1 Radiotherapy Plan Competition 2016 TomoTherapy Planning System Dmytro Synchuk Ukrainian Center of TomoTherapy, Kirovograd, Ukraine
2 Beam Geometry 6MV fan beam 3 jaw options 1.0, 2.5 and 5 cm 64 leaves binary collimator Leaf measurement at isocenter 0.65cm
3 Delivery regimes HELICAL Continuous gantry rotation at a constant rate Continuous couch movement in superior direction DIRECT Discrete 2-12 static directions Continuous couch movement in superior direction for each direction separately
4 Objective function F Plan F Targets F OARs where F ROI voxels (D prescr D calc ) 2 importance dose penalties DVH penalty ROI volume The smaller volume we can control the better
5 Beam blocking None (Default) Critical Structure Directional Blocking Critical Structure Complete Blocking Critical Structure Allowed Beam Direction Prohibited Beam Direction Target Target Target Directional Blocking only those beam directions that first go through the target are allowed Complete Blocking all beam directions that go through the completely blocked structure are prohibited
6 Troubleshootings: Time cumbersome Works only for PTV_BREAST Steel needed additional structures How I did my first Plan
7 Additional Contours Dmin2 Dmin2 Dmin1 Dmin1 Dmin1 0,5 cm margin around the PTV to create steep dose gradient around the target Dmin2 2,5-3 cm margin around the PTV to avoid high doses in body
8 Additional Contours Breast Right Block LUNG RIGHT BLOCK HEART BLOCK LUNG LEFT BLOCK LUNG LEFT Plan Brest Right Block 3 mm outer ring completely blocked to achieve no more than 2Gy dose inside right breast HEART BLOCK part of the hard that is further than 3 cm from PTV (completely blocked) HEART Plan - part of the hard that is closer than 3 cm to PTV (here I optimize) LUNG LEFT BLOCK part of the left lung that is further than 3 cm from PTV (completely blocked) Dmin Lung Left part of the left lung that is closer than 1.5 cm to PTV LUNG LEFT Plan part of the left lung that is closer than 3 cm to PTV excluding Dmin Lung Left
9 Additional contours Avoidances LUNG RIGHT LUNG LEFT Top LUNG RIGHT LUNG RIGHT BLOCK LUNG LEFT Top Emergence of PTV SC leads to: LUNG LEFT Top part of the left lung that is closer than 3 cm to PTV on the slices where PTV SC is present LUNG RIGHT BLOCK part of the right lung that is 1.5 cm lover than PTV SC Avoidances some volumes where some dose problems are predicted or found
10 Additional contours PTV Min BLOCK Block structure in the inferior part of the body made as the prolongation of the blocked regions PTV Min small part of the PTV made to shrink the volume of minimum dose
11 General Parameters Calculation Grid Fine resolution equal to imported planning images (256x256) Field width 2.5 cm was set according to the PTV length Beamlet Mode is a general mode in clinical practice - optimizer uses pre-computed dose volumes and simply re-weightings them Max beamlet open time MF Average beamlet open time for used beamlets Pitch dcouch/rotation field width
12 General Parameters Calculation Grid Fine resolution equal to imported planning images (256x256) Field width 2.5 cm was set according to the PTV length Beamlet Mode is a general mode in clinical practice - optimizer uses pre-computed dose volumes and simply re-weightings them Max beamlet open time MF Average beamlet open time for used beamlets Pitch dcouch/rotation field width
13 Starting restrictions
14 First Results First I protect OARs
15 Further optimization Now I work with Targets Don t forget about OARs
16 Further optimization
17 Final restrictions
18 Final Histogram Full Dose Final Dose
19 Final Dose Distribution
20 Main difference from my first plan
21 Evaluation Old Results New Results
22 Thank You!
Hugues Mailleux Medical Physics Department Institut Paoli-Calmettes Marseille France. Sunday 17 July 2016
Hugues Mailleux Medical Physics Department Institut Paoli-Calmettes Marseille France Sunday 17 July 2016 AGENDA 1. Introduction 2. Material 3. Optimization process 4. Results 5. Comments 6. Conclusion
More informationTomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS
Tomotherapy Physics Machine Twinning and Quality Assurance Emilie Soisson, MS Tomotherapy at UW- Madison Treating for nearly 5 years Up to ~45 patients a day on 2 tomo units Units twinned to facilitate
More informationMonaco VMAT. The Next Generation in IMRT/VMAT Planning. Paulo Mathias Customer Support TPS Application
Monaco VMAT The Next Generation in IMRT/VMAT Planning Paulo Mathias Customer Support TPS Application 11.05.2011 Background What is Monaco? Advanced IMRT/VMAT treatment planning system from Elekta Software
More informationTHE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!
THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER! Confidence in complex treatments Modern radiation therapy uses complex plans with techniques such as IMRT, VMAT and Tomotherapy.
More informationROBUST OPTIMIZATION THE END OF PTV AND THE BEGINNING OF SMART DOSE CLOUD. Moe Siddiqui, April 08, 2017
ROBUST OPTIMIZATION THE END OF PTV AND THE BEGINNING OF SMART DOSE CLOUD Moe Siddiqui, April 08, 2017 Agenda Background IRCU 50 - Disclaimer - Uncertainties Robust optimization Use Cases Lung Robust 4D
More informationAnalysis of RapidArc optimization strategies using objective function values and dose-volume histograms
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 1, WINTER 2010 Analysis of RapidArc optimization strategies using objective function values and dose-volume histograms Mike Oliver, a Isabelle
More informationTransitioning from pencil beam to Monte Carlo for electron dose calculations
Transitioning from pencil beam to Monte Carlo for electron dose calculations Jessie Huang-Vredevoogd (jyhuang4@wisc.edu) University of Wisconsin NCC AAPM October 12, 2019 1 Topics to cover Background RayStation
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More informationAcknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF
Page 1 Quality Assurance of IMRT Delivery Systems - Siemens Lynn J. Verhey, Ph.D. Professor and Vice-Chair UCSF Dept. of Radiation Oncology AAPM 22 Annual Meeting, Montreal Acknowledgments Ping Xia, Ph.D.,
More informationIMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays
IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays Sotiri Stathakis Outline Why IMRT/VMAT QA AAPM TG218 UPDATE Tolerance Limits and Methodologies for IMRT Verification QA Common sources
More information7/29/2017. Making Better IMRT Plans Using a New Direct Aperture Optimization Approach. Aim of Radiotherapy Research. Aim of Radiotherapy Research
Making Better IMRT Plans Using a New Direct Aperture Optimization Approach Dan Nguyen, Ph.D. Division of Medical Physics and Engineering Department of Radiation Oncology UT Southwestern AAPM Annual Meeting
More informationDose Calculation and Verification for Tomotherapy
Dose Calculation and Verification for Tomotherapy John P. Gibbons, PhD Chief of Clinical Physics Mary Bird Perkins Cancer Center Baton Rouge, LA Associate Professor Department of Physics and Astronomy
More informationMonaco Concepts and IMRT / VMAT Planning LTAMON0003 / 3.0
and IMRT / VMAT Planning LTAMON0003 / 3.0 and Planning Objectives By the end of this presentation you can: Describe the cost functions in Monaco and recognize their application in building a successful
More informationReal-time monitoring of linac performance using RT plans and logfiles
Real-time monitoring of linac performance using RT plans and logfiles NCCAAPM Wisconsin Dells 7 April 2017 Mark Wiesmeyer, PhD, DABR Technical Product Manager Technical talk with insights and results based
More informationIMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE
QA SOFTWARE FAST, PRECISE Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, VMAT, SRS and brachytherapy plans no film, no phantoms, no linac
More informationTreatment Planning Optimization for VMAT, Tomotherapy and Cyberknife
Treatment Planning Optimization for VMAT, Tomotherapy and Cyberknife Kerem Akartunalı Department of Management Science Strathclyde Business School Joint work with: Vicky Mak-Hau and Thu Tran 14 July 2015
More informationRadiation therapy treatment plan optimization
H. Department of Industrial and Operations Engineering The University of Michigan, Ann Arbor, Michigan MOPTA Lehigh University August 18 20, 2010 Outline 1 Introduction Radiation therapy delivery 2 Treatment
More informationUsing a research real-time control interface to go beyond dynamic MLC tracking
in partnership with Using a research real-time control interface to go beyond dynamic MLC tracking Dr. Simeon Nill Joint Department of Physics at The Institute of Cancer Research and the Royal Marsden
More informationADVANCING CANCER TREATMENT
The RayPlan treatment planning system makes proven, innovative RayStation technology accessible to clinics that need a cost-effective and streamlined solution. Fast, efficient and straightforward to use,
More informationThe MSKCC Approach to IMRT. Outline
The MSKCC Approach to IMRT Spiridon V. Spirou, PhD Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York, NY Outline Optimization Field splitting Delivery Independent verification
More informationBrilliance CT Big Bore.
