Agilent N5394A DVI Electrical Performance Validation and Compliance Software

Size: px
Start display at page:

Download "Agilent N5394A DVI Electrical Performance Validation and Compliance Software"

Transcription

1 Agilent N5394A DVI Electrical Performance Validation and Compliance Software Compliance Testing Methods of Implementation Agilent Technologies

2 Notices Agilent Technologies, Inc No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. Manual Part Number N Edition Fifth edition, August 2008 Printed in USA Agilent Technologies, Inc Garden of the Gods Road Colorado Springs, CO USA Warranty The material contained in this document is provided as is, and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license. Restricted Rights Legend If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as Commercial computer software as defined in DFAR (June 1995), or as a commercial item as defined in FAR 2.101(a) or as Restricted computer software as defined in FAR (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies standard commercial license terms, and non-dod Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR (c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR (June 1987) or DFAR (b)(2) (November 1995), as applicable in any technical data. Safety Notices CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

3 DVI Automated Testing At A Glance The Agilent N5394A DVI Electrical Performance Validation and Compliance Software helps you verify DVI transmitter device under test (DUT) compliance to specifications with the Agilent 54854A/54855A, and 90000A series Infiniium Digital Storage Oscilloscope. The DVI Electrical Performance Validation and Compliance Software: Lets you select individual or multiple tests to run. Lets you identify the device being tested and its configuration. Shows you how to make oscilloscope connections to the device under test. Automatically checks for proper oscilloscope configuration. Automatically sets up the oscilloscope for each test. Provides detailed information for each test that has been run and lets you specify the thresholds at which marginal or critical warnings appear. Creates a printable HTML report of the tests that have been run. NOTE The tests performed by the DVI Electrical Performance Validation and Compliance Software are intended to provide a quick check of the electrical health of the DUT. This testing is not a replacement for an exhaustive test validation plan. Compliance testing measurements are described in Section 4.7, Electrical Measurement Procedures, in the DVI Specification Revision 1.0. For more information, see the DDWG (Digital Display Working Group) web site at Required Equipment and Software In order to run the DVI automated tests, you need the following equipment and software: 54854A/54855A, or 90000A series Infiniium Digital Storage Oscilloscope. Option 001 (1M/ch memory upgrade) is recommended; this will greatly reduce Data Eye Pattern and Jitter test times. Version A or greater of the Infiniium software (54854A/54855A, and series Infiniium Digital Storage Oscilloscope) Version 1.0 of the baseline software (90000A series Infiniium Digital Storage Oscilloscope) N5394A DVI Electrical Performance Validation and Compliance Software. Probes and/or test fixtures. For more information on the specific probes and test fixtures required, refer to the chapters that describe tests. DVI Compliance Testing Methods of Implementation 3

4 Keyboard, qty = 1, (provided with the Agilent 54854A/54855A, and 90000A series oscilloscope). Mouse, qty = 1, (provided with the Agilent 54854A/54855A, and 90000A series oscilloscope). Precision 3.5 mm BNC to SMA male adapter, Agilent p/n , qty = 2 (provided with the Agilent 54854A/54855A, and 90000A series oscilloscope). 50- ohm Coax Cable with SMA Male Connectors 24- inch or less RG- 316/U or similar, qty = 2, matched length. 4 DVI Compliance Testing Methods of Implementation

5 In This Book This manual describes the tests that are performed by the DVI Electrical Performance Validation and Compliance Software in more detail; it contains information from (and refers to) the DVI Specification Revision 1.0, and it describes how the tests are performed. Chapter 1, Installing the DVI Electrical Performance Validation and Compliance Software shows how to install and license the automated test application software (if it was purchased separately). Chapter 2, Preparing to Take Measurements shows how to start the DVI Electrical Performance Validation and Compliance Software and gives a brief overview of how it is used. Chapter 3, Transmitter Data Eye Pattern Tests contains more information on the data eye pattern tests. Chapter 4, Transmitter Jitter Tests contains more information on the clock jitter test. Chapter 5, Transmitter Transition Time Tests contains more information on the risetime/falltime tests. Chapter 6, Transmitter Inter- Pair Skew Tests contains more information on the inter- pair skew tests. Chapter 7, Cable Assembly Tests contains more information on testing cable assemblies. Appendix A, Calibrating the Digital Storage Oscilloscope describes how to calibrate the oscilloscope in preparation for running the DVI automated tests. Appendix B, INF_SMA_Deskew.set Setup File Details describes a setup used when performing channel de- skew calibration. Appendix C, InfiniiMax Probing describes the 1132A/1134A probe amplifier and probe head recommendations for DVI testing. See Also The DVI Electrical Performance Validation and Compliance Software s online help, which describes: Creating or opening a test project. Selecting tests. Configuring selected tests. Connecting the oscilloscope to the DUT. Running tests. Viewing test results. Viewing/printing the HTML test report. Saving test projects. DVI Compliance Testing Methods of Implementation 5

6 6 DVI Compliance Testing Methods of Implementation

7 Contents DVI Automated Testing At A Glance 3 Required Equipment and Software 3 In This Book 5 See Also 5 1 Installing the DVI Electrical Performance Validation and Compliance Software Installing the Software 9 Installing the License Key 9 2 Preparing to Take Measurements Acquiring Test Fixtures 12 Soldering Damping Resistors into Test Fixtures 15 Calibrating the Oscilloscope 16 Starting the DVI Electrical Performance Validation and Compliance Software 17 Online Help Topics 18 3 Transmitter Data Eye Pattern Tests Probing for Data Eye Pattern Tests 22 D0/D1/D2 - Eye Pattern Test Method of Implementation 24 Test Definition Notes from the Specification 24 Test Procedure 24 PASS Condition 26 Measurement Algorithm 26 Test References 26 4 Transmitter Jitter Tests Probing for Jitter Tests 28 Clock - Jitter Test Method of Implementation 30 Test Definition Notes from the Specification 30 Test Procedure 30 PASS Condition 31 Measurement Algorithm 31 Test References 31 DVI Compliance Testing Methods of Implementation 7

8 5 Transmitter Transition Time Tests Probing for Transition Time Tests 34 Transition Time Test Method of Implementation 36 Test Definition Notes from the Specification 36 Test Procedure 36 PASS Condition 37 Measurement Algorithm 37 Test References 37 6 Transmitter Inter-Pair Skew Tests 7 Cable Assembly Tests Probing for InterPairSkew Tests 40 D0/D1, D0/D2, D1/D2 - Skew Test Method of Implementation 42 Test Definition Notes from the Specification 42 Test Procedure 42 PASS Condition 44 Measurement Algorithm 44 Test References 44 Probing for Cable Assembly Eye Pattern Tests 46 D0/D1/D2 - Eye Pattern Test Method of Implementation 48 Test Definition Notes from the Specification 48 Test Procedure 48 PASS Condition 50 Measurement Algorithm 50 Test References 50 A Calibrating the Digital Storage Oscilloscope Required Equipment for Calibration 51 Internal Calibration 52 Cable and Probe Calibration 58 Channel-to-Channel De-skew 67 B INF_SMA_Deskew.set Setup File Details C InfiniiMax Probing Index 8 DVI Compliance Testing Methods of Implementation

9 N5394A DVI Electrical Performance Validation and Compliance Software Compliance Testing Methods of Implementation 1 Installing the DVI Electrical Performance Validation and Compliance Software Installing the Software 9 Installing the License Key 9 If you purchased the N5394A DVI Electrical Performance Validation and Compliance Software separately, you need to install the software and license key. Installing the Software 1 Make sure you have version A or higher of the Infiniium oscilloscope software (54854A/54855A and Series oscilloscope) OR version 1.00 or greater of baseline software (90000A Series oscilloscope) by choosing Help>About Infiniium... from the main menu. 2 To obtain the DVI Electrical Performance Validation and Compliance Software, go to Agilent website: 3 The link for DVI Electrical Performance Validation and Compliance Software will appear. Double- click on it and follow the instructions to download and install the application software. Be sure to accept the installation of the.net Framework software; it is required in order to run the DVI Electrical Performance Validation and Compliance Software. Installing the License Key 1 Request a license code from Agilent by following the instructions on the Entitlement Certificate. You will need the oscillocope s Option ID Number, which you can find in the Help>About Infiniium... dialog. 2 After you receive your license code from Agilent, choose Utilities>Install Option License... Agilent Technologies 9

10 1 Installing the DVI Electrical Performance Validation and Compliance Software 3 In the Install Option License dialog, enter your license code and click Install License. 4 Click OK in the dialog that tells you to restart the Infiniium oscilloscope application software to complete the license installation. 5 Click Close to close the Install Option License dialog. 6 Choose File>Exit. 7 Restart the Infiniium oscilloscope application software to complete the license installation. 10 DVI Compliance Testing Methods of Implementation

11 N5394A DVI Electrical Performance Validation and Compliance Software Compliance Testing Methods of Implementation 2 Preparing to Take Measurements Acquiring Test Fixtures 12 Calibrating the Oscilloscope 16 Starting the DVI Electrical Performance Validation and Compliance Software 17 Before running the DVI automated tests, you need to acquire the appropriate test fixtures, and you should calibrate the oscilloscope. After the oscilloscope has been calibrated, you are ready to start the DVI Electrical Performance Validation and Compliance Software and perform measurements. Agilent Technologies 11

12 2 Preparing to Take Measurements Acquiring Test Fixtures DVI test fixtures can be acquired from a distributor of Silicon Image parts or from the DDWG (Digital Display Working Group) (see the web site at Currently available test fixtures, called TPA2 fixtures in this document, include a CRU (Clock Recovery Unit) board and TPA2- P (plug) and TPA2- R (receptacle) adapter boards. The raw clock test points for these fixtures are T1 and T2 on the CRU board. Figure 1 TPA2-P (Plug) Adapter Board Figure 2 TPA2-R (Receptacle) Adapter Board (Top and Bottom Views) 12 DVI Compliance Testing Methods of Implementation

13 Preparing to Take Measurements 2 Figure 3 CRU (Clock Recovery Unit) Board (for TPA2 Fixtures) NOTE The CRU board must be configured according to the clock frequency in use. DVI Compliance Testing Methods of Implementation 13

14 2 Preparing to Take Measurements Earlier DVI plug and receptacle test fixtures from Silicon Image, called TPA fixtures in this document, had the CRU (Clock Recovery Unit) built- in. The raw clock test points for these fixtures are T2 and T3. Figure 4 TPA-P (Plug) Test Point Access Board Figure 5 TPA-R (Receptacle) Test Point Access Board 14 DVI Compliance Testing Methods of Implementation

