1 8700 SERIES DIGITAL METER CALIBRATION FIXTURE OPERATION MANUAL narda Safety Test Solutions an L3 communications company P/N

2 narda Safety Test Solutions an L3 communications company USA: 435 Moreland Road Hauppauge, NY Tel Fax GERMANY: Sandwiesenstr 7 D Pfullingen Tel Fax support@narda-sts.de

3 Introduction The P/N calibration fixture is comprised of a cable with a connector that replicates a Narda 8700D series probe, a channel selector switch, input jacks for a separate power supply, and test jacks. A fuse and a zener diode are used to protect the meter from the accidental application of incorrect or reverse polarity voltages. This fixture is used to check that calibration of Narda 8700 Series Meters and to provide simple access to a meter s operating voltages. Theory of Operation The Narda 8700 Series of radio frequency radiation instruments is comprised of meters, probes, cables and fiber optic links that can be used in many different configurations to measure electric and magnetic fields. The system is modular with all components individually calibrated. The amplifiers in the handle of every prove are adjusted during the calibration process to provide the same DC output voltage at full scale. Every meter has two input channels a linear input and a nonlinear or compensated input. Full scale for linear output probes is equal to volts DC. Certain probe models have a somewhat non-linear response which is compensated for within the meter. Ultra-wideband probes those with a frequency response of 300 khz to GHz contain two different sensors. These probes utilize a twochannel amplifier one with a linear output and the other with an non-linear output. Application This calibration fixture can be used to check the calibration of all Narda 8700 series digital meters. The list of currently applicable meter models are: Models 8718 and 8718B Model 8715 Model Series 3

4 These digital meters do not have potentiometers that require periodic adjustment. The purpose of this calibration fixture is to provide a convenient means to apply a precise DC voltage to each of the meter s two input channels to confirm that the meter is functioning properly. A specific input voltage on each channel should provide a specific indicated measurement which is dependent on the probe model that the meter is set for. The calibration fixture also provides access to the operating voltages within the meter. These voltages can be checked to provide another indication that the meter is operating properly. Equipment Required The only equipment required in addition to the meter(s) to be checked are: An accurate, stable low voltage power supply. An accurate DC voltmeter Model cable, as required to connect the meter to the test fixture. Using the Calibration Fixture To use the calibration fixture: 1. Turn on the power supply and adjust it to approximately zero volts. Turn the power supply off and connect it to the two terminals on the top of the fixture. Make sure that the polarity matches the label for REF GND (reference ground) and V. 2. Connect the calibration fixture to the meter. The connector on the fixture plugs directly into meter Models 8712, 8715 and 8718B. A probe extension cable (Model ) is used to connect to the Model 8718 Meter and the 8717 Series Meters. 3. Set the toggle switch on the fixture to the CH A (channel A) position. 4. Turn the meter on. 4

5 5. On the Models 8718, 8718B and 8717 Series Meters set the meter to the model 8741D. On the Models 8712 and 8715 Meters, set the meter to the 20 mw/cm 2 range and for electric field probes. Note that the Model 8741D probe contains two sensors and uses a dual channel amplifier. NOTE The meter is now ready to be zeroed. However, the calibration fixture does not exactly replicate the probe and this test configuration will fail the normal criteria for zeroing a probe. Therefore, it is necessary to force a zero. This capability has been built into all the Narda 8700 Series Digital Meters. 6. Turn on the power supply and adjust the voltage at the terminals of the calibration fixture to V ±5 mv. 7. Force a zero on the meter (see Forcing a Zero on page 6). 8. Readjust the power supply to V ±5 mv. 9. Read the field strength on the meter. It should indicate 10 mw/cm 2 ±0.3 mw/cm 2. This is the linear input channel. 10. Move the toggle switch on the fixture to the CH B (channel B) position. 11. Read the field strength on the meter. It should indicate 12.5 mw/cm 2 ±0.375 mw/cm 2. This is the non-linear or compensated input channel. Updating the Calibration Date The meter is accurate within specifications providing that indicated field strength in steps 10 and 12 above are within the tolerances indicated. If the meter is operating properly with specification, the calibration date can be updated to the date it was checked using this fixture. 5

6 If the meter is not operating within specification, it should be returned to Narda for service. To update the calibration date on the Models 8712 and 8715 Meters: 1. Remove the old calibration label on the bottom of the meter. 2. Adhere a new calibration label to the bottom of the meter. Write in the calibration date. To update the calibration date on the models 8718, 8718B and 8717 series meters, it must be done electronically using the interface software supplied with the meter: 1. Connect the meter to a personal computer that has the interface software installed and establish communications using the procedures described in the meter operations manual. 2. Select Read Meter Data from the pull down menu under Diagnostics. 3. A screen will appear that states Read Any Meter Data. At this point type >CAL then press Enter on the computer. 4. A prompt will appear that states Type Cal Month. Type in a two digit number representing the month of calibration and press Enter on the computer. 5. A prompt will appear that states Type Cal Day. Type in a two digit number representing the day of calibration and press Enter on the computer. 6. A prompt will appear that states Type Cal Year. Type in a two digit number representing the year of calibration and press Enter on the computer. 7. A prompt will appear that indicates the calibration date that you have entered and gives three choices: Yes (Y) confirms that the date is correct and places the date in the meter s memory. No (N) is used to enter a different date. 6

