1 Flip-Chip-Aligner-Bonder Model M-8HP 1. Introduction DO NOT CHANGE THE SETTINGS OF THE VIDEO CAMERAS UNLESS YOU ARE AUTHORIZED TO CALIBRATE THE MACHINE AND KNOW WHAT YOU ARE DOING! The laminar flow hood causes some vibration which might be disadvantageous for the alignment procedure. If so turn it off for the time of the alignment but turn it on again when you are finished. In the following steps for calibration and bonding blue color indicates status messages on the video screen and red color buttons on the keypad. 2. Alignment and Bonding Turn on the machine, monitor and mechanical pump. Press any key. Press 1 on the keypad to initialize the machine. Press 2 on keypad to begin operation. Always follow this procedure when you start up the machine. In the following steps for calibration and bonding blue color indicates status messages on the video screen and red color buttons on the keypad. (1) Place pedestal on upper chuck Switch on Pedestal VAC (2) Lower VAC switch (as many rings as necessary) (3) Load samples KEYPAD LO/VAC KEYPAD UP/VAC (4) Flip over upper chuck; Slide left in frame; KEYPAD CLAMP 1
2 (5) Press/toggle joystick-button CAM SCAN (6) Set micrometer on left side of machine to 5 (safety distance to avoid a crash when lower chuck is risen). (7) Move in the camera system between the two chucks. Make sure there is enough space! (8) KEYPAD CAMERA, until CAMERA: UP/ (9) KEYPAD FOCUS, use joystick to focus (10) Press/toggle joystick-button CAM SCAN (11) Search alignment marks and store the position KEYPAD MEM A, (hit key twice to store position) KEYPAD MEM B, (hit key twice to store position) (12) KEYPAD CAMERA, until CAMERA: /LO focus with the focus screw on the right side of the machine (13) Press/toggle joystick-button LO SCAN Find the corresponding alignment marks. (14) KEYPAD CAMERA, until CAMERA: UP/LO (15) Press/toggle joystick-button CAM SCAN Check the alignement KEYPAD GOTO MEM A/B, to move to the marked positions. (13) Press/toggle joystick-button LO SCAN (14) Correct the lateral misalignment using the joystick (15) KEYPAD ROTATE Correct for rotational misalignment using the joystik (16) Repeat steps (14) and (15) until good alignment is achieved for both alignment marks. (17) KEYPAD Il/COL (Switch to the autocollimator mode) (18) KEYPAD P/ROLL Align the two green crosshairs with joystick. (19) KEYPAD Il/COL (Switch back to video mode, the two sample surfaces are now parallel) (20) Press/toggle joystick-button CAM SCAN 2
3 (21) Check alignment again. If not o.k. start over at step (13) (22) Joystick: Move camera to the far left and back corner (Load position, XY LIMIT on screen, otherwise bonding not possible!) (23) KEYPAD RISER Micrometer (left side of the machine): Rise the lower chuck until separation ~0.5 1mm between the samples! (24) KEYPAD PROG AUTO Enter the parameters for the bonding process. (25) KEYPAD * Exit the program menu. (26) KEYPAD AUTO, Start the bonding process Pressing any key during the process will stop the process. 3
4 3. Calibration Turn on the machine, monitor and mechanical pump. Press any key. Press 1 on the keypad to initialize the machine. Press 2 on keypad to begin operation. Always follow this procedure when you start up the machine. In the following steps for calibration and bonding blue color indicates status messages on the video screen and red color buttons on the keypad. (1) Flip over upper chuck (open position) (2) Place square target reticle on lower chuck such that the word up can be read and is in the lower right corner. Align edges approximately parallel to the frame and the crosshair to the center hole of the chuck. (3) KEYPAD LO/VAC (4) Place calibration reticle mount in slide frame for upper chuck and slide left. (5) KEYPAD CLAMP (6) Set micrometer on left side of machine to 5 (safety distance to avoid a crash when lower chuck is risen). (7) KEYPAD RISER (8) Manually rise bottom chuck with micrometer on left side. Leave a small gap between the targets (~0.5-1mm) (9) KEYPAD CALIB KEYPAD ENT With this the safety features are turned off (SAFETY OFF). (10) Press/toggle joystick-button LO SCAN Align crosshairs roughly with the eye using the joystick. (11) Put calibration bridge above upper slideframe (three feet into the corresponding holes). Switch power supply to lamp #1. Turn the autocollimator in position. You should see two green crosshairs through the microscope. 4
