NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation

Size: px
Start display at page:

Download "NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation"

Transcription

1 NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation The GVSU chemistry department owns 2 NMR spectrometers. A JEOL Eclipse 300 MHz and a Varian Inova 400 MHz. Due to their strong, constant magnetic fields NMR instruments have specific safety concerns. Please read the short safety section below. NMR Safety NMRs use strong magnetic fields. If you depend on a medical device that is sensitive to magnetic fields Do Not Enter the NMR Room. Do not approach within 5 feet of the magnet if you have iron or steel objects. Do not bring cell phones and magnetic media such as credit cards near the magnet; leave them in the lab or near the instrument computer. Sample Preparation is identical for either instrument but varies depending upon which nuclei you are investigating. NMR active isotopes are naturally present in different abundancies. The 100% abundance for 1 H nuclei allows for more dilute samples than 13 C, naturally present at 1.1% of carbon atoms. Preparing a HNMR sample For most HNMR samples you will want a 0.1M solution of your sample dissolved in a deuterated solvent often CDCl 3 (Chloroform D). This concentration corresponds to between 5 mg and 25 mg or 1-5 drops of sample in 0.7ml of deuterated solvent. Filter the solution if necessary to remove solids and transfer the solution to a 5mm NMR tube. The liquid level should be about 3 fingers or 50mm high in the NMR tube. Solvent height is important for obtaining a good spectrum. Preparing a CNMR sample Due to the lower abundance of C-13 isotope your solution should be more concentrated than a typical HNMR sample. For most purposes you will want a 0.5M solution (25 mg or 5-10 drops) of the sample of interest dissolved in 0.7ml of deuterated solvent - CDCl 3 - and transferred to a 5mm NMR tube. 13C is much less sensitive than 1H. Either make the sample concentrated or plan on more scans. Common steps used to obtain spectra are outlined below. More specific directions for use of each instrument follow. Steps for Obtaining NMR Spectra 1. Prepare a sample 2. Login to NMR computer and connect to spectrometer operating system 3. Insert your sample and lock on sample 4. Run the experiment 5. Process your spectrum 6. Eject your sample and reinsert instrument standard 7. Lock on the standard 8. Clean your NMR tube

2 INSTRUCTIONS FOR OBTAINING A JEOL 1H NMR SPECTRUM A. Sample preparation: Before you use the NMR 1. Prepare a sample as directed on page The NMR tube is lowered into the machine using a tube holder called a spinner. A graduated depth gauge will help you determine how deep to place the tube in the spinner. Put an empty spinner in the depth gauge. Push the center portion of the spinner down all the way; this is crucial to machine operation. Now, gently insert the NMR tube into the spinner, pushing the tube to the bottom of the depth gauge. Holding the spinner by its wider upper part remove spinner and tube from the gauge and clean with a Kimwipe. Be careful not to change the depth of the tube in the spinner as you clean. Recheck the tube depth before putting it in the auto sampler. The autosampler is an 8 tube arrangement at the top of the NMR. Place the spinner and tube in one of the empty slots in the side of auto sampler nearest the computer. Never place the spinner into the slot that is over the NMR probe (back of auto sampler)! It will just fall down on top of the sample currently there! Figure 1. JEOL Spinner, NMR tube and Depth Gauge Depth Gauge Spinner B. Connecting the NMR computer and your sample The JEOL NMR uses a program called Delta to run and process samples. Delta is UNIX based software. To make entries in Delta the computer cursor needs to be in the field where you are adding information. NMR tube 1. Computer Login and connection to NMR software Click on the appropriate login icon: See your research advisor/class instructor If you do not see an array of login icons go to switch user in the Start menu. Double click on the Delta icon on the desktop. This opens the Delta console. In Delta, select File-> Preferences. Under the Directory/Folder tab check that your data will be stored in the desired folder. Within the Delta console window click on the icon for the spectrometer control. This is the fourth icon in the second row and is circled in figure. A window for spectrometer control Figure 2. Delta window will open. You will likely see a line with the text "eclipse2 - FREE "

3 Within a brief moment the two computers will connect. In the yellow bar you should see Connect: eclipse2jeol.com". If it does not immediately connect: you may select the eclipse line (by clicking on it) and click on the yellow Connect button. Figure 3. Spectrometer Control 2. Locking the Sample. This procedure helps maintain a consistent magnetic field during the NMR experiment. In the Spectrometer control window Click on the button labeled Sample (lower left corner). The Sample window will open - see the figure 4 below. Under Sample State choose the auto sampler slot where your sample is by clicking on the arrows. Once you move your cursor out of this window the auto sampler will eject the current sample and lower your sample into the probe. Under Solvent select the deuterated solvent you used. Under Lock Control chose gradient shim and lock - the third icon to the right, 2nd row. Wait until the instrument completes the gradient shim and lock experiment before continuing. This will take some time. Figure 4. Sample Window Sample slot Solvent Lock Status

