Retrospective Spiral Respiratory Correlated Imaging with Varian RPM
|
|
- Zoe Lee Lyons
- 5 years ago
- Views:
Transcription
1 Retrospective Spiral Respiratory Correlated Imaging with Varian RPM This is a Quick Step Guide for Retrospective Spiral Respiratory Correlated Imaging (4D CT) using the Varian RPM device v1.7 with the Brilliance CT Big Bore v2.2.2 system and the Brilliance v For more detailed information, please refer to the Varian RPM operator manual and reference guide. Before starting, verify the following items are set correctly on the CT scanner and the RPM workstation. 1. Connect the Varian RPM Sensor Device to the gantry. 2. Ensure that the Varian RPM computer is turned on, is connected to the hospital network, and is fully booted up before logging in to the CT Host. 3. Wait at least 5 seconds after the login prompt appears before logging in to the CT Host. Note Restart the CT Host daily. 4. On the CT scanner, Select Preferences in the Directory tool panel. 5. Select Scanner from menu. 6. In the Pulmo Sensor Device option, select Varian. 7. Click OK to accept. 8. Place the marker block on the patient s chest or abdomen. 9. Review the procedure with the patient emphasizing the importance of maintaining a consistent breathing pattern. Refer to the Varian RPM reference guide for information about usage of Audio or Visual coaching. 1
2 Start Study on the CT scanner 1. Click Start Study to begin a new study. 2. Complete the patient data form. 3. Click Exam Protocol Groups. 4. Select the Thorax Protocol group. 5. Select Pulmo Retrospective Spiral Protocol. Launch the Varian RPM workstation 1. Login to the Varian RPM Workstation. 2. Double click the RPM Respiratory Gating Icon. 3. Close the Patient List. 4. Proceed with System Configuration. 5. Click on the View Menu. 6. Select System Configuration. 2
3 7. Click on the CT Interface Tab. CT Interface Tab 8. Select Trigger based on phase value. 9. Set Trigger At: to 0 degrees. 10. Set CT Trigger Length: to 0.50 sec. 11. Select Sensed via Digital I/O. 12. Set Binary Value indicating X-ray on: to [1] 13. Select No new pulse after and set to 180 sec. 14. Click File Export Tab 3
4 File Export Tab 1. Select Prompt to export a respiration data file after recording is finished. 2. Select Recalculate phase before respiration data export. Note Select Recalculate phase before respiration data export, only if you choose to recalculate phase before exporting. 3. Export filename format must be set to Patient ID_Exam Number_Series Number.vxp. 4. Export file location must be set to C:\Documents and Settings\VARIAN\Desktop\Gating\Breath (unless instructed otherwise by service or hospital personnel). 5. Click Chart Tab. Chart Tab 1. Set Recording Length to no higher than 9 minutes. Note Values greater than 9 minutes not supported. 2. Select Auto scale. 3. Click Tracking Options Tab. 4
5 Tracking Options Tab Note Options to be determined at installation by the field engineer (refer to the Varian manual for more information). 1. Once the System Configuration settings are confirmed click OK. Patient List 1. On the RPM System Click File. 2. Select Patient, patient list will open. 3. Click New Patient. 5
6 4. Complete the Patient Data Form, to match the CT scanner patient data form. 5. Click OK when complete. 6. Highlight the Patient Name. 7. Click Select. 8. In the New Treatment Field window, fill in the Treatment Field Label. 9. Select Appropriate Treatment Field. 10. Click OK. 6
7 Note A message will display: Do you want to create a reference session for selected option? 11. Click Yes. 12. RPM screen will display with Track Enabled. Learning Phase on the RPM System During the Learning Phase the RPM system will track the patient s respiratory motion and attempt to learn the patient s respiratory motion. Once the system has a good signal and learned the respiratory motion, the blue indicator bar at top left of the screen should display one or two bars. Acquiring Surview on the CT scanner On the CT scanner, acquire the Surview and any other preliminary scans desired. Note Some customers prefer a routine chest scan before the retrospective pulmonary acquisition. These must be completed before pressing the record button on the Varian RPM. 7
8 1. Confirm scan parameters. 2. Do not use Auto Voice. 3. Click Go to start the Surview. 4. On the CT set Scan Parameters. 5. On the Varian RPM computer Press Record. 6. On the RPM computer Click Enable Gating Button. Note Pulses will start to display on the RPM workstation after the enable gating is activated. 7. Use the table below to determine the pitch and rotation speed. Breath Rate (in breaths per minute) For 0.5 sec Rotation Time, use a pitch no higher than: For 0.44 sec Rotation Time, use a pitch no higher than: Once the parameters are set on the CT scanner click Go to start the scan. 8
