7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT
|
|
- Aldous Austin
- 5 years ago
- Views:
Transcription
1 CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance testing procedures for surface imaging systems Understand how surface imaging isocenter congruence testing must be incorporated into ongoing quality assurance of IGRT Describe the commissioning process of surface imaging systems for whole breast radiotherapy Video Positioning Video Positioning Re-record after each MV verification Video Positioning 3D Surface Imaging by VisionRT Advantages: Reproducing patient position from last filmed fraction thereby reducing filming frequency Real-time monitoring during treatment Disadvantages: Qualitative not quantitative feedback No interface with LINAC for beam-hold 1
2 Learning Objective Describe acceptance testing procedures for surface imaging systems Understand how surface imaging isocenter congruence testing must be incorporated into ongoing quality assurance of IGRT Describe the commissioning process of surface imaging systems for whole breast radiotherapy Acceptance Testing Acceptance test procedures provided by the vendor are typically designed to verify contractual system specifications for performance characteristics of the system. (TG-101) Acceptance Tests Note: on importing DICOM RT data as reference surface, VisionRT accepts no responsibility for the accuracy of such data which may be affected by data resolution and breathing artefacts. Such errors could influence positioning accuracy. Optimize camera positions Verify DICOM transfer from TPS Verify translational/rotational accuracy of registration algorithm Verify accuracy of retrospective gating Verify accuracy of prospective gating For both surface imaging modes: 1) Static capture 2) Real-time monitoring Additional Acceptance Tests? Acceptance: DICOM transfers Consistency of real-time data With static capture mode Over time Verify coincidence with treatment isocenter Verify beam hold-off capabilities 2
3 Acceptance: DICOM transfers Reference and captured surfaces are from 2 different patients! Acceptance: DICOM transfers Prone Orientation TPS AlignRT Acceptance: Translational & Rotational Accuracy Acceptance: Gating LAT Shift = 2.0cm VRT, LNG, LAT Shift = 2.0cm According to TG-76, QA of gating systems must address two fundamental sources of potential error in dose delivery: Determination of the tumor position as a function of time Calibration of the spatial relationship between the tracking coordinate system and the beam-delivery coordinate system 4D MOTION PHANTOM All Shifts = 0.0cm Rotation = 10 o Rotation = 45o Acceptance: Prospective Gating Acceptance: Real-Time Data Stability AlignRT Innacurate Tracking of Tumor Motion Accurate Tracking of Tumor Motion Philips BigBore: Table motion during scanning is out of bore Delay between trigger and acquisition Polaris Lights Off Lights On Data courtesy of Rodney Wiersma, Ph.D. at Univ. of Chicago 3
4 Acceptance: Isocenter Coincidence TG-142: Imaging & Treatment Isocenter Coincidence If individual errors are small by themselves, cumulative system accuracy for the procedure can be significant and needs to be characterized through an end-to-end test using phantoms with measurement detectors and imaging (TG-101) Learning Objective Describe acceptance testing procedures for surface imaging systems Understand how surface imaging isocenter congruence testing must be incorporated into ongoing quality assurance of IGRT Describe the commissioning process of surface imaging systems for whole breast radiotherapy IGRT QA Goals To develop an end-to-end test to characterize isocenter coincidence: Planning CT MV kv CBCT AlignRT To incorporate additional tests: DICOM transfer Translational & rotation accuracy Multiple Imaging Modality Isocentricity (MiMi) Phantom from Standard Imaging MiMi Phantom Makeover Easy Alignment due to Unique Design: The MIMI Phantom incorporates five bone equivalent rods uniquely set so that four of them intersect at 90 o angles when viewed in DRRs or a 2D projection image. The rods traverse the entire phantom making them visible in any image or slice allowing for easy 2D/2D and 3D/3D matching for fast verification of isocenter position. Test Integrated System Accuracy of: 3D Cone Beam MV/kV Lasers and Couch Table Adjustments Optical Guidance Systems Test Automatic Table Adjustments: Additional cross-hair markers that are offset known distances from the true isocenter allow for verification of the shifts prescribed by automatic table positioning systems. 4
