1 TLM_PRO_029 and Version No. 001 Page 1 of 6 Department: Clinical Sciences Unit: Tropical Laboratory Medicine Total WBC count and differential WBC count HemoCue WBC DIFF system Project/study: Not applicable 1. Scope and application This document provides instructions to ensure safe and accurate total WBC and differential WBC count determination using the HemoCue WBC DIFF system. This procedure describes the different steps to obtain the result. The device can be used in primary health care, in clinical trials, as a point-of-care testing device, etc. 2. Responsibilities Function Activities Lab Technician complies with this procedure keeps a record of all the data (direct print-out, transfer to pc, in lab note book) follows the Quality Assurance regulations and keeps a record of all the quality controls applies the maintenance procedures informs the lab manager when there is a problem with the device Lab Quality Manager checks that the internal quality procedures are followed correctly ensures that the quality control data is properly recorded subscribes the laboratory in an external quality program in order to evaluate the quality of the lab data Lab Manager checks that the lab technicians follow the procedure checks that maintenance is done at the right time takes action when reparation is needed formulates a maintenance contract with the supplier checks that data is properly recorded supervises the lab quality manager 3. References HemoCue WBC DIFF User Manual Osei-Bimpong A, et al. Point-of-care method for total white cell count: an evaluation of the HemoCue WBC device. Int Jnl Lab Hem 2009;31: De Jonge R. Evaluation of the HemoCue WBC analyzer to count leucocytes in body fluids. Ned Tijdschr Klin Chem Labgeneesk 2011;36: HemoCue WBC DIFF User Manual 4. Procedures 4.1 Principle of the test The red blood cells are lysed in the cuvette and the nuclei of the white cells are stained. 37 images are taken of the stained white cells and the cells are identified. The number of cells are counted by image analysis in the analyzer. 4.2 Safety All blood samples are potentially infectious. Follow standard safety precautions.
2 TLM_PRO_029 and Version No. 001 Page 2 of Specimen Special conditions for patient preparation: none Type of sample: capillary (finger stick) or anticoagulated venous blood (EDTA in solid form). Volume of sample: 10 µl of whole blood Sterility: not needed Time between collection and measurement: o o capillary: immediately venous: Store the samples at room temperature (15-35 C). Analyze them within 8 hours of sample taking. Rejection criteria: o blood samples with clots o cold agglutination or cyroglobulin o blood samples not or incorrectly labeled 4.4 Equipment and reagents HemoCue WBC DIFF analyzer HemoCue WBC DIFF or HemoCue WBC microcuvettes 6 type C batteries(1.5 V) and/or mains adapter Non sterile, single use gloves Tissue or gauze Alcohol swabs, sterile lancets Venous blood collection set (needles, syringe, sample tubes) R&D Systems HemoCue WBC control solutions WBC DIFF cleaner swab Pipette Parafilm or glass slide 4.5 Procedure Check the expiry date of the cuvettes. If expired, do NOT use the cuvettes. Take a new batch of cuvettes that are not expired. Pull out the cuvette holder into the loading position. Switch the machine on: Press and hold left button. The display is activated. Optronic unit is automatically checked: an hourglass is shown on the display The menu is displayed The analyzer is ready for use. You can use a keyboard and/or a barcode reader to insert patient data If you push the right button, you can adapt the settings of the analyzer (check operating manual): o language o time o date o the information you want to display on the result window/in the memory: patient/operator/lab/site/control ID o the units (10 9 /L and/or %) o sound signal o pin code
3 TLM_PRO_029 and Version No. 001 Page 3 of 6 Push the second button patient test and insert the required data (patient/operator/lab/site ID) with the keyboard or the barcode reader. Push the right button to confirm the inserted data If you want to change the data, push the left button until the required window is displayed The symbol insert cuvette is displayed Take a patient sample: o For fingerstick blood: Clean the fingertip with 70% alcohol and allow to dry Puncture the fingertip (middle or ring finger) with a sterile lancet Wipe away first 2-3 drops of blood Fill the microcuvette in one continuous process. The correct amount of blood (10 µl) is drawn into the microcuvette. The microcuvette should be completely filled. Do NOT refill!! o For venous blood/control: Mix the sample well (invert times by hand or place for 1-2 min on roller mixer) Place a drop of blood onto a hydrophobic surface (e.g. parafilm) or glass slide, using a pipette. Fill the microcuvette in one continuous process. The correct amount of blood (10 µl) is drawn into the microcuvette. The microcuvette should be completely filled. Do NOT refill!! Wipe away any excess of blood on the outside of the microcuvette tip Check for air bubbles in the filled microcuvette. If present, use a new microcuvette. Place the filled microcuvette into the cuvette holder and push the holder gently back into the measuring position (within 40 seconds after filling the cuvette).
