- Polly Johnson
- 6 years ago
- Views:
Transcription
1 Setup and analysis Mic qpcr Cycler (magnetic induction cycler; Bio Molecular Systems)
2 General remark Runs on Mic (Bio Molecular Systems) can be set up and analysed either manually (described in part 1 of the manual) or using pre-defined assays (part 2). Troubleshooting for problems that might occur when using FTD assays on Mic is described in the third part of this manual. All descriptions in this setup manual are based on micpcr software version v
3 1. Manual setup and analysis 3
4 How to set up a run on Mic qpcr Cycler 1. Switch on the cycler first, then start the micpcr software (v1.12.0) 2. Click New and select Run to set all parameters manually, or Run from Template, if the FTD run profile is already saved on the computer. 4
5 How to set up a run on Mic qpcr Cycler 3. In the sidebar on the left, select Run Setup Run Profile. 4. Edit the cycling profiles to match the cycling steps listed in the manual of FTD kit. 5
6 How to set up a run on Mic qpcr Cycler 5. Make sure acquisition is switched on for all channels in the assay by clicking on the channel name. A camera icon indicates data acquisition in the respective channel. Data acquisition needs to be set to the end of the elongation step (indicated by a camera symbol). 6. Adjust gain settings can be left in default settings (AutoGain, all tubes). 6
7 How to set up a run on Mic qpcr Cycler 7. Make sure Temperature Control is set to Standard TAQ (v3) (default setting). 8. Adjust the reaction volume: 15 µl for FTlyo products, 25 µl for FTD liquid product line. Optional: To save these settings as a template, select Save as Template 7
8 How to set up a run on Mic qpcr Cycler 9. Insert the tubes into the rotor and make sure the tube clamp is in place. Load tubes filled with water in unused wells. 10. Close the lid, select the instrument in the software and Start Run. 8
9 How to analyse a run on the Mic qpcr Cycler (without assay function) Qualitative analysis 1. Assigning samples: The sample list can be edited before, during or after the run. Click on Run Setup Samples in the left side bar to open the sample editor. Enter sample names, assign the sample type and, if desired, change the display colour for the amplification plot. 2. Add a new analysis by clicking on the + icon next to Analysis Cycling and select Non-Assay and the channel you want to analyse. The analysis item will appear under Cycling in the sidebar. Note: Separate analyses should be created for each PPMix in the run. You can create multiple analyses of the same channel by repeating the steps described here. 9
10 How to analyse a run on Mic qpcr Cycler (without assay function) 3. In the panel Sample: Select all samples to be included in the analysis. De-selected samples will be greyed out. To hide samples but still have them included for analysis and the calculation of e.g., the automatic threshold, click on the eye icon next to the sample name. 10
11 How to analyse a run on Mic qpcr Cycler (without assay function) 4. In the panel Parameters, change the Method to Dynamic. If background fluorescence occurs during the first cycles, increase Ignore Cycles Before. Other parameters can be used as default. Note: If you do not want to use the Auto Set Threshold function, deselect this function by unchecking the tick box, and adjust the threshold by moving the red line in the panel Cycling Analysis. 5. When all assays are analysed, export the results either by creating a Report (click on + next to report in the left sidebar) or export the data as Excel workbook (Save as Excel workbook) 11
12 How to analyse a run on Mic qpcr Cycler (without assay function) Quantitative analysis 1. Assigning samples and standard concentrations: Click on Run Setup Samples in the left side bar to open the sample editor. a. Enter sample names and assign the sample type. b. Select the sample type Standard for all wells containing quantification standards (QS). c. Click on the icon Toggle between singleplex and multiplex standard to enable the standard function for all channels in one well d. Specify the concentration of each QS for all channels as listed in the manual of the respective FTD kit. 12
13 How to analyse a run on Mic qpcr Cycler (without assay function) 2. Add a new analysis by clicking on the + icon next to Absolute Quantification and select Non-Assay [Channel] 3. Click on the analysis name in the side bar and a small triangle/arrow will appear to the left of the assay. Click on this icon to expand the Cycling menu. 4. Select all samples and QS to be analysed in this analysis in the sample editor. Note: When running several FTD PPMixes during one run, create a separate analysis for each PPMix! 5. In the panel Parameters, change the Method to Dynamic. If background fluorescence occurs during the first cycles, increase Ignore Cycles Before. Other parameters can be used as default. Note: If you do not want to use the Auto Set Threshold function, deselect this function by unchecking the tick box, and adjust the threshold by moving the red line in the panel Cycling Analysis. 13
14 How to analyse a run on Mic qpcr Cycler (without assay function) 6. Click on the assay name again to view the results of the quantification. This view contains a plot of the standard curve and sample concentrations, the standard curve characteristics and results, and the results for the selected samples. Review if slope, efficiency and correlation coefficient match the target criteria given in the manual of the respective FTD kit. 14
15 2. Setup and analysis using the Assay function 15
16 Creating an assay Assay Setup 1. Start the micpcr software, click on the New icon and select Assay. 2. In the left sidebar, select Assay Setup - Information. 3. In the drop down menu Chemistry Type, select Hydrolysis Probes. 4. Enter the Name and Flurophore (or channel) for the first target of the assay in the field 5 Modifier. The other fields are not required for setting up the run. The tick box Contains Alleles should be switched off. 5. Additional targets can be added by clicking on the + icon next to Targets. 16