9 9. Press Enable. 10. When patient s breathing is steady, press Manual to start the scan. Note The waveform displayed on the Varian system will turn green when x-rays are on and the Radiation On symbol will appear. Note Monitor the patient s breathing during the scanning process. You may choose to stop the scan if breathing is irregular through the area of interest. Note On the CT scanner you will be prompted to import the Varian File when the scan is completed. 11. Do Not click OK keep window open and proceed to the next step. Note The file has not been saved for export, it is still recording. 12. On the RPM Computer click the Stop button to save the file. 9
10 13. On the RPM Computer Type in 1 in the Exam Number field. 14. On the RPM Computer Type in 1 in the Series Number field. 15. On the RPM Computer click OK when complete. 16. On the RPM Computer click OK to New Export Step Verification. Note On the RPM Computer Respiration data file was exported successfully will display on the screen. 17. On the RPM Computer click OK. [again?] 18. On the CT scanner click OK on the Import Varian File window to complete the file import. 10
11 19. On the CT scanner review the wave form that was imported, and edit tags if desired. Note Tags represent where the reconstruction will define the beginning of the breath cycle (0%). Please be aware that there is no opportunity to edit tags sent from RPM TM during gated treatment. 20. On the CT scanner click OK to start the image reconstruction. Note May take up to 90 seconds for the first image 21. On the Varian RPM workstation click Close Patient. Philips Medical Systems is part of Royal Philips Electronics Philips Medical Systems Cleveland, Ohio Philips Medical Systems All rights reserved. (2007) Reproduction in whole or in part Is prohibited without prior consent of the copyright holder. Philips Medical Systems Printed in U.S.A. 11
Brilliance CT Big Bore.
1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.
More informationHelical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice Guido Hilgers, a Tonnis Nuver, André Minken Department
More informationLaser Systems for Patient Alignment
Laser Systems for Patient Alignment GALAXY Patient TOPOGRAPHY Laser System Page: 2 Date: 28 March 2009 The GALAXY System NEW METHOD FOR HIGHER ALIGNMENT ACCURACY. Patient Positioning System for Linac Simulator
More informationArtefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT
Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT Marcus Brehm 1,2, Thorsten Heußer 1, Pascal Paysan 3, Markus Oehlhafen 3, and Marc Kachelrieß 1,2 1 German Cancer Research Center
More informationVirtual Phantoms for IGRT QA
TM Virtual Phantoms for IGRT QA Why ImSimQA? ImSimQA was developed to overcome the limitations of physical phantoms for testing modern medical imaging and radiation therapy software systems, when there
More informationHow To Title: Scan and File into ICS Purpose: To scan and file information into the patient s medical record Audience: All Staff Date: June 21, 2013
There are two ways to import medical records into ICS: A. Scan paper document(s) utilizing a scanner B. Import documents from a file, folder or Right Fax Acquiring images through a scanner: 1. Check that
More informationImage Guidance and Beam Level Imaging in Digital Linacs
Image Guidance and Beam Level Imaging in Digital Linacs Ruijiang Li, Ph.D. Department of Radiation Oncology Stanford University School of Medicine 2014 AAPM Therapy Educational Course Disclosure Research
More informationAAPM Standard of Practice: CT Protocol Review Physicist
AAPM Standard of Practice: CT Protocol Review Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center September 11, 2014 2014 Texas Radiation Regulatory Conference Goals Understand purpose
More informationAn investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 An investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition Yildirim D. Mutaf
More information41126 Cognento (MODENA) Italy Via Bottego 33/A Tel: +39-(0) Internet: Fax: +39-(0)
QUICK ANALYZER User Guide Version 5.3 Index 1.0 Generality... 2 LICENSE AGREEMENT... 3 2.0 Channels Configuration... 4 2.1 IdroScan Data Log Management... 6 3.0 Test Results... 9 4.0 Excel Export... 10
More informationicatvision Quick Reference