5 (12) KEYPAD P/ROLL (13) Align the two green crosshairs using the joystick. The square target reticle on the lower chuck and the target in the upper slide frame should now be parallel! (14) Turn the NIKON objective on the calibration bridge in position. Switch to lamp #2 (15) Move the microscope with the manipulators such that the upper crosshair is focused and in the middle of the field of view. (16) Rise lower chuck manually (micrometer on left side of machine) until PRESS: >0. It can be ~ important is that the two parts do not touch. You should actually be able to see a very small gap. They have to be close enough to see both reticles with the microscope without changing the focus. Since you won t see both in focus if there is a gap between them you have to make a trade-off. So find a distance where the parts do not touch and you can still see both reticles with the microscope. THEY HAVE TO BE A LITTLE BIT OUT OF FOCUS. (17) Press/toggle joystick-button LO SCAN Align laterally with joystick KEYPAD ROTATE Repeat this step until you have an excellent alignment. (18) You now have aligned the to reticles perfectly and the plates are absolutely parallel. From now on, for the rest of the calibration procedure, do no further changes in the LO/SCAN, ROTATE or P/ROLL mode!!!! CHECK ON THE MONITOR WHICH MODE YOU ARE USING! (20) Remove the calibration bridge gently from the machine. (21) KEYPAD RISER, bring down lower chuck. (22) KEYPAD CALIB, turn on safety. (23) ONLY CHANGES TO THE VIDEO SYSTEM NOW! (24) Press/toggle joystick-button CAM SCAN, move in the camera and check if there is enough space between the two chucks to do so. If not you have to lower the lower chuck with the screw on the right side of the machine (focus screw) 5
6 (25) KEYPAD CAMERA, until CAMERA: UP/ Use joystick to search crosshair. (26) KEYPAD FOCUS, focus on the cross hair using the joystick. (27) KEYPAD CAMERA, until CAMERA: /LO focus with the focus screw on the right side of the machine (28) KEYPAD CAMERA, until CAMERA: UP/LO (29) Adjust the cameras with the CAMERA X and CAMERA Y micrometers behind the four lamps. (30) Rotational misalignment: follow (SECTION III page 13, step 39) in the manual. (31) KEYPAD IL/COL, switch to collimator (32) Align the two crosshairs using the X,Y screws (step 44, SECTION III page 14 in the manual) (END) 4. Use of low pressure load cell (a) Switch on the machine while pressing KEYPAD PROG (b) Change load cell max to 2268 (value for high pressure: 45450) switch control unit from high to low (c) Begin operation 6
SUSS MICROTEC MJB4 MASK ALIGNER DOCUMENTATION
SUSS MICROTEC MJB4 MASK ALIGNER DOCUMENTATION ANITHA & KARTHICK 1 Tool specification: Substrate size :1 up to 4 Mask size : Standard 2 x2 up to 5 x5 Alignment :X,Y-±5mm,theta-5⁰ Alignment Accuracy :Top
More informationSTEP-BY-STEP INSTRUCTIONS FOR BUILDING AN AUTOCOLLIMATOR. TECHSPEC Optical Cage System
STEP-BY-STEP INSTRUCTIONS FOR BUILDING AN AUTOCOLLIMATOR TECHSPEC Optical Cage System INTRODUCTION 2 What is an autocollimator? The Autocollimator is a precision instrument composed of an achromatic objective,
More information` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE
` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE Version: 1.0 April-2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research Center (NRC) TABLE OF CONTENTS 1. Introduction...2 1.1
More informationACCURACY, SPEED, RELIABILITY. Turnkey Production for: MEMS. Multi-Chip Modules. Semiconductor Packaging. Microwave Modules.
