Waters ACQUITY QDa Instrument Component Software version 1.54

Size: px
Start display at page:

Download "Waters ACQUITY QDa Instrument Component Software version 1.54"

Transcription

1 R E L E A S E N O T E S Waters ACQUITY QDa Instrument Component Software version 1.54 The ACQUITY QDa Detector is a single-quadrupole mass detector. The instrument component software (ICS) adds the capability of controlling the instrument by means of Waters' Empower software. These release notes explain important information about the system and software requirements as well as known defects in this release. System and software requirements... 2 Operating system... 2 Client and server compatibility... 2 Driver pack compatibility... 3 Empower Workstation and LAC/E 32 support... 3 Installing the ACQUITY QDa ICS... 3 Verifying the installation... 4 Uninstalling the ICS... 4 System considerations... 4 Updates in version Issues fixed in this release... 5 Fixed in version Fixed in version Fixed in version Fixed in version Known issues in this release... 8 Installation... 8 Console... 8 Instrument and instrument system management... 9 Acquisition Anti-virus considerations Empower, QDa, Waters, and "THE SCIENCE OF WHAT'S POSSIBLE." are registered trademarks of Waters Corporation. LAC/E 32 is a trademark of Waters Corporation. All other trademarks are the sole property of their respective owners. Copyright 2016 Waters Corporation * * *Rev.K* Page 1 of 12

2 Compliance recommendations Instrument update classification (minor update) Instrument requalification options after software/firmware change Requalification with Waters' Total Assurance Plans Contact Waters Technical Service System and software requirements The ACQUITY QDa ICS is supported by Empower 2, FR5 and later, and all versions of Empower 3. See also: The Empower 2 and Empower 3 release notes for further details about supported operating systems, service packs, hot fixes, and Internet Explorer compatibility. Operating system The following table lists the application software, feature release, and operating systems required for use with the ACQUITY QDa ICS. In this table, Windows XP means Microsoft Windows XP Professional and Windows 7 means Microsoft Windows 7 Professional; SP means service pack, and FR means feature release. Application software Minimum FR Operating system Empower 2 FR5 Windows XP SP3, 32-bit Empower 3 Supports all feature releases Windows 7, 64-bit Windows XP SP3, 32-bit Client and server compatibility The client and server applications must be version 1.53 or version See the table below for compatible versions. The LAC/E must have version 1.54 installed in order to have the issues fixed in this release. The issues fixed in version 1.54 are listed on page 5. Citrix LAC/E Page 2 of 12

3 Driver pack compatibility Requirement: The client and server must have the same version. The following driver packs are compatible with version 1.54: ACQUITY UPLC Systems June 2011 Driver Pack (DP3) ACQUITY UPC 2 System May 2014 Driver Pack ACQUITY Arc Driver Pack version 1.65 Waters Driver Pack 4 (DP4) through DP4 Supplemental Release 1 (DP4 SR1) See also: Driver Pack 4 Installation and Configuration Guide (part number ). Empower Workstation and LAC/E 32 support An Empower Workstation or LAC/E 32 can support four systems, where a total of one ACQUITY QDa Detector and one PDA detector are recommended. Installing the ACQUITY QDa ICS Recommendation: Back up all of your Empower projects, library information, and databases before installing this software. In addition to providing software for the QDa, this driver pack includes software for the isocratic solvent manager (ISM). The ACQUITY QDa ICS can be upgraded to version 1.54 from any previous version of ICS or firmware. You can install this software release from the software DVD. Alternatively, if you have a support plan, you can download the software from the Waters Web site. To do so, visit and click Services & Support > Support > Software Downloads. To install ACQUITY QDa ICS: 1. Save all data, exit all applications, and then restart the computer. 2. Log in to the computer using an account with local administrator privileges. 3. Insert the DVD-ROM into the drive. 4. If the DVD-ROM executes the Autorun program, go to step 5; otherwise, browse to the DVD drive in Windows Explorer, right-click Setup.exe, and select Run as administrator. Note: If the DVD or download includes a zip file, double-click it, extract the files to a temporary folder, and then run Setup.exe as described in step On the Welcome page of the Deployment Manager, click Next, and then select Install/Upgrade. 6. Confirm that you have read the release notes, and then click Next. 7. Accept the Waters Software License Agreement, and then click Next. 8. Select Typical. Result: A list of the instrument drivers that will be installed on the computer appear along with a list of the firmware files that will be deployed to the computer. Page 3 of 12

4 9. Review the list of instrument drivers for accuracy, and then click Next. Tip: To change settings before installing, click Back. 10. After installation is complete, click Finish. 11. Restart the computer. Verifying the installation To verify the installation, select Start > All Programs > Empower > Empower Installation Log. The installation of any option or service pack is documented in the checksum.txt file on your Empower computer. This file indicates the installation of any option or service pack. It also verifies the integrity of disk files by comparing their current CRC and sizes with original values recorded during installation of the base software and any installed option or service pack. The checksum.txt file is created when you run the Verify Files utility by selecting Start > Programs > Empower > Verify Files. To display the checksum.txt file, navigate to the \Empower\Script directory, and then double-click the checksum.txt file. Uninstalling the ICS To ensure successful file verification following the removal of the ICS, you must reboot the computer before you uninstall the ICS. To remove the ACQUITY QDa ICS from the system: 1. Reboot the computer. 2. Log on to the computer using an account that has local administrative privileges. 3. Insert the ACQUITY QDa ICS media in the media drive. 4. Follow the prompts to remove the control software from the system. The registry and new CRC checksums are updated for the Empower installation. No Oracle or system files are affected. System considerations When you click Stop Flow in the control panel for a module, only the flow for that module stops. You can stop flow for all pump modules in the system via the Console window. Page 4 of 12

