FlowAccess TM GETTING STARTED. Windows software for Skalar SAN plus systems. Version and up

Size: px
Start display at page:

Download "FlowAccess TM GETTING STARTED. Windows software for Skalar SAN plus systems. Version and up"

Transcription

1 FlowAccess TM V3 Windows software for Skalar SAN plus systems Version and up GETTING STARTED

2

3 FlowAccess TM V3 Windows software for Skalar SAN plus systems Version and up Getting started Publication No: B.US Issue date: May 2012 Copyright Skalar Analytical B.V. P.O Box DE Breda The Netherlands Tel: +31 (0) Fax: +31 (0) info@skalar.com Internet:

4

5 Welcome to the FlowAccess V3 TM Windows Software This Getting started manual is meant as a quick start-up tutorial with a description of a simple system set-up, the set-up of a sample table and a group for analysis, running the set-up analysis and the data processing of the run. A few aspects like communication settings, peak editing and table editing are described. This tutorial does not describe all features of the software. These are described in the User Manual. Skalar Analytical B.V. P.O. Box DE BREDA The Netherlands Tel: Fax: info@skalar.com Internet:

6

7 TABLE OF CONTENTS 1 INTRODUCTION TO THE FlowAccess V3 TM SOFTWARE New Software System requirements PC requirements Installation instructions Communication settings BThe FlowAccess V3 TM Main Window Login SYSTEM SETUP Defining the hardware of the system Connecting the hardware SETUP FOR ANALYSIS Introduction Method definition Group definition Setup of a sample table Hardware status settings ANALYSIS Starting an analysis Starting an analysis with time programming Real time viewing Viewing incoming results DATA PROCESSING Introduction PostAnalysis screen Peak editing Change of the table format Printing APPENDIX

8

9 FlowAccess V3 TM GETTING STARTED CHAPTER 1: INTRODUCTION Preface: This chapter includes general information, software installation and requirements and security options.

10

11 Chapter 1 Introduction to the FlowAccess V3 TM software INTRODUCTION TO THE FLOWACCESS V3 TM SOFTWARE This document is a step-by-step tutorial, which describes the basics how to create a simple system and how to work with it. The first step (Chapter 2) describes how to define a system from scratch. The second step (Chapter 3) describes how to define the methods and how to prepare an analysis table. The third step (Chapter 4) describes how to start and monitor the analysis. The fourth step (Chapter 5) describes the post-analysis data processing. 1.1 New Software The Skalar software package FlowAccess V3 TM will be preinstalled on the computer supplied by Skalar. Important note: If other programs then FlowAccess V3 TM are being installed Skalar does not guarantee the proper functioning of FlowAccess V3 TM If other computers then computers supplied by Skalar are used, Skalar will not guarantee the proper functioning of FlowAccess V3 TM System requirements If you install FlowAccess V3 TM as an update for an existing system, always first contact Skalar for investigation of your system, as update on your system may be required. Updates may be required for samplers (EPROM/motors), rinsing valves (220 volt, plugs), interfaces (EEPROM) and detectors (plugs) PC requirements Processor >2000 MHz RAM memory 4 GB Video memory 128 MB Hard disk drive at least 250 GB Available space 100 MB at least USB port 2.0 or higher COM port 1 x RS232 (9 pins) Parallel port 1 x (or printing on network) CD ROM drive 1 x (32 speed) Display monitor 19" TFT flat screen set 96 dpi Mouse, keyboard and printer Operating system MS Windows TM 7 Professional or Ultimate (recommended) 32/64 bit Note: OS must be installed completely, including screen resolution, fonts and original video drivers.

12 FlowAccess V3 Getting started Installation instructions Notes: 1. It is strongly recommended to exit all applications before starting the installation. 2. Be sure to be logged in with all (administrator) rights to the computer before installing the software and the user account settings in Action Center must be set at Never notify. 3. Windows Installer version 3.1 (V2) has to be present on the PC before installing the FlowAccess V3 TM software. This Windows Installer can be downloaded from the website link as follows, -5F56-4A38-B838-DE776FD4138C&displaylang=en 4. Read carefully the Last minute instructions and the FlowAccessV3 instructions 3.1.xx on the DVD before installation. The latter file contains information about the code to be inserted during installation. Installation: 1. Start Microsoft Windows. 2. Insert the DVD-ROM in the CD drive. After a few seconds the Autoplay screen is displayed. 3. Click Open folder to view files. Windows Explorer is opened with the DVD drive selected. 4. Double click on Setup.exe in the right part of the screen. The Install Wizard is opened and starts to prepare the installation.. 5. Follow the instructions on the screen for the installation of the software. Note: Make sure to select a convenient directory for the data. 6. When the software is installed the FlowAccess V3 TM program is displayed in the Programs group on the start bar and the FlowAccessV3 instructions manual is placed on the desktop. 7. Open the program by clicking the left mouse button. The program window as shown in Figure 1.4 will be displayed. Note: For the installation of an update refer to the User Manual. 1.2 Communication settings The communication settings for proper communication between the software program and the analyzer are set in the Communication Module. This program is present in the directory where FlowAccess V3 TM is installed. A. Log settings Activate the registration of the communication in a log file. This registration is important for locating errors in the communication. Each time the program is started a file called Communication Module.log starts to register every step in the communication between the system and the program. If any errors have occurred in the communication, this file can be viewed in Wordpad or Notepad or can be sent to Skalar for investigation. To activate the logging: 1. Open Windows Explorer and move to the directory where FlowAccess V3 TM is installed.

13 Chapter 1 Introduction to the FlowAccess V3 TM software Double click on the file Communication Module.exe. The program is opened (see Figure 1.1). Figure Select Logging from the File menu or click on the Logging icon (1 st left) on the toolbar. The Information Logging dialog box is opened (see Figure 1.2). Figure Select Single file logging for the Log mode. 5. Check the options Read Packets and Write Packets and click OK. The dialog box is closed. 6. Close the Communication Module program. The file is saved in the FlowAccess V3 TM directory when the program is ended, but is overwritten as soon as the program is started again. Important note: Always rename a log file that needs to be checked for errors in the communication before the FlowAccess V3 TM program is re-started. B. COM port settings By default the program assumes COM port 1 of the computer is used for the connection with the analyzer system. If you use another COM port for the connection, this has to be changed in the Communication module (Communication module.exe). 1. Open Windows Explorer 2. Move to the directory where FlowAccess V3 TM is installed. 3. Double click on the file Communication Module.exe. The program is opened. Select Properties from the File menu or click on the Properties icon (2 nd left) on the toolbar. The Communication Settings dialog box is opened (see Figure 1.3).

14 FlowAccess V3 Getting started 1-4 Figure Click the arrow for the port and select the proper COM port from the list. Check also if the option Connect to instrument is selected. 5. Click Apply and then OK to close the dialog box. 6. Close the Communication Module BThe FlowAccess V3 TM Main Window When the software is installed a FlowAccess V3 TM icon is displayed on the desktop. Open the program by double clicking the left mouse button. The following screen is displayed (Figure 1.4). Figure 1.4 Main window Login To be able to enter the program for the first time the master user name and password has to be entered. The master user name is Administrator and the password is admin. Warning: This password has to be changed into another (secret) password by the highest user (administrator) of the program to prevent unauthorized admission to the User Management window.

15 Chapter 1 Introduction to the FlowAccess V3 TM software Click on Change Password. The screen of Figure 1.5 is displayed. 2. Type the old password (admin) again. Figure 1.5 Change password 3. Then type a new password and confirm the new password. 4. Select Change. A message is displayed that the password is successfully updated. 5. Select OK. The Login screen (see Figure 1.4) is displayed again. 6. Type the new password and select Login. The screen is closed. Only when the Administrator has logged in (the user name Administrator cannot be removed) the options User Management (see User Manual) and Login Settings (see User Manual) are available from the Settings menu. Note: As soon as you are logged in a message is displayed whether you want to restore the Prefab system setup database or not. Refer to the User Manual to restore the Prefab system setup. In this Getting Started manual a new system setup is created. As soon as a choice is made to use the original database a new message pops up (see Figure 1.6). Figure 1.6 Click OK. The Preferences screen is displayed with the Path settings all set on C:\ (see Figure 1.7). Click the Browse icon and select for each item the required subdirectory of FlowAccessPost. Note: The location of FlowAccessPost was selected during installation.

