Equipment Specification

Size: px
Start display at page:

Download "Equipment Specification"

Transcription

1 MULTISLICE CT SCANNER ( 16 Slices ) Merk : Hitachi Japan Model : SUPRIA 5 MHU Price : Rp ,27 No Equipment Specification Scanner Gantry Object for scanning : Whole body including head Scanning time : rotation 0.75/ 1.0/ 1.5/ 2.0 sec Scanning region : 360 continuous scan Max. Number of slice : 16 slice/scan Gantry Opening/Bore diameter : 750 mm Effective FOV : 500 mm FOV Variations : ( 1mm step) Gantry Tilt angle : Forward tilt 30, backward tilt 30 Helical and/or sequential scanning : Yes, available Setting scan position : Light Localizer for: - Median sagital plane - Axial plane x 2 - Coronal plane x 2 Remote control : Yes Scan start/stop button : Standard (with delay time display) Measuring patient dose radiation : Yes Gantry width (front to rear) : 920 mm Patient Table Material of table top : Carbon fiber Table height (vertical speed) : 450-1,000 mm (CT-WT-20) Tabletop width : 400 mm Longitudinal Stroke mm (± 3 mm) Longitudinal Speed 150 mm ± 3mm/s Max. scanable range : 1,200 mm (1,500 with extended tabletop) Maximum Load : 180 kg Free float Standard (Button) X-Ray Generator/Detector X-Ray tube : 5 MHU Max. Output power : 48 kw (400 ma@120 kv) HV Generator Max output 51kW Max. cooling/heat dissipation : 748 KHU/min Tube voltage : 80/100/120/140 kv Tube current : ma (5 step)

2 4 ma Selection : ma Meaurement types 2 types, x16 and 1.25 x 16 Tube Focus Size : 0.7 x 0.8 mm and 1.2 x 1.4 mm CTDI (Surface and body) : CTDI 100 Body: 9.4mGy Spatial resolution : 17.1 lp/cm Detector type : Solid state detector Detector configuration : 880 ch x 32 rows (0.625 mm x 32 rows) Focus to isocenter distance : 606 mm Focus to detector distance : 1072 mm No. Elements per row : 880 ch Number of rows : 32 rows Number of elements : 28,160 No. of elements (usable) : (input) Fastest X-Ray tube Rot speed : 0.75 sec Max. No. of slice per rotation : 16 slice Physical detector size : 20 mm Physical detector use for scan : 20 mm Scanning Function Measurement slice thickness (Scanogram) : x 8 Collimation : x 32 (20 mm used for scan) Collimation at normal scan : mm x mm x mm x mm x mm x 16 Continuous scan time : sec (150 ma to 400 ma) Slice thickness : / 1.25 / 2.5 / 3.75 / 5 / 7.5 / 10 mm * Scanogram : - Max. scanable range : 1,500 mm (with extended tabletop) - Measurement slice thickness : x 8 * Normal Scan: - Scan time : 0.75 / 1.0 / 1.5 / 2.0 sec * Dynamic Scan: - Scan time : 0.75 / 1.0 / 1.5 / 2.0 sec * Volume scan: - Scan time : 0.75 / 1.0 / 1.5 sec - View rate : 1,200 view/sec - Reconstruction method : CORE (3D image reconstruction algorithm) Volume Pitch :

