Integration of Agilent OpenLAB CDS EZChrom Edition with OpenLAB ECM Compliance with 21 CFR Part 11

Size: px
Start display at page:

Download "Integration of Agilent OpenLAB CDS EZChrom Edition with OpenLAB ECM Compliance with 21 CFR Part 11"

Transcription

1 OpenLAB CDS Integration of Agilent OpenLAB CDS EZChrom Edition with OpenLAB ECM Compliance with 21 CFR Part 11 Technical Note Introduction Part 11 in Title 21 of the Code of Federal Regulations includes the US Federal guidelines for storing and protecting electronic records and applying electronic signatures. The intent of these guidelines is to ensure that applicable electronic records are reliable, authentic and maintained with high integrity. This technical note describes features and functionality of Agilent OpenLAB CDS EZChrom Edition, revision A or higher, in combination with Agilent OpenLAB Enterprise Content Manager (ECM) revision SP1 HF3 or higher, which enable them to meet the guidelines of 21 CFR Part 11. This document examines each section of 21 CFR Part 11 and provides a recommended remediation approach using the integration of OpenLAB CDS EZChrom Edition (revision A or higher) with Agilent OpenLAB ECM (revision SP1 HF3 or higher). Agilent OpenLAB CDS EZChrom Edition revision A provides a tight and seamless integration into Agilent OpenLAB ECM. This integrated solution provides the necessary controls for managing system access, user management, data transfer handling and audit trail functionality, while Agilent OpenLAB ECM ensures secure record keeping and data archival. Agilent OpenLAB ECM is an electronic library. It provides a secure compliant data repository, powerful search capabilities, data viewers and archiving features for all data in the lab independent from its data format. OpenLAB ECM is proven to satisfy compliance needs as mandated by regulations such as 21 CFR Part 11. It has been implemented by many leading life science companies for this reason. While the integration of OpenLAB CDS EZChrom Edition with OpenLAB ECM provides significant and extensive compliance tools to facilitate compliance, the ultimate responsibility rests with the enterprise employing the technology. Complete compliance therefore requires that the enterprise employing this functionality have the appropriate internal Standard Operating Procedures (SOP). Articles that can only be satisfied by an SOP are indicated as such in the detail below. Table 1 gives an overview of the different sections of the Part-11 regulations, and which of them are applicable for OpenLAB CDS EZChrom Edition with OpenLAB ECM.

2 21 CFR Part 11 sections addressed by the integration of Agilent OpenLAB CDS EZChrom Edition (revision A or higher) with Agilent OpenLAB ECM (revision SP1 HF3 or higher) Possible scenarios with EZChrom operated in a closed system Electronic record only (without signature) Electronic signature based upon User ID and Password 11.1, 11.2, 11.3 (Scope, implementation, definition) (Controls for closed systems) 3 3 The integration of OpenLAB CDS EZChrom Edition with OpenLAB ECM enables the operation of OpenLAB CDS EZChrom Edition in full support of all compliance requirements mandated by 21 CFR Part 11 for a closed system. In particular it ensures: Accurate and complete copies of records Versioning of all relevant records for traceability (Controls for open systems) (Signature manifestations) (Signature record linking) (e-sig general requirements) (e-sig not biometric) Preservation of records between their creation and their automatic transfer to OpenLAB ECM immediately after acquisition, reprocessing or interactive modifications Controlled copies of the data Mandatory login to OpenLAB ECM before allowing appropriate access to EZChrom Records of changes captured in user-independent time-stamped audit trails (b) (e-sig biometric) (a), (b), (d) (Controls for ID codes / passwords) (c), (e) (Token cards & other ID devices) 3 Table 1: Applicable sections of 21 CFR Part 11 for OpenLAB CDS EZChrom Edition/OpenLAB ECM operated in a closed system ( 3= applicable, = not applicable) These settings can be configured during or after installation to meet specific standard operation procedures and security guidelines. This includes customizable EZChrom security roles and privileges that provide restricted access to EZChrom functionality to individual users or groups. Changes to the security configuration can be made only by a dedicated system administrator. Details and recommendations for configuration of the system are outlined in the manual OpenLAB CDS Administration Guide, Part Number M , available on the OpenLAB CDS EZChrom Edition installation media. The integration of OpenLAB CDS EZChrom Edition with OpenLAB ECM was not specifically designed for operation in an open system. NOTE: In this technical note the term EZChrom (in the context of integration with OpenLAB ECM) always refers to OpenLAB CDS EZChrom Edition with the feature set for the integration with OpenLAB ECM enabled.