icatvision Quick Reference Navigating the i-cat Interface This guide shows how to: View reconstructed images Use main features and tools to optimize an image. REMINDER Images are displayed as if you are
More informationOptical Guidance. Sanford L. Meeks. July 22, 2010
Optical Guidance Sanford L. Meeks July 22, 2010 Optical Tracking Optical tracking is a means of determining in real-time the position of a patient relative to the treatment unit. Markerbased systems track
More informationMotion artifact detection in four-dimensional computed tomography images
Motion artifact detection in four-dimensional computed tomography images G Bouilhol 1,, M Ayadi, R Pinho, S Rit 1, and D Sarrut 1, 1 University of Lyon, CREATIS; CNRS UMR 5; Inserm U144; INSA-Lyon; University
More information3T/7T MRI FACILITY. 3T MRI System Start-Up and Shutdown
3T/7T MRI FACILITY SOP Number: 300.02 Title 3T MRI System Start-Up and Shutdown Revision Chronology Version Number Date Changes 300.01 01 March 2010 New 300.02 21 January 2013 Updated procedures Facility
More informationXRADIA microxct Manual
XRADIA microxct Manual Multiscale CT Lab Table of Contents 1. Introduction and Basics 1.1 Instrument Parts 1.2 Powering up the system 1.3 Preparing your sample 2. TXM Controller 2.1 Starting up 2.2 Finding
More informationFile Management Utility User Guide
File Management Utility User Guide Legal Notes Unauthorized reproduction of all or part of this guide is prohibited. The information in this guide is subject to change without notice. We cannot be held
More informationDeviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies
g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,
More informationDDX-R Quick Start Guide
DDX-R Quick Start Guide Imaging Support 1-855-726-9995 www.scilvet.com Contents USING DDX-R... 2 Opening a Patient File... 2 Opening a New Patient File... 3 Screen Layout... 3 ACQUIRING IMAGES ON A PATIENT...
More information7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT
CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance
More informationSecureGuard VMS. SecureGuard All-In-One Quick Install Guide V. 2.5
SecureGuard VMS Quick Install Guide V. 2.5 9/30/2018 2 Table of Contents Purpose... 3 Installation... 3 Setup... 4 Startup & Live View... 26 3 Purpose This is a step-by-step guide to installing the current
More informationMoving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region
Moving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region Andreas Hahn 1,2, Sebastian Sauppe 1,2, Michael Knaup 1, and Marc Kachelrieß 1,2 1 German Cancer Research Center (DKFZ),
More informationCT Protocol Review: Practical Tips for the Imaging Physicist Physicist
CT Protocol Review: Practical Tips for the Imaging Physicist Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center August 8, 2013 AAPM Annual Meeting Goals Understand purpose and importance
More informationCT Basics Principles of Spiral CT Dose. Always Thinking Ahead.
1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central
More informationJAZZ HARMONY User Manual
JAZZ HARMONY User Manual Copyright 2017 Imaging. All rights reserved. This manual and the software described herein are protected by copyright laws and international copyright treaties, as well as other
More informationSabre for Windows v4.5g INSTALLATION GUIDE
Sabre for Windows v4.5g INSTALLATION GUIDE Table of Contents Introduction... 3 Getting Help... 3 Upgrading the Software... 4 Standard Installation No Sabre Software... 7 Standard Installation Sabre Software
More informationPhilips SPECT/CT Systems
Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET
More informationThoracic target volume delineation using various maximum-intensity projection computed tomography image sets for stereotactic body radiation therapy
Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 8-2010 Thoracic target volume delineation using various maximum-intensity
More informationVITALSTIM PLUS Software Tutorial
VITALSTIM PLUS Software Tutorial 1 Index Opening software 3-5 Connect VITALSTIM PLUS device 6-10 Connect mouse EMG 11-12 Create a new patient 13-17 Open display session 18-36 Template training 37-50 Work/Rest
More informationImage Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD
Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image
More informationGuidelines of Ethernet for Using a Star Dvice. - Windows - Rev. 2.0
Guidelines of Ethernet for Using a Star Dvice - Windows - Rev. 2.0 Guidelines for Using a Star Ethernet Device on Windows The star dvice s IP address must be set in advance to use a star device that supports
More informationPURE. ViSION Edition PET/CT. Patient Comfort Put First.
PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion
More informationInitial Clinical Experience with 3D Surface Image Guidance
Initial Clinical Experience with 3D Surface Image Guidance Amanda Havnen-Smith, Ph.D. Minneapolis Radiation Oncology Ridges Radiation Therapy Center Burnsville, MN April 20 th, 2012 Non-funded research
More informationSICAT SUITE VERSION 1.0. Instructions for use SIDEXIS XG English
SICAT SUITE VERSION 1.0 Instructions for use SIDEXIS XG English Table of contents Table of contents 1 System requirements... 3 2 Overview of SICAT Suite... 5 3 Differences between full featured mode and
More informationSoarian Quick Start Guide for Office Staff
To Log in: 1. Go to the website: https://access.wakemed.org 2. Log in with your username and password. 3. Click on the link EPR Dashboard / Citrix Applications / Radiology PACS/ MUSE 4. Click on the EPR
More informationURGENT IMPORTANT FIELD SAFETY NOTIFICATION
Subject: Incorrect Movement of the Treatment Table Product: MOSAIQ Scope: Sites affected will be those: 1. Running MOSAIQ and, 2. Treating on linear accelerators with the RATM license Notification Released:
More informationAnnexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER
Annexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER A) Scanner Design X-Ray generator and tube: 1. Scanner: Whole body spiral CT scanner (16 slices) of latest technology.
More informationAnalysis of Acquisition Parameters That Caused Artifacts in Four-dimensional (4D) CT Images of Targets Undergoing Regular Motion
Original Article PROGRESS in MEDICAL PHYSICS Vol. 24, No. 4, December, 2013 http://dx.doi.org/10.14316/pmp.2013.24.4.243 Analysis of Acquisition Parameters That Caused Artifacts in Four-dimensional (4D)
More informationepact2 User Guide Accessing and Using Dashboards, Pre-built Reports & Quick Search Accessing the Dashboards... 1 Navigating the Dashboards...
epact2 User Guide Accessing and Using Dashboards, Pre-built Reports & Quick Search Contents Accessing the Dashboards... 1 Navigating the Dashboards... 3 Selecting a dashboard page... 3 Dashboard Criteria
More informationChapter 1 General Information
Chapter 1 General Information System Description Chapter 1 General Information 1.0 System Description The Affirm attaches to the Selenia Dimensions. A biopsy device attaches to the Affirm. X- and Y-axes
More informationUsing New Features in CDR DICOM 3.5 Software
Using New Features in CDR DICOM 3.5 Software Schick Technologies, Inc. 30-00 47 th Avenue Long Island City, NY 11101 (718) 937-5765 (718) 937-5962 (fax) PART NUMBER B1051053 REV. Copyright 2004 by Schick
More informationBrown County Avatar User Guide V1.0 Chapter 10 Order Entry
1. CLIENT PROFILE/PHYSICIAN ORDERS Menu Path: Avatar CWS > Order Entry > Order Entry Client Information> Client Profile/Physician Orders Note that in this section of the user guide, client ALEXIS GASTER
More informationUSING SOUNDMINER V4PRO - V4.5
F V T E C H D O C USING SOUNDMINER V4PRO V4.5 So you re looking for some sound effects for your movie... Here is an introduction to using Soundminer, which will help you navigate our sound effects library
More information1. Login as an administrator
ProQuest PartsManagerPro 4.x Installations There are 4 sections of installations for this product: I. Program Installation at TSE Server/Main Parts Machine II. Data Installation at TSE Server/Main Parts
More informationApplication Notes for Avaya IP Office Applications and CA Business Protection Suite Issue 1.0
Avaya Solution & Interoperability Test Lab Application Notes for Avaya IP Office Applications and CA Business Protection Suite Issue 1.0 Abstract These Application Notes describe the procedure for configuring
More informationSICAT SUITE VERSION 1.0. Instructions for use English
SICAT SUITE VERSION 1.0 Instructions for use English Table of contents Table of contents 1 System requirements... 4 2 Overview of... 6 3 Differences between full featured mode and viewer mode... 8 4 Safety
More informationSICAT SUITE VERSION 1.1. Instructions for use English
SICAT SUITE VERSION 1.1 Instructions for use English TABLE OF CONTENTS TABLE OF CONTENTS 1 Used icons and highlighting... 4 2 Safety information... 5 2.1 Identification of danger levels... 6 2.2 Qualification
