ST-565 Inline HST Fluorometer Probe Instruction Manual
|
|
- Lynne Berry
- 5 years ago
- Views:
Transcription
1 Pyxis Lab Inc. ST-565 Inline HST Fluorometer Probe Instruction Manual Version 1.1
2 Contents 1. INTRODUCTION UN-PACKAGING THE INSTRUMENT STANDARD ACCESSORIES OPTIONAL ACCESSORIES SPECIFICATION INSTALLATION CONNECT TO A CONTROLLER CONNECTION TO A COMPUTER PROBE CALIBRATION DOWNLOAD AND INSTALL PROBE CONFIGURATOR CALIBRATION VIA PROBE CONFIGURATOR Zero Point Calibration Slope Calibration ma Output Setup CALIBRATION ON THE CONTROLLER MODBUS TRU MAINTENANCE AND TROUBLESHOOTING GENERAL MAINTENANCE
3 1. Introduction The Pyxis ST-565 inline fluorometer probe measures the concentration of tolyltriazole (TTA) and halogen stable tolyltriazole (HST) in cooling tower water. The ST-565 probe excites the sample water with a deep UV light and measures a narrow band near UV fluorescent emission. It can be connected to any controller that accepts an isolated or non-isolated 4-20mA input. The ST-565 probe is a smart device. In addition to measuring fluorescence, the ST-565 probe has extra photoelectric components that monitor the color and turbidity of the sample water. This extra feature allows automatic color and turbidity compensation to eliminate interferences common in real-world samples. Other features of the ST-565 probe include: Menu-driven calibration can be carried out by a computer, a smart phone, or other Pyxis handheld devices via a USB or Wifi/Bluetooth connection. Diagnostic information (probe fouling, color or turbidity over range, failure modes) are available during calibrations. The water sample does not need to be acidified. Sample ph between 6.4 to 10.0 and the sample PTSA concentration (<300 ppb) do not affect the azole measurement. 2. Un-packaging the Instrument Remove the instrument and accessories from the shipping container and inspect each item for any damage that may have occurred during shipping. Verify that all accessory items are included. If any item is missing or damaged, please contact Pyxis Lab Customer Service at service@pyxis-lab.com Standard Accessories Bulkhead Extension Cable Instrument Manual, also available from Optional Accessories USB-RS485 Adapter (Item Number: MA-485) Wifi/Bluetooth Adapter (Item Number: MA-WB) Probe Configurator Software CD (Item Number: MA-120, free download from 2 ppm TTA Calibration Standard Solutions (Item Number: TTA-002) 3
4 3. Specification Items Specification HST 0-5 ppm, ±0.2 ppm precision Power Supply 24 VDC, ~1W Output 4~20 ma isolated, RS-485 isolated Connector IP67 water proof connector/cable 1.5 m (4.5 ft) Operation Pressure Up to 1.0 MPa (100 PSI) at 45 C (113 F) Installation ¾-inch NPT Storage temperature -20 C ~ 60 C (-4~140 F) Operational temperature 0 C ~ 40 C (32~104 F) Sample Water Flow 0 to 8 GPM Typical service life 2 years Housing material CPVC Dimension Shown in figure 1 in mm Weight 0.5 lbs 4. Installation The ST-565 probe is shipped in a PVC tee. The tee has two ¾ inch NPT ports (sample in and out) and can be fluidically connected to a sample panel. The orientation of the flow with respect to the gravity of earth is not critical. The tee can be vertically or horizontally installed in the sample pipe system. 4
5 Figure 1. Probe and Tee with dimensions in mm Figure 2. Probe and Tee set 5
6 4.1. Connect to a controller Follow the wiring table below to connect the ST-565 probe to a controller. Note that the 24V power ground and the negative 4-20 ma terminal in the ST-565 probe are internally connected. If a separate DC power supplier other than that from the controller is used, make sure that the output from the power supplier is within 22 to 26 VDC. Wire Color Designation Red 24 V + Black Power Ground White 4-20 ma + Green 4-20 ma -, internally connected to power ground Blue RS-485 A Yellow RS-485 B Clear Shield, solution ground (optional) 4.2. Connection to a Computer The ST-565 probe can be connected to a PC via the Pyxis USB-RS485 adapter (Item Number: MA- 485) or the Pyxis Wifi/Bluetooth adapter (Item: MA-485BW). The adapter has the same male and female bulkhead cable connectors as the probe and the probe extension cable. To connect the probe to a PC, plug the USB cable to a USB port in the PC and connect the adapter s connectors to the corresponding connectors from the probe and the probe extension cable. The probe extension cable must be connected to a 24V DC power source to power the probe and the adapter. 5. Probe Calibration 5.1. Download and Install Probe Configurator Download the Probe Configurator software package from Double click Probe Configurator Setup.msi file to install. Two Probe Configurator icons will be placed on the computer desktop for the USB connection and Wifi/Bluetooth connection respectively Calibration via Probe Configurator The probe is calibrated before shipping. It outputs 4 ma for zero ppm TTA/HST, 12 ma for 2.5 ppm TTA/HST, and 20 ma for 5 ppm. The probe can be recalibrated using the Probe Configurator Software. 6