5 MiMi Phantom Makeover Al 2 O 3 Al 2 O 3 Isocenter Coincidence Testing Axial CT Scanning (0.75mm) DICOM Transfer to TPS (Verify Geometric accuracy) Align Phantom to Lasers DICOM Transfer from TPS with DRRs and RT Structures MV 1MU Al 2 O 3 kv ZrO 2 Center Phantom in Radiation Isocenter by MV imaging at 4 orthogonal angles Measure Offset to kv Isocenter by imaging at 4 orthogonal angles Measure Offset to CBCT Isocenter Introduce Known Physical Shift & Measure Accuracy Measure Offset to AlignRT Isocenter Measure Offset to Laser Isocenter MiMi Phantom Axial CT Scans Center Phantom in MV Isocenter by Imaging at 4 Gantry Angles Measure Offset to kv Isocenter by 2D/2D Match at 4 Angles Measure Offset to CBCT Isocenter Dependent upon CBCT Technique! 5
6 Measure Offset to Laser Isocenter Root Mean Square Distances of IGRT Isocenter Offsets IGRT isocenters aligned to SRS isocenter at our institution! TG-142: Imaging & Treatment Isocenter Coincidence Learning Objective Describe acceptance testing procedures for surface imaging systems Understand how surface imaging isocenter congruence testing must be incorporated into ongoing quality assurance of IGRT Describe the commissioning process of surface imaging systems for whole breast radiotherapy Commissioning Commissioning tests should be developed by the institution s physics team to explore in detail every aspect of the system with the goal of developing a comprehensive baseline characterization of the performance of the system. (TG-101) Commissioning: Whole-Breast RT While the patient s surface is a good surrogate for the target, the quality of 3D surface registration could be compromised by deformation We investigated the reliability of 3D surface matching using AlignRT compared to positioning using skin marks followed by MV portal imaging for whole-breast radiotherapy (WBRT) 6
7 WBRT: Methods Supine Immobilization with both arms above the head: Breastboard (n=2) Alphacradle (n=3) No respiratory gating Alignment for breast positioning includes: Setup to skin marks using lasers daily MV verification & alignment weekly Orthogonal pair (AP & LAT) Tangential portal images Commissioning: WBRT 3D surface generated from their free-breathing CT scan was used as reference Absolute or Relative positioning? Absolute: Inter-fraction reproducibility? Could reduce frequency of filming? Relative: Intra-fraction reproducibility? Relies on use of other IGRT modality WBRT: Methods Commissioning: WBRT Average 3D distance: 3.8 ± 5.1 mm: table shifts after MV imaging 5.9 ± 4.1mm: AlignRT registration of all 5.6 ± 3.3mm: AlignRT registration of Breast 3D Surface from CT data All ROI Breast ROI Regression analysis indicated that only all registration correlated with table shifts Table shifts were detected accurately with AlignR for both ROIs (r 0.9) WBRT: Correlation with MV WBRT: Correlation with Table Shifts Results presented in Poster SU-E-T-228 7
8 Commissioning: WBRT Registration of Breast ROI Absolute table rotations calculated by AlignRT for consecutively acquired surfaces exhibited fluctuations that were significantly larger for registrations of: Breast : 0.62 o ±0.58 o All : 0.27 o ±0.18 o Registration of All yielded rotations of 1.32 o and 1.37 o. WBRT: Conclusions Breast surface registration depends upon the ROI A large ROI showed: Higher correlation with MV shifts Increased stability when calculating table rotations Before clinical use, we must investigate the sources of error that prevent a direct one-to-one correlation between AlignRT and MV 3D shifts including: Respiratory motion Misalignment between the MV and AlignRT isocenters Deformation of breast surface Misinterpretation errors due to MV imaging Acceptance & Commissioning: Conclusions Surface imaging systems must undergo extensive acceptance testing & commissioning Isocenter congruence must be verified for all imaging modalities simultaneously Comprehensive testing of VisionRT product revealed in-operational components Commissioning of AlignRT indicates that discrepancies are larger for: Registration to CT reference surface Registration of smaller ROIs Utility of Surface Imaging? Reduce filming frequency Requires absolute positioning? Improve patient safety Particularly for multiple isocenter treatments Improve intra-fraction positioning Real-time monitoring throughout treatment Improve throughput Utility of Surface Imaging? Variability in repositioning is dominated by the ability of therapists to make small, controlled changes in the position of the patient. (Milliken et al., Int J Rad Onc Biol Phys, 38(4): , 1997) Surface imaging does not preclude need for: Good immobilization Adequate PTV margins Common sense! 8