3 11.10 Controls for Closed Systems Section Question OpenLAB CDS EZChrom Edition/OpenLAB ECM Integration 11.10(a) 11.10(b) 11.10(c) 11.10(d) 11.10(e) 11.10(e) 11.10(e) 11.10(f) 11.10(g) 11.10(h) 11.10(i) 11.10(j) (k)(1) (k)(2) Has the system been validated in order to ensure accuracy, reliability, consistent intended performance, and the ability to discern invalid or altered records? Is the system capable of generating accurate and complete copies of all required records in both human readable and electronic form suitable for inspection, review and copying by the FDA? Are the records protected for accurate and convenient retrieval throughout the record retention period? Is system access limited to authorized individuals? Is there a secure, computer-generated audit trail that independently records the date and time of operator entries and actions that create, modify, or delete electronic records? When records are changed, is previously recorded information left unchanged? Are electronic audit trails saved at least as long as their subject electronic records and available for agency review and copying? Are operational system checks used to enforce permitted sequencing of steps and events? Are authority checks in place to ensure that only authorized individuals can use the system, electronically sign a record, access the operation or computer system input or output device, alter a record, or perform the current operation? Are device checks used to determine, as appropriate, the validity of the source of data or operational instruction? Do the persons who develop, maintain, or use electronic records and signature systems have the education, training, and experience to perform their assigned tasks? Have written policies that hold individuals accountable and responsible for actions initiated under their e-signatures in order to deter record and signature falsification been established and followed? Are there adequate controls over the distribution of, access to, and use of documentation for system operation and maintenance? Are there formal revision and change control procedures to maintain an audit trail that documents time-sequenced development and modification of systems documentation? Yes. Agilent Technologies has extensively validated the performance of EZChrom / OpenLAB ECM with tests written to specifically evaluate accuracy, reliability and consistent performance. The integrated solution of EZChrom with OpenLAB ECM incorporates the use of byte- order dependent check sums at each file transfer operation to ensure that records are valid and unaltered. Agilent Technologies recommends making use of Installation Qualification and Operation Qualification (IQ/OQ) service for EZChrom / OpenLAB ECM to validate the on-site system per compliance requirements. Yes. OpenLAB ECM stores all data types, from raw machine data to printable reports. All files are stored complete and unaltered in the original format. Raw, meta and result data generated by EZChrom are stored and managed in OpenLAB ECM. Printed reports can be stored as PDF files which can be made available for review without the source application being installed on the client machine. Depending on the data type, a wide selection of file viewers is available to view the original electronic record without the original application. Yes. Raw data, meta data and result data generated by EZChrom are stored and managed in OpenLAB ECM. Data stored in OpenLAB ECM resides in a protected storage location or archive media or both. When archived, the media may be on-line, near-line or off-line. Regardless of the physical location of the data, it remains searchable to all users with appropriate privileges. The individual users do not have access to the physical storage location of the files. The functionality is complemented by additional procedural controls that should be defined and implemented by the system administrator based on company-wide security policies. These policies should manage practices such as access to client computers and password renewal frequency. Yes. Each user is identified by a unique user ID and password combination. Access to EZChrom / OpenLAB ECM requires the entry of both identification components: user ID and password. In EZChrom / OpenLAB ECM, all file and software functionality access is controlled by privileges and roles assigned to individual users or groups of users. The system administrator determines levels of access. Yes. All activities in EZChrom / OpenLAB ECM are recorded in secure, computer-generated, time-stamped audit trails. Audit trails exist for system, instruments, methods, sequences, data and report templates. The data audit trail is embedded within the data itself to ensure long-term retention and association. Entries in the EZChrom / OpenLAB ECM audit trails are non-editable, non-deletable. Even the removal of records from OpenLAB ECM by an authorized user does not affect existing entries in the audit trail. Yes. All entries in the EZChrom / OpenLAB ECM audit trails are non-editable, non-deletable. Even the removal of records from OpenLAB ECM by an authorized user does not affect existing entries in the audit trail. Yes. Audit trail entries are stored in the OpenLAB ECM repository as part of a file s metadata and are kept throughout the electronic records retention period. The audit trail may be reviewed and printed. The OpenLAB ECM audit trails are unbreakably linked to the record. System-related activities such as logon events are unbreakably linked to the system. Yes. In all EZChrom / OpenLAB ECM functions, when a sequencing of events is required, system checks enforce it. In EZChrom users are stepped through sequences and events; data cannot be acquired unless parameters are within valid instrument and integration ranges. Users are prompted with an error message when steps are performed out of sequence. Yes. Users cannot gain access to the software modules of EZChrom / OpenLAB ECM without a valid user ID, password and account. Once logged in, that user s access to files and software functionality (including but not limited to signing a file, inputting values, or altering a record) is determined by the privileges assigned to the user. Yes. The integration with instruments through comprehensive automation and control provides various levels of device and validity checks depending on the instrument make and model. User entry fields provide feedback to the user about the entry types and ranges that are valid for that field in EZChrom. Yes. Records of the educational and employment history of Agilent Technologies employees are verified and kept with personnel records that can be made available during an on-site audit. In addition, all Agilent Technologies employees who work with regulations have attended training workshops for regulatory requirements. Users of EZChrom / OpenLAB ECM at a customer location will be required to show records or education, training and/or experience with the system. Agilent provides a basic familiarization during the installation of the product for system users. Training courses for administrators as well as users are available.. It is the responsibility of the organization implementing electronic signatures to develop written policies which ensure that individuals responsible for signing documents understand that their electronic signature is as equally binding as their handwritten signature.. While documentation is available for the integration of EZChrom / OpenLAB ECM for users and administrators, controls over the storage and distribution of this material are the responsibility of the organization that implements and uses the system. Yes. Agilent Technologies quality process includes formal written revision and change control procedures for system documentation. The process includes the use of OpenLAB ECM for development of system documentation. All revisions to the documents kept are time stamped and audit-trailed.

4 11.30 Controls for Open Systems Are there procedures and controls used to protect the authenticity, integrity and confidentiality of the electronic records from their creation point to the point of their receipt? Yes. When a file is transferred to or within OpenLAB ECM, a byte-order dependent checksum is calculated on the file in its source location. A copy of the file is made of the file in the destination location where a second checksum is calculated. The two values are compared and only if they are identical, is the transfer complete. If the values do not match, an error message is generated. The integration of EZChrom / OpenLAB ECM was not specifically designed for operation in an open system Are additional measures used to ensure the confidentiality of the electronic records from the point of their creation to the point of their receipt? Yes. OpenLAB ECM supports the use of Secure Socket Layer (SSL) encryption for security during data transmission. SSL breaks a single file into very small data packets. These data packets are individually encrypted with configurable 64-bit or 128-bit encryption before being transmitted. On the receiving side the data packets are decrypted and reassembled. The integration of EZChrom with OpenLAB ECM supports SSL encryption. The integration of EZChrom with OpenLAB ECM was not specifically designed for operation in an open system Signature Manifestation 11.50(a) Do the signed electronic records contain information associated with the signing that clearly indicates the following: printed name of signer date and time the signature was executed meaning associated with the signature? Yes. The EZChrom electronic signature manifestation includes: The user ID in addition to the full name of the signer The date and time that the signature was applied The user configurable meaning associated with the signature An esignature time-out is provided for added security. In addition, EZChrom result sets or SSIZip files can be electronically signed in OpenLAB ECM. The OpenLAB ECM electronic signature manifestation includes: The user ID in addition to the full name of the signer The signer s title. The date and time that the signature was applied The location where the signing occurred The user configurable meaning associated with the signature 11.50(b) Are these items part of any human readable form of the electronic record? Yes. Both the EZChrom machine-readable data and the human readable report contain the name, date and meaning of the EZChrom electronic signature. In addition, the optional esignature Plug-in for Adobe Acrobat places a visible signature manifestation on all human readable forms of the document, electronic display and printed form Signature/Record Linking Is the electronic signature linked to its respective electronic record to ensure that the signature cannot be excised, copied or otherwise transferred to falsify an electronic record by ordinary means? Yes. Signed records have a unique checksum that prevents signatures from being excised, copied or otherwise transferred. OpenLAB ECM will not recognize a signature that was applied outside its own electronic signature plug-ins. The EZChrom electronic signature is stored in the data file and is unbreakably linked to the EZChrom data even outside of OpenLAB ECM. The optional esignature Plug-in for Adobe Acrobat encrypts the signature within the document to prevent the signature from being excised or copied to another document Electronic Signatures: General Requirements (a) Is each electronic signature unique to one individual and not reused by, or reassigned to, anyone else? Yes. EZChrom as well as OpenLAB ECM uses the user ID / password combination unique to each user in the electronic signature feature. User names in EZChrom / OpenLAB ECM are required to be unique and cannot be reused or reassigned to another individual. Whether EZChrom / OpenLAB ECM uses the company s Windows NT logins to validate users or OpenLAB ECM administrated users, no two users can have the same user ID / password combination (b) (c) (c) (2) (c) (2) Are the identities of the individuals verified prior to the establishment, assignment, and certification of an individual s electronic signature or any element of an electronic signature? Has the company delivered its corporate electronic signature certification letter to the FDA? Is it in paper form with a traditional handwritten signature? Can additional certification or testimony be provided so that a specific electronic signature is the legally binding equivalent of the signer s handwritten signature?. This is the responsibility of the organization that plans, implements and operates the system. Such a verification process is a system requirement that is set before implementing electronic signature procedures or assigning electronic signature privileges to an individual.. It is the company s responsibility, before submitting electronically signed documentation to the FDA, to register their intent to use electronic signatures. In addition, training programs must be in place to ensure that users signing documents electronically understand the legal significance of their electronic signature.

