DICOM Conformance Statement

Size: px
Start display at page:

Download "DICOM Conformance Statement"

Transcription

1 DICOM Conformance Statement MOSAIC and MOSAIC HP PETView Version 9.40 Koninklijke Philips Electronics N.V All rights are reserved Rev. A April 2008

2 DICOM Conformance Statement Issued by: Philips Medical Systems (Cleveland), Inc. Philips Medical Systems 595 Miner Road Cleveland, Ohio USA Philips Medical Systems Nederland B.V. CTO C&S Interoperability Competence Center P.O. Box DA Best The Netherlands Internet: Document Number: Rev. A Date: April Rev. A April 2008

3 1 DICOM CONFORMANCE STATEMENT OVERVIEW 1.1 INTRODUCTION This DICOM Conformance Statement applies to MOSAIC systems using PETView software version PETView software version STORAGE SOP CLASSES All storage SOP Classes in Table 1-1 are supported. Table 1-1: Network and Print Services SOP Classes Instance Transfer User of Service (SCU) Provider of Service (SCP) Computed Tomography (CT) Image Storage Yes Stored and Viewed Magnetic Resonance (MR) Image Storage Yes Stored and Viewed Single-Frame Secondary Capture Image Storage Yes Stored and Viewed Multi-Frame Secondary Capture Image Storage Yes Stored and Viewed RT Structure Sets Yes Yes Storage Commitment Yes No Query-Retrieve Yes Yes Nuclear Medicine (NM) Image Storage Yes Stored and Viewed Positron Emission Tomography (PET) Image Storage Yes Stored and Viewed Raw Data Storage (for Philips PET raw data) Yes Yes Print Basic Grayscale Print Management Yes No Basic ColorPrint Yes No Rev. A April

4 MOSAIC DICOM Conformance Statement 1.3 ACRONYMS AND ABBREVIATIONS Table 1-2: Common Acronyms and Abbreviations ACR AE ANSI DICOM DIMSE-C DIMSE-N IOD ISO NEMA OSI PDU SCP SCU TCP/IP SOP UID American College of Radiology Application Entity American National Standards Institute Digital Imaging and Communication in Medicine DICOM Message Service Element-Composite DICOM Message Service Element-Normalized Information Object Definition International Standards Definition National Electrical Manufacturers Association Open Systems Interconnections Protocol Data Unit Service Class Provider (server) Service Class User (client) Transmission Control Protocol/Internet Protocol Service Object Pair Unique Identification Rev. A April 2008

5 2 TABLE OF CONTENTS 1 DICOM Conformance Statement Overview Introduction Storage SOP Classes Acronyms and Abbreviations Table of Contents Introduction Revision History Remarks Networking Implementation Model Application Data Flow Functional Definitions as AEs Import Server (ECHO-SCP and STORAGE-SCP) Export (ECHO-SCU and STORAGE-SCU) Print (PRINT-SCU) Query-Retrieve-SCU Query-Retrieve-SCP Sequencing of Real-World Activities AE Specifications Import Server (ECHO-SCP and STORAGE-SCP) SOP Classes Association Policies Association Acceptance Policy EXPORT (ECHO-SCU and Storage-SCU) SOP Classes Association Establishment Policies Association Initiation Policy Association Acceptance Policy Rev. A April

6 MOSAIC DICOM Conformance Statement Print (PRINT-SCU) SOP Classes Association Establishment Policies Association Initiation Policy Association Acceptance Policy Query-Retrieve-SCU SOP Classes Association Policies Association Initiation Policy Association Acceptance Policy Query-Retrieve-SCP SOP Classes Association Policies Association Initiation Policy Association Acceptance Policy Network Interfaces Physical Network Interface Additional Protocols Configuration AE Title/Presentation Address Mapping Local AE Titles Remote AE Title/Presentation Address Mapping Parameters Media Interchange Implementation Model Application Data Flow Functional Definition of AEs Functional Definition of File Set Creator for CDs Functional Definition of File Set Reader for CDs Sequencing of Real-World Activities File Meta Information Options AE Specifications File Set Creator for CDs File Meta Information for the Application Entity Rev. A April 2008

7 Real-World Activities File Set Reader for CDs Real-World Activities Augmented and Drive Application Profiles Media Configuration Support of character sets Overview Security Security Profiles Association Level Security Application Level Security Annexes IOD Contents Created SOP Instances PET Image IOD NM Image IOD Secondary Capture Image IOD Multi-Frame Grayscale Byte Secondary Capture Image IOD True-Color Secondary Capture Image IOD RT Structure Sets IOD Raw Data IOD Common Modules PET Modules NM Modules Secondary Capture Modules RT Structure Modules Usage of Attributes from Received IODs Attribute Mapping Coerced/Modified Fields Data Dictionaries Coded Terminology Context Groups Rev. A April

8 MOSAIC DICOM Conformance Statement Template Specifications Private Code Definitions Grayscale Image Consistency Standard Extended SOPS Private Transfer Syntax Rev. A April 2008

9 3 INTRODUCTION 3.1 REVISION HISTORY Table 3-1: Revision History Document Version Date of Issue Description Rev. A April 2008 Mosaic and Mosaic HP with PETView 9.40 release. 3.2 REMARKS Not applicable Rev. A April

10 MOSAIC DICOM Conformance Statement Rev. A April 2008

11 4 NETWORKING 4.1 IMPLEMENTATION MODEL Application Data Flow Rev. A April

12 MOSAIC DICOM Conformance Statement This implementation provides for simple transfer of PET, CT, NM, Secondary Capture, and MR images; RT Structure Sets; and PET Raw Data using the DICOM Storage Service Class as both a Service Class User (SCU) and a Service Class Provider (SCP). Transfers from the MOSAIC to a remote Application Entity (AE) are initiated by an operator by selecting the appropriate options from the File Manager. Query- Retrieve-SCU and Query-Retrieve-SCP are supported. No operator action is required on the MOSAIC to service Storage requests initiated by a remote Application Entry (AE). For diagnostic purposes, this implementation also provides for simple communication testing using the DICOM Verification Service Class, as both SCU and SCP. Verification of communications to a remote AE is initiated by an operator by clicking Test Connection on the DICOM Export window. Verification of communications from a remote AE is handled automatically and requires no operator actions Rev. A April 2008

13 4.1.2 Functional Definitions as AEs Import Server (ECHO-SCP and STORAGE-SCP) The Import Server waits in the background for a connection. It will accept Presentation Contexts for SOP Classes of the Verification Service Class, and will respond to echo requests. It will also accept associations with Presentation Contexts for SOP Classes of the Storage Service Class, and will store the received images to the local database where they may subsequently be listed and viewed through the user interface Export (ECHO-SCU and STORAGE-SCU) The Test Connection function provides an easy way to determine if the remote AE is available. When Test Connection is pressed, an association which includes a Presentation Context for Verification Class is proposed. A successful response indicates that the remote AE is available. The association is immediately closed. DICOM Export is activated through File Manager when a user selects one of the following image files: PET, CT, NM, Secondary Capture, MR images, RT Structure Sets, or PET Raw Data (xml) files from the local database and requests that they be sent to a remote AE (selected from a pre-configured list). Storage Commitment is supported during export, but the remote AE must be configured to accept it Print (PRINT-SCU) The Print function is available from within several processing functions. It prints to a DICOM-compatible network color printer or film imager using an implementation of the SCU role of DICOM Print Management Class. The Print function: 1. uses N-GET to request the printer status information. 2. starts a film session and film box and sets the attributes of an image/annotation box in the background according to the format defined by the application. 3. sends pixels/lut data to the service provider. 4. requests printing of a single film session. 5. requests cancellation of printing per user request Rev. A April

14 MOSAIC DICOM Conformance Statement Query-Retrieve-SCU QUERY-RETRIEVE-SCU AE initiates a new association for each query operation. Queries are initiated by the operator. When a QUERY-RETRIEVE-SCU C-MOVE request is made, the association remains open until all images have been transferred so that any error reports can be received if there are problems transferring the images. The actual image transfer occurs on a separate association initiated by the remote Application Entity Query-Retrieve-SCP The QUERY-RETRIEVE-SCP AE will never initiate Associations; it only accepts Association Requests from external DICOM AEs. The QUERY-RETRIEVE-SCP AE will accept Associations for Verification, C-FIND, and C-MOVE requests. In the case of a C-MOVE request, the The QUERY-RETRIEVE-SCP AE will issue a command to the STORAGE-SCU AE to initiate an Association with the Destination DICOM AE to send images as specified by the originator of the C-MOVE Request Sequencing of Real-World Activities All SCP (Import Server) activities are performed asynchronously in the background and are not dependent on any sequencing. All SCU (Export, Print, and Secondary Capture Image Export) activities are initiated using File Manager or a specific application Rev. A April 2008

15 4.2 AE SPECIFICATIONS Import Server (ECHO-SCP and STORAGE-SCP) SOP Classes The Import Server Application Entity provides Conformance to the following DICOM V3.0 SOP Classes. Table 4-1: SOP Classes Supported by Import Server SOP Class Name SOP Class UID Conformance Role CT Image Storage Standard SCP MR Image Storage Standard SCP NM Image Storage Standard SCP PET Image Storage Standard SCP Verification Standard SCP Storage Commitment PUSH Model Standard SCU RT Structure Set Storage Standard SCP Multi-Frame Grayscale Byte Secondary Capture Image Storage Multi-Frame True-Color Secondary Capture Image Storage Standard SCP Standard SCP Secondary Capture Standard SCP Raw Data Storage Standard SCP Rev. A April