17 Creating an assay Assay Setup 6. To edit the run profile, select Assay Setup Profile. 7. Edit the cycling profiles to match the cycling steps listed in the manual of the FTD kit. 8. Make sure acquisition is switched on for all channels in the assay by clicking on the channel name. A camera icon indicates data acquisition in the respective channel. Data acquisition needs to be set to for the end of the elongation step (indicated by a camera symbol). 9. Adjust gain settings can be left in default settings (AutoGain, all tubes). 10. Make sure Temperature Control is set to Standard TAQ (v3) (default setting). 11. Adjust the reaction volume: 15 µl for FTlyo products, 25 µl for FTD liquid product line. 17
18 Creating an assay Analysis setup 12. Click on the the + icon next to the target and select Cycling. 13. Default criteria for assays with hydrolysis probes should appear as below: Note: It can be useful to increase Ignore Cycles Before if background fluorescence is high in the first cycles of the run (described in 3. Troubleshooting). 3. Save the assay. To be accessible from the assay list in the software, assay files need to be saved locally in the folder Assays of the software (default: [User]/My Documents/Bio Molecular Systems/micPCR/Assays) 18
19 How to setup a run on Mic qpcr Cycler using Assays 14. Switch on the cycler first, then start the micpcr software (v1.12.0) 15. Click New to create a new run. Add a saved assay by clicking on the + icon next to Assays in the left sidebar and select the assay(s) to be added to the run. If assay settings are not compatible because of different run profiles, a warning message will appear. 16. Insert the tubes into the rotor and make sure the tube clamp is in place. Load tubes filled with water into unused wells. Close the lid, select the instrument in the software and start the run. 19
20 How to analyse data using the assay function Qualitative analysis: 1. Select Run Setup Samples to open the sample editor and assign sample names and types. In the field Assay: Select the assay to be used from the drop down menu. 2. Add an analysis by clicking on the + icon next to Cycling and select the target you want to analyse. An analysis will be opened that contains only the samples to which the assay is assigned in the sample editor, analysed with the settings saved with the assay. If necessary analysis parameters can be modified as described in the first part of this manual. 3. When all assays are analysed, export the results either by creating a pdf Report (click on + next to report in the left sidebar) or export the data as Excel workbook (Save as Excel workbook) 20
21 Data analysis using the assay function Quantitative analysis: 1. Select Run Setup Samples to open the sample editor and assign sample names and types. Change the sample type Standard to all quantification standards. Click on the icon Toggle between singleplex and multiplex standards and enter the copy numbers for each channel as given in the manual of the corresponding FTD kit. 2. In the field assay: Select the assay to be used from the drop down menu. 21
22 Data analysis using the assay function 3. Add an analysis by clicking on the + icon next to Absolute Quantification and select the target you want to analyse. An analysis will be opened including only those samples to which the assay is assigned in the sample editor, analysed with the settings saved with the assay. Additionally, the results of the quantification standards are shown. If necessary analysis parameters can be modified as described in the first part of this manual. 4. When all assays are analysed, export the results either by creating a Report (click on + next to report in the left sidebar) or export the data as Excel workbook (Save as Excel workbook) 22
23 Troubleshooting 23
24 Troubleshooting In some cases, a combination of high background fluorescence during the first cycles of the run, and low signal intensity can require a modification of the default analysis parameters. Such problems can best be identified in the raw fluorescence curves under Data Cycling [Channel]. Samples with Mix C: high initial fluorescence, low signal intensity Adjust analysis parameters Mixes A+B: low background fluorescence and high signal to noise ratio Default analysis parameters work well Parameters to adjust in such a case (can be found in the Parameters panel of the Analysis): 1. Increase of number of cycles to ignore 2. Exclusion criteria a. Decrease Fluorescence cutoff level or b. Switch off entireley: select Exclusion None 24
25 Sloping or uneven baseline If the baseline is not flat or you observe a slope in the baseline, this can be caused by the wrong settings in the analysis parameters. Analysis Method LinRegPCR For all FTD assays, the analysis method must be set to Dynamic! Analysis Method Dynamic 25
26 If you have any further questions or suggestions, please contact us at or check out the FAQ section on our website
Setup and analysis QuantStudio 5 Real-Time PCR System
Setup and analysis QuantStudio 5 Real-Time PCR System Applied Biosystems Software 1 Open software Click on the QuantStudio Design and Analysis Software icon to start the software 2 Software 2 Home window
More informationRoche LightCycler 480
Roche LightCycler 480 Instrument Setup Instructions for RT 2 Profiler PCR Array, TaqMan Version A. Preparation Before the Experiment (Presetting the Run Template will save time for your run): 1. Make sure
More informationRotor-Gene AssayManager v1.0 UDT Basic Plug-in User Manual
February 2018 Rotor-Gene AssayManager v1.0 UDT Basic Plug-in User Manual R3 QIAGEN GmbH, QIAGEN Strasse 1, 40724 Hilden, Germany Sample to Insight Contents 1 UDT Basic Plug-in User Manual...1-1 1.1 Safety
More informationHow to run a qpcr plate on QuantStudio 6
How to run a qpcr plate on QuantStudio 6 Important notes Check the compatibility of your mastermix with this instrument. In particular, mastermixes containing «high rox» concentrations should not be used.
