URGENT - Medical Device Correction IntelliVue MX40 Missing Warnings in IFU

Size: px
Start display at page:

Download "URGENT - Medical Device Correction IntelliVue MX40 Missing Warnings in IFU"

Transcription

1 Dear Customer, A problem has been found with the Philips IntelliVue MX40 Instructions for Use (IFU) for software revisions B.05, B.06 and B.06.5X. Your IntelliVue MX40 remains safe to use. These IFUs are missing warning statements related to monitoring paced patients and the interpretation of QT/QTc measurements that were present in earlier revisions of the IFU. A Warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to observe a warning may result in death or serious injury to the user or patient. This Field Safety Notice is intended to inform you about: what the problem is and under what circumstances it can occur the actions that should be taken by the customer/user in order to prevent risks for patients the actions planned by Philips to correct the problem. This document contains important information for the continued safe and proper use of your equipment Please review the following information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication. Please refer to the following pages, which provide information on the missing warnings and instructions for actions to be taken. Follow the Action to be taken by Customer/User section of this notice. This issue has been reported to the appropriate regulatory agencies. I sincerely regret the inconvenience that this may cause you. Your satisfaction with Philips products and with our response to this issue is very important to us. Please contact your local Philips representative under with questions or concerns about this correction. Sincerely, Kristen Phillips Head of Quality & Regulatory Affairs, Andover

2 AFFECTED PRODUCTS The Philips IntelliVue MX40 revisions B.05, B.06 and B.06.5X. Model Product Number IntelliVue MX PROBLEM DESCRIPTION HAZARD INVOLVED HOW TO IDENTIFY AFFECTED PRODUCTS ACTIONS PLANNED BY PHILIPS Five warning statements are missing from the IntelliVue MX40 IFU for software revisions B.05, B.06 and B.06.5X. If users are unaware of the hazards or limitations described in the missing warning statements, they may not properly assess or provide appropriate treatment to a patient being monitored using an IntelliVue MX40. IntelliVue MX40 Instructions for Use for software revisions B.05, B.06 and B.06.5X. Philips is voluntarily initiating a correction consisting of: Distribution of this Field Safety Notice (). Release of an IFU Errata that provides the missing warnings A representative will contact customers with affected IFU revisions to provide a copy of the errata. This FSN includes the Errata providing the missing warning statements. ACTION TO BE TAKEN BY CUSTOMER / USER The enclosed MX40 IFU Errata Sheet must be attached to the first page of Chapter 6 of the Instructions for Use for ready reference. Complete and return the attached Customer Reply Form.

3 Contact Name Telephone Number Address Facility Name Street Address City, State, Zip Customer Reply for IntelliVue MX40 IFU Please complete and fax to Please fax or this completed form to the number or address provided above. CUSTOMER ACKNOWLEDGEMENT: The MX40 IFU Errata must be attached to the first page of Chapter 6 to ensure that it is not misplaced and is stored with the Instructions for Use for ready reference. CUSTOMER NAME (please print) TITLE CUSTOMER SIGNATURE DATE Please fax or the completed reply form to customercare.ch@philips.com. If you experience difficulty carrying out the instructions contained in this communication, contact your local Philips representative:

4 IntelliVue MX40 Release B.05 or Later Instructions for Use Errata This Errata is a supplement to the IntelliVue MX40 Instructions for Use Release B.05 and later. It contains important safety information. Attach this Errata to the first page of Chapter 6 to ensure that it is not misplaced and is stored with the Instructions for Use for ready reference. Part Number Printed in the U.S.A. First Edition * *

5 6 ECG and Arrhythmia Monitoring English 6 ECG and Arrhythmia Monitoring ECG Safety Information Warning The device provides QT and QTc interval change information; the clinical significance of the QT and QTc interval change information should be determined by a clinician. For more information, see the QT Interval Monitoring Application Note, p/n For Paced Patients Warning The output power of the MX40 and other sources of radio frequency energy, when used in the proximity of a pacemaker, can be sufficient to interfere with pacemaker performance. Due to the shielding effects of the body, internal pacemakers are somewhat less vulnerable than external pacemakers. However, caution should be exercised when monitoring any paced patient. In order to minimize the possibility of interference, position electrodes, electrode wires, and the MX40 as far away from the pacemaker as possible. Consult the pacemaker manufacturer for information on the RF susceptibility of their products and the use of their products with the MX40. See the Patient Information Center Instructions for Use for additional information on monitoring paced patients. When an external pacemaker is being used on a patient, arrhythmia monitoring is severely compromised due to the high energy level in the pacer pulse. This may result in the arrhythmia algorithm's failure to detect pacemaker non-capture or asystole. Pacemakers that create fusion beats (pace pulse on top of the QRS complex) cannot be detected by the monitor's QRS detector. For paced patients who exhibit only intrinsic rhythm, the monitor can erroneously count pace pulses as QRS complexes when the algorithm first encounters them, resulting in missed detection of cardiac arrest. The risk of missing cardiac arrest can be reduced by monitoring these patients with the low heart rate limit at or slightly above the basic/ demand pacemaker rate. A low heart rate alarm notifies you when the patient begins pacing. Proper detection and classification of the paced rhythm can then be determined. Note During defibrillation, monitoring may be temporarily interrupted or distorted. It may take several seconds for the ECG trace to reappear on the screen. After defibrillation, the device will continue to monitor as before; the device settings will not be affected. 6-1