4 TLM_PRO_029 and Version No. 001 Page 4 of 6 The window measuring is displayed. After 5 minutes the results will be displayed: o the WBC DIFF result window (when using the WBC DIFF microcuvettes) o the WBC result window (when using the WBC microcuvettes) o the inserted data Push the second button to switch between the different windows. Read and record the result. Push the third or the first button to accept or refuse the result. The result is stored, refused results are also stored but are marked with a flag. You can now print out the result. The menu is displayed again. Remove and discard the microcuvette in the appropriate bio-hazard container. Push the cuvette holder back into the instrument. Don t measure the filled cuvette again. All results are stored in the analyzer (up to 600 results). You can view them on the display, print them out (immediately after each measurement) or transfer them to a computer (special software required). You can manually remove the results from the memory. (check the operating manual) 4.6 Quality control The analyzer has an internal electronic self-test, to check the optronic unit. This test is performed every time the analyzer is switched on and before every measurement. A daily quality control has to be done with the recommended controls: R&D Systems HemoCue WBC control Push the left button on the menu (QC-test), confirm, insert the required data and run the test (check operation manual).the QC data are stored in the memory (check operating manual) 4.7 Reporting results In 10 9 /L and/or % Measuring range: x10 9 /L. Higher results are reported as HHH, lower results as LLL. Differential counts are displayed when WBC count is within 1-30 x10 9 /L range. The normal ranges are: (depending on sex, daily variation, indigestion, stress, smoking, fitness)
5 TLM_PRO_029 and Version No. 001 Page 5 of 6 Children (mean ± 2SD) 3-6 months 1 year 2-6 years 6-12 years WBC count x 10 9 /L 12 ± 6 11 ± 5 10 ± 5 9 ± 4 Neutrophils x10 9 /L Lymphocytes x 10 9 /L Monocytes x 10 9 /L Eosinophils x 10 9 /L Adults (mean ± 2SD) 4.8 Storage and stability Adult Adult % WBC count x 10 9 /L Neutrophils x10 9 /L Lymphocytes x 10 9 /L Monocytes x 10 9 /L Eosinophils x 10 9 /L Basophils x 10 9 /L <1-2 Use the microcuvettes before the expiry date. Store the microcuvettes at C, < 90% non-condensing relative humidity. Do NOT refrigerate. An unopened capsule of microcuvettes can be stored for 4 weeks between 4 C - 50 C. Write down the date of opening on the vial. Allow the microcuvettes to reach C before use. Once the wrapping of a single package microcuvette is opened, it has to be used within 10 minutes. Store the analyzer at 4-50 C. Operate the analyzer at C,< 90% non-condensing relative humidity. 4.9 Maintenance Check that the analyzer is turned off Clean the cuvette holder every day with alcohol or mild detergent (check operating manual) Wait for 15 minutes before replacing the cuvette holder. Remove the cuvette holder to clean the optical unit with the WBC DIFF cleaner swab. (check operating manual) Wait for 15 minutes. Replace the cuvette holder. Clean the cover of the analyzer with alcohol or mild detergent 4.10 Causes of error Insufficient or non-uniform filling of the microcuvette (Rouleaux or agglutination): repeat the test. Presence of air bubbles in the microcuvette Incorrect mixing of the venous blood Excessive squeezing of the finger when collecting capillary blood Troubleshooting: (see operating manual for more troubleshooting)
6 TLM_PRO_029 and Version No. 001 Page 6 of 6 5. Records and archives Appendices & Forms for completion Number Title Not applicable - Revision Changes with respect to the previous version Initial version Name and function Date Signature Author Barbara Barbé 01/02/2012 Reviewed by An Vermaelen Approved by Jan Jacobs
Hemo Control Analyzer. Point of care device providing immediate, lab quality results for hemoglobin and hematocrit from one simple test
Hemo Control Analyzer Point of care device providing immediate, lab quality results for hemoglobin and hematocrit from one simple test Product features Easy to use Small sample size (ca. 8μl) Step by step
More informationSTANDARD OPERATING PROCEDURE ACCU-CHEK GO
Page 1 of 21 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the ACCU-CHEK Go in the Pharmacy Practice Resource Unit (PPRU) at the Pharmacy Department, University
More informationHemoPoint H2. Hemoglobin Analyzer. User Manual. Diagnostics for life
HemoPoint H2 Hemoglobin Analyzer User Manual Diagnostics for life HemoPoint H2 Hemoglobin Analyzer User Manual The copyright of this document is protected! Copying and distribution to third parties only
More informationSTANDARD OPERATING PROCEDURE
Document No: LAB/023/2 Page 1 of 7 Uncontrolled copy if not printed on yellow paper. Title: STANDARD OPERATING PROCEDURE Use of the Magtration System 8Lx Document No: LAB/023 Version No: 2 CopyNo/Holder:
More informationSTANDARD OPERATING PROCEDURE
Page 1 of 43 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the DCA 2000 Analyser in the Pharmacy Practice Resource Unit (PPRU) at the Pharmacy Department, University
More informationCoaguChek System. XS Plus. User Manual For meters shipped before January This is a CLIA waived system.