5 Electronic Signature Components and Controls Section Question OpenLAB CDS EZChrom Edition/OpenLAB ECM Integration (i) Does the e-signature employ at least two distinct identification components such as user ID and password? Yes. The EZChrom as well as the OpenLAB ECM electronic signature tools require that the user enters two distinct identification components; unique user ID and password. (1)(i) (1)(i) (1)(ii) (2) (3) (b) When an individual executes a series of signings during a single, continuous period of controlled system access, is the first signing executed using all the electronic signature components? When an individual executes a series of signings during a single, continuous period of controlled system access, is each subsequent signing executed using at least one electronic signature component that is only executable by, and designed to be used by, the individual? When an individual executes a series of signings not performed during a single, continuous period of controlled system access, does each signing executed require all signature components? Are controls in place to ensure that only their genuine owners can use the electronic signature? Are the electronic signatures to be administered and executed to ensure that the attempted use of an individual s electronic signature by anyone other than its genuine owner requires the collaboration of two or more individuals? Are electronic signatures based on biometrics designed to ensure that only their genuine owners can use them? Yes. When an individual signs the first of a series of documents during a single period of controlled access the user is required to enter both signature components: user ID and password. Yes. When an EZChrom / OpenLAB ECM user executes a series of continuous electronic signatures, which are defined as signatures executed within a period of time determined by the system administrator, they are required to enter user ID, password and reason on the first signature only. Each subsequent signature requires only the user s password, which is known only to the user. Yes. When an EZChrom / OpenLAB ECM user executes a series of non-continuous electronic signatures, which are defined as signatures executed outside of a period of time determined by the system administrator, they are required to enter user ID, password and reason on each signature. Yes. OpenLAB ECM can be configured such that an administrator can assign an initial password to a user for a new account or forgotten password, but the user is required to change that password on their first login. In this way the user ID / password combination is known only to the individual. Whether OpenLAB ECM uses the company s Windows NT logins to validate users or OpenLAB ECM administrated users, no two users can have the same user ID / password combination. Yes. EZChrom / OpenLAB ECM uses the user s user ID and password to initiate the electronic signature. An EZChrom / OpenLAB ECM user s password is stored encrypted within the database and is displayed as asterisks in all locations within the software. OpenLAB ECM can be configured such that an administrator can assign an initial password to a user for a new account or forgotten password, but the user is required to change that password on their first login. In this way the user ID / password combination is known only to the individual.. Neither EZChrom nor OpenLAB ECM support signatures based on biometrics at this time Controls for Identification Codes/Passwords Section Question OpenLAB CDS EZChrom Edition/OpenLAB ECM Integration (a) (b) (c) Are controls in place to ensure the uniqueness of each combined identification code and password maintained, such that no two individuals have the same combination of identification code and password? Are controls in place to ensure that the identification code and password issuance is periodically checked, recalled, and revised? Are there loss management procedures in place to electronically disable lost, stolen, missing, or otherwise potentially compromised tokens, cards, and other devices that bear or generate identification code or password information? Yes. EZChrom / OpenLAB ECM requires users to authenticate with user ID and password. Whether EZChrom / OpenLAB ECM uses the company s Windows NT logins to validate users or OpenLAB ECM administrated users, no two users can have the same user ID / password combination. Yes. If using Windows domain authentication, password renewal interval is configured as part of the Windows password policy setup. The administrator can define a time frame in which passwords are periodically revised automatically. Users are prevented from reusing passwords. If users are administrated in OpenLAB ECM, OpenLAB ECM can be configured such that user passwords are automatically, periodically revised.. Neither OpenLAB ECM nor EZChrom support devices that bear or generate identification codes, such as tokens or cards, at this time.

6 (d) (d) (e) Are transaction safeguards in place to prevent unauthorized use of passwords and/or identification codes? Are transaction safeguards in place to detect and report in an immediate and urgent manner, any attempts at their unauthorized use to the system security unit and, as appropriate, to organizational management? Are there controls in place to initially test devices that bear or generate identification code or password information to ensure that they function properly and have not been altered in an unauthorized manner? Yes. EZChrom / OpenLAB ECM can be configured such that only the user knows their user ID and password. Passwords are always displayed as asterisks and are stored encrypted within the database so that even an administrator cannot see them. Yes. OpenLAB ECM can be configured such that a configurable number of unauthorized access attempts lock out the user account and send an notification to a system administrator. The Windows security policy can be configured so that a user defined number of unauthorized access attempts locks out the user account and sends notification to a system administrator. The system audit trail documents general events such as logon attempts to the computer as well as application or user changes, in the Windows Event log as a central audit repository for all security information. This includes the system and computer ID along with the operator name and application identification, allowing for an immediate check of the potential security leak.. Neither EZChrom nor OpenLAB ECM support devices that bear or generate identification codes, such as tokens or cards, at this time. This information is subject to change without notice. Agilent Technologies, Inc., 2012 Published in the USA, October 15, EN

Integration of Agilent UV-Visible ChemStation with OpenLAB ECM

Integration of Agilent UV-Visible ChemStation with OpenLAB ECM Integration of Agilent UV-Visible ChemStation with OpenLAB ECM Compliance with Introduction in Title 21 of the Code of Federal Regulations includes the US Federal guidelines for storing and protecting

More information

Agilent ICP-MS ChemStation Complying with 21 CFR Part 11. Application Note. Overview

Agilent ICP-MS ChemStation Complying with 21 CFR Part 11. Application Note. Overview Agilent ICP-MS ChemStation Complying with 21 CFR Part 11 Application Note Overview Part 11 in Title 21 of the Code of Federal Regulations includes the US Federal guidelines for storing and protecting electronic

More information

OpenLAB ELN Supporting 21 CFR Part 11 Compliance

OpenLAB ELN Supporting 21 CFR Part 11 Compliance OpenLAB ELN Supporting 21 CFR Part 11 Compliance White Paper Overview Part 11 in Title 21 of the Code of Federal Regulations includes the US Federal guidelines for storing and protecting electronic records

More information

ChromQuest 5.0. Tools to Aid in 21 CFR Part 11 Compliance. Introduction. General Overview. General Considerations

ChromQuest 5.0. Tools to Aid in 21 CFR Part 11 Compliance. Introduction. General Overview. General Considerations ChromQuest 5.0 Tools to Aid in 21 CFR Part 11 Compliance Introduction Thermo Scientific, Inc. is pleased to offer the ChromQuest chromatography data system (CDS) as a solution for chromatography labs seeking

More information

SDA COMPLIANCE SOFTWARE For Agilent ICP-MS MassHunter Software

SDA COMPLIANCE SOFTWARE For Agilent ICP-MS MassHunter Software SDA COMPLIANCE SOFTWARE For Agilent ICP-MS MassHunter Software Part 11 in Title 21 of the US Code of Federal Regulations (commonly referred to as 21 CFR Part 11) governs food and drugs in the US, and includes

More information

Compliance Matrix for 21 CFR Part 11: Electronic Records

Compliance Matrix for 21 CFR Part 11: Electronic Records Compliance Matrix for 21 CFR Part 11: Electronic Records Philip E. Plantz, PhD, Applications Manager David Kremer, Senior Software Engineer Application Note SL-AN-27 Revision B Provided By: Microtrac,

More information

ISSUE N 1 MAJOR MODIFICATIONS. Version Changes Related Release No. PREVIOUS VERSIONS HISTORY. Version Date History Related Release No.

