VS-900. Vital Signs Monitor. Service Manual

Size: px
Start display at page:

Download "VS-900. Vital Signs Monitor. Service Manual"

Transcription

1 VS-900 Vital Signs Monitor Service Manual

2

3 Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this product and this manual. This manual may refer to information protected by copyrights or patents and does not convey any license under the patent rights of Mindray, nor the rights of others. Mindray does not assume any liability arising out of any infringements of patents or other rights of third parties.,, and are the registered trademarks or trademarks owned by Mindray in China and other countries. Revision History This manual has a revision number. This revision number changes whenever the manual is updated due to software or technical specification change. Contents of this manual are subject to change without prior notice. The version information of this manual is as follows: Version number: 1.0 Release time: May 2013 Copyright 2013 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved. I

4 Preface Manual Purpose This manual provides detailed information about the assembling, dissembling, testing and troubleshooting of the equipment to support effective troubleshooting and repair. It is not intended to be a comprehensive, in-depth explanation of the product architecture or technical implementation. This manual is based on the maximum configuration and therefore some contents may not apply to your product. If you have any question, please contact our Customer Service Department. Observance of the manual is a prerequisite for proper equipment maintenance and prevents equipment damage and personnel injury. Intended Audience This manual is for biomedical engineers, authorized technicians or service representatives responsible for troubleshooting, repairing and maintaining the equipment. Passwords A password may be required to access different modes of the following equipments. The passwords are listed below: User settings: (User adjustable) Factory maintenance: Demo mode: 2088 Configuration mode: (User adjustable) II

5 Contents 1 Introduction Information of This Manual Safety Information Dangers Warnings Cautions Notes Equipment Symbols Theory of Operation Overview Connectors for Peripheral Devices Main Unit Front Housing Assembly Main Board Rear Housing Assembly External Module Equipment Installation Unpacking the Equipment Preparation for Installation Preparation for Installation Site Environmental Requirements Electrical Requirements Equipment Installation Preparation for Power on Testing and Maintenance Introduction Test Equipment Test Report Preventative Maintenance Recommended Frequency Visual Inspection Power-on Test Module Performance Tests SpO2 test NIBP Test Temp Test Nurse Call Reply Performance Test Bar Code Scanner Test

6 4.5 Electric Safety Tests Touchscreen Calibration Recorder check Battery Check Factory Maintenance Accessing Factory Maintenance Menu Drawing Waves Software Version Monitor Information Calibrate NIBP Troubleshooting Overview Part Replacement Checking Equipment Status Checking Software Version Checking Technical Alarms Troubleshooting Guide Power On/Off Failure Display Failures Alarm Lamp Failures Button and Knob Failures Sound Failures Battery Failures Recorder Failures Output Interface Failure Data Storage Failure Wired Network Related Problems Wi-Fi Related Problems Module defective Software Upgrade Problems Technical Alarm Messages Disassembly and Repair Tools Required Preparations for Disassembly Disassembling the Main Unit Disassembling the Temp Module (Optional) Removing the Recorder Separating the Front and Rear Half of the Monitor Removing the Parameter Connector Panel Assembly Disassembling the Main Bracket Assembly Removing the Parameter Board (SpO2 Optional) and Power Management Board Disassembling Pumps and Valves (NIBP optional)

7 6.3.8 Disassembling AC/DC Power Board and Battery Converter Board Disassembling the Front Housing Assembly Removing the Touchscreen Control Board (Optional) Removing the Wi-Fi Module (Optional) Removing the Main Control Board Removing SD Card (Optional) Removing the Touchscreen (Optional) Disassembling the Display Removing the Keypad Removing the Encoder Removing the Alarm Lamp Disassembling the Temp Module (Optional) Removing the Temp Module PCBA and Temp Module Power Board PCBA Disassembling the Temp On-Position Detection Board PCBA Parts Introduction Main Unit Exploded View Parts List Front Housing Subassembly (Touchscreen) Exploded View Parts List Front Housing Subassembly (Non-Touchscreen) Exploded View Parts List Rear Housing Assembly Exploded View Parts List Main Bracket Assembly Exploded View Parts List Power Management Board Assembly Exploded View Parts List Parameter Connector Panel Assembly Exploded View Parts List Predictive Temp Assembly Exploded View Parts List Hardware and Software Upgrade Hardware Upgrade

8 8.1.1 Upgrade Package Upgrading Parameter Modules Upgrading NIBP Upgrading Temp Enabling Parameter Functions Upgrading Wireless Network Function Upgrading Touchscreen Upgrading Recorder Function Upgrading Nurse Call Function and DIAP Protocol Port Software Upgrade Installing Mindray Patient Monitor Software Upgrade Tool Software Upgrade Procedure A Electrical Safety Inspection... A-1 A.1 Power Cord Plug...A-1 A.2 Device Enclosure and Accessories...A-1 A.3 Device Labeling...A-2 A.4 Protective Earth Resistance...A-2 A.5 Earth Leakage Test...A-3 A.6 Patient Leakage Current...A-5 A.7 Mains on Applied Part Leakage...A-6 4

9 1 Introduction 1.1 Information of This Manual Version Revision History 1.0 New 1.2 Safety Information DANGER Indicates an imminent hazard that, if not avoided, will result in death or serious injury. WARNING Indicates a potential hazard or unsafe practice that, if not avoided, will result in death or serious injury. CAUTION Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage. NOTE Provides application tips or other useful information to ensure that you get the most from your product. 1-1

10 1.2.1 Dangers There are no dangers that refer to the product in general. Specific Danger statements may be given in the respective sections of this manual Warnings WARNING All installation operations, expansions, changes, modifications and repairs of this product are conducted by Mindray authorized personnel. There is high voltage inside the equipment. Never disassemble the equipment before it is disconnected from the AC power source or the battery. When you disassemble/reassemble a parameter module, a patient leakage current test must be performed before it is used again for monitoring. The equipment must be connected to a properly installed power outlet with protective earth contacts only. If the installation does not provide for a protective earth conductor, disconnect it from the power line and operate it on battery power, if possible. Disposal of the packaging material should observe the applicable waste control regulations and keeping it out of children s reach Cautions CAUTION Make sure that no electromagnetic radiation interferes with the performance of the equipment when preparing to carry out performance tests. Mobile phone, X-ray equipment or MRI devices are a possible source of interference as they may emit higher levels of electromagnetic radiation. Before connecting the receiver to the power line, check that the voltage and frequency ratings of the power line are the same as those indicated on the unit s label or in this manual. Protect the equipment from damage caused by drop, impact, strong vibration or other mechanical force during servicing Notes NOTE Refer to Operator's Manual for detailed operation and other information. 1-2

11 1.3 Equipment Symbols NOTE Not all the symbols listed below are applied to your equipment. Caution: Consult accompanying documents(this manual) Alternating current (AC) Power ON/OFF Battery indicator AUDIO PAUSED NIBP Start/Stop key Alarm Paused Record Admit a new patient DEFIBRILLATION PROOF TYPE CF APPLIED PART Insertion Direction Input/Output Network connector Equipotentiality USB connectors Manufacturer Serial number DATE OF MANUAFACTURE Electrostatic sensitive devices. Protection against fluid ingress The following definition of the WEEE label applies to EU member states only. This symbol indicates that this product should not be treated as household waste. By ensuring that this product is disposed of correctly, you will help prevent bringing potential negative consequences to the environment and human health. For more detailed information with regard to returning and recycling this product, please consult the distributor from whom you purchased it. * For system products, this label may be attached to the main unit only. 1-3

12 FOR YOUR NOTES 1-4

13 2 Theory of Operation 2.1 Overview The Monitor is intended for monitoring physiologic parameters, including SpO2, PR, NIBP and TEMP, on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. The equipment also: Provides audible and visual alarm indications in case of patient or equipment problems; Provides display, review, storage and printing of monitored information; Incorporates multiple input devices such as buttons, knob, and touchscreen; and, Enables program upgrade over the network. Monitor Nurse Call Wi Fi LAN Reserved port AP Other device CMS/Gateway 2-1 EMR Server The above figure shows a system consisting of the vital signs monitor and its peripheral devices. The vital signs monitor: Can be used for monitoring the physiological parameters, giving alarms and reviewing patient data, etc. Supports recorder. Supports nurse call function. Supports Wi-Fi module, wired network, remote view, and communication with the HyperVisor Central Monitoring System (hereinafter called CMS). Supports external AC power source and an internal battery. Supports clinical data acquisition, which has two ways: by SD card and by USB drive. The system software should support data output function, for SD card is a built-in device.

14 2.2 Connectors for Peripheral Devices Connector for Temp probe 2. Network connector: It is a standard RJ45 connector used to communicate with external devices, such as central monitoring system, e-gateway, or used to export data or upgrade the system. 3. Multi-function connector: connects to the hospital's nurse call system, or connects external devices through DIAP protocol. 4. USB connector: connects to barcode scanner or USB disk. 5. AC Power Input 6. Equipotential grounding terminal: When the equipment and other devices are to be used together, their equipotential grounding terminals should be connected together to eliminate the potential difference between them. 2.3 Main Unit The main unit of the vital signs monitor consists of three parts: Front housing assembly: main board, Wi-Fi module, keypad board assembly (knob), display, touchscreen, and alarm lamp board. Rear housing assembly: power module (AC/DC), power management and interface board (including SpO2 isolation power), recorder, speaker, battery, NIBP module, and SpO2 board (including three types of configuration, i.e. Mindray, Masimo, and Nellcor). External module: Temp module 2-2

15 The following figure shows the main unit architecture of the vital signs monitor. Front Housing Assembly Knob Alarm lamp board Power On/Off keypad (Indicator) SD card Main board Wi-Fi module I2C LVD S Backlight Touchscree n control board LCD display Touch screen Rear Housing Assembly Pump NIBP socket Speaker Power management and interface board NIBP module Recorder 2600mAh battery 15V AC-DC power module SpO2 isolation power SpO2 board (Standard Mindray, OEM optional) SpO2 socket External Module Predictive Temp module Temp probe AC-IN External interface RJ45/USB/Multifunctional (nurse call, extended ports) 2-3

16 2.4 Front Housing Assembly Alarm lamp board Touchscreen control board Wi-Fi antenna Main board Wi-Fi module Keypad Screws fixing the front and rear housing 2.5 Main Board The main board is the control center of the equipment. It provides communication, display, and data storage functions, including: Communication with SpO2 board, NIBP module, and Temp module through connector; Communication with power management board, keypad board, and recorder through connectors; Providing drive for display and backlight; Providing 2 USB connectors, a network connector, and a multifunctional connector; Communication with touchscreen control board through I2C; Providing drive for SD card; and, Providing drive for Wi-Fi module. 2-4

17 Keypad The keypad scans and detects the input of keys and knobs, integrates the power on/off key, and connects AC and battery indicators. Alarm Lamp Board The alarm lamp board is located at the top of front housing. It has two-color indicators, red and yellow. The alarm lamp board directly connects the main board through a cable. It is controlled directly by the main board. Touchscreen and Touchscreen Control Board The touchscreen control board drives the touchscreen and implements communication with the vital signs monitor. Wi-Fi Module The Wi-Fi enable the equipment to connect to g/n wireless network. 2.6 Rear Housing Assembly Speaker NIBP module Power management and interface board Recorder AC/DC power module SpO2 board Rear housing assembly consists of power module (AC/DC), power management and interface board (including SpO2 isolation power), recorder, speaker, battery, NIBP module, and SpO2 board (including three types of configuration, i.e. Mindray, Masimo, and Nellcor). 2-5

18 AC/DC Power Module The AC/DC power board transforms the input AC into DC power, which is the power source for all voltages in the equipment. Power Management and Interface Board The power management and interface board mainly provides the following functions: Charge and discharge of battery and charge detection; DC/DC conversion: outputs 12V and 5V DC power; Control over power on/off key and AC, BAT indicator; Communication transmission among parameter modules; Providing isolation power for Mindray/OEM SpO2 module; and, Providing external connectors, and also filter and protection for these connectors. Recorder The recorder receives data from the main board and sends them to the thermal printhead for printing. NIBP Module The NIBP module consists of blood pressure measurement board and pump and valve assembly, providing measurement acquisition of blood pressure data. The main functions of the NIBP module are: NIBP measurement; and, Data exchange with the main board through the serial ports. SpO2 Board The independently developed Mindray SpO2 board provides SpO2 measurement as good as other boards but with smaller size and lower consumption. The power management and interface board is also compatible with Nellcor NELL-1 SpO2 board and Masimo MS-2013 SpO2 board. The SpO2 board collects SpO2 signals, processes SpO2 algorithm and sends measurement results to the main board. The power management interface board provides isolation power for it. 2.7 External Module An external Temp module can be extended on the monitor. The independently developed Mindray Temp module consists of an isolation power board, Temp measurement board, and probes. The Temp measurement board collects Temp signals, processes algorithm and sends measurement results to the main board. 2-6

