1 Operating Instructions Controller for Automatic Mass Comparators from Software Version. for CC...- L / CCE...- L Model

2 Contents Page Practical Use Manual operation with cursor button Getting Started 5 Configuring the Controller 5 Setting the Language 6 Setting the Time for Averaging 6 Setting the Date and Time 7 Setting the Number of A B B A Cycles 8 Setting the Number of Pre - Cycles 9 Setting the Delay time before taken values 0 Setting the Scale - Timeout Setting the Results via AUX Setting the Start delay Operating the Controller Manual Operation of the Load Alternator Move Vertically Raise to Upper Position Move to Centring Position Lower toward Mass Comparator Move Horizontally 5 Starting the Mass Comparison 7 ) without start delay time is setting 7 Select the procedure for mass comparison 7 Begin Measurement 8 Continuous Function 9 Starting the Mass Comparison 0 ) when start delay time is setting 0 Select the procedure for mass comparison 0 The mass comparison start Continuous Function Data Interface ( Aux ) Values only Values and Times 5 Connecting cable to PC 5 Programming for external control from PC 6 Linking the PC to the control unit 6 External control commands 6 Vertical movements 6 Horizontal movements 6 Automatic weighing 7 Commands without movement 7 Error messages 7 Troubleshooting 8 Error messages 8 Maintenance 0 Replacing the Paper Roll 0

3 Practical Use The CC...-L / CCE..-L controller regulates the load alternator function sequences as well as the collection and evaluation of measured data. The individual values measured are output automatically to the built-in printer with the date and time of measurement and the calculated results. A modern, touch-screen display shows all function sequences and also serves as the user interface for operating the controller. User guidance prompts in plain English will guide you through the simple operation of the device. You can choose the following languages: - German - English - Spanish - French All of the function sequences are displayed and can be activated by pressing the corresponding field on the right hand side of the touch screen e. g. The number in this corresponding field is match marking to the number in front of the function sequence. The corresponding touch fields are not in the same line to the function sequences on the touch screen. Operators errors are virtually impossible! The controller recognizes the position of the load alternator at all times; if you select a function that cannot or should not be performed from a certain position, an error message is displayed and the function is not performed. This prevents damage to the mass comparator that could otherwise have been caused by operator error. You can choose the following measurement sequences: x A B B A - cycle 6 x A B B A - cycles 0-99 x A B B A - cycles With the -position load alternator, you can compare: - weights: N - - weights: N - / N - / N - - Combinations: N - / - / N - / - / N - / - The reset key is to stop the measurement also the emergency interrupt key. As soon as you press the reset key, the load alternator is stopped and remains in the position it was in when you pressed reset. Any measurement sequence running at the time is stop. The display indicates that the function was interrupted. You can also press reset to stop a measurement sequence after you have selected the continuous function.

4 Manual operation with the cursor buttons There are four cursor buttons on the front side. With this cursor buttons you can easy operate the load alternator, in it s basic movements. That means turning right and left, lifting or lowering the platform. This is helpful for loading the platform from the front side and other applications. The cursor buttons are only active if the main menu appears on the display. See page no. 5 Movement vertically: Movement horizontally: Four control light identify the current position and the activity of the load alternator. All lights on means that the platform is in the upper position. Movement in all directions is possible. A blinking control light identifies the actually operation relating to the direction of movement. A non active control light means that a movement in this direction is not possible. To start the movement by your choice, press the cursor button. Horizontal movement: - short press turn 90 degrees, - long press ( sec.) turn 80 degrees, Vertical movement: - short press move - from the top toward to centre, or - from the centre to the lower position, or - raise from the lower to the centre, or - raise from the centre to the upper position - long press move - direct from the upper to the lower, - from the lower to the upper position. Note: Two control lights on the printer indicate the capacity of the paper A green light means: enough paper A red light means: paper roll come to end If the paper roll in the printer runs out in the middle of a measuring sequence, the printer switches off. The individual values measured and the results calculated after the printer shuts off are stored in memory. For example: with 6 x A B B A selected, the memory can store data from up to 6 complete measurements. When the memory is full, this is indicated by a crossed-out printer symbol in the display. The load alternator continues according to the selected function sequence. Subsequent measurement data are not stored. To retain the data stored in memory: - Do not interrupt the measurement sequence - Do not press reset - Do not turn off the controller Place a new roll of paper in the printer (see "Maintenance") The stored data are output to the printer.

5 Getting Started After you turn on the controller, the Sartorius logo is displayed for approx. 5 second. Then the initial start up window is displayed. Initial System Start up load the weights. The continuous display in the upper right corner shows: - model no. of the connected comparator - software version, - language setting, - date. Press anywhere to access the main menu. Press anywhere to access the main menu. Configuring the Controller : Mass comparison Cycles : Move vertically : Move Horizontally : Configure All settings is stored in the memory when you turn off the controller. To configure the device settings, Press the numeric key "" for configuration. 5

6 The main configuration windows is displayed. Setting the Language : Language : Time for - averaging ( 0 ) : Date and time : More Configurations.. Press the numeric key "" to access the language menu. Press the numeric key on the touchscreen that corresponds to the desired language. : Deutsch : English : Español : Francais The display returns to the main configuration window in the desired language. Setting the Time for Averaging : Language : Time for - averaging ( 0 ) : Date and time : More Configuration.. The time for averaging is a period of time during which measured values are transferred to the controller continuously; i.e., each time a weight is placed on the mass comparator and stabilizes. The average calculated from these values is output for further processing. Press the numeric key "" to access the menu for setting the averaging time. 6

7 : Time for averaging Please enter the new value. ( 0 ) X 0 The following setting windows is displayed. Set a time from 05 to 99 seconds Then press to confirm new settings, or X to cancel new settings The display returns to the main configuration windows. Setting the Date and Time : Language : Time for averaging ( 0 ) : Date and time : More Configurations.. Press the numeric key "" to access the menu for setting the date and time. The following setting windows is displayed. : Hours ( 0 ) : Minutes ( 5 ) : Day - ( 0 ) : Month ( 07 ) 5: Year ( 05 ) 6: Set 5 6 The procedure for setting the date and time is self-explanatory. Press the number on the touch-screen that corresponds to the item you wish to set. Follow the instructions displayed. Press the numeric key "6" to complete the settings. When you have completed the settings, the display returns to the main configuration window, see following. 7

8 For more configuration: : Language : Time for averaging ( 0 ) : Date and time : More Configurations.. Press the numeric key "" to access to the second configurations windows. The second configuration windows is displayed. : Set number of A B B A - cycles ( 0 ) : Set number of Pre - Cycles ( 00 ) : Delay time before taken values ( 0 ) : More Configurations.. Setting the Number of A B B A Cycles You can set the desired number of A B B A cycles for a given measurement sequence. The factory setting is 0 x A B B A. To select a different number of ABBA - cycles, Press the numeric key "". The following setting windows is displayed. Set a number of cycles between 0 and 99. : Set number of - A B B A - cycles. Please enter the new value. ( 0 ) Then press to confirm new settings, or X 0 X to cancel new settings. The display returns to the second configuration window. 8

9 To optimize measuring sequences to meet the needs of the ambient conditions, the following settings are required. Setting the Number of Pre - Cycles : Set number of A B B A - cycles ( 0 ) : Set number of Pre - Cycles ( 00 ) : Delay time before taken values ( 0 ) : More Configurations.. This setting enables user-definable A B B A cycles to be performed immediately before the actual measurement. We recommend that you use this setting if, for instance, you plan to start a measurement right after placing weights on the load alternator platform. The factory setting is 00. To select a different number of Pre - cycles, Press the numeric key "". The following setting windows is displayed. : Set number of Pre - Cycles Please enter the new value. ( 00 ) 7 X Set a number of cycles between 00 and 99. Then press to confirm new settings, or X to cancel new settings. S o f t w a r e V.. The display returns to the second configuration window. : Set number of A B B A - cycles ( 0 ) : Set number of Pre - Cycles ( 00 ) : Delay time before taken values ( 0 ) : More Configurations.. 9

10 Setting the Delay Time Before Taken Values (XX) : Set number of A B B A - cycles ( 0 ) : Set number of Pre - Cycles ( 00 ) : Delay time before taken values ( 0 ) : More Configurations.. This setting defines a period of time before measured values are transmitted after the mass comparator has been loaded. This time begins as soon as the mass comparator is loaded and the load alternator motor shuts off. Because of, once the mass comparator has been loaded and the stability symbol g is displayed, the weight value may drift slightly until the weight value is recognized as a stable value in order to be transmitted. This time may differ from location to location and must be determined before performing the actual measurement. The factory setting is 0. To select a different number of Delay time, Press the numeric key "". The following setting windows is displayed : Delay time before taken values Please enter the new value. ( 0 ) 7 X Set a number of time between 00 and 99 sec Then press to confirm new settings, or X to cancel new settings. The display returns to the second configuration window. For more configurations : Set number of A B B A - cycles ( 0 ) : Set number of Pre - Cycles ( 00 ) : Delay time before taken values ( 0 ) : More Configurations.. Press the numeric key "". A third configuration windows is displayed. 0