5 Updates in version 1.54 This release adds support for a new embedded PC (EPC) in the QDa mass detector. The new EPC board can be identified by its part number, The obsolescence of the previous EPC prompts this update, which introduces no change in functionality. If an old QDa board is replaced by a board with the new EPC, you must upgrade the ICS to version Issues fixed in this release This section lists the software defects corrected in this release. The defects corrected in version 1.51, 1.52, and 1.53 are also shown, for completeness. The SCR numbers identify software issues that Waters personnel monitor within a system change request tracking tool. Fixed in version 1.54 SCR Number User mass scale calibration is now applied for SIR data channels instead of factory mass scale calibration The IntelliStart Resolution Check algorithm has been improved further in relation to fast scanning of peak widths Limits for the Critical Readback Turbo Power check are changed to accommodate hardware improvements made to the pumping block The Turbo Startup Healthcheck status is now displayed via the LED on the instrument s front panel The Low Turbo Power Healthcheck is now recorded in the console logs, if activated The resolution expiration date is removed when a resolution setup is required. Fixed in version 1.53 SCR Number Previously, when the QDa ran a sample list, the Operate control in the QDa instrument panel in Run Samples and in the console was enabled. Now the control is disabled Previously, the Recalibrate menu was enabled when the software reported operational problems involving the QDa (referred to as health problems). Now the Recalibrate menu is disabled when such a problem is reported Previously, the tooltip for the QDa image in the console showed the firmware version as unknown. Now the tooltip shows the correct firmware version Previously, Connections INSIGHT support for the QDa was unavailable. Now such support is available when Waters Driver Pack 4 is used Previously, the resolution- and calibration- check report was unavailable on Chinese, Japanese, and Korean systems. Now the report is available in [install drive]:\empower\instruments\log\ms Reports Previously, the QDa was not visible in the console when the SQ and TQ were also installed. Now the QDa appears in the console when the SQ and TQ are also installed Previously, the VxWorks folder was not included in Empower s Verify Files utility. Now the folder is included Previously, the QDa acquired data only when a dot was used as the decimal separator in the Windows regional settings. Now a comma can also be used. Page 5 of 12

6 30461 Previously, the message This is a message appeared in the QDa post-run report. Now the message does not appear Previously, recalibrating the instrument following an error state caused the embedded software to crash. Now the embedded software continues to run normally Previously, a printout of the Tune Page showed an incorrect value for the nominal rod polarity. Now the correct value appears Previously, when the QDa was in Standby mode, its status LED flashed red. Now the LED is unlit during standby mode Previously, the Turbo Power checks limit was 38W. Now the limit is 37W The WatersFileService Health Monitor Check has been added, to verify that the QDa starts up successfully Previously, when the instrument was allowed to operate at 600 C with no sample, a message stating Source Temperature Fail inappropriately appeared. Now the message does not appear Previously, when you saved the target registry, the nominal rod polarity value was not saved. Now that value is saved Previously, when the QDa was correcting errors, you could put the instrument in Standby mode. Now the Standby button is disabled when the instrument is correcting errors Previously, on Chinese, Japanese, and Korean computers, the image in the console showed an SQD, not a QDa. Now the image is correct Previously, the unit for the Set Rotation Speed was Hertz. Now the unit is revolutions per minute Previously, the High Voltage PSU LED in the Tune Page failed to show green when the QDa was in Operate mode. Now the LED shows green when the QDa is in Operate mode Previously, the Japanese channel name included a double-byte colon. Now the channel name includes a single-byte colon Previously, the Tune Page failed to display the user name, user group, and Empower project. Now these items appear in the window s title bar Previously, when an alternative calibrant was used, the calibration report stated only that an alternative calibrant was used. Now in addition to stating that an alternative was used, the report names the alternative calibrant Previously, the software did not recall the last successful alternative calibrant when switching from an alternative calibrant to the internal calibrant and, finally, to an alternative calibrant. Now the software recalls the last successful alternative calibrant Previously, when it acquired a single SIR data channel in mass-detector mode, the detector produced a noisy signal. Now the signal is acquired correctly Previously, following re-calibration of the QDa mass scale, the existing calibration continued to be applied. Now, the new mass scale calibration is applied. Page 6 of 12

7 Fixed in version 1.52 SCR Number Previously, when you rebooted the LAC/E 32 without rebooting the QDa and then attempted to run samples, the message Preparing for injection persisted indefinitely. Now, the system runs normally after a LAC/E 32 reboot. Fixed in version 1.51 SCR Number Previously, if you deselected Operate on the General tab of a QDa instrument method in Mass Detector mode, switched to Advanced mode, and then ran the method, the QDa transitioned to Standby mode, which stops flow and sample injections. Now, the QDa remains in the Operate mode, and sample injections can proceed Previously, if Empower software was installed to a drive other than C:\, the QDa did not function properly. Now, the QDa functions properly regardless of the drive to which Empower software is installed Previously, when you rebooted the QDa, the QDa settings could be erased or corrupted. The QDa settings are no longer susceptible to erasing or corrupting Previously, when you changed control software from MassLynx to Empower, the instrument settings could be erased or corrupted. The settings are no longer susceptible to erasing or corrupting Previously, when you rebooted the QDa, the Tune Page Ion Mode tab, used by Waters field service engineers, was sometimes unavailable. The tab is now available to service engineers Previously, for some QDa detectors, a problem involving the internal calibration beam saturated the internal calibration benchmark test that is performed when the QDa is powered-on. This prevented accurate Full Width Half Height (FWHH) measurement. Now, saturation no longer occurs when the FWHH measurements are taken Previously, the QDa did not receive the radio-frequency value during startup when the instrument was optimized to the default value, Mhz, causing radio-frequency amplitude failure. The defect is corrected Previously, settings specified in a QDa instrument method for the SIR cone voltage were ignored, and a voltage of 30 was universally applied. Now, the specified cone voltage settings are applied Previously data was, by default, acquired using the continuum mode. Now, the default for data acquisition is the centroid mode. Thus, when you develop a new method in the mass detector mode, the resulting data are acquired using centroid mode. If you run an existing method without changes, data acquisition continues in the continuum mode. When you save the method, however, the data collection mode is updated to centroid. You can nevertheless override the data collection mode in the QDa instrument method when the instrument is in the advanced mode. To do so, click a channel in the table, and then click Show Functions Details. The Function Details dialog box, which includes the Data field where you can specify continuum or centroid mode, appears. Note, however, that the continuum mode collects many more data points than the centroid mode, and continuum data require more time for the software to process. Page 7 of 12

8 30094 Previously, only the words Positive and Negative were used to identify ionization mode. Now, a plus sign ("+") or a minus sign ("-") appear in the channel description, for example, QDa Positive (+) Scan Previously, the source pressure test, performed when the QDa is powered on, sometimes incorrectly reported a failed result. Now, when the QDa checks the vacuum system, its reported results are correct Previously, when you updated an expired resolution and calibration, the error Instrument Calibration has expired persisted, even though the expiration date was properly updated. Now, the error does not appear after correctly updating the resolution and calibration. Known issues in this release The following tables contain a list of the known software issues and workarounds for this release of the ACQUITY QDa ICS. The SCR (software change request) numbers identify software issues that Waters personnel monitor within a system change request tracking tool. Installation SCR Number After installing the QDa Driver Pack, the autoloader opens and can occasionally crash. This issue occurs infrequently. If it occurs, relaunch the autoloader software After installing the UPC 2 and QDa driver packs, you cannot uninstall Empower software. Uninstall the ELS driver before the other drivers. Console SCR Number Whilst a sample set is running, the Operate control in the QDa instrument panel of the console remains available for use. Whilst a sample set is running, do not click Operate within the QDa instrument panel of the console. Switching the instrument out of operate during the running of a sample set will result in the sample set being aborted In the Chinese and Korean console, the plot x-axis label Mins is not displaying properly. None On the QDa Detector console page, when you click Maintain > Recalibrate, in some circumstances this does not recalibrate the system. Before attempting to perform a Recalibrate, ensure that there are no issues reported in the QDa health status. Page 8 of 12

