Research Report. Radio Frequency Identification Devices (RFIDs) Effect of RFIDs on lifesaving equipment

Size: px
Start display at page:

Download "Research Report. Radio Frequency Identification Devices (RFIDs) Effect of RFIDs on lifesaving equipment"

Transcription

1 Research Report Radio Frequency Identification Devices (RFIDs) Effect of RFIDs on lifesaving equipment Background It has been suggested in the Journal of the Medical Association that lifesaving equipment may be switched off by RFIDs (radio-frequency identification devices) which are used to track people and machines. Research Question: Research Report Apart from that published in the Journal of the Medical Association, is there any further published evidence that suggests that RFIDs may interfere with lifesaving equipment? Summary RFID is a general term that is used to describe a system which transmits the identity (in the form of a unique serial number) of an object wirelessly, using radio waves; this is sometimes referred to as contact-less technology. A typical RFID system is made up of three components: tags, readers and the host computer system. In the original research, medical equipment situated within metres of an RFID reader broadcasting a signal to nearby RFID tag was tested to see if any Electro-Magnetic (EMI) was detected. Tests carried out in a non-clinical environment showed that caused by RFIDs induced potentially hazardous incidents in medical devices. However, the scenarios in which the tests were conducted can be described as scenarios in which the devices were used at their highest output and in close proximity to medical devices. A preliminary literature review was carried out to ascertain if any further evidence could be found which corroborates or negates the original research. To date, no other literature has been published which corroborates or negates the work. Conclusions and recommendations for the use of RFIDs in healthcare environments are put forward. Version 1 Page 1 of 12

2 Abstract RFID is a general term that is used to describe a system that transmits the identity (in the form of a unique serial number) of an object wirelessly, using radio waves; this is sometimes referred to as contact-less technology. A typical RFID system is made up of three components: tags, readers and the host computer system In the original research article, medical equipment situated within metres of an RFID reader broadcasting a signal to nearby RFID tag was tested to see if any Electro- Magnetic (EMI) was detected. Tests carried out by van der Togt, van Lieshout et al, in a controlled non-clinical testing environment at the Academic Medical Centre, University of Amsterdam, showed that caused by RFIDs induced potentially hazardous incidents in medical devices. The scenarios in which the tests were conducted, can be described as worse case scenarios in which the devices were used at their highest output in close proximity to medical devices. A preliminary literature review was carried out to ascertain if any further evidence could be found which corroborates or negates the work carried out by van der Togt, van Lieshout et al. To date, no other literature has been published which corroborates or negates the work carried out by the University of Amsterdam Research Group. Method Literature review, and discussion with Medicines and Healthcare Product Regulatory Agency (MHRA). Findings from original research The University of Amsterdam researchers tested 41 pieces of medical equipment 3 times; each test was conducted within metres of an RFID reader broadcasting a signal to nearby RFID tag. Tests carried out showed that the equipment malfunctioned 34 times (28%). Malfunctions were deemed hazardous, significant or light, 22 of the malfunctions reported were noted as being hazardous, 2 were reported to be significant, and 10 were reported as being light. (Van der Togt et al., 2008) The 3 scales used in the study to describe the effect the EMI had on the critical care equipment are adapted from the critical care adverse event scale, the effects associated with each point on the scale are described in Table 1. In the Dutch study, types of hazardous malfunction reported included ventilators switching off or changing rates, syringe pumps stopping and external pacemakers malfunctioning. (Appendix 1 identifies the devices examined and details the malfunctions reported. Version 1 Page 2 of 12

3 Scale/Rating Light Significant Effect denoted Influence on monitoring without a significant level of attention needed, for example a disturbed display. Influence on monitoring with a significant level of attention needed, causing substantial distraction from patient care, for example an incorrect alarm or inaccurate monitoring of blood pressure. Direct physical influence on the patient by an Hazardous unintended change in equipment function, for example total stopping of ventilator or syringe pump. Table 1: Effects indicated by the critical care adverse effect scale Implications In areas containing sensitive medical devices, the potential risk suggested by the Dutch research team depends on the distance between the reader and any medical device. It should be noted that adverse occurred at separations of less than 1 metre, which is similar to that caused by mobile phones. However, the initial impression given by the research results was that the risk associated with RFID would be greater due to the fact that the median distance for all incidents established in the paper was 30cm and therefore more critical when compared with the median range of 3cm established with modern mobile phones. As mentioned previously, due to the fact that devices were tested in close proximity to critical care equipment at high power outputs, the scenarios described in the Dutch research paper are somewhat different to those commonly found within Healthcare Environments. To date no further evidence has been published which corroborates or negates the work carried out by van der Togt, van Lieshout et al. Therefore with the current level of information available it is difficult to ascertain what the likelihood of malfunction actually is. It should be noted that many pieces of communication technology, for example mobile phones, can interfere with medical equipment and therefore must be managed appropriately. Tags The Dutch study indicated that the main risk of does not come from tags, which are very low power devices. The low power associated with their use means that although RFID tags will be directly attached to medical devices, they are unlikely to interfere with their operation. Version 1 Page 3 of 12

4 Readers Readers tend to have higher powers, up to several watts. The Dutch research work has indicated that the main risk of EMI comes from readers; the risk of EMI is dependant on where the readers are used. The risk of is higher if the RFID is used close to medical devices which are being used for intensive care and therefore, may well be a problem. Consequently, it is not advisable to use readers in this situation, as is the case with mobile phones. In contrast there are many situations where tag readers could be used in complete safety. Risk Assessment Electro magnetic has only been detected at relatively close ranges, therefore those using RFIDs should assess the likelihood of RFID chips being placed within such a short range of hospital equipment. Until more information is gained about the use of RFIDs working out the risks of is not straightforward as there are many factors involved. These include: type/power rating of reader; type/power rating of tag; distance between reader and medical device; location of use: o store room = low/no risk; o operating theatre = higher risk. Although the tags are directly attached to medical devices, they are very low power devices and so are unlikely to cause. The main risk comes from the RFID readers that tend to be higher powered, up to several watts, putting them in the same risk bracket as mobile phones Reports of To date the MHRA (Medicines and Healthcare products Regulatory Agency) has not received any incident reports regarding RFID ; they are however proactive in monitoring the situation. The Federal Drugs Agency (FDA) has not received any incident reports of active implantable medical devices reacting to RFID tags or the readers. They too are monitoring the situation. Version 1 Page 4 of 12

