SRS501 User s manual

Size: px
Start display at page:

Download "SRS501 User s manual"

Transcription

1 SRS501 User s manual 1. Function Function of the device is measurement of angular rate projection. 1.1 Specification: in steady state power consumption: < 6 watt; time of functional ready no more 3 seconds; range of linear measured angular rate: ±250 deg/sec; range of measured angular rate with keep sign rate : ±500 deg/sec; the device mass: 0.4 kg. 1.2 Figure 1 shows the exterior of the device. Figure 1 2. Connection to external circuit 2.1 Connection of the device to external circuit implements over pins (Figure 2). Figure 2

2 2.2 Purposes of the device s pins. Purpose Pin Designation Description Input power supply 1, 2 +5V Power supply 5V ±5%. 3, 4 GND +5V Max current 3А. Serial interface 5 Line A Line A RS485 interface EIA Line B Line B RS485 interface EIA-485 Synchronization 7 TR+ Input for synchronization measurement (input RS485 8 TR- dif. signal EIA-485). Other 9 Case This pin connects to case of device Table 1 Attention! The device doesn t have protection from wrong connection. Connection error or over-voltage delivery can be the cause of device failure or its quick degradation. Time between device switching off and again switching on (and other) must be not less 1 minute. 2.3 Device power on and power off. In order to switch the device on need to supply power (5±0,25V) to the pins 1, 2 +5V and to the pins 3, 4 GND(+5V) of the device. In order to switch the device off need to remove power from the pins 1, 2 +5V and from the pins 3, 4 GND(+5V) of the device. It is recommended to use isolated from outside circuit power supply.

3 2.4 Figure 3 shows the scheme of device(s) connection. RS485 A+ B- 5 A+ 6 B- PC/CPU GND 3 GND 4 GND +5V 1 +5B 2 +5B SRS A+ 6 B- GND 3 GND 4 GND +5V 1 +5B 2 +5B SRS-501 Figure Figure 4 shows the scheme of device(s) connection with external synchronization signal input. RS485 A+ B- 5 A+ 6 B- T+ T- 7 TR+ 8 TR- T+ T- T+ T- GND 3 GND 4 GND PC/CPU +5V 1 +5B 2 +5B SRS A+ 6 B- 7 TR+ 8 TR- GND 3 GND 4 GND 1 +5B +5V 2 +5B Figure 4 SRS-501

4 3 Description of interface For connection to external equipment in the device use RS-485 interface. There are three modes of output data to external equipment in the device. Mode I: Output data through RS-485 interface with using exchange data protocol SSP 2.0. Mode II: Output data through RS-485 interface as response to a sync pulse. Mode III: Output data through RS-485 interface by a signal of internal timer with customizable frequency. Switching modes is performed automatically. If on entry synchronization signals are absent (sync pulse 1 -> 0 or DC 0 ) then protocol SSP 2.0 is working. When a sync pulse comes or on entry synchronization DC level 0 is hold, protocol SSP disable and device begin to transmit data in the form frame in compliance with table 2 or table Synchronization of measurement. In the mode II the measurement synchronization is performed by rising edge of sync pulse. Processing data is performed, data frame is generated and after constant time (Table 4) data frame is transmitted through interface RS-485. In the mode III clocking of measurement is performed by internal timer. Time from measurement to transmitting first byte of frame is constant (Table 4). Period of measurement can be changed by writing of coefficient by means of command PUT in mode I (address 34). In modes II and III baud rate of interface RS-485 can be changed from 9.6 Kbaud to Kbaud by writing of coefficient by means of command PUT in mode I (address 32). 3.2 Mode I description (protocol SSP 2.0) The device is attached to PC from asynchronous serial interface RS-485. Rate of exchange is baud only; 8 bit; 2 stop-bits without parity. Exchange data with device performs by protocol SSP 2.0 (Simple Serial Protocol). Maximum time of reaction on command is 1000 μsec (between receive and transmit). Maximum frequency of sampling is 200 Hz. The main purpose of this mode is configuring device s factory s and user s parameters. There is the simple description of the protocol SSP for a device TRS-500 in appendix Description of device operation in modes II and III The device TRS500 output data through serial interface RS-485 as response to an external sync pulse or by a signal of internal timer. Baud rate is set up in mode I (protocol SSP, table 7). Format: 8 bits, 2 stop-bits, no parity. In the mode II the measurement is performed by rising edge of sync pulse. Transmitting data begins after removal sync pulse.

5 There are formats of output packet (device work in mode II or III) in the table 2 and table 3. Table 2 shows the structure of base packet Table 2 Byte Name Description Size [bit] 0 Header Header: value equal 0xC0C0 (192d) 16 2 Angle rate Angle rate, deg/s. 32-bit value in format floating-point (float). In moment of coming 32 sync pulse. 6 CRC bit CRC-CCITT. Calculate by bytes 2 5 only 16 Table 3 shows the structure of extended packet Table 3 Byte Name Description Size [bit] 0 Header Header: value equal 0xC0C0 (192d) 16 2 Angle rate Angle rate, deg/s. 32-bit value in format floating-point (float). In moment of coming sync pulse. 32 [ 6 ] [ 6+2 T ] [ 6+2 T +2 FC ] Temperature case of the device Packet counter * Time-counter from start-up of the device LSB = 0.1ºC. 16-bit value in format signed short 16-bit unsigned integer value 16 LSB 8,68055 μsec (1/115200). 32-bit value in format unsigned integer 6+2 T +2 FC +4TM CRC bit CRC-CCITT. 16 T, FC, TM - bits of register for configuration of extended packet, see table 8. [] this field adds to base packet if only corresponded bit (T or FC or TM) set to one. 3.4 Time parameters of interface Figure 5 shows diagram of work interface

6 Figure 5 diagram of work interface. There are time parameters of interface in the table 4. Table 4 Designation Description Value min max t1 Time from transmission last byte request to start transmission first byte of packet, μsec 1000 t2 Time from falling edge of sync pulse to start transmission first byte of packet, μsec * t3 Time from giving fixed level on input synchronizing circuit to start transmission first byte of packet, msec 500 t4 Time from start of measurement to start transmission first byte of packet, μsec * t5 Time from start of measurement to start transmission first byte of packet, msec 3 t6 Time from falling edge of sync pulse to start of measurement, μsec 2 Tsync Period of external sync pulse See table 5 Tclock Period of signals internal timer See table 5 tp Sync pulse duration, μsec Тreq Request period (GET) when simultaneous request three parameters, μsec 5 * - For baud rate Kbaud. When baud rate decrease this time increase (for 9.6 Kbaud 375 μsec). Request frequency in mode I is no more 200 Hz. There is dependence request frequency from baud rate of interface RS485 in the table 5.

