1 R E L E A S E N O T E S Waters Driver Pack 4 Driver Pack 4 includes instrument drivers and firmware required to control Waters instruments, as well as helpful utility applications. The driver pack is compatible with Empower, MassLynx, and specific, third-party data systems. It is not compatible with ACQUITY UPLC IVD-marked systems. The instrument drivers and firmware in Driver Pack 4 support these ACQUITY UPLC systems: ACQUITY UPLC System ACQUITY UPLC H-Class System ACQUITY UPLC H-Class Bio System ACQUITY UPLC I-Class System ACQUITY UPLC Systems with 2D Technology Driver Pack 4 updates the June 2011 Driver Pack. It contains many new features and enhancements. See What s new in this release for detailed information about the improvements included in Driver Pack 4. For instructions about how to install Driver Pack 4, see Driver Pack 4 Installation and Configuration Guide. For information regarding the configuration of Empower Enterprise networks, see Empower 3 Installation, Configuration, and Upgrade Guide. Table of Contents What s new in this release... 2 Driver Pack 4 contents... 4 Instrument driver and firmware versions... 6 Supported Waters and third-party... 9 Defects addressed by this release Known issues in this release Anti-virus considerations Requalification guidelines Contact Waters Technical Service ACQUITY, ACQUITY UPLC, Connections INSIGHT, Empower, MassLynx, nanoacquity UPLC, THE SCIENCE OF WHAT S POSSIBLE. and Waters are registered trademarks and Auto Blend Plus, ecord, LAC/E 32, and Q-Tof are trademarks of Waters Corporation. All other trademarks are the sole property of their respective owners. Copyright 2013 Waters Corporation * * *Rev.A* Page 1 of 21
2 What s new in this release Driver Pack 4 implements new features, enhancements, and defect corrections to the instrument drivers, firmware, and utility applications included in the driver pack. New features Instrument or System Deployment Manager BSM and CM-A QSM QSM SM-FL and SM-FTN SM-FTN Console Console Utilities Description You can now run multiple driver pack versions on a single Empower Enterprise network. Previously, it was necessary to install the same driver pack on all Empower LAC/E 32 modules, Empower clients, and Citrix servers on the network. Driver Pack 4, however, can be installed on individual LAC/E 32 modules, as long as it is also installed on all of the Empower clients and Citrix servers on the network. This feature makes it possible to add new LAC/E 32 modules to the network without upgrading and requalifying the existing LAC/E 32 modules. For further information see the Driver Pack 4 Installation and Configuration Guide. New Verify Instrument Driver Files application, to verify driver installation on non-empower systems. 2D Repeat parameter added to instrument methods allows the gradient table and timed-event table to repeat throughout the run, allowing for two-dimensional chromatography. Flow rate range increased to 2.2 ml/min. Correspondingly, the flow/pressure envelope is now: psi from 0 ml/min to 1 ml/min psi to 7800 psi, linear, from 1 ml/min to 2.2 ml/min Ability to import to and export from the solvent library. New sample- compartment temperature test helps determine whether the sample heater/cooler is operating within specifications. The test can be found in the sample manager section of the console under the Troubleshoot menu. See the console s online help for a description of this test. Injection valve can now be cycled via the timed-event table, to help reduce carryover. Instrument driver versions on both the client and server, as well as the installed driver versions and firmware versions, now appear in the Help > About dialog. Maintenance summaries for the BSM, CM/CM-A, QSM, SM, SM-FL, and SM-FTN can be viewed and printed. New application allows a single workstation to switch between controlling an ACQUITY system and a nanoacquity system. Page 2 of 21
3 Enhancements Instrument or Deployment Manager Deployment Manager Multiple instruments BSM QSM QSM QSM QSM SM-FTN CM CM, CM-A Utilities Description Improved robustness and error handling (better detection, logging, visibility). Improved compatibility with the Empower Verify Files utility. Wide-area networks with latencies of up to 300 ms are now supported for the BSM, QSM, SM, SM-FL, SM-FTN, TUV and PDA. High latency networks previously caused their instrument drivers to freeze or terminate. When you run an Empower system that includes two BSMs, their order in the console tree now always matches that of the Empower system definition. Prime function enhanced by the ability to prime to a specific composition as well as apply a composition at the end of the prime. The QSM method print content is enhanced to include the buffer system details and the compositional percentages in Auto Blend Plus mode. The QSM will initiate the system s inject start to trigger any Ethernet detectors or column managers on its next intake stroke after receiving an inject-start, contact-closure, rear-panel event. This support permits operation without an ACQUITY Autosampler (SM, SM-FTN, or SM-FL) in the system, allowing for control with external devices. The QSM now supports a passive check valve option in addition to Waters Intelligent Intake Valve (i 2 Valve) operation, for use in certain applications. Autoaddition capabilities improved with the addition of the Mix Cycles and Mix Stroke Volume method parameters. The column manager post-run report is extended to include the part number, and any recorded ecord comments. The column manager now stops the solvent flow for any change in selected column position. Previously, the flow would not stop when pressure was less than 500 psi. Columns Calculator gradients can now be dragged-and-dropped to the BSM method editor. Page 3 of 21
4 Driver Pack 4 contents The Driver Pack 4 DVD includes these items: All current instrument drivers and firmware files for ACQUITY UPLC instruments and detectors Utilities for use with ACQUITY UPLC systems Driver Pack 4 Deployment Manager ACQUITY UPLC Console These supporting materials accompany the Driver Pack 4 DVD: Driver Pack 4 Installation and Configuration Guide on a documentation CD Driver Pack 4 release notes Pointer card with information about Waters website Instrument drivers and s Driver or Version 2489 Detector Detector 1.30 Binary Solvent Manager 1.60 Column Manager 1.60 Columns Calculator 1.60 Connections INSIGHT 3.2 ELS Detector (also supports 2420 and 2424) 1.40 FLR Detector 1.40 Local Console Controller 1.60 PDA Detector 1.60 Quaternary Solvent Manager 1.60 RI Detector 1.40 Sample Manager 1.60 Sample Manager-FTN 1.60 TUV Detector 1.60 Waters Pump Control Software 3.10 Waters SQ Detector 1.41 Waters TQ Detector 1.41 Deployment Manager 3.1 ICS Companion 1.60 Page 4 of 21
5 Firmware Firmware (SFT files) Version Checksum 2420 Detector x21A5BED Detector x21A5B8D Detector xBD0669D Detector xEDB36196 Binary Solvent Manager x36F220A9 Column Manager (Active) x1B55819D Column Manager x8E9D1DC0 ELS Detector x216F9370 FLR Detector xD99DEFB5 PDA Detector (1) xF7ED509A PDA Detector (2) xDCFE9340 Quaternary Solvent Manager xA9C8E138 RI Detector xAA39BF4C Sample Manager x47F616CE Sample Manager-FL x37CC5868 Sample Manager-FTN x35D5392B Sample Organizer x96FA9AB0 TUV Detector (1) xBC170FAF TUV Detector (2) xD71E0EDA Note: The installers for the ELS, FLR, RI, 2489, and 2998 detectors have been updated to address installation issues. Because their version numbers changed, they appear as upgrades during installation. Nevertheless the files they install remain identical to the currently released versions. Performing upgrades of those s is therefore harmless. Page 5 of 21
6 Instrument driver and firmware versions ACQUITY UPLC System Modules Part number Software version Firmware version ACQUITY UPLC BSM Binary Solvent Manager, ACQUITY UPLC SM , Sample Manager, Binary Solvent Manager, Sample Manager, ACQUITY UPLC HTCH N/A N/A ACQUITY UPLC Column Heater N/A N/A N/A ACQUITY UPLC CM ACQUITY UPLC Column Heater/Cooler Column Manager, N/A N/A Column Manager, v ACQUITY UPLC H-Class and H-Class Bio Systems Modules Part number Software version Firmware version ACQUITY UPLC H-Class QSM/ H-Class Bio QSM ACQUITY UPLC H-Class SM-FTN / H-Class Bio SM-FTN ACQUITY UPLC H-Class CH-A / H-Class Bio CH-A ACQUITY UPLC H-Class CM-A / H-Class Bio CM-A ACQUITY UPLC H-Class CM-Aux / H-Class Bio CM-Aux , , Quaternary Solvent Manager, Sample Manager- FTN, N/A N/A Column Manager, N/A N/A Quaternary Solvent Manager, Sample Manager- FTN, Column Manager (Active), Page 6 of 21