9 Acknowledgements Charles Pelizzar, Ph.D., Director of Physics Kamil Yenice, Ph.D., Clinical Physics Director Karl Farrey, M.S., Dosimetry/Physics Manager Rodney Wiersma, Ph.D., Assistant Professor y,, Hyejoo Kang, Ph.D., Research Associate Ji Li, Ph.D., Physics Resident Dimple Modgil, Ph.D., Physics Resident 9
1. Learn to incorporate QA for surface imaging
Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago ***No disclosures*** 1. Learn to incorporate QA for surface imaging into current QA procedures for IGRT. 2. Understand
More informationInitial Clinical Experience with 3D Surface Image Guidance
Initial Clinical Experience with 3D Surface Image Guidance Amanda Havnen-Smith, Ph.D. Minneapolis Radiation Oncology Ridges Radiation Therapy Center Burnsville, MN April 20 th, 2012 Non-funded research
More informationOptical Guidance. Sanford L. Meeks. July 22, 2010
Optical Guidance Sanford L. Meeks July 22, 2010 Optical Tracking Optical tracking is a means of determining in real-time the position of a patient relative to the treatment unit. Markerbased systems track
More information8/3/2016. Image Guidance Technologies. Introduction. Outline
8/3/26 Session: Image Guidance Technologies and Management Strategies Image Guidance Technologies Jenghwa Chang, Ph.D.,2 Department of Radiation Medicine, Northwell Health 2 Hofstra Northwell School of
More informationChoosing and Commissioning a Video Based Motion Management System
Choosing and Commissioning a Video Based Motion Management System David Shepard Swedish Cancer Institute 2 1 Acknowledgments Daliang Cao, PhD, SCI Mohammed Jermoumi, PhD, SCI Roger Xie, PhD, SCI Malin
More informationImage Guidance and Beam Level Imaging in Digital Linacs
Image Guidance and Beam Level Imaging in Digital Linacs Ruijiang Li, Ph.D. Department of Radiation Oncology Stanford University School of Medicine 2014 AAPM Therapy Educational Course Disclosure Research
More informationURGENT IMPORTANT FIELD SAFETY NOTIFICATION
Subject: Incorrect Movement of the Treatment Table Product: MOSAIQ Scope: Sites affected will be those: 1. Running MOSAIQ and, 2. Treating on linear accelerators with the RATM license Notification Released:
More informationThank-You Members of TG147 TG 147: QA for nonradiographic
Thank-You Members of TG147 TG 147: QA for nonradiographic localization and positioning systems Twyla Willoughby, M.S. Medical Physicist Clinical AAPM Meeting March 2013 Department of Radiation Oncology
More information8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques
Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery Lei Ren, Ph.D. Duke University Medical Center 2016 AAPM 58 th annual meeting, Educational Course, Therapy Track Disclosure I have
More informationAutomated Quality Assurance for Image-Guided Radiation Therapy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 1, WINTER 2009 Automated Quality Assurance for Image-Guided Radiation Therapy Eduard Schreibmann, a Eric Elder, Tim Fox Department of Radiation
More informationFeasibility study of performing IGRT system daily QA using a commercial QA device
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 3, summer 2011 Feasibility study of performing IGRT system daily QA using a commercial QA device Jean L.Peng, 1 Darren Kahler, 2 Jonathan
More informationReal Time Tumor Motion Tracking with CyberKnife
Real Time Tumor Motion Tracking with CyberKnife Martina Descovich, Ph.D University of California San Francisco July 16, 2015 Learning objectives Review the principles of real-time tumor motion tracking
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More informationLaser Systems for Patient Alignment
Laser Systems for Patient Alignment GALAXY Patient TOPOGRAPHY Laser System Page: 2 Date: 28 March 2009 The GALAXY System NEW METHOD FOR HIGHER ALIGNMENT ACCURACY. Patient Positioning System for Linac Simulator
More informationPerformance and characteristics of an IR localizing system for radiation therapy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 2, SPRING 2006 Performance and characteristics of an IR localizing system for radiation therapy Yulia Lyatskaya, 1 Hsiao-Ming Lu, 2 and Lee
More informationTHE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!
THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER! Confidence in complex treatments Modern radiation therapy uses complex plans with techniques such as IMRT, VMAT and Tomotherapy.
More informationImage Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD
Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image
More informationVersion 5.6. Quick Start Guide
WWW..COM Version 5.6 Quick Start Guide STANDARD IMAGING, INC. 3120 Deming Way Middleton, WI 53562-1461 Jul / 2018 2018 Standard Imaging, Inc. TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 DOC # 80714-05
More informationMR-guided radiotherapy: Vision, status and research at the UMC Utrecht. Dipl. Ing. Dr. Markus Glitzner
MR-guided radiotherapy: Vision, status and research at the UMC Utrecht Dipl. Ing. Dr. Markus Glitzner About myself Training Medizintechnik TU Graz PhD UMC Utrecht Clinical work Software implementation
More informationNew Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC
New Technology in Radiation Oncology James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC Technology s s everywhere From the imaging chain To the planning system To the linac To QA..it..it
More informationAutomated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator
Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4
More informationVirtual Phantoms for IGRT QA
TM Virtual Phantoms for IGRT QA Why ImSimQA? ImSimQA was developed to overcome the limitations of physical phantoms for testing modern medical imaging and radiation therapy software systems, when there
More informationEstimating 3D Respiratory Motion from Orbiting Views
Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct. 2005 Funding provided by NIH Grant P01 CA59827 Motivation Free-breathing
More informationTumor motion during liver SBRT
Tumor motion during liver SBRT - projects at Aarhus University Hospital - Per Poulsen, Esben Worm, Walther Fledelius, Morten Høyer Aarhus University Hospital, Denmark SBRT: Stereotactic Body Radiation
More informationBrilliance CT Big Bore.
1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.
More informationThe team. Disclosures. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge.
Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge Dimitre Hristov Radiation Oncology Stanford University The team Renhui Gong 1 Magdalena Bazalova-Carter 1
More informationCalibration of Video Cameras to the Coordinate System of a Radiation Therapy Treatment Machine
Calibration of Video Cameras to the Coordinate System of a Radiation Therapy Treatment Machine Scott W. Hadley, L. Scott Johnson, and Charles A. Pelizzari University of Chicago The Department of Radiation
More informationIntroduction. Quality Assurance for Image- Guided Radiation Therapy. Justification for IGRT. Image-Guided Radiation Therapy
Introduction Quality Assurance for Image- Guided Radiation Therapy Jean-Pierre Bissonnette, Ph.D., MCCPM Princess Margaret Hospital, Toronto, Canada IGRT What is it? Rationale Equipment Quality Assurance
More informationAcknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF
Page 1 Quality Assurance of IMRT Delivery Systems - Siemens Lynn J. Verhey, Ph.D. Professor and Vice-Chair UCSF Dept. of Radiation Oncology AAPM 22 Annual Meeting, Montreal Acknowledgments Ping Xia, Ph.D.,
More informationDigital Tomosynthesis for Target Localization
Digital Tomosynthesis for Target Localization Fang-Fang Yin, Devon Godfrey, Lei Ren Jacqueline Maurer, Jackie Q-L Wu Duke University Medical Center Acknowledgements Duke Radiation Oncology faculty and
More informationS. Guru Prasad, Ph.D., DABR
PURPOSE S. Guru Prasad, Ph.D., DABR Director of Medical Physics IAEA Consultant NorthShore University Health System and University of Chicago, Pritzker School of Medicine Current TPS utilize more information
More informationState-of-the-Art IGRT
in partnership with State-of-the-Art IGRT Exploring the Potential of High-Precision Dose Delivery and Real-Time Knowledge of the Target Volume Location Antje-Christin Knopf IOP Medical Physics Group Scientific
More informationICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588
ICARO Vienna April 27-29 2009 Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588 M. Saiful Huq, Ph.D., Professor and Director, Dept. of Radiation
More informationArtefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT
Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT Marcus Brehm 1,2, Thorsten Heußer 1, Pascal Paysan 3, Markus Oehlhafen 3, and Marc Kachelrieß 1,2 1 German Cancer Research Center
More informationDEVELOPING AND IMPLEMENTING A HIGH PRECISION SETUP SYSTEM
DEVELOPING AND IMPLEMENTING A HIGH PRECISION SETUP SYSTEM By LEE-CHENG PENG A DISSERTATION PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE
More information8/2/2017. PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA
PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA Jennifer R. Clark, M.S., DABR Conflict of Interest Statement: I am an employee of Sun Nuclear Corporation Outline: Platform overview
More informationMeasurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images
Measurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images T E Marchant, A M Amer 1 and C J Moore North Western Medical Physics, Christie Hospital NHS Foundation
More informationTomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS
Tomotherapy Physics Machine Twinning and Quality Assurance Emilie Soisson, MS Tomotherapy at UW- Madison Treating for nearly 5 years Up to ~45 patients a day on 2 tomo units Units twinned to facilitate
More informationOverview of Proposed TG-132 Recommendations
Overview of Proposed TG-132 Recommendations Kristy K Brock, Ph.D., DABR Associate Professor Department of Radiation Oncology, University of Michigan Chair, AAPM TG 132: Image Registration and Fusion Conflict
More informationImage Co-Registration II: TG132 Quality Assurance for Image Registration. Image Co-Registration II: TG132 Quality Assurance for Image Registration
Image Co-Registration II: TG132 Quality Assurance for Image Registration Preliminary Recommendations from TG 132* Kristy Brock, Sasa Mutic, Todd McNutt, Hua Li, and Marc Kessler *Recommendations are NOT
More information3D Surface Imaging for PBI Patient Setup
3D Surface Imaging for PBI Patient Setup G.T.Y. Chen 1, Ph.D., M. Riboldi 2, Ph.D. Christoph Bert 3, Ph.D., D.P. Gierga 1, Ph.D. 1 Massachusetts General Hospital Harvard Medical School 2 TBMLab - Department
More information7/31/ D Cone-Beam CT: Developments and Applications. Disclosure. Outline. I have received research funding from NIH and Varian Medical System.
4D Cone-Beam CT: Developments and Applications Lei Ren, PhD, DABR Department of Radiation Oncology Duke University Medical Center Disclosure I have received research funding from NIH and Varian Medical
More informationFacility Questionnaire PART I (General Information for 3DCRT and IMRT)
Facility Questionnaire PART I (General Information for 3DCRT and IMRT) The following items are required before you can enter cases on any RTOG protocol that requires data submission to the Image-Guided
More informationIMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays
IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays Sotiri Stathakis Outline Why IMRT/VMAT QA AAPM TG218 UPDATE Tolerance Limits and Methodologies for IMRT Verification QA Common sources
More informationRITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package.
So much is asked of medical physicists these days We Can Help Automated Radiation Therapy Phantom Analysis in Seconds QC for Therapy OBI s and CT simulators. Radia s Catphan/OBI module performs analysis
More informationADVANCING CANCER TREATMENT
3 ADVANCING CANCER TREATMENT SUPPORTING CLINICS WORLDWIDE RaySearch is advancing cancer treatment through pioneering software. We believe software has un limited potential, and that it is now the driving
More informationADVANCING CANCER TREATMENT
The RayPlan treatment planning system makes proven, innovative RayStation technology accessible to clinics that need a cost-effective and streamlined solution. Fast, efficient and straightforward to use,
More informationIAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques
IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques Report of the Coordinated Research Project (CRP) on Development of Procedures
More informationCommissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility
Technical Note Commissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility K. R. Muralidhar, Krishna Komanduri, Birendra Kumar Rout, K. K.