More informationMic qpcr. Fast. Accurate. Compact. Ultimate Performance for Real-Time PCR. Speed Accuracy Size Connectivity
Mic qpcr Ultimate Performance for Real-Time PCR Fast. Accurate. Compact. Speed Accuracy Size Connectivity Hardware Style meets substance - ultimate performance, beautiful design. Fast results without compromise
More informationqpcr Hand Calculations - Multiple Reference Genes
qpcr Hand Calculations - Multiple Reference Genes This word document will take you through the hand calculations needed to determine the Normalized Gene Expression, using the Pfaffl method for efficiency
More informationmiscript mirna PCR Array Data Analysis v1.1 revision date November 2014
miscript mirna PCR Array Data Analysis v1.1 revision date November 2014 Overview The miscript mirna PCR Array Data Analysis Quick Reference Card contains instructions for analyzing the data returned from
More informationLightCycler 480. Instrument and Software Training. Relative Quantification Module. (including Melting Curve)
LightCycler 480 Instrument and Software Training Relative Quantification Module (including Melting Curve) Agenda 1. Hardware demonstration (hands on) A quick look inside the instrument Xenon lamp exchange
More informationEcoStudy Software User Guide
EcoStudy Software User Guide FOR RESEARCH USE ONLY What is EcoStudy? 3 Setting Up a Study 4 Specifying Analysis Settings for your Study 6 Reviewing the Data in your Study 8 Exporting Study Data to a Report
More informationGuide to running KASP TM genotyping reactions on the Roche LC480-Series instruments
extraction sequencing genotyping extraction sequencing genotyping extraction sequencing genotyping extraction sequencing Guide to running KASP TM genotyping reactions on the Roche LC480-Series instruments
More informationAll About PlexSet Technology Data Analysis in nsolver Software
All About PlexSet Technology Data Analysis in nsolver Software PlexSet is a multiplexed gene expression technology which allows pooling of up to 8 samples per ncounter cartridge lane, enabling users to
More informationProNex DNA QC Assay Analysis Software
TECHNICAL MANUAL ProNex DNA QC Assay Analysis Software 9/17 TM512 ProNex DNA QC Assay Analysis Software All technical literature is available at: www.promega.com/protocols/ Visit the web site to verify
More informationUse the AccuSEQ Software v2.0 Custom Experiment mode
QUICK REFERENCE Use the AccuSEQ Software v2.0 Custom Experiment mode Publication Part Number 4425585 Revision Date 25 January 2013 (Rev. B) Custom Experiments Note: For safety and biohazard guidelines,
More informationData Collection Software Release Notes. Real-Time PCR Analysis Software Release Notes. SNP Genotyping Analysis Software Release Notes
PN 101-6531 D1 RELEASE NOTES Biomark/EP1 Software To download the latest version of the software for Biomark HD, Biomark, and EP1, go to fluidigm.com/software. For more information about updating the software,
More informationTraceFinder Analysis Quick Reference Guide
TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in Thermo TraceFinder analytical software. For detailed descriptions
More informationTraceFinder Analysis Quick Reference Guide
TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in the Thermo TraceFinder 3.0 analytical software. For detailed descriptions
More informationLEGENDplex Data Analysis Software Version 8 User Guide
LEGENDplex Data Analysis Software Version 8 User Guide Introduction Welcome to the user s guide for Version 8 of the LEGENDplex data analysis software for Windows based computers 1. This tutorial will
More informationIntroduction. Library quantification and rebalancing. Prepare Library Sequence Analyze Data. Library rebalancing with the iseq 100 System
Sequencing Library QC with the iseq System The iseq 100 System enables measurement of pooled library quality before a large-scale sequencing study on the NovaSeq 6000 System. Introduction This application
More informationInstructions manual Exporting Instrument Files
2013 Instructions manual Exporting Instrument Files Multid Analyses AB 5/22/2013 Contents 1. Exporting Instrument Files from LightCycler480... 3 2. Exporting Instrument Files from ABI7500... 8 3. Exporting
More informationGuide to running KASP genotyping on the BIO-RAD CFX-series instruments
