neonatal
More informationOperating Instructions. Panorama. Web Viewer
Operating Instructions Panorama Web Viewer Mindray is a U.S. registered trademark of Mindray DS USA, Inc. Panorama is a U.S. registered trademark of Mindray DS USA, Inc. Microsoft is a U.S. registered
More informationIntelliVue Patient Monitor. MP20/30/40/50/60/70/90 For monitor release B.1 with software revision B.10.xx. IntelliVue MP20/30/40/50/60/70/90
IntelliVue MP20/30/40/50/60/70/90 CONFIGURATION GUIDE IntelliVue Patient Monitor MP20/30/40/50/60/70/90 For monitor release B.1 with software revision B.10.xx Patient Monitoring Philips Medizin Systeme
More informationUrgent Medical Device Correction LIFEPAK 1000 Defibrillator
Urgent Medical Correction LIFEPAK 1000 Defibrillator Urgent Please bring this letter to the immediate attention of the person(s) responsible for maintaining/monitoring your LIFEPAK 1000 defibrillators.
More informationHTS820TC TRENSCenter Central Station Cardiac Monitor REHAB Operator/Technical Manual
Table of Contents Subject... Page Subject... Page Preface...2 Safety Summary... 3 Section 1 Operation Attaching Transmitter to Patient...4 Admitting a Group... 6 Admitting Individual Patients...7 Monitoring
More informationPCI Expansion Slot Kit for 6181P (1500P) Series D Integrated Display Computer
Installation Instructions PCI Expansion Slot Kit for 6181P (1500P) Series D Integrated Display Computer Catalog Number 6189V-2PCI15R Topic Page About This Publication 1 Important User Information 2 Safety
More information4P R. Wireless LAN Connecting Adapter <BRP069A81> Installation Manual
4P359542-4R Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com For details
More informationThe LifeVest Network Patient Data Management System Quick Start Guide. Information for Healthcare Professionals
The LifeVest Network Patient Data Management System Quick Start Guide Information for Healthcare Professionals Monitor and prioritize your patient s needs anywhere, anytime. The LifeVest Network is ZOLL
More information12. Mobile Devices and the Internet of Things. Blase Ur, May 3 rd, 2017 CMSC / 33210
12. Mobile Devices and the Internet of Things Blase Ur, May 3 rd, 2017 CMSC 23210 / 33210 1 Today s class Security and privacy for: mobile devices the IoT safety-critical devices Discuss midterm 2 Mobile
More informationIntelliVue Information Center
IntelliVue Information Center Demo Application User Guide Part Number 4535 642 63741 Printed in the U.S.A. June 2011 First Edition *453564263741* Notice Notice Proprietary Information Published by Philips
More information4P S. Wireless LAN Connecting Adapter <BRP069A41> Installation Manual
4P359542-1S Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com For details
More information4P Wireless LAN Connecting Adapter <BRP069B41> <BRP069B42> <BRP069B43> <BRP069B44> <BRP069B45> Installation Manual
4P481234-1 Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com
More informationDefibtech DDU-100 Semi-Automatic External Defibrillator
Defibtech DDU-100 Semi-Automatic External Defibrillator User Manual AHA /ERC 2010 U S E O N LY I F PA T I E N T I S : U N C O N S C I O U S U N R E S P O N S I V E N O T B R E A T H I N G 1 PUSH "ON" BUTTON
More informationOperating Instructions Vision Holter Analysis System Software Version 3.5
Operating Instructions Vision Holter Analysis System Software Version 3.5 Part No. 70-01234-01 Rev. A Copyright 2012 Cardiac Science Corporation All rights reserved. Cardiac Science Corporation N7 W22025
More informationMini Pulse Electronic Stimulator Model: PM-120
Mini Pulse Electronic Stimulator Model: PM-120 Operating Manual IMPORTANT: Please read all instructions before using this product. Retain this manual for future reference. www.santamedical.com IMPORTANT
More informationSave Face Radiofrequency Fact Sheet
Save Face Radiofrequency Fact Sheet Patient information Radiofrequency What is Radiofrequency? The field of body contouring and tissue tightening has grown very rapidly over the past several years, with
More informationPRODUCT SAFETY NOTICE WARNING OF POTENTIAL UNSAFE CONDITION
DATE: November 7, 2016 TO: FROM: Schneider Electric Distributors, and End-users using the Schneider Electric E-Stop 9001SKR16 David Collins US Product Management SUBJECT: Schneider E-Stop 30mm 9001SKR16