19 3 Equipment Installation 3.1 Unpacking the Equipment Open the package and take out the packing list. Check that all the articles included in the packing list are available and the quantity and specification are correct. Make sure that: All the optional parts purchased by the customer shall also be checked. Notify the supplier if provided components are not correct as compared to the packing list. In case of damage during transportation, keep the packing material and notify the supplier immediately. Keep the packing material till new equipment is accepted. 3.2 Preparation for Installation Preparation for Installation Site 1. Ensure that the site meets all safety, environmental and power requirements. 2. Check that required power sockets are available. 3. Check that a network connector is available if the equipment needs to be connected to network Environmental Requirements To avoid explosion hazard, do not use the equipment in the presence of flammable anesthetics, vapors or liquids. The environment where the equipment will be used should be reasonably free from vibration, dust and corrosive substances. If these conditions are not met, the system may not function normally. The environmental specification is as follows: Main Unit Item Temperature ( ) Operating environment Storage environment 0 to 40 (without Temp module) 5 to 40 (with Temp module) Relative humidity (noncondensing) Altitude (kpa) 15% to 95% 57.0 to to 70 10% to 95% 16.0 to NOTE The environmental specifications of unspecified parameters are the same as those of the main unit. 3-1

20 3.2.3 Electrical Requirements Check cables and power cords. Make sure that: 1. Check that the system cables, power cords, and power plugs are not damaged, and that the pins are not loose. In case of any damage, remove it from use. 2. The insulation of patient cables and leadwires is not damaged, and connectors are not loose. WARNING Only power sockets with protective grounding can be used. Use the supplied power cord only! If you doubt the completeness of the installation and cabling of the external protective cables, disconnect the equipment from the power line and operate it on battery. Otherwise, patient or operator might be shocked. Voltage 100 to 240V AC Current 1.1 to 0.5A Frequency 50/60 Hz 3.3 Equipment Installation Follow the procedure below to install the equipment: 1. Check the quantity of equipment and accessories as per the packing list. 2. Check for mechanical damages on the equipment and accessories. 3. Install the battery (optional). For detailed operations, please refer to the Operator's Manual of the vital signs monitor. 4. Connect AC power. 5. Connect the accessories. Installation Support The vital signs monitor can be mounted on a wall bracket or on a trolley support. The wall bracket or trolley support can be ordered optionally. Each type of mounting bracket is delivered with a complete set of mounting hardware and instructions. For detailed installation information, please refer to Wall-mount Bracket Instructions for Use (PN: ) and Rollstand Instructions for Use (PN: ). 3-2

21 CAUTION Use mounting brackets we supply or approve. If other compatible mounting bracket is used, be sure it can be safely used on the vital signs monitor. The mounting bracket should be installed by our qualified service personnel, or engineers who have adequate knowledge on it. If other mounting solution is used, the installation personnel and the customer should verify if it can be safely used on the equipment, and the customer assume the responsibility for any risk resulting from that Preparation for Power on 1. Before you start to make measurements, check the equipment for any mechanical damage and make sure that all external cables, plug-ins and accessories are properly connected. 2. Plug the power cord into the AC power source. If you run the equipment on battery power, ensure that the battery is sufficiently charged. 3. Press the Power ON/Off button ( ) on the front panel to turn on the equipment. 3-3

22 FOR YOUR NOTES 3-4

23 4 Testing and Maintenance 4.1 Introduction To ensure the equipment always functions normally, qualified service personnel should perform regular inspection, maintenance and test. This chapter provides a checklist of the testing procedures for the equipment with recommended test equipments and frequency. The service personnel should perform the testing and maintenance procedures as required and use appropriate test equipments. The testing procedures provided in this chapter are intended to verify that the equipment meets the performance specifications. If the equipment or a module fails to perform as specified in any test, repairs or replacements must be done to correct the problem. If you have any question, contact our Customer Service Department. CAUTION All tests should be performed by qualified service personnel only. Care should be taken to change the settings in [User Settings >>] and [Factory Maintenance >>] menus to avoid loss of data. Service personnel should acquaint themselves with the test tools and make sure that test tools and cables are applicable Test Equipment See the following sections Test Report Upon the completion of the tests, the table of preventative maintenance test reports and the table of maintenance test reports in this chapter should be kept properly Preventative Maintenance Below are preventative maintenance tests which need to be performed on the monitor. The recommended frequency of preventative maintenance is at least once per year. See the following sections for detailed maintenance procedures. Visual inspection NIBP test and calibration 4-1

24 4.1.4 Recommended Frequency Check/Maintenance Item Visual inspection Power-on test SpO2test NIBP test TEMP test Nurse call function test Electric Safety Tests Touchscreen Calibration Recorder check Battery check Pressure check Leakage test Calibration Refer to Appendix A Electrical Safety Inspection. Functional test Performance test Frequency 1. When first installed or reinstalled. 1. When first installed or reinstalled. 2. Following any repairs or replacement of any main unit parts. 1. If the user suspects that the measurement is incorrect. 2. Following any repairs or replacement of relevant module. 3. At least once every two years. Note: At least once a year is recommended for NIBP. If the user suspects that the nurse call function does not work well. 1. Following any repair or replacement of power module 2. After the monitor drops. 3. At least once every two years. 1. When the touchscreen appears abnormal. 2. After the touchscreen is replaced. Following any repair or replacement of the recorder. 1. When first installed. 2. Whenever a battery is replaced. Once per year or if the battery run time reduced significantly. 4.2 Visual Inspection Perform an overall inspection on the appearance of the equipment. The test is passed if the equipment has no obvious signs of damage. Follow these guidelines when inspecting the equipment: Carefully inspect the case, display screen, buttons, and knob for obvious signs of damage. Inspect all external connections for loose connectors, bent pins or frayed cables. Inspect all connectors on the equipment for loose connectors or bent pins. Make sure that safety labels and data plates on the equipment are clearly legible. 4-2

25 4.3 Power-on Test This test is to verify that the equipment can power up correctly. This test is passed if the equipment starts up by following this procedure: 1. Insert the battery in the battery compartment, and connect the equipment to the AC mains. The AC mains indicator and battery indicator light up. 2. Press the Power ON/Off button ( ) on the front panel to turn on the equipment. The work status indicator lights up inside the Power On/Off button. 3. The equipment gives a beep, which indicates that the selftest for alarm audio is passed, and the alarm lamp turns yellow, then red, and then off, which indicates that the selftest for alarm lamp is passed. 4. The welcome screen passes and the main interface is displayed. Now the equipment is correctly started. 4.4 Module Performance Tests SpO 2 test Test Method 1 Tool required: None 1. Connect SpO2 sensor for adult to the SpO2 connector of the monitor. Set [Patient Cat.] to [Adu]. 2. Apply the SpO2 sensor to your ring finger (assume that you stay healthy). 3. Check the Pleth wave and PR reading on the screen and make sure that the displayed SpO2 is within 95% and 100%. 4. Remove the SpO2 sensor from your finger and make sure that an alarm of SpO2 Sensor Off is triggered. Measurement validation The SpO2 accuracy has been validated in human studies against arterial blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically distributed, and only about two-thirds of the measurements can be expected to fall within the specified accuracy compared to CO-oximeter measurements. NOTE The SpO2 simulator can only be used to verify that the pulse oximeter operates properly. It cannot be used to verify the accuracy of the pulse oximeter or the SpO2 sensor. To verify the accuracy, clinical tests are required. 4-3

26 Test Method 2 Tool required: SpO2 simulator, Index-2 recommended 1. Connect the SpO2 sensor to the SpO2 simulator. 2. Selected the model and manufacturer of the SpO2 module to be tested on the simulator, and set the simulator as follows: SpO2 to 96% and PR to 80 bmp. 3. Set the patient type to [Adu], [Ped], and [Neo] respectively. Observe the monitor and make sure the displayed SpO2 and PR value fall in the following range. Manufacturer SpO2 PR Mindray Nellcor Masimo 96% ± 2% (Adult, pediatric) 96% ± 3% (Neonate) 96% ± 2% (Adult, pediatric) 96% ± 3% (Neonate) 96% ± 2% (Adult, pediatric) 96% ± 3% (Neonate) 80 ± 3 bpm 80 ± 3 bpm 80 ± 3 bpm NIBP Test Leakage Test NOTE Perform NIBP leakage test before any other NIBP concerned test and calibration. Tools required: NIBP cuff for adult patient Air tubing Cylinder Follow this procedure to perform the leakage test: 1. In the [Patient Demographics] menu, set [Patient Cat.] to [Adu]. 2. Connect the NIBP cuff to the NIBP connector on the monitor. 3. Apply the cuff to the cylinder as shown below. 4-4

27 Monitor Cylinder Connector for NIBP cuff Air tubing Cuff 4. Select [Main] [ Maintenance >>] [ NIBP Leakage Test]. The message [Leakage Testing ] is displayed in the NIBP parameter area. After about 20 seconds, the monitor will automatically deflate. This means the test is completed. If no message is displayed in the NIBP parameter area, it indicates that the system has no leakage. If the message [NIBP Pneumatic Leak] is displayed, it indicates that the system may have a leakage. Check the tubing and connections for leakages. If you ensure that the tubing and connections are all correct, perform a leakage test again. If the problem persists, contact your service personnel. You can either perform a manual leakage test: 1. Perform steps 1 to 4 in the NIBP Accuracy Test section. 2. Raise the pressure in the rigid vessel to 250 mmhg with the balloon pump. Then, wait for 5 s to let the measured values becoming stable. 3. Record the current pressure value, and meanwhile use a time counter to count the time. Then, record the pressure value after 60 s. 4. Compare the two pressure values and make sure the difference should not be greater than 6 mmhg. NIBP Accuracy Test Tool required: T-shape connector Appropriate tubing Balloon pump Rigid Vessel with volume 500 ± 25 ml Reference manometer (calibrated with accuracy equal to or better than 0.75 mmhg) Follow this procedure to perform the accuracy test: 1. Connect the equipments as shown below. 4-5

28 Monitor Standard sphygmomanometer Connector for NIBP cuff Appropriate tubing Balloon pump Rigid Vessel 2. Before inflation, check that the reading of the manometer is 0. If not, turn off the balloon pump to let the whole airway open to the atmosphere. Turn on the balloon pump after the reading is Select [Main] [ Maintenance >>] [ NIBP Accuracy Test]. 4. Check the manometer values and the monitor values. Both should be 0 mmhg. 5. Raise the pressure in the rigid vessel to 50 mmhg with the balloon pump. Then, wait for 10 seconds to let the measured values become stable. 6. Compare the manometer values with the monitor values. The difference between the manometer and displayed values should be ± 3 mmhg. If it is greater than ± 3 mmhg, calibrate the monitor as described in NIBP Calibration. 7. Raise the pressure in the rigid vessel to 200 mmhg with the balloon pump. Then, wait for 10 seconds to let the measured values become stable. Repeat step 6. NOTE You can use an NIBP simulator to replace the balloon pump and the reference manometer to perform the test. You can use an appropriate cylinder and a cuff instead of the rigid vessel. NIBP Calibration Tools required: T-shape connector Appropriate tubing Balloon pump Metal Vessel with volume 500 ± 25 ml Reference manometer (calibrated, with accuracy equal to or better than 0.75 mmhg) 1. Perform steps 1 to 4 in the NIBP Accuracy Test section. 2. Select [Main] [ Maintenance >>] [ Factory Maintenance >>] enter the required password [ Calibrate NIBP >>]. 4-6

29 3. In [NIBP Measurement Circuit], set [NIBP Pressure] to 150 mmhg. Raise the pump pressure to 150 mmhg. After the pressure value is stabilized, select the [Calibrate] button to start calibration. 4. In [Overpressure Protection Circuit], set [Patient Cat.] to [Adu/Ped]. Raise the pressure to 330 mmhg. After the pressure value is stabilized, select [Calibrate] to start calibration. 5. In the [Overpressure Protection Circuit], set [Patient Cat.] to [Neo]. Raise the pressure to 165 mmhg. When the pressure value is stabilized, select [Calibrate] to start calibration. 6. All calibration results are displayed in the [Calibrate NIBP] menu. If the calibration fails, check the test system for leakage and perform another calibration Temp Test Tool required: Resistance box (with accuracy above 0.1 Ω) 1. Connet the Temperature probe to Monitor. Connect the two wires (orange wires) of Temp probe connector to the two ends of the resistance box. 2. Set the resistance box to Ω (corresponding temperature is 25.3 ). 3. Verify displayed tempreture value on monitor is within 25.3± Set the resistance box to Ω (corresponding temperature is 37.0 ). 5. Verify displayed tempreture value on monitor is within 37.0± Set the resistance box to Ω (corresponding temperature is 43.7 ). 7. Verify displayed tempreture value on monitor is within 43.7±0.1. Please contact our service personnel if the Temp test fails. NOTE Due to the different application environment and the test object in vivo and vitro conditions, there is deviations in the measurement result. The maximum deviation of 2.5 may exist in predictive Temp measurement by liquid bath techniques Nurse Call Reply Performance Test Tool required: Oscilloscope 1. Connect the nurse call cable to the Multifunctional Connector of the equipment. 2. Select [Main] [ Maintenance >>] [ User Settings >>] enter the required password] Ok] [ Nurse Call >>]. 3. In the [Nurse Call Setup] menu, select the [Alarm Level] and [Alarm Category], and set [Contact Type] to [Normally Open, and [Signal Type] to [Continuous]. 4-7