CoaguChek System XS Plus User Manual For meters shipped before January 2013 This is a CLIA waived system. Revisions to this document are provided by Roche Diagnostics when necessary. No part of this document
More informationProducts for Health Care
www.medicacorp.com \EDIC@ Products for Health Care easy inside and out Medica s EasyStat analyzer measures ph, PCO 2, PO 2, Na +, K +, Ca ++, and Hct and calculates additional parameters. Patient parameters,
More informationFast Facts. for Operating the scil Vet abc Plus. Hematology Chemistry Digital Radiography Ultrasound Orthopedics Education
Fast Facts for Operating the scil Vet abc Plus Hematology Chemistry Digital Radiography Ultrasound Orthopedics Education SCIL Vet ABC Plus Fast Facts Table of Contents Start Up Cycle pg. 2 Reagents Status
More informationQuick Reference Guide for the i-stat 1 Handheld and i-stat Chemistry Cartridges
Quick Reference Guide for the i-stat 1 Handheld and i-stat Chemistry Cartridges For use with a CLIA Certificate of Waiver G 6+ EC4+ Crea CHEM8+ E3+ Quick Reference Guide for the i-stat 1 Handheld and i-stat
More informationEasyElectrolytes. With EasyElectrolytes, Medica has redefined electrolyte analyzer design
EasyElectrolytes With EasyElectrolytes, Medica has redefined electrolyte analyzer design Simple yes/no prompted operation, combined with a modular design makes EasyElectrolytes reliable and economical.
More informationMission Cholesterol Monitoring System
Mission Cholesterol Monitoring System Important Safety Instructions If the meter is powered by AC adapter, unplug the equipment immediately after use. Misuse of electrical equipment can cause electrocution,
More informationSkyla VB1 Quick Reference Guide
Skyla VB1 Quick Reference Guide 1 VB1 DM/JAC 2014 v1.0 Quick Reference Guide Contents External features of the analyser 3 VB1 Function Buttons 4 Testing a Sample 5 How to reconstitute a biochemistry control
More informationMagnetic Immunoassay Reader Operator Manual
Page 1 of 8 Magnetic Immunoassay Reader Operator Manual Table of contents 1 Introduction...1 2 Magnetic Immunoassay Reader...1 2.1 Contact information and spare parts...1 3 Contents of delivery...1 4 Technical
More informationWorld-class veterinary hematology system
World-class veterinary hematology system New Exigo H400 CBC from one drop of blood Easy-to-use interface 12 pre-installed animal profiles Robust and reliable Exigo when quality counts! Exigo H400 is a
More informationSTANDARD OPERATING PROCEDURE TONOPORT V
Page 1 of 24 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the Tonoport V in the Pharmacy Practice Resource Unit (PPRU) at the Pharmacy Department, University
More informationUniCel DxH 800 Coulter Cellular Analysis System Training Modules
Training Modules This document is not intended to replace the information in your Instrument Instructions for Use Manual (IFU). Information in the Instructions for Use Manual supersedes information in
More informationMedonic M-series M32 Innovation built on total quality
Medonic M-series M32 Innovation built on total quality FOR TODAY S HEMATOLOGY LABS Full CBC from one drop of blood Maintenance-free shear valve Unique walk-away Autoloader Make way for Medonic! Packed
More informationDPP Micro Reader. The DPP Micro Reader is maintenance-free, not configurable by the user and is operated with a single, multi-function button.
Read this User Guide completely before using the product. Storage conditions: Store between -20 to 80 C (-4 to 176 F) DPP Micro Reader For use under Emergency Use Authorization (EUA) only. For In Vitro
More informationVITROS TDM Performance Verifier is prepared from bovine serum to which therapeutic drugs, inorganic salts, and preservatives have been added.
VITROS Chemistry Products TDM Performance Verifier I, II, and III TDM Performance Verifier I, II, and III 896 2540 179 2357 818 2172 Rx ONLY Intended Use For in vitro diagnostic use only. VITROS Chemistry
More informationPatient Lens Handling & Care
Patient Lens Handling & Care Transform your vision with UltraHealth. What to expect: UltraHealth hybrid contact lenses provide consistently clear vision throughout the day and offer all-day comfort by
More informationPatient Lens Handling & Care
Patient Lens Handling & Care Transform your vision with UltraHealth. How to insert your lenses: Always wash your hands with mild soap and dry them well with a lint-free towel before handling the lenses.