1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.
More informationMonte Carlo methods in proton beam radiation therapy. Harald Paganetti
Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization
More informationIMRT site-specific procedure: Prostate (CHHiP)
IMRT site-specific procedure: Prostate (CHHiP) Scope: To provide site specific instructions for the planning of CHHIP IMRT patients Responsibilities: Radiotherapy Physicists, HPC Registered Therapy Radiographers
More informationTG-148 overview. Introduction. System Overview. System Overview. QA for Helical Tomotherapy: Report of the AAPM Task Group 148. Conflict of Interest:
QA or Helical Tomotherapy: Report o the AAPM Tas Group 148 Members: Conlict o Interest: r. John Balog owns TomoTherapy stoc. Katja Langen (Co-chair) Nio Papaniolaou (Co-chair) Walter Grant Richard Crilly
More informationSignificance of time-dependent geometries for Monte Carlo simulations in radiation therapy. Harald Paganetti
Significance of time-dependent geometries for Monte Carlo simulations in radiation therapy Harald Paganetti Modeling time dependent geometrical setups Key to 4D Monte Carlo: Geometry changes during the
More informationGPU applications in Cancer Radiation Therapy at UCSD. Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC
GPU applications in Cancer Radiation Therapy at UCSD Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC Conventional Radiotherapy SIMULATION: Construciton, Dij Days PLANNING:
More informationCreating a Knowledge Based Model using RapidPlan TM : The Henry Ford Experience
DVH Estimates Creating a Knowledge Based Model using RapidPlan TM : The Henry Ford Experience Karen Chin Snyder, MS, DABR AAMD Region V Meeting October 4, 2014 Disclosures The Department of Radiation Oncology
More informationDosimetric impact of the 160 MLC on head and neck IMRT treatments
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 6, 2014 Dosimetric impact of the 160 MLC on head and neck IMRT treatments Prema Rassiah-Szegedi, a Martin Szegedi, Vikren Sarkar, Seth Streitmatter,
More informationDosimetric Analysis Report
RT-safe 48, Artotinis str 116 33, Athens Greece +30 2107563691 info@rt-safe.com Dosimetric Analysis Report SAMPLE, for demonstration purposes only Date of report: ----------- Date of irradiation: -----------
More informationTOMOHD. Product Specifications
TOMOHD Treatment System Product Specifications 1. Complete Treatment System Dimensions Height Width Length Weight 8 ft., 3.25 in. (252.1 cm) 9 ft., 2.50 in. (280.6 cm) 15 ft., 2.25 in. (463.0 cm) 10,000
More informationHigh Throughput Computing and Sampling Issues for Optimization in Radiotherapy
High Throughput Computing and Sampling Issues for Optimization in Radiotherapy Michael C. Ferris, University of Wisconsin Alexander Meeraus, GAMS Development Corporation Optimization Days, Montreal, May
More informationImage Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD
Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image
More informationRaising the Bar in IMRT QA
MapCHECK 2TM Raising the Bar in IMRT QA The leader in quick and precise measurement of modulated radiotherapy beams Benefits Proven solution for film-less rotational delivery and IMRT QA - More than 1500
More informationAn optimization framework for conformal radiation treatment planning
An optimization framework for conformal radiation treatment planning Jinho Lim Michael C. Ferris Stephen J. Wright David M. Shepard Matthew A. Earl December 2002 Abstract An optimization framework for
More information8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques
Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery Lei Ren, Ph.D. Duke University Medical Center 2016 AAPM 58 th annual meeting, Educational Course, Therapy Track Disclosure I have
More informationATC Conformance Statement Overview
Advanced-Technology QA Consortium DICOM Workshop: ATC Conformance Statement Overview Walter R. Bosch, D.Sc. ATC DICOM Conformance Statement Overview DICOM Part 10 File Set Reader Application (ITC) Defines
More informationDose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them
Dose Distributions George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Purpose To familiarize the resident with dose distributions and the factors that affect them
More informationAnthropomorphic Verification of the Eclipse AAA Algorithm for Spine SBRT Treatments Involving Titanium Implants
Anthropomorphic Verification of the Eclipse AAA Algorithm for Spine SBRT Treatments Involving Titanium Implants Mike Grams Clinical Medical Physicist Mayo Clinic, Rochester, MN NCCAAPM Meeting, April 17,