7 Table 5 Baud rate RS-485 Sync pulse-repetition frequency, Sync pulse period, μsec by sync pulse, Hz Kbaud min max min max 9,6 25, ,4 20, ,2 12, ,4 6, ,6 5, ,2 2, ,4 1, ,8 0, ,6 0,

8 Appendix 1 Description of the protocol SSP For division of flow of data, the protocol SSP uses framing RFC1055 (SLIP). Every frame begins and ends with symbol FEND (0xс0). If packet of data included symbol FEND then this symbol is change to FESC TFEND (0xdb 0xdc). If packet of data included symbol FESC (0xdb) then this symbol is change to FESC TFESC (0xdb 0xdd). Any symbol following behind FESC, except TFEND or TFESC, is wrong. The symbols TFEND and TFESC is just data, if before them is absent symbol FECS. The protocol SSP doesn t support arbitration of bus between some master devices. Every frame include packet SSP. General format of packet SSP is: dest, srce, type,...data..., crc0, crc1 DEST SRCE TYPE DAT0.. DATN CRC0 CRC1 dest one byte is address of slave device, which transmit data. Value 0 for dest is reserve for broadcast. Value FEND or FESC for dest is wrong. Default address is 100(0x64). srce one byte is address of master device, which request data. Value 0 for dest is reserve for use in future. Value FEND or FESC for srce is wrong. type one byte include two fields: ss.pktype Low six bits (pktype) are type packet. High two bits are additional data, which can mean special data (if not use, ss=0 ). Some values of pktype specifies as standard values. crc0 и crc1 two bytes is CRC-16. CRC is received by low byte forward, and it includes all previous bytes starting with dest until last byte of data. Ignore any packet with wrong CRC. Example calculates CRC on C language follows below. Initial value shift register crc will must be equal 0xFFFF(hex). #define mask 0x1021 unsigned short updcrc(unsigned short crc, unsigned short c) { unsigned char i; c <<= 8; for (i = 0; i < 8; i++) { if ((crc ^ c) & 0x8000) crc = (crc << 1) ^ mask;

9 else crc <<= 1; c <<= 1; } return crc; } Description command of SSP protocol PING Command PING. Format > 0.0 (ss.pktype), 0x00 (Hex) DEST SRCE 0x00 CRC0 CRC1 Response of device is ACK. Format > 1.2 (ss.pktype) 0x42 (Hex) DEST SRCE 0x42 CRC0 CRC1 Description: test packet without data, for check device communication. Example: Master device with address 0x02 sends command PING by slave device with address 0х64. Request -> DEST SRCE PING CRC0 CRC SSP Dec 0x64 0x02 0x00 0x55 0xED SSP Hex 0xC0 0x64 0x02 0x00 0x55 0xED 0xC0 (SSP+RFC) Hex Response -> DEST SRCE ACK CRC0 CRC SSP Dec 0x02 0x64 0x42 0x94 0x0D SSP Hex 0xC0 0x02 0x64 0x42 0x94 0x0D 0xC0 (SSP+RFC) Hex INIT Command INIT. Format > 0.1 (ss.pktype), 0x01 (Hex) DEST SRCE 0x01 CRC0 CRC1 Response of device is ACK. Format > 0.2 (ss.pktype), 0x02 (Hex) DEST SRCE 0x02 CRC0 CRC1

10 Description: Optional command, use for compatibility with SSP2.0. Example: Master device (Master) with address 0х02 sends command INIT by slave device (Slave) with address 0х64. Request -> DEST SRCE INIT CRC0 CRC SSP Dec 0x64 0x02 0x01 0x74 0xFD SSP Hex 0xC0 0x64 0x02 0x01 0x74 0xFD 0xC0 (SSP+RFC) Hex Response -> DEST SRCE ACK CRC0 CRC SSP Dec 0x02 0x64 0x02 0x50 0x45 SSP Hex 0xC0 0x02 0x64 0x02 0x50 0x45 0xC0 (SSP+RFC) Hex ID Command ID. Format >0. 8 (ss.pktype), 0x08 (Hex) DEST SRCE 0x08 CRC0 CRC1 Response of device is ACK + ASCII string. Format > 0.2 (ss.pktype), 0x02 (Hex) DEST SRCE 0x02 ASCII ASCII CRC0 CRC1 Description: Identification device Example: Master device (Master) with address 0х02 sends command ID by slave device (Slave) with address 0х64. Slave device return message ACK+ PNSK Request -> DEST SRCE ID CRC0 CRC SSP Dec 0x64 0x02 0x08 0x5D 0x6C SSP Hex 0xC0 0x64 0x02 0x08 0x5D 0x6C 0xC0 (SSP+RFC) Hex Response -> DEST SRCE ACK ASCII ASCII ASCII ASCII ASCII ASCII SSP Dec 0x02 0x64 0x02 0x50 0x4E 0x53 0x4B 0x34 0x30 SSP Hex 0xC0 0x02 0x64 0x02 0x50 0x4E 0x53 0x4B 0x34 0x30 (SSP+RFC) ASCII ASCII ASCII ASCII CRC0 CRC SSP Dec 0x2D 0x30 0x31 0x38 0xEC 0x8C SSP Hex 0x2D 0x30 0x31 0x38 0xEC 0x8C 0xC0 (SSP+RFC)

11 WRITE The command WRITE. Format > 0.7 (ss.pktype), 0x07 (Hex) DEST SRCE 0x07 ADR DAT0 DATN CRC0 CRC1 4 byte 4 byte 4 byte The field ADR address array 32 bits long, it is transferred lower significant bit forward: BYT0 BYT1 BYT2 BYT3 The field DAT one field array data 32 bits long, it is transferred lower significant bit forward: BYT0 BYT1 BYT2 BYT3 Response of device is ACK. Format > 1.2 (ss.pktype), 0x42 (Hex) DEST SRCE ACK CRC0 CRC SSP Dec 0x02 0x64 0x42 0x94 0x0D SSP Hex 0xC0 0x02 0x64 0x42 0x94 0x0D 0xC0 (SSP+RFC) Hex Description: Writing of data array in the device. In product is used only for the definition of the new address. Example: The master device with address 0х02 send command WRITE the slave device for change of the address on 0х63. Request -> DEST SRCE WRITE ADR0 ADR1 ADR2 ADR SSP Dec 0x00 0x02 0x07 0x00 0x00 0x00 0x00 SSP Hex 0xC0 0x00 0x02 0x07 0x00 0x00 0x00 0x00 (SSP+RFC) Hex DAT0 DAT1 DAT2 DAT3 CRC0 CRC SSP Dec 0x63 0x00 0x00 0x00 0x20 0x79 SSP Hex 0x63 0x00 0x00 0x00 0x20 0x79 0xC0 (SSP+RFC) Hex Response -> DEST SRCE ACK CRC0 CRC SSP Dec 0x02 0x63 0x42 0x03 0x94 SSP Hex 0xC0 0x02 0x63 0x42 0x50 0x45 0xC0 (SSP+RFC) Hex The command with the memory address different from zero is considered an error.

