n o r d i c B r a i n E x Tutorial DSC Module

Size: px
Start display at page:

Download "n o r d i c B r a i n E x Tutorial DSC Module"

Transcription

1 m a k i n g f u n c t i o n a l M R I e a s y n o r d i c B r a i n E x Tutorial DSC Module Please note that this tutorial is for the latest released nordicbrainex. If you are using an older version please upgrade. NordicNeuroLab AS Møllendalsveien 1 N-5009 Bergen, Norway Phone: ProductSolutions@NordicNeuroLab.com

2 Contents 1 Loading of perfusion data Coregistration Interacting with the results Bolus arrival time General use Merge CBV for export to neuronavigation Saving DSC data Parametric maps Calculation of parametric maps Leakage correction Perfusion settings DSC settings Motion correction Leakage correction Vessel removal Normalization Auto-detect noise threshold Output maps References Date modified: Page 2 of 15

3 1 Loading of perfusion data To start the perfusion analysis, on structural and one DSC series have to be chosen. If the DSC type is unknown, right-click to set the type (Figure 1). It is possible to select several structural series, but only one DSC dataset for each analysis. To open DSC settings, right-click and select Edit DSC Settings. For more details about the DSC settings, see section 5 Perfusion settings on page 13. Hit Next to start analysis. Figure 1: To start perfusion analysis, select one structural and one DSC series. To set the type of DSC, right-click and select the correct type from the list. To open DSC settings, select Edit DSC Settings. Hit Next to start analysis. Date modified: Page 3 of 15

4 Figure 2: If you select series from different studies (time points), you will get a warning. Select yes, if you do want to load series from different studies. If you select series from different studies (time points) when loading data, you will get a warning (Figure 2). This is to assure that any loading of series from different studies is only done if intended. 2 Coregistration An important step in the analysis is coregistration of the DSC series to the structural series. This will ensure that perfusion maps used as overlays are displayed in the correct position. Please see Tutorial Visualization and Interaction for details about the coregistration. 3 Interacting with the results Figure 3: Perfusion analysis is done and the normalized BV map is added as an overlay. VOIs are added in the tumor tissue the contralateral normal appearing tissue, and the mean normalized CBV value is measured. The intensity curve and histogram of the VOI is displayed. Date modified: Page 4 of 15

5 When the perfusion analysis is completed, the perfusion maps are generated and shown as thumbnail in the left side of the panel. Add any map as overlay or underlay to investigate the output (Figure 3). 3.1 Bolus arrival time The bolus arrival time is calculated automatically by the program, but can be edited by sliding the red line along the time curve. Click to use the new time position (Figure 4). Figure 4: The bolus arrival time is automatically calculated, but can be changed by sliding the orange line along the time curve. Use recalculate to update with the new position. 3.2 General use Display perfusion output maps as overlays or underlays: - Right-click on the thumbnail Using the Volume-of-Interest (VOI) tool to: - Measure statistical values on the parametric maps (Figure 5). - Display histogram and intensity curves (Figure 5). VOI statistics, intensity curves and histograms are displayed by selecting the icons next to the VOI name in the VOI list (Figure 5). Change CBV overlay threshold (Figure 6) to remove low CBV values from the visualized overlay. Date modified: Page 5 of 15

6 Figure 5: Two VOIs are added, one in tumor tissue and one in normal appearing tissue. The statistical values are measured in the two regions, and the intensity curves and histograms for the two VOIs are displayed. Date modified: Page 6 of 15

7 3.3 Merge CBV for export to neuronavigation Similar to for BOLD, DTI and DCE results, it is possible to save the CBV result in a format that can be used in neuronavigation systems using Merge. Settings for merge export can be defined when choosing Settings and then Export results as... in the upper left corner of the Visualization interface. The BOLD activation maps can then be exported in a format compatible with viewing on certain neuronavigation systems. Color export This will export the BOLD activation maps as separate DICOM image series that can be viewed together with the structural volume on various neuronavigation workstations. White pixels on greyscale This will export the BOLD activation maps as DICOM image series where the activated regions are shown as white pixels in the structural dataset. Figure 6: CBV threshold can be changed in the DSC tab. Here, one can also use VOI filtering and merge to neuronavigation. In order to do so, adapt your workflow to the following steps: 1. Do perfusion analysis and add the CBV map as overlay. 2. Change the threshold of the overlay by sliding on the CBV threshold slider (Figure 6). 3. Locate the area(s) of interest and add VOIs Date modified: Page 7 of 15

8 4. Select VOI filtering and use include/exclude on VOI(s) to locate the region relevant for merge (it is possible to skip this step) 5. Verify the output in real time in the MPR. 6. Select Merge to create a new series. 7. Stop VOI filtering to continue. While in VOI filtering mode, some of the application s functionality is blocked. To access all functionality, VOI filtering mode must be stopped (hit Stop, see). Figure 7: Merge CBV for neuronavigation: Step 1: add the CBV map as overlay. Step 2: Threshold the CBV. Step 3: Locate the area of interest and create a VOI. Step 4: Select VOI filtering to include/exclude regions. Step 5: Verify the output in real in the MPR. Step 6: Select Merge to create new series. Note that multiple VOIs can be included or excluded to include/exclude multiple regions from the merged series. 3.4 Saving DSC data There are several different ways of saving the DSC data. Save the derived maps to the database: - Right-click on the thumbnail and select Save - Right-click on the thumbnail and select Save and send to send it to a remote entity (e.g. PACS) Create a new DICOM series with perfusion maps as overlay (see Figure 8): - Add a perfusion map (e.g. blood volume) as an overlay - Right-click in MPR and select Create snapshot, Slice selection or Slice all. - This will create a new secondary capture slice package that can be saved to the database and sent to PACS. If choosing Merge in the Interaction panel, the selected CBV overlaid on the structural datasets can be merged into a new DICOM series. By right clicking on any of the three planes of the current volume in the MPR, choosing Create snapshot will open the Slice editor window of the current slice. Create snapshot of MPR will open the Slice editor window of the three current planes of the MPR, as well as the visualized activations in the 3D viewer. Date modified: Page 8 of 15