16 MOSAIC DICOM Conformance Statement Association Policies General Import Server does not initiate associations. The Import Server AE waits for another application to connect at the presentation address configured for its Application Entity Title. The application that connects must be a DICOM application. Associations are accepted with Presentation Contexts for SOP Classes of the Storage Service Class, or the Verification Service Class. It will receive images on the Storage Service Class Presentation Contexts and create MOSAIC image files from them. It also stores a complete copy of the original DICOM message. This feature allows such images to be retransmitted exactly as they were received. Table 4-2: Maximum PDU Size Received as a SCP for Import Server Maximum PDU size received 32 KB NOTE: This option is configurable. Images including CT, MR, NM, PET, RT and SC SOP Instances can be read from a DICOM CD-R using QUERY-RETRIEVE-SCU. Images including CT, MR, NM, PET, and SC SOP Instances can also be read from non-dicom conformant CDs, but you must import the images manually using the import_dicom_file script Number of Associations Table 4-3: Number of Associations as a SCP for Import Server Maximum number of simultaneous associations Unlimited The Import Server AE will generate a separate process for each connection that is established Asynchronous Nature Import Server will only allow a single outstanding operation on an Association. Therefore, the import server will not perform asynchronous operations Implementation Identifying Information Table 4-4: DICOM Implementation Class for Import Server Implementation Class UID Rev. A April 2008

17 Association Initiation Policy Import Server does not initiate associations Association Acceptance Policy When Import Server accepts an association, it will respond to either echo requests or storage requests. If the called AE Title does not match the pre-configured AE Title for the import server, the association will be rejected Activity - Receive Echo Request Description and Sequencing of Activities When a Verification (C-ECHO) request is successfully received, it responds with a successful received message. When images are received, they are copied to the local file system, converted to PETView image files, and entered into the database. If the image received is a duplicate of a previously received instance, the old file and the database record will be overwritten with the new one Accepted Presentation Contexts Table 4-5: Acceptable Presentation Contexts for Import Server Requests Presentation Context Table Abstract Syntax Transfer Syntax Role Ext. Neg. Name UID Name List UID List Verification Implicit VR CT Image Implicit VR Explicit VR Explicit VR Big Endian MR Image Implicit VR Explicit VR Explicit VR Big Endian SCP None SCP None SCP None SCP None SCP None SCP None SCP None Rev. A April

18 MOSAIC DICOM Conformance Statement Table 4-5: Acceptable Presentation Contexts for Import Server Requests (Continued) NM Image Implicit VR Explicit VR Explicit VR Big Endian PET Image Implicit VR Explicit VR Explicit VR Big Endian SCP None SCP None SCP None SCP None SCP None SCP None Single- Frame Secondary Capture Multi-Frame True-Color Secondary Capture Multi-Frame Grayscale Byte Secondary Capture RT Structure Set Storage Implicit VR Explicit VR Explicit VR Big Endian Implicit VR Explicit VR Explicit VR Big Endian Implicit VR Explicit VR Explicit VR Big Endian Implicit VR Explicit VR Explicit VR Big Endian SCP None SCP None SCP None SCP None SCP None SCP None SCP None SCP None SCP None SCP None SCP None SCP None Storage Commitment PUSH Model Implicit VR SCU None Rev. A April 2008

19 Table 4-5: Acceptable Presentation Contexts for Import Server Requests (Continued) Raw Data Storage Implicit VR Explicit VR Explicit VR Big Endian SCP None SCP None SCP None Extended Negotiation No extended negotiation is performed SOP Specific Conformance SOP Specific Conformance to Verification SOP and Storage Classes Import Server provides standard conformance to the Verification Service and supported Storage Service Classes Presentation Context Acceptance Criterion Import Server will always accept a Presentation Context for the supported SOP Classes with the supported Transfer Syntaxes. More than one proposed Presentation Context will be accepted for the same Abstract Syntax if the Transfer Syntax is supported Transfer Syntax Selection Policies Import Server accepts the Transfer Syntaxes listed in Table 4-5. If offered a choice of Transfer Syntaxes in a Presentation Context, it will apply the following priority to the choice of Transfer Syntax: a. Explicit VR, if accepted b. Explicit VR Big Endian, if accepted c. Default Transfer Syntax Import Server will accept duplicate Presentation Contexts, that is, if it is offered multiple Presentation Contexts, each of which offers acceptable Transfer Syntaxes, it will accept all Presentation Contexts, applying the same priority for selecting a Transfer Syntax for each Rev. A April

20 MOSAIC DICOM Conformance Statement EXPORT (ECHO-SCU and Storage-SCU) SOP Classes The Export Application Entity provides Conformance to the following DICOM V3.0 SOP Classes. Table 4-6: SOP Classes Supported by Export SOP Class Name SOP Class UID Conformance Role CT Image Storage Standard SCU MR Image Storage Standard SCU NM Image Storage Standard SCU PET Image Storage Extended SCU Verification Standard SCU Storage Commitment PUSH Model Standard SCU RT Structure Set Storage Standard SCU Multi-Frame Grayscale Byte Secondary Capture Image Storage Multi-Frame True-Color Secondary Capture Image Storage Standard SCU Standard SCU Secondary Capture Standard SCU Raw Data Storage Standard SCU Association Establishment Policies General DICOM Send (Export) will attempt to establish an association each time any of its functions (Send and Test Destination) are invoked. When sending images, the association is maintained until all selected image files have been processed. Those files which cannot be transferred because the receiving AE does not support the required SOP Class are ignored. When sending the verification (Test Destination) request, the association is immediately closed as soon as the response is received. Each new request is made on a new association. Table 4-7: Maximum PDU Size Received as a SCU for Export Maximum PDU size received 4, 8, 16, or 32 KB NOTE: This option is configurable Rev. A April 2008

21 Number of Associations DICOM Send (Export) will initiate only one association at a time. However, since there can be more than one File Manager, it is possible for multiple copies to be invoked simultaneously. There is no synchronization attempted between multiple copies of the transfer software, so there may be a number of associations attempted simultaneously, limited only by the resources available. Table 4-8: Number of Associations as a SCU for Export Maximum number of simultaneous associations Unlimited Asynchronous Nature There is no asynchronous activity in this implementation Implementation Identifying Information Table 4-9: DICOM Implementation Class for Export Implementation Class UID Association Initiation Policy DICOM Send (Export) initiates an association each time you click the Export or Test Connection buttons Association Acceptance Policy DICOM Send (Export) application does not accept associations Activity - Export (ECHO-SCU and STORAGE-SCU) Description and Sequencing of Activities The Associated Real World Activity is the attempt to transfer a set of images. This occurs when the operator manually selects a set of images from File Manager and exports them using the DICOM Send (Export) function Rev. A April

22 MOSAIC DICOM Conformance Statement Proposed Presentation Contexts Table 4-10: Proposed Presentation Contexts for Export Presentation Context Table Abstract Syntax Transfer Syntax Role Ext. Neg. Name UID Name List UID List Verification Implicit VR CT Image Implicit VR Explicit VR Explicit VR Big Endian MR Image Implicit VR Explicit VR Explicit VR Big Endian NM Image Implicit VR Explicit VR Explicit VR Big Endian PET Image Implicit VR Explicit VR Explicit VR Big Endian SCU None SCU None SCU None SCU None SCU None SCU None SCU None SCU None SCU None SCU None SCU None SCU None SCU None Single-Frame Secondary Capture Implicit VR Explicit VR Explicit VR Big Endian SCU None SCU None SCU None Rev. A April 2008

23 Table 4-10: Proposed Presentation Contexts for Export (Continued) Multi-Frame True-Color Secondary Capture Multi-Frame Grayscale Byte Secondary Capture RT Structure Set Storage Raw Data Storage Implicit VR Explicit VR Explicit VR Big Endian Implicit VR Explicit VR Explicit VR Big Endian Implicit VR Explicit VR Explicit VR Big Endian Implicit VR Explicit VR Explicit VR Big Endian SCU None SCU None SCU None SCU None SCU None SCU None SCU None SCU None SCU None SCU None SCU None SCU None Storage Commitment PUSH Model Implicit VR SCU None Extended Negotiation Extended negotiation is not applicable SOP Specific Conformance Rev. A April

24 MOSAIC DICOM Conformance Statement SOP Specific Conformance to Verification and Storage SOP Class In the DICOM Export window, text appears next to each file listed indicating the transmission progress. After the image file is sent, the In Progress status changes to Done. Failed appears next to all the files if the DICOM Send (Export) function: is unable to determine the appropriate Abstract Syntax for a file. detects that the Abstract Syntax is not supported by the receiving AE. receives a failed, refused, or warning response to the C-STORE operation. The image transfer software does not attempt any extended negotiation Presentation Context Acceptance Criterion DICOM Send (Export) does not accept associations Transfer Syntax Selection Policies If the receiving system accepts more than one Transfer Syntax for any of the proposed Presentation contexts, then the DICOM Send (Export) function will select one using the following criterion: Print (PRINT-SCU) a. Explicit VR, if accepted b. Explicit VR Big Endian, if accepted c. Implicit VR SOP Classes This Application Entity provides Conformance to the following DICOM V3.0 SOP Classes. Table 4-11: Support SOP Classes SOP Class Name SOP Class UID Role Basic Grayscale Print Management Meta SCU Basic Color Print Management Meta SCU Rev. A April 2008

25 Association Establishment Policies General Table 4-12: Maximum PDU Size Received as a SCP for Print Maximum PDU size received 32 KB NOTE: This option is configurable Number of Associations Table 4-13: Number of Associations as a SCP for Print Maximum number of simultaneous associations Unlimited NOTE: This is configurable. The Print software will initiate only one association at a time. However, since there can be more than one File Manager or application, it is possible for multiple copies to be invoked simultaneously. There is no synchronization attempted between multiple copies of the print software, so there may be a number of associations attempted simultaneously, limited only by the resources available Asynchronous Nature There is no asynchronous activity in this implementation Implementation Identifying Information Table 4-14: DICOM Implementation Class and Version for Print Implementation Class UID Association Initiation Policy A new association is initiated for each print operation Association Acceptance Policy Print does not accept negotiations Rev. A April