URGENT Field Safety Notice Philips HeartStart MRx Monitor/Defibrillator

URGENT Field Safety Notice Philips HeartStart MRx Monitor/Defibrillator Philips Healthcare - 1/5 - FSN86100160A Dear Customer, Philips has identified issues that could impact the safety and/or performance of certain MRx monitor/defibrillators. These issues are further detailed

More information

This document contains important information for the continued safe and proper use of your equipment

This document contains important information for the continued safe and proper use of your equipment Philips Healthcare - 1/6 - FSN86100179A Dear Customer, Philips has identified the possibility of abnormal device behaviors on the HeartStart MRx Monitor/Defibrillator. Under exceptionally rare circumstances,

More information

URGENT FIELD SAFETY NOTICE

URGENT FIELD SAFETY NOTICE URGENT FIELD SAFETY NOTICE Affected Devices: Rythmic Evolution product family Type of Action: Field safety corrective action Ref.: FSN2018-01 Date: 12 March 2018 Attention: Distributors, Biomedical Professionals,

More information

This document contains important information for the continued safe and proper use of your equipment

This document contains important information for the continued safe and proper use of your equipment Philips Healthcare - 1/6 - FSN86100179A Dear Customer, Philips has identified the possibility of abnormal device behaviors on the HeartStart MRx Monitor/Defibrillator. Under exceptionally rare circumstances,

More information

Always in touch. IntelliVue Telemetry System with Smart-hopping technology, surveillance of ambulatory cardiac patients

Always in touch. IntelliVue Telemetry System with Smart-hopping technology, surveillance of ambulatory cardiac patients Always in touch with Smart-hopping technology, surveillance of ambulatory cardiac patients Real clinical network strength includes flexibility When you have mobile cardiac patients who need constant monitoring,

More information

URGENT FIELD SAFETY NOTICE Reserve battery and new instructions for users of the PAD 300/PAD 300P (Public Access Defibrillator)

URGENT FIELD SAFETY NOTICE Reserve battery and new instructions for users of the PAD 300/PAD 300P (Public Access Defibrillator) 14 th September 2012 URGENT FIELD SAFETY NOTICE Reserve battery and new instructions for users of the PAD 300/PAD 300P (Public Access Defibrillator) Dear Sir or Madam, The purpose of this letter is to

More information

Philips HeartStart XL Defibrillator Manual

Philips HeartStart XL Defibrillator Manual Philips HeartStart XL Defibrillator Manual Get an original copy of the Philips HeartStart XL Defibrillator Operators Guide for manufacturer information about service, available accessories and how to use

More information

CARELINK NETWORK UPDATES SUMMARY OF ENHANCEMENTS May 2017

CARELINK NETWORK UPDATES SUMMARY OF ENHANCEMENTS May 2017 CARELINK NETWORK UPDATES SUMMARY OF ENHANCEMENTS May 2017 Quick Look Update for Reveal LINQ ICM Patient Monitoring CareLink Network Experience With feedback from over 200 clinicians worldwide, Medtronic

More information

Important Medical Device Advisory

Important Medical Device Advisory Important Medical Device Advisory Battery Performance Alert and Cybersecurity Firmware Updates for Certain ICD & CRT-D Devices April 16, 2018 Dear Doctor, As part of a planned series of system updates

More information

Product codes with prefix (all variants): 8001, 8002, 8003, 8004, and ATTENTION: Clinical Personnel, Risk Managers, Biomedical Personnel

Product codes with prefix (all variants): 8001, 8002, 8003, 8004, and ATTENTION: Clinical Personnel, Risk Managers, Biomedical Personnel CareFusion Switzerland 317 Sàrl A-One Business Centre Zone d activités, Vers-la-Pièce No 10 Rolle, CH-1180 Switzerland carefusion.com Name Address URGENT FIELD SAFETY NOTICE Product Name: Alaris / Asena

More information

EC-12RT Resting ECG Machine. Technical Specification

EC-12RT Resting ECG Machine. Technical Specification 2017 ECG input Input impedance Input bias current Input dynamic range 10 MOhm 0,1μА from 0,03mV up to 10mV (peak-to-peak amplitude) Nonlinearity ±2% Electrode offset tolerance Common-mode rejection Internal

More information

Important Medical Device Advisory NM Implantable Pulse Generator Inaccurate Elective Replacement Indicator

Important Medical Device Advisory NM Implantable Pulse Generator Inaccurate Elective Replacement Indicator Important Medical Device Advisory NM Implantable Pulse Generator Inaccurate Elective Replacement Indicator 12 September 2017 Dear Physician, In an effort to keep you continually informed, and to ensure

More information

Use Of Mobile Communication Devices Within Healthcare Premises Policy

Use Of Mobile Communication Devices Within Healthcare Premises Policy Use Of Mobile Communication Devices Within Healthcare Premises Policy Co-ordinator: Director of Facilities Reviewer: Working Group chaired by Director of Facilities Approver: GAPF Signature Signature Signature

More information

!"# $ # # $ $ % $ &% $ '"# $ ()&*&)+(( )+(( )