>> Turnkey Production for: MEMS Multi-Chip Modules Semiconductor Packaging Microwave Modules Flip Chip Photonics Packaging ACCURACY, SPEED, RELIABILITY ULTRA-PRECISION ASSEMBLY WORK CELL Turnkey production
More informationOperating Procedure for Horiba Raman Microscope
Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch
More informationMicroscopic Imaging Research Station (MIRS) Assembly Guide. Version 1.0.0
Microscopic Imaging Research Station (MIRS) Assembly Guide www.adsyscontrols.com Adsys Controls, Inc.2012 Version 1.0.0 I. Assembly of the Adsys Controls MIRS system This document explains the assembly
More informationUser s Guide to the LMD Laser Micro-dissection. System
User s Guide to the LMD-6000 Laser Micro-dissection System Page 1 Glen MacDonald October 31, 2018 Start-up Procedure for Leica LMD-6000. 1. Turn on mercury lamp by pressing the rocker switch; a. beige
More information3 Indexer Installation For PRSalpha Tools
888-680-4466 ShopBotTools.com 3 Indexer Installation For PRSalpha Tools Copyright 2016 ShopBot Tools, Inc. page 1 Copyright 2016 ShopBot Tools, Inc. page 2 Table of Contents General Safety and Precautions...5
More informationFirst-Time Setup. Summary Steps CHAPTER
CHAPTER 4 Revised: May 9, 2012, This chapter describes the steps you take to configure the CTS 1300 to use it for the first time and includes the following sections: Required Tools and Accessories, page
More informationTheodolite Instructions
Theodolite Instructions Elizabeth Petrik ACME Experiment February 19, 2011 A theodolite is a telescope mounted to very sensitive horizontal and vertical protractors. It is capable of measuring angles and,
More informationHDAF800 HD 1080p Digital Microscope USB drive for Image Capture
HDAF800 HD 1080p Digital Microscope USB drive for Image Capture HDAF800-US Auto-Focus HD Camera System 1080p including Universal Stand with 15" Vertical Pole and integrated LED lighting and controls panels.
More informationEquipment Operating Procedure. Model Number: Dage-Series-4000PA Machine Name: Shear Tester Date: 8/17/2013 Serial Number:
Equipment Operating Procedure Manufacturer: Dage Precision Location: PRC Assembly Laboratory Industries Model Number: Dage-Series-4000PA Machine Name: Shear Tester Date: 8/17/2013 Serial Number: 1922708
More informationNov. 07, 2013 p. 5 - changed the B axis unit value to from Changed by Randy per Frank s request.
Correction notes Nov. 07, 2013 p. 5 - changed the B axis unit value to 45.1389 from 40.0000. Changed by Randy per Frank s request. Jan. 22, 2018 p. 5 - changed the B axis unit value and corresponding picture
More informationOperation Manual. Autocollimator Test Wedge for quick testing of Electronic Autocollimators
for quick testing of Electronic Autocollimators For Ident.-No.: 223 244 Version date 05.05.2004 1. Basics The is designed for quick testing of accuracy of electronic autocollimators. If a light beam passes
More information1551 S. Harris Court,. Anaheim, CA Phone: Fax:
1551 S. Harris Court,. Anaheim, CA 92806 Phone: 714.978.1551... Fax: 714.978.0431 Model 4KE / 4KEH Series: Semi-Automatic, Motorized Y/Z Axes, Wedge-Wedge, Ball- Wedge Triple-Convertible Wire Bonders Last
More informationBOTTOM SIDE ALIGNMENTSOP
BOTTOM SIDE ALIGNMENTSOP October 2013 The wafer is aligned to the mask using the bottom side alignment microscope (BSA). 1. Load mask.please refer to Step 1-3 in the previous SOP for loading a mask. 2.
More informationTable of Contents & Introduction 2 Product features & composition. 3 Functions Specifications
High Precision Wireless Bore Gauge B-100 User s Guide The contents of this manual could be different according to the software version and it can be changed without notice. Please use this good after reading
More informationMultiVision Operating Instructions
Innovation. Quality. Performance. Best in Industry. Worldwide. 1 MultiVision Operating Instructions MultiVision offers operators the option in completing various projects to use cameras for setting up
More informationfor R&D and Production Advanced Technology OAI is a Silicon Valley-based manufacturer of advanced precision equipment
Advanced Technology for R&D and Production OAI is a Silicon Valley-based manufacturer of advanced precision equipment for the MEMS, Semiconductor, Nanotechnology, Microfluidics, MEMS Micro TAS and Flat
More informationPortable Staging User Guide
Portable Staging User Guide Components: ExpressDeck was designed for quick and easy assembly: set up is a snap, even if you have no experience! So let s take a moment to meet the components of your new
More informationInstalling 6 Indexer: PRS Standard Tools
888-680-4466 ShopBotTools.com Installing 6 Indexer: PRS Standard Tools Copyright 2016 ShopBot Tools, Inc. page 1 Copyright 2016 ShopBot Tools, Inc. page 2 Table of Contents Overview...5 Installing the
More informationFurther Information can be found at
Below is a step by step guide to assembling the Hurricane-Rig. Remember that this is a precision optical instrument. Excessive force can bend critical parts. If treated well it should give many years of
More informationDV2. Alignment Procedure. Install DV2 on Microscope NOTE: PLEASE READ THE ENTIRE PROCEDURE BEFORE YOU BEGIN ALIGNMENT OF THE DV2. Alignment Procedure
H I G H - P E R F O R M A N C E E M C C D & C C D C A M E R A S F O R L I F E S C I E N C E S DV2 This document provides a straightforward, step-by-step outline of the alignment procedure for the Photometrics
More informationThe Nureva Span ideation system. Installation guide. Single panoramic system
The Nureva Span ideation system Installation guide Single panoramic system Important SAFETY WARNINGS Prior to the installation of this product, the installation instructions should be completely read and
More informationFirearms & Tool Marks Comparison Microscope LEEDS FORENSIC SYSTEMS. Discovery
Firearms & Tool Marks Comparison Microscope LEEDS FORENSIC SYSTEMS Discovery The Discovery, the world s premier firearms and tool marks comparison microscope, features Zeiss optics and provides the largest
More informationSamba Hair System User manual Jefferson Blvd. Culver City, CA 90230, USA Tel: (310) Fax: (310)
Samba Hair System User manual 11922 Jefferson Blvd. Culver City, CA 90230, USA Tel: (310) 577-8110 Fax: (310) 943-3280 http://www.bossanovatech.com Page 1/32 About This Manual This Manual describes SAMBA
More informationClearing paper jams. Identifying paper jam areas. Printing. Using color. Paper handling. Maintenance. Troubleshooting. Administration.