7 Connect the probe to a Windows computer using the USB-RS485 adapter (Item: MA-485) or the Wifi/Bluetooth adapter (Item: MA-WB). Double click the corresponding Probe Configurator icon to launch the program. If the probe is connected to the computer properly, the Probe Configurator software will detect the com port number and set up the communication parameters automatically. Click Devices tab to select ST565-HST as shown in figure 3. Click Connect button to connect the program to the probe. If the connection is successful, the probe system parameters will be filled as shown in the lower left in figure 4. Figure 3. Select Probe type ST565-HST 7
8 Figure 4. Probe being connected To connect the probe to the Probe Configurator software via the Wifi/Bluetooth adapter, please connect the Wifi/Bluetooth adapter between the probe and probe extension cable connectors. Click Probe Configurator Wifi icon to launch the program. Click Search to find the Wifi/Bluetooth adapter. Click Connect to connect Probe Configurator to the probe (figure 5). 8
9 Figure 5. Wifi Connection Zero Point Calibration Put the probe into deionized water. Click Zero Cal button to carry out the zero point calibration. If calibration is successful, the Calibration Successful message will be prompted Slope Calibration Put the probe into a solution with known TTA/HST concentration. Click Slope Cal button to carry out the slope calibration. The TTA/HST concentration in the calibration solution should be in the range of 1 to 5 ppm, ideally 2 to 4 ppm. If calibration is successful, the Calibration Successful message will be prompted. If errors are determined, an error message indicating the cause of the errors will be shown. 9
10 Figure 6. Slope calibration The zero point calibration and slope calibration can be carried out separately. It is recommended that the slope calibration be carried out after the zero point calibration. Direct sunlight or indoor light on the probe should be avoided although it is not necessary to completely shield the probe from the ambient light during both the zero point and slope calibrations ma Output Setup The default setup is 20 ma output for 5 ppm TTA/HST. Click on ma2ppb button to reset the current-to-ppm scale can be configured Calibration on the Controller If the probe is set to output 20 ma at 5 ppm, the controller should be set up to convert 20 ma to 5 ppm and 4 ma to 0 ppb correspondingly. Optionally, the probe can be calibrated by changing the conversion factor on the controller without being undergone the zero point and slope calibration steps. Please follow the controller generic 4-20mA analog input calibration procedure to calibrate the probe. The advantage of the calibration by re-scaling the 4-20mA input to a ppm value is a quick alternative. The TTA/HST ppb concentration in the water sample can be determined by using an offline fluorometer or other methods such as the treatment product concentration. One can force the controller to show the determined TTA/HST concentration without the need to calibrate the probe using a standard solution. The disadvantage is that this method is only valid in a small range of TTA/HST concentration. Potential probe operation problems may not be able identified as they would in the zero and slope calibration procedures. 10
11 5.4. Modbus TRU The ST-565 probe is a Modbus slave device. In addition to the ppm TTA/HST value, a range of operational parameters including warning and error message are available via Modbus TRU connection. Contact Pyxis Lab Customer Service (service@pyxis-lab.com) for more information. 6. Maintenance and Troubleshooting 6.1. General Maintenance The flow channel of the probe should be cleaned once a month using a soft brush. Deposit and debris should be removed. The probe cleanliness can be checked using the diagnosis function of Probe Configurator as show in figure 7. When contact Pyxis technical support, please save the diagnosis information to a file using Data Export and the file to service@pyxis-lab.com. The probe can be cleaned with mild acid solution such as 1% citric acid. Do not use organic solvents or strong acid/base to clean the probe. Carry out routine calibration check against a TTA standard. If necessary, carry out the zero point and slope calibration. Avoid long term storage at temperature over 100 F. In an outdoor installation, properly shield the probe from direct sunlight and precipitation. Figure 7. Diagnosis and cleanliness check 11
12 Pyxis Lab, Inc. 120 Jeffrey Ave Holliston, MA USA T service@pyxis-lab.com 12
Version 1.4 June 2018
1 Version 1.4 June 2018 Table Of Contents 1. Introduction 3 2. Unpacking the Instrument 4 2.1. Standard Accessories 2.2. Optional Accessories 3. Specifications 4 4. Installation 5 4.1. Quick 4-20 ma Start
More informationST-601 Inline ClO2 Concentration Monitoring Probe Instruction Manual
Pyxis Lab, Inc. ST-601 Inline ClO2 Concentration Monitoring Probe Instruction Manual Version 1.1 Pyxis Lab, Inc. 120 Jeffrey Ave Holliston, MA 01746 USA www.pyxis-lab.com service@pyxis-lab.com 1. Introduction
More informationSP-400 Portable Water Analyzer Operation Manual. Rev. A Firmware version 1.02
SP-400 Portable Water Analyzer Operation Manual Rev. A Firmware version 1.02 Pyxis Lab, Inc 120 Jeffrey Ave Holliston, MA 01746 USA www.pyxis-lab.com service@pyxis-lab.com 1 Contents 1 General Description...