15 Preparing to Take Measurements 2 Soldering Damping Resistors into Test Fixtures To get the best probed signal fidelity, the E2678A differential socket probe head for the 1132A/1134A InfiniiMax probe amplifier is recommended. This probing solution requires damping resistors to be soldered into test fixture boards at the data channel and raw clock test points. The proper damping resistors (and instructions for soldering them into a device under test) are included with the E2678A differential socket probe head. Figure 6 Damping Resistors Soldered into TPA2-P and CRU Boards DVI Compliance Testing Methods of Implementation 15

16 2 Preparing to Take Measurements Calibrating the Oscilloscope If you haven t already calibrated the oscilloscope, see Appendix A, Calibrating the Digital Storage Oscilloscope. NOTE If the ambient temperature changes more than 5 degrees Celsius from the calibration temperature, internal calibration should be performed again. The delta between the calibration temperature and the present operating temperature is shown in the Utilities>Calibration menu. NOTE If you switch cables between channels or other oscilloscopes, it is necessary to perform cable and probe calibration again. Agilent recommends that, once calibration is performed, you label the cables with the channel they were calibrated for. 16 DVI Compliance Testing Methods of Implementation

17 Preparing to Take Measurements 2 Starting the DVI Electrical Performance Validation and Compliance Software 1 From the Infiniium oscilloscope s main menu, choose Analyze>Automated Test Apps>DVI. Figure 7 The DVI Electrical Performance Validation and Compliance Software NOTE If DVI does not appear in the Automated Test Apps menu, the DVI Electrical Performance Validation and Compliance Software has not been installed (see Chapter 1, Installing the DVI Electrical Performance Validation and Compliance Software ). DVI Compliance Testing Methods of Implementation 17

18 2 Preparing to Take Measurements Figure 7 shows the DVI Electrical Performance Validation and Compliance Software main window. The task flow pane, and the tabs in the main pane, show the steps you take in running the automated tests: Select Tests Configure Connect Run Tests Results HTML Report Lets you select the tests you want to run. The tests are organized hierarchically so you can select all tests in a group. After tests are run, status indicators show which tests have passed, failed, or not been run, and there are indicators for the test groups. Lets you enter information about the device being tested and configure test parameters (like memory depth). This information appears in the HTML report. Shows you how to connect the oscilloscope to the device under test for the tests to be run. Starts the automated tests. If the connections to the device under test need to be changed while multiple tests are running, the tests pause, show you how to change the connection, and wait for you to confirm that the connections have been changed before continuing. Contains more detailed information about the tests that have been run. You can change the thresholds at which marginal or critical warnings appear. Shows a compliance test report that can be printed. You can choose between a verbose and compact report. Online Help Topics For information on using the DVI Electrical Performance Validation and Compliance Software, see its online help (which you can access by choosing Help>Contents... from the application s main menu). The DVI Electrical Performance Validation and Compliance Software s online help describes: Creating or opening a test project. Selecting tests. Configuring selected tests. Connecting the oscilloscope to the DUT. Running tests. Viewing test results. To show reference images and flash mask hits. To change margin thresholds. 18 DVI Compliance Testing Methods of Implementation

19 Preparing to Take Measurements 2 Viewing/printing the HTML test report. Saving test projects. DVI Compliance Testing Methods of Implementation 19

20 2 Preparing to Take Measurements 20 DVI Compliance Testing Methods of Implementation

21 N5394A DVI Electrical Performance Validation and Compliance Software Compliance Testing Methods of Implementation 3 Transmitter Data Eye Pattern Tests Probing for Data Eye Pattern Tests 22 D0/D1/D2 - Eye Pattern Test Method of Implementation 24 This section provides the Methods of Implementation (MOIs) for data eye pattern tests using an Agilent 54854A/54855A, or 90000A series Infiniium oscilloscope, 1132A/1134A probes, and the DVI Electrical Performance Validation and Compliance Software. Agilent Technologies 21

22 3 Transmitter Data Eye Pattern Tests Probing for Data Eye Pattern Tests When performing data eye pattern tests, the DVI Electrical Performance Validation and Compliance Software will prompt you to make the proper connections (also shown in Figure 8 or Figure 9). TPA2-P Adapter Board CRU Board DVI Source (DUT) DVI Connector Damping Resistors E2678A Differential Socket Probe Head 1132A/1134A Probe Amplifier + - D0/D1/D2 Test Point + - Raw Clock SMA to SMA Cables 54854A/54855A Oscilloscope + - Ch2 Ch3 Ch4 Recovered Clock SMA to BNC Cable Figure 8 Probing for Data Eye Pattern Tests (TPA2 Fixture) TPA-P Test Access Board DVI Source (DUT) DVI Connector D0/D1/D2 Test Point + - E2678A Differential Socket Probe Head Damping Resistors 1132A/1134A Probe Amplifier Recovered Clock 54854A/54855A Oscilloscope SMA to BNC Cable Ch2 Ch3 Ch4 Figure 9 Probing for Data Eye Pattern Tests (TPA Fixture) 22 DVI Compliance Testing Methods of Implementation

23 Transmitter Data Eye Pattern Tests 3 You can use any of the oscilloscope channels for the recovered clock and data test point. You identify the channels used for each signal in the Configuration tab of the DVI Electrical Performance Validation and Compliance Software. (The channels shown in Figure 8 and Figure 9 are just for example.) For more information on the 1132A/1134A probe amplifiers and differential probe heads, see Appendix C, InfiniiMax Probing, starting on page 75. DVI Compliance Testing Methods of Implementation 23

24 3 Transmitter Data Eye Pattern Tests D0/D1/D2 - Eye Pattern Test Method of Implementation Data eye pattern measurements are made as a differential measurements using the recovered clock signal as the oscilloscope trigger source. The transmitter drives the pseudo- random data test pattern and the test ensures that the eye diagram limits given by the DVI spec are not exceeded. Test Definition Notes from the Specification Test Procedure See section 4.7.6, Transmitter Eye, in the DVI Specification Revision Attach a DVI monitor to the DVI source (DUT). 2 Select a resolution for test, and display the pseudo random pattern in full screen mode. 3 Unplug the monitor, and connect the TPA2 plug adapter board directly to the output of the DVI source. 4 Confirm the frequency output of the DVI source by measuring RC+ and RCon the TPA2 plug board with the oscilloscope. 5 Set the CRU board to operate at the correct frequency. 6 Allow the CRU board to warm up for 5 minutes. 7 Attach the SMA to SMA cables from RC+ and RC- on the TPA2 plug board to the respective SMA connectors on the CRU board. 8 Start the automated testing application as described in Starting the DVI Electrical Performance Validation and Compliance Software" on page In the DVI Test application, click the Select Tests tab. 10 Navigate to the Data Eye Pattern group, and check the appropriate Eye Pattern test (D0, D1, or D2). 24 DVI Compliance Testing Methods of Implementation

25 Transmitter Data Eye Pattern Tests 3 Figure 10 Selecting Data Eye Pattern Tests 11 Follow the DVI Test application s task flow to set up configuration options, make oscilloscope connections, run the test, and view the test results. Table 1 Data Eye Pattern Test Configuration Options Configuration Option Device Description Test Fixture Recovered Clock Description Lets you enter information about the device under test, including the device ID and any comments that you wish to include. Identifies the revision of your test fixture. Agilent recommends test fixture TPA2. The recovered clock from TPA will have 2.5 times the frequency of the raw clock. TPA2 however will have the same frequency for both clocks. Identifies the oscilloscope channel probing the recovered clock. D0/D1/D2 Identifies the oscilloscope channels probing data channel links D0, D1, and D2. Test Freq [Screen Resolution] Mask Type Identifies the frequency of the DVI interface you are testing. You can select one of the frequencies from the drop-down list, or you can enter a frequency in the range from 25 to 165 MHz. Choose Normalized Eye Pattern Mask at TP2. DVI Compliance Testing Methods of Implementation 25

26 3 Transmitter Data Eye Pattern Tests Table 1 Data Eye Pattern Test Configuration Options (continued) Configuration Option # Unit Interval Available in Debug Mode, this option specifies the number of UI (Unit Intervals) to measure in the data eye pattern test. The DVI revision 1.0 specification recommends 100,000 acquisitions to achieve 99% confidence within 1% error of the mean value. For full compliance, a measurement of 1,000,000 acquisitions must be made. Note that increasing the number of UI has a negative impact on the run time of the data eye pattern tests. Value used in Compliance Mode: 100,000. Manual Mask Adjustment Composite Eye Worst Eye Location Description Specifies whether the manual mask adjustment is needed prior to the mask test. Available in Debug Mode, this option specifies whether all data bits in a pixel need to be overlapped in the eye diagram. Value used in Compliance Mode: Off. Available in Debug Mode, this option specifies the location of the worst eye in respect to the rising edge of the recovered clock. This parameter will be ignored if Composite Eye is turned ON. Choose auto to let the application automatically find the worst eye. The worst eye is defined as the bit having the smallest pulse width. Select only 1-4 for the TPA test fixture. Value used in Compliance Mode: Auto. PASS Condition Zero mask failures. Measurement Algorithm Test References 1 Trigger on the recovered clock; the data lane is being viewed in color grade form. 2 Repeat step 1 until the # Unit Intervals has been reached. 3 The test passes if none of the waveform sample points falls into the mask failure region. Section 4.3, Transmitter Electrical Specifications, in the DVI Specification Revision DVI Compliance Testing Methods of Implementation

27 N5394A DVI Electrical Performance Validation and Compliance Software Compliance Testing Methods of Implementation 4 Transmitter Jitter Tests Probing for Jitter Tests 28 Clock - Jitter Test Method of Implementation 30 This section provides the Methods of Implementation (MOIs) for jitter tests using an Agilent 54854A/54855A, or 90000A series Infiniium oscilloscope, probes, and the DVI Electrical Performance Validation and Compliance Software. If you are using the TPA2 fixtures, the raw clock test points are T1 and T2 on the CRU board. However, if you are using the TPA fixtures, the raw clock test points are T2 and T3. Agilent Technologies 27

28 4 Transmitter Jitter Tests Probing for Jitter Tests When performing the clock jitter test, the DVI Electrical Performance Validation and Compliance Software will prompt you to make the proper connections (also shown in Figure 11 or Figure 12). DVI Source (DUT) DVI Connector TPA2-P Adapter Board + - Raw Clock SMA to SMA Cables CRU Board Test Point Recovered Clock 54854A/54855A Oscilloscope Ch2 Ch3 Ch4 Damping Resistors E2678A Differential Socket Probe Head 1132A/1134A Probe Amplifier SMA to BNC Cable Figure 11 Probing for Jitter Tests (TPA2 Fixture) DVI Source (DUT) DVI Connector TPA-P Test Access Board Raw Clock Test Point - + Recovered Clock SMA to BNC Cable Damping Resistors E2678A Differential Socket Probe Head 1132A/1134A Probe Amplifier 54854A/54855A Oscilloscope Ch2 Ch3 Ch4 Figure 12 Probing for Jitter Tests (TPA Fixture) 28 DVI Compliance Testing Methods of Implementation