7 ACQUITY UPLC I-Class System Modules Part number Software version Firmware version ACQUITY UPLC I-Class BSM Binary Solvent Manager, ACQUITY UPLC I-Class SM-FTN Sample Manager- FTN, ACQUITY UPLC I-Class SM-FL Sample Manager, ACQUITY UPLC I-Class CH-A N/A N/A ACQUITY UPLC I-Class CM-A Column Manager, Detectors Binary Solvent Manager, Sample Manager- FTN, Sample Manager- FL, Column Manager (Active), Modules Compatible systems Part number Software version Firmware version ACQUITY UPLC TUV All UPLC systems TUV Detector, TUV Detector (1), ACQUITY UPLC TUV Thermally Enhanced All UPLC systems TUV Detector, TUV Detector (2), ACQUITY UPLC PDA All UPLC systems PDA Detector, PDA Detector (1), ACQUITY UPLC PDA Thermally Enhanced All UPLC systems PDA Detector, PDA Detector (2), ACQUITY UPLC epda All UPLC systems PDA Detector, ACQUITY UPLC epda Thermally Enhanced All UPLC systems PDA Detector, ACQUITY UPLC FLR All UPLC systems FLR Detector, v1.40 ACQUITY UPLC ELS All UPLC systems ELS Detector, v1.40 Waters 2420 ELS Detector Waters HPLC Systems ELS Detector, v1.40 PDA Detector (1), PDA Detector (2), FLR Detector, v1.42 ELS Detector, v Detector, v1.30 Page 7 of 21
8 Waters 2424 ELS Detector Waters HPLC Systems ELS Detector, v Detector, v1.30 Waters 2489 UV/Visible Detector Waters HPLC Systems Detector, v Detector, v1.41 Waters 2998 PDA Detector Waters HPLC Systems Detector, v1.30 SQ Detector All UPLC systems Waters SQ Detector, v1.41 TQ Detector All UPLC systems Waters TQ Detector, v Detector, v1.41 N/A N/A Other system modules Modules Compatible Systems Part Number Software Version Firmware Version Sample Organizer ACQUITY UPLC Systems ACQUITY UPLC 30-cm Column Heater/Cooler ACQUITY UPLC CH-30A ACQUITY UPLC I-Class / ACQUITY UPLC H-Class and H-Class Bio Systems Column Manager Auxiliary Units Local Console Controller SM, or SM-FTN, or SM-FL, All UPLC Systems N/A N/A ACQUITY UPLC H-Class and H-Class Bio Systems N/A N/A N/A N/A All UPLC systems Included with all Driver Pack 4 instrument drivers Sample Organizer, N/A Page 8 of 21
9 Supported Waters and third-party The table below shows which Waters and third-party operate with ACQUITY UPLC systems. System Empower 2: base release to SP H Hotfix 1 ACQUITY UPLC System (including 2D) ACQUITY UPLC H-Class System (including 2D) ACQUITY UPLC H-Class Bio System (including 2D) ACQUITY UPLC I-Class System (including 2D) Empower 3: base release to FR2 MassLynx (see list of SCNs, below) Standalone ACQUITY Console Analyst (see note, below) Yes Yes Yes Yes Version Yes Yes Yes Yes Version Yes Yes Yes Yes Version Yes Yes Yes Yes Version See also: Contact the appropriate vendor for compatibility information about other thirdparty control. The following MassLynx SCNs are compatible with this release: SCN 781, SCN 802, SCN 822, SCN 835, SCN 849, SCN 855, SCN 869, SCN 870, SCN 872, SCN 873, SCN 876, SCN 882, SCN 883, SCN 884, SCN 888, and SCN 896. Requirement: If an SQ, TQ, or 3100 detector is part of the system, you must install, at a minimum, Empower 2 Feature Release 4. Note that the Empower 2 System Suitability option is required. Note: Analyst supports 2D, but only with UPLC solvent managers that differ, such as one QSM and one BSM. Page 9 of 21
10 Defects addressed by this release General The table below lists problems that Driver Pack 4 addresses. The change request (SCR) numbers in the left-hand column identify problems that Waters monitors in its change request database. The right-hand column describes changes in Driver Pack 4. The improvements listed below take effect if you install Driver Pack 4 on both the Empower client, and on the LAC/E 32 module. If you install Driver Pack 4 on the Empower client, but run an earlier Driver Pack version on the LAC/E 32 module, some of the improvements do not take effect. SCR number Instrument or Description Console Equilibrations can once again be run in the standalone console with a CM in the system Console Pressure plots are now displayed in the units (psi, bar, kpa) selected by the user, rather than always psi Console When exporting multiple plot channels, the time base for the channels is now synchronized. Previously the time base of each channel could differ slightly, making plot comparison difficult Console The pressure units selected for display in the console now apply only to the current Windows user. Previously the pressure unit selection applied to all Windows users Deployment Manager The Local Console Controller driver can now be upgraded without affecting the BSM or CM instrument drivers Deployment Manager Deployment Manager 24873, Deployment Manager Deployment Manager Deployment Manager Deployment Manager Waters DHCP Server Fully installing the PDA instrument driver while upgrading other drivers no longer causes a blank PDA method editor. The Local Console Controller driver can now be upgraded without affecting the ELS or FLR drivers. The Empower Verify Files utility now displays the correct version number for the ELS instrument driver. The BSM instrument driver no longer corrupts the nanoacquity BSM instrument driver when they are installed on the same computer. The ICS Companion utility now runs correctly after installation on Citrix servers. The Java runtime engine (JRE) is now installed only if Connections INSIGHT is also installed. The Waters DHCP Server is modified to offer DHCP addresses only on the instrument LAN, not all network cards , Solvent Library The solvent library no longer allows multiple solvent aliases of the same name, a practice that can cause it to stop working. Page 10 of 21
11 SCR number Instrument or Column Calculator Multiple instruments Multiple instruments Multiple instruments Multiple instruments 28178, 26840, Multiple instruments Description Japanese version of the Columns Calculator utility correctly shows L/Dp ratio. Warning level alarms, such as Drawing sample rate excessive, no longer cause memory leaks. Systems with five or more instruments no longer cause a "Database error: unable to continue" error in Empower. Statistical data displayed in post-run reports, such as Maximum System Pressure, is now calculated correctly even without enabling a data channel in the associated instrument s method. Previously, when two of the same type of instrument were included in a system (for example, two BSMs) the instrument method for both instruments would display the same values in the method editor, even if the values of one method were changed. Now both instrument methods can have unique values. When the Empower system definition includes multiple instruments of the same type for example, two BSMs their relative precedence now carries through to the console. The BSM listed first in the Empower system definition now consistently appears first in the console tree, and so on. This relative order is most meaningful elsewhere in the console, where a simple numerical suffix, such as the 2 in BSM2, differentiates instruments of the same type. This change pertains only to instruments of the same type. For instruments that differ in type, the precedence rules of the console still apply to the order of the instruments displayed in the console tree. For example, solvent managers come first, followed by sample managers and then detectors. Previously, for systems that include two PDA detectors in previous driver packs, instrument status obtained for just one of the detectors showed identically in both PDA control panels. This defect is corrected. Page 11 of 21