4 C. Acquiring a Proton NMR Spectrum Figure 5. Lower Spectrometer Control window 1. Setting up an Experiment Once you have completed locking on your sample you may close the Sample window and return to the Spectrometer control window, figure 5. Select the Experiment box at the bottom of the window (shown circled). You will Figure 6. Open Experiment window see the Open Experiment window. 2. In the Open Experiment window set the type of experiment. If you do not see a list of files under Filename, you will need to click on the global folder: In the global folder scroll down under Filename to the file: single_pulse.exp Select this file and click on ok. The Experiment Tools window opens: 3. Within the Experiment Tools you will need to set some parameters, though most are set adequately by default. You should be in the Header tab. Put the Figure 7. Experiment Tools window cursor in the filename field and enter a filename. Your spectral data is stored in a folder based on your username and this filename. A good filename choice is your initials, followed by a brief experimental descriptor, eg: CM_s01_exp2_1HNMR When looking for this file remember UNIX is case sensitive. Enter additional info in the sample id field to convey more information about the sample or experiment. Click the auto_gain box Entries in other tabs are generally not needed. Click on Submit in the upper right of the window to start your experiment. The experiment is entered into the NMR s queue and will be run on a first-in, first-out basis. A pop up will appear and ask you to acknowledge addition of the experiment to the queue. Accept this by clicking Go. You can see the experiment queue in the Spectrometer Control window. In this window you may abort an experiment by clicking on the experiment in the queue and clicking on the Stop button. When the experiment is complete a data processing window will open with the spectrum.

5 D. Processing a Proton NMR Spectrum After measurement is complete, data is automatically forwarded from the spectrometer, and processing is automatically executed (If this is not done click on the 1H button and then click on the Process button). Figure 8. The 1D Processor window This section explains data processing in the 1-D Processor. The window usually displays two forms of the data. The upper display is the Free Induction Decay (FID) which is raw data from the spectrometer (display of the FID can be unselected in the preferences menu). The lower display is a more familiar spectrum processed by Fourier transformation (FT). Processing cursor tool On most spectra you will want to set a reference peak, peak pick important peaks, integrate the peaks and measure J-couplings. Short descriptions of these procedures are given here. The Delta manual has much more detail on processing. 1. Set a Reference Peak in the Display menu. Choose Display > Reference > Automatic Reference to set the reference. You will need to unselect the TMS radio button. Click on Process in the upper right. 2. Automatic integration and peak picking are obtained by clicking on the icon with an integral sign and octagon in the icon bar of the 1D Processor window. Figure 9. 1D Processor window icons 3. Determine coupling. The processor tab button in the upper right corner of the spectrum opens a series of processing options (see figure 8 of processor window above). In Peak mode of the cursor tool click on more peaks tool. On your spectrum select the 2 peaks where you want to measure a J value. A hashtag appears above the selected peak. Type "j" to assign a value. Type "u" to unselect those two peaks, repeat as necessary. More detailed information on processing spectra can be found in the Delta manual.

6 E. Finishing up 1. Eject your sample and return the standard to the NMR Click on the Sample button in lower left corner of Spectrometer Control window to go back to the Sample window. You need to retrieve your sample and replace it with the standard sample that was in the NMR. This is similar to loading your initial sample. In the Sample window place the cursor over the sample slot box Input or click arrows to indicate slot 1, the value for the slot holding the standard sample. Move the cursor out of the box. The instrument automatically ejects the current sample and loads the standard Figure 10. Sample window that is in auto sampler slot Lock on the standard sample in slot 1. Select the Solvent Sample usually it will be D2O slot Click on the icon to load system shims. Click on the icon for locking the sample. Close the sample window. 3. Disconnect NMR and Logoff In Spectrometer Control window click on the Unlink button. You are no longer in control of the NMR. Close the spectrometer control window. Record your experiment in the NMR log. Make note of any problems that occurred Log off the computer. Congratulations you have acquired an NMR spectrum. Lab Master IP address: Remove your sample and Clean your NMR tube After logoff retrieve your sample from the auto sampler and take it out of the spinner. Place the spinner back in foam spinner rack. Pour your NMR sample in the halogenated waste in the hood. Place your empty tube on the tube cleaner and rinse with acetone. Save the tube for reuse.

NMR Users Guide Organic Chemistry Laboratory

NMR Users Guide Organic Chemistry Laboratory NMR Users Guide Organic Chemistry Laboratory Introduction The chemistry department is fortunate to have a high field (400 MHz) Nuclear Magnetic Resonance (NMR) spectrometer. You will be using this instrument

More information

Varian Solution NMR Automation Procedure

Varian Solution NMR Automation Procedure Menu System System Tool Bar Command Line Graphics Tool Bar Vertical Panels Protocols NMR Data Display NMR Graphics Area Study Queue Horizontal Panels Activity Monitoring Varian Solution NMR Automation

More information

Varian Solution NMR Automation Procedure (VNMRS Machines running VNMRJ 4.2 under Red Hat Enterprise 5.1)

Varian Solution NMR Automation Procedure (VNMRS Machines running VNMRJ 4.2 under Red Hat Enterprise 5.1) Menu System System Tool Bar Command Line Graphics Tool Bar Vertical Panels Protocols NMR Data Display NMR Graphics Area Study Queue Horizontal Panels Activity Monitoring Varian Solution NMR Automation