ISSUE N 1 MAJOR MODIFICATIONS. Version Changes Related Release No. PREVIOUS VERSIONS HISTORY. Version Date History Related Release No. ISSUE N 1 MAJOR MODIFICATIONS Version Changes Related Release No. 01 First issue. 2.8.0 PREVIOUS VERSIONS HISTORY Version Date History Related Release No. N/A N/A N/A N/A APPROVAL TABLE Signatures below

More information

Exhibitor Software and 21 CFR Part 11

Exhibitor Software and 21 CFR Part 11 Exhibitor Software and 21 CFR Part 11 Subpart B Electronic Records 15 Columbia Drive Amherst, New Hampshire 03031-2334 No. 11.10 11.10(a) Controls for Closed Systems Validation of systems to ensure accuracy,

More information

Adobe Sign and 21 CFR Part 11

Adobe Sign and 21 CFR Part 11 Adobe Sign and 21 CFR Part 11 Today, organizations of all sizes are transforming manual paper-based processes into end-to-end digital experiences speeding signature processes by 500% with legal, trusted

More information

NucleoCounter NC-200, NucleoView NC-200 Software and Code of Federal Regulation 21 Part 11; Electronic Records, Electronic Signatures (21 CFR Part 11)

NucleoCounter NC-200, NucleoView NC-200 Software and Code of Federal Regulation 21 Part 11; Electronic Records, Electronic Signatures (21 CFR Part 11) NucleoCounter NC-200, NucleoView NC-200 Software and Code of Federal Regulation 21 Part 11; Electronic Records, Electronic Signatures (21 CFR Part 11) A ChemoMetec A/S White Paper March 2014 ChemoMetec

More information

COMPLIANCE. associates VALIDATOR WHITE PAPER. Addressing 21 cfr Part 11

COMPLIANCE. associates VALIDATOR WHITE PAPER. Addressing 21 cfr Part 11 VALIDATOR WHITE PAPER Addressing 21 cfr Part 11 Compliance Associates 1 1 INTRODUCTION 21 CFR Part 11 has been become a very large concern in the pharmaceutical industry as of late due to pressure from

More information

21 CFR Part 11 Module Design

21 CFR Part 11 Module Design 21 CFR Part 11 Module Design email: info@totallab.com web: www.totallab.com TotalLab Ltd Keel House Garth Heads Newcastle upon Tyne NE1 2JE UK Trademarks The following are either registered trademarks

More information

EZChrom Elite Chromatography Data System. Regulatory Compliance with FDA Rule of Electronic Records and Electronic Signatures (21 CFR Part 11)

EZChrom Elite Chromatography Data System. Regulatory Compliance with FDA Rule of Electronic Records and Electronic Signatures (21 CFR Part 11) EZChrom Elite Chromatography Data System Regulatory Compliance with FDA Rule of Electronic Records and Electronic Signatures (21 CFR Part 11) Scope On August 20, 1997 the final rule of the United States

More information

21 CFR PART 11 COMPLIANCE

21 CFR PART 11 COMPLIANCE 21 CFR PART 11 COMPLIANCE PRODUCT OVERVIEW ADD-ONS & INDIVIDUAL SOLUTIONS PLA SUPPORT CONTRACT TRAINING CONSULTING 21 CFR PART 11 COMPLIANCE PLA 3.0 Software For Biostatistical Analysis PLA 3.0 21 CFR

More information

TECHNICAL BULLETIN [ 1 / 13 ]

TECHNICAL BULLETIN [ 1 / 13 ] TECHNICAL BULLETIN [ 1 / 13 ] [Title] Guidelines on Compliance with FDA 21 CFR Part 11 for the GOT2000 and GOT1000 Series [Date of Issue] November 2014 (Ver. C: November 2017) [Relevant Models] GOT2000

More information

FDA 21 CFR Part 11 Compliance by Metrohm Raman

FDA 21 CFR Part 11 Compliance by Metrohm Raman FDA 21 CFR Part 11 Compliance by Metrohm Raman Norms and Standards 21 CFR Part 11 is the FDA rule relating to the use of electronic records and electronic signatures. Recognizing the increasing pact of

More information

Compliance of Shimadzu Total Organic Carbon (TOC) Analyzer with FDA 21 CFR Part 11 Regulations on Electronic Records and Electronic Signatures

Compliance of Shimadzu Total Organic Carbon (TOC) Analyzer with FDA 21 CFR Part 11 Regulations on Electronic Records and Electronic Signatures NT1D-1275 Compliance of Shimadzu Total Organic Carbon (TOC) Analyzer with FDA 21 CFR Part 11 Regulations on Electronic Records and Electronic Signatures TOC-Control L Ver.1 / LabSolutions DB/CS Ver.6 Part

More information

REGULATION ASPECTS 21 CFR PART11. 57, av. Général de Croutte TOULOUSE (FRANCE) (0) Fax +33 (0)

REGULATION ASPECTS 21 CFR PART11. 57, av. Général de Croutte TOULOUSE (FRANCE) (0) Fax +33 (0) REGULATION ASPECTS 21 CFR PART11 57, av. Général de Croutte - 31100 TOULOUSE (FRANCE) - +33 (0)5 34 47 40 00 - Fax +33 (0)5 34 47 43 01 Trademarks All names identified by are registered trademarks of the

More information

WHITE PAPER AGILOFT COMPLIANCE WITH CFR 21 PART 11

WHITE PAPER AGILOFT COMPLIANCE WITH CFR 21 PART 11 WHITE PAPER AGILOFT COMPLIANCE WITH CFR 21 PART 11 with CFR 21 Part 11 Table of Contents with CFR 21 Part 11 3 Overview 3 Verifiable Support for End-User Requirements 3 Electronic Signature Support 3 Precise

More information

Using "TiNet 2.5 Compliant SR1" software to comply with 21 CFR Part 11

Using TiNet 2.5 Compliant SR1 software to comply with 21 CFR Part 11 2003-08-08/dö Using "TiNet 2.5 Compliant SR1" software to comply with 21 CFR Part 11 The Title 21 Code of Federal Regulations Electronic Records; Electronic Signatures of the U.S. Food and Drug Administration,

More information

System Assessment Report Relating to Electronic Records and Electronic Signatures; 21 CFR Part 11. System: tiamo (Software Version 2.