11 : Scale - Timeout ( OFF ) : Results via AUX: ( values only ) : Swap keys. (OFF) : More Configurations.. Setting the Scale - Timeout This setting defines the time at which a measurement is to be automatically aborted in the event of an error, for example, if the stability symbol g is not displayed after the mass comparator has been loaded. This may happen if there is any interference due to vibration transferred to the surface on which the mass comparator rests, turbulent air currents or changes in the climatic conditions. Another source of this error is if the wrong substitution weight (F key on the display panel of the mass comparator) was selected for the weight to be measured. An additional source of this error is if the wrong number of weights was selected in the menu Type of comparisons before beginning the measurement; for example using weights when have actually been selected or visa versa. Scale - Timeout ( ON ) = waiting time hour Scale - Timeout ( OFF ) = waiting time infinity To change between ON / OFF press the numeric key. Note: After time out, the display will show this message: No correct value available! : Scale - Timeout ( OFF ) : Results via AUX: ( values only ) : Swap keys. (OFF) : More Configurations.. - J A N - 0 Setting the Results via AUX This setting defines the formats for receiving measurements data on PC (refer data interface page 9 and 0) Results via AUX: ( values only ) = measurement values only ( values and times ) = measurement values with To change between values only/values and times press the numeric key. For corrections or control of all settings please scroll by pressing the numeric key, or conclude the configurations by pressing the Sartorius logo to access to the main menu. A fourth configuration windows is displayed.

12 : Set start delay ( 000 ) : : : More configurations.. J A N - 00 Seting the start delay ( 000 ) This setting make it possible to start the measurements after the deceleration time. max. delay time : 56 Minute This setting is very useful for the of acclimatization of all weights in the same chamber. To set or change the delay time, Press the numeric key "" The following setting windows is displayed. J A N - 00 Set a number of time between 000 and 55 Minute : Set start delay Please enter the new value. ( 000 ) 7 X Then press to confirm new settings, or X to cancel new settings. The display returns to the fourth configuration window. : Set start delay ( 000 ) : : : More configurations.. J A N - 00 For corrections or control of all settings please scroll by pressing the numeric key, or conclude the configurations by pressing the Sartorius logo to access to the main menu. The main window is displayed.

13 Operating the Controller Manual Operation of the Load Alternator E.g.:. To control the position of the weight. To adjust the mass comparator. After interrupting a measurement Vertical Movement The main menu is displayed. If not, press the Sartorius logo. Always begin with the "Move vertically" item. : Mass comparison cycles : Move vertically : Move horizontally : Configure Press the number key "". Raise to Upper Position When you select this function, the load alternator goes through several position shifts until it recognizes the initial raised position. : Raise load alternator to upper position : Move load alternator to centering position : Load load alternator toward mass comparator : Move horizontally Always activate this function: - after turning on the mass comparator, - after canceling a measurement by pressing reset, and - before configuring the controller. When the menu illustrated on the left is displayed, press the number key "". The load alternator platform is raised into the upper initial position. : Raise load alternator to upper position : Move load alternator to centering position : Load load alternator toward mass comparator : Move horizontally : in progress. If the platform is already in this position when you select this function, it is lowered slightly and then returned to this position. ": in progress" is displayed while the platform is in motion.

14 Move to Centering Position : Raise load alternator to upper position : Move load alternator to centering position : Load load alternator toward mass comparator : Move horizontally : in progress. S o f t w a r e V.0 - J u l With this function, the load alternator Platform is parked in on intermeditate position as it is lowered. You can perform a visual inspection at this point. Press the number key "". ": in progress" is displayed during this procedure. Lower toward Mass Comparator : Raise load alternator to upper position : Move load alternator to centering position : Load load alternator toward mass comparator : Move horizontally When you activate this function, the load alternator platform is lowered toward the mass comparator and the weight is placed on the weighing pan. Press the number key "". You can select this function immediately following Function, if desired. When you press the "" key, the message ": Please comfirm by reselecting function" is displayed. : Raise load alternator to upper position : Move load alternator to centering position : Load load alternator toward mass comparator : Move horizontally : Please confirm by Re-selectiong function. Press the number key again "". The load alternator platform is lowered toward to the mass comparator.

15 : Raise load alternator to upper position : Move load alternator to centering position : Load load alternator toward mass comparator : Move horizontally : in progress. ": in progress" is displayed while the load alternator is in motion. After visual inspection or sensetivity adjustment of the mass comparator, return the load alternator platform to the upper position. Press the number key "". When the load alternator platform is in the upper position, touch the Sartorius logo on the screen to return to the main menu. Horizontal Movement : Mass Comparison cycles : Move vertically : Move horizontally : Configure C C U L - J A N 00 You can only activate this function when the load alternator is in the raised position. (If necessary, you can ensure that this is the case by selecting Function.) Press the number key "". When you press the "" key, another window is opened. : Turn 80 / Move to next defined position : Turn 90 : Turn 90 : Move vertically Press the number key that corresponds to the desired function: e.g., to rotate the alternator 80, press the number key "". 5

16 ": in progress" is displayed while the load alternator is in motion. : Turn 80 / Move to next defined position : Turn 90 : Turn 90 : Move vertically : in progress. Press the corresponding number key to activate other functions as desired. Then press the Sartorius logo on the display To return to the main menu The main menu is displayed. : Mass Comparison cycles : Move vertically : Move horizontally : Configure

17 Starting a Measurement ) without start delay time is setting Before placing a weight on the load alternator platform, make sure that the load alternator is in the initial raised position. See the section entitled "Vertical Movement." Center the reference and test weights within the marking on the load alternator platform. Always position the reference weight on the "N." Before beginning the measurement, the reference weight must be in position, on the "N" on the load alternator platform, above the weighing system. See the section entitled "Horizontal Movement." Select the procedure for mass comparison Select the "Mass comparison cycles" item from the main menu. To do this : Mass comparison cycles : Move vertically : Move horizontally : Configure Press the number key "". : A B B A - cycle : 6 A B B A - cycles : User-defined number of A B B A - cycles ( 0 ) : Mass comparison method When the start menu window opens, select the "Mass comparison method" item first. Press the number key "". : Mass comparison of two weights : Mass comparison of four weights : All 6 comparisons (N- / - / N- / - / N- / -) Mass comparison cycles - J A N 00 Select the desired method of mass comparison: e.g.: : Mass comparison of two weights Press the number key "". 7

18 : Mass comparison of two weights : Mass comparison of four weights : All 6 comparisons (N- / - / N- / - / N- / -) : Mass comparison cycles : in progress. ": in progress" is short displayed when you press this key. Press the number key "" to return to the start menu. Begin Measurement : A B B A - cycle : 6 A B B A - cycles : User-defined number of A B B A - cycles ( 0 ) : Mass comparison method : Please confirm by re-selecting function. When the start menu is displayed, press the number key that corresponds to the desired number of cycles to start the measurement procedure: e.g.: to select 6 A B B A cycles, press the number key"". The message : Please confirm by re-selecting function is displayed. Press the "" key again to start the measurement Start the selected measurement sequence, or cancel this input if you wish to select a different number of cycles by pressing another number key, or press the Sartorius logo to return to the main menu. : A B B A - cycle : 6 A B B A - cycles : User-defined number of A B B A - cycles ( 0 ) : Mass comparison method : in progress. To start the selected measurement sequence: Press the number key once more "". ": in progress" is displayed. 8

19 The measurement procedure starts The printer begins by printing the current time and date, the mass comparator model designation, the number of A B B A cycles and the number of weights. The load alternator begins by centering the weights. The first weight to be tested is centered first. With weights, this is the weight in position. The load alternator platform rotates 80 that the weight in position is above the mass comparator, then begins centering the weight. The weight is placed on the weighing pan and centered by the Centermatic times, in regular cycles. The procedure is then repeated using the reference weight. One ABBA - cycles end the centering procedure. Continuous Function With the continuous function option, you can repeat complete measurement cycles continuously. To stop the continuous function, press the reset key. Note: Do not press the reset key while the load alternator platform is in motion! Starting the Continuous Function : A B B A - cycle : 6 A B B A - cycles : User-defined number of A B B A - cycles ( 0 ) : Mass comparison method : in progress. Continuous function! Start the measurement sequence as described above. When the ": Please confirm by reselecting function" message is displayed, press the number key "" and hold it until the message : in progress. Continuous function! is displayed. The measurement sequence then starts. Press reset key to stop a measurement sequence. 9