More informationIntelliVue Patient Monitors
Hardware Upgrade Installation Guide IntelliVue Patient Monitors MX400/MX450 Patient Monitoring 1Table of Contents 1 Introduction 5 What is in this Installation Note 6 2 Checking your Tools and Kit Hardware
More informationResearch Report. Radio Frequency Identification Devices (RFIDs) Effect of RFIDs on lifesaving equipment
Research Report Radio Frequency Identification Devices (RFIDs) Effect of RFIDs on lifesaving equipment Background It has been suggested in the Journal of the Medical Association that lifesaving equipment
More informationMONITOR/DEFIBRILLATOR CONTROLS, CONNECTIONS AND INDICATORS
MONITOR/DEFIBRILLATOR CONTROLS, CONNECTIONS AND INDICATORS 1-10 Adult Dose CONTROLS, CONNECTIONS AND INDICATORS External Power SYNC Mark Event Lead Select Alarm Pause Event Summary Soft Keys Soft Keys
More informationR Series Configuration Manual
R Series Configuration Manual The issue date for the R Series Configuration Manual (REF 9650-0902-05 Rev. C) is June, 2012. If more than 3 years have elapsed since the issue date, contact ZOLL Medical
More informationPatient Manual. Confirm Rx Insertable Cardiac Monitor System
Confirm Rx Insertable Cardiac Monitor System Confirm Rx Insertable Cardiac Monitor Model DM3500 mymerlin Mobile Application Models APP1000, APP1001 Patient Manual CAUTION: Federal (USA) law restricts this
More information4P M. Wireless LAN Connecting Adapter <BRP069A45> Installation Manual
4P359542-3M Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com For details
More informationA Guide to Open-Access Databases and Open-Source Software on PhysioNet
A Guide to Open-Access Databases and Open-Source Software on PhysioNet George B. Moody Harvard-MIT Division of Health Sciences and Technology Cambridge, Massachusetts, USA Outline Background Open-Access
More informationHigh-resolution color display provides real-time preview of 12-lead ECG to ensure high quality data acquisiton at investigator site.
ELI 280 offers a choice of either the innovative WAM wireless acquisition module or the AM12 acquisition module. Both include replaceable lead wires, lead fail indicator, and remote control with buttons
More informationPolicies, Procedures, Guidelines and Protocols. John Snell - Head of Workforce Planning, Systems and Contributors
Policies, Procedures, Guidelines and Protocols Document Details Title Staff Mobile Phone Policy Trust Ref No 2036-39774 Local Ref (optional) N/A Main points the document Procurement, allocation and use
More informationTransTalk 9000 Digital Wireless System MDW 9040 Wireless Pocket Phone Quick Reference
TransTalk 9000 Digital Wireless System MDW 9040 Wireless Pocket Phone Quick Reference 503-801-191 Comcode 700060437 Issue 2, Feb. 2001 Mute Volume Control Display Softkeys/Column Buttons Feature/ Program
More informationMedical Device Usability
Medical Device Usability David Adams Global Head, Active Medical Devices Add logo on slide 4 here Topics What is usability? Why usability is so important The regulatory requirements EN 62366 Usability
More informationLATITUDE PROGRAMMING SYSTEM
OPERATOR S MANUAL LATITUDE PROGRAMMING SYSTEM REF 3300 Table of Contents INFORMATION FOR USE...1 Trademark Statement...1 Description and Use...1 Intended Use...1 Intended Audience...1 Required Expertise
More informationWireless LAN Connecting Adapter [BRP072B42] Installation Manual
Wireless LAN Connecting Adapter [BRP072B42] Installation Manual Wireless LAN Connecting Adapter [BRP072B42] Installation Manual Website: http://www.daikinthai.com/wlan3/modbus For details on specifications,
More informationM Series AHA/ERC Guidelines 2005 Software Configuration Overview
M Series AHA/ERC Guidelines 2005 Software Configuration Overview This page provides a summary of configuration options that have been added to support the American Heart Association and European Resuscitation
More informationMDLINK MANUAL PART NUMBER A Cardiac Science Corporation
MDLINK MANUAL PART NUMBER 70-00080-01 A 2007 Cardiac Science Corporation Notice of Rights 2007 Cardiac Science Corporation. All rights reserved. No part of this documentation may be reproduced or transmitted