extraction sequencing genotyping extraction sequencing genotyping extraction sequencing genotyping extraction sequencing genotypin Guide to running KASP genotyping on the BIO-RAD CFX-series instruments
More informationGetting Started Guide. HID Real-Time PCR Analysis Software Version 1.0 For 7500 Real-Time PCR System
Getting Started Guide HID Real-Time PCR Analysis Software Version 1.0 For 7500 Real-Time PCR System Getting Started Guide Get Started HID Real-Time PCR Analysis Software Version 1.0 For 7500 Real-Time
More informationcief Data Analysis Chapter Overview Chapter 12:
page 285 Chapter 12: cief Data Analysis Chapter Overview Analysis Screen Overview Opening Run Files How Run Data is Displayed Viewing Run Data Data Notifications and Warnings Checking Your Results Group
More informationMini8 Plus Real-Time PCR System
Mini8 Plus Real-Time PCR System Version:EN / 20160413 Coyote, highly specialized in research & development, offers integrate molecular diagnostic solutions by providing devices from sample pre to gene
More informationData acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.
Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.6 Contents 1. Description 1.1 Background information 1.2 Reagent and
More informationDART-PCR Version 1.0. Basic Use
RT-PR Version 1.0 Introduction RT-PR provides a simple means of analysing real-time PR data from raw fluorescence data. This allows an automated calculation of amplification kinetics, as well as performing
More informationQPCR User Guide. Version 1.8. Create Date Sep 25, 2007 Last modified Jan 26, 2010 by Stephan Pabinger
User Guide Version 1.8 Create Date Sep 25, 2007 Last modified Jan 26, 2010 by Table of content Table of content... 2 1 Introduction... 4 1.1 Purpose... 4 1.2 Hints for using the software... 4 1.3 Example
More informationUNC Flow Cytometry Core Facility LSR II Operation 2018
Reference: Startup BD FACSDiva 6 Online Course BD Training & elearning 1. Check Sheath and Waste containers: Empty Waste container in the sink Add bleach to the black mark ( ~10% final concentration).
More informationChromeleon software orientation
Chromeleon software orientation Upon opening of Chromeleon shortcut, a blue screen should appear (called control panel). If this does not occur, the green circled shortcut will open this screen. To ensure
More informationCEQ 8000 Series Fragment Analysis Training Guide
TM CEQ 8000 Series Fragment Analysis Training Guide PN A16039-AB September 2005 Beckman Coulter, Inc., 4300 N. Harbor Blvd., Fullerton, CA 92835 Copyright 2005 Beckman Coulter, Inc., Printed in U.S.A.
More informationData acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.
Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.8 Contents 1. Description 1.1 Background information 1.2 Reagent and
More informationSetup protocol for Firefly particles. On BD FACSCanto cytometers with single tube loader function
Setup protocol for Firefly particles On BD FACSCanto cytometers with single tube loader function Version 1 Last Updated May 2016 Setup protocol for Firefly particles Contents Introduction Cytometer requirements
More informationWhat is new in BioChek II Software version 2018
What is new in BioChek II Software version 2018 Elisa Test Create a plate layout Automatic registration of Customers and flocks/herds: Select Customers from a drop-down list Add Last Vaccination Quick
More informationAgilent 2100 Bioanalyzer. User s Guide for Molecular Assays
Agilent 2100 Bioanalyzer User s Guide for Molecular Assays ! WARNING For details of safety, see the Site Preparation and Safety Manual for the Agilent 2100 Bioanalyzer. The Agilent 2100 Bioanalyzer is
More informationUnscrew and remove the cap from the waste tank
Stanford Cytoflex HTS User Guide 11232015 Starting up 1. Check the sheath fluid level every time you use the cytometer. This ensures that you do not run out of sheath fluid during an experiment. Replenish
More informationUser Bulletin. Applied Biosystems StepOne Real-Time PCR System DRAFT. SUBJECT: StepOne System Updates. In This Bulletin Installation Updates...
Applied Biosystems StepOne Real-Time PCR System User Bulletin User Bulletin Applied Biosystems StepOne Real-Time PCR System January 2007 SUBJECT: StepOne System Updates In This Bulletin Installation Updates.................................................