More information(11-nor-9-carboxy- 9-THC) PCP Phencyclidine (Phencyclidine)
Profile-V Reader System Quick Reference Instructions Before performing the test, refer to the PROFILE -V MEDTOXScan Drugs of Abuse Test System Insert and MEDTOXScan User Manual for complete operating instructions
More informationCareFusion. Overview of the CareFusion Device. Press Red button on top right-hand corner to turn device on and off
Tim Hamill, M.D. Digitally signed by Tim Hamill, M.D. DN: cn=tim Hamill, M.D., o, ou=ucsf Clinical Laboratories, email=hamillt@labmed2.ucsf.edu, c=us Date: 2009.04.20 20:57:31-07'00' CareFusion Summary:
More informationAnaPrep 12. Instruction Manual. BioChain Institute Inc. Automated Nucleic Acid Preparation System
BioChain Institute Inc. Instruction Manual Automated Nucleic Acid Preparation System AnaPrep 12 Catalog No.: Z1321001 Rev. Date: April 30, 2015 Manual No. Z1001 Version: Z1321001UA (MV1.14) For General
More informationADVIA 360, ADVIA 560, and ADVIA 560 AL Systems
Siemens Solutions for Hematology ADVIA 360, ADVIA 560, and ADVIA 560 AL Systems Answers for life. The ADVIA Hematology Family Siemens can now offer a total solution with the ADVIA 360, ADVIA 560, ADVIA
More informationSwelab Alfa. A new definition of cell counter precision.
Swelab Alfa. A new definition of cell counter precision. Basic Swelab Alfa Basic gives you a big cell counter for the price of a small one. Featuring many of the attributes of the larger Alfa models, it
More informationBluelab ph Meter TM Instruction Manual
TM Instruction Manual www.getbluelab.com Table of Contents 1.0 Introduction to Bluelab ph Meter 2 2.0 Preparing the Bluelab ph Meter for use 3 3.0 Calibrating Meter 4 4.0 Measure ph Value 5 5.0 Cleaning
More informationIntroduction General description Technology - Mechanical Technology - Viability measurement
Instruction Manual Table of Contents Introduction General description Technology - Mechanical Technology - Viability measurement 4-6 4 5 6 Product description Packing list Front view of the ADAM-rWBC
More informationBD FACSPresto Near-Patient CD4 Counter
BD FACSPresto Near-Patient CD4 Counter 651000 Instructions For Use IVD 23-12808-02 11/2014 EC REP Becton, Dickinson and Company BD Biosciences 2350 Qume Drive San Jose, CA 95131 USA Benex Limited Pottery
More informationBrand Analyzers Portable Whole Blood Test System User Guide
Brand Analyzers Portable Whole Blood Test System User Guide For Self-Testing Use TABLE OF CONTENTS I. Introduction...3 II. Introduction to the CardioChek Test System...4 CardioChek Analyzer MEMo Chip CardioChek
More information4 plate fully automated ELISA analyzer Reagent system. Open (Human tests predefined) Throughput (max.) Up to 7 plates / run Reaction / reading system
Specifications & more > > System Specifications > > Consumables > > Accessories System Overview Cat. No. 16300 Analyzer type 4 plate fully automated ELISA analyzer Reagent system Open (Human tests predefined)
More informationVeterinary medicine All-in-one solutions for sample collection in veterinary medicine
Veterinary medicine All-in-one solutions for sample collection in veterinary medicine XXXX Blood XXXXXXX collection Small animals XXXX Blood XXXXXXX collection Small animals Today, laboratory testing for
More informationTesting Procedures for i-stat Chemistry Cartridges
Testing Procedures for i-stat Chemistry Cartridges These testing procedures are for use with the following CLIA-waived i-stat cartridges: CHEM8+, 6+, EC4+, E3+, Crea, and G. These cartridges include various
More informationOperating Instructions
Operating Instructions INTERCHANGEABLE LENS FOR DIGITAL CAMERA Model No. H-F007014 Before use, please read these instructions completely. E -1 Contents Information for Your Safety... 2 Precautions... 4
More informationWELCOME TO A member of
WELCOME TO A member of 1 BST Introduction Unit Consumables Berlin Unit Instruments Issum 2 BST Instruments - Laboratory Labtrend Consumables 3 BST POCT Instruments GLUKOMETER PRO LAC PRO 4 BST POCT Connectivity
More informationUSERS MANUAL DIGI-DOC.COM
USERS MANUAL DIGI-DOC.COM 800-518-1102 SAFETY INSTRUCTIONS THIS SYMBOL ALERTS THE USER THAT IMPORTANT LITERATURE CONCERNING THE OPERATIONS AND MAINTENANCE OF THIS UNIT HAVE BEEN INCLUDED OR IS AVAILABLE
More informationARRAYMATE. Operating Manual
ARRAYMATE Operating Manual Version 2.1 Alere Technologies 2013 Table of Contents: 1. 1.1 Introduction... 3 Instrument Overview... 3 2. 2.1 Safety Information... 4 Important Operation Notes... 4 2.2 Sign