16 FlowAccess V3 Getting started 1-6 Figure 1.7 Preferences Note: For the settings in the other tabs of the Preferences screen refer to the User Manual. Click OK to close the screen. The screen displays now the main window with a menu bar, a toolbar and a two-pane window. The left pane shows the main parts of the program: System, Data Processing, Reports (optional), Quality Control (optional) and General Settings. The System tab (see chapter 2) is active (and the options displayed in the upper part) when the program is opened. The toolbar displays the shortcut icons for the system. Click on Data Processing (see chapter 5), Reports (optional) Quality Control (optional) or General Settings to make that tab active. Browse through the tabs and change the settings as required for your application.

17 FlowAccess V3 TM GETTING STARTED CHAPTER 2: SYSTEM SETUP Preface: This chapter gives detailed instructions for a complete system setup.

18

19 Chapter 2 System Setup SYSTEM SETUP Before an analysis can be started the analyzer system has to be defined in all its aspects. In this Getting started manual a system setup for a system with system controller, consisting of a SA1050 sampler and 4 channels with digital detectors, will be explained. Note: For the setup of complicated systems and extended information about the setup, refer to the User Manual. 2.1 Defining the hardware of the system 1. Open the FlowAccess software and login as required.. 2. Select New System to create a new system. The Interface Settings screen (see Figure 2.1) is displayed (only when no system was yet defined) to set the hardware information of the system. 3. Select for each item the total number that is physically present on the system that is being setup. Figure 2.1 Interface Settings

20 FlowAccess V3 Getting started 2-2 Interface. For this system select interface type System Controllers. For this system select 1 system controller. Dual Detectors. For this system select 2 dual detectors. Density Meters. No density meter is selected for this system. Heaters. For this system select 2 heaters. External Valves. This option is for future use. Sampler Valves. This option is for future use. Sampler. For this system select 1 sampler. Single Detectors For this system select no single detectors. Rinsing Valves. For this system select 2 rinsing valves. External Rinsing Valves. This option is for future use. Channel Valves. This option is for future use. 4. Select Save and Continue. A message is displayed that the settings are successfully updated. 5. Select OK to close the screen. The Hardware Setup screen will be displayed (see Figure 2.2). Figure 2.2 Note: Make sure that the Add button is selected from the toolbar before the hardware is selected to be added. When any hardware has to be removed first select the Remove button and then click on the item to be removed. 6. Select the Sampler tab in the left pane. This tab has two sub-tabs: Samplers and Accessories. The Samplers sub-tab will be active. Then all available samplers are displayed. 7. Select the Add button (a + is added to the pointer) and click on the picture of the SA1050. The selected sampler is then displayed in the center pane. 8. Select the Accessories tab. The possible accessories (dilutor, valves and needles) are displayed.

21 Chapter 2 System Setup Click on the dilutor. The dilutor is added to the sampler in the center pane. 10. Click on the valve. The valve is added to the sampler. 11. Then select the Module tab in the left pane. The options are displayed (the option module valve is not yet available). 12. Click on the picture of the module (the module part in the center pane turns white) and click once on a row in the white area for each module that is physically present on the system. In this example 4 modules. 13. Click on the picture of the heater in the left pane and click in the module in the center pane, where a heater is used to add the heater to the module. In this example module 1 and Select the Detectors tab. This tab has two sub-tabs: Digital detectors and Analog detectors. The Digital detectors tab will be active. 15. Click on the picture of the dual digital detector and click in the white area behind the module 1. Automatically two detectors are placed behind module 1 and module 2. Click again behind module 3. Automatically two detectors are placed behind module 3 and module Select the Pumps & Valves tab. Click on the picture of the pump and then click on the white area in front of module Click on the picture of the rinsing valves and then click twice on the white area in front of the pump: before module 1 and module Now click on the pointer and then on the pump. Then click three times on the arrow up icon in the toolbar to extent the pump over all modules in the module holder. 19. Then click on the first rinsing valve to extent the valve over a second module. Do the same for the second rinsing valve. 20. Click twice with the pointer on module 1. Type a name for the module. Repeat this step for module 2, 3 and 4. The system should now look like Figure 2.3. Figure 2.3

22 FlowAccess V3 Getting started Connecting the hardware Now the hardware has to be connected. Note: Make sure to select the Connect tool each time a new part has to be connected. Always when the Connect button is pressed a + and tray is added to the pointer. As soon as then one of the parts to be connected is selected the + and tray is changed into an S until the connection part is selected. A Mismatch message is displayed when a connection is not possible. 1. Select the Sampler Connector tab in the right pane. 2. Select Connect in the toolbar. Click on the sampler in the center pane and then on the Sampler connector 01 in the right pane. The sampler and the connector get a pink lining to show that they are connected. 3. Select Connect. Click on the needle and then subsequently on module 1 to 4. The left front of the modules get a color lining with the same color as the needle to show the connection. 4. Select Connect. Click on the pump and then subsequently on module 1, 2, 3 and 4. The pump and the modules turn pink to show the connection. 5. Select the Detector connector tab. 6. Select Connect. Then click on module 1 and then on the detector behind the module. Repeat this step for module 2, 3 and 4. To show that they are connected the end of the module and the detector get a yellow lining. 7. Select Connect. Then click on the detector behind module 1 and then on dual connector 1,2 in the right pane. This means that also module 2 is automatically connected. Click on the detector behind module 3 and then on dual connector 3,4 in the right pane. This means also that module 4 is automatically connected. The connectors get a pink lining to show that they are connected. 8. Select the System controller tab. 9. Select Connect. Click on the pump in the center pane and then on the pump icon in the right pane. The icon turns green to show the connection. 10. Select Connect. Click on the first heater on the module and then on the Heater icon in the right pane. The icon turns green and the heater on the module gets a pink color. Repeat this step with the second heater. This heater also turns pink. Note: The digital detectors are already connected and are not separately connected to the system controller. This connection is for analogue detectors only. 11. Select the Rinsing Valve tab. 12. Select the Connect. Click on rinsing valve 1 and then on the connector 01. The rinsing valve turns pink and the connector gets a pink lining to show the connection. Click on rinsing valve 2 and then on connector 02. Also this valve turns pink and the connector gets a pink lining. 13. Select the Pointer button. Note: The Other Valves tab is for future use. The system should now look like Figure 2.4 (the names of the modules are just an example).

23 Chapter 2 System Setup 2-5 Figure 2.4 By means of the Properties tab all parts of the setup can be viewed with their status. Note: The status of the pump, heater, rinsing valves and detectors in the different modes (default, standby and online) can be set here as a standard for the system, but can be set also in the analysis window (see the User Manual). For this example we will set this in the analysis window. Click Save. Type the name of the group in the displayed dialog box and click Save again. The box is closed.. Note: If not all required connections were made a dialog box with warnings about the missing connections will be displayed. Make a not of the missing connections and close the box. Click Cancel in the dialog box and first connect the missing connections. Then save the system again. Click Exit to close the Hardware Setup screen. A confirmation is asked before the screen is closed. The Main screen will be displayed with the icon of the system just made. The system can now be opened for further setup in the Active System screen. Select the system in the Main screen and click on Open for Analysis.

24

25 FlowAccess V3 TM GETTING STARTED CHAPTER 3: SETUP FOR ANALYSIS Preface: This chapter gives detailed instructions for definition of the methods and groups and setup of a table.

26

27 Chapter 3 Setup for Analysis SETUP FOR ANALYSIS 3.1 Introduction The system we just defined will be displayed as shown in Figure 3.1. Figure 3.1 Active System window The system is always displayed in default mode when opened the first time (for more explanation about the modes refer to the User Manual). In this window the methods on the modules for the analysis are defined and the group of methods that will be analyzed. Here also the status for the detectors, rinsing valves, heaters and pump can be defined (this can also be done in the system setup). To be able to run an analysis, a sample table will be made. The run is started and the Real Time screen is opened from this screen. When the run is finished PostAnalysis can be started from the File menu. For this example we will define four methods and group these methods for an analysis. We will make a sample table and then start the analysis.

28 FlowAccess V3 Getting started Method definition There are several ways to open the method definition. The easiest way is to click on the toolbar icon Method and then on a channel in the displayed chemistry unit. The method definition can also be opened from the Method menu, but is then not attached to a channel. Four methods will be defined. 1. Select the Method tool and click on ch1 (Ammonia) in the chemistry unit. A dialog box is opened. 2. Select Create New Method and click OK. A dialog box is opened for the definition of the new method as shown in Figure 3.2 with the tab Peak Detection parameters displayed. Accept the default settings of this tab. Figure 3.2 Method Setup: Peak Detection parameters 3. Select the Peak Detection type tab and check if Normal is selected. 4. Select the Calibration tab. Figure 3.3 Method Setup: Calibration

29 Chapter 3 Setup for Analysis Check Calibrate, accept the default calibration type (1 st Order ISO ) or select the required calibration type. 6. Enter for S1 - S5 the required values for the measuring range; automatically the Select box for each standard will be checked. 7. Enter mg/l in the field for the concentration units. 8. The screen should now look as in Figure Select the Result tab. Check the Perform Wash correction and Perform Drift correction tick box if not checked (see Figure 3.4). Figure 3.4 Method Setup: Result 10. Select the Post Dilution tab. Check the tick box for Post Dilution and select Concentration for the Over range threshold. Figure Enter the threshold concentration and the post dilution factor.