More informationThermal Cycler Dice TM Real Time System Software. Operation Manual
Thermal Cycler Dice TM Real Time System Software Version 4.02 Code TP900/TP960 Operation Manual Attention: Thermal Cycler Dice TM Real Time System Software Operation Manual should be stored for reference
More informationSetup protocol for Firefly particles. On BD LSR II and LSRFortessa cytometers with single tube loader function
Setup protocol for Firefly particles On BD LSR II and LSRFortessa cytometers with single tube loader function Version 1 Last Updated May 2016 Setup protocol for Firefly particles Contents Introduction
More informationDiscover the CE you ve been missing.
Discover the CE you ve been missing. Designed for STR analysis, the Spectrum CE System is built to streamline and support your forensic and paternity workflow needs. Spectrum CE System combines state-of-the-art
More informationWorkflow Guide Slide(s) Topic 2-6 Importing Data and Labeling Samples 7-11 Processing Data Without an Allelic Ladder Processing Data With an
Workflow Guide Slide(s) Topic 2-6 Importing Data and Labeling Samples 7-11 Processing Data Without an Allelic Ladder 12-23 Processing Data With an Allelic Ladder 24-30 Reviewing Size and Allele Calls 31-37
More informationThermo Xcalibur Getting Started (Quantitative Analysis)
Thermo Xcalibur Getting Started (Quantitative Analysis) XCALI-97207 Revision B September 2010 2010 Thermo Fisher Scientific Inc. All rights reserved. Xcalibur, Surveyor, and Accela are registered trademarks
More informationOIW-EX Series of Oil In Water Analyzers MiView Handbook Document code: OIW-HBO-0013 Version: EX July 2013
OIW-EX Series of Oil In Water Analyzers MiView Handbook Document code: OIW-HBO-0013 Version: EX-004 17 July 2013 Tel: +44 (0) 28 9332 8922 Document History Change No. Version No. Description Month & Year
More informationHID Real-Time PCR Analysis Software
USER GUIDE HID Real-Time PCR Analysis Software Version 1.3 for use with: 7500 Real-Time PCR Instruments QuantStudio 5 Real-Time PCR Instruments (with 0.2-mL 96-Well Block) Quantifiler DNA Quantification
More informationAqueous GPC Instructions - Detailed
Aqueous GPC Instructions - Detailed Waste Bottle Columns PL Datastream Injection Port GPC Computer UV Detector GPC Pump Eluent Bottle The Aqueous GPC Set Up RI Detector Important The points below must
More informationABI PRISM 7700 Sequence Detection System: Routine Maintenance
ABI PRISM 7700 Sequence Detection System: Routine Maintenance Overview The ABI PRISM 7700 Sequence Detection System (SDS) and the Macintosh computer that runs the SDS need to be properly and routinely
More informationStandard Operating Procedure for the Horiba FluroMax-4
Standard Operating Procedure for the Horiba FluroMax-4 Adapted from Horiba Operations Manual Created by Michael Delcau, Modified by Brian Lamp The Fluoromax is capable of making a variety of measurements.
More informationAgilent 6400 Series Triple Quadrupole LC/MS System
Agilent 6400 Series Triple Quadrupole LC/MS System Quick Start Guide Where to find information 4 Getting Started 6 Step 1. Start the Data Acquisition software 7 Step 2. Prepare the LC modules 13 Step 3.
More informationLinRegPCR (11.0) Analysis of quantitative RT-PCR data
LinRegPCR (11.0) Analysis of quantitative RT-PCR data J. M. Ruijter S. van der Velden A. Ilgun Heart Failure Research Center Academic Medical Center Amsterdam, the Netherlands CONTENTS 1. About this manual...3
More informationQuick Start Guide Axxin T8-ISO. Instrument. In association with: E: W: D Version 1.5
Quick Start Guide Axxin T8-ISO Instrument In association with: E: support@axxin.com W: D006093 Version 1.5 Office DOCUMENT CONTROL DETAILS TITLE DOC No. CUSTOMER CONTACT QSG, Quick Start Guide, Axxin T8-ISO
More informationLinRegPCR. analysis of real-time quantitative PCR data J. M. Ruijter, C. Ramakers. manual under construction
LinRegPCR analysis of real-time quantitative PCR data J. M. Ruijter, C. Ramakers manual under construction 1. Update history Important changes in version 7.0 of LinRegPCR (February 2003): - Increased speed