5 Conclusions and recommendations Conclusion 1 Many pieces of communication technology can interfere with medical equipment and therefore must be managed appropriately and in some instances guidelines for their use within the healthcare environment are already established (e.g. Mobile phones). As readers have the similar power output and the distance at which is noted is similar to that of mobile phones, it can be said the risk associated with their use is likely to be in the same bracket as that of mobile phones. It may be surmised that the risk posed by and the way that the risk is managed should be similar to that of mobile phones. Therefore, the level of risk will depend on where they are used, for example; if RFIDs are used in storerooms and non-clinical areas the chances of causing adverse incidents are remote; whilst if they are used in operating theatres and intensive care wards the possibility of exists; with portal-type readers potential risks will depend on whether patients connected to critical care devices have to pass through them. Recommendation 1 As has only been detected at relatively close ranges, those using RFIDs should assess the likelihood of RFID chips being placed within such a short range of hospital equipment. Additionally it may be prudent to screen for potential mishaps at wider ranges. Conclusion 2 The study's authors did not recommend removing or banning the RFID technology from hospitals, noting that it does have a strong potential to help healthcare providers. They stated medical equipment makers should now develop ways to protect their medical devices from radio frequency. Recommendation 2 Those purchasing devices should require that manufacturers should re-engineer devices to withstand radio waves. Additional testing and efforts to create standards for how these devices should be used, may take years to establish. Conclusion 3 t enough is known about extent and mode of use of RFIDs within the UK. Version 1 Page 5 of 12

6 Recommendation 3 Due to the fact that at present not enough is known about the extent and mode of use of RFIDs in UK healthcare it is difficult to fully assess the potential risks caused by such devices. Until such information is known it is reasonable to state that RFID devices should be treated like mobile phones and not used in close proximity to operating electronic medical devices. Furthermore, use of readers in operating theatres and critical care areas should be avoided until further evidence is available. Future Work MHRA has published General Safety Information and Advice for use of RFIDs within the healthcare environment. Details can be found at the following web address: icalinformation/radiofrequencyidentification/index.htm further work is to be carried out by Health Facilities Scotland at this time. A periodic searching of the literature will continue and future findings or changes to guidance will be reported back to the service. Version 1 Page 6 of 12

7 Appendix 1 An overview of devices examined and findings. From Electronic Supplementary Material to the article: Effects of Electromagnetic from Radio Frequency Identification on Critical Care Equipments Classification: H = hazardous, S = significant, L = light Type Of Device (Distance In Cm) Signal (Type) Infusion/Syringe Pumps Classification Graseby MHz H Alaris IVAC MHz H B.Braun Compact (S) B.Braun Infusomat (P) B.Braun Infusomat (P) MHz L MHz L 5 125kHz H Alaris IVAC MHz H B.Braun perfusor ft B.Braun perfusor fm B.Braun Infusomat Space MHz H 0,1 868MHz L no Details Stops working, error messages, resetting not possible. Stops working, error messages, after reset pump works normally again. Beep heard from loudspeaker. Rattling noise heard from loudspeaker. Stops operating, error message. Stops operating, error messages. Stops operating, acoustic alarm. Lamp 'battery' switches off. Version 1 Page 7 of 12

8 Type Of Device Medtronic 5348 Single chamber Medtronic 5348 Single chamber Biotronik EDP20/A Medtronic 5388 Dual chamber Medtronic 5388 Dual chamber (Distance In Cm) Signal (Type) External Pacemakers Classification MHz H kHz H MHz H MHz H 5 125kHz H Mechanical Ventilators Hamilton Galileo MHz H Hamilton Galileo kHz S Drager Evita MHz H Hamilton Raphael color and Siemens Servo 300A and screen 390 Details On demand: pace pulses are given erroneously. On demand: pace pulses are given erroneously. Pace pulses are given erroneously. On demand: pace pulses are given erroneously in both atrium and ventricle. On demand: pace pulses are given erroneously in both atrium and ventricle. Ventilation triggers erroneously, higher ventilation frequency at flow curve at display. Stops operating, loss of settings, automatic restart. Version 1 Page 8 of 12

9 Type Of Device Gambro AK200 ULTRA S B.Braun Diapact CRRT Guidant Zoom Latitude programming system 3120 Guidant Zoom Latitude programming system 3120 Medtronic 2090 The Programmer Medtronic 2090 The Programmer Datascope System 97 (Distance In Cm) Signal (Type) Hemofiltration/Dialysis Devices Classification MHz H MHz H Pacemaker Programmers MHz S kHz H MHz H kHz H Intra-Aortic Balloon Pumps MHz H Details Stops operating (recoverable from operation panel), acoustic alarm. Can be continued. At 0 cm: stops working, sometimes treatment stops, reset Stops operating (recoverable from operation panel), acoustic alarm. in atrial and ventricular electrocardiogra m (EGM) Active tag, solely: in atrial and ventricular electrocardiogra m (EGM) Communication with ICD disturbed, LEDbar on programming head disrupted. Communication with ICD disturbed, LEDbar on programming head disrupted. Stops operating, settings lost. Version 1 Page 9 of 12

10 Type Of Device Arrow Acat 1 plus & Arrow AutoCat2Wave. Surgical company Fluido (Distance In Cm) Signal (Type) Fluid Warmer Classification MHz H Cardiopulmonary Bypass Device Sarns MHz H Haemonetics HaemoLite2Plus Agilent/HP CM- S Monitor with ECG module M1001A/B Agilent/HP CM- S Monitor with ECG module M1001A/B GE S/5 Avance, GE Aestiva/5, Philips Airway Gases M1026B Philips Heartstart Medtronic lifepak 20 Medtronic lifepak 12 Autologous blood recovery device 5 868MHz H Anaesthesia Devices MHz L kHz L Defibrillators MHz L 5 868MHz L Details Resets automatically and resumes operating normally when signal is removed. cardioplegy pump into band skid mode, no change in rotational speed. Change in rotational speed in ECG. in ECG. noise from loudspeaker. noise from loudspeaker. Version 1 Page 10 of 12

11 Type Of Device Marquette MAC leads ECG Marquette MAC leads ECG Philips Intellivue MP30 with module M3001A PhilipsIntellivue MP50 & Philips Intellivue MP90 Hill-Rom Avant Guard & Hill- Rom Total Care Maquet Alphamaquet Sarns/Terumo TCM II Laerdal LSU (Distance In Cm) Signal (Type) 12 leads electrocardiogram device Classification MHz L kHz L Monitors MHz L IC Bed Operating Table Hyper/Hypothermia Device Vacuum Pump Details in ECG. in ECG. ise from loudspeaker. References: Van der Togt R, van Lieshout EJ, Hensbroek R, et al. (2008) Electromagnetic from Radio Frequency Identification Inducing Potentially Hazardous Incidents in Critical Care Medical Equipment. Journal of the American Medical Association. Vol pp Electronic Supplementary Material to the article: Effects Of Electromagnetic from Radio Frequency Identification on Critical Care Equipment. Accessed 23/04/09. MHRA Radiofrequency Identification Version 1 Page 11 of 12