12 Solvent managers SCR number Instrument or Description BSM Previously, the BSM control panel occasionally failed to properly appear on some Empower Enterprise systems. The defect is corrected BSM Previously, the BSM was vulnerable to an erroneous Pump motor lost synchronization error and shutdown when running specific gradients with a 100% composition. The defect is corrected BSM Previously, the BSM would stop priming if there was an associated column change. The defect is corrected BSM Previously, the BSM did not fully stop the solvent flow if you specified doing so when the current flow rate was 0.01 ml/min. You can now fully stop a flow of 0.01 ml/min QSM Previously, the QSM lost its leak detector configuration after a power cycle. The defect is corrected QSM Previously, an error occurred under certain circumstances in the recommended ph range when you used the Auto Blend Plus function. The defect is corrected QSM Previously, curves 1 and 11 appeared in the method editor s graphical gradient representation as curve 6 (linear). The defect is corrected QSM The QSM leak test now properly primes the user-selected solvent, rather than always priming with the A solvent QSM Previously, the QSM might not completely reach a requested ml/min flow rate when slower method flow rate accelerations were specified. The defect is corrected QSM Stop flow buttons on the console and control panel now operate properly when flow is < 0.01 ml/min QSM The QSM to FTN system leak test has been modified to work successfully with larger ( 100 μl) extension loops , QSM Previously, a QSM pump could randomly cause a general failure error, especially with regard to applications in which flow rate or pressure changes rapidly, such as those that require starting and stopping the pump during a column change. The defect is corrected. Page 12 of 21
13 Sample managers SCR number Instrument or Description 24888, SM Previously, the XY mechanism move-needle functionality in the console did not move the needle. The defect is corrected SM, SM-FL Previously, the characterize Needle Seal diagnostic test showed incorrect values after calibration SM, SM-FL Previously, the results of the Calibrate XY and Zp Axes diagnostic test failed to appear in the console s log SM, SM-FL Previously, in Empower, the Run Samples Needle Wash sample queue function failed. The defect is corrected SM, SM-FL, SM- FTN Sample organizer Previously, during diagnostic tests, the sample manager failed to reliably warn of impending mechanical movements. The defect is corrected. SCR number Instrument or Description SO Previously, the sample organizer s LED lights failed to switch off when the door was closed, which could affect light-sensitive samples. The defect is corrected SO Previously, the sample organizer might not be visible in the console. The defect is corrected SO Previously, the shelf position immediately below the transfer location was not accessible in the sample organizer. The defect is corrected SO Error recovery is improved along with users ability to properly reset the sample organizer SO Previously, the sample organizer s configuration could be lost whenever the computer was restarted. Page 13 of 21
14 Column managers SCR number Instrument or Description CM-A The status of the CM-A leak sensor is now properly displayed CM-A Previously an error Preheater temperature sensor hardware fault was possible if the column temperature was set to ~90.0 without first stabilizing at a lower temperature. The defect is corrected CM, CM-A Previously, the ecord injection count and column history data was incorrectly updated in MassLynx open-architecture systems. The defect is corrected CM-A The CM-A s advanced-mode valve timing is improved to allow for more simultaneous valve operations CM Previously, the column history for Column 6 did not appear properly in the console. The defect is corrected CM, CM-A Previously, it was possible for ecord data to be incorrectly written to the device. The defect is corrected CM-A Previously, the console did not always adapt the list of available CM-A diagnostic channels according to supported hardware options or operating mode. The defect is corrected CM-A Previously, column switching errors were possible when transitioning between sample-queue and instrument-method column switching. Page 14 of 21
15 Third party SCR number Instrument or Description Analyst Previously, after loading the June 2011 driver pack, ACQUITY parameter settings appeared in Analyst acquisition methods, on the Software Application Properties tab. The defect is corrected , Analyst Previously, after loading the June 2011 driver pack on Windows 7, 64-bit machine, Analyst was unable to load plates. The defect is corrected Analyst System correctly collects data after a disabling or enabling the hardware profile. You must run StopProcesses before enabling a hardware profile Analyst Configuring the module in the console correctly updates ACQUITY method editor. Rebooting is required Analyst Column switching causes brief, expected flow delay in I-Class system under Analyst control Analyst ACQUITY AAO Service automatically restarts when the service stops working. The primary purpose of this improvement is to support StopAcquityProcesses and StopProcesses Analyst In mixed security mode, you can reliably log off and log back on. In integrated security mode, you cannot log off the computer while a batch is running. Detectors SCR number Instrument or Description PDA A small PDA memory leak that could cause problems if allowed to accumulate over thousands of injections is eliminated TUV Improved error detection and management of interrupted communications, which could have resulted in an unexpected change in data rate PDA Improved PDA error detection to guard against possible data overruns, should communications to the PDA be interrupted for an extended period of time PDA Channel name and description for MaxPlot and other complex, calculated channels, such as sum, difference, and ratio, are improved to show proper values and ranges. Page 15 of 21
16 SCR number Instrument or Description TUV Previously, a possible error involving the TUV method editor permitted saving a corrupt method, which can produce an incorrect data rate on the second channel. The defect is corrected Previously, the 2998 detector experienced data saturation or produced an incorrect auto exposure when next-injection delays exceed a duration of ten minutes. The defect is corrected PDA The PDA exposure calculation has been corrected to account for timed event wavelength changes to spectral regions of higher energy. Previously, it was possible for an incorrect exposure time to be calculated and executed, which could result in peak area discrepancies TUV Previously, the leak sensors in some older-model TUVs failed to function. The defect is corrected TUV Shutter idle configuration is correctly refreshed, and accurately shows current state PDA Previously, the 2998 server sometimes stopped working on startup or during resets of the server while under the control of MassLynx. The defect is corrected PDA Internal fans operate reliably when the detector is powered up in Maximum cooling mode , PDA Previously, under some circumstances, changing the PDA lamp could cause the console to stop working. The defect is corrected , PDA, TUV Improved PDA and TUV lamp and flow-cell ID communications, to reduce the likelihood of false identification failures , 28273, 28279, PDA, TUV Previously, when you changed a lamp, its memory device sometimes recorded incorrect data (hours, date, and time). The defect is corrected. Page 16 of 21
17 Known issues in this release The table below lists known issues in Driver Pack 4. Waters change request database uses the change request (SCR) numbers in the left-hand column to track these issues. SCR number Instruments or Description s SO If, in the console, you disable the Z-motor and then move the Z-axis manually, the device begins to beep and flash its frontpanel LED. Workaround: To recover, reset the device detector It is possible that the 2489 can experience a filter initialization failure in high temperature and humidity environments when calibrating. Workaround: Filter initialization under varying temperature and humidity conditions is optimized. If filter initialization fails, reboot the instrument to retry the initialization Analyst To ensure successful data acquisition, Analyst control requires methods to specify the same data rate Deployment Manager Occasionally the ICSCompanion service is not installed during normal installation. Workaround: Using the Deployment Manager, perform a typical removal, reboot the computer, and then perform the installation again Console After you specify which instruments will appear in the console, the method editor displays the previously specified instruments. Workaround: Reboot the computer after changing the instrument configuration in the console TUV The TUV s rear-panel switch output, or outputs, can momentarily pulse with each method setup on each injection, causing an extra valve movement if the outputs are used to control external valves Waters Pump Control It is possible that using the Windows Control Panel to uninstall Waters Pump Control released prior to Driver Pack 4 does not remove all files. Workaround: Uninstall Waters Pump Control using the Driver Pack 4 Deployment Manager. Page 17 of 21