30 4. Click to save the changes and quit the menu. 5. Select [Main] [ Maintenance >>] [ Demo >>] enter the required password [ Ok] to enter demo mode. 6. Make the monitor to generate an alarm and check that the oscillograph outputs continuous high level when there is an alarm. 7. Quit the demo mode, and repeat step In the [Nurse Call Setup] menu, select the [Alarm Level] and [Alarm Category], and set [Contact Type] to [Normally Open, and [Signal Type] to [Pulse]. 9. Repeat steps 4 and Make the monitor to generate an alarm and check that the oscillograph displays positive pulses of 1s width when there is an alarm Bar Code Scanner Test Tool required: None 1. Aim the bar code scanner at the target bar code. Adjust the field of view to capture the bar code. 2. Hold the trigger until the bar code scanner gives a beep, indicating the bar code is successfully decoded. Meanwhile, the indicator turns green and scanned characters are displayed on the monitor. Please contact our service personnel if the bar code scanner fails to work normally. 4.5 Electric Safety Tests See Appendix A Electrical Safety Inspection for electrical safety tests. 4.6 Touchscreen Calibration Tool required: None 1. Select [Main] [ Maintenance >>] [ Touchscreen Calibration]. 2. The symbol will appear at different positions of the screen. 3. Select, in turn, the central point of the symbol. 4. After the calibration is completed, the message [Screen Calibration Completed!]. Select [Ok] to confirm the completion of calibration. 4-8

31 4.7 Recorder check Tool required: None 1. Print SpO2 Pleth waveform. The recorder should print correctly and the printout should be clear. 2. Set the recorder to some problems such as out of paper, etc., and then the equipment should give corresponding prompt messages. After the problem is removed, the recorder should be able to work correctly. 3. Set the recorder to print trend data. Check that the recorder can give printouts accordingly. Please contact our service personnel if the recorder test fails. 4.8 Battery Check Tool required: None Functional Test 1. If the equipment is installed with a battery, remove the battery first. 2. Verify that the equipment works correctly when running powered form an AC source. 3. Insert the battery per the procedures provided in the Operator s Manual. 4. Remove the AC power cord and verify that the equipment still works correctly. Performance Test Perform the test by referring to the Battery chapter in the Operator's Manual (P/N: H ) and verify the operating time of the battery meets the product specification. 4-9

32 4.9 Factory Maintenance Accessing Factory Maintenance Menu Select [Main] [ Maintenance >>] [ Factory Maintenance >>] enter the required password [ Ok] to access the Factory Maintenance menu Drawing Waves There are two methods to draw waves: [Color] and [Mono]. Color: selecting Color will have smoother waveforms. Mono: selecting Mono will have a wider viewing angle. 4-10

33 4.9.3 Software Version Selecting the [Software Version >>] will show software version information, as shown below: System software version 2. Power software version 3. Module software version 4-11

34 4.9.4 Monitor Information Selecting [Monitor Information >>] will show the status of the monitor as shown below: NOTE If the main board is replaced, you need to check the label on the main unit and reconfigure the serial number of the monitor Calibrate NIBP For details, refer to section 4.4.2NIBP Test. 4-12

35 Customer name Customer address Servicing person Servicing company Equipment under test (EUT) Model of EUT SN of EUT Hardware version Software Version Maintenance and Test Report (See the above sections for detailed test procedures and contents) Test equipment Model/No. Effective date of calibration Test Contents Visual Inspection The case, display screen, buttons, knob, power cord, wall mount, and accessories have no obvious signs of damage. The external connecting cables are not frayed and the connector pins are not loose and bent. The external connectors are not loose or their pins are not bent. The safety labels and data plate are clearly legible. Power-on test The power-on test is passed. The power indicator and alarm system work correctly and the monitor start up properly. Performance test SpO2test Measure SpO2 on a healthy person s finger and a Pleth wave and PR value are displayed. The displayed SpO2 value is within 95% and 100%. NIBP test 4-13 Test Record Test Result (Pass/Fail)

36 The difference is within ±3 mm when 0, 50 or 200 mmhg is set for NIBP accuracy test. There is no leakage with NIBP, or the manual leakage test result does not exceed 6 mmhg/min. Temp test The displayed Temp value on the monitor is within 37 ± 0.1 ºC. Nurse Call Reply Performance Test When an alarm is reported on the monitor, a nurse call is sent out through the cable. Electric Safety Tests Refer to. All the electrical safety tests should be passed. Touchscreen Calibration The touchscreen is calibrated successfully. Recorder check The recorder can print correctly and the printout is clear. Set the recorder to some problems such as out of paper, etc. the monitor gives corresponding prompt messages. After the problem is removed, the recorder is able to work correctly. Automatic alarm recording for each parameter functions correctly when parameter alarms occur. Battery check The monitor can operates correctly from battery power when an AC power failure accidentally occurs. The monitor can operate independently on a single battery. Test conclusion Tested by: Test date: 4-14

37 5 Troubleshooting 5.1 Overview In this chapter, equipment problems are listed along with possible causes and recommended corrective actions. Refer to the tables to check the monitor, identify and eliminate the troubles. The troubles we list here are frequently arisen difficulties and the actions we recommend can correct most problems, but not all of them. For more information on troubleshooting, contact our Customer Service Department. 5.2 Part Replacement Printed circuit boards (PCBs), major parts and components in the monitor are replaceable. Once you isolate a PCB you suspect defective, follow the instructions in 6 Disassembly and Repair to replace the PCB with a known good one and check that the trouble disappears or the equipment passes all performance tests. Defective PCB can be sent to us for repair. If the trouble remains, exchange the replacement PCB with the original suspicious PCB and continue troubleshooting as directed in this chapter. To obtain information on replacement parts or order them, refer to 7 Parts. 5.3 Checking Equipment Status Some troubleshooting tasks may require you to identify the hardware version and status of your monitor. To check equipment status: 1. Select [Main] [Maintenance >>] [Monitor Information >>]. Then you can view the information on system start time, self check, etc. 2. Select [Main] [Maintenance >>] [ Factory Maintenance >>] enter the required password [ Ok] [Monitor Information >>]. You can also view the information on the monitor s current status. 5.4 Checking Software Version Some troubleshooting may involve software compatibility. Thus it requires you to know your monitor configuration and software version. For detailed information on version compatibility, please contact our Customer Service Department. To view information on the system configuration and system software version, Select [Main] [Maintenance >>] [Factory Maintenance >>] Enter the required password [ Ok] [Software Version >>]. You can also view the information on system software version and module software version. 5-1

38 5.5 Checking Technical Alarms Before troubleshooting the monitor, check for technical alarm messages. If an alarm message is presented, eliminate the technical alarm first. For detailed information on technical alarm message, possible cause and corrective action, refer to the monitor s Operation Manual. 5.6 Troubleshooting Guide Power On/Off Failure Symptoms Possible Cause Troubleshooting The monitor fails to start. AC mains not connected or battery too low Cable defective Power board defective Power management board defective The main board failed. Check that AC mains is properly connected or battery capacity is sufficient. 1. Check that the cable between the keypad board and main board is correctly connected. 2. Check that the cable between the power board and power management board is correctly connected. 3. Check that the cable between the main board and power management board is correctly connected. Replace the power board. Replace the power management board. Replace the main board Display Failures Symptoms Possible Cause Troubleshooting The display is blank or black. Images overlapped or Cable defective Main board defective Display defective Main board error 1. Check if the cable between the display and main board and the backlight cable are correctly connected. 2. Check that the cables and connectors are not damaged. Replace the main board. Replace the display. Replace the main board, or upgrade the main board with the upgrade software. 5-2

39 Symptoms Possible Cause Troubleshooting distorted Cable defective 1. Check if the cable between the display and main board and the backlight cable are correctly connected. Touchscreen does not respond. Touch position invalid Touchscreen disabled Cable defective Touchscreen control board defective Touchscreen defective The main board failed. Touchscreen not calibrated Check if there is a symbol displayed above the [Main] quickkey. If yes, press and hold the [Main Menu] quickkey for more than 3 seconds to enable the touchscreen. 1. Check that the cable between the touchscreen and touchscreen control board is correctly connected. 2. Check that the cable between the touchscreen control board and main board is correctly connected. Replace the touchscreen control board Replace the touchscreen. Replace the main board. Calibrate the touchscreen Alarm Lamp Failures Symptoms Possible Cause Troubleshooting The alarm lamp is not light or extinguished, or the alarm lamp illuminates abnormally. Cable defective Alarm lamp board defective The main board failed. 1. Check that the cable between the alarm lamp board and main board is correctly connected. 2. Check that the cables and connectors are not damaged. Replace the alarm lamp board Replace the main board Button and Knob Failures Symptoms Possible Cause Troubleshooting Buttons do not work Cable defective Keypad board failure Check that the cable between the keypad board and main board is correctly connected. Replace the keypad board. 5-3

40 Knob does not work Cable defective Knob failure Keypad board failure 1. Check that the cable between the knob and keypad board is correctly connected. 2. Check that the cable between the keypad board and main board is correctly connected. Replace the knob encoder. Replace the keypad board Sound Failures Symptoms Possible Cause Troubleshooting No hardkey or knob sound, or hardkey or knob sound abnormal No alarm sound or alarm sound abnormal The key volume is set to 0. Cable defective Speaker defective The main board failed. Power management board defective The alarm volume is set to 0. Cable defective Speaker defective The main board failed. Power management board defective Select [Main] [ General Setup >>] to adjust the key volume. Check that the cable between the speaker and interface board is properly connected. Replace the speaker. Replace the main board. Replace the power management board. Select [Main] [ Maintenance >>] ] User Settings >>] enter the required password [ Ok] [ Alarm Setup >>] and set the [Minimum Alarm Volume] to a proper level in the prompt menu. Select [Main] [ General Setup >>] to adjust the key volume. 1. Check that the cable between the speaker and interface board is properly connected. Replace the speaker. Replace the main board. Replace the power management board Battery Failures Symptoms Possible Cause Troubleshooting Battery defective Replace the battery. Battery cannot be charged Cable defective Power management board defective Battery interface board defective Check that the cable between the battery interface board and power management board is correctly connected. Replace the power management board. Replace the battery interface board. 5-4

41 NOTE When the battery module has a failure, it may cause problems to other components. In this case, troubleshoot the battery module per the procedure described in the table above. Components of the main unit are powered by the power module. In the event that a component malfunctions, check if the operating voltage is correct Recorder Failures Symptoms Possible Cause Troubleshooting No printout Poor print quality or paper not feeding properly Recorder module disabled Paper reversed Cable defective Recorder defective Paper roll not properly installed Print head dirty Recorder defective Output Interface Failure 1. Check if the recorder status indicator lights. 2. If yes, select [Main] [ Maintenance >>] [ Factory Maintenance >>] enter the required password [ Ok] [ Device Config >>] and check the recorder box in the prompt menu to enable the recorder. Otherwise, check for other possible causes. Re-install the paper roll. 1. Check that the cable between the recorder and main board is correctly connected. Replace the recorder. Stop the recorder and re-install the paper roll. 1. Check the thermal print head and the paper roller for foreign matter. 2. Clean the thermal print head with an appropriate clean solution. Replace the recorder. Symptoms Possible Cause Troubleshooting Unable to use the USB devices; USB drive data transfer failure Cable defective The connector board failed. Power management board defective The main board failed. 1. Check that the cable between the power management board and main board is correctly connected. 2. Check that the cable between power management board and interface board is correctly connected. Replace the connector board. Replace the power management board. Replace the main board. 5-5

42 Improper setup Select [Main] [ Transfer Data >>] [ Transfer Data to USB] Data Storage Failure Symptoms Possible Cause Troubleshooting Abnormal patient admitting Admit the patient properly. Fails to review SD card full; unavailable for Delete garbage patient data, remove the archived patient more patient data related alarm, and readmit the patient. data The main board failed. Replace the main board. SD card failure SD card not formatted SD card failure SD card is locked Main board defective Format the SD card. Replace the SD card. Unlock the SD card. Replace the main board Wired Network Related Problems Symptoms Possible Cause Troubleshooting Unable to connect the wired network The monitor is frequently off line or disconnects from the network. The monitor is connected to a LAN Incorrect LAN cable connection Incorrect IP address configuration Cable defective The connector board failed. Power management board defective The main board failed. Incorrect LAN cable connection Incorrect LAN cable connection Check LAN cable connection. LAN cable shall not be longer than 50 m. Check for IP address conflict. If yes, reconfigure the IP address. 1. Check that the cable between the power management board and main board is correctly connected. 2. Check that the cable between power management board and interface board is correctly connected. Replace the connector board. Replace the power management board. Replace the main board. Check LAN cable connection. LAN cable shall not be longer than 50 m. Check LAN cable connection. LAN cable shall not be longer than 50 m. 5-6