More informationLAKEWOOD INSTRUMENTS MODEL 1575e WATER TREATMENT CONTROLLER
LAKEWOOD INSTRUMENTS MODEL 1575e WATER TREATMENT CONTROLLER The Model 1575e uses the latest in microprocessor capability, giving the user a high level of application flexibility. A large illuminated graphics
More informationLittle Dipper 2 In-Line Fluorometer
Little Dipper 2 In-Line Fluorometer October 17, 2014 P/N 998-2821 Version A TURNER DESIGNS 845 W. Maude Avenue Sunnyvale, CA 94085 Phone: (408) 749-0994 FAX: (408) 749-0998 Table of Contents 1. Introduction
More informationKOBOLD TDA Series Digital Temperature Transmitter
KOBOLD TDA Series Digital Temperature Transmitter TDA-15 Series User Instructions KOBOLD Instruments Inc. 1801 Parkway View Drive Pittsburgh, PA 15205 Phone (412) 788-2830 Fax (412)-788-4890 www.koboldusa.com
More informationINSTRUCTION MANUAL HBPH DIFFERENTIAL 2-WIRE (4-20 ma) ph Sensor
INSTRUCTION MANUAL HBPH DIFFERENTIAL 2-WIRE (4-20 ma) ph Sensor TABLE OF CONTENTS 1. GENERAL INFORMATION 2 2. SPECIFICATIONS 2 3. INSTALLATION 3 3.1. General Instructions 3 3.2. HBPH 4-20 ma Sensor 3 4.
More informationph/orp 300 (Cat. No ) MICROCOMPUTER BASED OPERATION MANUAL ph /ORP/ Temperature CONTROLLER Phone: , Fax:
OPERATION MANUAL /ORP 300 (Cat. No. 0580212) MICROCOMPUTER BASED /ORP/ Temperature CONTROLLER 625 East Bunker CT, Vernon Hills, IL 60061-1844 Phone: 800-32-4340, Fax: 847-247-2929 E-mail: info@coleparmer.com,
More informationGeneral Purpose ph/orp Sensor
Instruction Sheet PN 51A-399VP/rev.J January 2011 Model 399VP General Purpose ph/orp Sensor SPECIFICATIONS Sensor Type: General purpose 399VP Measured Range: ph: ACCUGLASS 0-14 ORP: -1500 to +1500 mv Percent
More informationthe Interactive Catalog
Interactive Catalog Supplements Catalog PDFs If you need detailed product information, or help choosing the right product for your application, see our Interactive Catalog Use the Interactive Catalog to
More informationEngineering Note: EN0047 Commissioning a Hydro-Probe with a Command Alkon EZCal Station
Engineering Note: EN0047 Commissioning a Hydro-Probe with a Command Alkon EZCal Station Summary: Products affected: Commissioning a Hydro-Probe with a Command Alkon EZCal Manual Station Hydro-Probe Models
More informationKOBOLD PDA Series Digital Pressure Transmitter
KOBOLD PDA Series Digital Pressure Transmitter PDA-153 Series User Instructions KOBOLD Instruments Inc. 1801 Parkway View Drive Pittsburgh, PA 15205 Phone (412) 788-2830 Fax (412)-788-4890 www.koboldusa.com
More informationOperation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller
Operation Manual TX10 1/8 DIN Microcomputer Based /ORP Controller 11751 Markon Drive Garden Grove, CA. 92841 U.S.A. pg. 1 CONTENTS GENERAL INTRODUCTION.....3 INITIAL INSPECTION......3 USING THE Sensorex
More informationDescription, continued
DL424/425 DirectLine Sensor Module for DL5000 Dissolved Oxygen Probes Overview 70-82-03-48 4/02 Page of 8 Specification The DirectLine DL424/425 Sensor Modules are part of a family of products developed
More informationOperation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter
Operation Manual MODEL 2TX 2-wire Isolated ph/orp Transmitter 0 2TX CONTENTS INITIAL INSPECTION.....2 INTRODUCTION......2 ASSEMBLY...3 PREPARATION....4 CONNECTING THE ELECTRODE...4 CONNECTING THE TEMPERATURE
More informationEasy Flow Indicator Magnetisch-Induktiver Durchflussmesser. Type 8024 For Insertion Fittings. Analog Flow Rate Indicator, 1/2"-16"; P=87 PSI
Analog Flow Rate Indicator, 1/2"-16"; P=87 PSI Advantages / Benefits Reduced costs of ownership up to 6% Economic integration in pipe systems without any additional piping Shows flow rate with a clear,
More informationINSTALLATION MANUAL. ST-BT1000IR2812-G Varifocal IR Bullet Camera. v1.0 8/4/14 1
INSTALLATION MANUAL ST-BT1000IR2812-G Varifocal IR Bullet Camera v1.0 8/4/14 1 This package contains: PACKAGE CONTENTS One ST-BT1000IR2812-G varifocal infrared bullet color camera Mounting Hardware One
More informationUser s Manual. June 30, 2014 P/N Revision 2.4
User s Manual June 0, 0 P/N 998-00 Revision. TURNER DESIGNS 8 W. Maude Avenue Sunnyvale, CA 908 Phone: (08) 79-099 FAX: (08) 79-0998 Table of Contents. Introduction. Description. Inspection and Setup.