9 Copy will copy the current slice to the clipboard, so it can be pasted into other programs (like word etc.). See Figure 8. In the Slice editor window (See Tutorial Visualization and Interaction ), accessed by right clicking on any of the three planes in the MPR and choosing Slice selection (and clicking on Slice) or Slice all, slices can be: Saved to the database Saved and sent to a remote entity (for example PACS) Added to report (Select View slices by right-clicking on the new thumbnail after saving to database) Saved as AVI-file (Select View slices by right-clicking on the new thumbnail after saving to database) Send, accessed by right clicking on series thumbnail in the Data panel will send the series to a remote entity (for example PACS). Read more about saving sessions in Saving data to database and PACS in Tutorial Handling Image data. Add to report (bottom left of any plane in the MPR) can be chosen to add the current slice to Report (see Report in Tutorial Visualization and Interaction ). Saving the session can be done by choosing File -> Save Session. This will save the entire session (all loaded and acquired datasets) to the database. The session name that you entered will appear in the session table on the Select patient data window. Read more about saving sessions in Saving, loading and sending a Session in Tutorial Handling Image data. Date modified: Page 9 of 15

10 Figure 8: Slice selection from the axial MPR has been selecting, displaying the VOIs and the VOI statistics, as well as the histogram for the CBV and the intensity curves for the DSC dynamic series for the two VOIs. This frame can be send to PACS by selecting 'Save and send'. Date modified: Page 10 of 15

11 4 Parametric maps Figure 9: The parametric maps are displayed under derived data on the left side of the visualization screen. Right-click to interact (view as underlay / overlay, save to database, view slices etc.). 4.1 Calculation of parametric maps The resulting parametric maps appear as icons on the left side of the screen under Derived Data (Figure 9). The parametric maps that are obtained are calculated using standard formulas and assumptions. The first step is to obtain the contrast agent concentration as a function of time. It is obtained using the assumption that the * change in relaxation rate is proportional to the concentration of contrast agent: R = r CA with the relation between MR signal and relaxation rates, SI0 ln * SI( t) R2 ( t) =, TE 2. Combining this Date modified: Page 11 of 15

12 the concentrations can be found. Then the following parametric maps are obtained: Normalized blood volume (ncbv) * The pixel values are initially set equal to rcbv = R ( t) dt. If leakage correction is turned on, the pixel values are then leakage corrected (see explanation of leakage correction below). Finally, the map is normalized. Relative blood volume (rcbv) Same as ncbv, but not normalized. Uncorrected ncbv This map displays the normalized blood volume without leakage correction. Uncorrected rcbv This map displays the relative blood volume without leakage correction. Leakage map This map displays the degree and type of leakage. Pixels close to zero (black) represent low degree of leakage. Positive (red) or negative (blue) pixel values represent T1-dominant or T2-dominant leakage, respectively. Mean transit time (rmtt) 2 The relative MTT values are calculated from the following relation: Normalized Blood Flow (ncbf) rmtt t R = R * 2 * 2 dt dt The blood flow values are calculated based on this relation, followed by normalization. rcbf = rcbv rmtt Relative Blood Flow (rcbf) Same as ncbf, only without normalization. Time to peak (rttp) Relative time of maximum concentration expressed in seconds. Date modified: Page 12 of 15

13 Vessel Map Displays the vessels found during the vessel removal Leakage correction The method used in nbx for leakage correction is described in the article by Bjørnerud et al (2011). In brief, the method uses the tail of the residual function to estimate the leakage. The residual function is defined through the equation C(t) = AIF(t) R(t) where C is the measured concentration, AIF is the artierial input function and R is the residual function. The residual function describes the fraction of contrast agent that is left in a voxel at the time t after a pulse of contrast agent enters the voxel. The concentration is thereby equal to the convolution between the input concentration, the arterial input function, and the residual function. Basically, if the residual function does not approach 0 after a long time, there is leakage. The leakage may be a T1 type or a T2 type, giving a residual function ending above or below 0. Either way, the leakage correction is done by finding the mean value of the residual after a preset time T C and multiplying it with the time that has passed after T C. The value found is subtracted from the initial estimate of the CBV. In the case leakage correction is done, the flow is recalculated as the peak of the residual function, TTP is set equal to the time of the residual peak and MTT is recalculated as the ration between blood volume and flow. 5 Perfusion settings Figure 10: In perfusion settings, the user can decide which preprocessing and analysis settings to use, and which output maps to display. Date modified: Page 13 of 15