26 MOSAIC DICOM Conformance Statement Activity - Print (PRINT-SCU) Description and Sequencing of Activities UGM DICOM prints to a DICOM compatible network color printer or film imager using an implementation of the SCU role of DICOM Print Management Class in the following way: 1. uses the N-GET to request the printer status information. 2. starts a film session and film box and sets the attributes of an image/annotation box in the background according to the format defined by the application. 3. sends pixel/lut data to the service provider. 4. requests printing of a single film session. 5. requests cancellation of printing per user s request Proposed Presentation Contexts Table 4-15: Proposed Presentation Contexts Presentation Context Table Abstract Syntax Transfer Syntax Role Ext. Neg. Name UID Name List UID List Basic Grayscale Print Management Meta SOP Class Basic Color Print Management Meta SOP Class DICOM Implicit VR Transfer Syntax DICOM Implicit VR Transfer Syntax SCU None SCU None Extended Negotiation The Print software does not attempt any extended negotiation Rev. A April 2008

27 SOP Specific Conformance to Print SOP Class SOP Specific Conformance for the Film Session SOP Class Print supports the N-CREATE DIMSE operations for the Film Session SOP Class. Details of the supported attribute and status handling behavior are described in the following subsections Film Session SOP Class Operations (N-CREATE) The attributes supplied in an N-CREATE Request are listed below. Table 4-16: Film Session SOP Class N-CREATE Request Attributes Attribute Name Tag VR Value Presence of Value Source Number of Copies (2000,0010) IS ALWAYS Medium Type (2000,0030) CS BLUE FILM, CLEAR FILM, or PAPER ALWAYS The behavior of Print when encountering status codes in a N-CREATE response is summarized below. Table 4-17: Film Session SOP Class N-CREATE Response Status Handling Behavior Service Status Further Meaning Error Code Behavior Success Success 0000 The SCP has completed the operation successfully. * * Any other status code The Association is aborted SOP Specific Conformance for the Film Box SOP Class Print supports the following DIMSE operations for the Presentation LUT SOP Class: N-CREATE N-ACTION Details of the supported attributes and status handling behavior are described in the following subsections Rev. A April

28 MOSAIC DICOM Conformance Statement Film Box SOP Class Operations (N-CREATE) The attributes supplied in a N-CREATE Request are shown below. Table 4-18: Film Session SOP Class N-CREATE Request Attributes Attribute Name Tag VR Value Presence of Value Source Image Display Format Referenced Film Session Sequence >Referenced SOP Class UID (2010,0010) CS STANDARD\1,1 ALWAYS Auto (2010,0050) SQ ALWAYS Auto (0008,1150) UI ALWAYS AUTO >Referenced SOP Instance UID (0008,1155) UI From created Film Session SOP Instance ALWAYS Auto Film Orientation (2010,0040) CS PORTRAIT or LANDSCAPE ALWAYS User Film Size ID (2010,0050) CS 8 IN X 10 IN, 10 IN X 12 IN, 10 IN X 14 IN, 14 IN X 17 IN, 24 CM X 24 CM, 24 CM X 30 CM ALWAYS CONFIG Magnification Type (2010,0060) CS REPLICATE ALWAYS User Border Density (2010,0010) CS BLACK ALWAYS User Film Box SOP Class Operations (N-ACTION) An N-ACTION Request is issued to instruct the Print SCP to print the contents of the Film Box. The Action Reply argument in a N-ACTION response is not evaluated. The behavior of Print when encountering status codes in a N-ACTION response is summarized below. Table 4-19: Printer SOP Class N-ACTION Response Status Handling Behavior Service Status Further Meaning Error Code Behavior Success Success 0000 The SCP has completed the operation successfully. * * Any other status code The Association is aborted Rev. A April 2008

29 SOP Specific Conformance for the Image Box SOP Class Print supports the N-SET DIMSE operation for the Image Box SOP Class. Details of the supported attribute and status handling behavior are described in the following subsections Image Box SOP Class Operations (N-SET) The attributes supplied in an N-SET Request are listed below. Table 4-20: Film Session SOP Class N-SET Request Attributes Attribute Name Tag VR Value Presence of Value Source Image Position (2020,0010) US 1 ALWAYS AUTO Basic Grayscale Image Sequence (2020,0110) SQ ALWAYS AUTO >Samples Per Pixel (0028,0002) US 1 ALWAYS AUTO >Photometric Interpretation (0028,0004) US 1 ALWAYS AUTO >Rows (0028,0010) US ALWAYS AUTO >Columns (0028,0011) US ALWAYS AUTO >Pixel Aspect Ratio (0028,0034) IS 1\1 ALWAYS AUTO >Bits Allocated (0028,0100) US 8 ALWAYS AUTO >Bits Stored (0028,0101) US 8 ALWAYS AUTO >High Bit (0028,0102) US 7 ALWAYS AUTO >Pixel Representation (0028,0103) US 0 ALWAYS AUTO Pixel Data (7FE0,0010) OB Pixels rendered film sheet ALWAYS AUTO Rev. A April

30 MOSAIC DICOM Conformance Statement The behavior of Hardcopy AE when encountering status codes in a N-SET response is summarized below. Table 4-21: Printer SOP Class N-SET Response Status Handling Behavior Service Status Further Meaning Error Code Behavior Success Success 0000 The SCP has completed the operation successfully. Image successfully stored in Image Box. * * Any other status code The Association is aborted Transfer Syntax Selection Policies Print always uses default transfer syntax as shown in Table Query-Retrieve-SCU SOP Classes QUERY-RETRIEVE-SCU Application Entity provides Standard Conformance to the following SOP Classes. Table 4-22: SOP Classes Supported by QUERY-RETRIEVE-SCU SOP Class Name SOP Class UID Conformance Role Study Root QUERY-RETRIEVE-SCU Information Model - Find Study Root QUERY-RETRIEVE-SCU Information Model - Move Standard SCU Standard SCU Rev. A April 2008

31 Association Policies General QUERY-RETRIEVE-SCU initiates a new association for each query operation. Queries are initiated by the operator. When a QUERY-RETRIEVE-SCU C-MOVE request is made, the association remains open until all images have been transferred so that any error reports can be received if there are problems transferring the images. The actual image transfer occurs on a separate association initiated by the remote Application Entity. Images that are retrieved will automatically appear in the patient index. Table 4-23: Maximum PDU Size Received as a SCP for QUERY-RETRIEVE-SCU Maximum PDU size received 32K NOTE: This option is configurable. Images can also be read from a CD-R using the same user interface. Only CT, MR, NM, PET, SC, Multi-frame SC, and RT Structure Sets instances can be read. Refer to File Set Reader for CDs on page 60 in this manual Number of Associations Table 4-24: Number of Associations as a SCP for QUERY-RETRIEVE-SCU Maximum number of simultaneous associations Asynchronous Nature QUERY-RETRIEVE-SCU will only allow a single outstanding operation on an Association. Therefore, it will not perform asynchronous operations negotiations Implementation Identifying Information Table 4-25: DICOM Implementation Class and Version for QUERY-RETRIEVE-SCU Implementation Class UID Association Initiation Policy QUERY-RETRIEVE-SCU attempts to initiate a new association when the user performs the query action from the user interface. If this involves recursive queries for lower query levels in the hierarchy, these will be performed on the same association. The QUERY-RETRIEVE-SCU Application Entity initiates associations Rev. A April

32 MOSAIC DICOM Conformance Statement Activity - Query Remote AE Description and Sequencing of Activities A single attempt will be made to query the remote AE. If the query fails, for whatever reason, no retry will be performed. For the entity (study, series or instance) selected from the user interface to be retrieved, a single attempt will be made to retrieve it from the selected remote AE. If the retrieve fails, for whatever reason, no retry will be performed Proposed Presentation Contexts Table 4-26: Proposed Presentation Contexts for QUERY-RETRIEVE-SCU AE Presentation Context Table Abstract Syntax Transfer Syntax Role Ext. Neg. Name UID Name List UID List Study Root QUERY- RETRIEVE- SCU Information Model - Find Study Root QUERY- RETRIEVE- SCU Information Model - Move DICOM Implicit VR Transfer Syntax DICOM Implicit VR Transfer Syntax SCU None SCU None Extended Negotiation No extended negotiation is performed. In particular, relational queries are not supported Rev. A April 2008

33 SOP Specific Conformance SOP Specific Conformance to C-FIND-SCU QUERY-RETRIEVE-SCU provides standard conformance to the supported C-FIND SOP classes. Only a single information model, Study Root, is supported. All queries are initiated at the highest level of the information model (STUDY level), and then for each response received, recursively repeated at the next lower levels (SERIES and then IMAGE levels), in order to completely elucidate the tree of instances available on the remote AE (from which the user may subsequently request a retrieval at any level). No CANCEL requests are ever issued. Unexpected attributes returned in a C-FIND response (those not requested) are ignored. Requested return attributes not returned by the SCP are ignored. Nonmatching responses returned by the SCP due to unsupported (hopefully optional) matching keys are not filtered locally by the FIND-SCU and thus will still be presented in the browser. No attempt is made to filter out duplicate responses. Specific Character Set will always be included at every query level. If present in the response, Specific Character Set will be used to identify character sets other than the default character set for display of strings in the browser. STUDY Level Table 4-27: Study Root Request Identifier for C-FIND-SCU Name Tag Types of Matching Patient's ID (0010,0020) S,*,U Patient's Name (0010,0010) S,*,U Study ID (0020,0010) S,*,U Study Description (0008,1030) S,*,U Modalities In Study (0008,0061) NONE Study Date (0008,0020) S,*,U,R Study Time (0008,0030) NONE Accession Number (0008,0050) S,*,U Study Instance UID (0020,000D) UNIQUE Rev. A April