12 icalinformation/radiofrequencyidentification/index.htm Accessed 26/05/09 Version 1 Page 12 of 12

URGENT - Medical Device Correction IntelliVue MX40 Missing Warnings in IFU

URGENT - Medical Device Correction IntelliVue MX40 Missing Warnings in IFU Dear Customer, A problem has been found with the Philips IntelliVue MX40 Instructions for Use (IFU) for software revisions B.05, B.06 and B.06.5X. Your IntelliVue MX40 remains safe to use. These IFUs are

More information

Use Of Mobile Communication Devices Within Healthcare Premises Policy

Use Of Mobile Communication Devices Within Healthcare Premises Policy Use Of Mobile Communication Devices Within Healthcare Premises Policy Co-ordinator: Director of Facilities Reviewer: Working Group chaired by Director of Facilities Approver: GAPF Signature Signature Signature

More information

Global Service. Product Catalog Cardiac Assist

Global Service. Product Catalog Cardiac Assist Global Service Product Catalog Cardiac Assist Global Service Product Catalog Cardiac Assist 1 Intra-Aortic Balloon Pump MCare Service Plans In addition to the standard warranty provided with your IABP,

More information

Media centre Electromagnetic fields and public health: mobile phones

Media centre Electromagnetic fields and public health: mobile phones Media centre Electromagnetic fields and public health: mobile phones Fact sheet N 193 Reviewed October 2014 Key facts Mobile phone use is ubiquitous with an estimated 6.9 billion subscriptions globally.

More information

IntelliBridge EC10 and EC5

IntelliBridge EC10 and EC5 IntelliBridge EC10 and EC5 Philips 865115 & 865114 External Devices Configuration Guide A.01.12 Published in Germany 05/2012 IntelliBridge EC10/EC5 components Each IntelliBridge consists of three components:

More information

Medical Device Usability

Medical Device Usability Medical Device Usability David Adams Global Head, Active Medical Devices Add logo on slide 4 here Topics What is usability? Why usability is so important The regulatory requirements EN 62366 Usability

More information

Policies, Procedures, Guidelines and Protocols. John Snell - Head of Workforce Planning, Systems and Contributors

Policies, Procedures, Guidelines and Protocols. John Snell - Head of Workforce Planning, Systems and Contributors Policies, Procedures, Guidelines and Protocols Document Details Title Staff Mobile Phone Policy Trust Ref No 2036-39774 Local Ref (optional) N/A Main points the document Procurement, allocation and use

More information

URGENT Field Safety Notice Philips HeartStart MRx Monitor/Defibrillator

URGENT Field Safety Notice Philips HeartStart MRx Monitor/Defibrillator Philips Healthcare - 1/5 - FSN86100160A Dear Customer, Philips has identified issues that could impact the safety and/or performance of certain MRx monitor/defibrillators. These issues are further detailed

More information

This document contains important information for the continued safe and proper use of your equipment

This document contains important information for the continued safe and proper use of your equipment Philips Healthcare - 1/6 - FSN86100179A Dear Customer, Philips has identified the possibility of abnormal device behaviors on the HeartStart MRx Monitor/Defibrillator. Under exceptionally rare circumstances,

More information

IntelliBridge EC10 & EC5

IntelliBridge EC10 & EC5 IntelliBridge EC10 & EC5 Philips 865115 & 865114 External Devices Configuration Guide A.01.01 Printed in Germany 05/2011 IntelliBridge EC10/EC5 components Each IntelliBridge consists of three components.

More information

Technical Article. EMC f ilters for medical devices

Technical Article. EMC f ilters for medical devices Technical Article EMC f ilters for medical devices July 2014 EMC filters for medical devices Schaffner, the market leader in EMC filters and power quality devices, has consistently extended its proven

More information

Always in touch. IntelliVue Telemetry System with Smart-hopping technology, surveillance of ambulatory cardiac patients

Always in touch. IntelliVue Telemetry System with Smart-hopping technology, surveillance of ambulatory cardiac patients Always in touch with Smart-hopping technology, surveillance of ambulatory cardiac patients Real clinical network strength includes flexibility When you have mobile cardiac patients who need constant monitoring,

More information

Regulatory Aspects of Digital Healthcare Solutions

Regulatory Aspects of Digital Healthcare Solutions Regulatory Aspects of Digital Healthcare Solutions TÜV SÜD Product Service GmbH Dr. Markus Siebert Rev. 02 / 2017 02.05.2017 TÜV SÜD Product Service GmbH Slide 1 Contents Digital solutions as Medical Device

More information

Wireless Device Use & Patient Monitoring Equipment in Any Healthcare Delivery Setting: A Review of Safety & Guidelines

Wireless Device Use & Patient Monitoring Equipment in Any Healthcare Delivery Setting: A Review of Safety & Guidelines Wireless Device Use & Patient Monitoring Equipment in Any Healthcare Delivery Setting: A Review of Safety & Guidelines Mollie Butler, RN, PhD Regional Director Professional Practice Nursing Eastern Health

More information

TIP CARD MODEL TEMPORARY EXTERNAL PACEMAKER. Model Single Chamber Temporary Pacemaker

TIP CARD MODEL TEMPORARY EXTERNAL PACEMAKER. Model Single Chamber Temporary Pacemaker TIP CARD MODEL 53401 TEMPORARY EXTERNAL PACEMAKER Model 53401 Single Chamber Temporary Pacemaker PRE-USE What to do... What NOT to do... Battery Replace the battery for each new patient. Do NOT reuse battery.

More information

Securing Biomedical Devices. IT Challenges - A View from the Trenches

Securing Biomedical Devices. IT Challenges - A View from the Trenches Securing Biomedical Devices IT Challenges - A View from the Trenches Background Lead newly formed medical device security (MDS) team Previously clinical/research/teaching activities Extensively collaborated

More information

Medical Devices and Cyber Issues JANUARY 23, American Hospital Association and BDO USA, LLP. All rights reserved.