!# $ # # $ $ % $ &% $ '# $ ()&*&)+(( )+(( ) !"# # # % &% '"# ) !#, ' "# " "# -. / # 0 0 0 0 0 "0 "# " # 1 #! " " 0 0 0 0 0 0 2# 0 # # 3 ' 4 56 7-56 87 9# 5 6 7 6 & 0 " : 9 ; 4 " #! 0 - '% # % "# " "# " < 4 "! % " % 4 % % 9# 4 56 87 = 4 > 0 " %!#

More information

Important Medical Device Advisory NM Implantable Pulse Generator Inaccurate Elective Replacement Indicator

Important Medical Device Advisory NM Implantable Pulse Generator Inaccurate Elective Replacement Indicator Important Medical Device Advisory NM Implantable Pulse Generator Inaccurate Elective Replacement Indicator 12 September 2017 Dear Physician, In an effort to keep you continually informed, and to ensure

More information

4P S. Wireless LAN Connecting Adapter <BRP069A42> Installation Manual

4P S. Wireless LAN Connecting Adapter <BRP069A42> Installation Manual 4P359542-2S Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com For details

More information

URGENT MEDICAL DEVICE FIELD SAFETY NOTICE- REMOVAL

URGENT MEDICAL DEVICE FIELD SAFETY NOTICE- REMOVAL June 19, 2017 To: Subject: Surgeons/ Hospitals/ Clinics URGENT MEDICAL DEVICE FIELD SAFETY NOTICE- REMOVAL Affected Product: Trauma, Guide Wires 70cm Image 1. Guide wire with protector Image 2. 3mm*70cms

More information

IntelliVue Patient Monitor MX450, MX500, MX550. Remote Application Setup. Release K.2 with Software Revision K.2. Patient Monitoring

IntelliVue Patient Monitor MX450, MX500, MX550. Remote Application Setup. Release K.2 with Software Revision K.2. Patient Monitoring Installation and Configuration Guide IntelliVue Patient Monitor MX450, MX500, MX550 Remote Application Setup Release K.2 with Software Revision K.2 Patient Monitoring 1Table of Contents 1 Introduction

More information

URGENT: MEDICAL DEVICE RECALL

URGENT: MEDICAL DEVICE RECALL Wednesday, September 20, 2017 [Contact Name] Andrew Lindsay QM Manager [Business Name] [Customer Address] [City, State, Zip Code] Carl Zeiss Ltd. 509 Coldhams Lane Cambridge Cambridgeshire CB1 3JS U.K.

More information

FIELD CORRECTIVE ACTION NOTICE

FIELD CORRECTIVE ACTION NOTICE FIELD CORRECTIVE ACTION NOTICE Leica M525 F20 Microscope System FCA Identifier: CAPA-HER-MD-18-001 Enhancement of instructions for use and additional markings to prevent inadvertent tipping Customer Name

More information

cubestress your profession our mission

cubestress your profession our mission The CARDIOLINE wireless stress testing system combines combina the high performance of the software and of the acquisition unit clickecgbt. is the complete solution to manage ECG Stress tests. combines

More information

Urgent Field Safety Notice

Urgent Field Safety Notice Urgent Field Safety Notice Commercial name/model: WATO EX-65, WATO EX-55, A5 and A7 anesthesia machines FSCA-identifier: CP17-MZ0036 Type of action: Safety Notice and Device modification - June, 2017 Attention:

More information

SpO 2 Upgrade Instructions

SpO 2 Upgrade Instructions SpO 2 Upgrade Instructions M3535A/M3536A HeartStart MRx Notice About This Edition Edition 3 Printed in the USA Publication number M3535-90905 The information in this document applies to the HeartStart

More information

TIP CARD MODEL TEMPORARY EXTERNAL PACEMAKER. Model Single Chamber Temporary Pacemaker

TIP CARD MODEL TEMPORARY EXTERNAL PACEMAKER. Model Single Chamber Temporary Pacemaker TIP CARD MODEL 53401 TEMPORARY EXTERNAL PACEMAKER Model 53401 Single Chamber Temporary Pacemaker PRE-USE What to do... What NOT to do... Battery Replace the battery for each new patient. Do NOT reuse battery.

More information

Cardiostart. Biphasic Defibrillator Monitor

Cardiostart. Biphasic Defibrillator Monitor Cardiostart Biphasic Defibrillator Monitor Standard Accessories Carrying Case Rechargeable lithium ion battery Set of accessories Other accessories upon request Internal Pads Adult / Pediatric / Neonatal

More information

Operating Instructions. Panorama. Web Viewer

Operating Instructions. Panorama. Web Viewer Operating Instructions Panorama Web Viewer Mindray is a U.S. registered trademark of Mindray DS USA, Inc. Panorama is a U.S. registered trademark of Mindray DS USA, Inc. Microsoft is a U.S. registered

More information

IntelliVue Patient Monitor. MP20/30/40/50/60/70/90 For monitor release B.1 with software revision B.10.xx. IntelliVue MP20/30/40/50/60/70/90

IntelliVue Patient Monitor. MP20/30/40/50/60/70/90 For monitor release B.1 with software revision B.10.xx. IntelliVue MP20/30/40/50/60/70/90 IntelliVue MP20/30/40/50/60/70/90 CONFIGURATION GUIDE IntelliVue Patient Monitor MP20/30/40/50/60/70/90 For monitor release B.1 with software revision B.10.xx Patient Monitoring Philips Medizin Systeme

More information

Urgent Medical Device Correction LIFEPAK 1000 Defibrillator

Urgent Medical Device Correction LIFEPAK 1000 Defibrillator Urgent Medical Correction LIFEPAK 1000 Defibrillator Urgent Please bring this letter to the immediate attention of the person(s) responsible for maintaining/monitoring your LIFEPAK 1000 defibrillators.