By carefully selecting print media and loading them properly, you should be able to avoid most paper jams. If jams do occur, follow the steps outlined in this section. 1 To resolve the paper jam messages,
More informationWide Guy: Inverted Widefield Microscope
Wide Guy: Inverted Widefield Microscope Kyle Marchuk Adam Fries Jordan Briscoe August 2017 Contents 1 Introduction 2 2 Initial Setup 3 2.1 Hardware Startup........................................... 3
More informationMini Keyboard Controller (CKB-02 series) User Manual
Mini Keyboard Controller (CKB-02 series) User Manual Safety Notes Thank you for choosing our mini keyboard controller! When you open the box: 1. Check that the packing and the contents are not visibly
More informationWritten By: Ben Eisenman
iphone 3GS Rear Panel Replacement Replace a broken rear case on your iphone 3GS. Written By: Ben Eisenman ifixit CC BY-NC-SA www.ifixit.com Page 1 of 22 INTRODUCTION The plastic rear half of the iphone.
More informationStart-up system. Nikon Wide Field Microscope Ee1078. Complete manual
Start-up system Switch on the power socket located on the floor Switch devices on in the order of: o Mercury lamp control-unit switch on and ignite o Halogen lamp o Incubator control-box o Microscope stand
More informationCE 260 SURVEYING CHAPTER 5 THEODOLITES
CE 260 SURVEYING CHAPTER 5 THEODOLITES General Background: Theodolites or Transits are surveying instruments designed to precisely measure horizontal and vertical angles. 1 They are used to establish straight
More informationTitle: Heidelberg DWL66+ Semiconductor & Microsystems Fabrication Laboratory Revision: B Rev Date: 05/03/2017
Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Heidelberg DWL66+. All users are expected to have read and understood this document.
More informationMounting of large lenses in Infrared instruments
Mounting of large lenses in Infrared instruments Jean Louis Lizon, Gotthard Huster European Southern Observatory, Karl-Schwarzschild-Strasse 2, 85748 Garching bei München, Germany ABSTRACT Infrared instruments
More informationGoogle Pixel 2 Back Camera Lens Replacement
Google Pixel 2 Back Camera Lens Replacement How to replace the camera lens in the Google Pixel 2. Written By: Monica Andres ifixit CC BY-NC-SA www.ifixit.com Page 1 of 17 INTRODUCTION The camera is a component
More informationLens Mounting with Elastomeric Bonds Jon Blanchard University of Arizona, OPTI Dec08
Lens Mounting with Elastomeric Bonds Jon Blanchard University of Arizona, OPTI 521 05Dec08 Abstract In this paper, a precision method of lens mounting using elastomeric bonding is described. First a brief
More informationTRANSFORM KIT for 24 x 24 mm Phoenix Camera. TRANSFORM KIT for 24 x 24 mm Phoenix Camera
Phoenix S-mount, ix Connector Phoenix C-mount, RJ45 Phoenix NF-mount, RJ45 TRANSFORM KIT LUCID s Phoenix GigE Vision camera has a unique transformable design that can be easily configured to meet a wide
More informationWritten By: Andrew Bookholt
iphone 3GS Power Button Replacement Use this guide to replace the plastic power button cover on your iphone 3GS. Written By: Andrew Bookholt ifixit CC BY-NC-SA www.ifixit.com Page 1 of 16 INTRODUCTION
More informationLab 5 Microscopy. Introduction. Carrying the Microscope. Depth of Focus. Parts of the Compound Microscope. Magnification
Lab 5 Microscopy Introduction The microscope is an instrument that contains one or more lenses and is used to view objects that are too small be seen with the unaided eye. A magnifying glass is a simple
More informationWritten By: Ben Eisenman
iphone 3GS Battery Replacement Replace an old worn battery in your iphone3gs. Written By: Ben Eisenman ifixit CC BY-NC-SA www.ifixit.com Page 1 of 14 INTRODUCTION [video: http://www.youtube.com/watch?v=wuaht_jrso8]
More informationCut-True 29A & 29H Guillotine Cutters
Cut-True 29A & 29H Guillotine Cutters 11/2016 PROGRAMMING MANUAL First Edition Table of Contents TOPIC PAGE Calibration 1 Manual Mode 1 Absolute Mode 1 Incremental Mode 2 Manual Back Gauge Movement 2
More informationNXQ8000 Series Mask Aligner