More informationRaySafe Xi View. User Manual
RaySafe Xi View User Manual 2013.06 Unfors RaySafe, 5000151-C All rights are reserved. Reproduction or transmission in whole or in part, in any form or by any means, electronic, mechanical or otherwise,
More informationCarestream Vita user quick guide. Software version 3.2 From April 2012
Carestream Vita user quick guide Software version 3.2 From April 2012 1 Carestream Vita user quick guide Software version 3.2 from April 2012 1. To switch your Vita on Press the power button on the PC
More informationPatterson Imaging Multi-User Installation Instructions
Patterson Imaging Multi-User Installation Instructions Install the Patterson Bridge on the Server machine only 1. Please review all installation instructions before beginning the installation. You should
More informationCTA HEAD Perfusion AqONE without and with IV Contrast
CTA HEAD Perfusion AqONE without and with IV Contrast Patient Position Adult Contrast Adult Injection Rate Supine IOML perpendicular to table top. IV: 100 ml with helical head CTA 50 ml without helical
More informationMobile Workforce. Quick Start Guide
Mobile Workforce worksmart! LT Quick Start Guide Copyright Notice Mobile Workforce, Inc. has copyright ownership of the Mobile Workforce-workSmart! applications and other software related materials. This
More informationImage-In Direct Capture user guide
Image-In Direct Capture user guide For Omtool Healthcare Suite October 7, 2010 Omtool, Ltd. 6 Riverside Drive Andover, MA 01810 Phone: +1/1 978 327 5700 Toll-free in the US: +1/1 800 886 7845 Fax: +1/1
More informationRev: August 29, CIRS Motion Control User Manual
CIRS Motion Control User Manual 1 Table of Contents Introduction... 3 Installation... 3 Note... 3 Interface... 4 Using the Motion Library... 7 Running a Motion... 9 Adjusting Motion Parameters... 10 Motion
More informationDevelopment and Optimization of Fourdimensional Magnetic Resonance Imaging (4D-MRI) for Radiation Therapy
Development and Optimization of Fourdimensional Magnetic Resonance Imaging (4D-MRI) for Radiation Therapy by Yilin Liu Medical Physics Graduate Program Duke University Date: Approved: Jing Cai, co-advisor
More informationUvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication
UvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication Citation for published version (APA): Kruis, M. F. (2014). Motion compensation for 4D PET/CT General
More informationInstalling Cash Register Express
One Blue Hill Plaza, 16th Floor, PO Box 1546 Pearl River, NY 10965 1-800-PC-AMERICA, 1-800-722-6374 (Voice) 845-920-0800 (Fax) 845-920-0880 Installing Cash Register Express Below are instructions on installing
More informationCopyright 2012 Pulse Systems, Inc. Page 1 of 29
Use the CCD Control to receive and distribute a patient's "Continuity of Care Document" which contains the recorded medical history from a particular facility. Click anywhere to continue Copyright 2012
More informationManual. User Reference Guide. Analysis Application (EMG) Electromyography Analysis
Phone: (888) 765-9735 WWW.MINDWARETECH.COM User Reference Guide Manual Analysis Application Electromyography Analysis (EMG) Copyright 2014 by MindWare Technologies LTD. All Rights Reserved. 1 Phone: (614)
More informationIntelliVue Information Center
IntelliVue Information Center Research Data Export Viewer User Guide Part Number 4535 642 63711 Printed in the U.S.A. June 2011 First Edition *453564263711* Notice Notice Proprietary Information Published
More informationAN EVALUATION OF IMSIMQA AS A TOOL FOR COMMISSIONING 4DCT
AN EVALUATION OF IMSIMQA AS A TOOL FOR COMMISSIONING 4DCT e-poster: 12474 Congress: ESTRO 2011 Type: eposter Topic: 11th Biennial / QA: of software Authors: N. Whilde; Any information contained in this
More informationDRCR.NET Image Acquisition Protocol. Optical Coherence Tomography Using: Zeiss Cirrus OCT
DRCR.NET Image Acquisition Protocol Optical Coherence Tomography Using: Zeiss Cirrus OCT 1. PROCEDURE SUMMARY FOR EXPERIENCED USERS... 2 1.1. ENSURE THAT YOUR ZEISS CIRRUS SYSTEM MEETS SYSTEM REQUIREMENTS...