7 Quit (Q) is used to quit the calibration update portion of the interface software without entering a date. Forcing a Zero To force a zero on the Models 8712 and 8715 Meters: 1. Once the probe full scale measurement range and the field type have been entered and the word ZERO is flashing on the display, press the Auto Zero key as in normal operation 2. Within five seconds after pressing the Auto Zero key, press the Range key to force a zero. 3. The meter should almost immediately indicate that it has been zeroed. The ZERO indicator will no longer be flashing and the meter will be reading either zero or a very low level. To force a zero on the Models 8718, 8718B and 8717 Series Meters 1. Once the probe model has been entered and the screen indicates that the meter is ready to be zeroed, press the Auto Zero key as in normal operation. 2. Numbers will be displayed below the ZEROING prompt several seconds after pressing the Enter key. Press the 8 key for about two seconds immediately after these numbers appear. 3. The meter should indicate that it has been zeroed within about five seconds. Checking Operating Voltages The calibration fixture provides access to the operating voltages within the meter. These voltages can be checked to provide an increased level of confidence that the meter operation is completely normal. 7

8 The four operating voltages that can be checked are available via test jacks on the sides of the calibration fixture. These four voltages and the acceptable operating ranges are: Description Range +V os 0.0 V ±0.5 mv to +2.4 V ±0.5 mv -V os 0.0 V ±0.5 mv to -1.4 V ±0.5 mv +5 V V ±0.25 V -5 V V ±0.25 V To measure the operating voltages of the Narda 8700 Series digital meters: 1. The ±V os voltages are measured while the meter is zeroing. Connect a Digital Volt Meter (DVM) between the +V os jack and ground and set the range for millivolts. Turn the 8700 Series Meter off and on. Select a Model 8741 Probe (or a 20 mw/cm 2, E-field probe) and begin zeroing the meter. Observe and record the maximum and minimum voltages. 2. Connect a Digital Volt Meter (DVM) between the V os jack and ground and set the range for millivolts. Turn the 8700 Series Meter off and on. Select a Model 8741 Probe (or a 20 mw/cm 2, E-field probe) and begin zeroing the meter. Observe and record the maximum and minimum voltages. 3. Measure and record the +5 V and 5 V power supply voltages. narda Safety Test Solutions an L3 communications company USA: 435 Moreland Road Hauppauge, NY Tel Fax NardaSTS@L-3COM.com GERMANY: Sandwiesenstr 7 D Pfullingen Tel Fax support@narda-sts.de 8

Measuring static magnetic fields

Measuring static magnetic fields 3-AXIS HALL MAGNETOMETER THM1176 Wi A Complementary Product From Measuring static magnetic fields ranging from low fields up to 20 Tesla Non-directional measurement using an isotropic 3-axis HALL probe

More information

NBM-520 Narda Broadband Field Meter

NBM-520 Narda Broadband Field Meter NBM-520 Narda Broadband Field Meter Operating Manual Narda Safety Test Solutions GmbH Sandwiesenstraße 7 72793 Pfullingen, Germany 2007 Order no.: 2403/98.21 Edition: 01/07.03, A... Previous edition: --

More information

ELECTRONICS MANUFACTURE-The In-Circuit Test sequence

ELECTRONICS MANUFACTURE-The In-Circuit Test sequence ELECTRONICS MANUFACTURE-The In-Circuit Test sequence In-Circuit Test comprises several sections, each consisting of a series of tests on individual devices. By testing devices individually, failures can

More information

NRA & IDA Utility Software

NRA & IDA Utility Software Rackmount Analyzer NRA Interference and Direction Analyzer IDA Technical Note 1105 NRA & IDA Utility Software In combination with the NRA and the IDA, Narda STS provides a set of free, easyto-use software.

More information

ETHOS Auto Ranging Digital Multimeter

ETHOS Auto Ranging Digital Multimeter ETHOS 5020 Auto Ranging Digital Multimeter 1 1. SAFETY INFORMATION SAFETY SYMBOLS Warning! Dangerous Voltage (Risk of electric shock). Caution! Refer to the user s manual before using this Meter. Double

More information

User-generated maps in IDA 2

User-generated maps in IDA 2 Interference and Direction Analyzer IDA 2 Technical Note TN105 User-generated maps in IDA 2 IDA 2 and GIS software allow use of aerial photos and users own maps It is often useful to be able to use your

More information

Model INSTRUCTION MANUAL DIGITAL MULTIMETER

Model INSTRUCTION MANUAL DIGITAL MULTIMETER Model 57040 INSTRUCTION MANUAL DIGITAL MULTIMETER SAFETY INFORMATION This multimeter has been designed according to IEC 1010 concerning electronic measuring instruments with an overvoltage category (CAT

More information

OPERATING AND SERVICE MANUAL SPECIAL SUPPLEMENT

OPERATING AND SERVICE MANUAL SPECIAL SUPPLEMENT OPERATING AND SERVICE MANUAL SPECIAL SUPPLEMENT 11683A OPTION H01 POWER METER RANGE CALIBRATOR 11683-90010 Model Number: 11683A Option H01 Part Number: 11683-90010 Date Printed: June 5, 2012 2 11683A Option

More information

Strain gauge Measuring Amplifier GSV-1A8. Instruction manual GSV-1A8, GSV-1A8USB, GSV-1A16USB

Strain gauge Measuring Amplifier GSV-1A8. Instruction manual GSV-1A8, GSV-1A8USB, GSV-1A16USB Strain gauge Measuring Amplifier GSV-A8 Instruction manual GSV-A8, GSV-A8USB, GSV-A6USB GSV-A8USB SubD5 (front side) GSV-A8USB M2 (front side) GSV-A6USB (rear side) GSV-A8USB K6D (front side) Version:

More information

-auto Power Off- T1/T2/T1-T2

-auto Power Off- T1/T2/T1-T2 -auto Power Off- T1/T2/T1-T2 CONTENTS 1. Safety Information...1 2. Product Overview...3 3. Recognize Meter..... 4 3.1 Component...4 3.2 Display...5 3.3 Keys Description...6 4. Setting Meter...8 4.1 SETUP

More information

Advanced Test Equipment Rentals ATEC (2832) Technical Note. Power Meter Accuracy Test