20 Starting the Mass Comparison ) when the start delay time is setting, setting: see on page make sure that the load alternator is in the initial raised position, before placing a weight on the load alternator platform,. See the section entitled "Vertical Movement." Center the reference and test weights within the marking on the load alternator platform. Always position the reference weight on position mark "N." Before start the mass comparison, The load alternator platform, with the reference weight on the position "N" must be in position above the weighing system pan. See the section entitled "Horizontal Movement." Select the procedure for mass comparison Select the "Mass comparison cycles" item from the main menu. : Mass comparison cycles : Move vertically : Move horizontally : Configure Press the number key "". A new window will displayed. select : Please choos start option: : At once : After 0 minutes : Set start delay start the mass comparison without start delay time Press the number key "" follow the section entitled: starting the mass comparisson without start delay time start the mass comparison with setting start delay time, Press the number key "" and follow the next step below Press the number key "" to change the start delay time. : A B B A - cycle : 6 A B B A - cycles : User-defined number of A B B A - cycles ( 0 ) : Mass comparison method The start menu window will displayed. If you have press the number key "" Select the "Mass comparison method" item first. Press the number key "". 0

21 Select the desired method of mass comparison: e.g.: : Mass comparison of two weights : Mass comparison of two weights : Mass comparison of four weights : All 6 comparisons (N- / - / N- / - / N- / -) : Mass comparison cycles : in progress. Press the number key "". ": in progress" is short displayed Press the number key "" to return to the start menu. When the start menu is displayed, press the number key that corresponds to the desired number of cycles to start the mass comparison procedure: : A B B A - cycle : 6 A B B A - cycles : User-defined number of A B B A - cycles ( 0 ) : Mass comparison method e.g.: 6 A B B A cycles, press the number key "". The message : Please confirm by re-selecting function is displayed. : A B B A - cycle : 6 A B B A - cycles : User-defined number of A B B A - cycles ( 0 ) : Mass comparison method : Please confirm by re-selecting function. Confirm the selected mass comparison sequence, or cancel by press another number key, or press the Sartorius logo to return to the main menu. Confirm the selected mass comparison sequence, press the number key "" again to start the measurement.

22 : A B B A - cycle : 6 A B B A - cycles : User-defined number of A B B A - cycles ( 0 ) : Mass comparison of two weights in progress. 0 A number will displayed in the window. e. g. 0 This number, identicate the setting start delay time, in minute The time cont down to 0 before the mass comparison procedure start automatically The mass comparison starts The printer begins by printing the current time and date, the mass comparator model designation, the number of A B B A cycles and the number of weights. The load alternator begins by centering the weights. The first weight to be tested will centered first. as following: e.g. weights on the load alternator. The Testweight is on position. The load alternator platform rotates 80 so that the weight of position is above the mass comparator pan. The centering begins. The weight will loaded and unloaded two times on the weighing pan. The Centermatic center the weight in the center position. The load alternator platform rotates 80 back so that the weight of position N is above the mass comparator pan The same procedure for centering N start. One ABBA - cycles end the centering procedure.

23 Continuous Function With the continuous function option, you can repeat complete measurement cycles continuously. to stop the continuous function. Press the reset key Note: Do not press the reset key while the load alternator platform is in motion! Starting the Continuous Function : A B B A - cycle : 6 A B B A - cycles : User-defined number of A B B A - cycles ( 0 ) : Mass comparison method : in progress. Continuous function! Start the measurement sequence as described above. When the ": Please confirm by reselecting function" message is displayed, press the number key "" and hold it until the message : in progress. Continuous function! is displayed. The mass comparison starts. Press the reset key to stop the continous function,

24 Data Interface ( Aux ) General: The individual measurements printed out during the measurement process are also available as an ASCII string at the data interface (Aux) and can be received by the PC for further processing, The measurements are always output in a couple of a combined data string. Two different formats selectable. (refer configurations page 5 and results via AUX on page ). Results via AUX:. ( values only ) = data output of measurements values only. ( values and time ) = data output of measurements values with numbers of ABBA - cycles and time. Values only The measurements are always output in a couple of a combined data string comprising the measurements A and B or B and A (separated by a space). The number of characters to be transferred is always. The data output is as follows: Measurement A, space, measurement B and CR (carriage return no CR/LF). A line feed signal (LF) must be generated each time after data is received from the PC so that the next measured values to be received appear in a new line. Otherwise, the new values will overwrite the former values. Data output example: CCE000S-L Measurement A: , Measurement B: Data string: CR ± sign for measurement A (one space if sign is + ) - 0 measurement value A with decimal point space ± sign for measurement B (one space if sign is + ) - measurement value B with decimal point carriage return The Data string is characters long and to partition as follows. character = measurement A (B) character = measurement B (A) An example for receiving data on PC from the mass comparator and sorted in an excel file. PC receive: PC Comparator measurement: measurement A measurement B generated:. Measurement A: 0,000, B: 0, ,000 0,00056 LF. Measurement B: 0,000 59, A: 0, ,0006 0,00059 LF. Measurement A: 0,000 5, B: 0, ,0005 0,00057 LF. Measurement B: 0,000 59, A: 0,000 0,000 0,00059 LF 5. A: B:

25 values and times This means that the data string consists of the transaction counter (number of the ABBA cycle), time of value A, value A, time of value B and value B, or these data for B and A, respectively. The calculations of the differences and the final result does not belong to the output. The controller ( control unit ) recognizes different numbers of characters for the measured values so that the data string is correctly sorted with one setting on all automatic mass comparators model. The Data string is characters long and to partition as follows. The partition of the data string is follows: characters = number characters = space 8 characters = time measurement A (B) 0 characters = measurement A (B) characters = space 8 characters = time measurement B (A) 0 characters = measurement B (A) characters = space An example for receiving data on PC from the mass comparator and sorted in an excel file. PC receive: PC Comparator measurement: number measurement measurement time A value A time B value B generated:. Measurement A: 0,00056 B: 0,000 :56:9 0,000 :58: 0,00056 LF. Measurement B: 0,00056 A: 0,0006 5:0:09 0,0006 5:00:6 0,00056 LF. Measurement A: 0,0005 B: 0, :0:0 0,0005 5:05:55 0,00057 LF. Measurement B: 0,00059 A: 0,000 5:09: 0,000 5:07:9 0,00059 LF 5. Measurement A: 0,0008 B: 0, ::6 0,0008 5::9 0,00060 LF 6. Measurement B: A: Connecting cable to PC The two data protocols for the transfer require a semi duplex interface with transmit and receiver lines. As no handshake is required, the relevant handshake lines must be bridged. IBM PC Control Unit female connector (9-pin) male connector (5-pin) Shield Shield -- DCD I -- RXD I TXD TXD I RXD DTR o 5 -- GND I GND DSR o 8 -- CTS I 5

26 Programming for external operation with PC Introduction The load alternator can be operated from a PC via the AUX port. Measurement procedures can be freely programmed as required. As the data protocol runs in 7 bit / ASCII format, any terminal program on the PC is suitable for testing the functions. Interface parameters: 9600 Baud EVEN Parity start bit 7 data bits stop bit no handshake, local echo must be switched off Linking the PC to the control unit Switch on the control unit of the load alternator. Make sure that the main menu is displayed on the controller; if necessary key to the Sartorius logo. To link the PC to the control unit, send a CR ( Carriage Return only no CR/LF ) at intervals until the alternator responds with ready for operation??. The control unit display will bring up the message System controlled externally. This message will remain on the display as long as the external control condition persists. The built-in printer is disabled. On the PC monitor, two question marks?? will be displayed indicating the system s readiness to receive commands. If further data (parameters, confirmations) are required, this will be indicated by the appearance of a single question mark?. All command codes consist of a single letter or numeral. (No Carriage Return CR must be sent). The alternator sends back the character it receives (echo) by way of confirmation. Once the command has been executed,?? will be displayed on the screen. External control commands Vertical movements Locating top starting position. Lowering the centering position (middle position). a) Lowering to mass comparator: If load changer centred. Queries with?. - 0 confirms to lowering. b) Locating centering position: If load changer in weighing position. Executes command immediately it is entered. D (Down) Lowering to mass comparator. Executes command immediately it is entered. a) Lowering from the top starting position. b) Lowering from the centering position. Horizontal movements H Rotate 80 clockwise. Rotate 80 counter-clockwise. (CC50000S: Change the horizontal end position). 9 Rotate 90 clockwise. (CC50000S: Change the horizontal end/centre position). N Rotate 90 counter-clockwise.