43 but cannot view other patients under the View Others mode Excessive requests for viewing the monitor at the same time Incorrect IP address configuration A monitor can only be viewed by 4 other monitors at the same time under the View Others mode. The excessive view requests system will be ignored. Check for IP address conflict. If yes, reconfigure the IP address Wi-Fi Related Problems Symptoms Possible Cause Troubleshooting The monitor is frequently off line or disconnects from the Wi-Fi network. Unable to connect the Wi-Fi network. The Wi-Fi signal is unstable in the operating area. The monitor's Wi-Fi antenna is detached or not properly connect the Wi-Fi module. Antenna damaged Wi-Fi module defective Incorrect IP address configuration The Wi-Fi signal is unstable in the operating area. The monitor's Wi-Fi antenna is detached or not connect the Wi-Fi module. Antenna damaged Wi-Fi module defective Main board defective Check the signal quality of the hospital Wi-Fi network. Disassemble the monitor and fix the Wi-Fi antenna. Replace the Wi-Fi antenna. Replace the Wi-Fi module. Check for IP address conflict. If yes, reconfigure the IP address. Check the signal quality of the hospital Wi-Fi network. Fix the Wi-Fi antenna. Replace the Wi-Fi antenna. Replace the Wi-Fi module. Replace the main board Module defective Symptoms Possible Cause Troubleshooting Failed to connect the external parameter modules Module can be loaded, but "XX Module defective Main unit defective Cable defective inside the module 1. Check that the cable between the external converter board inside the module and the converter board is correctly connected. 2. Replace the converter board. 1. Check that the cable between the main board and power management board is correctly connected. 2. Replace the power management board. 3. Replace the main board. Check the cables connecting the converter board and corresponding parameter module. 5-7

44 communication stopped" is reported or some parameters cannot be used Parameter module defective Converter board defective inside the module Replace the corresponding module. "XX" indicates the configured modules, such as NIBP, SpO2 and Temp. Replace corresponding converter board Software Upgrade Problems Symptoms Possible Cause Troubleshooting Boot file upgrade fails Power failure or unintended power off during boot file upgrade Replace the main board. Program upgrade fails Battery abnormal after upgrading the power management program Incorrect network connection Wrong upgrade package has been downloaded Incorrect IP address configuration Fails to power cycle the monitor after upgrading the power management program 1. Check the network connector on the monitor. 2. Make sure that the hub or switch runs normally. Check that net twines are of the right type and have been connected correctly. Select package according to system requirement. Upgrade package shall be.pkg files. Configure a fixed IP address for the monitor. We recommend not to upgrade a program when the monitor is connected to a network with multiple PCs. Upgrade the power management software again and then power cycle the monitor Technical Alarm Messages Please refer to the Operator's Manual. 5-8

45 6 Disassembly and Repair 6.1 Tools Required To disassemble and replace the parts and components, the following tools may be required: Philips screwdrivers Tweezers Sharp nose pliers Clamp Slot-type screwdriver 6.2 Preparations for Disassembly Before disassembling the equipment, finish the following preparations: Stop monitoring, turn off the equipment, and disconnect all the accessories and peripheral devices. Disconnect the AC power source and remove the battery. WARNING Before disassembling the equipment, be sure to eliminate the static charges first. When disassembling the parts labeled with static-sensitive symbols, make sure you are wearing electrostatic discharge protection such as antistatic wristband or gloves to avoid damaging the equipment. Properly connect and route the cables and wires when reassembling the equipment to avoid short circuit. Select appropriate screws to assemble the equipment. If unfit screws are tightened by force, the equipment may be damaged and the screws or part may fall off during use, causing unpredictable equipment damage or human injury. Follow correct sequence to disassembly the equipment. Otherwise, the equipment may be damaged permanently. Be sure to disconnect all the cables before disassembling any parts. Be sure not to damage any cables or connectors. Be sure to place removed screws and disassembled parts properly, preventing them from being lost or contaminated. Place the screws and parts from the same module together to facilitate reassembling. To reassemble the equipment, first assemble the assemblies, and then the main unit. Carefully route the cables. Make sure that the waterproof material is properly applied during reassembling. 6-1

46 6.3 Disassembling the Main Unit NOTE The recorder can be disassembled separately. To disassemble the equipment, place the equipment on a work surface free from foreign material, avoiding damaging the antiglare screen, touchscreen and the knob. Be careful not to break the two cotters on the front ends of rear housing. All the operations should be performed by qualified service personnel only. Make sure to put on the insulating gloves during service operations. Operations relating to optional parts may not apply to your equipment Disassembling the Temp Module (Optional) 1. Lay the monitor on a table as shown below. Unscrew the three M3 6 screws, pull the Temp module up, and disconnect the Temp cable. Temp cable 6-2

47 6.3.2 Removing the Recorder Unscrew the two M3 6 screws and pull the clamps as indicated to take out the recorder. Then disconnect the cable. Clamps NOTE The recorder can be disassembled separately Separating the Front and Rear Half of the Monitor 1. Lay the monitor on a table as shown below. Unscrew the four M3 screws. 2. Remove the recorder or recorder cover, and pull the rear housing out as indicated below to separate the fastener that are fixing the front and rear housing. 6-3

48 3. Stand the monitor and separate the front housing assembly and rear housing assembly with caution. Disconnect the cable between the main board and power management board and then take off the front panel. Cable NOTE When reassembling the equipment, be sure to check if the front housing waterproof strip is correctly placed. 6-4

49 6.3.4 Removing the Parameter Connector Panel Assembly Lay the rear housing assembly of the equipment on a table. Pull out the cable for AC receptacle and SpO2 cable, and then the silicon tube. Cable for AC input receptacle SpO2 signal cable Parameter connector panel assembly Silicone tube 6-5

50 6.3.5 Disassembling the Main Bracket Assembly For monitors without multi-function connector, you need to prize the cover from the inside of the monitor with a slot-type screwdriver, and then take out the cover. Cover Disconnect the speaker cable, recorder cable, power cable, battery interface board cable, and NIBP cable. Unscrew the five PT3 8 screws and a M4 screw. Then take out the main bracket assembly. 6-6

51 Speaker cable Recorder cable Battery interface board cable Power cord NIBP cable 1 grounding screw 5 self-tapping screws 6-7

52 6.3.6 Removing the Parameter Board (SpO 2 Optional) and Power Management Board 1. Unscrew the three M3 6 screws and take the power management board assembly out from the main bracket assembly. 3 M3 screws 2. Unscrew the two M3 4 screws and take the parameter board out from the power management board assembly. Parameter Board 2 M3 screws 3. Unscrew the two M3 6 screws and take out the power management board assembly. 6-8

53 2 M3 screws Power management board Disassembling Pumps and Valves (NIBP optional) 1. Cut the two cable ties and take out the gas pump. Cable tie 2. Unscrew the three M3 6 screws as indicated and take out the valve. 3 M3 screws 6-9

54 6.3.8 Disassembling AC/DC Power Board and Battery Converter Board 1. Unscrew the four M3 6 screws as indicated and take out the AC/DC power board. 4 M3 screws 2. Unscrew the two M3 nuts to take out the battery interface board. 2 M3 screws 6.4 Disassembling the Front Housing Assembly NOTE To disassemble the equipment, place the equipment on a work surface free from foreign material, avoiding damaging the antiglare screen, touchscreen and the knob. Remember to install the screen support pad properly during reassembly. Operations relating to optional parts may not apply to your equipment. Position the touchscreen properly with the flexible cable facing down. 6-10

55 6.4.1 Removing the Touchscreen Control Board (Optional) Loose and unscrew the two M3 6 screws as shown below. Disconnect the cable between main board and touchscreen board, the touchscreen board, and the touchscreen control board cable, and then remove the touchscreen control board. Touchscreen cable Cable between the main board and touchscreen board 2 M3 screws Removing the Wi-Fi Module (Optional) 1. Take out the two antennas on the front panel from the slots as shown below: Antenna Antenna 2. Remove the antennas from the Wi-Fi module PCBA. Antenna Antenna 3. Push the clamps aside to remove the Wi-Fi module. 6-11

56 Clamps Clamps Removing the Main Control Board Disconnect the alarm lamp cable, cable for display backlight, display cable, and the cable between the main board and keypad board respectively. Unscrew the four M3 8 screws and take out the main board, as shown below: 4screws Cable Cable Cable 6-12

57 6.4.4 Removing SD Card (Optional) Take out the main board for VS-900, and push the SD card as indicated below to take out the SD card. SD card Removing the Touchscreen (Optional) Unscrew the seven PT3 8 screws as indicated below. Take out the touchscreen assembly and then the touchscreen. Touchscreen 6-13

58 6.4.6 Disassembling the Display Unscrew the four M3 6 screws indicated below to remove the screen Removing the Keypad Unplug the encoder cable and unscrew the three PT3 8 screws indicated below. Take out the keypad Removing the Encoder Poke the encoder knob out from the slot and loose the nut with a sharp nose plier. Take out the encoder. Knob 6-14

59 6.4.9 Removing the Alarm Lamp Unscrew the four PT2 6 screws indicated below and take out the alarm lamp board and alarm lamp. 6.5 Disassembling the Temp Module (Optional) Removing the Temp Module PCBA and Temp Module Power Board PCBA Loose the unscrew the two M3 6 screws and the two M3 screws. Remove the cover board and metal sheet. Disconnect the Temp board cable and the cable between Temp isolation power board and Temp board. Unscrew the four M3 6 screws, you can take out the Temp module PCBA. Then unscrew the three M3 6 screws, you can take out the Temp module power board PCBA. Screw for cover board Cover board Temp module power board PCBA 4 screws Temp board cable Cable between the isolation power board and Temp board 3 screws Temp module PCBA Metal sheet Screw for metal sheet 6-15

60 6.5.2 Disassembling the Temp On-Position Detection Board PCBA 1. Unscrew the four M3 6 screws as indicated and take out the temp module housing. 4 screws 2. Unscrew the four M2 screws as indicated below and take out the Temp on-position detection board PCBA. 4 screws Temp on-position detection board PCBA NOTE Remember to assemble the silicon button for the Temp on-position detection switch during reassembly. 6-16

61 7 Parts 7.1 Introduction This section contains the exploded views and parts lists of the main unit. It helps the engineer to identify the parts during disassembling the monitor and replacing the parts. This manual is based on the maximum configuration. Your equipment may not have same parts and the quantity of the screws or stacking sleeves etc. may be different with those included in the parts lists. NOTE The part number listed in the Parts List is only for checking the FRU part number which is also included in the Parts List. Please provide the FRU parts number if you want to purchase the spare parts. 7.2 Main Unit Exploded View 7-1

62 7.2.2 Parts List SN PN Description 1 / Front Housing Assembly for VS-900 FRU part number TR6F TR6F recorder TR6F / 3 / 4 / Rear Housing Assembly for VS-900 Screw, Pan head with washer, Phillips M / / 5 / Screw, pan head Phillips, M3 10 / / VS-900 Temp module / Remarks Touch screen Non-touch screen Cover assembly 7-2

63 7.3 Front Housing Subassembly (Touchscreen) Exploded View 7-3

64 7.3.2 Parts List SN PN Description FRU part number Remarks 1 / Font housing for VS / Silicon water-proof strip 4 / Alarm lamp for VS / 6 / Silicon water-proof strip 9 / Touchscreen position pad (8") Antenna cable / Alarm lamp pad for VS-900 / / Alarm Lamp Board / 8 / Cross recessed pan head self-tapping screw PT2X6 / / Touch-panel, resitive-type, 8.4" / 11 / 8" display-short 13 / 8" display-long / 14 / LCD TFT 8.4" 12 / Screw, pan head Phillips, M3 6 / / Touchscreen control board PCBA / 16 / Main board PCBA (6301) / 17 / Screw, Pan head w/washer / / 18 / Plate, mounting display (VS-900, touchscreen) / / 19 / Screw, pan head / / 20 / Keypad board adjusting sleeve / 21 / Keypad board cushion / VS-900 keypad PCBA / Encoder / VS-900 silicon buttons / Knob / 7-4

65 7.4 Front Housing Subassembly (Non-Touchscreen) Exploded View Parts List SN PN Description FRU part number Remarks 1 / Font housing for VS / Alarm lamp for VS-900 VS-900 Water-proof strip, front 7 / X1.5mm poron housing service kit Water-proof strip, 8 / 173.2X1.5mm poron Antenna cable / 4 / Alarm lamp pad for VS-900 / / 5 / Cross recessed pan head self-tapping screw PT2X6 7-5 / / Alarm Lamp Board / 9 / Screw, pan head Phillips, M3 6 / / 10 / LCD TFT 8.4" /

66 SN PN Description FRU part number Remarks 11 / / Plate, mounting display (VS-900, touchscreen) Touchscreen control board PCBA-6301 Screw, Pan head with washer, M3 6 / / / / / 14 / Main board PCBA (6301) / 15 / Naked screen water-proof plate for VS-900 / / 16 / Self-tapping screw / / 17 / Keypad board adjusting sleeve / 18 / Keypad board cushion / VS-900 keypad PCBA / Encoder / Knob / VS-900 silicon buttons / / Encoder cable / / / / / Cable between main board and display Cable between main board and keypad board (6012) Cable between the main board and display backlight board (VS-900) Cable between alarm lamp board and main board (VS-900) / / / / 7-6