System Assessment Report Relating to Electronic Records and Electronic Signatures; 21 CFR Part 11. System: tiamo (Software Version 2. Page 1 /15 System Assessment Report Relating to Electronic Records and Electronic Signatures; 21 CFR Part 11 System: tiamo (Software Version 2.5) Page 2 /15 1 Procedures and Controls for Closed Systems

More information

Sparta Systems TrackWise Solution

Sparta Systems TrackWise Solution Systems Solution 21 CFR Part 11 and Annex 11 Assessment October 2017 Systems Solution Introduction The purpose of this document is to outline the roles and responsibilities for compliance with the FDA

More information

Agilent Response to 21CFR Part11 requirements for the Agilent ChemStation Plus

Agilent Response to 21CFR Part11 requirements for the Agilent ChemStation Plus Agilent Response to 21CFR Part11 requirements for the Agilent ChemStation Plus 1. Preface This document describes which requirements of the FDA s rule for electronic records and electronic signature (21

More information

Sparta Systems Stratas Solution

Sparta Systems Stratas Solution Systems Solution 21 CFR Part 11 and Annex 11 Assessment October 2017 Systems Solution Introduction The purpose of this document is to outline the roles and responsibilities for compliance with the FDA

More information

The Impact of 21 CFR Part 11 on Product Development

The Impact of 21 CFR Part 11 on Product Development The Impact of 21 CFR Part 11 on Product Development Product development has become an increasingly critical factor in highly-regulated life sciences industries. Biotechnology, medical device, and pharmaceutical

More information

21 CFR Part 11 FAQ (Frequently Asked Questions)

21 CFR Part 11 FAQ (Frequently Asked Questions) 21 CFR Part 11 FAQ (Frequently Asked Questions) and Roles and Responsibilities for Assessment of METTLER TOLEDO STAR e Software Version 16.00, including: - 21 CFR 11 Compliance software option for Compliance

More information

Sparta Systems TrackWise Digital Solution

Sparta Systems TrackWise Digital Solution Systems TrackWise Digital Solution 21 CFR Part 11 and Annex 11 Assessment February 2018 Systems TrackWise Digital Solution Introduction The purpose of this document is to outline the roles and responsibilities

More information

System Assessment Report Relating to Electronic Records and Electronic Signatures; 21 CFR Part 11. System: StabNet (Software Version 1.

System Assessment Report Relating to Electronic Records and Electronic Signatures; 21 CFR Part 11. System: StabNet (Software Version 1. Page 1 /16 System Assessment Report Relating to Electronic Records and Electronic Signatures; 21 CFR Part 11 System: StabNet (Software Version 1.1) Page 2 /16 1 Procedures and Controls for Closed Systems

More information

Validation Checklist Appendix A WiZARD2 Secure and 21 CFR 11 Requirements

Validation Checklist Appendix A WiZARD2 Secure and 21 CFR 11 Requirements Appendix A Procedures and Controls for Closed Systems (check = yes) (check = yes) Customers may devise their own validation protocols that may or may not be compliant with 21 CFR 11 Is the system validated?

More information

Using "IC Net 2.2 " software to comply with 21 CFR Part 11

Using IC Net 2.2  software to comply with 21 CFR Part 11 CH-9101 Herisau/Switzerland E-Mail info@metrohm.com Internet www.metrohm.com Using "IC Net 2.2 " software to comply with 21 CFR Part 11 Compliance white paper 8.110.8273 CH-9101 Herisau/Switzerland E-Mail

More information

Metrohm White paper. FDA 21 CFR Part 11 Requirements for NIR Spectroscopy. Dr. N. Rühl

Metrohm White paper. FDA 21 CFR Part 11 Requirements for NIR Spectroscopy. Dr. N. Rühl FDA 21 CFR Part 11 Requirements for NIR Spectroscopy Dr. N. Rühl The prosperity of a society can be evaluated based on many criteria, and the focus is certainly different for each individual. However,

More information

INFORMATION. Guidance on the use of the SM1000 and SM2000 Videographic Recorders for Electronic Record Keeping in FDA Approved Processes

INFORMATION. Guidance on the use of the SM1000 and SM2000 Videographic Recorders for Electronic Record Keeping in FDA Approved Processes INFORMATION No. INF02/70 Issue 3 Date: October 2007 Product SM1000 and SM2000 Videographic Recorders Manuals IM/SM1000 and IM/SM2000 Guidance on the use of the SM1000 and SM2000 Videographic Recorders

More information

Meeting regulatory compliance guidelines with Agilent ICP-MS MassHunter and OpenLAB Server

Meeting regulatory compliance guidelines with Agilent ICP-MS MassHunter and OpenLAB Server Meeting regulatory compliance guidelines with Agilent ICP-MS MassHunter and OpenLAB Server White Paper Overview The United States Pharmacopoeia (USP) and the International Council for Harmonisation of

More information

System Assessment Report Relating to Electronic Records and Electronic Signatures; Final Rule, 21 CFR Part 11

System Assessment Report Relating to Electronic Records and Electronic Signatures; Final Rule, 21 CFR Part 11 Page 1 /16 System Assessment Report Relating to Electronic Records and Electronic Signatures; Final Rule, 21 CFR Part 11 System: Touch Control for Titrando (Software version 5.840.0150) Page 2 /16 1 Procedures

More information

Technical Information

Technical Information Technical Information TI 04L55B01-04EN SMARTDAC+ GM Advanced Security Functions White Paper for FDA 21 CFR Part 11 The contents of this Technical Information are subject to change without notice. Yokogawa

More information

White Paper Assessment of Veriteq viewlinc Environmental Monitoring System Compliance to 21 CFR Part 11Requirements

White Paper Assessment of Veriteq viewlinc Environmental Monitoring System Compliance to 21 CFR Part 11Requirements White Paper Assessment of Veriteq viewlinc Environmental Monitoring System Compliance to 21 CFR Part 11Requirements Introduction The 21 CFR Part 11 rule states that the FDA view is that the risks of falsification,

More information

21 CFR Part 11 LIMS Requirements Electronic signatures and records

21 CFR Part 11 LIMS Requirements Electronic signatures and records 21 CFR Part 11 LIMS Requirements Electronic signatures and records Compiled by Perry W. Burton Version 1.0, 16 August 2014 Table of contents 1. Purpose of this document... 1 1.1 Notes to version 1.0...

More information

Using the Titrando system to comply with 21 CFR Part 11

Using the Titrando system to comply with 21 CFR Part 11 06.2006/jb Using the Titrando system to comply with 21 CFR Part 11 The Title 21 Code of Federal Regulations Electronic Records; Electronic Signatures of the U.S. Food and Drug Administration, known as

More information

Assessment of Vaisala Veriteq viewlinc Continuous Monitoring System Compliance to 21 CFR Part 11 Requirements

Assessment of Vaisala Veriteq viewlinc Continuous Monitoring System Compliance to 21 CFR Part 11 Requirements / White PAPer Assessment of Vaisala Veriteq viewlinc Continuous Monitoring System Compliance to 21 CFR Part 11 Requirements The 21 CFR Part 11 rule states that the FDA view is that the risks of falsification,

More information

System Assessment Report Relating to Electronic Records and Electronic Signatures; Final Rule, 21 CFR Part 11. System: tiamo 2.3

System Assessment Report Relating to Electronic Records and Electronic Signatures; Final Rule, 21 CFR Part 11. System: tiamo 2.3 Page 1 /14 System Assessment Report Relating to Electronic Records and Electronic Signatures; Final le, 21 CFR Part 11 System: tiamo 23 052011 / doe Page 2 /14 1 Procedures and Controls for Closed Systems

More information

Using Chromeleon 7 Chromatography Data System to Comply with 21 CFR Part 11

Using Chromeleon 7 Chromatography Data System to Comply with 21 CFR Part 11 WHITE PAPER 80078 Using Chromeleon 7 Chromatography Data System to Comply with 21 CFR Part 11 Author Shaun Quinn, Marketing Manager Informatics and Chromatography Software, Thermo Fisher Scientific Keywords

More information

ComplianceQuest Support of Compliance to FDA 21 CFR Part 11Requirements WHITE PAPER. ComplianceQuest In-Depth Analysis and Review