14 Perfusion settings (Figure 10) can be opened from the loading frame by right-clicking on the DSC series, or from the visualization interface by right-clicking on the DSC source data thumbnail. In the perfusion settings, the user can decide which preprocessing and analysis settings to use, and which output maps to display. 5.1 DSC settings The settings for DSC are global settings, meaning that changes done at one time will be saved and applied to all future analysis until the settings again are changed Motion correction When checked this turns on motion correction of the DSC image series, meaning that all the image volumes are coregistered to the first image volume. These parameters are saved internally so that the user can display them at a later stage by right-clicking on the icon of the perfusion series and choose Motion Correction Results. The coregistration is performed using an iterative method based on a measure of similarity. Step sizes are calculated from the gradients of the similarity measure (the mutual information metric), and once the step size of the rotation and translation parameters is small enough, the iteration process is halted, and the estimated rotation and translation parameters are applied to the source volume using a cubic interpolation approach. Due to the limited geometrical resolution of such image series only in-plane rotations and translations for axial slices are done Leakage correction In tissue where contrast agent is leaking out of the vessels into the extra- or intracellular space the blood volume will not be correctly estimated. The remedy for this phenomenon is what is termed as leakage correction. If not corrected for, this leakage can lead to either under- or overestimation of the normalized blood volume. Leakage correction will correct the output maps mean transit time (MTT), time to peak (TTP), blood volume (CBV) and blood flow (CBF). Additional explanations can be found in Section 4 Parametric maps on page Vessel removal A patented method for vessel removal has been implemented in the software that identifies pixels that most likely are affected by the contrast agent in a vessel. These pixels will not be included in the formation of the output maps. The Vessel Map contains the pixels excluded Normalization Normalization is done by defining the normal tissue in the volume, and dividing all pixels on the value of the normal tissue. The normal tissue in a normalized CBV map should thus have an average value close to one. The normal tissue is found by segmenting the brain using a clustering algorithm. First, pixels with values lower than the noise level threshold are excluded. The remaining pixels are segmented based on their baseline signal intensity, and the pixels with the highest values are assumed to be lesion and are excluded. To exclude blood vessels from healthy brain tissue, a second segmentation is performed based on the already calculated BV values, where the highest values are excluded as blood. The output maps blood volume (CBV) and blood flow (CBF) are normalized. When normalization is done, the average CBV value of the normal tissue is equal to 1, while tumor tissue typically has a higher CBV value. Date modified: Page 14 of 15

15 If normalization is used, CBV and CBF output maps are indicated with a n prefix for normalized. All maps that have not been normalized are indicated with a r prefix for relative Auto-detect noise threshold If checked, an automatick noise threshold is performed, and if unchecked, the user can set the noise threhold manually using a slider (Figure 11). The noise threshold affects the performance of the normalization and leakage correction, so a warning will be given when manual noise threshold is selected. If brain tissue is removed by the nosie threshold, the normalization and leakage correction might not work optimally (i.e. noise threshold should not be used for segmenting the data). Figure 11: Noise-threshold can either be detected automatically, or the level can be changed using a slider. 5.2 Output maps In the Output maps tab (Figure 10), the user can select which output maps to display. Details on how these parametric maps are calculated are given in section 4 Parametric maps. 6 References Bjornerud A, Sorensen AG, Mouridsen K and Emblem KE (2011) T 1- and T 2* -dominant extravasation correction in DSC-MRI: Part I Theoretical considerations and implications for assessment of tumor hemodynamic properties. Journal of Cerebral Blood Flow & Metabolism 10: Date modified: Page 15 of 15

Tutorial BOLD Module

Tutorial BOLD Module m a k i n g f u n c t i o n a l M R I e a s y n o r d i c B r a i n E x Tutorial BOLD Module Please note that this tutorial is for the latest released nordicbrainex. If you are using an older version please

More information

n o r d i c B r a i n E x Tutorial DTI Module

n o r d i c B r a i n E x Tutorial DTI Module m a k i n g f u n c t i o n a l M R I e a s y n o r d i c B r a i n E x Tutorial DTI Module Please note that this tutorial is for the latest released nordicbrainex. If you are using an older version please

More information

Tutorial Visualization and Interaction

Tutorial Visualization and Interaction m a k i n g f u n c t i o n a l M R I e a s y n o r d i c B r a i n E x Tutorial Visualization and Interaction Please note that this tutorial is for the latest released nordicbrainex. If you are using

More information

Tutorial Handling image data

Tutorial Handling image data m a k i n g f u n c t i o n a l M R I e a s y n o r d i c B r a i n E x Tutorial Handling image data Please note that this tutorial is for the latest released nordicbrainex. If you are using an older version

More information

XI Conference "Medical Informatics & Technologies" VALIDITY OF MRI BRAIN PERFUSION IMAGING METHOD

XI Conference Medical Informatics & Technologies VALIDITY OF MRI BRAIN PERFUSION IMAGING METHOD XI Conference "Medical Informatics & Technologies" - 2006 medical imaging, MRI, brain perfusion Bartosz KARCZEWSKI 1, Jacek RUMIŃSKI 1 VALIDITY OF MRI BRAIN PERFUSION IMAGING METHOD Brain perfusion imaging

More information

Software Release Notes for nordicbrainex v1.1.3

Software Release Notes for nordicbrainex v1.1.3 Software Release Notes for nordicbrainex v1.1.3 Revision 1 Date: February 15 th 2013 Approved by: Sigvald Høyheim NOTE: The most current documentation for released products is available on http://www.nordicneurolab.com.

More information

Software Release Notes for nordicbrainex v2.3.2

Software Release Notes for nordicbrainex v2.3.2 Software Release Notes for nordicbrainex v2.3.2 Revision 1 Date: April 20 th, 2017 Approved by: VP Development NOTE: The most current documentation for released products is available on http://www.nordicneurolab.com.