30 FlowAccess V3 Getting started Group definition 12. Click the Save tool button, enter a name for the method in the displayed box and click OK. The name of the method is added to the channel in the left pane. 13. Click on channel 2 in the left pane. 14. Select the Calibration tab. 15. Check Calibrate and accept the default calibration type (1 st Order ISO) 16. Enter required the values for the standards. 17. Repeat step Click the Save tool button, enter a name for the method in the displayed box and click Save. 19. Perform step for channel 3 and 4. Enter for channel 3 the required values for S1-S5 and give the method a name. Enter for channel 4 the required values for S1-S5 and give the method a name. 20. The defined methods are now attached to the channels and the names of the methods are displayed in the left pane. 21. Click on the Exit button to close the Method Setup. Confirm the Method in the next dialog box by clicking Yes. The methods to be analyzed in one run must be grouped. 1. Select Group (a G icon is added to the pointer) and then click on the sampler. All channels turn white. 2. Click on all four channels. They will be colored green. 3. Click on the right mouse button in the chemistry unit and select Add groups from the pop-up menu. 4. Enter a name for the selected group (i.e. Demo ) and click OK. 5. The screen will now look like Figure 3.6. Figure Again click Group and then click with the right mouse button on the chemistry unit.

31 Chapter 3 Setup for Analysis Select Edit Group from the displayed menu. The Group Settings screen will be displayed (see Figure 3.7). Figure Select the Post Dilution settings tab. 9. Check Post Dilution and press interval settings to enter the settings for the batch. For detailed information refer to the User manual. 10. Enter the required information about resamples, cascade dilutions and whether the post dilution factor is auto-calculated for the whole group or set with a factor. This factor overrules the dilution factors in the individual methods. Check Use mixing and select the number of times to mix. 11. Click Save to close the screen. 3.4 Setup of a sample table To be able to run an analysis a sample table has to be defined. An example of a table for a complete run that includes a tracer, drifts, washes, 5 standards and 45 samples will be given. 1. Click Table. The Load Table screen is displayed. 2. Select Create New Table. The Select Trays screen is displayed. Only the trays that can be present in the sampler type of the displayed system are shown in this screen, as the table is automatically attached to the sampler in the displayed system. 3. Place a checkmark for the trays used on the sampler (see Figure 3.8). Only trays that compare to the configuration of the sampler are available. The default needle height (sampling height), pick-up height (for dilution), dispense height and end volume (after dilution) are displayed and can be changed if required.

32 FlowAccess V3 Getting started 3-6 Figure Click OK. The screen is closed and the Table Entry screen is opened (see Figure 3.9). Here the sample table, complete with standards, drifts etc. can be created. The easiest way to create a standard table is to use the Table Wizard. 5. Select the Table Wizard icon from the toolbar. Figure The Table wizard screen will be displayed with the Setup Tab selected. Here the tray selection is shown, but cannot be edited here. 7. Select Tracer or click Next to insert de settings for the tracer (see Figure 3.10).

33 Chapter 3 Setup for Analysis 3-7 Figure Enter the position of the Tracer in the sampler (i.e. ST5). 9. Select Drift/Wash or click Next to insert the settings for the Drift/Wash (see Figure 3.11). 10. Enter the position of the Drift in the sampler (i.e. ST4) and the interval of the drift/washes in the sample table (i.e. 10 positions). Figure Select Standards or click Next to insert the information about the standards (see Figure 3.12). Figure 3.12

34 FlowAccess V3 Getting started Enter the position of the first standard (ST1) and the total number of standards (5) and click Next or select Sample tab (see Figure 3.13). 13. Enter the number samples and the position of the first sample. Check the options Result selection and Post dilution. Figure Select Next three times and then Create. The Table Entry screen is displayed with the complete table (see Figure 3.14). If required the identity for the unknowns can be entered. 15. To insert a sample or standard: Select the row above which a sample or standard has to be inserted and click the right mouse button. Select Insert row. Type the position and select the type from the drop down menu. Then the information is automatically filled in. Figure Save the table and give the table a name. 17. Press Exit to return to the Active System screen. Note: If required, a pre-dilution factor and a pre-dilution position for the samples can be entered.

35 Chapter 3 Setup for Analysis Hardware status settings Now is the time to switch on the pump, heaters, rinsing valves and detectors to stabilize the system before the run is started. (When you are familiar with the system, you can switch on the hardware before creating your sample table, so that the system is stabilized and ready for analysis when the table is ready). 1. Double click on the traffic light ( Status ) tool. A message is displayed if you want to switch from Default status to Online status. Click Yes. The status changes to the Online status and the traffic light turns green (also in the right pane the status change is displayed). 2. Select Control. 3. Click on the rinsing valve for channel 1 and 2. A dialog box is displayed. Click on the pink Water button. It turns green while the text changes into Reagent. Click OK to accept the settings and to close the dialog box. Repeat this step for the second rinsing valve for channel 3 and 4. The numbers in the valves are green now. 4. Click on the pump. A dialog box is displayed. Click on the red Off button. It turns green and the text changes into On. Click OK to accept the settings and to close the dialog box. The number in the pump will be green now. 5. Click on a heater. A dialog box is displayed. Click on the red Off button. It turns green and the text changes into On. Click OK to accept the settings and to close the dialog box. (Both heaters will turn on as they are connected to the same power strip). The H of the heaters will be green now. 6. Click on the detector for channel 1 and 2.. A dialog box is displayed. Click on the red Off button. It turns green and the text changes into On. Click OK to accept the settings and to close the dialog box. Repeat this step for the detector for channel 3 and 4. The detectors will now show a little green box. 7. Now wait about 15 minutes until the system is stabilized. Then click Baseline and then check the tick box in front of the Demo group. Figure 3.15

36 FlowAccess V3 Getting started 3-10 Then all channels on the chemistry unit will be checked (see Figure 3.15). 8. Select Start. Now the baseline can be viewed in the RealTime screen. Open the RealTime screen to check if the baseline is stable. 9. If the baselines are stable, close the screen and return to the Active System window. 10. Select Stop and then uncheck the Demo group. A pop-up window is displayed with the question if you really want to stop a certain channel. 11. Click Yes to stop that channel. Each channel has to be stopped individually. 12. Now the system is ready for analysis. Note: If no system controller is present on the system you can leave the system online, when you close down the program. When you start the program the next time the system is immediately ready for analysis (after a stabilizing time).

37 FlowAccess V3 TM GETTING STARTED CHAPTER 4: ANALYSIS Preface: This chapter gives detailed instructions about starting and monitoring the analysis.

38

39 Chapter 4 Analysis ANALYSIS 4.1 Starting an analysis For a Quick start-up reference guide see Appendix Click on the (yellow) Start tool. A dialog box is displayed with the default name for the run file (run file format, as set in the Preferences screen). Note: If anything, necessary for the run, was not switched on, a message is displayed what is missing to start-up the analysis. (In this case you are asked if you want to proceed with the analysis while not all your rinsing valves or heaters are connected. Click Yes to continue). Accept the default name given to the file or type a new name. Click OK. The box is closed and the analysis will start. The yellow Start tool is grayed out and the Stop tool gets active (red). 4.2 Starting an analysis with time programming It is also possible to start an analysis with the aid of time programming. For more information about time programming and how to setup a time program, refer to the user manual. 4.3 Real time viewing During the run the on-going analysis can be displayed in Real-time. Select the Real-time tool on the toolbar. The Real time window is displayed by default with all 4 channels visible (see Figure 4.1). Figure 4.1

40 FlowAccess V3 Getting started 4-2 One channel view is active. To display one channel in the whole window double click in the title bar of a channel. The channel is then displayed with a menu and toolbar with all options for viewing and editing (see Figure 4.2). Figure 4.2 It is also possible to view the channels in a over layered view. Select Other views from the View menu. The Overlay display screen is displayed. Select the channels to be displayed. The view looks like Figure 4.3. Use the Exit button in the toolbar to return to the previous view. Figure 4.3

41 Chapter 4 Analysis Viewing incoming results During the run it is also possible to view the incoming results, as soon as the initial wash has been measured. Close or minimize the Real time window and return to the Analysis window. Select Results from the Analysis menu or click the Results tool. The Results screen is opened (see Figure 4.4) displaying all results. Figure 4.4 As soon as the drift and wash after the standards have been measured, the calibration curve is calculated and the results can be displayed by selecting the Calibration tab (see Figure 4.5). Figure 4.5

42 FlowAccess V3 Getting started 4-4 It is possible to view the Statistics of the run or to add a dummy channel for calculation, refer to the User manual for these options. When the run is finished, the Real-time screen and/or The Results screen is closed and the Active System window is displayed again. The group shows Stopped. Note: When the Postdilution option was selected, the system will automatically perform post dilutions on the over range samples and subsequently analyze these dilutions. From the time the last sample of the normal run is taken, until the performance of the post dilution, the sampler is off. However, the status of the system stays on Analysis and the Real time and/or Results screen stay open. The results of the post dilution analysis are automatically added to the results. The original sample table is adapted with the post dilution samples and their positions. Wait until the system shows Stopped. Close the Active System screen to return to the Main window. Here the Data processing tab can be selected. In the case you do not want to validate your results directly, close the program.