More information

HTS820TC TRENSCenter Central Station Cardiac Monitor REHAB Operator/Technical Manual

HTS820TC TRENSCenter Central Station Cardiac Monitor REHAB Operator/Technical Manual Table of Contents Subject... Page Subject... Page Preface...2 Safety Summary... 3 Section 1 Operation Attaching Transmitter to Patient...4 Admitting a Group... 6 Admitting Individual Patients...7 Monitoring

More information

PCI Expansion Slot Kit for 6181P (1500P) Series D Integrated Display Computer

PCI Expansion Slot Kit for 6181P (1500P) Series D Integrated Display Computer Installation Instructions PCI Expansion Slot Kit for 6181P (1500P) Series D Integrated Display Computer Catalog Number 6189V-2PCI15R Topic Page About This Publication 1 Important User Information 2 Safety

More information

4P R. Wireless LAN Connecting Adapter <BRP069A81> Installation Manual

4P R. Wireless LAN Connecting Adapter <BRP069A81> Installation Manual 4P359542-4R Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com For details

More information

The LifeVest Network Patient Data Management System Quick Start Guide. Information for Healthcare Professionals

The LifeVest Network Patient Data Management System Quick Start Guide. Information for Healthcare Professionals The LifeVest Network Patient Data Management System Quick Start Guide Information for Healthcare Professionals Monitor and prioritize your patient s needs anywhere, anytime. The LifeVest Network is ZOLL

More information

12. Mobile Devices and the Internet of Things. Blase Ur, May 3 rd, 2017 CMSC / 33210

12. Mobile Devices and the Internet of Things. Blase Ur, May 3 rd, 2017 CMSC / 33210 12. Mobile Devices and the Internet of Things Blase Ur, May 3 rd, 2017 CMSC 23210 / 33210 1 Today s class Security and privacy for: mobile devices the IoT safety-critical devices Discuss midterm 2 Mobile

More information

IntelliVue Information Center

IntelliVue Information Center IntelliVue Information Center Demo Application User Guide Part Number 4535 642 63741 Printed in the U.S.A. June 2011 First Edition *453564263741* Notice Notice Proprietary Information Published by Philips

More information

4P S. Wireless LAN Connecting Adapter <BRP069A41> Installation Manual

4P S. Wireless LAN Connecting Adapter <BRP069A41> Installation Manual 4P359542-1S Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com For details

More information

4P Wireless LAN Connecting Adapter <BRP069B41> <BRP069B42> <BRP069B43> <BRP069B44> <BRP069B45> Installation Manual

4P Wireless LAN Connecting Adapter <BRP069B41> <BRP069B42> <BRP069B43> <BRP069B44> <BRP069B45> Installation Manual 4P481234-1 Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com

More information

Defibtech DDU-100 Semi-Automatic External Defibrillator

Defibtech DDU-100 Semi-Automatic External Defibrillator Defibtech DDU-100 Semi-Automatic External Defibrillator User Manual AHA /ERC 2010 U S E O N LY I F PA T I E N T I S : U N C O N S C I O U S U N R E S P O N S I V E N O T B R E A T H I N G 1 PUSH "ON" BUTTON

More information

Operating Instructions Vision Holter Analysis System Software Version 3.5

Operating Instructions Vision Holter Analysis System Software Version 3.5 Operating Instructions Vision Holter Analysis System Software Version 3.5 Part No. 70-01234-01 Rev. A Copyright 2012 Cardiac Science Corporation All rights reserved. Cardiac Science Corporation N7 W22025

More information

Mini Pulse Electronic Stimulator Model: PM-120

Mini Pulse Electronic Stimulator Model: PM-120 Mini Pulse Electronic Stimulator Model: PM-120 Operating Manual IMPORTANT: Please read all instructions before using this product. Retain this manual for future reference. www.santamedical.com IMPORTANT

More information

Save Face Radiofrequency Fact Sheet

Save Face Radiofrequency Fact Sheet Save Face Radiofrequency Fact Sheet Patient information Radiofrequency What is Radiofrequency? The field of body contouring and tissue tightening has grown very rapidly over the past several years, with

More information

PRODUCT SAFETY NOTICE WARNING OF POTENTIAL UNSAFE CONDITION

PRODUCT SAFETY NOTICE WARNING OF POTENTIAL UNSAFE CONDITION DATE: November 7, 2016 TO: FROM: Schneider Electric Distributors, and End-users using the Schneider Electric E-Stop 9001SKR16 David Collins US Product Management SUBJECT: Schneider E-Stop 30mm 9001SKR16

More information

IntelliVue Patient Monitors

IntelliVue Patient Monitors Hardware Upgrade Installation Guide IntelliVue Patient Monitors MX400/MX450 Patient Monitoring 1Table of Contents 1 Introduction 5 What is in this Installation Note 6 2 Checking your Tools and Kit Hardware

More information

Research Report. Radio Frequency Identification Devices (RFIDs) Effect of RFIDs on lifesaving equipment