67 7.5 Rear Housing Assembly Exploded View Parts List SN PN Description FRU part number Remarks / Rear Housing Subassembly for VS-900 Battery door assembly for VS-900 AC socket assembly for VS-900 Parameter connector panel assembly for VS-900 Parameter connector board support for VS / Power management b df VS / / / Mindray SpO2, with no NIBP with no SpO / With multi-io

68 7 / / Pump 10 / Screw, Pan head w/washer, Phillips M3X6 Valve assembly for VS-900 Shock absorption cushion for pump 16 / Fixing strip 11 / Main bracket assembly for VS WO multi-io / / / / / 12 / Self-tapping screw / / 13 / Speaker laminate / / 14 / Speaker 15 / Speaker pad / / / Cable between the power management board and power board Cable between the interface board and main board Cable between NIBP module and power management board / / / / / Recorder cable / / multifunctional connector cap / ipm NIBP pump service kit 7-8

69 7.6 Main Bracket Assembly Exploded View Parts List SN PN Description FRU part number Remarks 1 / Screw, Pan head w/washer / / Power board insulator / Power board / Knob, Battery latch / 5 / Battery Spring (12 inch) / / 6 / Battery hook binder plate / / 7 / Screw, pan head / / 8 / Recorder bracket / / 9 / Screw, Flat Head Phillips, M3X6 / / 10 / Spring / / 11 / Nut with washer / / 12 / Battery spring / / 7-9

70 SN PN Description FRU part number Remarks 13 / 6301 battery interface PCBA / 14 / Battery compartment for VS-900 / / 7.7 Power Management Board Assembly Exploded View Parts List SN PN Description 1 / 2 / 7-10 FRU part number Power management board service kit for VS Interface board support for VS-900 / / 3 M Plastic hexagon nut M / SpO2 shield / Plastic hexagon bolt / Remarks with multifunctional connector without multifunctional connector

71 SN PN Description FRU part number 6 / Screw, pan head / / 7 / SpO2 board 8 / Screw, Pan head w/washer Parameter Connector Panel Assembly Exploded View / / Remarks Mindray SpO2 board service kit, 9008 V2.0 Nellcor SpO2 board (MDU) SpO2 board, Masimo MS Parts List SN PN Description FRU part number Remarks 1 / Parameter panel for VS / NIBP gas nozzle assembly 3 / SpO2 signal cable VS-900 panel assembly(wo SpO2) Mindray SpO2 module Masimo SpO2 module Nellcor SpO2 module / / Decorative NIBP socket / / / Decorative IBP socket /

72 7.9 Predictive Temp Assembly Exploded View Parts List SN PN Description FRU part number Remarks 1 / Screw, Pan head w/washer / / 2 M09A Predictive Temp module power board PCBA / 3 / Temp module insulator / / 4 / FOOT, "ENDEAVOUR" / Predictive Temp housing / Temp on-position detection board PCBA / 7 / Chassis / / External compartment for Temp /

73 SN PN Description FRU part number Remarks module (VS) Temp cover for VS / 10 M09A Silicon buttons M09A / Predictive Temp module PCBA / 12 / Screw, pan head cross recessed M2X6 / / 13 / Temp module support for VS-600 / / 14 / Screw, Flat Head Phillips, M3X8 / / 15 / Screw, pan head cross recessed M3X6 / / / / Silicone tube / / / / / VS-900 predictive Temp board cable VS-900 Cable between the Temp module and power management board VS-900 Cable between the isolation power board and Temp board / / / 7-13

74 FOR YOUR NOTES 7-14

75 8 Hardware and Software Upgrade 8.1 Hardware Upgrade The monitor supports upgrade of the following functions: NIBP measurement; SpO2 measurement; Temp measurement; Wireless network (Wi-Fi); Touchscreen; and, Recorder Upgrade Package Upgrade package SpO2 NIBP Description of upgrade package VS-900 Mindray SpO2 upgrade kit VS-900 Masimo SpO2 upgrade kit VS-900 Nellcor SpO2 upgrade kit VS-900 NIBP module upgrade kit PN of upgrade package Temp VS-900 Temp module / Wireless Network / ipm Wi-Fi module kit (Wi-Fi) Touchscreen Front housing assembly service kit for VS-900 (Touchscreen, without main board) Recorder TR6F recorder TR6F / Nurse call function and DIAP protocol port Power management board service kit (with multifunctional connector) Remarks / / / / / Upgrading multifunctional connector Note: measurement accessories are not included in the above upgrade packages. 8-1

76 8.1.2 Upgrading Parameter Modules Upgrading Mindray SpO2 and Nellcor SpO2 List of upgrade package: A Mindray SpO2 board or Nellcor SpO2 board; An SpO2 insulator; A connector panel assembly for Mindray SpO2 board or Nellcor SpO2 board; and, Two M3 6 screws 1. Remove the power management board and connector panel assembly as described in 6.3 Disassembling the Main Unit. 2. Assemble the SpO2 board and insulator onto the power management board as described in Removing the Parameter Board (SpO2 Optional) and Power Management Board. 3. Install the power management board assembly with the SpO2 board and the connector panel assembly in the service kit into the main unit as described in 6.3 Disassembling the Main Unit Upgrading Masimo SpO2 List of upgrade package: A power management board assembly with Masimo SpO2; and, A Masimo SpO2 connector panel assembly. 1. Remove the power management board, connector panel assembly and the stopple of multifunctional connector (if there is one) as described in 6.3 Disassembling the Main Unit. 2. Install the power management board assembly with the SpO2 board and the connector panel assembly in the service kit into the main unit as described in 6.3 Disassembling the Main Unit Upgrading NIBP List of upgrade package: An NIBP valve; A pump; Two cable ties; A cable between NIBP module and power management board; A parameter connector panel assembly; and, Two M3 6 screws 8-2

77 Disassemble the main bracket assembly and parameter connector panel assembly as described in 6.3 Disassembling the Main Unit, assemble the NIBP valve and pump onto the main bracket assembly, and then assemble the new connector panel assembly, cables, and tubes together into the main unit Upgrading Temp List of upgrade package: A Temp module with cables; and, Two M3 6 screws. Remove the decorative cover from the Temp module connector. Install the Temp module onto the main unit as described in 6.5 Disassembling the Temp Module (Optional) Enabling Parameter Functions 1. Select [Main] [ Maintenance >>] [ Factory Maintenance >>] enter the required password [ Ok] [ Device Config. >>]. 2. In the prompt menu, check the upgraded functions. 3. Click to save the changes and quit the menu. 4. Restart the monitor and the software for upgraded parameters are enabled Upgrading Wireless Network Function List of upgrade package: A Wi-Fi module; and, Two antenna cables. 1. Install the Wi-Fi module onto the main board and then connect the Wi-Fi antennas as described in Removing the Wi-Fi Module (Optional). 2. Select [Main] [ Maintenance >>] [ User Settings >>] enter the required password [ Ok] [ Network >>]. 3. In the Network Setup menu, set [Network Type] to [WLAN]. Set the wireless network as described in the Network section in the Operator's Manual. 4. Try to connect the equipment to a nearby wireless network and see if the connection succeeds Upgrading Touchscreen List of upgrade package: A front housing assembly service kit for VS-900 (Touchscreen, without main board) 8-3

78 1. Disassemble the main board and cables from the front housing, and install them to the front housing with touchscreen as described in 6.4 Disassembling the Front Housing Assembly. 2. Assemble the front housing assembly and rear housing assembly as described in 6.3 Disassembling the Main Unit. 3. Select [Main] [ Maintenance >>] [ Factory Maintenance >>] enter the required password [ Ok] [ Device Config. >>]. Check the touchscreen function in the menu to enable the software. Then restart the equipment Upgrading Recorder Function 1. Pry the recorder cover on the left side of the monitor with a little screwdriver, and take out the metal sheet used to seal the recorder as described in Removing the Recorder. Then install the recorder assembly. 2. When the recorder is installed, select [Main] [ Maintenance >>] [ Factory Maintenance >>] enter the required password [ Ok] [ Device Config >>]. Check the recorder option in the menu to enable the software. Then restart the equipment Upgrading Nurse Call Function and DIAP Protocol Port Remove the power management board, and the stopple of multifunctional connector on the rear panel of the equipment (if there is one) as described in 6.3 Disassembling the Main Unit. Then assemble the power management board with multifunctional connector. Then reassemble the main unit. Once the above functions are successfully upgraded, perform related tests on the equipment as described in 4.4 Module Performance Tests, 4.6 Touchscreen Calibration, and 4.7 Recorder check. 8.2 Software Upgrade You can upgrade system software and module software by installing and running the Mindray Patient Monitor Software Upgrade Tool on a PC with Windows operating system. Connect the monitor to be upgraded and a PC running the upgrade tool to the same network, or directly connect the monitor and the PC via a crossover network cable. Then configure the IP address of the PC. You can upgrade the following software: No. Type PN Description 1 System software package System software package Mindray NIBP function software 3 Module software Mindray SpO2 function software 4 5 Power Management Software Temp function software Power Management Software 8-4

79 Note: No specific sequence is required for the upgrading of above software. For detailed information, please refer to 8.2.2Software Upgrade Procedure Installing Mindray Patient Monitor Software Upgrade Tool 1. Find the installation program and double click it to start installation. 2. Select language. 3. Click [Ok] and the following screen is displayed. Click [Next] to go to the next step. 4. Enter User Name, Company name, and Serial Number " ". Then click [Next]. 8-5

80 5. Specify the destination folder for installing this program. Then select [Next]. 6. Select Program Folder. Then select [Next]. 7. Click [Finish] to complete installation. 8-6

81 8.2.2 Software Upgrade Procedure Check the version of the software to be upgraded as described in Software Version before upgrade. 1. Connect the monitor to be upgraded and a PC running the upgrade tool to the same network, or directly connect the monitor and the PC via a crossover network cable. 2. Set the IP address of the PC to " XX" and subnet mask to " ". 3. Run Mindray Patient Monitor Software Upgrade Tool on the PC and set Machine to [Vital sign]. 4. On the Mindray Patient Monitor Software Upgrade Tool screen, select [Select Package] and then the packages you want to upgrade. Then select [Start]. 5. Turn on the monitor to be upgrade. Press and hold (Silence) and (Admit Patient) buttons for 2 to 3 seconds, the monitor enters upgrade mode and starts software upgrade automatically, and corresponding prompt messages are displayed on both the monitor and PC. When software upgrade is finished, restart the monitor and check if the software is correctly upgrade. For details of software upgrade, please refer to help and instructions for use of Mindray Patient Monitor Software Upgrade Tool. CAUTION Disconnect the equipment from the patient and make sure the important data are saved before upgrade. Do not shut down or power off the equipment when upgrading the system software. Otherwise, it may cause the equipment to break down. Software upgrade should be performed by qualified service personnel only. Crossover network cable is recommended when a PC is connected for software upgrade. NOTE Make sure the version of the upgrade package is what you desired. To obtain the latest upgrade package, please contact Mindray Customer Service Department. 8-7

82 FOR YOUR NOTES 8-8

83 A Electrical Safety Inspection The following electrical safety tests are recommended as part of a comprehensive preventive maintenance program. They are a proven means of detecting abnormalities that, if undetected, could prove dangerous to either the patient or the operator. Additional tests may be required according to local regulations. All tests can be performed using commercially available safety analyzer test equipment. These procedures assume the use of a 601PROXL International Safety Analyzer or equivalent safety analyzer. Other popular testers complying with IEC used in Europe such as Fluke, Metron, or Gerb may require modifications to the procedure. Follow the instructions of the analyzer manufacturer. he electrical safety inspection should be periodically performed every two years.the safety analyzer also proves to be an excellent troubleshooting tool to detect abnormalities of line voltage and grounding, as well as total current loads. A.1 Power Cord Plug A.1.1 The Power Plug Test Item The power plug The power cord The power plug pins The plug body The strain relief The power plug Acceptance Criteria No broken or bent pin. No discolored pins. No physical damage to the plug body. No physical damage to the strain relief. No plug warmth for device in use. No loose connections. No physical damage to the cord. No deterioration to the cord. For devices with detachable power cords, inspect the connection at the device. For devices with non-detachable power cords, inspect the strain relief at the device. A.2 Device Enclosure and Accessories A.2.1 Visual Inspection Test Item The enclosure and accessories Acceptance Criteria No physical damage to the enclosure and accessories. No physical damage to meters, switches, connectors, etc. No residue of fluid spillage (e.g., water, coffee, chemicals, etc.). No loose or missing parts (e.g., knobs, dials, terminals, etc.). A-1