3 Image reconstruction time : 10 images/sec Scan time : (chest-lung field + : 7.2 sec (0.625 mm x 32/P : ) mediastinal space: 300 mm) Total examination time (scan time : Aprox. 50 sec (2.5mm/120 images x 2) + image reconstruction time Scan time : (chest-abdomen : 600 : 14.4 sec (0.625 mm x 32/P : ) Total (scan + image recon time) : Aprox. 25 sec (5.0 mm / 120 images) Confirmatio of images after scan : direct cine 5 Image Display/Proccessing Display monitor : 24 inch LCD Monitor Operating System (OS) : Windows embedded Standard 7 (64 bit) Reconstruction matrix : 512 x 512 Reconstruction time : max. 10 images/sec (ips) Reconstruction algorithm : CORE method (3D) Beam pitch : to Auto MPR : Standard Display matrix : 1,920 x 1,200 Spatial resolution : 17.1 Ip/cm Density resolution : 2.5 mm@ 0.25% Magnetic disck unit: - Storage image : 200,000 images or more - Storage Raw Data : 6,000 scans or more Archival storage : DVD Drive (CD-R, DVD-R) CD Approx. 1,100 images (650 MB) DVD Approx. 8,000 images (4.7 GB) Display grey scale : 256 levels Window level : Standard : -2, ,000 Extended : -32, ,767 Window width : Standard: 1-6,000 Extended: 1-32,767 Data display : patient name, Birthday, Sex, ID No, Slice thickness, tube voltage, tube current, slice position, state of contrast and others Window Process : - Window level/width adjustment - Black-white contrast revesal - Linear/Non linear window - Double window - Level detection Image Display process : - Multi-frame display - Magnification (real time, etc) - Image rotation

4 6 7 - Left-right inversion - Correction - Comment display - Cine display - Substraction/Addition - Edge enhancement/smoothing - Multi slice image addition Image analysis process : - Distance and angle measurement - Profile of CT value - Setting ROI * Shape: Oval, Free * Process: Area, Mean value of CT# * Display: up to 4 ROIs can be displayed * Control: Size, Position, Rotation - Histogram - Display of CT value - Scale Display - Volume calculation 3 D Image process : - MPR (SAG, COR, OBL, CURVE) - MIP, MINIP, RaySum display - Surface rendering - Volume rendering - MARP - Movie display - Perspective method Operator's/Main Console CPU : 64 bit Intel high end processor Operating System : Windows 7 based embedded Memory : 8 GB Hard disk : more than 250 GB Number of display : 1 Display with large LCD monitor 24 inch Control Methode : Keyboard, mouse and intercom Standard function : DICOM 3.0 DICOM Print DICOM Dose SR Predict Scan Intelly IP (Advanced) Intelly EC (SD-Mode) CEV-CPR Auto MPR Standard Acessories

5 Head rest (without arm rest) Head/chin band Patient belt Patient malt Patient fixing parts Instruction manual Note : Harga diatas tidak termasuk 1 Ongkos kirim 2 Dicom Printer 3 Injector 4 Renovasi 5 Ijin BAPETEN 6 Grounding 7 Panel 8 Cabling 9 Pelapis Dinding (PB) 10 Kaca PB 11 Fondasi/Penguatan 12 AC 13 Dehumidifier 14 Pencahayaan

PRODUCT DATA. Advanced 128

PRODUCT DATA. Advanced 128 PRODUCT DATA Advanced 128 Supria 16 Slice CT Puts You On The Path of High Quality, Cost-Effective CT Scanning Solid Capabilities Are Built Into the Supria Addressing the challenges of controlling healthcare

More information

Solid Capabilities Are Built Into the Supria Plus. Putting You On The Path of High Quality, Cost-Effective CT Scanning

Solid Capabilities Are Built Into the Supria Plus. Putting You On The Path of High Quality, Cost-Effective CT Scanning Specification Data Putting You On The Path of High Quality, Cost-Effective CT Scanning Solid Capabilities Are Built Into the Supria Plus Addressing the challenges of controlling healthcare organization

More information

Annexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER

Annexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER Annexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER A) Scanner Design X-Ray generator and tube: 1. Scanner: Whole body spiral CT scanner (16 slices) of latest technology.

More information

SCENARIA 64-Slice CT Putting Advanced CT Within Reach.

SCENARIA 64-Slice CT Putting Advanced CT Within Reach. SCENARIA 64-Slice CT Putting Advanced CT Within Reach. Advancing the technology of Computed Tomography for over 30 years, Hitachi is a recognized innovator of lower dose*, high diagnostic value CT solutions.