NXQ8000 Series Mask Aligner The NXQ8000 Production Mask Aligner and Front to Back Overlay Inspection System integrates the latest in Robotic Automation with state of the art next generation alignment stage
More informationAUET & AUET/MTM Desktop Torque Testers. Quick Start Guide. PO Box 16460, Portland, OR Fax
AUET & AUET/MTM Desktop Torque Testers Quick Start Guide PO Box 16460, Portland, OR 97292-0460 503-254-6600 Fax 503-255-2615 www.aimco-global.com See Product User Guide for: Safety Requirements Step By
More informationOperating instructions for DWL66 laser writer
Operating instructions for DWL66 laser writer Things to remember: 1. The DWL66 runs a 405-nm semiconductor laser. The laser beam path is covered and you must not remove any of the covers. To assure that
More informationSUSS MJB4. Manual Aligner For Research, Development and Operator Assisted Production October, 2009
SUSS MJB4 Manual Aligner For Research, Development and Operator Assisted Production October, 2009 Overview Product Portfolio Aligner MA/BA 8 MA200Compact LithoFab200 MJB4 MA300Plus MA/BA 6 MA150e LithoPack300
More informationProbe station system
Probe station system Manufacturer: Micromanipulator Mode: 450PM-HR Descriptions: The Model 450PM 8 probe station offers stable and reliable probing performance. It is based on a design of Micromanipulator
More informationTEL: / FAX:
Self-leveling Rotary Laser FRE-211 Operating Manual P.R.Engineering Ltd UK www.laser-level.co.uk TEL: 01246 269 777 / FAX: 01246 260 007 SAFETY PRECAUTIONS: During instrument operation, be careful not
More informationNanoBeam nb4 Standard Operating Procedure
NanoBeam nb4 Standard Operating Procedure These instructions are intended for reference only, and will not replace the thorough training required for proper system operation. Contact a clean room staff
More informationReplacing the TFM coupler
Repair manual Replacing the TFM coupler Instructions The TFM coupler is a key component of the hot end. It allows filament to flow smoothly through it from the Bowden tube and towards the nozzle. The coupler
More informationTitle: Karl Suss MA150 Semiconductor & Microsystems Fabrication Laboratory Revision: E Rev Date: 03/01/06
Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Karl Suss MA150 Mask Aligner. All users are expected to have read and understood
More informationPortable Staging User Guide
Portable Staging User Guide Components: Putting together your smart stage system was designed to be simple enough for anyone to do easily. Nonetheless, let s take a moment for a little meet and greet A.)
More informationRemoval and Installation8
8 Screw Types 8-4 Top Cover Assembly 8-5 Left Hand Cover 8-6 Right Hand Cover 8-10 Front Panel Assembly 8-14 Left Rear Cover 8-15 Right Rear Cover 8-16 Extension Cover (60" Model only) 8-17 Media Lever
More informationCondensed AFM operating instructions:
Condensed AFM operating instructions: 1. Log onto system at access controller 2. Take the parts you need to mount a probe out of the drawers. You need the appropriate probe holder, tweezers (these are
More informationSUSS SB6E BONDER SOP. October 2013
SUSS SB6E BONDER SOP October 2013 The SUSS SB6 bonder is capable of silicon fusion, anodic, and eutectic bonding. There are two separate heads; one for anodic bonding, and another for fusion bonding. Definitions
More informationINDEX. Quick Start Guide. Intermediate level guide. Digital Microscope. Displaying a scale. Split-screen display 3
Digital Microscope Quick Start Guide INDEX Displaying a scale Split-screen display 5 6 Using the mouse and operating the motorized XY stage Changing the lens Observing from an angle Additional uses for
More informationFORMTRACER SV-C3100/4100 Series
Form Measurement Surface Roughness/Contour Measuring System FORMTRACER SV-C3100/4100 Series Catalog No. E4304-525 Dual-purpose measurement and powerful analysis of surface roughness and contours combined
More informationDo not place or mount Ocli 1 in an area where it will be exposed to water. Ocli 1 is designed for indoor use.