12 GET The command GET. Format > 0.4 (ss.pktype), 0x04 (Hex) DEST SRCE 0x04 ADR1 ADRN CRC0 CRC1 2 byte 2 byte The field ADR address array 16 bits long, it is transferred lower significant bit forward: BYT0 BYT1 The field DAT variable by address ADR 32 bits long, it is transferred lower significant bit forward: BYT0 BYT1 BYT2 BYT3 Response of device is ACK. Format > 0.2 (ss.pktype), 0x02 (Hex) DEST SRCE 0x02 DAT1 DATN CRC0 CRC1 4 byte 4 byte 4 byte Description: Data read from the device The list of accessible addresses read on the command GET Address (dec) 0 Table 6 Description Unit Format Note Rate obtained on the last period of measurement. deg/sec float 3 Temperature case of the device LSB = 0.1ºC signed long 12 Setup a bandwidth Kf Time-counter from start-up of the device Code for setup baud rate of interface RS-485 in modes II and III Register for configuration of extended packet Devisor of internal timer frequency LSB 8,68055 μsec (1/115200) See table 7 See table 8 Div unsigned long unsigned long unsigned long unsigned long unsigned long Supplemental information Kf * coefficient of recursive filter, 1 < Kf < 1000 F average 1,5Kf Watch overflow Default set to zero f = /Div * zero value switch recursive filter to averaging in period between external or internal sync pulse ( Tsync or Tclock in table 4). This feature can be used in modes II and III. There is correspondence of code by baud rate Rs-485 in the table 7. Table 7 Code Baud rate RS-485 in modes II and III, KBaud Note 95 9, ,2

13 23 38, , ,2 Default value 3 230, , ,6 Table 8 shows structure register which used for configuration of extended packet. Table 8 Bit number Description Note 0 T 0- temperature data do not add to packet 1- temperature data add to packet 1 FC 0- count-packet data do not add to packet 2 TM 1- count-packet data add to packet 0- time-count data do not add to packet 1- time-count data add to packet 3 31 Not used On one command GET reading of value some variables is admitted. Call to not existing addresses is considered an error. Example: The master device with address 0х02 read information to addresses 24 and 0 from the slave device with address 0х64. The slave device return ACK+data( )= adr24+data( )=adr0. Request -> DEST SRCE GET ADR0.24 ADR1.24 ADR0.0 ADR1.0 CRC0 CRC SSP Dec 0x64 0x02 0x04 0x18 0x00 0x00 0x00 0x30 0x1F SSP Hex 0xC0 0x64 0x02 0x04 0x18 0x00 0x00 0x00 0x30 0x1F 0xC0 (SSP+RFC) Response -> DEST SRCE ACK DAT0.3 DAT1.3 DAT2.3 DAT SSP Dec 0x02 0x64 0x02 0x7D 0x9C 0x16 0x44 SSP Hex 0xC0 0x02 0x64 0x02 0x7D 0x9C 0x16 0x44 (SSP+RFC) DAT0.24 DAT1.24 DAT2.24 DAT3.24 CRC0 CRC SSP Dec 0x7D 0x03 0x4D 0x3B 0x78 0xAD SSP Hex 0x7D 0x03 0x4D 0x3B 0x78 0xAD 0xC0 (SSP+RFC) PUT The command PUT. Format > 0.5 (ss.pktype), 0x05 (Hex) DEST SRCE 0x05 ADR1 DAT CRC0 CRC1 2 byte 4 byte The field ADR address of variable 16 bits long, it is transferred lower significant bit forward:

14 BYT0 BYT1 The field DAT one field array of data 32-bit long, it is transferred lower significant bit forward: BYT0 BYT1 BYT2 BYT3 Response of device is ACK. Format > 0.2 (ss.pktype), 0x02 (Hex) DEST SRCE 0x02 CRC0 CRC1 Description: Write setup variables to device. There is the list of accessible for write addresses by command PUT in the table 9 Table 9 Address Description Note (dec) 12 Setup bandwidth 32 Code for setup baud rate of interface RS-485 in modes II and III 33 Register for configuration of extended packet 34 Devisor of internal timer frequency See table 6 Setup variables are saved in nonvolatile memory of the device. All this variables are unsigned long and transferred lower significant bit forward. Access to nonexistent address is error. Example: The master device with address 0х02 write code for setup baud rate RS Kbaud to address 32 of the slave device with address 0х64. Request -> DEST SRCE GET ADR0 ADR1 DAT0 DAT1 DAT2 DAT3 CRC0 CRC SSP Dec 0xC0 0x64 0x02 0x05 0x21 0x00 0x07 0x00 0x00 0x00 0x9c 0xee 0xC0 (SSP+RF) Response -> DEST SRCE ACK CRC0 CRC SSP Dec 0xC0 0x02 0x64 0x02 0x50 0x45 0xC0 (SSP+RFC)Hex Device ignore any misaddress and damage packet. Damage packet packet length lower 5 byte and/or CRC error.

CTT MODBUS-RTU COMMUNICATION PROTOCOL TEMPERATURE MONITOR DEVICE

CTT MODBUS-RTU COMMUNICATION PROTOCOL TEMPERATURE MONITOR DEVICE INSTRUCTION MANUAL IM149-U v0.92 CTT MODBUS-RTU COMMUNICATION PROTOCOL TEMPERATURE MONITOR DEVICE Firmware version: v3.0 or higher MODBUS PROTOCOL Modbus is a master-slave communication protocol able to

More information

DMTME Multimeters. Communication protocol. Technical specification V1.2 ABB

DMTME Multimeters. Communication protocol. Technical specification V1.2 ABB DMTME Multimeters Communication protocol ABB 1 Setting up DMTME serial communication.3 1.1 Serial network ID programming...3 1.2 RS-485 communication interface...3 1.3 Serial line connection...3 2 Communication

More information

Conto D2 COMMUNICATION PROTOCOL CONTENTS 1.0 INTRODUCTION

Conto D2 COMMUNICATION PROTOCOL CONTENTS 1.0 INTRODUCTION PR 121 rev. 0 11/11/2011 Pagina 1 di 9 ELECTRICITY ENERGY METER FIRMWARE 1.6 Conto D2 COMMUNICATION PROTOCOL CONTENTS 1.0 INTRODUCTION 2.0 DATA MESSAGE DESCRIPTION 2.1 Data field description 2.2 Data format

More information

SECTION 5 SMART PAYOUT MANUAL SET SOFTWARE IMPLEMENTATION GUIDE

SECTION 5 SMART PAYOUT MANUAL SET SOFTWARE IMPLEMENTATION GUIDE SECTION 5 SMART PAYOUT MANUAL SET SOFTWARE IMPLEMENTATION GUIDE Innovative Technology assume no responsibility for errors, omissions, or damages resulting from the use of information contained within this

More information

MTR-4. C8 Command to MODBUS Bridge User Manual F-1, No. 631, Chung Der Road, Sec 1, Taichung Taiwan.

MTR-4. C8 Command to MODBUS Bridge User Manual F-1, No. 631, Chung Der Road, Sec 1, Taichung Taiwan. MTR-4 C8 Command to MODBUS Bridge User Manual 404 19F-1, No. 631, Chung Der Road, Sec 1, Taichung Taiwan. TEL:886-4-2238-0698 FAX:886-4-2238-0891 Web Site:http://www.monicon.com.tw E-mail: sales@monicon.com.tw

More information

CE4DMID01 COMMUNICATION PROTOCOL CONTENTS 1.0 INTRODUCTION

CE4DMID01 COMMUNICATION PROTOCOL CONTENTS 1.0 INTRODUCTION 11/11/2011 Pagina 1 di 11 ELECTRICITY ENERGY METER FIRMWARE 1.3 CE4DMID01 COMMUNICATION PROTOCOL CONTENTS 1.0 INTRODUCTION 2.0 DATA MESSAGE DESCRIPTION 2.1 Data field description 2.2 Data format 2.3 Description

More information

M2M/DMTME Instruments Communication protocol. Technical specification V.2.1 2CSG445011D0201

M2M/DMTME Instruments Communication protocol. Technical specification V.2.1 2CSG445011D0201 M2M/DMTME Instruments Communication protocol 2CSG445011D0201 1 Setting up M2M/DMTME serial communication... 3 1.1 Serial network ID programming... 3 1.2 RS-485 communication interface... 3 1.3 Serial line

More information

COMMUNICATION MODBUS PROTOCOL

COMMUNICATION MODBUS PROTOCOL COMMUNICATION MODBUS PROTOCOL CE4DMID31 / CE4DMID21 CONTO D4 Pd MID PR123 20/10/2016 Pag. 1/9 Contents 1.0 ABSTRACT... 2 2.0 DATA MESSAGE DESCRIPTION... 3 2.1 Parameters description... 3 2.2 Data format...