More informationUser s Manual. May 31, 2017 P/N Revision 2.7. TURNER DESIGNS 1995 N. 1 st Street San Jose, CA Phone: (408) FAX: (408)
User s Manual May, 07 P/N 998-00 Revision.7 TURNER DESIGNS 99 N. st Street San Jose, CA 9 Phone: (08) 79-099 FAX: (08) 79-0998 Table of Contents. Introduction. Description. Inspection and Setup. Instrument
More informationMODEL 245 / 345 High Range
MODEL 245 / 345 High Range PRESSURE TRANSMITTER INSTALLATION DATA MANUAL 199 Fire Tower Drive Tonawanda, NY 14150 Toll Free: 1-800-688-0030 International: 716-629-3800 Fax: 716-693-9162 www.viatran.com
More informationVPINSTRUMENTS.COM VPFLOWSCOPE M. Your next step in gas flow measurement
VPINSTRUMENTS.COM VPFLOWSCOPE M Your next step in gas flow measurement The Internet of Things / industry 4.0 is rapidly transforming our world. Production becomes smart, intelligent and autonomous. Production
More informationB&W RearView Camera Installation & Operation
B&W RearView Camera Installation & Operation CA52 (Camera) FOR MORE INFORMATION WWW.STRATEGICVISTA.COM BEFORE OPERATING THIS SYSTEM, PLEASE READ THIS MANUAL THOROUGHLY AND RETAIN IT FOR FUTURE REFERENCE
More informationCole-Parmer Instrument Company
Operating Instructions Cole-Parmer ph/orp 400 ph/orp Controller Cole-Parmer Instrument Company 625 East Bunker Court, Vernon Hills, Illinois 60061 Tel: 1-847-549-7600 or Toll-free: 1-800-323-4340 Fax:
More informationInstruction Manual CTC-1
Troubleshooting & Specs CTC-1 Amps / Volts requirements Min / Max operating temperature Min / Max operating Humidity Temperature Measurement range Temperature Accuracy CO2 sensor type CO2 Measurement range
More informationCole-Parmer Instrument Company
Operating Instructions Cole-Parmer ph/orp 350 ph/orp Transmitter Cole-Parmer Instrument Company 625 East Bunker Court, Vernon Hills, Illinois 60061 Tel: 1-847-549-7600 or Toll-free: 1-800-323-4340 Fax:
More informationTEMPERATURE - HUMIDITY ELEKTRONIK EE08
TEMPERATURE - HUMIDITY ELEKTRONIK EE08 SENSOR MEASURE SIGNAL OUTPUT ELECTRICAL SUPPLY MODEL IN EOL MANAGER Temperature 0...1 V 12...24 V EE8 / ROTRONIC Humidity 0...1 V 12 24 V EE8 / ROTRONIC APPLICATION
More informationOPERATION MANUAL MODEL TT-LP
TURTLE TOUGH OPERATION MANUAL MODEL TT-LP 2-Wire Isolated ph/orp Transmitter turtletough.com.au CONTENT INITIAL INSPECTION 3 INTRODUCTION 3 ASSEMBLY 4 PREPARATION 5 Connection Diagram of Turtle Tough Sensors
More informationPDW30 Point-To-Point Wireless Bridge
PDW30 Point-To-Point Wireless Bridge WIRELESS WIreless System FEATURES Wire Replacement in a Rugged, Industrial Housing Inputs and Outputs on Both Field Units Simple to Order, Configure, & Install 1 Mile
More informationST-HD-TD W Varifocal IR Color HD-CVI Camera
ST-HD-TD2812-720-W Varifocal IR Color HD-CVI Camera V1.0 04/24/14 1 PACKAGE CONTENTS This package contains: One (1) ST-HD-TD2812-720-W varifocal infrared turret-dome color HD camera One (1) installation
More informationDistributor: Topac Inc 231 CJC Highway Cohasset MA Tel Fax
Distributor: Topac Inc 231 CJC Highway Cohasset MA 02025 Tel 781 740 8778 Fax 781 740 8779 www.topac.com email sales@topac.com 1. Introduction Congratulations! You have just purchased a quality Sentron
More informationSignet Conductivity Electrodes
Signet 2839-2842 Conductivity Electrodes Features ± 1% accuracy - Custom calibration certificate provided Dual-threaded Compact electrode length for easy in-line installation in small pipe sizes Triple
More informationHTZ. Description. Features. Smart HART Humidity and Temperature Transmitter
March 2018 Description The HTZ Smart HART Transmitter simultaneously measures humidity and temperature in industrial, commercial, manufacturing, and HVAC applications. Based on the humidity and temperature
More informationINSTRUCTION MANUAL. AquiStar GDL. Dissolved Oxygen Datalogger
INSTRUCTION MANUAL AquiStar GDL Dissolved Oxygen Datalogger Table of Contents 1 Introduction... 3 What is a GDL?... 3 Initial Inspection and Handling... 3 Do s and Don ts... 3 Getting Started... 4 GDL
More informationHigh Voltage DC Meter
High Voltage DC Meter Javelin D PD644 0-300 VDC input NEMA 4X, IP65 front Scale in engineering units Sunlight readable LED display 4-20 ma analog output Two form C 3 A relays option RS-485 serial communications
More informationECL150/ECL150e. Eddy-Current Displacement Sensor INSTRUCTION MANUAL. Measurement Systems from LION PRECISION
INSTRUCTION MANUAL ECL150/ECL150e Eddy-Current Displacement Sensor Measurement Systems from LION PRECISION Lion Precision 563 Shoreview Park Road St. Paul, Minnesota 55126-7014 Telephone: 651-484-6544
More informationRADIOMETER PHOTOMETER L-420 INSTRUCTION MANUAL
RADIOMETER PHOTOMETER L-420 INSTRUCTION MANUAL PPUH SONOPAN Sp. z o.o. ul. Ciołkowskiego 2/2 15-950 Białystok POLAND tel./fax: +48 85 742 36 62 http://www.sonopan.com.pl August 2012 CONTENTS 1. DESCRIPTION...