27 Automatic weighing 5 x A-B-B-A cycle. Queries with?. - Confirms to start with x A-B-B-A cycles. Queries with?. a) x 6 x A-B-B-A : - Confirms to start with 0. b) Permanent function 6 x A-B-B-A : - Enter 6 repeatedly. Queries with?. - Confirms to start with x A-B-B-A cycles freely selectable. Queries with? Enter required number of A-B-B-A measurements. Note! The number is entered as a single character whose ASCII value corresponds to the actual number required. (Enter in the terminal program: if applicable, using ALT and number on the keypad). - Executes command immediately it is entered. Commands without movement 0 (Zero) Set evaluation time in seconds Queries with? Enter evaluation time. Note! The number is entered as a single character whose ASCII value corresponds to the actual number required. (Enter in the terminal program: if applicable, using ALT and number on the keypad). 8 Compare mass of only weights. Entry necessary with only weights before start. A R V (All 6 comparisons) Mass comparison all 6 combinations. Only possible when weights are used. When required, entry before measurement starts. (READ) Take weight value from the mass comparator display. Possible only when weighing is in progress and the weight is shown with the stability symbol g. The weight is transferred after a delay, determined by the evaluation time entered. Note! The controller reports error and then stops the system if no correct weight is available. (Version) indicates the comparator model and the software version. Error messages NA (not available) Function not available from current position. SE (scale error) No correct weight available, control blocked. PE (position error) No response from position switch: alternator blocked. The measurements are sent to the PC as a couple of values an ASCII string. The measurements A and B are separated by: a space (hexadecimal 0), and end with: CR (hexadecimal 0D) A line feed signal (LF) must be generated each time after data is received from the PC so that the next measured values to be received appear in a new line. Otherwise, the new values sent will overwrite the former values. 7

28 Troubleshooting Error message: : Mass comparison cycles : Move vertically : Move horizontally : Configure : Function not possible from this position! : Function not possibile from this position! If you press the number key "" : Cause: The load alternator platform is in the centering or lowered position. Solution: Move the platform to the initial raised position. Press the number key "" and then press the key "" while the submenu is displayed. Error message: Mass comparison cycles : Move vertically : Move horizontally : Configure : Function not possible from this position! : Function not possible from this position! If you press the number key "" : Cause: The load alternator platform is in the centering or lowered position. Solution: Move the platform to the initial raised position. Press the number key "" and then press the key "" while the submenu is displayed. 8

29 Error message: No correct value available! Please check weights and number loaded - then press the reset button. After a measurement has begun: During measurement: Cause: The mass comparator has not been turned on. The controller is not connected to the mass comparator. The menu has not been configured for data transfer. You have started a measurement with weights. Data is not transferred to the controller because the display is not stable; i.e., the stability symbol is not displayed. Solution: Turn on the mass comparator. Connect the cable. Check the menu settings in the mass comparator. Select the comparison method before starting a measurement sequence. See "Starting a Measurement." Check for and correct unfavorable ambient conditions; e.g. vibrations, drafts or air turbulence, extreme temperature fluctuations. Error message: Initial System Startup Important! First raise load alternator to upper position, then load the weights. Press anywhere to access the main menu. After turning on the mass comparator or during measurement: Cause: The load alternator platform is in the centering or lowered position. The power supply has been interrupted. The reset key was pressed. Solution: Move the platform to the initial raised position. Restart measurement. 9

30 Maintenance Replacing the Paper Roll Turn on the controller. Lift up the printer cover. Locate the small lever to the right of the paper transport roll and move it towards the front. Insert the start of the paper roll with the inside towards the front into the retainer behind the paper transport roll and move the lever back to its original position. The paper is drawn in automatically. Press the grey push button to transport more paper, if is necessary. Lift back the printer cover. Order no.: 69 C 000 (set of 5 roll thermo paper) Note: If the paper roll in the printer runs out in the middle of a measuring sequence, the printer switches off. The individual values measured and the results calculated after the printer shuts off are stored in memory. With 6 x A B B A selected, for example, the memory can store data from up to 6 complete measurement. When the memory is full, this is indicated by a crossed-out printer symbol in the display. The load alternator continues according to the selected function sequence. Subsequent measurement data are not stored. To retain the data stored in memory: - Do not interrupt the measurement sequence - Do not press reset - Do not turn off the controller Place a new roll of paper in the printer as described above. When you return the lever to its original position, the stored data are printed automatically. If the printer symbol was not yet displayed, the subsequent measured data are printed automatically as before. If the printer symbol is displayed, it remains displayed after the stored data are printed. Subsequent data are neither printed nor stored in memory. Press reset to cancel the measurement sequence. The printer symbol is no longer displayed. You can now start a new measurement. 0

31

32 Sartorius Lab Instruments GmbH & Co. KG Weender Landstrasse Goettingen, Germany Phone Fax Copyright by Sartorius Lab Instruments GmbH & Co. KG, Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius. The status of the information, specifications and illustrations in this manual is indicated by the date given below. Sartorius reserves the right to make changes to the technology, features, specifications and design of the equipment without notice. Status: April 0, Sartorius Lab Instruments GmbH & Co. KG, Goettingen, Germany Specifications subject to change without notice. Printed in the EU on paper bleached without chlorine. HA_Controller for CC/CCE...-L KT Publication No.: WCC60-e005

Sartorius Comparator. Interface Description for the CC Model Series

Sartorius Comparator. Interface Description for the CC Model Series 98647-000-53 Sartorius Comparator Interface Description for the CC Model Series Contents Page General Information 4 General Specifications 5 Data Output Formats 6 Data Input Formats 11 Synchronization

More information

Operating Instructions. Sartorius. Description of the Interface for EA, EB, GD, GE and TE Balances/Scales

Operating Instructions. Sartorius. Description of the Interface for EA, EB, GD, GE and TE Balances/Scales Operating Instructions Sartorius Description of the Interface for EA, EB, GD, GE and TE Balances/Scales Contents Intended Use 2 Contents 2 Intended Use 2 Configuring the Interface Parameter Settings (Menu)

More information

Operating Instructions Sartorius

Operating Instructions Sartorius Operating Instructions Sartorius Description of the Interface for GD, GE and TE Balances/Scales Contents Intended Use 2 Contents 2 Intended Use 2 Configuring the Interface Parameter Settings (Menu) 3 Data

More information

Sartorius Research. R 160 P Electronic Semi-Microbalance

Sartorius Research. R 160 P Electronic Semi-Microbalance Sartorius Research. R 160 P Electronic Semi-Microbalance Installation and Operating Instructions 1 Power receptacle, fuse, 8 Level indicator voltage selector 9 ON/OFF key 2 Menu access switch 10 Print

More information

Table of Contents. Laboratory Balances. Instruction Manual - TORBAL AGC Series

Table of Contents. Laboratory Balances. Instruction Manual - TORBAL AGC Series Laboratory Balances Table of Contents Cautionary Notes and Precautions...3 1. Specification... 4 2. Keys and Display Indicators... 6 3. Commands and abbreviations...7 4. Parts Description...8 5. Unpacking

More information

AD-8923-BCD. Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD

AD-8923-BCD. Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD AD-8923-BCD Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD4002137 2010 A&D Company, Limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated

More information

Application Guide for the Secura, Quintix and Practum Series of Models

Application Guide for the Secura, Quintix and Practum Series of Models Application Guide for the Secura, Quintix and Practum Series of Models Getting Started Quickly Density Checkweighing Counting Statistics Weighing Components Mixing Weighing in Percent Peak Hold Unstable

More information

MT-SICS. Reference Manual for

MT-SICS.   Reference Manual for Reference Manual for METTLER TOLEDO Standard Interface Command Set (MT-SICS) MT-SICS 0 version 2.30 MT-SICS 1 version 2.20 MT-SICS 2 for Moisture Analyzer MJ33 version 2.30 MT-SICS 3 for Moisture Analyzer

More information

Sartorius PMA.Quality PMA7501, PMA

Sartorius PMA.Quality PMA7501, PMA Installation Instructions Sartorius PMA.Quality PMA7501, PMA7501-000 Electronic Paint-mixing Scales 98648-014-21 General View of the Equipment PMA7501 1 2 3 4 5 6 7 8 9 15 10 14 11 13 12 1 Display and

More information

Application Guide for the Secura, Quintix and Practum Series

Application Guide for the Secura, Quintix and Practum Series Application Guide for the Secura, Quintix and Practum Series Getting Started Quickly Density Checkweighing Counting Statistics Weighing Components Mixing Percentage Peak Hold Unstable Conditions Conversion

More information

BBA-Series Precision Balances

BBA-Series Precision Balances Schematic BBA-Series Precision Balances BBA-600/BBA-1200 Service Manual 11 Schematic 1 10 Troubleshooting You have purchased a quality precision weighing instrument that requires handling with care. Trouble

More information

Operating instructions. METTLER TOLEDO Bidirectional Data Interface for the J series balances. Option 018

Operating instructions. METTLER TOLEDO Bidirectional Data Interface for the J series balances. Option 018 Operating instructions METTLER TOLEDO Bidirectional Data Interface for the J series balances Option 018 2 Bidirectional Data Interface (CL/RS232C) for the J series balances, Option 018 METTLER TOLEDO J

More information

Multi-Function Smart Indicator User Manual

Multi-Function Smart Indicator User Manual Multi-Function Smart Indicator User Manual EC30006 Content 1 Notice... 2 2 Specification... 2 3 Main parameter... 3 4 How to fix... 3 5 Port... 4 6 On/Off... 4 7 Charging... 4 8 Display... 5 9 Screen keypad...