ComplianceQuest Support of Compliance to FDA 21 CFR Part 11Requirements WHITE PAPER. ComplianceQuest In-Depth Analysis and Review ComplianceQuest Support of Compliance to FDA 21 CFR Part 11 WHITE PAPER ComplianceQuest In-Depth Analysis and Review ComplianceQuest Support of Compliance to FDA is the FDA guideline that defines the criteria

More information

Electronic Data Processing 21 CFR Part 11

Electronic Data Processing 21 CFR Part 11 Live Webinar on How Does Compliance with 21 CFR Part 11 Ensure Data Integrity & Subject Safety in Clinical Research Wednesday, 19 June 2013 at 10:00 AM PST / 01:00 PM EST ByCharles H. Pierce, MD, PhD,

More information

21 CFR 11 Assistant Software. 21 CFR Part 11 Compliance Booklet

21 CFR 11 Assistant Software. 21 CFR Part 11 Compliance Booklet 21 CFR 11 Assistant Software 21 CFR Part 11 Compliance Booklet Notices Agilent Technologies, Inc. 2001-2004, 2009-2010 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Using Chromeleon Chromatography Management Software to Comply with 21 CFR Part 11

Using Chromeleon Chromatography Management Software to Comply with 21 CFR Part 11 Technical Note 54 Using Chromeleon Chromatography Management Software to Comply with 21 CFR Part 11 The Electronic Records and Signatures Rule 1, known as 21 CFR Part 11, was established by the U.S. Food

More information

Part 11 Compliance SOP

Part 11 Compliance SOP 1.0 Commercial in Confidence 16-Aug-2006 1 of 14 Part 11 Compliance SOP Document No: SOP_0130 Prepared by: David Brown Date: 16-Aug-2006 Version: 1.0 1.0 Commercial in Confidence 16-Aug-2006 2 of 14 Document

More information

Guidelines for applying FactoryTalk View SE in a 21 CFR Part 11 environment

Guidelines for applying FactoryTalk View SE in a 21 CFR Part 11 environment FactoryTalk View Site Edition (SE) Complying with 21 CFR Part 11: Electronic Records & Signatures Guidelines for applying FactoryTalk View SE in a 21 CFR Part 11 environment Doc ID FTALK-WP003C-EN-E Page

More information

Electronic Records and Signatures with the Sievers M9 TOC Analyzer and DataPro2 Software

Electronic Records and Signatures with the Sievers M9 TOC Analyzer and DataPro2 Software Water Technologies & Solutions fact sheet 21 CFR Part 11 Electronic Records and Signatures with the Sievers M9 TOC Analyzer and DataPro2 Software introduction Part 11 of Title 21 of the Code of Federal

More information

Introduction 2. History. Adapted to zenon version 6.20 (MH) January 13 th, 2006

Introduction 2. History. Adapted to zenon version 6.20 (MH) January 13 th, 2006 FDA 21 CFR Part 11 Introduction 2 History Date January 13 th, 2006 Comment Adapted to zenon version 6.20 (MH) 1994 COPA-DATA GmbH All rights reserved. Distribution and/or reproduction of this document

More information

Using the Titrando system to comply with 21 CFR Part 11

Using the Titrando system to comply with 21 CFR Part 11 01.2003/sn Using the Titrando system to comply with 21 CFR Part 11 The Electronic Records and Signatures Rule, known as 21 CFR Part 11, was established by the U.S. Food and Drug Administration (FDA) to

More information

Cell Therapy Data Management

Cell Therapy Data Management Cell Therapy Data Management Darin Sumstad, CLS (NCA, ASCP) System Administrator, Cell Therapy Clinical Laboratory University of Minnesota Medical Center, Fairview Cell Therapy Data Management Introduction

More information

Agilent Technologies Dissolution Workstation Software Electronic Records and Data Storage Background

Agilent Technologies Dissolution Workstation Software Electronic Records and Data Storage Background Agilent Technologies Electronic Records and Data Storage Background Page 1 of 20 Table of Contents Introduction... 3 User Administration... 4 System Administration... 7 Method Management... 11 Method Execution...

More information

Spectroscopy Configuration Manager (SCM) Software. 21 CFR Part 11 Compliance Booklet

Spectroscopy Configuration Manager (SCM) Software. 21 CFR Part 11 Compliance Booklet Spectroscopy Configuration Manager (SCM) Software 21 CFR Part 11 Compliance Booklet Notices Agilent Technologies, Inc. 2006-2007 and 2009-2011 No part of this manual may be reproduced in any form or by

More information

Technical Note. Introduction

Technical Note. Introduction Integration of ChemStation with Agilent OL ECM Technical Note Introduction Agilent Enterprise Content Manager (ECM) is a web-based electronic library that provides a storage place for all electronic data.

More information

Real World Examples for Part 11 Technical Controls

Real World Examples for Part 11 Technical Controls Wolfgang Winter Product Manager, Networked Data Systems 23. January 2003 Real World Examples for Part 11 Technical Controls Time: 3.00 p.m. Central European Time Telephone Number: +44 20 8240 8243 Chair

More information

Integration of GC/MSD ChemStation with OpenLAB ECM

Integration of GC/MSD ChemStation with OpenLAB ECM Integration of GC/MSD ChemStation with OpenLAB ECM Technical Overview Introduction OpenLAB Enterprise Content Manager (ECM) is a Web-based electronic library that provides a secure storage place for all

More information

System Assessment Report Relating to Electronic Records and Electronic Signatures; Final Rule, 21 CFR Part 11. System: tiamo 2.0

System Assessment Report Relating to Electronic Records and Electronic Signatures; Final Rule, 21 CFR Part 11. System: tiamo 2.0 Page 1 /14 Sstem Assessment Report Relating to Electronic Records and Electronic Signatures; Final le, 21 CFR Part 11 Sstem: tiamo 2.0 Page 2 /14 1 Procedures and Controls for Closed Sstems 1.1 11.10 (a)

More information

Guidance for a 21 CFR Part 11 implementation on Microsoft Office SharePoint Server 2007

Guidance for a 21 CFR Part 11 implementation on Microsoft Office SharePoint Server 2007 Guidance for a 21 CFR Part 11 implementation on Microsoft Office SharePoint Server 2007 May 2007 Microsoft Health and Life Sciences Microsoft Consulting Services Table of Contents Introduction... 3 Table

More information

Statement of 21 CFR Part 11 Validation Results

Statement of 21 CFR Part 11 Validation Results Statement of Validation Results Software application: Session Manager DB (TurboTag ) Sealed Air Corporation Version: 2.1.1.5 Status: Validated through version 2.1.1.5 (Original validation completed on

More information

Introduction. So what is 21 CFR Part 11? Who Should Comply with 21CFR Part 11?