More informationIMAGE STUDIO LITE. Tutorial Guide Featuring Image Studio Analysis Software Version 3.1
IMAGE STUDIO LITE Tutorial Guide Featuring Image Studio Analysis Software Version 3.1 Notice The information contained in this document is subject to change without notice. LI-COR MAKES NO WARRANTY OF
More informationFor the Odyssey Fc Near-Infrared Imaging System
Version 5.2 Quick Start Guide For the Odyssey Fc Near-Infrared Imaging System Page 2 - Odyssey Fc Near-Infrared Imaging System Quick Start Guide 1 Install Image Studio Software Note: Ensure you have the
More informationTutorial: FCAP Array Software with BD FACSArray Bioanalyzer
Tutorial: FCAP Array Software with BD FACSArray Bioanalyzer After completing this tutorial you will be able to: Create an Experiment with the Experiment Wizard in FCAP Array software. Export the Experiment
More informationFACSLSRFortessa SORP QUICK REFERENCE GUIDE
FACSLSRFortessa SORP QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Administrator Password: BDIS 2. Unlock the screen with your PPMS account (UTSW username
More informationWorkplan Distributor
Workplan Distributor User Guidance Manual Last updated in November 2009 2009 Adept Management Ltd. All rights reserved CHAPTER 1: INTRODUCTION...1 1.1 Welcome to the ADePT Workplan Distributor...1 1.2
More informationApplied Biosystems RapidFinder Software Version 1.0 and SDS Software Version Beta Release 10/05 Release Notes (PN Rev A)
Applied Biosystems RapidFinder Software Version 1.0 and SDS Software Version 1.3.2 Beta Release 10/05 Release Notes (PN 4370225 Rev A) Software Release notes for: Applied Biosystems RapidFinder Software
More informationWorkOut 2.0 User s Guide
WorkOut 2.0 User s Guide Copyright Dazdaq Solutions Ltd. 2004 i WorkOut 2.0 User s Guide Table Of Contents Getting started... 1 System requirements... 1 Minimum Requirements... 1 Recommended... 1 Installing...
More informationVNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS
VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS April 16, 2018 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they are shorter; you
More informationABI PRISM GeneMapper Software Version 3.0 SNP Genotyping
ABI PRISM GeneMapper Software Version 3.0 SNP Genotyping Tutorial ABI PRISM GeneMapper Software Version 3.0 SNP Genotyping Tutorial September 25, 2002 1:20 pm, 7x9_Title.fm Copyright 2002, Applied Biosystems.
More informationCFX Maestro Software for Mac
Real-Time PCR CFX Maestro Software for Mac User Guide Version 1.0 CFX Maestro Software for Mac User Guide Version 1.0 Bio-Rad Technical Support Department The Bio-Rad Technical Support department in the
More informationOlink NPX Manager. User Guide
Olink NPX Manager User Guide Table of contents INTRODUCTION 3 Process 3 Limitations 3 Requirements for analysis 3 List of abbreviations 3 TO CREATE A HEAT MAP CT CSV FILE 4 OLINK NPX MANAGER STEP BY STEP
More informationTutorial 7: Automated Peak Picking in Skyline
Tutorial 7: Automated Peak Picking in Skyline Skyline now supports the ability to create custom advanced peak picking and scoring models for both selected reaction monitoring (SRM) and data-independent
More informationNovoExpress Software Guide. Software Version 1.2.4
NovoExpress Software Guide Software Version 1.2.4 Table of Content 1. Introduction 6 1.1 Revision History...6 1.2 About the NovoExpress Software...6 1.3 Conventions...6 2. Installation 8 2.1 Installation
More informationApplied Biosystems TrueScience Aneuploidy STR Kits
Applied Biosystems TrueScience Aneuploidy STR Kits Software Setup and Data Analysis User Guide The results obtained from these or any other diagnostic kits should be used and interpreted only in the context
More informationDIAplex. DIAplex Pro 1.0 Software. Instructions for use. Fast Track Your Research. For research use only
DIAplex DIAplex Pro 1.0 Software Instructions for use For research use only Fast Track Your Research Table of Contents 1. Introduction... 2 2. DIAplex Pro 1.0 installation... 2 2.1. System Requirements...