More information

NMR Spectroscopy with VnmrJ. University of Toronto, Department of Chemistry

NMR Spectroscopy with VnmrJ. University of Toronto, Department of Chemistry NMR Spectroscopy with VnmrJ University of Toronto, Department of Chemistry Walk-up interface 1 Logging in 1 Starting VnmrJ 1 Inserting sample into the magnet or sample changer 1 Enter sample information

More information

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER DISCLAIMER Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 This document is intended to be a brief, bare-bones user s

More information

Varian Solution NMR Procedure

Varian Solution NMR Procedure System Tool Bar Command Line Vertical Panels Protocols Menu System User Tool Bar NMR Data Display Graphics Tool Bar NMR Graphics Area Study Queue Horizontal Panels Hardware Bar Varian Solution NMR Procedure

More information

c. Saturday and Sunday: 10 minute time blocks and unlimited time.

c. Saturday and Sunday: 10 minute time blocks and unlimited time. Scheduling 400 MHz A with 100 slot sample changer: a. Monday to Friday from 8 am to 8 pm: 10 minute blocks, 30min max. b. Monday to Friday from 8 pm to 8 am: 10 minute blocks, unlimited time. c. Saturday

More information

VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS

VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS April 16, 2018 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they are shorter; you

More information

General Information for Automated Inova 500 MHz NMR (Indy)

General Information for Automated Inova 500 MHz NMR (Indy) General Information for Automated Inova 500 MHz NMR (Indy) Primary Contacts: NMR Facility Manager: David VanderVelde, x3004, davidv@caltech.edu GLA: Pamela Tadross, x6131, 646-220-5789 (cell), 204 Church,

More information

Short instructions for the basic operation of the AVANCE II 400 MHz Bruker NMR Using ICON-NMR (PS751)

Short instructions for the basic operation of the AVANCE II 400 MHz Bruker NMR Using ICON-NMR (PS751) Dissimilar This manual is intended to be only a very brief introduction for using the Bruker 400 MHz NMR spectrometer (PS751) at the California State University LA NMR Facility. For complete information

More information

Vanderbilt Small Molecule NMR Center

Vanderbilt Small Molecule NMR Center Vanderbilt Small Molecule NMR Center Instructions for Using TOPSPIN version 2.0 DRX 400: Setting up 1H, 13C (1H decoupled), and 31P NMR Experiments Getting started: Sample Preparation: Make sure your sample

More information

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 January 10, 2019 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they are shorter; you

More information

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500 VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500 August 9, 2016 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they

More information

North Carolina State University Department of Chemistry Varian NMR Training Manual

North Carolina State University Department of Chemistry Varian NMR Training Manual North Carolina State University Department of Chemistry Varian NMR Training Manual by J.B. Clark IV & Dr. S. Sankar 1 st Edition 05/15/2009 Section 1: Essential Operations for Basic 1D Spectra Preparing

More information

NMR Training VNMRJ 4.2A/VNMRS 400

NMR Training VNMRJ 4.2A/VNMRS 400 NMR Training VNMRJ 4.2A/VNMRS 400 The VNMRS/new 400 MHz NMR uses VNMRJ 4.2A. Although VNMRJ is different than VNMRalmost ALL typed commands that you know from VNMR should work. LOGIN: Your username is

More information

Quick Guide to Topspin

Quick Guide to Topspin Quick Guide to Topspin 1. Log into the workstation. Right mouse click on the desktop anywhere and select new terminal. Type topspin [enter] (if this fails, create the launcher as described next). Alternatively,

More information

MNova Version Installation Instructions.

MNova Version Installation Instructions. MNova Version 12.0.0 Installation Instructions. Note: This Document contains information for a new NMR user. Carefully read and accomplish each section before moving to the next. If you can t finish a

More information

NMR Spectrometer Crib-Sheet

NMR Spectrometer Crib-Sheet NMR Spectrometer Crib-Sheet Sample Preparation: 1. Dissolve your sample in a deuterated solvent. For 1 H NMR use ~ 1 mm concentrations. For 13 C NMR, use ~ 10 mm. 2. Use clean dry tubes to avoid contaminating

More information

1D NMR Training Manual for Operating the Bruker 700 MHz Avance III HD NMR. Spectrometer without the Sample Changer and IconNMR

1D NMR Training Manual for Operating the Bruker 700 MHz Avance III HD NMR. Spectrometer without the Sample Changer and IconNMR 1D NMR Training Manual for Operating the Bruker 700 MHz Avance III HD NMR Spectrometer without the Sample Changer and IconNMR Scheduling NMR Time: Any use of the 700 MHz NMR spectrometer requires prior

More information

Agilent VnmrJ 3.2. Quick Start Guide

Agilent VnmrJ 3.2. Quick Start Guide Agilent VnmrJ 3.2 Quick Start Guide VnmrJ 3.2 Interface 3 Overview 4 Prepare Sample 5 Load Sample 6 Enter Sample Information 7 Build Study Queue 8 Run Study Queue 8 Process Data 9 Plot Data 10 This Quick

More information

Bruker Avance 400 MHz Instructions

Bruker Avance 400 MHz Instructions Bruker Avance 400 MHz Instructions General Policy: 1) All users must pass the training course before they will obtain their own access to the spectrometer. 2) A key to Bruker Avance 400 Room is accessible

More information

Delta V5 operation. Widetron Technologies Corp. Wu, Chung Ying. Progress through Synergy

Delta V5 operation. Widetron Technologies Corp. Wu, Chung Ying. Progress through Synergy Delta V5 operation Widetron Technologies Corp. Wu, Chung Ying Connect to spectrometer 1. Start Delta V5 (double click ). 2. Open Spectrometer Control (click ). 3. Select spectrometer on Available Instruments.