18 SCR number Instruments or s Local Console Controller (LCC), ELS, FLR Description During an upgrade from the 1.40 Driver Pack, if the Local Console Controller is being upgraded, the ELS and FLR detectors must be upgraded as well, to avoid failures of the ELS and FLR instrument drivers CM The Column Manager page in the console can sometimes display the message "No Function Text" Verify Instrument Driver Files Workaround: Ignore the message. If the Windows User Account Control (UAC) function is enabled, you must be logged in as an administrator to run the Verify Instrument Driver Files utility Console Closing the console after initiating the System Startup function cancels the function and resets the system modules. Workaround: Do not close the console until the System Startup function is completed Autoloader The Autoloader can sometimes attempt to load BSM firmware on the nanobsm device Deployment Manager Workaround: Power-off all connected nanoacquity modules before beginning an installation. Installing older instrument drivers can occasionally cause deletion of the Empower Verify Files checksumshell.bat file. Workaround: During installation and removal, the Deployment Manager creates a backup of checksumshell.bat in the C:\Empower\Instruments\Logs folder. If checksumshell.bat is deleted, it can be restored from that location to C:\Empower SM-FL Clearing the Disable Motors option in the console does not re-enable motors. Workaround: Reset the instrument via the console to reenable the motors FLR Aborting FLR diagnostic tests sometimes prevents them from being rerun. Workaround: Reboot instrument Waters Pump Control Waters Pump Control (WPC) 3.10 is not compatible with nanoacquity Workaround: If you run nanoacquity along with Driver Pack 4, choose a Custom installation in the Deployment Manager, and then deselect WPC 3.10 before you install. Page 18 of 21
19 SCR number Instruments or Description s ELS, FLR, RI, 2998 Although other Driver Pack 4 instruments support networks with latencies as high as 300 milliseconds, in some cases the ELS, FLR, RI, and 2998 do not. Performance will therefore vary SM Resuming an Empower sample set after an SM error can result in the erroneous message Waiting for injection. Workaround: Abort and restart sample set Verify Files The Verify Instrument Driver Files utility can display script errors with Internet Explorer 8 installed. Workaround: Install Microsoft hotfix KB Connection INSIGHT When you upgrade from the June 2011 Driver Pack to Driver Pack 4 on Windows 7, Connections INSIGHT installation fails occasionally. Workaround: Remove Connections INSIGHT, reboot the computer, install Connections INSIGHT D system In a MassLynx system using 515(s) with a PCM, it can be necessary to reboot the computer after changing the MassLynx instrument configuration Autoloader The Autoloader can stop working if run on a LAC/E 32 module connected to an e2695 separations module. Workaround: Reboot a LAC/E 32 module connected to an e2695 separations module before running the Autoloader QSM The QSM instrument driver must be upgraded when upgrading other instrument drivers to Driver Pack 4, otherwise the Auto Blend Plus feature, specified in the method editor, will stop working. Page 19 of 21
20 Anti-virus considerations If you are using real-time virus scanning, exclude the data system and instruments folders and sub-folders as shown below. Some real-time virus scanners can mistake normal data acquisition and instrument control for virus activity and thus interfere with proper operations. Full-system scans and live updates can be network-intensive, disk-intensive, and CPU-intensive, and they can also interfere with normal data acquisition. Schedule scans and updates for idle times when data acquisition does not occur. Exclude these folders: For Empower 2 and Empower 3 installations, the Empower installation folder, usually C:\Empower, and its sub-folders. For MassLynx installations, C:\Program Files\Waters Instruments, and its sub-folders ( C:\Program Files (x86)\waters Instruments on 64-bit machines). the MassLynx installation folder, usually C:\Masslynx, and its subfolders. For standalone console and other third party installations, C:\Program Files\Waters Instruments, and its sub-folders ( C:\Program Files (x86)\waters Instruments on 64-bit machines). Certain anti-virus program features like "intrusion prevention" and "tamper protection" can also interfere with normal operation. Disable them as well. Requalification guidelines Driver Pack 4 includes instrument driver and firmware used on your laboratory instruments, servers, and workstations. As with any change in a regulated environment, follow your organization's standard operating procedures for managing changes. This update has been designed not to alter the physical specifications of the instrument(s) or system(s), and Waters believes that methods run and data generated on instruments before and after the update will be identical. However, after a risk based review of the changes detailed in the release notes, users in a regulated environment may need to consider and document what, if any, requalification of the instrument modules, the chromatographic system, and/or the Chromatographic Data System (CDS) are required by their own change control SOPs. Use the appropriate qualification methods outlined below to verify installation and correct operation: To confirm that the driver pack installation has not negatively affected any critical system functions, and that the new files have loaded properly, you may consider performing a Installation Qualification (IQ). To confirm the firmware files installed on the modules are correctly installed, compare the checksum values in the product release notes to the checksum values displayed in the console. To confirm an instrument's operation with the newly loaded driver and firmware, you may consider performing an Operational Qualification (OQ) for the updated instrument or system. To confirm performance, control, and communications of the instruments in your system, you should consider conducting a Performance Qualification (PQ) (user or vendor PQ) or System Suitability test. Page 20 of 21
21 Where multiple identical systems are involved, a risk based approach to qualification activities may be an important consideration. Requalifying the instrument driver, firmware, and hardware when upgrading is covered under the Total Assurance Plan with System Qualification Option in the following cases: During yearly requalification under the plan. If loading this release is required for operation of a new module or system, where qualification of the new module or system is covered under the plan. Requalification of the CDS and computers after a driver upgrade may or may not be included in Assurance Plans from the CDS vendor. Review your Total Assurance Plan to determine which services are covered, and which are not covered. For situations not covered under the plan, Waters can still perform the qualification, but additional charges apply. Contact Waters Technical Service If you are located in the USA or Canada, report malfunctions or other problems to Waters Technical Service ( ). If you are located elsewhere, phone the Waters corporate headquarters in Milford, Massachusetts (USA), or contact your local Waters subsidiary. Our Web site includes phone numbers and addresses for Waters locations worldwide. Page 21 of 21
Alliance e2695 Separations Module Firmware Version 3.04
*716005261* *Rev.A* Alliance e2695 Separations Module Firmware Version 3.04 Release Notes 716005261 Revision A Copyright Waters Corporation 2018 All rights reserved General information Copyright notice
More informationWaters Driver Pack 2016 Release 1
Waters Driver Pack 2016 Release 1 Installation and Configuration Guide 715005197 Revision A Copyright Waters Corporation 2016 All rights reserved General information Copyright notice 2016 WATERS CORPORATION.