More information

Models: TD3000 Series. Table Displays. 2 by 20 character display USER MANUAL

Models: TD3000 Series. Table Displays. 2 by 20 character display USER MANUAL Models: TD3000 Series Table Displays 2 by 20 character display USER MANUAL i NOTICE The manufacturer of the POS table display makes no representations or warranties, either expressed or implied, by or

More information

CoolMaster Programmers Reference Manual (PRM)

CoolMaster Programmers Reference Manual (PRM) CoolMaster Programmers Reference Manual (PRM) CoolMaster Interface Adapter for VRV, VRF Air Conditioning Systems Cool Master 1000D Cool Master 2000S Cool Master 3000T Cool Master 4000M Version 3.0 8/12/2009

More information

Sartorius TS 6100B, TS 12 Electronic Over/Under Scales Installation and Operating Instructions

Sartorius TS 6100B, TS 12 Electronic Over/Under Scales Installation and Operating Instructions 98648-001-40 Sartorius TS 6100B, TS 12 Electronic Over/Under Scales Installation and Operating Instructions 1 Display head 2 Lock 3 Jack for AC power 4 Support arm 5 Weighing platform 6 Leveling foot 7

More information

Sartorius XQ 6100, XQ 6100 B.

Sartorius XQ 6100, XQ 6100 B. Sartorius XQ 6100, XQ 6100 B. Electronic Paint Mixing Scales Installation and Operating Instructions Sartorius XQ 6100, XQ 6100 B. Please read these installation and operating instructions carefully before

More information

Doug Fleenor Design, Inc. RS-232 to DMX512 Interface, 2 Generation. March 8, 2010 (Software V1.2)

Doug Fleenor Design, Inc. RS-232 to DMX512 Interface, 2 Generation. March 8, 2010 (Software V1.2) Doug Fleenor Design, Inc. nd RS-232 to DMX512 Interface, 2 Generation March 8, 2010 (Software V1.2) The second generation RS-232 to DMX512 interface has numerous features beyond the original device. The

More information

GxP-compliant Printing and Archiving of Initial Weights and Related Weighing Data. White Paper

GxP-compliant Printing and Archiving of Initial Weights and Related Weighing Data. White Paper GxP-compliant Printing and Archiving of Initial Weights and Related Weighing Data White Paper Contents 3 Introduction 4 Thermal Printing Processes 7 Synopsis 2 Introduction Sartorius Lab Instruments GmbH

More information

Manual DTM Digital Pressure Transmitter

Manual DTM Digital Pressure Transmitter TetraTec Instruments GmbH Gewerbestr. 8 71144 Steinenbronn Deutschland E-Mail: info@tetratec.de Tel.: 07157/5387-0 Fax: 07157/5387-10 Manual Digital Pressure Transmitter *** VERSION 1.1 *** Update: 22.11.2006

More information

Installation and operation manual ReciFlow Gas

Installation and operation manual ReciFlow Gas Installation and operation manual ReciFlow Gas 1 1. Measurement principle... 3 2. Installation... 5 3. Operation... 7 4. Electrical interfaces... 11 5. Communication protocol... 14 6. Software update and

More information

TABLE OF CONTENTS. Communication Functions

TABLE OF CONTENTS. Communication Functions TABLE OF CONTENTS Chapter 1: Chapter 2: Chapter 3: Chapter 4: General Features....................................................... 1-1 Functions......................................................

More information

IPC194. ipconv Protocol Converter

IPC194. ipconv Protocol Converter IPC194 General Operating, Maintenance and Installation Manual ipconv Protocol Converter D-91056 Erlangen Phone: +49 9131 7677 47 Fax: +49 9131 7677 78 Internet: http://www.ipcomm.de Email: info@ipcomm.de

More information

GSE Scale Systems M660 CUSTOM TRANSMIT

GSE Scale Systems M660 CUSTOM TRANSMIT M660 CUSTOM TRANSMIT A custom transmit is a sequence of characters, control codes and parameter values to be transmitted out a communication port to a peripheral device such as a printer, remote display,

More information

Practum. Start Weighing Right.

Practum. Start Weighing Right. Practum Practum. Start Weighing Right. Get best value for your money, without compromises in precision and reliability. Rely on consistent readings and excellent repeatability ensured by Sartorius quality

More information

The Sartorius Talent Series Your access to talented weighing technology

The Sartorius Talent Series Your access to talented weighing technology The Sartorius Talent Series Your access to talented weighing technology The Sartorius Talent series the affordable weigh to get started with Sartorius weighing technology The performance specifications

More information

The New Practum. Start Weighing Right.

The New Practum. Start Weighing Right. The New Practum. Start Weighing Right. The New Practum. Start Weighing Right. 3 Get best value for your money, without compromises in precision and reliability. Rely on consistent readings and excellent

More information

CyberComm Pro 2.4 Data Acquisition Software Installation & User Guide. CyberScan DO 1500

CyberComm Pro 2.4 Data Acquisition Software Installation & User Guide. CyberScan DO 1500 CyberComm Pro 2.4 Data Acquisition Software Installation & User Guide CyberScan DO 1500 Bench Dissolved Oxygen Meter Technology Made Easy... 68X292341 Rev.0 01/04 PREFACE Thank you for selecting the CyberScan

More information

Installation Manual. WKC Weigh Modules

Installation Manual. WKC Weigh Modules Installation Manual Weigh Modules Table of Contents 1 Installation 3 1.1 Mechanical installation... 3 1.1.1 Support Interface (Support Surface)... 3 1.1.2 Weighing Interface (Weighing Platform)... 3 1.2

More information

Datapanel. VT100 Terminal Emulator for Datapanels. Operator Interface Products. User's Manual

Datapanel. VT100 Terminal Emulator for Datapanels. Operator Interface Products. User's Manual Datapanel Operator Interface Products VT100 Terminal Emulator for Datapanels User's Manual GFK-1698 July 1999 Warnings, Cautions, and Notes as Used in this Publication GFL-002 Warning Warning notices are

More information

Operating instructions. METTLER TOLEDO Differential weighing software with LC-BCR Bar-Code Reader for G balances

Operating instructions. METTLER TOLEDO Differential weighing software with LC-BCR Bar-Code Reader for G balances Operating instructions METTLER TOLEDO Differential weighing software with LC-BCR Bar-Code Reader for G balances General balance settings Your will find detailed information on the general balance settings

More information

MODEL : LK-T200 Receipt Printer User s Manual

MODEL : LK-T200 Receipt Printer User s Manual SEWOO TECH CO.,LTD. Doosung BD, 689-20, Geumjeong-dong, Gunpo-si, Gyeonggi-do, 435-862, Korea TEL : +82-31-459-8200 FAX : +82-31-459-8880 www.miniprinter.com MODEL : LK-T200 Receipt Printer User s Manual

More information

X-Series System Manual Part 7: Interfaces and data communication

X-Series System Manual Part 7: Interfaces and data communication X-Series System Manual Part 7: Interfaces and data communication Version check: Version No. Date Version 3.1 April 2012 Version 3.0 April 2011 Version 2.0 October 2007 Version 1.0 March 2006 Mettler-Toledo

More information

Sartorius Expert Sartorius Master pro Strong signals for the laboratory

Sartorius Expert Sartorius Master pro Strong signals for the laboratory Sartorius Expert Sartorius Master pro Strong signals for the laboratory The Sartorius Expert Series The Sartorius Expert series the range of balances made for experts The new Sartorius Expert series offers

More information

D-500s Indicator D LED. User Instructions. English Issue September SangR. Page 2. User s manual of Sang D-500s LED

D-500s Indicator D LED. User Instructions. English Issue September SangR. Page 2. User s manual of Sang D-500s LED D-500s Indicator D - 500 LED User Instructions English 8921-004210 Issue September 2010 1 2-2 0 1 0 D 6 0 0 4 6 7 5 Page 2 D-500s Indicator Page 2 Table of Content Section 1: Introduction. 1.1 Introduction