More information

Troubleshooting Acquisition Related Problems

Troubleshooting Acquisition Related Problems Troubleshooting Acquisition Related Problems This handout contains pertinent information to help you when a problem arises. Please look through the topics below to identify and find resolutions to the

More information

CHEMISTRY DEPARTMENT Category Work Instruction

CHEMISTRY DEPARTMENT Category Work Instruction No. of pages: 1 of 10 1. PURPOSE The purpose of this procedure established is to provide clear instructions for steps or orders on how to run sample analysis using JEOL NMR ECX 400MHz SYSTEM JOEL equipment.

More information

TopSpin QuickStart. Default State TopSpin should be running at all times. If this is not the case, you may need to Log In and Start TopSpin.

TopSpin QuickStart. Default State TopSpin should be running at all times. If this is not the case, you may need to Log In and Start TopSpin. TopSpin QuickStart Preliminaries Sign Up Sign up for instrument time at calendar.yahoo.com/hmcnmr The password is nmr. During the summer, 1100-1200 and 1600-1700 are reserved for walk-on use (15 min maximum

More information

Bruker NMR Topspin Manual

Bruker NMR Topspin Manual Bruker NMR Topspin Manual Topspin is an icon driven software package. This means that instead of reading words and typing commands in the software most of the things will be carried out by clicking icons

More information

Acquiring Data in VnmrJ 4.2A

Acquiring Data in VnmrJ 4.2A Acquiring Data in VnmrJ 4.2A Linux Primer Initial Steps Changing Password Disabling Screen Lock Reservations Reservation Terminals Rules and Proper Etiquette System Identification Working with VnmrJ 4.2A

More information

Running Multiple Experiments on the Same Sample Automatically. (multizg, spooler/queue, iconnmr)

Running Multiple Experiments on the Same Sample Automatically. (multizg, spooler/queue, iconnmr) Running Multiple Experiments on the Same Sample Automatically (multizg, spooler/queue, iconnmr) Multiple Experiment Setup If you have one sample that requires full characterization (e.g. 1 H, 13 C, COSY,

More information

EQNMR Facility Rules of Use

EQNMR Facility Rules of Use New User Package for East Quad 600 NMR Date: 01/25/08 Welcome to the East Quad NMR facility. This document contains information for new users using the 600MHz NMR at our facility. In this package, you

More information

PINMRF. Varian 300 MHz NMR Spectrometers Training Guide for Basic 1D NMR Spectroscopy

PINMRF. Varian 300 MHz NMR Spectrometers Training Guide for Basic 1D NMR Spectroscopy PINMRF Varian 300 MHz NMR Spectrometers Training Guide for Basic 1D NMR Spectroscopy INCLUDING: Inova-300-1 w/ 5mm 4-nucleus probe 365 WTHR Inova-300-2 w/ 5mm 4-nucleus probe 4100 BRWN Table of Contents

More information

Walkup NMR. Varian NMR Spectrometer Systems With VNMR 6.1C Software. Pub. No , Rev. A0800

Walkup NMR. Varian NMR Spectrometer Systems With VNMR 6.1C Software. Pub. No , Rev. A0800 Walkup NMR Varian NMR Spectrometer Systems With VNMR 6.1C Software Pub. No. 01-999159-00, Rev. A0800 Walkup NMR Varian NMR Spectrometer Systems With VNMR 6.1C Software Pub. No. 01-999159-00, Rev. A0800

More information

DCIF NMR Training Guide 401 MHz Bruker AVANCE III HD b401 last edit 3/11/2015

DCIF NMR Training Guide 401 MHz Bruker AVANCE III HD b401 last edit 3/11/2015 DCIF NMR Training Guide 401 MHz Bruker AVANCE III HD b401 last edit 3/11/2015 The Bruker AVANCE III HD 401 has a broadband (BBO) probe with a 1 H Channel and X-Channel tunable from (30-300 MHz). Both channels

More information

Bruker AVANCE III HD 400 MHz Spectrometer Operation

Bruker AVANCE III HD 400 MHz Spectrometer Operation Bruker AVANCE III HD 400 MHz Spectrometer Operation The spectrometer is located in Room 312 of Physical Sciences I. It is equipped with a SampleXpress sample changer for full automation. To utilize the

More information

DCIF NMR Training Guide 400 MHz Bruker DRX B401 Bruker DPX B400

DCIF NMR Training Guide 400 MHz Bruker DRX B401 Bruker DPX B400 1 DCIF NMR Training Guide 400 MHz Bruker DRX B401 Bruker DPX B400 400: The Bruker DPX 400 has a broadband probe with 1 H Channel and X-Channel tunable from (30-300 MHz). Both Channels may be tuned researchers.