More informationWaters ACQUITY QDa Instrument Component Software version 1.54
R E L E A S E N O T E S Waters ACQUITY QDa Instrument Component Software version 1.54 The ACQUITY QDa Detector is a single-quadrupole mass detector. The instrument component software (ICS) adds the capability
More informationACQUITY UPLC Systems June 2011 Driver Pack and Firmware Updates
R E L E A S E N O T E S ACQUITY UPLC Systems June 2011 Driver Pack and Firmware Updates What s new in this release... 3 Reasons for installing this software release... 3 Installation Notes... 3 Recommended
More informationWaters Driver Pack 2018 Release 1
Waters Driver Pack 2018 Release 1 Installation and Configuration Guide 715005659 Revision B Copyright Waters Corporation 2018 All rights reserved General information Copyright notice 2018 WATERS CORPORATION.
More informationMassLynx 4.2 SCN 989 Release Notes
MassLynx 4.2 SCN 989 Release Notes Page 1 of 20 Waters, THE SCIENCE OF WHAT S POSSIBLE., ACQUITY, Alliance, MassLynx, nanoacquity, and Xevo are registered trademarks and OpenLynx, Quanpedia, and TargetLynx
More informationFor additional information, see these documents on the Breeze 2 software DVD:
R E L E A S E N O T E S Waters Breeze 2 Read these release notes before you install and operate Breeze 2. They contain information necessary to ensure successful liquid chromatography results, including:
More informationMassLynx 4.1 SCN 804. Release Notes. SQ Detector 2
R E L E A S E N O T E S MassLynx 4.1 SCN 804 Release Notes SQ Detector 2 43464 Rev 01 Copyright Waters Corporation 2011. All rights reserved. Copyright 2011 Waters Corporation 43464_1. Waters is a registered
More informationACQUITY UPLC Systems with 2D Technology Capabilities Guide
ACQUITY UPLC Systems with 2D Technology Capabilities Guide Revision A Copyright Waters Corporation 2012 All rights reserved Copyright notice 2012 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA
More informationAgilent 7890 GC Instrument Control Software Version 2.6
R E L E A S E N O T E S Agilent 7890 GC Instrument Control Software Version 2.6 Version 2.6 of the instrument control software (ICS) for the Agilent 7890 GC addresses several issues, provides support for
More informationWaters 2707 Autosampler ICS v1.20 Installer
R E L E A S E N O T E S Waters 2707 Autosampler ICS v1.20 Installer Read these release notes before you install and operate the Waters 2707 autosampler instrument control software (ICS). They contain information
More informationEmpower 3 Feature Release 3 Hotfix 1 Release Notes
*716004786* *Rev.A* Empower 3 Feature Release 3 Hotfix 1 Release Notes 716004786 Revision A Copyright Waters Corporation 2016 All rights reserved General information Copyright notice 2016 WATERS CORPORATION.
More informationSCN 848 Release Notes
R E L E A S E N O T E S SCN 848 Release Notes 43198 Rev 1 Copyright Waters Corporation 2011. All rights reserved. 43198 Rev 1. Waters is a registered trademark, Waters the Science of What s Possible and
More informationACQUITY UPLC Local Console Controller Instructions
ACQUITY UPLC Local Console Controller Instructions Contents Topic Page Product description and function Operation in a GXP regulated environment Preparing to install the LCC 3 Installing the LCC 3 Navigating
More informationAgilent 7890 GC Instrument Control Software Version 2.5
R E L E A S E N O T E S Agilent 7890 GC Instrument Control Software Version 2.5 The Agilent 7890 GC Instrument Control Software version 2.5 enables dual-tower injection capability and provides support
More informationWaters Empower 2 Service Pack G
R E L E A S E N O T E S Waters Empower 2 Service Pack G Empower 2 Service Pack G enhances LDAP capabilities. These release notes also address the following information and instructions: System and software
More informationHITACHI LACHROM ELITE
HITACHI LACHROM ELITE Clarity Control Module ENG Code/Rev.: M133/60D Date: 7/16/2018 Phone: +420 251 013 400 DataApex Ltd. Fax: +420 251 013 401 Petrzilkova 2583/13 clarity@dataapex.com 158 00 Prague 5
More informationAgilent Instrument Control Framework (ICF) Update A Driver Update 1 (DU1) Hot Fix 2 (HF2)
R E L E A S E N O T E S Agilent Instrument Control Framework (ICF) Update A.02.03 Driver Update 1 (DU1) Hot Fix 2 (HF2) These release notes explain the installation of the updated Agilent Instrument Control
More informationEmpower Software and Security Software Interactions
TECN134790604 Rev. 00 Page 1 of 7 Empower Software and Security Software Interactions This Technical Note discusses how Empower software functions and examines potential interactions with anti-virus software
More informationAgilent 7890 GC Instrument Control Software Version 2.01
R E L E A S E N O T E S Agilent 7890 GC Instrument Control Software Version 2.01 These release notes explain how to install Waters Instrument Control Software (ICS), version 2.01, for the Agilent 7890
More informationWaters Empower Software A1100 System Configuration Instructions for Ethernet Control
Waters Empower Software A1100 System Configuration Instructions for Ethernet Control In order to configure an A1100 system for Ethernet control or to acquire A1100 DAD 3D data in Empower Software, it is
More informationBreeze 2 Installation and Configuration Guide
Breeze 2 Installation and Configuration Guide 715001951/Revision B Copyright Waters Corporation 2008 2009 All rights reserved Copyright notice 2008 2009 WATERS CORPORATION. PRINTED IN THE UNITED STATES
More informationAgilent OpenLAB CDS Waters Alliance HPLC Instrument Control Add-On. Installation, Upgrade and Migration Guide
Agilent OpenLAB CDS Waters Alliance HPLC Instrument Control Add-On Installation, Upgrade and Migration Guide Notices Agilent Technologies, Inc. 2013, 2014 No part of this manual may be reproduced in any
More informationWaters Headspace Control Option, Version 3.0
R E L E A S E N O T E S Waters Headspace Control Option, Version 3.0 Version 3.0 of the Headspace Control Option (HCO) contains instrument drivers for Agilent Headspace Samplers and a toolkit program that
More informationMassLynx TM 4.1 SCN 916 Release Notes
MassLynx TM 4.1 SCN 916 Release Notes Page 1 of 19 46154_1 Waters is a registered trademark, Waters the Science of What s Possible and MassLynx are trademark of Waters Corporation All other trademarks
More informationThermo. EASY-nLC User Guide for LC Devices
Thermo EASY-nLC 1200 User Guide for LC Devices 60053-97273 Revision A September 2015 2015 Thermo Fisher Scientific Inc. All rights reserved. AFC, EASY-nLC, EASY-Column, EASY-Spray, Foundation, LC TriPlus,
More informationValidation Sample Set Generator
Validation Sample Set Generator General Information This tool is an example program which demonstrates some capabilities of the Empower Toolkit. It was not created using Waters Software Development Lifecycle.