34 MOSAIC DICOM Conformance Statement Table 4-27: Study Root Request Identifier for C-FIND-SCU (Continued) SERIES Level Series Number (0020,0011) S,*,U Series Description (0008,103E) S,*,U Modality (0008,0060) S,*,U Series Instance UID (0020,000E) UNIQUE IMAGE Level Instance Number (0020,0013) S,*,U Image ID (0054,0400) S,*,U SOP Instance UID (0008,0018) UNIQUE COMMON to all query levels Name Tag Types of Matching Specific Character Set (0008,0005) NONE The Types of Matching supported by the C-FIND SCU include the following: S indicates the identifier attribute uses Single Value Matching R indicates Range Matching U indicates Universal Matching L indicates UID lists are sent NONE indicates no matching is supported, but values for this Element are requested to be returned (i.e. universal matching) UNIQUE indicates this is the Unique Key for that query level, in which case Universal Matching or Single Value Matching is used depending on the query level. * indicates Wildcard Matching Presentation Context Acceptance Criterion QUERY-RETRIEVE-SCU does not accept associations Rev. A April 2008

35 Response Status FIND-SCU will behave as described in the table below in response to the status returned in the C-FIND and C-MOVE response command message(s). Table 4-28: Response Status for QUERY-RETRIEVE-SCU and Query Remote AE Request Service Status Further Meaning Status Codes Behavior Refused Out of Resources A700 Current query is terminated. Error Identifier does not match SOP Class A900 Current query is terminated. Unable to process Cxxx Current query is terminated. Cancel Matching terminated due to Cancel request FE00 Ignored (should never occur, since cancels never issued). Success Matching is complete - No final Identifier is supplied 0000 Current query is terminated. Pending Matches are continuing - Current Match is supplied and any Optional Keys were supported in the same manner as Required Keys FF00 Identifier used to populate browser. Matches are continuing - Warning that one or more Optional Keys were not supported for existence and/or matching for this Identifier FF01 Identifier used to populate browser SOP Specific Conformance to C-MOVE-SCU QUERY-RETRIEVE-SCU provides standard conformance to the supported C-MOVE SOP Classes. Only a single information model, Study Root, is supported. A retrieval will be performed at the STUDY, SERIES or IMAGE level depending on what level has been selected by the user in the browser. No CANCEL requests are ever issued Rev. A April

36 MOSAIC DICOM Conformance Statement STUDY level The retrieval is performed from the AE that was specified in the Retrieve AE attribute returned from the query performed by FIND-SCU. The instances are retrieved to the current application s local database by specifying the destination as the AE Title of the STORE-SCP AE of the local application. This implies that the remote C-MOVE SCP must be preconfigured to determine the presentation address corresponding to the STORE-SCP AE. The STORE-SCP AE will accept storage requests addressed to it from anywhere, so no pre-configuration of the local application to accept from the remote AE is necessary (except in so far as it was necessary to configure FIND- SCU). Table 4-29: Study Root Request Identifier for MOVE-SCU Name Tag Unique, Matching or Return Key Study Instance UID (0020,000D) U SERIES level Series Instance UID (0020,000E) U IMAGE level SOP Instance UID (0008,0018) U Presentation Context Acceptance Criterion QUERY-RETRIEVE-SCU does not accept associations Rev. A April 2008

37 Response Status MOVE-SCU will behave as described in Table 4-30 in response to the status returned in the C-MOVE response message(s). Table 4-30: Response Status for MOVE-SCU and Retrieve from Remote AE Request Service Status Further Meaning Status Codes Related Fields Behavior Refused Out of Resources - Unable to calculate number of matches A701 (0000,0902) Retrieval is terminated Out of Resources - Unable to perform sub-operations A702 (0000,1020) (0000,1021) (0000,1022) (0000,1023) Retrieval is terminated Move Destination unknown A801 (0000,0902) Retrieval is terminated Failed Identifier does not match SOP Class A900 (0000,0901) (0000,0902) Retrieval is terminated Unable to process Cxxx (0000,0901) (0000,0902) Retrieval is terminated Cancel Sub-operations terminated due to Cancel Indication FE00 (0000,1020) (0000,1021) (0000,1022) (0000,1023) Retrieval is terminated (should never occur, since cancels never issued) Warning Sub-operations Complete - One or more Failures B000 (0000,1020) (0000,1022) (0000,1023) Retrieval is terminated Success Sub-operations Complete - No Failures 0000 (0000,1020) (0000,1021) (0000,1022) (0000,1023) Retrieval is terminated Pending Sub-operations are continuing FF00 (0000,1020) (0000,1021) (0000,1022) (0000,1023) Retrieval continues Rev. A April

38 MOSAIC DICOM Conformance Statement Sub-operation Dependent Behavior Since the C-MOVE operation is dependent on completion of C-STORE suboperations that are occurring on a separate association, the question of failure of operations on the other association(s) must be considered. MOVE-SCU completely ignores whatever activities are taking place in relation to the STORAGE-SCP AE that is receiving the retrieved instances. Once the C-MOVE has been initiated it runs to completion (or failure) as described in the C-MOVE response command message(s). There is no attempt by MOVE-SCU to confirm that instances have actually been successfully received or locally stored. Whether or not completely or partially successful retrievals are made available in the local database to the user is purely dependent on the success or failure of the C- STORE sub-operations, not on any explicit action by MOVE-SCU. Whether or not the remote AE attempts to retry any failed C-STORE sub-operations is beyond the control of MOVE-SCU. If the association on which the C-MOVE was issued is aborted for any reason, whether or not the C-STORE sub-operations continue is dependent on the remote AE. The local STORAGE-SCP will continue to accept associations and storage operations regardless Association Acceptance Policy QUERY-RETRIEVE-SCU does not accept associations Rev. A April 2008

39 4.2.5 Query-Retrieve-SCP SOP Classes QUERY-RETRIEVE-SCP AE provides Standard Conformance to the following DICOM V3.0 SOP Classes: Table 4-31: SOP Classes Supported by QUERY-RETRIEVE-SCP SOP Class Name SOP Class UID Conformance Role Patient Root QUERY-RETRIEVE-SCP Information Model - Find Patient Root QUERY-RETRIEVE-SCP Information Model - Move Study Root QUERY-RETRIEVE-SCP Information Model - Find Study Root QUERY-RETRIEVE-SCP Information Model - Move Patient/Study Only QUERY-RETRIEVE- SCP Information Model - Find Patient/Study Only QUERY-RETRIEVE- SCP Information Model - Move Standard SCP Standard SCP Standard SCP Standard SCP Standard SCP Standard SCP Association Policies General The QUERY-RETRIEVE-SCP AE will never initiate Associations; it only accepts Association Requests from external DICOM AEs. The QUERY-RETRIEVE-SCP AE will accept Associations for Verification, C-FIND, and C-MOVE requests. In the case of a C-MOVE request, the The QUERY-RETRIEVE-SCP AE will issue a command to the STORAGE-SCU AE to initiate an Association with the Destination DICOM AE to send images as specified by the originator of the C-MOVE Request. The DICOM standard Application Context Name for DICOM 3.0 is always accepted: Table 4-32: DICOM Application Context for QUERY-RETRIEVE-SCP Application Context Name Rev. A April

40 MOSAIC DICOM Conformance Statement Number of Associations The QUERY-RETRIEVE-SCP AE can support multiple simultaneous Associations. Each time the QUERY-RETRIEVE-SCP AE receives an Association, a child process will be spawned to process the Verification, Query, or Retrieval request. The maximum number of child processes, and thus the maximum number of simultaneous Associations that can be processed is limited only by system resources. Table 4-33: Number of Simultaneous Associations as a SCP for QUERY-RETRIEVE-SCP Maximum number of simultaneous associations limited only by system resources Asynchronous Nature QUERY-RETRIEVE-SCP does not support asynchronous communication (multiple outstanding transactions over a single association). All association requests must be completed and acknowledged before a new operation can be initiated. Table 4-34: Asynchronous Nature as a SCP for QUERY-RETRIEVE-SCP Maximum number of outstanding asynchronous transactions Implementation Identifying Information The implementation information for the Application Entity is: Table 4-35: DICOM Implementation Class and Version for QUERY-RETRIEVE-SCP Implementation Class UID Note that the STORAGE-SCU AE and QUERY-RETRIEVE-SCP AE use the same Implementation Class UID Association Initiation Policy QUERY-RETRIEVE-SCP does not initiate associations Rev. A April 2008

41 Association Acceptance Policy Activity - Handling Query and Retrieval Requests Description and Sequencing of Activities The QUERY-RETRIEVE-SCP AE accepts Associations only if they have valid Presentation Contexts. If none of the requested Presentation Contexts are accepted, then the Association Request itself is rejected. If QUERY-RETRIEVE-SCP AE receives a query (C-FIND) request, then the response(s) will be sent over the same Association used to send the C-FIND- Request. If QUERY-RETRIEVE-SCP AE receives a retrieval (C-MOVE) request, then the responses will be sent over the same Association used to send the C-MOVE- Request. The QUERY-RETRIEVE-SCP AE will notify the STORAGE-SCU to send the requested SOP Instances to the C-MOVE Destination. The STORAGE-SCU AE notifies the QUERY-RETRIEVE-SCP AE of the success or failure of each attempt to send a Composite SOP Instance to the peer C-MOVE Destination AE. The QUERY- RETRIEVE-SCP AE then sends a C-MOVE Response indicating this status after each attempt. Once the STORAGE-SCU AE has finished attempting to transfer all the requested SOP Instances, the QUERY-RETRIEVE-SCP AE sends a final C-MOVE Response indicating the overall status of the attempted retrieval Rev. A April

42 MOSAIC DICOM Conformance Statement. Figure 4-1: Sequencing of Activity - Handling Query and Retrieval Requests The following sequencing constraints illustrated in Figure 4-1 apply to the QUERY- RETRIEVE-SCP AE for handling queries (C-FIND-Requests): 1. Peer AE opens an Association with the QUERY-RETRIEVE-SCP AE. 2. Peer AE sends a C-FIND-RQ Message. 3. QUERY-RETRIEVE-SCP AE returns a C-FIND-RSP Message to the peer AE with matching information. A C-FIND-RSP is sent for each entity matching the identifier specified in the C-FIND-RQ. A final C-FIND-RSP is sent indicating that the matching is complete. 4. Peer AE closes the Association. Note that the peer AE does not have to close the Association immediately. Further C-FIND or C-MOVE Requests can be sent over the Association before it is closed Rev. A April 2008