Medical Devices and Cyber Issues JANUARY 23, American Hospital Association and BDO USA, LLP. All rights reserved. Medical Devices and Cyber Issues JANUARY 23, 2018 AHA and Cybersecurity Policy Approaches Role of the FDA FDA Guidance and Roles Pre-market Post-market Assistance during attack Recent AHA Recommendations

More information

TITLE: Wireless Device Use and Patient Monitoring Equipment in Any Healthcare Delivery Setting: A Review of Safety and Guidelines

TITLE: Wireless Device Use and Patient Monitoring Equipment in Any Healthcare Delivery Setting: A Review of Safety and Guidelines TITLE: Wireless Device Use and Patient Monitoring Equipment in Any Healthcare Delivery Setting: A Review of Safety and Guidelines DATE: 18 August, 2016 CONTEXT AND POLICY ISSUES Wireless communication

More information

Tender no /TNMSC/ENGG/2015, DT: Results of Evaluation. Sl. No. Name of the bidders 1. M/s. Akas Medical, Chennai.

Tender no /TNMSC/ENGG/2015, DT: Results of Evaluation. Sl. No. Name of the bidders 1. M/s. Akas Medical, Chennai. Tender for Supply and Installation of Pulse Oximeter, Defibrillator, Volumetric Infusion Pump, Syringe Infusion Pump and Computerized ECG Machine to Dept. of Anesthesiology, Govt. Rajaji Hospital, Madurai.

More information

Mobile Phone Policy General Use

Mobile Phone Policy General Use Mobile Phone Policy General Use Author(s) Daniel Clifford, IT and Systems Project Manager Version 2.1 Version Date July 2013 Implementation/approval Date July 2013 Review Date July 2016 Review Body Policy

More information

ETSI IN SUPPORT OF EHEALTH

ETSI IN SUPPORT OF EHEALTH ETSI IN SUPPORT OF EHEALTH Presented by Dr Saad Mezzour Medtronic ETSI TG30 Chairman ETSI 2014. All rights reserved ETSI is Independent, not for profit organization, created in 1988 A recognized EuropeanStandards

More information

TITLE: Wireless Device Use and Patient Monitoring Equipment in Any Healthcare Delivery Setting: A Review of Safety and Guidelines

TITLE: Wireless Device Use and Patient Monitoring Equipment in Any Healthcare Delivery Setting: A Review of Safety and Guidelines TITLE: Wireless Device Use and Patient Monitoring Equipment in Any Healthcare Delivery Setting: A Review of Safety and Guidelines DATE: 17 January 2014 CONTEXT AND POLICY ISSUES Wireless technologies are

More information

DESIGN AND REALISATION OF A FRAMEWORK FOR DEVICE ENDCOMMUNICATION ACCORDING TO THE IEEE STANDARD

DESIGN AND REALISATION OF A FRAMEWORK FOR DEVICE ENDCOMMUNICATION ACCORDING TO THE IEEE STANDARD DESIGN AND REALISATION OF A FRAMEWORK FOR DEVICE ENDCOMMUNICATION ACCORDING TO THE IEEE 11073-20601 STANDARD Frohner M 1, Urbauer P 1, Bauer M 1, Gerbovics F 1, Mense A 2, Sauermann S 1 Abstract The standardization

More information

!"# $ # # $ $ % $ &% $ '"# $ ()&*&)+(( )+(( )

!# $ # # $ $ % $ &% $ '# $ ()&*&)+(( )+(( ) !"# # # % &% '"# ) !#, ' "# " "# -. / # 0 0 0 0 0 "0 "# " # 1 #! " " 0 0 0 0 0 0 2# 0 # # 3 ' 4 56 7-56 87 9# 5 6 7 6 & 0 " : 9 ; 4 " #! 0 - '% # % "# " "# " < 4 "! % " % 4 % % 9# 4 56 87 = 4 > 0 " %!#

More information

User Guide. CG666 Page 2. Open the back cover and remove the battery. Insert the SIM card correctly in the card slot as shown below.

User Guide. CG666 Page 2. Open the back cover and remove the battery. Insert the SIM card correctly in the card slot as shown below. GET TO KNOW YOUR PHONE... 2 INSERT SIM AND MEMORY CARD... 2 CHARGE THE BATTERY VIA POWER SOCKET OR USB... 4 HELPFUL KEYS AND SHORTCUTS...5 SWITCH PHONE ON/OFF... 6 LOCK/UNLOCK THE PHONE... 6 MAKE A CALL...

More information

Enabling Breakthroughs in Medical Electronics. Karthik Vasanth, Ph.D General Manager, Medical and High Reliability Business Unit

Enabling Breakthroughs in Medical Electronics. Karthik Vasanth, Ph.D General Manager, Medical and High Reliability Business Unit Enabling Breakthroughs in Medical Electronics Karthik Vasanth, Ph.D General Manager, Medical and High Reliability Business Unit Outline Trends driving the medical market Semiconductors in healthcare Diverse

More information

Copyright 2018 by Boston Scientific, Inc.. Permission granted to INCOSE to publish and use. #hwgsec

Copyright 2018 by Boston Scientific, Inc.. Permission granted to INCOSE to publish and use. #hwgsec Balancing Safety, Security and Usability in the Design of Secure Medical Devices Ken Hoyme Director, Product Security Boston Scientific Ken.hoyme@bsci.com Copyright 2018 by Boston Scientific, Inc.. Permission

More information

17 th Edition (BS 7671:2008) incorporating Amendment No 1

17 th Edition (BS 7671:2008) incorporating Amendment No 1 20 17th Edition 17 th Edition (BS 7671:2008) incorporating Amendment No 1 Installations designed from the 1 st January 2012 are to comply with Amendment number 1. Geoff Cronshaw The first amendment to

More information

Table of contents Page

Table of contents Page User Guide Prog-Box Thank you for purchasing a Hacker Brushless Motors product. The MASTER Prog-Box was developed in close cooperation with JETI and incorporates the latest technological developments.

More information

EO-BG935. User Manual. 1 /2016. Rev.1.

EO-BG935. User Manual. 1 /2016. Rev.1. EO-BG935 User Manual 1 /2016. Rev.1. www.samsung.com Table of Contents Français Deutsch Italiano Español Magyar Polski Română Български Hrvatski Srpski Português Latviešu Lietuvių kalba Eesti Nederlands

More information

Product codes with prefix (all variants): 8001, 8002, 8003, 8004, and ATTENTION: Clinical Personnel, Risk Managers, Biomedical Personnel

Product codes with prefix (all variants): 8001, 8002, 8003, 8004, and ATTENTION: Clinical Personnel, Risk Managers, Biomedical Personnel CareFusion Switzerland 317 Sàrl A-One Business Centre Zone d activités, Vers-la-Pièce No 10 Rolle, CH-1180 Switzerland carefusion.com Name Address URGENT FIELD SAFETY NOTICE Product Name: Alaris / Asena

More information

Data exchange between remote monitoring databases and local electronic patient record system Implementation based on international standards