More informationPACS ADMIN. Quick Reference Guide
Quick Reference Guide PACS ADMIN Creating Domain Users.......................................................... Adding NOVARAD Users......................................................... Adding Console
More informationOUTPUTS AND INTERFACES CARDIOSCAN HOLTER ECG. VERSION 76a
OUTPUTS AND INTERFACES CARDIOSCAN HOLTER ECG VERSION 76a INTRODUCTION: CardioScan (10, 11, and 12) has evolved into a significant upgrade with the introduction of Version 76a. In addition to many new features
More informationAutomatic Intrinsic Cardiac and Respiratory Gating from Cone-Beam CT Scans of the Thorax Region
Automatic Intrinsic Cardiac and Respiratory Gating from Cone-Beam CT Scans of the Thorax Region Andreas Hahn 1, Sebastian Sauppe 1, Michael Lell 2, and Marc Kachelrieß 1 1 German Cancer Research Center
More information2015 Optima Healthcare Solutions Page 2
Viewing Documents Logging into Rehab Optima 1. To log onto Rehab Optima, go to www.rehaboptima.com. To place an icon on your desktop for future use, right click in the light blue box and choose create
More informationIntelliVue Information Center
IntelliVue Information Center Demo Application User Guide Part Number 4535 642 63741 Printed in the U.S.A. June 2011 First Edition *453564263741* Notice Notice Proprietary Information Published by Philips
More informationReal World Experience: Developing Dose and Protocol Monitoring from Scratch
Real World Experience: Developing Dose and Protocol Monitoring from Scratch Ingrid Reiser, PhD DABR Department of Radiology The University of Chicago Outline CT protocol monitoring Let s build a protocol
More informationSecure Access Authentication / Restriction of workflows based on User
Secure Access Authentication / Restriction of workflows based on User SMARTdocument Travel v5.0 with EIP Connect 2.0 Prepared by Anthony Vigliotti rev 1.0 January 2010 Table of Contents Enable Standard
More informationLower Dose with Confidence
Lower Dose with Confidence The Next Generation SCENARIA CT is Here. 64 or 128 Slice Capability The SCENARIA SE delivers lower dose technologies*, enhanced patient access, fast workflow speed and added
More informationRoutine excellence. Philips Ingenuity Core specifications
Routine excellence Philips Ingenuity Core specifications Table of contents 1 Introduction 3 2 User interface 4 2.1 ipatient key benefits 4 2.2 ExamCards 4 2.3 ScanRuler 4 3 DoseWise 5 3.1 DoseRight Index
More informationCardiac Dual Energy CT: Technique
RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de
More informationHitachi Medical Systems America, Inc. Hitachi Medical Corporation
Advanced 128 Advanced 128 128 Slice Capability Enlarged 75cm Aperture Thin Gantry Profile 88cm Intelli IP Advanced, Iterative Reconstruction IntelliCenter Auto-Lateral Shifting Table, up to 16cm Range
More informationMobile POS Tablet. Barcode Scanner DATA CAPTURE MODULE OPERATION GUIDE. Peripheral Scanning Tool. To Connect Barcode Scanner Module
Barcode Scanner Follow below steps to check the scanner module with. 1. Click Start All Programs Utilities Peripheral Scanning Tool. 2. Select POS Tab. 3. Check the Scan Scanner from checkbox.. Press Auto
More informationVivoSense. User Manual - LifeShirt Import Module. Vivonoetics, Inc. San Diego, CA, USA Tel. (858) , Fax. (248)
VivoSense User Manual - VivoSense LifeShirt Import Version 3.0 Vivonoetics, Inc. San Diego, CA, USA Tel. (858) 876-8486, Fax. (248) 692-0980 Email: info@vivonoetics.com; Web: www.vivonoetics.com Cautions