Introduction. So what is 21 CFR Part 11? Who Should Comply with 21CFR Part 11? Introduction The following guide is an explanation of the term 21 CFR Part 11, and gives some background into the tools/features that Comark includes in its 21 CFR Part 11 products to aid compliance with

More information

21 CFR PART 11 FREQUENTLY ASKED QUESTIONS (FAQS)

21 CFR PART 11 FREQUENTLY ASKED QUESTIONS (FAQS) 21 CFR PART 11 FREQUENTLY ASKED QUESTIONS (S) The United States Food and Drug Administration (FDA) defines the criteria under which electronic records and electronic signatures are considered trustworthy,

More information

MicroLab FTIR Software 21 CFR Part 11 Compliance

MicroLab FTIR Software 21 CFR Part 11 Compliance MicroLab FTIR Software 21 CFR Part 11 Compliance Technical Overview Introduction Electronic data submitted to the United States FDA must comply with specifications set forth in the Code of Federal Regulations,

More information

OM-MICROLITE-8 AND OM- MICROLITE-16 DATA LOGGERS AND OM-MICROLAB

OM-MICROLITE-8 AND OM- MICROLITE-16 DATA LOGGERS AND OM-MICROLAB User s Guide Shop online at omega.com e-mail: info@omega.com For latest product manuals: omegamanual.info OM-MICROLITE-8 AND OM- MICROLITE-16 DATA LOGGERS AND OM-MICROLAB And DatPass Administration Software

More information

Automation Change Management for Regulated Industries

Automation Change Management for Regulated Industries Automation Change Management for Regulated Industries Achieving Part 11 Compliance A White Paper Synopsis This whitepaper provides information related to FDA regulation 21 CFR Part 11 (Part 11) for organizations

More information

Achieving 21 CFR Part11 Compliance using Exaquantum/Batch Authored by Stelex

Achieving 21 CFR Part11 Compliance using Exaquantum/Batch Authored by Stelex Technical Information TI 36J04B11-01E Achieving 21 CFR Part11 Compliance using Exaquantum/Batch Authored by Stelex Yokogawa Electric Corporation 2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan

More information

Agilent ChemStation OpenLAB Option

Agilent ChemStation OpenLAB Option Agilent ChemStation OpenLAB Option Concepts Guide ChemStation OpenLAB Option Concepts Guide Agilent Technologies Notices Agilent Technologies, Inc. 2008-2009, 2010 No part of this manual may be reproduced

More information

EU Annex 11 Compliance Regulatory Conformity of eve

EU Annex 11 Compliance Regulatory Conformity of eve White Paper EU Annex 11 Compliance Regulatory Conformity of eve Franco Berz, Head of Quality Management INFORS HT Dr. Britta Abellan, Computer System Validation Manager INFORS HT 1. Introduction More and

More information

MySign Electronic Signature

MySign Electronic Signature MySign Electronic Signature Advisory Circular Compliance Matrix FAA AC 120 78A Dated 06/22/16 1 Table of Contents Table of Contents 2 Purpose 3 FAA Acceptance 3 Non Requirement for Approval 3 2-2 Electronic

More information

eprost System Policies & Procedures

eprost System Policies & Procedures eprost System Policies & Procedures Initial Approval Date: 12/07/2010 Revision Date: 02/25/2011 Introduction eprost [ Electronic Protocol Submission and Tracking ] is the Human Subject Research Office's

More information

SECURITY & PRIVACY DOCUMENTATION

SECURITY & PRIVACY DOCUMENTATION Okta s Commitment to Security & Privacy SECURITY & PRIVACY DOCUMENTATION (last updated September 15, 2017) Okta is committed to achieving and preserving the trust of our customers, by providing a comprehensive

More information

testo Comfort Software CFR 4 Instruction manual

testo Comfort Software CFR 4 Instruction manual testo Comfort Software CFR 4 Instruction manual 2 1 Contents 1 Contents 1 Contents... 3 2 Specifications... 4 2.1. Intended purpose... 4 2.2. 21 CFR Part 11 and terminology used... 5 3 First steps... 9

More information

DFARS Requirements for Defense Contractors Must Be Satisfied by DECEMBER 31, 2017

DFARS Requirements for Defense Contractors Must Be Satisfied by DECEMBER 31, 2017 DFARS 252.204-7012 Requirements for Defense Contractors Must Be Satisfied by DECEMBER 31, 2017 As with most government documents, one often leads to another. And that s the case with DFARS 252.204-7012.

More information

Achieving 21 CFR Part 11 Compliance using CENTUM VP

Achieving 21 CFR Part 11 Compliance using CENTUM VP Achieving 21 CFR Part 11 Compliance using CENTUM VP Authored by: Achieving 21 CFR 11 Compliance Table Of Contents 1 ABSTRACT...3 2 21 CFR PART 11 GENERAL REQUIREMENTS...3 2.1 WHY ELECTRONIC RECORDS & ELECTRONIC

More information

Good Laboratory Practice GUIDELINES FOR THE ARCHIVING OF ELECTRONIC RAW DATA IN A GLP ENVIRONMENT. Release Date:

Good Laboratory Practice GUIDELINES FOR THE ARCHIVING OF ELECTRONIC RAW DATA IN A GLP ENVIRONMENT. Release Date: AGIT: Swiss Working Group on Information Technology in a GLP Environment Good Laboratory Practice GUIDELINES FOR THE ARCHIVING OF ELECTRONIC RAW DATA IN A GLP ENVIRONMENT Release Date: 31.01.2018 Version:

More information

Secure Workstation for OpenLAB CDS ChemStation Edition User's Guide

Secure Workstation for OpenLAB CDS ChemStation Edition User's Guide Secure Workstation for OpenLAB CDS ChemStation Edition User's Guide Secure Workstation for OpenLAB CDS ChemStation Edition User's Guide Agilent Technologies Notices Agilent Technologies, Inc. 2014-2017

More information

Criminal Justice Information Security (CJIS) Guide for ShareBase in the Hyland Cloud

Criminal Justice Information Security (CJIS) Guide for ShareBase in the Hyland Cloud Criminal Justice Information Security (CJIS) Guide for ShareBase in the Hyland Cloud Introduction The Criminal Justice Information Security (CJIS) Policy is a publically accessible document that contains

More information

Information Security Policy

Information Security Policy April 2016 Table of Contents PURPOSE AND SCOPE 5 I. CONFIDENTIAL INFORMATION 5 II. SCOPE 6 ORGANIZATION OF INFORMATION SECURITY 6 I. RESPONSIBILITY FOR INFORMATION SECURITY 6 II. COMMUNICATIONS REGARDING

More information

Premium HMI and FDA 21 Part 11 regulations TN0009. Version Description Date 1 First emission 10/04/2012

Premium HMI and FDA 21 Part 11 regulations TN0009. Version Description Date 1 First emission 10/04/2012 Premium HMI and FDA 21 Part 11 regulations Introduction This document contains a brief explanation of the FDA CFR21 Part 11 regulations. It describes then the procedures and actions to take, in order to

More information

Chapter 9 Section 3. Digital Imaging (Scanned) And Electronic (Born-Digital) Records Process And Formats

Chapter 9 Section 3. Digital Imaging (Scanned) And Electronic (Born-Digital) Records Process And Formats Records Management (RM) Chapter 9 Section 3 Digital Imaging (Scanned) And Electronic (Born-Digital) Records Process And Formats Revision: 1.0 GENERAL 1.1 The success of a digitized document conversion