More informationscil Technical Support
Vet ABC Hematology Analyzer Calibration ** Before Calibration is performed run a concentrated cleaning and make sure your Photometer is within range. After both are completed run a Startup by pressing
More informationEquipment Requirements for a MCH Centre Laboratory
www.phclab.com The Public Health Care Laboratory The website for laboratory personnel in developing countries Equipment Requirements for a MCH Centre Laboratory The following list has been prepared based
More informationELECTIVES MODULES. Implement. Electives. Setup. Troubleshooting. Basics
ELECTIVES MODULES Implement Electives Setup Troubleshooting Basics This document is not intended to replace the information in your Instrument Instructions for Use Manual (IFU). Information in the Instructions
More informationLIGHT BRICK - 12 X 3W TRI-COLOUR
www.prolight.co.uk LIGHT BRICK - 12 X 3W TRI-COLOUR (Order code: LEDJ162) USER MANUAL Safety WARNING FOR YOUR OWN SAFETY, PLEASE READ THIS USER MANUAL CAREFULLY BEFORE YOUR INITIAL START-UP! CAUTION! Keep
More informationCellaVision. Ola Andersson Peter Wilson. Siemens Belgium
CellaVision Ola Andersson Peter Wilson Siemens Belgium CellaVision in short Headquarters in Lund, Sweden Around 80 employees globally Market support offices in the Nordic countries, the US, Canada, Japan,
More informationRichter Optica. Instructions for Model: S-2 Dual Power Stereo Microscope
Richter Optica 6122 Innovation Way Carlsbad, California 92009 (800) 942-0528 US TOLL FREE (760) 438-0528 INTERNATIONAL Instructions for Model: S-2 Dual Power Stereo Microscope Eyepieces Post Diopter Adjustment
More informationEco D-ELS-4 Stereo Microscope. Instruction Manual
Eco D-ELS-4 Stereo Microscope Instruction Manual Contents 1. Application 2 2. Nomenclature 2 3. Specifications 3 4. Operation 4 4.1 Working Environment 4 4.2 Light Intensity Control 4 4.3 Select a Work
More information9 Maintenance. Agilent Technologies. This chapter describes the maintenance of the module. InfinityLab LC Series Diode Array Detectors User Manual
InfinityLab LC Series Diode Array Detectors User Manual 9 Maintenance Introduction to Maintenance 158 Warnings and Cautions 159 Overview of Maintenance 161 Cleaning the Module 162 Remove and Install Doors
More informationManual for the MicroSIA Instrument
Manual for the MicroSIA Instrument Version 1.1 1 FIAlab Instruments, Inc., 2151 N. Northlake Way, Suite 100, Seattle, WA 98103 USA Table of Contents 1 INTRODUCTION... 3 1.1 GENERAL... 3 1.2 OPERATION PRINCIPLE...
More information1 plate fully automated ELISA analyzer Reagent system. Open (Human tests predefined) Throughput (Max.) 1 plate /run Barcode reader
Specifications & more Elisys Uno > > System Specifications > > Consumables > > Accessories System Overview Cat.. 17350 Analyzer type 1 plate fully automated ELISA analyzer Reagent system Open (Human tests
More informationUser Manual. Trinocular Stereo Microscope
User Manual Trinocular Stereo Microscope Model V434 XV434 Series MicroscopeNet.com Table of Contents i. Caution.. 1 ii. Care and Maintenance... 2 1. Component Illustration... 3 2. Installation 4 3. Operation....6
More informationSOP:07:VL:015:03:NIBT PAGE 1 of 16
SOP:07:VL:015:03:NIBT PAGE 1 of 16 Northern Ireland Blood Transfusion Service STANDARD OPERATING PROCEDURE (Operational Copy) Document Details Document Number: SOP:07:VL:015:03:NIBT Supersedes Number:
More informationSTANDARD OPERATING PROCEDURE TENSOVAL COMFORT
Page 1 of 13 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the Tensoval Comfort in the Pharmacy Practice Resource Unit (PPRU) at the Pharmacy Department, University
More informationRichter Optica. Instructions for Models: UX1B Binocular and UX1T Trinocular Laboratory and University Biological Microscope
Richter Optica info@richter-optica.com Instructions for Models: UX1B Binocular and UX1T Trinocular Laboratory and University Biological Microscope Interpupillary Adjustment Scale Eyepieces Diopter Adjustment
More informationContact Lens Wearer's Guide PERSONAL CLEANLINESS FOR LENS HANDLING AND INSERTION
Contact Lens Wearer's Guide PERSONAL CLEANLINESS FOR LENS HANDLING AND INSERTION 1. Preparing The Lens For Wearing It is essential that you learn and use good hygienic methods in the care and handling
More informationUSER S MANUAL TULIP DIAGNOSTICS (P) LTD. Gitanjali, Tulip Block, Dr. Antonion De Rego Bagh, Alto Santacruz Bambolim P.O GOA.