Research Report. Radio Frequency Identification Devices (RFIDs) Effect of RFIDs on lifesaving equipment Research Report Radio Frequency Identification Devices (RFIDs) Effect of RFIDs on lifesaving equipment Background It has been suggested in the Journal of the Medical Association that lifesaving equipment

More information

MONITOR/DEFIBRILLATOR CONTROLS, CONNECTIONS AND INDICATORS

MONITOR/DEFIBRILLATOR CONTROLS, CONNECTIONS AND INDICATORS MONITOR/DEFIBRILLATOR CONTROLS, CONNECTIONS AND INDICATORS 1-10 Adult Dose CONTROLS, CONNECTIONS AND INDICATORS External Power SYNC Mark Event Lead Select Alarm Pause Event Summary Soft Keys Soft Keys

More information

R Series Configuration Manual

R Series Configuration Manual R Series Configuration Manual The issue date for the R Series Configuration Manual (REF 9650-0902-05 Rev. C) is June, 2012. If more than 3 years have elapsed since the issue date, contact ZOLL Medical

More information

Patient Manual. Confirm Rx Insertable Cardiac Monitor System

Patient Manual. Confirm Rx Insertable Cardiac Monitor System Confirm Rx Insertable Cardiac Monitor System Confirm Rx Insertable Cardiac Monitor Model DM3500 mymerlin Mobile Application Models APP1000, APP1001 Patient Manual CAUTION: Federal (USA) law restricts this

More information

4P M. Wireless LAN Connecting Adapter <BRP069A45> Installation Manual

4P M. Wireless LAN Connecting Adapter <BRP069A45> Installation Manual 4P359542-3M Wireless LAN Connecting Adapter Installation Manual Wireless LAN Connecting Adapter Installation Manual Homepage: http://www.onlinecontroller.daikineurope.com For details

More information

A Guide to Open-Access Databases and Open-Source Software on PhysioNet

A Guide to Open-Access Databases and Open-Source Software on PhysioNet A Guide to Open-Access Databases and Open-Source Software on PhysioNet George B. Moody Harvard-MIT Division of Health Sciences and Technology Cambridge, Massachusetts, USA Outline Background Open-Access

More information

High-resolution color display provides real-time preview of 12-lead ECG to ensure high quality data acquisiton at investigator site.

High-resolution color display provides real-time preview of 12-lead ECG to ensure high quality data acquisiton at investigator site. ELI 280 offers a choice of either the innovative WAM wireless acquisition module or the AM12 acquisition module. Both include replaceable lead wires, lead fail indicator, and remote control with buttons

More information

Policies, Procedures, Guidelines and Protocols. John Snell - Head of Workforce Planning, Systems and Contributors

Policies, Procedures, Guidelines and Protocols. John Snell - Head of Workforce Planning, Systems and Contributors Policies, Procedures, Guidelines and Protocols Document Details Title Staff Mobile Phone Policy Trust Ref No 2036-39774 Local Ref (optional) N/A Main points the document Procurement, allocation and use

More information

TransTalk 9000 Digital Wireless System MDW 9040 Wireless Pocket Phone Quick Reference

TransTalk 9000 Digital Wireless System MDW 9040 Wireless Pocket Phone Quick Reference TransTalk 9000 Digital Wireless System MDW 9040 Wireless Pocket Phone Quick Reference 503-801-191 Comcode 700060437 Issue 2, Feb. 2001 Mute Volume Control Display Softkeys/Column Buttons Feature/ Program

More information

Medical Device Usability

Medical Device Usability Medical Device Usability David Adams Global Head, Active Medical Devices Add logo on slide 4 here Topics What is usability? Why usability is so important The regulatory requirements EN 62366 Usability

More information

LATITUDE PROGRAMMING SYSTEM

LATITUDE PROGRAMMING SYSTEM OPERATOR S MANUAL LATITUDE PROGRAMMING SYSTEM REF 3300 Table of Contents INFORMATION FOR USE...1 Trademark Statement...1 Description and Use...1 Intended Use...1 Intended Audience...1 Required Expertise

More information

Wireless LAN Connecting Adapter [BRP072B42] Installation Manual

Wireless LAN Connecting Adapter [BRP072B42] Installation Manual Wireless LAN Connecting Adapter [BRP072B42] Installation Manual Wireless LAN Connecting Adapter [BRP072B42] Installation Manual Website: http://www.daikinthai.com/wlan3/modbus For details on specifications,

More information

M Series AHA/ERC Guidelines 2005 Software Configuration Overview

M Series AHA/ERC Guidelines 2005 Software Configuration Overview M Series AHA/ERC Guidelines 2005 Software Configuration Overview This page provides a summary of configuration options that have been added to support the American Heart Association and European Resuscitation

More information

MDLINK MANUAL PART NUMBER A Cardiac Science Corporation

MDLINK MANUAL PART NUMBER A Cardiac Science Corporation MDLINK MANUAL PART NUMBER 70-00080-01 A 2007 Cardiac Science Corporation Notice of Rights 2007 Cardiac Science Corporation. All rights reserved. No part of this documentation may be reproduced or transmitted

More information

Hardware and Components: Philips Monitors in Critical Care

Hardware and Components: Philips Monitors in Critical Care Hardware and Components: Philips Monitors in Critical Care Intellivue MX 800 (the monitor): This is the new bedside monitor. It is a touch screen. Touch will work even when double gloved. The monitor WILL