43 FlowAccess V3 TM GETTING STARTED CHAPTER 5: DATA PROCESSING Preface: This chapter gives detailed instructions for handling the results of the analysis.

44

45 Chapter 5 Data Processing DATA PROCESSING 5.1 Introduction In the Data Processing program the measured data are calculated and the results displayed. The results can be displayed in a flexible report format and printed. Also the statistics can be displayed. The layout of the results can be changed. The method set-up can be viewed and edited. This means that when a standard shows an outlying result due to wrong standard preparation, this result can be deselected, where after the results are recalculated. Peaks that are measured wrong due to disturbed peak shapes can be edited in the real-time screen and recalculated. 5.2 PostAnalysis screen The analysis run from chapter 5 can be opened here. 1. Start the FlowAccessV3 TM program. 2. Select the Data Processing tab in the left pane. 3. Select the desired run from the list in the right pane and then on PostAnalysis in the left pane. The PostAnalysis screen is opened with the layout of the system as used for the analysis (see Figure 5.1). Figure 5.1

46 FlowAccess V3 Getting started Select Results. The Results screen is displayed (see Figure 5.2). The screen is the same as during analysis (see Figure 4.4), now with all results. Figure 5.2 To view the Calibration curves select the Calibration tab (see Figure 5.3). The calculated regression coefficient for each curve is shown. A good calibration curve in the first order has an r of at least 0.999xxx. Figure 5.3 To view the statistics or to edit the table or method, select the required tab. It is possible to edit the peaks and the format of the results.

47 Chapter 5 Data Processing Peak editing To edit a peak close or minimize the Results screen to return to the PostAnalysis screen. 1. Select Real Time. The Real Time screen displays all channels of the run in separate windows (see Figure 4.1). 2. To be able to edit a peak, double click in the title bar of the channel where a peak is to be edited. This channel is then displayed in full screen with all options (see Figure 5.4). Figure Select Peak Information from the Tools menu or the icon Peak Information from the toolbar. The Peak information screen is displayed (see Figure 5.5). The left pane shows the picking point of the initial wash, which is also encircled in the Realtime screen. Information about this point is shown in the right pane. 4. Select Forward. The next peak picking point (Tracer) is shown in the left pane and information about the peak in the right pane. The peak is encircled in the Realtime screen. In this way all peaks can be viewed. 5. Move to the peak to be edited and select Modify. The peak gets a blue Figure 5.5

48 FlowAccess V3 Getting started 5-4 cross. Select Backward or Forward or use the spacebar (when Backward or Forward is selected) until the blue cross is at the correct place of the peak. 6. Select Save and Close (the Modify button changed into Save as soon as the button was clicked). The new peak picking point is displayed in the Realtime screeen. 7. Close the Realtime screen. 8. Open the Results screen. The results are already recalculated with the edited values. 5.4 Change of the table format The default table format can be changed. In the example the weight factor is added to the table. 1. Select Results to display the results. 2. Select Format. The Results Format and Custom Chemistries screen is displayed (see Figure 5.6). Figure Select Add. The items in the screen become available. 4. Check e.g, Weight in the Table Column. 5. Select Save. The Save View dialog box will be displayed. 6. Enter a name for the view and select Save. The box is closed. 7. Select the just created view from the drop down list in Available Views and select Show View. 8. The Custom Result screen is closed and the Weight factor is inserted as column in the results table.

49 Chapter 5 Data Processing Printing The results calibration curves, the statistics and the table can be printed. To print the Results: 1. Select the tab Detailed results and press the Print button. 2. A Print Preview is opened (see Figure 5.7). 3. Press the Print button in the Print preview. The results are printed. Figure 5.7 To print the Calibration curves: 1. Select the tab Calibration. 2. Click on the calibration curve to be printed (each calibration curve is printed separately). 3. Press the Print button in the toolbar. The calibration curve is printed. To print the Statistics: 1. Select the tab Statistics. 2. Select the chemistry of which the statistics has to be printed in the toolbar. 3. Click in the graph where the required information has to be displayed and then select the required type of graph in the left pane. The information will be displayed. 4. Select Print Preview from the toolbar. The Print preview is displayed. 5. Set here the desired print options and press Print. The graph will be printed. To print the Table: 1. Select the tab Edit table and method. 2. Select the table to be printed from the Table entry list and click Edit. The table will be displayed. 3. Select Print in the toolbar. A screen is displayed where the columns to be printed can be selected/deselected. 4. When the selection is made click OK. The table is printed.

50

51 Appendix APPENDIX Quick start-up reference guide A. If no system controller is present: 1. Prepare fresh reagents and standards if required, fill the containers and place the inlet tubing in the containers. 2. Close the pump decks. 3. Switch on the computer and wait until the desktop is displayed. 4. Switch on the interface. 5. Switch on the sampler and let it initialize. 6. Switch on the analyzer, other hardware and gas supply (if present). 7. Let the system stabilize for about 30 minutes. 8. Open FlowAccess while the system is stabilizing. 9. Open the active system. 10. Prepare a sample table according the samples in the sampler. 11. Select the channels that are used for the analysis. 12. When the stabilizing time has passed, check the baseline in the Real Time screen. When the baseline is stable, an analysis can be started. 13. Close the Real time screen, stop the channels. 14. Select Analysis, select the group to be analyzed and select Start to start the analysis. B When a system controller is present: Two options are possible: 1. The system was shut down in a time program 2. The system is off. When the system was shut down in a time program: To start an analysis: 1. Prepare fresh reagents and standards if required and fill the containers. 2. Close the pump decks. 3. Wake up the system and set the system online with the Mode button. 4. Let the system stabilize for about 30 minutes. 5. Prepare during stabilizing a sample/system table according the samples in the sampler. 6. Select the channels that are used for the analysis. 7. When the stabilizing time has passed, open the Real time window and check the baseline. 8. When the baseline is stable, an analysis can be started. 9. Close the Real time screen and start the analysis.

52 FlowAccess V3 Getting started To start an analysis with a new time program: 1. Prepare fresh reagents and standards if required and fill the containers. 2. Wake up the system. 3. Prepare a sample table and set a time program for the analysis. When the system is off: To start an analysis: 1. Prepare fresh reagents and standards if required and fill the containers. 2. Switch on the computer and wait until the desktop is displayed. 3. Switch on the sampler and let it initialize. 4. Switch on the interface. 5. Start FlowAccess. 6. Open the active system. 7. Set the system online with the Mode button. 8. Let the system stabilize for about 30 minutes. 9. In the meantime prepare a sample table according the samples in the sampler. 10. Select the channels that are used for the analysis. 11. When the stabilizing time has passed, open the Real time window and check the baseline. 12. When the baseline is stable, an analysis can be started. 13. Close the Real time screen. 14. Start the analysis. To start an analysis with a new time program: 1. Prepare fresh reagents and standards if required and fill the containers. 2. Switch on the computer and wait until the desktop is displayed. 3. Switch on the sampler and let it initialize 4. Switch on the interface.. 5. Start FlowAccess. 6. Open the active system. 7. Prepare a sample table according the samples in the sampler. 8. Set a time program.

53 Appendix Quick shutdown reference guide A. When no system controller is present: 1. Place the reagent lines in distilled water and rinse the system for at least 20 minutes. 2. Switch off the hardware on the analyzer and the analyzer. 3. Close the FlowAccess program. 4. Switch off the sampler. 5. Switch off the interface. 6. Switch off the computer. 7. Open the pump decks. B. When a system controller is present: 1. Set the status on Standby to rinse the system. 2. After about 20 minutes set the system on Default. 3. Close the FlowAccess program. 4. Switch off the sampler. 5. Switch off the interface. 6. Switch off the computer. 7. Open the pump decks.