USER S MANUAL TULIP DIAGNOSTICS (P) LTD. Gitanjali, Tulip Block, Dr. Antonion De Rego Bagh, Alto Santacruz Bambolim P.O GOA. Pin: 403202 Phone: +91 832-2458546/51; Fax: +91 832-2458544; Email: sales@tulipgroup.com
More informationMT5000-DSS 4_Rev _ _user manual_en.docx. Digital Slide Scanner MT5000/DSS. User manual
Digital Slide Scanner MT5000/DSS 4 User manual Contents 1 Introduction... 3 1.1 MT5000/DSS 4 contents... 3 1.2 About the user manual... 3 1.3 Intended use... 3 1.4 Specifications... 4 1.5 System requirements...
More informationINSTRUCTION MANUAL. Automatic Blood Pressure Monitor with Arm Cuff. Model HEM-711AC
INSTRUCTION MANUAL Automatic Blood Pressure Monitor with Arm Cuff Model HEM-711AC ENGLISH ESPAÑOL CONTENTS Before Using the Monitor Introduction..........................................2 Important Safety
More informationNCSBI Forensic Biology Section DNA SOP Effective Date: January 30, 2007 Title: ABI PRISM 7000: DNA Quantitation Revision 02
1. Preparing the DNA Standards 1.1. Label eight microcentrifuge tubes to be used for dilution of standards (i.e. Std. 1, Std. 2, Std. 3, etc.) 1.2. Add required amount of TE/glycogen to each tube (See
More informationOPERATING INSTRUCTIONS
OPERATING INSTRUCTIONS Automatic Tissue Processor MTP INS5500GB 2011-07-08 Instructions MTP 1 Instructions MTP 2 CONTENTS 1. INTENDED USE... 4 2. SYMBOLS... 4 3. SAFETY NOTES... 4 4. COMPONENTS... 5 5.
More informationViral Load Scale-Up Facility Readiness Assessment Checklist
Viral Load Scale-Up Facility Readiness Assessment Checklist Facility Name: Region: Date: Assessors Names: 1. 2. Interview Start Time: Interview End Time: Instructions: Please read and make sure you understand
More informationElectrolyte analyzer i-smart 30
Electrolyte analyzer i-smart 30 Overview i-sens electrolyte analyzer, i-smart 30 uses 60 μl of blood sample to measure ions such as Na+, K+, Cl- and hematocrit automatically. Unlike other competitors products,
More informationBD FACSDiva Software Quick Reference Guide for BD FACSCanto Systems with Loader Option
BD FACSDiva Software Quick Reference Guide for BD FACSCanto Systems with Loader Option This guide contains instructions for using BD FACSDiva software version 6 with BD FACSCanto and BD FACSCanto II systems
More informationThe Care And Cleaning Of Your Manatee County Schools ibook
The Care And Cleaning Of Your Manatee County Schools ibook The ibook is fun to use, but it is not a toy! The ibooks belong to Manatee County Schools. They are on loan to you to use as a tool for learning.
More informationKeylines DO NOT print 4-Color Process. with TIPS CARD. See System Operating Manual for complete instructions including warnings and cautions.
Keylines DO NOT print 4-Color Process with TIPS CARD See System Operating Manual for complete instructions including warnings and cautions. 444-04829-001 PAGE 1 OF 9 DEFINITION OF TERMS Hospira MedNet
More informationCOULTER STKS Analyzer with Reticulocyte Analysis. Operator s Guide. PN B (March 1995) COULTER CORPORATION Miami, Florida 33196
COULTER STKS Analyzer with Reticulocyte Analysis Operator s Guide PN 4237188B (March 1995) COULTER CORPORATION Miami, Florida 33196 READ ALL PRODUCT MANUALS AND CONSULT WITH COULTER-TRAINED PERSONNEL BEFORE
More informationCustomer Responsibilities
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationPERSONAL MEDICATION ASSISTANT
PERSONAL MEDICATION ASSISTANT Instructions for Use Cat. FUSE-001 Table of Contents Preface...1 Purpose...1 Help Information...1 System Overview...2 System Description...2 Intended Use...2 Contraindications...3
More informationNHS Health Checks Quality Control testing guide
NHS Health Checks Quality Control testing guide Contents HbA1c Quality Control Page 2-3 (Hybrid Internal and External) Lipid Quality Control Page 4 (Hybrid Internal and External) QC Schedule (June November
More informationSTAINING PROTOCOL FOR CYTOMETRY (Réf LabEx #1)
STAINING PROTOCOL FOR CYTOMETRY (Réf LabEx #1) Aim To provide detailed description of all steps, material and methods used for fluorescent surface staining and preparation of whole blood cells for analysis
More informationF Printed in Japan
83-0040F Printed in Japan DOC.No.OM-28573 REV.6 03/02 DOC.No.OM-28573 REV.6 03/02 SPOTCHEM EZ SP-4430 OPERATING MANUAL Premise This manual contains important information on the functions of the SPOTCHEM
More informationCH142 Spring Spectrophotometers with Vernier Data Acquisition Software
Spectrophotometers with Vernier Data Acquisition Software The absorbance of a sample is given as A = log I o I, where I o is the intensity without sample present and I is the intensity with the sample
More informationA+B INK CARTRIDGES INSTALLATION INSTRUCTIONS
Tel: 02 80655614 1300 213 589 www.mir-aus.com www.mir-aus.com.au info@mir-aus.com A+B INK CARTRIDGES INSTALLATION INSTRUCTIONS 1. Keep out of reach of children. Seek medical advice if taken. 2. Avoid exposure
More informationGTA PREVENTIVE MAINTENANCE CHECKS AND SERVICES ON THE ZOOM STEREOSCOPE 95R
PREVENTIVE MAINTENANCE CHECKS AND SERVICES ON THE ZOOM STEREOSCOPE 95R DISTRIBUTION RESTRICTION: Approved for public release; distribution is unlimited. DISTRIBUTION:... US Army Training and Audiovisual
More information1. Review the printer safety information in the Regulatory Compliance document provided with your printer.