More information

SCENARIA Hitachi Innovation at Work for You.

SCENARIA Hitachi Innovation at Work for You. PRODUCT DATA SCENARIA Hitachi Innovation at Work for You. Continuing its commitment to the SCENARIA scalable CT platform, Hitachi introduces cutting-edge 64/128-slice technology with the SCENARIA SE. The

More information

Fits you like no other

Fits you like no other Fits you like no other BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase in room

More information

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead.

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead. 1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central

More information

SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER

SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER A) Scanner Design X-Ray generator and tube: 1. Scanner: Whole body spiral CT scanner (16 slices) of latest technology. 2. X-Ray

More information

Optimisation of Toshiba Aquilion ONE Volume Imaging

Optimisation of Toshiba Aquilion ONE Volume Imaging Optimisation of Toshiba Aquilion ONE Volume Imaging Jane Edwards, RPRSG Royal Free London NHS Foundation Trust Dr Mufudzi Maviki, Plymouth Hospitals NHS Trust Background In 2011/12 Radiology at RFH was

More information

RADIOLOGY AND DIAGNOSTIC IMAGING

RADIOLOGY AND DIAGNOSTIC IMAGING Day 2 part 2 RADIOLOGY AND DIAGNOSTIC IMAGING Dr hab. Zbigniew Serafin, MD, PhD serafin@cm.umk.pl 2 3 4 5 CT technique CT technique 6 CT system Kanal K: RSNA/AAPM web module: CT Systems & CT Image Quality

More information

Whole Body X-ray CT System Supria. DICOM Conformance Statement

Whole Body X-ray CT System Supria. DICOM Conformance Statement Whole Body X-ray CT System Supria DICOM Conformance Statement Tokyo, Japan E1E-HC0001-02 Hitachi, Ltd. 2016. All rights reserved. ii E1E-HC0001 History Revision Description Date Rev. 1.0 Initial. 2016/4/1

More information

SRI GURU RAMDAS INSTITUTE OF MEDICAL SCIENCES & RESEARCH MEHTA ROAD, VALLAH, SRI AMRITSAR

SRI GURU RAMDAS INSTITUTE OF MEDICAL SCIENCES & RESEARCH MEHTA ROAD, VALLAH, SRI AMRITSAR SRI GURU RAMDAS INSTITUTE OF MEDICAL SCIENCES & RESEARCH MEHTA ROAD, VALLAH, SRI AMRITSAR TECHNICAL SPECIFICATIONS FOR PURCHASE OF CT SIMULATOR FOR DEPARTMENT OF RADIOTHERAPY 1. MANDATORY REQUIREMENTS

More information

Medical Image Processing: Image Reconstruction and 3D Renderings

Medical Image Processing: Image Reconstruction and 3D Renderings Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,

More information

Fits you like no other

Fits you like no other Fits you like no other Philips BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase

More information

Whole Body Submillimeter Scan

Whole Body Submillimeter Scan 128 Whole Body Submillimeter Scan The real value of 64ch/128slice CT systems does not come from the ability to perfom cardiac scans but the capability to scan all parts of the body in high definition submillimeter

More information

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA Computed Tomography Principles, Design, Artifacts, and Recent Advances THIRD EDITION Jiang Hsieh SPIE PRESS Bellingham, Washington USA Table of Contents Preface Nomenclature and Abbreviations xi xv 1 Introduction

More information

Whole Body X-ray CT System Supria. DICOM Conformance Statement

Whole Body X-ray CT System Supria. DICOM Conformance Statement Whole Body X-ray CT System Supria DICOM Conformance Statement Tokyo, Japan (SN-C633E) Copyright Hitachi Medical Corporation. 2013, 2014. All rights reserved. History Revision Description Date Rev. 1.0

More information

Z-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL

Z-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL Control-X Medical Z-MOTION Compact design, low ceiling height requirement Motorized and manual movement capability Wide motion / SID range Best-in-class image quality Flexible connectivity to PACS systems

More information

Fujifilm DR Solution. FDR AcSelerate. The new pinnacle in diagnostic imaging from Fujifilm ISS. CsI. Dynamic Visualization. Technology.