More informationHardware and Components: Philips Monitors in Critical Care
Hardware and Components: Philips Monitors in Critical Care Intellivue MX 800 (the monitor): This is the new bedside monitor. It is a touch screen. Touch will work even when double gloved. The monitor WILL
More informationVITAL TIMER RELAY ASSEMBLY, A
INSTALLATION & INSTRUCTION" VITAL TIMER RELAY ASSEMBLY, A44999-02 JULY 2008, REVISED JULY 2014 DOCUMENT NO. SIG-00-08-02 VERSION A.1 Siemens Industry, Inc., Rail Automation 9568 Archibald Ave., Suite 100,
More informationEmbedded Systems. Cristian Rotariu
Embedded Systems Cristian Rotariu Dept. of of Biomedical Sciences Grigore T Popa University of Medicine and Pharmacy of Iasi, Romania cristian.rotariu@bioinginerie.ro May 2016 Introduction An embedded
More informationBeneVision DMS Workstation Default Settings
Hospital, City, ST: Unit: Clinical/Sales: Service/CSS: SW Version: 03.02 Date: Signature/Title: System Setup General Alm Vol (1-10) High: Alm Vol + 0 1 2 Reminder Vol (1-10) Display (Password protected)
More informationLXM32. Explanation for detected error E 733F. Expert Support Machine Solution
LXM32 Explanation for detected error E 733F Expert Support Machine Solution The information provided in this documentation contains general descriptions and/or technical characteristics of the performance
More informationA1. Multilin EPM 4600 Meter Displays DIS3500/DIS5700. Quick Start Guide. GE Digital Energy
GE Digital Energy Multilin EPM 4600 Meter Displays DIS3500/DIS5700 Quick Start Guide Product version: 1.0x GE publication code: 1601-0500-A1 (GEK-119644) 1601-0500-A1 Copyright 2014 GE Multilin Inc. All
More informationSmart Wireless THUM Adapter
Reference Manual Smart Wireless THUM Adapter www.rosemount.com Reference Manual Smart Wireless THUM Adapter Smart Wireless THUM Adapter Smart Wireless THUM Adapter Hardware Revision 1 HART Device Revision
More informationExpert support and Reach back activities
GICNT- Nuclear Detection Working Group Magic Maggiore Ispra 28 March 2017 Expert support and Reach back activities Thierry PELLETIER Nuclear Security Division Safety and Security department International
More informationBD Alaris System User Manual Addendum
BD Alaris System User Manual Addendum For BD Alaris TM PC Unit, Model 8015 and BD Alaris TM Pump Module, Model 8100 January 2018 DME: 10000318809-00 Part Number: P00000272 2018 CareFusion Corporation or
More informationECG100S. General Information. Product Code
General Information Product Name General Name Product Code 80508096 Manufacturer Device Description Intended Use Cardioline Spa Registered Office and Factory: Via Linz, 19-20-21 Zona Ind. Spini di Gardolo
More informationCleanroom Microbiology for Non-Microbiologists
Microrite, Inc. brings you this unique learning experience in understanding cleanroom microbiology concepts and techniques Cleanroom Microbiology for Non-Microbiologists A course designed specifically
More informationF-1 OPTIONAL PRACTICAL TRAINING STEM EXTENSION
Global Education Office (GEO) 2120 Mesa Vista Hall MSC06 3850 1 University of New Mexico Albuquerque, NM 87131-1056 Phone (505) 277-4032 Fax (505) 277-1867 Email geo@unm.edu Web: www.geo.unm.edu/ F-1 OPTIONAL
More informationAdditional Details About the AbioCor System
Additional Details About the AbioCor System The AbioCor System is an implantable artificial heart intended to replace the patient s natural heart. It is designed for patients whose hearts are irreparably
More informationBanff Mineral Springs Hospital LOSS OF TELEPHONE EMERGENCY RESPONSE PLAN
Facilities Maintenance & Engineering Switchboard / Designate Area Supervisor / Designate Aware of Loss of Telephone services Banff Mineral Springs Hospital LOSS OF TELEPHONE EMERGENCY RESPONSE PLAN LOSS
More informationComputer Security Incident Response Plan. Date of Approval: 23-FEB-2014
Computer Security Incident Response Plan Name of Approver: Mary Ann Blair Date of Approval: 23-FEB-2014 Date of Review: 31-MAY-2016 Effective Date: 23-FEB-2014 Name of Reviewer: John Lerchey Table of Contents
More informationOracle Argus Safety. 1 Configuration. 1.1 Configuring a Reporting Destination for the emdr Profile. emdr Best Practices Document Release 8.0.