9 SCR Number The console plots can occasionally fail to update. The instrument data collection is not affected and the plots will resume as normal after a short period Clicking between the ACQUITY log entries can cause an error in the Console. None Cancelling a print of the ACQUITY logs can cause an error in the Console. None Printing from the ACQUITY logs will print all entries. To print a single log, it should first be printed to PDF and then the relevant log selected. Instrument and instrument system management SCR Number You cannot differentiate between the ACQUITY Binary Solvent Manager and ACQUITY Isocratic Solvent Manager data channels in the method editor. Avoid using the same data channels for the two pumps. If all the channels are required for diagnostic testing, it is recommended to use the instrument console application If the ACQUITY Isocratic Solvent Manager (ISM) is configured as part of a system being run, but is not included in the instrument method used, the ISM automatically turns on at a flow rate of 1 ml/min when the run begins. It is recommended the Empower instrument method editor feature for turning the instrument off is not used for any pump defined within the system. If you want to remove the instrument from participating in the acquisition, it is recommended that you remove the instrument from the chromatographic system s definition When you select the wet prime function when running a sample set, both the ACQUITY Binary Solvent Manager and ACQUITY Isocratic Solvent Manager pumps prime. You cannot specify which pump to prime. None The QDa Detector status is shown as Unknown. This issue occurs infrequently. To recover, reboot the QDa Detector Non OS Admin users cannot run ACQUITY Plot Viewer from C:\Program Files (x86)\waters Instruments\Bin folder. None. Page 9 of 12

10 SCR Number The Rear Panel Event Out 2 will not pulse, it will only allow the output to be closed or opened when triggered. None If a Source Pressure Test fails during the startup process then the Critical Readback Check will not be carried out. If the Critical Readback Check had previously failed, it will still show as having failed once the Source Pressure Test has passed. At this point the instrument will need to be rebooted to allow for a new Source Pressure Test When you are operating a LAC/E 32 acquisition server using a standard Windows user account, two error messages indicating that the AcquityISMServer and the KestrelUI application stopped working can appear when you submit an ihelp request using the newly supported Connections INSIGHT facility (see 29551). You can dismiss these messages without adversely affecting data acquisition. Use an administrator account on a LAC/E 32 to submit an ihelp request The newly supported Connection INSIGHT facility (see 29551) does not report error messages on Japanese installations During evaluation of version 1.53, it was found that the QDa would not start up correctly when connected to a particular computer running the Japanese version of Windows XP. Other computers did not exhibit this problem. If this issue occurs, re-install version You cannot properly save the QDa event table in Mass Detector mode. Use Advanced mode for timed events. Alternatively, you can save the method events in Advanced mode, select Mass Detector mode, and then reenter the other method parameters. Acquisition SCR Number During a long run, an error message appears stating Unhandled Exception has occurred in a component in your application. If you click Continue, the application will ignore this error and attempt to continue. Out of Memory. This issue occurs infrequently. If it occurs, close the Run samples window and then reopen it. Page 10 of 12

11 Anti-virus considerations If you are using real-time virus scanning, exclude the data system and instruments folders and subfolders; usually C:\Empower and its sub-folders for Empower 2 and Empower 3 installations. Some realtime virus scanners can mistake normal data acquisition and instrument control for virus activity and thus interfere with proper operations. Full-system scans and live updates can be network-intensive, diskintensive, and CPU-intensive, and they can also interfere with normal data acquisition. Schedule scans and updates for idle times when data acquisition does not occur. Certain anti-virus program features like "intrusion prevention" and "tamper protection" can also interfere with normal operation. Disable them as well. Compliance recommendations Any time you install, change, or uninstall software or system modules in a regulated environment, Waters recommends that you follow your organization's approved standard operating procedures. A risk-based review may assist you in a regulated environment to evaluate changes detailed in the release notes. Using company SOPs, determine if any documentation updates and requalification of the system modules, chromatographic system, or chromatographic data system (CDS) are required. Instrument update classification (minor update) Waters considers this update to be a minor change from previous versions. This update is designed to not alter the physical specifications of instruments or systems. Waters expects that methods run and data generated on instruments before and after the update will be consistent. As with any change, you should evaluate the impact of this update to your instruments or systems. Instrument requalification options after software/firmware change Consider using the requalification options outlined below to verify hardware installation and correct instrument operation: To confirm that the firmware files on the system modules were installed correctly, compare the checksum values in the product release notes to the checksum values displayed in the console. To confirm instrument system operation with any newly loaded software, driver, or firmware, consider performing an operational qualification for the updated instrument system. To confirm performance, control, and communications of the instrument system, consider conducting a performance qualification (user or vendor) or system suitability test. When multiple, identical instruments are involved, consider a risk-based approach to qualification activities. Requalification with Waters' Total Assurance Plans The Waters' Total Assurance Plan (TAP) with System Qualification Option covers upgrades and requalification of the instrument driver, software, firmware, or hardware in these cases: During yearly requalification, as provided in the plan. If installing this release is required for operation of a new module or system, where qualification of the new module or system is covered by the plan. Requalification of the CDS software and computers after a driver upgrade may or may not be included in your TAP. Review your TAP to determine which services are covered and which are not covered. For situations not covered by the plan, Waters can perform the qualification, but additional charges will apply. Page 11 of 12

12 Contact Waters Technical Service If you are located in the USA or Canada, report malfunctions or other problems to Waters Technical Service ( ). If you are located elsewhere, phone the Waters corporate headquarters in Milford, Massachusetts (USA), or contact your local Waters subsidiary. Our Web site includes phone numbers and addresses for Waters locations worldwide. Page 12 of 12

Alliance e2695 Separations Module Firmware Version 3.04

Alliance e2695 Separations Module Firmware Version 3.04 *716005261* *Rev.A* Alliance e2695 Separations Module Firmware Version 3.04 Release Notes 716005261 Revision A Copyright Waters Corporation 2018 All rights reserved General information Copyright notice

More information

MassLynx 4.1 SCN 804. Release Notes. SQ Detector 2

MassLynx 4.1 SCN 804. Release Notes. SQ Detector 2 R E L E A S E N O T E S MassLynx 4.1 SCN 804 Release Notes SQ Detector 2 43464 Rev 01 Copyright Waters Corporation 2011. All rights reserved. Copyright 2011 Waters Corporation 43464_1. Waters is a registered

More information

Waters Driver Pack 2016 Release 1

Waters Driver Pack 2016 Release 1 Waters Driver Pack 2016 Release 1 Installation and Configuration Guide 715005197 Revision A Copyright Waters Corporation 2016 All rights reserved General information Copyright notice 2016 WATERS CORPORATION.