Advanced Test Equipment Rentals ATEC (2832) Technical Note. Power Meter Accuracy Test Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Technical Note Power Meter Accuracy Test To Calibrate EPM Series Power Meters to Option G12 or Option H12 Specifications

More information

Mounting DIGITAL DISPLAY

Mounting DIGITAL DISPLAY Mounting The Class digital display mounts in a. by. cutout. Overall area necessary for installation is. by.. Two 0.0 diameter holes are provided for mounting screws..0".0..0.900 Ø 0.0" () HOLES."." DIGITAL

More information

MODEL 801R REMOTE CONTROL MICROPHONE PREAMPLIFIER

MODEL 801R REMOTE CONTROL MICROPHONE PREAMPLIFIER OWNERS MANUAL MODEL 801R REMOTE CONTROL MICROPHONE PREAMPLIFIER Software version 1.4 PO Box 204 Boulder, CO 80306 tel:303.443.7454 fax: 303.444.4634 http://www.gracedesign.com Revision B January, 2000

More information

HAND-HELD THERMOCOUPLE THERMOMETER SERVICE MANUAL. CATALOG NUMBERS , and , and PROPRIETARY

HAND-HELD THERMOCOUPLE THERMOMETER SERVICE MANUAL. CATALOG NUMBERS , and , and PROPRIETARY HAND-HELD THERMOCOUPLE THERMOMETER SERVICE MANUAL CATALOG NUMBERS 600-1000, 600-1010 and 600-1020 91100-00, 91100-10 and 91100-20 PROPRIETARY Information contained in this manual is proprietary to COLE-PARMER

More information

OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER MAS830 MAS830L

OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER MAS830 MAS830L OPERATOR S INSTRUCTION MANUAL MAS830 MAS830L DIGITAL MULTIMETER SAFETY INFORMATION This multimeter has been designed according to IEC -1010 concerning electronic measuring instruments with an overvoltage

More information

Model R5005. Instruction Manual. True RMS Industrial Multimeter. reedinstruments. www. com

Model R5005. Instruction Manual. True RMS Industrial Multimeter. reedinstruments. www. com Model R5005 True RMS Industrial Multimeter Instruction Manual reedinstruments com Table of Contents Safety... 4 Features... 5 Specifications...5-6 Instrument Description...7-8 Operating Instructions...9-13

More information

NHT-310. Broadband EM Field Meter: DC 40 GHz. HAVSCo LTD. Key Features:

NHT-310. Broadband EM Field Meter: DC 40 GHz. HAVSCo LTD. Key Features: Key Features: Broadband measurement capability from DC up to 40 GHz utilizing a range of probes Interchangeable probes for measuring Electric, Magnetic and Electromagnetic fields Shell in aluminum alloy

More information

AN-730 APPLICATION NOTE One Technology Way P.O. Box 9106 Norwood, MA Tel: 781/ Fax: 781/

AN-730 APPLICATION NOTE One Technology Way P.O. Box 9106 Norwood, MA Tel: 781/ Fax: 781/ APPLICATION NOTE One Technology Way P.O. Box 9106 Norwood, MA 02062-9106 Tel: 781/329-4700 Fax: 781/326-8703 www.analog.com Theory of Operation for AD8555 Evaluation Hardware and Software by Reza Moghimi

More information

AC/DC Current Probe Model SL201

AC/DC Current Probe Model SL201 AC/DC Current Probe Model SL201 USER MANUAL I ZERO 1 V/A 1 mv/a OFF Limited Warranty The AC/DC Current Probe Model SL201 is warranted to the owner for a period of two years from the date of original purchase

More information

ICS Regent. Guarded Analog Output Modules. 4 to 20 ma (T3480) PD-6026

ICS Regent. Guarded Analog Output Modules. 4 to 20 ma (T3480) PD-6026 ICS Regent PD-6026 Guarded Analog Output Modules 4 to 20 ma (T3480) Issue 1, March, 06 Guarded analog output modules provide 4 to 20 ma current outputs for a maximum of six user loads per module. These

More information

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter S-14 Compact Digital Multimeter S-14 Compact Digital Multimeter SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation at this meter

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE HVA-100 HIGH VOLTAGE AMPLIFIER MODULE FOR EPMS SYSTEMS

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE HVA-100 HIGH VOLTAGE AMPLIFIER MODULE FOR EPMS SYSTEMS OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE HVA-100 HIGH VOLTAGE AMPLIFIER MODULE FOR EPMS SYSTEMS VERSION 1.7 npi 2015 npi electronic GmbH, Bauhofring 16, D-71732 Tamm, Germany Phone +49 (0)7141-9730230;

More information

SA SERIES Arlingate Lane, Columbus, Ohio (614)

SA SERIES Arlingate Lane, Columbus, Ohio (614) SA SERIES 2080 Arlingate Lane, Columbus, Ohio 43228 (614) 850-5000 Sensotec, Inc. SA Series Instruction Manual Sensotec Part Number: 008-0131 - 00 Rev. A: January 1996 Copyright Notice: 2080 Arlingate

More information

Micro Force V+F3 ver. 1.1

Micro Force V+F3 ver. 1.1 Micro Force V+F3 ver. 1.1 Preston Cinema Systems 1659 Eleventh Street Santa Monica, CA 90404 fig 1. V+F3 Controls LEMO receptacle Threaded well for bracket Camera Run Switch Zap Switch Zoom Sensitivity

More information

Service Manual. P GHz Differential Probe

Service Manual. P GHz Differential Probe Service Manual P7330 3.5 GHz Differential Probe 071-1057-00 Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you

More information

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter Operation Manual MODEL 2TX 2-wire Isolated ph/orp Transmitter 0 2TX CONTENTS INITIAL INSPECTION.....2 INTRODUCTION......2 ASSEMBLY...3 PREPARATION....4 CONNECTING THE ELECTRODE...4 CONNECTING THE TEMPERATURE