More informationResolution selectable from 0.01 and ph. Range ph. Accuracy ± ph
ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand manual log on stability Interval logging Temperature readout ( C or
More informationMP 101A HUMIDITY TEMPERATURE PROBE INSTRUCTION MANUAL
MP 101A HUMIDITY TEMPERATURE PROBE INSTRUCTION MANUAL -2- CONTENTS DESCRIPTION...3 OPERATION...3 INSTALLATION...4 MAINTENANCE...7 CALIBRATION BASICS...7 CALIBRATION PROCEDURE...8 SPECIFICATIONS...11 PLEASE,
More informationDigital pressure gauge Data Sheet: DMM2.431.R1.EN
Data Sheet: DMM.431.R1.EN www.aep.it Digital pressure gauge DMM is a digital pressure gauge made according to the most modern technologies to ensure a high level of reliability, versatility and practicality
More informationScanner 2000 microefm QuickStart. Installing the Scanner Remote Mount. Direct Mount NUFLO. Part No , Rev. A
NUFLO Part No. 30165024, Rev. A Scanner 2000 microefm QuickStart Installing the Scanner 2000 H L H L Flow Direct Mount To install the Scanner 2000 microefm using a direct mount to an orifice or cone meter
More informationMMT330 Series Moisture and Temperature Transmitters for Oil
MMT330 Series Moisture and Temperature Transmitters for Oil Features Continuous online measurement of moisture in oil Ball-valve installation no need to shut down the process or drain the oil Proven Vaisala
More informationP522 Signal Processor
P522 Signal Processor INSTALLATI INSTRUCTIS Self-Check Contact Ratings: Max switching power - 60 W 125Vac Max switching voltage - 220VDC, 250Vac Max switching current - 2A DC, AC Analog Flame Signal: User
More informationThermo Scientific Orion Star A214 Benchtop ph/ise Meter. Instruction Sheet
English ph / ISE Thermo Scientific Orion Star A214 Benchtop ph/ise Meter Instruction Sheet Preparation Power Source 1. Power adapter (included with meter) a. Select the appropriate wall socket plug plate.
More informationUniversal IV Water Cut Meter
DREXELBROOK DATA SHEET Universal IV Water Cut Meter with Density Compensation Use the Best For over 50 years, Drexelbrook has been the world s leader in capacitive based measurements by providing reliable
More informationMULTI PARAMETER WATER QUALITY SENSORS
MULTI PARAMETER WATER QUALITY SENSORS IRRIGATION AUTOMATION WATER MEASUREMENT REMOTE MONITORING CLOUD BASED DATA PENTAIR ENVIRONMENTAL SYSTEMS 47.0 MP47 (UP TO 4 PARAMETERS) MP65 (UP TO 6 PARAMETERS) MULTI
More informationManual of Electrical Installation for Pressure Probes KS and KS-SIL2 series
Manual of Electrical Installation for Pressure Probes KS and KS-SIL2 series GEFRAN SAFETY GUARANTEE Code 8594A Edition 2/204 Contents. General precautions pag 2 2. Transmitters with amplified analog output
More informationHygroClip M22 Series. 2-wire loop powered humidity temperature transmitters INSTRUCTION MANUAL
HygroClip M22 Series 2-wire loop powered humidity temperature transmitters INSTRUCTION MANUAL 20060329 CONTENTS Overview... 3 Operation... 5 Power supply... 5 HygroClip S digital probe... 5 Temperature
More informationContents. Coat-Endure User Manual & Datasheet V 2.3
1 Contents Revision History................................................... 4 1 Introduction.................................................... 5 2 Operating Principle................................................
More informationSIGNET ph/orp Transmitter
SIGNET 8750- ph/orp Transmitter ENGLISH 3-8750.090- D-4/03 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.