More information

Bruker Topspin NMR Training Manual

Bruker Topspin NMR Training Manual Bruker Topspin NMR Training Manual Rev 180521 Perry Pellechia University of South Carolina NMR Facility Subject Page Number Overview of Operation... 1 Gauging Samples in Spinners... 2 Starting and logging

More information

VnmrJ Walkup. Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No , Rev. 0604

VnmrJ Walkup. Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No , Rev. 0604 VnmrJ Walkup Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No. 01-999266-00, Rev. 0604 VnmrJ Walkup Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No. 01-999266-00, Rev. 0604 VnmrJ Walkup

More information

Nicolet is50 FTIR User s Booklet

Nicolet is50 FTIR User s Booklet Nicolet is50 FTIR User s Booklet I. Getting started 1) Start your session by launching the LSA Chemistry Recharge software. This software keeps track of the amount of time that you are using the instrument.

More information

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software Spectrophotometers with Vernier Data Acquisition Software The absorbance of a sample is given as A = log I o I, where I o is the intensity without sample present and I is the intensity with the sample

More information

Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer

Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer Dr. Zhenming Du Email: zdu@gsu.edu Phone: 3-5538 Revised 04/02/2015 LC=Left click; DC=double

More information

DCIF NMR Training Guide 400 MHz Bruker AVANCE III HD b400 last edit 3/9/2015

DCIF NMR Training Guide 400 MHz Bruker AVANCE III HD b400 last edit 3/9/2015 DCIF NMR Training Guide 400 MHz Bruker AVANCE III HD b400 last edit 3/9/2015 The Bruker AVANCE III HD 400 has a broadband (BBO) probe with a 1 H Channel and X-Channel tunable from (30-300 MHz). Both channels

More information

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP)

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP) 500 MHz Solution-state NMR Procedure (Bruker AVANCE Machines running TopSpin under WINDOWS XP) Jerry Hu, x7914, jghu@mrl.ucsb.edu Shamon Walker, x6079 shamonwalker@mrl.ucsb.edu Materials Research Laboratory

More information

DCIF FAQ: Frequently Asked Questions Varian Last Update (2/21/2005)

DCIF FAQ: Frequently Asked Questions Varian Last Update (2/21/2005) DCIF FAQ: Frequently Asked Questions Varian Last Update (2/21/2005) How do I get help? Please review the material here and in the training guide. For night and weekend support to fix instrument acquisition

More information

NMR Data Acquisition and Processing Procedure

NMR Data Acquisition and Processing Procedure NMR Data Acquisition and Processing Procedure Dr. Jianfeng Zhu (Research Officer) Please DO NOT remove from NMR lab! Things to Avoid when Using NMR Lab Here's a list of stuff that I have seen that you

More information

TraceFinder Analysis Quick Reference Guide

TraceFinder Analysis Quick Reference Guide TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in the Thermo TraceFinder 3.0 analytical software. For detailed descriptions

More information

Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300

Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300 1 Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300 Please note: Under no circumstances move the magnet or the automatic sampler table. Do not attempt to use the auto sampler

More information

DCIF NMR Training Guide Varian Mercury 300 MHz Last Update

DCIF NMR Training Guide Varian Mercury 300 MHz Last Update DCIF NMR Training Guide Varian Mercury 300 MHz Last Update 10-12-07 1 This document is intended as an introductory guide to the Varian VNMR NMR processing / acquisition software. It is by no means intended

More information

TopPoint User's Guide

TopPoint User's Guide TopPoint User's Guide LT2152/ 2000 Gilson, Inc. All rights reserved August 2000 Table of Contents 1 Introduction 2 Operation 3 Troubleshooting Steps to Using TopPoint... 1-2 Accessing Online Help... 1-3...

More information

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP)

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP) 500 MHz Solution-state NMR Procedure (Bruker AVANCE Machines running TopSpin under WINDOWS XP) Jerry Hu, x7914, jghu@mrl.ucsb.edu Shamon Walker, x3248 shamonwalker@mrl.ucsb.edu Materials Research Laboratory

More information

Jasco 4700 FTIR. Laser Spectroscopy Labs. Jasco 4700 FTIR. Operation instructions. Laser Spectroscopy Labs, UCI

Jasco 4700 FTIR. Laser Spectroscopy Labs. Jasco 4700 FTIR. Operation instructions. Laser Spectroscopy Labs, UCI Laser Spectroscopy Labs Jasco 4700 FTIR Operation instructions Turn the system on 1. Turn on the computer in case it has been switched off. Normally computer stays on. 2. The instrument should be always

More information

ONE DIMENSIONAL NMR SPECTROSCOPY. DPX 200 and DPX 300

ONE DIMENSIONAL NMR SPECTROSCOPY. DPX 200 and DPX 300 1 1H, 13C, 15N, 19F and 31P ONE DIMENSIONAL NMR SPECTROSCOPY (Xwinnmr on PC, W2Kpro, version 3.5) on DPX 200 and DPX 300 Department of Chemistry University of Oslo February 2005 Version 1 (To be used from