84 A.2.2 Contextual Inspection Test Item The enclosure and accessories Acceptance Criteria No unusual noises (e.g., a rattle inside the case). No unusual smells (e.g., burning or smoky smells, particularly from ventilation holes). No taped notes that may suggest device deficiencies or operator concerns. A.3 Device Labeling Check the labels provided by the manufacturer or the healthcare facility are present and legible. Main unit label Integrated warning labels A.4 Protective Earth Resistance Protective Earth Resistance is measured using the RED test lead attached to the DUT Protective Earth terminal or enclosure. Select the test current by pressing SOFT KEY 3 to toggle between 1AMP, 10AMP, and 25AMP. The front panel outlet power is turned off for this test. The following conditions apply: L1 and L2 Open. Preparation 1. First select the test current that will be used for performing the Protective Earth Resistance test by pressing AMPERES (SOFT KEY 3). 2. Connect the test lead(s) between the RED input jack and the GREEN input jack. 3. Press CAL LEADS. The 601PRO will measure the lead resistance, and if less than Ohms, it will store the reading and subtract it from all earth resistance readings taken at the calibrated current. If the calibration fails, the previously stored readings will be used until a passing calibration has occurred.: WARNING During Earth Resistance testing, the DUT must be plugged into the 601PRO front outlet. If the DUT fails Earth Resistance, discontinue tests and label the device defective. A-2

85 To Perform the Test 1. From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PRO front panel outlet. 2. Attach the 601PRO RED input lead to the device s Protective Earth terminal or an exposed metal area. 3. Press shortcut key 3. The Protective Earth Resistance test is displayed. 4. Press SOFT KEY 3 to select a test current (1AMP, 10AMP, or 25AMP). The selected test current is displayed in the upper right corner of the display. 5. Press START TEST to start the test. The test current is applied while resistance and current readings are taken. This takes approximately 5 seconds. 6. Press the print data key at any time to generate a printout of the latest measurement(s). NOTE When "Over" is displayed for Ohms, this signifies that a valid measurement was not obtained because either an open connection was detected or that the measurement was not within range. Readings greater than Ohms will be displayed as Over. In Case of Failure Once it reaches the limitation, stop using and inform the Customer Service Engineer for analysis and disposal. LIMITS ALL COUNTRIES R = 0.2 Ω Maximum A.5 Earth Leakage Test Run an Earth Leakage test on the device being tested before performing any other leakage tests. Leakage current is measured the following ways: Earth Leakage Current, leakage current measured through DUT outlet Earth Earth Leakage Current AP-EARTH (ALL Applied Parts connected to Earth), leakage current measured through DUT outlet Earth There is no need to attach a test lead; the 601PRO automatically connects the measuring device internally. A-3

86 To Perform the Test 1. From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PRO front panel outlet, and turn on the device. 2. Attach the device's applied parts to the 601PRO applied part terminals if applicable. 3. Press shortcut key 4.The Earth Leakage test appears on the display, and the test begins immediately: SOFT KEY 1 toggles the DUT outlet Polarity from Normal to Off to Reverse. SOFT KEY 2 toggles the DUT outlet from Earth to No Earth. SOFT KEY 3 toggles the DUT outlet from L2 to No L2. SOFT KEY 4 toggles the AP to Earth to No AP to Earth. 4. Press the print data key at any time to generate a printout of the latest measurement. In Case of Failure Check any broken of the enclosure. Replace any defective part. Inspect wiring for bad crimps, poor connections, or damage. Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or owner to correct any deviations. As a work around, check the other outlets to see if they could be used instead. Change another probe to confirm if the fail is caused by console. If the leakage current measurement tests fail on a new unit and if situation can not be corrected, submit a Safety Failure Report to document the system problem. Remove unit from operation. If all else fails, stop using and inform the Customer Service Engineer for analysis and disposal. LIMITS For UL , 300 μa in Normal Condition 1000 μa in Single Fault Condition For IEC , 500 μa in Normal Condition 1000 μa in Single Fault Condition A-4

87 A.6 Patient Leakage Current Patient leakage currents are measured between a selected applied part and mains earth. All measurements have a true RMS only response. Preparation Perform a calibration from the Mains on Applied Part menu. The following outlet conditions apply when performing this test: Normal Polarity, Earth Open, Outlet ON Normal Polarity, Outlet ON Normal Polarity, L2 Open, Outlet ON Reversed Polarity, Outlet ON Reversed Polarity, Earth Open, Outlet ON Reversed Polarity, L2 Open, Outlet ON WARNING If all of the applied parts correspond to the instrument type, the applied parts will be tied together and one reading will be taken. If any of the applied parts differ from the instrument type, all applied parts will be tested individually, based on the type of applied part. This applies to Auto and Step modes only. To Perform the Test 1. From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PRO front panel outlet, and turn on the device. 2. Attach the applied parts to the 601PRO's applied part terminals. 3. Press shortcut key 6. The Patient Leakage test is displayed, and the test begins immediately. 4. Press APPLIED PART (SOFT KEY 4) at any time to select the desired applied part leakage current. 5. Modify the configuration of the front panel outlet by pressing the appropriate SOFT KEY on the 601PRO. 6. Press the print data key at any time to generate a printout of the latest measurement. In Case of Failure Check any broken of the enclosure. Replace any defective part. Inspect wiring for bad crimps, poor connections, or damage. A-5

88 LIMITS Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or owner to correct any deviations. As a work around, check the other outlets to see if they could be used instead. Change another probe to confirm if the fail is caused by console. If the leakage current measurement tests fail on a new unit and if situation can not be corrected, submit a Safety Failure Report to document the system problem. Remove unit from operation. If all else fails, stop using and inform the Customer Service Engineer for analysis and disposal. For CF For BF applied parts 10μA in Normal Condition 50μA in Single Fault Condition applied parts 100μA in Normal Condition 500μA in Single Fault Condition A.7 Mains on Applied Part Leakage The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage, through a limiting resistance, to selected applied part terminals. Current measurements are then taken between the selected applied part and earth. Measurements are taken with the test voltage (110% of mains) to applied parts in the normal and reverse polarity conditions as indicated on the display. The following outlet conditions apply when performing the Mains on Applied Part test. Normal Polarity; Reversed Polarity Preparation To perform a calibration from the Mains on Applied Part test, press CAL (SOFT KEY 2). 1. Disconnect ALL patient leads, test leads, and DUT outlet connections. 2. Press CAL to begin calibration, as shown: If the calibration fails, the previously stored readings will be used until a passing calibration has occurred. Also, the esc/stop key has no effect during calibration. A-6

89 3. When the calibration is finished, the Mains on Applied Part test will reappear. WARNING A 2-beep-per-second signal indicates high voltage present at the applied part terminals while a calibration is being performed. High voltage is present at applied part terminals while measurements are being taken. To Perform the Test 1. From the MAIN MENU, or with the outlet unpowered, plug the DUT into the Attach the applied parts to the 601PRO applied part terminals. 3. Attach the red terminal lead to a conductive part on the DUT enclosure. 4. Press shortcut key 7. The Mains on Applied Part test is displayed. 5. Select the desired outlet configuration and applied part to test using the appropriate SOFT KEYS: 6. Press START TEST (SOFT KEY 1) to begin the test. 7. Press the print data key to generate a printout of the latest measurement. NOTE If all of the applied parts correspond to the instrument type, the applied parts will be tied together and one reading will be taken. If any of the applied parts differ from the instrument type, all applied parts will be tested individually, based on the type of applied part. This applies to Auto and Step modes only. In Case of Failure Check any broken of the enclosure. Replace any defective part. Inspect wiring for bad crimps, poor connections, or damage. Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or owner to correct any deviations. As a work around, check the other outlets to see if they could be used instead. Change another probe to confirm if the fail is caused by console. If the leakage current measurement tests fail on a new unit and if situation can not be corrected, submit a Safety Failure Report to document the system problem. Remove unit from operation. If all else fails, stop using and inform the Customer Service Engineer for analysis and disposal. A-7

VS-600. Vital Signs Monitor. Service Manual

VS-600. Vital Signs Monitor. Service Manual VS-600 Vital Signs Monitor Service Manual Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this

More information

imec12/imec10/imec8 Patient Monitor Service Manual

imec12/imec10/imec8 Patient Monitor Service Manual imec12/imec10/imec8 Patient Monitor Service Manual Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights

More information

ipm-9800 Patient Monitor Service Manual

ipm-9800 Patient Monitor Service Manual ipm-9800 Patient Monitor Service Manual Revision History This manual has a revision number. This revision number changes whenever the manual is updated due to software or technical specification change.

More information

ipm 12/iPM 10/iPM 8 ipm 7/iPM 6/iPM 5 Patient Monitor Service Manual

ipm 12/iPM 10/iPM 8 ipm 7/iPM 6/iPM 5 Patient Monitor Service Manual ipm 12/iPM 10/iPM 8 ipm 7/iPM 6/iPM 5 Patient Monitor Service Manual Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual

More information

BeneHeart R12/BeneHeart R12A Electrocardiograph. Service Manual

BeneHeart R12/BeneHeart R12A Electrocardiograph. Service Manual BeneHeart R12/BeneHeart R12A Electrocardiograph Service Manual Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property

More information

Intellectual Property Statement

Intellectual Property Statement Intellectual Property Statement Mindray DS USA, Inc. (hereinafter called Mindray DS) owns the intellectual property rights to this product and this manual. This manual may refer to information protected

More information

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide Model P4017 Single Channel USB Oscilloscope Quick Start Guide General Warranty BNC warrants that the product will be free from defects in materials and workmanship for 3 years from the date of purchase

More information

4P S. Wireless LAN Connecting Adapter <BRP069A42> Installation Manual

4P S. Wireless LAN Connecting Adapter <BRP069A42> Installation Manual 4P359542-2S Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com For details

More information

Installing and Configuring Rialto Analytic Appliances

Installing and Configuring Rialto Analytic Appliances Installing and Configuring Rialto Analytic Appliances Important Safety Information This manual provides installation and operation information and precautions for the use of this camera. Incorrect installation

More information

POCKET MULTIMETER Model No: MM18

POCKET MULTIMETER Model No: MM18 INSTRUCTIONS FOR: POCKET MULTIMETER Model No: MM18 Thank you for purchasing a Sealey product. Manufactured to a high standard this product will, if used according to these instructions and properly maintained,

More information

HITACHI. EH-150 series PLC EH-RTD8 Resistance Temperature Detective input module Instruction manual. Safety precautions

HITACHI. EH-150 series PLC EH-RTD8 Resistance Temperature Detective input module Instruction manual. Safety precautions HITACHI EH-150 series PLC Resistance Temperature Detective input module Instruction manual Thank you for purchasing a Hitachi Programmable Logic Controller. To operate it safely, please read this instruction

More information

Installation and Operation Back-UPS BR1000G-IN / BR1500G-IN

Installation and Operation Back-UPS BR1000G-IN / BR1500G-IN Installation and Operation Back-UPS BR1000G-IN / BR1500G-IN Important Safety Information Read the instructions carefully to become familiar with the equipment before trying to install, operate, service

More information

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9310/9350/9510/9550/9500 QUICK START GUIDE

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9310/9350/9510/9550/9500 QUICK START GUIDE AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9310/9350/9510/9550/9500 QUICK START GUIDE Thank you for purchasing a TSI AeroTrak Portable Airborne Particle Counter (particle counter). This guide will

More information

Network Camera. Quick Guide DC-B1203X. Powered by

Network Camera. Quick Guide DC-B1203X. Powered by Network Camera Quick Guide DC-B1203X Powered by Safety Precautions English WARNING RISK OF ELECTRIC SHOCK DO NOT OPEN WARNING: TO REDUCE THE RISK OF ELECTRIC SHOCK, DO NOT REMOVE COVER (OR BACK). NO USER-SERVICEABLE

More information

Manual Version: V1.00. Video Decoder Quick Guide

Manual Version: V1.00. Video Decoder Quick Guide Manual Version: V1.00 Video Decoder Quick Guide Thank you for purchasing our product. If there are any questions, or requests, please do not hesitate to contact the dealer. Copyright Copyright 2016 Zhejiang

More information

4P M. Wireless LAN Connecting Adapter <BRP069A45> Installation Manual

4P M. Wireless LAN Connecting Adapter <BRP069A45> Installation Manual 4P359542-3M Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com For details

More information

4P S. Wireless LAN Connecting Adapter <BRP069A41> Installation Manual

4P S. Wireless LAN Connecting Adapter <BRP069A41> Installation Manual 4P359542-1S Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com For details

More information

4P Wireless LAN Connecting Adapter <BRP069B41> <BRP069B42> <BRP069B43> <BRP069B44> <BRP069B45> Installation Manual

4P Wireless LAN Connecting Adapter <BRP069B41> <BRP069B42> <BRP069B43> <BRP069B44> <BRP069B45> Installation Manual 4P481234-1 Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com

More information

Line Interactive 1000VA/1400VA/2000VA Uninterruptible Power System

Line Interactive 1000VA/1400VA/2000VA Uninterruptible Power System USER MANUAL Line Interactive 1000VA/1400VA/2000VA Uninterruptible Power System 614-06762-00 IMPORTANT SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS This manual contains important instructions for Line Interactive

More information

CS2 DMX Distribution Splitter

CS2 DMX Distribution Splitter CS2 DMX Distribution Splitter User Manual Order code: BOTE72 Safety advice WARNING FOR YOUR OWN SAFETY, PLEASE READ THIS USER MANUAL CAREFULLY BEFORE YOUR INITIAL START-UP! Before your initial start-up,

More information

TA12 full-rangespeaker. user manual

TA12 full-rangespeaker. user manual TA12 full-rangespeaker user manual Musikhaus Thomann e.k. Treppendorf 30 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 email: info@thomann.de Internet: www.thomann.de 30.11.2011 Table of contents

More information

Emerson Network Power provides customers with technical support. Users may contact the nearest Emerson local sales office or service center.