More informationDigital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities
sup176_pc_newrtradiations.doc: Sup 176: 2 nd Generation RT Additional RT Treatment Modalities Page 1 2 4 Digital Imaging and Communications in Medicine (DICOM) 6 Supplement 176: Second Generation Radiotherapy
More informationTechnical Publication. DICOM Conformance Statement. iplan RT Dose 4.1. Document Revision 1. December 4, Copyright BrainLAB AG
Technical Publication DICOM Conformance Statement iplan RT Dose 4.1 Document Revision 1 December 4, 2008 2008 Copyright BrainLAB AG IPlan RT Dose 4.1 1 Conformance Statement Overview This document details
More informationDigital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities
sup17_13_newrtradiations.doc: Sup 17: nd Generation RT Additional RT Treatment Modalities Page 1 Digital Imaging and Communications in Medicine (DICOM) Supplement 17: Second Generation Radiotherapy Additional
More informationA fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study
A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study Indrin J. Chetty, a) Mihaela Rosu, Neelam Tyagi, Lon H. Marsh, Daniel
More informationADVANCING CANCER TREATMENT
3 ADVANCING CANCER TREATMENT SUPPORTING CLINICS WORLDWIDE RaySearch is advancing cancer treatment through pioneering software. We believe software has un limited potential, and that it is now the driving
More informationDICOM Conformance Statement. Leksell GammaPlan & Leksell SurgiPlan 10.2
Name of document DICOM Conformance Statement Regarding Date Leksell GammaPlan & Leksell SurgiPlan 2012-10-15 1 (57) DICOM Conformance Statement Leksell GammaPlan & Leksell SurgiPlan 10.2 2 (57) Contents
More information7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT
CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance
More informationOp#miza#on of a Fast CT System for More Accurate Radiotherapy. Erica Cherry, Meng Wu, and Rebecca Fahrig
Op#miza#on of a Fast CT System for More Accurate Radiotherapy Erica Cherry, Meng Wu, and Rebecca Fahrig Problem: mo#on during treatment Two overlapping thorax views during normal breathing liver lung spine
More informationMachine and Physics Data Guide
WWW..COM Machine and Physics Data Guide STANDARD IMAGING, INC. 3120 Deming Way Middleton, WI 53562-1461 May / 2008 2008 Standard Imaging, Inc. TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 www.standardimaging.com
More informationA closer look at CT scanning
Vet Times The website for the veterinary profession https://www.vettimes.co.uk A closer look at CT scanning Author : Charissa Lee, Natalie Webster Categories : General, Vets Date : April 3, 2017 A basic
More informationConformal Radiation Therapy. with. Cobalt-60 Tomotherapy
Conformal Radiation Therapy with Cobalt-60 Tomotherapy by Sandeep Kaur Dhanesar A thesis submitted to the Department of Physics, Engineering Physics, and Astronomy in conformity with the requirements for
More informationInitial Clinical Experience with 3D Surface Image Guidance
Initial Clinical Experience with 3D Surface Image Guidance Amanda Havnen-Smith, Ph.D. Minneapolis Radiation Oncology Ridges Radiation Therapy Center Burnsville, MN April 20 th, 2012 Non-funded research
More informationHow would, or how does, the patient position (chin extended) affect your beam arrangement?
1 Megan Sullivan Clinical Practicum II Parotid Lab July 29, 2016 PLAN 1: IPSILATERAL WEDGE PAIR TECHNIQUE The ipsilateral wedge pair technique consisted of an anterior oblique field at 45 degrees and a
More informationDAILY LINAC QA BEAM QA
BEAM QA DAILY LINAC QA The QA BeamChecker Plus allows for fast, reliable, and uncomplicated daily QA of Varian, Elekta, Siemens, and Accuray Treatment Machines. The QA BeamChecker Plus is specifically
More informationPosition accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon
Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon WHITE PAPER Introduction An image guidance system based on Cone Beam CT (CBCT) is included in Leksell
More informationTEPZZ Z754_7A_T EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (51) Int Cl.: A61N 5/10 ( )
(19) TEPZZ Z74_7A_T (11) EP 3 07 417 A1 (12) EUROPEAN PATENT APPLICATION (43) Date of publication: 0..16 Bulletin 16/ (1) Int Cl.: A61N / (06.01) (21) Application number: 16163147.8 (22) Date of filing:
More informationBasic Radiation Oncology Physics
Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides
More informationVolumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT
Volumetric Modulated Arc Therapy - Clinical Implementation Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA Acknowledgement David M. Shepard, Ph.D. Muhammad K. N. Afghan, Ph.D. Fan Chen, Ph.D.