29 Transmitter Jitter Tests 4 You can use any of the oscilloscope channels for the recovered clock and raw clock test point. You identify the channels used for each signal in the Configuration tab of the DVI Electrical Performance Validation and Compliance Software. (The channels shown in Figure 11 and Figure 12 are just for example.) DVI Compliance Testing Methods of Implementation 29

30 4 Transmitter Jitter Tests Clock - Jitter Test Method of Implementation The jitter measurement in the raw clock is performed as a differential measurement of the rising edge of the clock signal (clk+ minus clk- ). The clock signal from the clock recovery unit must be used as a trigger source. Test Definition Notes from the Specification Test Procedure See section 4.7.7, Jitter Measurement, in the DVI Specification Revision Start the automated testing application as described in Starting the DVI Electrical Performance Validation and Compliance Software" on page In the DVI Test application, click the Select Tests tab. 3 Navigate to the Jitter group, and check the Clock - Jitter test. Figure 13 Selecting Jitter Tests 4 Follow the DVI Test application s task flow to set up configuration options, make oscilloscope connections, run the test, and view the test results. 30 DVI Compliance Testing Methods of Implementation

31 Transmitter Jitter Tests 4 Table 2 Clock Jitter Test Configuration Options Configuration Option Device Description Test Fixture Recovered Clock Raw Clock Test Freq [Screen Resolution] Number of edges (Jitter) Description Lets you enter information about the device under test, including the device ID and any comments that you wish to include. Identifies the revision of your test fixture. Agilent recommends test fixture TPA2. Identifies the oscilloscope channel probing the recovered clock. Identifies the oscilloscope channel probing the raw clock. Identifies the frequency of the DVI interface you are testing. You can select one of the frequencies from the drop-down list, or you can enter a frequency in the range from 25 to 165 MHz. Specifies the number of differential clock signal acquisitions to use when performing the jitter measurement. PASS Condition 0.25 T bit The bit time T bit is one tenth of the pixel time. The bit time is also referred to as one Unit Interval, or UI, in the jitter and eye diagram specification. Measurement Algorithm Test References 1 Trigger on the recovered clock; the raw clock signal is being viewed in color grade form. 2 Repeat step 1 until "Number of edges" has been acquired. 3 Jitter is the width the of signal at 50% crossing. (Measure using histogram.) Section 4.6, Jitter Specifications, in the DVI Specification Revision 1.0. DVI Compliance Testing Methods of Implementation 31

32 4 Transmitter Jitter Tests 32 DVI Compliance Testing Methods of Implementation

33 N5394A DVI Electrical Performance Validation and Compliance Software Compliance Testing Methods of Implementation 5 Transmitter Transition Time Tests Probing for Transition Time Tests 34 Transition Time Test Method of Implementation 36 If you are using the TPA2 fixtures, the raw clock test points are T1 and T2 on the CRU board. However, if you are using the TPA fixtures, the raw clock test points are T2 and T3. Agilent Technologies 33

34 5 Transmitter Transition Time Tests Probing for Transition Time Tests When performing transition time tests, the DVI Electrical Performance Validation and Compliance Software will prompt you to make the proper connections (also shown in the figures that follow). TPA2-P Adapter Board DVI Source (DUT) DVI Connector D0/D1/D2 Test Point + - Damping Resistors E2678A Differential Socket Probe Head 1132A/1134A Probe Amplifier 54854A/54855A Oscilloscope Ch2 Ch3 Ch4 Figure 14 Probing Data Channels for Transition Time Tests (TPA2 Fixture) DVI Source (DUT) DVI Connector TPA2-P Adapter Board + - Raw Clock SMA to SMA Cables CRU Board Test Point Damping Resistors E2678A Differential Socket Probe Head 1132A/1134A Probe Amplifier 54854A/54855A Oscilloscope Ch2 Ch3 Ch4 Figure 15 Probing Raw Clock for Transition Time Tests (TPA2 Fixture) 34 DVI Compliance Testing Methods of Implementation

35 Transmitter Transition Time Tests 5 TPA-P Test Access Board DVI Source (DUT) DVI Connector D0/D1/D2 Test Point + - E2678A Differential Socket Probe Head Damping Resistors 1132A/1134A Probe Amplifier 54854A/54855A Oscilloscope Ch2 Ch3 Ch4 Figure 16 Probing Data Channels for Transition Time Tests (TPA Fixture) TPA-P Test Access Board DVI Source (DUT) DVI Connector Raw Clock Test Point - + Damping Resistors E2678A Differential Socket Probe Head 1132A/1134A Probe Amplifier 54854A/54855A Oscilloscope Ch2 Ch3 Ch4 Figure 17 Probing Raw Clock for Transition Time Tests (TPA Fixture) You can use any of the oscilloscope channels for the data and raw clock test points. You identify the channels used for each signal in the Configuration tab of the DVI Electrical Performance Validation and Compliance Software. (The channels shown in the preceding figures are just for example.) DVI Compliance Testing Methods of Implementation 35

36 5 Transmitter Transition Time Tests Transition Time Test Method of Implementation Risetime/falltime is a differential measurement across the outputs of a differential pair. The transition is defined as the time interval between the normalized 20% and 80% amplitude levels. Test Definition Notes from the Specification Test Procedure See section 4.7.4, Transmitter Rise/Fall Time, in the DVI Specification Revision Start the automated testing application as described in Starting the DVI Electrical Performance Validation and Compliance Software" on page In the DVI Test application, click the Select Tests tab. 3 Navigate to the Transition Time group and the Clock or Data group, and check the appropriate risetime/falltime test. Figure 18 Selecting Transition Time Tests 36 DVI Compliance Testing Methods of Implementation

37 Transmitter Transition Time Tests 5 4 Follow the DVI Test application s task flow to set up configuration options, make oscilloscope connections, run the test, and view the test results. Table 3 Transition Time Test Configuration Options Configuration Option Device Description Test Fixture Raw Clock Description Lets you enter information about the device under test, including the device ID and any comments that you wish to include. Identifies the revision of your test fixture. Agilent recommends test fixture TPA2. Identifies the oscilloscope channel probing the raw clock. D0/D1/D2 Identifies the oscilloscope channels probing data channel links D0, D1, and D2. Test Freq [Screen Resolution] Number of edges (Transition Time) Identifies the frequency of the DVI interface you are testing. You can select one of the frequencies from the drop-down list, or you can enter a frequency in the range from 25 to 165 MHz. Specifies the number of acquisitions to use when performing the rise/fall time measurements. PASS Condition 75 ps Risetime/Falltime 0.4 T bit The bit time T bit is one tenth of the pixel time. The bit time is also referred to as one Unit Interval, or UI, in the jitter and eye diagram specification. Measurement Algorithm Test References The oscilloscope is set up to look for a waveform with 7 bits of 0 followed by 7 bits of 1 inside the pseudo- random pattern (using advanced comm triggering). When this waveform is found, it is used to determine V top (100%) and V base (0%) for the rise/fall time measurements. All subsequent risetime/falltime measurements are performed using simple edge triggering only. Section 4.3, Transmitter Electrical Specifications, in the DVI Specification Revision 1.0. DVI Compliance Testing Methods of Implementation 37

38 5 Transmitter Transition Time Tests 38 DVI Compliance Testing Methods of Implementation

39 N5394A DVI Electrical Performance Validation and Compliance Software Compliance Testing Methods of Implementation 6 Transmitter Inter-Pair Skew Tests Probing for InterPairSkew Tests 40 D0/D1, D0/D2, D1/D2 - Skew Test Method of Implementation 42 Agilent Technologies 39

40 6 Transmitter Inter-Pair Skew Tests Probing for InterPairSkew Tests When performing inter- pair skew tests, the DVI Electrical Performance Validation and Compliance Software will prompt you to make the proper connections (also shown in Figure 19 or Figure 20). NOTE Be sure to match the polarity of the differential probe to the polarity of the probed signals; otherwise, the tests will not run correctly. TPA2-P Adapter Board DVI Source (DUT) DVI Connector Data Channel + Test Points E2678A Differential Socket Probe Head 1132A/1134A Probe Amplifier Damping Resistors E2678A Differential Socket Probe Head 1132A/1134A Probe Amplifier 54854A/54855A Oscilloscope Ch2 Ch3 Ch4 Figure 19 Probing for Inter-Pair Skew Tests (TPA2 Fixture) 40 DVI Compliance Testing Methods of Implementation

41 Transmitter Inter-Pair Skew Tests 6 TPA-P Test Access Board DVI Source (DUT) DVI Connector + Data Channel - Test Points + - E2678A Differential Socket Probe Head Damping Resistors 1132A/1134A Probe Amplifier 54854A/54855A Oscilloscope Ch2 Ch3 Ch4 Figure 20 Probing for Inter-Pair Skew Tests (TPA Fixture) You can use any of the oscilloscope channels for the data test points. You identify the channels used for each signal in the Configuration tab of the DVI Electrical Performance Validation and Compliance Software. (The channels shown in Figure 19 and Figure 20 are just for example.) DVI Compliance Testing Methods of Implementation 41

42 6 Transmitter Inter-Pair Skew Tests D0/D1, D0/D2, D1/D2 - Skew Test Method of Implementation The transmitter skew is the time difference between the two differential signals measured at the normalized 50% crossover point. The trigger source can be either one of the differential signals. Channel- to- channel de- skew must be performed on the two oscilloscope channels used for this measurement (see Channel- to- Channel De- skew" on page 67). Test Definition Notes from the Specification Test Procedure See section 4.7.5, Transmitter Skew Measurement, in the DVI Specification Revision Start the automated testing application as described in Starting the DVI Electrical Performance Validation and Compliance Software" on page In the DVI Test application, click the Select Tests tab. 3 Navigate to the InterPairSkew group, and check the appropriate skew test. 42 DVI Compliance Testing Methods of Implementation

43 Transmitter Inter-Pair Skew Tests 6 Figure 21 Selecting InterPairSkew Tests 4 Follow the DVI Test application s task flow to set up configuration options, make oscilloscope connections, run the test, and view the test results. Table 4 InterPairSkew Test Configuration Options Configuration Option Device Description Test Fixture Description Lets you enter information about the device under test, including the device ID and any comments that you wish to include. Identifies the revision of your test fixture. Agilent recommends test fixture TPA2. D0/D1/D2 Identifies the oscilloscope channels probing data channel links D0, D1, and D2. Test Freq [Screen Resolution] Number of edges (InterPairSkew) Identifies the frequency of the DVI interface you are testing. You can select one of the frequencies from the drop-down list, or you can enter a frequency in the range from 25 to 165 MHz. Specifies the number of acquisitions to use when performing the inter-pair skew measurements. DVI Compliance Testing Methods of Implementation 43