More information

COMMUNICATION MODBUS PROTOCOL

COMMUNICATION MODBUS PROTOCOL COMMUNICATION MODBUS PROTOCOL CE4DT36 CONTO D4 Pd (3-single phase) PR134 20/10/2016 Pag. 1/11 Contents 1.0 ABSTRACT... 2 2.0 DATA MESSAGE DESCRIPTION... 3 2.1 Parameters description... 3 2.2 Data format...

More information

COMMUNICATION MODBUS PROTOCOL MF96001 / 021 NEMO 96HD

COMMUNICATION MODBUS PROTOCOL MF96001 / 021 NEMO 96HD COMMUNICATION MODBUS PROTOCOL MF96001 / 021 NEMO 96HD PR106 20/10/2016 Pag. 1/31 Contents 1.0 ABSTRACT... 2 2.0 DATA MESSAGE DESCRIPTION... 3 2.1 Parameters description... 3 2.2 Data format... 4 2.3 Description

More information

DULCOMETER Multi-parameter Controller dialog DACa

DULCOMETER Multi-parameter Controller dialog DACa Software manual DULCOMETER Multi-parameter Controller dialog DACa Modbus RTU EN Valid only in combination with the operating instructions for the Multi-parameter Controller dialog DACa. A2100 Please carefully

More information

MCW Application Notes 24 th February 2017

MCW Application Notes 24 th February 2017 MCW Application Notes 24 th February 2017 www.motorcontrolwarehouse.co.uk Document number MCW-HEDY-001 Revision 0.1 Author Gareth Lloyd Product HEDY HD700 Title Summary HEDY HD700 Modbus Serial Communications

More information

IF96017 MODBUS COMMUNICATION PROTOCOL

IF96017 MODBUS COMMUNICATION PROTOCOL CONTENTS 1.0 ABSTRACT 04/07/14 Pagina 1 di 9 MULTIFUNCTION FIRMWARE 1.00 COMMUNICATION PROTOCOL IF96017 MODBUS COMMUNICATION PROTOCOL 2.0 DATA MESSAGE DESCRIPTION 2.1 Parameters description 2.2 Data format

More information

COMMUNICATION MODBUS PROTOCOL

COMMUNICATION MODBUS PROTOCOL COMMUNICATION MODBUS PROTOCOL BOZZA_V04 Conto D6-Pd 05/12/2017 Pag. 1/15 CONTENTS 1.0 ABSTRACT... 2 2.0 DATA MESSAGE DESCRIPTION... 3 2.1 Parameters description... 3 2.2 Data format... 4 2.3 Description

More information

Specification E2 Interface

Specification E2 Interface Specification E2 Interface Version 4.1 Name Date Created: Robert Mayr. 15.04.2011 Checked: Haider A. 15.04.2011 Approved: Reason for change: Text corrections TABLE OF CONTENTS 1 INTRODUCTION... 3 1.1 Overview..................................................................................................................

More information

TBS Unify Pro / SmartAudio

TBS Unify Pro / SmartAudio TBS Unify Pro / SmartAudio 02.12..2017 Revision 08 SmartAudio uses single wire half duplex communication based on UART. In Idle mode both host and VTX define the SmartAudio pin as input. If the host want

More information

SCYLAR INT 7. Communication description. Software Version 01

SCYLAR INT 7. Communication description. Software Version 01 SCYLAR INT 7 Communication description Software Version 01 Changes to be reserved 1 Contents 1 Introduction...3 2 Communication interfaces...3 2.1 Communication priorities...3 2.2 Telegram formats...3

More information

COMMUNICATION MODBUS PROTOCOL

COMMUNICATION MODBUS PROTOCOL COMMUNICATION MODBUS PROTOCOL MFD4E06 - NEMO-D4e PR146 08/03/2018 Pag. 1/16 CONTENTS 1.0 ABSTRACT... 2 2.0 DATA MESSAGE DESCRIPTION... 3 2.1 Parameter description... 3 2.2 Data format... 4 2.3 Description

More information

Conto D1 MODBUS COMMUNICATION PROTOCOL

Conto D1 MODBUS COMMUNICATION PROTOCOL ENERGY METER Conto D1 MODBUS COMMUNICATION PROTOCOL 4/03/15 Pagina 1 di 7 FIRMWARE CONTENTS 1.0 ABSTRACT 2.0 DATA MESSAGE DESCRIPTION 2.1 Parameters description 2.2 Data format 2.3 Description of CRC calculation

More information

Peripheral Sensor Interface for Automotive Applications

Peripheral Sensor Interface for Automotive Applications I Peripheral Sensor Interface for Automotive Applications Substandard Powertrain Technical _spec_v2d2_powertrain.doc 08/2016 II Contents 1 1 Introduction 1 2 2 System Setup & Operation Modes 1 3 4 5 6

More information

SHARKY BR773. Communication Description. Software Version 28

SHARKY BR773. Communication Description. Software Version 28 SHARKY BR773 Communication Description Software Version 28 06.03.2007 1 Autor: Edgar v. Schloß CONTENTS 1 INTRODUCTION... 3 2 COMMUNICATION INTERFACES... 3 2.1 COMMUNICATION PRIORITIES... 3 2.2 TELEGRAM

More information

Interface Card, Interface Box, Datalogger & Interface. Bedienungsanleitung. Kommunikationsprotokoll. Operating Instructions. Communications protocol

Interface Card, Interface Box, Datalogger & Interface. Bedienungsanleitung. Kommunikationsprotokoll. Operating Instructions. Communications protocol Interface Card, Interface Box, Datalogger & Interface D GB Bedienungsanleitung Kommunikationsprotokoll Operating Instructions Communications protocol 42,0410,1318 022005 Table of contents General remarks...

More information

Revision 1.2. July 24, COM Protocol Manual. for MDC and ADC N 11th St - San Jose CA

Revision 1.2. July 24, COM Protocol Manual. for MDC and ADC N 11th St - San Jose CA Revision 1.2 July 24, 2017 COM Protocol Manual for MDC and ADC www.mountztorque.com - 1080 N 11th St - San Jose CA 95112-408.292.2214 1 1 Overview and Communication Specifications 1.1 Overview If the PC

More information

GNetPlus Communication Protocol

GNetPlus Communication Protocol Basic Package (BINARY VERSION) Master Query Package (HOST) Field Header Address Query Function Data length DATA BYTES Error Check Desc SOH 0~255 0~255 0~255 CRC16_Low CRC16_Hi Size 1 BYTE 1 BYTE 1 BYTE

More information

Real Time Clock with Temperature Sensor and RS485/Modbus Comunications

Real Time Clock with Temperature Sensor and RS485/Modbus Comunications Real Time Clock with Temperature Sensor and RS485/Modbus Comunications April 29, 2014 Power 8 20 VDC @ less than 100 MA. Battery connect jumper RS485 Bus Load Jumpers Model: RTC-TI2C Page 1 of 6 Features:

More information

Technical Specification. Third Party Control Protocol. AV Revolution

Technical Specification. Third Party Control Protocol. AV Revolution Technical Specification Third Party Control Protocol AV Revolution Document AM-TS-120308 Version 1.0 Page 1 of 31 DOCUMENT DETAILS Document Title: Technical Specification, Third Party Control Protocol,

More information

B Interface description 12.01/

B Interface description 12.01/ B 95.3530.2 Interface description 12.01/00340396 Contents 1 Introduction 1.1 Preface... 3 1.2 Typographical conventions... 4 1.2.1 Warning signs... 4 1.2.2 Note signs... 4 1.2.3 Presentation... 4 2 Protocol

More information

RS485 MODBUS Module 8I8O

RS485 MODBUS Module 8I8O Expansion Module 8 digital inputs, 8 digital outputs Version 2.2 12/01/2014 Manufactured for Thank you for choosing our product. This manual will help you with proper support and proper operation of the

More information

Input Channels: 4 differential or four single-ended. Address / Baud rate / range configurable by the user

Input Channels: 4 differential or four single-ended. Address / Baud rate / range configurable by the user DAM-Series User Manual Amazing Electronic (HK) Limited User s Manual Overview: DAM module is a new generation data acquisition and control system based on modular embedded systems. Adopting standard DIN35

More information

Motors I Automation I Energy I Transmission & Distribution I Coatings. Modbus RTU CFW701. User s Manual

Motors I Automation I Energy I Transmission & Distribution I Coatings. Modbus RTU CFW701. User s Manual Motors I Automation I Energy I Transmission & Distribution I Coatings Modbus RTU CFW701 User s Manual Modbus RTU User s Manual Series: CFW701 Language: English Document Number: 10001538593 / 00 Publication

More information

HDLC-ETH. Serial Ethernet Converter. Rev. Dec 20, Datasheet. Website:

HDLC-ETH. Serial Ethernet Converter. Rev. Dec 20, Datasheet.   Website: HDLC-ETH Serial Ethernet Converter Rev. Dec 20, 2017 HDLC-ETH Datasheet Email: yacer@yacer.cn Website: www.yacer.cn 1 Overview... 3 1.1 Introduction... 3 1.2 Features... 3 1.3 Applications... 3 1.4 Technical

More information

CAN / RS485. Product Description. Technical Reference Note. Interface Adapter. Special Features

CAN / RS485. Product Description. Technical Reference Note. Interface Adapter. Special Features CAN / Interface Adapter For SHP Series Total Power: < 1 Watts Input Voltage: 5V Internal Outputs: CAN,, USB, I 2 C Special Features Input Protocols: 1) using Modbus 2) CAN using modified Modbus Output

More information

IFC 100 Supplementary instructions

IFC 100 Supplementary instructions IFC 100 Supplementary instructions Signal converter for electromagnetic flowmeters Description of Modbus interface Electronic Revision: ER 3.0.xx Modbus version: 1.0.xx KROHNE CONTENTS IFC 100 1 Important

More information

CONTO IMP PROTOCOL COMMUNICATION CONTENTS 1.0 INTRODUCTION

CONTO IMP PROTOCOL COMMUNICATION CONTENTS 1.0 INTRODUCTION 7/02/12 Pagina 1 di 14 PULSE MANAGER FIRMWARE 1.01 CONTO IMP PROTOCOL COMMUNICATION CONTENTS 1.0 INTRODUCTION 2.0 DATA MESSAGE DESCRIPTION 2.1 Data field description 2.2 Data format 2.3 Description of

More information

VPGate Manual PROFIBUS to serial

VPGate Manual PROFIBUS to serial VPGate Manual PROFIBUS to serial Important information Purpose of the Manual This user manual provides information how to work with the VPGate PROFIBUS to serial. Document Updates You can obtain constantly

More information

Logosol Joystick Node LS-731

Logosol Joystick Node LS-731 Features 2 and 3 axis models Travel ±20 deg Non contact hall effect joystick Mechanical MTBF 15,000,000 cycles 3 pushbuttons Up to 2 stick pushbuttons 8 LEDs Member of Logosol s distributed motion control

More information

Technical Information. Command overview of Vision Systems

Technical Information. Command overview of Vision Systems Technical Information Command overview of Vision Systems Image analysis command Grab image 0x01 X X X X Shutter speed 0x07 X X X X Synchronous flash 0x49 X X X X Video mode 0x00 X X Display 0x05 X X X

More information

WiMOD LR Base Plus Host Controller Interface

WiMOD LR Base Plus Host Controller Interface WiMOD LR Base Plus Host Controller Interface Specification Version 1.2 Document ID: 4000/40140/0125 IMST GmbH Carl-Friedrich-Gauß-Str. 2-4 47475 KAMP-LINTFORT GERMANY Introduction Document Information

More information

Modbus RTU CFW100. User s Manual. Phone: Fax: Web: -

Modbus RTU CFW100. User s Manual. Phone: Fax: Web:  - Modbus RTU CFW100 User s Manual Modbus RTU User s Manual Series: CFW100 Language: English Document Number: 10002909455 / 00 Publication Date: 06/2014 CONTENTS CONTENTS... 3 ABOUT THIS MANUAL... 5 ABBREVIATIONS

More information

HDLC-USB. Portable Protocol Converter. Rev. Dec 25, Datasheet. Website:

HDLC-USB. Portable Protocol Converter. Rev. Dec 25, Datasheet.   Website: HDLC-USB Portable Protocol Converter Rev. Dec 25, 2017 HDLC-USB Datasheet Email: yacer@yacer.cn Website: www.yacer.cn 1 Overview... 3 1.1 Introduction... 3 1.2 Features... 3 1.3 Applications... 3 1.4 Technical

More information

MODBUS Protocol. The ecode Encoders contain both single register (16 bit) and double register (32 bit) values.

MODBUS Protocol. The ecode Encoders contain both single register (16 bit) and double register (32 bit) values. ecode Series MODBUS Protocol Preface This document describes the implementation of the MODBUS protocol in the Scancon ecode Series of communicating encoders. It is not intended to be a description of the

More information

I 2 C Application Note in Protocol B

I 2 C Application Note in Protocol B I 2 C Application Note in Protocol B Description This document is a reference for a possible coding method to achieve pressure, temperature, and status for SMI part readings using I 2 C. This SMI Protocol

More information

on a 35 mm top-hat rail (in accordance with DIN EN TH35) Ambient temperature Operation: C Storage: C

on a 35 mm top-hat rail (in accordance with DIN EN TH35) Ambient temperature Operation: C Storage: C Data sheet PCD1.B1020-A20 E-Line RIO 16DI, 4Rel Starting with FW 1.08.xx The L-Serie E-Line RIO modules are controlled via the RS-485 serial communication protocols S-Bus and Modbus for decentralised automation

More information

JMY504M User's Manual

JMY504M User's Manual JMY504M User's Manual (Revision 3.42) Jinmuyu Electronics Co. LTD 2011/6/28 Please read this manual carefully before using. If any problem, please mail to: Jinmuyu@vip.sina.com Contents 1 Product introduction...

More information

ECE251: Thursday November 8

ECE251: Thursday November 8 ECE251: Thursday November 8 Universal Asynchronous Receiver & Transmitter Text Chapter 22, Sections 22.1.1-22.1.4-read carefully TM4C Data Sheet Section 14-no need to read this A key topic but not a lab

More information

HDLC-PCIE. Synchronous Serial Card. Rev. Dec 22, Datasheet. Website:

HDLC-PCIE. Synchronous Serial Card. Rev. Dec 22, Datasheet.   Website: HDLC-PCIE Synchronous Serial Card Rev. Dec 22, 2017 HDLC-PCIE Datasheet Email: yacer@yacer.cn Website: www.yacer.cn 1 Overview... 3 1.1 Introduction... 3 1.2 Features... 3 1.3 Driver Support... 3 1.4 Applications...