More informationQuality assurance of a helical tomotherapy machine
INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 49 (2004) 2933 2953 PHYSICS IN MEDICINE AND BIOLOGY PII: S0031-9155(04)71892-8 Quality assurance of a helical tomotherapy machine J D Fenwick 1,2,WATomé
More informationDigital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities
Modalities Page 1 2 4 Digital Imaging and Communications in Medicine (DICOM) 6 Supplement 176: Second Generation Radiotherapy Additional RT Treatment Modalities 8 10 12 14 16 DICOM Standards Committee,
More informationMINIMAX OPTIMIZATION FOR HANDLING RANGE AND SETUP UNCERTAINTIES IN PROTON THERAPY
MINIMAX OPTIMIZATION FOR HANDLING RANGE AND SETUP UNCERTAINTIES IN PROTON THERAPY Albin FREDRIKSSON, Anders FORSGREN, and Björn HÅRDEMARK Technical Report TRITA-MAT-1-OS2 Department of Mathematics Royal
More informationph fax Deming Way Middleton WI USA
www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 3120 Deming Way Middleton WI 53562-1461 USA 2007 Standard Imaging, Inc. 1239-21 (03.07) beam qa LINAC AND TOMOTHERAPY QA The NEW
More informationDose Calculations: Where and How to Calculate Dose. Allen Holder Trinity University.
Dose Calculations: Where and How to Calculate Dose Trinity University www.trinity.edu/aholder R. Acosta, W. Brick, A. Hanna, D. Lara, G. McQuilen, D. Nevin, P. Uhlig and B. Slater Dose Calculations - Why
More informationTwo Effective Heuristics for Beam Angle Optimization in Radiation Therapy
Two Effective Heuristics for Beam Angle Optimization in Radiation Therapy Hamed Yarmand, David Craft Department of Radiation Oncology Massachusetts General Hospital and Harvard Medical School, Boston,
More informationMapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays
MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.
More informationEvaluation of 3D Gamma index calculation implemented in two commercial dosimetry systems
University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2015 Evaluation of 3D Gamma index calculation implemented
More informationIterative regularization in intensity-modulated radiation therapy optimization. Carlsson, F. and Forsgren, A. Med. Phys. 33 (1), January 2006.
Iterative regularization in intensity-modulated radiation therapy optimization Carlsson, F. and Forsgren, A. Med. Phys. 33 (1), January 2006. 2 / 15 Plan 1 2 3 4 3 / 15 to paper The purpose of the paper
More informationComputer-Tomography II: Image reconstruction and applications
Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de
More information4 Measurement. and Analysis. 4.1 Overview and Underlying Principles 4-1
Measurement and Analysis.1 Overview and Underlying Principles.1.1 Introductory Remarks The physics and setup for film dosimetry have been described in the previous chapters. The measurement setup for IMRT
More informationMapCHECK 2 & 3DVH The Gold Standard for 2D Arrays
MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.