44 6 Transmitter Inter-Pair Skew Tests PASS Condition 0.20 T pixel The pixel time T pixel is the time period of the raw clock. Measurement Algorithm Test References 1 Trigger on the rising edge of the lowest data lane number. (D0/D1 use D0, etc.). 2 Measure the time from the first signal's rising edge to the second signal s rising edge at 50% crossing. 3 Repeat step 2 for the "Number of edges", record the worst value. 4 Repeat steps 1 through 3 for the falling edge. 5 Report the worst rising skew, falling skew, and the worst among both. Section 4.3, Transmitter Electrical Specifications, in the DVI Specification Revision DVI Compliance Testing Methods of Implementation

45 N5394A DVI Electrical Performance Validation and Compliance Software Compliance Testing Methods of Implementation 7 Cable Assembly Tests Probing for Cable Assembly Eye Pattern Tests 46 D0/D1/D2 - Eye Pattern Test Method of Implementation 48 This section provides the Methods of Implementation (MOIs) for data eye pattern tests using an Agilent 54854A/54855A, or 90000A series Infiniium oscilloscope, 1132A/1134A probes, and the DVI Electrical Performance Validation and Compliance Software. Agilent Technologies 45

46 7 Cable Assembly Tests Probing for Cable Assembly Eye Pattern Tests When performing cable assembly eye pattern tests, the DVI Electrical Performance Validation and Compliance Software will prompt you to make the proper connections (also shown in Figure 22 or Figure 23). TPA2-R Adapter Board CRU Board DVI Source (DUT) DVI Receptacle DVI Plug DVI Cable Assembly DVI Plug DVI Receptacle Damping Resistors E2678A Differential Socket Probe Head + - D0/D1/D2 Test Point + - Raw Clock SMA to SMA Cables A/54855A Oscilloscope Recovered Clock SMA to BNC Cable 1132A/1134A Probe Amplifier Ch2 Ch3 Ch4 Figure 22 Probing for Cable Assembly Eye Pattern Tests (TPA2 Fixture) TPA-R Test Access Board DVI Source (DUT) DVI Receptacle DVI Plug DVI Cable Assembly DVI Plug DVI Receptacle D0/D1/D2 Test Point - + Recovered Clock SMA to BNC Cable Damping Resistors E2678A Differential Socket Probe Head 1132A/1134A Probe Amplifier 54854A/54855A Oscilloscope Ch2 Ch3 Ch4 Figure 23 Probing for Cable Assembly Eye Pattern Tests (TPA Fixture) 46 DVI Compliance Testing Methods of Implementation

47 Cable Assembly Tests 7 You can use any of the oscilloscope channels for the recovered clock and data test point. You identify the channels used for each signal in the Configuration tab of the DVI Electrical Performance Validation and Compliance Software. (The channels shown in Figure 22 and Figure 23 are just for example.) DVI Compliance Testing Methods of Implementation 47

48 7 Cable Assembly Tests D0/D1/D2 - Eye Pattern Test Method of Implementation Data eye pattern measurements are made as differential measurements using the recovered clock signal as the oscilloscope trigger source. Test Definition Notes from the Specification Test Procedure See section 4.7.6, Transmitter Eye, in the DVI Specification Revision Attach a DVI monitor to the DVI source (DUT). 2 Select a resolution for test, and display the pseudo random pattern in full screen mode. 3 Unplug the monitor, and connect the TPA2 plug adapter board directly to the output of the DVI source. 4 Confirm the frequency output of the DVI source by measuring RC+ or RCon the TPA2 plug board with the oscilloscope. 5 Set the CRU board to operate at the correct frequency. 6 Allow the CRU board to warm up for 5 minutes. 7 Attach the SMA to SMA cables from RC+ and RC- on the TPA2 plug board to the respective SMA connectors on the CRU board. 8 Start the automated testing application as described in Starting the DVI Electrical Performance Validation and Compliance Software" on page In the DVI Test application, click the Select Tests tab. 10 Navigate to the Data Eye Pattern group, and check the appropriate Eye Pattern test (D0, D1, or D2). 48 DVI Compliance Testing Methods of Implementation

49 Cable Assembly Tests 7 Figure 24 Selecting Data Eye Pattern Tests 11 Follow the DVI Test application s task flow to set up configuration options, make oscilloscope connections, run the test, and view the test results. Table 5 Cable Assembly Test Configuration Options Configuration Option Device Description Test Fixture Recovered Clock Description Lets you enter information about the device under test, including the device ID and any comments that you wish to include. Identifies the revision of your test fixture. Agilent recommends test fixture TPA2. The recovered clock from TPA will have 2.5 times the frequency of the raw clock. TPA2 however will have the same frequency for both clocks. Identifies the oscilloscope channel probing the recovered clock. D0/D1/D2 Identifies the oscilloscope channels probing data channel links D0, D1, and D2. Test Freq [Screen Resolution] Mask Type Identifies the frequency of the DVI interface you are testing. You can select one of the frequencies from the drop-down list, or you can enter a frequency in the range from 25 to 165 MHz. Choose either Cable Test Low-amplitude Eye Mask or Cable Test High-amplitude Eye Mask. DVI Compliance Testing Methods of Implementation 49

50 7 Cable Assembly Tests Table 5 Cable Assembly Test Configuration Options Configuration Option # Unit Interval Available in Debug Mode, this option specifies the number of UI (Unit Intervals) to measure in the data eye pattern test. The DVI revision 1.0 specification recommends 100,000 acquisitions to achieve 99% confidence within 1% error of the mean value. For full compliance, a measurement of 1,000,000 acquisitions must be made. Note that increasing the number of UI has a negative impact on the run time of the data eye pattern tests. Value used in Compliance Mode: 100,000. Manual Mask Adjustment Composite Eye Worst Eye Location Description Specifies whether the manual mask adjustment is needed prior to the mask test. Available in Debug Mode, this option specifies whether all data bits in a pixel need to be overlapped in the eye diagram. Value used in Compliance Mode: Off. Available in Debug Mode, this option specifies the location of the worst eye in respect to the rising edge of the recovered clock. This parameter will be ignored if Composite Eye is turned ON. Choose auto to let the application automatically find the worst eye. The worst eye is defined as the bit having the smallest pulse width. Select only 1-4 for the TPA test fixture. Value used in Compliance Mode: Auto. PASS Condition Zero mask failures. Measurement Algorithm Test References 1 Trigger on the recovered clock; the data lane is being viewed in color grade form. 2 Repeat step 1 until the # Unit Intervals has been reached. 3 The test passes if none of the waveform sample points falls into the mask failure region. Section 4.5, Cable Assembly Specifications, in the DVI Specification Revision DVI Compliance Testing Methods of Implementation

51 N5394A DVI Electrical Performance Validation and Compliance Software Compliance Testing Methods of Implementation A Calibrating the Digital Storage Oscilloscope Required Equipment for Calibration 51 Internal Calibration 52 Cable and Probe Calibration 58 Channel-to-Channel De-skew 67 This appendix describes the Agilent 54854A/54855A, or 90000A series digital storage oscilloscope calibration procedures. Required Equipment for Calibration To calibrate the 54854A/54855A, or 90000A series oscilloscope in preparation for running the DVI automated tests, you need the following equipment: Keyboard, qty = 1, Mouse, qty = 1, Precision 3.5 mm BNC to SMA male adapter, Agilent p/n , qty = 2 Calibration cable BNC shorting cap 50- ohm Coax Cable with SMA Male Connectors 24- inch or less RG316/U or similar, qty = 2, matched length. SMA T- adapter. BNC to SMA male adapter, qty = 1. Figure 25 below shows an example of the above mentioned cables and connectors needed. Agilent Technologies 51

52 A Calibrating the Digital Storage Oscilloscope Figure 25 Additional Cables and Adapters Internal Calibration This will perform an internal diagnostic and calibration cycle for the oscilloscope. For the Agilent oscilloscope, this is referred to as Calibration. This Calibration will take about 20 minutes. Perform the following steps: 1 Set up the oscilloscope with the following steps: a Connect the keyboard, mouse, and power cord to the rear of the oscilloscope. b If SigTest is being used on the oscilloscope, then connect a second monitor to the VGA connector located near the LAN port, on the rear of the oscilloscope. c Plug in the power cord. d Turn on the oscilloscope by pressing the power button located on the lower left of the front panel. e Allow the oscilloscope to warm up at least 30 minutes prior to starting the calibration procedure in step 3 below. 2 Locate and prepare the accessories that will be required for the internal calibration: a Locate the BNC shorting cap. b Locate the calibration cable. c Locate the two Agilent precision SMA/BNC adapters. d Attach one SMA adapter to one end of the calibration cable - hand tighten snugly. e Attach the other SMA adapter to the other end of the calibration cable - hand tighten snugly. 52 DVI Compliance Testing Methods of Implementation

53 Calibrating the Digital Storage Oscilloscope A 3 Referring to Figure 26 below, perform the following steps: a Click on the Utilities>Calibration menu to open the Calibration window. Click here to open the calibration window. Figure 26 Accessing the Calibration Menu. 4 Referring to Figure 27 below, perform the following steps to start the calibration: a Uncheck the Cal Memory Protect checkbox. b Click the Start button to begin the calibration. DVI Compliance Testing Methods of Implementation 53

54 A Calibrating the Digital Storage Oscilloscope Uncheck this first Then click here to start Figure 27 Oscilloscope Calibration Menu. 5 Follow the on- screen instructions: a b c You will be prompted to disconnect everything from all the inputs, click the OK button. Then, you will be prompted to connect BNC shorting cap to a specified input. Install the BNC shorting cap by pressing it on the specified input BNC, and turning right. Click the OK button after moving the BNC cap to each specified channel. Then you will be prompted to connect the calibration cable with SMA adapters between the Aux Out and a specified input, as shown in the example in Figure 28 below. Install the SMA adapter by pressing it on input BNC, and hand tightening the outer ring turning right. Click the OK button after connecting the cable as prompted. 54 DVI Compliance Testing Methods of Implementation

55 Calibrating the Digital Storage Oscilloscope A Precision SMA Adapter on Aux Out Precision SMA Adapter on Channel 1 Input Calibration Cable Figure 28 Calibration Cable Connection Example. d e f Early during the calibration of channel 1, you will be prompted to perform a Time Scale Calibration, as shown in Figure 29 below. Click on the Default button to continue the calibration, using the Factory default calibration factors. When the calibration procedure is complete, you will be prompted with a Calibration Complete message window. Click the OK button to close this window. DVI Compliance Testing Methods of Implementation 55