FlowAccess TM V3 USER MANUAL. Windows software for Skalar SAN plus systems. Version and up

FlowAccess TM V3 USER MANUAL. Windows software for Skalar SAN plus systems. Version and up FlowAccess TM V3 Windows software for Skalar SAN plus systems Version 3.1.11 and up USER MANUAL FlowAccess TM V3 Windows software for Skalar SAN plus systems Version 3.1.11 and up User Manual Publication

More information

UniPoint System Software User s Guide

UniPoint System Software User s Guide UniPoint System Software User s Guide LT2137 1998 Gilson, Inc. All rights reserved Exercise 3-Creating operations list... 4-22 Create and name operations list... 4-22 Set up steps... 4-23 Identify sample

More information

ChromQuest 5.0 Quick Reference Guide

ChromQuest 5.0 Quick Reference Guide ChromQuest 5.0 Quick Reference Guide This guide contains an overview of the ChromQuest chromatography data system, with topics organized by workflow. For more information, refer to the ChromQuest User

More information

Tutorial EN

Tutorial EN Tutorial 8.102.8001EN Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com Tutorial 8.102.8001EN 07.2009 ek Teachware Metrohm AG CH-9101

More information

Chromeleon software orientation

Chromeleon software orientation Chromeleon software orientation Upon opening of Chromeleon shortcut, a blue screen should appear (called control panel). If this does not occur, the green circled shortcut will open this screen. To ensure

More information

Omnion 3.0 Software USER MANUAL. June 2006, Edition 3. Hach Company, All rights reserved. Printed in the U.S.A. te/dk

Omnion 3.0 Software USER MANUAL. June 2006, Edition 3. Hach Company, All rights reserved. Printed in the U.S.A. te/dk 02134 Omnion 3.0 Software USER MANUAL June 2006, Edition 3 Hach Company, 2006. All rights reserved. Printed in the U.S.A. te/dk Visit us at www.lachatinstruments.com Table of contents Section 1 Introduction...

More information

Aqueous GPC Instructions - Detailed

Aqueous GPC Instructions - Detailed Aqueous GPC Instructions - Detailed Waste Bottle Columns PL Datastream Injection Port GPC Computer UV Detector GPC Pump Eluent Bottle The Aqueous GPC Set Up RI Detector Important The points below must

More information

Video Management Software

Video Management Software Video Management Software User Manual Web: europesecurity.eu Mail: info@europesecurity.eu Tel.: 0541 352 952 User s Manual 2 Contents CHAPTER : XMS CENTRAL USER MANUAL 1 PC REQUIREMENT 3 2 INSTALL 3 3

More information

Windows Me Navigating

Windows Me Navigating LAB PROCEDURE 11 Windows Me Navigating OBJECTIVES 1. Explore the Start menu. 2. Start an application. 3. Multi-task between applications. 4. Moving folders and files around. 5. Use Control Panel settings.

More information

SoftWare. M.Wave Professional PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

SoftWare. M.Wave Professional PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Operation Manual SoftWare M.Wave Professional PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.011.14 Contents 1.

More information

QAM-I-102 Operation and Calibration of the Autoanalyzer

QAM-I-102 Operation and Calibration of the Autoanalyzer 1. Applicability and Purpose a. This procedure applies to the operation and calibration of the LaChat Quikchem 8000 Automated Ion Analyzer. An equivalent system may be used with proper revision of this

More information

Sun VirtualBox Installation Tutorial

Sun VirtualBox Installation Tutorial Sun VirtualBox Installation Tutorial Installing Linux Mint 5 LTS Guest OS By Dennis Berry Welcome to the world of virtualization and Linux. This tutorial is intended to help users who are new to the world

More information

Model FP-6500 Spectrofluorometer Instruction Manual. FP-6500 for Windows

Model FP-6500 Spectrofluorometer Instruction Manual. FP-6500 for Windows Model FP-6500 Spectrofluorometer Instruction Manual FP-6500 for Windows P/N: 0302-9999 April 2000 Contents Safety Considerations...i Regulatory Statements... iii Preface... iv Installation Conditions...v

More information

Frontline Test System

Frontline Test System Frontline Test System Serialtest Async Serialtest Spy for Windows 9x/NT Manual Technical Support Frontline Test Equipment, Inc. PO Box 7507 Charlottesville, VA 22906-7507 USA Voice: (804) 984-4500 Fax:

More information

CyberComm Pro 2.4 Data Acquisition Software Installation & User Guide. CyberScan DO 1500

CyberComm Pro 2.4 Data Acquisition Software Installation & User Guide. CyberScan DO 1500 CyberComm Pro 2.4 Data Acquisition Software Installation & User Guide CyberScan DO 1500 Bench Dissolved Oxygen Meter Technology Made Easy... 68X292341 Rev.0 01/04 PREFACE Thank you for selecting the CyberScan

More information

IT Essentials v6.0 Windows 10 Software Labs

IT Essentials v6.0 Windows 10 Software Labs IT Essentials v6.0 Windows 10 Software Labs 5.2.1.7 Install Windows 10... 1 5.2.1.10 Check for Updates in Windows 10... 10 5.2.4.7 Create a Partition in Windows 10... 16 6.1.1.5 Task Manager in Windows

More information

Welcome In this chapter

Welcome In this chapter 1 Welcome In this chapter Welcome to ProWORX NxT 2.10 2 Installing ProWORX NxT 3 Starting ProWORX NxT 5 Authorizing ProWORX NxT 8 Customer support 11 ProWORX NxT User s Guide Welcome to ProWORX NxT 2.10

More information

Operating Instructions METTLER TOLEDO Calibry Software for Pipette Calibration Calibry

Operating Instructions METTLER TOLEDO Calibry Software for Pipette Calibration Calibry Operating Instructions METTLER TOLEDO Calibry Software for Pipette Calibration Calibry Version 3.1 Contents Contents 1 Introduction... 5 1.1 Short description of Calibry... 5 1.2 Important information

More information

Release Notes Life Technologies Attune NxT Software v2.3

Release Notes Life Technologies Attune NxT Software v2.3 Release Notes Life Technologies Attune NxT Software v2.3 In the following pages you will find instructions describing: New software features Known software/system issues with troubleshooting guidance Software

More information

Operating Manual. Version 7.3.4

Operating Manual. Version 7.3.4 Operating Manual Version 7.3.4 March 2011 Index Page 1. Installation instructions 4 1.1 System requirements 4 1.2 Installation procedure 4 1.3 Meitav-tec USB adaptor installation 7 2. Open Maxinet and

More information

CyberComm Pro Data Acquisition Software Installation & User Guide

CyberComm Pro Data Acquisition Software Installation & User Guide CyberComm Pro 2.2.3 Data Acquisition Software Installation & User Guide ph 1100 and ph 2100 Bench ph and Bench ph/ion Meter Technology Made Easy... 68X090822 rev 1 Aug 2002 2 PREFACE Thank you for selecting

More information

FOODSERVICE. Owner s Manual. Read And Save These Instructions CUSTOM BLENDING. Software. for all. Programmable Vita-Mix Machines

FOODSERVICE. Owner s Manual. Read And Save These Instructions CUSTOM BLENDING. Software. for all. Programmable Vita-Mix Machines FOODSERVICE Owner s Manual Read And Save These Instructions CUSTOM BLENDING Software for all Programmable Vita-Mix Machines Table of Contents Support Information................................................3

More information

Discovering Computers & Microsoft Office Office 2010 and Windows 7: Essential Concepts and Skills

Discovering Computers & Microsoft Office Office 2010 and Windows 7: Essential Concepts and Skills Discovering Computers & Microsoft Office 2010 Office 2010 and Windows 7: Essential Concepts and Skills Objectives Perform basic mouse operations Start Windows and log on to the computer Identify the objects

More information

Data setting software MEXE02

Data setting software MEXE02 HM-40143 Data setting software MEXE02 OPERATING MANUAL Before Use Thank you for purchasing an Oriental Motor product. This operating manual describes product handling procedures and safety precautions.

More information

Multivariate Calibration Quick Guide

Multivariate Calibration Quick Guide Last Updated: 06.06.2007 Table Of Contents 1. HOW TO CREATE CALIBRATION MODELS...1 1.1. Introduction into Multivariate Calibration Modelling... 1 1.1.1. Preparing Data... 1 1.2. Step 1: Calibration Wizard

More information

OPERATION MANUAL. MV-410HS Layout Editor. Version higher. Command

OPERATION MANUAL. MV-410HS Layout Editor. Version higher. Command OPERATION MANUAL MV-410HS Layout Editor Version 3.0 - higher Command Command Table of Contents 1. Setup... 1 1-1. Overview... 1 1-2. System Requirements... 1 1-3. Operation Flow... 1 1-4. Installing MV-410HS

More information

DRAWings 8 and Wings modular 8 Embroidery Software INSTALLATION GUIDE

DRAWings 8 and Wings modular 8 Embroidery Software INSTALLATION GUIDE DRAWings 8 and Wings modular 8 Embroidery Software INSTALLATION GUIDE We thank you for purchasing DRAWings 6 and Wings modular 5 software. We are committed to providing you with the most innovative technology.