QUICK REFERENCE This Quick Reference contains supply loading information and general care and maintenance procedures for the Monarch Pathfinder Ultra Platinum 6039 printer. For more detailed information,
More informationQuick Reference Guide. For Export Only. Lactate Hospital Meter
Lac Quick Reference Guide L A C Refer to the StatStrip Lactate Hospital Meter Instructions for Use Manual and Package Inserts for complete instructions for use, indications, precautions, and limitations
More informationintellipath Research Software Manual
intellipath Research Software Manual www.biocare.net Biocare Medical Technical Support: 800-799-9499 Option 3 Customer Service: 800-799-9499 Fax: 925-603-8080 Corporate Office: 4040 Pike Lane Concord,
More informationProtein Quantification Kit (BCA Assay)
Protein Quantification Kit (BCA Assay) Booklet Item NO. KTD3001 Product Name Protein Quantification Kit (BCA Assay) ATTENTION For laboratory research use only. Not for clinical or diagnostic use Version
More informationQuo-Test. User Manual SCIENCE MADE SIMPLE
Quo-Test User Manual SCIENCE MADE SIMPLE Copyright 2008-2015 Quotient Diagnostics Ltd No part of this publication may be reproduced, transmitted, transcribed, stored in any retrieval system or translated
More informationMultiTherm Shaker. Operations Manual Item: H5000-H and H5000-HC
MultiTherm Shaker Operations Manual Item: H5000-H and H5000-HC Foreword Thank you for purchasing a MultiTherm Shaker. This operations manual contains instructions for the care of this Instrument. In order
More informationQuick Instruction Guide for Operating the scil Vet abc Plus
Quick Instruction Guide for Operating the scil Vet abc Plus Hematology Chemistry Digital Radiography Ultrasound Orthopedics Education Table of Contents Start Up Cycle... pg. 2 Reagents....pg.2-3 User Profiles......pg.4-5
More informationUnicel DxH 600 System
SETUP MODULES Setup Unicel DxH 600 System Troubleshooting System Overview Additional Topics This document is not intended to replace the information in your Instrument Instructions for Use Manual (IFU).
More informationRichter Optica. Instructions for Model: MDS2 Middle School Microscope
Richter Optica info@richter-optica.com Instructions for Model: MDS2 Middle School Microscope Eyepiece 360 Rotatable Head Revolving Nosepiece Carrying Handle Objectives Rack Stop Stage Stage Clips Disc
More informationUser Guide. Quality Check Messages and Codes (continued) (01) Message Cause Action
Quality Check Messages and Codes (continued) Message Cause Action i-stat Analyzer Error, Use Electronic Simulator Analyzer Error, See Manual No display Cartridge Locked does not disappear after test cycle
More informationDC100 Digital Dispenser Operating Manual
DC100 Digital Dispenser Operating Manual 2015 Fisnar Phone: (262)253-8600 E-mail: info@fisnar.com Table of Contents Product Safety Statement 4 Specifications 6 Accessories 6 External Controls 7 Machine
More informationTABLE OF CONTENTS. Table of Contents. Table of Figures
TABLE OF CONTENTS Table of Contents Table of Figures 1 2 1 Introduction 3 2 Installation 4 2.1 Installation Precautions 4 2.2 Environmental Requirements 4 3 Specifications 4 4 Precautions 4 5 Materials
More informationepoc Blood Gas Analysis System
epoc Blood Gas Analysis System Objectives Introduce EPOC Blood Gas Analysis System Give system overview of device Explain system operation Demonstrate sample analysis Describe pre/post sample analysis
More informationP OWERBUDS. The perfect accessory for anyone who wants to be truly wireless PLEASE READ BEFORE OPERATING THIS EQUIPMENT
P OWERBUDS The perfect accessory for anyone who wants to be truly wireless PLEASE READ BEFORE OPERATING THIS EQUIPMENT HALO POWERBUDS Thank you for choosing HALO. The HALO POWERBUDS are the perfect accessory
More informationTable of Contents. Chapter 1. Safety Precautions Notice Chapter 2. Components and accessories Chapter 3. Camera components...