More informationWaters Fraction Collector III with Empower Instrument Control Software Version 1.00
R E L E A S E N O T E S Waters Fraction Collector III with Empower Instrument Control Software Version 1.00 These release notes explain how to install Waters Instrument Control Software (ICS), version
More informationStandard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017
Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017 Introduction This document is only for operation of the HPLC and ICP-MS instruments to obtain data. It contains no maintenance
More informationA. About this Manual
Section A. About this Manual A. About this Manual 1. Table of Contents 1. Table of Contents... 1 2. How to Use this User Manual... 2 3. General Introduction to the PAL3 System... 3 4. PAL3 System Specifications...
More informationFiery proserver User Guide
Fiery proserver User Guide Congratulations on your purchase of the Fiery proserver, a powerful production solution for your printer. Before you begin Make sure that you have: About this document Installed
More informationWhat s New in Empower 3
What s New in Empower 3 Revision A Copyright Waters Corporation 2010 All rights reserved Copyright notice 2010 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IN IRELAND. ALL RIGHTS RESERVED.
More informationRelease Notes. Analyst TF Software
Release Notes Analyst TF 1.5.1 Software Release Date: April 2011 This document is provided to customers who have purchased AB SCIEX equipment to use in the operation of such AB SCIEX equipment. This document
More informationManual HPLC-4 Alliance-UV
Manual HPLC-4 Alliance-UV MAKE SURE NEVER TO: 1. pump particles or air through the column. 2. inject when the injector is dry. 3. inject solutions containing particles. 4. apply large (sudden) pressure
More informationChromeleon-PA Analyzer User s Guide Dionex Corporation
Chromeleon-PA Analyzer User s Guide 2003 Dionex Corporation Document No. 031964 Revision 01 October 2003 2003 by Dionex Corporation All rights reserved worldwide. Printed in the United States of America.
More informationEmpower 3 Installation, Configuration, and Upgrade Guide /Revision A Copyright Waters Corporation All rights reserved
Empower 3 Installation, Configuration, and Upgrade Guide 715005081/Revision A Copyright Waters Corporation 2002-2016 All rights reserved Page ii General Information Copyright notice 2002-2016 WATERS CORPORATION.
More informationEASY-nLC Series User Guide for the Xcalibur Data System (Xcalibur version 2.1 or later)
Thermo EASY-nLC Series User Guide for the Xcalibur Data System (Xcalibur version 2.1 or later) P/N 60053-97230 Revision B January 2013 2013 Thermo Fisher Scientific Inc. All rights reserved. EASY-nLC,
More informationFiery proserver User Guide
Fiery proserver User Guide Congratulations on your purchase of the EFI Fiery proserver, a powerful production solution for your printer. Before you begin Make sure that you have: About this document Installed
More informationImpact Analysis for Software changes in OpenLAB CDS A.01.04
Impact Analysis for Software changes in OpenLAB CDS A.01.04 Document Information: Filename Product Identifier Product Revision Project Identifier Document Revision Impact_Analysis_OpenLAB_CDS_A0104 OpenLAB
More informationTA Instruments New Features in ITCRun TM Software
TA Instruments New Features in ITCRun TM Software Notice The material contained in this manual, and in the online help for the software used to support TA Instruments products, is believed adequate for
More informationRelease Note for Agilent LC and CE drivers Revision A.02.14
Release Note for Agilent LC and CE drivers Revision A.02.14 Introduction This release note provides important information for the release of Agilent LC and CE drivers A.02.14. The 1260 Infinity II LC is
More informationLTQ Velos SP3 Release Notes
LTQ Velos 2.6.0 SP3 Release Notes These release notes briefly list changes in the LTQ Velos 2.6.0 SP3 experiment setup, tuning, and calibration application; fixes to previously reported issues; and known
More informationChromQuest 5.0 Quick Reference Guide
ChromQuest 5.0 Quick Reference Guide This guide contains an overview of the ChromQuest chromatography data system, with topics organized by workflow. For more information, refer to the ChromQuest User
More informationCliquid Software for Routine Analysis
Cliquid 3.0.1 Software for Routine Analysis Release Notes Table of Contents Known Issues... 2 LC Systems...2 Initial Log On... 2 Continuous Usage Limitation... 2 Acquisition Fails due to Locked Windows
More informationOperating Manual. High Performance Liquid Chromatograph. Scientific Equipment Center, Prince Of Songkla University
Operating Manual High Performance Liquid Chromatograph Agilent 1100 ; VWD, DAD, FLD and RID Scientific Equipment Center, Prince Of Songkla University Operating Manual High Performance Liquid Chromatograph
More informationWaters 600E Multisolvent Delivery System Quick Start Guide
Waters 600E Multisolvent Delivery System Quick Start Guide 34 Maple Street Milford, MA 01757 WAT174-04TP, Revision 3 NOTICE The information in this document is subject to change without notice and should
More informationVarian 356-LC Control Software
OPERATOR S MANUAL FOR Varian 356-LC Control Software Version 1.0 November 2007 Polymer Laboratories Ltd, Now a part of Varian, Inc., Essex Road, Church Stretton, Shropshire SY6 6AX, UK Tel +44 (0)1694
More informationFlexar FX-20 LC Pump: Ultimate Performance and Flexibility for UHPLC and HPLC
P R O D U C T N O T E Liquid Chromatography Key Features Improved productivity compatible with all UHPLC columns, the Flexar FX-20 Pump can operate at pressures as high as 18,000 psi, providing throughput
More informationEmpower 2 System Administrator s Guide
Empower 2 System Administrator s Guide 71500109808/Revision B Copyright Waters Corporation 2008 All rights reserved Copyright notice 2008 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND
More informationFlexar FX-20 LC Pump: Ultimate Performance and Flexibility for UHPLC and HPLC
P R O D U C T N O T E Liquid Chromatography Key Features Improved productivity compatible with all UHPLC columns, the Flexar FX-20 Pump can operate at pressures as high as 18,000 psi, providing throughput
More information3 Starting the chiral HPLC and logging into LC Real Time Analysis.