1 Manual: Ocli 1 with Observeye Viewer App for Android Setting Up Ocli 1 Live Viewing with Ocli 1 Recording Video and Taking Snapshots Using Advanced Features Setting Up Ocli 1 Place or Mount the Camera
More informationHighly Efficient Assembly of Lenses with. OptiCentric and OptiCentric Cementing
Highly Efficient Assembly of Lenses with OptiCentric and OptiCentric Cementing Centration Measurement and Lens Alignment Contents OptiCentric Cementing Centration Measurement and Lens Alignment 3 4 Centration
More informationiphone 3G Headphone Jack Replacement Replace a broken audio port in an iphone 3G. Written By: irobot ifixit CC BY-NC-SA
iphone 3G Headphone Jack Replacement Replace a broken audio port in an iphone 3G. Written By: irobot ifixit CC BY-NC-SA www.ifixit.com Page 1 of 18 INTRODUCTION No audio? Replace the headphone jack! TOOLS:
More informationWELCOME. For customer support or any inquiries, please visit our web site at or contact us at
WELCOME Congratulations on purchasing the GBF Smart Four Wire Intercom System. Our factory engineers were the first to enable multiple security cameras being monitored through a smart mobile device and
More informationTotal Station & Distomat
UNIT 6 Total Station & Distomat Learning Objectives After studying this unit, the student will be able to Understand the parts of Total station Temporary adjustments of total station Measurement of horizontal
More informationForm Measurement ROUNDTEST RA-1600
Form Measurement ROUNDTEST RA-1600 A new PC-compliant Roundness and Cylindricity form measuring instrument with extensive analysis features to enable measurement of a wide variety of workpieces. Powerful
More informationHow to Transport a Path Program Release: How to Transport a Path Program-C PVA Support Hub. How to Transport a Path Program
PVA Support Hub How to Transport a Path Program This guide shows how to transport a path program to a new workcell. This procedure should be used with the PathMaster and Workcell Manuals and is not intended
More informationCylinders in Vs An optomechanical methodology Yuming Shen Tutorial for Opti521 November, 2006
Cylinders in Vs An optomechanical methodology Yuming Shen Tutorial for Opti521 November, 2006 Introduction For rotationally symmetric optical components, a convenient optomechanical approach which is usually
More informationLaser Alignment. For the Epilog Fusion CO2 10/10/2013
EPILOG LASER 16371 Table Mountain Parkway Golden, Colorado 80403 Phone 303-215-9171 FAX 303-277-9669 www.epiloglaser.com For the Epilog 10/10/2013 Requirements: Tools: 5/32 Allan Wrench 3/32 Allan Wrench
More informationOMM200 and OMM200T. Metallurgical Microscope
OMM200 and OMM200T Metallurgical Microscope Instruction Manual (Please read the manual carefully before using the microscope) Contents Safety ----------------------------------------------------------------------------
More informationWarranty 3 Auditory feedback and headphones 11. General care 3 Access options Grid Pad Go 4-7 Shoulder strap 14
Grid Pad Start the conversation Warranty 3 Auditory feedback and headphones 11 General care 3 Access options 12-13 Grid Pad Go 4-7 Shoulder strap 14 Grid Pad Go 8 and Grid Pad 11 overview 4-6 Flip out
More informationIoline CrystalPress. Quick Start Guide
Ioline CrystalPress Quick Start Guide The Quick Start Guide is intended to help a new user of the Ioline CrystalPress get everything setup and running quickly. Please note that there are important notices,
More informationGP-B Telescope Image Divider Assembly (IDA) Position Left and Right Relay Lenses on IDA Channel B P0294 Rev -
Stanford University W.W. Hansen Experimental Physics Laboratory Gravity Probe B Relativity Mission Stanford, California 94305-4085 SUGP-B P0294 Rev - GP-B Telescope Image Divider Assembly (IDA) Position
More informationVisitron Spinning Disk Unit 1
Startup Procedure and Notes (Last update: Sept. 2014) 6 4 2 5 1 3 7 8 9 1. Switch on the fluorescence illumination for widefield viewing (HXP lamp) 2. Switch on power-bar on the left side of the shelf