More information

The instrument drivers and firmware in Driver Pack 4 support these ACQUITY UPLC systems: ACQUITY UPLC Systems with 2D Technology

The instrument drivers and firmware in Driver Pack 4 support these ACQUITY UPLC systems: ACQUITY UPLC Systems with 2D Technology R E L E A S E N O T E S Waters Driver Pack 4 Driver Pack 4 includes instrument drivers and firmware required to control Waters instruments, as well as helpful utility applications. The driver pack is compatible

More information

Waters Headspace Control Option, Version 3.0

Waters Headspace Control Option, Version 3.0 R E L E A S E N O T E S Waters Headspace Control Option, Version 3.0 Version 3.0 of the Headspace Control Option (HCO) contains instrument drivers for Agilent Headspace Samplers and a toolkit program that

More information

Waters Empower 2 Service Pack G

Waters Empower 2 Service Pack G R E L E A S E N O T E S Waters Empower 2 Service Pack G Empower 2 Service Pack G enhances LDAP capabilities. These release notes also address the following information and instructions: System and software

More information

Empower 3 Feature Release 3 Hotfix 1 Release Notes

Empower 3 Feature Release 3 Hotfix 1 Release Notes *716004786* *Rev.A* Empower 3 Feature Release 3 Hotfix 1 Release Notes 716004786 Revision A Copyright Waters Corporation 2016 All rights reserved General information Copyright notice 2016 WATERS CORPORATION.

More information

Agilent 7890 GC Instrument Control Software Version 2.5

Agilent 7890 GC Instrument Control Software Version 2.5 R E L E A S E N O T E S Agilent 7890 GC Instrument Control Software Version 2.5 The Agilent 7890 GC Instrument Control Software version 2.5 enables dual-tower injection capability and provides support

More information

Agilent 7890 GC Instrument Control Software Version 2.01

Agilent 7890 GC Instrument Control Software Version 2.01 R E L E A S E N O T E S Agilent 7890 GC Instrument Control Software Version 2.01 These release notes explain how to install Waters Instrument Control Software (ICS), version 2.01, for the Agilent 7890

More information

Agilent 7890 GC Instrument Control Software Version 2.6

Agilent 7890 GC Instrument Control Software Version 2.6 R E L E A S E N O T E S Agilent 7890 GC Instrument Control Software Version 2.6 Version 2.6 of the instrument control software (ICS) for the Agilent 7890 GC addresses several issues, provides support for

More information

Waters Fraction Collector III with Empower Instrument Control Software Version 1.00

Waters Fraction Collector III with Empower Instrument Control Software Version 1.00 R E L E A S E N O T E S Waters Fraction Collector III with Empower Instrument Control Software Version 1.00 These release notes explain how to install Waters Instrument Control Software (ICS), version

More information

For additional information, see these documents on the Breeze 2 software DVD:

For additional information, see these documents on the Breeze 2 software DVD: R E L E A S E N O T E S Waters Breeze 2 Read these release notes before you install and operate Breeze 2. They contain information necessary to ensure successful liquid chromatography results, including:

More information

MassLynx 4.2 SCN 989 Release Notes

MassLynx 4.2 SCN 989 Release Notes MassLynx 4.2 SCN 989 Release Notes Page 1 of 20 Waters, THE SCIENCE OF WHAT S POSSIBLE., ACQUITY, Alliance, MassLynx, nanoacquity, and Xevo are registered trademarks and OpenLynx, Quanpedia, and TargetLynx

More information

Waters Driver Pack 2018 Release 1

Waters Driver Pack 2018 Release 1 Waters Driver Pack 2018 Release 1 Installation and Configuration Guide 715005659 Revision B Copyright Waters Corporation 2018 All rights reserved General information Copyright notice 2018 WATERS CORPORATION.

More information

Waters Empower Software A1100 System Configuration Instructions for Ethernet Control

Waters Empower Software A1100 System Configuration Instructions for Ethernet Control Waters Empower Software A1100 System Configuration Instructions for Ethernet Control In order to configure an A1100 system for Ethernet control or to acquire A1100 DAD 3D data in Empower Software, it is

More information

ACQUITY UPLC Local Console Controller Instructions

ACQUITY UPLC Local Console Controller Instructions ACQUITY UPLC Local Console Controller Instructions Contents Topic Page Product description and function Operation in a GXP regulated environment Preparing to install the LCC 3 Installing the LCC 3 Navigating

More information

Waters 2707 Autosampler ICS v1.20 Installer

Waters 2707 Autosampler ICS v1.20 Installer R E L E A S E N O T E S Waters 2707 Autosampler ICS v1.20 Installer Read these release notes before you install and operate the Waters 2707 autosampler instrument control software (ICS). They contain information

More information

SCN 848 Release Notes

SCN 848 Release Notes R E L E A S E N O T E S SCN 848 Release Notes 43198 Rev 1 Copyright Waters Corporation 2011. All rights reserved. 43198 Rev 1. Waters is a registered trademark, Waters the Science of What s Possible and

More information

LTQ Velos SP3 Release Notes

LTQ Velos SP3 Release Notes LTQ Velos 2.6.0 SP3 Release Notes These release notes briefly list changes in the LTQ Velos 2.6.0 SP3 experiment setup, tuning, and calibration application; fixes to previously reported issues; and known

More information

Agilent Instrument Control Framework (ICF) Update A Driver Update 1 (DU1) Hot Fix 2 (HF2)

Agilent Instrument Control Framework (ICF) Update A Driver Update 1 (DU1) Hot Fix 2 (HF2) R E L E A S E N O T E S Agilent Instrument Control Framework (ICF) Update A.02.03 Driver Update 1 (DU1) Hot Fix 2 (HF2) These release notes explain the installation of the updated Agilent Instrument Control

More information

ACQUITY UPLC Systems June 2011 Driver Pack and Firmware Updates

ACQUITY UPLC Systems June 2011 Driver Pack and Firmware Updates R E L E A S E N O T E S ACQUITY UPLC Systems June 2011 Driver Pack and Firmware Updates What s new in this release... 3 Reasons for installing this software release... 3 Installation Notes... 3 Recommended