Fujifilm DR Solution. FDR AcSelerate. The new pinnacle in diagnostic imaging from Fujifilm ISS. CsI. Dynamic Visualization. Technology. Fujifilm DR Solution FDR AcSelerate The new pinnacle in diagnostic imaging from Fujifilm CsI Scintillator ISS Technology Dynamic Visualization Welcome to the X-ray room of the future! A streamlined solution

More information

Symbia T Series System Specifications Answers for life.

Symbia T Series System Specifications Answers for life. www.siemens.com/mi System Specifications Answers for life. 2 Diagnostic Excellence Today s healthcare institutions have increasingly diverse patient needs, greater budget limitations and the need for higher

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

Lower Dose with Confidence

Lower Dose with Confidence Lower Dose with Confidence The Next Generation SCENARIA CT is Here. 64 or 128 Slice Capability The SCENARIA SE delivers lower dose technologies*, enhanced patient access, fast workflow speed and added

More information

Computer-Tomography II: Image reconstruction and applications

Computer-Tomography II: Image reconstruction and applications Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement SCENARIA Tokyo, Japan E1E-HC0002-01 Hitachi, Ltd. 2016. All rights reserved. History Revision Description Date Rev. 1.0 Initial. 2016/04/01 ii E1E-HC0002 Contents History...

More information

DX-D 800 DX-D 800 DYNAMIC, 3-IN-1 DIRECT RADIOGRAPHY SYSTEM OFFERING REAL TIME IMAGES FOR FLUOROSCOPY, GENERAL RADIOGRAPHY AND DIRECT EXPOSURES.

DX-D 800 DX-D 800 DYNAMIC, 3-IN-1 DIRECT RADIOGRAPHY SYSTEM OFFERING REAL TIME IMAGES FOR FLUOROSCOPY, GENERAL RADIOGRAPHY AND DIRECT EXPOSURES. D I R E C T R A D I O G R A P H Y S Y S T E M DX-D 800 The DX-D 800, part of Agfa HealthCare s family of Direct Radiography (DR) solutions, provides enhanced flexibility in diagnostics through its direct

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

Hitachi Medical Systems America, Inc. Hitachi Medical Corporation

Hitachi Medical Systems America, Inc. Hitachi Medical Corporation Advanced 128 Advanced 128 128 Slice Capability Enlarged 75cm Aperture Thin Gantry Profile 88cm Intelli IP Advanced, Iterative Reconstruction IntelliCenter Auto-Lateral Shifting Table, up to 16cm Range

More information

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed

More information

O-ARM IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE

O-ARM IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE SYSTEM FEATURES CATEGORY PHYSICAL DIMENSIONS IMAGING MODALITY PERFORMANCE SPECIFICATION Length Width Height Weight Gantry Opening Bore Diameter Single Plane

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement ECLOS Tokyo, Japan (SN-C357E) Copyright Hitachi Medical Corporation. 2007, 2008. All rights reserved. History Revision Description Date Rev. 1.0 Initial. 2007/02/01 Rev 2.0

More information

International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN:

International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: COMPARISON OF RADIATION DOSE AND IMAGE QUALITY BETWEEN THE SINGLE DETECTOR CT AND MULTI DETECTOR CT ON HEAD EXAMINATION: STUDY PHANTOM Samburi Master of Physics, University of Diponegoro, Indonesia Agency

More information

icatvision Quick Reference

icatvision Quick Reference icatvision Quick Reference Navigating the i-cat Interface This guide shows how to: View reconstructed images Use main features and tools to optimize an image. REMINDER Images are displayed as if you are

More information

PURE. ViSION Edition PET/CT. Patient Comfort Put First.