More information

The Advanced Neuro MR Processing & Visualization Solution

The Advanced Neuro MR Processing & Visualization Solution The Advanced Neuro MR Processing & Visualization Solution Optimized Workflow Automated Processing Provides Answers Faster and More Efficiently The DynaSuite Neuro Workstation is a high performance advanced

More information

better images mean better results

better images mean better results better images mean better results A better way for YOU and YOUR patient brought to you by Advanced Neuro analysis with access to studies wherever you need it Advanced Neuro from Invivo Advancements in

More information

CTA HEAD Perfusion AqONE without and with IV Contrast

CTA HEAD Perfusion AqONE without and with IV Contrast CTA HEAD Perfusion AqONE without and with IV Contrast Patient Position Adult Contrast Adult Injection Rate Supine IOML perpendicular to table top. IV: 100 ml with helical head CTA 50 ml without helical

More information

syngo.mr Neuro 3D: Your All-In-One Post Processing, Visualization and Reporting Engine for BOLD Functional and Diffusion Tensor MR Imaging Datasets

syngo.mr Neuro 3D: Your All-In-One Post Processing, Visualization and Reporting Engine for BOLD Functional and Diffusion Tensor MR Imaging Datasets syngo.mr Neuro 3D: Your All-In-One Post Processing, Visualization and Reporting Engine for BOLD Functional and Diffusion Tensor MR Imaging Datasets Julien Gervais; Lisa Chuah Siemens Healthcare, Magnetic

More information

PMOD Features dedicated to Oncology Research

PMOD Features dedicated to Oncology Research While brain research using dynamic data has always been a main target of PMOD s developments, many scientists working with static oncology data have also found ways to leverage PMOD s unique functionality.

More information

syngo.pet&ct Oncology CT Tutorial

syngo.pet&ct Oncology CT Tutorial www.siemens.com/healthcare syngo.pet&ct Oncology CT Tutorial Storyboard VA30 Answers for life. Table of contents 1 What's New in the Latest Upgrade of syngo.pet&ct Oncology? 5 2 Introduction 7 2.1 Prerequisites

More information

EPILOG PREOP. Viewer Interpretation Guideline

EPILOG PREOP. Viewer Interpretation Guideline EPILOG PREOP Viewer Interpretation Guideline Starting the viewer Launch the viewer for a patient by clicking this icon. A new browser tab or window will open. Data loading To ensure smooth switching between

More information

SIVIC GUI Overview. SIVIC GUI Layout Overview

SIVIC GUI Overview. SIVIC GUI Layout Overview SIVIC GUI Overview SIVIC GUI Layout Overview At the top of the SIVIC GUI is a row of buttons called the Toolbar. It is a quick interface for loading datasets, controlling how the mouse manipulates the

More information

Advanced MRI Techniques (and Applications)

Advanced MRI Techniques (and Applications) Advanced MRI Techniques (and Applications) Jeffry R. Alger, PhD Department of Neurology Ahmanson-Lovelace Brain Mapping Center Brain Research Institute Jonsson Comprehensive Cancer Center University of

More information

fmri/dti analysis using Dynasuite

fmri/dti analysis using Dynasuite fmri/dti analysis using Dynasuite Contents 1 Logging in 2 Finding patient session 3 Viewing and adjusting images 4 Checking brain segmentation 5 Checking image registration 6 Seeing fmri results 7 Saving

More information

CP Generalize Concepts in Abstract Multi-dimensional Image Model Component Semantics. David Clunie.

CP Generalize Concepts in Abstract Multi-dimensional Image Model Component Semantics. David Clunie. CP-1390 - Generalize Concepts in Abstract Multi-dimensional Image Model Semantics Page 1 STATUS Date of Last Update Person Assigned Submitter Name Submission Date Assigned 2014/06/09 David Clunie mailto:dclunie@dclunie.com

More information

Automatic Determination of Arterial Input Function for Dynamic Contrast Enhanced MRI in Tumor Assessment

Automatic Determination of Arterial Input Function for Dynamic Contrast Enhanced MRI in Tumor Assessment Automatic Determination of Arterial Input Function for Dynamic Contrast Enhanced MRI in Tumor Assessment Jeremy Chen, Jianhua Yao, and David Thomasson Diagnostic Radiology Department, Clinical Center,

More information

Patient-Motion Analysis in Perfusion Weighted MRI

Patient-Motion Analysis in Perfusion Weighted MRI Patient-Motion Analysis in Perfusion Weighted MRI Thesis submitted in partial fulfillment of the requirements for the degree of MS by Research in Computer Science and Engineering by ROHIT GAUTAM 200702035

More information

Correction of Partial Volume Effects in Arterial Spin Labeling MRI

Correction of Partial Volume Effects in Arterial Spin Labeling MRI Correction of Partial Volume Effects in Arterial Spin Labeling MRI By: Tracy Ssali Supervisors: Dr. Keith St. Lawrence and Udunna Anazodo Medical Biophysics 3970Z Six Week Project April 13 th 2012 Introduction

More information

Manual image registration in BrainVoyager QX Table of Contents

Manual image registration in BrainVoyager QX Table of Contents Manual image registration in BrainVoyager QX Table of Contents Manual image registration in BrainVoyager QX......1 Performing manual alignment for functional to anatomical images......2 Step 1: preparation......2

More information

4DM Packages. 4DM Packages & License Types. Information to help you order the appropriate licenses for your site.

4DM Packages. 4DM Packages & License Types. Information to help you order the appropriate licenses for your site. 4DM Packages 4DM Packages & License Types. Information to help you order the appropriate licenses for your site. Nuclear Cardiac Quantification, Review, and Reporting Select Your 4DM Package and corresponding

More information

AORTA CTA VPMC-12419

AORTA CTA VPMC-12419 AORTA CTA VPMC-12419 Workflow Overview: The Aorta can be post-processed in various ways. Auto Bone Removal and Vessel Pick provide a quick overview of the entire Aorta. Vessel Probe creates a centerline

More information

ASAP_2.0 (Automatic Software for ASL Processing) USER S MANUAL

ASAP_2.0 (Automatic Software for ASL Processing) USER S MANUAL ASAP_2.0 (Automatic Software for ASL Processing) USER S MANUAL ASAP was developed as part of the COST Action "Arterial Spin Labelling Initiative in Dementia (AID)" by: Department of Neuroimaging, Institute