More information

SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.31: PSS

SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.31: PSS SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.31: PSS0051-05 Introduction This document details the installation and known issues with OMNISEC v10.31. This release is a maintenance release to

More information

Empower 2 Feature Release 5 Upgrade Guide

Empower 2 Feature Release 5 Upgrade Guide Empower 2 Feature Release 5 Upgrade Guide 715001799/Revision A Copyright Waters Corporation 2008 All rights reserved Copyright notice 2008 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND

More information

SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.30: PSS

SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.30: PSS SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.30: PSS0051-04 Introduction This document details the installation and known issues with OMNISEC v10.30. This release is a major release to fix a

More information

Breeze 2 Installation and Configuration Guide

Breeze 2 Installation and Configuration Guide Breeze 2 Installation and Configuration Guide 715001951/Revision B Copyright Waters Corporation 2008 2009 All rights reserved Copyright notice 2008 2009 WATERS CORPORATION. PRINTED IN THE UNITED STATES

More information

What s New in Empower 3

What s New in Empower 3 What s New in Empower 3 Revision A Copyright Waters Corporation 2010 All rights reserved Copyright notice 2010 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IN IRELAND. ALL RIGHTS RESERVED.

More information

Agilent OpenLAB CDS Waters Alliance HPLC Instrument Control Add-On. Installation, Upgrade and Migration Guide

Agilent OpenLAB CDS Waters Alliance HPLC Instrument Control Add-On. Installation, Upgrade and Migration Guide Agilent OpenLAB CDS Waters Alliance HPLC Instrument Control Add-On Installation, Upgrade and Migration Guide Notices Agilent Technologies, Inc. 2013, 2014 No part of this manual may be reproduced in any

More information

Software Update Notification

Software Update Notification Software Update Notification PSS0002-31: Mastersizer 2000 v5.60 software Abstract This document details the release of software PSS0002-31. This is version 5.60 of software for the Mastersizer 2000 laser

More information

Analyst QS. Administrator s Guide. Part Number: B July 2004

Analyst QS. Administrator s Guide.   Part Number: B July 2004 Analyst QS Administrator s Guide Part Number: 1001930 B July 2004 http://www.appliedbiosystems.com This document is provided to customers who have purchased MDS Sciex equipment to use in the operation

More information

Upgrading your IntelliTrax System Firmware

Upgrading your IntelliTrax System Firmware Changes in v06502 This maintenance release is the first version that includes support for Dynamic Polarization (DP New!). It also improves tracking accuracy for small aperture reading heads and small colorbar

More information

Empower Software and Security Software Interactions

Empower Software and Security Software Interactions TECN134790604 Rev. 00 Page 1 of 7 Empower Software and Security Software Interactions This Technical Note discusses how Empower software functions and examines potential interactions with anti-virus software

More information

Fiery proserver User Guide

Fiery proserver User Guide Fiery proserver User Guide Congratulations on your purchase of the EFI Fiery proserver, a powerful production solution for your printer. Before you begin Make sure that you have: About this document Installed

More information

xponent Version 4.2 RELEASE NOTES

xponent Version 4.2 RELEASE NOTES xponent Version 4.2 RELEASE NOTES Copyright Luminex Corporation, 1996-2013. All rights reserved. Part Number: 89-30000-00-462 Disclaimer Subject to change without notice. Although prepared to ensure accuracy,

More information

Abila MIP. Human Resource Management Installation Guide

Abila MIP. Human Resource Management Installation Guide Human Resource Management Installation Guide This is a publication of Abila, Inc. Version 2017.2 2017 Abila, Inc. and its affiliated entities. All rights reserved. Abila, the Abila logos, and the Abila

More information

Symantec pcanywhere 12.5 SP4 Release Notes

Symantec pcanywhere 12.5 SP4 Release Notes Symantec pcanywhere 12.5 SP4 Release Notes Symantec pcanywhere 12.5 SP4 Release Notes The software described in this book is furnished under a license agreement and may be used only in accordance with

More information

ZENworks 2017 Full Disk Encryption Pre-Boot Authentication Reference. December 2016

ZENworks 2017 Full Disk Encryption Pre-Boot Authentication Reference. December 2016 ZENworks 2017 Full Disk Encryption Pre-Boot Authentication Reference December 2016 Legal Notice For information about legal notices, trademarks, disclaimers, warranties, export and other use restrictions,

More information

Rotor-Gene AssayManager v1.0 Quick-Start Guide

Rotor-Gene AssayManager v1.0 Quick-Start Guide September Rotor-Gene AssayManager v1.0 Quick-Start Guide For installation and setup of Rotor-Gene AssayManager v1.0 Sample to Insight Contents Installing Rotor-Gene AssayManager v1.0...3 Minimum computer

More information

ChromQuest 4.2 Chromatography Data System

ChromQuest 4.2 Chromatography Data System ChromQuest 4.2 Chromatography Data System Installation Guide CHROM-97200 Revision A April 2005 2006 Thermo Electron Corporation. All rights reserved. Surveyor is a registered trademark and ChromQuest is

More information

Agilent 6400 Series Triple Quadrupole LC/MS System

Agilent 6400 Series Triple Quadrupole LC/MS System Agilent 6400 Series Triple Quadrupole LC/MS System Quick Start Guide Where to find information 4 Getting Started 6 Step 1. Start the Data Acquisition software 7 Step 2. Prepare the LC modules 13 Step 3.

More information

Fiery proserver User Guide

Fiery proserver User Guide Fiery proserver User Guide Congratulations on your purchase of the Fiery proserver, a powerful production solution for your printer. Before you begin Make sure that you have: About this document Installed

More information

Pearl Echo.Suite Quick Start Installation Guide

Pearl Echo.Suite Quick Start Installation Guide WORKSTATION SERVER Pearl Echo.Suite Quick Start Installation Guide Your initial test setup should take between 5 and 15 minutes. For detailed automated deployment instructions and configuration settings

More information

Waters Empower Software Build 1154 Toolkit Service Pack Release Notes

Waters Empower Software Build 1154 Toolkit Service Pack Release Notes Waters Empower Software Build 1154 Toolkit Service Pack Release Notes These release notes are intended as supplemental information and should be used in conjunction with the Empower Software Release Notes.

More information

Copyright Autodesk, Inc.