PURE. ViSION Edition PET/CT. Patient Comfort Put First. PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

Routine excellence. Philips Ingenuity Core specifications

Routine excellence. Philips Ingenuity Core specifications Routine excellence Philips Ingenuity Core specifications Table of contents 1 Introduction 3 2 User interface 4 2.1 ipatient key benefits 4 2.2 ExamCards 4 2.3 ScanRuler 4 3 DoseWise 5 3.1 DoseRight Index

More information

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist CT Protocol Review: Practical Tips for the Imaging Physicist Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center August 8, 2013 AAPM Annual Meeting Goals Understand purpose and importance

More information

CT: Physics Principles & Equipment Design

CT: Physics Principles & Equipment Design CT: Physics Principles & Equipment Design James Kofler, Ph.D Radiology Mayo Clinic Rochester, MN June 27, 2012 Disclosures Nothing to disclose Learning Objectives Understand fundamental concepts of - CT

More information

Best inspection system for after reworking of the BGA, and other electronic parts.

Best inspection system for after reworking of the BGA, and other electronic parts. MSX1000HR X Ray Inspection System High resolution micro-focus system, by new developed X ray generator. Easy and safety operation system, by the original radiation prevention cabinet. Best inspection system

More information

Philips SPECT/CT Systems

Philips SPECT/CT Systems Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET

More information

Athena Radiology Medical Workstation

Athena Radiology Medical Workstation High productivity and integration Athena DICOM Viewer is being designed according with the suggestions and the necessities of medical radiologists. From the design of its interface to the advanced image

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

AIDR 3D Iterative Reconstruction:

AIDR 3D Iterative Reconstruction: Iterative Reconstruction: Integrated, Automated and Adaptive Dose Reduction Erin Angel, PhD Manager, Clinical Sciences, CT Canon Medical Systems USA Iterative Reconstruction 1 Since the introduction of

More information

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing ImPACT Information Leaflet No. 1: CT Scanner Acceptance Testing Version 1.02, 18/05/01 CONTENTS: 1. SCOPE OF LEAFLET 2. GENERAL PRINCIPLES OF ACCEPTANCE AND COMMISSIONING 2.1 PHANTOMS 2.2 EXPOSURE AND

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60601-2-44 2001 AMENDMENT 1 2002-09 Amendment 1 Medical electrical equipment Part 2-44: Particular requirements for the safety of X-ray equipment for computed tomography Amendement

More information

A closer look at CT scanning

A closer look at CT scanning Vet Times The website for the veterinary profession https://www.vettimes.co.uk A closer look at CT scanning Author : Charissa Lee, Natalie Webster Categories : General, Vets Date : April 3, 2017 A basic

More information

The protocols used to scan the musculoskeletal system are tailored to each patient and

The protocols used to scan the musculoskeletal system are tailored to each patient and Chapter 22. Musculoskeletal Protocols The protocols used to scan the musculoskeletal system are tailored to each patient and region being examined. The clinical indication for the examination will also

More information

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging

More information

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon WHITE PAPER Introduction An image guidance system based on Cone Beam CT (CBCT) is included in Leksell

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

Micro-CT Methodology Hasan Alsaid, PhD

Micro-CT Methodology Hasan Alsaid, PhD Micro-CT Methodology Hasan Alsaid, PhD Preclinical & Translational Imaging LAS, PTS, GlaxoSmithKline 20 April 2015 Provide basic understanding of technical aspects of the micro-ct Statement: All procedures

More information

Baby Explorer 2002 Software Description

Baby Explorer 2002 Software Description Baby Explorer 2002 Baby Explorer is a cost effective software only solution that allows sonographers to quickly and easily generate fetal images for expectant parents months before the baby is born. The

More information

CTA HEAD Perfusion AqONE without and with IV Contrast

CTA HEAD Perfusion AqONE without and with IV Contrast CTA HEAD Perfusion AqONE without and with IV Contrast Patient Position Adult Contrast Adult Injection Rate Supine IOML perpendicular to table top. IV: 100 ml with helical head CTA 50 ml without helical