Edition: 1.2 Date: 7-Oct-11 Page 1 of 5 1 Responsibilities. First person responsible: Mariëlle Delville. Second person responsible: Helene I.V. Amatdjais-Groenen. 2 Definitions. Read the information on
More informationRelease Notes for Agilent Drivers for Thermo Scientific Chromeleon 7 (Rev. 1.1)
Release Notes for Agilent Drivers for Thermo Scientific Chromeleon 7 (Rev. 1.1) Release Notes for Agilent Drivers for Thermo Scientific Chromeleon 7 (Rev. 1.1) This document provides information on the
More informationApplied Biosystems RapidFinder Software Version 1.0 and SDS Software Version Beta Release 10/05 Release Notes (PN Rev A)
Applied Biosystems RapidFinder Software Version 1.0 and SDS Software Version 1.3.2 Beta Release 10/05 Release Notes (PN 4370225 Rev A) Software Release notes for: Applied Biosystems RapidFinder Software
More informationxponent Version 4.2 RELEASE NOTES
xponent Version 4.2 RELEASE NOTES Copyright Luminex Corporation, 1996-2013. All rights reserved. Part Number: 89-30000-00-462 Disclaimer Subject to change without notice. Although prepared to ensure accuracy,
More informationServer Edition. V8 Peregrine User Manual. for Microsoft Windows
Server Edition V8 Peregrine User Manual for Microsoft Windows Copyright Notice and Proprietary Information All rights reserved. Attix5, 2015 Trademarks - Microsoft, Windows, Microsoft Windows, Microsoft
More informationAvigilon Control Center 6 System Integration Guide
Avigilon Control Center 6 System Integration Guide for Paxton Net2 Access Control Systems 2018, Avigilon Corporation. All rights reserved. AVIGILON, the AVIGILON logo, AVIGILON CONTROL CENTER and ACC are
More informationSpark Holland introduces
BETTER SAMPLE CARE Spark Holland introduces ALIAS All-round, state-of-the-art autosampler for HPLC Spark Holland introduces ALIAS All-round, state-of-the-art autosampler for HPLC ALIAS All-round, state-of-the-art
More informationEquipment Overview: Safety:
PerkinElmer Series 200 HPLC: Operating Procedure 12/2/11 R.C Equipment Overview: This document outlines the basic procedure for using the Sedex light scattering detector and the PE series 200 HPLC system
More informationTaurus Super-S LCM. Dual-Bay RAID Storage Enclosure for two 3.5 Serial ATA Hard Drives. User Manual July 27, v1.2
Dual-Bay RAID Storage Enclosure for two 3.5 Serial ATA Hard Drives User Manual July 27, 2009 - v1.2 EN Introduction 1 Introduction 1.1 System Requirements 1.1.1 PC Requirements Minimum Intel Pentium III
More informationNew features and Workflow Enhancements Empower 3 Software
New features and Workflow Enhancements Empower 3 Software Mogens Hallas Nordic User Training 2011 Långvik, September 6th to 9th, 2011 2011 Waters Corporation 1 Empower 3 Architecture and operating systems
More informationABI PRISM 3100-Avant and 3100 Data Collection v2.0 Software Frequently Asked Questions (FAQ)
ABI PRISM 3100-Avant and 3100 Data Collection v2.0 Software Frequently Asked Questions (FAQ) Table of Contents General Information...2 Access Control/Auditing Features...2 Instrument/Plate Setup...3 Spectral
More informationTA Instruments New Features in DSCRun TM Software
TA Instruments New Features in DSCRun TM Software Notice The material contained in this manual, and in the online help for the software used to support TA Instruments products, is believed adequate for
More informationServer Edition USER MANUAL. For Microsoft Windows
Server Edition USER MANUAL For Microsoft Windows Copyright Notice & Proprietary Information Redstor Limited, 2016. All rights reserved. Trademarks - Microsoft, Windows, Microsoft Windows, Microsoft Windows
More informationHigh Motor Current Shutdown High Motor Current Warning Low Motor Current Shutdown Compressor Auxiliary Shutdown
Important: To upgrade from Quantum HD version 10.21 or earlier to Quantum HD Unity version 10.30 or later requires the flashcard to be changed. These are available through the Baltimore Parts Center, part
More informationSOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.31: PSS
SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.31: PSS0051-05 Introduction This document details the installation and known issues with OMNISEC v10.31. This release is a maintenance release to
More informationMaxum Workstation Update Release Notes
1 August 9, 2017 Maxum Workstation 5.30.01 Update Release Notes Requirements Maxum Workstation Software will operate best on computers equipped with a multi-core processor, with at least 4 GB RAM (8+ GB
More informationCliquid Software. Release Notes. Where to Get Help. Changes from Version 3.0.x to Version 3.2. Other Documentation. Support
Cliquid Software Software for Routine Analysis Version 3.2 Where to Get Help Other Documentation Cliquid 3.2 Software Site Planning Guide Cliquid 3.2 Software Installation Guide for Upgrades Cliquid 3.2
More informationTechnical Overview. Introduction
Performance characteristics of the Agilent 12 Infinity Series Variable Wavelength Detectors Faster results, improved sensitivity and absolute data security Technical Overview Introduction The Agilent 12
More informationSOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.30: PSS
SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.30: PSS0051-04 Introduction This document details the installation and known issues with OMNISEC v10.30. This release is a major release to fix a
More informationEmpower 2 Feature Release 5 Upgrade Guide
Empower 2 Feature Release 5 Upgrade Guide 715001799/Revision A Copyright Waters Corporation 2008 All rights reserved Copyright notice 2008 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND
More informationServer Edition. V8 Peregrine User Manual. for Linux and Unix operating systems
Server Edition V8 Peregrine User Manual for Linux and Unix operating systems Copyright Notice and Proprietary Information All rights reserved. Attix5, 2015 Trademarks - Red Hat is a registered trademark
More informationUniPoint System Software User s Guide
UniPoint System Software User s Guide LT2137 1998 Gilson, Inc. All rights reserved Exercise 3-Creating operations list... 4-22 Create and name operations list... 4-22 Set up steps... 4-23 Identify sample
More informationDesktop & Laptop Edition
Desktop & Laptop Edition USER MANUAL For Mac OS X Copyright Notice & Proprietary Information Redstor Limited, 2016. All rights reserved. Trademarks - Mac, Leopard, Snow Leopard, Lion and Mountain Lion
More informationChromQuest 4.2 Chromatography Data System
ChromQuest 4.2 Chromatography Data System Installation Guide CHROM-97200 Revision A April 2005 2006 Thermo Electron Corporation. All rights reserved. Surveyor is a registered trademark and ChromQuest is
More informationWindows NT 4.0 and Windows 2000 Service Pack and Hotfix Deployment Best Practices
Windows NT 4.0 and Windows 2000 Service Pack and Hotfix Deployment Best Practices Standards and Guidelines for Management and Deployment Prepared by: Al Guevara Rev. 04/02/03 I. Purpose The purpose of