More information

MT-SICS. Reference Manual. METTLER TOLEDO Standard Interface Command Set MT-SICS 0 version 2.2x MT-SICS 1 version 2.2x

MT-SICS. Reference Manual. METTLER TOLEDO Standard Interface Command Set MT-SICS 0 version 2.2x MT-SICS 1 version 2.2x Reference Manual METTLER TOLEDO Standard Interface Command Set MT-SICS 0 version 2.2x MT-SICS 1 version 2.2x MT-SICS 2 for Halogen Moisture Analyzers HR83 and HG63 version 2.3x MT-SICS 3 for Halogen Moisture

More information

THERMAL PRINTER SPEC BAS Basic Features. 1.) Type : PANEL Mounting or DESK top type. 2.) Printing Type : THERMAL PRINT

THERMAL PRINTER SPEC BAS Basic Features. 1.) Type : PANEL Mounting or DESK top type. 2.) Printing Type : THERMAL PRINT THERMAL PRINTER SPEC BAS - 6026 1. Basic Features 1.) Type : PANEL Mounting or DESK top type 2.) Printing Type : THERMAL PRINT 3.) Printing Speed : 25mm / SEC 4.) Printing Column : 24 COLUMNS 5.) FONT

More information

MODEL : PR-T25 Receipt Printer User s Manual

MODEL : PR-T25 Receipt Printer User s Manual MODEL : PR-T25 Receipt Printer User s Manual All specifications are subject to change without notice Table of Contents 1. Parts Identifications 1. Parts Identifications 3 2. Setting up the printer 4 2.1

More information

Pinnacle Series interfaces: RS232C, USB and LAN. Pinnacle balances are unique in offering three interfaces for external communication.

Pinnacle Series interfaces: RS232C, USB and LAN. Pinnacle balances are unique in offering three interfaces for external communication. Pinnacle Pinnacle Series interfaces: RS232C, USB and LAN. Pinnacle balances are unique in offering three interfaces for external communication. These will enable you to create a network of analytical laboratory

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Sartorius Laboratory. L 610, L2200S, L2200P Electronic Precision Balances Installation and Operating Instructions 1

More information

MODEL : AP-8220 U Receipt Printer User s Manual

MODEL : AP-8220 U Receipt Printer User s Manual MODEL : AP-8220 U Receipt Printer User s Manual AP-8220 U Rev.A 10/14 All specifications are subject to change without notice Table of Contents 1. Parts Identifications 1. Parts Identifications 3 2. Setting

More information

MODEL : TRP-100-II Receipt Printer User s Manual

MODEL : TRP-100-II Receipt Printer User s Manual MODEL : TRP-100-II Receipt Printer User s Manual All specifications are subject to change without notice Table of Contents 1. Parts Identifications 3 2. Setting up the printer 4 2.1 Unpacking 4 2.2 Connecting

More information

Unique Micro Design. Advanced Thinking Products. Model 164 Quad Coupler User Manual

Unique Micro Design. Advanced Thinking Products. Model 164 Quad Coupler User Manual Advanced Thinking Products Model 164 Quad Coupler User Manual Document reference : UMD Part Number : 6-0164-993-4 Issue 1 - First release Revision 2/12/92 Advanced Thinking Products Model 164 Quad Coupler

More information

GENIUS Innovative in Function and Design

GENIUS Innovative in Function and Design GENIUS Innovative in Function and Design GENIUS. So N.I.C.E. So F.A.S.T. GENIUS Its functional design sets it apart. GENIUS offers the highest level of ergonomic operation for optimal functionality. Designed

More information

Fics-RT1. User s Manual

Fics-RT1. User s Manual Fics-RT1 User s Manual Contents 1: Overview... 1 2: Terminal Specifications... 1 2-1: Standard Specifications... 1 2-2: Customer Defined Features:... 1 3: Terminal RS232C Communication Link Details...

More information

Operating instructions METTLER TOLEDO BD balances

Operating instructions METTLER TOLEDO BD balances Operating instructions METTLER TOLEDO BD balances 0.00 g Startup procedure 1 Unpacking, standard equipment The standard equipment contains: BD balance, Weighing pan, permanently mounted with BD6000, AC

More information

Reference Guide METTLER TOLEDO WMH Weighing Platforms

Reference Guide METTLER TOLEDO WMH Weighing Platforms Reference Guide METTLER TOLEDO Weighing Platforms Mettler-Toledo GmbH 2005 Doc.No. KA021, 42908713C Weighing Platforms Reference Guide Contents 1 Initial Setup... 4 1.1 Installation... 4 1.2 Electrical

More information

Explorer Pro Instruction Manual

Explorer Pro Instruction Manual Explorer Pro Instruction Manual * WEIGHING 01/01/03 03:05:45: PM UNITS 2nd UNITS 0g 0.0 Compliance Compliance to the following standards is indicated by the corresponding mark on the product. Marking

More information

MRZJW3- SETUP154E. General-Purpose AC Servo Servo Configuration Software MODEL INSTALLATION GUIDE

MRZJW3- SETUP154E. General-Purpose AC Servo Servo Configuration Software MODEL INSTALLATION GUIDE General-Purpose AC Servo Servo Configuration Software MODEL MRZJW3- SETUP154E INSTALLATION GUIDE Thank you for choosing the Mitsubishi general-purpose AC servo Servo Configuration Software. To optimize

More information

ACS-LCD-128x64. LCD Graphic Display Terminal. General Description. Features. Typical Applications. Specifications. 22-Feb-08

ACS-LCD-128x64. LCD Graphic Display Terminal. General Description. Features. Typical Applications. Specifications. 22-Feb-08 6 2 3 3 E. S a w g ra s s R d S a ra s o ta, F L. 3 4 2 4 0 (9 4 1 )3 7 7-5 7 7 5 F A X(9 4 1 )3 7 8-4 2 2 6 www.acscontrol.com 22-Feb-08 ACS-LCD-128x64 LCD Graphic Display Terminal General Description

More information

MODEL : SLK-TE20X Series Receipt Printer User s Manual

MODEL : SLK-TE20X Series Receipt Printer User s Manual MODEL : SLK-TE20X Series Receipt Printer User s Manual TE20X Series Rev.F 03/16 All specifications are subject to change without notice Table of Contents 1. Parts Identifications 1. Parts Identifications

More information

The New Practum. Start Weighing Right.

The New Practum. Start Weighing Right. The New Practum. Start Weighing Right. Rugged Weigh Cell for Reliable Results Every Time, All the Time High-resolution, overload protected ed weigh cells, made in Germany, ensure durable and reliable performance

More information

Sartorius MP8 Series Balances and Scales

Sartorius MP8 Series Balances and Scales Sartorius MP8 Series Balances and Scales Making Changes to the Balance Operating Program 1) The Program Locking Switch Before any changes can be made to the balance operating program you must first unlock

More information

MOBILE THERMAL PRINTER

MOBILE THERMAL PRINTER MOBILE THERMAL PRINTER MODEL CMP-20 series Technical Manual Rev. 1.00 TABLE OF CONTENTS 1. General Specifications 1.1 Printing Specifications 1.2 Character Specifications 1.3 Paper Specification 1.4 Printable

More information

USER S MANUAL. PH232Ex1. #1 RS-232 Serial Port to Ethernet, Terminal Server/Client. Doc No: PH232Ex1-UM-001 IPEX. (IP Electronix)

USER S MANUAL. PH232Ex1. #1 RS-232 Serial Port to Ethernet, Terminal Server/Client. Doc No: PH232Ex1-UM-001 IPEX. (IP Electronix) USER S MANUAL PH232Ex1 Doc No: PH232Ex1-UM-001 #1 RS-232 Serial Port to Ethernet, Terminal Server/Client IPEX (IP Electronix) Contents 1. INTRODUCTION... 3 2. SPECIFICATIONS... 3 3. PACKAGE CHECKLIST...