More information

rht+ RHTPlus CalComm Calibration Unit with Analogue Outputs

rht+ RHTPlus CalComm Calibration Unit with Analogue Outputs rht+ RHTPlus CalComm Calibration Unit with Analogue Outputs Skye In stru ments Ltd., 21Ddole En ter prise Park, Llandrindod Wells, Powys LD1 6DF UK. Tel: +44 (0) 1597 824811 Fax: +44 (0) 1597 824812 General

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE PTC-20 PELTIER TEMPERATURE CONTROL SYSTEM

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE PTC-20 PELTIER TEMPERATURE CONTROL SYSTEM OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE PTC-20 PELTIER TEMPERATURE CONTROL SYSTEM VERSION 2.11 npi 2002 npi electronic GmbH, Hauptstrasse 96, D-71732 Tamm, Germany Phone +49 (0)7141-601534;

More information

Strain gauge Measuring Amplifier GSV-1A8. Instruction manual GSV-1A8, GSV-1A8USB, GSV-1A16USB

Strain gauge Measuring Amplifier GSV-1A8. Instruction manual GSV-1A8, GSV-1A8USB, GSV-1A16USB Strain gauge Measuring Amplifier GSV-1A8 Instruction manual GSV-1A8, GSV-1A8USB, GSV-1A16USB GSV-1A8USB SubD1 (front side) GSV-1A8USB M12 (front side) GSV-1A16USB (rear side) GSV-1A8USB K6D (front side)

More information

Sheet Resistance Meter. Model SRM 232 MODE SELECT STORE SURFACE RESISTIVITY METER SRM-232 POWER. Operator s Guide

Sheet Resistance Meter. Model SRM 232 MODE SELECT STORE SURFACE RESISTIVITY METER SRM-232 POWER. Operator s Guide Sheet Resistance Meter Model SRM 232 MODE SELECT STORE SURFACE RESISTIVITY METER SRM-232 POWER Operator s Guide Manufactured by Guardian Mfg. For sales, contact: Bridge Technology Tel: 480-988-2256 Fax:

More information

AC/DC Current Probe Model SL206 USER MANUAL

AC/DC Current Probe Model SL206 USER MANUAL AC/DC Current Probe Model SL206 USER MANUAL Limited Warranty The AC/DC Current Probe Model SL206 is warranted to the owner for a period of 2 years from the date of original purchase against defects in

More information

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER SAFETY INFORMATION The following safety information must be observed to ensure maximum personal safety during the operation of this meter: This meter is

More information

IXIAN. Reads. Range / Accuracy. Calibration. 1, 2, 3 point remotely through PLC or directly on board.

IXIAN. Reads. Range / Accuracy. Calibration. 1, 2, 3 point remotely through PLC or directly on board. V 3.0 Revised 12/13/17 ph IXIAN Transmitter Reads ph Range.001 14.000 Accuracy Calibration Supported probes Temp probe +/ 0.002 1, 2, 3 point remotely through PLC or directly on board Any type & brand

More information

DSTCAL Software User Manual 09/16/09 WD1037 Rev E Revised 09/09/14

DSTCAL Software User Manual 09/16/09 WD1037 Rev E Revised 09/09/14 DSTCAL Software User Manual 09/16/09 WD1037 Rev E Revised 09/09/14 Wahl Instruments, Inc. 234 Old Weaverville Road Asheville, NC 28804 Toll Free 800-421-2853 Phone 828-658-3131 Fax 828-658-0728 www.palmerwahl.com

More information

User Manual. Single-Input Thermometers. with NIST-Traceable Calibration. Thermocouple Meter, Type J/K/T Input Model

User Manual. Single-Input Thermometers. with NIST-Traceable Calibration. Thermocouple Meter, Type J/K/T Input Model User Manual Single-Input Thermometers with NIST-Traceable Calibration Thermocouple Meter, Type J/K/T Input Model 20250-91 Thermistor Meter Model 20250-93 RTD Meter Model 20250-95 THE STANDARD IN PRECISION

More information

Smart Fieldmeter Digital Specification

Smart Fieldmeter Digital Specification Smart Fieldmeter Digital Specification Smart Fieldmeter Digital (SFD): Electromagnetic field monitor having flexible and expandable architecture, state of the art design with convenience and simplicity

More information

OPERATING INSTRUCTION

OPERATING INSTRUCTION OPERATING INSTRUCTION AUTORANGING MULTIMETER MAX Ω F C 10A MAX every 15 min. COM V SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation

More information

NBM-520 Narda Broadband Field Meter

NBM-520 Narda Broadband Field Meter NBM-520 Narda Broadband Field Meter Operating Manual Narda Safety Test Solutions GmbH Sandwiesenstraße 7 72793 Pfullingen, Germany Names and Logo are registered trademarks of Narda Safety Test Solutions

More information

CAN EGT-8 Amplifier. About:

CAN EGT-8 Amplifier. About: CAN EGT-8 Amplifier About: 1. The amplifier supports up to eight K-Type temperature probes. 2. EGT temperatures are sent to engine management ECU via CAN Bus. 3. One volt input that can be used to send

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TC-20 TEMPERATURE CONTROL SYSTEM

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TC-20 TEMPERATURE CONTROL SYSTEM OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TC-20 TEMPERATURE CONTROL SYSTEM VERSION 3.8 npi 2016 npi electronic GmbH, Bauhofring 16, D-71732 Tamm, Germany Phone +49 (0)7141-9730230; Fax: +49

More information

UV LIGHT METER UVC-254

UV LIGHT METER UVC-254 UV LIGHT METER UVC-254 Storage for the " UV sensor " UV sensor is with extremely precise structure. Once don't use it, be sure to store it in the dry environment. For example, put the whole sensor including