Data exchange between remote monitoring databases and local electronic patient record system Implementation based on international standards Data exchange between remote monitoring databases and local electronic patient record system Implementation based on international standards Enno T. Van der Velde, FESC Medical Physicist, dept of Cardiology

More information

CPRD Aurum Frequently asked questions (FAQs)

CPRD Aurum Frequently asked questions (FAQs) CPRD Aurum Frequently asked questions (FAQs) Version 1.0 Date: 6 December 2017 Author: Helen Booth and Daniel Dedman, CPRD, UK Documentation Control Sheet During the course of the project it may be necessary

More information

Certification for Meaningful Use Experiences and Observations from the Field June 2011

Certification for Meaningful Use Experiences and Observations from the Field June 2011 Certification for Meaningful Use Experiences and Observations from the Field June 2011 Principles for Certification to Support Meaningful Use Certification should promote EHR adoption by giving providers

More information

State of US Telemedicine Industry

State of US Telemedicine Industry State of US Telemedicine Industry Prepared for ETSI Sophia Antipolis, France 6 May 2014 Agenda Current State of Healthcare in the US Industry dynamics Adoption of telemedicine/mhealth/digital health Healthcare

More information

UDI Implementation Update. GS1 UK Healthcare Conference - 22 November 2017 John Wilkinson OBE Medicines and Healthcare Products Regulatory Agency

UDI Implementation Update. GS1 UK Healthcare Conference - 22 November 2017 John Wilkinson OBE Medicines and Healthcare Products Regulatory Agency UDI Implementation Update GS1 UK Healthcare Conference - 22 November 2017 John Wilkinson OBE Medicines and Healthcare Products Regulatory Agency 2 Why new European medical device and IVD regulations? Old

More information

#

# INSTALLATION MANUAL Contents Getting To Know Your AVANT HIT+ TM... 3 Setting up the System... 4 Software Installation... 5 Driver Installation Windows 7... 8 Driver Installation Windows XP... 11 Accessories...

More information

EMS Guide M C S. January 2016/17 ECHANICAL IRCULATORY UPPORT O RGANIZATION

EMS Guide M C S. January 2016/17 ECHANICAL IRCULATORY UPPORT O RGANIZATION EMS Guide January 2016/17 M C S ECHANICAL IRCULATORY UPPORT O RGANIZATION This guide is produce by MCSO The Mechanical Circulatory Support Organization It is produced by VAD Coordinators from some of the

More information

Quick Start Guide CJB1950ALAAB

Quick Start Guide CJB1950ALAAB Quick Start Guide 1 CJB1950ALAAB www.sar-tick.com This product meets applicable national SAR limits of 2.0W/kg. The specific maximum SAR values can be found in the section of this user guide. When carrying

More information

Safety issues in medical devices

Safety issues in medical devices Safety issues in medical devices Prof. Cinzia Bernardeschi Dipartimento di Ingegnera dell Informazione Università di Pisa cinzia.bernardeschi@unipi.it 1 Medical devices Interactive medical systems used

More information

Indirect Effects: Interference With Medical Devices

Indirect Effects: Interference With Medical Devices Latin American Experts Committee on High Frequency Electromagnetic Fields and Human Health Scientific Review on Non-Ionizing Electromagnetic Radiation in the Radiofrequency Spectrum and its Effects on

More information

Context Monitoring Of A Patient Using Wireless Networks

Context Monitoring Of A Patient Using Wireless Networks Context Monitoring Of A Patient Using Wireless Networks N.Keerthana ECE Department Excel college of engineering and technology Abstract The paper describes a network that has been intended for context

More information

Context. Findings. Objectives. Canadian Hospital Policies on Wireless Communication Device Use

Context. Findings. Objectives. Canadian Hospital Policies on Wireless Communication Device Use Canadian Hospital Policies on Wireless Communication Device Use Issue 31 November 2011 Context Reliance on the use of cellular telephones and similar wireless communication devices has become pervasive

More information

CARELINK NETWORK UPDATES SUMMARY OF ENHANCEMENTS May 2017

CARELINK NETWORK UPDATES SUMMARY OF ENHANCEMENTS May 2017 CARELINK NETWORK UPDATES SUMMARY OF ENHANCEMENTS May 2017 Quick Look Update for Reveal LINQ ICM Patient Monitoring CareLink Network Experience With feedback from over 200 clinicians worldwide, Medtronic

More information

Medical Devices Cybersecurity? Introduction to the Cybersecurity Landscape in Healthcare

Medical Devices Cybersecurity? Introduction to the Cybersecurity Landscape in Healthcare May 5 & 6, 2017 Medical Devices Cybersecurity? Introduction to the Cybersecurity Landscape in Healthcare Marc Schlessinger, RRT, MBA, FACHE Senior Associate Applied Solutions Group Evolution of the Connected

More information

How to choose the right power supply for your design starting with accurate DC/DC converter measurements

How to choose the right power supply for your design starting with accurate DC/DC converter measurements Press Release Date By 10/8/18 Florian Haas, Director Marketing TRACO POWER GROUP How to choose the right power supply for your design starting with accurate DC/DC converter measurements One of the topics

More information

ABSTRACT 1. INTRODUCTION 2. SYSTEM ARCHITECTURE. Ching-Sung Wang. JBiSE

ABSTRACT 1. INTRODUCTION 2. SYSTEM ARCHITECTURE. Ching-Sung Wang. JBiSE J. Biomedical Science and Engineering, 2010, 3, 47-51 doi:10.4236/jbise.2010.31007 Published Online January 2010 (http://www.scirp.org/journal/jbise/). Mobile and wireless technologies applying on sphygmomanometer

More information

HAZARD NOTICE By arrangement with the Scottish Executive Health Department

HAZARD NOTICE By arrangement with the Scottish Executive Health Department HAZARD NOTICE By arrangement with the Scottish Executive Health Department CRITIKON DINAMAP PRO 100-400 & PRO 1000 BLOOD PRESSURE AND VITAL SIGNS MONITORS: RISK OF ELECTRIC SHOCK HAZ(SC)02/04 Page 1 of

More information

Important Medical Device Advisory

Important Medical Device Advisory Important Medical Device Advisory Battery Performance Alert and Cybersecurity Firmware Updates for Certain ICD & CRT-D Devices April 16, 2018 Dear Doctor, As part of a planned series of system updates

More information

Utilizing a Test Lab For Wireless Medical Devices

Utilizing a Test Lab For Wireless Medical Devices Utilizing a Test Lab For Wireless Medical Devices Allie Paquette Changes in healthcare suggest that more and more medical devices will begin to utilize the information technology (IT) network infrastructure