More informationSetup for Omnia assays on SpectraMax Paradigm Microplate Detection Platform with SoftMax Pro 6 software
Page 1 of 20 Setup for Omnia assays on SpectraMax Paradigm Microplate Detection Platform with SoftMax Pro 6 software The Molecular Devices SpectraMax Paradigm Microplate Detection Platform was tested for
More informationddpcrquant Hands-on Wim Trypsteen Matthijs Vynck Doctoral schools specialist course 1 september 2016, t Pand Ghent
ddpcrquant Hands-on Wim Trypsteen Matthijs Vynck Doctoral schools specialist course 1 september 2016, t Pand Ghent 1 Day 1 (01-09-2016) Day 2 (02-09-2016) 08:00-9:30 Introduction to digital PCR Ward De
More informationBD Accuri C6 with single tube loader protocol. For multiplex assays
BD Accuri C6 with single tube loader protocol For multiplex assays BD Accuri C6 with single tube loader protocol This protocol contains instructions for setting up an Accuri C6 with Tube Loader. If you
More informationUser guide F A CARTO. Explore 3.0
User guide F-9930-1008-07-A CARTO Explore 3.0 Legal information Safety Before using the laser system, please consult the laser safety information booklet. Disclaimer Renishaw has made considerable efforts
More informationGeneious Microsatellite Plugin. Biomatters Ltd
Geneious Microsatellite Plugin Biomatters Ltd November 24, 2018 2 Introduction This plugin imports ABI fragment analysis files and allows you to visualize traces, fit ladders, call peaks, predict bins,
More informationGPR Analyzer version 1.23 User s Manual
GPR Analyzer version 1.23 User s Manual GPR Analyzer is a tool to quickly analyze multi- species microarrays, especially designed for use with the MIDTAL (Microarray Detection of Toxic ALgae) chip. It
More informationXcalibur. QuickQuan. User Guide. XCALI Revision C July For Research Use Only Not for use in Diagnostic Procedures
Xcalibur QuickQuan User Guide XCALI-97195 Revision C July 2007 For Research Use Only Not for use in Diagnostic Procedures 2007 Thermo Fisher Scientific Inc. All rights reserved. Microsoft, Access, Excel,
More informationMaintenance and Troubleshooting Guide
Applied Biosystems 7900HT Fast Real-Time PCR System Maintenance and Troubleshooting Guide Performing Calibration and Verification Runs Maintaining the 7900HT Instrument Hardware Maintaining the Computer
More informationQuick Guide NC-200. NucleoCounter. to the easiest cell count ever!
Quick Guide to the easiest cell count ever! NucleoCounter NC-00 How to get started Dear NucleoCounter Customer! Thank you for purchasing the NucleoCounter NC-00 offering not only the easiest one-step viability
More informationData acquisition and analysis of MACSPlex Exosome Kit using the MACSQuant Analyzer Express Mode in MACSQuantify Software version 2.
Data acquisition and analysis of MACSPlex Exosome Kit using the MACSQuant Analyzer Express Mode in MACSQuantify Software version 2.8 Contents 1. Description 1.1 Background information 1.2 Reagent and instrument
More informationStandard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017
Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017 Introduction This document is only for operation of the HPLC and ICP-MS instruments to obtain data. It contains no maintenance
More informationBD Lyoplate Human Screen Analysis Instructions For analysis using FCS Express or FlowJo and heatmap representation in Excel 2007
BD Biosciences Technical Resources Page 1 For use with the BD Lyoplate Human Cell Surface Marker Screening Panel (Cat. No. 560747). Please check that your catalog numbers for the FCS Express Excel templates
More informationUsing Excel for Graphical Analysis of Data
EXERCISE Using Excel for Graphical Analysis of Data Introduction In several upcoming experiments, a primary goal will be to determine the mathematical relationship between two variable physical parameters.
More informationVolocity ver (2013) Standard Operation Protocol
Faculty Core Facility Volocity 6.3.0 (2013) SOP A-1 Volocity ver. 6.3.0 (2013) Standard Operation Protocol Faculty Core Facility Volocity 6.3.0 (2013) SOP A-2 A. Content Overview. 3 Start up. 3 Change
More informationwgmlst typing in the Brucella demonstration database
BioNumerics Tutorial: wgmlst typing in the Brucella demonstration database 1 Introduction This guide is designed for users to explore the wgmlst functionality present in BioNumerics without having to create
More informationQuantiFERON -TB Gold In-Tube (v2.17.3*) Analysis Software Instructional Guide
QuantiFERON -TB Gold In-Tube (v2.17.3*) Analysis Software Instructional Guide QuantiFERON-TB Gold In-Tube Analysis Software. is a PC-based application for calculating QuantiFERON-TB Gold In-Tube (QFT )
More information1. What s New in Bio-Plex Manager 6.1
Bio-Rad Laboratories, Inc. Release Notes Bio-Plex Manager Software, Version 6.1 September 26, 2011 This document has the following sections: 1. What s New in Bio-Plex Manager 6.1 2. Resources for New Users
More informationMicrocycler TM. 8-Well Thermal Cycler. User Manual
8-Well Thermal Cycler IFU-MC-0001 Revision 1 26 Oct 2017 Table of Contents... 1 Introduction... 2 Description... 2 Safety and Regulatory Compliance... 3 Safety warning labels... 3 Set up... 4 Operation...