56 A Calibrating the Digital Storage Oscilloscope Click Default Figure 29 Time Scale Calibration Menu. 6 Referring to Figure 30 below, perform the following steps: a b c d Confirm that the Vertical and Trigger Calibration Status for all Channels passed. Click the Close button to close the calibration window. The internal calibration is completed. Read NOTE below. 56 DVI Compliance Testing Methods of Implementation

57 Calibrating the Digital Storage Oscilloscope A Delta Calibration Temperature Verify Calibration Passed Close this menu Figure 30 Calibration Status Screen. NOTE These steps do not need to be performed every time a test is run. However, if the ambient temperature changes more than 5 degrees Celsius from the calibration temperature, this calibration should be performed again. The delta between the calibration temperature and the present operating temperature is shown in the Utilities>Calibration menu. DVI Compliance Testing Methods of Implementation 57

58 A Calibrating the Digital Storage Oscilloscope Cable and Probe Calibration Perform a 50- ohm direct- coupled input calibration for the SMA interface of channel 1 and channel 3. This calibration compensates for gain, offset, and skew errors in cables and probes. Perform the following steps. 1 Referring to the Figure 31 below, perform the following steps: a b c d e f Locate and connect one of the Agilent precision SMA adapters to the Channel 1 oscilloscope input. Locate and connect the other Agilent precision SMA adapter to the Channel 3 oscilloscope input. Locate and connect one end of one of the RG- 316 cables to the SMA adapter on Channel 1. Locate and connect one end of the other RG- 316 cable to the SMA adapter on Channel 3. Locate and connect the non- Agilent SMA/BNC adapter to the Aux Out BNC on the oscilloscope. Connect the other end of the cable attached to Channel 1 to the SMA adapter on the Aux Out. Channel 1 Channel 3 Aux Out Figure 31 Vertical Input Calibration Connections (Cable on Channel 3 not shown). 58 DVI Compliance Testing Methods of Implementation

59 Calibrating the Digital Storage Oscilloscope A 2 Referring to Figure 32 below, perform the following steps: a b Click on the Setup>Channel 1 menu to open the Channel Setup window. Click the Probes button in the Channel Setup window, to open the Probe Setup window. Click Setup Channel 1 Click here for Probe Setup Menu Figure 32 Channel Setup Window. DVI Compliance Testing Methods of Implementation 59

60 A Calibrating the Digital Storage Oscilloscope 3 Referring to Figure 33 below, perform the following steps: a Click the Configure Probing System button, and then click on User Defined Probes. Click here Then click here Figure 33 Probe Setup Window. 60 DVI Compliance Testing Methods of Implementation

61 Calibrating the Digital Storage Oscilloscope A 4 Referring to Figure 34 below, perform the following steps: a Click on the Calibrate Probe button to open the Probe Calibration window. Click Here Figure 34 User Defined Probe Window. 5 Referring to Figure 35 below, perform the following steps: a b Select the Calibrated Atten/Offset Radio Button Click the Start Atten/Offset Calibration Button to open the Calibration window. DVI Compliance Testing Methods of Implementation 61

62 A Calibrating the Digital Storage Oscilloscope Select Calibrated Atten/Offset Then Click Here Figure 35 Probe Calibration Window. 6 Referring to Figure 36 shown below, perform the following steps: a Ignore the instructions shown in the dialog box. b Click the OK button on the Calibration window. c The calibration should complete in about 10 seconds. Click OK Figure 36 Calibration Window. 62 DVI Compliance Testing Methods of Implementation

63 Calibrating the Digital Storage Oscilloscope A 7 Referring to Figure 37 below, perform the following steps: a Click OK to close the Probe Calibration Done window. Click to close this window Figure 37 Probe Calibration Done Window. 8 Referring to Figure 38 below, perform the following steps: a b Select the Calibrated Skew Radio button in the Probe Calibration window Click the Start Skew Calibration button Select Calibrated Skew Then Click Here Figure 38 Probe Calibration Window. 9 Referring to Figure 39 shown below, perform the following steps: a Ignore the instructions shown in the dialog box. b Click the OK button on the Calibration window. c The calibration should complete in about 10 seconds. DVI Compliance Testing Methods of Implementation 63

64 A Calibrating the Digital Storage Oscilloscope Click OK Figure 39 Calibration Window. 10 Referring to Figure 40 below, perform the following steps: a Click OK to close the Probe Calibration Done window. Click to close this window Figure 40 Calibration Window. 64 DVI Compliance Testing Methods of Implementation

65 Calibrating the Digital Storage Oscilloscope A 11 Referring to Figure 41 below, perform the following steps: a Click the Close button to close this window. Click Close Figure 41 Calibration Window. DVI Compliance Testing Methods of Implementation 65

66 A Calibrating the Digital Storage Oscilloscope 12 Referring to Figure 42 below, perform the following steps: a Click on the Channel 3 tab. Click on Channel 3 Tab Figure 42 Calibration Window. 13 Referring to Figure 31 on page 58, perform the following steps: a Disconnect the RG- 316 cable connected to the SMA adapter on the Aux Out. b Connect the other end of the RG- 316 cable connected to the SMA adapter on Channel 3, to the SMA adapter on the Aux Out. 14 Repeat steps 3 through 11 of this section to calibrate the cable on Channel Click the Close button on the Probe Setup window (Figure 42) to close this window. 16 Click the Close button on the Channel Setup window (Figure 32 on page 59) to close this window. 17 The Cable and Probe calibration is complete. 18 Read the NOTE below. 66 DVI Compliance Testing Methods of Implementation

67 Calibrating the Digital Storage Oscilloscope A NOTE Each cable is now calibrated for the oscilloscope channel it is connected to. Do not switch cables between channels or other oscilloscopes, or it will be necessary to calibrate them again. It is recommended that the cables be labeled with the channel they were calibrated for. Channel-to-Channel De-skew This procedure ensures that the timing skew errors between channel 1 and channel 3 are minimized. Perform the following steps: 1 Referring to Figure 43 below, perform the following steps: a b c d e f Do not disconnect the RG- 316 cables from either the Channel 1 or Channel 3 SMA adapters. If not already installed, install the non- Agilent SMA adapter on the oscilloscope Aux Out. Disconnect any cable connected to the SMA adapter on the Aux Out. Locate and connect the middle branch of the SMA Tee to the SMA adapter on the Aux Out BNC. Connect the far end of the cable from the Channel 1 SMA adapter, to one branch of the SMA Tee on the Aux Out. Connect the far end of the cable from the Channel 3 SMA adapter, to the other branch of the SMA Tee on the Aux Out. DVI Compliance Testing Methods of Implementation 67

68 A Calibrating the Digital Storage Oscilloscope SMA Tee on Aux Out Channel 1 Channel 3 Figure 43 De-skew Connection. 2 Referring to Figure 44 below, perform the following steps: a b c d Select the File>Load>Setup menu to open the Load Setup window. Navigate to the directory location that contains the INF_SMA_Deskew.set setup file. If the setup file is not available, it can be created by following the instructions in Appendix B, INF_SMA_Deskew.set Setup File Details. Select the INF_SMA_Deskew.set setup file by clicking on it. Click the Load button to configure the oscilloscope from this setup file. 68 DVI Compliance Testing Methods of Implementation

69 Calibrating the Digital Storage Oscilloscope A 1. Click File Load Setup 2. Then find and select INF_SMA_Deskew.set 3. Then click here to load setup file. Figure 44 Load De-skew Setup. The oscilloscope display should look similar to Figure 45 below. A falling edge of the square wave is shown in a 200 ps/div horizontal scale. The upper portion of the screen shows channel 1 (yellow trace) and channel 3 (purple trace) superimposed on one another. The lower portion of the screen is the differential signal (green trace) of channel 1 minus channel 3. The top two traces provide for visual inspection of relative time skew between the two channels. The bottom trace provides for visual presentation of unwanted differential mode signal resulted from relative channel skew (and to a much lesser extent from other inevitable channel mismatch parameters like gain and non- linearity). Figure 45 is an example of exaggerated skew between channel 1 and channel 3, measured to be about 50 ps with the cursor. DVI Compliance Testing Methods of Implementation 69

70 A Calibrating the Digital Storage Oscilloscope Skew between Channel 1 and Channel 3 Differential signal not flat, indicating mismatch in skew. Figure 45 Channel Skew. Figure 46 below shows the desired effect of no skew between the cables. Note that the channel 1 (yellow trace), channel 3 (purple trace) traces overlap, and the differential signal (green trace) is flat. If this is not the case, then perform the following steps to reduce the skew between channels 1 and DVI Compliance Testing Methods of Implementation

71 Calibrating the Digital Storage Oscilloscope A Channel 1 and 3 traces overlap, indicating no skew error. Flat differential signal, indicating no skew error. Figure 46 Skew Minimized. 3 Referring to Figure 47, perform the following steps to de- skew the channels: a Click on the Setup>Channel 1 menu to open the Channel Setup window. b c d e f g Move the Channel Setup window to the left so you can see the traces. Adjust the Skew by clicking on the < or > arrows, to achieve the flattest response on the differential signal (green trace). Click the Close button on the Channel Setup window to close it. The de- skew operation is complete. Disconnect the cables from the Tee on the Aux Out BNC. Leave the cables connected to the Channel 1 and Channel 3 inputs. Read the NOTE below. DVI Compliance Testing Methods of Implementation 71

72 A Calibrating the Digital Storage Oscilloscope Click Setup Channel 1 to open the Channel Setup window. Then adjust the Skew left or right to maximize flatness of green trace. Then click Close when done. Figure 47 De-skewing Procedure. NOTE Each cable is now calibrated for the oscilloscope channel it is connected to. Do not switch cables between channels or other oscilloscopes, or it will be necessary to calibrate them again. It is recommended that the cables be labeled with the channel they were calibrated for. 72 DVI Compliance Testing Methods of Implementation

73 N5394A DVI Electrical Performance Validation and Compliance Software Compliance Testing Methods of Implementation B INF_SMA_Deskew.set Setup File Details If the INF_SMA_Deskew.set file is not available, you can create it by following these instructions. 1 Start from a default setup by pressing the Default Setup key on the front panel. Then configure the following settings: Acquisition Channel 1 Channel 3 Time base Trigger Averaging on number of averages 16 Interpolation on Scale mv/ Offset 350mV Coupling DC Impedance 50 Ohms Turn Channel On; Scale mv/ Offset 350m V Coupling DC Impedance 50 Ohms Scale 200 ps/sec Trigger level 173mV Slope falling Function 2 Turn on and configure for channel 1 subtract channel 3, Vertical scale 50 mv/ Offset mv Agilent Technologies 73