More information

A COMPUTER-CONTROLLED SYRINGE DRIVER FOR USE DURING ANAESTHESIA

A COMPUTER-CONTROLLED SYRINGE DRIVER FOR USE DURING ANAESTHESIA Br.J. Anaesth. (1986), 58, 670-675 A COMPUTER-CONTROLLED SYRINGE DRIVER FOR USE DURING ANAESTHESIA A Modification of the Graseby MS16A A. T. COHEN The continuous infusion of drugs during the course of

More information

IIC Testbed Overview. Stan Schneider, PhD CEO, Real-Time Innovations (RTI) Vice Chair, Industrial Internet Consortium (IIC)

IIC Testbed Overview. Stan Schneider, PhD CEO, Real-Time Innovations (RTI) Vice Chair, Industrial Internet Consortium (IIC) IIC Testbed Overview Stan Schneider, PhD CEO, Real-Time Innovations (RTI) Vice Chair, Industrial Internet Consortium (IIC) What is the Industrial IoT Really About? 2018 Real-Time Innovations, Inc. Permission

More information

Study ON Remote Medical Monitoring System Based ON MSP430 AND CC2530

Study ON Remote Medical Monitoring System Based ON MSP430 AND CC2530 2nd International Conference on Electrical, Computer Engineering and Electronics (ICECEE 2015) Study ON Remote Medical Monitoring System Based ON MSP430 AND CC2530 SONG Aijuan, SI Guangyuan, GU Qiongchan

More information

pat hways Medtech innovation briefing Published: 26 October 2016 nice.org.uk/guidance/mib85

pat hways Medtech innovation briefing Published: 26 October 2016 nice.org.uk/guidance/mib85 pat hways Needle-free arterial non-injectable connector Medtech innovation briefing Published: 26 October 2016 nice.org.uk/guidance/mib85 Summary The technology described in this briefing is the needle-free

More information

Phil Raymond Wireless Architect, Philips Healthcare Steven D Baker, PhD Sr. Principal Engineer, Welch Allyn

Phil Raymond Wireless Architect, Philips Healthcare Steven D Baker, PhD Sr. Principal Engineer, Welch Allyn Phil Raymond Wireless Architect, Philips Healthcare Steven D Baker, PhD Sr. Principal Engineer, Welch Allyn Gap between medical device mfr (MDM) testing and hospital IT testing Using risk analysis in designing

More information

Electromagnetic Radiation from Mobile Phones and their base stations: Health Effects

Electromagnetic Radiation from Mobile Phones and their base stations: Health Effects Regional Development Forum 2008 Bridging the Standardization Gap in Developing Countries Electromagnetic Radiation from Mobile Phones and their base stations: Health Effects Joseph K. Amoako Radiation

More information

Below we ve highlighted several of the key points from the final guidance document.

Below we ve highlighted several of the key points from the final guidance document. On September 6, 2017, the Food and Drug Administration (FDA or the Agency) released its final guidance document, Design Considerations and Pre-Market Submission Recommendations for Interoperable Medical

More information

Powering On/Off 4. Appendix 5

Powering On/Off 4. Appendix 5 PUCK USER MANUAL TABLE OF CONTENTS What s included? 1 Powering On/Off 4 Components 2 Bluetooth Connection 4 Puck Wired Earphones Charging 3 Puck Wired Earphones Appendix 5 Solutions for Troubleshooting

More information

AliveCor s Kardia Mobile ECG

AliveCor s Kardia Mobile ECG AliveCor s Kardia Mobile ECG General Practice user guide Dr John Marszal; Dr Ruth Chambers 21 July 2016 1 Contents Introduction 3 Information governance 3 Warnings and contraindications to AliveCor mobile

More information

Continuity of Operations During Disasters: Electronic Systems and Medical Records

Continuity of Operations During Disasters: Electronic Systems and Medical Records Idaho Health Care Association Continuity of Operations During Disasters: Electronic Systems and Medical Records Philip Niemer, MBA, MS, HEM Director Operational Continuity & Emergency Management Children

More information

The Next Frontier in Medical Device Security

The Next Frontier in Medical Device Security The Next Frontier in Medical Device Security Session #76, February 21, 2017 Denise Anderson, President, NH-ISAC Dr. Dale Nordenberg, Executive Director, MDISS 1 Speaker Introduction Denise Anderson, MBA

More information

Dialock DT Lite. Operating instructions...2. Operating instructions

Dialock DT Lite. Operating instructions...2. Operating instructions Operating instructions Dialock DT Lite Operating instructions... Operating instructions. Notes on these operating instructions. Contents and target group of the instructions It is essential to pay attention

More information

Frequently Asked Questions. My life. My healthcare. MyChart.

Frequently Asked Questions. My life. My healthcare. MyChart. Frequently Asked Questions My life. My healthcare. MyChart. My life. My healthcare. MyChart. What is MyChart? MyChart offers patients personalized and secure online access to portions of their medical

More information

Philip E. Plantz. Application Note. SL-AN-08 Revision C. Provided By: Microtrac, Inc. Particle Size Measuring Instrumentation

Philip E. Plantz. Application Note. SL-AN-08 Revision C. Provided By: Microtrac, Inc. Particle Size Measuring Instrumentation A Conceptual, Non-Mathematical Explanation on the Use of Refractive Index in Laser Particle Size Measurement (Understanding the concept of refractive index and Mie Scattering in Microtrac Instruments and

More information

Mapping Your Success 2013 BSI Healthcare Roadshow: Mobile Health Software, Mobile Phones and Telemedicine Under the EU Approach

Mapping Your Success 2013 BSI Healthcare Roadshow: Mobile Health Software, Mobile Phones and Telemedicine Under the EU Approach Mapping Your Success 2013 BSI Healthcare Roadshow: Mobile Health Software, Mobile Phones and Telemedicine Under the EU Approach Telemedicine & mhealth Definition of Telemedicine: Use of telecommunication

More information

Medical Device Cybersecurity: FDA Perspective

Medical Device Cybersecurity: FDA Perspective Medical Device Cybersecurity: FDA Perspective Suzanne B. Schwartz MD, MBA Associate Director for Science and Strategic Partnerships Office of the Center Director (OCD) Center for Devices and Radiological

More information

SMARTSCAN INFORMATION T4 SERIES LIGHT CURTAINS HANDBOOK

SMARTSCAN INFORMATION T4 SERIES LIGHT CURTAINS HANDBOOK SMARTSCAN INFORMATION T4 SERIES LIGHT CURTAINS HANDBOOK Smartscan Ltd, Pywell Road, CORBY, NN17 5XJ, UK, Tel: +44 (0) 1536 401313, Fax: +44 (0) 1536 268954, Email: sales@smartscan.com, www.smartscan.com