More information

VITAL TIMER RELAY ASSEMBLY, A

VITAL TIMER RELAY ASSEMBLY, A INSTALLATION & INSTRUCTION" VITAL TIMER RELAY ASSEMBLY, A44999-02 JULY 2008, REVISED JULY 2014 DOCUMENT NO. SIG-00-08-02 VERSION A.1 Siemens Industry, Inc., Rail Automation 9568 Archibald Ave., Suite 100,

More information

Embedded Systems. Cristian Rotariu

Embedded Systems. Cristian Rotariu Embedded Systems Cristian Rotariu Dept. of of Biomedical Sciences Grigore T Popa University of Medicine and Pharmacy of Iasi, Romania cristian.rotariu@bioinginerie.ro May 2016 Introduction An embedded

More information

BeneVision DMS Workstation Default Settings

BeneVision DMS Workstation Default Settings Hospital, City, ST: Unit: Clinical/Sales: Service/CSS: SW Version: 03.02 Date: Signature/Title: System Setup General Alm Vol (1-10) High: Alm Vol + 0 1 2 Reminder Vol (1-10) Display (Password protected)

More information

LXM32. Explanation for detected error E 733F. Expert Support Machine Solution

LXM32. Explanation for detected error E 733F. Expert Support Machine Solution LXM32 Explanation for detected error E 733F Expert Support Machine Solution The information provided in this documentation contains general descriptions and/or technical characteristics of the performance

More information

A1. Multilin EPM 4600 Meter Displays DIS3500/DIS5700. Quick Start Guide. GE Digital Energy

A1. Multilin EPM 4600 Meter Displays DIS3500/DIS5700. Quick Start Guide. GE Digital Energy GE Digital Energy Multilin EPM 4600 Meter Displays DIS3500/DIS5700 Quick Start Guide Product version: 1.0x GE publication code: 1601-0500-A1 (GEK-119644) 1601-0500-A1 Copyright 2014 GE Multilin Inc. All

More information

Smart Wireless THUM Adapter

Smart Wireless THUM Adapter Reference Manual Smart Wireless THUM Adapter www.rosemount.com Reference Manual Smart Wireless THUM Adapter Smart Wireless THUM Adapter Smart Wireless THUM Adapter Hardware Revision 1 HART Device Revision

More information

Expert support and Reach back activities

Expert support and Reach back activities GICNT- Nuclear Detection Working Group Magic Maggiore Ispra 28 March 2017 Expert support and Reach back activities Thierry PELLETIER Nuclear Security Division Safety and Security department International

More information

BD Alaris System User Manual Addendum

BD Alaris System User Manual Addendum BD Alaris System User Manual Addendum For BD Alaris TM PC Unit, Model 8015 and BD Alaris TM Pump Module, Model 8100 January 2018 DME: 10000318809-00 Part Number: P00000272 2018 CareFusion Corporation or

More information

ECG100S. General Information. Product Code

ECG100S. General Information. Product Code General Information Product Name General Name Product Code 80508096 Manufacturer Device Description Intended Use Cardioline Spa Registered Office and Factory: Via Linz, 19-20-21 Zona Ind. Spini di Gardolo

More information

Cleanroom Microbiology for Non-Microbiologists

Cleanroom Microbiology for Non-Microbiologists Microrite, Inc. brings you this unique learning experience in understanding cleanroom microbiology concepts and techniques Cleanroom Microbiology for Non-Microbiologists A course designed specifically

More information

F-1 OPTIONAL PRACTICAL TRAINING STEM EXTENSION

F-1 OPTIONAL PRACTICAL TRAINING STEM EXTENSION Global Education Office (GEO) 2120 Mesa Vista Hall MSC06 3850 1 University of New Mexico Albuquerque, NM 87131-1056 Phone (505) 277-4032 Fax (505) 277-1867 Email geo@unm.edu Web: www.geo.unm.edu/ F-1 OPTIONAL

More information

Additional Details About the AbioCor System

Additional Details About the AbioCor System Additional Details About the AbioCor System The AbioCor System is an implantable artificial heart intended to replace the patient s natural heart. It is designed for patients whose hearts are irreparably

More information

Banff Mineral Springs Hospital LOSS OF TELEPHONE EMERGENCY RESPONSE PLAN

Banff Mineral Springs Hospital LOSS OF TELEPHONE EMERGENCY RESPONSE PLAN Facilities Maintenance & Engineering Switchboard / Designate Area Supervisor / Designate Aware of Loss of Telephone services Banff Mineral Springs Hospital LOSS OF TELEPHONE EMERGENCY RESPONSE PLAN LOSS

More information

Computer Security Incident Response Plan. Date of Approval: 23-FEB-2014

Computer Security Incident Response Plan. Date of Approval: 23-FEB-2014 Computer Security Incident Response Plan Name of Approver: Mary Ann Blair Date of Approval: 23-FEB-2014 Date of Review: 31-MAY-2016 Effective Date: 23-FEB-2014 Name of Reviewer: John Lerchey Table of Contents

More information

Oracle Argus Safety. 1 Configuration. 1.1 Configuring a Reporting Destination for the emdr Profile. emdr Best Practices Document Release 8.0.