More information

LC8 CONTROLLER BOARD INSTALLATION AND MAINTENANCE MANUAL (IMM) p/n 52280T rev 11/08/05

LC8 CONTROLLER BOARD INSTALLATION AND MAINTENANCE MANUAL (IMM) p/n 52280T rev 11/08/05 LC8 CONTROLLER BOARD INSTALLATION AND MAINTENANCE MANUAL (IMM) p/n 52280T rev 11/08/05 3101 111th ST SW #F Everett Washington 98204 Telephone: 425-353-7552 Fax: 425-353-3353 1 LC8 CONTROLLER BOARD INSTALLATION

More information

User Manual. Dual-Input Data Logging Thermometers. with NIST-Traceable Calibration. Thermocouple Meter, Type J/K/T Input Model

User Manual. Dual-Input Data Logging Thermometers. with NIST-Traceable Calibration. Thermocouple Meter, Type J/K/T Input Model User Manual Dual-Input Data Logging Thermometers with NIST-Traceable Calibration Thermocouple Meter, Type J/K/T Input Model 20250-92 Thermistor Meter Model 20250-94 RTD Meter Model 20250-96 THE STANDARD

More information

OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER

OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER SAFETY INFORMATION This multimeter has been designed according to IEC 1010 concerning electronic measuring instruments with an overvoltage category (CATⅡ)

More information

CII: Current Input, Isolated

CII: Current Input, Isolated Safety 1. Application CII Driver Card 2. Installation 2.1 Services 2.2 Driver Card Installation 2.3 Driver Wiring 3. Configuration - Operation 3.1 Diagnostics 3.2 Calibration 4. Specifications Safety Electrical

More information

HI-POT TESTER. User s Manual

HI-POT TESTER. User s Manual HI-POT TESTER 7620 User s Manual Contents 1. Before Use... 1.1Electric Shock Avoidance... 4 1.2 Grounding... 4 1.3 AC Power Supply... 4 1.4 Connecting Test Leads... 4 1.5 Warm Up... 4 1.6 External Control...

More information

Explorer V1.20. Features

Explorer V1.20. Features V1.20 Multi-function USB I/O Expander and Controller Features Dual h-bridge 1.3A motor drive with PWM speed control 4.6V to 10.8V input range USB communication 4x digital inputs 2x analogue inputs 7x 100mA

More information

SCIENCEWORKSHOP 750 INTERFACE. Instruction Sheet for the PASCO Model CI Introduction

SCIENCEWORKSHOP 750 INTERFACE. Instruction Sheet for the PASCO Model CI Introduction Instruction Sheet for the PASCO Model CI-7500 012-06772A 4/98 $1.00 SCIENCEWORKSHOP 750 INTERFACE ventilation louvres POWER power light ScienceWorkshop 750 1 2 3 4 A B C DIGITAL CHANNELS ANALOG CHANNELS

More information

SOUND LEVEL METER DVM1351 INSTRUCTION MANUAL HANDLEIDING MANUEL D UTILISATEUR DVM1351 SOUND LEVEL METER 1

SOUND LEVEL METER DVM1351 INSTRUCTION MANUAL HANDLEIDING MANUEL D UTILISATEUR DVM1351 SOUND LEVEL METER 1 SOUND LEVEL METER DVM1351 INSTRUCTION MANUAL HANDLEIDING MANUEL D UTILISATEUR DVM1351 SOUND LEVEL METER 1 SAFETY INFORMATION l Read the following safety information carefully before attempting to operate

More information

MAINTENANCE MANUAL KTS 153 TEST SET

MAINTENANCE MANUAL KTS 153 TEST SET MAINTENANCE MANUAL KTS 153 TEST SET MANUAL NUMBER 006-15631-0007 REVISION 7 FEBRUARY, 2002 WARNING Prior to the export of this document, review for export license requirement is needed. COPYRIGHT NOTICE

More information

Isolated Wideband Voltage Input 3B40 / 3B41 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated Wideband Voltage Input 3B40 / 3B41 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM Isolated Wideband Voltage Input 3B40 / 3B41 FEATURES Interfaces, amplifies, protects& filters wide-bandwidth (h0 khz) single-channel analog voltage inputs. Module provides simultaneous precision voltage

More information

and Multiparameter Meters

and Multiparameter Meters and Multiparameter Meters 54650-23 ph Menu Options Table 1 How to Change ph Menu Options Setup Setup 1 Time Setup 2 Date Setup 3 Year Setup 4 Temperature Units Setup 5 Display Lock Setup 6 Measurement

More information

MOTOR ARMATURE TESTER

MOTOR ARMATURE TESTER MOTOR ARMATURE TESTER 6920 ENGLISH MANUAL 1 st Oct 2008 CATALOG BEFORE USE...1 1.1 Product Introduction...1 1.2 Product Characteristic...1 2 SPEC INSTRUCTION...2 3 SAFETY INSTRUCTION...4 4 INSTALLATION...6

More information

Operation Manual for OPM-1315 Optical Power Meter

Operation Manual for OPM-1315 Optical Power Meter Operation Manual for OPM-1315 Optical Power Meter www.tonercable.com Table of Contents 1. Introduction... 1,2 1.1 Summary... 1 1.2 Product Features... 2 1.3 Technical Parameters... 2 1.4 Main Applications...