More informationJUMO ecotrans ph 03 Microprocessor transmitter/ switching device for ph/redox voltage and temperature
Page 1/7 JUMO ecotrans 03 Microprocessor transmitter/ switching device for /Redox voltage and temperature with a 2-line LCD for mounting on a 35 mm DIN rail Brief description Depending on the configuration,
More informationTechnical data. General specifications. factory setting: 125 ms Memory Non-volatile memory Write cycles Indicators/operating means
Model Number Single head system Features IO-link interface for service and process data Programmable via DTM with PACTWARE Switch output and analog output Selectable sound lobe width Synchronization options
More informationGOLANDER PUMP BQ80S Microflow Variable-Speed Peristaltic Pump Operation Manual
GOLANDER PUMP BQ80S Microflow Variable-Speed Peristaltic Pump Operation Manual info@golanderpump.com http://golanderpump.com 1-678-587-8806 Contents 1 Safety Cautions... 1 2 Introduction... 1 3 Functions
More informationMODEL 8000MP LEVEL SENSOR
1 MODEL 8000MP LEVEL SENSOR INSTRUCTIONS FOR INSTALLATION, OPERATION & MAINTENANCE VISIT OUR WEBSITE SIGMACONTROLS.COM 2 SERIES 8000MP LEVEL SENSOR 1. DESCRIPTION The Model 8000MP Submersible Level Sensor
More informationHIGH-VOLTAGE DC METER
HIGH-VOLTAGE DC METER Javelin D PD644 0-300 VDC input NEMA 4X, IP65 front Scale in engineering units Sunlight readable LED display 4-20 ma analog output Two form C 3 A relays option RS-485 serial communications
More informationThermometer / Barometer System
TBX USER S MANUAL Thermometer / Barometer System 616TA Pg.2of12 TABLE OF CONTENTS QUICK START GUIDE... 3 INTRODUCTION... 6 SPECIFICATIONS... 6 TERMINAL COMPARTMENT LAYOUT... 7 FEATURES... 8 OPERATION...
More information955S BRIK. Series 955S Brik. Smart BRIK LDT ABSOLUTE PROCESS CONTROL KNOW WHERE YOU ARE... REGARDLESS. Linear Displacement Transducer
Series 955S Brik Linear Displacement Transducer Installation Manual 955S BRIK Smart BRIK LDT ABSOLUTE PROCESS CONTROL KNOW WHERE YOU ARE... REGARDLESS Introduction The 955S Smart Brik is an accurate programmable,
More informationDREXELBROOK. Datasheet Universal IV In-Line Water Cut Meter. with Spool Piece Mounting USE THE BEST APPLICATIONS. Eliminate Routine Maintenance
USE THE BEST For over 50 years, Drexelbrook has been the world s leader in capacitive based measurements by providing reliable and accurate products at a reasonable cost. We offer the highest pressure
More informationSubmersion/Insertion ph/orp Sensors
Instruction Sheet PN 51A-396VP/rev.G January 2011 Models 396VP and 398VP Submersion/Insertion ph/orp Sensors For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/.
More informationSee below for software screens to guide you through the configuration and calibration of your sensor
Modbus sensor addendum Rev 3 1/4/19 Introduction Thank you for purchasing Sensorex Modbus water quality sensors. This instruction manual is an addendum to all product instructions that only covers calibration
More information1. PRODUCT INTRODUCTION Description 1-2. Feature...
CONTENTS PAGE 1. PRODUCT INTRODUCTION... 1-1. Description 1-2. Feature... 1 1 1 2. TECHNICAL SPECIFICATIONS 3 3. PRECAUTIONS BEFORE OPERATION.... 3-1. Unpacking the Instrument... 3-2. Checking the Line
More informationDIABLOSPORT PREDATOR REVISION UPDATE INSTRUCTIONS
DIABLOSPORT PREDATOR REVISION UPDATE INSTRUCTIONS This page contains instructions that will guide you through the process of updating the DiabloSport Predator to the latest software revision available.
More informationEchotouch, EchoSpan & EchoSwitch Ultrasonic Liquid
Echotouch, EchoSpan & EchoSwitch Ultrasonic Liquid The Echotouch, EchoSpan and EchoSwitch are innovative ultrasonic liquid level sensor families that replace float, conductance and pressure sensors that
More informationMIST Version 2.0 USER S GUIDE. Preliminary Version. INTRODUCTION...2 General Description...2 Models Available...2
1 MIST Version 2.0 USER S GUIDE Preliminary Version INTRODUCTION...2 General Description...2 Models Available...2 INSTALLATION...2 Unpacking...2 Mechanical Installation...3 Electrical Wiring...3 Stand-alone
More informationSimplified information where and how you want it.
ConsolIdator 4 & 8 Multi-channel controllers Simplified information where and how you want it. ConsoliDator 4 Model PD940 ConsoliDator ConsoliDator 4 Features Four 4-20 Inputs Four 4-20 Outputs ConsoliDator
More informationHD B. [ GB ] Precision Barometer
HD 9408.3B [ GB ] Precision Barometer [ GB ] [ GB ] Characteristics High accuracy temperature-compensated sensor with excellent stability Wide temperature operating range Various output options available
More informationPTU300 Combined Pressure, Humidity and Temperature Transmitter
PTU300 Combined Pressure, Humidity and Temperature Transmitter For Demanding Applications Features Barometric pressure, humidity, and temperature measurement in one transmitter RS-232C serial interface
More informationEE360. High-End Moisture in Oil Transmitter. Typical applications. Features EE360 EE360. (+34)
is dedicated for reliable monitoring of lubrication, hydraulic and insulation oils as well as diesel fuel. In addition to highly accurate measurement of water activity (a w ) and temperature (T), calculates
More informationProduct User Guide PR2000. PR2000 Pressure Data Logger with LCD
Product User Guide PR2000 PR2000 Pressure Data Logger with LCD PR2000 Table of Contents Quick Start Steps... 3 Product Overview... 4 Software Installation... 5 Device Operation... 5 Computer Interface...