More information

Fiery proserver User Guide

Fiery proserver User Guide Fiery proserver User Guide Congratulations on your purchase of the EFI Fiery proserver, a powerful production solution for your printer. Before you begin Make sure that you have: About this document Installed

More information

TouchKit TouchScreen Controller User Guide for Windows 2000 / XP Version: 3.2.4

TouchKit TouchScreen Controller User Guide for Windows 2000 / XP Version: 3.2.4 TouchKit TouchScreen Controller User Guide for Windows 2000 / XP Version: 3.2.4 TouchKit Guide for Win2000/XP v3.2.4 0 CONTENT CHAPTER 1. TOUCH PANEL CONTROLLER...2 1.1 CONTROLLER...2 1.2 SPECIFICATIONS

More information

DriveWizard Plus Instruction Manual

DriveWizard Plus Instruction Manual DriveWizard Plus Instruction Manual To properly use the product, read this manual thoroughly. MANUAL NO. TOEP C730600 20C Table of Contents Safety Symbols and Markings...4 Manual Overview...5 Related Manuals...5

More information

Trademark and Copyrights

Trademark and Copyrights Trademark and Copyrights Océ Imagistics is a registered trademark of Océ Imagistics Inc. The Océ Imagistics logo is a registered trademark of Océ Imagistics Inc. Microsoft, Windows, Windows 3.1, Windows

More information

Finalizing the Project

Finalizing the Project Item#: MB7920 Finalizing the Project Use the steps in this section to prepare your project to send to Yearbook Interactive for duplication. This process is required before project submission. It may take

More information

NET-DYN USB Dual Band (Mediatek) Installation Guide. This manual is divided into three parts: Windows XP, Windows 7 / 8 / 8.

NET-DYN USB Dual Band (Mediatek) Installation Guide. This manual is divided into three parts: Windows XP, Windows 7 / 8 / 8. Installation Guide NET-DYN USB Dual Band (Mediatek) Installation Guide This manual is divided into three parts: Windows XP, Windows 7 / 8 / 8.1 /10, and Mac 1.Windows XP Please do the following steps to

More information

Provide Accurate Results. Operational Manual. Visible Spectrophotometer. SP-880 Mate. PC Software. Metertech Inc. Version 1.2.3

Provide Accurate Results. Operational Manual. Visible Spectrophotometer. SP-880 Mate. PC Software. Metertech Inc. Version 1.2.3 Provide Accurate Results Operational Manual Visible Spectrophotometer SP-880 Mate PC Software Metertech Inc. Version 1.2.3 Metertech Inc. provides this publication as is without warranty of any kind, either

More information

VANGUARD LOAD TAP CHANGER ANALYZER (LTCA) VERSION 2.xx SOFTWARE MANUAL. LTCA-10, LTCA-40, WRM-10P, WRM-40, TRM-20, TRM-40, TRM-203, and TRM-403

VANGUARD LOAD TAP CHANGER ANALYZER (LTCA) VERSION 2.xx SOFTWARE MANUAL. LTCA-10, LTCA-40, WRM-10P, WRM-40, TRM-20, TRM-40, TRM-203, and TRM-403 VANGUARD LOAD TAP CHANGER ANALYZER (LTCA) VERSION 2.xx SOFTWARE MANUAL For Use with Vanguard s LTCA-10, LTCA-40, WRM-10P, WRM-40, TRM-20, TRM-40, TRM-203, and TRM-403 Load Tap Changer Analyzers and Winding

More information

67 Series Spectrophotometer PC Software

67 Series Spectrophotometer PC Software 67 Series Spectrophotometer PC Software Instruction Manual 670 004/Rev D/04-17 IMPORTANT Please ensure the SD Card is fitted into the socket as detailed in the image below. -------------------------------------------------------------------------------------------------------------------------

More information

PC RateLink (Open-i) Guide

PC RateLink (Open-i) Guide Version 2.2.2 - for other versions click here PC RateLink (Open-i) Guide Click on the chapter titles below to view them: Chapter 1: Introducing RateLink Chapter 2:Installing PC RateLink Chapter 3: Installing

More information

Macintosh Supplement

Macintosh Supplement Macintosh Supplement Read this manual carefully before you use this machine and keep it handy for future reference. For safe and correct use, be sure to read the Important Information in User s Guide before

More information

Fiery proserver User Guide

Fiery proserver User Guide Fiery proserver User Guide Congratulations on your purchase of the Fiery proserver, a powerful production solution for your printer. Before you begin Make sure that you have: About this document Installed

More information

QUICK START. DevCom2000 User Manual

QUICK START. DevCom2000 User Manual QUICK START DevCom2000 uses Device Descriptions (DDs) to access data stored in the memory of the smart field device. These DDs are developed by the manufacturer for their products and, in turn, distributed

More information

Standard Plus Player. User Guide. i-tech Company LLC TOLL FREE: (888) WEB:

Standard Plus Player. User Guide. i-tech Company LLC TOLL FREE: (888) WEB: Standard Plus Player User Guide i-tech Company LLC TOLL FREE: (888) 483-2418 EMAIL: info@itechlcd.com WEB: www.itechlcd.com 1. INTRODUCTION OF THE Standard Plus PLAYER... 3 2. MAIN MENU... 4 2.1 START

More information

User Manual. Computer Program Pomiar Win. Version: 4.0.X. Manual number: ITKU A MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS

User Manual. Computer Program Pomiar Win. Version: 4.0.X. Manual number: ITKU A MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS User Manual Computer Program Pomiar Win Manual number: ITKU-05-02-09-10-A Version: 4.0.X MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG Wagi Elektroniczne, 26-600 Radom 28 Bracka Street - POLAND

More information

Computer Essentials Session 1 Lesson Plan

Computer Essentials Session 1 Lesson Plan Note: Completing the Mouse Tutorial and Mousercise exercise which are available on the Class Resources webpage constitutes the first part of this lesson. ABOUT PROGRAMS AND OPERATING SYSTEMS Any time a

More information

UNIVERSAL SOFTWARE. Universal Software. Data Sheet

UNIVERSAL SOFTWARE. Universal Software. Data Sheet Universal Software Data Sheet System Requirements: The minimum requirements for using the Software are: 1). Windows XP/Vista/7 2). A minimum of 512 MB RAM 3). 1 GB of hard disk space 4). Microsoft Office

More information

Quick Start Guide: Fiery proserver for EFI VUTEk

Quick Start Guide: Fiery proserver for EFI VUTEk Quick Start Guide: Fiery proserver for EFI VUTEk This document is a step-by-step guide on how to set up and use the EFI Fiery proserver with your EFI VUTEk printer. It deals with the following topics:

More information

1 - The Windows XP Desktop

1 - The Windows XP Desktop 1 - The Windows XP Desktop Above in fig 1.1 you can see an example of the Windows XP desktop, when you first switch on your machine and Windows XP loads it usually looks similar to that shown in our example,

More information

User Manual. Central Monitoring Software for TER-series recorders. CenPro

User Manual. Central Monitoring Software for TER-series recorders. CenPro EN User Manual Central Monitoring Software for TER-series recorders CenPro Contents CENPRO USER MANUAL 1 PC REQUIREMENT 3 2 INSTALL 3 3 UNINSTALL 5 4 BASIC OPERATION 7 4.1 Log In 7 4.2 Screen Layout 8

More information

NETWORK PRINT MONITOR User Guide

NETWORK PRINT MONITOR User Guide NETWORK PRINT MONITOR User Guide Legal Notes Unauthorized reproduction of all or part of this guide is prohibited. The information in this guide is subject to change for improvement without notice. We

More information

TouchKit Touch Panel User manual for Windows9X/ME Version: 3.1.4

TouchKit Touch Panel User manual for Windows9X/ME Version: 3.1.4 TouchKit Touch Panel User manual for Windows9X/ME Version: 3.1.4 TouchKit Touch Panel v3.1.4 0 CONTENT CHAPTER 1. TOUCH PANEL CONTROLLER... 2 1.1 CONTROLLER... 2 1.2 SPECIFICATIONS AND FEATURES... 3 CHAPTER

More information

Procom PPC Software User s Manual

Procom PPC Software User s Manual Procom PPC Software User s Manual Version 1.1 Procom Engineering Inc. 66 Maxwell, Irvine, CA 92618, USA Tel: 949-748 6338 Fax: 949-748 6339 Email: info@procomengineering.com www.procomengineering.com Contents

More information

GUARD1 PLUS Documentation. Version TimeKeeping Systems, Inc. GUARD1 PLUS and THE PIPE are registered trademarks

GUARD1 PLUS Documentation. Version TimeKeeping Systems, Inc. GUARD1 PLUS and THE PIPE are registered trademarks GUARD1 PLUS Documentation Version 3.02 2000-2005 TimeKeeping Systems, Inc. GUARD1 PLUS and THE PIPE are registered trademarks i of TimeKeeping Systems, Inc. Table of Contents Welcome to Guard1 Plus...