More information

LASER PLUS OPERATING INSTRUCTIONS

LASER PLUS OPERATING INSTRUCTIONS LASER PLUS OPERATING INSTRUCTIONS Phone: (209) 586-1022 (800) 545-1022 Fax: (209) 586-1026 E-Mail: salessupport@olsontech.com 025-370403 REV X9 www.olsontech.com 12/16/03 TABLE OF CONTENTS SYSTEM OVERVIEW

More information

INSTRUCTION MANUAL HIGH CURRENT DRIVER BOARD STAND ALONE

INSTRUCTION MANUAL HIGH CURRENT DRIVER BOARD STAND ALONE Instruction Manual High Current Driver Board Stand Alone 1 INSTRUCTION MANUAL HIGH CURRENT DRIVER BOARD STAND ALONE Models: 1 AMP: HCDBSA1_0AMP 1.5 AMP: HCDBSA1_5AMP 2.0 AMP: HCDBSA2_0AMP 2.5 AMP: HCDBSA2_5AMP

More information

OBSOLETE. Isolated, Linearized, Thermocouple Input 3B47 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

OBSOLETE. Isolated, Linearized, Thermocouple Input 3B47 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM FEATURES Interfaces, amplifies, filters, isolates, & linearizes analog input voltages from a J, K, T, E, R, S or B-type thermocouple Thermocouple input signal is internally linearized High accuracy internal

More information

ACTIONI/Q Q498. DC Powered DC Input Field Configurable Isolator with Math Functions

ACTIONI/Q Q498. DC Powered DC Input Field Configurable Isolator with Math Functions ACTIONI/Q Q498 DC Powered DC Input Field Configurable Isolator with Math Functions Q498-0000 Provides 2 Isolated Analog Inputs, 1 Analog Output & 1 Frequency Output, as well as a Discrete Input & Output

More information

FIELD CALIBRATION PROCEDURE FOR ALTEK MODEL 222A

FIELD CALIBRATION PROCEDURE FOR ALTEK MODEL 222A FIELD CALIBRATION PROCEDURE FOR ALTEK MODEL 222A DRAWING NO. 1-827 Date Rev. Revision Record Dr. Review Approve 8FEB94 A Initial Release DJN PBG RPC 25AUG98 B Corrected Specifications DJN AC RPC 26AUG98

More information

Model ST-FT1 DIGITAL FORCE / TORQUE INDICATOR. User s Guide

Model ST-FT1 DIGITAL FORCE / TORQUE INDICATOR. User s Guide Model ST-FT1 DIGITAL FORCE / TORQUE INDICATOR Thank you Thank you for purchasing a Mesa Labs Model ST-FT1 digital force / torque indicator, designed for use with a remote torque sensor. With proper usage,

More information

LVDT Test Device. Product Manual (Revision B) Original Instructions. Testing and Calibration of LVDT Feedback Devices on PG Governors

LVDT Test Device. Product Manual (Revision B) Original Instructions. Testing and Calibration of LVDT Feedback Devices on PG Governors Product Manual 55036 (Revision B) Original Instructions Testing and Calibration of LVDT Feedback Devices on PG Governors Operation Manual DEFINITIONS This is the safety alert symbol. It is used to alert

More information

L200 User Manual. Operating Instructions for L200-TC & L200-PT Temperature Monitor. RoHS compliant

L200 User Manual. Operating Instructions for L200-TC & L200-PT Temperature Monitor. RoHS compliant L200 User Manual Operating Instructions for L200-TC & L200-PT Temperature Monitor RoHS compliant Contents 1. About the L200................................................................... 3 2. Installation.......................................................................

More information

2 in 1 LAN Tester and Multimeter Model:

2 in 1 LAN Tester and Multimeter Model: 2 in 1 LAN Tester and Multimeter Model: 72-8495 1 IMPORTANT SAFETY INFORMATION Please read these instructions carefully before use and retain for future reference. This instrument is designed and manufactured

More information

Autoranging True RMS Multimeter User Manual

Autoranging True RMS Multimeter User Manual Autoranging True RMS Multimeter User Manual Please read this manual before switching the unit on. Important safety information inside. Contents Page 1. Safety Information... 4 2. Safety Symbols... 5 3.

More information

2 in1 LAN Tester & Multimeter. Model: PCE-LT 1

2 in1 LAN Tester & Multimeter. Model: PCE-LT 1 www.pce-industrial-needs.com Tursdale Technical Services Ltd Unit N12B Tursdale Business Park Co. Durham DH6 5PG United Kingdom Phone: +44 ( 0 ) 191 377 3398 Fax: +44 ( 0 ) 191 377 3357 info@tursdaletechnicalservices.co.uk

More information

HAKKO FG-101 Calibration Instructions Revision

HAKKO FG-101 Calibration Instructions Revision HAKKO FG-101 Calibration Instructions Revision 2013-10 Tools Required: Room Temperature Thermometer Yokogawa CA-71 Portable Calibrator or equivalent Yokogawa 2786 Decade Resistance Box or equivalent Fluke

More information

PMM IMMUNITY SUITE. User Manual DATA ACQUISITION PROGRAM FOR RADIATED & CONDUCTED IMMUNITY TESTS

PMM IMMUNITY SUITE. User Manual DATA ACQUISITION PROGRAM FOR RADIATED & CONDUCTED IMMUNITY TESTS http://www.narda-sts.it NARDA Safety Test Solutions S.r.l. Socio Unico Sales & Support: Via Leonardo da Vinci, 21/23 20090 Segrate (MI) - ITALY Tel.: +39 02 2699871 Fax: +39 02 26998700 Manufacturing Plant:

More information

Strain Gauge Measurement by Applying Displacement using Strain Indicator

Strain Gauge Measurement by Applying Displacement using Strain Indicator University Diploma Program Electronic Equipment Maintenance Lab Instructor: Muhammad Ajmal Khan EET-027, Experiment # 8 Strain Gauge Measurement by Applying Displacement using Strain Indicator Objectives:

More information

TC100 Precision Thermocouple Calibrator

TC100 Precision Thermocouple Calibrator TC100 Precision Thermocouple Calibrator Table of Contents 1. Introduction.....................1 2. Accessories....................5 3. Set-Up Basic/Advanced...........5 4. Operating Procedure.............6