43 The following sequencing constraints illustrated in Figure 4-1 apply to the QUERY- RETRIEVE-SCP AE for handling retrievals (C-MOVE-Requests): 1. Peer AE opens an Association with the QUERY-RETRIEVE-SCP AE. 2. Peer AE sends a C-MOVE-RQ Message. 3. QUERY-RETRIEVE-SCP AE notifies the STORAGE-SCU AE to send the Composite SOP Instances to the peer C-MOVE Destination AE as indicated in the C-MOVE-RQ. 4. After attempting to send a SOP Instance, the STORAGE-SCU AE indicates to the QUERY-RETRIEVE-SCP AE whether the transfer succeeded or failed. The QUERY-RETRIEVE-SCP AE then returns a C-MOVE-RSP indicating this success or failure. 5. Once the STORAGE-SCU AE has completed all attempts to transfer the SOP Instances to the C-MOVE Destination AE, or the first failure occurred, the QUERY-RETRIEVE-SCP AE sends a final C-MOVE-RSP indicating the overall success or failure of the retrieval. 6. Peer AE closes the Association. Note that the peer AE does not have to close the Association immediately. Further C-FIND or C-MOVE Requests can be sent over the Association before it is closed Rev. A April

44 MOSAIC DICOM Conformance Statement The QUERY-RETRIEVE-SCP AE may reject Association attempts as shown in Table The Result, Source and Reason/Diag columns represent the values returned in the corresponding fields of an ASSOCIATE-RJ PDU (see PS 3.8, Section 9.3.4). The following abbreviations are used in the Source column: a. 1 DICOM UL service-user b. 2 DICOM UL service-provider (ASCE related function) Table 4-36: Association Rejection Reasons Result Source Reason/Diag Explanation 1 - rejected-permanent a 2 - applicationcontext-namenot-supported 1 - rejected-permanent a 7 - called-aetitle-not-recognized 1 - rejected-permanent b 1 - no-reasongiven The Association request contained an unsupported Application Context Name. An association request with the same parameters will not succeed at a later time. The Association request contained an unrecognized Called AE Title. An Association request with the same parameters will not succeed at a later time unless configuration changes are made. This rejection reason normally occurs when the Association initiator is incorrectly configured and attempts to address the Association acceptor using the wrong AE Title. The Association request could not be parsed. An Association request with the same format will not succeed at a later time Rev. A April 2008

GEMINI DICOM Conformance Statement 16 Power PET/CT Imaging System Configuration

GEMINI DICOM Conformance Statement 16 Power PET/CT Imaging System Configuration GEMINI DICOM Conformance Statement 16 Power PET/CT Imaging System Configuration Positron Emission Tomography (PET) Technical Information Software Version 9.1 1 DICOM CONFORMANCE STATEMENT OVERVIEW 1.1

More information

Punctual Dicom Workstation

Punctual Dicom Workstation DICOM Conformance Statement Punctual Dicom Workstation Company Name Product Name Product Version 2.0 Document number 2008-1001 Date June 1, 2009 Punctual Software Inc. Punctual Dicom Workstation TABLE

More information

Application Launcher 2.2 DICOM Conformance Statement

Application Launcher 2.2 DICOM Conformance Statement Contents mm761-0 Application Launcher 2.2 DICOM Conformance Statement 1 Introduction... 3 1.1 Integration and Features... 3 1.2 Definitions... 3 2 NETWORKING... 4 2.1 Implementation Model... 4 2.1.1 Application

More information

DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM MODEL SSA-640A V5.0

DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM MODEL SSA-640A V5.0 DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM TM MODEL SSA-640A V5.0 TOSHIBA MEDICAL SYSTEMS CORPORATION 2015 ALL RIGHTS RESERVED Trademarks Viamo is a trademark of Toshiba Medical Systems

More information

AVIA (Dx MM) DICOM 3.0 Conformance Statement

AVIA (Dx MM) DICOM 3.0 Conformance Statement AVIA (Dx MM) DICOM 3.0 Conformance Statement AVIA (Dx MM) is a product of Medasys Digital Systems. Table of Contents 1INTRODUCTION... 4 1.1SCOPE AND AUDIENCE... 4 1.2REFERENCES... 4 1.3ACRONYMS AND ABBREVIATIONS...

More information

1 CONFORMANCE STATEMENT OVERVIEW

1 CONFORMANCE STATEMENT OVERVIEW PMOD Technologies Ltd PMOD v3.9 Version 1.0 Date: 2017.11.02 1 CONFORMANCE STATEMENT OVERVIEW This document is a DICOM Conformance statement for PMOD application version 3.9xx, where xx denotes the build

More information

DICOM CONFORMANCE STATEMENT FOR PET/CT SCANNER. Celesteion TM V6.4 OR LATER (PCA-9000A)

DICOM CONFORMANCE STATEMENT FOR PET/CT SCANNER. Celesteion TM V6.4 OR LATER (PCA-9000A) No. 2G985-031EN*A DICOM CONFORMANCE STATEMENT FOR PET/CT SCANNER Celesteion TM V6.4 OR LATER (PCA-9000A) CANON MEDICAL SYSTEMS CORPORATION 2017-2018 ALL RIGHTS RESERVED No. 2G985-031EN*A Trademarks Celesteion

More information

DICOM CONFORMANCE STATEMENT

DICOM CONFORMANCE STATEMENT Document No.: S517-1087 Revision: B DAR-7500 DICOM CONFORMANCE STATEMENT MEDICAL SYSTEMS DIVISION Revision History History Date Revision Description 1 st Edition July 2007 - A July.2008 Entirely recreated.

More information

AG Mednet Agent DICOM Conformance Statement Version 1.3

AG Mednet Agent DICOM Conformance Statement Version 1.3 AG Mednet Agent DICOM Conformance Statement Version 1.3 AG Mednet, Inc. Page 2 CONTENTS 1 INTRODUCTION...5 1.1 REVISION HISTORY...5 1.2 AUDIENCE...5 1.3 REMARKS...5 2 NETWORKING...7 2.1 IMPLEMENTATION

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement BrightView X and XCT Acquisition Software 2.0 Koninklijke Philips Electronics N.V. 2010 All rights are reserved. Issued by: Philips Healthcare Nuclear Medicine B/L 3860 North

More information

Echology Server. DICOM Conformance Statement Volume 1. <Storage and Query/Retrieve Server> Revision 1.0 Software Version 5.

Echology Server. DICOM Conformance Statement Volume 1. <Storage and Query/Retrieve Server> Revision 1.0 Software Version 5. Echology Server DICOM Conformance Statement Volume 1 Revision 1.0 Software Version 5.0 and after Date: January 17, 2008 Prepared for Page 1 Copyright 2005-2008 EBM Technologies

More information

DICOM. Conformance Statement. Panorama Release 6. Panorama 0.6T Panorama 0.23T Panorama 0.23T I/T Panorama 0.23T R/T

DICOM. Conformance Statement. Panorama Release 6. Panorama 0.6T Panorama 0.23T Panorama 0.23T I/T Panorama 0.23T R/T DICOM Conformance Statement Panorama Release 6 Panorama 0.6T Panorama 0.23T Panorama 0.23T I/T Panorama 0.23T R/T Koninklijke Philips Electronics N.V. 2005 All rights are reserved. This DICOM Conformance

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER. Aquilion TM 32/64 (TSX-101A/D, TSX-101A/E)

DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER. Aquilion TM 32/64 (TSX-101A/D, TSX-101A/E) No. MIICT0041EA DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER Aquilion TM 32/64 (TSX-101A/D, TSX-101A/E) TOSHIBA MEDICAL SYSTEMS CORPORATION 2005 ALL RIGHTS RESERVED Trademarks Aquilion

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL DRAD-3000A, DRAD-3000E

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL DRAD-3000A, DRAD-3000E DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL DRAD-3000A, DRAD-3000E TOSHIBA MEDICAL SYSTEMS CORPORATION 2009 ALL RIGHTS RESERVED Trademarks This document may include trademarks

More information

1 Introduction Scope and audience References Acronyms and abbreviations 3. 2 Implementation model... 4

1 Introduction Scope and audience References Acronyms and abbreviations 3. 2 Implementation model... 4 NAME FONCTION VISA AUTHOR Alain Battistini Senior Engineer CONFORM TO ORIGINAL REVIEW Jean-François Hoh Senior Engineer CONFORM TO ORIGINAL APPROVAL Pascal Gilbert-Jeantet CONFORM TO ORIGINAL Revision

More information

Sep, th Edition 897N101668H

Sep, th Edition 897N101668H DICOM Conformance Statement Console Advance DR-ID 300CL/700CL 800CL/900CL for DICOM Storage DICOM Storage Commitment DICOM MWM DICOM MPPS DICOM Print DICOM Query / Retrieve DICOM Media Storage DICOM Dose

More information

FUSION RIS 3.30 DICOM Conformance Statement

FUSION RIS 3.30 DICOM Conformance Statement FUSION RIS 3.30 DICOM Conformance Statement FUSION RIS 3.30 DICOM Conformance Statement Copyright 2007 by. All rights reserved. Unauthorized use, reproduction, or disclosure is prohibited. 6737 W. Washington

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER. Aquilion ONE TM,Aquilion TM V4.63 OR LATER (TSX-301A,TSX-301B)

DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER. Aquilion ONE TM,Aquilion TM V4.63 OR LATER (TSX-301A,TSX-301B) No.MIICT0055EAA DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER Aquilion ONE TM,Aquilion TM V4.63 OR LATER (TSX-301A,TSX-301B) TOSHIBA MEDICAL SYSTEMS CORPORATION 2010 ALL RIGHTS RESERVED