More informationAlpha Cycler Endpoint PCR range
A Bibby Scientific Company Alpha Cycler Endpoint PCR range Speed. Confidence. Value. Sensitivity. Performance. PCRmax Alpha Cyclers The PCRmax Alpha Cyclers are developed to deliver not only the best quality
More informationImage Analysis and Morphometry
Image Analysis and Morphometry Lukas Schärer Evolutionary Biology Zoological Institute University of Basel 1 13. /15.3.2013 Zoology & Evolution Block Course Summary Quantifying morphology why do we need
More informationOperating Manual. High Performance Liquid Chromatograph. Scientific Equipment Center, Prince Of Songkla University
Operating Manual High Performance Liquid Chromatograph Agilent 1100 ; VWD, DAD, FLD and RID Scientific Equipment Center, Prince Of Songkla University Operating Manual High Performance Liquid Chromatograph
More informationNCSBI Forensic Biology Section DNA SOP Effective Date: January 30, 2007 Title: ABI PRISM 7000: DNA Quantitation Revision 02
1. Preparing the DNA Standards 1.1. Label eight microcentrifuge tubes to be used for dilution of standards (i.e. Std. 1, Std. 2, Std. 3, etc.) 1.2. Add required amount of TE/glycogen to each tube (See
More informationChromeleon-PA Analyzer User s Guide Dionex Corporation
Chromeleon-PA Analyzer User s Guide 2003 Dionex Corporation Document No. 031964 Revision 01 October 2003 2003 by Dionex Corporation All rights reserved worldwide. Printed in the United States of America.
More informationTraceFinder Shortcut Menus Quick Reference Guide
TraceFinder Shortcut Menus Quick Reference Guide This quick reference guide describes the right-click shortcut menus available in the Thermo TraceFinder application. Contents Acquisition Mode Analysis
More informationCOLOCALISATION. Alexia Loynton-Ferrand. Imaging Core Facility Biozentrum Basel
COLOCALISATION Alexia Loynton-Ferrand Imaging Core Facility Biozentrum Basel OUTLINE Introduction How to best prepare your samples for colocalisation How to acquire the images for colocalisation How to
More informationRelease Notes. Bio-Plex Manager Software, Version 5.0 October 18, 2007
Release Notes Bio-Plex Manager Software, Version 5.0 October 18, 2007 This document has the following sections: 1. About Bio-Plex Manager 5.0 2. Resources for New Users 3. System Requirements 4. Bio-Plex
More informationS800 Spectrawave Visible Diode Array Spectrophotometer
6 SCANNING VISIBLE INSTRUMENT FOR EDUCATION OPTIMISED FOR THE TEACHING LABORATORY S800 Spectrawave Visible Diode Array Spectrophotometer ABSORBANCE, % TRANSMISSION, CONCENTRATION AND KINETICS LARGE, EASY
More informationProtein Quantification Kit (BCA Assay)
Protein Quantification Kit (BCA Assay) Booklet Item NO. KTD3001 Product Name Protein Quantification Kit (BCA Assay) ATTENTION For laboratory research use only. Not for clinical or diagnostic use Version
More informationVeriti Thermal Cycler
product brochure PCR Systems Veriti Thermal Cycler The Power of MORE Applied Biosystems, the most trusted name in PCR instruments, introduces a new line of innovative thermal cyclers The Veriti Thermal
More informationNEW IMPROVED SCREENING PLATFORM MATERNAL FETAL HEALTH
Introducing the new AutoDELFIA immunoassay system for prenatal screening Brochure not for distribution in the USA NEW IMPROVED MATERNAL FETAL HEALTH SCREENING PLATFORM ALREADY 60 MILLION PREGNANCIES SCREENED
More informationSkyline Targeted MS/MS
Skyline Targeted MS/MS Skyline now supports several methods of extracting chromatography-based quantitative measurements from the raw data files of full-scan mass spectrometers, such as ion trap and Q-TOF
More informationBIOMA USER MANUAL VERSION
BIOMA USER MANUAL VERSION 2.2.1304 VIENNA UNIVERSITY OF TECHNOLOGY, JULY 2013 Vienna University of Technology Institute for Water Quality, Resources and Waste Management A-1040 Wien, Karlsplatz 13/226
More informationOlink Wizard for GenEx
Olink Wizard for GenEx USER GUIDE Version 1.1 (Feb 2014) TECHNICAL SUPPORT For support and technical information, please contact support@multid.se, or join the GenEx online forum: www.multid.se/forum.php.
More informationCOLOCALISATION. Alexia Ferrand. Imaging Core Facility Biozentrum Basel
COLOCALISATION Alexia Ferrand Imaging Core Facility Biozentrum Basel OUTLINE Introduction How to best prepare your samples for colocalisation How to acquire the images for colocalisation How to analyse
More informationReport Generator User Guide
12 Tra Report Generator User Guide Agilent Seahorse XF Glycolysis Stress Test For research use only. 1 Table of Contents Introduction... 3 Parameter Calculations... 3 XF Report Generator Overview... 4
More informationSoftware installation
Table of contents 1 Introduction...4 2 Software installation...4 2.1 Protection...4 2.2 Minimum recommended configuration...4 2.3 Installation...4 3 Uninstall the application...4 4 Software presentation...5
More information