Oracle Argus Safety. 1 Configuration. 1.1 Configuring a Reporting Destination for the emdr Profile. emdr Best Practices Document Release 8.0. Oracle Argus Safety emdr Best Practices Document Release 8.0.1 E68590-01 November 2015 This document provides information on using emdr to switch from MedWatch Device reporting, and special considerations

More information

Application for Lead-based Paint Activities Training Program Accreditation

Application for Lead-based Paint Activities Training Program Accreditation ALABAMA LEAD-BASED PAINT ACTIVITIES ACCREDITATION PROGRAM Application for Lead-based Paint Activities Training Program Accreditation A completed application shall consist of Parts I - III of this application

More information

3T/7T MRI FACILITY. 3T MRI System Start-Up and Shutdown

3T/7T MRI FACILITY. 3T MRI System Start-Up and Shutdown 3T/7T MRI FACILITY SOP Number: 300.02 Title 3T MRI System Start-Up and Shutdown Revision Chronology Version Number Date Changes 300.01 01 March 2010 New 300.02 21 January 2013 Updated procedures Facility

More information

TraceMasterVue ECG Management System INSTALLING TRACEMASTERVUE AND CONFIGURING COMMUNICATION

TraceMasterVue ECG Management System INSTALLING TRACEMASTERVUE AND CONFIGURING COMMUNICATION TraceMasterVue ECG Management System INSTALLING TRACEMASTERVUE AND CONFIGURING COMMUNICATION Notice About This Edition Publication number 453564051211 Edition 1 Printed in USA Edition History Edition 1,

More information

Wiring Instructions for Replacement of 2301A Load Sharing and Speed Controls with 2301D Load Sharing and Speed Controls

Wiring Instructions for Replacement of 2301A Load Sharing and Speed Controls with 2301D Load Sharing and Speed Controls Application Note 51242 (Revision NEW) Original Instructions Wiring Instructions for Replacement of 2301A Load Sharing and Speed Controls with 2301D Load Sharing and Speed Controls General Precautions Read

More information

NHS Gloucestershire Clinical Commissioning Group. Business Continuity Strategy

NHS Gloucestershire Clinical Commissioning Group. Business Continuity Strategy NHS Gloucestershire Clinical Commissioning Group 1 Document Control Title of Document Gloucestershire CCG Author A Ewens (Emergency Planning and Business Continuity Officer) Review Date February 2017 Classification

More information

User s Manual. LumiVision series. A Designer s X-ray film illuminator. Walnut, CA U.S.A. By LCD TV Technology

User s Manual. LumiVision series. A Designer s X-ray film illuminator. Walnut, CA U.S.A. By LCD TV Technology User s Manual A Designer s X-ray film illuminator By LCD TV Technology series Point A Technologies, Inc. Walnut, CA U.S.A. FCC RF INTERFERENCE STATEMENT NOTE: This equipment has been tested and found to

More information

BHG Operational Awareness Program May 8, 1998 Surveillance Guide NSS-18.1 Revision 0 Criticality Safety Page 1 of 8 CRITICALITY SAFETY

BHG Operational Awareness Program May 8, 1998 Surveillance Guide NSS-18.1 Revision 0 Criticality Safety Page 1 of 8 CRITICALITY SAFETY Criticality Safety Page 1 of 8 1.0 Objective The objective of this surveillance is to ensure that effective programs have been developed and implemented to protect the public and DOE's workers from unplanned

More information

Input Power Selector Switch. Product Manual (Revision NEW) Original Instructions for use with Generator Set Control Assembly

Input Power Selector Switch. Product Manual (Revision NEW) Original Instructions for use with Generator Set Control Assembly Product Manual 82017 (Revision NEW) Original Instructions 9900-471 for use with 8272-713 Generator Set Control Assembly Operation Manual General Precautions Read this entire manual and all other publications

More information

Faros Explorer Manual

Faros Explorer Manual Faros Explorer Manual Date of issue: May 30, 2017 Mega Electronics Ltd, Pioneerinkatu 6, FI-70800 Kuopio, Finland, http://www.megaemg.com Contents 1. Introduction... 1 1.1. Intended use... 1 1.2. Symbols...

More information

Delta 1600 User Manual

Delta 1600 User Manual Delta 1600 User Manual Intended Use: Delta 1600 AED Tester is a precision instrument that simulates various cardiac waveforms and measures discharged energy. It assures AEDs detect the waveforms, audibly

More information

FOCUS - ON - PATIENTS. corpuls aed

FOCUS - ON - PATIENTS. corpuls aed FOCUS - ON - PATIENTS corpuls aed FOCUS - ON - PATIENTS corpuls aed Advanced early defibrillation for specialists and non-specialists The new corpuls aed safely guides experts and non-experts through life-saving

More information

Information Security Guide

Information Security Guide Broadband Kit Information Security Guide Broadband Kit ISO 27001 IS 541672 Broadband Kit Information Security Guide CONTENTS 1. Introduction... 3 1.1 Target audience...3 1.2 Overview of St. Jude Medical

More information

Analog Input Installation Manual

Analog Input Installation Manual Analog Input Installation Manual August 2011 Part Number: 144-23917 Copyright 2011 Magnetek 1. Preface and Safety Magnetek manufactures products used as components in a wide variety of industrial systems