Oracle Argus Safety emdr Best Practices Document Release 8.0.1 E68590-01 November 2015 This document provides information on using emdr to switch from MedWatch Device reporting, and special considerations
More informationApplication for Lead-based Paint Activities Training Program Accreditation
ALABAMA LEAD-BASED PAINT ACTIVITIES ACCREDITATION PROGRAM Application for Lead-based Paint Activities Training Program Accreditation A completed application shall consist of Parts I - III of this application
More information3T/7T MRI FACILITY. 3T MRI System Start-Up and Shutdown
3T/7T MRI FACILITY SOP Number: 300.02 Title 3T MRI System Start-Up and Shutdown Revision Chronology Version Number Date Changes 300.01 01 March 2010 New 300.02 21 January 2013 Updated procedures Facility
More informationTraceMasterVue ECG Management System INSTALLING TRACEMASTERVUE AND CONFIGURING COMMUNICATION
TraceMasterVue ECG Management System INSTALLING TRACEMASTERVUE AND CONFIGURING COMMUNICATION Notice About This Edition Publication number 453564051211 Edition 1 Printed in USA Edition History Edition 1,
More informationWiring Instructions for Replacement of 2301A Load Sharing and Speed Controls with 2301D Load Sharing and Speed Controls
Application Note 51242 (Revision NEW) Original Instructions Wiring Instructions for Replacement of 2301A Load Sharing and Speed Controls with 2301D Load Sharing and Speed Controls General Precautions Read
More informationNHS Gloucestershire Clinical Commissioning Group. Business Continuity Strategy
NHS Gloucestershire Clinical Commissioning Group 1 Document Control Title of Document Gloucestershire CCG Author A Ewens (Emergency Planning and Business Continuity Officer) Review Date February 2017 Classification
More informationUser s Manual. LumiVision series. A Designer s X-ray film illuminator. Walnut, CA U.S.A. By LCD TV Technology
User s Manual A Designer s X-ray film illuminator By LCD TV Technology series Point A Technologies, Inc. Walnut, CA U.S.A. FCC RF INTERFERENCE STATEMENT NOTE: This equipment has been tested and found to
More informationBHG Operational Awareness Program May 8, 1998 Surveillance Guide NSS-18.1 Revision 0 Criticality Safety Page 1 of 8 CRITICALITY SAFETY
Criticality Safety Page 1 of 8 1.0 Objective The objective of this surveillance is to ensure that effective programs have been developed and implemented to protect the public and DOE's workers from unplanned
More informationInput Power Selector Switch. Product Manual (Revision NEW) Original Instructions for use with Generator Set Control Assembly
Product Manual 82017 (Revision NEW) Original Instructions 9900-471 for use with 8272-713 Generator Set Control Assembly Operation Manual General Precautions Read this entire manual and all other publications
More informationFaros Explorer Manual
Faros Explorer Manual Date of issue: May 30, 2017 Mega Electronics Ltd, Pioneerinkatu 6, FI-70800 Kuopio, Finland, http://www.megaemg.com Contents 1. Introduction... 1 1.1. Intended use... 1 1.2. Symbols...
More informationDelta 1600 User Manual
Delta 1600 User Manual Intended Use: Delta 1600 AED Tester is a precision instrument that simulates various cardiac waveforms and measures discharged energy. It assures AEDs detect the waveforms, audibly
More informationFOCUS - ON - PATIENTS. corpuls aed
FOCUS - ON - PATIENTS corpuls aed FOCUS - ON - PATIENTS corpuls aed Advanced early defibrillation for specialists and non-specialists The new corpuls aed safely guides experts and non-experts through life-saving
More informationInformation Security Guide
Broadband Kit Information Security Guide Broadband Kit ISO 27001 IS 541672 Broadband Kit Information Security Guide CONTENTS 1. Introduction... 3 1.1 Target audience...3 1.2 Overview of St. Jude Medical
More informationAnalog Input Installation Manual
Analog Input Installation Manual August 2011 Part Number: 144-23917 Copyright 2011 Magnetek 1. Preface and Safety Magnetek manufactures products used as components in a wide variety of industrial systems
More informationKBMG MULTI-SPEED BOARD
TM INSTALLATION AND OPERATION MANUAL KBMG MULTI-SPEED BOARD KB Part No. 8833 Multi-Speed Board for KBMG Series Regenerative Drive Pending! See Safety Warning on Page 1 The information contained in this
More informationApplication Note (Revision NEW) Original Instructions EGCP-3 LS Using an EGCP-3 in a Single-Phase Zig Zag Generator Application
Application Note 51247 (Revision NEW) Original Instructions EGCP-3 LS 8406-113 Using an EGCP-3 in a Single-Phase Zig Zag Generator Application General Precautions Read this entire manual and all other
More informationQuick Start Reference Guide Series 9.x and 10.x Software
Quick Start Reference Guide Series 9.x and 10.x Software NOTE: This quick reference should only be used by persons who have read the Software User Guide and received training on the use of the System.