More information

SISCOM (Subtraction Ictal SPECT CO-registered to MRI)

SISCOM (Subtraction Ictal SPECT CO-registered to MRI) SISCOM (Subtraction Ictal SPECT CO-registered to MRI) Introduction A method for advanced imaging of epilepsy patients has been developed with Analyze at the Mayo Foundation which uses a combination of

More information

Learn Image Segmentation Basics with Hands-on Introduction to ITK-SNAP. RSNA 2016 Courses RCB22 and RCB54

Learn Image Segmentation Basics with Hands-on Introduction to ITK-SNAP. RSNA 2016 Courses RCB22 and RCB54 Learn Image Segmentation Basics with Hands-on Introduction to ITK-SNAP RSNA 2016 Courses RCB22 and RCB54 RCB22 Mon, Nov 28 10:30-12:00 PM, Room S401CD RCB54 Thu, Dec 1 2:30-4:30 PM, Room S401CD Presenters:

More information

Supplementary Figure 1

Supplementary Figure 1 Supplementary Figure 1 BOLD and CBV functional maps showing EPI versus line-scanning FLASH fmri. A. Colored BOLD and CBV functional maps are shown in the highlighted window (green frame) of the raw EPI

More information

IMPAX Volume Viewing Configuration

IMPAX Volume Viewing Configuration Getting started guide IMPAX Volume Viewing Configuration This guide describes the various configuration settings for IMPAX Volume Viewing. Some of these settings are only relevant if a license for one

More information

RT_Image v0.2β User s Guide

RT_Image v0.2β User s Guide RT_Image v0.2β User s Guide RT_Image is a three-dimensional image display and analysis suite developed in IDL (ITT, Boulder, CO). It offers a range of flexible tools for the visualization and quantitation

More information

MB-EPI PCASL. Release Notes for Version February 2015

MB-EPI PCASL. Release Notes for Version February 2015 MB-EPI PCASL Release Notes for Version 1.0 20 February 2015 1 Background High-resolution arterial spin labeling (ASL) imaging is highly desirable in both neuroscience research and clinical applications

More information

Diffusion Mapping with FireVoxel Quick Start Guide

Diffusion Mapping with FireVoxel Quick Start Guide Diffusion Mapping with FireVoxel Quick Start Guide Medical image analysis tool developed by Artem Mikheev and Henry Rusinek Radiology Department, NYU School of Medicine Original version prepared by Jinyu

More information

Brain Perfusion Imaging - Performance and Accuracy

Brain Perfusion Imaging - Performance and Accuracy Brain Perfusion Imaging - Performance and Accuracy Fan Zhu E H U N I V E R S I T Y T O H F R G E D I N B U Doctor of Philosophy Centre for Intelligent Systems and their Applications School of Informatics

More information

DIFFUSION TENSOR IMAGING ANALYSIS. Using Analyze

DIFFUSION TENSOR IMAGING ANALYSIS. Using Analyze DIFFUSION TENSOR IMAGING ANALYSIS Using Analyze 2 Table of Contents 1. Introduction page 3 2. Loading DTI Data page 4 3. Computing DTI Maps page 5 4. Defining ROIs for Fiber Tracking page 6 5. Visualizing

More information

Imaging Guide Pancreatic Imaging

Imaging Guide Pancreatic Imaging Imaging Guide Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Imaging System Ver 1.0 Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Imaging System Course Objectives: This guide

More information

Data Loading & 3D Visualization

Data Loading & 3D Visualization Neuroimage Analysis Center Data Loading & 3D Visualization Sonia Pujol, Ph.D. Surgical Planning Laboratory Harvard Medical School Leonardo da Vinci (1452-1519), Virgin and Child Alte Pinakothek, München

More information

Functional MRI. Jerry Allison, Ph. D. Medical College of Georgia

Functional MRI. Jerry Allison, Ph. D. Medical College of Georgia Functional MRI Jerry Allison, Ph. D. Medical College of Georgia BOLD Imaging Technique Blood Oxygen Level Dependent contrast can be used to map brain function Right Hand Motor Task Outline fmri BOLD Contrast

More information

NA-MIC National Alliance for Medical Image Computing fmri Data Analysis

NA-MIC National Alliance for Medical Image Computing   fmri Data Analysis NA-MIC fmri Data Analysis Sonia Pujol, Ph.D. Wendy Plesniak, Ph.D. Randy Gollub, M.D., Ph.D. Acknowledgments NIH U54EB005149 Neuroimage Analysis Center NIH P41RR013218 FIRST Biomedical Informatics Research

More information

VueBox. Quantification Toolbox. Instructions for use. Copyright 2015 Bracco Suisse SA. Bracco Suisse SA Software Applications VBO.8.

VueBox. Quantification Toolbox. Instructions for use. Copyright 2015 Bracco Suisse SA. Bracco Suisse SA Software Applications VBO.8. VueBox Quantification Toolbox Instructions for use Copyright 2015 Bracco Suisse SA Bracco Suisse SA Software Applications VBO.8.11 1 This publication may not be reproduced, stored in a retrieval system,

More information

This exercise uses one anatomical data set (ANAT1) and two functional data sets (FUNC1 and FUNC2).