More information

TopSpin 1.3 patch level 10. DPX 200, DPX 300 and DRX 500. Windows XP

TopSpin 1.3 patch level 10. DPX 200, DPX 300 and DRX 500. Windows XP KJM 5280 1 H nmr introductory user manual I Version 1.5 TopSpin 1.3 patch level 10 DPX 200, DPX 300 and DRX 500 Windows XP F. Rise May 2018 1 Welcome to the wonderful and amazing world of Nuclear Magnetic

More information

Varian NMR Instructions - 1D. Department of Chemistry NMR Facility University of Minnesota

Varian NMR Instructions - 1D. Department of Chemistry NMR Facility University of Minnesota Varian NMR Instructions - 1D Department of Chemistry NMR Facility University of Minnesota June 17, 2008 Table of Contents 1. General Information...2 2. References...2 3. Rules...3 4. Scheduling...3 5.

More information

PROCESSING 2D SPECTRA USING VNMRJ JB Stothers NMR Facility Materials Science Addition 0216 Department of Chemistry Western University

PROCESSING 2D SPECTRA USING VNMRJ JB Stothers NMR Facility Materials Science Addition 0216 Department of Chemistry Western University PROCESSING 2D SPECTRA USING VNMRJ JB Stothers NMR Facility Materials Science Addition 0216 Department of Chemistry Western University 1. INTRODUCTION...1 1.1. About this Worksheet... 1 1.2. A Very Brief

More information

Processing data with Bruker TopSpin

Processing data with Bruker TopSpin Processing data with Bruker TopSpin This exercise has three parts: a 1D 1 H spectrum to baseline correct, integrate, peak-pick, and plot; a 2D spectrum to plot with a 1 H spectrum as a projection; and

More information

MALDI Software Users Guide (simplified version)

MALDI Software Users Guide (simplified version) MALDI Software Users Guide (simplified version) Central Analytical Lab Ekeley E266 The Acquisition program is called Voyager Control Panel. Once you shoot your sample and determine it looks good, you transfer

More information

Agilent VnmrJ 3.2 Automation

Agilent VnmrJ 3.2 Automation Agilent VnmrJ 3.2 Automation User Guide Agilent Technologies Notices Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in any form or by any means (including electronic storage and

More information

VnmrJ Walkup. Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software Pub. No , Rev. A 0808

VnmrJ Walkup. Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software Pub. No , Rev. A 0808 VnmrJ Walkup Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software Pub. No. 01-999377-00, Rev. A 0808 VnmrJ Walkup Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software

More information

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Section 1: Starting an experiment 1.1 Opening a file 1. Open the File menu and select Open Activity. 2. In the Open dialog box, navigate to

More information

3 Starting the chiral HPLC and logging into LC Real Time Analysis.

3 Starting the chiral HPLC and logging into LC Real Time Analysis. Edition: 1.2 Date: 7-Oct-11 Page 1 of 5 1 Responsibilities. First person responsible: Mariëlle Delville. Second person responsible: Helene I.V. Amatdjais-Groenen. 2 Definitions. Read the information on

More information

Basic Procedure 1D NMR Data Collection

Basic Procedure 1D NMR Data Collection Basic Procedure 1D NMR Data Collection Log into your UNIX account and start XWIN- NMR program At the prompt login:, type your (e.g. jdw, mmj, mjt) or click on your group icon (in some spectrometers

More information

Gemini-300, Manual Operation (version 1.0, 4Feb95) 1

Gemini-300, Manual Operation (version 1.0, 4Feb95) 1 Gemini-300, Manual Operation (version 1.0, 4Feb95) 1 I. GETTING STARTED (LOGGING IN, LOGGING OUT, USING THE NMR PROGRAM). A. Logging into the Gemini To login onto the Sun workstations, hit the return key.

More information

CyAn ADP Guide. Starting Up

CyAn ADP Guide. Starting Up CyAn ADP Guide Starting Up 1. Check the sheath and waste fluids, if needed, fill and empty: 1. To fill the Sheath Tank: a. Unscrew the top of the tank labeled Sheath Tank, and set it on top of the tank.

More information

TraceFinder Analysis Quick Reference Guide

TraceFinder Analysis Quick Reference Guide TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in Thermo TraceFinder analytical software. For detailed descriptions

More information

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer Operating instructions - Cary 100 Bio UV-visible Spectrophotometer This document describes how to power up the spectrophotometer, set its measurement parameters, insert one or more of samples for analysis

More information

Sciex QTrap Operational Steps for Trained Personnel

Sciex QTrap Operational Steps for Trained Personnel Sciex 6500+ QTrap Operational Steps for Trained Personnel Last Updated 09172017 1. If any of the following instructions does not make sense to you or was not covered during your hand-on training, stop

More information

Waste Pickup Form User s Guide

Waste Pickup Form User s Guide Waste Pickup Form User s Guide Updated: 3/13/12 Introduction: Welcome to the Cal State University Fullerton Online Waste Pickup Form User s Guide. In this guide you will learn what you can use the form