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA SUPERCONDUCTING MRI SYSTEMS VANTAGE-TITAN TM

DICOM CONFORMANCE STATEMENT FOR TOSHIBA SUPERCONDUCTING MRI SYSTEMS VANTAGE-TITAN TM DICOM CONFORMANCE STATEMENT FOR TOSHIBA SUPERCONDUCTING MRI SYSTEMS VANTAGE-TITAN TM TOSHIBA MEDICAL SYSTEMS CORPORATION 2011 ALL RIGHTS RESERVED Trademarks VANTAGE-TITAN is a trademark of Toshiba Medical

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER

DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER No. MIICT0044EA DICOM CONFORMANCE STATEMENT FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER Aquilion TM /Asteion TM V1.61 (TSX-101A/G, TSX-101A/F, TSX-101A/7, TSX76-101A/9, TSX-201A, TSX-021B/4, TSX-021B/5) TOSHIBA

More information

DICOM Conformance Statement. Allura FC II

DICOM Conformance Statement. Allura FC II DICOM Conformance Statement Allura FC II Koninklijke Philips N.V. 2017 All rights are reserved. Document Number: PIIOffc.0001274 Issued by: Philips Medical Systems Nederland BV, a Philips Healthcare company,

More information

DICOM Conformance Statement. Fusion RIS Version 3.1

DICOM Conformance Statement. Fusion RIS Version 3.1 DICOM Conformance Statement Fusion RIS Version 3.1 571 Boston Mills Road, Suite 500 Hudson, OH 44236 330-342-4800 (Main) 866-747-5644 (Toll Free) 330-342-4848 (Fax) http://www.merge-emed.com/ Page 1 of

More information

QPC XSCAN32 DICOM 3.0 Conformance Statement Document Version October 2004

QPC XSCAN32 DICOM 3.0 Conformance Statement Document Version October 2004 QPC XSCAN32 DICOM 3.0 Conformance Statement Document Version 1.3.0 28 October 2004 QPC XSCAN32 DICOM 3.0 CONFORMANCE STATEMENT Revision 1.3.0 2 8 October 2004 Page 1 of 22 Table of Contents 0 INTRODUCTION...

More information

XPACS DICOM 3.0 Conformance Statement Document Version October 2004

XPACS DICOM 3.0 Conformance Statement Document Version October 2004 XPACS DICOM 3.0 Conformance Statement Document Version 1.2.0 28 October 2004 XPACS DICOM 3.0 CONFORMANCE STATEMENT Revision 1.3.0 2 8 October 2004 Page 1 of 25 Table of Contents 0 INTRODUCTION... 4 0.1

More information

Digital Lightbox 1.0

Digital Lightbox 1.0 Technical Publication DICOM Conformance Statement 1.0 Document Revision 1 June 3, 2008 2008 Copyright BrainLAB AG 1 Conformance Statement Overview This is a conformance statement for the BrainLAB software.

More information

POP-Net Essential DICOM Conformance Statement

POP-Net Essential DICOM Conformance Statement POP-Net Essential DICOM Conformance Statement Revision 4.0 Software Version 4.0 and after Date: February 9, 2005 Prepared for Page 1 Copyright 2000-2005 EBM Technologies, Inc. Head office Japan office

More information

DICOM Conformance Statement. CR-VW 674 for DICOM Storage DICOM Print DICOM MWM DICOM MPPS DICOM Media Storage

DICOM Conformance Statement. CR-VW 674 for DICOM Storage DICOM Print DICOM MWM DICOM MPPS DICOM Media Storage DICOM Conformance Statement CR-VW 674 for DICOM Storage DICOM Print DICOM MWM DICOM MPPS DICOM Media Storage December, 2011 7th Edition Copyright FUJIFILM Corporation, Japan 1 CONTENTS 1... CONFORMANCE

More information

Copyright FUJIFILM Corporation, Japan. August, th Edition 897N100760F

Copyright FUJIFILM Corporation, Japan. August, th Edition 897N100760F DICOM Conformance Statement ***** FDR-1000AWS CR-IR363AWS for DICOM Storage DICOM Storage Commitment DICOM MWM DICOM MPPS DICOM Print DICOM Query / Retrieve DICOM Media Storage (Standard) August, 2010

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement EP Navigator R4.0 All rights are reserved. DICOM Conformance Statement Page 2 of 43 Issued by: Philips Medical Systems Nederland B.V. CTO/ C&S - Interoperability Competence

More information

DICOM CONFORMANCE STATEMENT FOR VANTAGE-GALAN TM VERSION V5.0

DICOM CONFORMANCE STATEMENT FOR VANTAGE-GALAN TM VERSION V5.0 DICOM CONFORMANCE STATEMENT FOR VANTAGE-GALAN TM VERSION V5.0 CANON MEDICAL SYSTEMS CORPORATION 2018 ALL RIGHTS RESERVED Trademarks VANTAGE-GALAN is trademark of Canon Medical Systems Corporation. This

More information

No. MIIMS0009EA DICOM CONFORMANCE STATEMENT FOR MODEL TFS-3000 (MIIMS0009EA) TOSHIBA CORPORATION 2001 ALL RIGHTS RESERVED

No. MIIMS0009EA DICOM CONFORMANCE STATEMENT FOR MODEL TFS-3000 (MIIMS0009EA) TOSHIBA CORPORATION 2001 ALL RIGHTS RESERVED DICOM CONFORMANCE STATEMENT FOR MODEL TFS-3000 (MIIMS0009EA) TOSHIBA CORPORATION 2001 ALL RIGHTS RESERVED IMPORTANT! (1) No part of this manual may be copied or reprinted, in whole or in part, without

More information

Version 7 DICOM Conformance Statement. Document Version 3.02, June 2009

Version 7 DICOM Conformance Statement. Document Version 3.02, June 2009 Version 7 DICOM Conformance Statement Document Version 3.02, June 2009 1 Conformance Statement Overview The application described in this Conformance Statement VEPRO EMR Manager PACS is a collection of

More information

Dx Server for Windows NT DICOM 3.0 Conformance Statement

Dx Server for Windows NT DICOM 3.0 Conformance Statement Dx Server for Windows NT DICOM 3.0 Conformance Statement NAME FONCTION VISA AUTHOR Alain Battistini Senior Engineer CONFORM TO ORIGINAL REVIEW Jean-François Hoh Senior Engineer CONFORM TO ORIGINAL APPROVAL

More information

Dx Server for Windows NT DICOM 3.0 Conformance Statement

Dx Server for Windows NT DICOM 3.0 Conformance Statement Dx Server for Windows NT DICOM 3.0 Conformance Statement NAME FONCTION VISA AUTHOR Alain Battistini Senior Engineer CONFORM TO ORIGINAL REVIEW Jean-François Hoh Senior Engineer CONFORM TO ORIGINAL APPROVAL

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement Brilliance Workspace Portal V1 All rights are reserved. MAR 2007 Issued by: Philips Medical Systems Nederland B.V. CTO C&S Interoperability Competence Center P.O. Box 10.000

More information

StellarPACS DICOM Conformance Statement. Version 1.3, August 2008 SoftTeam Solutions

StellarPACS DICOM Conformance Statement. Version 1.3, August 2008 SoftTeam Solutions StellarPACS DICOM Conformance Statement Version 1.3, August 2008 SoftTeam Solutions www.softteam.com Table of Contents 1 INTRODUCTION... 3 1.1 REFERENCE... 3 1.2 DEFINITIONS... 3 1.3 ACRONYMS, ABBREVIATIONS

More information

Medical Imaging Consultants, Inc.

Medical Imaging Consultants, Inc. IA2000 Conformance Statement 1 Medical Imaging Consultants, Inc. Conformance Statement IA2000 Document: MICI-210-01 Revision: 6.0 Document Status: Approved When printed, this is NOT a controlled copy Reviewers

More information

Technical Publication. DICOM Conformance Statement. Patient Browser 2.1. Document Revision 2. April 13, Copyright BrainLAB AG

Technical Publication. DICOM Conformance Statement. Patient Browser 2.1. Document Revision 2. April 13, Copyright BrainLAB AG Technical Publication DICOM Conformance Statement 2.1 Document Revision 2 April 13, 2011 2011 Copyright BrainLAB AG 1 Conformance Statement Overview This is a conformance statement for the BrainLAB software.

More information

OASIS V4.0 DICOM Conformance Statement Rev. 3

OASIS V4.0 DICOM Conformance Statement Rev. 3 0BMR Imaging System OASIS V4.0 DICOM Conformance Statement Rev. 3 Tokyo, Japan E1E-BM7403-3 Copyright Hitachi Medical Corporation. 2011. All rights reserved. Revision History Revision Date Change Description

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM. Infinix Celeve-i series / Infinix-i series Model XIDF-3DP801 AND ANGIO WORKSTATION

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM. Infinix Celeve-i series / Infinix-i series Model XIDF-3DP801 AND ANGIO WORKSTATION No. MIIXR0024EA DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM Infinix Celeve-i series / Infinix-i series Model XIDF-3DP801 AND ANGIO WORKSTATION Model XIDF-AWS801 TOSHIBA MEDICAL SYSTEMS

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement Brilliance Workspace Portal V2 All rights are reserved. JUN 2007 Issued by: Philips Medical Systems Nederland B.V. Medical IT, Interoperability Building QV-282 P.O. Box 10.000

More information

DBSWIN DICOM Conformance Statement. DBSWIN DD-DICOM Interface. DICOM Conformance Statement V2.1

DBSWIN DICOM Conformance Statement. DBSWIN DD-DICOM Interface. DICOM Conformance Statement V2.1 DBSWIN DD-DICOM Interface DICOM Conformance Statement V2.1 DÜRR DENTAL Page 1 / 23 Conformance Statement Overview The DBSWIN software controls the digital Dürr Dental imaging products. The DD-DICOM Interface

More information

Aloka ProSound DICOM

Aloka ProSound DICOM 11/18/09 Page 1 of 17 Aloka ProSound DICOM Aloka ProSound C3 and C3cv Ultrasound Imaging System ProSound Versions 1.0 and 1.1 DICOM 11/18/09 Page 2 of 17 Table of Contents 1 Introduction...4 1.1 Scope...4