More information

CyberComm Pro Data Acquisition Software Installation & User Guide

CyberComm Pro Data Acquisition Software Installation & User Guide CyberComm Pro 2.2.3 Data Acquisition Software Installation & User Guide ph 1100 and ph 2100 Bench ph and Bench ph/ion Meter Technology Made Easy... 68X090822 rev 1 Aug 2002 2 PREFACE Thank you for selecting

More information

MODEL : LK-T210 Receipt Printer User s Manual

MODEL : LK-T210 Receipt Printer User s Manual SEWOO TECH CO.,LTD. Doosung BD, 689-20, Geumjeong-dong, Gunpo-si, Gyeonggi-do, 435-862, Korea TEL : +82-31-459-8200 FAX : +82-31-459-8880 www.miniprinter.com MODEL : LK-T210 Receipt Printer User s Manual

More information

User s Manual Models CD5220 / CD6220. Vacuum Fluorescent Customer Display

User s Manual Models CD5220 / CD6220. Vacuum Fluorescent Customer Display User s Manual Models CD5220 / CD6220 Vacuum Fluorescent Customer Display INDEX 1. FEATURES...3 2. TYPE CLASSIFICATION...4 3. GENERAL SPECIFICATIONS...5 4. INTERFACE SPECIFICATIONS...6 5. FUNCTION SELECTION...9

More information

CUSTOM MANUFACTURED IN U.S.A. BY RADIO SHACK, A DIVISION OF TANDY CORPORATION CAT. NO

CUSTOM MANUFACTURED IN U.S.A. BY RADIO SHACK, A DIVISION OF TANDY CORPORATION CAT. NO CUSTOM MANUFACTURED IN U.S.A. BY RADIO SHACK, A DIVISION OF TANDY CORPORATION CAT. NO. 26-6050 TRS-80 DT-1 OWNER'S MANUAL Contents Introduction 1 1/ Description of the DT-1 3 2/ Setting Up the DT-1 5

More information

DS-800. ::: Receipt Printer User s manual :::

DS-800. ::: Receipt Printer User s manual ::: DS-800 ::: Receipt Printer User s manual ::: All specifications are subjected to change without notice TABLE OF CONTENTS 1. Parts Identifications 2 2. Setting up the printer 3 2.1 Unpacking 3 2.2 Connecting

More information

PRODUCT INFORMATION. MPP5500 Series PANEL-MOUNT PRINTERS

PRODUCT INFORMATION. MPP5500 Series PANEL-MOUNT PRINTERS PRODUCT INFORMATION Series PANEL-MOUNT PRINTERS Applications Datasheet Features Easy load paper feature High speed, high resolution printing 5Vdc, 10-35Vdc or low power Quiet, non-impact system Maintenance-free

More information

MultiView Matrix 8x8 CAT5 Switch

MultiView Matrix 8x8 CAT5 Switch PRELIMINARY MultiView Matrix 8x8 CAT5 Switch Quick Reference & Setup Guide Magenta Research Ltd 934B Federal Road, Brookfield, CT 06804 USA (203) 740-0592 FAX (203) 740-0596 www.magenta-research.com Magenta

More information

THT-500-A/R (C5) 1. System Configuration COMMUNICATION INSTRUCTION MANUAL. No. THT5CE

THT-500-A/R (C5) 1. System Configuration COMMUNICATION INSTRUCTION MANUAL. No. THT5CE COMMUNICATION INSTRUCTION MANUAL THT-500-A/R (C5) No. THT5CE2 2018.04 This manual contains instructions for communication functions of the THT-500-A/R. 1. System Configuration 1.1 When Using USB Communication

More information

Operating Instructions SECUTEST SI /12.10

Operating Instructions SECUTEST SI /12.10 Operating Instructions SECUTEST SI+ 3-349-613-15 1/12.10 6 5 4 3 2 1 MENU SECUTEST SI+ 7 6 8 without Function 9 10 11 12 13 14 Standard Equipment 1 Input and memory module SECUTEST SI+, 1 USB connector

More information

MODEL : TRP100-III Receipt Printer User s Manual

MODEL : TRP100-III Receipt Printer User s Manual MODEL : TRP100-III Receipt Printer User s Manual All specifications are subject to change without notice A software tool is available to configure the printer settings on AURES technical website: www.aures-support.fr

More information

COMPLETE MANUAL. WEIGHING INDICATOR TI-1200 and TI-1200-S TI-1200 with rechargeable battery.

COMPLETE MANUAL. WEIGHING INDICATOR TI-1200 and TI-1200-S TI-1200 with rechargeable battery. COMPLETE MANUAL WEIGHING INDICATOR TI-1200 and TI-1200-S TI-1200 with rechargeable battery. TI-1200-S without rechargeable battery and with a smaller cabinet. Note: When TI-1200 is shipped as a part of

More information

AD Leak Tester INSTRUCTION MANUAL 1WMPD

AD Leak Tester INSTRUCTION MANUAL 1WMPD AD-1690 Leak Tester INSTRUCTION MANUAL 1WMPD4002043 2009 A&D Company Ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in

More information

CONTENTS. INTRODUCTION 2 HOW IT WORKS 3 USING THE OPTIONS MENU 6 THE MENU TREE 7 VIEWING DATA 8 PRINTING DATA 9 ENGINEER SET-UP FACTORY SET-UP

CONTENTS. INTRODUCTION 2 HOW IT WORKS 3 USING THE OPTIONS MENU 6 THE MENU TREE 7 VIEWING DATA 8 PRINTING DATA 9 ENGINEER SET-UP FACTORY SET-UP CONTENTS. INTRODUCTION 2 HOW IT WORKS 3 USING THE OPTIONS MENU 6 THE MENU TREE 7 VIEWING DATA 8 PRINTING DATA 9 SET-UP SETTINGS 11 CALIBRATE ZERO 14 CLEAR DATA LOG 15 ENGINEER SET-UP SETTINGS 16 PRINT

More information

MODEL : LK-TL200 Receipt Printer User s Manual

MODEL : LK-TL200 Receipt Printer User s Manual SEWOO TECH CO.,LTD. 28-6, Gajangsaneopdong-ro, Osan-si, Gyeongi-do, 447-210, Korea TEL : +82-31-459-8200 FAX : +82-31-459-8880 www.miniprinter.com MODEL : LK-TL200 Receipt Printer User s Manual TL200 Rev.

More information

Models: LD9000 Series. Customer Displays. 2 by 20 character display USER MANUAL

Models: LD9000 Series. Customer Displays. 2 by 20 character display USER MANUAL Models: LD9000 Series Customer Displays 2 by 20 character display USER MANUAL i NOTICE The manufacturer of the POS pole display makes no representations or warranties, either expressed or implied, by or

More information

AD Remote Controller INSTRUCTION MANUAL WM+PD A

AD Remote Controller INSTRUCTION MANUAL WM+PD A AD-8922 Remote Controller INSTRUCTION MANUAL WM+PD4000929A 2005 A&D Company Ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language

More information

Precision Balance W3200 Series Operating Manual

Precision Balance W3200 Series Operating Manual Precision Balance W3200 Series Operating Manual Version 9-19-16 1 2 CONTENTS 1. INTRODUCTION. 4 2. INSTALLATION... 4 2-1. Unpacking.... 5 2-2. Specifications.... 5 2-3. Environmental Requirements... 5

More information

Printer Control Command Set

Printer Control Command Set Printer Control Command Set Technical Manual Revision: November 1, 2017 Omniprint Inc. 1923 East Deere Ave., Santa Ana, California 92705, U.S.A. T: 949.833.0080 :: F: 949.833.0040 www.omniprintinc.com

More information

DSP860 Operation Manual

DSP860 Operation Manual DSP860 Operation Manual 1. Information 2. Introduction 3. Installation 4. Pin Assignment 5. DIP Switch Setting 6. Character Tables 7. Software Control (Command List) 8. Specifications 9. Instruction of

More information

USB Interface Instruction Manual

USB Interface Instruction Manual RANGER 3000 RANGER COUNT 3000 VALOR 7000 USB Interface Instruction Manual 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com USB

More information

Mounting Dimensions / Viewing 2 Mounting Options 3. Wiring Configuration 4. Quick Set up Procedure 5. Changing Intensity 6.

Mounting Dimensions / Viewing 2 Mounting Options 3. Wiring Configuration 4. Quick Set up Procedure 5. Changing Intensity 6. Section Mounting Dimensions / Viewing 2 Mounting Options 3 Section 2 Wiring Configuration 4 Section 3 Quick Set up Procedure 5 Section 4 Changing Intensity 6 Section 5 Option Summary 7 Section 6 Option

More information

ENGINEERING G. M. B. H. ENGINEERING MAKES THE DIFFERENCE

ENGINEERING G. M. B. H. ENGINEERING MAKES THE DIFFERENCE ENGINEERING G. M. B. H. ENGINEERING MAKES THE DIFFERENCE CLERMONT-FERRAND-ALLEE 36 D-93049 REGENSBURG TEL 0941/29638 0, FAX 90 EMAIL info@bauer-eng.de INTERNET http://www.bauer-eng.de POF MPX (Plastic

More information

DSP840 Operation Manual Version 3.0

DSP840 Operation Manual Version 3.0 DSP840 Operation Manual Version 3.0 1. Information 2. Introduction 3. Installation 4. Pin Assignment 5. DIP Switch Setting 6. Character Tables 7. Software Control (Command List) 8. Specifications 9. Instruction

More information

FEC-80T Receipt Printer User s Manual

FEC-80T Receipt Printer User s Manual FEC-80T Receipt Printer User s Manual All specifications are subject to change without notice TABLE OF CONTENTS 1. Parts Identifications 3 2. Setting up the printer 4 2.1 Unpacking 4 2.2 Connecting the