This exercise uses one anatomical data set (ANAT1) and two functional data sets (FUNC1 and FUNC2). Exploring Brain Anatomy This week s exercises will let you explore the anatomical organization of the brain to learn some of its basic properties, as well as the location of different structures. The human

More information

Dosimetric Analysis Report

Dosimetric Analysis Report RT-safe 48, Artotinis str 116 33, Athens Greece +30 2107563691 info@rt-safe.com Dosimetric Analysis Report SAMPLE, for demonstration purposes only Date of report: ----------- Date of irradiation: -----------

More information

Pre-processing of ASL data T CT

Pre-processing of ASL data T CT Wed October 2, 2013 Image Processing Pre-processing: motion correction, denoising, outlier detection Alessandra Bertoldo Pre-processing of ASL data T CT C T C Single TI ASL T T T T C CCC average Pre-processing

More information

Visage 7 Clinical Training Basic Features

Visage 7 Clinical Training Basic Features Visage 7 Clinical Training Basic Features Contents Overview... 4 Usage... 4 Client Server Architecture... 5 Client Login... 6 Study Browser... 7 Query Section... 8 Study Labels... 10 Query Labeled Studies...

More information

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging 1 CS 9 Final Project Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging Feiyu Chen Department of Electrical Engineering ABSTRACT Subject motion is a significant

More information

Voxar 3D ColonMetrix. Reference Guide

Voxar 3D ColonMetrix. Reference Guide Voxar 3D ColonMetrix Reference Guide The software described in this document is furnished under a license, and may be used or copied only according to the terms of such license. Toshiba means, Toshiba

More information

multimodality image processing workstation Visualizing your SPECT-CT-PET-MRI images

multimodality image processing workstation Visualizing your SPECT-CT-PET-MRI images multimodality image processing workstation Visualizing your SPECT-CT-PET-MRI images FUSION FUSION is a new visualization and evaluation software developed by Mediso built on state of the art technology,

More information

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis Clinical Importance Rapid cardiovascular flow quantitation using sliceselective Fourier velocity encoding with spiral readouts Valve disease affects 10% of patients with heart disease in the U.S. Most

More information

IMAGE STUDIO LITE. Tutorial Guide Featuring Image Studio Analysis Software Version 3.1

IMAGE STUDIO LITE. Tutorial Guide Featuring Image Studio Analysis Software Version 3.1 IMAGE STUDIO LITE Tutorial Guide Featuring Image Studio Analysis Software Version 3.1 Notice The information contained in this document is subject to change without notice. LI-COR MAKES NO WARRANTY OF

More information

6 credits. BMSC-GA Practical Magnetic Resonance Imaging II

6 credits. BMSC-GA Practical Magnetic Resonance Imaging II BMSC-GA 4428 - Practical Magnetic Resonance Imaging II 6 credits Course director: Ricardo Otazo, PhD Course description: This course is a practical introduction to image reconstruction, image analysis

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

ThE ultimate, INTuITIVE Mr INTErFAcE

ThE ultimate, INTuITIVE Mr INTErFAcE ThE ultimate, INTuITIVE Mr INTErFAcE Empowering you to do more The revolutionary Toshiba M-power user interface takes Mr performance and flexibility to levels higher than ever before. M-power is able to

More information

RSNA, /rg

RSNA, /rg RSNA, 2015 10.1148/rg.2015140320 Appendix As noted in the main article, DICOM image files cannot be used directly for 3D printing; further steps are necessary to make them readable by 3D printers. The

More information

Stroke Quantification Tool (Sonia) Ver User Manual

Stroke Quantification Tool (Sonia) Ver User Manual Stroke Quantification Tool (Sonia) Ver. 1.0 User Manual English. 12/2016 Rev. 1.0 www.wakeup-stroke.eu 1 Table of Contents 1. Introduction...3 2. Installation...4 3. Data Import...5 4. Registration...7

More information

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Outline High Spectral and Spatial Resolution MR Imaging (HiSS) What it is How to do it Ways to use it HiSS for Radiation

More information

Chapter 11: Time Dependent Analysis(TDA) for Flow Analysis

Chapter 11: Time Dependent Analysis(TDA) for Flow Analysis Chapter 11: Time Dependent Analysis(TDA) for Flow Analysis The optional Flow workflow performs quantitative flow analysis on phase-contrast MR images. It provides semi-automatic contour detection for the

More information

CS/NEUR125 Brains, Minds, and Machines. Due: Wednesday, April 5

CS/NEUR125 Brains, Minds, and Machines. Due: Wednesday, April 5 CS/NEUR125 Brains, Minds, and Machines Lab 8: Using fmri to Discover Language Areas in the Brain Due: Wednesday, April 5 In this lab, you will analyze fmri data from an experiment that was designed to

More information

COBRE Scan Information

COBRE Scan Information COBRE Scan Information Below is more information on the directory structure for the COBRE imaging data. Also below are the imaging parameters for each series. Directory structure: var/www/html/dropbox/1139_anonymized/human:

More information

Detecting White Matter Lesions in Lupus

Detecting White Matter Lesions in Lupus Slicer3 Training Compendium Detecting White Matter Lesions in Lupus Version 2.1 6/25/2009 H. Jeremy Bockholt Mark Scully -1- Learning objective This tutorial demonstrates an automated, multi-level method

More information

Exploration and Study of MultiVolume Image Data using 3D Slicer

Exploration and Study of MultiVolume Image Data using 3D Slicer Exploration and Study of MultiVolume Image Data using 3D Slicer Meysam Torabi and Andriy Fedorov torabi@bwh.harvard.edu, fedorov@bwh.harvard.edu Surgical Navigation and Robotics Lab and Surgical Planning

More information

MR Advance Techniques. Vascular Imaging. Class III

MR Advance Techniques. Vascular Imaging. Class III MR Advance Techniques Vascular Imaging Class III 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology

More information

MSI 2D Viewer Software Guide

MSI 2D Viewer Software Guide MSI 2D Viewer Software Guide Page:1 DISCLAIMER We have used reasonable effort to include accurate and up-to-date information in this manual; it does not, however, make any warranties, conditions or representations

More information

TARDIS Working Practice Document No: 035

TARDIS Working Practice Document No: 035 TARDIS Working Practice Document No: 035 Uploading CT or MR Images All stroke patients should have a baseline scan taken before randomisation. Any other CT scans or MRIs, taken up until completion of the

More information

NIH Public Access Author Manuscript Proc IEEE Int Symp Biomed Imaging. Author manuscript; available in PMC 2014 November 15.