More information

VJ Technologies Inspection Services Division. By Bob Maziuk, February 2015

VJ Technologies Inspection Services Division. By Bob Maziuk, February 2015 White Paper on the Inspection of Underground Electrical Conductors, Insulation, and Splices in Oil-Filled Piping Using High Energy Computed Tomography (CT) VJ Technologies Inspection Services Division

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

AAPM Standard of Practice: CT Protocol Review Physicist

AAPM Standard of Practice: CT Protocol Review Physicist AAPM Standard of Practice: CT Protocol Review Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center September 11, 2014 2014 Texas Radiation Regulatory Conference Goals Understand purpose

More information

VXRE Reconstruction Software Manual

VXRE Reconstruction Software Manual VXRE Reconstruction Software Manual Version 1.7.8 3D INDUSTRIAL IMAGING 3D Industrial Imaging Co.,Ltd. Address :#413 Institute of Computer Technology, Seoul National University, Daehak-dong, Gwanak-gu,

More information

S. Guru Prasad, Ph.D., DABR

S. Guru Prasad, Ph.D., DABR PURPOSE S. Guru Prasad, Ph.D., DABR Director of Medical Physics IAEA Consultant NorthShore University Health System and University of Chicago, Pritzker School of Medicine Current TPS utilize more information

More information

abc Report Wide bore CT scanner comparison report version 12 February 2005 best choice best practice nww.medical-devices.nhs.

abc Report Wide bore CT scanner comparison report version 12 February 2005 best choice best practice  nww.medical-devices.nhs. abc February 2005 Report 05015 Wide bore CT scanner comparison report version 12 best choice best practice www.mhra.gov.uk nww.medical-devices.nhs.uk About MHRA evaluation reports. What you can expect.

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

DX-D 100 with wireless detector

DX-D 100 with wireless detector M O B I L E D R S O L U T I O N DX-D 100 with wireless detector Patients who most need imaging exams may lack the mobility necessary to move to the X-ray room or to position themselves properly for optimum

More information

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Improvement of Efficiency and Flexibility in Multi-slice Helical CT J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): 408 412 DOI: 10.1007/s12204-008-0408-x Improvement of Efficiency and Flexibility in Multi-slice Helical CT SUN Wen-wu 1 ( ), CHEN Si-ping 2 ( ), ZHUANG Tian-ge

More information

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice Guido Hilgers, a Tonnis Nuver, André Minken Department

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

GPU implementation for rapid iterative image reconstruction algorithm

GPU implementation for rapid iterative image reconstruction algorithm GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial

More information

Lorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc.

Lorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc. Lorad FFDM QC Procedures for Medical Physicist Tao Wu, Ph.D. Hologic, Inc. Lorad Selenia FFDM Tests Following Lorad QC Manual Collimation Assessment Artifact Evaluation System Resolution Phantom Image

More information

product data Raybow dr mobile digital radiographic system with wireless detector product data

product data Raybow dr mobile digital radiographic system with wireless detector product data Raybow dr mobile digital radiographic system with wireless detector Integrating all the advanced technologies, the ergonomic and high quality system Raybow dr lets you perform digital radiographies anywhere:

More information

ImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners

ImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners ImPACT Imaging Performance Assessment of CT Scanners A Medical Devices Agency Evaluation Group CT dosimetry and a data base for CTDI values EFOMP Workshop at ECR 2002 S. Edyvean, ImPACT St George s Hospital,

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

BLUEPRINT TM 3D Planning Software - v1.4 scan protocol

BLUEPRINT TM 3D Planning Software - v1.4 scan protocol S H O U L D E R Solutions by Tornier 3 D P L A N N I N G S O F T W A R E B L U E P R I N T T M 3 D P L A N N I N G S O F T W A R E S C A N P R O T O C O L BLUEPRINT TM 3D Planning Software - v1.4 scan