More information

Microsoft Windows 10. Quick Reference. Watsonia Publishing 47 Greenaway Street Bulleen VIC 3105 Australia

Microsoft Windows 10. Quick Reference. Watsonia Publishing 47 Greenaway Street Bulleen VIC 3105 Australia Watsonia Publishing 47 Greenaway Street Bulleen VIC 3105 Australia www.watsoniapublishing.com info@watsoniapublishing.com Quick Reference Course Code: INF1440 Table of Contents Chapter 1: Starting With

More information

Windows 7 Professional 64 bit Installation and Configuration for MassLynx or Empower Controlled Ethernet Instrument Communication

Windows 7 Professional 64 bit Installation and Configuration for MassLynx or Empower Controlled Ethernet Instrument Communication Windows 7 Professional 64 bit Installation and Configuration for MassLynx or Empower Controlled Ethernet Instrument Communication May 2014 Version 4 PLEASE READ BEFORE CONTINUING: This document applies

More information

User Manual JofraCal for JOFRA Temperature, Pressure and Signal Calibrators. Copyright 2004 AMETEK DENMARK A/S

User Manual JofraCal for JOFRA Temperature, Pressure and Signal Calibrators. Copyright 2004 AMETEK DENMARK A/S User Manual JofraCal for JOFRA Temperature, Pressure and Signal Calibrators Copyright 2004 AMETEK DENMARK A/S 2 124918 01 LIST OF CONTENTS Page 1. INTRODUCTION... 11 1.1 General information... 12 1.2 Hardware

More information

xponent Version 4.2 RELEASE NOTES

xponent Version 4.2 RELEASE NOTES xponent Version 4.2 RELEASE NOTES Copyright Luminex Corporation, 1996-2013. All rights reserved. Part Number: 89-30000-00-462 Disclaimer Subject to change without notice. Although prepared to ensure accuracy,

More information

Ansur Index 2XL. Users Manual. Plug-In

Ansur Index 2XL. Users Manual. Plug-In Ansur Index 2XL Plug-In Users Manual April 2010, Rev. 1 2010 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies. Table of Contents Chapter Title Page

More information

Logger Pro 3. Quick Reference

Logger Pro 3. Quick Reference Logger Pro 3 Quick Reference Getting Started Logger Pro Requirements To use Logger Pro, you must have the following equipment: Windows 98, 2000, ME, NT, or XP on a Pentium processor or equivalent, 133

More information

Instruction Manual XLComm Data Acquisition Software For XL Series Meters

Instruction Manual XLComm Data Acquisition Software For XL Series Meters Data Acquisition Software For XL Series Meters (XL 15, 20, 25, 30, 40, 50 and 60 meters) 68X090832 Rev. 0 JAN 07 Preface This manual serves to explain the use of the Acquisition Software. The manual functions

More information

SequencePro Data Analysis Application. User Guide

SequencePro Data Analysis Application. User Guide SequencePro Data Analysis Application User Guide SequencePro Data Analysis Application User Guide DRAFT October 31, 2001 12:52 pm, Title_page.fm Copyright 2001, Applied Biosystems. All rights reserved.

More information

Quick Start Guide. Quick Start Guide I

Quick Start Guide. Quick Start Guide I Quick Start Guide Quick Start Guide I II Quick Start Guide Contents of Package The Multi-Converter package contains the following items: > One (1) Multi-Converter Device > One (1) USB Cable > One (1) Installation

More information

FaxFinder Fax Servers

FaxFinder Fax Servers FaxFinder Fax Servers Models: FF130 FF230 FF430 FF830 Client User Guide FaxFinder Client User Guide Fax Client Software for FaxFinder Series PN S000460B, Version B Copyright This publication may not be

More information

Computer Basics. Hardware. This class is designed to cover the following basics:

Computer Basics. Hardware. This class is designed to cover the following basics: Computer Basics This class is designed to cover the following basics: computer hardware software computer s operating system different kinds of storage devices you can use to save files using a mouse what

More information

TopPoint Administrator User's Guide

TopPoint Administrator User's Guide TopPoint Administrator User's Guide LT2153/ 2000 Gilson, Inc. All rights reserved December 2000 Table of Contents 1 Introduction 2 Installation 3 Operation System Requirements... 1-2 Minimum... 1-2 Recommended...

More information

Podium Plus Data Analysis Software. User Manual. SWIS10 Version

Podium Plus Data Analysis Software. User Manual. SWIS10 Version SWIS10 Version Issue 1.10 February 2005 Contents 1 Introduction 6 1.1 What is Podium Plus? 6 1.2 About This Manual 6 1.3 Typographical Conventions 7 1.4 Getting Technical Support 7 2 Getting Started 8

More information

SoftTools Suite 7.0. Flowserve Corporation

SoftTools Suite 7.0. Flowserve Corporation SoftTools Suite 7.0 Quick Start Guide for SoftTools 7.0 with support for the HART Communications Protocol on the Logix 1200/1200e, 3200IQ, and 500/500si Series Positioners Flowserve Corporation Flow Control

More information

GX-2009 Data Logger Management Program Operator s Manual

GX-2009 Data Logger Management Program Operator s Manual GX-2009 Data Logger Management Program Operator s Manual Part Number: 71-0163RK Revision: P1 Released: 4/30/09 www.rkiinstruments.com Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by

More information

Table of Contents. 1. Intended Use Overview Materials Required System Requirements User Mode 4

Table of Contents. 1. Intended Use Overview Materials Required System Requirements User Mode 4 Table of Contents 1. Intended Use.. 3 2. Overview.. 3 2.1 Materials Required. 3 2.2 System Requirements. 3 2.3 User Mode 4 3. Installation Instructions.. 5 3.1 Installing the On Call Diabetes Management

More information

MAXTOR ONETOUCH III MINI EDITION. User s Guide

MAXTOR ONETOUCH III MINI EDITION. User s Guide MAXTOR ONETOUCH III MINI EDITION User s Guide MAXTOR ONETOUCH III MINI EDITION USER S GUIDE Contents Windows Installation Introduction.................................................... 1 About this Guide...............................................

More information

UPDATING THE SYSTEM MANAGER SD Updating the System Manager with Prism 2

UPDATING THE SYSTEM MANAGER SD Updating the System Manager with Prism 2 Updating the System Manager with Prism 2 Updating the System Manager SD with Prism 2 Version 4.5.0 and higher 7. Click on the icon. The Job-Sites Window will appear. In the Type of CommLink

More information

GX-6000 Data Logger Management Program Operator s Manual

GX-6000 Data Logger Management Program Operator s Manual GX-6000 Data Logger Management Program Operator s Manual Part Number: 71-0383 Revision: B Released: 7/27/17 www.rkiinstruments.com Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us

More information

ibidi PumpControl Instruction Manual Version 1.5.0

ibidi PumpControl Instruction Manual Version 1.5.0 ibidi PumpControl Instruction Manual Version 1.5.0 II ibidi GmbH 2011, version 1.5.0, revision 2, 2011-10-26, Andreas Rainer Table of contents 1 General Information... 1 1.1 System Requirements... 1 1.2

More information

Remote Deposit Service

Remote Deposit Service Remote Deposit Service Start Up Guide *Screen shots are shown for illustrative purposes only and may vary based on the version of Windows you are currently using. Step 1: Confirm your PC meets system

More information

XDR Installation. USB Driver Installation and Testing

XDR Installation. USB Driver Installation and Testing XDR Installation The installation and configuration of Sopro inter-oral cameras with XDR version 3.2.1 is a three step process. The first step is to install the USB 2.0 driver and test the camera functionality.