More informationAqueous GPC Instructions - Detailed
Aqueous GPC Instructions - Detailed Waste Bottle Columns PL Datastream Injection Port GPC Computer UV Detector GPC Pump Eluent Bottle The Aqueous GPC Set Up RI Detector Important The points below must
More informationTaurus - USB. User Manual. Dual-Bay Storage Enclosure for 3.5 Serial ATA Hard Drives. (English)
Dual-Bay Storage Enclosure for 3.5 Serial ATA Hard Drives User Manual (English) v1.0 June 21, 2007 EN Table of Contents CHAPTER 1 - INTRODUCTION 1 CHAPTER 3 - SYSTEM SET UP 6 ICON KEY 1 THE TAURUS USB
More informationIntroduction to the HPLC ChemStation and Acquisition
Introduction to the HPLC ChemStation and Acquisition In This Section, We Will Discuss: How to work in the Microsoft Windows Environment The structure of the ChemStation Software. How to set up an acquisition
More informationAgilent OpenLAB ECM Intelligent Reporter
Agilent OpenLAB ECM Intelligent Reporter Installation and Configuration Guide Agilent Technologies Notices Agilent Technologies, Inc. 2007-2016 No part of this manual may be reproduced in any form or by
More informationDionex AS-AP Sample Conductivity and ph Accessory Setup and Operation Guide
Dionex AS-AP Sample Conductivity and ph Accessory Setup and Operation Guide Document No. 065470 Revision 02 February 2012 2012 by Thermo Fisher Scientific Inc. All rights reserved. Chromeleon is a registered
More informationRelease Notes P/N REV A01 August 11, 2010
EMC Unisphere Analyzer Version 1.0.0 Release Notes P/N 300-011-113 REV A01 August 11, 2010 These release notes contain supplemental information about EMC Unisphere Analyzer, version 1.0.0. Topics include:
More informationChromatography Software Training Materials. Contents
Chromatography Software Training Materials This document contains information on how to build a method, start the instrument to acquire data, and then process the data using the Galaxie Program. You will
More informationIntel Entry Storage System SS4000-E
Intel Entry Storage System SS4000-E Software Release Notes January 2007 Storage Systems Technical Marketing Engineering Document Revision History Intel Entry Storage System SS4000-E Document Revision History
More informationDominion SX Release Notes
Dominion SX Release Notes Dominion SX Release Notes and Compatibility Release 3.1.7 April 14, 2009 Document updated June 10, 2010 Contents Applicability... 2 Release Status... 2 Expiration Date of Content...
More informationTaurus Super-S3 LCM. Dual-Bay RAID Storage Enclosure for two 3.5-inch Serial ATA Hard Drives. User Manual March 31, 2014 v1.2
Dual-Bay RAID Storage Enclosure for two 3.5-inch Serial ATA Hard Drives User Manual March 31, 2014 v1.2 www.inxtron.com EN Table of Contents Table of Contents 1 Introduction... 1 1.1 Technical Specifications...
More informationESCA 2000 UPGRADE V ESCA 2000 V102 is compatible with NT 4.0, Windows 2000 and XP. It comes in five versions.
ESCA 2000 UPGRADE V102.02 ESCA 2000 V102 is compatible with NT 4.0, Windows 2000 and XP. It comes in five versions. ESCA 2000 A - ESCA 2000 B - ESCA 2000 D - ESCA 2000 E - ESCA 2000 G - Standalone Data
More informationRelease Notes Life Technologies Attune NxT Software v2.3
Release Notes Life Technologies Attune NxT Software v2.3 In the following pages you will find instructions describing: New software features Known software/system issues with troubleshooting guidance Software
More informationAvigilon Control Center 5.10 Release Notes
Version 5.10.2.2 Released Tuesday October 12 th, 2016 Release Summary This is a service release of Avigilon Control Center 5.10. Users experiencing issues listed in the Issues Fixed list should consider
More informationXenClient Enterprise Release Notes
XenClient Enterprise Release Notes Version 4.1 June 25, 2012 XenClient Enterprise Synchronizer Version 4.1 XenClient Enterprise Engine Version 4.1 1 About XenClient Enterprise XenClient Enterprise is a
More informationNT2 U3. 2-Bay RAID Storage Enclosure. User Manual May 18, 2010 v1.1
2-Bay RAID Storage Enclosure User Manual May 18, 2010 v1.1 EN Introduction 1 Introduction 1.1 System Requirements 1.1.1 PC Requirements Minimum Intel Pentium III CPU 500MHz, 128MB RAM Windows Vista, Windows
More informationSoftware Update Notification
Software Update Notification PSS0002-31: Mastersizer 2000 v5.60 software Abstract This document details the release of software PSS0002-31. This is version 5.60 of software for the Mastersizer 2000 laser
More informationSciex QTrap Operational Steps for Trained Personnel
Sciex 6500+ QTrap Operational Steps for Trained Personnel Last Updated 09172017 1. If any of the following instructions does not make sense to you or was not covered during your hand-on training, stop
More informationNOTICE: This document contains references to Varian. Please note that Varian, Inc. is now part of Agilent Technologies. For more information, go to
NOTICE: This document contains references to Varian. Please note that Varian, Inc. is now part of Agilent Technologies. For more information, go to www.agilent.com/chem. Varian MS Workstation 6.9 Service
More informationMicro-Master Touch Release Notes
Micro-Master Touch Release Notes 1 Micro-Master Touch Release 1.7 1 2 Version Information 1 2.1 Current... Released Versions 1 3 Front Panel 1 3.1 3.1.1 3.1.2 3.1.2.1 3.1.3 3.1.3.1 3.1.3.2 3.1.3.3 3.1.3.4
More informationAgilent ICF Support v3.0
*716004817* *Rev.A* Agilent ICF Support v3.0 Release Notes 716004817 Revision A Copyright Waters Corporation 2018 All rights reserved General information Copyright notice 2018 WATERS CORPORATION. PRINTED
More informationRelease Note for Agilent LC and CE Drivers Revision A.02.17
Release Note for Agilent LC and CE Drivers Revision A.02.17 Introduction This release note provides important information for the release of Agilent LC and CE drivers A.02.17. The Agilent LC and CE Driver
More informationAgilent 1290 Infinity with ISET
Agilent 1290 Infinity with ISET User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in any form or by any means (including electronic storage
More informationData Acquisition with CP-2002/2003 Micro-GC Control
Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA 94598 Star Chromatography Workstation Version 6 Data Acquisition with CP-2002/2003 Micro-GC Control Operation Manual Varian, Inc. 2002
More informationANALYTICAL SYSTEMS. L3320 Autosampler with excellent accuracy and linearity. System configuration examples
ANALYTICAL SYSTEMS Modular HPLC system can be configured according to customer needs from components as follows: PDA Detector with wavelength range 200-800 nm with scanning possibility or UV-VIS Detector
More informationDionex SRD-10 Suppressor Regenerant Detector Installation Instructions
Dionex SRD-10 Suppressor Regenerant Detector Installation Instructions Document No. 065355 Revision 03 January 2012 2012 Thermo Fisher Scientific Inc. All rights reserved. Chromeleon and SRS are registered
More information