More information

Wide Bandwidth Strain Gage Input 3B18 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Wide Bandwidth Strain Gage Input 3B18 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM Wide Bandwidth Strain Gage Input 3B18 FEATURES Wideband (20 khz) single-channel signal conditioning module. Module Bandwidth is user-selectable between 20 khz and 100Hz, with user-supplied filter caps

More information

User s Guide 200 ma Current Source Module CSM-39020

User s Guide 200 ma Current Source Module CSM-39020 User s Guide 200 ma Current Source Module CSM-39020 ILX Lightwave Corporation 31950 Frontage Road Bozeman, MT, U.S.A. 59715 U.S. & Canada: 1-800-459-9459 International Inquiries: 406-556-2481 Fax 406-586-9405

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) ACF-3000 AK 3000A Flexible Current Transducer For use with multimeters or any instrument with AC milli-volt input Measures

More information

FIBEROPTIC DISPLACEMENT SENSOR. with Analog Output

FIBEROPTIC DISPLACEMENT SENSOR. with Analog Output USER MANUAL FOR FIBEROPTIC DISPLACEMENT SENSOR with Analog Output TYPE RC REFLECTANCE COMPENSATED PHILTEC www.philtec.com Fiberoptic Sensors for the Measurement of Distance, Displacement and Vibration

More information

Colorado Series. CR-10 Portable ph / mv / Temperature Meter Operation Manual

Colorado Series. CR-10 Portable ph / mv / Temperature Meter Operation Manual Colorado Series CR-10 Portable ph / mv / Temperature Meter Operation Manual 2 Contents Page General Introduction and Initial Inspection.............................. 4 Safety Infomation..................................................

More information

OWNER S MANUAL 9908-TE. HIGH PRECISION AUTO-RANGING DC/True RMS AC BENCH-TOP DIGITAL MULTIMETER

OWNER S MANUAL 9908-TE. HIGH PRECISION AUTO-RANGING DC/True RMS AC BENCH-TOP DIGITAL MULTIMETER OWNER S MANUAL 9908-TE HIGH PRECISION AUTO-RANGING DC/True RMS AC BENCH-TOP DIGITAL MULTIMETER IMPORTANT! Read and understand this manual before using the instrument. Failure to understand and comply with

More information

User's Guide. 800 Amp Clamp Meters. EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter

User's Guide. 800 Amp Clamp Meters. EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter User's Guide 800 Amp Clamp Meters EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter Introduction Congratulations on your purchase of the EX710, EX720, or EX730 Clamp DMM.

More information

JR3 EXTERNAL SENSOR ELECTRONICS WITH SERIAL DATA OUTPUT. JR3, Inc. 22 Harter Ave. Woodland, CA 95776

JR3 EXTERNAL SENSOR ELECTRONICS WITH SERIAL DATA OUTPUT. JR3, Inc. 22 Harter Ave. Woodland, CA 95776 JR3 EXTERNAL SENSOR ELECTRONICS WITH SERIAL DATA OUTPUT JR3, Inc. 22 Harter Ave. Woodland, CA 95776 5963B 13 October, 2003 TABLE OF CONTENTS CHAPTER 1 OVERVIEW General...1-1 Sensor...1-1 Electronic System...1-1

More information

Blue Point Engineering

Blue Point Engineering Blue Point Engineering Board - Pro Module (E) Instruction Pointing the Way to Solutions! Controller I Version 2.1 The Board Pro E Module provides the following features: Up to 4 minutes recording time

More information

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: 57314

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: 57314 Instruction Manual 2in1 LAN Tester & Multimeter Model: 57314 1 Contents Introduction... Features... Safety Precautions.. Meter Description... Electrical Specification... Operation.. AutoRanging Multimeter.

More information

The 5700A/5720A High Performance Multifunction Calibrators. Taking Accuracy to a Whole New Level

The 5700A/5720A High Performance Multifunction Calibrators. Taking Accuracy to a Whole New Level The 5700A/5720A High Performance Multifunction Calibrators Taking Accuracy to a Whole New Level The 5700A/5720A Calibrators Set the Highest Standards for Accuracy and Performance Growing global trade is

More information

Digital Power Probes, ACL120, ACL220, ACL Internal, and DCL5 (#30015, #30023, #30024, #30025)

Digital Power Probes, ACL120, ACL220, ACL Internal, and DCL5 (#30015, #30023, #30024, #30025) Digital Power Probes, ACL120, ACL220, ACL Internal, and DCL5 (#30015, #30023, #30024, #30025) This series of digital power probes by Networked Robotics is used to record a yes/no indication of the presence

More information

Service Manual. P GHz Active Probe

Service Manual. P GHz Active Probe Service Manual P7240 4 GHz Active Probe 071-1056-00 Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified

More information

VersaPoint I/O Module

VersaPoint I/O Module Output 24VDC Positive Logic 0.5A 14 Points Module is used to output 24VDC digital signals. Module Specifications Housing dimensions (width x height x depth) Connection style Operating temperature 48.8mm

More information

Model User Manual Revision E 04/29/99. OEM Gaging System. Part Number R01

Model User Manual Revision E 04/29/99. OEM Gaging System. Part Number R01 04/29/99 Model 3800 OEM Gaging System User Manual Revision E Part Number 028585-R01 Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted

More information

FL MC 2000E (SM40) LC

FL MC 2000E (SM40) LC IEC 61850 fiber optic converter with LC fiber optic connection (1310 nm) to convert 100Base-Tx to single- or multi-mode fiber glass Data sheet 3205_en_C 1 Description PHOENIX CONTACT 2014-04-04 2 Features

More information

WalkLAB ph mv/orptemperature

WalkLAB ph mv/orptemperature 16,1,16,1,2,15,2,15,14,3,14,3,4,13,4,13,12,5,12,5,6,11,6,11,10,7,10,7,8,9,8,9 WalkLAB ph mv/orptemperature meter HP9000 Operations Manual TABLE OF CONTENTS PAGE 1. INTRODUCTION 1 1.1 Features 1 1.2 Technical