Emerson Network Power provides customers with technical support. Users may contact the nearest Emerson local sales office or service center. Liebert PSA iton User Manual Version: V2.8 Revision date: November 14, 2005 Emerson Network Power provides customers with technical support. Users may contact the nearest Emerson local sales office or

More information

General Warranty. For more details, please refer to the user manual on the supplied CD, it can also be downloaded at

General Warranty. For more details, please refer to the user manual on the supplied CD, it can also be downloaded at General Warranty OWON warrants that the product will be free from defects in materials and workmanship for a period of 3 years (1 year for accessories) from the date of purchase of the product by the original

More information

Druck DPI 620 advanced modular calibrator

Druck DPI 620 advanced modular calibrator GE Sensing & Inspection Technologies A1.1 DPI 620: Channel 1 (CH1) Druck DPI 620 advanced modular calibrator safety and quick reference guide - K0454 Measure (M) / Source (S) / Power (P) a ±30 V (M) b

More information

UV Power Flood. User Manual. Order code: EQLED029

UV Power Flood. User Manual. Order code: EQLED029 UV Power Flood User Manual Order code: EQLED029 Safety advice WARNING FOR YOUR OWN SAFETY, PLEASE READ THIS USER MANUAL CARE- FULLY BEFORE YOUR INITIAL START-UP! Before your initial start-up, please make

More information

General Warranty. For more details, please refer to the user manual on the supplied CD, it can also be downloaded at

General Warranty. For more details, please refer to the user manual on the supplied CD, it can also be downloaded at General Warranty Lilliput warrants that the product will be free from defects in materials and workmanship for a period of 3 years (1 year for accessories) from the date of purchase of the product by the

More information

TPC25 active crossover. user manual

TPC25 active crossover. user manual TPC25 active crossover user manual Musikhaus Thomann Thomann GmbH Hans-Thomann-Straße 1 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de 25.09.2017,

More information

PAR56 10MM UV LED PAR. user manual

PAR56 10MM UV LED PAR. user manual PAR56 10MM UV LED PAR user manual Musikhaus Thomann e.k. Treppendorf 30 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de 03.12.2012 Table of contents

More information

Operating and Installation Manual. EASYLAB Expansion modules Type EM-TRF for 230 V AC mains voltage

Operating and Installation Manual. EASYLAB Expansion modules Type EM-TRF for 230 V AC mains voltage Operating and Installation Manual EASYLAB Expansion modules Type EM-TRF for 230 V AC mains voltage Type EM-TRF-USV for 230 V AC mains voltage; provides uninterruptible power supply (UPS) The art of handling

More information

TA18 active subwoofer. user manual

TA18 active subwoofer. user manual TA18 active subwoofer user manual Musikhaus Thomann e.k. Treppendorf 30 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 email: info@thomann.de Internet: www.thomann.de 30.11.2011 Table of contents

More information

Obtaining Documentation and Submitting a Service Request, page xvii Safety Warnings, page xvii Safety Guidelines, page xx

Obtaining Documentation and Submitting a Service Request, page xvii Safety Warnings, page xvii Safety Guidelines, page xx Preface Obtaining Documentation and Submitting a Service Request, page xvii Safety s, page xvii Safety Guidelines, page xx Obtaining Documentation and Submitting a Service Request For information on obtaining

More information

T1 Patient Monitor. Service Manual

T1 Patient Monitor. Service Manual T1 Patient Monitor Service Manual Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this product

More information

Blitzer Strobe. User Manual. Order code: EQLED366

Blitzer Strobe. User Manual. Order code: EQLED366 Blitzer Strobe User Manual Order code: EQLED366 Safety advice WARNING FOR YOUR OWN SAFETY, PLEASE READ THIS USER MANUAL CAREFULLY BEFORE YOUR INITIAL START-UP! Before your initial start-up, please make

More information

CM E FUJITSU SEMICONDUCTOR CONTROLLER MANUAL F 2 MC-16L/16LX EMULATION POD MB HARDWARE MANUAL

CM E FUJITSU SEMICONDUCTOR CONTROLLER MANUAL F 2 MC-16L/16LX EMULATION POD MB HARDWARE MANUAL FUJITSU SEMICONDUCTOR CONTROLLER MANUAL CM42-00411-2E F 2 MC-16L/16LX EMULATION POD MB2145-507 HARDWARE MANUAL F 2 MC-16L/16LX EMULATION POD MB2145-507 HARDWARE MANUAL FUJITSU LIMITED PREFACE Using the

More information

USER MANUAL. Uninterruptible Power Supply Line-interactive VCL Series UPS VA. GE Critical Power

USER MANUAL. Uninterruptible Power Supply Line-interactive VCL Series UPS VA. GE Critical Power Critical Power USER MANUAL Uninterruptible Power Supply Line-interactive VCL Series UPS 400 600 800 1000 1500 VA GE Consumer & Industrial SA General Electric Company CH 6595 Riazzino (Locarno) Switzerland

More information

RIGOL. Quick Guide. DP800A Series Programmable Linear DC Power Supply. Aug RIGOL Technologies, Inc.

RIGOL. Quick Guide. DP800A Series Programmable Linear DC Power Supply. Aug RIGOL Technologies, Inc. Quick Guide DP800A Series Programmable Linear DC Power Supply Aug. 2013 RIGOL Technologies, Inc. Guaranty and Declaration Copyright 2012 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information

More information

4P R. Wireless LAN Connecting Adapter <BRP069A81> Installation Manual

4P R. Wireless LAN Connecting Adapter <BRP069A81> Installation Manual 4P359542-4R Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com For details

More information

Kinetix 300 Memory Module Programmer

Kinetix 300 Memory Module Programmer Kinetix 300 Memory Module Programmer Catalog Number 2097-PGMR Topic About the Memory Module Programmer 1 Parts List 3 Batteries Operation 4 Using Memory Module Programmer 6 Switch On/Off Memory Module

More information

Always there to help you. Register your product and get support at BT5500. Question? Contact Philips.

Always there to help you. Register your product and get support at   BT5500. Question? Contact Philips. Always there to help you Register your product and get support at www.philips.com/support Question? Contact Philips BT5500 User manual Contents 1 Important 2 2 Your Bluetooth speaker 3 Introduction 3

More information

Installation Guide. QBox-V6. Standalone/Spare V6 SDI QBox. Standalone/Spare V6 SDI QBox. Part No. A

Installation Guide. QBox-V6. Standalone/Spare V6 SDI QBox. Standalone/Spare V6 SDI QBox. Part No. A Installation Guide Standalone/Spare V6 SDI QBox QBox-V6 Standalone/Spare V6 SDI QBox Part No. A9009-0004 EN www.autocue.com Copyright 2017 All rights reserved. Original Instructions: English All rights

More information

Pyrit 212 Sub A active subwoofer. user manual

Pyrit 212 Sub A active subwoofer. user manual Pyrit 212 Sub A active subwoofer user manual Musikhaus Thomann e.k. Treppendorf 30 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de 22.01.2013 Table

More information

H Series PLC. ! : Indicates Compulsion. EH-150 Analog input module EH-AXH8M Instruction manual. Safety precautions DANGER CAUTION COMPULSION

H Series PLC. ! : Indicates Compulsion. EH-150 Analog input module EH-AXH8M Instruction manual. Safety precautions DANGER CAUTION COMPULSION H Series PLC EH-150 Analog input module EH-AXH8M Instruction manual Thank you for purchasing a Hitachi Programmable Logic Controller. To operate it safely, please read this instruction manual and all the

More information

User Manual AIMB-C200. Economical Embedded Chassis for Mini-ITX Motherboard

User Manual AIMB-C200. Economical Embedded Chassis for Mini-ITX Motherboard User Manual AIMB-C200 Economical Embedded Chassis for Mini-ITX Motherboard Copyright The documentation and the software included with this product are copyrighted 2010 by Advantech Co., Ltd. All rights

More information

D610H, D610H MKII, D610S, D610T dimmer pack. user manual

D610H, D610H MKII, D610S, D610T dimmer pack. user manual D610H, D610H MKII, D610S, D610T dimmer pack user manual Musikhaus Thomann Thomann GmbH Hans-Thomann-Straße 1 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de

More information

Operating instructions. Switching amplifier DN0210 DN / / 2015

Operating instructions. Switching amplifier DN0210 DN / / 2015 Operating instructions Switching amplifier DN0210 DN0220 UK 80011079 / 00 01 / 2015 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5

More information

MOBILE CONNECTOR - GEN 2 OWNER'S MANUAL

MOBILE CONNECTOR - GEN 2 OWNER'S MANUAL MOBILE CONNECTOR - GEN 2 OWNER'S MANUAL UNITED STATES Contents Safety Information... 2 Save These Important Safety Instructions... 2 Warnings...2 Cautions...3 General Information... 4 Mobile Connector

More information

BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL

BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL August 2016 This product is designed and manufactured by: ASL Intercom B.V. Zonnebaan 42 3542 EG Utrecht The Netherlands Phone: +31 (0)30 2411901 Fax: +31

More information

Achat 112 MA Achat 112 MA WH full-range speaker. user manual

Achat 112 MA Achat 112 MA WH full-range speaker. user manual Achat 112 MA Achat 112 MA WH full-range speaker user manual Musikhaus Thomann Thomann GmbH Hans-Thomann-Straße 1 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet:

More information

DMX Splitter 4 MKIII 19" DMX splitter. user manual

DMX Splitter 4 MKIII 19 DMX splitter. user manual DMX Splitter 4 MKIII 19" DMX splitter user manual Musikhaus Thomann Thomann GmbH Hans-Thomann-Straße 1 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de

More information

PowerMust Office Uninterruptible Power System

PowerMust Office Uninterruptible Power System USER MANUAL E PowerMust Office Uninterruptible Power System 614-05737-05 28-2PRO000001 IMPORTANT SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS This manual contains important instructions for Models PowerMust

More information

User Manual IDS-3106 Series. 6.5" Industrial Open Frame Monitor

User Manual IDS-3106 Series. 6.5 Industrial Open Frame Monitor User Manual IDS-3106 Series 6.5" Industrial Open Frame Monitor Copyright The documentation and the software included with this product are copyrighted 2014 by Advantech Co., Ltd. All rights are reserved.

More information

Economy Single Channel Output DC Power Supply

Economy Single Channel Output DC Power Supply Economy Single Channel Output DC Power Supply User Manual www.owon.com.cn Feb. 2019 edition V1.1.0 Copyright LILLIPUT Company. All rights reserved. The LILLIPUT's products are under the protection of the

More information

TVAC25100 TVAC25110 User manual

TVAC25100 TVAC25110 User manual TVAC25100 TVAC25110 User manual Version 11/2010 Original user manual. Keep for future use. 12 Introduction Dear Customer, Thank you for purchasing this product. This product meets the requirements of the

More information

HP4 headphone amplifier. user manual

HP4 headphone amplifier. user manual HP4 headphone amplifier user manual Musikhaus Thomann Thomann GmbH Hans-Thomann-Straße 1 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de 03.11.2015,

More information

Gobo Projector XP 80W

Gobo Projector XP 80W Gobo Projector XP 80W User Manual Order code: EQLED084 Safety advice WARNING FOR YOUR OWN SAFETY, PLEASE READ THIS USER MANUAL CARE- FULLY BEFORE YOUR INITIAL START-UP! Before your initial start-up, please

More information

BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL

BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL Issue 2011 ASL Intercom BV DESIGNED & MANUFACTURED BY: ASL Intercom B.V. Zonnebaan 42 3542 EG Utrecht The Netherlands Tel: +31 (0)30 2411901 Fax: +31 (0)30

More information

USER MANUAL. Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX310

USER MANUAL. Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX310 USER MANUAL Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX310 Introduction Congratulations on your purchase of the Extech EX310 MultiMeter. The EX310 offers AC/DC Voltage, AC/DC Current,

More information

Multi-touch Touch-Screen. Installation Guide

Multi-touch Touch-Screen. Installation Guide Multi-touch Touch-Screen Installation Guide Table of Contents Read before use...3 Safety and troubleshooting information...4 Installing Software...8 Connecting Cables... 13 Calibrating the Touch Screen...

More information

Wireless LAN Connecting Adapter [BRP072B42] Installation Manual

Wireless LAN Connecting Adapter [BRP072B42] Installation Manual Wireless LAN Connecting Adapter [BRP072B42] Installation Manual Wireless LAN Connecting Adapter [BRP072B42] Installation Manual Website: http://www.daikinthai.com/wlan3/modbus For details on specifications,

More information

ADC7520 SERIES. 1600W Battery Chargers and Power Supplies

ADC7520 SERIES. 1600W Battery Chargers and Power Supplies ADC7520 SERIES 1600W Battery Chargers and Power Supplies Wide output adjustment range 0 72VDC Analog control by external 0-5VDC voltage Temp.comp charging, sense as on option Power fail relay alarm Master-Slave

More information

ENVIRONMENTAL PROTECTION

ENVIRONMENTAL PROTECTION ENVIRONMENTAL PROTECTION Do not dispose of this product with the normal household waste at the end of its life cycle. Return it to a collection point for the recycling of electrical and electronic devices.