More information

Fall Test 21 due 12/11/2012

Fall Test 21 due 12/11/2012 Fall Test 21 due 12/11/2012 Multiple Choice Identify the choice that best completes the statement or answers the question. 1. Emergency system transfer equipment, including automatic transfer switches,

More information

Philip E. Plantz. Application Note. SL-AN-08 Revision B. Provided By: Microtrac, Inc. Particle Size Measuring Instrumentation

Philip E. Plantz. Application Note. SL-AN-08 Revision B. Provided By: Microtrac, Inc. Particle Size Measuring Instrumentation A Conceptual, Non Mathematical Explanation on the Use of Refractive Index in Laser Particle Size Measurement: (Understanding the concept of refractive index and Mie Scattering in Microtrac Instruments

More information

Table of Content. Sales, Service, Parts, and Accessories. Defibrillator... 3 Philips DVT Therapy... 4 Covidien... 4

Table of Content. Sales, Service, Parts, and Accessories. Defibrillator... 3 Philips DVT Therapy... 4 Covidien... 4 Table of Content Defibrillator... 3 Philips... 3 DVT Therapy... 4 Covidien... 4 ECG/EKG... 5 Burdick Cardiac Science... 5 GE Marquette... 6 Mortara Instrument... 6 Enteral Feeding Pump... 8 Abbott... 8

More information

Proximity Card and Pin Reader Installation Manual

Proximity Card and Pin Reader Installation Manual Multi Prox Proximity Card and Pin Reader Installation Manual PUBLICATION INFORMATION 60A9 - Draft Release Version 0.1.2 71D0 - Version 1.0.5 CONTENTS Introduction... 1 Legend... 2 Terminology... 2 Mounting...

More information

2015 HFMA What Healthcare Can Learn from the Banking Industry

2015 HFMA What Healthcare Can Learn from the Banking Industry 2015 HFMA What Healthcare Can Learn from the Banking Industry Agenda Introduction- Background and Experience Healthcare vs. Banking The Results OCR Audit Results Healthcare vs. Banking The Theories Practical

More information

Hacking Medical Devices

Hacking Medical Devices Hacking Medical Devices All your vital signs are belong to us #2 Blog: Conference: Florian Grunow Security Analyst ERNW GmbH in Heidelberg Team Lead: Pentests Research: Medical Devices #3 Agenda Motivation

More information

Connected Medical Devices

Connected Medical Devices Connected Medical Devices How to Reduce Risks Inherent in an Internet of Things that Can Help or Harm Laura Clark Fey, Esq., Principal, Fey LLC Agenda Overview of the Internet of Things for Healthcare

More information

Design Considerations and Premarket. Recommendations for Interoperable Medical Devices Guidance for Industry and Food and Drug Administration Staff

Design Considerations and Premarket. Recommendations for Interoperable Medical Devices Guidance for Industry and Food and Drug Administration Staff Design Considerations and Premarket Submission Recommendations for Interoperable Medical Devices Guidance for Industry and Food and Drug Administration Staff Document issued on: September 6, 2017 The draft

More information

RS TechMedic BV the Netherlands T: vision.com

RS TechMedic BV the Netherlands T: vision.com Dyna Vision Remote Patient Monitoring: Demonstration of Real Time 12 lead ECG Acquisition and Waveform Analysis using Integrated Cellular Transmission Telemonitoring Session Med e Tel 2010 Definitions

More information

Design and Implementation of Remote Medical Monitoring System for. Homecare

Design and Implementation of Remote Medical Monitoring System for. Homecare 2nd International Conference on Electronics, Network and Computer Engineering (ICENCE 2016) Design and Implementation of Remote Medical Monitoring System for Homecare Juan Zhong1, a, * Hua Liao2, b 1 College

More information

MedDRA Update. MedDRA Industry User Group Meeting. 28 September 2018

MedDRA Update. MedDRA Industry User Group Meeting. 28 September 2018 MedDRA Update MedDRA Industry User Group Meeting 28 September 2018 Topics MedDRA Users Profile MedDRA Translations MSSO Email Distribution List Opt-In Retrieving MedDRA Unzip Passwords Device terms in

More information

Wireless Wearable Body Area Network (WWBAN) for Elderly People Long-Term Health-Monitoring

Wireless Wearable Body Area Network (WWBAN) for Elderly People Long-Term Health-Monitoring Wireless Wearable Body Area Network (WWBAN) for Elderly People Long-Term Health-Monitoring Institution: Arab Academy for Science, Technology & Maritime Transport (AASTMT), Abu Kir Campus, P.O. Box 1029,

More information

NEW REQUIREMENTS FOR MEDICAL EMC

NEW REQUIREMENTS FOR MEDICAL EMC IEC 60601 1 2 EDITION 4 NEW REQUIREMENTS FOR MEDICAL EMC Products & Retail White Paper intertek.com/medical CONTENTS Introduction 3 4th Edition vs. 3rd Edition 4 Environments of Intended Use 5 Risk Management

More information

SAMPLE POLICY. Current State Assessment Criteria. 1. That EPHI that is transmitted electronically is not vulnerable to interception; and

SAMPLE POLICY. Current State Assessment Criteria. 1. That EPHI that is transmitted electronically is not vulnerable to interception; and Documentation of HIPAA Security Implementation Standards The HIPAA Privacy regulations required the adoption of formal policies and procedures. For the HIPAA Security Standards, the documentation is even

More information

UNIT 4: DATA COLLECTION AND TRANSMISSION

UNIT 4: DATA COLLECTION AND TRANSMISSION UNIT 4: DATA COLLECTION AND TRANSMISSION This page intentionally left blank. UNIT 4: DATA COLLECTION AND TRANSMISSION Visual 1 OBJECTIVES Visual 2 At the end of this unit participants will be able to:

More information

PERFORMANCE EVALUATION OF LOW RATE WPANS FOR MEDICAL APPLICATIONS

PERFORMANCE EVALUATION OF LOW RATE WPANS FOR MEDICAL APPLICATIONS PERFORMANCE EVALUATION OF LOW RATE WPANS FOR MEDICAL APPLICATIONS Nada Golmie, David Cypher, and Olivier Rebala National Institute of Standards and Technology Gaithersburg, Maryland ABSTRACT In this article

More information

Bluetooth mobile solutions APPLICATION NOTE / FAQ. Page 1 on 24

Bluetooth mobile solutions APPLICATION NOTE / FAQ. Page 1 on 24 Bluetooth mobile solutions APPLICATION NOTE / FAQ Page 1 on 24 Table of Contents I. Introduction... 5 II. Bluetooth Smart technology General principles... 5 III. Frequently Asked Questions... 5 A. STid

More information

When a Clinical Trial Lasts Forever: A Reporting Solution for Post-Market Surveillance of Cardiac Rhythm Devices

When a Clinical Trial Lasts Forever: A Reporting Solution for Post-Market Surveillance of Cardiac Rhythm Devices PharmaSUG 2018 - Paper MD-07 When a Clinical Trial Lasts Forever: A Reporting Solution for Post-Market Surveillance of Cardiac Rhythm Devices ABSTRACT Becky DeBus, Medtronic Post-market surveillance data

More information

RAD system. C- arm. Price 337,500 SDG. Mobile C-arm: Model OPESCOPE Activo. 2 monitors. Easy to use. Quick c-arm positioning. Simple operation.