More informationAUTOMATION. Operator s Manual PG-7. Full Access. Doc Part Rev B, 07/18. Automation Products Group, Inc.
AUTOMATION P R O D U C T S GROUP, INC. Operator s Manual PG-7 Full Access Doc. 9003312 Part 200180 Rev B, 07/18 Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com E-mail: sales@apgsensors.com PG7
More informationHMT330 Series Humidity and Temperature Transmitters for Industrial Applications
HMT330 Series Humidity and Temperature Transmitters for Industrial Applications The HMT330 transmitter family has the solution for demanding industrial humidity measurements. Six models for demanding industrial
More informationTP32MTT.03 TP32MTT [ GB ] Probes for soil thermal profile measurement
TP32MTT.03 [ GB ] Probes for soil thermal profile measurement [ GB ] [ GB ] Description Temperature measurement at 7 levels (TP32MTT.03) or 6 levels () In accordance with the requirements of the World
More informationSeries ULT Ultrasonic Level Transmitter
Bulletin L-45 Series ULT Ultrasonic Level Transmitter Specifications - Installation and Operating Instructions 3 29/32 [99.21] 4 9/32 [108.74] 2X 1/2 NPT ø3 21/64 [84.53] 8 1/16 [204.79] Series ULT Ultrasonic
More informationElectrodes ph EC DO /14 dge 2 P/L e
ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand Manual log on stability Interval logging Temperature readout ( C or
More informationMODEL 3150 SUSPENDED SOLIDS ANALYZER MODEL 35/35L OPTICAL SUSPENDED SOLIDS SENSOR
MODEL 3150 SUSPENDED SOLIDS ANALYZER MODEL 35/35L OPTICAL SUSPENDED SOLIDS SENSOR REVISION 22 August 2016 Insite Instrumentation Group, Inc 80 Whisperwood Blvd. Suite 107 Slidell, LA 70458 Phone: (985)
More informationThe Model 464-BC Monitor controls and stores flow data from Marsh- McBirney Flo-Dar sensors. A portable computer and Flo-Ware software is used to
Chapter 5 Model 464-BC Monitor Operation The Model 464-BC Monitor controls and stores flow data from Marsh- McBirney Flo-Dar sensors. A portable computer and Flo-Ware software is used to setup the flowmeter
More informationGMC-206/215/230 Digital Desktop Weighing Scale Operation Manual
GMC-206/215/230 Digital Desktop Weighing Scale Operation Manual Version 2.1 Gamma Scale Tel: (905) 455-8333 Fax: (905) 455-0658 E-mail: info@gammascale.com Website: www.gammascale.com 1. Technical Parameters
More informationD115 The Fast Optimal Servo Amplifier For Brush, Brushless, Voice Coil Servo Motors
D115 The Fast Optimal Servo Amplifier For Brush, Brushless, Voice Coil Servo Motors Ron Boe 5/15/2014 This user guide details the servo drives capabilities and physical interfaces. Users will be able to
More informationNIR Moisture Analyzer KB-30
NIR Moisture Analyzer KB-30 Operation Manual Safety Precautions FPL950701 Improper use of the NIR moisture analyzer in violation of the following safety notes may result in death, injury or damage to property
More informationSerial Connection of HC900 Hybrid Controller to 900CS Control Station
Note: Ethernet connections will provide faster performance than RS-485 HC900 1. Remove HC900 CPU and set S2 Dip Switches for RS-485 unterminated Replace CPU & follow instructions per Installation and User
More informationFluoroSELECT Fluorometer User s Manual
FluoroSELECT Fluorometer User s Manual Version 4.4.A Sigma-Aldrich Chemie GmbH Industriestrasse 25 CH-9470 Buchs Switzerland EurTechServ@sial.com techservice@sial.com (US and Canada) sigma-aldrich.com
More informationmanual microflu-blue release: TriOS GmbH Werftweg Oldenburg Germany fon: +49 (0) fax: +49 (0)
manual miniature fluorometer for phycocyanin detection in blue algae (cyanobacteria) release: 2007-04-18 TriOS GmbH Werftweg 15 26135 Oldenburg Germany fon: +49 (0) 441 48598 0 fax: +49 (0) 441 48598 20
More informationFeatures. Introduction
MPM8 Intelligent Pressure Switch MPM8 Intelligent Pressure Switch Features Pressure measurement for gas and liquid Optional for PNP, valtage and current output Peak value display Ranges drift function
More informationNeon pr. ph/ ORP measurement. Applications. Description
Easy handling by graphical menue Fitting for your application by modular setup Simple firmware update and configuration using SD-card Applications Process Water Drinking Water / Beverages Waste Water Treatment
More informationUsers Manual Ocean Sensor Systems Wave Staff III
Users Manual Ocean Sensor Systems Wave Staff III Introduction: The Ocean Sensor Systems Wave Staff III has been designed to provide a high-resolution measurement of liquid surface height at a data rate