More information

DICOM Conformance Statement. CT AV Series Product Line. Document Number February 1997

DICOM Conformance Statement. CT AV Series Product Line. Document Number February 1997 Philips Medical Systems DICOM Conformance Statement CT AV Series Product Line Document Number 4522 983 65911 28 February 1997 Copyright Philips Medical Systems Nederland B.V. 1997 Philips Medical Systems

More information

DX-D WITH DX-D 40/45 DETECTOR FAMILY MOBILE DR SOLUTION

DX-D WITH DX-D 40/45 DETECTOR FAMILY MOBILE DR SOLUTION DX-D 100 + WITH DX-D 40/45 DETECTOR FAMILY MOBILE DR SOLUTION WITH ITS EXCELLENT IMAGE QUALITY AND FLEXIBLE HANDLING, THE MOBILE DX-D 100 + WITH WIRELESS DETECTOR OFFERS FAST IMAGING THAT CAN BE VALIDATED

More information

Image Post-Processing, Workflow, & Interpretation

Image Post-Processing, Workflow, & Interpretation Image Post-Processing, Workflow, & Interpretation Richard L. Hallett, MD Chief, Cardiovascular Imaging Northwest Radiology Network Indianapolis, IN Adjunct Assistant Professor Stanford University Stanford,

More information

Some reference material

Some reference material Some reference material Physics reference book on medical imaging: A good one is The Essential Physics of Medical Imaging, 3 rd Ed. by Bushberg et al. ($170! new). However, there are several similar books

More information

xorantech.com Suite of DR Products

xorantech.com Suite of DR Products xorantech.com Suite of DR Products 2 / xorantech.com Xoran provides unsurpassed, white-glove customer service and high quality, reliable products that are user- and patient-friendly Xoran is the pioneer

More information

Fundamentals of CT imaging

Fundamentals of CT imaging SECTION 1 Fundamentals of CT imaging I History In the early 1970s Sir Godfrey Hounsfield s research produced the first clinically useful CT scans. Original scanners took approximately 6 minutes to perform

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

LAB DEMONSTRATION COMPUTED TOMOGRAPHY USING DESKCAT Lab Manual: 0

LAB DEMONSTRATION COMPUTED TOMOGRAPHY USING DESKCAT Lab Manual: 0 LAB DEMONSTRATION COMPUTED TOMOGRAPHY USING DESKCAT Lab Manual: 0 Introduction This lab demonstration explores the physics and technology of Computed Tomography (CT) and guides the student and instructor

More information

Fischerscope X Ray CIRCULARS 07/01

Fischerscope X Ray CIRCULARS 07/01 Fischerscope X Ray CIRCULARS 07/01 New instrument Fischerscope X-Ray XDVM-P As we informed in December, both the instruments Fischerscope X-RAY XDVM-T7-W 602-636 (measuring head 602-557) and Fischerscope

More information

Comparison of the Imaging Performance of CT Scanners, Issue 11

Comparison of the Imaging Performance of CT Scanners, Issue 11 Imaging Performance Assessment of CT Scanners Comparison of the Imaging Performance of CT Scanners, Issue 11 Comparison data is presented on the following scanners : Elscint Twin Flash IGE HiSpeed Advantage

More information

First CT Scanner. How it Works. Contemporary CT. Before and After CT. Computer Tomography: How It Works. Medical Imaging and Pattern Recognition

First CT Scanner. How it Works. Contemporary CT. Before and After CT. Computer Tomography: How It Works. Medical Imaging and Pattern Recognition Computer Tomography: How t Works Medical maging and Pattern Recognition Lecture 7 Computed Tomography Oleh Tretiak Only one plane is illuminated. Source-subject motion provides added information. 2 How

More information

DX-D 100 WITH WIRELESS DETECTOR

DX-D 100 WITH WIRELESS DETECTOR DX-D 100 WITH WIRELESS DETECTOR MOBILE DR SOLUTION WITH ITS EXCELLENT IMAGE QUALITY AND FLEXIBLE HANDLING, THE MOBILE DX-D 100 WITH WIRELESS DETECTOR OFFERS FAST IMAGING THAT CAN BE VALIDATED IMMEDIATELY.