Copyright Autodesk, Inc. Customer and Administration Building, Ditzingen, Germany - Design by Barkow Leibinger Architects Copyright 2003-2008 Autodesk, Inc. Table of Contents 1 Introduction... 3 What is Facility Manager?... 3

More information

ZENworks 2017 Update 1 Full Disk Encryption Pre-Boot Authentication Reference. July 2017

ZENworks 2017 Update 1 Full Disk Encryption Pre-Boot Authentication Reference. July 2017 ZENworks 2017 Update 1 Full Disk Encryption Pre-Boot Authentication Reference July 2017 Legal Notice For information about legal notices, trademarks, disclaimers, warranties, export and other use restrictions,

More information

SAS Installation Instructions Windows 2003, XP, 2000, NT. Workstation Installation Guidelines

SAS Installation Instructions Windows 2003, XP, 2000, NT. Workstation Installation Guidelines UCit Instructional and Research Computing, Software Distribution Office, 303B Zimmer Hall, Cincinnati, OH 45221-0088. Phone: (513) 556 9068 Email: Software@uc.edu SAS 9.1.3 Installation Instructions Windows

More information

WEBSEWSS SINGLE SERVER INSTALLATION INSTRUCTIONS

WEBSEWSS SINGLE SERVER INSTALLATION INSTRUCTIONS 2300 East 14th Street Tulsa, Oklahoma 74104 USA Phone: (918) 749-1119 Fax: (918) 749-2217 e-mail: info@statsoft.com WEB: http://www.statsoft.com Notes: WEBSEWSS SINGLE SERVER INSTALLATION INSTRUCTIONS

More information

Xcalibur. QuickQuan. User Guide. XCALI Revision C July For Research Use Only Not for use in Diagnostic Procedures

Xcalibur. QuickQuan. User Guide. XCALI Revision C July For Research Use Only Not for use in Diagnostic Procedures Xcalibur QuickQuan User Guide XCALI-97195 Revision C July 2007 For Research Use Only Not for use in Diagnostic Procedures 2007 Thermo Fisher Scientific Inc. All rights reserved. Microsoft, Access, Excel,

More information

LUXRIOT EVO Installation Guide

LUXRIOT EVO Installation Guide LUXRIOT EVO Installation Guide October 18, 2016 A&H Software House, Inc. Luxriot EVO S Installation Guide 1 Table of Contents 1. Introduction 2 2. Getting Started 3 3. Prerequisites 4 4. Installation 5-10

More information

Employee Web Services. Installation Guide

Employee Web Services. Installation Guide Employee Web Services This is a publication of Abila, Inc. Version 2017.x 2016 Abila, Inc. and its affiliated entities. All rights reserved. Abila, the Abila logos, and the Abila product and service names

More information

pclamp Electrophysiology Data Acquisition and Analysis Software

pclamp Electrophysiology Data Acquisition and Analysis Software pclamp Electrophysiology Data Acquisition and Analysis Software Version 10.5.1 Software Release Notes December 2014 www.moleculardevices.com This document is provided to customers who have purchased Molecular

More information

Frequently Asked Questions UNIFI v1.9 Service Release 4 Upgrade

Frequently Asked Questions UNIFI v1.9 Service Release 4 Upgrade Frequently Asked Questions UNIFI v1.9 Service Release 4 Upgrade Q. What are the features and benefits of upgrading to UNIFI v1.9 SR4? A. UNIFI v1.9 SR4 incorporates improvements across all application

More information

Empower 3 Installation, Configuration, and Upgrade Guide /Revision A Copyright Waters Corporation All rights reserved

Empower 3 Installation, Configuration, and Upgrade Guide /Revision A Copyright Waters Corporation All rights reserved Empower 3 Installation, Configuration, and Upgrade Guide 715005081/Revision A Copyright Waters Corporation 2002-2016 All rights reserved Page ii General Information Copyright notice 2002-2016 WATERS CORPORATION.

More information

Nortel Media Application Server 5.1. Installation NN

Nortel Media Application Server 5.1. Installation NN NN44450-300 Document status: Standard Document issue: 01.03 Document date: 15 October 2008 Product release: Release 5.1 Job function: and Commissioning Type: Technical document Language type: English All

More information

MapMarker Plus Desktop Installation Guide

MapMarker Plus Desktop Installation Guide MapMarker Plus 11.0 Desktop Installation Guide Information in this document is subject to change without notice and does not represent a commitment on the part of the vendor or its representatives. No

More information

Relius Administration Version 16.0 (and higher) Component Installation and Configuration. July 6, 2011

Relius Administration Version 16.0 (and higher) Component Installation and Configuration. July 6, 2011 Relius Administration Version 16.0 (and higher) Component Installation and Configuration July 6, 2011 Table Of Content Section Subject 1 Overview 2 Preliminary Steps 3 Installing the Oracle Client 4 Installing

More information

MapMarker Plus Developer Installation Guide

MapMarker Plus Developer Installation Guide MapMarker Plus 11.1 Developer Installation Guide Information in this document is subject to change without notice and does not represent a commitment on the part of the vendor or its representatives. No

More information

NETePay POSPAD. Moneris Canadian EMV Host. Installation & Configuration Guide V5.07. Part Number:

NETePay POSPAD. Moneris Canadian EMV Host. Installation & Configuration Guide V5.07. Part Number: NETePay POSPAD Installation & Configuration Guide Moneris Canadian EMV Host V5.07 Part Number: 8660.83 NETePay Installation & Configuration Guide Copyright 2006-2017 Datacap Systems Inc. All rights reserved.

More information

Features - SRM Windows File System Agent

Features - SRM Windows File System Agent Page 1 of 53 Features - SRM Windows File System Agent Table of Contents Overview System Requirements - SRM Windows File System Agent Installation Install the SRM Windows File System Agent Install the SRM

More information

Silk Performance Manager Installation and Setup Help

Silk Performance Manager Installation and Setup Help Silk Performance Manager 18.5 Installation and Setup Help Micro Focus The Lawn 22-30 Old Bath Road Newbury, Berkshire RG14 1QN UK http://www.microfocus.com Copyright 2004-2017 Micro Focus. All rights reserved.