More information

MediPlus TM PACS&RIS Version 2.6

MediPlus TM PACS&RIS Version 2.6 Revision 2.0, Publication #7143-202 MediPlus TM PACS&RIS Version 2.6 DICOM 3.0 Conformance Statement June 2007 Revision 2.0, Publication #7143-202 MediPlus PACS&RIS V2.6 DICOM Conformance Statement Revision

More information

ClearCanvas Workstation DICOM Conformance Statement. Document Version: 1.5

ClearCanvas Workstation DICOM Conformance Statement. Document Version: 1.5 ClearCanvas Workstation DICOM Conformance Statement Document Version: 1.5 Product Name(s): Release: ClearCanvas Workstation 2.0 SP1 Date: February 22, 2010 1 CONFORMANCE STATEMENT OVERVIEW 1.1 OVERVIEW

More information

DICOM Conformance Statement * /02* /02 MADE IN GERMANY

DICOM Conformance Statement * /02* /02 MADE IN GERMANY EN DICOM 9000-608-109/02 *9000-608-109/02* MADE IN GERMANY Overview Vet-Exam Intra controls the digital DÜRR MEDICAL imaging products. The Vet-Exam Intra DICOM Interface translates between DICOM applications

More information

CoActiv, LLC CoActiv Medical Business Solutions EXAM-PACS Conformance Statement

CoActiv, LLC CoActiv Medical Business Solutions EXAM-PACS Conformance Statement CoActiv EXAM-PACS DICOM Conformance Statement Page 1 of 40 0. COVER PAGE CoActiv, LLC CoActiv Medical Business Solutions EXAM-PACS Conformance Statement The information contained in this document is subject

More information

Version 9 DICOM Conformance Statement. Version 3.05, September 2015

Version 9 DICOM Conformance Statement. Version 3.05, September 2015 Version 9 DICOM Conformance Statement Version 3.05, September 2015 1 Conformance Statement Overview The application described in this conformance statement, VEPRO EMR Manager PACS, is a collection of processes

More information

Philips Medical Systems. Xcelera R1.1L1. Document Number XPR July 2003

Philips Medical Systems. Xcelera R1.1L1. Document Number XPR July 2003 Philips Medical Systems DICOM CONFORMANCE STATEMENT Xcelera R1.1L1 Document Number XPR 080-030073 11 July 2003 Copyright Philips Medical Systems Nederland B.V. 2003 All rights reserved 3 DICOM Conformance

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM. Infinix Celeve-i series / Infinix-i series Model DFP-8000 series V4.

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM. Infinix Celeve-i series / Infinix-i series Model DFP-8000 series V4. No. MIIXR0010EAL DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM Infini Celeve-i series / Infini-i series Model DFP-8000 series V4.40 TOSHIBA MEDICAL SYSTEMS CORPORATION 2011 ALL RIGHTS

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL DI Station TM ADR-1000A /E2,RF TOSHIBA MEDICAL SYSTEMS CORPORATION 1998-2007 ALL RIGHTS RESERVED Trademarks DI Station is a trademark

More information

Technical Publications

Technical Publications Technical Publications Direction DOC0913087 Revision 1.0 VOLUSON Voluson S6/S8 11.x.x Copyright 2011 by General Electric Co. Do not duplicate GE Medical Systems THIS PAGE LEFT INTENTIONALLY BLANK REVISION

More information

Printlink III-ID/IV. Revision History. Date Version Description 20/08/2004 Ver First edition

Printlink III-ID/IV. Revision History. Date Version Description 20/08/2004 Ver First edition Revision History Date Version Description 20/08/2004 Ver. 1.00 First edition NOTE: Descriptions in this document are subject to change without prior notice. Printlink III-ID/IV DICOM 3.0 Conformance Statement

More information

Philips Medical Systems. Intera Document Number April 2003

Philips Medical Systems. Intera Document Number April 2003 Philips Medical Systems DICOM CONFORMANCE STATEMENT Intera 10.1 Document Number 4522 131 88821 10 April 2003 Copyright Philips Medical Systems Nederland B.V. 2003 All rights reserved DICOM Conformance

More information

SecurView RT/TechMate Workstation DiCom Conformance statement

SecurView RT/TechMate Workstation DiCom Conformance statement breast imaging solutions DICOM Standard Interface SecurView RT/TechMate Workstation DiCom Conformance statement Software Version 8.4 man-03818 revision 002 SecurView RT/TechMate 8.4 Workstation DICOM

More information

Technical Publications

Technical Publications GE HEALTHCARE DIRECTION 5184517-100REV 3.0 LOGIQ A5 3.X.X Technical Publications Direction 5184517-100 Revision 3.0 LOGIQ A5 version 3.x.x for DICOM Copyright 2009 By General Electric Co. Do not duplicate

More information

CHILI CHILI PACS. Digital Radiology. DICOM Conformance Statement CHILI GmbH, Heidelberg

CHILI CHILI PACS. Digital Radiology. DICOM Conformance Statement CHILI GmbH, Heidelberg CHILI Digital Radiology CHILI PACS DICOM Conformance Statement 2010-10 2013 CHILI GmbH, Heidelberg Manufacturer CHILI GmbH Friedrich-Ebert-Str. 2 D-69221 Dossenheim/Heidelberg Tel. (+49) 6221-18079-10

More information

DE32-DCM DentalEye 3.2. DICOM conformance statement

DE32-DCM DentalEye 3.2. DICOM conformance statement DentalEye 3.2 Table of contents 1. INTRODUCTION... 4 1.1. DentalEye Enterprise Application... 4 1.2. About this Document... 4 1.3. Integration and Validation... 4 1.4. Future Evolution... 5 1.5. Sources

More information

Technical Publications

Technical Publications g GE HEALTHCARE Technical Publications Direction 00-888080-01 Revision A GE OEC 1Kx1K Elite Workstation Copyright 2005 by General Electric Co. Do not duplicate g GE Healthcare g GE HEALTHCARE THIS PAGE

More information

X-Porte DICOM Conformance Statement

X-Porte DICOM Conformance Statement 21919 30th Dr. SE, Bothell, WA 98021-3904 USA Telephone 1.425.951.1200 Facsimile 1.425.951.1201 www.sonosite.com Document Number: Revision: Title: D10711 D X-Porte DICOM Conformance Statement CHANGE HISTORY:

More information

Technical Publications

Technical Publications DIRECTION 5177444-100 REV 02 Technical Publications Direction 5177444-100 Revision 02 LOGIQ e version 4.x.x for DICOM Copyright 2006 By General Electric Co. g Do not duplicate GE Ultrasound DIRECTION 5177444-100

More information

Dx Server for Windows NT DICOM 3.0 Conformance Statement

Dx Server for Windows NT DICOM 3.0 Conformance Statement Dx Server for Windows NT DICOM 3.0 Conformance Statement NAME FONCTION VISA AUTHOR Alain Battistini Senior Engineer CONFORM TO ORIGINAL REVIEW Jean-François Hoh Senior Engineer CONFORM TO ORIGINAL APPROVAL

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement DICOM Pro Plus as an SCU Array Corporation Index 0. INTRODUCTION...3 1. IMPLEMENTATION MODEL...4 1.1 APPLICATION DATA FLOW DIAGRAM...4 1.2 FUNCTIONAL DEFINITION OF APPLICATION

More information

M517-E092C Feb Digital Radiography System SDR-150C. DICOM Conformance Statement

M517-E092C Feb Digital Radiography System SDR-150C. DICOM Conformance Statement M517-E092C Feb.2016 Digital Radiography System SDR-150C DICOM Conformance Statement This page is intentionally left blank. Conformance Statement Overview The application software of SDR-150C is called

More information

DICOM 3.0 Conformance Statement. DICOM PRI v. 4.64

DICOM 3.0 Conformance Statement. DICOM PRI v. 4.64 DICOM 3.0 Conformance Statement DICOM PRI v. 4.64 Version 11.0 September 2011 Summary : This document presents the conformance statement of DICOM PRI that implements as Service Class Provider (SCP) the

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement for (For Rev.01.01.00 or later) Apr. 2015 NO TEXT 2 Overview: This conformance statement details the compliance to DICOM 3.0 of Side Station i3 for SONIALVISION G4 system. Table

More information

Philips Medical Systems DICOM Conformance Statement. Inturis Suite R2.2

Philips Medical Systems DICOM Conformance Statement. Inturis Suite R2.2 Philips Medical Systems DICOM Conformance Statement Inturis Suite R2.2 Document Number XZR 080-010044 8 February 2002 Copyright Philips Medical Systems Nederland B.V. 2002 Page ii DICOM Conformance Statement

More information

Technical Publications

Technical Publications DIRECTION 2325843-100REV 6.01 LOGIQ 9 Technical Publications Direction 2325843-100 Revision 6.01 LOGIQ 9 CONFORMANCE STATEMENT for DICOM Do not duplicate Copyright 2005 By GE Healthcare DIRECTION 2325843-100REV

More information

Technical Publications

Technical Publications Technical Publications Direction 2316173-100 Revision 8.00 LOGIQ 7 CONFORMANCE STATEMENT for DICOM Do not duplicate Copyright ª 2008 By GE Healthcare THIS PAGE LEFT INTENTIONALLY BLANK REVISION HISTORY

More information

Technical Publication. DICOM Conformance Statement. BrainSUITE NET 1.5. Document Revision 1. March 5, Copyright BrainLAB AG

Technical Publication. DICOM Conformance Statement. BrainSUITE NET 1.5. Document Revision 1. March 5, Copyright BrainLAB AG Technical Publication DICOM Conformance Statement Document Revision 1 March 5, 2009 2009 Copyright BrainLAB AG 1 Conformance Statement Overview This is a conformance statement for the BrainLAB software

More information

Technical Publications

Technical Publications Technical Publications Direction FC250039 Revision 16 Vivid7 version 6 for DICOM Copyright 2007 By General Electric Co. g Do not duplicate GE Ultrasound THIS PAGE LEFT INTENTIONALLY BLANK TABLE OF CONTENTS

More information

Technical Publications

Technical Publications Technical Publications Direction DOC0637451 Revision 1 Vivid i version 10 for DICOM Copyright 2009 By General Electric Co. g Do not duplicate GE Ultrasound THIS PAGE LEFT INTENTIONALLY BLANK TABLE OF CONTENTS

More information

Technical Publications

Technical Publications GE HEALTHCARE DIRECTION 5178040-100REV 2.0 LOGIQ P5 2.X.X Technical Publications Direction 5178040-100 Revision 2.0 LOGIQ P5 version 2.x.x for DICOM Copyright ª 2007 By General Electric Co. g Do not duplicate

More information

CONFORMANCE STATEMENT

CONFORMANCE STATEMENT Philips Medical Systems CONFORMANCE STATEMENT BV Family XA Document Number XPR 080-020061 July 15, 2002 Copyright Philips Medical Systems Nederland B.V. 2002 All rights reserved DICOM Conformance Statement

More information

Cover will be supplied by PMG MARCOM group.