More informationMobile POS Tablet Barcode Scanner
DT398C/ DT398B OPERATION GUIDE Mobile POS Tablet Barcode Scanner Installation The Barcode Scanner is preinstalled as an option for the DT398C/ DT398B. Barcode Scanner Trigger Button Peripheral Scanning
More informationEagleSoft Set-up Instructions for Denoptix Version: Date Created: 6/10/05 Last Modified: 10/24/05 Filename: Denoptix ES Setup Instructions
EagleSoft Set-up Instructions for Denoptix Version: 12.00 Date Created: 6/10/05 Last Modified: 10/24/05 Filename: Denoptix ES Setup Instructions Required materials: EagleSoft Digital Integration CD Denoptix
More informationTutorial: set up of measurement recipe with database
Tutorial: set up of measurement recipe with database This tutorial explains 1. how to create a recipe to perform repeated measurements at a given place on the wafer 2. how to make a data analysis routine
More informationThe Near Future in Cardiac CT Image Reconstruction
SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de
More informationHESKA DCU (DATA CAPTURE UTILITY) VERSION 2010 AND HIGHER
HESKA DCU (DATA CAPTURE UTILITY) VERSION 2010 AND HIGHER 855-478-7920 www.avimark.net support@avimark.net POWERING SUCCESSFUL PRACTICES TM VETERINARY SOLUTIONS Table of Contents Heska DCU...3 Installation...3
More informationGetting Started with Crazy Talk 6
Getting Started with Crazy Talk 6 Crazy Talk 6 is an application that generates talking characters from an image or photo, as well as facial animation for video. Importing an Image Launch Crazy Talk and
More informationAbstract. Apteryx Imaging (XrayVision, XVLite ) Installation Guide. Abstract Abstract
Apteryx Inc. 313 S. High St. Suite 200 Akron, OH 44308 330-376-0889 voice 330-376-0788 fax sales@apteryx.com www.apteryx.com Apteryx Imaging (XrayVision, XVLite ) Installation Guide Abstract Abstract Abstract
More informationFOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION
Philips J. Res. 51 (1998) 197-201 FOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION This special issue of Philips Journalof Research includes a number of papers presented at a Philips
More informationAbstract. Apteryx Imaging (XrayVision, XVLite ) Installation Guide. Abstract Abstract
Apteryx Inc. 313 S. High St. Suite 200 Akron, OH 44308 330-376-0889 voice 330-376-0788 fax sales@apteryx.com www.apteryx.com Apteryx Imaging (XrayVision, XVLite ) Installation Guide Abstract Abstract Abstract
More informationD1.5a Create Access Master Files
D1.5a Create Access Master Files Step 1. Open the file in Audacity a) Double- click on the Audacity program icon. Wait for Audacity to fully load. Click ok to dismiss Welcome to Audacity screen if it pops
More informationNVMS User Manual. Version 2.1.0
NVMS-1000 User Manual Version 2.1.0 Contents 1 Software Introduction... 1 1.1 Summary... 1 1.2 Operation Environment... 1 1.3 Install and Uninstall... 2 1.3.1 Install the Software... 2 1.3.2 Uninstall
More informationUnder SOAP Notes icon/ Progress Notes icon
Under SOAP icon/ Progress icon SOAP notes Daily Progress notes On the function bar - Blue File Folder icon Show all progress notes will show you a history of any SOAP notes you have already written On
More informationFour-Dimensional MVCT Imaging and Spatial Translation of Helical Tomotherapy Dose Distributions Using Sinogram Modification
University of Tennessee, Knoxville Trace: Tennessee Research and Creative Exchange Masters Theses Graduate School 12-2006 Four-Dimensional MVCT Imaging and Spatial Translation of Helical Tomotherapy Dose
More informationState-of-the-Art IGRT