NIH Public Access Author Manuscript Proc IEEE Int Symp Biomed Imaging. Author manuscript; available in PMC 2014 November 15. NIH Public Access Author Manuscript Published in final edited form as: Proc IEEE Int Symp Biomed Imaging. 2013 April ; 2013: 748 751. doi:10.1109/isbi.2013.6556583. BRAIN TUMOR SEGMENTATION WITH SYMMETRIC

More information

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different

More information

Modern Medical Image Analysis 8DC00 Exam

Modern Medical Image Analysis 8DC00 Exam Parts of answers are inside square brackets [... ]. These parts are optional. Answers can be written in Dutch or in English, as you prefer. You can use drawings and diagrams to support your textual answers.

More information

Quantitative MRI of the Brain: Investigation of Cerebral Gray and White Matter Diseases

Quantitative MRI of the Brain: Investigation of Cerebral Gray and White Matter Diseases Quantities Measured by MR - Quantitative MRI of the Brain: Investigation of Cerebral Gray and White Matter Diseases Static parameters (influenced by molecular environment): T, T* (transverse relaxation)

More information

QMass 8.1. Quick Start Manual. June 13, v2.0

QMass 8.1. Quick Start Manual. June 13, v2.0 QMass 8.1 Quick Start Manual June 13, 2017 9.10.250.81.3 v2.0 Medis medical imaging systems bv Schuttersveld 9, 2316 XG Leiden, the Netherlands http://www.medis.nl Medis medical imaging systems bv Medis

More information

EMSegmenter Tutorial (Advanced Mode)

EMSegmenter Tutorial (Advanced Mode) EMSegmenter Tutorial (Advanced Mode) Dominique Belhachemi Section of Biomedical Image Analysis Department of Radiology University of Pennsylvania 1/65 Overview The goal of this tutorial is to apply the

More information

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial

More information

Copyright 2012 Pulse Systems, Inc. Page 1 of 22

Copyright 2012 Pulse Systems, Inc. Page 1 of 22 Use the Pediatric Growth Chart Control to monitor a patient's measurements over an extended time vs. the national percentiles in a graphical output. Click anywhere to continue Copyright 2012 Pulse Systems,

More information

Image Registration. Prof. Dr. Lucas Ferrari de Oliveira UFPR Informatics Department

Image Registration. Prof. Dr. Lucas Ferrari de Oliveira UFPR Informatics Department Image Registration Prof. Dr. Lucas Ferrari de Oliveira UFPR Informatics Department Introduction Visualize objects inside the human body Advances in CS methods to diagnosis, treatment planning and medical

More information

Automated Determination of Arterial Input Function for DCE-MRI of the Prostate

Automated Determination of Arterial Input Function for DCE-MRI of the Prostate Automated Determination of Arterial Input Function for DCE-MRI of the Prostate Yingxuan Zhu a, Ming-Ching Chang b, Sandeep N. Gupta b a Dept. of EECS, Syracuse University, Syracuse, NY 12304 USA b GE Global

More information

TRANSFERRING MRI DATA FROM THE WORKSTATION

TRANSFERRING MRI DATA FROM THE WORKSTATION 1 TRANSFERRING MRI DATA FROM THE WORKSTATION There are various ways transferring your data from the MR workstation onto your computer or home directory on the network. Which way you choose will depend

More information

Getting Started Guide

Getting Started Guide Getting Started Guide Version 2.5 for BVQX 1.9 Rainer Goebel, Henk Jansma and Jochen Seitz Copyright 2007 Brain Innovation B.V. 2 Contents Preface...4 The Objects Tutorial...5 Scanning session information...5

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

Supplementary methods

Supplementary methods Supplementary methods This section provides additional technical details on the sample, the applied imaging and analysis steps and methods. Structural imaging Trained radiographers placed all participants

More information

Basic fmri Design and Analysis. Preprocessing

Basic fmri Design and Analysis. Preprocessing Basic fmri Design and Analysis Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial filtering

More information

Automatic Cerebral Aneurysm Detection in Multimodal Angiographic Images

Automatic Cerebral Aneurysm Detection in Multimodal Angiographic Images Automatic Cerebral Aneurysm Detection in Multimodal Angiographic Images Clemens M. Hentschke, Oliver Beuing, Rosa Nickl and Klaus D. Tönnies Abstract We propose a system to automatically detect cerebral

More information

Politecnico di Torino. Porto Institutional Repository

Politecnico di Torino. Porto Institutional Repository Politecnico di Torino Porto Institutional Repository [Proceeding] Motion artifact correction in ASL images: automated procedure an improved Original Citation: Di Cataldo S., Ficarra E., Acquaviva A., Macii

More information

Brain Mask User s Guide

Brain Mask User s Guide Brain Mask User s Guide Segmentation of the brain from three-dimensional MR images is a crucial pre-processing step in morphological and volumetric brain studies. BrainMask software implements a fully

More information

A Study of Medical Image Analysis System

A Study of Medical Image Analysis System Indian Journal of Science and Technology, Vol 8(25), DOI: 10.17485/ijst/2015/v8i25/80492, October 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 A Study of Medical Image Analysis System Kim Tae-Eun

More information

Functional MRI in Clinical Research and Practice Preprocessing

Functional MRI in Clinical Research and Practice Preprocessing Functional MRI in Clinical Research and Practice Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization

More information

Normalization for clinical data

Normalization for clinical data Normalization for clinical data Christopher Rorden, Leonardo Bonilha, Julius Fridriksson, Benjamin Bender, Hans-Otto Karnath (2012) Agespecific CT and MRI templates for spatial normalization. NeuroImage

More information

QFlow 5.6. User Manual. February 18, v1c0

QFlow 5.6. User Manual. February 18, v1c0 QFlow 5.6 User Manual February 18, 2016 9.07.400.56.3 v1c0 Medis medical imaging systems bv Schuttersveld 9, 2316 XG Leiden, the Netherlands http://www.medis.nl Medis medical imaging systems bv Medis medical

More information

Measuring baseline whole-brain perfusion on GE 3.0T using arterial spin labeling (ASL) MRI

Measuring baseline whole-brain perfusion on GE 3.0T using arterial spin labeling (ASL) MRI Measuring baseline whole-brain perfusion on GE 3.0T using arterial spin labeling (ASL) MRI Revision date: 09/15/2008 Overview This document describes the procedure for measuring baseline whole-brain perfusion

More information

Analysis of Functional MRI Timeseries Data Using Signal Processing Techniques

Analysis of Functional MRI Timeseries Data Using Signal Processing Techniques Analysis of Functional MRI Timeseries Data Using Signal Processing Techniques Sea Chen Department of Biomedical Engineering Advisors: Dr. Charles A. Bouman and Dr. Mark J. Lowe S. Chen Final Exam October

More information

EPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing

EPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing Functional Connectivity Preprocessing Geometric distortion Head motion Geometric distortion Head motion EPI Data Are Acquired Serially EPI Data Are Acquired Serially descending 1 EPI Data Are Acquired

More information

Transform Introduction page 96 Spatial Transforms page 97

Transform Introduction page 96 Spatial Transforms page 97 Transform Introduction page 96 Spatial Transforms page 97 Pad page 97 Subregion page 101 Resize page 104 Shift page 109 1. Correcting Wraparound Using the Shift Tool page 109 Flip page 116 2. Flipping

More information

How to make color-coded 3D models for structure thickness in CTVox

How to make color-coded 3D models for structure thickness in CTVox How to make color-coded 3D models for structure thickness in CTVox Method note Page 1 of 10 2 Bruker microct method note: Color-coded structure thickness in CTVox Introduction When running a 3D analysis

More information

Beginner s Guide to Vevo 2100

Beginner s Guide to Vevo 2100 PN 12090 Beginner s Guide to Vevo 2100 December 2008 This Beginner s Guide to Vevo 2100 provides a quick overview of the Vevo 2100 imaging modes and provides key tips to allow users to start a study, acquire

More information

MR IMAGE SEGMENTATION

MR IMAGE SEGMENTATION MR IMAGE SEGMENTATION Prepared by : Monil Shah What is Segmentation? Partitioning a region or regions of interest in images such that each region corresponds to one or more anatomic structures Classification

More information

Using TeraRecon intuition Viewer with STAT Quick Reference Guide

Using TeraRecon intuition Viewer with STAT Quick Reference Guide Using TeraRecon intuition Viewer with STAT Quick Reference Guide 1. Launching intuition 2. 3D Assessment of Coronary Arteries for Stenosis 3. Time Volume Analysis (TVA) for Determination of Left Ventricular

More information

Image Segmentation and Registration

Image Segmentation and Registration Image Segmentation and Registration Dr. Christine Tanner (tanner@vision.ee.ethz.ch) Computer Vision Laboratory, ETH Zürich Dr. Verena Kaynig, Machine Learning Laboratory, ETH Zürich Outline Segmentation

More information

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,

More information

Introduction to Neuroimaging Janaina Mourao-Miranda

Introduction to Neuroimaging Janaina Mourao-Miranda Introduction to Neuroimaging Janaina Mourao-Miranda Neuroimaging techniques have changed the way neuroscientists address questions about functional anatomy, especially in relation to behavior and clinical

More information

Centre for Mathematical Sciences, Mathematics, LTH, December 2015

Centre for Mathematical Sciences, Mathematics, LTH, December 2015 Centre for Mathematical Sciences, Mathematics, LTH, December 2015 FMNA30 - Medical Image Analysis, Assignment 4 1 Introduction The purpose of this assignment is to give hands-on experience with handling

More information

SPM8 for Basic and Clinical Investigators. Preprocessing

SPM8 for Basic and Clinical Investigators. Preprocessing SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial

More information

A benchmarking tool to evaluate computer tomography perfusion infarct core predictions against a DWI standard

A benchmarking tool to evaluate computer tomography perfusion infarct core predictions against a DWI standard Original Article A benchmarking tool to evaluate computer tomography perfusion infarct core predictions against a DWI standard Journal of Cerebral Blood Flow & Metabolism 0(00) 1 10! Author(s) 2015 Reprints

More information

Image Processing Guideline for TMU 7T MRI

Image Processing Guideline for TMU 7T MRI Image Processing Guideline for TMU 7T MRI Chia Feng Lu Laboratory of NeuroImage Biomarker Analysis, Translational Imaging Research Center, TMU 08/18/2015, version 1.0 Section 1: Installation of ImageJ

More information

BrainVoyager TM. Getting Started Guide. Version 3.0. for BV 21

BrainVoyager TM. Getting Started Guide. Version 3.0. for BV 21 BrainVoyager TM Getting Started Guide Version 3.0 for BV 21 Rainer Goebel, Henk Jansma, Caroline Benjamins, Judith Eck, Hester Breman and Armin Heinecke Copyright 2018 Brain Innovation B.V. Contents About

More information