74 B INF_SMA_Deskew.set Setup File Details 74 DVI Compliance Testing Methods of Implementation

75 N5394A DVI Electrical Performance Validation and Compliance Software Compliance Testing Methods of Implementation C InfiniiMax Probing Figure A InfiniiMax Probe Amplifier Agilent recommends either the E2678A differential socket probe head. Figure 49 Table 6 Recommended Probe Head for the DVI Testing Probe Head Characteristics Probe Head Model Number Differential Measurement (BW, input C, input R) Single-Ended Measurement (BW, input C, input R) Differential socket E2678A 7 GHz, 0.34 pf, 50 kohm 7 GHz, 0.56 pf, 25 kohm Agilent Technologies 75

Agilent N5410A Fibre Channel Automated Test Application

Agilent N5410A Fibre Channel Automated Test Application Agilent N5410A Fibre Channel Automated Test Application Compliance Testing Methods of Implementation Agilent Technologies Notices Agilent Technologies, Inc. 2005 No part of this manual may be reproduced

More information

Agilent N5393B PCI Express Automated Test Application

Agilent N5393B PCI Express Automated Test Application Agilent N5393B PCI Express Automated Test Application Compliance Testing Methods of Implementation Agilent Technologies Notices Agilent Technologies, Inc. 2004-2009 No part of this manual may be reproduced

More information

Agilent N5393C PCI Express Automated Test Application

Agilent N5393C PCI Express Automated Test Application Agilent N5393C PCI Express Automated Test Application Compliance Testing Methods of Implementation Agilent Technologies Notices Agilent Technologies, Inc. 2004-2010 No part of this manual may be reproduced

More information

Agilent InfiniiMax II 1168A/1169A Probes

Agilent InfiniiMax II 1168A/1169A Probes Agilent InfiniiMax II 1168A/1169A Probes Handling Guide Agilent Technologies Notices Agilent Technologies, Inc. 2008 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Agilent N2739A 1000 Series Oscilloscope Rack Mount Kit

Agilent N2739A 1000 Series Oscilloscope Rack Mount Kit Agilent N2739A 1000 Series Oscilloscope Rack Mount Kit Installation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2008 No part of this manual may be reproduced in any form or by any means

More information

Agilent Series Logic Analysis System

Agilent Series Logic Analysis System Agilent 16900-Series Logic Analysis System Quick Start Guide for 16700-Series Users Agilent Technologies Notices Agilent Technologies, Inc. 2003-2004 No part of this manual may be reproduced in any form

More information

Keysight N5392B/N5392C Ethernet Compliance Test Application. Methods of Implementation

Keysight N5392B/N5392C Ethernet Compliance Test Application. Methods of Implementation Keysight N5392B/N5392C Ethernet Compliance Test Application Methods of Implementation Notices Keysight Technologies 2015-2017 No part of this manual may be reproduced in any form or by any means (including

More information

Keysight N5990A DisplayPort Extended Tests Embedded DisplayPort

Keysight N5990A DisplayPort Extended Tests Embedded DisplayPort Keysight N5990A DisplayPort Extended Tests Embedded DisplayPort Calibration and Test Procedure Descriptions User Guide Notices Keysight Technologies 2018 No part of this manual may be reproduced in any

More information

Agilent N5411B SATA6G Electrical Compliance Test Application. Online Help

Agilent N5411B SATA6G Electrical Compliance Test Application. Online Help Agilent N5411B SATA6G Electrical Compliance Test Application Online Help Notices Agilent Technologies, Inc. 2005-2010 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Agilent Technologies E5385A 100-Pin Probe

Agilent Technologies E5385A 100-Pin Probe Agilent Technologies E5385A 100-Pin Probe Installation Note The Agilent Technologies E5385A 100-pin probe provides a convenient way to connect two Agilent Technologies logic analyzer probe cables to a

More information

Agilent OpenLAB. Data Store. Backup and Restore Guide

Agilent OpenLAB. Data Store. Backup and Restore Guide Agilent OpenLAB Data Store Backup and Restore Guide Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval

More information

Agilent 1260 Infinity Purification Solution

Agilent 1260 Infinity Purification Solution Agilent 1260 Infinity Purification Solution Administrator's Quick Start Guide Agilent Technologies Notices Agilent Technologies, Inc. 2017 No part of this manual may be reproduced in any form or by any

More information

Agilent N2916B 6000 and 5000 Series Oscilloscope Rack Mount Kit

Agilent N2916B 6000 and 5000 Series Oscilloscope Rack Mount Kit Agilent N2916B 6000 and 5000 Series Oscilloscope Rack Mount Kit Installation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2005, 2007 No part of this manual may be reproduced in any form

More information

Keysight N7015A/16A Type-C Test Kit. User Guide

Keysight N7015A/16A Type-C Test Kit. User Guide Keysight N7015A/16A Type-C Test Kit User Guide Notices Keysight Technologies 2015, 2016 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval

More information

Agilent OpenLAB Chromatography Data System (CDS)

Agilent OpenLAB Chromatography Data System (CDS) Agilent OpenLAB Chromatography Data System (CDS) EZChrom Edition Reusing EZChrom Data in OpenLAB CDS with Data Store or ECM Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced

More information

Agilent CSV Export Utility

Agilent CSV Export Utility Agilent CSV Export Utility User Guide Agilent Technologies Notices Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in any form or by any means (including electronic storage and

More information

Setting Up Probes. Online Help

Setting Up Probes. Online Help Setting Up Probes Online Help Notices Agilent Technologies, Inc. 2001-2009 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation

More information

Agilent 85130C NMD 3.5 mm to Type-N Adapter Kit

Agilent 85130C NMD 3.5 mm to Type-N Adapter Kit Agilent 85130C NMD 3.5 mm to Type-N Adapter Kit Operating and Service Manual Agilent Technologies Notices Agilent Technologies, Inc. 1987, 2004, 2008 No part of this manual may be reproduced in any form

More information

Agilent Technologies E5346A 38-Pin Probe and E5351A 38-Pin Adapter Cable

Agilent Technologies E5346A 38-Pin Probe and E5351A 38-Pin Adapter Cable Agilent Technologies E5346A 38-Pin Probe and E5351A 38-Pin Adapter Cable Installation Note The 38-pin probe and adapter cable provide a convenient way to connect two Agilent Technologies logic analyzer

More information

Agilent's FlexRay Physical Layer Conformance Test Application. Online Help

Agilent's FlexRay Physical Layer Conformance Test Application. Online Help Agilent's FlexRay Physical Layer Conformance Test Application Online Help s1 Notices Agilent Technologies, Inc. 2006-2012 No part of this manual may be reproduced in any form or by any means (including

More information

Agilent ChemStation Plus

Agilent ChemStation Plus Agilent ChemStation Plus Getting Started Guide Agilent Technologies Notices Agilent Technologies, Inc. 2004 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Agilent OBSAI Protocol Tester

Agilent OBSAI Protocol Tester Agilent OBSAI Protocol Tester Hardware Reference Guide Agilent Technologies Notices Agilent Technologies, Inc. 2008 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Keysight N108xA-Series IEEE Ethernet KR/CR Compliance and Debug Application

Keysight N108xA-Series IEEE Ethernet KR/CR Compliance and Debug Application Keysight N108xA-Series IEEE 802.3 Ethernet KR/CR Compliance and Debug Application Getting Started Notices Keysight Technologies, Inc. 2014 No part of this manual may be reproduced in any form or by any

More information

Agilent OpenLAB. Data Store. Maintenance Guide

Agilent OpenLAB. Data Store. Maintenance Guide Agilent OpenLAB Data Store Maintenance Guide Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or

More information

Keysight 85130F NMD 2.4 mm to 3.5 mm Adapter Kit

Keysight 85130F NMD 2.4 mm to 3.5 mm Adapter Kit Keysight 85130F NMD 2.4 mm to 3.5 mm Adapter Kit Operating and Service Manual Notices Keysight Technologies 1988-2015 No part of this manual may be reproduced in any form or by any means (including electronic

More information

PCI Express Signal Quality Test Methodology

PCI Express Signal Quality Test Methodology PCI Express Signal Quality Test Methodology Users Guide LeCroy SDA 6000 October 2003 Revision 0.7 Document Number: XXXX DISCLAIMER OF WARRANTIES THIS SPECIFICATION IS PROVIDED AS IS AND WITH NO WARRANTIES

More information

Keysight 11878A 50 Ohm 3.5 mm Adapter Kit

Keysight 11878A 50 Ohm 3.5 mm Adapter Kit Keysight 11878A 50 Ohm 3.5 mm Adapter Kit Operating and Service Manual Notices Keysight Technologies 1989, 2004, 2013, 2014 No part of this manual may be reproduced in any form or by any means (including

More information

86100 Series Infiniium DCA Oscilloscope

86100 Series Infiniium DCA Oscilloscope 86100 Series Infiniium DCA Oscilloscope This manual provides the documentation for the following instruments 86100D 86100C User s Manual Notices Keysight Technologies, Inc. 20xx-2014 No part of this manual

More information

Agilent ChemStation Plus

Agilent ChemStation Plus Agilent ChemStation Plus Getting Started Guide Agilent Technologies Notices Agilent Technologies, Inc. 2004, 2006-2008 No part of this manual may be reproduced in any form or by any means (including electronic

More information

BenchCel Workstations Software

BenchCel Workstations Software BenchCel Workstations Software User Guide For Research Use Only. Not for use in diagnostic procedures. Original Instructions Notices Agilent Technologies, Inc. 2017 No part of this manual may be reproduced

More information

Keysight 85130D NMD 3.5 mm to 3.5 mm Adapter Kit

Keysight 85130D NMD 3.5 mm to 3.5 mm Adapter Kit Keysight 85130D NMD 3.5 mm to 3.5 mm Adapter Kit Operating and Service Manual Notices Keysight Technologies 1987-2015 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) User s Guide Publication number E2625-97003 March 2002 For Safety information and Regulatory information, see the pages

More information

Agilent OpenLAB Chromatography Data System (CDS)

Agilent OpenLAB Chromatography Data System (CDS) Agilent OpenLAB Chromatography Data System (CDS) EZChrom Edition Compatibility Guide for Agilent Instrument Controllers Agilent Technologies Notices Agilent Technologies, Inc. 2010 No part of this manual

More information

Agilent EZChrom SI. Startup Guide

Agilent EZChrom SI. Startup Guide Agilent EZChrom SI Startup Guide Notices Copyright Scientific Software, Inc 2003 Agilent Technologies, Inc. 2006. No part of this manual may be reproduced in any form or by any means (including electronic