More informationarxiv: v3 [physics.med-ph] 6 Dec 2011
Multicriteria VMAT optimization David Craft, Dualta McQuaid, Jeremiah Wala, Wei Chen, Ehsan Salari, Thomas Bortfeld December 5, 2011 arxiv:1105.4109v3 [physics.med-ph] 6 Dec 2011 Abstract Purpose: To make
More informationDigital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities
sup17_15_newrtradiations.doc: Sup 17: nd Generation RT Additional RT Treatment Modalities Page 1 Digital Imaging and Communications in Medicine (DICOM) Supplement 17: Second Generation Radiotherapy Additional
More informationViewRay System Commissioning
ViewRay System Commissioning Kyle R. Padgett, PhD, DABR Assistant Professor University of Miami Viewray Team: Matt Studenski, PhD Chet Ford, PhD Yidong Yang, PhD Nesrin Dogan, PhD Medical Physics Residents:
More informationComputational modelling and analysis: IOP Medical Physics Group
Computational modelling and analysis: IOP Medical Physics Group Using Bespoke Matlab Software for the Collection and Processing of Dynalog Files from the Varian Millennium 120 MLC system following delivery
More informationA secondary monitor unit calculation algorithm using superposition of symmetric, open fields for IMRT plans
Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2013 A secondary monitor unit calculation algorithm using superposition of symmetric, open fields for IMRT plans Adam
More informationAn experimental investigation on the effect of beam angle optimization on the reduction of beam numbers in IMRT of head and neck tumors
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 An experimental investigation on the effect of beam angle optimization on the reduction of beam numbers in IMRT of head and neck tumors
More informationDose Calculation and Optimization Algorithms: A Clinical Perspective
Dose Calculation and Optimization Algorithms: A Clinical Perspective Daryl P. Nazareth, PhD Roswell Park Cancer Institute, Buffalo, NY T. Rock Mackie, PhD University of Wisconsin-Madison David Shepard,
More informationFAST, precise. qa software
qa software FAST, precise Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, rotational or brachytherapy plans no film, no phantoms, no linac
More informationATTENTION! The whole content of the lecture with all the animations can be dowloaded from:
ATTENTION! The whole content of the lecture with all the animations can be dowloaded from: http://www.onko.szote.uszeged.hu/letoltes/radioter_phys_basis/radiother_phys_basis.zip Size > 300 M!!! The dowloaded.zip
More informationTRAJECTORY- BASED TREATMENT PLANNING AND DELIVERY FOR CRANIAL RADIOSURGERY
TRAJECTORY- BASED TREATMENT PLANNING AND DELIVERY FOR CRANIAL RADIOSURGERY JAMES ROBAR, PHD, FCCPM LEE MACDONALD, MSC CHRISTOPHER THOMAS, PHD, MCCPM DALHOUSIE UNIVERSITY, CANADA CLASS SOLUTIONS IN SRS/SRT
More informationradiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li NCAAPM Spring Meeting 2010 Madison, WI
GPU-Accelerated autosegmentation for adaptive radiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li agodley@mcw.edu NCAAPM Spring Meeting 2010 Madison, WI Overview Motivation Adaptive
More informationA Fully-Automated Intensity-Modulated Radiation Therapy Planning System
A Fully-Automated Intensity-Modulated Radiation Therapy Planning System Shabbir Ahmed, Ozan Gozbasi, Martin Savelsbergh Georgia Institute of Technology Ian Crocker, Tim Fox, Eduard Schreibmann Emory University
More informationInteractive Treatment Planning in Cancer Radiotherapy
Interactive Treatment Planning in Cancer Radiotherapy Mohammad Shakourifar Giulio Trigila Pooyan Shirvani Ghomi Abraham Abebe Sarah Couzens Laura Noreña Wenling Shang June 29, 212 1 Introduction Intensity
More informationFour-Dimensional MVCT Imaging and Spatial Translation of Helical Tomotherapy Dose Distributions Using Sinogram Modification
University of Tennessee, Knoxville Trace: Tennessee Research and Creative Exchange Masters Theses Graduate School 12-2006 Four-Dimensional MVCT Imaging and Spatial Translation of Helical Tomotherapy Dose
More informationHelical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice Guido Hilgers, a Tonnis Nuver, André Minken Department
More informationA comparative analysis for verification of IMRT and VMAT treatment plans using a 2-D and 3-D diode array
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2014 A comparative analysis for verification of IMRT and VMAT treatment plans using a 2-D and 3-D diode array
More informationDelta 4 PT. True Volumetric Pre-treatment Verification. The difference is clear