More information

NMR Spectroscopy with VNMRJ & EZ NMR

NMR Spectroscopy with VNMRJ & EZ NMR NMR Spectroscopy with VNMRJ & EZ NMR Locally developed EZ NMR for VNMRJ versions 2.2D/2.3A Dr. Albin Otter Edited by: Mark Miskolzie Department of Chemistry University of Alberta Edmonton, Alberta T6G

More information

Open Access Manual for Superusers

Open Access Manual for Superusers NMR-Service LOC Tel: +41 1 632 29 33 Email: nmrservice@org.chem.ethz.ch Open Access Manual for Superusers 1. Sample preparation Make a homogeneous solution of ca. 20mg in exactly 0.7ml of deuterated solvent,

More information

CONTROL ROOM SOFTWARE

CONTROL ROOM SOFTWARE CONTROL ROOM SOFTWARE Starting QTWatch... Main screen... Control Monitor Screen... Action page... Site Map page... Standard Operations page... Handling calls on Actions... How to log a manual/remote call...

More information

Fitting NMR peaks for N,N DMA

Fitting NMR peaks for N,N DMA Fitting NMR peaks for N,N DMA Importing the FID file to your local system Any ftp program may be used to transfer the FID file from the NMR computer. The description below will take you through the process

More information

MestReC Cheat Sheet. by Monika Ivancic, July 1 st 2005

MestReC Cheat Sheet. by Monika Ivancic, July 1 st 2005 MestReC Cheat Sheet by Monika Ivancic, July 1 st 2005 This Cheat Sheet is to be used at UW-Madison as a quick guide to processing using the MestReC NMR software. You may find more help at the MestReC homepage

More information

CSImage Tutorial v August 2000

CSImage Tutorial v August 2000 CSImage Tutorial v 1.0 8 August 2000 To use CSImage you need to have Java2 installed. Development was done using JDK 1.2.2 from Sun Microsystems. The program has been run on Digital Unix (Compaq Tru64

More information

Getting low-resolution GC-MS data on the JEOL GCMate Introduction

Getting low-resolution GC-MS data on the JEOL GCMate Introduction Getting low-resolution GC-MS data on the JEOL GCMate Introduction GC-MS is a technique that combines gas phase separation and mass-based detection/characterization of analytes. It s a very sensitive method

More information

Lab - Share Resources in Windows

Lab - Share Resources in Windows Introduction In this lab, you will create and share a folder, set permissions for the shares, create a Homegroup and a Workgroup to share resources, and map a network drive. Due to Windows Vista lack of

More information

2 Dimensional NMR User s Brief Guide

2 Dimensional NMR User s Brief Guide 2D NMR Handout 7/17/18 Weiguo Hu 2 Dimensional NMR User s Brief Guide (This handout presumes your thorough familiarity with the 1D handout commands and their use!) Options There are a vast number of 2D

More information

Operating Procedure for Horiba Raman Microscope

Operating Procedure for Horiba Raman Microscope Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch

More information

Visual Physics Introductory Lab [Lab 0]

Visual Physics Introductory Lab [Lab 0] Your Introductory Lab will guide you through the steps necessary to utilize state-of-the-art technology to acquire and graph data of mechanics experiments. Throughout Visual Physics, you will be using

More information

Standard Operating Procedure for the Horiba FluroMax-4

Standard Operating Procedure for the Horiba FluroMax-4 Standard Operating Procedure for the Horiba FluroMax-4 Adapted from Horiba Operations Manual Created by Michael Delcau, Modified by Brian Lamp The Fluoromax is capable of making a variety of measurements.

More information

ChromQuest 5.0 Quick Reference Guide

ChromQuest 5.0 Quick Reference Guide ChromQuest 5.0 Quick Reference Guide This guide contains an overview of the ChromQuest chromatography data system, with topics organized by workflow. For more information, refer to the ChromQuest User

More information

TopPoint Administrator User's Guide

TopPoint Administrator User's Guide TopPoint Administrator User's Guide LT2153/ 2000 Gilson, Inc. All rights reserved December 2000 Table of Contents 1 Introduction 2 Installation 3 Operation System Requirements... 1-2 Minimum... 1-2 Recommended...

More information

ChemSense Studio Client Version 3.0.7

ChemSense Studio Client Version 3.0.7 Quick Start Guide: ChemSense Studio Client Version 3.0.7 January 5, 2005 Comments/Questions/Bug Report? E-mail: chemsense-contact@ctl.sri.com Background The ChemSense Studio Client software supports the

More information

UC DAVIS NMR FACILITY VNMRJ SHORT MANUAL FOR VARIAN/AGILENT NMR SPECTROMETERS

UC DAVIS NMR FACILITY VNMRJ SHORT MANUAL FOR VARIAN/AGILENT NMR SPECTROMETERS UC DAVIS NMR FACILITY VNMRJ SHORT MANUAL FOR VARIAN/AGILENT NMR SPECTROMETERS TABLE OF CONTENTS Disclaimer... 2 Conventions... 2 Routine 1D NMR Experiments Procedure for acquiring 1D Proton spectrum...