More information

Achieving 21 CFR Part 11 Compliance using CENTUM VP

Achieving 21 CFR Part 11 Compliance using CENTUM VP Achieving 21 CFR Part 11 Compliance using CENTUM VP Authored in cooperation with: Achieving 21 CFR 11 Compliance Table of Contents 1 ABSTRACT... 3 2 21 CFR PART 11 GENERAL REQUIREMENTS... 3 2.1 WHY ELECTRONIC

More information

State of Colorado Cyber Security Policies

State of Colorado Cyber Security Policies TITLE: State of Colorado Cyber Security Policies Access Control Policy Overview This policy document is part of the State of Colorado Cyber Security Policies, created to support the State of Colorado Chief

More information

Comply with Data Integrity Regulations with Chromeleon CDS Software

Comply with Data Integrity Regulations with Chromeleon CDS Software Comply with Data Integrity Regulations with Chromeleon CDS Software Anna Severoni Sales Support Specialist for Chromatography Thermo Fisher Scientific, Rodano (MI) The world leader in serving science Introduction

More information

21 CFR Part 11 Compliance DELMIA Apriso 2018 Implementation Guide

21 CFR Part 11 Compliance DELMIA Apriso 2018 Implementation Guide 21 CFR Part 11 Compliance DELMIA Apriso 2018 Implementation Guide 2017 Dassault Systèmes. Apriso, 3DEXPERIENCE, the Compass logo and the 3DS logo, CATIA, SOLIDWORKS, ENOVIA, DELMIA, SIMULIA, GEOVIA, EXALEAD,

More information

Impact Analysis for Software changes in OpenLAB CDS A.01.04

Impact Analysis for Software changes in OpenLAB CDS A.01.04 Impact Analysis for Software changes in OpenLAB CDS A.01.04 Document Information: Filename Product Identifier Product Revision Project Identifier Document Revision Impact_Analysis_OpenLAB_CDS_A0104 OpenLAB

More information

How to get your Movicon 11 project FDA 21 CFR Part 11 ready. Document: RAC-4105 Released: Updated: Rel. Movicon: 11.

How to get your Movicon 11 project FDA 21 CFR Part 11 ready. Document: RAC-4105 Released: Updated: Rel. Movicon: 11. How to get your Movicon 11 project FDA 21 CFR Part 11 ready Document: RAC-4105 Released: 29-04-2005 Updated: 22-08-2008 Rel. Movicon: 11.0 or later Summary 1 Document s Aim 3 1.1 Introduction 3 2 General

More information

Access to University Data Policy

Access to University Data Policy UNIVERSITY OF OKLAHOMA Health Sciences Center Information Technology Security Policy Access to University Data Policy 1. Purpose This policy defines roles and responsibilities for protecting OUHSC s non-public

More information

NIST Compliance Controls

NIST Compliance Controls NIST 800-53 Compliance s The following control families represent a portion of special publication NIST 800-53 revision 4. This guide is intended to aid McAfee, its partners, and its customers, in aligning

More information

21 CFR Part 11 Fundamentals

21 CFR Part 11 Fundamentals 21 CFR Part 11 Fundamentals By Praxis Life Sciences 1925 West Field Court, Suite 125, Lake Forest, IL 60045 praxislifesciences.com +1(847) 295-7160 validationcenter.com 21 CFR Part 11 Fundamentals Copyright

More information

e-authentication guidelines for esign- Online Electronic Signature Service

e-authentication guidelines for esign- Online Electronic Signature Service e-authentication guidelines for esign- Online Electronic Signature Service (Issued under Electronic Signature or Electronic Authentication Technique and Procedure Rules, 2015) Version 1.3 April 2017 Controller

More information

Red Flags/Identity Theft Prevention Policy: Purpose

Red Flags/Identity Theft Prevention Policy: Purpose Red Flags/Identity Theft Prevention Policy: 200.3 Purpose Employees and students depend on Morehouse College ( Morehouse ) to properly protect their personal non-public information, which is gathered and

More information

Password Standard Version 2.0 October 2006

Password Standard Version 2.0 October 2006 Password Standard Version 2.0 October 2006 TABLE OF CONTENTS 1.1 SCOPE 2 1.2 PRINCIPLES 2 1.3 REVISIONS 3 2.1 OBJECTIVE 4 3.1 POLICY 4 3.2 PROTECTION 4 3.3 LENGTH 4 3.4 SELECTIONS 4 3.5 EXPIRATION 5 3.6

More information

90% 191 Security Best Practices. Blades. 52 Regulatory Requirements. Compliance Report PCI DSS 2.0. related to this regulation

90% 191 Security Best Practices. Blades. 52 Regulatory Requirements. Compliance Report PCI DSS 2.0. related to this regulation Compliance Report PCI DSS 2.0 Generated by Check Point Compliance Blade, on April 16, 2018 15:41 PM O verview 1 90% Compliance About PCI DSS 2.0 PCI-DSS is a legal obligation mandated not by government

More information

Rev.1 Solution Brief

Rev.1 Solution Brief FISMA-NIST SP 800-171 Rev.1 Solution Brief New York FISMA Cybersecurity NIST SP 800-171 EventTracker 8815 Centre Park Drive, Columbia MD 21045 About EventTracker EventTracker delivers business critical

More information

ORA HIPAA Security. All Affiliate Research Policy Subject: HIPAA Security File Under: For Researchers

ORA HIPAA Security. All Affiliate Research Policy Subject: HIPAA Security File Under: For Researchers All Affiliate Research Policy Subject: HIPAA File Under: For Researchers ORA HIPAA Issuing Department: Office of Research Administration Original Policy Date Page 1 of 5 Approved by: May 9,2005 Revision

More information

Leveraging ALCOA+ Principles to Establish a Data Lifecycle Approach for the Validation and Remediation of Data Integrity. Bradford Allen Genentech

Leveraging ALCOA+ Principles to Establish a Data Lifecycle Approach for the Validation and Remediation of Data Integrity. Bradford Allen Genentech Leveraging ALCOA+ Principles to Establish a Data Lifecycle Approach for the Validation and Remediation of Data Integrity Bradford Allen Genentech 1 Agenda Introduction Data Integrity 101 Review What is

More information

Agilent OpenLAB Chromatography Data System

Agilent OpenLAB Chromatography Data System Agilent OpenLAB Chromatography Data System EZChrom Edition EZChrom Elite and ICM Migration Guide Agilent Technologies Notices Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in

More information

Tennessee Technological University Policy No Password Management

Tennessee Technological University Policy No Password Management Tennessee Technological University Policy No. 852 Password Management Effective Date: January 1, 2014 Policy No: 852 Policy Name: Password Management Policy Policy Subject: Password Management Date Revised:

More information

Agilent OpenLAB Data Store Administration. Guide for Administrators

Agilent OpenLAB Data Store Administration. Guide for Administrators Agilent OpenLAB Data Store Administration Guide for Administrators Notices Agilent Technologies, Inc. 2013 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Management: A Guide For Harvard Administrators

Management: A Guide For Harvard Administrators E-mail Management: A Guide For Harvard Administrators E-mail is information transmitted or exchanged between a sender and a recipient by way of a system of connected computers. Although e-mail is considered

More information

Electronic Signature Policy

Electronic Signature Policy Electronic Signature Policy Definitions The following terms are used in this policy. Term Definition Electronic Signature An electronic signature is a paperless method used to authorize or approve documents

More information