in partnership with State-of-the-Art IGRT Exploring the Potential of High-Precision Dose Delivery and Real-Time Knowledge of the Target Volume Location Antje-Christin Knopf IOP Medical Physics Group Scientific
More informationZmachine Data Viewer Software Quick Start
Zmachine Data Viewer Software Quick Start Reviewing Zmachine Insight/Insight+ Data Copyright 2016 by General Sleep Corporation All rights reserved About this Document You are advised to read and understand
More informationAn automated method for comparing motion artifacts in cine four-dimensional computed tomography images
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 An automated method for comparing motion artifacts in cine four-dimensional computed tomography images Guoqiang Cui, 1 Brian Jew,
More information8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition
Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare
More informationHardware: Acquiring Data and Communicating with Instruments
Hardware: Acquiring Data and Communicating with Instruments 4 Acquiring a Signal This chapter introduces you to the Express VIs you use to acquire data and communicate with instruments on Windows. These
More informationUsing Scantron ParLAN 6.5 for the First Time:
Page 1 of 19 Using Scantron ParLAN 6.5 for the First Time: Last updated: Monday, December 02, 2013 Nine Basic Steps To request a Scantron Account, please contact the Academic Technology Center. Step One:
More informationQuestionmark Assessments Taking an Assessment
1 IMPORTANT! Before taking your readiness assessment, you must be prepared to print your Certificate of Completion either to your printer or as a PDF. We recommend that you print a PDF of the certificate
More informationTumor motion during liver SBRT
Tumor motion during liver SBRT - projects at Aarhus University Hospital - Per Poulsen, Esben Worm, Walther Fledelius, Morten Høyer Aarhus University Hospital, Denmark SBRT: Stereotactic Body Radiation
More informationATTACHING MULTIPLE IMAGES TO A DOCUMENT POWERING SUCCESSFUL PRACTICES TM VETERINARY SOLUTIONS
ATTACHING MULTIPLE IMAGES TO A DOCUMENT 855-478-7920 www.avimark.net support@avimark.net POWERING SUCCESSFUL PRACTICES TM VETERINARY SOLUTIONS Table of Contents Attaching Multiple Images to a Document...3
More information7/31/ D Cone-Beam CT: Developments and Applications. Disclosure. Outline. I have received research funding from NIH and Varian Medical System.
4D Cone-Beam CT: Developments and Applications Lei Ren, PhD, DABR Department of Radiation Oncology Duke University Medical Center Disclosure I have received research funding from NIH and Varian Medical
More informationABAXIS VETSCAN CHEMISTRY VERSION AND HIGHER
ABAXIS VETSCAN CHEMISTRY VERSION 2013.1 AND HIGHER 855-478-7920 www.avimark.net support@avimark.net POWERING SUCCESSFUL PRACTICES TM VETERINARY SOLUTIONS Table of Contents Interfacing AVImark with the
More informationPVDOMICS. Chest CT Scanning and Data Transmission. Jason Lempel, MD Lung Imaging Core Data Coordinating Center Cleveland Clinic Nov.
PVDOMICS Chest CT Scanning and Data Transmission Jason Lempel, MD Lung Imaging Core Data Coordinating Center Cleveland Clinic Nov. 7, 2016 CT Imaging Options Routine contrast enhanced CT CT pulmonary angiogram
More informationWeb Forms: Erasmus+ & ESC How to complete the form
Web Forms: Erasmus+ & ESC How to complete the form When completing the form, there are certain elements that are common in every section. This page covers these basic functionalities and highlights the
More information