Delta 4 PT True Volumetric Pre-treatment Verification The difference is clear ScandiDos innovators in advanced volumetric dosimetry solutions ScandiDos ScandiDos is the innovative company that introduced
More informationA Mimetic Algorithm for Simultaneous Multileaf Collimator Aperture Shape and Dosimetric Optimization in CyberKnife Robotic Radiosurgery
A Mimetic Algorithm for Simultaneous Multileaf Collimator Aperture Shape and Dosimetric Optimization in CyberKnife Robotic Radiosurgery Matthew R. Witten, Owen C. Clancey, Division of Radiation Oncology,
More informationCHAPTER 9 INFLUENCE OF SMOOTHING ALGORITHMS IN MONTE CARLO DOSE CALCULATIONS OF CYBERKNIFE TREATMENT PLANS: A LUNG PHANTOM STUDY
148 CHAPTER 9 INFLUENCE OF SMOOTHING ALGORITHMS IN MONTE CARLO DOSE CALCULATIONS OF CYBERKNIFE TREATMENT PLANS: A LUNG PHANTOM STUDY 9.1 INTRODUCTION 9.1.1 Dose Calculation Algorithms Dose calculation
More informationA Fully-Automated Intensity-Modulated Radiation Therapy Planning System
A Fully-Automated Intensity-Modulated Radiation Therapy Planning System Shabbir Ahmed, Ozan Gozbasi, Martin Savelsbergh Georgia Institute of Technology Tim Fox, Ian Crocker, Eduard Schreibmann Emory University
More informationDosimetry Simulations with the UF-B Series Phantoms using the PENTRAN-MP Code System
Dosimetry Simulations with the UF-B Series Phantoms using the PENTRAN-MP Code System A. Al-Basheer, M. Ghita, G. Sjoden, W. Bolch, C. Lee, and the ALRADS Group Computational Medical Physics Team Nuclear
More informationAutomated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator
Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4
More informationGood Morning! Thank you for joining us
Good Morning! Thank you for joining us Deformable Registration, Contour Propagation and Dose Mapping: 101 and 201 Marc Kessler, PhD, FAAPM The University of Michigan Conflict of Interest I receive direct
More informationApplied Optimization Application to Intensity-Modulated Radiation Therapy (IMRT)
Applied Optimization Application to Intensity-Modulated Radiation Therapy (IMRT) 2008-05-08 Caroline Olsson, M.Sc. Topics Short history of radiotherapy Developments that has led to IMRT The IMRT process
More informationTrajectory optimization for dynamic couch rotation during volumetric modulated arc radiotherapy
Trajectory optimization for dynamic couch rotation during volumetric modulated arc radiotherapy Gregory Smyth 1, Jeffrey C Bamber 1, Philip M Evans 2 and James L Bedford 1 1 Joint Department of Physics,
More informationPerformance of a hybrid Monte Carlo-Pencil Beam dose algorithm for proton therapy inverse planning
Performance of a hybrid Monte Carlo-Pencil Beam dose algorithm for proton therapy inverse planning Ana Marıa Barragan Montero, a) and Kevin Souris Universite catholique de Louvain, Institut de Recherche
More informationCONTOURING ACCURACY. What Have We Learned? And Where Do We Go From Here? BEN NELMS, PH.D. AUGUST 15, 2016
CONTOURING ACCURACY What Have We Learned? And Where Do We Go From Here? BEN NELMS, PH.D. AUGUST 15, 2016 FIRST THINGS FIRST Happy Medical Dosimetrist s Week! OUTLINE 1. Objectives 2. The importance of
More informationPatient Set-ups and Tumor Localizations
Patient Set-ups and Tumor Localizations Amy S. Harrison Patient Positioning Prior to starting any localization or simulation procedure patients need to be positioned and immobilized Patients disease location
More informationGoals. Scope/exclusions. Outline: follow outline of MPPG (plus rationale & some implementation experiences)
AAPM MEDICAL PHYSICS PRACTICE GUIDELINE # : Commissioning and QA of Treatment Planning Dose Calculations: Megavoltage Photon and Electron Beams Jennifer Smilowitz (Chair), University of Wisconsin-Madison
More informationCommissioning, beam matching and small field dosimetry experience with Elekta Versa
Commissioning, beam matching and small field dosimetry experience with Elekta Versa Raj Varadhan, PhD Director of Physics, Technology Minneapolis Radiation Oncology Outline Elekta Versa/Infinity platform
More informationGrupo Física Médica de Sevilla (CTS-233) Antonio Leal Plaza alplaza@us.es Dept. Fisiología Médica y Biofísica, Fac. Medicina. Univ. Sevilla http://grupos.us.es/medicalphysics/ Overview Introduction to
More informationA Genetic Algorithm approach to Full Beam Configuration Inverse Planning in Coplanar Radiotherapy
A Genetic Algorithm approach to Full Beam Configuration Inverse Planning in Coplanar Radiotherapy Vitoantonio Bevilacqua Giuseppe Mastronardi Giuseppe Piscopo bevilacqua@poliba.it mastrona@poliba.it giuseppe.piscopo@katamail.com
More information