More information

Agilent N4219B Packet Analysis Probe for Serial ATA. User s Guide

Agilent N4219B Packet Analysis Probe for Serial ATA. User s Guide Agilent N4219B Packet Analysis Probe for Serial ATA User s Guide A Notices Agilent Technologies, Inc. 2002-2005 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Agilent Dissolution Tester Driver

Agilent Dissolution Tester Driver Agilent Dissolution Tester Driver Agilent Technologies Notices Agilent Technologies, Inc. 2011-2012 No part of this manual may be reproduced in any form or by any means (including electronic storage and

More information

Agilent OpenLAB Chromatography Data System (CDS)

Agilent OpenLAB Chromatography Data System (CDS) Agilent OpenLAB Chromatography Data System (CDS) ChemStation Edition Reusing Data from ChemStation B.04.03 SP1 and SP2 in OpenLAB CDS ChemStation Edition Notices Agilent Technologies, Inc. 2012-2013, 2014

More information

Keysight N5990A Type-C Extension. User Guide

Keysight N5990A Type-C Extension. User Guide Keysight N5990A Type-C Extension User Guide Notices Keysight Technologies 2017 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation

More information

Agilent OpenLAB Chromatography Data System

Agilent OpenLAB Chromatography Data System Agilent OpenLAB Chromatography Data System EZChrom Edition EZChrom Elite and ICM Migration Guide Agilent Technologies Notices Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in

More information

Firmware Guide. Keysight PXIe Chassis Family

Firmware Guide. Keysight PXIe Chassis Family Firmware Guide Keysight PXIe Chassis Family Notices Keysight Technologies, Inc. 2018 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or

More information

Agilent OpenLAB Chromatography Data System (CDS)

Agilent OpenLAB Chromatography Data System (CDS) Agilent OpenLAB Chromatography Data System (CDS) ChemStation Edition AIC (M8304AA) Setup Guide Agilent Technologies Notices Agilent Technologies, Inc. 2010-2014, 2015 No part of this manual may be reproduced

More information

Agilent OpenLAB Data Analysis Upload Download Tool. User s Guide

Agilent OpenLAB Data Analysis Upload Download Tool. User s Guide Agilent OpenLAB Data Analysis Upload Download Tool User s Guide Notices Agilent Technologies, Inc. 2014 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Agilent OpenLAB Chromatography Data System

Agilent OpenLAB Chromatography Data System Agilent OpenLAB Chromatography Data System EZChrom Elite Edition Migration Guide Agilent Technologies Notices Agilent Technologies, Inc. 2010 No part of this manual may be reproduced in any form or by

More information

Agilent OpenLAB Data Store. Maintenance Guide

Agilent OpenLAB Data Store. Maintenance Guide Agilent OpenLAB Data Store Maintenance Guide Notices Agilent Technologies, Inc. 2013 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or

More information

Keysight 11636B APC-3.5 mm Power Divider DC to 26.5 GHz

Keysight 11636B APC-3.5 mm Power Divider DC to 26.5 GHz Keysight 11636B APC-3.5 mm Power Divider DC to 26.5 GHz Operating and Service Manual Notices Keysight Technologies 1985, 2013, 2014 No part of this manual may be reproduced in any form or by any means

More information

Quick Start/Installation Guide

Quick Start/Installation Guide Agilent Technologies E5850A Time Correlation Fixture Quick Start/Installation Guide The Agilent E5850A time correlation fixture allows you to make time-correlated measurements between a 1680/90 or 16700

More information

BIT-3000 Dynamic Sequencing Generator and Analyzer. Datasheet 1.11

BIT-3000 Dynamic Sequencing Generator and Analyzer. Datasheet 1.11 BIT-3000 Dynamic Sequencing Generator and Analyzer Datasheet 1.11 BitifEye Digital Test Solutions GmbH Herrenberger Strasse 130 71034 Boeblingen, Germany info@bitifeye.com www.bitifeye.com Notices BitifEye

More information

Agilent Technologies. Connectivity Guide. USB/LAN/GPIB Interfaces. Agilent Technologies

Agilent Technologies. Connectivity Guide. USB/LAN/GPIB Interfaces. Agilent Technologies Agilent Technologies USB/LAN/GPIB Interfaces Connectivity Guide Agilent Technologies Notices Agilent Technologies, Inc. 2003-2006 No part of this manual may be reproduced in any form or by any means (including

More information

Agilent 89600B VSA. Software Installation Guide

Agilent 89600B VSA. Software Installation Guide Agilent 89600B VSA Software Installation Guide Notices Agilent Technologies, Inc. 2000-2011 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval

More information

Agilent Lab Advisor. IT Administrator's Guide. Agilent Technologies

Agilent Lab Advisor. IT Administrator's Guide. Agilent Technologies Agilent Lab Advisor IT Administrator's Guide Agilent Technologies Notices Agilent Technologies, Inc. 2016 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

U4431 MIPI M-PHY Protocol Analyzer. User Guide

U4431 MIPI M-PHY Protocol Analyzer. User Guide U4431 MIPI M-PHY Protocol Analyzer User Guide Notices Agilent Technologies, Inc. 2014 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or

More information

Keysight Second Source, Combiner, and Mechanical Switches Upgrade Kit

Keysight Second Source, Combiner, and Mechanical Switches Upgrade Kit Keysight Second Source, Combiner, and Mechanical Switches Upgrade Kit To Upgrade PNA-X N5241A, N5242A, or N5249A Option 419 to Option 423 Upgrade Kit Order Number: N5241AU- 927, N5242AU- 927, and N5249AU-

More information

Agilent E2094M IO Libraries

Agilent E2094M IO Libraries Agilent E2094M IO Libraries Agilent IO Libraries Installation and Configuration Guide Agilent Technologies Notices Agilent Technologies, Inc. 2003 No part of this manual may be reproduced in any form or

More information

Agilent Technologies N5393A PCI Express Electrical Performance Validation and Compliance Software for Infiniium 54855A or Series Oscilloscopes

Agilent Technologies N5393A PCI Express Electrical Performance Validation and Compliance Software for Infiniium 54855A or Series Oscilloscopes Agilent Technologies N5393A PCI Express Electrical Performance Validation and Compliance Software for Infiniium 54855A or 80000 Series Oscilloscopes Data Sheet Verify and debug your PCI Express designs

More information

Microscan Barcode Reader

Microscan Barcode Reader Microscan Barcode Reader Device Driver Guide For Research Use Only. Not for use in diagnostic procedures. Original Instructions Notices Agilent Technologies, Inc. 2017 No part of this manual may be reproduced

More information

Agilent ChemStation for UV-visible Spectroscopy

Agilent ChemStation for UV-visible Spectroscopy Agilent ChemStation for UV-visible Spectroscopy Understanding Your Biochemical Analysis Software Agilent Technologies Notices Agilent Technologies, Inc. 2000, 2003-2008 No part of this manual may be reproduced

More information

DDR Setup Assistant. Online Help

DDR Setup Assistant. Online Help DDR Setup Assistant Online Help Notices Agilent Technologies, Inc. 2014 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into

More information

Agilent E2943A/E2944A ATCA Probes for Advanced Switching Interconnect

Agilent E2943A/E2944A ATCA Probes for Advanced Switching Interconnect Agilent E2943A/E2944A ATCA Probes for Advanced Switching Interconnect Hardware Setup Guide Agilent Technologies Notices Agilent Technologies, Inc. 2005 No part of this manual may be reproduced in any form

More information

Agilent N5430A User Defined Function

Agilent N5430A User Defined Function Agilent N5430A User Defined Function User s and Programmer s Guide Agilent Technologies Notices Agilent Technologies, Inc. 2008 No part of this manual may be reproduced in any form or by any means (including

More information

Getting Started with the Agilent Serial BERT N4906B. You only need a few minutes to get started with the Serial BERT.

Getting Started with the Agilent Serial BERT N4906B. You only need a few minutes to get started with the Serial BERT. Getting Started with the Agilent Serial BERT N4906B You only need a few minutes to get started with the Serial BERT. This Getting Started Brochure helps you to quickly understand the operating principles

More information

Keysight Technologies N5394A DVI Electrical Performance Validation and Compliance Software for Infiniium Oscilloscopes. Data Sheet

Keysight Technologies N5394A DVI Electrical Performance Validation and Compliance Software for Infiniium Oscilloscopes. Data Sheet Keysight Technologies N5394A DVI Electrical Performance Validation and Compliance Software for Infiniium Oscilloscopes Data Sheet 02 Keysight N5394A DVI Electrical Performance Validation and Compliance

More information

Keysight Add Source and Receiver Attenuators Upgrade Kit

Keysight Add Source and Receiver Attenuators Upgrade Kit Keysight Add Source and Receiver Attenuators Upgrade Kit To Upgrade PNA N5221A or N5222A Option 401 to Option 417 Upgrade Kit Order Numbers: N5221AU-417 or N5222AU-417 Keysight Kit Number: N5222-60107

More information

Installation Note. Source Attenuators and Bias Tees Upgrade Kit. For E8362B/C PNA Series Microwave Network Analyzers. Network Analyzer Model Number

Installation Note. Source Attenuators and Bias Tees Upgrade Kit. For E8362B/C PNA Series Microwave Network Analyzers. Network Analyzer Model Number Installation Note Source Attenuators and Bias Tees Upgrade Kit For E8362B/C PNA Series Microwave Network Analyzers Network Analyzer Model Number Upgrade Kit Part Number E8362B/C E8362-60115 Agilent Part

More information

Keysight E7515A UXM Firmware Update Tool

Keysight E7515A UXM Firmware Update Tool Keysight E7515A UXM Firmware Update Tool For the E7515A UXM Wireless Test Set Installation & User Guide Notices Keysight Technologies, Inc. 2014-2017 No part of this manual may be reproduced in any form

More information

Agilent Gb/s Serial BERT

Agilent Gb/s Serial BERT Agilent 3.125 Gb/s Serial BERT N5980A Programming Guide Notices Agilent Technologies, Inc. 2007 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval

More information

Agilent OpenLAB Chromatography Data System (CDS)

Agilent OpenLAB Chromatography Data System (CDS) Agilent OpenLAB Chromatography Data System (CDS) Hardware and Software Requirements Agilent Technologies Notices Agilent Technologies, Inc. 2013-2014 No part of this manual may be reproduced in any form

More information

Agilent Technologies E5339A 38-Pin Low-Voltage Probe

Agilent Technologies E5339A 38-Pin Low-Voltage Probe Agilent Technologies E5339A 38-Pin Low-Voltage Probe Installation Note The 38-pin low-voltage probe provides a convenient way to connect two Agilent Technologies logic analyzer probe cables to a small

More information

Agilent 34826A BenchLink Data Logger for 34980A. Getting Started Guide. Agilent Technologies