Cover will be supplied by PMG MARCOM group. Cover will be supplied by PMG MARCOM group. DICOM Conformance Statement Philips Medical Systems DICOM CONFORMANCE STATEMENT ViewForum 2003 Release 3.2 Document Number 4522 170 02292 21 January 2004 Copyright

More information

DICOM CONFORMANCE STATEMENT FOR ZIOBASE 4.0

DICOM CONFORMANCE STATEMENT FOR ZIOBASE 4.0 DICOM CONFORMANCE STATEMENT FOR ZIOBASE 4.0 DICOM Conformance Statement, Ziobase 4.0 1 Copyright 2006-2012, Ziosoft, Inc. 0 DICOM CONFORMANCE STATEMENT OVERVIEW... 4 1 IMPLEMENTATION MODEL... 5 1.1 APPLICATION

More information

DICOM 3.0 Conformance Statement DICOMlink for DELTAmanager Send

DICOM 3.0 Conformance Statement DICOMlink for DELTAmanager Send DICOM 3.0 Conformance Statement DICOMlink for DELTAmanager Send Mac OS Part number: 500-505 Windows Part number: 500-514 Document #: Copyright 1996-2008 MedImage, Inc. All Rights Reserved. MedImage, Inc.

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA SUGICAL C-ARM SYSTEM. SXT series. Model SXT-2000A ( CXVIEW )

DICOM CONFORMANCE STATEMENT FOR TOSHIBA SUGICAL C-ARM SYSTEM. SXT series. Model SXT-2000A ( CXVIEW ) DICOM CONFORMANCE STATEMENT FOR TOSHIBA SUGICAL C-ARM SYSTEM SXT series Model SXT-2000A ( CXVIEW 1.2.1 ) TOSHIBA MEDICAL SYSTEMS CORPORATION 2012 ALL RIGHTS RESERVED Trademarks SXT-2000A is a trademark

More information

HS40 Ultrasound System

HS40 Ultrasound System HS40 Ultrasound System DICOM Conformance Statement Revision 1.0 System Version 1.00 Created: Apr 17, 2017 Last Updated: Apr 17, 2017 HS40 DICOM Conformance Statement 0 COVER PAGE Company Name: SAMSUNG

More information

FusionXD 2.1 DICOM Conformance Statement

FusionXD 2.1 DICOM Conformance Statement FusionXD 2.1 DICOM Conformance Statement Contents 1 Introduction... 3 1.1 References... 3 1.2 Integration and Features... 3 1.3 Definitions, Acronyms and Abbreviations... 3 2 Services... 4 3 Networking...

More information

Matrix Server and Client DICOM Conformance Statement

Matrix Server and Client DICOM Conformance Statement Matrix Server and Client DICOM Conformance Statement Revision: 3.0 October 2005 All information contained in this document is accurate to the best knowledge of Merge emed as of date of publication and

More information

Technical Publications

Technical Publications DIRECTION 2390126-100REV2 Technical Publications Direction 2390126-100 Revision 2 LOGIQ3 /Expert/Pro version 4.x.x for DICOM Copyright 2005 By General Electric Co. g Do not duplicate GE Ultrasound DIRECTION

More information

Technical Publications

Technical Publications g GE Healthcare Technical Publications Direction DOC1629000 Revision 2 LOGIQ P9/P7 R2.x Copyright 2016 by General Electric Co. g GE Healthcare Do not duplicate THIS PAGE LEFT INTENTIONALLY BLANK CONFORMANCE

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement TOPCON Corporation TOPCON SPECULAR MICROSCOPE SP-1P Version 1.10 Document Version: 1 Date: 2014-02-17 1 CONFORMANCE STATEMENT OVERVIEW This document declares conformance to

More information

Infinity Medical Soft, Inc ,Higashinaganuma,Inagishi, Tokyo JAPAN

Infinity Medical Soft, Inc ,Higashinaganuma,Inagishi, Tokyo JAPAN DICOM 3.0 Conformance Statement MIE-IAS v 8.1(.Net Version) Part number: 0300001-0001 02/05/2003 Weiping-Ren Part number: 0401001-0001 10/11/2003 Makoto-Okazaki Part number: 0402003-0001 01/30/2004 Makoto-Okazaki

More information

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL HDR-08A

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL HDR-08A DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL HDR-08A TOSHIBA MEDICAL SYSTEMS CORPORATION 2008-2010 ALL RIGHTS RESERVED Trademarks This document may include trademarks or registered

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement. UNIQ Release 1.0 Koninklijke Philips N.V. 2016 All rights are reserved. ICAP-T-030001.09b Corresponds to ICAP-W-030001.02 Issued by: Philips Medical Systems Nederland BV, a

More information

SCORE 3D Workstation DICOM CONFORMANCE STATEMENT

SCORE 3D Workstation DICOM CONFORMANCE STATEMENT Document No.: S517-E105 Revision No.: A DIGITAL ANGIOGRAPHY SYSTEM DAR-9500f SCORE 3D Workstation DICOM CONFORMANCE STATEMENT Revision History Rev. Date Content of Change First 2013.09 Newly issued. A

More information

CADstream DICOM Conformance Statement

CADstream DICOM Conformance Statement CADstream 5.2.9 DICOM Conformance Statement CADstream 5.2.9 DICOM Conformance Statement This document has been prepared by Cedara Software Corp. doing business as Merge OEM (hereinafter referred to as

More information

Technical Publications

Technical Publications Technical Publications Direction 5141741-100 Revision 01 LOGIQ5 Expert version 4.x.x for DICOM Copyright 2005 By General Electric Co. g Do not duplicate GE Ultrasound THIS PAGE LEFT INTENTIONALLY BLANK

More information

DICOM Conformance Statement. Philips IntelliSpace Portal V6.0

DICOM Conformance Statement. Philips IntelliSpace Portal V6.0 DICOM Conformance Statement Philips IntelliSpace Portal V6.0 Koninklijke Philips Electronics N.V. 2013 All rights are reserved. Document Number: PIIOffc.0001070 Issued by: Philips Medical Systems Nederland

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement SCENARIA Tokyo, Japan E1E-HC0002-01 Hitachi, Ltd. 2016. All rights reserved. History Revision Description Date Rev. 1.0 Initial. 2016/04/01 ii E1E-HC0002 Contents History...

More information

FusionXD 2.0 DICOM Conformance Statement

FusionXD 2.0 DICOM Conformance Statement FusionXD 2.0 DICOM Conformance Statement 1 Introduction 1.1 Revision History See the Revision History in the beginning of this document 1.2 References Reference REF1 Document Digital Imaging and Communications

More information

SecurView DX Workstation DICOM Conformance Statement

SecurView DX Workstation DICOM Conformance Statement breast imaging solutions DICOM Standard Interface SecurView DX Workstation DICOM Conformance Statement Software Version 8.2 MAN-03404 Revision 001 SecurView DX 8.2 Workstation DICOM Conformance Statement

More information

DICOM Conformance Statement. for ES9410 DICOM

DICOM Conformance Statement. for ES9410 DICOM DICOM Conformance Statement. for ES9410 DICOM (DICOM Embedded Printer) 1 DICOM 3.0 Conformance Statement Summary: This document is the DICOM Conformance Statement of the Print Service Class Provider (SCP)

More information

NumaStore 1.0 DICOM 3.0 Conformance Statement

NumaStore 1.0 DICOM 3.0 Conformance Statement External Distribution Revision 2 February 19, 2001 NumaStore 1.0 Revision 2 Date February 19, 2001 Author Bill Kilgore, George Panousis Revision History Revision Date By Comment 1 January 13, 2001 Bill

More information

DICOM Conformance Statement Product Name HCP DICOM Net Version

DICOM Conformance Statement Product Name HCP DICOM Net Version Title DICOM Conformance Statement Product Name HCP DICOM Net Version 5.00.00 Copyright - 2000-2013 SoftLink International Pvt. Ltd. SoftLink International Pvt. Ltd. Page 1 of 37 2, Anand Park, Aundh, Pune

More information

efx Software DICONDE Conformance Statement

efx Software DICONDE Conformance Statement efx Software DICONDE Conformance Statement Version 1.2 9/29/2015 North Star Imaging, Inc. 1 DICOM conformance statement overview The North Star Imaging s efx Software is an imaging viewing software with

More information

Technical Publications

Technical Publications Technical Publications Direction EP250409 Revision 21 EchoPAC PC version 5 for DICOM (From version 5.2.2) Copyright 2008 By General Electric Co. Do not duplicate GE Ultrasound THIS PAGE LEFT INTENTIONALLY

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement ECLOS Tokyo, Japan (SN-C357E) Copyright Hitachi Medical Corporation. 2007, 2008. All rights reserved. History Revision Description Date Rev. 1.0 Initial. 2007/02/01 Rev 2.0

More information