More informationOlympus cellsens Live/TIRF Unit
Version 4 (July 2017) Startup (Widefield Mode) Switch on power strips #1-4 (behind the monitor/lasers) in the order as they are labeled. Switch on #2 (TIRF lasers) only if you need one or more lasers for
More informationMotorized ZDeck Quick Adjust Platform System
For Nikon FN1/Ni, Olympus BX51WI/BX61WI and Zeiss Axio Examiner Microscopes Zero current for electrophysiology applications Large top plate surface for mounting micromanipulators, 6mm holes on 25mm centers
More informationMAIN OPERATING INSTRUCTIONS
MAIN OPERATING INSTRUCTIONS VPS-1 Pedestal Including Dolly options & accessories 1. INTRODUCTION You have bought a Pedestal out of the expanding new Cambo range of Video products. We expect that you will
More information5. Office furniture. 5. Office furniture. Laboratory furniture. Flexible and customised
5. Office Flexible and customised 5. Office Laboratory Your laboratory staff needs a thoughtfully planned office or work area and we have just the you have been looking for. Functional and customised to
More informationAppendix C. Vernier Tutorial
C-1. Vernier Tutorial Introduction: In this lab course, you will collect, analyze and interpret data. The purpose of this tutorial is to teach you how to use the Vernier System to collect and transfer
More informationMPI TS150-THZ 150 mm Manual Probe System
MPI TS150-THZ 150 mm Manual Probe System Industry s first explicitly designed probe system for accurate measurements at mm-wave and sub-mm wave (THz) frequency range FEATURES / BENEFITS Variety of Applications
More informationAdvances in Metrology for Guide Plate Analysis
Advances in Metrology for Guide Plate Analysis Oxford Lasers Ltd Overview Context and motivation Latest advances: Automatic entrance hole measurement Hole shape analysis Debris detection File format We
More informationSetup Information Panosaurus May 3, 2011
Setup Information Panosaurus 2.0 www.gregwired.com May 3, 2011 Please take the time to read all of the setup information to ensure success and ease of use of this tripod head. Much of the setup is a one
More informationSuss Aligner SOP. Suss Aligner SOP Page 1 of 6 Revision Scope 1.1 This document provides the procedure for operating the Suss Aligner.
Suss Aligner SOP Page 1 of 6 Suss Aligner SOP 1. Scope 1.1 This document provides the procedure for operating the Suss Aligner. 2. Table of Contents 1. Scope... 1 2. Table of Contents... 1 Suss Aligner
More informationPanowin F1. User Manual
Panowin F1 User Manual 1 PANOWIN TECHNOLOGIES CO.,LTD. WARNING power outlet. CAUTION: In case of emergency unplug the Panowin F1 from the WARNING: Carefully monitor the Panowin F1 during operation. Do
More informationBLIPS perfectly aligned cleaned by any impurity Gently press the lens
BLIPS - TIPS & TRICKS BLIPS, the thinnest Macro and Micro lenses for smartphones in the world, can guarantee an excellent picture quality for photos and videos, but they need some simple tricks to get
More information3ds Max Cottage Step 1. Always start out by setting up units: We re going with this setup as we will round everything off to one inch.
3ds Max Cottage Step 1 Always start out by setting up units: We re going with this setup as we will round everything off to one inch. File/Import the CAD drawing Be sure Files of Type is set to all formats
More informationHARP-NEF Front End Assembly
Science Fibers HARP-NEF Front End Assembly Calib Fibers HARPN Front End Assembly ( w/ Partial Cover Plates ) HARPN Front End Assembly 1 HARP-NEF Front End Assembly Roger Eng December 6, 2007 2 HARP-NEF
More informationModel 90. thermwood. Thermwood CNC Machining Centers. Machine Features Shown
22017 MADE IN USA 5 Axis Model 90 Thermwood, the industry leader in 5 Axis applications with more five-axis systems in operation than any other company. These 5 Axis CNC routers have become the accepted
More informationDanalock V3 SCANDI Mounting guide.