More information

UPS WorldShip Install on a Workgroup Remote

UPS WorldShip Install on a Workgroup Remote PRE-INSTALLATION INSTRUCTIONS: Install UPS WorldShip on the Workgroup Admin. Temporarily disable any virus scan software that you may have installed. Request access to the network share drive created by

More information

Total Protection Service

Total Protection Service User Guide McAfee Total Protection Service for Microsoft Windows Home Server COPYRIGHT Copyright 2008 McAfee, Inc. All Rights Reserved. No part of this publication may be reproduced, transmitted, transcribed,

More information

Installation Instructions. Release Version 14.0 August 15 th, 2008

Installation Instructions. Release Version 14.0 August 15 th, 2008 Release Version 14.0 August 15 th, 2008 ARGUS Software: ARGUS Valuation - DCF The contents of this document are considered proprietary by ARGUS Software, the information enclosed and any portion thereof

More information

Fibre Channel Adapter STOR Miniport Driver for Windows. Table of Contents

Fibre Channel Adapter STOR Miniport Driver for Windows. Table of Contents Fibre Channel Adapter STOR Miniport Driver for Windows QLogic Corporation All rights reserved Table of Contents 1. Package Contents 2. OS Support 3. Supported Features 4. Using the Driver 4.1 Creating

More information

Trace Debug Tools Version 1.2 Installation Guide

Trace Debug Tools Version 1.2 Installation Guide Trace Debug Tools Version 1.2 Installation Guide Copyright 2000-2002 ARM Limited. All rights reserved. Proprietary Notice Words and logos marked with or are registered trademarks or trademarks owned by

More information

Empower 3 FR3 System Administrator's Guide / Revision A Copyright Waters Corporation All rights reserved

Empower 3 FR3 System Administrator's Guide / Revision A Copyright Waters Corporation All rights reserved Empower 3 FR3 System Administrator's Guide 715005082 / Revision A Copyright Waters Corporation 2002-2016 All rights reserved Page ii Empower 3 FR3 System Administrator's Guide Copyright notice 2002-2016

More information

Release Notes. Analyst TF Software

Release Notes. Analyst TF Software Release Notes Analyst TF 1.5.1 Software Release Date: April 2011 This document is provided to customers who have purchased AB SCIEX equipment to use in the operation of such AB SCIEX equipment. This document

More information

SoftPro Installation Guide

SoftPro Installation Guide SoftPro Installation Guide UPGRADING TO STANDARD EDITION VERSION 12.0 MICROSOFT JET DATABASE ENGINE 4800 FALLS OF NEUSE ROAD, SUITE 400 / RALEIGH, NC 27609 phone: 800 848 0143 / fax: 919 755 8350 / e mail:

More information

Sage Fixed Assets Single User

Sage Fixed Assets Single User Single User 2018.0 Installation guide October 2017 Last updated October 17, 2017 2017 The Sage Group plc or its licensors. All rights reserved. Sage, Sage logos, and Sage product and service names mentioned

More information

Impact Analysis for Software changes in OpenLAB CDS A.01.04

Impact Analysis for Software changes in OpenLAB CDS A.01.04 Impact Analysis for Software changes in OpenLAB CDS A.01.04 Document Information: Filename Product Identifier Product Revision Project Identifier Document Revision Impact_Analysis_OpenLAB_CDS_A0104 OpenLAB

More information

Spectrometer Visible Light Spectrometer V4.4

Spectrometer Visible Light Spectrometer V4.4 Visible Light Spectrometer V4.4 Table of Contents Package Contents...3 Trademarks...4 Manual Driver and Application installation...5 Manual Application Installation...6 First Start of the Application...8

More information

User Guide 701P Wide Format Solution Wide Format Scan Service

User Guide 701P Wide Format Solution Wide Format Scan Service User Guide 701P44865 6204 Wide Format Solution Wide Format Scan Service Xerox Corporation Global Knowledge & Language Services 800 Phillips Road Bldg. 845-17S Webster, NY 14580 Copyright 2006 Xerox Corporation.

More information

Contents. 4. Getting Started 4.1 Connecting your 2200 TapeStation. 5. Getting Help 5.1 Using Online Help 5.2 Getting Support

Contents. 4. Getting Started 4.1 Connecting your 2200 TapeStation. 5. Getting Help 5.1 Using Online Help 5.2 Getting Support README Agilent 2200 TapeStation Software Contents 1. Introduction 1.1 New features with A.02.01(SR1) 1.2 New features with A.02.01 1.3 New features with A.01.05 1.4 New features with A.01.04 1.5 New features

More information

PROFESSIONAL. Users Guide. Version 3.5

PROFESSIONAL. Users Guide. Version 3.5 PROFESSIONAL Users Guide Version 3.5 Table of Contents 1 System Overview... 3 2 System Requirements... 3 3 Contents in ECMTUNE System Box... 3 4 Installation Information... 4 5 Registration Information...

More information

Anti-Executable Enterprise User Guide

Anti-Executable Enterprise User Guide 1 2 Last modified: October, 2015 1999-2015 Faronics Corporation. All rights reserved. Faronics, Deep Freeze, Faronics Core Console, Faronics Anti-Executable, Faronics Device Filter, Faronics Power Save,

More information

EASY-nLC Virtual Instrument for Xcalibur 2.1.x Quick Reference Guide

EASY-nLC Virtual Instrument for Xcalibur 2.1.x Quick Reference Guide EASY-nLC Virtual Instrument for Xcalibur 2.1.x Quick Reference Guide This guide describes how to install, configure, and use the Thermo EASY-nLC Virtual Instrument (VI) software with the Thermo Xcalibur

More information

Full User Manual and Quick Start Guide

Full User Manual and Quick Start Guide Full User Manual and Quick Start Guide 2 W hile every precaution has been taken in the preparation of this manual, we assume no responsibility for errors or omissions. Neither, is any liability assumed

More information

SIMATIC Automation License Manager Manual 02/2008 A5E

SIMATIC Automation License Manager Manual 02/2008 A5E s Contents SIMATIC Automation License Manager Product Overview 1 Installation 2 Working with the Automation License Manager 3 Glossar Index Manual 02/2008 A5E02128430-01 Safety Guidelines This manual contains

More information

QUICK START. DevCom2000 User Manual

QUICK START. DevCom2000 User Manual QUICK START DevCom2000 uses Device Descriptions (DDs) to access data stored in the memory of the smart field device. These DDs are developed by the manufacturer for their products and, in turn, distributed

More information

EMC ApplicationXtender Reports Management 6.0

EMC ApplicationXtender Reports Management 6.0 EMC ApplicationXtender Reports Management 6.0 Administrator s Guide 300-008-283 REV A01 EMC Corporation Corporate Headquarters: Hopkinton, MA 01748-9103 1-508-435-1000 www.emc.com Copyright 1994-2009 EMC

More information

New features and Workflow Enhancements Empower 3 Software

New features and Workflow Enhancements Empower 3 Software New features and Workflow Enhancements Empower 3 Software Mogens Hallas Nordic User Training 2011 Långvik, September 6th to 9th, 2011 2011 Waters Corporation 1 Empower 3 Architecture and operating systems

More information

ER/Studio Business Architect

ER/Studio Business Architect Product Documentation ER/Studio Business Architect Installation Guide 2nd Edition Version 1.6.1 Published August 2010 2010 Embarcadero Technologies, Inc. Embarcadero, the Embarcadero Technologies logos,

More information

Agilent MassHunter Easy Access Software

Agilent MassHunter Easy Access Software Agilent MassHunter Easy Access Software Quick Start Guide Where to Find More Information 3 Installation 4 Sharing Configurations with Multiple Instruments 10 Installed Files 11 Using Easy Access Software