More information

HiRES TRANSIENT RECORDER

HiRES TRANSIENT RECORDER HiRES TRANSIENT RECORDER n Measurement of voltage, current and other values n Exact measurements even in environments with very strong EMI n Reliable signal transmission via fiber optic cables n Sophisticated

More information

OPERATION MANUAL MODEL TT-LP

OPERATION MANUAL MODEL TT-LP TURTLE TOUGH OPERATION MANUAL MODEL TT-LP 2-Wire Isolated ph/orp Transmitter turtletough.com.au CONTENT INITIAL INSPECTION 3 INTRODUCTION 3 ASSEMBLY 4 PREPARATION 5 Connection Diagram of Turtle Tough Sensors

More information

Manual 2.1. Installation of HEKA Hardware and Software

Manual 2.1. Installation of HEKA Hardware and Software Manual 2.1 Installation of HEKA Hardware and Software HEKA Elektronik Phone +49 (0) 6325 / 95 53-0 Dr. Schulze GmbH Fax +49 (0) 6325 / 95 53-50 Wiesenstrasse 71 Web Site www.heka.com D-67466 Lambrecht/Pfalz

More information

JUMO dtrans T02. programmable. transmitter. B Operating Instructions 5.00/

JUMO dtrans T02. programmable. transmitter. B Operating Instructions 5.00/ JUMO dtrans T02 programmable transmitter B 95.6521 Operating Instructions 5.00/00380664 Overview of operation JUMO dtrans T02 Standard accessory 1 Type designation (1) Basic version 956521 programmable

More information

ANALOG STORAGE OSCILLOSCOPE

ANALOG STORAGE OSCILLOSCOPE ANALOG STORAGE OSCILLOSCOPE An ISO 9001:2008 Company Model KM-20-5030(30MHz) / KM-20-5060(60MHz) FEATURES : DC-30MHZ / DC-60MHz Dual channels /dual traces, X-Y mode 6 display, high brightness Toshiba oscilloscope

More information

CLEVELAND CONTROLS MODEL 6650 DIGITAL MANOMETER INSTRUCTION MANUAL DM

CLEVELAND CONTROLS MODEL 6650 DIGITAL MANOMETER INSTRUCTION MANUAL DM CLEVELAND CONTROLS MODEL 6650 DIGITAL MANOMETER INSTRUCTION MANUAL DM-6650.02 1 Instruction Manual DM-6650.02 TABLE OF CONTENTS I.0 INTRODUCTION 3 1.1 General Description... 3 1.2 Specifications... 3 1.3

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS TT-568 /a INSTALLATION INSTRUCTIONS Original Issue Date: 2/ Model: 4/20RES and 4/20RESL Generator Sets Market: Residential/Commercial Subject: RDC and DC Controller Replacement Kits GM79884 and GM79885

More information

MODEL 9250 Preliminary data sheet

MODEL 9250 Preliminary data sheet Delivery: ex stock Warranty: 24 months Universal Instrumentation Amplifier for strain gage, potentiometric, DC/DC and incremental sensors MODEL 9250 Preliminary data sheet NEW burster TEDS Highlights Ultra-fast

More information

PSI DITHERLESS BIAS CONTROLLER USER GUIDE

PSI DITHERLESS BIAS CONTROLLER USER GUIDE PSI-0404-11 DITHERLESS BIAS CONTROLLER USER GUIDE Rev D Photonic Systems, Inc. Tel: 978-670-4990 900 Middlesex Turnpike Fax: 978-670-2510 Building #5 www.photonicsinc.com Billerica, MA 01821 1 DOCUMENT

More information

RS Stock No Instruction Manual RS Input Data Logging Thermometer

RS Stock No Instruction Manual RS Input Data Logging Thermometer RS Stock No. 730-0458 Instruction Manual RS-1384 4 Input Data Logging Thermometer EN FR IT DE ES TABLE OF CONTENTS / EN TITLE TABLE OF CONTENTS PAGE 1. INTRODUCTION FEATURE... 1 2. SPECIFICATIONS... 2

More information

Operation Manual. MODEL ph6175. Microcomputer Based ph/orp/temperature Benchtop Meter

Operation Manual. MODEL ph6175. Microcomputer Based ph/orp/temperature Benchtop Meter Operation Manual MODEL ph6175 Microcomputer Based ph/orp/temperature Benchtop Meter ph6175 0 CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 POWER INPUT.......2 INSTALLING THE BATTERIES.....3

More information

OTEB-CO-P PRE-AMP EDFA INSTRUCTION MANUAL

OTEB-CO-P PRE-AMP EDFA INSTRUCTION MANUAL OTEB-CO-P PRE-AMP EDFA INSTRUCTION MANUAL Phone: (209) 586-1022 (800) 545-1022 Fax: (209) 586-1026 E-Mail: salessupport@olsontech.com www.olsontech.com 025-000521 Rev. X1 10/01/2012 SAFETY WARNINGS LASER

More information

INSTRUCTION MANUAL. ph/mv CONTROLLER/RECORDER

INSTRUCTION MANUAL. ph/mv CONTROLLER/RECORDER INSTRUCTION MANUAL ph/mv CONTROLLER/RECORDER LIMITED ONE YEAR WARRANTY Manufacturer warranties all instruments (excluding batteries, damage caused by batteries, probes, standards, buffers) against defects

More information

Digital Multimeter Instruction Manual

Digital Multimeter Instruction Manual 126 Digital Multimeter Instruction Manual TABLE OF CONTENTS page A. INTRODUCTION 1. Congratulations..................3 2. Product Description..............3 3. EC Declaration of Conformity........4 B.

More information