P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046 P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046 Table of Contents Chapter 1. Safety Precautions Notice...
More informationOM-A Revision history Rev. Date Edited by Sections affected Approved by
OM-A380-01-115 Revision history Rev. Date Edited by Sections affected Approved by 1.0 2010.05.14 Gabor Farkas Initial revision 1.1 2010.07.27. Gabor Farkas Reagent Lock System added, Profile settings added,
More information6 x 6 Peltier-Thermostatted Multicell Holder Series II Accessory
6 x 6 Peltier-Thermostatted Multicell Holder Series II Accessory Overvoltage Category II Pollution Degree 2 Equipment Class III Safety information Introduction Before using this accessory, you must read
More informationA1H Portable Humidity Temperature Indicator Instruction Manual
A1H Portable Humidity Temperature Indicator Instruction Manual -2- CONTENTS Description...3 Operation...3 Measurement...4 Maintenance...4 Probe Installation and Removal...5 Output Connector...6 Accessories
More informationOMM200 and OMM200T. Metallurgical Microscope
OMM200 and OMM200T Metallurgical Microscope Instruction Manual (Please read the manual carefully before using the microscope) Contents Safety ----------------------------------------------------------------------------
More informationContents Safety.. 1 General. 1 Electrical.. 1 Warning 1 Working Principle 2 Unpacking Instructions 2 Specifications 3 Installation.. 3 Introduction..
Contents Safety.. General. Electrical.. Warning Working Principle 2 Unpacking Instructions 2 Specifications 3 Installation.. 3 Introduction.. 4 Description of keys and operating method.. 5 Operating Procedures.
More informationOperating Instructions
Operating Instructions INTERCHANGEABLE LENS FOR DIGITAL CAMERA Model No.H-F008 Before use, please read these instructions completely. E Contents Information for Your Safety... 2 Precautions... 5 Supplied
More informationBC Auto Hematology Analyzer. Pleasant Experience in Hematology Testing
BC-3600 Auto Hematology Analyzer Pleasant Experience in Hematology Testing BC-3600 Auto Hematology Analyzer Labs today are facing a multitude of challenges including limited laboratory budgets, increasing
More informationSOP:07:VL:018:03:NIBT PAGE 1 of 7
SOP:07:VL:018:03:NIBT PAGE 1 of 7 Northern Ireland Blood Transfusion Service STANDARD OPERATING PROCEDURE (Operational Copy) Document Details Document Number: SOP:07:VL:018:03:NIBT Supersedes Number: 07:VL:018:02:NIBT
More information9 Maintenance and Repair
Agilent 1200 Infinity II Series RID User Manual 9 Maintenance and Repair Introduction to Maintenance 120 Warnings and Cautions 121 Overview of Maintenance 123 Cleaning the Module 124 Storage of the Detector
More informationDigital Overhead Stirrer
A Geno Technology, Inc. (USA) brand name Digital Overhead Stirrer Cat. No. BT1021 1-800-628-7730 1-314-991-6034 info@btlabsystems.com Thanks for choosing BT1021 Digital Overhead Stirrer. This operation
More informationTACTIVO for iphone 4S and iphone 4. User Reference Guide
TACTIVO for iphone 4S and iphone 4 User Reference Guide Trademarks: Tactivo and PRECISE BIOMETRICS Design: registered community design and pending design patent in USA Patent: patents pending in EU and
More informationELECTRONIC PULSE MASSAGER ZX-581 USER MANUAL. Care for Your Loved Ones facebook.com/nursalonline
ELECTRONIC PULSE MASSAGER ZX-581 USER MANUAL Care for Your Loved Ones www.nursal.co/warranty facebook.com/nursalonline ACTIVATE YOUR 12 MONTH WARRANTY & GET EXCLUSIVE GIFT Register within 2 weeks after
More informationDate Name Signature Issued 06-MAY-2011 Csaba Magyar Checked and approved 17-MAY-2011 Gábor Farkas
Revision history OM-ACS30-01-1110 Rev. Date Edited by Sections affected Approved by 1.0 2009-06-02 Arpad Gyetvai Initial revision Csaba Magyar 1.1 2010-03-25 Gabor Farkas First review, new GUI screenshots
More informationUniCel DxH 600 Coulter Cellular Analysis System Training Modules
UniCel DxH 600 Coulter Cellular Analysis System Training Modules This document is not intended to replace the information in your Instrument Instructions for Use Manual (IFU). Information in the Instructions
More information