More information

*M C* Philips A1 and A3 Patient Monitors M3926A M3923A M3924A M3927A M3928A M3929A. Service Guide. Part Number M C

*M C* Philips A1 and A3 Patient Monitors M3926A M3923A M3924A M3927A M3928A M3929A. Service Guide. Part Number M C Philips A1 and A3 Patient Monitors A1 Models: A3 Models: M3921A M3922A M3923A M3924A M3926A M3927A M3928A M3929A Service Guide Part Number M3921-9000C Printed in the U.S.A. February 2002 First Edition

More information

Achat 206 passive full-range speaker. user manual

Achat 206 passive full-range speaker. user manual Achat 206 passive full-range speaker user manual Musikhaus Thomann Thomann GmbH Hans-Thomann-Straße 1 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de

More information

FengMi Wemax One Laser Projection TV

FengMi Wemax One Laser Projection TV FengMi Wemax One Laser Projection TV User`s Manual About electrical ground Transportation Use this device only with a compulsory grounding condition. It is recommended that you use the original packaging

More information

Product Documentation

Product Documentation Product Documentation Emanate PowerPath TM TempTag PPT-300 Date: April 12, 2016 Document Number: PPT200-001 R1.4 Emanate Wireless, Inc. 11145 Windsor Rd. Ijamsville, MD 21754 Telephone: 844-EMANATE Email:

More information

Part No. Z , IA Mar OPERATION MANUAL. High Voltage Digitalmeter A

Part No. Z , IA Mar OPERATION MANUAL. High Voltage Digitalmeter A Part No. Z1-109-820, IA001704 Mar. 2011 OPERATION MANUAL High Voltage Digitalmeter 149-10A Use of Operation Manual Please read through and understand this Operation Manual before operating the product.

More information

OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS

OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS GB Contents Page 1 Introduction 4 2 Safety 5 3 Use 6 4 Checking on Delivered Equipment 6 5 General Specification and Dimensions 7 6 Positioning 10 7 Operating

More information

Part No. Z , IA Jul OPERATION MANUAL. High Voltage Digitalmeter A

Part No. Z , IA Jul OPERATION MANUAL. High Voltage Digitalmeter A Part No. Z1-109-820, IA001705 Jul. 2016 OPERATION MANUAL High Voltage Digitalmeter 149-10A Use of Operation Manual Please read through and understand this Operation Manual before operating the product.

More information

Model 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1

Model 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1 Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.keithley.com Model 2460-KIT Screw Terminal Connector Kit Description The Model 2460-KIT Screw Terminal Connector Kit

More information

PowerPlant Tuner tuner pedal with power distribution. user manual

PowerPlant Tuner tuner pedal with power distribution. user manual PowerPlant Tuner tuner pedal with power distribution user manual Musikhaus Thomann Thomann GmbH Hans-Thomann-Straße 1 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet:

More information

Flexoled FTP3HP RGB LED DMX Driver

Flexoled FTP3HP RGB LED DMX Driver Flexoled FTP3HP RGB LED DMX Driver User Manual + 12-24V DMX INPUT DMX OUTPUT DC IN (12A MAX) DMX ADDRESS & DIMMER SETTINGS SWITCH 10 11 12 DMX MODE DIMMING MODE BUILT-IN MODE LOAD 4A PER CHANNEL (12A MAX)

More information

RGB Power Batten. User Manual. Order code: EQLED032

RGB Power Batten. User Manual. Order code: EQLED032 RGB Power Batten User Manual Order code: EQLED032 Safety advice WARNING FOR YOUR OWN SAFETY, PLEASE READ THIS USER MANUAL CARE- FULLY BEFORE YOUR INITIAL START-UP! Before your initial start-up, please

More information

USER GUIDE. AXIS T8120 Midspan 15 W 1-port ENGLISH

USER GUIDE. AXIS T8120 Midspan 15 W 1-port ENGLISH USER GUIDE AXIS T8120 Midspan 15 W 1-port ENGLISH Legal Considerations Video and audio surveillance can be prohibited by laws that vary from country to country. Check the laws in your local region before

More information

Startup Guide C01

Startup Guide C01 Startup Guide 4012988-00 C01 Startup Guide English Where to Find Information........................ 2 Safety Instructions.............................. 4 Important Safety Instructions...........................

More information

SINAMICS G130. Terminal Module 150 (TM150) Operating Instructions 03/2013 SINAMICS

SINAMICS G130. Terminal Module 150 (TM150) Operating Instructions 03/2013 SINAMICS SINAMICS G130 Operating Instructions 03/2013 SINAMICS s Safety information 1 General information 2 SINAMICS SINAMICS G130 Mechanical installation 3 Electrical installation 4 Technical specifications 5

More information

EC Series. Safety Guide

EC Series. Safety Guide EC Series Safety Guide R5906018/02 01/07/2017 Barco Inc, Image Processing 3078 Prospect Park Drive, Rancho Cordova, CA, 95670, USA Phone: +1 916 859-2500 Fax: +1 916 859-2515 Support: www.barco.com/en/support

More information

MA100 full-range speaker. user manual

MA100 full-range speaker. user manual MA100 full-range speaker user manual Musikhaus Thomann Thomann GmbH Hans-Thomann-Straße 1 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de 10.05.2016,

More information

LED Colour Flood 10W / 30W RGB LED floodlight. user manual

LED Colour Flood 10W / 30W RGB LED floodlight. user manual LED Colour Flood 10W / 30W RGB LED floodlight user manual Musikhaus Thomann e.k. Treppendorf 30 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de

More information

Manual Version: V1.00. Video Decoder User Manual

Manual Version: V1.00. Video Decoder User Manual Manual Version: V1.00 Video Decoder User Manual Thank you for purchasing our product. If there are any questions, or requests, please do not hesitate to contact the dealer. Copyright Copyright 2016 Zhejiang

More information

PF LED Profile. User Manual

PF LED Profile. User Manual PF LED Profile User Manual Order codes: ELUM094-100W 3200K WW Version ELUM095-100W 6000K CW Version ELUM096-150W 3200K WW Version ELUM097-150W 6000K CW Version Safety advice WARNING FOR YOUR OWN SAFETY,

More information

Crossfire. User Manual. Order code: EQLED061

Crossfire. User Manual. Order code: EQLED061 Crossfire User Manual Order code: EQLED061 Safety advice WARNING FOR YOUR OWN SAFETY, PLEASE READ THIS USER MANUAL CAREFULLY BEFORE YOUR INITIAL START-UP! Before your initial start-up, please make sure

More information

INSPECTION TOOL. Inspection Video Camera with Color LCD Monitor. EU Environmental Protection. Model: 8802LE,8803LE. Model: 8802LE,8803LE

INSPECTION TOOL. Inspection Video Camera with Color LCD Monitor. EU Environmental Protection. Model: 8802LE,8803LE. Model: 8802LE,8803LE INSPECTION TOOL Inspection Video Camera with Color LCD Monitor Model: 8802LE,8803LE EU Environmental Protection Waste electrical products should not be disposed of with household waste. Please recycle

More information

Always there to help you. Register your product and get support at BT5580 BT5880. Question? Contact Philips.

Always there to help you. Register your product and get support at  BT5580 BT5880. Question? Contact Philips. Always there to help you Register your product and get support at www.philips.com/support Question? Contact Philips BT5580 BT5880 User manual Contents 1 Important 2 2 Your Bluetooth speaker 3 Introduction

More information

The following symbols are used to show dangerous operation or handling. Make sure you understand them before reading the guide.

The following symbols are used to show dangerous operation or handling. Make sure you understand them before reading the guide. Safety Instructions Before use Thank you very much for purchasing this product. This product is an interface box called "Connection & Control Box" for EPSON short throw projectors. For your safety, read

More information

xbrick Full-Colour 16 3W LED floodlight user manual

xbrick Full-Colour 16 3W LED floodlight user manual xbrick Full-Colour 16 3W LED floodlight user manual Musikhaus Thomann e.k. Treppendorf 30 96138 Burgebrach Germany Telephone: +49 (9546) 9223-0 email: info@thomann.de Internet: www.thomann.de 15.11.2011

More information

User Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at

User Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at User Manual 400Amp AC Clamp Meter + NCV Model MA430 Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of this Extech MA430 Clamp Meter. This

More information

INSTALLATION MANUAL. ST-BTWD650IR2812 B or W Weatherproof Day/Night Infrared Color Camera

INSTALLATION MANUAL. ST-BTWD650IR2812 B or W Weatherproof Day/Night Infrared Color Camera INSTALLATION MANUAL ST-BTWD650IR2812 B or W Weatherproof Day/Night Infrared Color Camera PACKAGE CONTENTS This package contains: One ST-BTWD650IR2812 camera with Black or White housing One extension tube

More information

Always there to help you. Register your product and get support at DS3480. Question? Contact Philips.

Always there to help you. Register your product and get support at   DS3480. Question? Contact Philips. Always there to help you Register your product and get support at www.philips.com/welcome Question? Contact Philips DS3480 User manual Contents 1 Important 2 Safety 2 Notice 2 2 Your Docking Speaker 3

More information

ZF LED Zoom Fresnel. User Manual

ZF LED Zoom Fresnel. User Manual ZF LED Zoom Fresnel User Manual Order codes: ELUM090-100W 3200K WW Version ELUM091-100W 6000K CW Version ELUM092-150W 3200K WW Version ELUM093-150W 6000K CW Version Safety advice WARNING FOR YOUR OWN SAFETY,

More information

FT-20 Fire Training Fog Machine User Manual English

FT-20 Fire Training Fog Machine User Manual English FT-20 Fire Training Fog Machine User Manual English 2018 Antari Lighting and Effects Ltd. 1 User Manual - English Safety Information Please read the following safety information carefully before operating

More information

Installation Manual. 65 Interactive LED/LCD. Model: HILF65101 (64.56 )

Installation Manual. 65 Interactive LED/LCD. Model: HILF65101 (64.56 ) Installation Manual 65 (64.56 ) Model: HILF65101 65 Interactive LED/LCD QUICK SETUP GUIDE For further information, see the user manual. Please contact directly if you have questions on the use of the touch

More information

USP-070-B08 USP-104-B10, USP-104-M10 USP-156-B10

USP-070-B08 USP-104-B10, USP-104-M10 USP-156-B10 UniStream HMI Panel Installation Guide USP-070-B10, USP-070-B08 USP-104-B10, USP-104-M10 USP-156-B10 Unitronics UniStream platform comprises control devices that provide robust, flexible solutions for

More information

SB 115A active subwoofer. user manual

SB 115A active subwoofer. user manual SB 115A active subwoofer user manual Musikhaus Thomann e.k. Treppendorf 30 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de 27.09.2012 Table of contents

More information

icore Kiosk system Installation Guide

icore Kiosk system Installation Guide icore Kiosk system Installation Guide The reproduction, transmission or use of this document or its contents is not permitted without express authority. Offenders will be liable for damages. All rights,

More information

DS1000B Series Digital Oscilloscope

DS1000B Series Digital Oscilloscope Quick Guide RIGOL Publication number QGA04114-1110 Feb. 2014 DS1000B Series Digital Oscilloscope DS1074B, DS1104B, DS1204B All Rights Reserved Copyright All Rights Reserved. RIGOL products are protected

More information

M910q. User Guide and Hardware Maintenance Manual. Overview. Computer locks. Specifications. Replacing FRUs. Replacing CRUs.

M910q. User Guide and Hardware Maintenance Manual. Overview. Computer locks. Specifications. Replacing FRUs. Replacing CRUs. M90q User Guide and Hardware Maintenance Manual Machine Type (MT): 0MU, 0MV, 0MW, 0MX, 0QN Energy Star MT: 0MU, 0MV, 0MW, 0MX, 0QN Overview Computer locks Specifications Locations of indicators, connectors,

More information

D1210H dimmer pack. user manual

D1210H dimmer pack. user manual D1210H dimmer pack user manual Musikhaus Thomann Thomann GmbH Hans-Thomann-Straße 1 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de 13.08.2015,

More information

Always there to help you. Register your product and get support at AJ6200DBZ. Question? Contact Philips.

Always there to help you. Register your product and get support at   AJ6200DBZ. Question? Contact Philips. Always there to help you Register your product and get support at www.philips.com/support Question? Contact Philips AJ6200DBZ User manual Contents 1 Important 2 2 Your clock radio 3 Introduction 3 What's

More information

DS-1H05 Ethernet-over-Coax Extender. User Manual

DS-1H05 Ethernet-over-Coax Extender. User Manual DS-1H05 Ethernet-over-Coax Extender User Manual Thank you for purchasing our product. If there is any question or request, please do not hesitate to contact dealer. This manual is applicable to DS-1H05-T,

More information

Fixed Wireless Phone. User Manual

Fixed Wireless Phone. User Manual Fixed Wireless Phone User Manual V1.0 Content 1. Introduction... 1 2. Important Information... 2 3. Dos & Don ts... 2 4. Views... 4 5 Key board Introduction... 4 6. Installation... 5 6.1 SIM card installation...

More information