More information

KBMG MULTI-SPEED BOARD

KBMG MULTI-SPEED BOARD TM INSTALLATION AND OPERATION MANUAL KBMG MULTI-SPEED BOARD KB Part No. 8833 Multi-Speed Board for KBMG Series Regenerative Drive Pending! See Safety Warning on Page 1 The information contained in this

More information

Application Note (Revision NEW) Original Instructions EGCP-3 LS Using an EGCP-3 in a Single-Phase Zig Zag Generator Application

Application Note (Revision NEW) Original Instructions EGCP-3 LS Using an EGCP-3 in a Single-Phase Zig Zag Generator Application Application Note 51247 (Revision NEW) Original Instructions EGCP-3 LS 8406-113 Using an EGCP-3 in a Single-Phase Zig Zag Generator Application General Precautions Read this entire manual and all other

More information

Quick Start Reference Guide Series 9.x and 10.x Software

Quick Start Reference Guide Series 9.x and 10.x Software Quick Start Reference Guide Series 9.x and 10.x Software NOTE: This quick reference should only be used by persons who have read the Software User Guide and received training on the use of the System.

More information

The Learner can: 1 Follow recommended safe practices

The Learner can: 1 Follow recommended safe practices Unit Title: Communicating Information Using ICT OCR unit number: 3 Level: Entry 3 Credit value: 3 Guided learning hours: 15 Unit reference number: H/502/1197 Learning Outcomes Assessment Criteria The Learner

More information

DATA BREACH POLICY [Enniskillen Presbyterian Church]

DATA BREACH POLICY [Enniskillen Presbyterian Church] DATA BREACH POLICY [Enniskillen Presbyterian Church] Enniskillen Presbyterian Church is committed to complying with data protection legislation and will take appropriate technical and organisational measures

More information

Model No. CF-19/CF-30 Series

Model No. CF-19/CF-30 Series Supplementary Instructions. for Wireless WAM Modem Personal Computer Model No. CF-19/CF-30 Series This manual contains supplementary instructions regarding the use of the computer. It explains the cases

More information

InView Firmware Update

InView Firmware Update Installation Instructions InView Firmware Update Topic Page Hazardous Voltage 3 Change EPROM on 2706-P72, 2706-P74 Display 3 Change EPROM on 2706-P42, 2706-P44 Displays 5 Firmware Upgrade Kit 7 2 InView

More information

DEFRIBILLATORS PARAMEDIC CU-ER1

DEFRIBILLATORS PARAMEDIC CU-ER1 DEFRIBILLATORS PARAMEDIC CU-ER1 (Automated External Defibrillator) Someone goes down in cardiac arrest, resuscitation is just a hand away Patented e-cube Biphasic measure impedance and deliver efficient

More information

Banner AG4 Series Safety Laser Scanner Checkout Procedures

Banner AG4 Series Safety Laser Scanner Checkout Procedures Shift/Daily Checkout Perform the Daily Checkout procedure at every shift change, power-up and machine set-up change and at intervals not to exceed 24 hours during continuous machine run periods. Record

More information

Reliable and Real-time Wireless Sensor Networks: Protocols and Medical Applications

Reliable and Real-time Wireless Sensor Networks: Protocols and Medical Applications Reliable and Real-time Wireless Sensor Networks: Protocols and Medical Applications Octav Chipara University of Iowa https://cs.uiowa.edu/~ochipara 1 Detecting clinical deterioration at low cost Clinical

More information

Multilin EPM 4600 Metering Solution

Multilin EPM 4600 Metering Solution GE Digital Energy Multilin EPM 4600 Metering Solution Multi-feed Power and Energy Metering Solution Quick Start Guide GE publication code: 1601-0297-A2 (GEK-119590A) 1601-0297-A2 Copyright 2014 GE Multilin

More information

Argentine Manufacturer of Hospital Equipment. Electrocardiograph Pulse Oximeters Vital Signs Monitors

Argentine Manufacturer of Hospital Equipment. Electrocardiograph Pulse Oximeters Vital Signs Monitors Electrocardiograph Pulse Oximeters Vital Signs Monitors ELECTROCARDIOGRAPH Model: RG-401plus Model: RG-401 ELECTROCARDIOGRAPH Model: RG-401 Technical Specifications Accessories Carrying bag Patient cable

More information

Lakeshore Technical College Official Policy

Lakeshore Technical College Official Policy Policy Title Original Adoption Date Policy Number Information Security 05/12/2015 IT-720 Responsible College Division/Department Responsible College Manager Title Information Technology Services Director

More information

Proximity Warning. Site Zone PWS. System. Personal protection & identification to help safeguard personnel from vehicle & machinery hazards

Proximity Warning. Site Zone PWS. System. Personal protection & identification to help safeguard personnel from vehicle & machinery hazards Site Zone PWS System Personal protection & identification to help safeguard personnel from vehicle & machinery hazards Creating a safer workplace Site Zone PWS Site Zone PWS key facts Reduces the risk

More information

Important Information KX-NTV150. Communication IP Camera. IP Video Doorphone. Model No. Model No. KX-NTV160

Important Information KX-NTV150. Communication IP Camera. IP Video Doorphone. Model No. Model No. KX-NTV160 Important Information Communication IP Camera Model No. KX-NTV150 IP Video Doorphone Model No. KX-NTV160 Thank you for purchasing this Panasonic product. Please read this Important Information carefully

More information