More information

Preliminary Data MOS IC. Type Ordering Code Package SDA Q67100-H5092 P-DIP-8-1

Preliminary Data MOS IC. Type Ordering Code Package SDA Q67100-H5092 P-DIP-8-1 Nonvolatile Memory 1-Kbit E 2 PROM SDA 2516-5 Preliminary Data MOS IC Features Word-organized reprogrammable nonvolatile memory in n-channel floating-gate technology (E 2 PROM) 128 8-bit organization Supply

More information

Basics of UART Communication

Basics of UART Communication Basics of UART Communication From: Circuit Basics UART stands for Universal Asynchronous Receiver/Transmitter. It s not a communication protocol like SPI and I2C, but a physical circuit in a microcontroller,

More information

Rev 2.00 NA-9286 (EtherCAT) Page 1 of 31. FnIO S Series: NA EtherCAT Adapter

Rev 2.00 NA-9286 (EtherCAT) Page 1 of 31. FnIO S Series: NA EtherCAT Adapter Rev 2.00 NA-9286 (EtherCAT) Page 1 of 31 FnIO S Series: NA-9286 EtherCAT Adapter Rev 2.00 NA-9286 (EtherCAT) Page 2 of 31 DOCUMENT CHANGE SUMMARY REV. PAGES REMARKS DATE Editor N/A New Draft release 2012/6/13

More information

XS S ERIES TM PMB US TM O PTION C ARD

XS S ERIES TM PMB US TM O PTION C ARD XS Series PMBus Option Card XS S ERIES TM PMB US TM O PTION C ARD Document: 40110r01 1 Contents 1 Introduction 4 2 Option Card Connectors 4 2.1 PMBus Address..............................................

More information

Optris CT/ CTlaser/ CTvideo communication interface

Optris CT/ CTlaser/ CTvideo communication interface Optris CT/ CTlaser/ CTvideo communication interface Serial interface parameters Protocol Baud rate: 9600 115200, set by user (factory default: 115200) Data bits: 8 Parity: ne Stop bits: 1 Flow control:

More information

FnIO S-Series. FnIO MODBUS Adapter Specification Rev 1.00 NA-9473 (MODBUS/RS485) Page 1 of 30. NA-9473 (MODBUS/RS485) Adapter

FnIO S-Series. FnIO MODBUS Adapter Specification Rev 1.00 NA-9473 (MODBUS/RS485) Page 1 of 30. NA-9473 (MODBUS/RS485) Adapter Rev 1.00 NA-9473 (MODBUS/RS485) Page 1 of 30 FnIO S-Series NA-9473 (MODBUS/RS485) Adapter Rev 1.00 NA-9473 (MODBUS/RS485) Page 2 of 30 DOCUMENT CHANGE SUMMARY REV. PAGES REMARKS DATE Editor Draf t#0 First

More information

OEM Proximity Reader with Keypad Manual REV1

OEM Proximity Reader with Keypad Manual REV1 484-52 OEM Proximity Reader with Keypad Manual REV1 Overview The 484-52 proximity keypad reader provides pin code identification to be used together with a proximity identification card. The 484-52 consists

More information

RS-232 Control of the Advantage DRI

RS-232 Control of the Advantage DRI RS-232 Control of the Advantage DRI Biamp Systems, 14130 N.W. Science Park, Portland, Oregon 97229 U.S.A. (503) 641-7287 an affiliate of Rauland-Borg Corp. Introduction This document contains technical

More information

RS485 MODBUS Module 16I-M

RS485 MODBUS Module 16I-M Version 2.0 12/02/2013 Manufactured for Thank you for choosing our product. This manual will help you with proper support and proper operation of the device. The information contained in this manual have

More information

Planar Simplicity Series

Planar Simplicity Series Planar Simplicity Series RS232 PROTOCOL Document 020-1285-00 1. INTRODUCTION 1.1 Purpose The purpose of this document is to explain in detail the commands and steps that can be used to control a Planar

More information

Flexi Soft T E L E G R A M M L I S T I N G

Flexi Soft T E L E G R A M M L I S T I N G T E L E G R A M M L I S T I N G This document is protected by the law of copyright, whereby all rights established therein remain with the company SICK AG. Reproduction of this document or parts of this

More information

RS232C Serial Interface for Pirani Diaphragm and Pirani Standard Gauges. Caution. binary 8 data bits 1 stop bit no parity bit no handshake

RS232C Serial Interface for Pirani Diaphragm and Pirani Standard Gauges. Caution. binary 8 data bits 1 stop bit no parity bit no handshake General Information The Serial Interface allows the communication of the digital Agilent Pirani Capacitance Diaphragm Gauges (PCG-750, PCG-752) and the digital Agilent Pirani Standard Gauges (PVG-550,

More information

SPI Lasers UK Limited. Serial Command Reference for the PRISM Laser Platform

SPI Lasers UK Limited. Serial Command Reference for the PRISM Laser Platform SPI Lasers UK Limited Serial Command Reference for the PRISM Laser Platform 1 of 89 Table of Contents Introduction... Connecting to the OEM Fibre Laser... Data Link Layer... Frame Structure... Data Element...

More information

The RS-485 user manual for B800 series communication

The RS-485 user manual for B800 series communication The user manual of B800 Series Rs-485 The RS-485 user manual for B800 series RS-232 inbuilt inside the main board of B800 series frequency inverter, we can effect RS-485 through fitting board externally.

More information

Golander Peristaltic Pump MODBUS Communication Instruction

Golander Peristaltic Pump MODBUS Communication Instruction Golander Peristaltic Pump MODBUS Communication Instruction 1 Introduction... 1 2 Modbus Protocol... 2 2.1 Modbus Protocol Model... 2 2.2 Byte Format... 2 2.3 MODBUS Message Timing... 2 2.4 Field... 3 2.5

More information

Preliminary Data MOS IC. Type Ordering Code Package SDA Q67100-H5096 P-DIP-8-1

Preliminary Data MOS IC. Type Ordering Code Package SDA Q67100-H5096 P-DIP-8-1 Nonvolatile Memory 4-Kbit E 2 PROM with I 2 C Bus Interface SDA 2546-5 Preliminary Data MOS IC Features Word-organized reprogrammable nonvolatile memory in n-channel floating-gate technology (E 2 PROM)

More information

E2 Modbus RTU Register Map Revision History Version Comments Author Date 1.02 Previous version PAE 11/06/ Revised to new format PAE 09/03/09

E2 Modbus RTU Register Map Revision History Version Comments Author Date 1.02 Previous version PAE 11/06/ Revised to new format PAE 09/03/09 Application Note Title AN-ODE-01 E2 Modbus RTU Register Map Revision History Version Comments Author Date 1.02 Previous version PAE 11/06/08 1.03 Revised to new format PAE 09/03/09 General This document

More information

EM210 COMMUNICATION PROTOCOL. Version 3 Revision 3

EM210 COMMUNICATION PROTOCOL. Version 3 Revision 3 EM210 COMMUNICATION PROTOCOL Version 3 Revision 3 November 13 th, 2015 Index 1.1 Introduction... 3 1.2 MODBUS functions... 3 Function 03h (Read Holding Registers)... 3 Function 04h (Read Input Registers)...