More information

Skyline Targeted MS/MS

Skyline Targeted MS/MS Skyline Targeted MS/MS Skyline now supports several methods of extracting chromatography-based quantitative measurements from the raw data files of full-scan mass spectrometers, such as ion trap and Q-TOF

More information

Tecan crystallisation robot user guide

Tecan crystallisation robot user guide Tecan crystallisation robot user guide The TECAN crystallisation robot in RRB S9 provides an automated facility for setting up crystallisation trays. This user guide is meant as online support for trained

More information

A Brief Introduction to the ProClarity Enrollment Reports

A Brief Introduction to the ProClarity Enrollment Reports ACCESS TO THE REPORTS First Step: Open the recommended browser, Internet Explorer, go to the Compatibility View Settings on the Tools menu & select the Display all websites in Compatibility View checkbox.

More information

FlowAccess TM GETTING STARTED. Windows software for Skalar SAN plus systems. Version and up

FlowAccess TM GETTING STARTED. Windows software for Skalar SAN plus systems. Version and up FlowAccess TM V3 Windows software for Skalar SAN plus systems Version 3.1.0.4 and up GETTING STARTED FlowAccess TM V3 Windows software for Skalar SAN plus systems Version 3.1.0.4 and up Getting started

More information

Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted)

Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted) Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted) Once FTIR data has been acquired for the gases evolved during your DSC/TGA run, you will process using the OPUS software package. Select

More information

Adobe Premiere: Getting Started

Adobe Premiere: Getting Started Prepared by Ansel Herz for the UW Department of Communication http://www.com.washington.edu/tech/irc/ Questions? Comments? Contact Kristina Bowman at kriscb@uw.edu. Adobe Premiere: Getting Started This

More information

Welcome to the NMR Center

Welcome to the NMR Center 1 Welcome to the NMR Center The NMR center in chemistry consists of six Bruker NMR instruments Field/Name Location Comments Operating Software 300nb (narrow bore) 119 Havemeyer walk-on, routine 1 H Topspin

More information

Protein Deconvolution Quick Start Guide

Protein Deconvolution Quick Start Guide Protein Deconvolution Quick Start Guide The electrospray ionization (ESI) of intact peptides and proteins produces mass spectra containing a series of multiply charged ions with associated mass-to-charge

More information

Fluorimeter User s Booklet

Fluorimeter User s Booklet Fluorimeter User s Booklet 1) Verify that the instrument power and the computer are both off. Then turn on the lamp. The red indicator light will glow when the lamp is on. Turn on lamp switch here. This

More information

Malvern Zetasizer Nano Basic Operating Instructions 08/14/2015 S.V.

Malvern Zetasizer Nano Basic Operating Instructions 08/14/2015 S.V. 1234 Hach Hall 515-294-5805 www.cif.iastate.edu Malvern Zetasizer Nano Basic Operating Instructions 08/14/2015 S.V. Location: Contact: 1238 Hach Hall Steve Veysey, 1234 Hach Hall Safety All researchers

More information

Nanosight NS300 NBTC User Instructions

Nanosight NS300 NBTC User Instructions Nanosight NS300 NBTC User Instructions Preliminary notes: -The laser module is delicate and expensive. Handle it with care and do not leave it on the bench unattended. Only leave it in place in the instrument

More information

Agilent GC/MSD Instructions

Agilent GC/MSD Instructions Agilent GC/MSD Instructions GC/MS SAMPLE PREPARATION: Sample Components to Avoid Completely: The following should never be injected: metals, strong acids or bases, salts, oligomeric and polymeric material.

More information

T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised

T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised 4-21-2015 Relaxation times are divided into two types: longitudinal, which concerns change in magnetization along the

More information

Investigator Site OC RDC PDF User Guide

Investigator Site OC RDC PDF User Guide Investigator Site OC RDC PDF User Guide Version 1.0 Page 1 of 40 TABLE OF CONTENTS Accessing OC RDC Steps for Access 3 Logging On 4 Change Password 4 Laptop and System Security 5 Change Study 5 Navigating

More information

Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection

Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT For assistance in

More information

Agilent MassHunter Qualitative Data Analysis

Agilent MassHunter Qualitative Data Analysis Agilent MassHunter Qualitative Data Analysis Qualitative Navigator B.08.00 Presenters: Kevin Costalunga Stephen Harnos With Matt Leyden & Kevin Costalunga 1 MassHunter Webinar Series MassHunter Qualitative

More information

Visual Physics - Introductory Lab Lab 0

Visual Physics - Introductory Lab Lab 0 Your Introductory Lab will guide you through the steps necessary to utilize state-of-the-art technology to acquire and graph data of mechanics experiments. Throughout Visual Physics, you will be using

More information

UC DAVIS NMR FACILITY BRUKER SHORT MANUAL FOR TOPSPIN 3.X

UC DAVIS NMR FACILITY BRUKER SHORT MANUAL FOR TOPSPIN 3.X TABLE OF CONTENTS UC DAVIS NMR FACILITY BRUKER SHORT MANUAL FOR TOPSPIN 3.X VERSION 3 February, 2016 Disclaimer... 2 Conventions... 2 Routine 1D NMR Experiments Procedure for acquiring 1D Proton spectrum...

More information