More information

MODEL : SLK-TL100 Receipt Printer User s Manual

MODEL : SLK-TL100 Receipt Printer User s Manual J. STEPHEN Lab., Ltd. 28-6, Gajangsaneopdong-ro, Osan-si, Gyeongi-do, 447-210, Korea TEL : +82-31-459-8200 FAX : +82-31-459-8880 www.miniprinter.com MODEL : SLK-TL100 Receipt Printer User s Manual TL100

More information

Model 7620U. Bench Scale. User s Manual

Model 7620U. Bench Scale. User s Manual Model 7620U Bench Scale User s Manual CAUTION Risk of electrical shock. Do not remove cover. No user serviceable parts inside. Refer servicing to qualified service personnel. Weigh-Tronix reserves the

More information

805HP. Handheld Digital Weight Indicator Operations Manual (V1612) Anyload Transducer Co. Ltd Website:

805HP. Handheld Digital Weight Indicator Operations Manual (V1612) Anyload Transducer Co. Ltd Website: 805HP Handheld Digital Weight Indicator Operations Manual (V1612) Anyload Transducer Co. Ltd Website: www.anyload.com Email: info@anyload.com TABLE OF CONTENTS 1. Introduction and Product Features 3 2.

More information

RTD-500 Precision RTD Simulator. Operations Manual.

RTD-500 Precision RTD Simulator. Operations Manual. RTD-500 Precision RTD Simulator Operations Manual. Page 1 of 16 Table of Content. Table of Content....1 1 Scope...3 2 Package Content....3 3 Technical Data....3 4 Preparation for use....4 4.1 Switching

More information

Subset for HB43 MT-SICS. Reference Manual for

Subset for HB43 MT-SICS. Reference Manual for 0-100 Reference Manual for METTLER TOLEDO Standard Interface Command Set (MT-SICS) MT-SICS 0 version 2.30 MT-SICS 1 version 2.20 MT-SICS 2 for Halogen Moisture Analyzer version 2.30 MT-SICS 3 for Halogen

More information

805HP. Handheld Digital Weight Indicator Operations Manual (V1612) Anyload Transducer Co. Ltd Website:

805HP. Handheld Digital Weight Indicator Operations Manual (V1612) Anyload Transducer Co. Ltd Website: 805HP Handheld Digital Weight Indicator Operations Manual (V1612) Anyload Transducer Co. Ltd Website: www.anyload.com Email: info@anyload.com TABLE OF CONTENTS 1. Introduction and Product Features 3 2.

More information

Hardware Manual. PCMCIA 1 Port RS EDITION APRIL 1999

Hardware Manual. PCMCIA 1 Port RS EDITION APRIL 1999 Hardware Manual 232 232232 PCMCIA 1 Port RS232 1.1 EDITION APRIL 1999 Guarantee. FULL 36 MONTHS GUARANTEE. We guarantee your interface card for a full 36 months from purchase, parts and labour, provided

More information

CAD-CARE TROUBLESHOOTING GUIDE

CAD-CARE TROUBLESHOOTING GUIDE CAD-CARE TROUBLESHOOTING GUIDE CAD-Care is a stable and error free system. The biggest problem encountered with CAD-Care is when something stops CAD-Care during a system sort. Windows Screen Savers have

More information

AMT203 ABSOLUTE ENCODER Demo Kit User Guide DISCONTINUED

AMT203 ABSOLUTE ENCODER Demo Kit User Guide DISCONTINUED AMT203 ABSOLUTE ENCODER Demo Kit User Guide PART NUMBER: AMT203 AMT-DMBD Demo Board DESCRIPTION: date 2 of 13 DEMO KIT USER GUIDE The AMT-DMBD Demo Board is designed for testing and configuring the AMT

More information

Integrity Instruments

Integrity Instruments Integrity Instruments P.O. Box 451 Order Phone 800-450-2001 Pine River Minnesota Fax Phone 218-587-3414 56474 USA Tech Phone 218-587-3120 http://www.integrityusa.com 232M300C-CNT 232M300CT-CNT Counter

More information

InfoTag KE28xx Communications for 186 CPU Firmware Version 4

InfoTag KE28xx Communications for 186 CPU Firmware Version 4 InfoTag KE28xx Communications for 186 CPU Firmware Version 4 *KE28xx models include: KE2800, KE2852, KE2853, KE2856 This document applies to printer firmware versions 4.x only. Note that changes made to

More information

IDEA. User s Guide. Part No IDE02 Issue 1-0, April 1995 Printed in U.S.A. (183)

IDEA. User s Guide. Part No IDE02 Issue 1-0, April 1995 Printed in U.S.A. (183) IDEA User s Guide Part No. 17500IDE02 Issue 1-0, April 1995 Printed in U.S.A. (183) 4 FOREST PARKWAY, SHELTON, CONNECTICUT 06484 TEL: 203-926-5400 FAX: 203-929-0535 This manual has been developed by Nitsuko

More information

AD Leak Tester INSTRUCTION MANUAL 1WMPD B

AD Leak Tester INSTRUCTION MANUAL 1WMPD B AD-1690 Leak Tester INSTRUCTION MANUAL 1WMPD4002043B 2009 A&D Company Ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language

More information

AD-8923-CC. Remote Controller (CC-Link) INSTRUCTION MANUAL 1WMPD B

AD-8923-CC. Remote Controller (CC-Link) INSTRUCTION MANUAL 1WMPD B AD-8923-CC Remote Controller (CC-Link) INSTRUCTION MANUAL 1WMPD4002124B 2013 A&D Company, Limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated

More information

VFD Series Vacuum Fluorescent Customer Display Instruction Manual ESC/POS commands Sets Supports

VFD Series Vacuum Fluorescent Customer Display Instruction Manual ESC/POS commands Sets Supports VFD Series Vacuum Fluorescent Customer Display Instruction Manual ESC/POS commands Sets Supports Ver. 1.1 CONTENTS I. PRODUCT OVERVIEW... 3 Specification... 4 Package... 8 II. INSTALLING YOUR VFD CUSTOMER

More information

ISDA/ISDA4 Protocol Driver Manual. Table of Contents

ISDA/ISDA4 Protocol Driver Manual. Table of Contents ISDA/ISDA4 Protocol Driver Manual Table of Contents ISDA 1 Functional Overview... 3 1.1 Master Serial Port(s)... 3 1.2 Module Internal Database... 4 1.2.1 ISDA Serial Port Driver Access to Database...

More information

Customer Pole Display. Model: WD-304 Version: 1.12

Customer Pole Display. Model: WD-304 Version: 1.12 Customer Pole Display Model: WD-304 Version: 1.12 INDEX 1 FEATURES... 3 2 GENERAL SPECIFICATIONS... 4 3 UNPACKING... 5 3.1 PARTS LIST (WITH BASE PCB)... 5 3.2 PARTS LIST (WITHOUT BASE PCB)... 6 4 CONFIGURATION...

More information

This document hosted by: Extension Load Cell Cable (OP-02)

This document hosted by: Extension Load Cell Cable (OP-02) This document hosted by: www.oldwillknottscales.com 17.1. Extension Load Cell Cable (OP-02) This cable is used for installing the indicator away from the base unit. This extension load cell cable is 5

More information

QUICK START GUIDE PACKAGING. 1 x 220 V/12 VDC 1A mains adapter. Accompanying documentation. 1 x K3 weight indicator

QUICK START GUIDE PACKAGING. 1 x 220 V/12 VDC 1A mains adapter. Accompanying documentation. 1 x K3 weight indicator QUICK START GUIDE PACKAGING 1 x 220 V/12 VDC 1A mains adapter. Accompanying documentation 1 x K3 weight indicator KEYBOARD & DISPLAY LCD DISPLAY Displays the weight on the scale pan. In HOLD mode, the

More information

MPP6800 series - PANEL MOUNT PRINTERS. Options

MPP6800 series - PANEL MOUNT PRINTERS. Options series - PANEL MOUNT PRINTERS Page 1 of 12 MPP6810 MPP6820 MPP6840 Introduction 5-8Vdc, 4A peak 5-8Vdc, 2A peak 10-35Vdc 24Vdc Features Easy load paper feature High resolution thermal printing 5-8Vdc standard,

More information

Calibration. RS 232 and other out puts. Q: Will a standard RS232 cable work with an Ohaus balance?

Calibration. RS 232 and other out puts. Q: Will a standard RS232 cable work with an Ohaus balance? Calibration Q: What is the difference between a Span and a Linearity calibration? A: A span calibration sets the balance at 2 points Usually zero and full capacity. A linearity calibration sets the balance

More information