Danalock V3 SCANDI Mounting guide www.danalock.com Table of contents Tools needed page 4 Parts involved page 5 1. Secure the front lock page 6 2. Remove the inside thumb-turn page 7 3. Normal tailpiece
More informationIEEM 215. Manufacturing Processes I Introduction to the ARIX CNC milling machine
IEEM 215. Manufacturing Processes I Introduction to the ARIX CNC milling machine The image below is our ARIX Milling machine. The machine is controlled by the controller. The control panel has several
More informationMPI TS150-AIT 150 mm Manual Probe System
MPI TS150-AIT 150 mm Manual Probe System Industry s first explicitly designed 150 mm probe system providing accurate tests for mm-wave, THz, and automated impedance tuner applications FEATURES / BENEFITS
More informationFRDM Replacement Procedure
Operating Documentation Direction 5193574-800, Rev# 21 LightSpeed 7.X Service Methods Module Version 11.0, Version Date 3/20/14 FRDM Replacement Procedure (Click Here for Summary Version) 1 Overview This
More informationINSTALLATION AND OPERATIONAL INSTRUCTIONS FOR THE OFITE MODEL 1000 VERSION 1.0
OFI Testing Equipment Model 1000 HPHT Viscometer (130-75) Instructions Page 1 of 7 INSTALLATION AND OPERATIONAL INSTRUCTIONS FOR THE OFITE MODEL 1000 VERSION 1.0 INSTALL DIRECTIONS FOR OFITE SOFTWARE.
More informationPrism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis
Start Up: Upon entering the laser room turn on the wall mounted Laser Power Button by pulling it away from the wall. Turn on Shutter controllers (toggle switch on back of unit). There should be a U in
More information,M,R,!iilll""M,:3 MEMS. ~ IliJctanill:5 Pac~ RSI. YSTEMS
,M,R,!iilll""M,:3 U!i!.,TIRAw.'";P,RElCISIONI,F'LEXIBLE W,I!:IIR:I;C,ICELL, MEMS ~ IliJctanill:5 Pac~.... RSI. YSTEMS ULTRA-PRECISION flexible ASSEMBLY WORK CELL Turnkey production for: MEMS Multi-Chip
More informationTable of Contents. iii
Thermal Analysis Table of Contents Performing Thermal Analyses... 1 Overview... 1 Simplifying Assumptions... 2 Approach... 2 Common Analysis Parameters... 2 Specialized Parameters... 3 Modifying Set Up
More informationSteps to Print (without supports) This will take you from Meshmixer to the PSU PrintDepot
Steps to Print (without supports) This will take you from Meshmixer to the PSU PrintDepot Read this first to understand some of the dilemmas a model can have and ways to avoid printing problems: http://www.3dshook.com/2016/05/how-to-optimize-a-3d-model-for-3dprinting/#normal
More informationpsee Control Software
psee Control Software Installing psee Software psee is not currently available on the Web and can only be downloaded by submitting a request via e-mail. To obtain the psee URL, send an e-mail to RVision
More informationFor documentation the instrument is available with a trinocular head which permits mounting of a video or digital camera system.
1. Introduction The PriorLux MET upright compound metallurgical microscope is a high quality instrument equipped with high resolution, chromatically corrected optics for excellent image quality in reflected
More informationFive-Axis Fiber Aligner Model 9091 and Accessories
USER S GUIDE Five-Axis Fiber Aligner Model 9091 and Accessories Includes Model 9131-FS-FC and Model 9131-FS-ST Fiber-Aligner Kits U.S. Patents: #5,282,393 & #5,400,674 5215 Hellyer Ave. San Jose, CA 95138-1001
More information1: 4 Zoom Ratio Stereomicroscope
1: 4 Zoom Ratio Stereomicroscope SMZ-140 SCHEMATIC DIAGRAMS/SPECIFICATIONS SMZ-140/143 SCHEMATIC DIAGRAMS SMZ-140-N2GG Binocular Stereomicroscope SMZ-143-N2GG Trinocular Stereomicroscope Unit: mm Unit:
More informationX-series C-series. english
X-series C-series english The X-series The Euromex X-series microscopes are perfect for use at schools and in laboratories. The nice, ergonomically designed stand is very comfortable in use. All movable
More informationToucan LT board printer
Toucan LT board printer Setup and Operating instructions Unpack the Toucan LT board printer as you would any Toucan LT. Follow all cautions associated with installing a standard Toucan LT. Change in wash
More informationOXY 4 Air Sterilizer. Ultra Violet Advanced Air Treatment System
TM OXY 4 Air Sterilizer Ultra Violet Advanced Air Treatment System EXTENDED SURFACE AREA WITH CATALYTIC REACTIONS FOR ADVANCED PLASMA OXIDIZERS USB PROGRAMMING PORT OXY 4 LED SMART LIGHT MONITOR PROGRAMMABLE
More information