More information

Creating Interactive PDF Forms

Creating Interactive PDF Forms Creating Interactive PDF Forms Using Adobe Acrobat X Pro for the Mac University Information Technology Services Training, Outreach, Learning Technologies and Video Production Copyright 2012 KSU Department

More information

KIMOTO K I M O S E T T E R R I P. Kimosetter RIP User Guide 1. Revised: February 2015 U SER GUID E (FOR W INDOWS 7 )

KIMOTO K I M O S E T T E R R I P. Kimosetter RIP User Guide 1. Revised: February 2015 U SER GUID E (FOR W INDOWS 7 ) KIMOTO K I M O S E T T E R R I P U SER GUID E (FOR W INDOWS 7 ) Revised: February 2015 Kimosetter RIP User Guide 1 COPYRIGHT AND TRADEMARKS Kimosetter RIP User Guide Copyright Notices for the Software

More information

INSTALLING THE PS3 XBOX READY SOFTWARE:

INSTALLING THE PS3 XBOX READY SOFTWARE: INSTALLING THE PS3 XBOX READY SOFTWARE: 1. Insert the Installation CD to CD-ROM drive and execute Ready_Setup.exe NOTE: If it is the first time for the target USB disk using under this software, the software

More information

Centralised monitoring software. (version v2.35) Program installation and configuration

Centralised monitoring software. (version v2.35) Program installation and configuration Centralised monitoring software (version v2.35) Program installation and configuration The purpose of the document This document describes the installation and configuration process of program package

More information

Provides access to Worldspan Reservations Manager.

Provides access to Worldspan Reservations Manager. The Worldspan Go! SM Home Page Windows 95 Quick Reference The home page displays when you start your computer. This screen is continuously updated with new information. The figure below is intended to

More information

Install Manual. P2 Software ENGLISH M0509AT5091 -FJ VQT2G26-4

Install Manual. P2 Software ENGLISH M0509AT5091 -FJ VQT2G26-4 Install Manual P2 Software M0509AT5091 -FJ ENGLISH VQT2G26-4 Contents Operating Precautions When Using Windows 1. Before Installing................................................... 3 2. Precautions in

More information

Tremblay Harrison Inc Dufferin St Toronto, ON M6H 3M Version 1.2

Tremblay Harrison Inc Dufferin St Toronto, ON M6H 3M Version 1.2 Tremblay 1684 Dufferin St Toronto, ON M6H 3M1 1 866 829-7926 Version 1.2 Introduction The EZ Health Oracle Diabetes Management System is a personal computer (PC) software application that allows users

More information

KMnet Viewer. User Guide

KMnet Viewer. User Guide KMnet Viewer User Guide Legal Notes Unauthorized reproduction of all or part of this guide is prohibited. The information in this guide is subject to change for improvement without notice. We cannot be

More information

Windows 2000 Professional

Windows 2000 Professional The American University in Cairo Academic Computing Services Windows 2000 Professional prepared by Soumaia Ahmed Al Ayyat 4 August 2003 Table of Contents Starting Up the Computer Windows Environment Start

More information

Kimosetter RIP. User Guide (for Windows) Kimosetter RIP User Guide. Revised: July, 2007

Kimosetter RIP. User Guide (for Windows) Kimosetter RIP User Guide. Revised: July, 2007 Kimosetter RIP User Guide (for Windows) Revised: July, 2007 1 Copyright and Trademarks (Win version) July, 2007 The Copyright Notices for the Software and Documentation: Copyright Lucid Dream Software,

More information

Central Management Software. Cam Viewer 3 Lite. User Manual

Central Management Software. Cam Viewer 3 Lite. User Manual Central Management Software Cam Viewer 3 Lite User Manual Version 1.0.0 Table of Contents 1. System Requirement...4 2. Software Installation...5 3. Configuration Wizard...9 3.1 Channel...10 3.1.1 Adding

More information

CompuLEAD: Installing LeadLINK Plus!

CompuLEAD: Installing LeadLINK Plus! CompuLEAD: Installing LeadLINK Plus! LEADlink Plus! Sales Lead Management Software, by CompuSystems, is a new, innovative and creative way of capturing sales leads through a CompuLEAD data recorder into

More information

Lab - Data Migration in Windows

Lab - Data Migration in Windows Lab - Data Migration in Windows Introduction When a new computer is purchased or a new operating system is installed, it is often desirable to migrate a user s data to the new computer or OS. Windows has

More information

PAL Installation Instructions

PAL Installation Instructions PAL Installation Instructions ADMINISTRATOR (NETWORK) INSTALLS These instructions are for PCs that are connected by a network, for the purpose of sharing PAL data and notes. PAL can be installed using

More information

Legal Notes. Regarding Trademarks KYOCERA MITA Corporation

Legal Notes. Regarding Trademarks KYOCERA MITA Corporation Legal Notes Unauthorized reproduction of all or part of this guide is prohibited. The information in this guide is subject to change without notice. We cannot be held liable for any problems arising from

More information

Quick Start. Chapter. In This Chapter:

Quick Start. Chapter. In This Chapter: Chapter Quick Start In This Chapter: Getting to Know Windows.... - Installation of irectsoft 6.... - Getting Started.... -8 Welcome to irectsoft00.... - Begin Editing a Program.... -4 Establish the Communication

More information

For additional information, see these documents on the Breeze 2 software DVD:

For additional information, see these documents on the Breeze 2 software DVD: R E L E A S E N O T E S Waters Breeze 2 Read these release notes before you install and operate Breeze 2. They contain information necessary to ensure successful liquid chromatography results, including:

More information

FastLook. LAN License. This guide is intended to help you install the FastLook LAN version.

FastLook. LAN License. This guide is intended to help you install the FastLook LAN version. FastLook LAN License This guide is intended to help you install the FastLook LAN version. Please refer to the electronic help file for more details on the operation of FastLook. Contents Overview...1 Server

More information

METS methane sensor SOFTWARE MANUAL UNDERWATER SENSORS FOR DISSOLVED GASES

METS methane sensor SOFTWARE MANUAL UNDERWATER SENSORS FOR DISSOLVED GASES METS methane sensor SOFTWARE MANUAL UNDERWATER SENSORS FOR DISSOLVED GASES - METS - Software Manual Release: Software v4.1.3, August 2006 Manual v4.1.5, July 2014 CONTENTS 1 SOFTWARE SYSTEM REQUIREMENTS

More information

INSTALLATION AND USER GUIDE

INSTALLATION AND USER GUIDE INSTALLATION AND USER GUIDE Introduction 1 Introduction About 6140 Agent Portal Mitel Networks 6140 Agent Portal allows an agent to receive call information via a number of different applications that

More information

TouchKit TouchScreen Controller User Manual for Windows NT4 Version: 3.4.0

TouchKit TouchScreen Controller User Manual for Windows NT4 Version: 3.4.0 TouchKit TouchScreen Controller User Manual for Windows NT4 Version: 3.4.0 1 CONTENT CHAPTER 1. TOUCH PANEL CONTROLLER 2 1.1 Controller 2 1.2 Specifications and Features 3 CHAPTER 2. INSTALLING TOUCHKIT

More information

User s Manual CAP 531*1.5 Configuration and Programming tool

User s Manual CAP 531*1.5 Configuration and Programming tool User s Manual CAP 531*1.5 Configuration and Programming tool This manual belongs to: Contents Chapter Page About this manual 1 Introduction 3 Technical descriptions 81 References 177 Customer feedback

More information

idaq Device Manager User Manual

idaq Device Manager User Manual idaq Device Manager User Manual REV. 1 JULY 2018 Worldwide technical support and product information: www.toolsforsmartminds.com TOOLS for SMART MINDS Corporate headquarter Via Padania, 16 Castel Mella

More information

CITO2 Installation & User Instructions

CITO2 Installation & User Instructions CITO2 Installation & User Instructions DD 56107 Stoneridge Electronics Ltd 1. Installation...4 1.1. System Requirements...4 1.2. Installing CITO2...4 1.3. Uninstalling CITO2...4 2. Starting and closing

More information

This section provides you with information for installing P-CAD products and modifying a P-CAD installation.

This section provides you with information for installing P-CAD products and modifying a P-CAD installation. P-CAD 2001 Installation Instructions Network Edition ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

More information

Operating Instructions. Windows. Video Wall Adjustment & Calibration Software. Video Wall Manager Ver. 1.8

Operating Instructions. Windows. Video Wall Adjustment & Calibration Software. Video Wall Manager Ver. 1.8 Operating Instructions Video Wall Adjustment & Calibration Software Windows Video Wall Manager Ver. 1.8 Thank you for purchasing this Panasonic product. Before using this software, please read the instructions

More information

Introduction to Windows

Introduction to Windows Introduction to Windows Naturally, if you have downloaded this document, you will already be to some extent anyway familiar with Windows. If so you can skip the first couple of pages and move on to the

More information