More informationThe Learner can: 1 Follow recommended safe practices
Unit Title: Communicating Information Using ICT OCR unit number: 3 Level: Entry 3 Credit value: 3 Guided learning hours: 15 Unit reference number: H/502/1197 Learning Outcomes Assessment Criteria The Learner
More informationDATA BREACH POLICY [Enniskillen Presbyterian Church]
DATA BREACH POLICY [Enniskillen Presbyterian Church] Enniskillen Presbyterian Church is committed to complying with data protection legislation and will take appropriate technical and organisational measures
More informationModel No. CF-19/CF-30 Series
Supplementary Instructions. for Wireless WAM Modem Personal Computer Model No. CF-19/CF-30 Series This manual contains supplementary instructions regarding the use of the computer. It explains the cases
More informationInView Firmware Update
Installation Instructions InView Firmware Update Topic Page Hazardous Voltage 3 Change EPROM on 2706-P72, 2706-P74 Display 3 Change EPROM on 2706-P42, 2706-P44 Displays 5 Firmware Upgrade Kit 7 2 InView
More informationDEFRIBILLATORS PARAMEDIC CU-ER1
DEFRIBILLATORS PARAMEDIC CU-ER1 (Automated External Defibrillator) Someone goes down in cardiac arrest, resuscitation is just a hand away Patented e-cube Biphasic measure impedance and deliver efficient
More informationBanner AG4 Series Safety Laser Scanner Checkout Procedures
Shift/Daily Checkout Perform the Daily Checkout procedure at every shift change, power-up and machine set-up change and at intervals not to exceed 24 hours during continuous machine run periods. Record
More informationReliable and Real-time Wireless Sensor Networks: Protocols and Medical Applications
Reliable and Real-time Wireless Sensor Networks: Protocols and Medical Applications Octav Chipara University of Iowa https://cs.uiowa.edu/~ochipara 1 Detecting clinical deterioration at low cost Clinical
More informationMultilin EPM 4600 Metering Solution
GE Digital Energy Multilin EPM 4600 Metering Solution Multi-feed Power and Energy Metering Solution Quick Start Guide GE publication code: 1601-0297-A2 (GEK-119590A) 1601-0297-A2 Copyright 2014 GE Multilin
More informationArgentine Manufacturer of Hospital Equipment. Electrocardiograph Pulse Oximeters Vital Signs Monitors
Electrocardiograph Pulse Oximeters Vital Signs Monitors ELECTROCARDIOGRAPH Model: RG-401plus Model: RG-401 ELECTROCARDIOGRAPH Model: RG-401 Technical Specifications Accessories Carrying bag Patient cable
More informationLakeshore Technical College Official Policy
Policy Title Original Adoption Date Policy Number Information Security 05/12/2015 IT-720 Responsible College Division/Department Responsible College Manager Title Information Technology Services Director
More informationProximity Warning. Site Zone PWS. System. Personal protection & identification to help safeguard personnel from vehicle & machinery hazards
Site Zone PWS System Personal protection & identification to help safeguard personnel from vehicle & machinery hazards Creating a safer workplace Site Zone PWS Site Zone PWS key facts Reduces the risk
More informationImportant Information KX-NTV150. Communication IP Camera. IP Video Doorphone. Model No. Model No. KX-NTV160
Important Information Communication IP Camera Model No. KX-NTV150 IP Video Doorphone Model No. KX-NTV160 Thank you for purchasing this Panasonic product. Please read this Important Information carefully
More information