More informationCode of practice: Create a verification plan for OCTAVIUS Detector 729 in Philips Pinnacle³
Technical Note D655.208.03/00 Code of practice: Create a verification plan for OCTAVIUS Detector 729 in There are basically three ways to carry out verification of IMRT fluences. For all options, it is
More informationProtocol. Technical evaluation of X-ray tomographic image-guided radiotherapy devices CEP10070
Protocol Technical evaluation of X-ray tomographic image-guided radiotherapy devices CEP10070 March 2010 Contents 2 Introduction... 3 Protocol design and validation... 4 General information... 7 Technical
More informationCommissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility
0 0 0 0 JMP R Commissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility K. R. Muralidhar, Krishna Komanduri, Birendra Kumar Rout, K. K. D.
More informationQuick Reference Datasheet For All RIT113 Packages
Quick Reference Datasheet For All RIT113 Packages For Rotational Therapies, IMRT & TG142 Highlights and selected product information only. A complete TG142 brochure is available. For more information on
More informationPosition accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon
Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon WHITE PAPER Introduction An image guidance system based on Cone Beam CT (CBCT) is included in Leksell
More informationAdvanced Radiotherapy
Advanced Radiotherapy IBA Concept for IMRT Verification Dr. Lutz Müller 1 IMRT Verification BIS BIS ADAS Fluence Model Delivered fluence Target Volume Constraints Inverse Backprojection Fluence Map Leaf
More informationA novel phantom and procedure providing submillimeter accuracy in daily QA tests of accelerators used for stereotactic radiosurgery *
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 A novel phantom and procedure providing submillimeter accuracy in daily QA tests of accelerators used for stereotactic radiosurgery
More informationUsing a research real-time control interface to go beyond dynamic MLC tracking
in partnership with Using a research real-time control interface to go beyond dynamic MLC tracking Dr. Simeon Nill Joint Department of Physics at The Institute of Cancer Research and the Royal Marsden
More informationImprovement and Evaluation of a Time-of-Flight-based Patient Positioning System
Improvement and Evaluation of a Time-of-Flight-based Patient Positioning System Simon Placht, Christian Schaller, Michael Balda, André Adelt, Christian Ulrich, Joachim Hornegger Pattern Recognition Lab,
More informationUse of TrueBeam developer mode for imaging QA
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Use of TrueBeam developer mode for imaging QA Gilmer Valdes, 1,2a Olivier Morin, 1 Yanisley Valenciaga, 3 Niel Kirby, 1 Jean Pouliot,
More informationClinical implementation of photon beam flatness measurements to verify beam quality
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Clinical implementation of photon beam flatness measurements to verify beam quality Simon Goodall, a Nicholas Harding, Jake Simpson,
More informationEvaluation of AutoQA Lite TM Image Quality Measurement Software
Evaluation of AutoQA Lite TM Image Quality Measurement Software Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4 2XU Phone: 0131 537 1161
More informationradiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li NCAAPM Spring Meeting 2010 Madison, WI
GPU-Accelerated autosegmentation for adaptive radiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li agodley@mcw.edu NCAAPM Spring Meeting 2010 Madison, WI Overview Motivation Adaptive
More informationComparison of Two Phantoms for End-to-End SRS/SBRT Testing. Vikren Sarkar, Ph.D. Associate Professor University of Utah Huntsman Cancer Center
Comparison of Two Phantoms for End-to-End SRS/SBRT Testing Vikren Sarkar, Ph.D. Associate Professor University of Utah Huntsman Cancer Center Disclosure Travel and Speaker Honorarium from Sun Nuclear Corp.