More informationSALES NUMBER NUMBER OF BEAMS PROTECTIVE HEIGHT
SECTION 21 182 Safety Light Curtains Type: SLC-F Finger (14mm) FEATURES: Resolution: 14mm Protective height: 160mm to 1040mm Type 4 according to IEC61496-1 and -2 ORDERING INFORMATION: SALES OF BEAMS PROTECTIVE
More informationMULTI PARAMETER WATER QUALITY SENSORS PENTAIR ENVIRONMENTAL SYSTEMS
MULTI PARAMETER WATER QUALITY SENSORS PENTAIR ENVIRONMENTAL SYSTEMS 47.0 MP47 (UP TO 4 PARAMETERS) MP65 (UP TO 6 PARAMETERS) MULTI PARAMETER WATER QUALITY SENSORS DESCRIPTION The Greenspan MP47 and MP65
More informationOperation Manual CX10 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller
Operation Manual CX10 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller 11751 Markon Drive Garden Grove, CA. 92841 U.S.A. pg. 1 CONTENTS GENERAL INTRODUCTION.....3 INITIAL INSPECTION......3
More informationPressure transmitter for highest pressure applications Up to 217,500 psi (15,000 bar) Model HP-2
Electronic pressure measurement Pressure transmitter for highest pressure applications Up to 217,500 psi (15,000 bar) Model HP-2 WIKA data sheet PE 81.53 Applications Test bench construction Water jet
More informationEchotouch, EchoSpan & EchoSwitch Ultrasonic Liquid
Echotouch, EchoSpan & EchoSwitch Ultrasonic Liquid The Echotouch, EchoSpan and EchoSwitch are innovative ultrasonic liquid level sensor families that replace float, conductance and pressure sensors that
More informationZenith DS9000 Dispensing System
Page Date: 04/2009 Zenith DS9000 Dispensing System Installation & Operation Manual Page 2 Table of Content Introduction... 3 Control Specification... 4 Wiring Diagram... 5 Wiring Instructions... 6 Field
More informationDEFECTOMETER High detection sensitivity to surface flaws (even underneath paint) Simple, microprocessor-controlled operation
DEFECTOMETER 2.837 High detection sensitivity to surface flaws (even underneath paint) Simple, microprocessor-controlled operation Automatic lift-off and zero compensation Several alternatives for signal
More informationWeighing Indicator USER S GUIDE. globalindustrial.com. globalindustrial.com. 11 Harbor Park Drive Port Washington, NY 11050
Weighing Indicator USER S GUIDE 0713 11 Harbor Park Drive Port Washington, NY 11050 Table of Contents I Main Specifications.............................3 II Display and Keyboard..........................4
More informationInterface Manual Chemical Injection System
1 Interface Manual Chemical Injection System SignalFire Numbers: CIJ-xLine (Local Modbus control) CIJ-xLine-Wireless (SignalFire radio) SignalFire Chemical Injection Controller Module The SignalFire Chemical
More informationVIP-55. Operation Guide. Video Inspection Probe V
Video Inspection Probe V.9.6.17 Operation Guide Table of Contents 1. VIP-55 Video Inspection Probe...3 FB Connect...4 2. Highlights...4 3. APP installation and use...5 App installation...5 4. USB Cable
More informationIndustrial Pressure Transmitter
NAT 8 Industrial Pressure Transmitter Swiss based Trafag is a leading international supplier of high quality sensors and monitoring instruments for measurement of and temperature. The industrial transmitter
More informationF-1000 DIGITAL PADDLEWHEEL FLOWMETER
F-1000 DIGITAL PADDLEWHEEL FLOWMETER F1000 with Machined In-line Fitting High accuracy digital paddlewheel technology with very low pressure drop. 3/8, 1/2, 3/4, 1, 1-1/2, and 2 female pipe threads. Flow
More informationInterface Level Analyzers & Sensors
products dedicated to solving problems Algae Control Ammonia Interface Level Analyzers & Sensors Calibration Chlorine Conductivity Cooling Tower Monitoring Dissolved Oxygen Interface Level Nitrate/Nitrite
More informationGE Energy. HYDRAN 201i System Technical Specifications General Electric Company. All Rights Reserved GEA (06/04)
GE Energy HYDRAN 201i System Technical Specifications g 2004 General Electric Company. All Rights Reserved GEA-13763 (06/04) Contents HYDRAN 201Ti Intelligent Transmitter........................... 1
More informationGMM220 CO 2. Transmitter Modules. Introduction. Mounting
Introduction GMM220 CO 2 Transmitter Modules Vaisala GMM220 transmitter modules are versatile instruments for measuring CO 2 in industrial applications. The CARBOCAP sensor is silicon based and its operation
More informationControllers. Metal Finishing. and Surface Treatment
Metal Finishing and Surface Treatment Controllers Walchem integrates its advanced sensing, instrumentation, fluid handling and communications technologies to deliver reliable and innovative solutions to
More information