Agilent 34826A BenchLink Data Logger for 34980A. Getting Started Guide. Agilent Technologies Agilent 34826A BenchLink Data Logger for 34980A Getting Started Guide Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or by any means (including

More information

Automated Plate Labeling

Automated Plate Labeling Automated Plate Labeling Protocol Guide For Research Use Only. Not for use in diagnostic procedures. Original Instructions Notices Agilent Technologies, Inc. 2017 No part of this manual may be reproduced

More information

Keysight Technologies N6468A SFP+ Electrical Performance Validation and Conformance Software

Keysight Technologies N6468A SFP+ Electrical Performance Validation and Conformance Software Keysight Technologies N6468A SFP+ Electrical Performance Validation and Conformance Software For Infiniium Oscilloscopes Data Sheet 02 Keysight N6468A SFP+ Electrical Performance Validation and Conformance

More information

Agilent Protocol Analyzer and Jammer for USB. Quick Start Guide

Agilent Protocol Analyzer and Jammer for USB. Quick Start Guide Agilent Protocol Analyzer and Jammer for USB Quick Start Guide Notices Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Keysight DSOXT3NFC/ DSOX4NFC Near Field Communication Test Application

Keysight DSOXT3NFC/ DSOX4NFC Near Field Communication Test Application Keysight DSOXT3NFC/ DSOX4NFC Near Field Communication Test Application Online Help Notices Keysight Technologies, Inc. 2006-2016 No part of this manual may be reproduced in any form or by any means (including

More information

Keysight M9393A PXIe Vector Signal Analyzer

Keysight M9393A PXIe Vector Signal Analyzer Security Guide Keysight M9393A PXIe Vector Signal Analyzer. Notice: This document contains references to Agilent. Please note that Agilent s Test and Measurement business has become Keysight Technologies.

More information

Keysight E5864A Removable Hard Drive for Series Logic Analyzers. Installation Guide

Keysight E5864A Removable Hard Drive for Series Logic Analyzers. Installation Guide Keysight E5864A Removable Hard Drive for 16850-Series Logic Analyzers Installation Guide Notices Keysight Technologies 2013-2014 No part of this manual may be reproduced in any form or by any means (including

More information

Keysight N5990A-155 Display Port. User Guide

Keysight N5990A-155 Display Port. User Guide Keysight N5990A-155 Display Port User Guide Notices Keysight Technologies 2017 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation

More information

Agilent Softtouch 2.0 Midbus Probe

Agilent Softtouch 2.0 Midbus Probe Agilent Softtouch 2.0 Midbus Probe User Guide Agilent Technologies Notices Agilent Technologies, Inc. 2008 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Agilent ChemStation. ECM Interface Guide. Agilent Technologies

Agilent ChemStation. ECM Interface Guide. Agilent Technologies Agilent ChemStation ECM Interface Guide Agilent Technologies Notices Agilent Technologies, Inc. 2004, 2005-2007 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Installation Note. Removable Hard Disk Drive Upgrade Kit. For All PNA Series RF Network Analyzers. Network Analyzer Model Number

Installation Note. Removable Hard Disk Drive Upgrade Kit. For All PNA Series RF Network Analyzers. Network Analyzer Model Number Installation Note Removable Hard Disk Drive Upgrade Kit For All PNA Series RF Network Analyzers Network Analyzer Model Number E8356A, E8357A, E8358A E8801A, E8802A, E8803A N3381A, N3382A, N3383A Upgrade

More information

Agilent OpenLAB. Agilent Instrument Controller Installation and Configuration Guide

Agilent OpenLAB. Agilent Instrument Controller Installation and Configuration Guide Agilent OpenLAB Agilent Instrument Controller Installation and Configuration Guide Notices Copyright Scientific Software, Inc 1997-2005 Agilent Technologies, Inc. 2005-2007. No part of this manual may

More information

Agilent CytoGenomics 2.5

Agilent CytoGenomics 2.5 Agilent CytoGenomics 2.5 Installation and Administration Guide Research Use Only. Not for Diagnostic Procedures. Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may

More information

Keysight Second Source, Combiner, and Mechanical Switches Upgrade Kit

Keysight Second Source, Combiner, and Mechanical Switches Upgrade Kit Keysight Second Source, Combiner, and Mechanical Switches Upgrade Kit To Upgrade PNA-X N5241A or N5242A Option 219 to Option 224 For Analyzers with Serial Numbers Prefixed MY/SG/ US5201 and Above Upgrade

More information

Keysight N8814A 10GBASE-KR Ethernet Backplane Electrical Performance Validation and Conformance

Keysight N8814A 10GBASE-KR Ethernet Backplane Electrical Performance Validation and Conformance Keysight N8814A 10GBASE-KR Ethernet Backplane Electrical Performance Validation and Conformance For Infiniium Oscilloscopes Data Sheet 02 Keysight N8814A 10GBASE-KR Ethernet Backplane Electrical Performance

More information

DDR/LPDDR Custom Configuration Creator Tool. User s Guide

DDR/LPDDR Custom Configuration Creator Tool. User s Guide DDR/LPDDR Custom Configuration Creator Tool User s Guide Notices Keysight Technologies 2008-2014 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval

More information

Nimsoft Monitor. proxy Guide. v3.1 series

Nimsoft Monitor. proxy Guide. v3.1 series Nimsoft Monitor proxy Guide v3.1 series Legal Notices Copyright 2012, CA. All rights reserved. Warranty The material contained in this document is provided "as is," and is subject to being changed, without

More information

Agilent W2635A and W2636A

Agilent W2635A and W2636A Agilent W2635A and W2636A DDR3 BGA Probe Adapters for Oscilloscopes User s Guide Notices Agilent Technologies, Inc. 2007, 2008 No part of this manual may be reproduced in any form or by any means (including

More information

G5584A Labware MiniHub

G5584A Labware MiniHub G5584A Labware MiniHub User Guide For Research Use Only. Not for use in diagnostic procedures. Original Instructions Notices Agilent Technologies, Inc. 2017 No part of this manual may be reproduced in

More information

Installation Note. IF Access Upgrade Kit. For PNA Series Microwave Network Analyzers. Network Analyzer Model Number. Upgrade Kit Part Number

Installation Note. IF Access Upgrade Kit. For PNA Series Microwave Network Analyzers. Network Analyzer Model Number. Upgrade Kit Part Number Installation Note IF Access Upgrade Kit For PNA Series Microwave Network Analyzers Network Analyzer Model Number E8361A/C, E8362B/C, E8363B/C, E8364B/C Upgrade Kit Part Number E8362-60116 Agilent Part

More information

Agilent OpenLAB ECM Intelligent Reporter

Agilent OpenLAB ECM Intelligent Reporter Agilent OpenLAB ECM Intelligent Reporter Installation and Configuration Guide Agilent Technologies Notices Agilent Technologies, Inc. 2007-2016 No part of this manual may be reproduced in any form or by

More information

Agilent Genomic Workbench 6.0

Agilent Genomic Workbench 6.0 Agilent Genomic Workbench 6.0 Standard Edition Installation Guide Notices Agilent Technologies, Inc. 2010 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Keysight Software

Keysight Software Keysight 89600 Software Notice: This document contains references to Agilent. Please note that Agilent s Test and Measurement business has become Keysight Technologies. For more information, go to www.keysight.com.

More information

Keysight N5990A Test Automation Software Platform

Keysight N5990A Test Automation Software Platform Keysight N5990A Test Automation Software Platform Installation Guide Notices Copyright Notice Keysight Technologies 2016 No part of this manual may be reproduced in any form or by any means (including

More information

Keysight Technologies N6467A BroadR-Reach Automotive Ethernet Electrical Compliance Application

Keysight Technologies N6467A BroadR-Reach Automotive Ethernet Electrical Compliance Application Ihr Spezialist für Mess- und Prüfgeräte Keysight Technologies N6467A BroadR-Reach Automotive Ethernet Electrical Compliance Application For Infiniium Oscilloscopes Data Sheet Easy and accurate BroadR-Reach

More information

OpenLAB CDS ChemStation Edition

OpenLAB CDS ChemStation Edition OpenLAB CDS ChemStation Edition Guide for Administrators Guide for Administrators Agilent Technologies Notices Agilent Technologies, Inc. 2010-2017 No part of this manual may be reproduced in any form

More information

Agilent Technologies 16910/11A Logic Analyzers

Agilent Technologies 16910/11A Logic Analyzers Service Guide Publication number 16910-97000 April 2004 For Safety and Regulatory information, see the pages at the end of the book. Copyright Agilent Technologies 2001-2004 All Rights Reserved. Agilent

More information

Keysight U7231B, U7231C DDR3 and LPDDR3 Compliance Test Application For Infiniium Series Oscilloscopes DATA SHEET

Keysight U7231B, U7231C DDR3 and LPDDR3 Compliance Test Application For Infiniium Series Oscilloscopes DATA SHEET Keysight U7231B, U7231C DDR3 and LPDDR3 Compliance Test Application For Infiniium Series Oscilloscopes DATA SHEET Test, Debug and Characterize Your DDR3 and LPDDR3 Designs Quickly and Easily The Keysight

More information

E2655C Probe Deskew and Performance Verification Kit

E2655C Probe Deskew and Performance Verification Kit Keysight E2655C Probe Deskew and Performance Verification Kit User s Guide Calibration for the Solder-in and Socketed Probe Heads 3 Calibration for Hand-held Browser Probe Heads 10 Calibrating the InfiniiMax

More information

Nimsoft Monitor. ocs_monitor Guide. v1.3 series

Nimsoft Monitor. ocs_monitor Guide. v1.3 series Nimsoft Monitor ocs_monitor Guide v1.3 series Legal Notices Copyright 2012, Nimsoft Corporation Warranty The material contained in this document is provided "as is," and is subject to being changed, without

More information

N1014A SFF-8431 (SFP+)

N1014A SFF-8431 (SFP+) DATA SHEET N1014A SFF-8431 (SFP+) Compliance and Debug Application for 86100D DCA-X and N109X DCA-M Oscilloscopes Be Confident With Compliant Measurements Easy-to-use oscilloscope application that lets

More information

Agilent SureCall 3.5. Installation Guide. Agilent Technologies. For Research Use Only. Not for use in diagnostic procedures.

Agilent SureCall 3.5. Installation Guide. Agilent Technologies. For Research Use Only. Not for use in diagnostic procedures. Agilent SureCall 3.5 Installation Guide For Research Use Only. Not for use in diagnostic procedures. Agilent Technologies Notices Agilent Technologies, Inc. 2016 No part of this manual may be reproduced

More information

Agilent U4154A Logic Analyzer

Agilent U4154A Logic Analyzer Agilent U4154A Logic Analyzer Service Guide Agilent Technologies Notices Agilent Technologies, Inc. 2013 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information