More information

1 SETTING UP GENERAL FUNCTION TYPE PROTOCOL MESSAGES COMMAND RETURN CODES... 6 TRANSACTION EXAMPLES...

1 SETTING UP GENERAL FUNCTION TYPE PROTOCOL MESSAGES COMMAND RETURN CODES... 6 TRANSACTION EXAMPLES... 1 SETTING UP... 3 1.1 RS232 CONTROL CABLES...3 1.2 RS232 SETTINGS...3 2 GENERAL... 4 3 FUNCTION TYPE... 4 4 PROTOCOL MESSAGES... 4 4.1 MESSAGE FORMATS...4 4.1.1 Message Head...4 4.1.2 Message Body...5

More information

RS485 MODBUS Module 8I8RO

RS485 MODBUS Module 8I8RO Expansion Module 8 digital inputs, 8 relay outputs Version 1.0 3.12.2014 Manufactured for Thank you for choosing our product. This manual will help you with proper support and proper operation of the device.

More information

RS-232 Control of the Advantage EQ281/8, EQ282/8 and Advantage SMS200

RS-232 Control of the Advantage EQ281/8, EQ282/8 and Advantage SMS200 RS-232 Control of the Advantage EQ281/8, EQ282/8 and Advantage SMS200 Biamp Systems, 14130 N.W. Science Park, Portland, Oregon 97229 U.S.A. (503) 641-7287 an affiliate of Rauland-Borg Corp. Introduction

More information

Application Note BDLxxxx RS232 SERIAL INTERFACE COMMUNICATION PROTOCOL (SICP V1.82)

Application Note BDLxxxx RS232 SERIAL INTERFACE COMMUNICATION PROTOCOL (SICP V1.82) Application Note BDLxxxx RS232 SERIAL INTERFACE COMMUNICATION PROTOCOL (SICP V1.82) Table of Contents 1. INTRODUCTION... 1 1.1 PURPOSE... 1 1.2 DEFINITIONS, ABBREVIATIONS AND ACRONYMS... 1 2. COMMAND PACKET

More information

SC2004MBS 20x4 Characters MODBUS RTU Slave LCD

SC2004MBS 20x4 Characters MODBUS RTU Slave LCD SC004MBS 0x4 Characters MODBUS RTU Slave SC004MBS is a MODBUS slave device that receives data from a Master MODBUS device and display them on the panel. The is 0 x 4 characters in size and each character

More information

Motors I Automation I Energy I Transmission & Distribution I Coatings. Modbus RTU CFW300. User s Manual

Motors I Automation I Energy I Transmission & Distribution I Coatings. Modbus RTU CFW300. User s Manual Motors I Automation I Energy I Transmission & Distribution I Coatings Modbus RTU CFW300 User s Manual Modbus RTU User s Manual Series: CFW300 Language: English Document Number: 10003806158 / 02 Publication

More information

Cab Bus Communications Protocol

Cab Bus Communications Protocol Page 1 Overview of Controller Bus The controller bus connects up to 63 Hand-Held Cab Controllers to the Command Station. Data on the bus runs at 9600 bps with 8 data bits, no parity, 2 stop bits and meets

More information

Motors I Automation I Energy I Transmission & Distribution I Coatings. Modbus RTU CFW500. User s Manual

Motors I Automation I Energy I Transmission & Distribution I Coatings. Modbus RTU CFW500. User s Manual Motors I Automation I Energy I Transmission & Distribution I Coatings Modbus RTU CFW500 User s Manual Modbus RTU User s Manual Series: CFW500 Language: English Document Number: 10002253377 / 00 Publication

More information

Number Representations

Number Representations Simple Arithmetic [Arithm Notes] Number representations Signed numbers Sign-magnitude, ones and twos complement Arithmetic Addition, subtraction, negation, overflow MIPS instructions Logic operations MIPS

More information

Protocol of SOJI Fuel level sensor FJ-RS232/FJ-RS485 Version 1.0.2

Protocol of SOJI Fuel level sensor FJ-RS232/FJ-RS485 Version 1.0.2 Protocol of SOJI Fuel level sensor FJRS232/FJRS485 Version 1.0.2 Document preparion Full Name Title Prepared by Nguyen Duy Tan Engineer Approved by Quang Nguyen Project manager Date 1 October 2016 1 October

More information

Computer Control of the Advantage DRC 4+4

Computer Control of the Advantage DRC 4+4 Computer Control of the Advantage DRC 4+4 Biamp Systems, 14130 N.W. Science Park, Portland, Oregon 97229 U.S.A. (503) 641-7287 an affiliate of Rauland-Borg Corp. Introduction This document contains technical

More information

LCM-160. Low-Power Embedded Communication Module. Rev. Dec 26, LCM-160 Datasheet. Website:

LCM-160. Low-Power Embedded Communication Module. Rev. Dec 26, LCM-160 Datasheet.   Website: LCM-160 Low-Power Embedded Communication Module Rev. Dec 26, 2017 LCM-160 Datasheet Email: yacer@yacer.cn Website: www.yacer.cn 1 Overview... 3 1.1 Introduction... 3 1.2 Features... 3 1.3 Applications...

More information

Using the USB and Serial Interfaces

Using the USB and Serial Interfaces Using the USB and Serial Interfaces PulsON 300 and 400 Series TIME DOMAIN Cummings Research Park 4955 Corporate Drive Suite 101 Huntsville, AL 35805 USA http://www.timedomain.com Tel: +1 256.922.9229 +1

More information

NHD-0216K3Z-NS(RGB)-FBW-V3

NHD-0216K3Z-NS(RGB)-FBW-V3 NHD-0216K3Z-NS(RGB)-FBW-V3 Serial Liquid Crystal Display Module NHD- Newhaven Display 0216-2 Lines x 16 Characters K3Z- Model N- Transmissive S(RGB)- Side LED Backlights (Red-Green-Blue) F- FSTN(-) B-

More information

Heat meter PolluStat M-bus communication protocol description

Heat meter PolluStat M-bus communication protocol description Heat meter PolluStat M-bus communication protocol description 1 1. General structure of protocol 1.1. General features of protocol Meter uses a M-bus protocol. Default baud rate: 2400 bps, Even, 1 Stop.

More information

OEM API Specification

OEM API Specification OEM API Specification For Wasatch Photonics OEM Spectrometers WasatchDevices.com Revised 2016-08-26 Page 1 Revision Log Revision Date By Reason 1.0 2016-08-29 J. Traud Initial Release Contents General

More information

Forlong MODBUS Protocol Specification V1.0

Forlong MODBUS Protocol Specification V1.0 Forlong MODBUS Protocol Specification V1.0 2009 / 05 MODBUS APPLICATION PROTOCOL SPECIFICATION V1.0 CONTENTS 1 Introduction. 1.1 Scope of this document.. 1.2 Protocol overview.. 1.3 Contacts... 2 Modbus

More information

NHD 0216K3Z FL GBW. Serial Liquid Crystal Display Module

NHD 0216K3Z FL GBW. Serial Liquid Crystal Display Module NHD 0216K3Z FL GBW Serial Liquid Crystal Display Module NHD Newhaven Display 0216 2 lines x 16 characters K3Z Model F Transflective L Yellow/Green LED backlight G STN Gray B 6:00 view W Wide Temperature

More information

M-BUS PROTOCOL for ML 311

M-BUS PROTOCOL for ML 311 OPERATING MANUAL OPERATING MANUAL M-BUS PROTOCOL for ML 311 Release number: 311_MBUS_EN_01.docx INDEX INTRODUCTION... 3 HARDWARE CONNECTION... 3 SETTING ML 311 TO COMMUNICATE WITH M-BUS PROTOCOL... 4 ABBREVIATION

More information

LCM-200/400 Datasheet

LCM-200/400 Datasheet LCM-200/400 Low-Power Embedded Communication Module Rev. Dec 26, 2017 LCM-200/400 Datasheet Email: yacer@yacer.cn Website: www.yacer.cn 1 Overview... 3 1.1 Introduction... 3 1.2 Features... 3 1.3 Applications...