More information

Cliquid Software for Routine Analysis

Cliquid Software for Routine Analysis Cliquid 3.0.1 Software for Routine Analysis Release Notes Table of Contents Known Issues... 2 LC Systems...2 Initial Log On... 2 Continuous Usage Limitation... 2 Acquisition Fails due to Locked Windows

More information

bbc Adobe Central Output Server Getting Started for Microsoft Windows Version 5.7

bbc Adobe Central Output Server Getting Started for Microsoft Windows Version 5.7 bbc Adobe Central Output Server Version 5.7 Getting Started for Microsoft Windows Getting Started for Microsoft Windows Edition 4.0, March 2009 2009 Adobe Systems Incorporated All rights reserved. As of

More information

License Server

License Server License Server 11.13.0.2 Installation Guide Isograph Copyright 2015 Isograph Limited. All rights reserved. This document and the associated software contains proprietary information which is protected

More information

ChromQuest 5.0 Quick Reference Guide

ChromQuest 5.0 Quick Reference Guide ChromQuest 5.0 Quick Reference Guide This guide contains an overview of the ChromQuest chromatography data system, with topics organized by workflow. For more information, refer to the ChromQuest User

More information

SOFTWARE UPDATE NOTIFICATION MASTERSIZER 2000 SOFTWARE v5.61: PSS

SOFTWARE UPDATE NOTIFICATION MASTERSIZER 2000 SOFTWARE v5.61: PSS SOFTWARE UPDATE NOTIFICATION MASTERSIZER 2000 SOFTWARE v5.61: PSS0002-34 Introduction This document details the release of software PSS0002-34: version 5.61 of the software for the Mastersizer 2000 laser

More information

Citrix SCOM Management Pack for XenServer

Citrix SCOM Management Pack for XenServer Citrix SCOM Management Pack for XenServer May 21, 2017 Citrix SCOM Management Pack 2.25 for XenServer Citrix SCOM Management Pack 2.24 for XenServer Citrix SCOM Management Pack 2.23 for XenServer Citrix

More information

Fiery Command WorkStation 5.8 with Fiery Extended Applications 4.4

Fiery Command WorkStation 5.8 with Fiery Extended Applications 4.4 Fiery Command WorkStation 5.8 with Fiery Extended Applications 4.4 Fiery Extended Applications (FEA) v4.4 contains Fiery software for performing tasks using a Fiery Server. This document describes how

More information

REVISIONS * The manual number is given on the bottom left of the back cover.

REVISIONS * The manual number is given on the bottom left of the back cover. REVISIONS * The manual number is given on the bottom left of the back cover. Print Date * Manual Number Revision Apr., 2002 SH (NA) -080270-A First edition Jun., 2002 SH (NA) -080270-B Correction Section

More information

Osprey PCI Series Multimedia Capture Driver Release Notes Driver Version

Osprey PCI Series Multimedia Capture Driver Release Notes Driver Version June 2015 Osprey PCI Series Multimedia Capture Driver Release Notes Driver Version 5.0.1.93 VIDEO CAPTURE CARD Osprey 100, 210, 230, 440, & 530 This driver has been tested on the following operating systems:

More information

AnySite. Australia. Install Guide. AnySite v9.1 Install Guide

AnySite. Australia. Install Guide. AnySite v9.1 Install Guide AnySite Australia Install Guide AnySite v9.1 Install Guide TABLE OF CONTENTS Chapter 1: Introduction... 4 Product Inventory Chapter 2: Before You Begin... 6 Permissions System Requirements Default Install

More information

Release Notes for Avaya Proactive Contact 5.0 Supervisor. Release Notes for Avaya Proactive Contact 5.0 Supervisor

Release Notes for Avaya Proactive Contact 5.0 Supervisor. Release Notes for Avaya Proactive Contact 5.0 Supervisor Release Notes for Avaya Proactive Contact 5.0 Supervisor Copyright@2011 Notice: While reasonable efforts were made to ensure that the information in this document was complete and accurate at the time

More information

KYOCERA Net Admin Installation Guide

KYOCERA Net Admin Installation Guide KYOCERA Net Admin Guide Legal Notes Unauthorized reproduction of all or part of this guide is prohibited. The information in this guide is subject to change without notice. We cannot be held liable for

More information

An Apple Subsidiary. This software addresses an issue where the OpenSSL library used by FileMaker Server 13.0v1 was vulnerable to the Heartbleed bug.

An Apple Subsidiary. This software addresses an issue where the OpenSSL library used by FileMaker Server 13.0v1 was vulnerable to the Heartbleed bug. An Apple Subsidiary FileMaker Server 13.0v2 Installer and Updater Release Notes Software Description The installer software installs FileMaker Server 13.0v2 for Windows and OS X. The updater software updates

More information

Workstation Configuration

Workstation Configuration Workstation Configuration Version 8.2 May 2, 2014 For the most recent version of this document, visit our documentation website. Table of Contents 1 Workstation configuration 3 1.1 Custom level Internet

More information

Remote Watchman Device Server User Manual

Remote Watchman Device Server User Manual Remote Watchman Device Server User Manual Software Version 4.0 Manual P/N 071-0043 Rev 8 September 29, 2016 Copyright 2000-2016 Sensorsoft Corporation, All rights reserved. Sensorsoft, SCOM and Remote

More information

ver Sound Editor for MAGICSTOMP ver is for units using version 2.10 firmware or greater.

ver Sound Editor for MAGICSTOMP ver is for units using version 2.10 firmware or greater. ver. 2.10 Sound Editor for ver. 2.10 is for units using version 2.10 firmware or greater. SPECIAL NOTICES The software and this are exclusive copyrights of Yamaha Corporation. Copying of the software or

More information

ViewNow X Server Setup Guide

ViewNow X Server Setup Guide ViewNow X Server 9.6.4 Setup Guide Micro Focus The Lawn 22-30 Old Bath Road Newbury, Berkshire RG14 1QN UK http://www.microfocus.com Copyright Micro Focus 1984-2014. All rights reserved. MICRO FOCUS, the

More information

Setting up FT3000 and version on Windows 7. This is a step by step guide for setting up the FT3000 on a new Windows 7 or 8 PC

Setting up FT3000 and version on Windows 7. This is a step by step guide for setting up the FT3000 on a new Windows 7 or 8 PC Setting up FT3000 and version 3.0.1 on Windows 7 This is a step by step guide for setting up the FT3000 on a new Windows 7 or 8 PC PROBLEM: Computer being updated from XP to WIN 7 or 8. New PC does not

More information