More informationBasic Radiation Oncology Physics
Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides
More informationEvaluation of the Elekta Synergy concept for patient positioning in image guided radiotherapy
Master of Science Thesis Evaluation of the Elekta Synergy concept for patient positioning in image guided radiotherapy Johan Renström Supervisor: Per Nilsson, PhD and Tommy Knöös, PhD Medical Radiation
More informationSimple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Simple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field Hideharu Miura, 1a Shuichi
More informationDelta 4 PT. True Volumetric Pre-treatment Verification. The difference is clear
Delta 4 PT True Volumetric Pre-treatment Verification The difference is clear ScandiDos innovators in advanced volumetric dosimetry solutions ScandiDos ScandiDos is the innovative company that introduced
More informationIMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE
QA SOFTWARE FAST, PRECISE Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, VMAT, SRS and brachytherapy plans no film, no phantoms, no linac
More informationPET-CT in Radiation Treatment Planning
PET-CT in Radiation Treatment Planning TA van de Water, Radiotherapeutic Institute Friesland, Leeuwarden JA van Dalen, Isala, Zwolle ACKNOWLEDGEMENTS A. van der Schaaf (University of Groningen, University
More informationDigital phantoms for the evaluation of a software used for an automatic analysis of the Winston-Lutz test in image guided radiation therapy
Author manuscript, published in "Medical Imaging 008: Physics of Medical Imaging, San Diego, CA, USA : United States (008)" DOI : 10.1117/1.768668 Digital phantoms for the evaluation of a software used
More informationMRI-LINAC Dynamic Phantom
MRI-LINAC Dynamic Phantom Model 008M IMAGE ACQUISITION TREATMENT PLANNING DOSE DELIVERY Strict QA procedures for the imaging, planning and delivery of radiotherapy using respiratory management devices
More informationIMRT site-specific procedure: Prostate (CHHiP)
IMRT site-specific procedure: Prostate (CHHiP) Scope: To provide site specific instructions for the planning of CHHIP IMRT patients Responsibilities: Radiotherapy Physicists, HPC Registered Therapy Radiographers
More informationPatient Set-ups and Tumor Localizations
Patient Set-ups and Tumor Localizations Amy S. Harrison Patient Positioning Prior to starting any localization or simulation procedure patients need to be positioned and immobilized Patients disease location
More informationREAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT
REAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT Anand P Santhanam Assistant Professor, Department of Radiation Oncology OUTLINE Adaptive radiotherapy for head and
More information8/3/2016. MR-guided RT: Commissioning and Quality Control. Topics for MR-guided RT system commissioning & QC
MR-guided RT: Commissioning and Quality Control T. Stanescu, PhD, MCCPM Medical Physicist, Princess Margaret Cancer Centre, UHN Assistant Professor, Radiation Oncology, University of Toronto Affiliated
More informationTG 132: Use of Image Registration and Fusion in RT
TG 132: Use of Image Registration and Fusion in RT Kristy K Brock, PhD, DABR, FAAPM Associate Professor Department of Radiation Oncology, University of Michigan Chair, AAPM TG 132: Image Registration and
More informationReal-time monitoring of linac performance using RT plans and logfiles
Real-time monitoring of linac performance using RT plans and logfiles NCCAAPM Wisconsin Dells 7 April 2017 Mark Wiesmeyer, PhD, DABR Technical Product Manager Technical talk with insights and results based
More information2018/7/31. Automation and Artificial Intelligence for Precision Radiation Therapy QA. Bin Han Ph.D. July 31, 2018 AAPM Annual Meeting
Automation and Artificial Intelligence for Precision Radiation Therapy QA Bin Han Ph.D. July 31, 2018 AAPM Annual Meeting Department of Radiation Oncology School of Medicine Disclosure I have no conflicts
More informationMapCHECK 2 & 3DVH The Gold Standard for 2D Arrays
MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.
More information4 Measurement. and Analysis. 4.1 Overview and Underlying Principles 4-1
Measurement and Analysis.1 Overview and Underlying Principles.1.1 Introductory Remarks The physics and setup for film dosimetry have been described in the previous chapters. The measurement setup for IMRT
More informationVolumetric Cine Imaging for On-board Target Localization in Radiation Therapy. Wendy Beth Harris. Graduate Program in Medical Physics Duke University
Volumetric Cine Imaging for On-board Target Localization in Radiation Therapy by Wendy Beth Harris Graduate Program in Medical Physics Duke University Date: Approved: Lei Ren, Supervisor Fang-Fang Yin,
More informationCARS 2008 Computer Assisted Radiology and Surgery
Online External Beam Radiation Planning and Training Felix Hamza-Lup a, Ivan Sopin a, Omar Zeidan b a Computer Science, Armstrong Atlantic State University, Savannah, Georgia, USA b MD Anderson Cancer
More informationA Study of Motion Tracking Accuracy of Robotic Radiosurgery Using a Novel CCD Camera Based End-to-end Test System
A Study of Motion Tracking Accuracy of Robotic Radiosurgery Using a Novel CCD Camera Based End-to-end Test System Lei Wang 1, Brett Nelson 2,Youming Yang 1 1. Department of Radiation Oncology, Stanford
More information3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH
3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk
More informationTable of Contents. About AQUA
User Manual About AQUA Table of Contents 1 About AQUA... 6 1.1 Radiation Therapy... 6 1.1.1 Workflow manager... 6 1.1.2 Reports... 7 1.1.3 Tests... 7 1.1.4 Many Ways to Work... 7 1.2 Technology... 7 1.3
More informationRaising the Bar in IMRT QA
MapCHECK 2TM Raising the Bar in IMRT QA The leader in quick and precise measurement of modulated radiotherapy beams Benefits Proven solution for film-less rotational delivery and IMRT QA - More than 1500
More informationImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing
ImPACT Information Leaflet No. 1: CT Scanner Acceptance Testing Version 1.02, 18/05/01 CONTENTS: 1. SCOPE OF LEAFLET 2. GENERAL PRINCIPLES OF ACCEPTANCE AND COMMISSIONING 2.1 PHANTOMS 2.2 EXPOSURE AND
More informationMapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays
MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.