More information

FnIO S Series: NA MODBUS/TCP(UDP) Adapter. with Two RJ-45

FnIO S Series: NA MODBUS/TCP(UDP) Adapter. with Two RJ-45 Preliminary NA-9289 Page 1 of 34 FnIO S Series: NA-9289 MODBUS/TCP(UDP) Adapter with Two RJ-45 Preliminary NA-9289 Page 2 of 34 DOCUMENT CHANGE SUMMARY REV. PAGES REMARKS DATE Editor Released May.2 1,

More information

NOVOtechnik. Content. TIM CANopen Gebrauchsanleitung TIM CANopen user manual SIEDLE GRUPPE

NOVOtechnik. Content. TIM CANopen Gebrauchsanleitung TIM CANopen user manual SIEDLE GRUPPE Content 9 CANopen 2 9.1 EDS Files 2 9.2 Features 2 9.2.1 Basic information 2 9.2.2 Basics based on CiA DS-301, V4.02 2 9.2.3 Basics based on CiA DSP-406, V3.2 3 9.2.4 Basics SDO communication 3 9.2.5 Basics

More information

RS 485 Mini Modbus 1AO

RS 485 Mini Modbus 1AO RS 485 Mini Modbus 1AO Version 1.0 14/08/2014 Manufactured for Thank you for choosing our product. This manual will help you with proper support and proper operation of the device. The information contained

More information

GIGAVAC Contactors J1939 Protocol

GIGAVAC Contactors J1939 Protocol GIGAVAC Contactors J1939 Protocol Documentation Revision: 7 See the last page for different revisions change history and apply to the firmware revisions to each update. Product Models: GXCAN15/16, MXCAN15/16.

More information

DATA SHEET. article numbers P125-1b & P125-1g

DATA SHEET. article numbers P125-1b & P125-1g LC 16.2 TREND DATA SHEET article numbers P125-1b & P125-1g Pottery House, Pottery Road, Dun Laoghaire, Co. Dublin, Ireland. Tel: +353-1-2350279 Fax: +353-1-2350361 Am Gneisenauflöz 6, 66538 Neunkirchen,

More information

E2 Modbus RTU Register Map

E2 Modbus RTU Register Map Application Note AN ODE 01 E2 Modbus RTU Register Map Author: Peter Evans, Invertek Drives Ltd Revision: 1.02 11 June 2008 Software Version: 1.02 General This document details the Modbus RTU memory mapping

More information

OPTRIS CT/CTL communication interface

OPTRIS CT/CTL communication interface OPTRIS CT/CTL communication interface Serial interface parameters Protocol Baud rate: 9600 115200, set by user (factory default: 9600) Data bits: 8 Parity: ne Stop bits: 1 Flow control: off The protocol

More information

QBridge. I2C, SPI, CAN Control Software User s Manual. Date: Rev 1.3

QBridge. I2C, SPI, CAN Control Software User s Manual. Date: Rev 1.3 QBridge I2C, SPI, CAN Control Software User s Manual Date: 9-10-2005 Rev 1.3 1. Introduction...1 1.1. What QBridge can do?... 1 1.2. Disclaimer... 1 1.3. Operational Format... 1 1.4. QBridge-V2... 1 2.

More information

Appendix to the Operating Instructions RMx621 with ModBus Interface V Connection to ModBus-IDA System

Appendix to the Operating Instructions RMx621 with ModBus Interface V Connection to ModBus-IDA System Operating Instructions Appendix to the Operating Instructions RMx621 with ModBus Interface V3.03.00 Connection to ModBus-IDA System BA231R/09/en/08.09 71027031 MS Word Endress+Hauser 2 Contents 1 General...

More information

JDICON 400/500 JDICON 401/501. B Interface Description. Universal process controller

JDICON 400/500 JDICON 401/501. B Interface Description. Universal process controller JDICON 400/500 Universal process controller JDICON 401/501 Universal profile controller Universal profile generator B 70.3570.2 Interface Description 10.99/00371593 Contents 1 Introduction 5 1.1 Preface...

More information

EVShield Interface Specifications

EVShield Interface Specifications EVShield Advanced Development Guide v1.0 EVShield Interface Specifications Power Specs: EVShield can be powered from external power supply. Max Power Rating: 10.5 Volts DC Minimum 6.6 Volts DC needed to

More information

MODEL CIO-EN PROGRAMMING GUIDE MODBUS/TCP, MODBUS/RTU I/O MODULE. Revision A1 Rapid City, SD, USA, 09/2008

MODEL CIO-EN PROGRAMMING GUIDE MODBUS/TCP, MODBUS/RTU I/O MODULE. Revision A1 Rapid City, SD, USA, 09/2008 PROGRAMMING GUIDE MODEL CIO-EN MODBUS/TCP, MODBUS/RTU I/O MODULE Revision A1 Rapid City, SD, USA, 09/2008 2880 North Plaza Drive, Rapid City, South Dakota 57702 (800) 843-8848 (605) 348-5580 fax (605)

More information

AquaCER with TTL OUTPUT

AquaCER with TTL OUTPUT INSTRUCTION MANUAL AquaCER with TTL OUTPUT DESCRIPTION: The AquaCER TTL is a hydrostatic pressure transmitter, with a capacitive ceramic sensor, and can withstand very high burst pressures. The applied

More information

IMU06WP. What is the IMU06?

IMU06WP. What is the IMU06? IMU06 What is the IMU06? The IMU06 is a compact 6 degree of freedom inertial measurement unit. It provides 3 axis acceleration (maximum 10G) and angular velocities (maximum 300 degrees/s) on both CAN and

More information

Revision 1.3. Sept 26, COM Protocol Manual. for MDC and ADC N 11th St - San Jose CA

Revision 1.3. Sept 26, COM Protocol Manual. for MDC and ADC N 11th St - San Jose CA Revision 1.3 Sept 26, 2018 COM Protocol Manual for MDC and ADC www.mountztorque.com - 1080 N 11th St - San Jose CA 95112-408.292.2214 1 1 Overview and Communication Specifications 1.1 Overview If the PC

More information

VISY-Command. Technical Documentation. with Modbus Protocol. Edition: Version: 1 Art. no.:

VISY-Command. Technical Documentation. with Modbus Protocol. Edition: Version: 1 Art. no.: Technical Documentation with Modbus Procol Edition: 2017-04 Version: 1 Art. no.: 350253 FAFNIR GmbH Schnackenburgallee 149 c 22525 Hamburg Germany Tel.: +49 / 40 / 39 82 07-0 Fax: +49 / 40 / 390 63 39

More information