RAD system. C- arm. Price 337,500 SDG. Mobile C-arm: Model OPESCOPE Activo. 2 monitors. Easy to use. Quick c-arm positioning. Simple operation. RAD system SHIMADZU is one of the world's Japanese leading manufacturers of advanced imaging systems and equipment for use in medical diagnosis and treatment. NASIF Medical imported and distributed all

More information

Running head: Personal Medical Devices 1. Personal Medical Devices. Ann Funk. East Carolina University

Running head: Personal Medical Devices 1. Personal Medical Devices. Ann Funk. East Carolina University Running head: Personal Medical Devices 1 Personal Medical Devices Ann Funk East Carolina University Personal Medical Devices 2 ABSTRACT Technology in any field has been growing at a rapid pace, but one

More information

DOD Medical Device Cybersecurity Considerations

DOD Medical Device Cybersecurity Considerations Enedina Guerrero, Acting Chief, Incident Mgmt. Section, Cyber Security Ops Branch 2015 Defense Health Information Technology Symposium DOD Medical Device Cybersecurity Considerations 1 DHA Vision A joint,

More information

FHD Driving Recorder E272S/S272W. Quick Start Guide

FHD Driving Recorder E272S/S272W. Quick Start Guide FHD Driving Recorder E272S/S272W Quick Start Guide 400-8401-030 www.polaroidcarcam.com 1 Introduction 1.1 Package Contents The package contains the following items. In case there is any missing or damaged

More information

Cybersecurity and Hospitals: A Board Perspective

Cybersecurity and Hospitals: A Board Perspective Cybersecurity and Hospitals: A Board Perspective Cybersecurity is an important issue for both the public and private sector. At a time when so many of our activities depend on information systems and technology,

More information

i 4 imm Compact Critical Care Monitor GE Healthcare Modular monitoring for critical care

i 4 imm Compact Critical Care Monitor GE Healthcare Modular monitoring for critical care GE Healthcare i 4 Modular monitoring for critical care imm Compact Critical Care Monitor imm imm Compact Critical Care Monitor modularity and versatility iconnect i 4 icentral ipon imm the innovative,

More information

NHS GREATER GLASGOW AND CLYDE

NHS GREATER GLASGOW AND CLYDE NHS GREATER GLASGOW AND CLYDE Policy on safe use of mobile communication devices, two-way radios, wireless routers, wireless computers within Healthcare premises. LEAD MANAGER: K. Fleming Head of Health

More information

TransTalk 9000 Digital Wireless System MDW 9040 Wireless Pocket Phone Quick Reference

TransTalk 9000 Digital Wireless System MDW 9040 Wireless Pocket Phone Quick Reference TransTalk 9000 Digital Wireless System MDW 9040 Wireless Pocket Phone Quick Reference 503-801-191 Comcode 700060437 Issue 2, Feb. 2001 Mute Volume Control Display Softkeys/Column Buttons Feature/ Program

More information

Life Sciences Applications: Modeling and Simulation for Biomedical Device Design SGC 2013

Life Sciences Applications: Modeling and Simulation for Biomedical Device Design SGC 2013 Life Sciences Applications: Modeling and Simulation for Biomedical Device Design Kristian.Debus@cd-adapco.com SGC 2013 Modeling and Simulation for Biomedical Device Design Biomedical device design and

More information

UK COSMOS PARTICIPANT NEWSLETTER FEBRUARY 2014

UK COSMOS PARTICIPANT NEWSLETTER FEBRUARY 2014 UK COSMOS PARTICIPANT NEWSLETTER FEBRUARY 14 Welcome Another year has passed and we are pleased to be back in touch with our participants to share the exciting progress of UK COSMOS to date! COSMOS remains

More information

LEGISLATIVE SITUATION IN MALAYSIA

LEGISLATIVE SITUATION IN MALAYSIA ATTACHMENT 13 LEGISLATIVE SITUATION IN MALAYSIA Pirunthavany Muthuvelu Ministry of Health Malaysia (MOH) Engineering Services Division BACKGROUND There were public concerns over the health effects of base

More information

Condition Report. Sale /10/ :00:00 - Medical Equipment Auction - PART :00. Sale Date

Condition Report. Sale /10/ :00:00 - Medical Equipment Auction - PART :00. Sale Date Condition Report Sale Sale Date Location: Viewing: Removal: 1820-09/10/2018 10:00:00 - Medical Equipment Auction - PART 2 09-10-2018 10:00 Malmesbury, Wiltshire, UK Friday 5th October 12pm - 5pm, Monday

More information

Creating Solutions for Health through Technology Innovation

Creating Solutions for Health through Technology Innovation Creating Solutions for Health through Technology Innovation Karthik Vasanth General Manager, Medical and High Reliability Business Unit Texas Instruments, Inc Outline Healthcare trends and opportunities

More information

EMF Risk Communication Five Cardinal Rules

EMF Risk Communication Five Cardinal Rules Mitglied der Helmholtz-Gemeinschaft EMF Risk Communication Five Cardinal Rules Peter Wiedemann 13. Februar 2009 Overview Introductionary remarks 5 cardinal rules Summary Societal Worries There is a public

More information

GROUP 17 COLLEGE OF ENGINEERING & COMPUTER SCIENCE. Senior Design I Professor: Dr. Samuel Richie UNIVERSITY OF CENTRAL FLORIDA

GROUP 17 COLLEGE OF ENGINEERING & COMPUTER SCIENCE. Senior Design I Professor: Dr. Samuel Richie UNIVERSITY OF CENTRAL FLORIDA GROUP 17 COLLEGE OF ENGINEERING & COMPUTER SCIENCE Senior Design I Professor: Dr. Samuel Richie UNIVERSITY OF CENTRAL FLORIDA Anh Loan Nguyen John E. Van Sickle Jordan Acedera Christopher Spalding December

More information