More informationDosimetric Analysis Report
RT-safe 48, Artotinis str 116 33, Athens Greece +30 2107563691 info@rt-safe.com Dosimetric Analysis Report SAMPLE, for demonstration purposes only Date of report: ----------- Date of irradiation: -----------
More informationEvaluation report. X-ray tomographic image guided radiotherapy systems CEP10071
Evaluation report X-ray tomographic image guided radiotherapy systems CEP10071 March 2010 Contents 2 Summary... 3 Introduction... 6 Product description... 9 Methods... 14 Technical performance... 23 Purchasing...
More informationISO ISO ISO OHSAS ISO
ISO 9001 ISO 13485 ISO 14001 OHSAS 18001 ISO 27001 Pro-NM Performance 08-101 - standard version 08-103 - version with the PET Lid Phantom for NM and PET systems performance evaluation (collimator, artifacts,
More informationMoving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region
Moving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region Andreas Hahn 1,2, Sebastian Sauppe 1,2, Michael Knaup 1, and Marc Kachelrieß 1,2 1 German Cancer Research Center (DKFZ),
More informationDetermination of rotations in three dimensions using two-dimensional portal image registration
Determination of rotations in three dimensions using two-dimensional portal image registration Anthony E. Lujan, a) James M. Balter, and Randall K. Ten Haken Department of Nuclear Engineering and Radiological
More informationThe IORT Treatment Planning System. radiance. GMV, 2012 Property of GMV All rights reserved
The IORT Treatment Planning System radiance Property of GMV All rights reserved WHY RADIANCE? JUSTIFICATION Property of GMV All rights reserved ADVANTAGES OF IORT PRECISION: RT guided by direct vision.
More informationDesign and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon
Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization
More informationUse of image registration and fusion algorithms and techniques in radiotherapy: Report of the AAPM Radiation Therapy Committee Task Group No.
Use of image registration and fusion algorithms and techniques in radiotherapy: Report of the AAPM Radiation Therapy Committee Task Group No. 132 Kristy K. Brock a) Department of Imaging Physics, The University
More informationIs deformable image registration a solved problem?
Is deformable image registration a solved problem? Marcel van Herk On behalf of the imaging group of the RT department of NKI/AVL Amsterdam, the Netherlands DIR 1 Image registration Find translation.deformation
More informationNovel methods and tools for MR Commissioning and Quality Control
Novel methods and tools for MR Commissioning and Quality Control T. Stanescu, PhD, MCCPM Medical Physicist, RMP, Princess Margaret Cancer Centre Assistant Professor, Radiation Oncology, University of Toronto
More informationAn investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 An investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition Yildirim D. Mutaf
More informationDevelopment of Video Image Guided Setup (VIGS) System for Tomotherapy: Preliminary Study
Development of Video Image Guided Setup (VIGS) System for Tomotherapy: Preliminary Study Jin Sung Kim*, Sang Gyu Ju*, Chae Seon Hong*, Jaewon Jeong, Kihong Son*, Jung Suk Shin*, Eunheak Shin*, Sung Hwan
More information