More information

Computed Tomography Imaging: CT Protocol Management. Caveat 8/3/2017

Computed Tomography Imaging: CT Protocol Management. Caveat 8/3/2017 Computed Tomography Imaging: CT Protocol Management Mark P. Supanich, Ph.D., DABR AAPM Annual Meeting 3 rd August, 2017 Slides at: goo.gl/k8n8jf Rush is an academic health system comprising Rush University

More information

The 3D Accuitomo 80 offers a minute voxel. size of just 80 μm (micrometer). This super-fine. voxel combined with the unit s 13 bit grayscale

The 3D Accuitomo 80 offers a minute voxel. size of just 80 μm (micrometer). This super-fine. voxel combined with the unit s 13 bit grayscale 2 Thinking ahead. Focused on life. REALIZED: Groundbreaking Resolution of 80 µm Voxel 05 The Crystallization of cutting-edge mechatronics and High Definition display algorithm The 3D Accuitomo 80 offers

More information

DICOM CONFORMANCE STATEMENT MEDIA STORAGE FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A. with XIDF-037A and XIDF-038A or SRS-1000A

DICOM CONFORMANCE STATEMENT MEDIA STORAGE FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A. with XIDF-037A and XIDF-038A or SRS-1000A DICOM CONFORMANCE STATEMENT MEDIA STORAGE FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A with XIDF-037A and XIDF-038A or SRS-1000A (MIIXR0002EAC) 2001 IMPORTANT! (1) No part of this manual may

More information

Brilliance CT Big Bore.

Brilliance CT Big Bore. 1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.

More information

CT Dose Profiler. Probe for evaluation of CT systems. CT Dose Profiler User's Manual - English - Version 6.4A. RTI article number:

CT Dose Profiler. Probe for evaluation of CT systems. CT Dose Profiler User's Manual - English - Version 6.4A. RTI article number: CT Dose Profiler Probe for evaluation of CT systems - English - Version 6.4A RTI article number: 9630512-00 CT Dose Profiler The CT Dose Profiler probe makes it possible to evaluate the performance of

More information

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications 4th International Symposium on NDT in Aerospace 2012 - Tu.3.A.1 Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications Malte KURFISS 1, Gerd STRECKENBACH 2 1 YXLON International

More information

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. 2-D D Dose-CT Mapping in Geant4 Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. Table of Contents Background & Purpose Materials Methods

More information

LAMBDA DATA SHEET. ANALOGIC mammography unit. ANALOGIC mammography unit. Rev. n.05 of 15/01/13

LAMBDA DATA SHEET. ANALOGIC mammography unit. ANALOGIC mammography unit. Rev. n.05 of 15/01/13 NOTE The manufacturer reserves the right to make further improvements while keeping main features unchanged MAIN CHARACTERISTICS Line voltage 220/230/240 VAC 10% @ 50/60 Hz Power 6.6 kva (0.5 kva stand-by)

More information

A study on image quality provided by a kilovoltage cone-beam computed tomography

A study on image quality provided by a kilovoltage cone-beam computed tomography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 1, 2013 A study on image quality provided by a kilovoltage cone-beam computed tomography Julia Garayoa a and Pablo Castro Servicio de Radiofísica,

More information

DICOM Enhanced XA/XRF Object New Dimensions for X-Ray Projection Imaging. (DICOM Supplement 83)

DICOM Enhanced XA/XRF Object New Dimensions for X-Ray Projection Imaging. (DICOM Supplement 83) DICOM Enhanced XA/XRF Object New Dimensions for X-Ray Projection Imaging (DICOM Supplement 83) Authors: Heinz Blendinger Bas Revet Francisco Sureda Rainer Thieme Speaker: Heinz Blendinger Siemens Medical

More information